Beruflich Dokumente
Kultur Dokumente
MATERIALS JOINING
TECHNOLOGY
PRACTICAL RADIOGRAPHY
(NDT20)
CONTENTS
Section Subject
8.1.2.1
GEOMETRIC UNSHARPNESS
8.1.2.2
INHERENT UNSHARPNESS
8.1.2.2.1
FILM (EFFECT ON INHERENT
UNSHARPNESS)
8.1.2.2.2 QUALITY OF RADIATION (EFFECT ON
INHERENT UNSHARPNESS)
8.1.2.2.3 INTENSIFYING SCREENS (EFFECT ON
INHERENT UNSHARPNESS)
8.1.2.3 RELATIVE MOVEMENT DURING
EXPOSURE
8.2 RADIATION SCATTERING AND SCATTER CONTROL
8.2.1 SCATTERING MECHANISMS – THE CAUSES OF
SCATTER
8.2.1.1. THE PHOTOELECTRIC EFFECT
8.2.1.2. COMPTON SCATTERING
(INCOHERENT SCATTERING)
8.2.1.3. PAIR PRODUCTION
8.2.1.4. TOTAL SCATTER AT DIFFERENT
PRIMARY BEAM ENERGIES
8.2.2 TYPES OF SCATTER
8.2.2.1 SIDE SCATTER
8.2.2.2 BACK SCATTER
8.2.2.3 SELF-SCATTER
8.2.3 SCATTER CONTROL
8.2.3.1 COLLIMATION
8.2.3.2 DIAPHRAGMS
8.2.3.3 MASKING OR BLOCKING
8.2.3.4 GRIDS
8.2.3.5 FILTERS
Practical Radiography. Rev 2 Oct 2005 WORLD CENTRE FOR
Copyright © 2005, TWI Ltd MATERIALS JOINING
TECHNOLOGY
Section Subject
8.2.3.6
METALLIC FOIL SCREENS
8.2.3.7
HIGHER RADIATION ENERGY
8.2.3.8
CHANGE FROM X-RAY TO GAMMA
RAY RADIOGRAPHY
8.2.3.9 REDUCING THE FOCUS OR SOURCE
TO FILM DISTANCE
8.3 DETERMINING THE CORRECT EXPOSURE: EXPOSURE
CHARTS
8.3.1 EXPOSURE CHARTS
8.3.1.1 USING EXPOSURE CHARTS (X-RAY)
8.3.1.1.1 FOCUS TO FILM DISTANCE
8.3.1.1.2 TUBE VOLTAGE
8.3.1.1.3 CHANGING THE FILM DENSITY
8.3.1.1.4 CHANGING THE FILM TYPE
8.3.1.1.5 RADIOGRAPHY OF OTHER
MATERIALS
8.3.1.1.6 COMPENSATING FOR THE USE OF A
FILTER
8.3.1.1.7 OTHER POSSIBLE CHANGES
8.3.1.2 GAMMA RAY EXPOSURES
9.0 SENSITIVITY
9.1 RADIOGRAPHIC SENSITIVITY
9.2 CONTROLLING RADIOGRAPHIC QUALITY
9.3 BS EN 462-1 WIRE TYPE IQIs
9.4 OTHER WIRE TYPE IQIs
9.5 BS EN 462-2 STEP–HOLE TYPE IQIs
9.6 ASTM E 1025 PLAQUE TYPE PENETRAMETERS
9.7 IQI SENSITIVITY
11.6.14 RETICULATION
11.6.15 FILM FOGGING BY X OR GAMMA RAYS
11.6.16 LIGHT FOGGING
11.6.17 FILM FOGGING DUE TO INADEQUATE STORAGE
CONDITIONS
11.6.18 SOLARISATION
11.6.19 A FINAL WORD ON ARTEFACTS
11.7 INTERPRETATION OF WELD RADIOGRAPHS
11.7.1 RADIOGRAPHIC INDICATIONS DUE TO SURFACE
GEOMETRY
11.7.1.1 EXCESSIVE ROOT PENETRATION
11.7.1.2 ROOT CONCAVITY
11.7.1.3 INCOMPLETELY FILLED GROOVE
(LACK OF FILL)
11.7.1.4 LACK OF REINFORCEMENT
11.7.1.5 UNDERCUT
11.7.1.6 SPATTER
11.7.1.7 EXCESSIVE DRESSING / GRINDING
MARKS
11.7.1.8 HAMMER MARKS (TOOL MARKS)
11.7.1.9 TORN SURFACE
11.7.1.10 SURFACE PITTING
11.7.2 INTERNAL DEFECTS
11.7.2.1 CRACKS
11.7.2.2 LACK OF FUSION
11.7.2.3 INCOMPLETE ROOT PENETRATION
11.7.2.4 NON-METALLIC INCLUSIONS
11.7.2.5 METALLIC INCLUSIONS
11.7.2.6 GAS PORES: POROSITY
11.7.2.7 ELONGATED CAVITIES (HOLLOW
BEAD)
11.7.2.8 WORM HOLES
11.7.2.9 CRATER PIPES & CRATER CRACKS
11.8 INTERPRETATION OF CASTING RADIOGRAPHS
11.8.1 VOIDS
11.8.1.1 MACROSHRINKAGE
11.8.1.2 FILAMENTARY SHRINKAGE (ALSO
CALLED SPONGINESS)
11.8.1.3 MICROPOROSITY / MICROSHRINKAGE
Practical Radiography. Rev 2 Oct 2005 WORLD CENTRE FOR
Copyright © 2005, TWI Ltd MATERIALS JOINING
TECHNOLOGY
11.8.1.4 PINHOLE POROSITY
11.8.1.5 GASHOLES
11.8.1.6 AIRLOCKS (ENTRAPPED AIR)
11.8.2 CRACKS
11.8.2.1 HOT TEARS
11.8.2.2 STRESS CRACKS
11.8.3 COLD SHUTS
11.8.4 INCLUSIONS
11.8.5 SEGREGATIONS
12.0 LOCALISATION
12.1 90° METHOD
12.2 TUBE (SOURCE) SHIFT METHOD
12.3 TUBE (SOURCE) SHIFT METHOD WITH LEAD MARKERS
17.0 APPENDIX 1
The physical characteristics of the absorber (in particular its density and atomic
number).
The radiograph that would result from the set-up in figure above
(Note that in film radiography thin sections appear darker while thicker sections
appear lighter. The opposite is true if a fluorescent screen rather than a photographic
film is used as a radiation detector)
(Note that the radiograph cannot be used to determine the through thickness position
of the voids.)
Two important things to keep in mind when viewing a radiographic image are:
and
(2) The lack of side fusion will not appear as an image on the radiograph because
A - (D + E) = 0.01 mm which is much too small to be detected by the
technique used.
Radiography of a weld
In the equilibrium state the number of orbital electrons is equal to the number of
protons and there is no net electrical charge. When there is inequality between the
numbers of protons and electrons then there is a net electrical charge and the atom is
said to be ionised. Ions may be negatively charged if the number of electrons
exceeds the number of protons or positively charged if the converse is true. So called
electropositive elements, a group which includes all metals, ‘like’ to form positive ions
while the electronegative elements such as oxygen, phosphorous, chlorine and
sulphur ‘like’ to form negative ions.
The orbital electrons exist in fixed energy levels or shells. Each shell can
contain a fixed maximum number of electrons. The shells are identified by the letters
K, L, M, N and so on. The lowest energy level is represented by the K-shell; this is the
innermost of the electron shells and it can contain a maximum of 2 electrons. The L-
shell can contain up to 8 electrons while the M-shell contains a maximum of 18 and
the N-shell contains a maximum of 32. The maximum total number of electrons in
each shell is equal to 2n2 where ‘n’ is the shell number counting the K-shell as 1, L-
shell as 2 etc. Within the M, N and other shells certain groupings of electrons produce
greater stability, elements having an even number of electrons tend to less chemically
1 2
H He
1.008 4.003
3 4 5 6 7 8 9 10
Li Be B C N O F Ne
6.940 9.012 10.81 12.01 14.01 16.00 19.00 20.17
11 12 13 14 15 16 17 18
Na Mg Al Si P S Cl Ar
22.99 24.30 26.98 28.09 30.97 32.06 35.45 39.95
19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36
K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
39.10 40.08 44.96 47.90 50.94 52.00 54.94 55.85 58.93 58.71 63.55 65.38 69.74 72.59 74.92 78.96 79.90 83.80
37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54
Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe
85.47 87.62 88.91 91.22 92.91 95.94 98.91 101.1 102.9 106.4 107.9 112.4 114.8 118.7 121.8 127.6 126.9 131.3
55 56 57 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86
Cs Ba La Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
132.9 137.3 138.9 178.5 181.0 183.9 186.2 190.2 192.2 195.1 197.0 200.6 204.4 207.2 209.0 (209) (210) (222)
87 88 89
Fr Ra Ac
(223) 226.0 (227)
58 59 60 61 62 63 64 65 66 67 68 69 70 71
Lanthanide Series Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
140.1 140.9 144.2 (145) 150.4 152.0 157.3 158.9 162.5 164.9 167.3 168.9 173.0 175.0
(The numbers above and below each chemical symbol are the atomic number and the atomic weight of
each element. Note that the atomic weight differs slightly from the atomic mass number)
All types of electromagnetic radiation travel at the same velocity (v), the velocity
of light, which is about 2.998 x 108 ms-1 (186,000 miles per second in old money), but
differ in terms of their wavelength ( ) and frequency (f). Wavelength can be defined
as the distance travelled during one complete field oscillation while frequency is the
total number of oscillations occurring in one second.
v = f
As scientific knowledge advanced it became clear that in some circumstances
light behaved not so much like a waveform, but more like a particle. Considering such
behaviour in 1900 a scientist called Max Planck first put forward the theory that light
had, what he called, a ‘quantum’ nature. Planck postulated that electromagnetic
energy could not exist in amounts (‘quantum’ being Latin for ‘amount’) smaller than a
given very small amount of energy and that all larger amounts of electromagnetic
energy were exact multiples of this amount to which he gave the name “photon”.
Planck believed that the photon energy of any form of electromagnetic radiation would
be equal to a constant multiplied by its frequency. In later years Planck’s hypothesis
was proved to be true and the constant in question became known as ‘Planck’s
Constant’, usually abbreviated as ‘h’.
E = hf
–23
Where ‘h’ is Planck’s constant (= 6.63 x 10 Js) and ‘E’ is the photon energy
of electromagnetic radiation of frequency ‘f’.
1. The electron volt (eV) is a unit of energy, which is equal to the kinetic energy that
an electron obtains when it accelerates through an electric field of 1 volt. A Mega
electron volt (MeV) is equal to the kinetic energy of an electron that has
accelerated through an electric field of 1 million volts. On electron volt is equal to
1.6 x 10-19 Joules.
2. The relationship between wavelength and photon energy on which the diagram
above has been based is approximate.
and
(ii) the relative efficiency of the process of stopping the incident electron –
does this occur in a single large event or in a series of events of varying
magnitude?
X-ray production
The value of the minimum wavelength, which will be produced by an x-ray tube having
an accelerating voltage V, can be estimated using the following formula:
The maximum intensity (the number of photons per square metre) in the continuous
spectrum produced occurs at approximately 2 x λmin .
The ability of the x-rays to penetrate matter depends on their photon energy.
The shorter the wavelength, the higher the photon energy, the more penetrating the
radiation. The penetrating power of the x-rays can be controlled by increasing or
decreasing the accelerating voltage. The greater the accelerating voltage, the more
penetrating the radiation. In an x-ray set the accelerating voltage is the tube voltage.
The two characteristic peaks shown in above X-ray spectrum figure are caused
by target material inner shell electrons jumping to a higher energy level, then falling
1. A source of electrons.
High velocity electrons cannot travel far in air; therefore the process of
acceleration must take place in a high vacuum.
The anode consists of a heavy section of high conductivity copper with a small
tungsten (or other high melting point high atomic number metal) insert which is called
the target. The anode has a positive electrical potential with respect to the cathode.
The body of the anode is always copper because copper is an efficient conductor of
heat. This property is necessary because approximately 95% of the kinetic energy of
the impacting electrons is converted to heat at the anode.
