Sie sind auf Seite 1von 12

Biofouling

The Journal of Bioadhesion and Biofilm Research

ISSN: 0892-7014 (Print) 1029-2454 (Online) Journal homepage: http://www.tandfonline.com/loi/gbif20

Biofilm formation in an experimental water


distribution system: the contamination of
non-touch sensor taps and the implication for
healthcare

Ginny Moore, David Stevenson, Katy-Anne Thompson, Simon Parks, Didier


Ngabo, Allan M. Bennett & Jimmy T. Walker

To cite this article: Ginny Moore, David Stevenson, Katy-Anne Thompson, Simon Parks, Didier
Ngabo, Allan M. Bennett & Jimmy T. Walker (2015) Biofilm formation in an experimental water
distribution system: the contamination of non-touch sensor taps and the implication for
healthcare, Biofouling, 31:9-10, 677-687, DOI: 10.1080/08927014.2015.1089986

To link to this article: http://dx.doi.org/10.1080/08927014.2015.1089986

Published online: 07 Oct 2015.

Submit your article to this journal

Article views: 575

View related articles

View Crossmark data

Full Terms & Conditions of access and use can be found at


http://www.tandfonline.com/action/journalInformation?journalCode=gbif20

Download by: [Universidad de Buenos Aires] Date: 07 June 2016, At: 11:40
Biofouling, 2015
Vol. 31, Nos. 9–10, 677–687, http://dx.doi.org/10.1080/08927014.2015.1089986

Biofilm formation in an experimental water distribution system: the contamination of


non-touch sensor taps and the implication for healthcare
Ginny Moore*, David Stevenson, Katy-Anne Thompson, Simon Parks, Didier Ngabo, Allan M. Bennett and
Jimmy T. Walker
Biosafety Investigation Unit, Public Health England, Salisbury, UK
(Received 3 July 2015; accepted 28 August 2015)

Hospital tap water is a recognised source of Pseudomonas aeruginosa. UK guidance documents recommend measures to
control/minimise the risk of P. aeruginosa in augmented care units but these are based on limited scientific evidence. An
experimental water distribution system was designed to investigate colonisation of hospital tap components. P. aeruginosa
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

was injected into 27 individual tap ‘assemblies’. Taps were subsequently flushed twice daily and contamination levels
monitored over two years. Tap assemblies were systematically dismantled and assessed microbiologically and the effect of
removing potentially contaminated components was determined. P. aeruginosa was repeatedly recovered from the tap
water at levels above the augmented care alert level. The organism was recovered from all dismantled solenoid valves
with colonisation of the ethylene propylene diene monomer (EPDM) diaphragm confirmed by microscopy. Removing the
solenoid valves reduced P. aeruginosa counts in the water to below detectable levels. This effect was immediate and
sustained, implicating the solenoid diaphragm as the primary contamination source.
Keywords: Pseudomonas aeruginosa; tap water contamination; non-touch sensor taps; biofilm; model water distribution
system

Introduction Several outbreaks of P. aeruginosa have been attribu-


Water and water systems harbour a diversity of microor- ted to contaminated tap water (Ferroni et al. 1998;
ganisms and whilst the majority are considered harmless Vianelli et al. 2006; Aumeran et al. 2007; Durojaiye
to healthy individuals, some, such as Pseudomonas et al. 2011). Once established within a faucet or else-
aeruginosa, can cause serious opportunistic infections, where within the water system, P. aeruginosa is difficult
particularly in individuals with a weakened immune sys- to eradicate and this is partially attributable to the bio-
tem (Lyczak et al. 2000). There are reports to suggest film forming properties of the organism (Høiby et al.
that waterborne P. aeruginosa may account for 3% of all 2001). Biofilms are complex, structured, specialised
healthcare-associated infections (Anaissie et al. 2002; communities of adherent microorganisms surrounded by
Hidron et al. 2008). Consequently, hospitals and other extracellular polymeric substances. They are fundamen-
healthcare facilities need to control the risk posed by tally different from populations in suspension in terms of
water systems to at-risk patients. cell surface characteristics, metabolism and chemical
Neonates in whom the immune system is not yet structure (Donlan 2002). These physiological changes
developed are particularly vulnerable (Cipolla et al. together with delayed penetration of chemical agents
2011; Jefferies et al. 2012). Between November 2011 through the biofilm matrix result in an increased toler-
and January 2012, 25 babies admitted to four different ance to antimicrobial agents, including disinfectants and
hospitals in Northern Ireland acquired P. aeruginosa other chemical biocides (Bridier et al. 2011).
whilst being treated in the neonatal intensive care unit Following the incidents in Northern Ireland, 30 repre-
(NICU); nine babies developed a symptomatic infec- sentative tap assemblies were removed from the neonatal
tion and four died (Troop 2012). Each of the four units and sent to Public Health England (PHE) Porton,
incidents was associated with a different strain (or where they were dismantled and analysed. Scanning
strains) of P. aeruginosa, indicating that transmission electron microscopy revealed extensive biofilm formation
between units had not occurred. However, 18 (72%) on sections of the plastic flow straighteners (Walker
of the 25 babies acquired the same strain of et al. 2014). These flow straighteners, in comparison to
P. aeruginosa as that recovered from water samples other tap components, were the most likely to be con-
and/or swabs taken from handwash basins located taminated with P. aeruginosa and were associated with
within their NICU. the highest P. aeruginosa counts, implicating them as a

