Sie sind auf Seite 1von 4

1

Mechanical characterization of surgical templates


used for guided dental implant placement
Narayanan Kakkad, Katrien Baeck and Jos Vander Sloten


Abstract— This study evaluated the mechanical strength
characteristics of surgical templates used for guided dental
implant placement, based on the location, angle of drill and
materials. The template was evaluated using both theoretical
calculations based on analytical formulas and finite element
analysis. The template was evaluated for different loading cases
and procedures. A detailed mesh convergence analysis was
performed to determine the optimal mesh. Four different cases
were studied comprehensively. The results showed that the
maximum stresses produced in the templates for each of the cases
for the two surgical procedures; drilling and implant insertion
were lower than the flexural strength of the material (95MPa). The
current material used for production was found to be superior to Fig. 1 - (a) Case 1; (b) Case 2; (c) Case 3; (d) Case 4. The red circle indicate
the other two available biomaterials used for template production. the loading position.
The change in inclination of drill did not have any major impact
on the stress produced in the template during drilling procedure. A. Assumptions
Thus the current design used for template production was In order to adapt the loading cases to beam theory, several
determined to be safe in the cases studied under the loads applied. assumptions were made, which are listed below:
Index Terms—Guided dental implant placement, Surgical
template, Nobel Biocare, Beam theory, Mesh convergence 1. The template is assumed to be a beam of uniform
analysis, Finite element analysis. rectangular cross section, the dimensions of the rectangular
cross section are as given below. The template is made out of
I. INTRODUCTION Renshape, for which young’s modulus, E= 2500MPa [5].
2. Breadth, b = 10mm.
R ECENTLY the conventional two staged treatment procedure
for dental implant placement has been replaced by
immediate loading of implants using guided implant placement
3. Depth, d = 2.5mm (template thickness=2.5mm) [3].
4. Moment of Inertia, I = bd3/12 = 13.02mm4.
techniques [1] . Clinicians use computer aided surgical planning 5. Section Modulus, Z = bd2/6 = 10.41mm3.
tools provided by Nobel Biocare to determine the patient 6. Supports are assumed to be the sharp edges of the tooth.
pathology and to come up with a surgical plan and also decide 7. The original template has metallic sleeves and drill guides
the optimal implant placement [2]. supporting the drilling and implants insertion steps. This makes
Finally the surgical planning data is translated into the template a composite beam. Since this would make the
operating room through aid of surgical templates which have calculations too complex which is not possible using analytical
metallic sleeves for guiding the drilling and implant insertion solutions, the metallic sleeves and drill guides are not taken into
procedures [3]. The templates have been used extensively in consideration.
recent past. However, there has been no comprehensive study 8. Thus, the entire load is assumed to be taken by the acrylic
performed to determine the mechanical strength characteristics part of the template.
of these surgical templates. 9. For the drilling procedure a load of 100 N was used as the
The objective of the present study was therefore to evaluate maximum force acting on the template [4].
the mechanical strength characteristics of surgical templates 10. For the implant insertion procedure, it was assumed that,
used for guided dental implant placement, based on the the clinician would make an extra half turn of the implant
location, angle of drilling and materials. resulting in a deflection of the template by a distance half the
pitch [3] of the implant thread (0.375mm).
II. THEORETICAL CALCULATIONS B. Calculations
There were four different partially edentulous cases chosen Each of the cases discussed previously were simplified so as
for study (Fig. 1). For each of these cases, the drilling procedure to apply analytical equations of beam theory. The simplified
and the implant insertion procedures were separately studied. models for each of the cases have been depicted in the Fig. 2.
Initially the study involved application of beam theory using For determining the maximum stress during drilling procedure,
various analytical formulae to determine the stresses involved in case1-4 the equations (1), (2), (3) and (4) respectively were
in each case. used [6].
2

