JHM-2028A Haemodialysis
Device
User’s Manual
Catalog
1 Safety guidelines 3
2 Brief introductions 6
2.1 Introductions on the HD device 6
2.2 Main technical specifications 8
3 Definition and terms 12
4 Structure and function of HD device 13
4.1 Overall structure of HD device 13
4.2 Components 18
4.2.1 Structure of blood pump/spare pump 18
4.2.2 Structure of syringe stand 19
4.3 Introduction on extracorporeal blood circulation 20
4.4 Introduction on the water way system 22
4.5 Introduction on hydrogen carbonate dialysis 27
4.5.1 Connection of concentrate tube 27
4.5.2 Supply method of the RO water 30
4.5.2.1 RO water tube supply 30
4.5.2.2 Container supply (selective device) 30
4.6 Description of combination for Na+ curve graph and UF curve graph 31
4.7 Introduction on the cleaning procedures 31
4.8 Menu description 34
4.8.1 Screen comments 34
4.8.2 Menu description 36
Select self-test mode 36
Self-test menu 37
Prime rinse menu 37
Dialysis menu 37
Curve display men 38
Set the selective menu 39
Dialysis setting menu 39
Advanced setting menu 40
Working mode setting menu 40
Time setting menu 40
HCO3- setting menu 43
Other setting menu 43
Software type of Shut off menu 43
Cleanness menu 44
Cleanness program menu(such as disinfection) 44
4.9 Using menu-parameter setting 44
-I-
4.9.1 Set flux of blood pump 45
4.9.2 Dialysis setting menu 45
Set dialysis time 45
Set heparin time 46
Set heparin flux 46
Set UF volume 46
+
Set Na basic value 47
+
Set Max. value of Na curve 47
+
Set Na conc. curve and UF curve 47
4.9.3 Advanced setting menu 48
Set V.P. span 48
Set A.P. span 49
Set TMP span 49
Set conductivity span 49
Set upper temp. alarm 50
Set lower temp. alarm 50
Set dialysate flux 50
Set pump dia. of blood pump(spare pump) 51
Alarm volume adjustment 51
Display the record stored by conductivity 51
Display the record stored by temperature 53
Set tube diameter of heparine pump 54
4.9.4 Work mode setting menu 55
Set work mode 55
Set Disinfect mode 56
Set of rinse mode 56
Set Hot Dis. mode 57
Set single needle treatment 58
Set ISO UF parameter 58
Pipe connection diagram in HD (ISO UF) mode 59
4.9.5 Time setting menu 60
Setting date 60
Setting time 60
Setting up limit 61
Setting low limit 61
Setting bypass 61
4.9.6 HCO3- setting menu 62
Setting HCO3- concentration 62
4.9.7 Kt/V setting menu 63
Set dialysis volume measure 63
Kt/V Calibration 63
-II-
URR 63
SRI2pool 64
nPCR 64
TACurea 65
4.9.8 Other setting menu 66
Setting dialysate temp. 66
Setting bridging flux 66
Setting flux of spare pump 66
Spare pump acts as blood pump 67
4.9.9 Shut off menu 67
Operate the software type of Shut off 67
Restart 68
4.9.10 Priming menu 68
Modify the range of alarm limit (i.e.: modify V.P. alarm limit) 68
4.10 Tone mute function 69
4.11 UF test 69
5 Treatment 70
5.1 Preparation work 70
5.1.1 Start to run hemodialysis equipment 70
5.1.1.1 A+B common dialysis mode 70
5.1.2 Self-test of function 71
5.1.3 The supply of the dialysate 73
5.1.4 The installation of the injector 74
5.1.5 The installation of the blood pipe system 75
5.1.6 Adjust the injection bracket 77
5.1.7 The installation of the adjust-liquor connection 77
5.1.8 The setting of parameter 78
5.1.9 Na+ curve graph and UF curve graph 79
5.1.10 The prime before the treatment 79
5.1.10.1 The preparation work for prime rinse 79
5.2 The operation of treatment 80
5.2.1 Dialysis treatment 80
5.2.2 Isolated UF treatment 84
6 Cleanness 86
6.1 Introduction 86
6.2 Exterior cleanness 87
6.3 The basic condition to undergo the process of interior cleanness 87
6.4 Interior cleanness 87
6.4.1 Pipeline connection during the disinfection, hot disinfection and rinse 87
6.4.2 Operation 89
6.4.2.1 Operation of disinfection 89
-III-
6.4.2.2 Operation of hot disinfection 91
6.4.2.3 Operation of rinse 94
7 Alarms show and the disposal of unusual thing 96
7.1 Alarms show 96
7.2 The alarm of blood loop 97
7.2.1 TMP alarm 97
7.2.2 A.P. alarm 97
7.2.3 V.P. alarm 98
7.2.4 Blood leakage alarm 98
7.2.5 Blood level alarm 99
7.2.6 Alarm of air bubble 99
7.2.7 Alarm of heparin 100
The alarm of heparin clog 100
Alarm of heparin empty 100
Alarm of heparin time up 100
7.3 The alarm of dialysate loop 101
7.3.1 The alarm of conductivity 101
7.3.2 Alarm of temp. 101
7.4 Power failure alarm 102
7.5 Battery alarm 102
7.6 Warming/notice 102
7.7 Disposal of unusual thing 103
7.7.1 The disposal principle for the unusual thing of hemodialysis device 103
7.7.2 The disposal of unusual things of hemodialysis device 104
7.7.2.1 The disposal of warning 104
Spot disposal 105
Maintenance disposal 111
7.7.2.2 The disposal of unusual things for the main parts 112
8 Maintenance and storage 113
9 Instruction for selective device 114
9.1 Instruction of blood-pressure meter(optional) 114
9.2 Instruction of back-up power supply 119
Appendix A 120
Appendix B 121
-IV-
JHM-2028 A Hemodialysis Device User’s Manual
This user’s manual is part of the documents of JHM-2028A Hemodialysis Device, whic h contains
some necessary information in the operation of JHM-2028A Hemodialysis Device.
Warning:
It w arns the operators not to conduct some operation. Otherwis e the
device may be damaged, or hazard may be generated toward the
operator or patient.
Note:
It reminds the operator to operate according to the instructions.
Otherwis e you may not get the intended result.
Equal potential
-1-
JHM-2028 A Hemodialysis Device User’s Manual
Please carefully read through this user’s manual. Please ref er to the actual device f or the real appearance of the device.
Fax: 0086-20-37020037
-2-
JHM-2028 A Hemodialysis Device User’s Manual
1 Safety guidelines
1.1 Requirements on the operator
1.2.1 The operator should carefully read through the user’s manual before the operation of the
1.2.2 The operator should verify the parameter that he or she inputs, that’s to say, the operator should
check whether the input parameter value is correct or not. If there is discrepancy betw een the
optimal parameter value and the displayed parameter value, the setting value should be
1.2.3 Make sure the power supply system (cir cuit, voltage and cable) meets the electricity load
1.2.4 JHM-2028A Hemodialysis Device is a Category I and Type b devic e, please plug the power cable
into a three core socket with grounding wire, making sure the grounding is appropriate.
1.2.5 During the dialysis period, if the protective system alarms, the operator should find out the alarm
reason timely and solv e the abnormality quickly before the dialysis continues. In case of alarms
that fail to resume normal in short time, or serious abnormality, the dialysis treatment should be
terminated.
1.2.6 It is recommended to select the syringe w ith same brand and same model when changing the
1.2.7 The exposed live parts should be earthed by means of the protective earth for the individual
safety and the prevention from indirect electric shock. The good protectiv e earth may ensure the
personal safety in the case of the electric leakage against the injury or death caused by the
electric shock.
1.2.8 No item unrelated to the devic e is allowed to be placed on the top of devic e for the sake of the
-3-
JHM-2028 A Hemodialysis Device User’s Manual
1.5 Only the manufacture or those personnel authoriz ed by the manufacturer can assemble, debug or
1.6 In case of malfunctions, you can try the normal troubleshooting methods, In case of failure, please
contact the company. Non-professional personnel are not allowed to dis assemble the device.
1.7 It is the responsibility of the users to comply with the regulations related to the waste fluid discharge
1.8 The surface of waste pipe should not be contacted with that of the wastewater treatment system or
the bacteria may enter the hemodialysis devic e along the waste pipe. The use of the waste pipe
1.9 The abnormal substance, whic h reaches the blood chamber of dialyzer from the dialysate chamber,
may result in the risk that the blood of patient is contaminated by bacteria and endotoxin.
1.10 The protection system w ith the measurement of venous pressure against the loss of blood in vitro
can only reduce in part the hazardousness which may remain such as the ris k condition of the
blood loss of the patient in vitro. The measures to minimize the risk: the blood pump stops and the
ultrafiltration is lowered to the min value when the preset value of alarm is exceeded.
1.11 Any narrow part of the external tube (such as the small blood tubing set or intubation) may lead to
the risk of hemolysis whic h might not be detected by the protection system.
1.12 The failure in the function of ultrasonic air detector may be resulted from the ultrasonic glue or clot.
1.13 The air may enter the external tube through the joint at the bottom of air detector; the occurrence of
such event may be caused by the use of single needle or central venous catheter in the case of
negative pressure.
1.14 When the pre-pump arterial pressure is low, the blood flow and the efficacy of treatment may be
compromised, thus the blood flow and the pressure range of pre-pump arterial pressure should be
maintained.
1.15 Contraindic ations: patient w ith the contraindic ations of hemodialysis such as shock or hypotension,
and severe hemorrhage trend; severe heart-brain complications such as cardiac enlargement
accompanied with cardiac hypofunction or cerebral hemorrhage, and severe arrhythmia, etc.
1.16 Only one type of potential equalization conductor can be used in the device.
1.17 There are certain interferences during the operation of the devic e, such as high-power motor,
solenoid valv e, induction heating equipment, high frequency welder, external radio, X-ray device,
high frequency physiotherapy device and ignition system of internal combustion engine, whic h may
-4-
JHM-2028 A Hemodialysis Device User’s Manual
affect the safe operation of the devic e. Do not place the dialy zer in the environment w ith
electromagnetic interference.
1.18 The operator should pay attention to the safety ris k related to the connection/dis connection with
patient; as well as the safety risk caused by the incorrect connection of external pipeline.
Note:
When the outdoor temperature is below 0℃, the device should
be laid aside for one hour before installation after moving it in from
-5-
JHM-2028 A Hemodialysis Device User’s Manual
2 Brief introductions
2.1 Introductions on the HD device
2.1.4 Application: JHM-2028A Hemodialysis Device is used for Hemodialysis on patients w ith acute or
chronic kidney failure.
2.1.5 Performance, structure and components of the product: the product consists of dialysate supply
system (dialysis pump, ultrafiltration pump, proportion preparation unit, liquid constant
temperature unit, transmembrane pressure monitoring and conductivity monitoring) and blood
circulation system (blood pump, heparin pump, arterial-venous pressure monitoring, air bubble
monitoring and blood leakage monitoring), without the dis posal sterile consumables used with
2.1.6 JHM-2028A Hemodialysis Device can be used in various medical centers and for clinical use.
When JHM-2028A Hemodialysis Devic e is conducting HD treatment, it runs and monitors the
dialysate cir culation and extracorporeal blood circulation.
2.1.7 All the operation menu and running status of JHM-2028A Hemodialysis Device is displayed on the
color LCD screen.
2.1.8 JHM-2028A Hemodialysis Device uses hydrogen carbonate and acetate for HD treatment.
2.1.9 JHM-2028A Hemodialysis Device With single needle dialysis function.
2.1.9 JHM-2028A Hemodialysis Device should be connected to the RO w ater treatment
2.1.10 In the circulation, after heating and air removing, the dialysate is conveyed to the dialyzer, whic h
keeps the flow-in fluid and flow-out fluid equal in volume. According to the selected UF rate, the
necessary UF volume for treatment can be achieved.
2.1.11 In the extracorporeal blood circulation, the blood goes through the dialy zer. The air inspector can
prevent the air entering. The blood leakage inspector and venous pressure inspector can prevent
the blood loss.
2.1.12 The mixing proportion of hydrogen carbonate concentrate and RO water is 1:1.225:32.775
(A:B:water). The manufacturer can customize the mixing proportion according to the user’s
demand.
+
2.1.13Based on the requirement of treatment, the concentration of Na can be adjusted in the range of
136mmol/L~145 mmol/L.
2.1.14 JHM-2028A Hemodialysis Devic e can conduct pure UF (ISO UF)/ single needle treatment/double
-6-
JHM-2028 A Hemodialysis Device User’s Manual
needle treatment .
2.1.15 JHM-2028A Hemodialysis Devic e can conduct rinse, hot disinfection, and dis infection procedures
to complete automatic cleaning and disinfection.
2.1.16 The flux of dialy sate can be adjusted within the range of 300 ml/min~800 ml/min.
2.1.17JHM-2028A Hemodialysis Device is equipped with a safety monitor system that ensures the safety
of patients. Extracorporeal blood-losing protection system meets with GB9706.2 “The part 2~16 for
Medical Electronic Device: safety requirement for hemodialysis , hemodiafiltration and
hemofiltration”.
-7-
JHM-2028 A Hemodialysis Device User’s Manual
Width: About 34 cm
z Weight: About 90 kg
z Electric safety
z Pow er supply
z Fuse: 2×10A(Ф6×30)
z Working condition
0572-2005.
