Mark V and Mark V Plus Injection Systems Controt Unit Disassembly 20-2 Control Unit Tools Required: + Medium-blade screwdriver + 9/64 (3.5 mm) Hex Key Pedestal System: 1. Remove power from the injector. 2. Loosen and remove the two screws at the rear of the Consoie. 8. Ifthe injector does not have ECG trigger and scope, continue with step 4, NOTE: Ifthe injector has ECG trigger and scope, remove the four screws retaining the bottom cover of the main Control Panel, then remove the cover. Disconnect the ribbon cable from underneath the chassis. 4. Grasp the front bezel and pull it forward. 5. Ifthe unit is a 20000 series Mark V Injector, skip to step 9. Ifthe unit is above series 60000, proceed with step 6. 6. Remove the PC cards. 7. Disconnect the two (2) hamess connectors at the Motherboard. 8. Ifthe unit contains the ECG/Scope option, remove the connector from J44. 9. Disconnect harness mount clamp and ground wire. 10. Slide the chassis out of the case. Control Room Console: 1. Remove power from the injector. 2. Loosen the two screws (either slotted or 9/64 (3.5 mm) hex key ‘screws) at the top of the Control Panel. 3. Tilt the Control Panel downward. Disassembly and Reassembly Control Unit Reassembly Control Unit - (continued) Pedestal System: 1. Slide the chassis into the case, aligning the chassis with the sides. of the case. 2. Ifthe unit is a 20000 series Mark V Injector, skip to step 6. If the unit is above series 60000, proceed with step 3. 3. Connect the chassis hamess connectors to the Motherboard. 4. Connect the ground wire and J44 connector (if unit contains ECG/ Scope option). 5. Fasten hamess mounting clamp, and install the PC cards. 6. Push the chassis in until the rear of the bezel is flush with the front of the case. 7. Install and tighten the two screws at the rear of the Console. 8. If the injector contains the ECG/Scope option, install the ribbon cable connecting the Control Panel and the chassis. Replace the bottom cover on the Control Panel and retain with the four screws previously removed. Control Room Console: 1. Tilt the Control Panel upward, and close. 2. Tighten the two screws (either slotted or hex key) at the top of the Control Panel. 20-3 Mark V and Mark V Plus Injection Systems PC Card Removal PC Card Replacement Control Unit- Pe cards 1. (1. GRC units only) Remove the chassis shield plate by removing the two thumbscrews. 2. In Pedestal Systems, remove the card retainer from across the top ofthe cards. In a Control Room Console, remove the card retainer brackets from across the edges of the cards. 8. Ifthe CPU or /O Card is being removed, disconnect the ribbon cable connector from the VO Card. 4. Lift the inside edge of the ejector tabs on the card being removed, then slide the card out. 1. Identity the proper slot for reinsertion it more than one card was removed. Card Ejector Tab color CPU Card White VO Card Violet Servo Control Card Green PPI Card Yellow ISI Card (or dummy card) Red (none on dummy card) 2. Align the card with the proper PC card guides. + When facing the front of a Pedestal System, components are located on the right side of the PC cards. + Ina Control Room Console, card components face upward. 3. Insert the card into its respective slot, and align card edge connec- tor fingers with connectors on Motherboard. Push the card firmly to ensure that itis securely seated in the Motherboard. 4. Ifthe CPU or VO Cards are being replaced, fully insert the /O rib bon cable into connector. 5. In Pedestal Systems, reposition the card retainer across the top of the cards. In Control Room Console systems, reposition the card retainers across the front edges of the cards. 6. (CRC units only) Replace and secure the shield plate with the ‘two thumbscrews previously removed.

Das könnte Ihnen auch gefallen