Sie sind auf Seite 1von 20

Comparison of Large-Core

Vacuum-Assisted Breast 2
Biopsy and Excision Systems

Robin Wilson and Sanjay Kavia

expert centers. Even then, false-negative results


2.1 for sampling in situ disease represented by
Introduction microcalcifications and for certain types of
invasive breast cancer were disappointing. For
The past 30 years have seen dramatic changes in this reason, in the 1990s there was a trend away
the diagnosis and management of breast prob- from the use of FNAC to needle core biopsy
lems. Much of this change has been driven by techniques. Core biopsy provides histological
the quest for early diagnosis and by the wide- material for morphological as well as cellular
spread use of imaging in the diagnosis of symp- assessment and the skills required for interpreta-
tomatic breast disease and for breast cancer tion are much more widely available. The sensi-
screening. The traditional approach of surgical tivity for the more elusive disease, such as
biopsy for diagnosis has now been replaced by lobular invasive carcinoma, is also significantly
needle biopsy techniques that provide both better. In addition, when sampling microcalcifi-
accurate and reliable diagnosis. The aims are to cations, core samples can be imaged to prove
prevent unnecessary surgery for benign proc- retrieval of representative tissue. Overall, com-
esses and to provide detailed information on pared to cytology, core biopsy histology pro-
borderline and malignant processes that allow vides significantly better sensitivity and positive
for prospective fully informed treatment plan- predictive values for both benign and malignant
ning (Teh et al. 1998). To achieve these aims, disease and has a reduced rate of false-negative
specialized techniques for needle biopsy of the results. Automated core biopsy is now accepted
breast have been developed that facilitate the as the preferred technique for breast tissue
removal of sufficient amounts of tissue required sampling (Bassett et al. 1997; Teh et al. 1998;
to ensure accurate diagnoses. Litherland 2001; Parker et al. 1996; Schueller
In the 1980s, the predominant technique for et al. 2008).
needle sampling of the breast was fine-needle Using core biopsy, the vast majority of breast
aspiration for cytology (FNAC). The results abnormalities can be accurately diagnosed and
achieved with FNAC were encouraging but the most breast lesions are amenable to ultrasound-
required reliability and acceptable sensitivity guided biopsy (Philpotts et al. 2003). However,
and specificity proved to be only achievable in particularly for screen-detected impalpable
lesions, there remain around 5–10% of abnor-
malities where core biopsy either does not pro-
vide sufficient material for accurate diagnosis
Robin Wilson ()
King’s College Hospital or does not target the lesion accurately (Parker
Denmark Hill, London SE5 9RS UK and Burbank 1996). For these abnormalities,
e-mail: robinwilson@nhs.net either more tissue or more accurate targeting is

Renzo Brun del Re (Ed.), Minimally Invasive Breast Biopsies, Recent Results in Cancer Research 173, 23
Doi: 10.1007/978-3-540-31611-4_2, © Springer-Verlag Berlin Heidelberg 2009
24 R. Wilson and S. Kavia

