Sie sind auf Seite 1von 25

OMNI-100/200 Guidelines Page 1 of 25

Table of Contents
Title Page #(s)
Connecting OMNI tester to computer …………………………….... 2

Configuring the Tester ………………………………………………. 3-4

Basic Steps for Acquiring ET Data ………………………………. 5

Basic Steps for Acquiring RFT Data ………………………………. 6

Graphical Steps Needed to Setup OMNI Tester ………………. 7-15

SAMPLE High Frequency OMNI-100 Config ………………………. 16

SAMPLE Low Frequency OMNI-100 Config ………………………. 17

SAMPLE RFT OMNI-100 Config ………………………………………. 18

SAMPLE High Frequency RFT OMNI-100 Config ………………. 19

SAMPLE High Frequency External Reference OMNI-200 Config .. 20

SAMPLE High Frequency Internal Reference OMNI-200 Config .. 21

SAMPLE Low Frequency OMNI-200 Config ……………………….. 22

SAMPLE Low Frequency Simultaneous Injection OMNI-200 Config .. 23

SAMPLE RFT OMNI-200 Config ……………………………………….. 24

SAMPLE High Frequency RFT OMNI-200 Config ……………….. 25


OMNI-100/200 Guidelines Page 2 of 25

Connecting OMNI tester to computer


• Connect computer to OMNI tester unit via Ethernet patch cable. Any port will work
• Connect ET pigtail to EDDY CURRENT connector or the RFT pigtail to the DRIVER
PICKUP connector (only use the corresponding pigtail that is meant for the tester.
Example: DO NOT USE OMNI-100 pigtails on the OMNI-200)
• Turn on the tester
• Ensure that the computer has a compatible IP address (not identical)
The Acquisition expects the tester IP address to match the value displayed in the
upper right part of the TESTER CONFIG screen
Tester IP address can be changed by selecting UTIL – IP SETUP from TESTER
CONFIG menu
o Make sure that this value is the same as the one displayed in the upper
right side of the TESTER CONFIG screen
o If you must change the IP value on the OMNI-200, edit the IP address
field and toggle the SET field to YES to confirm the change and then
press OK
Computer IP address should contain the first three octets as the tester
(192.1.6.xxx)
Subnet mask on computer must be 255.255.255.0
• Select TEST LINK button in the TESTER CONFIG dialog box
Green bar at bottom should appear saying “TEST LINK GOOD”

NOTE: ALWAYS POWER DOWN TESTER BEFORE REMOVING OR CONNECTING


PIGTAILS
OMNI-100/200 Guidelines Page 3 of 25

Configuring the Tester

Coils
o Coils 1 and 3 are used for ET on the OMNI-100
o Coils 1 and 2 are used for ET on the OMNI-200
o Coils 2 and 4 are used for RFT on the OMNI-100
o Coils 5 and 6 are used for RFT on the OMNI-200
o Must choose DIFF or ABS by right clicking on the coil number button
• A single Coil must be either all DIFF or ABS

Frequencies
o In the multiplexed mode there can be up to 8 timeslots for the OMNI-100
and 16 timeslots for the OMNI-200
o Using simultaneous injection (OMNI-200 Only) allows up to 4 frequencies
per timeslot
o Changing frequencies can be accomplished by:
• Left or right clicking on the frequency button
• Left click increments and right click decrements
• SHIFT left or right inc/decrements by 10
• CTRL left or right inc/decrements by 100
• SHIFT/CTRL or right inc/decrements by 1,000
• The left side of the frequency button changes the frequency value on
the left side of the decimal
• The right side of the frequency button changes the frequency value on
the right side of the decimal

Coil Gains (actually coil excitation voltage)


o In the ET mode the percentage represents a percentage of the coil
excitation voltage. The max value (100%) is 12 Vpp for the OMNI-100
and 20 Vpp for the OMNI-200

