These instructions describe the procedure for converting an Ultraview 1030 (90367), 1050 (90369), 1500 (90363),
1600 (90364), 1700 (90387), or Universal Clinical Workstation (UCW®) (90385) to use AUI (10Base5) instead of
10BaseT cabling.
CAUTION:
• Printed circuit boards in this equipment contain static sensitive devices;
only handle at a static-safe workstation.
!
• This kit is applicable for use with the following models:
90363 — all
90364 — all
90367 — all
90369 — all
90385 — all units with serial numbers 385-3xxxxxx or above
90387 — all units with serial numbers 387-1xxxxxx or above
Contents
Verify that all parts are available in the kit before beginning this installation.
075-0147-00 Rev. E 1
Installation Instructions
!
• There are many possible configurations in which this kit may be installed. A suggested
configuration for a 90369 or 90367 monitor is shown in Figure 1. Questions about the installation
for a particular configuration should be directed to the local Customer Service Representative.
nurse alert
connector
power cable
(P/N 012-0590-02)
8. Attach the 10BaseT Ethernet cable (P/N 175-0951-00) and power cable (P/N 012-0590-02) to the media
converter, and attach the other end of the 10BaseT Ethernet cable to the Ethernet connector on the rear of
the monitor.
9. Coil the 10BaseT Ethernet cable and use the remaining three cable ties (P/N 346-0004-00) to secure the
loops that have been made.
10. To complete the installation, plug in an AUI (10Base5) cable to the media converter and power ON all related
equipment.
Ultraview and UCW are trademarks of Spacelabs Medical, Inc.
2 075-0147-00 Rev. E