necessary to achieve the tissue needed for reli- stereotactic (upright and prone table), and MR
able histopathological assessment (Kettritz et imaging guidance; the single intact biopsy sys-
2 al. 2003). With these two factors in mind, in the tem is described as being suitable for ultrasound
early 1990s techniques were developed to pro- and X-ray stereotactic-guided biopsy. All the
vide both directional sampling capability and systems are suitable for use in the out-patient
retrieval of larger volumes of tissue (Parker setting with local anesthesia.
et al. 1994; Burbank 1993). These have been
further developed and refined over the past 15
years and are now in routine use for both diag- 2.2.1
nosis and therapeutic excision. Single Large-Core Biopsy System
Two different approaches for large-core
biopsy have been developed: multiple contigu- The Intact Breast Lesion Excision System
ous large-bore cores retrieved with the assistance (BLES; Intact Medical Corporation Inc.) is a
of suction (vacuum-assisted mammotomy; VAM) breast excision system that combines the use of
and single very-large-bore core (SLCB). Both vacuum and radiofrequency (RF) cutting to
methods can be used under X-ray (stereotactic remove the targeted lesion as a single specimen
guidance) and ultrasound, but currently only (Intact Medical Corporation 2008). The probe
VAM is recommended for magnetic resonance (or wand) (Fig. 2.1) is available in four sizes,
(MR)-guided biopsy. With these techniques, it is designed to retrieve specimens that are 10, 12,
now possible to retrieve sufficient tissue using 15, and 20 mm in diameter (Fig. 2.2). The
image-guided biopsy such that 98–99% accuracy Intact BLES probe is positioned under imaging
of nonsurgical diagnosis can be achieved. In fact, guidance (ultrasound or X-ray stereotaxis)
so much tissue can be removed that these tech- through a 6- to 8-mm skin incision and
niques are now being used for total excision of advanced to the periphery of the area to be
certain breast abnormalities. Compared to core excised. A cutting RF wire is activated and
biopsy, VAM and very-large-core biopsy reduce advanced to cut and ensnare the target lesion
by half understaging of pathology (atypical by means of four insulated struts that first
hyperplasia and DCIS), on average from 20 to expand and then contract to surround the lesion
10%. This means that repeat biopsy and further (Fig. 2.3). The single large sample is then
surgery for diagnosis and treatment are required withdrawn intact through the same tract.
half as often (Liberman et al. 2000). Vacuum- Vacuum is used to minimize the extent of the
assisted biopsy is the technique of first choice RF effect on the sample excised and into the
for MR-guided breast biopsy (Liberman et al. surrounding breast tissue and to extract any
2005; Kuhl 2007). bleeding that may occur during the procedure.
The Intact BLES system is said to have the
advantage over VAM of retaining the full
histological architecture and potentially clear
2.2 margins around the area of interest and with
Large-Core Biopsy Systems: Overview little RF artifact on histology (Sie et al. 2006).
It has also been reported to be associated with
Currently one single large-core radiofrequency reduced understaging compared to VAM (Sie
biopsy system and four vacuum-assisted multi- et al. 2006; Killebrew and Oneson 2006).
ple-core biopsy systems are in routine use for However, the RF function does limit its use for
breast diagnosis and excision. All of the VAM lesions close to the skin or the chest wall and
systems are suitable for use under ultrasound, for lesions in small breasts.
2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 25

Fig. 2.1  The Intact disposable wand (probe) and driver

Fig. 2.2  Intact whole-tissue samples showing the size of samples achieved with the 10-, 12-, 15-, and
20-mm wands

Fig. 2.3  Diagram of how the Intact system operates


26 R. Wilson and S. Kavia

2.2.2 rotation of the needle within the driver. The


Vacuum-Assisted Mammotomy Systems EnCor system has fully automated and program-
2 mable directional functions. Each system is
There are four systems in routine use based on described here in greater detail.
the principle of single or multiple cores using 14
to 7 French gauge needles or probes. Although
the principles of operation are similar for all of 2.2.2.1
these VAM systems, there are significant differ- Mammotome System
ences in their design, method of operation, and
attributes. Table 2.1 shows a comparison of the Mammotome (Breast Care, Ethicon ­Endo-
various attributes of these systems. All of these Surgery), the first VAM system to be developed
systems are directional, allowing sampling and originally marketed under the Biopsys name,
around a full 360-degree arc either by manual has been available since 1995, and has been
rotation of the driver or manual or automated upgraded several times since. It is a ­well-proven

Table 2.1  Comparison of VAM systems

Attribute Mammotome Vacora Atec EnCor


Drivers required Separate for US, Same for all Same for all Separate for MRI
X-ray, and MRI
Command unit Same for all Self-contained Various options Same for all
Vacuum adjustment No No Requires different Yes
units
Manual vacuum +++ No + (With lavage) +++
control
Needle gauge 11 and 8 14 and 10 12 and 9 10 and 7
Multiple core retrieval Yes No Yes Yes
Cutting method Rotating Rotating Rotating Scissor
Needle sharpness ++ + ++ +++
Core sample size ++ ++ ++ +++
Volume of tissue per ++ + +++ +++
minute
Open or closed tissue Open Open Closed Closed
collection
Smaller chamber size No No Requires different Selectable with the
Choice probe same probes
Speed of tissue ++ + +++ +++
retrieval
Needle rotation Manual only Manual only Manual only Manual or
automated
Lavage No No Full Sample chamber
Programmable ++ No No +++
functions
Biopsy site marker Yes No Yes Yes
system
Local anesthetic ++ No ++ +++
function
Probe offset Yes Yes No Yes
2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 27

