SLW Liquid Oil Flow meter Digital Display Operation

OPERATION

1. DISPLAY

The display is divided into four function display


First zone: Flow range, pressure and temperature indication area,Q,P
Second Zone: Battery indication in the right side of the display
Third Zone: instaneous flow, 6 figures in first line, the decimal points are floating.
Fourth Zone: Totalized flow, 8 figures in second line, the decimal point is fixed two decimal
places.
Display unit:
1.
Instaneous flow: M3/h, Totalized flow M3,
2.
Instaneous flow: L/h, Totalized flow L

2. Keyboard Operation
1
http://www.silverinstruments.com

The display is shown as following:

There are 4 keys: “Ent”、“→”、“↑”、“Esc”


“Ent”:Scroll the menus; save keys when confirming the modification.
“Esc”:Exit without modification, Exit when wrongly setting.
“→”:Change cursor position
“↑”:To add value

Modification Operation: when move “→”key, the corresponding number will flash, then you
can change by “↑”.

First Display: Normal display, no unit display, M3 or L unit is decided by function options.
The first line shows instaneous flow, the decimal points are floating.
The first line shows Totalized flow,the decimal point is fixed two decimal places.
Battery indication is shown in the right side.

Second Display: other display items


First line: battery voltage: first decimal point place, it will show in “U x.x”format.
Progress bar of voltage in the right side: display: ≥3V display S1-S14;between 2.3V-3V,
One check represents 0.05V.
Second line shows signal frequency: one decimal, it will show in “F x.x”format.

2
http://www.silverinstruments.com

Third Display: Parameter setting:


First line: no display
Second line: “- - - -”, password is needed
Enter password “1234”:browse and modify \parameter settings;
Enter password “5678”: save the factory setting and return to main display screen.
Enter password “1111”: return to factory setting.
When no entering any password or enter wrong password, you can still browse the menu but
you have to right to modify any parameter setting

When the password is correct, press “Ent”and the correct prompt is “0000”.
When re-confirm, ① enter into “Fourth display”, starting to browse and modify the settings.
②③ exit to normal display, ④ enter into “Fourth Display”, start to scroll but without any
setting. Press“Esc”to exit.

Note: password is to prevent wrong setting from users.

Fourth Display: Display the compensation factor unit


First line:Display function code 01
Second line: Display compensation factor unit
1: when entering M3 factor, it will display M3 flow.
2: when entering L factor, it will display L flow.

3
http://www.silverinstruments.com

3: when entering L factor, it will display M3 flow. (This option is for large M3 factor, but the
requirement is to display M3 flow. To choose with “↑”, the flow will show the real time
exchange, when exchange the unit into “L”, only 8 figures are displayed. Others are only saved
without any display.
4: enter L factor and display L/min flow.

Fifth Display:
Display “ liter” pulse equivalent.( M3 unit is set by DIP Switch)
First line shows function code 02
The second line is shown:
100: means 100L
10: means: 10L
1: means: 1L (change by “↑” key).

Sixth Display: Damping time


First line shows Function code 03
Second line shows damping time, the unit is “second”,1-10 second.

4
http://www.silverinstruments.com

Seventh Display: It shows the max flow


First line shows Function code 04
Second line shows the max flow (8 digits)
No unit indication, M3or L is determined by function option.
Accuracy: 0.01 M3/h or 0.01L/h

Eighth Display: It shows the min flow of liquid turbine flow meter
First line shows Function code 05
Second line shows the min flow (8 digits)
No unit indication, M3 or L is determined by function option.
Accuracy: 0.01 M3/h or 0.01L/h

5
http://www.silverinstruments.com

Ninth Display: It shows the upper limit of the output frequency:


First line shows Function code 06
Second line: Frequency upper limit
0-3000Hz, accuracy is 0.1Hz

Tenth Display: Baud rate display


First line shows Function code 07
Second lines shows baud rate display: 1200、2400、4800、9600、19200;
Verification: n ( no verification), o(odd verification),E(even verification)
Data length: 7,8
Stopping numbers: 1, 2

Eleventh Display
First line shows Function code 08
Second line shows communication address: 01~99

6
http://www.silverinstruments.com

Twelfth Display
First line shows Function code 09
Second line shows frequency output mode:
1: Original frequency
2: frequency after treating by program.

Thirteenth Display
First line shows Function code 10
Second line shows totalized flow (changing all the time), you can modify to any value you want
(unit is same as totalized flow), 2 decimals.
Press “Ent”save and press “Esc” to quit.

7
http://www.silverinstruments.com

Fourteenth Display: display the first point of flow meter factor compensation “P1”.
The first line shows first point frequency, no decimals
The second line shows first point factor compensation with 2 decimals.
Frequency point (1-3000) and compensation factor (flow meter factor is the same as factor
setting) are needed to be set at the same time.

Fifteenth Display: display the second point of flow meter factor compensation “P2”
The first line shows second point frequency, no decimals
The second line shows second point factor compensation with 2 decimals.
The displaying format is the same as above.

Sixteenth Display: display the third point of flow meter factor compensation “P3”
The first line shows third point frequency, no decimals
The second line shows third point factor compensation with 2 decimals.
The displaying format is the same as above.

Seventeenth Display: display the fourth point of flow meter factor compensation “P4”
The first line shows fourth point frequency, no decimals
The second line shows fourth point factor compensation with 2 decimals.
The displaying format is the same as above. Liquid turbine flow meter is a kind of diesel flow
meter

Eighteenth Display: display the fifth point of flow meter factor compensation “P5”
The first line shows fifth point frequency, no decimals
The second line shows fifth point factor compensation with 2 decimals.
The displaying format is the same as above.

Nineteenth Display: display the average flow meter factor


The first line shows the modified frequency with one decimal.
The second line shows average factor with two decimals.

8