Beruflich Dokumente
Kultur Dokumente
-
3 4 5 East 47th Street, New York, N.Y. 10017
Liquid Flow in
Weighing Method
Closed Conduits by
Measurement of
ASME/ANSI MFC-9M-1988
This Standard will be revised when the Society approves the issuance of a new edition.
There will be no addenda or written interpretations of the requirements of this Standard
issued to thisedition.
This code or standard was developed under procedures accredited as meeting the criteria for
American National Standards. The Consensus Committee that approved the code or standard was
balanced to assure that individuals from competent and concerned interests have an opportunity
had
t o participate. The proposed codeor standard was made available for public review and comment
whichprovides an opportunityforadditionalpublicinputfromindustry, academia, regulatory
agencies, and the public-at-large.
ASME does not '.'approve," "rate," or "endorse" any item, construction, proprietary device, or
activity.
ASME does not take any position with respect t o the validity of any patent rights asserted in
connection with any items mentioned in this document, and does not undertake t o insure anyone
utilizing a standard against liability for infringement of
any applicable Letters Patent, nor assume any
such liability. Users of a codeor standard are expressly advised that determination of the validity of
any such patent rights, and the riskof infringement of such rights, is entirely their
own responsibility.
Participation by federal agency representative(s) or person(s) affiliated with industry is not t o be
interpreted as government or industry endorsement of this code or standard.
ASME accepts responsibility for only those interpretations issued in accordance with governing
ASMEproceduresandpolicies which precludetheissuanceofinterpretationsbyindividual
volunteers.
Copyright 0 1989 by
THE AMERICAN SOCIETY OF MECHANICAL ENGINEERS
All Rights Reserved
Printed in U.S.A.
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FOREWORD
This Standard was prepared by the ASME Committee on Measurement of Fluid Flow in
Closed Conduits (MFC). It is based on and closely parallels the InternationalOrganization
for Standardization(ISO) International StandardIS0 4185-1980, incorporating U.S. prac-
tices and terminology where they differ.
This Standard was approved by the American National Standards Institute(ANSI) as an
American National Standard on December 15, 1988.
iii
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME STANDARDS COMMITTEE MFC
Measurement of Fluid Flow in Closed Conduits
(The following is theroster of the Committee atthe time of approval of this Standard.)
OFFICERS
R. W. Miller, Chairman
W. F. 2 . Lee, Vice Chairman
C. J. Gomez, Secretary
COMMITTEE PERSONNEL
R. B. Abernethy A. G. Ferron W. F. 2 . Lee
N. A. Alston R. L. Galley E. D. Mannherz
H. P. Bean D. Halmi G. E. Mattingly
S.R. Beitler Z. D. Husain R. W. Miller
M. Bradner B. T. Jeffries M. H. November
E. E.Buxton E. H. Jones, Jr. W. M. Reese, Jr.
J. S.Castorina L. J. Kemp P. G. Scott
G. P. Corpron C. A. Kemper H. E. Snider
D. G. Darby D. R. Keyser D. W. Spitzer
R. H. Dieck C. P. Kittredge D. A. Sullivan
R. B. Dowdell C . G. Langford R. G. Teyssandier
G. E. Mattingly, Chairman
G. P. Corpron
R. B. Dowdell
D. Halmi
G. A. Lenz
R. W. Miller
P. G. Scott
R. G. Teyssandier
V
t
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled whe
CONTENTS
Foreword ...............................................................
...
111
Standards Committee Roster ............................................... v
1 General ............................................................ 1
1.1 Scope and Fieldof Application ................................... 1
1.2 References ..................................................... 1
1.3 Definitions .................................................... 1
1.4 Symbols ....................................................... 2
2 Principles .......................................................... 2
2.1 Statement of the Principles ...................................... 2
2.2 Accuracy of the Method ......................................... 2
3 Apparatus .......................................................... 2
3.1 Diverter ....................................................... 2
3.2 Time-Measuring Apparatus ...................................... 9
3.3
Weighing Tank ................................................. 9
3.4
Weighing
Device ............................................... 10
3.5 Auxiliary Measurements ......................................... 10
4 Procedure .......................................................... 10
4.1 Static Weighing Method ......................................... 10
4.2 Dynamic Weighing Method ...................................... 10
4.3 Common Provisions ............................................ 11
5 Calculation of Flow Rate ............................................. 11
5.1 Calculation ofMassFlow Rate ................................... 11
5.2 Calculation ofVolumeFlow Rate ................................. 11
6 Uncertainties in the Measurement of Flow Rate ........................ 11
Figures
1A Diagram of an Installation for Calibration by Weighing (Static
Method. Supply by a Constant Level Head Tank) ...................... 4
1B Diagram of an Installation for Flow Rate Measure by Weighing
(Used for a Hydraulic Machine Test; Static Method. Supply
by aConstant Level Head Tank) .................................... 5
1C Diagram of an Installation for Calibration by Weighing (Static
Method. Direct Pumping Supply) .................................... 6
1D Diagram of an Installation for Calibrationby Weighmg (Dynamic Method.