*Corresponding author. Email: ginny.moore@phe.gov.uk

© Crown Copyright 2015. Reproduced with the permission of the Controller of Her Majesty’s Stationery Office and Public Health England
678 G. Moore et al.

Twenty-seven tap ‘assemblies’ each comprising a


thermostatic mixer valve (Pegler Prestex P402 15 mm
TMV3; Pegler Yorkshire Group Ltd, Doncaster,
UK), solenoid valve (Phoenix Contact Ltd, Telford, UK),
15 mm gate valve (Screwfix Direct Ltd, Yeovil, UK),
binder-style test point (Macro Construction, Portsmouth,
UK) and stainless steel non-touch sensor tap (Aquarius
deck mounted short neck tap with copper tails; Dart
Valley Systems, Paignton, UK) were connected to the
circulating hot and cold water supply via 15 mm diame-
ter copper piping (Figure 1). To allow automation of the
EWDS, each solenoid valve (disconnected from its
associated infra-red sensor) was connected to a bespoke
30-channel Siemens electronic control panel (AJK Ser-
vices Ltd, Tidworth, UK) programmed to sequentially
Figure 1. The experimental water distribution system open each solenoid valve for 30 s.
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

(EWDS).

Operation of experimental water distribution system


possible source of the neonatal infections (Walker et al. Twice daily (Monday–Friday), the hot and cold bypass
2014). As a consequence, the UK Department of Health valves were opened to allow water to circulate the
commissioned PHE to investigate the formation of bio- EWDS. When the temperature of the returning hot water
film on tap components similar in type to those removed was ~ 60°C, the control panel was manually switched on
from the neonatal units in Northern Ireland. The aim of and each tap assembly automatically flushed for 30 s.
this study was to design and build an experimental water The TMV and gate valve (pre-set during installation)
distribution system and, with particular reference to ensured the temperature and flow rate of the water from
P. aeruginosa, investigate the contamination of different each outlet was ~ 43°C and ~ 4.5 l min−1 respectively.
tap components, including different types of flow
straightener.
Contamination of experimental water distribution
system
Materials and methods
The EWDS was contaminated with Pseudomonas
Design of experimental water distribution system aeruginosa PA14 (variable number tandem repeat
(EWDS) (VNTR) profile 12, 2, 1, 5, 5, 2, 4, 5, 12) (Martin et al.
Borehole-supplied water (total hardness: 248 mg 2013). An overnight culture of P. aeruginosa was diluted
CaCO3 l−1; total chlorine: 0.25–0.61 mg l−1) was stored 100-fold using sterile deionised water. Ten ml of the
in two domestic-style copper indirect combination cylin- resulting suspension (~107 colony forming units (CFU)
ders (capacity: 115 l (bottom tank), 40 l (top tank); RM ml−1) were injected directly into each tap assembly via its
Cylinders, Castleford, UK). A heating element in one of associated binder point. To encourage P. aeruginosa
the tanks constantly heated the water to 60°C; the other colonisation of the distribution system, the water within
held water at room temperature. The water storage tanks the EWDS was allowed to stagnate for five days, after
were connected to a centrifugal, negative head, supply- which time the twice daily flushing regimen was resumed.
on-demand shower pump (Salamander Pumps, Sunder-
land, UK) which, operating at 3 bar, supplied the hot
and cold water to the EWDS via two separate copper Collection and microbiological analysis of water
lines (diameter 22 mm). The water system incorporated samples
bypass valves to allow for the flushing of each line with- Water samples were taken to assess colonisation and to
out the use of the taps, both to drain and to allow recir- monitor contamination levels over time. In the UK, aug-
culation of unmixed hot and cold water back to the mented care units are advised that if test results are satis-
respective storage tanks. Recirculation allowed a constant factory (ie if P. aeruginosa is not detected) water sampling
supply of water at a controlled temperature to be main- can be undertaken on a six-monthly basis. For the pur-
tained throughout the system, preventing the constant on/ poses of this study and to assess how contamination levels
off operation of the water pump, which can lead to may fluctuate between any two sampling occasions, water
inconsistent supply pressures and flow rates. samples were taken over a two year period.
Biofouling 679

