1-1 M aximum input voltage vs frequency derating curve for C H 1 OR Χ, CH 2 O R Υ , and ΕΧΤ I NPUT connectors . . 1-7
2-3 P ower an d d isp lay co ntrols and indicators and PR O BE ADJU ST out p ut . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Table Page
Th e following electrical characteristics (Table 1-1) are mance c haracteristics for which no absolute limits are speci-
valid for th e 2213Α whe n it h as been adjusted at an am b ient fied, or c haracteristics that are imp ractical to check.
temperature betwee n +20°C and +30°C, h as h ad α warm-
up period of at least 20 minutes, and is operating at an
am bient temperature between Ο °C and +50°C (unless ot h- En vironmental c h aracte ristics are given in Ta ble 1-2. The
erwise noted) . 2213Α meets the req uireme n ts of MIL-Τ -288000, para-
graph s 4.5 .5 .1 .3, 4 .5 .5 .1 .4, and 4.5 .5 .1 .2 .2 for type ΙΙΙ ,
Class 5 equi pme nt, exce pt where ot herwise noted.
Items listed in the " Performance Req ui remen ts" column
are verifiable qu alitative o r quan titative limits, while items
liste d in th e "S upp leme ntal Information" column are either Ph ysical characteristics of the instrument are listed in
explanatory notes, calibration set up descri ption s, perfor- Table 1-3.
Ta ble 1-1
Electrical Characte ristics
C h a racte ristics Performance Requ irements Supplemental Info rmation
0°C to 50°C
2 mV per Division to 5 V per 7.0 ns or less .
Division
B andwidth (-3 d B) M easu red with α vertically centered
6-division reference sign al from α 50 Ω
0°C to +35°C
so urce d riving α 50 Ω coaxial cable
2 mV per Division Dc to at least 50 MH z. that is terminated in 50 Ω, bot h at
the input co nnector an d at t h e probe
5 mV per Division to 5 V per Dc to at least 60 MHz . input, with t he VOLTS/DIV V ariable
Division control in the CA L detent .
0°C to +50°C
M aximum Safe Input Voltage A See Fig ure 1-1 for derating curve.
DC Co u p led 400 V ( d c + pea k ac) or 800 V ac ρ-ρ
to 10 kH z or less .
Common-Mode Rejection Ratio At least 20 to 1 at 25 MHz. Chec ked at 10 mV per division for
(C MRR) common-mode signals of 6 divisions or
less with VOLTS/DIV Varia ble cont rol
adjusted fo r best CMRR at 50 kH z.
Trace Shift with Attenuator Rotation 0.75 division or less . V O LTS/DIV Variable co ntrol in CAL
detent .
Extern al 40 mV 150 mV
EXTI NPUT
M aximum Input Voltage A 400 V (dc + peak ac) or 800 V ac ρ-ρ See F igure 1-1 for derating curve.
at 10 kH z or less .
In p ut Capacitan ce 20 pF ±2 .5 pF .
Delay Time
PROBE AD JUST
Z-AXIS INPU T
POWER SO UR CE
Standard Phosph or Ρ 31 .
NOTE
The instrument meets the requirements of MIL-T-288000, paragraphs 4.5 .5.1 .3,
4.5.5.1 .4, and 4.5.5.1 .2.2 for Type /Ι/, Class 5 equipment, except where other-
wise noted.
Temperature
Altitude
H umidity (Operating and N onoperating) 5 cycles (120 h ours) refe renced to M I L-Τ-288000 paragraph 4.5 .5 .1 .2 .2 for Ty pe
III, Class 5 instruments. Operati ng a n d non-operating at 95% +0% to -5%
relative h umidity. Operating at +50°C and +30°C. Non-operating at +30°C to
+60°C.
Vibration (Operating) 15 mi nutes along each of 3 major axes at α total d isplacementof 0.015 inch ρ-ρ
(2 .4 g's at 55 H z) with freq ue ncy varied from 10 H z to 55 H z to 10 H z in 1-mi nute
swee ps. H old for 10 mi nutes at 55 H z in eac h of the 3 major axes . All major
resonances must be above 55 Hz .
Shoc k (Operati ng a nd N onoperating) 30 g's, h alf-si ne, 11-ms d uratio n , 3 shocks per axis each di rection, for α total of
18 shocks.
EMI M eets ra d iated and conducted emission requ irements per V DE 0871 Class Β.
