Customer Installation Instructions

90496 Ultraview® Command Module Noninvasive Blood Pressure (NIBP)


Performance Upgrade Kit (P/N 040-1487-00)

Caution
Observe precautions for handling electrostatic-sensitive devices.

Note:
• Never touch electrostatic-sensitive electronic components without following proper anti-static procedures,
including the use of an ESD wrist band and mat. An electrostatic discharge from your fingers can
permanently damage electronic components.
• All static-sensitive electronic components are packaged in static-shielding bags. Retain the bag for
repackaging the component should you need to store it or return it to Spacelabs Medical for any reason.

Kit Contents

Verify that the following parts are available before beginning the installation.

Quantity Description Part Number

1 PCBA, CPU/NIBP, 90496 670-0882-04

1 Label, Module, SW version, CE 334-2495-xx

Installation instructions, 90496


Ultraview Command Module
1 075-0464-00
Noninvasive Blood Pressure (NIBP)
Performance Upgrade Kit

Documents Required

Ultraview Command Module Service Manual (P/N 070-0752-xx), located on CD-ROM P/N 084-0700-xx.

Installation Instructions

1 Remove the screws from the outer covers and remove the module side covers.
2 Carefully separate the PCBAs.
3 Remove any interconnecting cables.
4 Remove and set aside the CPU PCBA and metal shield.
5 Install the new CPU PCBA into the module.
6 Reassemble the module.
7 Follow the Performance Verification procedures specified in the Ultraview Command Module Service Manual
to confirm the functionality of the module.
8 Return the removed CPU PCBA and shield to:
Spacelabs Medical, 22011 SE 51st St, Issaquah, WA 98029.

Ultraview is a trademark of Spacelabs Medical, Inc.

075-0464-00 Rev. B 1 of 1