Coil Input Gain (OMNI-200 only)


o This value represents the final amplification stage
o Typical values for ET are 14-20DB for coil 1 (DIFF) and 8-14DB for coil 2
(ABS)
o Typical values for RFT are 8DB for coil 5 (DIFF) and 2DB for coil 6 (ABS)

Dynamic Gain (OMNI-200 only)


o Allows OMNI-200 to function as a 21 bit tester (32-bit mode eneabled)
o Most effective when 32 bit data mode is used
o Requires longer delay times
OMNI-100/200 Guidelines Page 4 of 25

Delays and over sampling


o Delays and over sampling affect signal quality
o Auto Config should be used to set the delays, filter, and over sampling
for the OMNI-100
o At a minimum, the delay on the OMNI-200 should be set high enough to
allow for settling time
o As much over-sampling as possible, this reduces noise
• Lower frequencies do not permit as much over-sampling due to time
constraints (OMNI-200)
o The value will turn RED if there is not enough time in the time-slot
(OMNI-200)
o Press the EXPLAIN button when the values turn RED, it will describe
problem and tell you what to set the values at (OMNI-200)

Internal Reference (OMNI-200 Only)


o This is an optional feature that generates absolute reference data without
requiring an absolute probe
o It is enabled by activating the toggle button on the upper left side of the
TESTER CONFIG dialog box called Internal Reference
o REF NULL must be selected once the probe is in a defect area of tubing
o Some noise may be present at high frequencies
• The noise is easily eliminated by creating a channel with a smoothing
or bandpass filter
OMNI-100/200 Guidelines Page 5 of 25

Basic Steps for Acquiring ET Data


• Power down tester and connect ET pigtail to the EDDY CURRENT connector on
the front of the tester
• Connect test probe and reference probe to ET pigtail. If using the internal
reference feature of the OMNI-200, there is no need to connect a reference probe.
• Perform a TEST LINK function to verify that the tester is connected to the
computer
• Adjust frequencies to desired values
• Make sure that driver 1 is selected for all time slots (OMNI-200)
• OMNI-100 will use coils 1(diff) & 3(abs)
• OMNI-200 will use coils 1(diff) & 2(abs)
• Put the probe in a clean section of tube and press BALANCE (OMNI-200)
• If you have selected to use the internal reference feature of the OMNI-200, put the
probe in a clean section of tube and press REF NULL – wait for the software to
null all channels
• Save configuration to file (use FILE-SAVE AS from menu)
• Close the TESTER CONFIG dialog box and start the tester
OMNI-100/200 Guidelines Page 6 of 25

Basic Steps for Acquiring RFT Data


• Power down tester and connect RFT pigtail to the DRIVER PICKUP connector on
the front of the tester
• Perform a TEST LINK function to verify that the tester is connected to the
computer
• OMNI-100 will use coils 2(diff) & 4(abs)
• OMNI-200 will use coils 5(diff) & 6(abs)
• Make sure that driver 3 is selected for all time slots (OMNI-200)
• Adjust frequencies to desired values
• Save configuration to file (use FILE-SAVE AS from menu)
• Close the TESTER CONFIG dialog box and start the tester
OMNI-100/200 Guidelines Page 7 of 25

Graphical Steps Needed to Setup OMNI Tester

The operator must ensure that the correct tester is selected for use. Select Edit
from the main toolbar in the Acquisition screen. Then select Screens …
Then in the Tester drop down dialog box in the Acquisition section, select the
tester type being used
OMNI-100/200 Guidelines Page 8 of 25

Shown above is the TESTER CONFIG screen for v6.2 and the OMNI-200. This is
where the IP Address in the upper right must match the OMNI-200 IP Address.