Supply by aConstant Level HeadTank) .............................. 7
2 Examples of Diverter Design .......................................... 8
3 Operational Lawof Diverter .......................................... 8
4 Time Metering for a Diverter. the Operation Law of Which Is Identical
in Both Directions ................................................. 9
vii
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
Table
Symbols
1 ............................................................ 3
Appendices
ACorrections on the Measurement of Filling Time ......................... 13
B Density of Pure
Water ............................................... 17
Figure
A1 Plotting of Results of DiverterTimer Actuator .......................... 14
...
Vlll
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
ASME/ANSI MFC-9M-1988
1
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
MEASUREMENT OF LIQUID FLOW IN
ASME/ANSI MFC-9M-1988 CLOSED CONQUITS BY WEIGHING METHOD
conduit being supplied with liquid; for example, a stantaneous values are properly time-averaged during
constant level head tank, the level of liquid which is the whole filling period.
controlled by a weir of sufficient capacity
static weighing - method in which the net mass of 2 . 2 Accuracy of the Method
liquid collected is deduced from tare andgross weigh-
ings made before and after the liquid has been diverted 2 . 2 . 1 OverallUncertaintyontheWeighing
for a measured time interval into the weighing tank Measurement. The weighing method gives an abso-
lute measurement of flow which, in principle, requires
only mass and time measurements. Provided that the
1 . 4 Symbols precautions listed in para. 2.2.2 are taken,this method
Table 1reproduces the symbols that areused inthis may be considered as one of the most accurate of all
Standard. flow rate measuring methods, and for this reason it is
often used as a calibration method.When the installa-
tion is carefully constructed, maintained, andused, an
uncertainty of +0.1% (with 95% confidence limits
2 PRINCIPLES for the random part of that uncertainty) can be
2.1 Statement of the Principles achieved.
ASME/ANSI MFC-9M-1988
MEASUREMENT OF LIQUID FLOW IN CLOSED CONDUITS
BY WEIGHING METHODS
Page3
(1) In Table 1 for Symbol Qs, changeQuantity J?+mz
Density of air (at 20°Cand 1 bar*) to read:
Density of air
(2) Deleteasterisked
footnote.
TABLE 1 SYMBOLS
Dimension SI (Metric) U.S. (Customary)
Units Units
ntity Symbol (111
... ... ...
B Bias
m Mass M kg Ibm
lbmlsec
Qrn Mass flow rate MT- kg/s
t Time T S sec
... ...
t95 Two-tailed
Student's t ...
, ,<.. .
ft3 m3
V Volume .L3 -
. .
uRSS
u95 Uncertainty
95%
the
at confidence ...
. ,
- level **,., .
:. .
! ,I
\
. ,
., *:.
UADD , I ,
, , .
, -
. .-
-~ -. ' .
ML-3
Ibm/ft3 kg/m3
ea Density
of air 20"C;and
(at 1 bar")i . . ......
"3 ' .
.. ,ML t 3 Ibm/ft3 kg/m3
Density of standardweights. 1'
ep .. _.
.
:
. . . .
...
__
c
r
__
......... -- I .
.. ,_..-.. F r .
* I bar = lo6 Pa
3
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled whe
CLOSED CONDUITS BY WEIGHING METHOD
MEASUREMENT OF LIQUID FLOW IN
4
ASME/ANSI MFC-9M-1988
FIG. 1A
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
MEASUREMENT OF LIQUID FLOW IN
CLOSED CONDUITS BY WEIGHING METHOD ASME/ANSI MFC-9M-1988
-- -
Sump
-- - -
5
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
MEASUREMENT OF LIQUID FLOW IN
CLOSED CONDUITS BY WEIGHING METHOD
--
valve
6
Pump
ASME/ANSI MFC-9M-1988
FIG. 1C
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
MEASUREMENT OF LIQUID FLOW IN
CLOSED CONDUITS BY WEIGHING METHOD ASMElANSl MFC-9M-1988
Flow control
valve
II II Device under
calibration
Weighing
Timer
Overflow - - -
H I Switches
--- - - -- - -
Pump _-- Sump
---
7
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
MEASUREMENT OF LIQUID FLOW IN
ASME/ANSI MFC-9M-1988 CLOSED CONDUITS BY WEIGHING METHOD
Flow input
\ Flow output
-
slot should not exceed about 20 kPa (3 psi) to avoid
splashing, air entrainment,' and flow across the di- I
Time
verter and turbulence in the weighing tank. This mo-
tion of the-diverter can be generated by various ne of diverter motion
*---
electrical or mechanical devices - for example, by a
spring or torsion bar or by electrical or pneumatic ac-
FIG. 3 OPERATIONAL LAW OF DIVERTER
tuators. The diverter should in no way influence the
flow in the circuit during any phase of the measuring
procedure.
