Beruflich Dokumente
Kultur Dokumente
January 2014
00813-0100-4750, Rev FA
Rosemount reliability in a customized offering specific to the Water and Wastewater industries
Line sizes 42-in. (1050 mm) and 48-in. (1200 mm) available Globally
Options for:
- Diagnostic Suite for improved maintenance practices
- Diagnostic Suite for simplified meter verification
- Submersible to IP68
- NSF Drinking Water Certification
Rosemount 8750WA January 2014
The operating principle of the magnetic flowmeter system is The field mount Rosemount 32ES transmitter has a die cast
based upon Faraday’s Law of electromagnetic induction, which aluminum housing for excellent reliability. With an optional
states that a voltage will be induced in a conductor moving through backlit 2 line by 16 character local operator interface, the
a magnetic field. transmitter can be configured by optical switches to simplify
adjustments without removing the cover.
Faraday’s Law: E=kBDV
Content
Rosemount 8750WA specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 9
2 www.rosemount.com
January 2014 Rosemount 8750WA
Flanged sensors
Flanged sensors are fabricated from stainless and carbon steel
and welded to provide a hermetic seal that protects against
moisture and other contaminants. Sizes range from 1/2-in. (15
mm) to 48-in. (1200 mm) standard (larger sizes available -
consult factory). The sealed housing ensures maximum sensor
reliability by protecting all internal components and wiring from
the most hostile environments.
Wafer sensors
The flangeless design of the wafer sensor makes it an
economical, compact, and lightweight alternative to flanged
magnetic flowmeters. Alignment spacers provided with every
sensor assist with centering the sensor in the process line
making installation easier.
www.rosemount.com 3
Rosemount 8750WA January 2014
With Rosemount Magmeter diagnostics enabled, users can change their practices to improve plant availability and output, and
reduce costs through simplified installation, maintenance and troubleshooting.
Access Diagnostics through the LOI for quicker installation, maintenance, and meter verification
Rosemount Magmeter Diagnostics are available through the LOI to make maintenance of every magmeter easier.
Access Diagnostics through AMS™ Suite: Intelligent Device Manager for the Ultimate Value
The value of the Diagnostics increases significantly when AMS Device Manager is used. AMS Device Manager provides a simplified
screen flow and procedures for how to respond to the Diagnostic messages.
4 www.rosemount.com
January 2014 Rosemount 8750WA
www.rosemount.com 5
Rosemount 8750WA January 2014
SMART Meter Verification checks transmitter Deviation from baseline values are reported. Meter Calibration is verified.
and sensor characteristics.
Verification report
A verification report that can be filled out manually is available on Rosemount.com.
http://www2.emersonprocess.com/siteadmincenter/PM%20Rosemount%20Documents/00816-0200-4727.pdf
6 www.rosemount.com
January 2014 Rosemount 8750WA
www.rosemount.com 7
Rosemount 8750WA January 2014
LOI indicates high process noise is detected. Signal-to-noise ratio (SNR) is viewed in Diagnostic Improved SNR and signal stability by moving coil
menu. If the SNR <25, Diagnostic is tripped. drive frequency from 5 Hz to 37 Hz.
Diagnostic in AMS Device Manager
AMS status screen indicates high process noises detected and shows SNR at both coil drive frequencies.
AMS help provides procedure for adjusting mag coil drive frequency to improve signal stability.
8 www.rosemount.com
January 2014 Rosemount 8750WA
Service Per ASME B16.5 and ASME B16.47 for the flange selected.
1 wafer: 10 - 18 V
/2 -in. to 48-in. (15 mm to 1200 mm)
Neoprene lining
PTFE lining
(2) 30- to 48-in. AWWA C207 Class D and Class E are rated to 150 psi at atmospheric temperature.
www.rosemount.com 9
Rosemount 8750WA January 2014
Diagnostic status:
Activates switch closure output when the transmitter (2) The enclosure is suitable for constant submersion in water under given
conditions which are determined by the manufacturer (submersion).
detects a condition that meets the configured criteria of this
output.