a b

Fig. 2.4  Mammotome biopsy probes (a) ultrasound EX system and (b) stereotactic ST system

device with more than 3 million biopsy procedures


performed worldwide and more than 200 papers
reporting various aspects of its use published in the
medical literature.
This system uses three different drivers for Fig. 2.5  Close-up view of the Mammotome probe
tip showing the double lumen system with fenes-
stereotactic, ultrasound, and MR-guided biopsy
trations used to suck in the samples for cutting by
(Fig. 2.4a, b). All three drivers use the same
the rotating inner trocar. A scalpel-type blade is
unique double-lumen probe, one lumen used for embedded in the tip to ease insertion of the probe
providing the suction (at 23–25 mmHg) that
draws the sample into the biopsy chamber and
the other through which the internal rotating procedure to extract any bleeding or hematoma.
cutting trocar moves to cut the sample and for Additional local anesthetic can be delivered
retrieval of the specimen (Fig. 2.5). All three through a port in the vacuum tubing during a
drivers are controlled by the same command biopsy. The needle bore and the biopsy can also
module (Fig. 2.6) that can be programmed to have lavage applied if blockage occurs by injec-
individual preference for automated or semi- tion of saline through the same vacuum tubing
automated function with variable pause periods port.
between each sample, application of suction For stereotactic biopsy, the drive for the
sequences, and “clear” mode for “dry tap” rotation and retraction of the cutting trocar is
events (Fig. 2.7). The control module can be via torsion cables driven from the command
operated via a foot switch or a handheld control- unit, while for MR- and ultrasound-guided
ler for sterotactic and MR-guided biopsy and via procedures, the drive for the cutter is incor-
buttons on the driver itself for MR- and ultra- porated into the handheld device, allowing
sound-guided procedures (Fig. 2.4). The com- for more flexible use (Fig. 2.5). The stereo-
mand module provides the suction via plastic tactic driver incorporates a spring-loaded
tubing for all three uses and provides real-time system that allows the needle to be fired for-
pictorial feedback of all the functions, including ward 20 mm into the breast while it is held in
the position of the cutting trocar and where in the stereotactic holder (Fig. 2.8). In all appli-
the system suction is being applied. Suction can cations, the mammotome is an open biopsy
be applied manually at any time during a biopsy system and each core sample must be retrieved
28 R. Wilson and S. Kavia

Fig. 2.6  The Encor (left), Atec (middle), and Mammotome (right) control modules

Fig. 2.7  The Mammotome control module monitor showing the operating functions

from the sample collection chamber after scalpel embedded into the probe tip (Fig.
each biopsy (Fig. 2.9). The probes are dispos- 2.5). The 11-G probe provides approximately
able and are available in two sizes (11 and 8 100 mg and the 7-G probe approximately 175
G) for ultrasound, stereotactic, and MR use mg of tissue per core sample. Special part-
and are available with or without a cutting ceramic probes, guides, and introducers are
2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 29

a b

Fig. 2.8  The Mammotome ST system a showing the ital stereotactic system (GE Medical Systems) with
set-up for use with a lateral arm and b for biopsy of the patient in the lateral decubitus position
the inferior part of the breast using an upright dig-

Fig. 2.9  The Mammotome ST system in use for biopsy of the upper breast in the craniocaudal position
using an upright stereotactic system (GE Medical Systems) showing manual retrieval of a core sample

available to facilitate MR-guided biopsy 2.2.2.2


using standard lateral grid MR biopsy coils. Vacora
Various types of ultrasound and X-ray-visible
biopsy site markers are available for use with the The Vacora system (C.R. Bard Inc.) was the sec-
Mammotome probes, which are inserted through ond device to become available for VAM (origi-
the probe directly into the biopsy cavity. nally marketed as the BIP VacuFlash). It is
30 R. Wilson and S. Kavia

a b
2

Fig. 2.10  The Bard Vacora vacuum biopsy system: a driver, b the system ready for use, and c a close-up
the handheld device showing insertion of the nee- view of the operating panel
dle and the self-contained vacuum system into the