Microbiological analysis of tap components


Flow straighteners/faucet aerators
Each of the 27 taps was fitted with a flow straightener or
one of two faucet aerators (Figure 2; Neoperl Ltd,
Malvern, UK). After remaining in situ for two years, the
external surface (outlet face) of each flow straightener/
aerator was sampled (whilst still in situ) using a sterile
cotton-tipped swab (Sterilin Ltd, Newport, UK) moist-
ened with sterile thiosulphate Ringers (Oxoid). The swab
was snapped into 1 ml sterile thiosulphate Ringers and
vortexed for 1 min to release any bacteria from the swab
bud. The resulting suspension was diluted 10-fold and
100 μl aliquots of each dilution were plated on to non-
selective agar (R2A) and agar selective for P. aeruginosa
(C-N agar) and Pseudomonas spp. (C-F-C agar; Oxoid).
Each flow straightener/aerator (and associated rubber
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

seal) was removed from the tap outlet and rinsed with
50 ml of sterile irrigation water to remove any loosely
attached bacteria. The flow straightener/aerator was then
placed in a plastic container containing 5 ml of sterile
thiosulphate Ringers and 10 sterile glass beads and vor-
texed for 2 min. The resulting suspension was diluted
100-fold and 100 μl aliquots of each dilution were plated
on R2A, C-N and C-F-C agar plates. The remaining
(undiluted) suspension was passed through a membrane
filter (0.2 μm) which was then placed on the surface of a
C-N agar plate. Finally, each flow straightener/aerator
was transferred to 10 ml of nutrient broth (Oxoid) and
incubated at 37°C for 48 h. An aliquot of broth was then
spread over the surface of a C-N agar plate.

Figure 2. Tap outlets were fitted with a flow straightener (A)


or one of two faucet aerators (B and C).

The first sample of water (500 ml) delivered from each


tap was collected in sterile, disposable polyethylene bottles
containing sodium thiosulphate (20 mg l−1) to neutralise
chlorine residues. The presence of P. aeruginosa was
determined by passing 100 ml of each water sample
through a polyethersulfone membrane disc filter (47 mm,
0.2 μm; Pall Life Sciences, Portsmouth, UK) and transfer-
ring the filter to Pseudomonas agar supplemented with
cetrimide and sodium nalidixate (C-N agar, Oxoid Ltd,
Basingstoke, UK). To determine the heterotrophic plate
count, the water samples were diluted 100-fold and ali-
quots (100 μl) of each dilution plated onto non-selective
Figure 3. Break down of the solenoid valve into its component
agar (R2A; Oxoid). Water taken directly from the hot and parts: (1) inlet port; (2) outlet port; (3) EPDM diaphragm;
cold water tanks was sampled in the same manner. (4) armature, push rod and seal; (5) outer casing for magnetic coil.
680 G. Moore et al.

Binder points and solenoid valves Kruskal–Wallis one way analysis of variance in
The binder points and solenoid valves associated with 10 combination with Dunn’s multiple comparison test.
tap assemblies (randomly selected using Microsoft Excel Sample data were considered statistically significant
(2012) (RANDBETWEEN function)) were removed and when p < 0.05.
dismantled using manual hand tools (Figure 3). Each
component part was rinsed with sterile irrigation water
Results
(to remove loosely attached bacteria) before being sam-
pled using two sterile, pre-moistened cotton-tipped Colonisation of the experimental water distribution
swabs. Each pair of swabs was transferred to a plastic system
container containing 10 ml of sterile water and 10 sterile Following contamination of the EWDS (and after the
glass beads and vortexed for 2 min to release any bacte- water had been allowed to stagnate for five days), the
ria from the swab bud. The resulting suspension was median number of P. aeruginosa colonies recovered from
diluted 100-fold and 100 μl aliquots of each dilution the first sample of water delivered from each tap was
were plated on to R2A and C-N agar plates. The remain- 1.1 × 103 CFU 100 ml−1. All taps tested positive for
ing (undiluted) suspension was passed through a mem- P. aeruginosa with counts ranging from 1 CFU 100 ml−1
brane filter (0.2 μm) which was then placed on the to 4.5 × 105 CFU 100 ml−1. Sampling was next carried
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