Working environment: Temperature 10℃~38℃, with relative humidity not greater than
Temporary idle period: It is recommended to disinfect the device before the HD device
Stability: 10º
-8-
JHM-2028 A Hemodialysis Device User’s Manual
z Working procedure
Function test: Automatic test on the safety system after powering on.
Rinse the system lines: In “Rinse” status, the device automatically rinse the system lines.
Rinse mode includes the options of cold rinse and hot rinse
Hot disinfection mode is devided into “Heat + Citric acid” and “Heat”.
1:1.830:34.000
1:1.830:34.000
1:2.100:34.000
-9-
JHM-2028 A Hemodialysis Device User’s Manual
The concentration of NaCl contained in liquid B between the second ratio and the thir d ratio is
different.
z Cleaning procedure:
Hot disinfection: Time: 42 min for working mode “heating + citric acid”
+10%
Error: -5%
-300mmHg ~ 0mmHg)
Equipped w ith a spare pump, the technic data is the same with those whic h
-10-
JHM-2028 A Hemodialysis Device User’s Manual
Syringe size: 10 ml / 20 ml / 30 ml
The cycle of sound alarm is one per second, the intermittent silence
is 0.5 second.
z Blood way lines and dialysate are universal w ith Fresenius and Toray.
24V(2×12V)/ 7Ah
z Hemomanometer The value indicates the measured pressured value and pulses.
-11-
JHM-2028 A Hemodialysis Device User’s Manual
Air trap: Air device installed in the blood line system to separate the gas that is
irresolvable in the blood, including blood level inspector and air bubble
inspector.
Venous line clamp: The device that automatically clamps the blood line system.
Blood measurement line: Line that connects the blood way system and the pressure monitor device.
Rotation pin: Pump head of the blood pump and spare pump, w hic h is used to rollingly
-12-
JHM-2028 A Hemodialysis Device User’s Manual
HD Device ——Appearance
-13-
JHM-2028 A Hemodialysis Device User’s Manual
Illustration:
1. Indicator
Red blinks—Alarm
Yellow on—Ready
Green on—Running
2. LCD display
3. Blood pressure measurement
display screen (optional device,
refer to 9.1)
4. The adaptor of venous pressure
5. Spare pump
6. Sw itch of Spare pump
7. Venous pot with ultrasonic sensor
8. Sw itch of venous line clamp
9. Venous line clamp
10. Air bubble inspector (Optical
inspection)
11. A adapter group (red adapter)
12. Suction tube of A concentrate
13. Suction tube of B concentrate
14. Pow er supply indicator
15. Fluid stand
16. Switch of power supply
17. The adapter of arterial pressure
18. Blood pump
19. Switch of blood pump
20. Arterial pot plank
21. Heparin pump
22. B adapter group (blue adapter)
-14-
JHM-2028 A Hemodialysis Device User’s Manual
Illustration:
1. Ventilation fan
2. Pow er box
3. Pow er switch
4. Ventilation port
5. ①Sampling port 1 of UF fluid
6. ②Disinfection solution connector
7. ③Sampling port 2 of UF fluid
9. Alarm beeper
RF1-30
10. Fuse (2X10A) F10AL250V
-15-
JHM-2028 A Hemodialysis Device User’s Manual
Illustration:
1. Blood level adjustment button of
venous pot
Raise the height of blood
level of venous pot
Lower the height of blood
level of venous pot
2. Blood pressure measurement
port (select devic e)
-16-
JHM-2028 A Hemodialysis Device User’s Manual
Illustration:
1. Clamp of dialyzer
2. Supply tube of dialysate (blue
adapter)Filtration adapter
3. Box
4. Reflux tube of dialysate (red
adapter)
3 5. Bobber observation devic e
6. Filter
1
4
5
-17-
JHM-2028 A Hemodialysis Device User’s Manual
4.2 Components
Illustration:
1. Pump cover
2. Rotation pin
3. Pump case
4. Clamp
5. Pump tube fixation
slot
Pull out the spanner of the rotation pin, rotate the spanner until the handle is vertical against the
rotation pin fix ation slot, then the rotation pin is able to be released from the pump case
Illustration:
1. Rotation post (one on each side)
2. Guiding post
3. Spanner
4. Adjustment screw (one on each side)
5. Rotation pin fix ation slot
-18-
JHM-2028 A Hemodialysis Device User’s Manual
Illustration:
1. Stopper
1 2. Stopper adjustment button
3. Fixation slot
2
4. Syringe clamp
5. Pump holder
-19-
JHM-2028 A Hemodialysis Device User’s Manual
Connect the artery and vein of the patients w ith the blood pump, dialyzer and monitoring device, to
During the dialysis , the heparin pump injects heparin into the blood way on a regular basis to prevent
the blood from coagulation. The heparin flow rate can be adjusted according to the treatment
The patient blood was led out of the body by the blood pump and enters the venous pot after going
through the dialy zer. In the venous pot, the venous pressure is measured and strictly monitored w ithin the
preset range. The venous pot is located in the ultrasonic fluid level inspector to prevent the introduction of
air. Once the fluid level is lowered and bubble blood appears in the venous pot, the blood alarm is
triggered. The blood pump stops immediately, the venous line clamp automatically clamps the venous
blood way to block blood reflux, alarm w ith light and sound begins, and the UF ends.
After going through the venous pot, the blood flows through the air bubble inspector, whic h detects the
existence of air bubble in the blood circuit. Once the air bubble is found, the alarm w ith light and sound is
triggered. The blood pump and the UF pump stop immediately, the venous line clamp automatically clap
-20-
JHM-2028 A Hemodialysis Device User’s Manual
B2
B1 B3 B4
Artery End
B11
B14 B12
Connect
B16 B10 To
B6
B15 B13 Dialysate
B7 Circut
B8
Vein End B9
Illustration:
-21-
JHM-2028 A Hemodialysis Device User’s Manual
1. Supply by the RO water tube, the adapter (W1) connects to the RO w ater tube.
2. Supply by the container, the RO water container supplies the fluid, the adapter(W1) connects to
The RO w ater enters the water way through the adapter W1, and it’s heated in the heater system (W5)
after getting through W2 ,W3 and W4, meanwhile, air will be removed in W5 by the air-removing device
(W6).
The heat exchanger (W3) is to transfer the heat from the used dialy sate that runs out of the balance
After the air is removed, the RO w ater gets to the confecting system (W7), and the concentrate is
pumped up into the confecting system by the ceramic pump at the same time, and then they get mixed
into dialysate. The dialysate is driv en by the pre-dialysis pump (W8), and runs through the mix
temperature sensor (W9) and the mix conductivity sensor (W10) into the balance system (W11), then it
runs through another temperature sensor (W12) and conductiv ity sensor (W13), if the temperature or the
conductiv ity doesn’t meet the preset values, the pre-dialysis valve (W14) and the post-dialysis valve
(W22) w ill be closed and the bypass valv e (W23) will be opened to get the unqualif ied dialysate around
the dialyzer (W18). If the temperature and the conductiv ity meet the preset values, the bypass valve
(W23) w ill be closed, and the pre-dialysis valv e (W14) and the post-dialysis valv e (W22) will be opened
Note:When bic art cartridge is loaded, tube B should connect back to the connector Block of
Concentrate B.
The used dialysate that come out from the dialyzer runs through a pressure sensor (W21), the
pressure measured here can be regarded as the value for TMP calculation.
The used dialy sate runs through the air-removing device (W24) again in order to ensure the volume of
the fresh dialysate is exactly the same as the used dialysate that is replaced all along. Then it is driven
-22-
JHM-2028 A Hemodialysis Device User’s Manual
by the post-dialysis pump (W28), and runs through the balance system (W11), the heat exchanger (W3)
and the blood leakage inspector (W35). The blood leakage inspector is to monitor if there is blood
infiltrating the dialysate cir cuit due to the damage of dialysis membrane. After that, the used dialysate
runs through evacuation valv e (W36) and gets out of the water way from outlet of evacuation (W37).
The UF pump (W34) draws out the dialysate from the fluid infusion system at the preset rate. As the
system is airtight, the ultrafiltrate can be drawn out at the same volume from the blood through the
dialyzer.
-23-
W 27
Waterway sketch map
W 26
W 31
W 25
W 32
W 24
W 30
W4 2 W29 Ar ti al
W 34
W 40
W 48
W2 2 W2 1
W20 W19 W 49
W4 1
W2 8
W 38 W 18
W 23
W9 W1 2 W 14
W8 W 15 W17
W1 0 W 13 W 16
W7
W1 1 W50
Connect blood loop
-24-
W 51 W 53 W52 Ve no us
W 43 W 39
W 33
W 44 W 46
W6 W 35 W3 6
W5
W4 5 W 47
W 37
W3
W2 W1
B l o o d wa y
Water way
JHM-2028 A Hemodialysis Device User’s Manual
JHM-2028 A Hemodialysis Device User’s Manual
Artial
Venous
W52
W53
W48
W51
W49
W50
Connect blood loop
W18
W17
W19
W16
W37
W1
W33
W32
W20
W15
W36
W22 W21
W14
W2
W35
W30
W34
W31
W13
W23
W12
W29
W3
W26 W25
W11
W27
W24
W28
W10
W9
W6
W8
W7
W5
W38
W39
W40
W47
W46
Blood way
Water way
W43
W44
W42
W45
W41
-25-
JHM-2028 A Hemodialysis Device User’s Manual
Illustration:
W17. Fast adapter sleeve of dialy zer inlet (blue) W44. Concentrate A adapter
W19. Fast adapter sleeve of dialy zer inlet (red) W46. Connector Block of Concentrate A
W24. Fluid air-removing devic e W51. Venous pot with ultrasonic sensor
-26-
JHM-2028 A Hemodialysis Device User’s Manual
The concentrate is sucked by two pumps, and is conveyed to the HD devic e to mix w ith the RO water.
The suction tube of concentrate is connected to the container accordingly.
To minimize misusing of concentration, concentration container should have special mark.
To avoid the misconnection with container, sucking pump should als o have special mark.
Warning:
You must use the container with specif ic mark and matched suction
tube, to avoid wrong connection.
The mixing process is based on volume and proportion. That is , the concentrate pump mix certain
volume of concentrate into the water of fix ed volume.
Only by using the concentrate matching the mixing system, you can get a correct ion concentration in
the dialysate.
By measurement on the conductiv ity and the bypass function, the system can prevent the dialysate
with wrong concentration from entering the dialyzer.
In case of existence of wrong components in the dialysate, the conductiv ity will exceed the alarm range
and triggers the alarm. The HD device then opens the bypass, accompanied w ith light and sound alarm.
The condensate of dialy sate is required to comply with the standard YY 0598-2006.
The components of dialy sate include: sodium chloride, potassium chloride, calcium chloride,
The dialytic powder should be diluted as per the ratio specif ied in the instruction. The sample of
dialysate is collected from the device and the resultant ionic concentration of dialysate should be
The sodium (Na) is 97.5%-102.5% of labeled amount;acetic acid (or acetate) is 90%-110% of
labeled amount;
Calcium (Ca), kalium (K), Magnesium (Mg), Chlorine (Cl) and bicarbonate (HCO3)should be
-27-
JHM-2028 A Hemodialysis Device User’s Manual
Specific measurement method: When the device is running in Hemodialysis mode, collect the
sample of dialy sate as the devic e become stable. The test is conduced according to the description in
Item 6.5.1~6.5.4 on Page 4 of YY0598-2006 and the result should be consistent w ith the requirements
described above.
Warning:
The hydrogen carbonate concentrate and acetate concentrate cannot
be stored again once opened.
Warning:
Please make sure the container has sufficient hydrogen carbonate
concentrate and acetate concentrate to complete the treatment.
The single fault condition defined based on the alarm limit of protection system:
Upper limit alarm of mixed conductivity: the error among each component of dialy sate is ±0.5%;
low er limit alarm of mixed conductivity: the error among each component of dialy sate is ±0.5%.
Upper limit of outflow conductiv ity: the error among each component of dialysate is ±0.5%; lower
limit alarm of outflow conductivity: the error among each component of dialysate is ±0.5%.
-28-
JHM-2028 A Hemodialysis Device User’s Manual
-29-
JHM-2028 A Hemodialysis Device User’s Manual
Only the RO water can be used when confecting the dialy sate.
The requirement of the RO water's quality meets with the standard YY0572.
Connect to
dialysate
container
-30-
JHM-2028 A Hemodialysis Device User’s Manual
4.6 Description of combination for Na+ curve graph and UF curve graph
JHM-2028A HD devic e possesses the function of combining Na+ curve graph with UF curve graph.
Na+ curve graph: 0 ~ 7.
UF curve graph: 0 ~ 7.
