CAUTION: For proper operation, follow the instruction manual when using the instrument.
Specifications in this catalog are subject to change with or without notice, as Hitachi High-Technologies Corporation continues to develop the latest
technologies and products for our customers.
Yellow background: ON
Maintenance buttons
method of analysis). The L-8800 also uses the rection column Pump purge button
system to provide the highest sensitivity among the amino acid
Linear Velocity
analyzers using the ninhydrin reaction method.
*1 To achieve enhanced sensitivity, the reaction column is filled with special particles, which
B N2 gas automatic purge button
are used to suppress peak dispersion for better separation. In addition, the loading
pressure has been reduced to improve the reaction efficiency. Automatic operation of solenoid valve
A
4 6 (new function)
1 2
DATA
Cys
Resin: #2622
Ser
Lys
Val
Thr
His
Asp
Glu
Met
Ala
・N2 gas automatic purge *2
Phe
Tyr
NH3
・Leak sensor
Gly
Ile
Arg
・Degasser
Leu
・Accommodates up to 6 buffers *3
・Max. temp: 140 ° C (Reaction unit)*4
Trp
・Max. temp: 80 ° C (Column oven)*5
Pro
・New and improved pump
・New, direct injection autosampler *6
0 2 4 6 8 10 12 14 16 18 20
Retention time (min)
*2: Automatic operation of solenoid valve
*3,*4 &*5: Together the new hardware features support new
applications.
*6: The new autosampler reduces the impact on the injection
selector valve produced when the sample is injected to about
1/7 of that with the existing autosamplers. As a result, it
provides improved accuracy in determination of the peak that ● PH (Protein Hydrolysate)
appears within 5 minutes after the start of separation. Standard analysis, 30 minutes
Column size: ?.6 mmID x 60 mm
Resin: #2622
Cys
Lys
Val
Ser
Thr
Asp
Glu
His
Met
Tyr
Phe
NH3
Gly
Arg
Ile
Leu
Ala
Trp
0 10 20 30
Retention time (min)
● High-speed physiological
35
fluid analysis, 120minutes
HCys
Asp
Column size: 4.6 mmID x 60 mm
ASA-Anhy1
30
Thr
Resin: #2622
P-Ser
Orn
GluNH2
Ser
Lys
Impact reduced to approx. 1/7 Data acquired using an existing autosampler
P-Ser
25
Gly
AEC
Ala
Cit
Cysthi
Signal strength (mv)
1Mehis
20
Glu
Hylys
Phe
g-ABA
His
Hylys
3Mehis
Met
Val
Cys
Urea
Arg
15
PEA
AspNH2
a-ABA
ASA
Sar
Leu
IIe
EOHNH2
HCit
0.8mV 10
NH3
MetSOX
a-AAA
b-AiBA
Tyr
Car
b-Ala
Cys-HCys
HCysH
Saccha
ALevA
AIIo-IIe
Trp
Ans
Pipeco
5
Hypro
Pro
3 4
DATA SPECIFICATIONS
Val
ninemia
Acetyl-Lys
25 Detection limit 3 pmol (S/N = 2, Asp)
Cystathionin-
uria Phenylketonuria Analyzer
Column 4.6 mm ID ' 60 mm L ※1
Met
Signal strength (mv)
20
AIIo-IIe
Packed with Hitachi custom ion exchange resin
Leu
IIe
Phe
2
15 Pump Delivery pressure; 0 to 20 MPa (0 to 306 kgf/cm )
Cysthi
● Biological sample partial Flow rate range; 0.000 to 0.999 mL/min
Tyr
10 analysis, 26 minutes Autosampler Injection methods; Direct injection method
Sample vial capacity; 1500 µL
Column size: 4.6 mmID x 60 mm
No. of vials accommodated; 120 (80 when optional cooler installed)
5
Resin: #2622
Sample injection volume; 0.1 to 100 µL
Column oven System; Peltier
0
Temperature setting range; 20 to 85 °
C (in 1 °
C steps)
0 2 4 6 8 10 12 14 16 18 20 22 24 26
Reaction unit System; Electronic heating (reaction column)
Retention time (min)
Temperature setting; 50 to 140 °
C (in 1 °
C steps)
Photometer Spectrophotometer; Aplanatic concave diffraction grating
Wavelength; 570 nm, 440 nm
EZChrom Elite for HITACHI Amino Acid Analyzer
※2
CPU 32 bit OS, Windows XP ®
● Biological sample partial CRT 17 inch color display
analysis, 10 minutes Dimensions (excluding PC) 900(W) ' 650(D) ' 1050(H) mm
Weight (excluding PC) Approx. 250 kg
Column size: 4.6 mmID x 60 mm
Operating temperature range 15 to 35 °
C
Resin: #2622
Power supply 100 ∼115 V AC / 220 ∼240 V AC, 800 VA and over, w/in 50/60 Hz ± 0.5 Hz
N2 gas N2 gas source must be prepared.
Tyr
Phe
※2 Windows is a registered trademark of Microsoft Corp. in the U.S. and other countries.
■ Set Layout
100∼
200 or greater 900 300 700
400 or greater
Power supply N
2
& grounding gas
0 1 2 3 4 5 6 7 8 9 10 2m 5m
Retention time (min)
650
PC table
L-8900 or
Main unit PC rack
1,000 or greater
This space is unnecessary
Cys
Orn
Gly
g-ABA
Lys
Ala
Cit
Asp
Hylys
Cysthi
Tau
Phe
GluNH2
Leu Ile
Thr
Glu
Ser
3Mehis
b-AiBA
1Mehis
Arg
b-Ala
His
a-ABA
Tyr
Urea PEA
EOHNH2
AspNH2
Trp
a-AAA
Sar
Car
NH3
Ans
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70
Retention time (min)
5 6