The target material is usually tungsten because this has a very high melting
point (3370°C). This reduces the chances that it will be vaporised by the large amount
of heat generated. Tungsten has a high atomic number and therefore a large number
of electrons. This makes it a relatively efficient material for converting kinetic energy to
x-ray energy which in turn helps to reduce the amount heat produced as a proportion
of the total output of energy. Sometimes the target is constructed from Tantalum
(melting point 2996°C) and less frequently from other refractory metals.
Nearly all anodes are ‘hooded’ (see 2 figures below) the hood is a high
conductivity copper shroud which is designed to intercept stray electrons and to
prevent them from hitting the tube walls. The hood has a ‘window’ in the form of a
beryllium insert or a thinned section of copper, which permits x-rays to exit without
unduly increasing ‘inherent filtration’. Inherent filtration is the term used to describe
removal of x-rays from the primary beam due to absorption by the materials used in x-
ray head construction. The reason that a beryllium window is used in many x-ray
heads is that beryllium has a very low absorption factor and this minimises inherent
filtration whilst still affording the tube walls protection from stray electrons.
Directional anode
The target is generally set at an angle of about 70° to the electron beam as
shown in both figures above. This produces a small effective focus size whilst
maintaining a large actual focal spot size. The large actual focus size helps to
dissipate the heat generated more efficiently. Therefore higher tube currents can be
sustained without the risk of damaging the target. This design feature is known as
“Benson” or “Line” focusing. See figure below:
The cathode and anode are mounted in an evacuated glass (or in modern
tubes metal-ceramic) envelope as shown in figure below. The tube may be provided
with shielding to absorb any unwanted radiation that is not already shielded out by the
natural geometry of the anode. Directional type tubes produce a useful beam of
radiation that is usually in the form of a cone with a dihedral angle of around 40°. X-
ray tubes fitted with a panoramic anode produce a useful beam of radiation through an
angle of 360° about the tube axis.
The x-ray beam produced is filtered by the wall of the glass (or metal-ceramic)
envelope. This reduces the useful quantity of x-rays produced, with the low energy
components of the spectrum being particularly effected. Therefore it is common in
glass tubes that the tube wall will be ground thinner in the region of the useful beam in
order to minimise the x-ray energy lost due to self-filtration. Metal-ceramic x-ray tubes
(and low kilovoltage glass tubes) may have Beryllium inserts (usually called windows)
in order to minimise the filtration effect of the tube wall. Beryllium is used for this
purpose because it has a very low x-ray absorption coefficient and because it is
mechanically strong enough to contain the necessary vacuum.
X-ray tubes are invariably mounted inside some form of ‘tank’. This is usually a
metal cylinder that may be fitted with a beryllium or plastic window in order to minimise
self-filtration of the x-rays produced. The tank contains a coolant, which may be oil or
some type of gas. It provides high voltage insulation and mechanical protection. In
portable equipment the high voltage transformer is mounted inside the tank.
Most older type portable x-ray sets were half-wave self-rectified. This produced
a considerable weight saving compared with the earlier types of constant potential
unit. Most modern portable units are constant potential and use lightweight solid state
rectifiers to produce what is effectively DC current.
The metal ceramic tubes used in modern equipment are able to safely
withstand a greater potential difference between the anode / cathode and the tube
wall. This permits the use of “grounded anode” type circuitry, which in turn permits
direct water cooling of the anode. In an older type unit operating at say 200 kV the
cathode voltage would have been minus100 kV while the anode voltage would have
been plus 100 kV, giving a maximum potential difference between the electrodes and
the glass tube wall of 100 kV. With modern grounded anode circuitry it is safe to hold
the cathode at minus 200 kV with the anode at zero volts to produce the same 200 kV
potential difference. An anode held at zero volts can be safely cooled by water.
Water is a very efficient coolant and direct water cooling of the anode permits
operation at greatly increased tube currents. For example the maximum tube current
for an older type 200 kV oil cooled head was typically 5 mA self-rectified. With modern
portable equipment maximum constant potential tube currents of 15 or 20 mA are not
unusual for a 200 kV head.
The radiation produced by the x-ray tube can be varied in quantity and
penetrating power (or quality) by controlling the electrical supplies to the tube.
(1) Voltage control: This alters the tube voltage (kV) by varying the low voltage
supply to the high voltage transformer. Note that high voltage is not generated
in the control panel. This minimises the hazard to personnel.
(2) Milliampere control: An ammeter incorporated into the control panel measures
(albeit indirectly) the current flowing across the tube. This is proportional to the
number of electrons flowing from the cathode to the anode per unit time. In
order to increase the supply of electrons the heating current to the filament is
increased using the milliampere control. Note that the ammeter measures
the current flowing between the anode and the cathode, not the current
flowing in the filament (i.e. the heating current).
(3) Timer: Since the quantity of x-rays produced is proportional to the length of
time during which the tube is energised it is convenient to incorporate a time-
switch into the control panel of the x-ray set which automatically terminates the
exposure at a preset time.
This is because the amount of radiation produced is the same in each case.
Obviously it would be desirable to use a high value of mA, in order to reduce the
Practical Radiography. Rev 2 Oct 2005 5.11 WORLD CENTRE FOR
X-Rays or Bremstrahlung MATERIALS JOINING
5.1.6.1 BETATRONS
Van der Graaf generators generate a high voltage charge of static electricity by
mechanical means - friction. The static charge produced can be used to accelerate
electrons for x-ray production. Van der Graaf generators can produce short intense
pulses of x-ray energy. They have therefore found some application in the field of
‘ballistic radiography.’
Standard x-ray equipment has an effective focus size usually in the range from
1 to 4 mm. This is small enough to provide adequate image quality for most standard
techniques. Microfocus x-ray equipment may have an effective focus size as small as
0.1 mm. Using such a small focus size geometric enlargement techniques are
possible whilst still producing an adequately sharp image.
In a rod anode tube the target is at the end of a copper or aluminium tube which
is usually less than 20 mm outside diameter and may be up to a metre long. The
target is invariably of the panoramic variety. Grounded anode circuitry is essential for
this type of tube. The anode can be positioned inside small diameter pipes in order to
carry out panoramic radiography of girth welds; it can also be positioned in many other
otherwise inaccessible locations. Rod anode tubes are most often used in aerospace
applications.
The disintegration process is fixed for each radioactive isotope and as a result the
gamma ray energies produced are also fixed.
Thus in beta emission there is no loss from the atomic mass number whilst the
atomic number increases by 1 (see the example above). Beta radiation is more
penetrating than alpha. It can penetrate the outer layers of the skin and lead to fatal
skin burns. The damage caused is very similar to sunburn, but much more severe.
Many of the early victims of the Chernobyl disaster died as a result of skin burns
caused by exposure to high intensities of beta radiation. If beta emitters are ingested
they will often lead to cancer.
All gamma sources in use today are man-made. They are manufactured by
neutron bombardment of non-radioactive raw materials in the core of a small nuclear
reactor. The sources in use are all beta emitters, gamma rays being produced as a
by-product of beta emission. In order to prevent beta emission or contamination
hazard the sources used in industrial radiography are invariably sealed sources. The
fissile material is encapsulated in a high integrity titanium or stainless steel shell. Beta
radiation is not capable of penetrating the walls of the capsule, and the capsule further
precludes any possible contamination hazard so long as it remains intact.
This figure shows the typical encapsulation arrangement for Iridium 192 and
Cobalt 60. Some isotopes such as Caesium 137 are double encapsulated. In the
case of Caesium 137 this is because the caesium is in the form of caesium chloride
which is highly corrosive and highly water-soluble (but this is still an improvement on
caesium metal which causes an explosion on contact with water).
The becquerel, which is the SI unit of radioactivity, is a very small unit in terms
of what is required for industrial radiography. The curie is therefore generally
preferred. If the becquerel is used at all then it is usually in the form of gigabecquerels
(GBq). One gigabecquerel is equal to one thousand million (109) becquerels. One
curie is equal to 37 gigabecquerels (37 GBq). In the great majority of cases gamma
ray exposures are expressed in curie-hours, curie-minutes or curie-seconds; this in
Example:
All other factors being equal the exposure time would therefore be:
or
or
If the half-life of an isotope is known then the source activity at a given time can
be calculated if at some point previously the source activity was measured.
St = S0 2-(t/h)
Alternatively the activity of a source can be estimated using a decay chart. The
figure below shows the decay chart for Iridium 192.
(1) To contain the radioactive isotope and reduce the emitted intensity of radiation
to a level which allows for safe transportation and storage.
(2) To allow the radioactive isotope to be safely exposed in order that it may be
used for radiography.
All modern isotope containers are designed to be operated by cable (see figure
26). They are of two basic types (see following 2 figures). Of the two types depicted
the “S” tube type is intrinsically safer but around 30% heavier than the equivalent
shutter type. Older types of isotope container did not provide for remote operation.
The output of radiation per unit time is variable for x-ray equipment up to the
maximum mA rating of the tube. The output of radiation from a radioactive isotope is
fixed by the source activity. The output of radiation produced by x-ray equipment is
generally much greater than that produced by radioactive isotopes.
Maximum
practically
Practically
achievable
achievable
Density activity for 3 mm
Isotope maximum specific
(g/cm3) diameter, 3 mm
activity
long cylindrical
(Curies per gram)
pellet
(Curies)
Cobalt 60 50 8.9 10
Iridium 192 350 22.4 166
Caesium 137 25 3.5 (note 1) 2
Thulium 170 1,000 4 (note 2) 85
The output of radiation from a typical x-ray machine is much greater than the
output of radiation from a typical gamma source. This means that in x-radiography the
use of long focal to film distances is more economically feasible than is the case in
gamma radiography. Thus, even though the focus is similar in physical size when
compared with the average gamma source, it is generally the case that geometric
unsharpness is better for x-ray techniques than for gamma.
The exposure time for gamma radiography tends to be longer. This is because
the output of radiation (in photons per second) is generally much less. Gamma ray
exposure times are usually in the range from about 30 seconds to 1 hour, but
exposure times exceeding 24 hours are not unheard of. The required exposure time
for a gamma ray source increases as the source activity reduces with time.
X-ray sets require power from a mains supply or mobile generator. Usually a
4.5 kW generator will provide sufficient power to operate a 300 kV self-rectified set.
Gamma radiography can in general be carried out without the need for a power
supply.
An Iridium 192 isotope with a source activity of up to 100 curies can safely be
stored in a container weighing approximately 15-20 kg which has outside dimensions
of approximately 200 x 400 x 100 mm. Such isotopes are useful for the radiography of
A typical self-rectified 300 kV rated x-ray set (which is useful for the
radiography of steel sections of up to 60 mm thick) is on the other hand considerably
less portable and less manoeuvrable. A typical 300 kV SR tube head could weigh 55
kg and measure 300 x 300 x 750 mm while the associated control panel might weigh
as much as 30 kg and measure 450 x 350 x 250 mm. Low kilovoltage equipment
offers improved portability and manoeuvrability but this has to be offset against the
reduced penetrating power.
(i) The acetate or polyester base material is considerably thicker than is the case
with photographic film.
(ii) The emulsion is applied to both sides of the film. This effectively doubles the
film density (i.e. degree of darkness) for the same exposure to radiation and
thereby doubles the film speed.
(iii) The emulsion tends to be thicker (usually around 0.025 mm) than that used in
photographic films, in order to further increase the film speed.
Two types of radiographic film are used for industrial radiography, these being:
Direct type film, where the principal cause of image formation is the ionising radiation
itself. This may be coupled with the effect of “secondary electrons” emitted from
metallic foil intensifying screens.
and
Screen type film, where the principal cause of image formation is light emitted from
fluorescent image intensifying screens under the action of ionising radiation.
Some radiographic film will produce good results either as a direct type or as a
screen type film. The film emulsion in screen type films usually has a matt finish so as
to avoid reflecting the light produced by fluorescence. Radiographic film is generally
further divided into three categories, these being:
(BS EN 584-1 lists a total of 6 film classifications, C1 to C6. BS EN 1435, however, refers to only 3 of
these 6, these being C3, C4 and C5. Class C3 corresponds to ultrafine grain (e.g. Agfa D4). Class C5
corresponds to fine grain (e.g. Agfa D7 or Fuji 100) while C4 is intermediate between the two (e.g. Agfa
D5, Kodak MX or Fuji 80).
fine grain (medium speed film) e.g. Agfa D7, Fuji 100 or Kodak CX
and
Coarse grained film requires a shorter exposure time than does fine grain film
because each grain of silver halide needs only to receive as few a single photon of
radiation or single ‘secondary electron’ in order to become ‘sensitised.’ When a
sensitised grain contacts the developer solution the entire grain, regardless of its size
is converted to image forming metallic silver. Large grains of silver will block out more
light than small grains so a coarse grained film will appear darker after processing
than will a fine grained film even though the exposure conditions were exactly the
same. The table below lists the direct type radiographic films produced by various
manufacturers in order of their relative film speed.