surface of a C-N agar plate. out two days after the twice-daily flushing regimen had
Segments of the different surface materials associated been resumed. On this occasion, there was no detectable
with an unused flow straightener, aerator and solenoid P. aeruginosa in water taken from seven of the 27 taps
valve were analysed using a Phillips XL30 FEG (ie P. aeruginosa counts were <1 CFU 100 ml−1). These
scanning electron microscope (SEM) (FEI UK Ltd, seven tap outlets tested negative for P. aeruginosa for the
Cambridge). Corresponding segments taken from compo- duration of the study period. The remaining 20 tap
nents removed from the EWDS were also analysed. assemblies had become colonised. The median hetero-
trophic plate count was 3.5 × 105 CFU 100 ml−1 whilst
median P. aeruginosa counts one week and six months
Removal and replacement of tap components after contamination of the EWDS were 690 CFU
The 10 binder points and five of the 10 solenoid valves 100 ml−1 and 304 CFU 100 ml−1 respectively (Figure 4).
that had been removed for sampling were replaced with The type of flow straightener/aerator fitted within the out-
new components. The remaining five solenoid valves let had a small but significant effect upon contamination
and all 27 flow straighteners/aerators were permanently levels (p = 0.006). The median number of P. aeruginosa
removed from the EWDS; the former replaced with colonies recovered from water delivered via the flow
copper piping (15 mm). The twice daily flushing regimen straightener was 197 CFU 100 ml−1 (n = 210) whilst the
was resumed and weekly water samples were taken to more complex aerator (outlet fitting B; Figure 2) was
assess any immediate and continual effect of removing associated with the lowest P. aeruginosa counts
or replacing potentially contaminated tap components. (145 CFU 100 ml−1; n = 208; p < 0.05). P. aeruginosa
persisted within the EWDS and continued to be recovered
from the 20 tap outlets at levels above the alert value for
Analysis of results augmented care (ie > 10 CFU 100 ml−1). One year after
All C-N plates were incubated at 37°C for 48 h. contamination of the EWDS, median P. aeruginosa
Presumptive P. aeruginosa colonies were subcultured counts were 267 CFU 100 ml−1. At no point during the
and confirmed using the API 20NE biochemical test kit investigation was P. aeruginosa detected in the source
(bioMérieux, Basingstoke, UK). Representative isolates water.
were sent to the Antimicrobial Resistance and Healthcare
Associated Infections Reference Unit (London, UK) for
Microbiological analysis of tap components
typing. VNTR analysis at nine loci was carried out as
previously described (Turton et al. 2010). Flow straighteners/faucet aerators
All C-F-C and R2A plates were incubated at 30°C The median number of bacteria recovered from the swab
for three to five days, after which time the number of samples (ie the external surface (outlet face) of the flow
colony forming units were enumerated. Predominant straighteners/faucet aerators) was 10 CFU (n = 27). Het-
colonies were subcultured and identified using the API erotrophic plate counts ranged from <10 CFU to 4.72 ×
20NE test kit. 104 CFU. P. aeruginosa (10 CFU) was recovered from a
Data analysis was performed using SigmaPlot for single surface swab. The type of flow straightener/aerator
Windows (version 12.0). Intergroup comparisons were had no significant effect upon heterotrophic plate count
made using the Mann–Whitney rank sum test or the (p > 0.05) or the prevalence of P. aeruginosa.
Biofouling 681
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

Figure 4. Median (IQR) recovery of P. aeruginosa from water delivered from the EWDS. The alert value for augmented care is also
illustrated (- - - -).

The median number of bacteria recovered from the In all cases, the most predominant colony was identified
flow straighteners/aerators (together with the associated as being Sphingomonas paucimobilis. Stenotrophomonas
rubber seal) was 2.55 × 105 CFU (n = 27). Heterotrophic maltophilia (50 CFU) was recovered from one of the
plate counts ranged from <50 CFU to 1.60 × 106 CFU faucet aerators. However, no presumptive P. aeruginosa
with significantly fewer bacteria being recovered from colonies were recovered from any of the direct or filtered
the flow straighteners than from the aerators (p = 0.019). samples, although P. aeruginosa was recovered from one

Figure 5. Median (IQR) number of bacteria recovered from the surface of different parts of the solenoid valve.
682 G. Moore et al.

of the enriched flow straighteners. The type of flow (p < 0.05; Figure 5). Isolates colonising the solenoid valve
straightener/aerator had no significant effect upon had the same VNTR profile (ie were the same strain) as
prevalence of P. aeruginosa (p > 0.05). that recovered from the water. P. aeruginosa was not
recovered from any of the binder points.

Binder points and solenoid valves


The heterotrophic and P. aeruginosa plate counts from Removal and replacement of tap components
different component parts of the solenoid valve are pre- Removing the flow straightener/aerator from the tap
sented in Figure 5. The results have been standardised by outlet had no significant effect upon the median number
dividing the number of colonies recovered by the surface of P. aeruginosa colonies recovered from the tap water
area sampled and are expressed as CFU cm−2. The med- (p > 0.05). P. aeruginosa continued to be recovered from
ian number of bacteria recovered from the solenoid valves tap assemblies in which the original solenoid valve had
was 3.4 × 104 CFU cm−2. Heterotrophic plate counts ran- remained in situ and, two years after contamination of the
ged from 2.9 × 103 CFU cm−2 to 5.1 × 105 CFU cm−2. EWDS, P. aeruginosa counts associated with eight of 10
The diaphragm was significantly more contaminated such taps were still above the alert level for augmented
(1.7 × 105 CFU cm−2) than the outer casing for the mag- care (Figure 6). In contrast, removing the solenoid valve
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

netic coil (5.75 × 103 CFU cm−2; p < 0.05). The median and replacing it with either copper piping or a new
number of P. aeruginosa colonies recovered from the component immediately reduced the P. aeruginosa count
solenoid valves was 8.7 CFU cm−2. Counts ranged from in water delivered from the tap outlet to below detectable
0.3 CFU cm−2 to 1.25 × 103 CFU cm−2 with significantly levels (ie < 1 CFU 100 ml−1; Figure 6). Removing the
more P. aeruginosa being recovered from the diaphragm solenoid valve from a tap assembly already negative for
(79 CFU cm−2) than from the inlet and outlet ports P. aeruginosa (tap 16) had no immediate effect on the