Table 1-3
Physical Characteristics
Characteristics Description
Weight With Power Cord
Height
Width
Depth
L I NE VO LTAG E
The inst rument is capa ble of continuous operation using
input voltages that r ange from 90 V to 250 V nominal at
fre qu encies from 48 H z to 440 H z.
POWER CO R D
Α d etac hable t hree-wire power co rd with α three-con tact
plug is provided with each instrument to pe rmit connection
to both t he power source and protective ground. The p lug
protective-ground contact connects (through t he protective-
grou nd co nductodto the accessible metal p arts of the instru-
ment. For electrical-shock protection, insert this plug only
into α power outlet that has α securely grou nded protective-
grou nd contact .
This inst ru men t is shipped with the requ ired powe r cord
as or dered by the customer . Available powe r-cord informa-
tion is ill ust rated in Figure 2-1, an d part numbers are listed in
Section 5 of this ma nual . Contact your Tektronix represen-
tative or local Tektro nix Field Office fo r add itio nal powe r-
cord informatio n. 2. P ress in a nd slig htly rotate the fuse-hol der cap coun-
terclockwise to release it .
1 . Unplug the power cord f rom the power-input source 5. Reinstall the fuse (o r re placement fuse) and the fuse-
(if app licable) . h older ca p .
Always maintain adequate instrument cooling. The venti-
lation holes on both sides of the instrument cabinet and on
the rear panel must remain free of obstruction .
T he following descri ption s ar e intended to familiarize t h e PR OBE AD JU ST Co nn ecto r-Provides an a pproxi-
operator wit h t h e location , operation , a n d fu n ctio n of t h e mately 0.5 V, negative-going, sq uare-wave voltage (at
instrument's controls, connectors, an d indicators . approximately 1 kH z) that permits an operator to
compensate voltage probes and to c hec k operation of
th e oscilloscope vertical system . It is not intended for
DIS PL AY, PO WER, A N D PR O BE AD JU ST verifying the acc u racy of th e vertical gain or time-base
circuitry.
Refer to Fig ure 2-3 for locatio n of items 1 t h rough 8.
Internal Gratic ule- Elimi n ates parallax viewing error BEAM F IND Switch-Wh en h eld in, compresses the
between t he t race and g raticule lines . R ise-time am pli- display to within the graticule area and provides α visi-
tud e an d meas ureme nt points are ind icated at th e left ble viewing intensity to ai d i n locating off-screen
edge of th e g raticule . dis plays .
i
POWER Switch-Turn s n strument power on and off.
Press in for ON ; press again for OFF . TRAC E ROTATIO N Contr ol-Screwdriver adjust-
ment used to align the crt trace with t h e horizontal
Power Indicator-An LE D that illuminates when the graticule li n es .
inst ru ment is operating.
F OC US Co ntrol-Adjusts for optimum d isplay INTEN SITY Co ntrol-Determines the brig htness of
definitio n. the sweep trace.
j2 Input Coupling (AC-GND-DC) Switches-Three-po-
sition switches that select the method of cou p ling the
R efer to Figure 2-4 for location of items 9 through 16 .
input signals to the instrument d eflection system .
CH 1 V OLTS/DIV a nd CH 2 VOLTS/DIV Switches- AC-Input signal is capacitively coupled to the ver-
Used to select the vertical deflection factor in α 1-2-5 tical amplifier. The do component of the input sig-
sequence . To obtain α calibrated d eflection factor, the
nal is b locked . Low-fre q uency limit (-3 dB point) is
V OLTS/DIV variable control must be in the calibrated approximately 10 Hz .
(CAL) d etent (fully clockwise) .
G ND-The input of the vertical amplifier is
1X-Indicates the deflection factor selected when
grounded to p rovide α zero (ground) reference-volt-
using either α 1 Χ probe or α coaxial cable.
age dis play ( d oes not ground the input signal). This
switch position allows precharging the input cou-
10Χ PROBE-Indicates t he deflection factor se-
pling ca p acitor .
lected when using α 10 Χ probe.
DC-All frequency components of the input signal
are cou p led to the vertical deflection systems.
VOLTS/DIV Variable Controls-When rotated coun-
terclockwise out of their calibrate d detent positions,
these controls provide continuously variable,
uncalibrated deflection factors between the calibrated Ί3 CH 1 OR Χ a n d CH 2 OR Υ Input Connectors-
settings of the VOLTS/DIV switches . Provide for application of external signals to the verti-
cal d eflection system or for an Χ-Υ display. In the Χ-Υ
mode (SEC/DIV switch set to Χ-Υ), the signal con-
POSITION Controls-Use d to vertically p osition the nected to the CH 1 OR Χ input connector p rovides
display on the crt. When t he SEC/DIV switch is set to horizontal deflection (X-axis) and the signal connecte d
Χ-Υ, the Channel 2 POSITION control moves the dis- to the CH 2 OR Υ input connector provides vertical
play vertically (Y-axis), and t he H orizontal POSITION deflection (y-axis) .
control moves the d isplay h orizontally (X-axis) .