For the most part, industrial radiography is carried out using fine grain film such
as Kodak CX, Agfa D7, Cronex NDT 70 or Fuji 100. Ultrafine grained film is used
where adequate sensitivity cannot be achieved using fine-grained film. Coarse-
grained film is rarely used where except very rapid results have to be obtained.
Radiographic film
When film is exposed to x-rays, gamma rays, or light, an invisible change called
a latent image is produced in the film emulsion. The areas so exposed become dark
when the film is immersed in a developing solution, the degree of darkening
depending on the amount of exposure.
Photography utilising film emulsion similar to that which is in use today has
been with us since around 1839 but the mechanisms involved in latent image
formation remained a mystery until 1938 when the ‘Gurney – Mott Theory’ was first
put forward. Although this theory is now generally accepted there remain areas of
speculation.
Formation of a latent image involves very subtle change in the silver halide
grain. It is known to involve the absorption of only one or a few photons of radiation.
Because of the small amount of energy involved it is obvious that only a few atoms,
out of the ten thousand million or so atoms in a typical silver halide grain, can actually
be affected. To date it has proved impossible to detect either the physical or the
chemical nature of the tiny changes involved. Against this, however, much can be
deduced about what the physical nature of these changes must be. For one thing we
know that the substance which forms the radiographic image must be metallic silver.
We also know that the latent image is localised at certain discrete sites within the
silver halide grain. The evidence for this is shown in figure below. This figure is an
electron micrograph of a section of film emulsion that has been exposed to light
followed by a brief contact with developer. Note how tiny amounts of silver have
appeared (the dark areas) within each grain of silver halide. Further it is known that
prolonged exposure to light will darken the film emulsion even without development.
Therefore the mechanism of latent image formation will by itself cause the release of
silver from a silver halide grain under extreme conditions (see figure on next page).
Film cassettes must be handled with care, flexible cassettes are particularly
easy to rupture during loading and unloading of the film. Cassettes which leak light
can add considerably to the cost of radiography if they lead to a radiograph having to
be retaken. Therefore it is good practice to inspect cassettes prior to use. Leaky
cassettes can often be satisfactorily repaired using opaque adhesive tape.
Practical Radiography. Rev 2 Oct 2005 7.4 WORLD CENTRE FOR
Methods of Producing a Radiographic Image MATERIALS JOINING
For most purposes lead foil screens 0.125 mm thick are used but thicker
screens are used for high-energy radiography. Copper screens tend to be used only
for extremely high-energy techniques (above 1 MeV). The lead screens found in pre-
packed film are only a few microns thick, they produce a strong intensifying effect but
have a much reduced effect on the scattered radiation as compared with standard re-
useable lead screens. Pre-packed film is available either in individual disposable
cassettes or as ‘roll-pack’ where a long narrow length of film is supplied complete with
lead screens in a protective light proof sheath. Roll-pack film can be cut to any desired
length. The cut ends have to be light sealed with suitable adhesive tape. Roll-pack is
commonly used on pipelines in conjunction with the panoramic technique.
BS EN 1435 tables 2 and 3 specify metallic foil screens of lead, copper, steel
and tantalum. The specified thickness range and screen material change for different
x-ray tube voltages and for different isotopes.
These screens, which attempt to combine the advantages of lead screens with
those of salt screens are occasionally used in industrial radiography when there are
strong financial pressures for a reduction in exposure time. One such application is on
offshore pipe laying barges. They are even more expensive than salt screens at
around £70 for a pair of 10 x 40 cm screens and they are just as easily damaged.
They do not provide quite the same reduction in exposure time as do salt screens but
the image quality is considerably improved (although still inferior to that produced
using lead screens). Fluorometallic screens consist of a cardboard backing material to
which a layer of lead foil is attached, attached to the lead foil is a layer of calcium
tungstate or other fluorescent crystalline material suspended in a suitable binding
material.
The first stage in film processing is development. During this stage a reducing
agent such as hydroquinone or metol reduces the sensitised silver halide grains in the
film emulsion to metallic silver. Development whether manual or automatic must be
carried out within the temperature range recommended by the developer
manufacturer otherwise image quality will be severely impaired.
Developers for manual processing are usually designed for use at 20°C, for
automatic processing this will usually be increased to around 27°C. Films should
always be developed for the optimum processing time of about 5 minutes for manual
development and under 2 minutes for automatic development.
Film developed for a time which is not within the developer manufacturer’s
recommendations will have impaired image quality. Films must be agitated thoroughly
in the first 20 - 30 seconds of development when the reaction with the developer is
rapid and at regular intervals (usually 10 seconds per minute) throughout the
remainder of allotted development time. Failure to properly agitate the film will result
in a streaky radiograph and inferior image contrast. The developer solution undergoes
chemical changes during film processing and it must be replenished regularly in order
to maintain its effectiveness. Exposure to air must be minimised because developer is
readily oxidised. Contamination of the developer with foreign material, especially
metal particles, is likely to lead to unwanted images being produced on the film.
The chemicals which are used to ‘fix’ the image and ‘harden’ the emulsion are
normally combined together in a single chemical bath. Both types of chemical have to
be protected by an acid buffer solution. In acid solution the active ingredient in the
fixer will dissolve only silver halide from the film; if the solution becomes alkaline it will,
Practical Radiography. Rev 2 Oct 2005 7.7 WORLD CENTRE FOR
Methods of Producing a Radiographic Image MATERIALS JOINING
The process of removing excess silver halides is called “fixing”. The chemical
used to achieve this is sodium (or ammonium) thiosulphate (sometimes called hypo).
The gelatine binder which holds the silver and silver halides becomes soft and
spongy in the developer. ‘Hardening’ the film serves to get rid of some of this
sponginess and gives the film better resistance to the formation of water marks during
drying.
7.1.4.4 WASHING
After fixing the film must be thoroughly washed so as to remove all traces of the
fixer chemicals from the emulsion. Insufficient washing will result in the formation of
brownish yellow stains while over-washing can cause water marks or even frilling (see
above). Adequate wash times in a running water wash vary from 10 minutes at 30°C
to 30 minutes at 10°C. Most film manufacturers recommend that the wash
temperature should not be more than 25°C. Film can be washed successfully in a still
water bath provided that the water is changed regularly.
7.1.4.5 DRYING
The application of a wetting agent to the film prior to drying will help the film to
dry quickly / evenly and without watermarks. If the films are to be dried using a warm
The image quality can approach that of film radiography (limiting sensitivity
about 2% as opposed to less than 1% for film radiography) and permanent records
can be made using a computer, photography or video.
Solid state imaging panels are now available. Such panels can form a very
bright image without reducing the size of the image. Panels measuring 250 mm x 250
mm are available which can produce an image about 50 times brighter than a
conventional fluoroscopic screen. Using such panels a resolving power of around 300
lines per inch can be achieved which is very close to that of radiographic film.
In still more recent developments similar panels have been used to form
digitised electronic images which can be viewed via a computer. This enables the
images to be processed by the computer in order to increase contrast, thereby
equalling the sensitivity produced in film radiography. Such systems can form
radiographic images from extremely low doses of radiation. This makes them ideal for
use with micro focus x-ray equipment. This in turn allows the image size to be
magnified simply by increasing the object to detector distance without a detrimental
increase in penumbra. So-called x-ray tomography is possible using this type of
equipment. Small components can be rotated through 360° while recording the
images formed using the computer. Equipped with right software computers can
convert this information into a three dimensional image. Such systems are in use in
the aerospace industry.
8.1.1 CONTRAST
Contrast can be defined as the ease with which it is possible to distinguish
between two adjacent areas of different film density. The chief factor that determines
whether or not the two areas will be clearly defined is the degree of difference in film
density.
and
The factors affecting film and subject contrast are discussed below:
The logarithm to the base 10 of a number is just the power of 10 that will
produce the number itself.
For example:
The previous 2 figures show how film density affects film contrast. Film density
does not affect subject contrast. The gradient of the film characteristic curve is a good
measure of film contrast. The gradient for all films increases with increasing film
density. If the gradient is steep then a small change in radiation intensity or exposure
time will produce a large change in film density. The gradient of all of the film
characteristic curves becomes shallow at film densities of less than 1.5, indicating that
film contrast will be poor at low film densities. In view of this all relevant national
standards stipulate a minimum film density of about 2.0 for industrial radiography.
The film attains about 80% of its final density in the first 30 seconds of
development. During the remaining 3½ to 5½ minutes of the standard development
time radiographic developers are designed to increase film contrast. The developer
works more vigorously in areas where a lot of metallic silver has already been
released than in areas where the converse is true. Thus film contrast gradually
improves during the final minutes of the development process. This is why
radiographs which have been ‘pulled’ – intentionally underdeveloped in an effort to
produce acceptable film density, invariably show poor film contrast. If the film is
allowed to remain in the developer for too long, however, the developer will begin to
attack all areas of the film and contrast will begin to suffer.
Nature of incident Half value layer Half value layer Half value layer
radiation (steel) / mm (aluminium) / mm (lead) / mm
Iridium 192 γ-rays 13 35 4.8
Cobalt 60 γ-rays 22 70 12.5
100 kV x-rays 2.4 15 0.1
150 kV x-rays 4.5 18.5 0.3
200 kV x-rays 6 21 0.65
300 kV x-rays 8 25 1.6
and
Unsharpness
(The unsharpness of the boundary between light and dark increases from left to right)
Geometric Unsharpness
Inherent unsharpness depends on three factors: the type film, the type of
intensifying screen and the quality (or photon energy) of the radiation.
Each silver halide grain in a fine grain film is around 1 m in size. To give an
idea of just how small this is, the pixel size on a computer screen at a resolution of
It is now known that most silver halide grains in a direct type radiographic film
are not sensitised directly by the penetrating radiation itself. They are for the most part
sensitised by the secondary electrons released by the intensifying screens and by
secondary electrons generated within the film emulsion itself. The greater the distance
the secondary electrons are able to travel within the emulsion the greater the resulting
unsharpness.
Salt screens fluoresce strongly under the influence of X-rays and produce very
large increases in film speed. They are not generally used in industrial radiography
due to the large increase in inherent unsharpness associated with their use.
Matter which has absorbed ionising radiation and which has therefore reached
an unstable energy state will emit energy in the form of radiation as it returns to a
stable energy state. Some of this radiation will be in the form of heat, in a few
specialised cases it will be in the form of light and in many cases x-rays will be
produced. Such x-rays are termed scattered radiation and they can very adversely
effect radiographic quality. Control of scattered radiation is therefore essential if high
quality radiographs are to be produced.
Scattered radiation
Photoelectric effect
( An incident X-ray photon, energy E0 collides with an outer shell electron which is ejected from the
atom with energy E0 – Eb where Eb is the binding energy)
Compton scattering
Pair production
Radiation may be scattered by parts of the object that are not within the
diagnostic area of the radiograph or by the walls of the exposure room. This is termed
‘side scatter’.
This type of scatter can be reduced by collimating the beam such that only the
area to be examined is subjected to the primary beam and by the use of lead
masking, diaphragms or grids. In x-radiography the use of a filter may help to reduce
side scatter.
Side scatter causes ‘undercutting’ of the radiographic image around the edges
of a component where these can be seen on the radiograph or at any site where there
is a large change in section thickness (e.g. a bolt hole). Undercutting causes a lack of
sharpness and may mask possible defect indications.
This type of scatter can easily be reduced by shielding the back of the film
cassette with a sheet of lead. A sheet of lead approximately 2 mm thick is adequate
Note: If a dark image of the letter ‘B’ appears this is not an indication of excessive
back scatter. It merely indicates scatter caused by the letter ‘B’ itself.
Should back scatter be detected then the thickness of the lead sheet shielding
the back of the film cassette must be increased.