OMNI-200 IP address can be changed by selecting UTIL – IP SETUP


from menu.
OMNI-100/200 Guidelines Page 9 of 25

Computer IP Address should have the same first three octets


as the tester. In this example it is 192.1.6.50

The Subnet mask on the computer must be: 255.255.255.0

Select TEST LINK button in the TESTER CONFIG dialog box. Green bar at bottom should
appear saying “TEST LINK GOOD”.
OMNI-100/200 Guidelines Page 10 of 25

Changing the configuration is done in the TESTER CONFIG dialog box. Shown above
is an OMNI-200 tester config. Coils 1 and 2 are used for ET impendence probe data.
Coils 5 and 6 are used for RFT data. Change frequencies to as listed in the ETSS.

Changing frequencies can be accomplished by:


• Left or right clicking on the frequency button
• Left click increments and right click decrements
• SHIFT left or right inc/decrements by 10
• CTRL left or right inc/decrements by 100
• SHIFT/CTRL or right inc/decrements by 1,000
• The left side of the frequency button changes the frequency
value on the left side of the decimal
• The right side of the frequency button changes the frequency
value on the right side of the decimal
OMNI-100/200 Guidelines Page 11 of 25

The operator must choose the DIFF or ABS by right clicking on the coil number button. A single
coil must be either all DIFF or ABS. Up to 4 ET impedance coils can be used.

Coil gains (actually coil excitation voltage) in the ET mode, the percentage represent a percentage
of the coil excitation voltage. The max value (100%) is 12 Vpp for the OMNI-100 and 20 Vpp for
the OMNI-200
OMNI-100/200 Guidelines Page 12 of 25

The value of the COIL INPUT GAIN represents the final amplification stage for the
OMNI-200. Typical values for ET are 14-20DB for coil 1 (DIFF) and 8-14DB for coil 2
(ABS). In the example above, they are set to 14DB for coil 1 and 8DB for coil 2. This
is the best starting point.

Typical values for RFT are 8DB for coil 5 (DIFF) and 2DB for coil 6 (ABS).
OMNI-100/200 Guidelines Page 13 of 25

Dynamic Gain allows the OMNI-200 to function as a 21 bit tester. It requires 32 bit
data mode and longer delay times.

Delay times and over sampling can affect data quality if the operator is not careful.
Ideally enough delay is needed to use the dynamic gain feature. Use as much over
sampling as possible. This will help reduce noise.
OMNI-100/200 Guidelines Page 14 of 25

The DELAY value will turn RED if there is not enough time in the time-slot.

Press the EXPLAIN button when the values turn RED and it will describe the
problem and tell you what to set the DELAY values at.

Press OK and then enter in the delay numbers required in the appropriate time slots.
OMNI-100/200 Guidelines Page 15 of 25

Select TEST LINK button in the TESTER CONFIG dialog box. Green bar at bottom
should appear saying “TEST LINK GOOD”.

If you are using the Internal Reference Mode, you must select that option from the
toggle buttons on the upper left side of the TESTER CONFIG dialog box
OMNI-100/200 Guidelines Page 16 of 25

SAMPLE High Frequency OMNI-100 Config


OMNI-100/200 Guidelines Page 17 of 25

SAMPLE Low Frequency OMNI-100 Config


OMNI-100/200 Guidelines Page 18 of 25

SAMPLE RFT OMNI-100 Config


OMNI-100/200 Guidelines Page 19 of 25

SAMPLE High Frequency RFT OMNI-100 Config


OMNI-100/200 Guidelines Page 20 of 25

SAMPLE High Frequency External Reference OMNI-200 Config


OMNI-100/200 Guidelines Page 21 of 25

SAMPLE High Frequency Internal Reference OMNI-200 Config


OMNI-100/200 Guidelines Page 22 of 25

SAMPLE Low Frequency OMNI-200 Config


OMNI-100/200 Guidelines Page 23 of 25

SAMPLE Low Frequency Simultaneous Injection OMNI-200 Config


OMNI-100/200 Guidelines Page 24 of 25

SAMPLE RFT OMNI-200 Config


OMNI-100/200 Guidelines Page 25 of 25

SAMPLE High Frequency RFT OMNI-200 Config

Das könnte Ihnen auch gefallen