The below diagram will show the possible combination of Na+ curve graph and UF curve graph:
+
Na curv e graph
0 1 2 3 4 5 6 7
UF 1
curv e
graph 2
4.7.3 Disinfection
4.7.3.1 Hot disinfection
In the specif ic period of disinfection, the surface temperature of dialy sate pipeline in the device
(1) Run the chemical disinfection program, connect the pipe of Liquid B to the dis infectant cartridge
following the prompt of the device, and then start the time measurement whic h will be ended upon the
completion of disinfection, the contact time of dis infectant should be calculated on the basis of the
following formula;
The sample is collected on the supply port at 10 minutes after the chemical infection program is run,
the content of citric acid is measured with high-performance liquid chromatography (refer to appendix B)
and the content of peracetic acid is determined w ith quantitative test paper, the result should be
Follow ing the dis infection, the sampling is done at the supply port, and the content of citric acid is
measured w ith high-performance liquid chromatography (refer to appendix B) and the content of
peracetic acid is determined with quantitativ e test paper, the result should be consistent with the
(2) Run the hot dis infection program, connect the pipe of Liquid B to the disinfectant cartridge
following the prompt of the device, and then start the time measurement whic h will be ended upon the
completion of disinfection, the contact time of dis infectant should be calculated on the basis of the
following formula;
The sample is collected on the supply port at 10 minutes after the hot infection program is run, the
content of citric acid is measured with high-performance liquid chromatography (refer to appendix B) and
the content of peracetic acid is determined w ith quantitative test paper, the result should be consis tent
Follow ing the dis infection, the sampling is done at the supply port, and the content of citric acid is
measured w ith high-performance liquid chromatography (refer to appendix B) and the content of
peracetic acid is determined with quantitativ e test paper, the result should be consistent with the
-33-
JHM-2028 A Hemodialysis Device User’s Manual
2 3 4 5
Prime
Stopped
Dialysis Setting KEY
<- ->
UF Curve 0 Na+ Cur ve 0
Return OK
8 7
z Screen description:
1. Menu area
2. Current mode
4. Mute button
-34-
JHM-2028 A Hemodialysis Device User’s Manual
2 3 4 5
Dialysis Low er Conductivity Alam
Stopped Blood Level Alarm
TMP. A.P. V.P. Battery Blood Leak Blood Level Air Bubble Hepar in Clogging Temp. Cond.
6
-100
00000 ml 13.5
100 50
Blood pump Spare Pump 13
0 -200 0
11 12
z Screen description:
1. Menu area
2. Current mode
4. Mute button
9. Curve button (simultaneously display Na+ curve graph and UF curve graph)
10. Running key (blood pump, heparin pump, UF pump, spare pump and bypass)
11. Function key (run key, bridging key, return key, setting key)
12.Time area
-35-
JHM-2028 A Hemodialysis Device User’s Manual
Note:
Location of operator: 10-20 cm directly away from LCD.
JHM-2028A
t reatm ent .
5: Place the suct ion tubes of liquid A& B into concentrate container respect ively prior to
running sel f-test m ode 1. 1.Blue (running time)
6: Please turn to repairer if self -t est failed.
2.White (remained time)
Automatically enter into self -test mode 1 if not
Self-Test 1 Self-Test 2 Skip select self-test mode within the waiting time.
1 2
-36-
JHM-2028 A Hemodialysis Device User’s Manual
z Self-test menu
Lower Temp alarm
Self-test Lower Conductivity alarm Self test.
Blood Level Normal Spare Pump Normal
Test Gear
pump Tes t In-Pro.
The below color marks the status for every
test:
Air B ubble Test C ond.
Normal Heparin Pump Normal sy stem Test In-Pro.
test status color
V.P. Normal W at er Work
preparation
Test Temp.
Test In-pro. syst em Tes t In-Pro. test in-process yellow
normal green
A.P. Normal Tes t Balance
valve
Test in-pro
failed red
TMP Normal Test UF
pump Test in-pro The info.of current testing displays in text area.
Blood Pump Normal Test A/B
Test in-pro
Screen w ill enter into priming menu
pump
treatment.
UF Rate Hepari n Time Time Left Temp.
6 00 40 0 3 00
0010 ml/h 3:30 3:59 23.0 ℃ 1 5.5
Use ▲or▼ to modify the alarm range of
5 00 30 0 2 50
UF Volume Hepari n Flu x Na+ HCO3-
4 00 20 0 2 00
0000 ml 10.0 ml/h 138 mmol /l 032 mmol /l
15
TMP/AP/VP/Cond., modify the flux of blood
1 4.5
- 10 0 -3 00 -50
200 ml/min 200 ml/min 1 2.5
disinfection and setting menu.
mmHg mmHg mmHg mS/cm
Blood p. Hepar in UF P. Spar e P. Bypass Use menu frame to select dialysis, dis infect,
04.17
Pr ime Treatment Disinfect Setting MON
15:03 and set menu and start.
Note: For the prime, press “Bypass” button to
enter “Solution-saving prime mode”.
z Dialysis menu
6 00 40 0 3 00
UF Rate Hepari n Time Time Left Temp. It display the parameters during treatment,
0010 ml/h 3:30 3:59 23.0 ℃
curve key display the preset Na+ cond. curve and
1 5.5
5 00 30 0 2 50
UF Volume Hepari n Flu x Na+ HCO3 - 15
4 00 20 0 2 00
0000 ml 10.0 ml/h 138 mmol /l 032 mmol /l
3 00 100 1 50
1 4.5 UF volume curve by graph
UF Goal 14
2 00 0
-1 00
1 00
0000 ml 1 3.5 Use ▲or▼ to modify the alarm limit of
10 0 50
Blood pump Spare Pump
TMP/AP/VP/Cond. modify flux of blood ump and
13
0 -2 00 0
- 10 0 -3 00 -50
200 ml/min 200 ml/min 1 2.5
04.17
Pr ime Treatment Disinfect Setting 15:03
MON
-37-
JHM-2028 A Hemodialysis Device User’s Manual
3000
TMP - 100 m mHg
2000
Na + 140 m mol/l If dialysia treatment time is more than 5 hours,
1000
Co nd . 00.0 m S/cm
machine w ill not go into curve display manual
Te mp . 00.0 ℃
0
through curve key-press of menu.
0 1 2 3 4 5 6
T ime (h )
Exit
Pri me
Stopped Press the key of “Na+ profile” and it will display
U F pr of ile Na + p r ofile Kt/V HCO 3-
Na + p r ofile
Time L e ft 4:00
menu of the Na+ curve.
Na+ pro file m ea ns u se d fi fer ent sodiu m c once ntr ation au tom atically
UF Ra te 2000 ml/h
150
TMP -100 mmHg
140
Na + 140 mmol/l
Co nd . 00.0 mS/cm
130
Temp . 00.0 ℃
120
0 1 2 3 4 5 6
T im e (h ) Exit
Prim e
St opped Press the key of “Kt/V” and it w ill display menu
UF p ro file Na + pr o file Kt/ V HCO 3 -
4:00
If the resu lt af ter dialysis is >1 .2, it me ans dialysis ade qua cy.
Time Le ft
of the Kt/V curve.
UF Rat e 2000 ml/h
K t/V
3.0 UF Go al 0000 ml
0.0
2.5
UF Volu me 0000 ml
1.0
C on d. 00.0 mS /cm
0.5
Te mp. 00.0 ℃
0.0
0 1 2 3 4 5 6
T ime (h ) Exit
-38-
JHM-2028 A Hemodialysis Device User’s Manual
HCO 3- p ro file
Time L eft 4:00
menu of HCO3- curve.
U F Ra te 2000 ml/h
40
V.P. 0003 mm Hg
TMP -100 mm Hg
35 N a+ 140 mm ol/l
C on d. 00.0 mS /cm
30
Te mp. 00.0 ℃
28
0 1 2 3 4 5 6
Tim e (h )
Exit
-39-
JHM-2028 A Hemodialysis Device User’s Manual
——Work mode;
A : B : H2O Worke Mode Disinfect Mode Hot Dis. Mode
1:1.225:32.775 HD ——ISO UF setting;
Citric acid Heat+Citri c acid
1:1.830:34.000 ISO UF
1:1.830:34.000 Peraceti c acid Heat ——Conn. Draw.
1:2.100:34.000 Rinse Mode The working mode includes HD and ISO UF.
Acetate
Cold rinse
The disinfection mode includes citric acid
Na:1 38 K:2.0 Ca:1.7 5 Mg:0.5
Hot rinse
Ac:3.0 HCO3:32 B)NaCl:0g/l and peracetic acid.
Blood Mode
The rinse mode includes clod rinse and hot
A+B Mode
A+B
(Needle)
Double
rinse.
Single Return OK The hot dis infection mode includes “heating
+ citric acid” and “heating”.
-40-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Cond. Compare menu.
Time Setting KE Y
The scope of both the monitor upper limit and
Time Set C ond comp Temp c omp Other Comp Result
0 1
16.5 16.5 16.5 16.0
the monitor lower limit is 0.5 ~2mS/cm. The
Pre.Cond. Mi x Cond.
16.0
20.6
16.0 16.0
00.1
15.0 2 3 default value in the systerm is 1.0mS/cm.
15.5 15.5 15.5 14.0
15.0
Post Cond. 15.0 15.0
Cond. A 13.0
4 5 According to the operator’s setting, system w ill
00.1 00.1
14.5
14.0
14.5
14.0
14.5
14.0
12.0
11.0
6 7 automatically process the pre-dialysis and
Up Limit (A:B:H2O)
13.5
01.0
13.5 13.5
1 :1.2 25: 32.7 75
10.0
8 9 post-dialysis conuctivity compare. If compare
13.0 13.0 13.0 1 :1.2 60: 32.7 40 9.0
12.5
Low Limit
12.5 12.5 1 :1.8 30: 34.0 00
1 :2.1 00: 34.0 00
8.0
<- -> fails for 3 times continousely, it w ill display alarm
12.0 12.0 12.0 7.0
01.0
11.5
mS/cm
11.5
m S/ cm
11.5
mS/cm
6.0
mS/cm
Return OK message:”Cond of pre-dia and post-dia different,
please check.” If one of the 3 times compares
success, it won’t alarm and won’t choose
bypass. The compare will process during the
dialysis process.circularly.
The process information of this compare w ill
display in “Comp. Result” menu.
P rime
Stopped Temp comp. menu.
Tim e Setti ng KEY
System w ill process pre-dialysis temparature
Ti me Set Cond c omp Temp c omp Other Com p Re sult
0 1 and post-dialy sis temp.compare automatically,
41.0 41.0 41.0 41.0
Pre Temp Mix T em p
40.0 40.0 40.0 40.0 2 3 and display the information of pre-dialysis temp.
16.4 16.9
39.0 39.0 39.0 39.0
-41-
JHM-2028 A Hemodialysis Device User’s Manual
Pr ime
S topped Other menu.
Time Setting KEY
This menu displays the temperature of heating
Tim e Set C on d co m p Te m p co mp Other Comp Resu lt
0 1
chamber, goal of heating chamber, average of
Pum p A Data
Heat Ch. Temp 023. 0 ℃
ml/10 0Round 01198 ml
2 3 outlet Temp., goal Cond. of A , blood leakage
G oa l o f
h ea t Ch . 021. 9 ℃
Pum p B Data
ml/10 0Round 01192 ml 4 5 feedback, communication ok, Pump A data,
Dia. Fluid
Ave ra g e o f
O u tle t Te m p.
004. 0 ℃
( ml/9 Cavit y)
00709 ml 6 7 pump B data, dialysis fluid feedback, UF Pump
G oa l C on d. o f A
( mS/c m) 000. 0 UF pum p
( 3.5 ms /Round)
00000 8 9 feedback, UF P Reference and communication
Bloo d .le ak 00035 UF P. Ref 22634300 <- -> failed.
Co m m OK 003066 Comm Faild 000013
Return OK
Prime
S topped
Comp. Result menu
Tim e Setting K EY This menu displays process information during
Tim e S et C ond comp Temp c omp Other Co mp Re sult
0 1 the compare in “cond. Compare menu”
Chec k Phase 000 Pre & Post C ond.
comp. bypass or not
Forbid
OK
2 3 “Check phase” means whih times of compare
Comp. time 000 S 4 5 it is processing;
Conduc tivit y
Dif. time
000 S 6 7 “Comp. Time” means how long it will take for
P-c ond. Chec k
D if. times 000 8 9 the compare;
Comp. Result
of Pre. & Post
Not Compare <- -> ”Conductiv ity Dif. Time” means the time
Waterw ay
Bypas s St atus
Normal
Return OK beyond the error scope in this compare result;
”P-cond.Check Dif. Times: means the times
that compare result beyond the error scope
during process of the compare;
”Comp. Result of Pre.&Post.” Means the result
of this compare;
”Waterway Bypass Status” : “Normal” means in
the status of waterway; “abnormal” means in the
status of bypass.
“Pre&Post Cond. Compa. Bypass or not”:
“Forbid” means won’t carry out bypass even if
compare failed; “OK” means carrying out bypass
when compare failed.
-42-
JHM-2028 A Hemodialysis Device User’s Manual
-43-
JHM-2028 A Hemodialysis Device User’s Manual
z Cleanness menu
Dis infection
S topped List the information for cleanness description.
TMP. A.P. V. P. Battery Blood Leak B lood Level Air Bubble Heparin C logging Temp. C ond.
6 00 400 300
liqu id( B)
Co nn ec t d ialysa te su pply /re tu rn tub es of dia lyze r to th e por t o f mac hin e
function key to select and start cleanness mode
1 5. 5
500 300 25 0
40 0 20 0 20 0
Rins e t ime is 1 0 min s
Ple as e d on 't sto p it d ur ing ho t d isinfe ctio n o r disin fec tion pr oc ess
15 needed.