8.2.2.3 SELF-SCATTER
If the radiation source is an x-ray tube then the use of a copper filter can help to
reduce the effects of this type of scatter. A copper filter significantly reduces the
proportion of low energy radiation within the primary beam. Since it is the low energy
radiation which is chiefly responsible for scatter the use of such a filter can reduce the
overall amount of scatter occurring and in this way improve image quality. Filters
made from lead, steel or other metals may be used in a similar way.
Metallic foil intensifying screens made from lead or other metals reduce the
effects of self-scatter for both x-ray and gamma ray radiography. They filter out the
low energy scattered radiation and prevent it from reaching the film.
8.2.3.1 COLLIMATION
Probably the single most effective way of reducing scatter is to collimate the
radiation beam. Collimators shield out most of the radiation that is not travelling in the
useful direction. X-ray equipment is always to some extent self-collimated – the
geometry of the hooded anode shields out much of the unwanted radiation produced,
but some x-ray heads may contain additional shielding. In gamma radiography
collimators consisting of hollowed out blocks of lead weighing around 2.5 kg are
common. More effective (but more expensive) collimators for gamma radiography are
made from tungsten or tantalum.
The principle of collimation is simply that if there is less radiation then there will
be proportionally less scatter.
8.2.3.2 DIAPHRAGMS
Using a diaphragm
Practical Radiography. Rev 2 Oct 2005 8.16 WORLD CENTRE FOR
Production of a Radiograph (Film Radiography) MATERIALS JOINING
8.2.3.4 GRIDS
The use of a grid is generally limited to medical radiography. A grid consists of
a matrix of parallel metal bars which is set in oscillation during exposure such that the
grid itself does not produce a radiographic image. The use grid is a very effective
method of reducing the effects of side scatter, but grids are very rarely a practical
option for industrial situations. In order to be effective the grid must be placed as close
as possible to the film. In microfocus x-radiography it may be placed between the film
and the object.
8.2.3.5 FILTERS
The previous figure in section 8.2.1.4 shows how the percentage of scattered
radiation is high when the radiation energy is low. Placing a thin sheet (typically 1 to 2
mm) of copper or other metal in the primary beam, close to the source of radiation,
greatly reduces the amount of low energy radiation while permitting most of the higher
energy radiation to pass through. If there is less low energy radiation there will be less
scatter, although it is possible that film contrast will be reduced. The use of a filter to
reduce scatter is limited to x-radiography because gamma ray sources do not produce
long wavelength low energy radiation.
Oscillating grid
In addition to this both screens intensify the effect of radiation, the energy of
which, exceeds 120 keV. The screens do this by producing secondary electrons to
which the film emulsion is sensitive. Most of the radiation exceeding 120 keV will be
part of the primary beam. Thus the effect of the primary beam is amplified at the
expense of the unwanted scattered radiation.
(* Note that in film radiography it is better to think of intensity as photons per square metre
rather than energy per square metre. This is because radiographic film has a fairly flat
response to changes in photon energy right across the full spectrum of x and gamma rays. In
terms of film density 1 photon of 150 keV radiation has much the same effect as 1 photon of
20 MeV radiation.)
The factors listed in below table will affect either the film speed or the amount of
radiation reaching the film and have to be taken into account when determining the
correct exposure for film radiography. In addition to these factors the required film
density obviously has an impact upon the required exposure time (see same table).
Taking the example given in the figure just above let’s suppose that we wish to
achieve a geometric unsharpness of 0.25 mm or better. This would be sufficient to
satisfy the requirements of most national codes or standards including BS EN 1435
class A.
but the OFD can be taken as being equal to the material thickness, in this case 18
mm, and the FFD is equal to the FOD plus the material thickness so:
If we choose to use the Pantak 200 CP the effective focus size on the broad focus
setting is about 4 mm, so:
Note that in BS EN 1435 ‘standard techniques’ are ‘class A’, while ‘enhanced techniques’ are ‘class B’.
The exposure chart for the Pantak 200 CP is drawn for a source to film distance
of 914 mm. The maximum tube current is 14 mA. Let’s use 10 mA. We would like to
achieve an exposure time of between 1 and 5 minutes, giving an exposure of between
10 and 50 mA-mins. The focus to film distance that we wish to use is 400 mm. The
exposure chart has been constructed using FFD = 914. Using the inverse square law
(see figure 55) we can see that an exposure of 10 to 50 mA-mins at 400 mm FFD is
equivalent to an exposure at 914 mm FFD of between 10 x 9142 / 4002 = 52.2 mA-
mins and 50 x 9142 / 4002 = 261 mA-mins.
Looking at the exposure chart (see the 3rd figure in section 8.4.1) a density of
2.2 will be achieved using Kodak CX with an exposure of about 150 mA-mins at 120
kV or with an exposure of about 40 mA-mins at 140 kV. These values are for an FFD
of 914 mm. The equivalent exposures at an FFD of 400 mm will be:
Thus for Kodak CX and a film density of 2.2 these exposures should work:
The 120 kV exposure should produce the best film contrast while the 140 kV exposure
will be more economic.
Intensity = number of photons per square metre. At 2D the same number of photons that
passed through one square at D now passes through 4 squares. Thus the intensity at 2D is
one quarter of what it was at D. Intensity is proportional to 1/(distance)2. Radiographic
exposure time is proportional to 1/(intensity), thus exposure time is proportional to (distance)2.
If for example a film density of 2.2 was thought to be too light, then the film
characteristic curve (see figure in section 8.1.1.1) can be used to find the correct
amount of exposure compensation. For example, let’s say that we wanted to achieve
a film density of 2.5.
Using the logarithm from above mentioned figure (relative exposure) for a film
density of 2.2 on CX = 1.3 while for a film density of 2.5 logarithm (relative exposure)
= 1.38
1.38 – 1.3 = 0.08
Thus to change from CX to MX we divide by the film factor for CX (= 2.50) and
multiply the result by the film factor for MX (= 10.0).
Therefore using MX film, for a film density of 2.5 our exposures become 140
mA-mins at 120 kV or 37 mA-mins at 140 kV. The exposure at 120 kV is now rather
long (14 minutes at 10mA) so an exposure using 140 kV of around 3 minutes at 12.1
mA would be preferred.
All of the exposure charts in the figures above have been constructed for steel.
If radiography has to be carried out on materials other than steel then the exposure
time will have to be adjusted to compensate for the difference in radiation absorption.
This can be done using ‘half value layers.’ The half value layer of a given material for
a given incident radiation energy being the thickness of the material which reduces the
intensity of the incident radiation by a factor of 2. However the simplest way is to use
‘equivalence factors’, some examples of which are listed in this table.
Equivalence factors
Suppose that we wish to radiograph our 18 mm thick steel weld using the
Pantak 200 at 140 kV with a focus to film distance of 400 mm. The exposure required
for a film density of 2.5 without a filter was calculated above as 9.2 mA-mins if using
CX film. The equivalence factor for copper at 140 kV is about 0.64. Therefore 1.0 mm
of copper will be radiographically equivalent to 1.0 ÷ 0.64 x 1.0 = 1.6 mm of steel. To
find the correct exposure (for a copper filter thickness of 1 mm) we simply need to add
this amount to the steel thickness which is being radiographed :
18 + 1.6 = 19.6 mm
The 3rd figure in section 8.4.1 gives an exposure of about 50 mA-mins for 19.6
mm of steel at a focus to film distance of 914 mm, which is an increase of 25%
compared to the exposure which was required under the same conditions without the
filter. Thus the exposure required with a FFD of 400 mm using CX film for a film
density of 2.5 increases by 25% from 9.2 to 11.5 mA-mins.
These days the type of IQI most commonly in use is the wire type but other
types exist, two examples being the plaque type and the step hole type.
Note that it is fairly easy to remember the wire diameters: if you can remember the diameters
of the first three, 3.2, 2.5 and 2.0 mm you can arrive at all other wire diameters by halving as
shown just below.
Looking along each row the wire diameters are successively halved, e.g. 3.2, 1.6, 0.8…..
The EN 462-1 material groupings are as follows: the FE designated IQIs are
made from low alloy steel and cover all ferrous materials; the CU designated IQIs are
made from copper and cover copper, tin, zinc and their alloys; the AL designated IQIs
are made from aluminium and cover aluminium and its alloys; the TI designated IQIs
are made from titanium and cover titanium and its alloys. Special IQIs can be used for
materials lying outside these 4 groups, or the contracting parties could agree to use
one of the four normal designations.
Several other wire type IQIs are in common usage. Amongst these are the BS
3971, the DIN 54 109 and the ASTM E 747. The BS3971 series consisted of 21 wires
where wire number 21 was the thickest at 3.2 mm and wire number 1 was the thinnest
at 0.032 mm, the order of numbering was basically the reverse of that used by EN
462-1 with wire number 11 being the same diameter in both series (0.32 mm).
The DIN 54 109 series consisted of 16 wires corresponding exactly to the first
16 wires in the EN 462-1 series. The ASTM E 747 series consists of 21 wires ranging
from 0.08 mm to 8.1 mm in diameter; there are 4 overlapping groups of 6 wires, each
designated by a letter (A, B, C or D), see table below, and a large number of material
groupings each designated by a number with ferrous being ‘1’.
IQI
WIRE DIAMETERS
Designation
A 0.08 0.1 0.13 0.16 0.2 0.25
B 0.25 0.33 0.4 0.5 0.63 0.81
C 0.81 1.0 1.27 1.6 2.0 2.5
D 2.5 3.2 4.0 5.1 6.3 8.1
BS EN 462-2 IQIs consist of stepped blocks of material with each step having a
through drilled hole or pair of through drilled holes. Step thicknesses of 0.8 mm or less
have two drilled holes, while the thicker steps have a single hole. In each case the
step thickness and the hole diameter are equal.
IQI sensitivity is usually defined as the thickness of the thinnest wire, plaque or
step that is visible on the radiograph expressed as a percentage of the specimen
thickness. Exactly what is meant by specimen thickness varies from standard to
standard and from technique to technique. BS EN 1435 contains tables of ‘essential
wires’ for class A and class B techniques for IQI placed source or film side. ASME V
article 2 also permits the use of wire type IQIs and takes a similar ‘essential wire’
approach. BS EN 1435 bases its requirements on nominal thickness while ASME V
article 2 bases its requirements on the actual weld throat thickness.
It used to be common good practice to place the IQI in the least favourable
position within the diagnostic area of the radiograph. This would usually have meant
placing the IQI upon the source side of the specimen and towards the extremities of
the diagnostic area because this is where the contrast and definition would tend to be
at their least favourable (highest value of geometric unsharpness and lowest film
density).
In the past it was not uncommon for national codes or standards to specify an
overall requirement for a radiographic sensitivity of 2% or better. This was easy to
achieve on thicker sections but often impossible to achieve on thinner sections of
material. Modern radiographic standards take account of the fact that the best
achievable sensitivity for a given situation and technique is not a fixed quantity but a
variable which depends upon such factors as the type of radiation source, the
technique and the thickness of the specimen. Such standards specify a minimum
It is important that IQIs are placed source or film side and at a position within
the diagnostic film length (DFL) in accordance with the requirements of the contract
specification.
As a general rule, wherever possible, the IQI should be placed source side.
IQIs placed source side are affected both by radiographic contrast and by geometric
unsharpness. Film side IQIs indicate radiographic contrast only, thus source side IQIs
give a more accurate measure of the overall radiographic quality.
It used to be standard good practice to place wire type IQIs towards the end of
the diagnostic area, with the thinner wires toward the outside of the DFL; the wires
were invariably placed across the weld and sensitivity was assessed on the weld
allowing for any weld reinforcement present. This way of working would still meet
ASME V article 2 requirements, although this document does not specify where within
the DFL the IQI should be placed. In Europe matters are different. When working in
accordance with BS EN 1435 sensitivity should generally be assessed at the centre of
the DFL on the parent material.
Plaque type and step hole type IQIs should (preferably) always be placed at the
centre of the diagnostic area on the parent material. Should the image of these IQI
types encroach on the weld area the radiograph should be re-taken.
If working with a wire type IQI in accordance with ASME V article 2 sensitivity
would probably be measured on the weld. ASME V article 2 then has a requirement
that the film density through the diagnostic length shall not vary by more than + 30%
or – 15% from that measured at the IQI. The same allowable density variation applies
to plaque type IQIs, but these, of course, must be placed alongside, not on the weld.
Plaque type IQIs may be shimmed to compensate for any weld reinforcement. If a
technique produces a wide range of film density the placement of several IQIs may be
necessary in order to meet the allowable density variation requirement.