𝑊𝑎𝑏
𝑠𝑚𝑎𝑥 = − (1) body as the cast will undergo no deformation compared to the
𝑍𝑙
template. In this study, four-node, first-order, one integration
𝑊𝑙
𝑠𝑚𝑎𝑥 = − (2) point tetrahedral elements of type C3D4 were used [7]. A
4𝑍
rigorous mesh-sensitivity study was carried out to obtain a
𝑊𝑐
𝑠𝑚𝑎𝑥 = (3) computationally accurate finite-element mesh. In the current
𝑍
model, we are interested in determining the stress developed in
𝑊𝑐
𝑠𝑚𝑎𝑥 = (4) the template for each of procedures [8] [9]. Thus, the history of
𝑍
the stress-time signal is used as the criterion of convergence.
where, smax is the resulting maximum stress produced in the All the results are presented based on simulations performed
template, W the load of 100 N, a, b, c and l are dimensions with an optimized mesh. Renshape is the material used to
related to the beam length and point of action of load with produce templates at Nobel Biocare. Thus a young’s modulus
respect to the supports of the beam and Z is the section modulus of 2500MPa and Poisson’s ratio of 0.3 was defined for the
which has already been defined. template.
For determining the maximum stress during implant In this model it is very difficult to define a separate part with
insertion procedure as well, for case 1-4 the equations (1), (2),
properties of the sleeve and then to define a surface to surface
(3) and (4) respectively were used. In each of those cases the
interaction between the template and the sleeve, as the two parts
load W was determined from the equation of deflection using
the equations (5), (6), (7) and (8) respectively [6]. in real are glued to each other. Due to the complex nature of
interaction between the two parts this has been avoided and
𝑦𝑚𝑎𝑥 = 𝑊𝑎2 𝑏 2 /3𝐸𝐼𝑙 (5)
instead the nodes inside the drill hole of the template were
𝑦𝑚𝑎𝑥 = 𝑊𝑙 3 /48𝐸𝐼 (6) selected and constrained as a rigid body using tie nodes.
A surface to surface contact was defined between the two
𝑦𝑚𝑎𝑥 = 𝑊𝑐 2 (𝑐 + 𝑙)/3𝐸𝐼 (7) interacting surfaces. The cast was defined to be the master and
𝑦𝑚𝑎𝑥 = 𝑊𝑐 2 (𝑐 + 𝑙)/3𝐸𝐼 (8) template the slave surface as cast is a rigid body [10]. Finite
sliding algorithm has been implemented as recommended for
Where, ymax is the resulting maximum deflection due to the extra static analysis. A tangential behavior has been defined between
half turn of the implant (0.375mm), W the resulting load and E the two surfaces with a friction coefficient of 0.3 [11]. Similarly
the young’s modulus of the material. Thus using these a normal behavior of hard contact has been defined.
equations the corresponding loads produced for each case was
determined.

Fig. 2 – Beam theory applied to every case. (a) Case 1; (b) Case 2; (c) Case 3
and Case 4. Fig. 3 – Loading in every case. (a) Case 1; (b) Case 2; (c) Case 3; (d) Case 4.
The pink colored loads represent the clinician and anchor load and the orange
colored load represents the drilling load.
III. FINITE ELEMENT MODELLING
The finite element modelling and analysis was performed in
an Intel Xeon CPU E5-2600 with a clock speed of 2.20GHz, B. Boundary conditions and loads
with a RAM of 32GB and 64-bit operating system. In this study The cast, earlier constrained to be a rigid body, was fixed on
the meshing tool in use is ANSA (version: 15.2.3). ANSA is a all of its degrees of freedom. The major loads applied were the
meshing software which is capable of creating mesh of high clinician load, anchor load, drilling load and implant insertion
quality from an STL file and create an input file for Abaqus load (Fig. 3). The clinician load of 25N was applied which
software. The FE calculations have been performed using the represents the force with which the clinician presses the
Abaqus 6.13 software package. template using his/her fingers [12]. An anchor load of 25N was
applied, which is normally achieved using the anchor pins
A. Material properties and interactions
which stabilizes the template. For the drilling step a drill load
The cast in the current analysis has been assumed to be a rigid of 100N was applied [4]. For implant insertion a displacement
3

boundary condition was applied on the sleeve of 0.35mm [3]. of angle the applied load makes with the face of the drilling
guide to see its effects on the maximum stress resulting in each
IV. RESULTS case and to come up with a comparison [15]. Even though there
For all the four cases, results were obtained using theoretical was a slight degree of increase in the stress, from the results it
as well as finite element analysis. The results from the was clear that the increase was not significant. So, the effect of
theoretical calculations are listed in the Table I. From the results the angle of inclination of the applied load is not of a big issue.
it can be inferred that the template would remain safe under More importantly the maximum stresses produced are all within
most of the cases except for the drilling procedure in case-4, the safe limits and the template is safe in all cases.
TABLE I
RESULTS FROM THEORETICAL CALCULATIONS
Smax Smax
Case
Drilling (MPa) Implant insertion(MPa)

Case-1 48.4 27.2


Case-2 48.03 35.18
Case-3 48.03 46.9
Case-4 144.09 9.4

since the stress is above the ultimate flexural strength of the


material (95MPa) [5].
Finite element analysis was performed for each of the four
cases for the two procedures under study. Both quantitative and
qualitative results were determined. For quantitative results
extraction, python scripting in Abaqus was performed. The
results have been quantified in Table II, Fig. 4 and Fig. 5. The
TABLE II
RESULTS FROM FINITE ELEMENT ANALYSIS
Smax Smax
Case Fig. 5 – Implant insertion results (MPa). (a) Case 1; (b) Case 2; (c) Case 3; (d)
Drilling (MPa) Implant insertion(MPa)
Case 4.

Case-1
A study was also performed to compare the results with
41.4 8.32
Case-2 different materials. In this study, the case-1 was used with a
56.7 32.2
Case-3 32.3 2.09 different material property in each case. The two materials with
Case-4 44.63 2.09 which the current design was compared were- Accura
ClearVueTM [16] and DS 3000 [17]. When compared to the
other two materials used in the industry Renshape was found to
maximum stresses were determined in a region of interest,
be superior in its performance from the results obtained as
which was defined close to the vicinity of the applied load.
maximum stress produced in both cases were close to the value
obtained with Renshape, but the flexural strength of both the
materials were low as compared to Renshape [5].

V. DISCUSSION
There was no previous work related to study of mechanical
strength characteristics of surgical templates found in the
literature review. Due to the lack of previous work in this field
a new methodology had to be defined. Previous studies in the
field of determining drill force was used as a reference in the
present study. For the implant insertion step it was assumed that
an extra half rotation of the implant would be made by the
clinician.
As far as the theoretical calculations are concerned, beam
theory was applied to these situations. The template was the
beam and the drilling and implant insertion loads were assumed
to be acting as point loads over the template. A lot of
assumptions were made to simplify the complex loading
scenario, to apply beam theory. Thus the results obtained from
Fig. 4 – Drilling results (MPa). (a) Case 1; (b) Case 2; (c) Case 3; (d) Case 4.
these analytical equations were far from the real results. Even
A study using the case-1 was performed to study the effect
then these results would show a general trend.
4

A Finite Element study was performed as well. A mesh REFERENCES


convergence was performed to determine the optimal mesh size [1] Z. G. A. T. Brånemark P-I, Tissue integrated prosthesis. Osseointegration
[13] [14]. The model was assumed to be a static model. Since in Clinical Dentistry, Chicago: Quintessence Publishing Co., 1985.
[2] Chatzistavrou M, Felton DA, Cooper LF. Immediate loading of dental
it’s a static model the maximum load during drilling and
implants in partially edentulous patients: a clinical report. J Prosthodont.
implant insertion is applied in on single step, but using a ramp. 2003.
Since a surface to surface contact definition has been applied [3] NobelGuide®: Treatment workflow for the partially edentulous patient.
with a tangential behavior defined, some non-linearities are NobelGuide® Procedures manual. 2014.
[4] Wendong Wang, Yikai Shi, Ning Yang, Xiaoqing Yuan. Experimental
expected in the model [10]. Therefore accordingly the analysis of drilling process in cortical bone. Medical Engineering &
incrementation of the step module needs to be controlled. Physics 2014; 36: 261– 266.
The model used constrains to define the steel sleeves inserted [5] RenShape™ SL 7570 datasheet. Huntsman. Last reviewed: 09 August
2015. [Online]. Available at:
as guides for the drills. For future models a proper sleeve should http://www.rpc.msoe.edu/datasheets/sla/Huntsman_RenShape_SL_7870
be modelled with an interaction defined between the sleeve and .pdf
the template. The cast was fixed at all its degrees of freedom. [6] Structural beam deflection and stress formula and calculation. Last
reviewed: 09 August 2015. [Online]. Available at:
The template was loaded using pressure load, where the load https://www.engineersedge.com/beam_calc_menu.shtml
would be applied at angle of 90 degrees to the drill surface. This [7] R. Huiskes, E.Y.S. Chao. A survey of finite element analysis in orthopedic
case would be an ideal case. In real life some minor deviation biomechanics: The first decade. Journal of Biomechanics 1983; 16(6):
385–409.
from this is expected. Thus a study with different angles were [8] Cifuentes, A.O., and Kalbag, A. A performance study of tetrahedral and
also performed, giving an idea of how a change in angle would hexahedral elements in 3-D finite element structural analysis. Finite
affect the stresses developed in the template. It was found that Elements in Analysis and Design, 1992; 12: 313-318.
[9] A comparison of all-Hexahedra and all Tetrahedral Finite Element
though there is a correlation between the two, it is not a critical
Meshes for elastic & elastoplatic analysis. Proceedings 4th International
factor as the variation in the stress with every 1 degree change Meshing Round table Sandia National Labs, Oct. 1995; 179-181.
in angle was estimated to be around 0.1MPa. [10] Abaqus 6.13 documentation. Last reviewed: 09 August 2015. [Online].
The results obtained from the simulations showed that, the Available at: http://129.97.46.200:2080/v6.13/books/usb/default.htm
[11] Parekh J, Shepherd DET, Hukins DWL, Hingley C, Maffulli N. In vitro
templates would remain safe under the applied loads and investigation of friction at the interface between bone and a surgical
boundary conditions. The drilling and implant insertion process instrument; Journal of Engineering in Medicine 2013; 712-718.
invariably resulted in a maximum stress which was less than [12] Finger force capability: measurement and prediction using
anthropometric and myoelectric measures. Last reviewed: 09 August
ultimate flexural strength of the material. In most of the cases 2015. [Online].
this was less than half the strength of the material. In both the Available at: http://scholar.lib.vt.edu/theses/available/etd-01132000-
procedures, case 2 showed highest maximum stress values. The 12570058/unrestricted/thesis_astin.pdf
[13] The Importance of Mesh Convergence 1. Last reviewed: 09 August 2015.
stress produced due to drilling increased with an increase in [Online].
distance from the support in case 3 and 4. For implant insertion Available at: http://www.nafems.org/join/resources/knowledgebase/001/
though with the same displacement the stress decreased with an [14] The Importance of Mesh Convergence 2. Last reviewed: 09 August 2015.
[Online].
increase in distance from the nearest support. The study with Available at: http://www.nafems.org/join/resources/knowledgebase/002/
two other materials (Accura ClearVueTM and DS3000) used in [15] Marcelo Michele Novellino, Newton Sesma, Dalva Cruz Laganá, Glais
the market was also performed. The results from that study also Ferrari. Linear and Angular Deviations of Implants Placed in
Experimental Casts with Stereolithographic Drill Guides Fixed by O’ring
validated that, the current material is superior in its mechanical
Ortho Implant Devices. Brazilian Dental Journal 2013; 24(4): 391-396.
strength characteristics. [16] Accura® ClearVue™ Material. 3D Systems. Last reviewed: 09 August
Thus the finite element model created provides a very robust 2015. [Online].
and dependable option which takes into account all possible Available at:
http://www.paramountind.com/pdfs/PI_3DS_Accura_Clearvue_US.pdf
complexities if the real life case. Though some assumptions [17] DS 3000 datasheet. DWS Systems. Last reviewed: 09 August 2015.
have been made, they are taken with due consideration. The [Online].Available at: http://www.dwssystems.com/wp-
present work provides a good model for reference for any future content/uploads/documents/ds_3000_d_technical_datas.pdf.
work in this field.

Das könnte Ihnen auch gefallen