1 4. 5
300 1 00 1 50 The d isinfe ctio n solu tion u sed in disin fec tion an d ho t dis infe ctio n is
con fec ted by th e d efin ed fo rmula 14
200 0 1 00 N ot a llow ed to re tur n to oth er mod es if th e d isinf ect o
i n or ho t d isinf ectio n
h as no t b ee n co mple ted 1 3. 5
1 00 -1 00 50
Th e mac hin e mus t g o th ro ug h d isin fec tion or h ot d isin fec tion pr oc ess 13
0 -2 00 0 a fre r eve ry tre at me nt eve ry time
1 2. 5
-1 00 -3 00 -5 0
mmHg mmHg mmHg mS/cm
04.17
Ri nse Disinfect Hot Di s. Return MON 15:03
04.17
Rinse Disinfect Hot Dis. Return MO N 15:03
1 2
-1 00
0000 ml 1 3.5
10 0 50
-2 00
Blood pump Spare Pump 13
0 0
- 10 0 -3 00 -50
200 ml/min 200 ml/min 1 2.5
-44-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Press the fig. (flux of blood pump) which need
Setting Select K EY modifying, the fig. Color shows black background
0 1 and white font.
Dialysis Setting Advanced Setting
2 3 Press again the fig. in keyboard, input the fig.
Work Mode Time Setting
4 5 of flux demanded.
HCO3- Kt/V 6 7 Selective range for flux of blood pump:
Pipe dia.ф6mm,20ml/min~460ml/min;
Other Setting Shut Off 8 9
Pipe dia.ф8mm,20ml/min~620ml/min.
Common S etting <- ->
BP Vol ume
Blood Pump 210 ml /min 000.0 L
Return OK
Prime
Stopped Set selection menu.
Setting Select K EY Press “Dialysis Setting” key, the screen w ill
0 1
Dialysis Setting Advanced Setting display dialysis setting menu.
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Common S etting <- ->
BP Vol ume
Blood Pump 210 ml /min 000.0 L
Return OK
P rime
Stopped The default dialysis time is 4:00.
Dialys is Setting KEY
Press the fig. (Dialysis Time) w hich need
Dialysi s Time 4 : 00 Heparin Time 3 : 30 0 1 modifying, the fig. color displays black
2 3 background and white font.
UF Goal 0000 ml Heparin Fl ux 03.0 ml/h
4 5 Re-press the fig. in keyboard, input the
UF Max 0000 ml/h Base Na+ 140 mmol/l 6 7 demanded time.
8 9 Selective range for dialysis time:
UF Volume 0000 ml Na+ Max 150 mmol/l
-45-
JHM-2028 A Hemodialysis Device User’s Manual
<- ->
UF Curve 0 Na+ Curve 0
Return OK
z Set UF volume
P rime
Stopped Press the fig. (UF Goal) w hic h needs
Dialys is Setting KEY
modifying, the fig. color displays black
4 : 00 3 : 30 0 1
Dialysi s Time Heparin Time
background and white font.
2 3 Re-press the fig. in keyboard, input the
UF Goal 0000 ml Heparin Fl ux 03.0 ml/h
4 5 demanded UF volume.
UF Max 0000 ml/h Base Na+ 140 mmol/l 6 7 Selective range for UF goal:
8 9 0~16000ml.
UF Volume 0000 ml Na+ Max 150 mmol/l
<- ->
UF Curve 0 Na+ Curve 0
Return OK
-46-
JHM-2028 A Hemodialysis Device User’s Manual
-47-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Set selection menu.
Setting Select KEY Press “Advanced Setting” key, the screen w ill
0 1 display dialysis setting menu.
Dialysis Setting Advanced Setting
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Common Setting <- ->
BP Vol ume
Blood Pump 210 ml /min 000.0 L
Return OK
Prime
Stopped Passport login menu.
Password Logon KEY
Input the preset passport in the keyboard,
0 1
Passwor d: press “OK” key, and screen display senior set
2 3
menu.
4 5
Code: 3721
6 7
8 9
<- ->
Return OK
-48-
JHM-2028 A Hemodialysis Device User’s Manual
-49-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Press the fig. (Upper temp.) whic h need
Advanced Setting KEY
modifying, the fig. color displays black
Adv. set Cond. save Temp. save Reserve Tube of Hep
0 1
mmHg
background and white font.
V.P. Span 200 Upper T emp. 38.5 ℃
2 3
Re-press the fig. in keyboard, input the
A.P. Span 200 mmHg Lower Temp 34.5 ℃
4 5
demanded upper temp. value.
Blood Pump 8mm 6 7 Selective range for upper temp. alarm:
TMP. Span 400 mmHg
Pipe dia. 6mm
8 9 37.5℃~40℃.
Spare Pump 8mm
Cond. Span 1.0 mS/cm
Pipe dia.
6mm <- ->
Di alysis Fl ux 500 ml/min Alarm Volume
Return OK
-50-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Press “○” beside the set fig. of pipe dia., it
Advanced Setting KEY
displays“⊙”, whic h means it has been set.
Adv. set Cond. save Temp. save Reserve Tube of Hep
0 1
mmHg
The selective range of pipe dia. for blood
V.P. Span 200 Upper T emp. 38.5 ℃
2 3
pump spare pump: ф6mm
A.P. Span 200 mmHg Lower Temp 34.5 ℃
4 5
ф8mm
Blood Pump 8mm 6 7
TMP. Span 400 mmHg
Pipe dia. 6mm
8 9
Spare Pump 8mm
Cond. Span 1.0 mS/cm
Pipe dia.
6mm <- ->
Di alysis Fl ux 500 ml/min Alarm Volume
Return OK
Prime
Stopped This menu can display the profile records
Advanced Setting KEY
information of pre-dialy sis conductiv ity,
Adv. set Cond. save T emp. save Reserve Tube of Hep
0 1
1 6.0
Pr e. post-dialysis conductivity, mixture conductiv ity
1 5.5 Co nd .
2 3
1 5.0
and Na+.
Bac k 4 5
1 4.5 Co nd . White: Pre-dialysis conductivity;
1 4.0
M ix.
6 7 Dark blue: Port-dialysis conductivity;
1 3.5 Co nd .
1 3.0
8 9 Green: Mixture conductiv ity;
1 2.5 Na +
<- -> Light blue: Na+ concentration;
1 2.0
( mS/cm ) 1 2 3 4 5 6 ( Hour )
Date
Dat e: 2008. 12.01 09:30
Pre. record Next record
Tim es: 0001
Reserve Reserve
Return OK Conductivity Scope: 12.0~16.0 mS/cm;
Time scope: 0~6 hour.
Press “Pre. record” or “Next record” can
display last or next conductiv ity profile records.
The menu w ill be displayed as above if “Data”
is pressed.
-52-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped This menu can display the profile records
Advanced Setting KEY
Adv. set Cond. save Tem p. save Reserve Tube of Hep
information of pre-dialysis temperature,
0 1
3 9.5
Pre . post-dialysis temperature, mixture temperature
3 9.0 Tem p.
2 3
3 8.5
and temperature A.
3 8.0
Back 4 5
3 7.5 Tem p. White: Pre-dialysis temperature;
3 7.0
6 7
3 6.5
3 6.0
M ix.
Tem p.
Dark blue: Port-dialysis temperature;
3 5.5 8 9 Green: Mixture temperature;
3 5.0
3 4.5
A
Tem p. <- -> Light blue: Temperature A;
3 4.0
1 2 3 4 5 6 ( Hou r)
℃
Date
Date: 2008.12. 01 09: 30
Pre. record Next record
Ti mes: 0001
Reserve Reserve
Return OK Temperature Scope: 34.0~39.5 ℃;
Time scope: 0~6 hour.
Press “Pre. record” or “Next record” can
display last or next temperature profile records.
The menu w ill be displayed as above if “Data”
is pressed.
-53-
JHM-2028 A Hemodialysis Device User’s Manual
Ch oose au to.
2 3 manual and choose automatically.
4 5 Press “○” beside the set fig. of pipe dia., it
10ml
Choose heparin
tube diameter
by manual
20m l 6 7 displays“⊙”, whic h means it has been set.
30ml
8 9 Heparin pump tube diameter size has 10ml,
<- -> 20ml and 30ml.
Return OK
Return OK
Prime
Stopped Set selection menu.
Setting Select KEY Press “Work Mode” key, screen displays work
0 1 mode set menu.
Dialysis Setting Advanced Setting
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Common Setting <- ->
Blood Pum p 210 ml /m in BP Vol ume
000.0 L
Return OK
-54-
JHM-2028 A Hemodialysis Device User’s Manual
1:2.100:34.000
Rinse Mode
changed on this menu.
Acetate
Cold rinse When the work mode , disinfect mode, hot dis .
Na:1 38 K:2.0 Ca:1.7 5 Mg:0.5 Hot rinse
Ac:3.0 HCO3:32 B)NaCl:0g/l Mode or bloodway mode is changed, the menu
A+B Mo de
Blood Mode will display clue on relativ e confirmation or
(Needle)
A+B Double
operation error as followed.
Single Return OK
To confirm the modific ation, click "OK" button,
otherwis e click "Cancel" button.
Prime
Stopped Press “ISO UF ‘O’ ”, it displays “⊙”, whic h
Work Mode Setting
Wor k Mo de ISO UF Set Con n. Dra w Re serve d Rese rved Reser ved Re serve d means “ISO UF” mode has been selected.
A : B : H2 O Wor ke M ode Disinfect Mode Hot Dis. Mode
1:1 .225:32.775 HD
Citric acid Heat+Ci tric acid
1 :1.830:34.000 ISO UF
Peraceti c acid Heat
1:1.830:3 4.00 0
1:2.100:3 4.00 0
Rinse Mode
Acetate
Cold rinse
Na:1 38 K:2 .0 Ca:1.75 M g:0.5
Hot rinse
A c:3.0 HCO3:32 B )NaCl: 0g/l
Blood Mod e
A+B Mode (Need le)
A+B Do uble
Sing le R eturn OK
-55-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Confirmation on work mode modification.
Work Mode Setting
Work Mode ISO UF Set Conn. Draw Reserved Reserved Reserved Reserved Click “OK” to confirm modicafic ation,
A : B : H2O Work Mode Disinfect Mode Hot Dis. Mode
otherwis e click "Cancel" button.
1:1.225 :32.775 HD Citric acid Heat+Citric acid
1:1.260 :32.740 ISO UF Confi rm
1:1.830 :34.000 Are you sure to ch ange 'Work Mode'?
1:2.100 :34.000
? Hemodial ysis mode (HD)
Change to b e:
Ace tate
Na:138 K:2.0 Ca:1.75 Mg:0 .5 ISO Utrafil tration(ISO UF)
Ac:3.0 HCO3:32 B)NaCl:0g /l OK Cancel
Blood Mod e
A+B Mode (Nee dle)
A+B Doub le
Sing e
l Return OK
Prime
Stopped Press “Peracetic acid ‘O’ ”, it displays “⊙”,
Work Mode Setting
Wo rk M ode I SO UF Set Con n. Dr aw Reser ved Reser ved Rese rved Reser ved whic h means “Peracetic acid” mode has been
A : B : H2O Wor ke M ode Disinf ect Mode Hot Dis. Mode selected.
1: 1. 225: 32.7 75 HD
Cit ric acid Heat+Citric acid
1:1 .83 0:34 .000 ISO UF
Peracetic acid Heat
1 :1. 830: 34. 000
1 :2. 100: 34. 000
Rinse M ode
Ace tat e
Col d rinse
Na:138 K:2.0 Ca:1 .75 Mg: 0.5
Hot rinse
Ac:3.0 HCO3:3 2 B) Na Cl:0g /l
Blood Mod e
A+ B M ode (Need le)
A+ B Do uble
Sing le Retu rn OK
Pr ime
Stopped Confirmation on disinfect mode modification.
Work Mode Setting
Work Mode I SO UF Set Conn. Draw Reserved Reserved Reserved Reserved
Click “OK” to confirm modicafic ation,
A : B : H2O W ork M ode Disinfect Mode Hot Di s. Mode otherwis e click "Cancel" button.
1: 1.225 :32. 775 HD
Citric aci d Heat+Citric aci d
1: 1.830 :34. 000 ISO UF Confirm
1: 1.830 :34. 000 Are you su re to change 'Disi nfect Mode'?
1: 2.100 :34. 000
Acet ate
? Citri c acid
Change to be:
Peracetic acid
Na:13 8 K:2. 0 Ca :1.75 Mg:0. 5
OK Can cel
Ac:3.0 HCO3 :32 B) Na Cl:0g/l
Bloo d Mo de
A+B M ode (Ne edle)
A+B Doub le
Single Re turn OK
-56-
JHM-2028 A Hemodialysis Device User’s Manual
Blo od M ode
A+B Mode ( Need e
l )
A+B Dou ble
Single Return OK
Prime
Stopped Confirm again the information to change
Work Mode Setting
Work Mode ISO UF Set Conn. Draw Reserved Reserved Reserved Reserved
the rinse mode:
A : B : H 2O Work Mode Disinfect Mode Hot Di s. Mode To confirm the change, press “Confirm”
1:1.225 :32 .77 5 HD
1:1.830 :34 .00 0 ISO UF C onfirm
Citric acid Heat+Citric acid
button; the device is switched to hot rinse
1:1.830 :34 .00 0 Are yo u sure to cha nge 'rinse Mode'?
