Beruflich Dokumente
Kultur Dokumente
EXPERIMENTAL SECTION
The direct experimental determination of vapour – liquid equilibrium means
analyzing the concentration of both liquid and vapour phase which are in equilibrium which
each other. There are some systems for which equilibrium relation can be calculated from
Raoult’s and Dalton’s laws from the known dependences of the vapour pressure of the pure
compounds on temperature, but there are a large number of industrially important systems
whose equilibrium relations can’t be predicted from purely theoretical considerations. Vapour
liquid equilibrium data can be generated either at constant temperature or at constant pressure.
For distillation calculations the isobaric data are more important and their determination is also
simpler. Classification of the methods that can use to find equilibrium relation experimentally
(Hala et al. 1967).
Distillation method,
Static method,
Flow method,
Circulation method,
The distillation method is the oldest method in which small amount of liquid is
distil off from boiling flask. It is very simple method. Major disadvantage is that large amount
of liquid is required and major error can be obtain due to condensation of boiling liquid on the
cold surface of the distillation flask at the beginning of the experiment(Hala et al. 1967).
In static method, binary mixture is charged to a closed, heated chamber and mixes
until equilibrium is established. Major disadvantage is that, small change in pressure and
volume can have significant effect on the system due to that it is very difficult to take a sample
for analysis without disturbing the equilibrium. This method is generally use for high pressure
because for low and moderate pressure other easier methods are available (Scholten 1997).
13 | P a g e
The circulation method is the most common and widely accepted method. In
circulation method vapors coming off from the boiling mixture are condensed in condenser and
this condensate is returned to liquid chamber, creating a continuous cycle. Constant pressure
and temperature in the system indicate that pseudo – equilibrium steady state is achieved.
Setup 1:
For the VLE data generation chemicals (pure component as well as binary mixture)
were fed into the modified ebulliometer. Ebulliometer has provision of drop counter to ensure
equilibrium conditions more over data can be generated with minimum quantity of chemical
requirement. The drop counter facility is very unique and provides additional resource to ensure
equilibrium condition along with the temperature. The bottom section of the ebulliometer was
wound with external heating element. The detail of the experimental setup given in Table 3.1.
14 | P a g e
Figure 3.2 Schematic Experimental Setup 1.
1. Ebulliometer, 1 a. Feed outlet, 1 b. Boiling chamber, 1 c. Cottrell Pump, 1 d. Drop counter, 1 e. Thermowell,
1 f. Feed inlet 2. Mercury in glass thermometer, 3. External belt heater, 4. U – Tube Mercury manometer, 5.
Jacketed internal coil condenser, 5 a. Cooling water inlet, 5 b. Cooling water outlet, 6. SS Ballast tank, 7. Bypass
line, 8. Belt Driven oil ring Vacuum pump, 9. Nitrogen cylinder, 10. Dimmer stat.
15 | P a g e
Setup 2:
Setup 1 was failed due to high temperature gradient between condensed vapor and
boiling chamber create an crack in setup so that new setup of stainless steel – 316 was prepared
as shown in figure 3.3. Design parameter of SS-316 ebulliometer is unchanged and as same as
mention in figure 3.1. The detail of the experimental setup given in Table 3.2. Experimental
procedure for this setup is also same as setup 1. Small but thick glass tube is provided as drop
counter.
16 | P a g e
Setup 3:
Pure component data for n-butanol was successfully generated in Setup 2 but it
failed to generate data for butyl glycolate as due to high wall temperature effect leading to
decomposition of butyl glycolate. To resolve this issue third setup was build based on indirect
heating concept. A cylindrical glass vessel with provision of tilted thermowell was connected
to double coiled glass condenser with O ring vacuum trap joint. System assembly is shown in
figure 3.4. Change detail of the experimental setup given in Table 3.2. Steps for experimental
procedure are mention below.
Pure or mixture solution and magnetic stirrer was initially taken in the glass vessel.
Vessel was deep in oil bath,
Condenser was then connected using O ring connector,
Start cooling water supply and magnetic stirrer,
Oil bath temperature was set at specific temperature,
Wait till thermometer shows constant temperature,
17 | P a g e
Setup 4:
3.2 CHEMICALS
Pure component vapor pressure data as well as partial vapor – liquid equilibrium data was
generated for n-butanol – butyl glycolate system. The details about chemicals used for this
work is given in Table 3.4. The GC analysis of all these species showed absence of any
additional peak hence they were used directly without any purification.
18 | P a g e
Butly Glycolate Sigma – Aldrich > 90% GC
Chemie Gmbh
3.4 Analysis
The mixture of ethylene dichloride and n-butanol was analyzed on Chemito Gas
Chromatograph using 1.8 m Chromatopak column. The parameters set for the analysis are
given in Table 3.3.
19 | P a g e
Figure 3.3 Calibration plot for Ethylene Dichloride (EDC) and n-Butanol (BuOH)
20 | P a g e