unique in that it is a self-contained system incor-


porating the driver for the rotating cutting trocar
and the suction unit within the handheld unit
(Fig. 2.10a, b). The disposable probes are single-
lumen and contain a rotating cutting trocar. The
probes are available in 14- and 10-G sizes, pro-
viding approximately 50 and 150 mg of tissue
per core sample, respectively (Fig. 2.11). The
needle is directional but needs to be rotated
manually in the holder to achieve different
biopsy directions for stereotactic biopsy; for
ultrasound- and MR-guided biopsy, the com-
plete holder can be rotated for multidirectional
sampling (Fig. 2.12). The Vacora is a single-
sample system and the probe must be with-
drawn from the breast to retrieve each core
sample. For this reason a plastic guiding canula Fig. 2.11  Comparative core sample sizes achieved
is best inserted into the breast first to facilitate with 14-, 10-, and 7-G vacuum core needles
repeated insertion of the probe to the same site
in the breast. This single-sample function means extracted. It is not possible to apply manual
that it is not suitable for therapeutic excision of suction, and this limits its use when bleeding
anything other than small lesions (10 mm or occurs at the biopsy site. Additional local
less). The suction is only applied while the core anesthetic injection and biopsy site marking
is being taken and again when each core is must be delivered through the guiding plastic
2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 31

Fig. 2.12  The Bard Vacora system in use for MRI-guided breast biopsy. The black line on the back of the
device indicates the direction of sampling

canula. The system incorporates a spring-loaded able in 9- and 12-cm lengths, both with 20-mm
firing mechanism that allows the needle to be sampling chambers. The larger 9-G needles are
fired forward into the breast. The system is available in 9-, 12-, and 14-cm lengths with
simple to use because its functions are all preset the two shorter lengths available with either
and are activated using the control panel on the 20- or 12-mm sampling chambers (Fig. 2.14).
side of the device; changes to its operation The sampling processes are preset and not
cannot be programmed by the user. programmable by the user. All handpieces include
a closed sample collection system (Fig. 2.14).
There are two operation modes that are used
2.2.2.3 for all methods of image guidance (Fig. 2.15). In
ATEC (Automatic Tissue Extraction and Collection) the lavage mode the sample chamber is open
and saline is continuously instilled through the
The Automatic Tissue Extraction and Collection system into the biopsy area and through the
(ATEC) system (Suros Inc.) is similar in its sampling filter (Fig. 2.15). The ATEC is the only
function to the Mammotome ST system in that it system that uses lavage of the biopsy area.
is driven by cables from the command unit and Manual suction can also be applied (Fig. 2.15).
uses an internal rotating cutter. However, it is a In the biopsy mode, the sample chamber is
single-bore system and the whole driver unit is closed in the resting position until the foot pedal
disposable (Fig. 2.13). There are three different is used to trigger multiple rapid retrievals of
command modules available that deliver differing samples (averaging 150–175 mg per core). The
suction levels depending or required usage (MR ATEC is the fastest-acting VAM system,
only, ultrasound and stereotactic only, and all although it does not deliver more tissue per time
three). The disposable handpieces are available unit than the EnCor system (see below).
in two needle sizes in several lengths and sampling Directional sampling is achieved by rotating the
chamber sizes. Twelve-gauge needles are avail- handpiece manually (Fig. 2.16). The needle of
32 R. Wilson and S. Kavia

a b
2

Fig. 2.13  The Suros ATEC biopsy device a showing the component parts with the disposable driver and
detached closed sampling chamber and b close-up of the closed sampling chamber in place

for dense tissue and more prolonged use. The


driver contains the mechanisms for driving the
cutting device and for rotating the needle (Fig.
2.17). The driver has an electrical cable and plas-
tic vacuum tubing that connects it to the com-
mand module. The same command module is
used for ultrasound-, X-ray stereotactic-, and
MR-guided biopsy (Fig. 2.6). The functions are
fully programmable by the user, including auto-
mated rotation of the biopsy sampling chamber,
which can be set for single, three (180°), six
Fig. 2.14  The Suros ATEC needles showing the 9-
(270°), and eight (360°) rotations in any direction
and 14-cm lengths (Fig. 2.18). The automated function of the probe
rotation is particularly useful for stereotactic and
the Suros ATEC system is not offset and this MR-guided biopsy, as the sample selection direc-
means that there may be restrictions on its use tion and extent of rotation can be preset. The
under stereotactic and MR guidance for lesions driver unit is lightweight and handheld for ultra-
close to the chest wall and in breasts with a small sound-guided procedures (Fig. 2.19). The same
compression thickness. driver is attached to a holder for either prone
table or upright stereotactic use (Fig. 2.20). The
stereotactic probe holder has a spring-loaded
2.2.2.4 mechanism that allows the driver and probe to be
EnCor fired forward 20 mm into the breast.
The needle probes are available in 10- and
The EnCor system (Senorx Corporation), like the 7-G sizes. Both have the patented tri-concave tip
Vacora and ATEC, is a single-bore VAM system. design that renders the probe extremely sharp so
However, it differs from the other VAM systems that it passes through all but the most dense
in many ways. The inner cutting trocar, rather breast tissue with ease (Fig. 2.21). The 7-G probe
than rotating to cut the sample, uses a scissor provides the largest samples of all the VAM
oscillating action that is said to be more effective systems at 300 mg per core sample. The sample
2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 33