All national codes and standards require the use of location markers, usually in
the form of lead letters or numbers that appear in the radiograph as a radiographic
image. It is very important that the markers are placed in such a way as to prove
coverage of the weld where a multiple exposure technique is used. Three general
rules apply:
(1) When performing radiography of welds in flat plate location markers must be
placed source side. Film side markers will not prove coverage because of
‘parallax.’
All national codes and standards require unique and permanent identification of
radiographs. In general this can be applied by any suitable means although in some
cases identification using lead numbers that appear as radiographic images is
required. Where not prohibited by the contract specification ‘flashing’ the radiographic
identification is a good method. he required identification is written on a scrap of white
paper, the radiograph is suitably masked and the scrap of paper is placed on the
unmasked area. The radiograph is then flashed with a suitable light source and the
identification becomes visible during subsequent film processing. Exactly what
constitutes an acceptable unique identification varies widely from specification to
specification, but the minimum is a unique number. ASME V article 2 requires a
unique weld number, the date and the manufacturer’s name or symbol. Most codes
require radiographs of repair welds to be marked with R1, R2, R3 etc depending on
the number of repair attempts. RW is commonly used to identify a complete re-weld.
Items such as heat treatment condition, welder number and welding procedure
reference may also be required.
BS EN 1435 specifies the maximum x-ray tube voltage that may be used based
on the component thickness (see the figure in section 5.1.5.1 above). BS EN 1435
also specifies the minimum and maximum thickness on which each type of gamma
ray isotope may be used (see table just below). ASME V article 2 specifies a
recommended minimum steel thickness for iridium 192 of 19 mm and a minimum of
38 mm for cobalt 60, a minimum of 63 mm of aluminium is specified for iridium 192.
BS EN 1435 applicable thickness ranges for gamma ray sources and high
energy x-rays
The minimum source to film distance for BS EN 1435 is calculated using the
formula: f/d ≤ kb2/3 where f is the source to object distance, d is the effective source or
focus size, b is the object to film distance and k is a constant equal to 7.5 for class A
techniques and 15 for class B. BS EN 1435 also includes a nomogram for the less
mathematically minded. The first figure in section 8.4.1.1.1 above shows the BS EN
1435 requirements graphically.
The figure above shows a typical set-up for exposure of a butt weld in flat plate.
The captions refer to BS EN 1435 requirements. The source should be positioned on
the centre line of the weld, directly above the centre of the diagnostic area.
This technique is commonly used for pipeline welds where specially designed,
remotely operated, devices known as crawlers are often used. These machines can
travel distances of up to several kilometres along the inside of the pipeline in order to
reach the desired position to radiograph a particular weld. The typical battery life for
an x-ray crawler will usually allow about 100 exposures to be made between
successive battery charges. Gamma ray crawlers are also used.
This technique may also be used for examining girth welds in cylindrical
pressure vessels. Using Thulium 170 isotopes boiler tube welds which may have an
outside diameter of only 40 mm are occasionally examined by this technique.
The major advantage of this technique is that it can radiograph an entire girth
weld in a single exposure. With this technique location marker placement is not
critical, but it usually more convenient to place the markers film side. In most cases it
will be impractical to place the IQI source side for this technique, although source side
IQIs would be preferred if access is not a problem. Film side IQIs are therefore
generally used. Comparitor radiographs having IQIs placed source and film side can
be used to establish sensitivity requirements for film side IQIs. In most cases three
IQIs are placed at 120° intervals around the circumference, although some
specifications require more or less than this. The radiograph may consist of a number
of overlapping films or it may be a single length of ‘rollpack’ film. Identification of the
film may be included as a radiographic image but it may also be added later. Where
several overlapping films are used each film must be uniquely and permanently
identified.
Practical Radiography. Rev 2 Oct 2005 10.5 WORLD CENTRE FOR
Radiographic Techniques (for Welds in Plate and Pipe) MATERIALS JOINING
Single wall single image source internal & offset technique (BS EN 1435)
Required number of exposures: see figure next page for BS EN 1435 requirements
Single wall single image film inside, source outside (BS EN 1435)
Practical Radiography. Rev 2 Oct 2005 10.6 WORLD CENTRE FOR
Radiographic Techniques (for Welds in Plate and Pipe) MATERIALS JOINING
In the double wall single image technique the source of radiation is usually
placed at the minimum possible distance from the film. The reason for this is that as
the source to film distance increases so does the number of exposures needed to
cover the entire circumference of the weld. In addition, any improvement in image
quality due to the reduced geometric unsharpness associated with an increase in SFD
or FFD has to be offset against a reduction in image quality due to increased scatter.
X-ray tubes are bulky and the minimum achievable FFD will usually be about
125 mm plus the outside diameter of the pipe. A minimum of 4 exposures per weld is
therefore required when using an x-ray source for this technique.
Being able to place the source of radiation in close contact with the pipe gives
gamma ray techniques another significant advantage over x-ray techniques
particularly on smaller pipe diameters. Less offset is needed with gamma ray sources
in order to ensure that the image of the source side portion of the weld is not
superimposed upon the film side part of the weld. This can increase the chance of
finding vertical defects such as lack of root fusion in the weld being radiographed.
As the wall thickness to diameter ratio increases the double wall single image
technique becomes increasingly difficult to apply, the number of exposures required
increases and the quality of the radiographs produced diminishes. For these reasons
double-wall-double-image (superimposed) techniques tend to be preferred for heavy
wall small diameter pipes.
Because in general there will be no access to the inside of the pipe when this
technique is employed the location markers and IQI are always placed film side.
For pipes with a wall thickness to outside diameter ratio in excess of about 0.15
the double-wall-double-image (superimposed) technique is to be preferred.
Long source to film distances are needed because the minimum value of object
to film distance is equal to the outside diameter of the pipe. Exposure times for this
technique therefore tend to be rather long especially in the case of gamma ray
techniques.
BS EN 1435 requires wire type IQI s to be placed on the parent material with
their wires parallel to the weld axis (see previous figure above). Special BS EN 462-1
having wires just 10 mm long are available for this purpose. Working in accordance
with ASME V article 2 standard wire type IQIs should be placed with their wires across
the weld at 90° to the weld axis.
In the double wall single image technique the film is wrapped around the pipe
so as to remain as close as possible to the weld. Conversely, in the double wall
double image technique the film should be kept as flat as possible (see previous
figure).
This technique has the same range of application as the elliptical technique, but
is preferred when the thickness to outside diameter ratio exceeds 0.15. Welds having
difficult geometry that may prevent them from being radiographed using the elliptical
technique can generally be radiographed successfully using this technique.
As the image of the source side part of the weld is superimposed on the image
the film side part of the weld it is often not possible to accurately locate a weld defect
when using this technique. This is not usually much of a handicap because small
diameter welds tend to be cut-out and re-welded rather than being repaired locally.
A single location marker per exposure is usually sufficient when using this
technique and it may be placed either source or film side. IQIs should always be
placed source side.
The double wall double image superimposed technique may be more likely
(than the elliptical technique) to successfully detect lack of root fusion due to the more
favourable angle of incidence of the primary beam.
11.1 INTRODUCTION
Competent interpretation of radiographs is a skill which can only be gained
through long experience. This chapter is intended to give the reader a guide to
radiographic interpretation and should be regarded as a base upon which to build.
(a) The film emulsion is swollen with water and the images are not as clear as
when the film is dry.
(b) The emulsion is very delicate and any attempt to mask the film will result in
scratches or marks on the film, effectively ruining it.
11.3 REPORTING
ASME V article 2 requires that the film density within the area of interest must
not vary by more than minus 15% or plus 30% from the value measured through the
body of the IQI. If necessary additional IQIs can be used in order to satisfy this
requirement for exceptional areas.
Occasionally an upper limit is specified for film density. ASME V article 2, for
example, specifies an upper limit of 4.0.
These are produced by careless film handling. If the film is crimped or buckled
either before or after exposure crescent shaped images in the processed radiograph
will result. Light marks indicate crimping before exposure. Dark marks indicate
crimping after exposure but before film processing.
It is usually possible to identify crimp marks by viewing the film in reflected light.
They should appear as indentations in the surface of the film. Lead screens that have
been crimped should be discarded.
These will appear as dark spots on the film and indicate poor dark room
practice. Such marks are usually visible in reflected light.
These will appear as light spots on the film and again indicate poor dark room
practice. Such marks are usually visible in reflected light.
These are easily seen on the radiograph in both transmitted and reflected light
and are due to uneven drying. They commonly occur where a dry or partially dry film
is wetted locally either by splashing or by excess water running down from a film clip.
The appearance of water marks can be reduced or eliminated by the use of a
squeegee to remove excess water or by the use of a final wash that contains a small
amount of detergent (i.e. a wetting agent).
Radiographs affected by dichroic fog will appear reddish when viewed using
transmitted light and greenish in reflected light. Dichroic means two-coloured. This
artefact is caused when the development process continues during the fixing process.
It happens when the fixer solution has become insufficiently acidic to stop the
development process. The use of an acidic stop bath between the development and
fixing processes will in general prevent the occurrence of this seldom seen artefact.
11.6.14 RETICULATION
This artefact appears on the radiograph as an orange peel like mottling effect.
It is caused when the film emulsion is subjected to a temperature shock at any stage
during the film processing. A sudden change in temperature causes the film emulsion
to wrinkle. It will not generally occur as long as the sudden change in temperature is
less than 10°C.
Film stored at too high a temperature or which has been exposed to chemical
fumes may become fogged. The fog level of all film increases with age, even under
ideal storage conditions, therefore all film boxes are marked with an expiry date.
High-speed films deteriorate more quickly than do slower films.
11.6.18 SOLARISATION
It should be stressed again that artefacts are cause for rejecting the film only if
they interfere with interpretation. A large number of artefacts present on the
radiographs indicate poor practice and the interpreter should take time to inspect the
radiographic facilities and review darkroom procedures.
Root concavity
This is a continuous or intermittent channel along the edge of the weld, due to
insufficient weld material. Incompletely filled groove is a fusion defect and should not
be confused with lack of reinforcement or undercutting. It produces an image in the
radiograph of a straight edged (on one side at least) dark band. Incompletely filled
groove is caused by poor welding practice.
This is a concave area of the weld cap where the weld is locally thinner,
sometimes thinner than the parent material. In the radiograph it appears as a dark
area towards the centre of the weld which has diffuse edges. Lack of reinforcement is
caused by poor welding practice.
Lack of reinforcement
11.7.1.5 UNDERCUT
This is an irregular groove at the toe the weld in the parent material due to
burning away during welding. It appears in the radiograph as a dark / irregular
/intermittent band in a position adjacent to either the cap or root weld toe or between
adjacent capping runs. It may therefore appear inside or outside the weld image on
the radiograph. The major causes of undercutting are excessive amperage and poor
welding technique. Welds in the vertical or horizontal – vertical position tend to be
prone to undercutting.
Practical Radiography. Rev 2 Oct 2005 11.11 WORLD CENTRE FOR
Interpretation of Radiographics MATERIALS JOINING
11.7.1.6 SPATTER
Spatter consists of globules of molten filler metal expelled during arc welding on
to the surface of the parent material or weld. Spatter appears in the radiograph as
small light spots. The major causes of spatter are incorrect polarity and welding
current too high. Spatter particularly affects MIG, MAG, MMA and FCAW, spatter is
highly unlikely to be seen in association with welds made by TIG or SAW. In pipe
welding spatter is possible on both the external and internal surfaces.
Weld spatter
Excessive dressing
This is a surface imperfection, usually of the parent material but also the weld
metal where a component has been in service. It usually takes the form of small
depressions resulting from localised corrosion. Pitting appears in a radiograph as
small dark rounded images. It is possible to mistake this for a welding defect, its
appearance in the radiograph can be identical to that of porosity.
11.7.2.1 CRACKS
Cracks are often invisible on radiographs but when they are detected they
appear in the radiograph as dark, fine often branching lines which are usually diffuse
or discontinuous. The ability of the radiographic technique to detect a crack is
dependent on the crack’s orientation relative to the direction of the radiation. The
figure below shows how even a slight deviation from the optimum orientation will
greatly reduce the change in section thickness that the radiation experiences due to a
planar defect such as a crack. In the case shown a variation from optimum incidence
of just ± 1° will reduce the change in penetrated thickness from 10 mm to 1 mm for a
planar defect measuring 10 mm by 17 m.