Figure 6. Recovery of P. aeruginosa from the water delivered from the EWDS: the immediate and continual effect of removing or
replacing the solenoid valve.
Biofouling 683

presence/absence of P. aeruginosa. However, removing became colonised and the water delivered from the tap
or replacing the solenoid valve ensured that there was no outlets was contaminated with P. aeruginosa at a level
detectable P. aeruginosa in water taken from any such that, if detected in an augmented care unit, would result
tap for the remainder of the study. in the isolation of the outlet and a requirement for
remediation work. The absence of P. aeruginosa in both
the source and recirculated water suggested that the
Discussion source of the contamination, rather than being systemic,
A range of model systems, including chemostats (Rogers was at or close to the outlet. This was similar to that
et al. 1994), Propella reactors (Manuel et al. 2007) and observed in the neonatal units where despite high counts
rotating disc reactors (Murga et al. 2001) have been used in pre-flush samples, P. aeruginosa was not detected in
to investigate the formation of biofilm on different samples collected post-flush, indicative of a problem
plumbing materials. Such laboratory studies are control- related to the local water outlet(s).
lable and reproducible but do not necessarily replicate Flow straighteners taken from the neonatal units were
‘in-use’ conditions. Recreating a water distribution sys- shown to be heavily colonised with P. aeruginosa (median
tem under laboratory conditions can be challenging but CFU: 5.2 × 104; n = 41) and those described as being
can provide a realistic model in which to investigate ‘complex’ were associated with significantly higher
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

microbial contamination and control (Muraca et al. 1987; P. aeruginosa counts than those more simplistic in design
Waines et al. 2011). (Walker et al. 2014). In the current study, the type of outlet
The experimental water distribution system (EWDS) fitting had a small but significant effect upon the number
used in the current study was designed to simulate that of of P. aeruginosa colonies recovered from the water. The
a hospital ward and was based upon tap components aerator considered the most complex (outlet fitting B;
similar in type to those removed from the neonatal units Figure 2) was associated with the lowest P. aeruginosa
in Northern Ireland (Troop 2012; Walker et al. 2014). counts (p < 0.05) suggesting that bacteria may have
Following contamination of the EWDS, there was no become trapped and retained within the complex internal
detectable P. aeruginosa in water taken from seven of the configuration of the fitting, potentially leading to the
27 taps, reflecting perhaps the natural variability observed formation of biofilm on and/or within the aerator.
in hospital water systems. However, 20 tap assemblies Certainly, SEM analysis of the two types of fitting showed

Figure 7. SEM images (magnification 4,000×) of biofilm on a section of the plastic flow straightener (A) and aerator (B) after two -
years in situ. Higher numbers of bacteria were recovered from the aerator (B). Images of an unused flow straightener (C) and aerator
(D) are included for comparison.
684 G. Moore et al.

marked differences in terms of biofilm structure (Figure 7)


and when removed from the outlet, significantly more
bacteria were recovered from aerators than from the less
complex flow straighteners (p = 0.019). However, despite
the presence of high numbers of P. aeruginosa in the water
delivered from the outlets and despite remaining in situ for
two years, only two of the 27 outlet fittings were contami-
nated with P. aeruginosa (≤ 10 CFU) and both isolates
had a VNTR profile (8, 2, 5, 3, 4, 2, 7, 2, 10) that differed
from those recovered from the water.
It is known that P. aeruginosa can enter a viable but
non-culturable state (Bédard et al. 2014), thus, the num-
ber of colonies recovered from the outlet fittings may
not represent the actual level of colonisation. However,
the weakness of any laboratory model is the inherent
difficulty in extrapolating results to the ‘real-world’ envi-
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

ronment (Muraca et al. 1987). The visual appearance of


the flow straighteners removed from the EWDS differed
considerably from those taken from the neonatal units
after four months in situ (Mayes et al. 2014) (Figure 8).
Local environmental factors including the concentration
of carbon, nitrogen and phosphorus in the water (Frias
et al. 2001; Chu et al. 2005), the disinfection process
(Lund and Ormerod 1995) and the condition of the pipe
material (Pederson 1990) can impact bacterial growth
and biofilm formation. Fundamentally however, whilst a
laboratory model can simulate a hospital water distribu-
tion system, it is not possible to recreate the clinical
environment.
In the current study, colonisation of the EWDS was
Figure 8. Visual appearance of outlet fittings taken from a achieved by injecting P. aeruginosa directly into each tap
neonatal unit (A) and the EWDS (B) after remaining in situ for assembly. In the ‘real-world’ environment, retrograde
four months and two years, respectively. Images demonstrate a
contamination of faucets by patients may be a more likely
significant difference in biofilm formation.