Fig. 2.15  The Suros ATEC control system

Fig. 2.16  The Suros ATEC system in use for ultrasound-guided biopsy

chamber length (10 or 20 mm) for the two needle larly dense tissue is encountered. The module
sizes can be set at the control module and avoids also has a preset anesthetic function that allows
the need to select a different needle if a short for delivery of local anesthetic 360° around the
core length is required. The strength of the vac- biopsy site either before or during the biopsy
uum applied can also be doubled when particu- procedure. The same system is used to deploy
34 R. Wilson and S. Kavia

a b
2

Fig. 2.17  The EnCor biopsy system: a the driver


unit and disposable needle, b close-up of the
closed sample retrieval component, in place and
c detached

Fig. 2.18  The Senorx EnCor control display


2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 35

Fig. 2.19  The Suros EnCor system set up and ready for handheld use

Fig. 2.21  Close-up of the Tri-concave EnCor needle


tip

can also be used for sample radiography. There


is an optional lavage system that uses saline to
cleanse the tissue samples in the collection
chamber as they are retrieved. Like the
Mammotome and ATEC systems, the EnCor
device can be used with MRI-specific introducers
and trocars (Fig. 2.23a, b)

2.3
Fig. 2.20  The Suros EnCor system in place for ster- Indications and Limitations
eotactic-guided biopsy using a prone table
There are a number of diagnostic and therapeu-
gel and clip markers at the biopsy site (Fig. 2.22). tic situations where VAM or SLCB should be
The closed sample collection system retrieves considered as the primary technique. These
the samples in an easily removable basket that include:
36 R. Wilson and S. Kavia

a a
2

Fig. 2.23  The EnCor system: a MRI trocar and can-


Fig. 2.22  The Encor a anesthetic control display and
ula guides and b in use for MRI-guided biopsy
b method of injection of anesthetic through the
probe system to the biopsy site

Diagnostic biopsy: – Radial scar/complex sclerosing lesion


– Sentinel lymph node
– Equivocal or failed core biopsy
– Small lesion (sub 5 mm)
– Architectural distortion
– Clustered microcalcifications 2.3.1
– Diffuse nonspecific abnormality Limitations
– Complex cyst
– Intraductal lesion The various differences in the attributes of the
– Abscess drainage large-core systems mean that some are not
suited for all of the possible indications. The
Therapeutic excision:
Vacora system has the most limitations because
– Fibroadenoma and other biopsy-proven it is a single-core device that needs to be
benign lesions removed from the breast to retrieve each sample
– Papillary and mucocele-like lesions and cannot be used to aspirate the biopsy area.
2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 37

2.3.2 specific changes on imaging. All of the systems


Diagnostic Biopsy described here are potentially suitable for these
indications, as previously described. The Intact
When conventional automated core biopsy has system is not suitable for superficial lesions and
either failed to target the lesion or there is a bor- the Vacora system is not ideal if more than
derline pathological result, VAM or SLCB will 10–15 cores are likely to be necessary.
usually provide the material required to either Similarly, the Intact and Vacora systems are not
avoid unnecessary surgery for benign lesions or ideal for sampling complex cysts, particularly
facilitate single-stage therapeutic surgery for those that appear to contain a mass component.
malignant disease. In certain circumstances These are best biopsied by VAM devices where
where there is a high chance of failed sampling the probe remains in position in the breast
with core biopsy, such as small clusters of suspi- throughout the procedure and manual vacuum can
cious microcalcifications and very small mass be applied. The same is true for removal of an
lesions, it is usually better to use VAM or SLCB intraductal lesion when manual vacuum is an
as the primary sampling technique without important factor, as is the ability to move the sam-
attempting core biopsy first (Fig. 2.24a–c). pling chamber around the area without removing
The same is true for circumstances where the probe from the breast (Fig. 2.25a, b).
from the outset it is recognized that larger vol- VAM systems are most widely used for
umes of tissue will be required for diagnosis, stereotactic biopsy of microcalcifications iden-
such as suspicion of radial scar and diffuse non- tified on mammography, and all of those