Any crack that lies across the weld axis is called a transverse crack. Basically
there are two distinct types of transverse crack. Both types usually occur when the
compressive strength of the parent material is significantly greater than the tensile
strength of the weld metal.
The first type is a shrinkage or solidification crack. Cracks of this type usually
occur at 90° to the weld axis, often affecting the root pass of single sided welds. In
nature they are very similar to centreline cracks, but the source of restraint is different.
They are relatively easy to detect by radiography.
The second type is a chevron crack. This type of crack occurs at an angle of
about 45° to the weld axis, usually at some time after the completion of welding.
Chevron cracks are a special type of hydrogen induced crack; the stress that causes
the crack being due to an excessive amount of dissolved hydrogen in the weld metal.
They are sometimes detected by radiography, but in situations where there is a known
problem, other NDT methods with a higher probability of detection should be used.
Various mechanisms can lead to cracking in the heat affected zone of a weld.
Heat affected zone cracks will often start at or run to the toe of the weld since there is
always a high stress concentration at this point. In ferrous welds the hardest, most
martensitic, most brittle microstructure is usually to be found in the heat affected zone.
It is this ‘susceptible grain structure’ that makes the heat affected zone a prime site for
cracking.
Heat affected zone cracks are usually caused by one of two mechanisms.
The first of these involves dissolved hydrogen. Molten iron has a very high
solubility for hydrogen while solid iron has a very low solubility. Thus as the metal
freezes hydrogen will attempt to leave solution and escape from the weld pool but this
process is slow compared to the process of freezing, therefore most of the hydrogen
becomes trapped in the solidified metal. The trapped hydrogen then diffuses through
the metal crystals and begins to build up an internal pressure at points of weakness,
A second type of cracking that can occur in the heat affected zone of a weld is
sometimes called weld decay. This type of cracking can affect stainless steels and is
caused by the precipitation of brittle material at the grain boundaries. The brittle
material is chromium carbide. All stainless steels contain a small proportion of carbon
which is generally held in solution within the austenitic grains. The heat from welding
can cause this carbon to combine with the chromium which is present forming
chromium carbide which is an extremely brittle material. Weld decay can be avoided
by reducing the carbon content of the parent material and filler wire. Cracking caused
by weld decay is unlikely to be detected by radiography.
Crater cracks
This type of crack occurs when the heat source is removed too suddenly at the
end of a weld run. The cracking mechanism is the same as that for centreline
cracking. The major dimension of a crater crack is usually less than 5 mm. They often
have a star shaped appearance in a radiograph and they are relatively easy to detect.
Many welding standards will permit this type of cracking provided that it does not
exceed a specified maximum dimension.
Lack of fusion is an area where the solid material immediately adjacent to the
molten weld pool failed to become molten during the welding process leading to a lack
of union between the molten weld material and the adjacent solid material. The ability
of radiographic techniques to successfully detect lack of fusion is strongly dependent
on the orientation of the defect with respect to the incident beam of radiation (see
figure in section 11.7.2.1). Given favourable orientation lack of fusion with the parent
material will appear in the radiograph as a fine dark straight line which may be
continuous or intermittent. Unfavourably orientated lack of fusion with the parent
material may sometimes still be detected due to the presence of associated slag
inclusions or porosity. A slag inclusion with a straight edge normally indicates lack of
fusion and gas escaping from an area lack of fusion during the deposition of a
subsequent welding run may lead to a line of ‘linear porosity’.
Lack of fusion between subsequent layers of weld material will generally not be
detected by radiography unless it is associated with some other type of defect such as
slag.
This can be described as the failure of the weld material to extend into the root
of a joint. Incomplete penetration is a fusion defect, not to be confused with root
concavity (see section 11.7.1.2 above).
Where welds are deposited without a root gap lack of penetration may appear
as a single continuous or intermittent dark line. It should be noted that root gaps
frequently close during welding so even in cases where there should have been a root
gap lack of penetration may appear in the radiograph as a single dark line.
Most welding slag and other possible sources of non-metallic inclusions are
radiographically much less absorbing than the surrounding metallic material, therefore
they appear in the radiograph as dark images.
Although very rarely used, some types of covered welding electrode have a
high barium content in the flux coating. These electrodes produce a slag which is
radiographically denser than steel. In this case, therefore, a slag inclusion may appear
as a light image.
Dependent upon the nature of the welding process it is possible for foreign
metallic material to become entrapped within the molten weld material. Associated
with the gas tungsten arc welding process, tungsten inclusions are probably the
most commonly encountered form of metallic inclusion. They are caused by the break-
up of the non-consumable tungsten electrode during welding. Since tungsten has a
melting point well in excess of 3000°C particles of tungsten falling into the weld pool
do not become molten. Tungsten is radiographically extremely dense, therefore
tungsten inclusions always appear as bright - light images which tend to be angular.
They are usually quite small - typically around 0.5 mm. Copper inclusions can occur
particularly with the submerged arc or other welding process where the consumable
electrode is fed through a copper contact. If the copper contact gets too near to or if it
touches the weld pool molten copper (melting point about 900°C) will become
included in the weld pool. Copper is radiographically more absorbing than most other
materials including steel so copper inclusions may produce light rounded images with
extremely diffuse edges. Copper inclusions in ferritic steel welds usually cause severe
transverse cracking.
Metallic inclusions are quite common in aluminium welds, where such welds
are not properly segregated from their steel counterparts. Aluminium melts at around
660°C, so particles of steel or iron oxide falling into the weld pool will not become
molten (the melting point of steel is about 1400°C). Contamination can easily occur if
tools such as grinding disks which have been used for steel are used on aluminium.
Steel inclusions in aluminium appear as very bright angular shapes with sharp edges.
Slag inclusions
Tungsten inclusion
The solubility for gas of the molten weld material is many times that of the solid
weld material, thus as the material freezes there is a tendency for any dissolved gases
to precipitate from solution causing gas pores or porosity in the finished weld. Gas
pores are extremely easy to detect by radiography since they are not sensitive to the
direction of radiation and the gas which fills them is many times less radiographically
dense than the surrounding material. Gas pores appear on a radiograph as sharply
defined dark circular spots. They may be isolated, grouped or evenly distributed.
Aligned porosity is usually an indication of lack of fusion (see 11.7.2.2 above). Evenly
distributed porosity generally indicates that the electrode was in some way faulty,
group porosity usually occurs at restarts and is due to poor welding technique.
These will generally only occur in the root run of welds deposited by manual
metallic arc. Welds deposited using cellulosic coated electrodes (AWS E6010, 7010
etc.) are more likely to suffer from this defect than welds deposited with other types of
electrode.
Hollow bead can be caused by holding the arc at too shallow an angle with
respect to the work piece or by a strong draught of air along the inside of the pipe
during welding. On the radiograph it has the appearance very similar to that of slag.
The radiographic indication usually has rounded ends and it is always situated along
the centre of the root bead.
These occur due to shrinkage at the end of a weld run where the source of heat
was removed too suddenly causing the weld pool to freeze too rapidly. This defect is
quite common when the welding process is gas tungsten arc but it may also occur
with shielded metallic arc and other welding processes.
A crater pipe will appear in the radiograph with an image very similar to that of
a wormhole. It can only be distinguished from a wormhole by its position in the weld.
Crater cracks are shrinkage cracks and as such have a relatively greater volume than
do most other cracks. They often have a star like appearance in a radiograph. Their
radiographic image rarely measures more than 3 or 4 mm.
(i) Voids.
(ii) Cracks.
(iii) Cold Shuts.
(iv) Segregation.
(v) Inclusions.
11.8.1 VOIDS
Voids in castings are formed by gases dissolved in the molten material
precipitating from solution during the solidification process or by shrinkage caused by
inadequate feeding.
This is a large cavity formed during the solidification process that occurs to lack
of sufficient feed material. With good mould design macroshrinkage (also called
piping) should be confined to the feeder heads.
This is a coarse form of shrinkage that has smaller physical dimensions than a
macroshrinkage cavity. These cavities may be extensive and branching in nature.
Filamentary shrinkage occurs at the point in a casting that freezes last. Theoretically
this should always be at the centre of a section but this is not always the case. On
some occasions the defect may actually extend to the surface of the casting.
This is a very fine form of filamentary shrinkage due to lack of sufficient feed
metal or gas or both, in which numbers of cavities occur either round the grain
boundaries or between the dendrite arms (a dendrite is a material crystal which in the
initial stages of growth is tree-like). These cavities tend to link up in a three
dimensional network throughout the material.
In the radiograph the images of these cavities are superimposed and generally
produce a mottled or cloudy effect. In non-ferrous alloys, particularly magnesium
based alloys, microshrinkage may occur in layers and produce dark streaks in the
radiograph.
11.8.1.5 GASHOLES
A gashole is a discrete cavity greater than about 1.5 mm diameter caused by
gas evolved from the material as it freezes. It may also arise from gas evolved from
the core or mould (in which case the defects are called blowholes).
The radiographic image appears as a dark area of smooth outline that may be
circular or elongated and can be associated with pinhole porosity. Gasholes
occasionally become elongated as they try to rise to the surface of the molten material
during cooling, in this form they are known as wormholes. The radiographic image of a
wormhole may vary from a circular to an extremely elongated image depending upon
the angle of view.
In pressure die casting air may be injected with the material. In this instance the
defect is usually more severe in the runners and may assume an angular form. In
pressure and gravity die castings this defect may occur in clusters or as strings of
small voids. In investment casting it may appear as small rounded voids.
11.8.2 CRACKS
The defects are not necessarily continuous, they may exist in groups and will
often terminate at the surface. Hot tears may sometimes be referred to as a pulls.
Radiographically hot tears are revealed as wavy, ragged dark lines which are
often discontinuous, with areas appearing as approximately parallel dark lines which
may possibly be overlapping. Generally the ends of the indication taper to become
fine.
These are well defined and approximately straight cracks formed after the
material has become completely solid, quite large stresses being required to cause
fracture. Distinctions are sometimes drawn between types depending upon the time at
which fracture occurred.
In the radiograph stress cracks are often revealed as clearly defined smooth
dark lines - thus differing from the ragged appearance of a hot tears.
In the radiograph these defects usually appear as a dark lines. They may be difficult to
distinguish from a hot tears except by the typical involute appearance of the end of the
defects. The shape of an unfused chaplet or unfused chill in a radiograph is
dependent upon orientation of the beam. A cold shut resulting from a splash may
appear as a dark crescent or circle.
11.8.4 INCLUSIONS
These consist of foreign matter entrapped in the casting. This may be of many
types including sand, slag, flux, or dross.
11.8.5 SEGREGATIONS
These result from local concentrations of any of the constituents of an alloy.
They may be classified as general, localised or banded.
This is the simplest method of localisation, but it is rather limited in its field of
application. A typical test object, where this method might be useful would be a small
to medium sized casting that has a fairly simple cross section. The figure below shows
how this method would work on a small cylindrical casting. Two radiographs are taken
with primary beam mutually at 90° to each other. In an ideal situation the component
would be placed on some kind of turntable so that it could be moved accurately
keeping the two exposures in the same plane relative to the axis of the component.
The apparent defect position in each radiograph can be measured relative to
convenient datum point and the results plotted on a sketch. The defect position is then
deduced by triangulation.
2nd figure below shows how this method could be used to locate a slag
inclusion in a butt weld. In order for this method to work well a high degree of
dimensional accuracy is needed. The source to object distance, the object to film
distance and the distance that the source is moved between the successive
exposures must all be accurately measured and controlled. Two ‘half ‘exposures are
made from different source positions using the same radiographic film to produce two
flaw images. The distance between the two images, ‘m’ is then measured and the flaw
depth is calculated as shown. The source shift, which is usually about one sixth of the
source to film distance, has been exaggerated in the figure.
But:
so:
Refer to same figure below. The three triangles in the enlarged view will be
very close to similar as long as the source or focus to film distance is long in
comparison with the thickness. The calculation below assumes that the triangles are
similar.
so: and:
and:
so:
and:
t, the thickness of the plate is known and x, y and z can be measured on the
radiograph. Therefore d can be calculated.