For large flow rates which could involve excessive
stresses, however, a diverter with a proportionately
slow performance rate (1 s to 2 s, for example) can be Care shall be taken when designing the mechanical
used, provided the operatinglaw isconstant (see para. parts of the device and the diverter, as well as during
3.2 and Fig. 3) and the variation of the flow rate distri- frequent checks in service, that no leak or splash of
bution as a functionof the diverter stroke is preferably liquid occurs either toward the outside or from one
linear and is in any case known and can be verified. diverter channel to the other.
Besides a thin flat liquid stream, other shapes of
liquid stream are permissible in the diverter duct, if the
21n certain designs of nozzle slot, however,special vents to allow air
ingress to the fluid jetmay be necessary to ensure stable flow within necessary corrections for the diverting time are
the test circuit. applied as indicated in Appendix A.
8
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
MEASUREMENT OF LIQUID FLOW IN
CLOSED CONDUITS BY WEIGHING METHOD ASME/ANSI MFC-9M-1988
3.2 Time-Measuring Apparatus tion should be made in accordance with the directions
of Appendix A.
The time of discharge into the weighing tank is
normally measured by an electronic counter with a
built-in accurate time reference - for example, a 3.3 WeighingTank
quartz crystal. The diversion period can thus be read The weighing tank shall be of sufficient capacity so
to 0.01 s or better. The error arising from this source that the error in timing is negligible. Taking account
can be regarded as negligible, provided the resolution of what is stated in paras. 3.1 and 3.2, the filling time
of the timer display is sufficiently high and the equip- for thehighest expectedflow rate shall be at least 30 s.
ment is checked periodically against a national time Nevertheless, this time may be reduced provided that
standard - for example, the frequency signals trans- it is possible to determine experimentally, according
mitted by certain radio stations. to procedures such as described in Appendix A, that
The timer shall be actuated by the motion of the the required accuracy is achieved.
diverter itself through aswitch fitted on the diverter - The tankmay be ofany shape butit is essential that
for example, optical or magnetic. Strictly speaking, it is perfectly leak-tight, and care should be taken to
the time measurement shall be started (or stopped) at avoid liquid spillage. Internal walls or baffles may be
the instant when the hatched areas in Fig. 3, which required to reduce oscillations of the liquid in the tank
represent flow variation with time, are equal. In prac- and to improve structural rigidity.
tice, however, it is generally accepted that this point The tank may be suspended from the weighing de-
corresponds to the mid-travel position of the diverter vice or may constitute the platform of the latter, or
in the fluid jet. The errorwill be negligible, provided may be placed on one of the platforms. To prevent
the time of passage of the diverter through the stream sudden overloads detrimental to the weighing appa-
is negligible in comparison with the period of diver- ratus, it may be necessaryto lock the tank in position
sion to the tank. on the scale during filling.
If a diverter is used, then the operating law of the The tank may be drained by different means:
diverter should be identical in both directions (see Fig. (a) by a valve at the base, the leak-tightness of
4), and thetimer may bestarted andstopped at the in- which shall be capable ofbeing verified (free dis-
stant when the motion of the diverter is started in each charge, transparent hose, or leak detection circuit); or
direction; this is the case particularly when the time- (b) by a siphon fitted with an efficient and check-
flow rate law is linear. able siphon break; or
If the error in the filling time measurement arising (c) by a self-priming or submersible pump.
from the operation of the diverter and starting and The rateof draining shall be sufficiently high so that
stopping of the timer is not negligible, then a correc- test runs can follow each other at short intervals.