a b

Fig. 2.24  Stereotactic procedure radiograph showing a an EnCor probe in place for biopsy of calcifica-
tions, b core specimens in the sample retrieval tray, and c radiography showing calcifications success-
fully sampled
38 R. Wilson and S. Kavia

a particularly because equivocal pathology results


and understaging of disease are much less fre-
2 quent, making repeat procedures and the need
for surgical biopsy much less common. Some
clinicians prefer to map the samples retrieved in
order to document where each sample has come
from in the breast. It is not clear what the benefit
of doing this is because it does not affect the
b subsequent management of the result and imag-
ing provides the information needed if the first
sets of samples do not contain sufficient mate-
rial. If sample mapping is required the Vacora
and Mammotome systems need to be used. For
sampling lesions close to the chest wall or when
using the lateral approach in women with small
breasts, access to the breast requires a probe
with the needle offset to avoid snagging on the
biopsy table or the chest wall. The ATEC system
does not have an offset needle.
Aspiration of breast abscess is preferred to
surgical drainage and is usually achieved by
manual suction applied through a standard nee-
Fig. 2.25  Ultrasound images showing a an intraduct dle. However, VAM systems with continuous
lesion and b an intracystic lesion, which are ideal suction capability provide the means of dealing
for VAM excision
with larger and more organized breast abscesses
when surgery would otherwise be required. The
described here are suitable for this purpose. All Vacora is not suitable for this use.
of the upright and prone biopsy mammography
systems can accommodate the VAM and Intact
systems described here (Georgian-Smith et al. 2.3.3
2002). VAM is particularly suited for this situation, Therapeutic Excision
as the vacuum assistance means that sampling is
directional and can be used to retrieve tissue at a VAM and SLCB provide an alternative to surgery
distance from the actual site of the needle. For for the removal of known benign lesions such as
core biopsy, the needle must pass through the fibroadenomas and focal fibrous lesions (Tennant
lesion for successful sampling, but with VAM et al. 2008). Large-core techniques are also being
the probe only needs to be placed close to the increasingly used instead of surgery to widely
target area. The vacuum effect is used to pull sample borderline lesions such as radial scars and
the tissue into the sampling chamber. This is papillary lesions shown on previous core biopsy
particularly useful for lesions close to the chest to have no evidence of epithelial atypia (Rosen et
wall, and behind the nipple and where the al. 2002; Carder et al. 2008). The Vacora system
cluster is more scattered. In many units now, all is only suitable for removing small lesions less
stereotactic biopsy procedures are carried out than 10 mm in diameter for the reasons outlined
using VAM systems because the retrieval results above. Similarly, the Intact system is limited by
are significantly better than core biopsy, and the size of its retrieval system, which has a maxi-
2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 39

