OPX/105

TECHNICAL DATA GENERAL INFORMATIONS


When mounting tube inserts adopt proper caution, in order to avoid glass bulb breaking
and fragments projection. Please use protective gloves and glasses.
Voltage Tube insert connected to H.V. supply is a radiation source: be sure to take all necessary
nominal 110* safety cautions
kV
maximum for test 121
Focal spot ((IEC60336:2005) 0.5 mm  Wash thoroughly with alcohol the external surface of tube insert (care of fire risk).
Remove any labels on the glass bulb. Avoid contact of dirty surfaces with cleaned
4.0 ÷ 8.0 V
Filament characteristics tube insert.
2.5 ÷ 4.0 A
Anode material tungsten  Clamp system inside housing or self-contained units must not mechanically stress the
Target angle 5 ° tube.
Anode heat storage capacity 30000 J
 After installation, check the right working of the tube (no fluctuation of tube current
Maximum anode cooling rate 250 W nor crackling)
Nominal anode input power at 0.1 s (DC) 2000 W
Inherent filtration 0.5 mm Al  Comply with insert thermal parameters, planning and programming the exposure
parameters and cooling pauses. Housing or self-contained units must be provided
Maximum diameter 42 mm
with an adequate thermic protection.
Overall length 159 max mm
 Voltages indicated in charts are valid for transformer supplied with ground center.
* Maximum allowed exposure time for 110 kV is 25 s
 Tube inserts contain environment polluting materials, particularly lead liner tubes.
Please apply to qualified operator for waste disposal, according to local regulation
requirements.

INCIDENT REPORT ACCORDING TO 93/42/EEC MEDICAL DEVICES DIRECTIVE


In order to comply with CE marking requirements, end users have to report to local Competent
Authority all the informations about possible incidents involving the device, regarding any
deterioration in its characteristics and performances, as well as any inaccuracies in this
documentation, which might lead to or might have led to the death of patient / user or a
deterioration in his state of health. This information must be promptly reported also to C.E.I. in order
to start manufacturer reporting, as per above mentioned directive.

this mark assures device conformity to


EC Directive 93/42 on Medical Devices Safety.
0051

Rev.N - 0514 - Original text: italian Page 1 of 4 SKANX RADIOLOGY DEVICES Srl – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OPX/105
DIMENSIONS

Rev.N - 0514 - Original text: italian Page 2 of 4 SKANX RADIOLOGY DEVICES Srl – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OPX/105
650 W 550 W 450 W 300 W
THERMAL
30 CURVES 250 W EMISSION
25 CHARACTERISTICS 50 FILAMENT
8.0 CHARACTERISTICS
kJ
7.5
25
200 W 20 40
7.0

V
mA
20
6.5
15 30

filament voltage
15 6.0

anode current
10 20 5.5
70 kV
10
5.0
5 10
5
4.5

0 0 0 4.0
2.5 3.0 3.5 4.0 4.5
0 1 2 3 4 5 6 7 8 min 9 2.5 3.0 3.5 4.0 4.5
filament current A filament current A

RATING CHARTS

40 kV
30 40 40 kV
50 kV 35
25 50 kV
anode current (mA)

anode current (mA)


60 kV 30 60 kV
20 70 kV
70 kV 25
80 kV 80 kV
15 20 90 kV
90 kV
100 kV 15 100 kV
10 105 kV
105 kV
10
5 5

0 0
0,1 1 10 0,1 1 10
exposure time (sec)
exposure time (sec)
Rev.N - 0514 - Original text: italian Page 3 of 4 SKANX RADIOLOGY DEVICES Srl – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OPX/105
EMISSION CHARACTERISTICS DC

50 40 kV 40 kV 18

45 16

40 50 kV
anodic current (mA)

50 kV 14

35 60 kV

mA
12
30 70 kV 60 kV
70 kV 10
25 80 kV

anode current
80 kV 100 kV
90 kV
20 100 kV 90 kV 8 60 kV
100 kV
15 110 kV 105 kV
6

10 4

5 2

0 0

0,1 1 10 100 1000 3 3,1 3,2 3,3 3,4 3,5 3,6 3,7

Filament current A
exposure time (sec)

Rev.N - 0514 - Original text: italian Page 4 of 4 SKANX RADIOLOGY DEVICES Srl – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy