Sie sind auf Seite 1von 3

Designation: C 1580 – 05

Standard Test Method for


Water-Soluble Sulfate in Soil1
This standard is issued under the fixed designation C 1580; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope 4. Apparatus
1.1 This test method is for the determination of water- 4.1 Photometer—One of the following, given in order of
soluble sulfate in soils. preference:
1.2 This test method was developed for concentrations of 4.1.1 Nephelometer or turbidimeter,
water-soluble sulfate in soils between 0.02 and 3.33 % sulfate 4.1.2 Spectrophotometer for use at 420 nm with light path of
by mass. 4 to 5 cm, and
1.3 This test method does not determine sulfur in any form 4.1.3 Filter photometer with a violet filter having a maxi-
except as sulfate. mum near 420 nm and a light path of 4 to 5 cm. Filter
1.4 Some governing bodies regulate the movement of soils photometers and photometric practices prescribed in this test
from one area to another. It is up to the sampler and laboratory method shall conform to Practice E 60; spectrophotometer
to comply with all regulations. practices shall conform to Practice E 275.
1.5 The values stated in SI units are to be regarded as the 4.2 Stopwatch, readable to 0.1 minutes.
standard. 4.3 Measuring Spoon, capacity 0.2 to 0.3 mL.
1.6 This standard does not purport to address all of the 4.4 Drying oven, capable of continuously heating at 110 6
safety concerns, if any, associated with its use. It is the 5º C.
responsibility of the user of this standard to establish appro- 4.5 Balance, shall be capable of reproducing results within
priate safety and health practices and to determine the 0.0002 g with an accuracy of 60.0002 g. Direct-reading
applicability of regulatory limitations prior to use. balances shall have a sensitivity not exceeding 0.0001 g.
Conventional two-pan balances shall have a maximum sensi-
2. Referenced Documents bility reciprocal of 0.0003 g. Any rapid weighing device that
2.1 ASTM Standards: 2 may be provided, such as a chain, damped motion, or heavy
C 114 Test Methods for Chemical Analysis for Hydraulic riders, shall not increase the basic inaccuracy by more than
Cement 0.0001 g at any reading and with any load within the rated
D 1193 Specification for Reagent Water capacity of the balance.
E 11 Specification for Wire and Cloth Sieves for Testing 4.6 Stirrer, magnetic variable speed, with a TFE-
Purposes fluorocarbon coated magnetic stirring rod or an overhead stirrer
E 60 Practice for Analysis of Metals, Ores, and Related with a propeller.
Materials by Molecular Absorption Spectrometry
E 275 Practice for Describing and Measuring Performance 5. Reagents and Materials
of Ultraviolet, Visible, and Near Infrared Spectrophotom- 5.1 Purity of Reagents—All reagents shall conform to the
eters specifications of the Committee on Analytical Reagents of the
American Chemical Society.3
3. Significance and Use 5.2 Purity of Water—Unless otherwise indicated, reference
3.1 This test method can be used to determine if soils could to water shall be understood to mean reagent water conforming
have an adverse reaction with hydraulic cement concrete. to Specification D 1193, Type I. Other reagent water types (See
Note 1) may be used, provided it is first ascertained that the
1
This test method is under the jurisdiction of ASTM Committee C09 on
Concrete and Aggregates and is the direct responsibility of Subcommittee on C09.69
3
Miscellaneous Tests. Reagent Chemicals, American Chemical Society Specifications, American
Current edition approved March 1, 2005. Published April 2005. Chemical Society, Washington, DC. For suggestions on the testing of reagents not
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or listed by the American Chemical Society, see Analar Standards for Laboratory
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
Standards volume information, refer to the standard’s Document Summary page on and National Formulary, U.S. Pharmaceutical Convention, Inc. (USPC), Rockville,
the ASTM website. MD.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
C 1580 – 05
water is of sufficiently high purity to permit its use without using pH paper and, if needed, neutralize each filtrate to pH of
adversely affecting the precision and bias of the test method. 7 6 1 using either 0.1N HCl or 0.1N NaOH.
8.2 Adjust the temperature of the solution in the volumetric
NOTE 1—Type II water was specified at the time of round robin testing
of this test method. flasks to between 20 and 25 °C [68 to 77 °F].
8.3 Use an aliquot of 10 mL for specimen A1, 20 mL for
5.3 Barium Chloride—Crystals of barium chloride specimen A2, 10 mL for specimen B1, and 20 mL for specimen
(BaCl2·2H2O) screened to 850 to 600 µm. To prepare in the B2. Dilute to a volume of 100 mL in a volumetric flask and
laboratory, spread crystals over a large watch glass, desiccate follow the procedure below or Test Methods C 114. Pour the
for 24 h, screen to remove any crystals that are not 850 to 600 100 mL into a 250 mL beaker. Add 5.0 mL of conditioning
µm, and store in a clean, dry jar. reagent. Mix by stirring with a stir bar and magnetic stirrer.
5.4 Conditioning Reagent—Place 30 mL of concentrated Add 0.3 g of the prepared BaCl2 crystals using a measuring
hydrochloric acid (HCl, sp gr 1.19), 300 mL reagent water, 100 spoon and start the timer.
mL 95 % ethanol or isopropanol and 75 g sodium chloride 8.4 Stir exactly 1.0 min, at the same constant speed for all
(NaCl) in a container. Add 50 mL glycerol and mix. determinations.
5.5 Sulfate Solution, Standard (1 mL = 0.100 mg SO4)— 8.5 Remove the beaker from the stirrer, pour solution into
Dissolve 0.1479 g of anhydrous sodium sulfate (Na2SO4) in the cell, and put the cell into the photometer. Record the
water, and dilute with water to 1 L in a volumetric flask. turbidity at 4.0 min after pouring the solution into the cell.
5.6 pH paper, readable to at least 0.5 pH units. 8.6 Determine the mg of SO4 from the calibration curve.
5.7 Sodium Hydroxide, (0.1N)—Dissolve 4 g of sodium
hydroxide (NaOH) in water and dilute with water to 1L 9. Calculation or Interpretation of Results
5.8 Hydrochloric Acid (0.1N)—Dilute 8.6 mL of concen- 9.1 Calculate the sulfate content (% by mass) for each
trated hydrochloric acid (HCl sp gr 1.19) to 1L with water specimen as follows:
·
6. Permissible Variations 2.5*M
P 5 A*W (1)
6.1 Both the referee method and qualified methods as
described in Test Methods C 114 are permissible as alternatives
to the turbidimetric procedure. Where:
P = % SO4 in dried soil
M = mg/L of SO4 calculated from calibration curve
7. Calibration and Standardization W = mass of soil placed into 400 mL beaker, in g
7.1 Prepare standards by diluting 0.0, 2.0, 5.0, 10.0, 15.0, A = volume of aliquot specimen, in mL
20.0, 30.0, and 40.0 mL of standard sulfate solution with water 9.2 Limits of values for results to be valid. (See Note 3)
to 100-mL volumes in volumetric flasks. These solutions will Specimen Lower Limit Upper Limit
have sulfate ion concentrations of 0.0, 2.0, 5.0, 10.0, 15.0, 20.0, A1 0.04% 0.33%
A2 0.02% 0.17%
30.0, and 40.0 mg/L, respectively. Follow the procedure given B1 0.41% 3.33%
in Section 8 using appropriate amounts of the standard sulfate B2 0.21% 1.67%
solution prepared in accordance with 5.5 and prepare a
1. Determine which results fall within the above limits
calibration curve showing sulfate ion content in mg/L plotted
2. Discard those that do not
against the corresponding photometer readings.
3. Average all results that comply
7.2 A separate calibration curve must be prepared for each
photometer and a new curve must be prepared if it is necessary NOTE 3—These values were derived from the limits on the sulfate test
to change the cell, lamp, or filter, or if any other alterations of of 5 to 40 mg/L, and a solubility of sulfate of 1.6 g/L at pH 7 when calcium
instrument or reagents are made. Check the curve with each sulfate is present. The method is not applicable for soils outside this range.
For those samples, see the Annex.
series of tests by running two or more solutions of known
sulfate concentrations in the range of 5 to 40 mg SO4/L. 10. Report
NOTE 2—The slope of the curve may not be linear below 5 or above 40 10.1 Report the average obtained as % sulfate by mass of
mg SO4/L for this method. sample. If milligrams sulfate in kilograms of soil (mg SO4/kg
soil) is desired, multiply % sulfate by 10 000. Report results to
8. Procedure nearest 0.01 % or 100 mg/kg.
8.1 Collect at least 100 g of representative soil for analysis.
Dry the sample for 18 to 24 h at 110 °C. Crush to pass a 600 11. Precision and Bias
µm sieve. In a 400 mL beaker labeled Sample A, put a sample 11.1 Precision—The standard deviation was found to vary
consisting of approximately 30 g. In a second 400 mL beaker with concentration, so therefore, a coefficient of variation was
labeled Sample B, place a sample of approximately 3 g. Record used for the precision statement. The single operator coefficient
the mass of each sample to the nearest 0.001 g. Add 250 mL of of variation was found to be 4.4 %. Therefore, results of two
deionized water to each beaker, and stir on a magnetic stirrer properly conducted tests by the same operator on the same
for one h. Filter the extraction through two dry medium texture material should not differ by more than 12.4 % of their
filter papers under suction. Do not wash the residue. Repeat the average. The multilaboratory coefficient of variation was found
filtration if solution is not clear. Check the pH of the filtrate to be 21.2 %. Therefore, results of two different laboratories on

2
C 1580 – 05
identical samples of a material should not differ from each 12. Keywords
other by more than 60 % of their average.
12.1 hydraulic cement; soil; Sulfate; turbidimeter
11.2 Bias—Since there is no accepted reference material
suitable for determining bias, bias can not be determined.

ANNEX

(Mandatory Information)

A1. REDUCING THE NUMBER OF SUB-SAMPLES AND SAMPLES OUT OF COMPLIANCE

A1.1 This procedure was written to accommodate labora- sample masses and/or dilutions can be used. If this procedure
tories that are not familiar with the soil under test. If prior is followed, then the mg SO4 from the turbidimeter reading
knowledge is available, a lesser number of sub-samples, and/or shall be maintained within 5 to 40 mg/L. If another procedure
dilutions may be used. is used, then validation shall be provided showing that the
A1.2 If a sample falls above 3.33 % sulfate, then different results are within the linear range of the calibration.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

Das könnte Ihnen auch gefallen