Spirometers
k To outline the characteristics of currently available types of spirometer. Lungfunction Lab. Poli 12
k To explain the requirements a spirometer should fulfil. Medisch Spectrum Twente
Haaksbergerstraat 55
7500 KA Enschede
Summary The Netherlands
Fax: 31 534872658
There are many spirometers on the market. They can be divided roughly into a group that E-mail:
f.h.dejongh@amc.uva.nl
primarily measures volume and a group that primarily measures flow. Quality control and
calibration should be performed on a regular basis to ensure that a device (still) functions
appropriately. The types have different characteristics and are optimal for different cir-
cumstances (from a simple peak-flow meter to monitor asthma at home to an ultrasonic
device to measure a patient in a pulmonary function lab). In this article, the characteris- Competing interests
tics of the currently available spirometers are discussed, as well as the requirements they None declared
should fulfil.
Provenance
j From cheap and easy devices usually seen
in primary care settings, to the sophisti-
cated and expensive devices used in pul-
dry spirometers).
• Flow-measurement devices (e.g. pneumo-
tachographsystems, mass flow meters).
Adapted from an ERS School
Course
monary function labs, the range of spirometers In the first case, the flow is calculated by
on the market is huge. Simple versions are the time derivative of the volume signal. In the
capable of measuring basic values such as vital second case, volume is calculated by integra-
capacity (VC), forced expiratory volume in one tion over time of the flow signal.
second (FEV1) and peak expiratory flow (PEF), An American Thoracic Society
while others are capable of displaying almost (ATS)/European Respiratory Society task force
any lung function parameter imaginable. There on the standardisation of lung function testing
are many different ways to measure the flows has written articles containing a great deal of
and/or volumes generated by the respiratory detailed information about spirometry. The
system. An anatomical cast of a lung shows tables and some of the information in the pres-
the complexity of the system that we want to ent article are taken from the second paper in
quantify in a few parameters which are of max- the series [1]. In brief, the requirements of an
imal clinical relevance (figure 1). Since the val- acceptable spirometer are:
ues of these parameters often change slowly • Spirometers must be able to accumulate
with age, the measurement system (i.e. the volume for ≥15 s.
spirometers) needs to measure as accurately • The measuring volume should be ≥8 L (body
and reproducibly as possible. temperature and pressure, saturated).
Spirometers can be divided into two basic • The accuracy of reading should be at least
groups. ±3% (or ±0.05 L) with flows from 0–14 L
• Volume-measurement devices (e.g. wet and per s.
Table 1 Range and accuracy recommendations specified for forced expiratory manoeuvres
Test Range/accuracy (BTPS) Flow range Time s Resistance and Test signal
L per s back pressure
VC 0.5–8 L, ±3% of reading or 0–14 30 3-L calibration syringe
±0.050 L, whichever is greater
FVC 0.5–8 L, ±3% of reading or 0–14 15 <1.5 cmH2O per L per s 24 ATS waveforms,
±0.050 L, whichever is greater (0.15 kPa per L per s) 3-L calibration syringe
FEV1 0.5–8 L, ±3% of reading or 0–14 1 <1.5 cmH2O per L per s 24 ATS waveforms
±0.050 L, whichever is greater (0.15 kPa per L per s)
Time zero The time point from which Back extrapolation
all FEVt measurements are taken
PEF Accuracy: ±10% of reading or 0–14 Mean resistance at 200, 400, 26 ATS flow waveforms
±0.30 L per s (20 L per min), 600 L per min (3.3, 6.7, 10 L per s)
whichever is greater; repeatability: must be <2.5 cmH2O per L per s
±5% of reading or ±0.15 L per s (0.25 kPa per L per s)
(10 L per min), whichever is greater
Instantaneous Accuracy: ±5% of reading or 0–14 <1.5 cmH2O per L per s Data from manufacturers
flows (not PEF) ±0.200 L per s, whichever is greater (0.15 kPa per L per s)
FEF25–75% 7.0 L per s, ±5% of reading or ±14 15 Same as FEV1 24 ATS waveforms
±0.200 L per s, whichever is greater
MVV 250 L per min at VT of 2 L within ±14 (±3%) 12–15 <1.5 cmH2O per L per s Sine wave pump
±10% of reading or ±15 L per min, (0.15 kPa per L per s)
whichever is greater
BTPS: body temperature and ambient pressure saturated with water vapour; FVC: forced vital capacity; FEVt: forced expiratory volume in t seconds;
FEF25–75%: mean forced expiratory flow between 25% and 75% of FVC; MVV: maximum voluntary ventilation; VT: tidal volume. Reproduced from [1].