1:2.100 :34 .00 0
mode follow ing the confirmation; otherwise
Acetate
? 'col d ri nse' Change tobe:'hot ri nse'
Blo od M od e
A+B Mo de ( Need le)
A+B Dou ble
Sing e
l Return OK
Prime
Stopped Press“Heat ‘O’ ”, it displays “⊙”, whic h
Work Mode Setting
Wor k Mo de ISO UF Set Con n. Dra w Re serve d Rese rved Reser ved Re serve d means “Heat” mode has been selected.
A : B : H2 O Wor ke M ode Disinfect Mode Hot Dis. Mode
1:1 .225:32.775 HD
Citric acid Heat+Ci tric acid
1 :1.830:34.000 ISO UF
Peraceti c acid Heat
1:1.830:3 4.00 0
1:2.100:3 4.00 0
Rinse Mode
Acetate
Cold rinse
Na:1 38 K:2 .0 Ca:1.75 M g:0.5
Hot rinse
Ac:3.0 HCO3:32 B)NaCl: 0g/l
Blood Mod e
A+B Mode (Need le)
A+B Do uble
Sing le R eturn OK
-57-
JHM-2028 A Hemodialysis Device User’s Manual
Pr ime
Stopped Confirmation on hot dis. mode modification.
Work Mode Setting
Work M ode ISO UF Set Conn. Draw Reserved Reserved Reserved Reserved
Click “OK” to confirm modicafic ation,
A : B : H2O Work Mo de Di sinfect Mode Hot Dis. Mode otherwis e click "Cancel" button.
1:1.2 25:32.7 75 HD
Citri c acid Heat+Citri c acid
1:1.8 30:34.0 00 ISO UF Confirm
1:1.8 30:34.0 00 Are you su re to ch ange 'Hot Dis. Mode'?
1:2.1 00:34.0 00
? Heat+Ci tric acid
Chan ge to be:
Acetate
Heat
Na:138 K:2.0 Ca:1.75 M g:0.5
OK Ca ncel
Ac:3. 0 HCO 3:32 B )NaCl:0g/l
Bloo d M ode
A+B Mode (Ne edle )
A+B Dou ble
Single Re turn OK
Prime
Stopped
Select “Single ‘O’ ”, it displays “⊙”, that means
Work Mode Setting
Wor k Mo de Re serve d
single needle mode has been selected.
ISO UF Set Conn . Draw Re serve d Re serve d Rese rved
Bloo d M ode
A+B Mo de ( Ne edle )
A+ B Dou ble
Single Return OK
Prime
Stopped Confirmation on bloodway mode modification.
Work Mode S etting
Work Mode IS O UF Set Conn. Draw Res erv ed Res erv ed Res erv ed Reserved Click “OK” to confirm modicafic ation,
A : B : H 2O Work Mode Dis infec t M ode Hot Di s. Mode
otherwise click "Cancel" button.
1 :1.225:32.77 5 HD
Cit ric ac id Heat+Citric ac id
1 :1.830:34.00 0 IS O UF Con firm
1 :1.830:34.00 0 Are you sure to chan ge 'Blood Mode'?
1 :2.100:34.00 0
? Double Needle Trea tment
Change to be:
Ace tate
Na:138 K :2.0 Ca:1.75 M g:0.5
Single Needle Trea tment
OK Can cel
Ac:3.0 HCO3:32 B )NaCl:0g/l
B loo d Mo de
A+B Mode (Nee dle)
A +B Doub e
l
S ingle Return OK
-58-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Click "ISO UF setting" button, ISO UF time,
Work Mode Setting
Work Mode ISO UF Set Conn. Draw Reserved Reserved Reserved Reserved
double needle ISO UF volume and single
needle ISO UF volume can be set in this menu.
0 1
Optional range of ISO UF: 0 ~ 9:59;
ISO UF Time 0 : 30 2 3
Optional range of double needle ISO UF: 0 ~
Double Needle
0000 ml 4 5
ISO UF Goal
8000ml;
Sin gle Nee dle
6 7
ISO UF Goal 0000 ml
Optional range of single needle ISO UF: 0 ~
8 9
ISO UF Volume 0000 ml 2000ml.
Software ed iti on is: V9.68 EN Return OK Press the data needed to be modified and the
data will be displayed in white and with black
background, then press the number in the
keyboard for input. After setting, click "OK"
button to confirm.
Prime
Stopped
When the set volume of the double needle
Work Mode Setting
ISO UF exceeds Max. residue volume, the
Work Mode ISO UF Set Conn. Draw Reserved Reserved Reserved Reserved
8 9
ISO UF Volume
0000 ml
Software edi tion is: V9.68 EN
Return OK
Reserved
Di alyzer
Spare P.
Hot Disfect.
Air Bubble
Reserved
-59-
JHM-2028 A Hemodialysis Device User’s Manual
P rim e
S topped The Left diagram is the pipe diagram in rinse
W or k M ode S etting
Work M ode ISO UF Set Conn. Draw Reserved Reserved Reserved Reserved
and hot dis infection
A Port B Port suppl y port return port
To machine To m achine To machine T o machine HD(ISO UF)
Reserved
Reserved
Hot disfect.
Reserved
Prime
Stopped Set selection menu.
Setting Select K EY Press “Time Setting” key, screen time setting
0 1
Dialysis Setting Advanced Setting menu.
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Comm on S etting <- ->
Blood Pum p 210 ml /m in BP Vol ume
000.0 L
Return OK
z Setting date
Prim e
Stopped Press the date which need to revis e,the color
Time Setting KEY
of it dis play black background.and white font.
T ime Set Co n d c om p T em p c om p Other Co mp Result
0 1
Then press the figure of the keyboard,and
Year M on th Dat e
2 3
input the needed date.
20 08 01 01
4 5
Hour M i nut e D ay
6 7
00 03 7
8 9
Ac cunulativ e Running T im e
<- ->
0000042 Hou r 16 m inu te
R etu rn OK
-60-
JHM-2028 A Hemodialysis Device User’s Manual
z Setting time
Prime
Stopped Press the time and week which need to revis e,
Time Setting KEY
the color of it display black background, and
T ime Set Cond c omp Temp c omp Other Comp Result
0 1
white font.
Year Mon th Date
2 3
20 08 01 01 Then press the figure of the keyboard, and
4 5
Hour Mi nute D ay input the needed time and week.
6 7
00 03 7
8 9
Ac cunulativ e Running T ime
<- ->
0000042 Hou r 16 minu te
R etu rn OK
z Setting up limit
Prime
Stopped Press the date of Up Limit needed to be
Time Setting KEY
modified and input necessary value.
T ime Set Cond c omp Temp comp O ther Comp Result
0 1
16.5 16.5 16.5 16.0
Scope of Up Limit: 0.5~2mS/cm.
Pre.Cond. Mix Cond.
16.0
20.6
16.0 16.0
00.1
15.0 2 3
15.5 15.5 15.5 14.0
Up Limit (A:B:H2O)
13.5
01.0
13.5 13.5
1: 1.22 5:32 .77 5
10.0
8 9
13.0 13.0 13.0 1: 1.26 0:32 .74 0 9.0
12.5
Low Li mit
12.5 12.5
1: 1.83 0:34 .00 0
1: 2.10 0:34 .00 0
8.0 <- ->
12.0 12.0 12.0 7.0
01.0
11.5 11.5 11.5 6.0
m S/ cm m S/cm mS/cm m S/c m Return OK
-61-
JHM-2028 A Hemodialysis Device User’s Manual
P rime
Stopped Press the date of Low Limit needed to be
Time Setting KEY
modified and input necessary value.
Ti me Set Co nd c om p Tem p co m p Other Com p Re sult
0 1
16.5 16.5 16.5 16.0
Scope of Low Limit: 0.5~2mS/cm.
Pre.Cond. Mix Cond.
16.0
20.6 16.0 16.0
00.1 15.0 2 3
15.5 15.5 15.5 14.0
Up Lim it (A:B:H2O)
13.5
01.0
13.5 13.5
1:1 .22 5:32 .77 5
10.0
8 9
13.0 13.0 13.0 1:1 .26 0:32 .74 0 9.0
12.5
Low Lim it
12.5 12.5 1:1 .83 0:34 .00 0
1:2 .10 0:34 .00 0
8.0
<- ->
12.0 12.0 12.0 7.0
01.0
11.5 11.5 11.5 6.0
m S/c m mS/cm m S/cm m S/c m Return OK
z Setting bypass
Prime
S topped
Press “OK” means it will go bypass when
Time Setting K EY comparis on between pre-dialysis conductiv ity
Time Set C on d co m p T em p c om p Other Co mp Re sult
0 1 and post-dialy sis conductiv ity is failing.
F or bid
Ch ec k Ph a se 000 Pre & Po st C on d .
co m p. b yp ass o r no t OK
2 3
Co m p. tim e 000 S
4 5
Co nd uc tivit y
Dif. tim e 000 S
6 7
P-c on d. Ch ec k
D if. tim es 000 8 9
Co m p. Re su lt
of Pr e. & Po st
Not Co mp are <- ->
Wa te rw ay
Nor ma l
Byp as s St atu s
Return OK
Prime
Stopped Set selection menu.
Setting Select K EY Press “HCO3-” key, screen displays HCO3-
0 1
Dialysis Setting Advanced Setting set menu.
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Comm on S etting <- ->
Blood Pum p 210 ml /m in BP Vol ume
000.0 L
Return OK
-62-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Set selection menu.
Setting Select K EY Press “Kt/V” key, screen displays Kt/V set
0 1 menu.
Dialysis Setting Advanced Setting
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Comm on S etting <- ->
BP Vol ume
Blood Pump 210 ml /min 000.0 L
Return OK
Kt /V UR R SP1 2 po o l nP CR TACu re a
Pa tie n t d a ta and weight can be set through keyboard, and
He igh t W e igh t
Ca libr a tio n sh ou ld b e d o ne w he n Kt/ V t es ts ina cc ur a tel y.
In ca lib ra tio n , m a ch ine s ho u ld be in p r im e Kt/V function
Ban 170 cm 060 k g sex can be selected.
st at us . Pr e co nd itio n o n su cce ss fu l ca lib ra tio n : Pr e -d ia lysis OK
M a le
co n du ct ivity a nd p o st- dia ly sis co nd u ctiv ity sh ou ld b e in th e ra ng e o f G en d er
Fe ma le
13 .0 -1 5 .0 m S/c m , t he e rr o r be tw ee n th em sh ou ld b e les s t ha n 1 .0 m S/c m .
k ey bo a rd
Be g in Ca libr a tio n 0%
1 2 3
4 5 6
Pre-dialysis 00.0 m S/ cm 7 8 9
m S/ cm
0 OK
Pos t-dialysis 00.0
Ex it
-63-
JHM-2028 A Hemodialysis Device User’s Manual
z Kt/V Calibration
Prim e
S topped Press key of “Kt/V” and it will display Kt/V
dialysis volume measure
Kt /V U RR SP 12 po ol n PCR TACu re a
Pa tien t d a ta
He igh t We igh t
calibration menu.
Ca libr at ion sh ou ld b e do ne wh en Kt/ V te sts ina ccu r ate ly.
In ca lib ra tion , ma chin e sh ou ld b e in pr ime Kt/ V fu ncti on
Ban
OK
170 cm 060 k g
Press“○”in front of “OK” key to start Kt/V
s tat us . Pr eco nd itio n on su cce ssfu l c alib ra tion : Pr e- d ialys is Ma le
c on du ctiv ity a nd p os t-d ialy sis c on du ctiv ity s ho uld b e in th e ra ng e o f
1 3.0 -1 5. 0 mS/ cm, th e er ro r be twe en th em s ho uld be le ss tha n 1.0 mS/cm.
Ge nd er
Fe ma le function. If it dis plays “⊙”, it means Kt/V function
ke yb oa rd
is already started.
Be gin Ca libr at ion 0% 1 2 3
This menu displays pre-dialy sis and
4 5 6
P re-d ialysi s 00.0 mS/cm 7 8 9 post-dialysis value.
P ost-di alysi s 00.0 mS/cm
0 OK In priming condition, press “start calibration”
Exi t
key to process Kt/V calibration.
The precondition of calibration successful is
showed as this menu.
z URR
Prime
St opped Press the key of “URR” and URR menu w ill be
dialysi s volum e m easure
z SRI2pool
-64-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
St opped Press“SRI2pool” key and it w ill display the
dialysis volume measure
z nPCR
P rime
S topped Press the key of “nPCR”, it w ill display the
dialysis volum e measure
z TACurea
-65-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
St opped Press the key of “TACurea”, it w ill
dialysis volume measure
Prime
Stopped Set selection menu.
Setting Select KEY Press “Other Setting” key, screen displays
0 1 other setting menu.
Dialysis Setting Advanced Setting
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Common Setting <- ->
B lood Pum p 210 ml /m in BP Vol ume
000.0 L
Return OK
-67-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Press “○” for yes , it displays “⊙”,whic h
Other Setting KEY
means that spare pump acts as blood pump.
Dialysis Temp. 36.5 ℃ Spare P. Fl ux 200 ml/ min
0 1
Prime
Stopped Set selection menu.
Setting Select KEY Press “Shut Off” key, screen dis plays shut off
0 1
Dialysis Setting Advanced Setting menu.