mum diameter of 20 mm. The Mammotome (7 breast throughout the biopsy procedure and are
G), ATEC (9 G), and EnCor (7 G) are all ideal for therefore easier to use in this situation. All of the
excision of large lesions. The Mammotome sys- systems are light enough to be easily handheld
tem takes longer to complete the task simply for ultrasound-guided use. The sharpness of the
because each sample has to be retrieved from the tri-concave tip of the Encor system means that it
sample chamber while with the other two the is more easily advanced through the breast to the
samples are automatically collected by their target than the other systems and is more easily
closed sampling systems. Many clinicians restrict sited in the ideal position for ultrasound-guided
the size of lesion they are willing to attempt to VAM immediately behind the lesion (Fig. 2.21).
remove with VAM to around 20–30 mm. Both the This is particularly apparent in the dense and
EnCor and ATEC systems will retrieve more than fibrous breast.
1 g of tissue per minute and can be readily used to
excise lesions up to 50–60 mm in diameter.
There are also some reports of these tech- 2.3.4
niques being used to sample sentinel nodes prior Image-Guided Biopsy Technique
to surgery or primary chemotherapy treatment.
However, this indication must be considered Anyone familiar with the technique for image-
experimental at present. Similarly, SLCB and guided fine-needle aspiration and conventional
VAM should not be used for excision of known core biopsy will be able to adapt easily to the
malignant lesions or borderline lesions associ- technique required for large-core biopsy because
ated with a significant risk of breast cancer the basic principles are the same.
except in exceptional circumstances (Eby et al. Particular attention must be paid to adminis-
2008; Lee et al. 2008). In some patients who are tration of sufficient local anesthetic for these
not medically fit for conventional treatment large-bore procedures. Significantly larger
(surgery and chemotherapy), SLCB and VAM doses are usually required than for conventional
can be considered. To ensure a reasonable exci- core biopsy, particularly for excision proce-
sion margin, the Intact system should be con- dures and procedures done under ultrasound
fined to lesions no more than 10–25 mm in guidance. The larger size and vacuum assist-
diameter if a clear excision margin is to be ance both mean that vessel damage is more
achieved. The VAM system can be used for likely with these techniques. The use of local
larger lesions but it must be recognized that the aesthetic combined with adrenaline is preferred
excision margins will be suspect, even if sample because this reduces the chances of hematoma
mapping is used. and increases the time that the anesthesia is
As with all breast biopsies, ultrasound guid- effective. Plain local anesthetic may be pre-
ance is the preferred guidance method whenever ferred for the skin and subcutaneous tissues in
possible. All of the systems described here can older patients and those with compromised
be used for ultrasound-guided biopsy. The skin. For stereotactic X-ray-guided procedures,
Vacora system is usually used with a trocar. For care should be taken not to inject large volumes
stereotactic biopsy, this is satisfactory because of anesthetic since this can significantly displace
the breast is fixed by compression. For ultra- the target area. The biopsy-targeting process is
sound, the trocar guide system can be less effec- the same as for core biopsy with the aim of
tive, particularly in the large breast; since the passing the probe directly through the area to
breast is not fixed, it can be difficult to reintro- be sampled and then biopsy around 360°.
duce the needle to the same site for successive However, unlike core biopsy, successful sampling
biopsies. The other VAM systems remain in the can be achieved if the lesion is not transfixed
40 R. Wilson and S. Kavia

using the vacuum and directional capabilities of provide the means for minimally invasive ther-
the VAM systems. apeutic lesion excision of benign and border-
2 For ultrasound-guided sampling, the anes- line lesions. Accuracy approaching 99% can be
thetic must be infiltrated to surround the lesion achieved, thus avoiding the need for diagnostic
being targeted and can be used to dissect the tis- surgical open biopsy in the vast majority of
sue down to the lesion and to assist in separating cases and providing tissue samples in quanti-
the target from the deep and superficial tissues. ties sufficient to allow for detailed treatment
Deep tissue anesthesia is particularly important, planning. The choice of vacuum biopsy system
as for ultrasound-guided sampling the VAM will depend on workload, the image guidance
probe is best positioned behind the lesion. Some methods that are used, and whether lesion exci-
also advocate the use of longer-acting local sion is required.
anesthetic in the deeper tissues around the target
lesion to reduce postprocedure anesthesia.
All but the Vacora system allow for further
References
injection of local anesthetic through the biopsy
probe into the target area (Fig. 2.22b). This can Bassett L, Winchester DP, Caplan RB et al (1997)
be done as a matter of routine before the biopsy Stereotactic core needle biopsy of the breast: a
is commenced, particularly for stereotactic pro- report of the joint task force of the American
cedures when the probe has been placed and dis- College of Surgeons and College of American
placement of the lesion is then less likely to Pathologists. CA Cancer J Clin 47:171
Burbank F (1993) Stereotactic breast biopsy: com-
occur. The EnCor device has a specific program parison of 14- and 11-guage mammotome probe.
for delivering local anesthetic around the area to Acta Cytol 37:461–471
be biopsied. Carder PJ, Khan T, Burrows P, Sharma N (2008)
MR-guided biopsy can be achieved with all Large volume mammotome biopsy may reduce
of the VAM systems described (Figs. 2.12 and the need for diagnostic surgery in papillary
lesions of the breast. J Clin Pathol 61:928–933
2.23b). All of the manufacturers provide MR Eby PR, Ochsner JE, DeMartini WB, Allison KH,
biopsy packs that contain the necessary materi- Peacock S, Lehman CD (2008) Is surgical exci-
als to carry out the procedure using most of the sion necessary for focal atypical ductal hyperpla-
currently available MR biopsy coil systems sia found at stereotactic vacuum-assisted biopsy.
(Fig. 2.23a). VAM is recommended for all Annu Surg Oncol 15;3232–3238
Georgian-Smith D, D’Orsi C, Morris E, Clark CF,
MR-guided biopsies; these lesions are only vis- Liberty E, Lehman CD (2002) Stereotactic
ible on MR and therefore the method most biopsy of the breast using an upright unit, a
likely to successfully retrieve tissue from the vacuum suction needle, and a lateral arm sup-
targeted area should be used (Lee et al. 2008). port. AJR 178:1017–1024
Intact Medical Corporation (2008) Percutaneous
contiguous electrosurgical breast biopsy devices.
http://www.intactmedical.com/pdf/ML087_
Rev00.pdf
Kellebrew LK, Oneson RH (2006) Comparison of
2.4 the diagnostic accuracy of a vacuum-assisted
Conclusions percutaneous intact specimen sampling device to
a vacuum-assisted core needle sampling device
There are a number of well-designed devices for breast biopsy: initial experience. Breast
12:302–308
available for vacuum biopsy and excision Kettritz U et al (2003) Stereotactic vacuum-assisted
biopsy. These devices enable the radiologist to breast biopsy in 2874 patients: a multicentre
deliver a high level of diagnostic accuracy and study. Cancer 100:245–251
2  Comparison of Large-Core Vacuum-Assisted Breast Biopsy and Excision Systems 41