9
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled whe
MEASUREMENT OF LIQUID,FLOW !N
ASME/ANSI MFC-SM-1988 CLOSED CONDUITS BY WEIGHING METHOD
10
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
MEASUREMENT 0.F LIQUID:F.LOW:IN
CLOSED,CONDUlTSBY WEIGHING METHOD
In all casesit shall be carefully checked that nopipe about ambient temperature,an accuracy of 0.5"C
connections or electric wire links existiyhich are likely ( 1°F)is enough t o ensure less than lop4error on den-
to transmit stresses between the weighing tank and the sity evaluation.. ' '' . :
fixed parts of the installation;indispensable links shall '.. If, however, the purity of the liquid is in doubt, it is
''. "
therefore be extremely flexible and their .flexibility essential to -measureits density. To this end, a sample
verified during the calibrationof the wetgqing
. '. device. can be collected and its density measured either by a
.. . .,
*?'. .* ,'\, <., '! ,
direct method,,by,weighipg graduated in a cylinder on
,i ~..'\ \ ,. '..f
., ', an analytical ,balance, or$~yan indirect method - for
_,'
.?\ x, x.,
q =-=
m, - ma 1 - -
m
X-
ep
ea atures. (In exceptional cases, it may be necessary to
measure the density directly.) .
m t t ea
1 --
e
e = 1 OOO kg/m3
11
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
MEASUREMENT OF LIQUID FLOW IN
ASME/ANSI MFC-9M-1988 CLOSED CONDUITS BY WEIGHING METHOD
12
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
APPENDIX A
CORRECTIONS ON THE MEASUREMENT OF FILLING TIME
(This Appendix contains supplementary information for theconvenience of the reader. It is not partOf
ASME/ANSI MFC-9M-1988.)
The results obtainedshould be fitted into the If a significant value of At is obtained, the diverter
following equation, in which At is the required timing timer actuator should be adjusted to minimize the
error of the diverter system: value of the error as shown by repeated testing.
The procedure should be repeated at a few lower
flow rates to examine whether or not the value of At
(qi - qn) - (4ir - GnJ obtained is significantly flow rate dependent. If signif-
icant changes in the At value are obtained, it will be
4n
necessary to improve the operationof the diverter sys-
tem or tointroduce avariable correction time At to be
where applied to the diversion time.
tqi = diversion time for a particular short test
tqn = diversion time for the normal length test oc-
A2 DYNAMIC WEIGHING METHOD
curring nearest in diurnal time in the testing
sequence This procedure involves movement of the beam of
q; = flow rate calculated for the particular diver- the weighing device just prior to both start and stop
sion time t,; actuations of the timer.
q, = flow rate calculated for the normal diversion Four important dynamic phenomena take place
time t,, occurring nearest in diurnal time in during the dynamic weighing cycle, namely:
the testing sequence (a) a change in the impact force of the falling liquid
-
qit = average flow rate meter reading time tqi between the initial and final weighing points;
-
qnr= average flow rate meter reading during time (b) collection of an extra amountof liquid from the
falling column by the rising level in the tank;
The values obtained for the right-hand side of this (c) forces due to waves in the tank;
equation should be plotted against ( l / t q j - l/tq,) as (d)a change in the inertia of the weighing device
shown in Fig. A l . The points should define a straight and liquid in the weighing tank, with a resultant
line passing through the origin, the slope of which is change of time required to accelerate the balance
equal to At. beam to the timer actuation point.
14
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
Generally, the decrease in impact force is equal and where
opposite to the additionalweight of liquidcollected so La = distancetravelledby the end of a balance
that these two effects cancel each other. beam of length L deflected through an angle
Oscillation of liquid within the weighing tank may (Y from rest to the timing point
have a seriousinfluence on theprecisionofthe M I ordinarily will include the masses of the weigh-
method. Devices prescribed in para. 3.3 can reduce, ing tank and initial liquid therein, and possibly other
but not eliminate completely, this undesirable masses depending on the weighing device used.
phenomenon, which is always most pronounced at The correctedcollection time in thiscase is ( t - At).
higher rates of flow. This errorAt can be reduced in conventional weigh-
Changes in inertia between the initial and final ing applications by limiting the deflection(Y.Alterna-
weighing points can affect indicated flow rate by up tively, static weighing experiments can be compared
to 0.5% if the error At in measured time t is not with those using the dynamic technique to determine
accounted for. This error is approximately:’ At; the results can then be used to test the above equa-
tion for applicability and to evaluate the constants
therein. On smaller dynamic weighing systems, the in-
ertia effect can be practically eliminated by using a
substitute weighing technique.
15
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
APPENDIX B
DENSITY OF PURE WATER
(This Appendix contains supplementary information for the convenience of the reader. It is not Part Of
ASME/ANSI MFC-9M-1988.)
Density
"C kg1m3 "F lbmlft
17
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
STANDARDS FOR MEASUREMENT OF FLUID IN CLOSED CONDUITS
TITLE OF STANDARD