Optional digital input function
Security lockout
10 www.rosemount.com
January 2014 Rosemount 8750WA
Polyurethane
1.5
Process wetted materials
1.0
0.5% 0.25%
Lining
0.5
Polyurethane, Neoprene, and PTFE
0 Electrodes
0 3 6 12 18 24 30 316L SST and Nickel Alloy 276 (UNS N10276)
(1) (2) (4) (6) (8) (10)
Velocity in ft/s (m/s) Process connections
Wafer sensor accuracy ASME B16.5
System accuracy is ±0.5% of rate from 3 to 30 ft/s (1 to 10 1
/2-in. to 24-in. (Class 150)
m/s); between 0.04 and 3.0 ft/s (0.01 and 0.3 m/s), the 1/2-in. to 24-in. (Class 300)
system has an accuracy of ±0.015 ft/s (0.005 m/s). Optional
high accuracy is ±0.25% of rate from 3 to 30 ft/s (1 to 10 AWWA C207
m/s). 30-in. to 48-in. (Class D)
2.5
42-in. to 48-in. (Class E)
EN 1092-1
2.0
48-in. (1200 mm) PN6, PN10
% of Rate
1.5
AS2129
1.0 48-in. (1200 mm) Table D, Table E
0.5% 0.25%
0.5 AS4087
48-in. (1200 mm) PN16, PN21
0
0 3 6 12 18 24 30
(1) (2) (4) (6) (8) (10)
Wafer sensors
Velocity in ft/s (m/s)
Non-wetted materials
Repeatability Sensor body
303 SST (ASTM A-743)
±0.1% of reading
Coil housing
Response time (analog output)
ASTM A732 Grade 1A
50 ms seconds maximum response to step change in input CS 1026 per ASTM A519
Stability CS 1020, 1026 per ASTM A513, A519
www.rosemount.com 11
Rosemount 8750WA January 2014
34
32
Electrical connections 30
28
1/2–14 NPT connections with number 8 screw terminals are
Two 26
24
provided in the terminal enclosure for electrical wiring. 22
20
Process reference electrode 80 100 120 140 160 180 200 220 240 250
Grounding rings are installed between the flange and the sensor
Figure 3. DC current requirements
face on both ends of the sensor. Single ground rings can be 1
installed on either end of the sensor. They have an I.D. slightly
Supply Current (Amps)
larger than the sensor I.D. and an external tab to attach ground
0.75
wiring. Grounding rings are available in 316L SST and Nickel
Alloy 276 (UNS N10276).
0.5
Lining protectors - (optional)
Lining protectors are installed on the flanges at both ends of the 0.25
sensor. The leading edge of lining material is protected by the
lining protector; lining protectors cannot be removed once they 0
are installed. Lining protectors are available in 316L SST and 12 18 24 30 36 42
Nickel Alloy 276 (UNS N10276). Power Supply (Volts)
Dimensions
Weight
12 www.rosemount.com
January 2014 Rosemount 8750WA
0 Mounting
10.8 30 Integrally mounted transmitters do not require additional
Power Supply (Volts)
remote cables. The local display and transmitter can be rotated
in 90° increments. Remote mounted transmitters require only a
Rmax = 31.25 (Vps – 10.8) single conduit connection to the sensor.
Vps = Power Supply Voltage (Volts)
Transmitter weight
Rmax = Maximum Loop Resistance (Ohms)
Approximately 7 pounds (3.2 kg). Add 0.5 pounds (0.5 kg) for
Figure 5. 12ES DC load limitations
local display.
1000 12ES
Materials of construction
Load (Ohms)
Housing
Operating Low-copper aluminum, NEMA 4X and IEC 60529 IP66
Region
0 Paint
Polyurethane
10.8 30
Power Supply (Volts)
Cover gasket
Silicone Rubber
www.rosemount.com 13
Rosemount 8750WA January 2014
14 www.rosemount.com
January 2014 Rosemount 8750WA
Dimensional drawings
Figure 6. Rosemount 12ES Transmitter
9.01
(229)
4.31 2.81 3.11 0.44
(109) (71) (79) (11)
12.02
(305)
11.15
(283)
LOI Keypad
Cover
2.96
(75)
1/2–14 NPT
Conduit Connection
(4 Places)
Ground Lug
www.rosemount.com 15
Rosemount 8750WA January 2014
7.49
(190)
6.48
1.94 (165)
LOI Cover
(49)
1/2-14 NPT
Electrical Conduit Connections*
5.10 3.00 (3 places)
(130) (76)
8.81
10.5 (224)
(267)
3.07
(78)
5.10
3.43 (130)
(87)
1/2-14 NPT Conduit
Connections*
(2 places)
11.02
(280)
5.82
(148)
* Conduit connections are also available with M20 connections with the use of conduit threaded adapters.