The SI unit of ionising effect is the ‘coulomb per kilogram’, the quantity of
ionising radiation that produces a total electric charge of 1 coulomb per kilogram (Ckg-
1
) of air at STP.
or
1 Ckg-1 = 3876 R
The SI unit of absorbed dose is the ‘gray’ (Gy). The gray is defined as the
quantity of ionising radiation which releases 1 joule of energy per kilogram of
absorber.
The cgs unit of absorbed dose is the ‘roentgen absorbed dose’ (rad). The rad is
defined as the quantity of ionising radiation which releases 100 ergs of energy per
gram of absorber.
1 Gy = 100 rad
The units of radiation absorbed dose can be approximately related to the units
of ionising effect as follows:
1R = 0.88 rad
The conversions above are approximate since the relationship between the
roentgen and the rad or the coulomb per kilogram and the gray varies to some extent
with radiation energy.
Practical Radiography. Rev 2 Oct 2005 13.1 WORLD CENTRE FOR
Units Used in Radiography MATERIALS JOINING
The effect which ionising radiation has on our bodies varies with the type of
radiation and also, to some extent, with radiation energy. In order to compensate for
this a quality factor (QF) is introduced. Quality factors for several types of ionising
radiation are listed in table below.
Quality factors
In the cgs system multiplying the dose in rad by the appropriate quality factor
gives the dose in ‘roentgen equivalent man’ (Rem) where 1 Rem is the amount of
ionising radiation which has the same biological effect as 1 rad of x-rays.
In the SI system multiplying the dose in gray by the appropriate quality factor
gives the dose in ‘sievert’ (Sv) where 1 Sv is the amount of ionising radiation which
has the same biological effect as 1 Gy of x-rays.
Thus:
1 Sv = 100 Rem
or
1 Rem = 0.01 Sv
In the cgs system dose rate is generally measured in millirem per hour
(mRem/h) but may be measured in Rem per hour (Rem/h).
1 mRem/h = 10 Sv/h
or
The cgs unit of source strength or activity is the curie (Ci). One curie is equal to
3.7 x 1010 disintegrations per second.
1 Ci = 37 GBq
or
1 GBq = 0.027 Ci
The output of a source of ionising radiation is the dose rate per hour at some
fixed distance, usually 1 metre, from the source. For radioactive isotopes it is useful to
state output in grays, sieverts, rads or Rems per hour per curie at one metre. Table
below gives some examples.
OUTPUT
ISOTOPE NAME
(mSv per Ci)
Thulium 170 0.026
Ytterbium 169 1.25
Selenium 75 1.8
Iridium 192 4.8
Cobalt 60 13.0
The output of radiation from a typical 200 kV industrial constant potential x-ray
machine is as much as 1,000 mSv per milliampere of tube current at a distance of 1
metre from the focal spot.
Survey metres give a real time measurement of dose rate. They are of 5 basic
types, ionisation chambers, proportional counters, Geiger counters, scintillation
counters and solid state devices. Each of these is discussed and described in the
sections below.
Neither the Geiger counter below, nor the ionisation chamber above can give
any information as to the photon energy of the ionising radiation received. The best
that can be achieved with these instruments is to shield the chamber such that alpha
and beta radiation is excluded from the measurement. The gas chamber used in a
proportional counter often contains multiple electrodes. Proportional counters operate
in a voltage range intermediate between the ionisation chamber and the Geiger
Phosphors have been used as radiation detectors since the very early days of
the discovery of ionising radiation, both Roentgen and Becquerel used them. The
amount of light produced can be quite small so phosphors are always used in
conjunction with a light amplification system such as the photomultiplier tube. Modern
instruments use ‘charge coupled devices’ or CCDs in conjunction with a radiation
sensitive phosphor. A CCD is at the heart of any modern digital camera. The CCDs
used for radiation detection measure the intensity of light emitted from the
phosphorescent layer under the influence of ionising radiation. Whichever system is
used scintillation counters relate the intensity of light produced by the phosphor to the
intensity of the ionising radiation received. In general they give a reading in counts per
second but occasionally they will be calibrated to read directly in microsieverts or
millisieverts per hour.
Scintillation counters are extremely sensitive, they can detect very low levels of
ionising radiation. They are direction sensitive instruments and are very useful when
searching for radioactive contamination. They are used in industrial radiography to
check for leakage of fissile material from a sealed source.
Survey meters, with a few exceptions, give a real time reading of dose rate but
do not integrate this to give a total dose received over a given period of time. Several
types of device exist which are capable of integrating the dose received over a period
of time. One convenient use of such a device is for monitoring the total dose that a
person receives during the course of his or her working day. When used in this way
such devices are referred to as ‘personal monitors’. Four types of personal monitor
are commonly used in industrial radiography.
Film Badge
The principle of a film badge is that when exposed to ionising radiation followed
by developing under tightly controlled conditions the film density produced can be
related to the radiation dose received. Film badges (as shown in figure 91) can be
used to detect x, gamma and beta radiation. Coupled with the right type of
intensification screen radiographic film can be used to detect and measure other types
of ionising radiation.
The film badge of the type shown in the figure above contains a section of
carefully manufactured radiographic film having two emulsions, one fast and one slow.
The use of two emulsions extends the measurement range of the badge. The badge
holder is equipped with various filters that extend the range of measurement and
additionally enable the badge to give some information as to the type and photon
energy of the ionising radiation received.
The film badge has in large part, been replaced by the thermoluminescent
dosimeter (TLD) (see below). This table gives a comparison of typical film badge and
TLD specifications.
TLD badges are worn in specially designed plastic holders similar to those used
for film badges. The addition of plastic or aluminium filters extends the measurement
range of the badge facilitates the obtaining of information concerning the photon
energy and type of radiation.
These devices are still widely used in the USA where in many states they are
mandatory wear for all personnel involved in working with ionising radiation. In the UK
they used to be popular for use inside nuclear power plants but they have now largely
been replaced by more reliable, more accurate solid state devices.
The quartz fibre electrometer (QFE) is a gaseous detector like the ionisation
chamber, proportional counter and Geiger counter described above. When raised to
the light a scale like the one on the left of figure above can be seen through the lens of
the instrument. The vertical line is the quartz fibre. When a static electrical charge is
applied to the instrument the quartz fibre moves to the zero point of the scale. As the
gas inside the QFE becomes ionised the static charge is gradually dissipated and the
fibre begins to move to the right. The corresponding total dose received can be read
on the upper scale.
The QFE has quite a narrow measuring range, typically 0 to 50 mSv or less.
The example shown above has a measurement range of 0 to 200 mRem which is
equivalent to 0 to 2 mSv. The QFE is sensitive to x and gamma radiation in the photon
energy range 45 keV to 3.5 MeV. The QFE is a very convenient means for checking
how radiation doses are accumulating during a working day but it suffers from fragility
and is very easily damaged.
15.1 PRECAUTIONS
Such exposure booths should be regularly monitored to ensure that the radiation
dose rate is within safe limits in the areas outside the booth where personnel can
move freely.
Exposure booths should be equipped with switches inside the x-ray compound
which can be operated in order to prevent the operation of the x-ray or gamma ray
equipment should any personnel be accidentally trapped inside.
Areas which will be irradiated at greater than the legal limit must be cleared of
all non classified personnel prior to any exposure. Audible and visible warnings must
be given before any exposure takes place.
The barrier should be monitored with an efficient radiation detector and should
be guarded by classified personnel during exposure.
15.1.3 SCATTER
15.2.1 DOSIMETERS
For work in radioactive environments (i.e. nuclear reactors) personnel must be
equipped with direct reading dosimeters which will display immediately the
accumulated dose received. Personnel working in these locations must take particular
care to avoid ingesting radioactive particles. Tightly fitting breathing masks are
required and protective clothing should be worn.
The figures given in the paragraphs below relate to Statutory Instrument 1999
Number 3232, ‘Ionising Radiation Regulations 1999’. These regulations exclude
radiation doses received due to medical reasons.
The maximum permitted dose rate for personnel equipped with film badges (or
TLDs) is 20 mSv per year. This is approximately equivalent to a constant dose rate of
10 Sv/hr for a 40 hour working week if a 48 working week year is assumed.
Controlled area
Unclassified personnel must be excluded from any area where radiation dose is
deemed likely to exceed three tenths of the annual allowable dose for a classified
worker (6 mSv). The maximum permissible dose rate at the boundary of a controlled
area is 7.5 Sv/hr.
Supervised area
The dose rate from a source of ionising radiation reduces in proportion to the
reciprocal of the square of the distance from the source.
Thus:
If we take 7.5 Sv/hr to be the ‘safe’ dose rate then we can calculate the ‘safe’
distance using the formula below:
Where ‘C’ is a constant for each isotope. For Thulium 170 C = 1.86, for Ytterbium 169
C = 12.91, for Selenium 75 C = 15.49, for Iridium 192 C = 25.30 and for Cobalt 60 C =
41.63.
15.4.1 SHIELDING
Where:
Absorbed Dose
(of ionising radiation) The energy per unit mass imparted to the irradiated material.
Absorbed dose is measured in Grays [Gy] (1 Gy = J kg-1)
Absorption
The reduction in intensity of a beam of radiation during its passage through matter.
Absorption Coefficient
Alpha Radiation
A type of ionising radiation consisting of high velocity charged particles emitted from
the nucleus of heavy radioactive isotopes. The alpha particle consists of 2 protons
and 2 electrons – a helium nucleus, and has a positive charge of 2. Alpha radiation
has very low penetrating power, but it is very strongly ionising.
Anode
The positive electrode of a discharge tube. In an x-ray tube the anode carries the
target. (see ‘target’)
Atom
The total number of protons plus neutrons in the nucleus of an atom, standard
abbreviation A.
A measure of atomic weight, roughly speaking the proton has a weight of 1 amu while
that of the neutron is marginally greater than 1 amu. The weight of the electron is
roughly 0.00054 amu.
Atomic Number
Atomic weight
The weight of an atom expressed in atomic mass units, it is approximately equal to the
atomic mass number.
Back-Scatter
Scattered radiation caused by the presence of objects behind the film or radiation
detector. (see ‘scatter’)
Background Count
The ionising radiation dose rate due to natural causes (see ‘background radiation’).
Background Radiation
Ionising radiation which is present at any given site which is due to natural causes:
sunlight contains a proportion of ionising radiation; some natural rocks such as
granite are weakly radioactive.
Becquerel
Ionising radiation consisting of very high velocity electrons emitted from the nucleus of
a radioactive isotope. In beta emission a neutron converts to a proton while emitting a
very high speed electron. Beta radiation has low penetrating power but must not be
ignored when assessing radiation safety. Beta radiation causes severe skin burns
and can lead to fatality.
Betatron
An apparatus in which electrons are accelerated along a spiral path by means of the
electric force associated with a varying magnetic field.
The reciprocity law, which basically states that the film density produced by a dose of
ionising radiation is independent of the radiation dose rate (i.e. a low dose rate for a
long exposure time will produce the same film density as a high dose rate for a short
exposure time so long as [dose rate] x [time] remains constant).
Build-up
Calcium Tungstate
A complex salt of calcium which fluoresces in the blue part of the visible light spectrum
during exposure to ionising radiation. Calcium tungstate is the common base material
for salt or fluorometallic intensifying screens.
Cassette
A light proof container for holding an x-ray film during exposure. Cassettes may be
rigid or flexible. They must be designed so as to maintain good contact between the
x-ray film and intensifying screens.
Casting
The formation, by pouring molten base material into a mould, of a useful product
shape, or any component produced by such a process. The great majority of metallic
components begin life as a casting.
Characteristic Curve
The curve, for a given photographic film (or x-ray film) which relates the logarithm to
the base of 10 of the relative amount of radiation exposure (i.e. radiation intensity x
exposure time) to the achieved photographic density under specified processing
conditions. Also known as Hunter & Driffield curve, H&D curve or sensiometric curve.
The process by which a characteristic curve is produced is called sensiometry.
Compton Scattering
Constant Potential
Contrast:
The difference in brightness to the human eye of two adjacent areas in a radiographic
image.
Curie
The unit of radioactivity for any radioactive isotope. A radioactive isotope with a
source strength of 1 Curie is decaying at the rate of 3.7 x 1010 disintegrations per
second. (abbreviation: Ci). The SI unit of radioactivity is the Becquerel. (see
‘Becquerel’)
Curie-Hours or Curie-Minutes
gamma ray exposures are usually expressed in curie hours or curie minutes because
the intensity of radiation is proportional to the source strength and the total amount of
radiation received is proportional to the source strength multiplied by the exposure
time.