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Common Setting <- ->
B lood Pum p 210 ml /m in BP Vol ume
000.0 L
Return OK
z Restart
-68-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Press “Restart” key, machine can restart
Setting Select KEY automatically.
0 1
Dialy sis Setting Advanced Setting
2 3
Shut off system g
4 5
Shut Off Restart Cancel
6 7
Other Setting Shut Off 8 9
Common Setting <- ->
B lood P ump 210 m l/m in B P V olum e 000.0 L
Return OK
Prime
Stopped Set selection menu.
Setting Select KEY Press “Return” key, screen displays priming
0 1 menu.
Dialysis Setting Advanced Setting
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Common Setting <- ->
B lood Pum p 210 ml /m in BP Vol ume
000.0 L
Return OK
- 10 0
0000 m l 13 .5
1 00 50
- 20 0
Blood pump Spare Pump 13
0 0
-10 0 - 30 0 - 50
200 m l/ min 200 ml/ mi n 12 .5
-69-
JHM-2028 A Hemodialysis Device User’s Manual
40 0 20 0 2 00
Confirm 15 “Cancel” not to start prime as shown in left
It will shif t to Prime, But syst em det ect s that: 14 .5
30 0 10 0 15 0 1.Conductivity i s incorrect .
figure.
20 0 0 10 0
? 2. Temperature is i ncorrect .
T herefore, It f orbi ds to pum p blood and treat.
14
13 .5
1 00 -1 00 50
OK Cancel 13
0 -2 00 0
-1 00 -3 00 - 50
200 m l/m in 200 m l/ m in 12 .5
mm Hg mm Hg m mHg mS / cm
alarm voice.
The alarm info. won’t be displayed in text area
until the malfunction is resolved .
4.11 UF test
-70-
JHM-2028 A Hemodialysis Device User’s Manual
5 Treatment
5.1 Preparation work
Note:
Please make sure the follow ing preparation work has been done
before using the HD device.
1) The connection betw een the HD device and the RO water supply
system has been done.
2) The connection between the HD device and the drainage system
has been done.
3) The connection betw een the HD device and the power supply
system has been done.
When the hemodialysis machine is running, don't run the other device
whic h will transmit electromagnetic wave or radio wave as this kind of
device will bring work malfunction to hemodialysis machine.
-71-
JHM-2028 A Hemodialysis Device User’s Manual
t reatment.
5:P lace the s uc tion t ubes of liquid A &B into c onc entrate c ont ainer respec tiv el y prior to
running self -t est mode 1.
-72-
JHM-2028 A Hemodialysis Device User’s Manual
Note:
In the procedure of test-self , if the system inspects that system didn’t
undergo the procedure of disinfection or it was not finis hed last time, the
system will run this procedure automatically, until it is finished.
- 10 0
0000 m l 13 .5
1 00 50
- 20 0
Blood pump Spare Pump 13
0 0
-10 0 - 30 0 - 50
200 m l/ min 200 ml/ mi n 12 .5
-73-
JHM-2028 A Hemodialysis Device User’s Manual
40 0 2 00 2 00
Confirm 15
When the conductivity or temperature is
will close the whole wat er way, 14 .5
30 0 10 0 15 0 i t wi ll t ake long t ime to resume incorrect, it is not allow ed to perform the blood
0 10 0 ? 14
drawing or treatment.
20 0
13 .5
- 10 0
10 0 50
OK Cancel 13
0 - 20 0 0
-1 00 - 30 0 - 50
200 m l/ mi n 200 ml/ m in 12 .5
mm Hg m mHg m mH g m S/ cm
Note:
If the system didn’t pass the function self-test, please turn off the machine
and re-press “power” button for re-test.
Warning:
Do undergo the disinfection before the first treatment.
Put the connector of A concentrate water tube (red) into the A concentration water container.
Put the connector of B concentrate water tube (blue) into the B concentration water container.
-74-
JHM-2028 A Hemodialysis Device User’s Manual
Note :
Wings of injection should be in the fixed slot.
-75-
JHM-2028 A Hemodialysis Device User’s Manual
Look the 4.2.1 to have a look at the name of parts f or ref erence.
-76-
JHM-2028 A Hemodialysis Device User’s Manual
Note:
1. Make sure that the pump pipe can be fully pressed by the two ball
bearings.
2. After the pump pipe is set up, take on the cover of pump, otherwise
the rotor won’t move.
Warning:
All the connection between the blood pipelines should be
operated in asepsis.
The blood pipe should not twisted when the pipes are put into the
system.
-77-
JHM-2028 A Hemodialysis Device User’s Manual
Note:
All the connection between the blood pipelines should be operated in
asepsis .
The hydronium in the dialysate is the main factor to decide the penetrativ e pressure of the dialysate,
whic h effects directly the concentration of the blood serum of the patient and the penetrative pressure. In
the process of the haemodialysis, adjust the concentration of the hydronium of the dialy sate according to
-78-
JHM-2028 A Hemodialysis Device User’s Manual
Note:
During the treatment, the set parameters can be modified.
-79-
JHM-2028 A Hemodialysis Device User’s Manual
The below chart shows the possible combination of Na+ curve graph and UF curve graph:
+
Na curv e graph
0 1 2 3 4 5 6 7
1
UF
curv e 2
graph
3
(3)Connect the blood pipe on the both ends of dialyzer to the bag of physiological saline.
(4)Remove the air in the blood pipe.
-80-
JHM-2028 A Hemodialysis Device User’s Manual
set temperature ±0.5℃ of dialysate and the conductivity within the target conductivity ±0.3Sm/cm prior to
- 10 0
0000 m l 13 .5
1 00 50
- 20 0
Blood pump Spare Pump 13
0 0
-10 0 - 30 0 - 50
200 m l/ min 200 ml/ mi n 12 .5
20 0 200
UF Volume He pari n Flu x Na+ HCO3- 15 clamp opens and blood begins to circulate.
40 0
0000 ml 10.0 m l/ h 138 mm ol/ l 032 m mol/ l 1 4. 5
30 0 1 00 1 50
UF Goal 14
In bridging, the upper and lower limits of
20 0 0 1 00
0000 ml 1 3. 5
transmembrane pressure, arterial pressure and
100 -1 00 50
Blood pump Spare Pump 13
-2 00
venous pressure reach the maximal value.
0 0
-1 00 -3 00 -50
200 ml/ m in 200 m l/m in 1 2. 5
mm Hg mm Hg m m Hg mS / cm
-81-
JHM-2028 A Hemodialysis Device User’s Manual
Note:
The system only monitors air bubble and liquid level in bridging
state.
Note:
Under the mode of dialysis and blood returning, entering into bridging
status can avoid system monitor, but need the supervis ion by people.
- 100
0000 ml 1 3.5
10 0 50
Blood pump Spare Pump 13
0 - 200 0
- 100 - 300 - 50
200 ml/min 200 ml/mi n 1 2.5
04.17
Run Bridging Return Setting 15:03
M ON
-82-
JHM-2028 A Hemodialysis Device User’s Manual
Do ub le N eed le
When the set dialysis time counts down to
D ia lysis Tim eU p
Di aly sis Ru nn ing
0:00, system will raise alarm voic e.
Information displays in the text area.
1 00 - 10 0 50
0000 ml 13 .5 button “Blood pump” at this point to re-transfuse
Blood pump Spare Pump 13
0 - 20 0 0
200 ml/min 200 ml/mi n 12 .5
the blood to the patient.
-10 0 - 30 0 - 50
mmHg mmHg mmHg mS/cm
Blood p. Evacuate UF P. Spare P. Bypass
04.17
Run Bld Rtn Return Setting 15:03
MON
Blood R et urn
Running
Low er Conductiv ity Alarm
Blood Lev el alarm
When the air bubble is detected by the
TMP. A. P. V .P. Battery Blood Leak Blood Lev el Air Bubble H eparin Clogging Temp. C ond.
system, it w ill prompt if the blood return
60 0 40 0 30 0
UF Rate Hep arin Time Time Left Temp. continues. Press “Confirm” to continue the blood
0010 ml/h 3:30 0:00 23.0 ℃ 1 5.5
5 00 30 0 25 0
Confirm 15
return while the system will not detect the air
4 00 20 0 20 0
1 4.5
bubble; press “Cancel” to stop the return.
3 00
2 00
100
0
150
100
? Air bubbl e alarm! Continue to return the blood? 14
1 3.5
10 0 -1 00 50 OK Cancel
13
0 -2 00 0
- 100 -3 00 -5 0
200 ml/min 200 ml/min 1 2.5
-83-
JHM-2028 A Hemodialysis Device User’s Manual
Blood Return
Runnig
Lower C onduct ivity Alarm
B lood Level alarm
When the returning of blood is completed,
TMP. A.P. V.P. Batt ery Blood Leak Blood Level A ir B ubble Heparin C logging Temp. Cond.
the arrangement of dialyzer is changed to the
He parin Time Time Left Temp.
UF Rate one that the top end acts as the supply connector
6 00 4 00 30 0
0010 ml/h 3:30 0:00 23.0 ℃ 1 5. 5
500 3 00 2 50
UF Volume He parin Flu x Na+ HCO3- 15
whic h is connected back to the device while the
400 2 00 2 00
0000 ml 10.0 ml/h 138 mmol/l 032 mmol/l
300 10 0 15 0
1 4. 5
low er end as the return connector.
UF Goal 14
200 0 10 0
1 00 - 10 0 50
0000 ml 1 3. 5 Press “Drain” button to drain the liquid in the
Blood pump Spare Pump 13
0 - 20 0 0
200 ml/mi n 200 ml /min 1 2. 5
dialyzer.
-10 0 - 30 0 - 50
mmHg mmHg mmHg mS/cm
Blood p. Evacuate UF P. Spare P. Bypass
04.17
Run Bld Rtn Retur n Setting MON 15:03
Warning:
The “Drain” button can only be used after the separation of body and device
to drain the water in the dialyzer.
Warning:
When return the blood, take care of supervis ing not to put air into
patient’s body.
Note:
In the bypass mode, dialysis is ineffic acy.
-84-
JHM-2028 A Hemodialysis Device User’s Manual
Prime
Stopped Menu of working mode setting.
Work Mode Setting
Work Mode I SO UF Set Conn. Draw Reserved Reserved Reserved Reserved
Select ISO UF, the system w ill enter
A : B : H2O Worke Mode Disinfect Mode Hot Dis. Mode automatically “ISO UF setting”.
1:1.225:32.775 HD
Citric acid Heat+Citric acid
1:1.830:34.000 ISO UF
Peracetic acid Heat
1:1.830:34.000
1:2.100:34.000
Acetate
Blood Mode
A+B Mode (Needle)
A+B Double
Single Return OK
Prime
Stopped
Set volume and time of ISO UF (refer to the
Work Mode Setting
menu of ISO UF setting in section 4.10.4).
Work Mode ISO UF Set Conn. Draw Reserved Reserved Reserved Reserved
SingleNeedle
6 7
ISO UF Goal 0000 ml
8 9
ISO UF Volume 0000 ml
Softwareeditionis: V9.68EN Return OK
Prime
Stopped Press the button “Return” to show the
Setting Select KEY Prime menu.
0 1
Dialysis Setting Advanced Setting
2 3
Work Mode Time Setting
4 5
HCO3- Kt/V 6 7
Other Setting Shut Off 8 9
Common Setting <- ->
BP Vol ume
Blood Pump 210 ml /min 000.0 L
Return OK
-85-
JHM-2028 A Hemodialysis Device User’s Manual
-1 00
0000 ml 1 3.5
10 0 50
-2 00
Blood pump Spare Pump 13
0 0
- 10 0 -3 00 -50
200 ml/min 200 ml/min 1 2.5
Note:
In ISO UF, the conductivity and temperature are not monitored while
the indicator remains green.
-86-
JHM-2028 A Hemodialysis Device User’s Manual
6 Cleanness
6.1 Introduction
Exterior cleanness is the cleanness of the outer part of the hemodialysis device after treatment.
Interior cleanness includes hot dis infection, dis infection and rinse, that is the cleanness of the
The process of rinse: the system w ill rinse the waterway of the hemodialysis device.
The process of hot disinfection: the system will remove the calc ium and rinse with hot water to the
The process of the disinfection: the system w ill perform the dis infection and iron-removing to the
Warning:
The system should be undergoing the interior cleanness after the
dialysis treatment.
Note:
If the system is left unused for more than 72 hours, it is proposed to
under go the interior cleanness before use.
Note:
If the hemodialysis devic e is left unused as long as 5 minutes after the
process of the cleanness, the system w ill be power-off automatically.
Note:
After the completion of rinse procedure, the hemodialysis device
may be powered off automatically if it idles for 15 minutes.
-87-
JHM-2028 A Hemodialysis Device User’s Manual
Every time after the treatment, use the pledget to clean the water and stain on the surface of the
hemodialysis devic e with the alc ohol or the hypochlorite. Keep the cleanness and the dryness of the
hemodialysis devic e. If there is the phenomenon of leaking water or dialysate, it should be cleanup and
Do examine the colander web on the connector termly. If with the phenomena of jam, the web
should be cleaned or changed. The discarded filtering web should be dis posed properly to prevent
6.4.1 Pipeline connection during the disinfection, hot disinfection and rinse
-88-
JHM-2028 A Hemodialysis Device User’s Manual
RO water
supply system
-89-
JHM-2028 A Hemodialysis Device User’s Manual
Warning:
In the process of interior cleanness, make sure that there is enough
disinfection liquor in the dis infection container to undergo the disinfection
process.