Figure 9. Macroscopic and microscopic (magnification 4,000×) images of biofilm on an EPDM diaphragm removed from the
EWDS. Arrows indicate individual bacterial cells within the biofilm.
Biofouling 685

scenario (Reuter et al. 2002; Rogues et al. 2007). Conclusions


Colonisation of taps via splash contamination from the The aim of this investigation was to build an experimen-
hand basin (Pitten et al. 2001) and/or during the washing tal water distribution system to investigate the formation
of soiled hands (Grundmann et al. 1993) has also been of biofilm on different tap components. Following
reported. A tap outlet associated with a hand basin fre- artificial contamination of the EWDS, the solenoid valve
quently used for rinsing patient care items or for disposing became colonised with P. aeruginosa and the EPDM
patient secretions is also more likely to be contaminated diaphragm was identified as a major weakness within a
than an outlet associated with a basin that is not misused non-touch tap system. These findings support recom-
(Balm et al. 2013). Such observations have resulted in a mendations that sensor taps be removed from augmented
recommendation that aerators/flow straighteners be care units. However, there are benefits to installing auto-
removed, replaced and/or regularly disinfected (Weber matic taps and manufacturers should be made aware of
et al. 1999; Kappstein et al. 2000; Balm et al. 2013). the potential risks associated with EPDM and challenged
However, if, as in the current study, the outlet fitting is not to propose new design solutions incorporating alternative
the primary or only source of P. aeruginosa (or other materials. The EWDS was operated over a two year per-
potential pathogen), such measures may have little effect iod and during this time high numbers of P. aeruginosa
in terms of microbial or infection control. were repeatedly recovered from the water delivered from
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

Tap handles can become heavily contaminated and the outlets. Despite this, under laboratory conditions,
non-touch sensor taps have been introduced in many colonisation of the flow straighteners and aerators was
hospitals both for water economy and in the belief that minimal, implying clinical practice and/or behaviour may
avoidance of hand contact will reduce cross infection play a major role in the contamination of outlet fittings.
(Zingg and Pittet 2012). However, it has been reported Thus, in addition to engineering interventions, strategies
that in comparison to manually operated taps, the water should be developed to reduce the potential for retrograde
delivered from sensor taps is more likely to be contami- contamination and these should be effectively communi-
nated with P. aeruginosa and/or other waterborne patho- cated to clinical and non-clinical staff. P. aeruginosa is a
gens (Halabi et al. 2001; Merrer et al. 2005; Sydnor et al. ubiquitous environmental microorganism and its control
2012) and be associated with higher bacterial counts within the healthcare environment should be the responsi-
(Hargreaves et al. 2001). It has been suggested that the bility of all.
comparatively low flow rate coupled with a lack of ther-
mal control and the presence of synthetic rubber materials
Acknowledgements
may facilitate the survival of bacteria and encourage bio-
The authors would like to thank Paul Day, Trevor Williams
film growth (Halabi et al. 2001; Walker et al. 2014). and Liam Oliver for their contribution to the design and build
In a previously reported study, water samples taken configuration of the EWDS; Dart Valley Systems for supplying
from different sections of non-touch sensor taps implied the non-touch sensor taps and solenoids; Neoperl Ltd for sup-
the outlet and the solenoid valve were highly contaminated plying the outlet fittings; Howard Tolley for carrying out the
with P. aeruginosa (Halabi et al. 2001). In the current scanning electron microscopy; Jane Turton for carrying out the
VNTR analysis and Caroline Lymbery for additional technical
study, individual tap components were systematically assistance.
removed from the EWDS, dismantled and sampled
directly. High numbers of bacteria were recovered from
the solenoid valve, particularly from the diaphragm. Conflict of interest disclosure statement
Macroscopic, microscopic and microbiological analysis of No potential conflict of interest was reported by the authors.
the ethylene propylene diene monomer (EPDM) surface
confirmed the presence of biofilm (Figure 9) and the
Funding
colonisation by P. aeruginosa (Figure 5).
Replacing sensor taps with conventional manual taps This study was funded by the Department of Health. The views
expressed in this publication are those of the authors and not
has eliminated P. aeruginosa from hospital tap outlets necessarily those of PHE or any other Government Agency.
(Merrer et al. 2005; Berthelot et al. 2006) and resulted in
the termination of outbreaks (Durojaiye et al. 2011). The
solenoid valve is integral to a non-touch tap system and, References
in the current study, its removal from a tap assembly Anaissie EJ, Penzak SR, Dignani MC. 2002. The hospital water
reduced the P. aeruginosa count in water delivered from supply as a source of nosocomial infections: a plea for
the tap outlet to below detectable levels. This effect was action. Arch Intern Med. 162:1483–1492.
immediate and sustained and implicated the solenoid dia- Aumeran C, Paillard C, Robin F, Kanold J, Baud O, Bonnet R,
Souweine B, Traore O. 2007. Pseudomonas aeruginosa
phragm as the source of the contamination, confirming and Pseudomonas putida outbreak associated with contami-
EPDM as an unsuitable material for use in water distribu- nated water outlets in an oncohaematology paediatric unit.
tion systems (Moritz et al. 2010; Waines et al. 2011). J Hosp Infect. 65:47–53.
686 G. Moore et al.