Definition
Densitometer
Density Strip
A strip of film exposed to form gradations of film density. Once calibrated using a
suitable densitometer density strips form a convenient means of comparing film
densities. Alternatively density strips of known density can be used to check the
calibration of densitometers.
Developer
The chemical solution used in the development of a radiographic film. All developers
are reducing agents which reduce the sensitised silver halide grains in the film
emulsion to metallic silver thereby producing an image on the film.
Development
The chemical process by which a latent image is converted to form a visible image.
(see ‘developer’).
Die Casting
A casting process for producing small to medium sized components, mainly applicable
to low melting point alloys, involving the use of a reusable mould generally
constructed from steel. In pressure die casting the molten charge is forced into the
mould under pressure. Die castings have a better surface finish and better
mechanical properties than equivalent sand castings.
Dose Rate
The total quantity of radiation energy per unit time. It usually is expressed in Sieverts
or Rems per unit time. (see ‘sievert’ and ‘Rem’)
Electromagnetic Radiation
Light waves, radio waves and ionising radiation (x or gamma rays) are all forms of
‘electromagnetic radiation’. All electromagnetic radiation travels at the same velocity
(299,274,000 metres per second in a vacuum), the different types of electromagnetic
radiation differ only in their wavelengths.
Electron
Very tiny, negatively charged fundamental particle. The mass of the electron is
roughly 0.00054 times that of a proton.
A unit of energy which is equal to the amount kinetic energy acquired by an electron
when it is accelerated through a potential difference of one volt. (abbreviated as eV 1
keV = 1,000 eV, 1 MeV = 1,000,000 eV). Electron volts are a convenient unit for
expressing ionising radiation energies.
Emulsion
Exposure Chart
A chart which relates the required radiographic exposure for a given radiographic film
and film density to the penetrated thickness of a specified material.
Filament
(in an x-ray tube) The heated cathode which usually consists of a thin tungsten wire
through which a heating current is passed in order to stimulate the thermionic
emission of electrons.
Film Badge
A piece of photographic film used as a radiation monitor. Film badges are usually
partially shielded so as to increase the effective measuring range.
Film Contrast
This is the degree to which a particular radiographic film when viewed by the human
eye can differentiate between two adjacent areas of different radiation exposure. Film
contrast is related to the change in film density per unit increase in radiation exposure.
This increases with the achieved film density.
Filter
Material (usually thin copper sheet) interposed in the path of radiation in order to
reduce selectively the intensity of radiation of a certain range of wavelengths or
energies (usually the lower range of energies). Filters are useful for reducing the
effect of scattered radiation in x-radiography.
Fixer
The chemical removal of unused silver halides from an emulsion after development.
(see ‘fixer’)
Flaw Sensitivity
The ability of a radiographic technique to detect flaws. This is not easy to quantify but
is expressed as the minimum detectable thickness of a specific flaw measured in the
direction of the radiation beam, expressed as a percentage of the total thickness of a
specimen of specified homogeneous material.
Fluorescence
Fluorescent Screen
Fluorometallic Screen
Focal Spot
The area of the target on which the electron stream impinges and from which X-rays
are emitted. The effective focal spot size in an x-ray tube is usually less than the
actual focal spot size.
Focus-to-film Distance
The distance from the focus of an x-ray tube to a film set up for radiographic
exposure. (abbreviation: FFD).
Gamma Radiation
Gray
Half-Life
A fixed property of any radioactive isotope. Half-life is the time taken for the number
of radioactive atoms in a given sample to reduce by half. It can be anything from a
few seconds to millions of years dependent on the isotope. The decay process itself
is random but the half-life is fixed because each radioactive atom has the same
probability of decay – averaged over a very large number of radioactive atoms this
gives rise to a fixed half-life.
Half-value Thickness
(half-value layer)
The thickness of a specified substance which when introduced into the path of a given
beam of radiation, reduces the radiation intensity by half. It may be used as an
indication of the quality of the beam or the opacity of the substance. (abbreviation:
HVL)
Identification Marker
Image Intensifier
Intensifying Factor
The ratio of the exposure time without intensifying screens to that when screens are
used, all other conditions being the same.
Intensifying Screen
A layer of suitable material, e.g. lead foil, which when placed in close contact with
photographic emulsion, adds to the photographic effect of the incident radiation.
Injection Moulding
Die casting process whereby molten raw material is forced into a mould under
pressure. The process of injection moulding is common for plastics and low melting
point metal alloys, especially those of zinc, magnesium and aluminium.
Investment Casting
Also called the lost wax process. In investment casting a wax model of the required
item is made and mould is formed around this using some type of refractory material.
After firing to harden the mould and burn out the wax molten metal is poured to
produce the desired component. Investment casting is an expensive method but it
produces the best surface finish and the best material properties of all of the casting
processes.
Isotopes
Nuclides having the same atomic number but different mass number (i.e. the nucleus
contains the same number of protons but a different number of neutrons). Some
isotopes are stable while others undergo nuclear fission thus producing emissions of
radiation.
Latitude
Macroradiography
Radiography of thin sections of material in such a way that the resulting image may be
enlarged to reveal microstructure.
Practical Radiography. Rev 2 Oct 2005 16.9 WORLD CENTRE FOR
A Glossary of Terms (Used in Radiographic Testing) MATERIALS JOINING
Metal active gas welding, often referred to as CO2 welding, mainly applicable to
carbon steel. MAG is an automatic or semi-automatic arc welding process involving a
reel fed consumable electrode; the arc is shielded by an active gas, usually carbon
dioxide. Porosity and lack of sidewall fusion are common defects in the MAG process.
MIG
Metal inert gas welding. MIG is an automatic or semi-automatic arc welding process
involving a reel fed consumable electrode; the arc is shielded by an inert gas, usually
argon. MIG welding is applicable to most metals and alloys. Porosity and lack of
sidewall fusion are common defects in the MIG process.
MMA
Manual metallic arc welding. An arc is struck between a flux coated consumable
electrode and the work piece. The flux coating decomposes to form a shielding gas,
usually carbon dioxide and a molten slag which protects the hot metal.
Neutron
The fundamental particle having a mass slightly greater than 1 amu and zero electrical
charge.
Neutron Radiography
Neutrons emitted by nuclear reactors and some radioactive isotopes are a form of
penetrating radiation and can be used to perform radiography. Neutrons are heavily
absorbed by substances such as water or plastic which contain significant amounts of
hydrogen. Neutrons pass easily through metals such as steel or aluminium. Neutron
radiography is useful for the detection of water ingress into aeroplane wing structures
and other similar applications.
Pair Production
The partial shadow extending beyond the edges of the main shadow (umbra) of an
object due to the finite size of the radiation source: the width of this partial shadow.
Photoelectric effect
Positron
Basically an electron but with opposite electrical charge. Positrons are emitted during
pair production which is an important scattering mechanism in high energy x-
radiography.
Pressure Mark
Primary Radiation
Radiation which is incident on the absorber and which continues unaltered in photon
energy and in direction after passing through the absorber.
Processing
A series of operations, such as developing, fixing and washing, associated with the
conversion of a latent image into a stable, visible image.
Quality Factor
In order to take account of the fact that for instance 1 Gy or 100 R of alpha radiation is
biologically much more damaging than 1 Gy or 100 R of x or gamma radiation a
quality factor is used. The dose in Sieverts or Rem is then equal to the dose in Grays
or Rads multiplied by the quality factor, 1 Sievert or 100 Rem of any type of ionising
radiation has the same biological effect.
Rad
The old unit of radiation absorbed dose. The SI equivalent is the Gray.
Radiograph
Radiographic Exposure
The subjection of an emulsion to radiation for the purpose of producing a latent image;
commonly expressed in milliampere-minutes or curie-hours.
Reciprocity Law
A law which states that, all other conditions remaining constant, the time of exposure
required to produce a given density is inversely proportional to the intensity of the
radiation. (see Bunsen-Roscoe Law)
Rem
Roentgen Equivalent Mammal: The old unit of man mammal equivalent absorbed
radiation dose. 1 Roentgen of alpha radiation has a much greater biological effect
than does 1 Roentgen of x or gamma rays whereas 1 Rem has the same biological
effect whatever the type of ionising radiation. Despite the name the Rem is arrived at
by multiplying the dose in rad (radiation absorbed dose) by a quality factor. The SI
equivalent of this unit is the Sievert.
Resolution
Reticulation
Rod-Anode Tube
A type of unipolar (grounded anode) X-ray tube in which the target is situated near the
end of a long tubular anode.
The old unit of exposure or ionising effect. The SI equivalent is the coulomb per
kilogram.
Salt Screen
Sand Casting
Casting process where a molten charge is poured into a mould formed from
compressed sand. Sand casting is the most versatile of all the casting processes, but
suffers from coarse grain structure and poor surface finish.
SAW
Scattering
The redirection of radiation, with or without a change in photon energy (but usually
with a reduction in photon energy), during its passage through matter.
Screen-Type Film
X-ray film designed for use with salt screens. It is sensitive to the fluorescent light
emitted by such screens under the action of x-rays.
Secondary Radiation
Sievert
Source-to-Film Distance
The distance from a source of radiation to a film set up for a radiographic exposure.
(abbreviation: SFD)
The amount of radioactive material per unit mass of a sample. Usually expressed in
Curies/gram.
Speed
TIG
Tungsten inert gas welding. A manual or fully automatic arc welding process. TIG is
extremely versatile but requires a high degree of operator skill. The heat source is an
arc struck under a shield of argon or helium gas between a non-consumable tungsten
electrode and the work piece. Filler wire may be fed into the arc, or in some
circumstances a weld may be produced without the need for filler wire.
NOTES
1) Where special circumstances apply – an employer is able to show that the annual limit of 20 mSv is
impractical, up to 50 mSv can be received in a single calendar year but not more than 100 mSv
over any five year period. Any employee exceeding these limits is likely to be suspended from work
pending an investigation by the HSE.
2) Where person in question – note 1 above, is a pregnant female the dose shall not exceed 13 mSv
in any period of 3 months.
3) Where the recorded dose exceeds 20 mSv in one year the employer is required to make a formal
investigation to determine whether the dose limitations of note 1 are likely to be complied with. The
employer must report the matter to the HSE and must put in place a program to ensure that the
dose limitations of note 1 are not exceeded.
4) For persons who act as carers to others who receiving exposure to ionising radiation for medical
purposes the dose limit is 5 mSv in any five year period.
7) ANNUAL DOSE The dose limitations, items 1 to 6 above are additional to
LIMITATIONS – exposure to radiation for medical purposes.
GENERAL.
Item Details Requirement(s)
8) RADIATION MONITORS Radiation monitors must be 1) RADIATION
checked before use to MONITORING
ensure correct functioning; EQUIPMENT SHALL BE
typically this would involve a PROPERLY
battery check and a check to MAINTAINED.
see that a reading is
produced when the 2) RADIATION
instrument is exposed to a MONITORING
source of ionising radiation. EQUIPMENT SHALL BE
Radiation monitors must FIT FOR THE
have a scale appropriate to DESIGNATED
the magnitude of the doses PURPOSE.
being measured. General
good practice is to have 3) RADIATION
portable monitors calibrated MONITORING
on an annual basis. EQUIPMENT SHALL BE
Appropriate calibration CALIBRATED AT
periods can vary dependent APPROPRIATE
on the type and usage of the INTERVALS.
radiation monitor in question.
TWI Training & Examination Services TWI Technology Centre (North East)
Granta Park Aurora Court
Barton Road
Chapter 1 Gt Abington Riverside Park
Middlesbrough
Cambridge TS2 1RY, UK
CB1 6AL, UK
Tel: +44 (0)1642 210512
Fax: +44 (0)1642 252218
Tel: +44 (0)1223 899000 Email: twinorth@twi.co.uk
Fax: +44 (0)1223 891630
Email: trainexam@twi.co.uk
TWI Training & Certification (SE Asia) TWI Training & Examination Services
No. 8 Jalan TSB 10 P O Box 52721
Sg Buloh Ind. Park Abu Dhabi
47000 Sg Buloh
Selangor Darul Ehsan
Malaysia
Tel. : 603-61573526/28 Tel: +971 2 6270750
Fax : 603-61572378 Fax: +971 2 6270424
EMail : inquiry@twisea.com Email: twi_uae@eim.ae