6.4.2 Operation
Blood Return
Runnig
Lower C onduct ivity Alarm
B lood Level alarm Dialysis treatment finished.
TMP. A.P. V.P. Batt ery Blood Leak Blood Level A ir B ubble Heparin C logging Temp. Cond.
Blood return(stop status ).
UF Rate He parin T ime Time Left Temp.
6 00 4 00 30 0
0010 ml/h 3:30 0:00 23.0℃ Press “Return” key, the screen w ill return to
1 5. 5
500 3 00 2 50
2 00 2 00
UF Volume He parin F lu x Na+ HCO3- 15 display prime menu (stop status).
400
0000 ml 10.0 ml/h 138mmol/l 032 mmol/l 1 4. 5
300 10 0 15 0
UF Goal 14
200 0 10 0
- 10 0
0000 ml 1 3. 5
1 00 50
- 20 0
Blood pump Spare Pump 13
0 0
-10 0 - 30 0 - 50
200 ml/mi n 200ml /min 1 2. 5
-90-
JHM-2028 A Hemodialysis Device User’s Manual
Pr ime
S top p ed
Lo we r Co nd uc tivit y Ala r m
Blo od L eve l a la rm Prime rinse menu (stop status).
TMP . A.P . V. P. Bat te ry Blo o d L e ak Bloo d L eve l Air Bub b le He p ar in Clo gg ing Te mp . Co n d.
-1 00 -3 00 -5 0
200 ml /min 200ml/min 12 .5
Disinfection
Stopped Cleanness menu.
Press “Disinfect” key, and according to
TMP. A.P. V. P. Ba tte ry Blo od L ea k B loo d Le ve l Air Bu b ble He pa rin C log g ing Te mp. C on d .
04.17
Ri nse Disinfect Hot Di s. Return MON 15:03
Disinfection
Stopped Confirm disinfection mode, system w ill start to
TMP. A. P. V.P. Batt ery Blood Leak Blood Level Air Bubble H eparin C logging Temp. Cond.
undergo disinfection process.
Ac cor din g t o su gg est ion ,do op er atio n on suc tion tu be s of liqu id( A) a nd
liq uid( B)
60 0 40 0 30 0
Con ne ct d ialysa te su pply /re tur n tub es of dialy zer to th e p or t o f mach ine
15 .5
50 0 30 0 25 0 Rin se time is 10 min s
Co nfi rm 15
4 00 20 0 20 0
Disinfecti on selected: Citric acid 14 .5
30 0 10 0 15 0
2 00 0 10 0
? Press 'ok' to continue
Press 'cancel' to re-sel ect, cha nge mode
14
13 .5
10 0 - 100 50
OK Can ce l 13
0 - 200 0
12 .5
- 100 - 300 - 50
mmHg mmHg mmHg mS/cm
04.17
Rinse Disinfect Hot Di s. Return 15:03
MON
-91-
JHM-2028 A Hemodialysis Device User’s Manual
Disinfection
Running Disinfection program menu.
TMP . A. P. V.P. Ba tt er y Blo o d Le a k Blo o d L e ve l Air Bub b le H ep ar in Cl og gin g Temp . C on d.
100 - 100 50
13. 5
The menu displays the conductiv ity, remained
13
0 - 200 0 Rins e Dis infec t
12. 5 time and temperature in dis infect.
-1 00 - 300 - 50 1%
1%
mmHg mmHg mmHg mS/cm
When time counts down to 0:00, dis infection
04.17 finished and alarm voic e raised. Press “Confirm”,
Rinse Disinfect Hot Dis. Return MO N 15:03
the system prompts “Auto-Power off 5 minutes
after no operation on the device”.
Press “Return” button and the system enters
automatically the self -test mode.
Disinfection
Disinfec tion T ime Up “Disinfection time out” info. will display in text
Stopped
area.
-92-
JHM-2028 A Hemodialysis Device User’s Manual
Dis infection
S topped Cleanness menu.
TMP. A.P. V. P. Ba tte ry Blo od L ea k B loo d Le ve l Air Bu b ble He pa rin C log g ing Te mp. C on d .
Press “Hot Dis.” Key, system will start hot
disinfection process.
Acc or ding to su gg est ion ,do op er at ion on su ction tu be s o f liq uid( A) an d
liqu id( B)
6 00 400 300
Co nn ec t d ialysa te su pply /re tu rn tub es of dia lyze r to th e por t o f mac hin e
1 5. 5
500 300 25 0 Rins e t ime is 1 0 min s
15
40 0 20 0 20 0 Ple as e d on 't sto p it d ur ing ho t d isinfe ctio n o r disin fec tion pr oc ess
1 4. 5
300 1 00 1 50 The d isinfe ctio n solu tion u sed in disin fec tion an d ho t dis infe ctio n is
con fec ted by th e d efin ed fo rmula 14
200 0 1 00 N ot a llow ed to re tur n to oth er mod es if th e d isinf ect o
i n or ho t d isinf ectio n
h as no t b ee n co mple ted 1 3. 5
1 00 -1 00 50
Th e mac hin e mus t g o th ro ug h d isin fec tion or h ot d isin fec tion pr oc ess 13
0 -2 00 0 a fre r eve ry tre at me nt eve ry time
1 2. 5
-1 00 -3 00 -5 0
mmHg mmHg mmHg mS/cm
04.17
Ri nse Disinfect Hot Di s. Return MON 15:03
Disinfection
Stopped Confirm hot disinfection mode, system w ill
TMP. A. P. V.P. Batt ery Blood Leak Blood Lev el Air Bubble H eparin C logging Temp. Cond.
start to undergo disinfection process.
Ac cor din g t o s ugg es tion ,do o per at ion on su ction tu be s o f liqu id( A) a nd
liq uid (B)
60 0 4 00 30 0
Con ne ct d ialys ate su pp ly/re tu rn tub es of dia lyzer to th e p or t o f mac hine
15 .5
5 00 3 00 2 50 Rin se time is 10 min s
15
4 00 2 00 2 00 C onfi rm
Hot D is. selected: He at + Citric a cid 14 .5
3 00 10 0 15 0
2 00 0 10 0
? Press 'ok' to continue
Press 'cancel' to re -sel ect, ch ange mode
14
13 .5
10 0 - 100 50
OK Can ce l 13
0 - 200 0
12 .5
- 100 - 300 - 50
mmHg mmHg mmHg mS/cm
04.17
Rinse Disinfect Hot Dis. Return 15:03
MON
-93-
JHM-2028 A Hemodialysis Device User’s Manual
Disinfection
Running Hot Disinfection process menu.
TMP . A. P. V.P. B att er y Blo o d Le a k Blo o d Le ve l Air Bu bb le H ep a rin C log gi ng Te mp. C on d.
04.17
Rinse Disinfect Hot Dis. Return MON
15:03
-94-
JHM-2028 A Hemodialysis Device User’s Manual
Warning:
It maybe damage this device if using the improper disinfection
solution (concentration).
Dis infection
S topped Cleanness menu.
Press “Rinse” key, system starts to undergo
TMP. A.P. V. P. Ba tte ry Blo od L ea k B loo d Le ve l Air Bu b ble He pa rin C log g ing Te mp. C on d .
6 00 400 300
Acc or ding to su gg est ion ,do op er at ion on su ction tu be s o f liq uid( A) an d
liqu id( B) rinse process.
Co nn ec t d ialysa te su pply /re tu rn tub es of dia lyze r to th e por t o f mac hin e
1 5. 5
500 300 25 0 Rins e t ime is 1 0 min s
15
40 0 20 0 20 0 Ple as e d on 't sto p it d ur ing ho t d isinfe ctio n o r disin fec tion pr oc ess
1 4. 5
300 1 00 1 50 The d isinfe ctio n solu tion u sed in disin fec tion an d ho t dis infe ctio n is
con fec ted by th e d efin ed fo rmula 14
200 0 1 00 N ot a llow ed to re tur n to oth er mod es if th e d isinf ect o
i n or ho t d isinf ectio n
h as no t b ee n co mple ted 1 3. 5
1 00 -1 00 50
Th e mac hin e mus t g o th ro ug h d isin fec tion or h ot d isin fec tion pr oc ess 13
0 -2 00 0 a fre r eve ry tre at me nt eve ry time
1 2. 5
-1 00 -3 00 -5 0
m m Hg m m Hg mm Hg m S /cm
04.17
Ri nse Disinfect Hot Di s. Return MON 15:03
-95-
JHM-2028 A Hemodialysis Device User’s Manual
Note:
When the rinse mode is selected as hot rinse, the system can only
execute the rinse procedure as the temperature reaches the hot rinsing
temperature preset on the adjustment interface.
Rinse
Running Rinse process menu.
TMP. A .P. V.P. B at tery Bl ood Leak Blood Level Air Bubble H eparin Clogging Temp. Cond.
Current mode displays rinse info.
6 00 4 00 3 00
Cond. Time Left Temp.
Work time of rinse process is 10 mins.
15 .5
5 00 3 00 2 50
00.0 mS/cm 0:09 34.3 ℃ 15 The menu displays the conductiv ity, remained
4 00 2 00 2 00
3 00 10 0 15 0
14 .5
time and temperature in rinse.
14
2 00 0 10 0
10 0 - 10 0 50 Rin se
13 .5 When time counts down to 0:00, rinse finished
13
1%
0 - 20 0 0
12 .5
and alarm voic e raised. Press “Confirm”, the
- 10 0 - 30 0 - 50
mmHg mmHg mmH g mS/cm
system prompts “Auto-Power off 15 minutes after
04.17 no operation on the device”.
Rinse Disinfect Hot Dis. Return MON 15:03
Press “Return” button and the system enters
automatically the self -test mode.
Disinfection
Rinse Time Up “Rinse time out” info. will display in text area .
Stopped
The rinse program can be stopped via
pressing “Rinse” key in rinse.
-96-
JHM-2028 A Hemodialysis Device User’s Manual
Dialysis
Running
Air bubble alarm Alarm info. displays in text area.
TMP. A.P. V. P. Battery Blood Leak Blood Level Air Bubble H eparin C logging Temp. Cond.
Such as: Air bubble alarm.
Corresponding alarm indicator shines red and
send off voice alarm.
Note:
The mute cycle of audio alarm is 100s.
Note:
If the problem occurred is not be solved but the alarm being cleared
repeatedly, there is potential risk whic h should be monitored by the
personnel, especially the items that may result in the personal injury
such as the venous pressure, liquid level, air bubble and blood leakage.
-97-
JHM-2028 A Hemodialysis Device User’s Manual
200 0 100
100 -100 50
0 -200 0
-100 -300 - 50
m m Hg m m Hg m m Hg
6 00 400 300
Cause:
5 00 300 25 0
A.P exceeds alarm limit.
4 00 200 20 0
3 00 1 00 150
2 00 0 100
10 0 -10 0 50
0 -20 0 0
- 100 -30 0 -5 0
mmHg mmHg mmHg
-98-
JHM-2028 A Hemodialysis Device User’s Manual
TMP. A.P. V. P.
V.P. alarm indicator (red) shines.
600 400 3 00
Cause:
500 300 250
V.P exceeds alarm limit.
400 200 200
300 1 00 1 50
200 0 1 00
1 00 -10 0 50
0 -20 0 0
-10 0 -30 0 -50
m m Hg m m Hg m m Hg
Ba ttery Bl ood Le ak Blo od Leve l Air bubb le Hep arin cl ogg ing Blood leakage alarm indicator (red) shines.
Cause:
Blood exis ts in the dialysate loop.
-99-
JHM-2028 A Hemodialysis Device User’s Manual
Ba ttery Bl ood Le ak Blo od Leve l Air bubb le Hep arin cl ogg ing Blood level alarm indicator (red) shines.
Cause:
Blood level in venous pot is below detector
or there are foamed blood.
Ba ttery Blo od Le ak Blo od Leve l Air b ubb le Hep arin cl oggi ng Air-bubble alarm indicator (red) shines.
Cause:
There is air bubble in blood loop.
-100-
JHM-2028 A Hemodialysis Device User’s Manual
Ba ttery Bl ood Le ak Blo od Leve l Air b ubb le Hep arin cl oggi ng The indicator (red) for the alarm of heparin
block will shine.
Cause:
Heparin pump clog.
Alarm cause:
Dialysis
He parin e mpty alarm
Ru nning No injection solution in heparin syringe.
The info. of “heparin empty alarm” displays in
word area.
Send off alarm voice.
Alarm cause:
Dialysis
He parin t ime up
Ru nning The time of heparin injection counts down to
0:00.
The info. of “heparin time up” displays in word
area.
Send off alarm voice.
-101-
JHM-2028 A Hemodialysis Device User’s Manual
C on d .
Indicator (red) of conductiv ity alarm shines.
Cause:
1 5.5
Actual value of conductiv ity exceeds alarm
15
14.5
limit.
14
13.5
13
12.5
m S /c m
Bl ood Le vel Air bub ble Hep arin cl oggi ng Temp. C ond. Indicator (red) of temp. alarm shines.
Cause:
Actual temp. of dialysate exceeds alarm
limit.
Over-heat limit
Dialysis
Overh eat alarm Alarm cause:
Ru nning
Actual temp. of dialysate is abouve 41℃.
