VISIBLE and UV - VIS SPECTROPHOTOMETER
USER'S IUIAHUAL
{
working Environment and lnstalration Requirements
t Rated voltage is 94V-240V AC, 50 / 60Hz*1H2.
O Rated voltage is 110 / 2z0v+10% AC,60 /50H2*1Hz (for model UV-vis cnly).
O lnstrument should be put in a dry room with temperature at 5'C-35'C, and relative humidity
is not than 85 %.
I The instrument should be placed on the soiid stable working platform, no strong
vibration or continuous vibration.
{l The room should no hydrogen sulfide, nii.ess acid fluoride and other eorrosive gasses.
O The instrument should stay awav from higr-intensitv magnetic field, electric field and
produced high-frequency \^/ave device.
O The instrument should avoid a direct hit by Sl-i-:3 wifld.
t The instrument should avoid strcnq sunshine.
I Grounding well is necessary.
t Warm up the host at least 3Onrins aiie. aci,,rer on.
t Standard Configuration
1) Spectrophotometer
2) Power cable
3 ) User's manual & packing list
Vis UV-Vis
Display LCD
Wavelength Range (nm)
340-1000 330-1000 I 325-1000 320- 1 000 200-1 000
tungsten-halogen
Lamp tungsten-halogen
deuterium
Detector
Silicon Photocell
Spectral Bandwidth 4nm 2nm 4nm 2nm
I Model -
16 17
r Steps
Linear Accuracy.
Note: Effective linear concentration range is 1-1999, pls dilution samples if it is out of range.
ffi
+ Remove 4pcs screws on the left and right side of instrument cover
r Open instrument cover (care for connection cable)
+ Press the spring which fixed halogen{ungsten lamp lightly, pulithe hair:g*n irlrG$!r.,
ijliij;r.
out of the seat.
t lnsert the new halogen-tungsten lamp into the seat, and fix it with the spring.
o Watch the light spot emitted by the halogen-tungsten iar-np on the entrance slit,.check
whetherthe position is correct. lf ihe value on LCD lTrans) is unstable, adjustthe position
of lamp slight until the reading is stable.
ffiffi
e Remove 4pcs screws on the left and right side of instrument cover
+ Open instrument cover (care for connection mble)
+ Loosen three color wires of deuterium lamp by screwdriver (pls remember connected positions of
color wires, this is very important ! ).
+ Take away deuterium lamp (together with the lamp holder) from the seat (this can be done only
by unscrewing the two bolts in the lamp holder), and loosen the clamp bolt of deuterium lamp.
+ Put on the new lamp, fix the deuterium lamp (together with the deuterium lamp holder) on the
seat, connect the colorwires (wires should be connected as before)
Caution: pls connect color wires just as before!
+ Adjust the deuterium lamp hole around 55mm away from the bottom plate, the position of light
hole is toward the spherical reflection minor in the lamp chamber
. Power on, keep lamp selecting switch at deuterium side.
+ Watch the light spot emitted by the deuterium lamp on the entrance siit, to check whether its
position is correct. lf the reading (Trans value) is unstable, readjust the position of the
deuterium lamp in a slight way until the reading is stable, and then tighten the clamp bolt.