16 www.rosemount.com
January 2014 Rosemount 8750WA
Figure 8. Rosemount Flanged Sensors, typical of 1/2-in. to 4-in. (15 mm to 100 mm) line sizes
4.02 5.00
(102) (127)
2.00
(50.8)
2.6
1.8 (66)
2.6 (46)
(66)
1 H
/2-14 NPT
Conduit
Connection C
A L
Figure 9. Rosemount Flanged Sensors, typical of 6-in. to 48-in. (150 mm to 1200 mm) line sizes
4.02
(102)
2.00 5.00
(51) (127)
1.8
2.6 2.6 (66)
(46)
(66)
www.rosemount.com 17
Rosemount 8750WA January 2014
(1) When 2 grounding rings are specified, add 0.25-in. (6.35 mm) for 1/2-in. to 14-in. (15 mm to 350 mm) sensors, add 0.50-in. (12.7 mm) for 16-in. (400 mm) and larger. When
lining protectors are specified, add 0.25- in. (6.35 mm) for ½-in to 12-in. (15 mm to 300 mm) sensors, add 0.50- in. (12.7 mm) for 14-in. to 36-in. (350 mm to 900 mm) sensors.
18 www.rosemount.com
January 2014 Rosemount 8750WA
Figure 10. Rosemount Wafer Sensors, typical of 1/2-in. and 1-in. (15 mm and 25 mm) line sizes
Figure 11. Rosemount Wafer Sensors, typical of 11/2-in. to 8-in. (40 mm to 200 mm) line sizes
www.rosemount.com 19
Rosemount 8750WA January 2014
20 www.rosemount.com
January 2014 Rosemount 8750WA
Example: SI units
Magmeter Size: 100 mm (factor from Table 7 = 492.78)
Normal Flow Rate: 800 L/min
800 (L/min)
Velocity =
492.78
www.rosemount.com 21
Rosemount 8750WA January 2014
22 www.rosemount.com
January 2014 Rosemount 8750WA
Note
For the 32E transmitters with a Local Operator Interface
(LOI), the lower ambient temperature limit is -20 °C.
NH FM Ordinary Locations
General Purpose Fire and Shock
Safety Requirements - ANSI/ISA 61010-1 (FM/US)
Enclosure Type 4X, IP66
12ES (-40 °C Ta 40 °C)
32ES (-50 °C Ta 60 °C)
www.rosemount.com 23
Rosemount 8750WA January 2014
24 www.rosemount.com
January 2014 Rosemount 8750WA
www.rosemount.com 25
Rosemount 8750WA January 2014
Ordering information
Model Product description
8750WA Magnetic Flowmeter System for Water Industries
Code Transmitter class
32ES Traditional Transmitter with Aluminum Housing - 0.5% Ref Acc with Option for 0.25%
12ES Wall Mount Transmitter with Premier LOI Capabilities - 0.5% Ref Acc with Option for 0.25%
0000 Spare sensor only - Unit shipped without a transmitter.
Code Transmitter mount
T Integral Mount(1)
R Remote Mount
Code Power supply
1 AC Power Supply (90-250 V AC, 50–60 Hz)
2 DC Power Supply (12–42 V DC)
0 Spare sensor only - Unit shipped without a transmitter.
Code Outputs
A 4-20mA Output with Digital HART Protocol & Scalable Pulse Output
0 Spare sensor only - Unit shipped without a transmitter.
Code Conduit entry
1 1/2-14 NPT, Standard Conduits (12ES,Qty=4; 32ES, Qty=2)
2 CM20, Standard Conduits (12ES,Qty=4; 32ES, Qty=2)
4 1/2-14 NPT, Additional Conduits (12ES,N/A; 32ES, Qty=3)