Balm MN, Salmon S, Jureen R, Teo C, Mahdi R, Seetoh T, bacteraemia in paediatric oncology patients. J Hosp Infect.
Teo JT, Lin RT, Fisher DA. 2013. Bad design, bad prac- 44:27–30.
tices, bad bugs: frustrations in controlling an outbreak of Lund V, Ormerod K. 1995. The influence of disinfection
Elizabethkingia meningoseptica in intensive care units. J processes on biofilm formation in water distribution sys-
Hosp Infect. 85:134–140. tems. Water Res. 29:1013–1021.
Bédard E, Charron D, Lalancette C, Déziel E, Prévost M. Lyczak JB, Cannon CL, Pier GB. 2000. Establishment of Pseu-
2014. Recovery of Pseudomonas aeruginosa culturability domonas aeruginosa infection: lessons from a versatile
following copper- and chlorine-induced stress. FEMS opportunist. Microbes Infect. 2:1051–1060.
Microbiol Lett. 356:226–234. Manuel CM, Nunes OC, Melo LF. 2007. Dynamics of drinking
Berthelot P, Chord F, Mallaval F, Grattard F, Brajon D, water biofilm in flow/non-flow conditions. Water Res.
Pozzetto B. 2006. Magnetic valves as a source of faucet 41:551–562.
contamination with Pseudomonas aeruginosa? Intens Care Martin K, Baddal B, Mustafa N, Perry C, Underwood A, Con-
Med. 32:1271–1271. stantidou C, Loman N, Kenna DT, Turton JF. 2013. Clus-
Bridier A, Briandet R, Thomas V, Dubois-Brissonnet F. 2011. ters of genetically similar isolates of Pseudomonas
Resistance of bacterial biofilms to disinfectants: a review. aeruginosa from multiple hospitals in the UK. J Med
Biofouling. 27:1017–1032. Microbiol. 62:988–1000.
Chu C, Lu C, Lee C. 2005. Effects of inorganic nutrients on Mayes C, McCracken G, Rooney P. 2014. Minimising risk
the regrowth of heterotrophic bacteria in drinking water from Pseudomonas aeruginosa using tap design. Arch Dis
distribution systems. J Environ Manage. 74:255–263. Child Fetal Neonatal Ed. 99:A40–A40.
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