Table 2 Recommended minimum scale factors for time, volume and flow on
graphical output
as FEV1 and the VC. They are simple, under- on the mesh. To (partially) avoid this problem, the
standable (good for teaching), no electronics are mesh is often heated and a filter can be used.
needed and there is a direct overview of the val- Spirometers based on the principle of a rotat-
ues measured. On the other hand, these devices ing vane do not need to be heated, which should
are no longer produced, have a limited range of ensure that better linear and high ranges of flows
parameters (e.g. maximum expiratory flow at are possible. Such devices are often used for
25% of VC is very difficult to obtain) and may office spirometry and also during maximal
have problems caused by the inertia of the mov- ergometry. A disadvantage is that they have mov-
ing drum, which may lead to an overshoot of the ing parts (the rotating vane), which are fragile
volumes measured. and easily broken. Their low-flow characteristics
Dry spirometers (e.g. dry rolling seal or bel- are dubious (it takes a certain flow-rate before the
lows) generally have better frequency characteris- vane begins to rotate) and at high flows the vane
tics and are more easy to automate, but they too may continue rotating due to inertia after the
are almost unobtainable. flow has stopped.
Mass flow sensors (anemometers) are based
Flow-measurement spirometers on heated wires that are cooled by the flow of air
The most widely used system in pulmonary func- generated by the patient. They have no moving
tion labs is the pneumotachograph spirometer. parts, are independent of air temperature, mois-
Early systems had a Fleisch or Lilly pneumo- ture or viscosity, have good frequency character-
tachograph, while nowadays the pressure drop istics and are simple to automate. However they
over a resistance (a heated fine mesh) is meas- are non-linear (requiring a difficult software cor-
ured. The pressure drop divided by the resistance rection) and often unstable. The sensor-resistance
of the pneumotachograph yields the flow, which is connected in series with cables and connectors,
can be transformed into a volume by time inte- and changes in those components are also meas-
gration. ured as if they were air-flows. These devices can
These devices have no moving parts, have be used in humid, pressurised environments (for
good frequency characteristics and are simple to instance, for ventilated patients), whereas other
automate, but since the volume is calculated by spirometers have difficulties under these condi-
time integration, calibration is difficult (the cali- tions.
bration signal is normally a syringe, which gener- The most modern spirometers often contain
ates a fixed volume rather than a known flow). an ultrasonic flow-head. They have no moving
Also, linearity may be a problem if the flow is not parts, have no resistance (nothing is blocking the
laminar, and the devices can become contami- flow) and have good frequency characteristics.
nated by sputum or moisture, which condenses The signal depends in general on the air-mixture,
Filters
A common question is whether spirometers
temperature, moisture and pressure. This is both should be used in combination with filters. When
a difficulty and an advantage, since a lot of other a patient inhales from a device, everything should
measurements may be performed with the same be cleaned (flush a closed system with at least
sensor (especially when using gases with a differ- five times the maximal volume of that system).
ent molecular weight like SF6). However, by clever The tubing, mouthpiece and resistive device
choices the flow can be calculated independently should be replaced. The alternative is to use a
of gas composition, temperature, humidity and low-resistance one-way valve, or to use disposable
pressure. A drawback is that these sensors are filters and/or disposable (flow) sensors or inserts.
often more expensive. Manufacturers have a responsibility to make
devices that are easy to dismantle and to clean.
The purchaser should always ask the manufac-
Accuracy turer how a device can be cleaned.
References
1. Miller MR, Hankinson J, Brusasco V, et al. Standardisation of spirometry. Eur Respir J 2005; 26: 319–338.
2. American Thoracic Society. Standardization of spirometry, 1994 update. Am J Respir Crit Care Med 1995; 152: 1107–1136.