Kuhl C (2007) The current status of breast MR imag- Parker SH, Dennis MA, Stavros AT, Johnson KK
ing. Part 1. Choice of technique, image interpre- (2006) A new breast biopsy technique. J Diagn
tation, diagnostic accuracy, and transfer of Med Sonogr 12:113–118
clinical practice. Radiology 244:356–378 Philpotts LE, Hooley RJ, Lee CH (2003) Comparison
Lee K-M, Kaplan JB, Murray MP, Liberman L of automated versus vacuum-assisted methods
(2008) Complete excision of the MRI target for sonographically guided core biopsy of the
lesion at MRI-guided vacuum-assisted biopsy of breast. AJR 180:347–351
breast cancer. AJR 191:1198–1202 Rosen EL, Bentley RC, Baker JA, Soo MS (2002)
Liberman L (2000) Percutaneous image-guided core Imaging-guided core needle biopsy of papillary
biopsy: state of the art at the millennium. AJR lesions of the breast. AJR 179:1185–1192
174:1191–1199 Schueller G et al (2008) US-guided 14 gauge core
Liberman L, Bracero N, Morris E, Thornton C, needle breast biopsy: results of a validation study
Dershaw DD (2005) MRI-guided 9-guage vac- in 1352 cases. Radiology 248:406–413
uum assisted breast biopsy: initial clinical expe- Sie A et al (2006) Multi-center evaluation of the
rience. AJR 185:183–193 breast lesion excision system, a percutaneous,
Litherland J (2001) The role of needle biopsy in vacuum-assisted intact-specimen breast biopsy
the diagnosis of breast lesions. Breast 10: device. Cancer 107:945–949
383–387 Teh W, Evans AJ, Wilson ARM (1998) Editorial.
Parker SH, Burbank F (1996) A practical approach Definitive non-surgical breast diagnosis: the role
to minimally invasive breast biopsy. Radiology of the radiologist. Clin Radiol 53:81–84
200:11–20 Tennant SL, Evans AJ, Hamilton LJ, James J, Lee AH,
Parker SH, Burbank F, Jackman J, Aucreman CJ, Hodi Z, Ellis IO, Rakha EA, Wilson AR (2008)
Cardenosa G, Clink TM et al (1994) Percutaneous Vacuum-assisted excision of breast lesions of
large-core breast biopsy: a multi-institutional uncertain malignant potential (B3) - an alternative
study. Radiology 3:359–363 to surgery in selected cases. Breast 17(6):546–549
http://www.springer.com/978-3-540-31403-5

Das könnte Ihnen auch gefallen