5 CM20, Additional Conduits (12ES,N/A; 32ES, Qty=3)
0 Spare sensor only - Unit shipped without a transmitter.
Code Sensor style
F Flanged Style
W Wafer Style(2)(3)
0 Spare transmitter only - Unit shipped without a sensor.
Code Lining material
P Polyurethane
N Neoprene
T PTFE
0 Spare transmitter only - Unit shipped without a sensor.
Code Electrode material
S 316L Stainless Steel
H Nickel Alloy 276 (UNS N10276)
0 Spare transmitter only - Unit shipped without a sensor.
Code Electrode type
A 2 Measurement Electrodes
B 2 Bulletnose Measurement Electrodes
E 2 Measurement Electrodes plus 1 Reference Electrode
0 Spare transmitter only - Unit shipped without a sensor.
26 www.rosemount.com
January 2014 Rosemount 8750WA
www.rosemount.com 27
Rosemount 8750WA January 2014
Code Options
Displays
M4 Local Operator Interface
M5 LCD Display only
Remote Mounting Options
B6 Stainless Steel 4-Bolt Kit for 2-in. Pipe Mount (Transmitter Class 12ES only)
Wafer Mounting Kit Options
MK2 Carbon Steel Mounting Studs, and Nuts(5)
Ground Rings
G1 (2) 316L SST Ground Rings
G2 (2) Ni Alloy C-276 Ground Rings
G5 (1) 316L SST Ground Ring
G6 (1) Ni Alloy C-276 Ground Ring
Lining Protectors(4)
L1 (2) 316L SST Lining Protectors
Calibration
D1 High Accuracy (0.25%) -(must order transmitter with sensor)
Custom Software Configuration
C1 Custom Software Configuration (requires CDS form)
Tagging
DT Heavy Duty Tagging
Paint
V1 Coal Tar Paint (Submersible/Direct Burial), does not meet Type 4x. Remote mount only
American Recovery and Reinvestment Act
US ARRA Compliance Certificate
Other Agency Approvals
DW NSF Drinking Water Approval(10)
Submergence Protection (Sensor Only)(11)(12)
SA Potted Junction Box with 50 feet of Combo Cable
SB Potted Junction Box with 100 feet of Combo Cable
SC Potted Junction Box with 150 feet of Combo Cable
SD Potted Junction Box with 200 feet of Combo Cable
SE Potted Junction Box with 250 feet of Combo Cable
SF Potted Junction Box with 300 feet of Combo Cable
Q Certs
Q4 Calibration Certificate per ISO 10474 3.1B / EN 10204 3.1
Typical Model Number: 8750WA 32ES T A 1 F P S A 030 C A1 NH DA2 M4 DW
28 www.rosemount.com
January 2014 Rosemount 8750WA
(10) Only available with Lining Materials: PTFE (all line sizes) or Polyurethane (4-in. or larger) and Electrode Materials: 316L SST or Ni-Alloy 276.
(12) Provides pre-wired potted remote junction box, sealed conduit gland, and conduit plug. Requires the use of sealed conduit to meet IP68 requirements.
www.rosemount.com 29
Rosemount 8750WA January 2014
Additional configurations require the C1 option and a (1) For remote mount installations, combination electrode and coil drive
completed Configuration Data Sheet (CDS) for custom cable should be limited to less than 330 ft (100 m).
30 www.rosemount.com
January 2014 Rosemount 8750WA
www.rosemount.com 31
Rosemount 8750WA Product Data Sheet
00813-0100-4750 Rev FA January 2014
Emerson Process Management Emerson Process Management Asia Pacific Emerson Process Management Latin America
Rosemount Inc. Private Limited Multipark Office Center
8200 Market Boulevard 1 Pandan Crescent Turrubares Building, 3rd & 4th floor
Chanhassen, MN 55317 USA Singapore 128461 Guachipelin de Escazu, Costa Rica
www.rosemount.com T (65) 6777 8211 T+(506) 2505-6962
T (USA) +1 800-406-5252 F (65) 6777 0947 international.mmicam@emersonprocess.com
T (International) +1 (303) 527 5200 Enquiries@AP.EmersonProcess.com
F +1 (303) 530 8459