Time Left Temp
Over-heat alarm info. Displays in word area.
3:00 41.5℃
Current temp. displays on temp. screen.
Na+ Concentration
-102-
JHM-2028 A Hemodialysis Device User’s Manual
Ba ttery Bl ood Le ak Blo od Leve l Air b ubb le Hep arin cl oggi ng The signal light (red) for battery alarm shine.
Cause:
The storage battery inside machine is not
suffic ient to support one minute’s alarm.
Dialysis
Battery alarm info. displays in text area.
Low er battery alarm
Sto pped Generate audio alarm.
7.6 Warning/notice
Warning/notice show
Dialysis
The info. of warning/notic e or the relative
Co nnect Su pply t ube to machine
Sto pped
content display in the text area.
i.e.: Connect Supply tube to machine.
Generate auto alarm.
-103-
JHM-2028 A Hemodialysis Device User’s Manual
7.7.1 The disposal principle for the unusual thing of hemodialysis device
2. If there is the malfunction about the conductivity, perform the “rinse” in the allowed short time, then
examine and check.
3. If the hemodialysis device can’t resume, please transfer the dialysis patient to another good
hemodialysis devic e to receive the treatment.
2. If the warning voic e can’t be eliminated at once, do reinput the blood to the patient’s body and pause
the treatment, then undergo the examination and check. If it is resumed, continue the treatment,
otherwis e change the hemodialysis device or continue the treatment after the device is resumed.
3. During the treatment of dialysis , if there is malfunction that it is impossible to use the dialysis device
to return the blood for the patient, do turn the spanner on clockw is e by hand to make the blood
transport to the patient. Continue the treatment after the devic e is resumed or change another
device.
4.If the reverse ultrafiltration occurs during dialysis treatment, the water whic h enters into the body may
increase the blood pressure and weight of the patient. When confirm that this is not caused by the
malfunction of devic e, you can increase the blood flow or the ultrafiltration volume such that the
-104-
JHM-2028 A Hemodialysis Device User’s Manual
Note:
With the alarm voice, press the key of “alarm mute” to eliminate the
warning voic e,but the warning indicator and warning words will still
display on the screen. At this time, operator should inspect the
parameter of protection system which the alarming is cancelled.
Warning:
The power supplied by the backup battery may maintained for 15
min after the power supply of the alarm system is failed; and the log will
be retained in this period.
Note:
The record will be retained even if the power of alarm system fails (mains
and internal supply ).
Note:
The power w ill be supplied by backup battery after the mains is
failed, and the alarm settings will not be changed during the period of the
power-off and recovery.
Warning:
If the device can’t be resume by the outer adjustment, do report to the
Engineers to repair the device.
Warning:
If the same or similar devices are used in any separate area such as
ICU or cardiac operating room, there may be potential hazardness when
the different alarm preset is applied.
-105-
JHM-2028 A Hemodialysis Device User’s Manual
Spot disposal:
Physiolog Alarm
Alarm Priorit
ical alarm delay Possible cause Solution
inform ationin b y
Yes/No (sec)
Upper limit alarm Excessiv e Lower the flow of
of Middle ultrafiltration volume ultrafiltration pump
No 2
transmembrane
Clogged pipe Unclog the pipe
pressure
Lower limit alarm
of Increase the flow of
Middle No 2 Insuffic ient blood flow
transmembrane blood pump
pressure
Clogged pipe Unclog the pipe
Upper limit alarm
Venous pipeline
of venous High Yes 4
tw is ted or under reorganize the pipeline
pressure
pressure
increase flow of blood
low blood flow
Lower limit alarm pump
of venous High Yes 4 Loose of intravenous Reconnect the
pressure needle intravenous needle
Clogged dialyzer Replace dialyzer
The conductivity of Check if there is no liquid
concentrate A and B in w ater pipeline
Upper limit alarm does not comply with Replace concentrate A
Middle No 2
of conductivity the requirement and B
Air leakage in the fast Replace the fast joint of
joint of dialy zer dialyzer
The conductivity of Check if there is no liquid
Low limit alarm of concentrate A and B in w ater pipeline
Middle No 2
conductiv ity does not comply with Replace concentrate A
the requirement and B
Press Bridging button for
3 seconds to start the
blood pump. When the
Alarm of air Air entering the blood
High Yes <0.12 air bubbles are queued
bubble pipeline
to the opening of vein,
pull out the needle to
remove the air bubbles,
-106-
JHM-2028 A Hemodialysis Device User’s Manual
-107-
JHM-2028 A Hemodialysis Device User’s Manual
Concentrate A does
Error of
Middle No 2 not comply with Replace the concetrate
concentrate A
requirement
Concentrate B does
Error of
Middle No 2 not comply with Replace the concetrate
concentrate B
requirement
No disinfectant can be
Put the disinfectant
aspirated by the
suction tube into the
disinfectant suction
Error of disinfectant container
Middle No 2 tube
disinfectant
Disinfectant cannot
comply with the Replace the dis infectant
requirement
Connect the The circuit of
Put the connector of
pipeline of Liquid concentrate A is not
Middle No 2 Liquid A (red) into the
A to the supply connected to the tank
tank of concentrate A
kettle of concentrate A
The connector of
Put connector of Liquid A
Connect the Liquid A (red) is not
(red) into the socket of
pipeline of Liquid Middle No 2 put in the socket of
Liquid A in the device
A to the device Liquid A in the devic e
and screwed up
and screwed up.
Connect the The circuit of
Put the connector of
pipeline of Liquid concentrate B is not
Middle No 2 Liquid B (blue) into the
B to the supply connected to the tank
tank of concentrate B
kettle of concentrate B
The connector of
Put connector of Liquid B
Connect the Liquid B (blue) is not
(blue) into the socket of
pipeline of Liquid Middle No 2 put in the socket of
Liquid B in the device
B to the device Liquid B in the devic e
and screwed up
and screwed up.
Connect the supply pipe
of dialysate (blue) to the
Connect the The return pipeline is
low er end of dialyzer and
return pipe to Middle No 2 not connected to
the return pipe of
dialyzer dialyzer
dialysate (red) to the
upper end of dialy zer
-108-
JHM-2028 A Hemodialysis Device User’s Manual
-109-
JHM-2028 A Hemodialysis Device User’s Manual
Dialyzed Patient’s
Supplied by backup
Alarm of leaving device to be
Middle No <1 supply after the mains
power-off arranged as soon as
failed
possible
When set ultrafiltration
Time-out of Time-out of the set time elapses, instruct the
High No <0.12
dialysis time dialysis time patient leaving the
device
as the time-out of ISO
Time-out of ISO Time-out of set ISO
High No <1.1 UF, the device switches
UF UF
to dialysis state
As the time-out of the
Time-out of the devic e
Time-out of rinse High No <1.1 device rinse, the water
rinse
pipeline stops
As the time-out of the
Time-out of Time-out of the devic e
High No <1.1 device dis infection, the
disinfection disinfection
water pipeline stops
As the time-out of the
Time-out of hot Time-out of the devic e
High No <1.1 device hot dis infection,
disinfection hot disinfection
the water pipeline stops
Incorrect setting of
Upper limit of re-set the upper/lower
High No 4 upper/lower limit of
arterial pressure limit of arterial pressure
arterial pressure
The arterial end of
blood pipeline under Unclog the pipeline
the pressure
Blood pump runs too Decrease the flow of
fast blood pump
Arterial stenosis or
Lower limit of Re-adjust the puncture
High No 4 incorrect position of
arterial pressure needle
arterial needle
Insuffic ient flow of the Adjust the upper/lower
patient limit of arterial pressure
Incorrect setting of
re-set the upper/lower
upper/lower limit of
limit of arterial pressure
arterial pressure
No liquid in w ater No supply of reverse Check if the switch of
High No 2
pipeline osmosis water reverse osmosis water is
-110-
JHM-2028 A Hemodialysis Device User’s Manual
opened or not
Note: Alarm of highest priority, the alarm indicator flickers in red, “beep, beep, beep…beep, beep…beep,
beep, beep…beep, beep…”.
Alarm of middle priority, the alarm indicator flic kers in yellow, “beep, beep, beep…beep, beep…beep,
beep, beep…beep, beep…”.
-111-
JHM-2028 A Hemodialysis Device User’s Manual
Maintenance disposal
Note:
The following is for maintenance engineer’s reference.
Communicaton There are sth wrong w ith the Ask for professional personel to solve the problem.
error for connection between main Non-professional person is not allowed to
host-computer board and waterway board disassemble it
and water way
-112-
JHM-2028 A Hemodialysis Device User’s Manual
Warning:
If the hemodialysis device can’t be resumed to normal status by
exterior adjustment, pleaseinform maintenance engineer.
Unusual
Unusual status The way to check and solve
part
-113-
JHM-2028 A Hemodialysis Device User’s Manual
8.1 Termly check the device whether the supply of water (such as the water purify unit) and electric ity
(such as the manostat) is good or not. And check whether those devices work properly or not, the
voltage is stable or not. For the area whic h the voltage is alw ays higher or lower, it is better to use the
automatic manostat.
8.2 Check the precision of the ultrafiltration per half a year or a year, have a look of the operation 4.12 for
reference.
8.3 Termly check the filtration web in the hemodialysis device, it there exists the situation of blocks, clear
up or change it in time.
Note:
After clearing up or changing the web, take care the airproof when
installing to avoid the connector of web w ill leak air.
8.4 If with the built-in power supply, the user should charge the battery according to follow ing ways when
Connect the hemodialysis device with the electricity source system by the electricity wire.
Connect with the electric ity source, the devic e should be charged for about 20 hours.
If the device have not used for a long time, repeat the process every 6 months.
8.5 The device can’t be impacted to avoid the damage to the parts in it.
8.6 The device should be put upright in the house w ith the relative humidity not more than 80%, no
8.7 This device has the guarantee of “return, replace and repair” for one year,please read the guarantee
card for reference.
-114-
JHM-2028 A Hemodialysis Device User’s Manual
power sw itch
display
screen
start switch
Start
-115-
JHM-2028 A Hemodialysis Device User’s Manual
-116-
JHM-2028 A Hemodialysis Device User’s Manual
mmHg
kPa
mmHg
kPa
-117-
JHM-2028 A Hemodialysis Device User’s Manual
Safety precautions
-118-
JHM-2028 A Hemodialysis Device User’s Manual
minutes, the cuff can only function as expected when such period is expired.
9.1.4 There is potential risk when the balloon is maintained in over-inflation state.
9.1.5 For the occurrence of the general arrhythmia (such as APB, VPB and AF), the measured value
may not be correct or the blood pressure cannot be measured.
Relevant statement
9.1.6 Although the precision of pressure sensor of cuff has been tested, it is recommended to check and
calibrate the sensor annually in order to ensure the normal function and precis e measurement of
the hemomanometer. Use of pressure calibration instrument for verif ic ation is suggested.
70kPa-106kPa.
Storage temperature/humidity: relative humidity ≤80%, in the room without caustic gas and w ith good
ventilation.
9.1.8 The blood pressure value measured w ith this hemomanometer is equiv alent to the measured value
of stethoscopy w ith the error consistent with the requirement of YY0670.
Warning:
To measure according to the recorded instruction and the standard
of storage temperature/humidity, or the correct measurement may not be
realized.
Note:
Any measurement of blood pressure may be subject to the posture
and/or physical conditions of the measured indiv idual. The error of
unwanted movement may occur if there is any motion of body in the
measurement.
-119-
JHM-2028 A Hemodialysis Device User’s Manual
Dialysis
In the power cut status, system can start
Power off alarm
Running
spare power supply automatically, battery
displays surplus capacity.
-120-
JHM-2028 A Hemodialysis Device User’s Manual
Citric acid 36
Peracetic acid 31
Heating+Citric acid 42
Heating 32
Introduction:
z In normal condition, all the Pre-rinse Time in every model should be 1min. While if the water
conductiv ity is higher, time maybe extended till the conductivity is reduced.
z Disinfection Mode—> citric acid: it can disinfect the machine and remove the calcium.
z Hot Dis. Mode—> heat + citric acid: it can dis infect the machine and remove calc ium. In this mode,
the time of heating water till 65℃ is not included in both cycling disinfec tion time and Total Time.
z Hot Disi. Mode—> heat: It can disinfect the machine. In this mode, the time of heating water till 65℃
-121-
JHM-2028 A Hemodialysis Device User’s Manual
Use of sulfonic acid anion exchange resin (Rezex ROA- Organic Acid H+ ( 8%,150 mm ×7.8
mm, 8µm), or column w ith similar nature) as filler, and 0.015mol/L sulphuric acid solutionas the
mobile phase, flow rate 0.6ml/min, detection w avelength 210nm, column temperature 40℃. The
chromatogram conditions and suitability test should be consistent w ith the requirements of Appendix
1.0 g citr ic acid control w eighted precisely is dissolved w ith dialysis w ater to volume in the
volumetric flask until 10.0g/L citric acid control is obtained. Take and pour 20μL of such control in the
liquid chromatography for determination and the calculation of the peak area. Take the test solution
and dilute based on the appropriate dilution ratio w ith the dialysis w ater as the solution to be tested.
This solution is filtered w ith 0.45μm membrane and 20μL of w hich is taken and poured in the liquid
chromatography for determination and the calculation of the peak area. The concentration of citric
acid in the test article is calculated w ith the external standard method.
-122-