Cipolla D, Giuffrè M, Mammina C, Corsello G. 2011. Prevention Merrer J, Girou E, Ducellier D, Clavreul N, Cizeau F, Legrand
of nosocomial infections and surveillance of emerging resis- P, Leneveu M. 2005. Should electronic faucets be used in
tances in NICU. J Matern-Fetal Neonatal Med. 24:23–26. intensive care and hematology units? Intens Care Med.
Donlan R. 2002. Biofilms: microbial life on surfaces. Emerg 31:1715–1718.
Infect Dis. 8:881–890. Moritz MM, Flemming H-C, Wingender J. 2010. Integration of
Durojaiye OC, Carbarns N, Murray S, Majumdar S. 2011. Out- Pseudomonas aeruginosa and Legionella pneumophila in
break of multidrug-resistant Pseudomonas aeruginosa in an drinking water biofilms grown on domestic plumbing mate-
intensive care unit. J Hosp Infect. 78:154–155. rials. Int J Hyg Environ Health. 213:190–197.
Ferroni A, Nguyen L, Pron B, Quesne G, Brusset M, Berche P. Muraca P, Stout JE, Yu VL. 1987. Comparative assessment of
1998. Outbreak of nosocomial urinary tract infections due chlorine, heat, ozone, and UV light for killing Legionella
to Pseudomonas aeruginosa in a paediatric surgical unit pneumophila within a model plumbing system. Appl Envi-
associated with tap-water contamination. J Hosp Infect. ron Microbiol. 53:447–453.
39:301–307. Murga R, Forster TS, Brown E, Pruckler JM, Fields BS,
Frias J, Ribas F, Lucena F. 2001. Effects of different nutrients Donlan RM. 2001. Role of biofilms in the survival of
on bacterial growth in a pilot distribution system. Antonie Legionella pneumophila in a model potable-water system.
van Leeuwenhoek. 80:129–138. Microbiology. 147:3121–3126.
Grundmann H, Kropec A, Hartung D, Berner R, Daschner F. Pederson K. 1990. Biofilm development on stainless steel and
1993. Pseudomonas aeruginosa in a neonatal intensive care PVC surfaces in drinking water. Water Res. 24:239–243.
unit: reservoirs and ecology of the nosocomial pathogen. J Pitten F-A, Panzig B, Schröder G, Tietze K, Kramer A. 2001.
Infect Dis. 168:943–947. Transmission of a multiresistant Pseudomonas aeruginosa
Halabi M, Wiesholzer-Pitt M, Schober J, Mittermayer H. 2001. strain at a German university hospital. J Hosp Infect.
Non-touch fittings in hospitals: a possible source of Pseu- 47:125–130.
domonas aeruginosa and Legionella spp. J Hosp Infect. Reuter S, Sigge A, Wiedeck H, Trautmann M. 2002. Analysis
49:117–121. of transmission pathways of Pseudomonas aeruginosa
Hargreaves J, Shireley L, Shannon H, Bren V, Gordon F, between patients and tap water outlets. Crit Care Med.
Lacher C, Esslinger V, Watne T. 2001. Bacterial contamina- 30:2222–2228.
tion associated with electronic faucets: a new risk for Rogers J, Dowsett A, Dennis P, Lee J, Keevil C. 1994. Influ-
healthcare facilities. Infect Control Hosp Epidemiol. ence of plumbing materials on biofilm formation and
22:202–205. growth of Legionella pneumophila in potable water sys-
Hidron A, Edwards J, Patel J, Horan T, Sievert D, Pollock D, tems. Appl Environ Microbiol. 60:1842–1851.
Fridkin S. 2008. Antimicrobial-resistant pathogens associ- Rogues AM, Boulestreau H, Lasheras A, Boyer A, Gruson D,
ated with healthcare-associated infections: annual summary Merle C, Castaing Y, Bebear CM, Gachie JP. 2007. Con-
of data reported to the national healthcare safety network at tribution of tap water to patient colonisation with Pseu-
the centers for disease control and prevention, 2006–2007. domonas aeruginosa in a medical intensive care unit. J
Infect Control Hosp Epidemiol. 29:996–1011. Hosp Infect. 67:72–78.
Høiby N, Krogh Johansen H, Moser C, Song Z, Ciofu O, Sydnor ER, Bova G, Gimburg A, Cosgrove SE, Perl TM,
Kharazmi A. 2001. Pseudomonas aeruginosa and the Maragakis LL. 2012. Electronic-eye faucets: legionella spe-
in vitro and in vivo biofilm mode of growth. Microbes cies contamination in healthcare settings. Infect Control
Infect. 3:23–35. Hosp Epidemiol. 33:235–240.
Jefferies JM, Cooper T, Yam T, Clarke SC. 2012. Pseudomonas Troop P. 2012. Independent review of incidents of Pseu-
aeruginosa outbreaks in the neonatal intensive care unit - a domonas aeruginosa infection in neonatal units in Northern
systematic review of risk factors and environmental Ireland. Belfast (Northern Ireland): The Regulation and
sources. J Med Microbiol. 61:1052–1061. Quality Improvement Authority.
Kappstein I, Grundmann H, Hauer T, Niemeyer C. 2000. Aera- Turton JF, Turton SE, Yearwood L, Yarde S, Kaufmann ME,
tors as a reservoir of Acinetobacter junii: an outbreak of Pitt TL. 2010. Evaluation of a nine-locus variable-number
Biofouling 687

tandem-repeat scheme for typing of Pseudomonas aerugi- Walker JT, Jhutty A, Parks S, Willis C, Copley V, Turton JF,
nosa. Clin Microbiol Infect. 16:1111–1116. Hoffman PN, Bennett AM. 2014. Investigation of health-
Vianelli N, Giannini MB, Quarti C, Sabattini MB, Fiacchini M, care-acquired infections associated with Pseudomonas
de Vivo A, Graldi P, Galli S, Nanetti A, Baccarani M. aeruginosa biofilms in taps in neonatal units in Northern
2006. Resolution of a Pseudomonas aeruginosa outbreak Ireland. J Hosp Infect. 86:16–23.
in a hematology unit with the use of disposable sterile Weber DJ, Rutala WA, Blanchet CN, Jordan M, Gergen MF.
water filters. Haematologica. 91:983–985. 1999. Faucet aerators: a source of patient colonization with
Waines PL, Moate R, Moody AJ, Allen M, Bradley G. 2011. Stenotrophomonas maltophilia. Am J Infect Control.
The effect of material choice on biofilm formation in a 27:59–63.
model warm water distribution system. Biofouling. Zingg W, Pittet D. 2012. Electronic-eye faucets - curse or
27:1161–1174. blessing? Infect Control. 33:241–242.
Downloaded by [Universidad de Buenos Aires] at 11:40 07 June 2016

Das könnte Ihnen auch gefallen