D
swee p and t he beginning of the intensifie d zone .
16 SER IAL a nd Mod Slots-The SER IA L slot is im-
p rinted with the instrume nt's se rial number . The M od Range Selector Switch--This three-positio n
slot contains the option nu mber that is installed in the switch selects 0.4 ms, 20 μs, and 1 .0 μs of delay
inst rument. time . To increase the sweep delay from the cali-
brated setting of the Range Selector Switch, rotate
the MULTI PL IER control clockwise .
TV FIEL D- Press in bot h Ρ-Ρ A UTO and NORM ΙΝΤ Switch-Selects the so urce of the internal trig-
2$
buttons . Permits triggering on television fiel d sig- gering signal when the SO UR C E switch is set to INT.
nals. TRIGGER LEVEL control s hould be rotated
fully counterclockwise when triggering on TV sig- CH 1-The signal applied to the CH 1 O R Χ in p ut
nals with negative going sync and cloc kwise for connector is the so urce of the trigger sign al .
p ositive going sync .
VERT M OD E-The inte rn al trigger source is deter-
SGL SWP RE SET- Press in th e sp ring-ret urn but- mi ned by the signals selected for dis play by the
ton momentarily to arm the trigger circuit for α sin- VERTICA L M OD E switches . See Table 2-1 for
gle-sweep display. In this mode, the trigger system VERT M OD E trigger source .
operates the same as N ORM, except only one
sweep is displayed for each trigger sign al . Anot her
sweep cannot b e displayed u ntil the SG L SWP
RE SET button is momentarily pressed in agai n to
reset the trigger circuit. This mode is useful for dis-
playing and p hotographing either nonrepetitive sig-
nals or signals that cause u nstable conventional
,.
displays (e .g ., signals that vary in amplit ude,
shape, or time).
D
swee p and t he beginning of the intensifie d zone .
16 SER IAL a nd Mod Slots-The SER IA L slot is im-
p rinted with the instrume nt's se rial number . The M od Range Selector Switch--This three-positio n
slot contains the option nu mber that is installed in the switch selects 0.4 ms, 20 μs, and 1 .0 μs of delay
inst rument. time . To increase the sweep delay from the cali-
brated setting of the Range Selector Switch, rotate
the MULTI PL IER control clockwise .
I NPUT CO UPL I N G
level difference is more than 10 times the VOLTS/DIV 3. Connect the probe tip to the signal source and wait
switch setting: several seconds for the input coupling capacitor to charge .
2 . Insert the probe tip into the oscilloscope GND connec- 4. Set the Input Coupling switch to AC . The display will
tor and wait several seconds for the input coupling capacitor remain on the screen, and the ac component of the signal
to discharge . can be measured in the normal manner .
O PERATOR'S ADJU ST MENTS
I NT ROD UCTION TRIGGER
VAR HOLDO FF NORM
To verify the operatio n and accuracy of yo ur instrument
Mode Ρ-Ρ AUTO
b efore ma king meas ureme nts, perform the followi ng adj ust- OUT
S LO PE
ment procedu res. If adjustments are required beyond the
LEVEL Midrange
scope of the operators's ad justments, refer th e instrument VERT M OD E
ΙΝΤ
to α qualifie d service technician. ΙΝΤ
SO UR CE
ΕΧΤ CO UPLING AC
BAS EL I NE T RAC E
T RAC E ROTATIO N
First obtain α baseline trace, using the following
proced ure .
Normally, the resulting trace will be pa rallel to the center
horizontal graticule line, and the Trace Rotation ad ju stment
s houl d not be req uired . If ad ju stment is needed, perform t h e
1 . Preset the inst ru ment front- panel controls as follows:
following procedure :
Display
I NTEN SITY Fu lly countercloc kwise 1 . Preset instrument controls and obtain α baseline
FOC U S M idrange trace.
V ertical (Bot h Chann els) 2. Use the Channel 1 POSITIO N control to move the
baseline trace to the center horizontal graticule line .
POSITIO N Midrange
VERTICA L MOD E CH 1
BW LI MIT Off (butto n out)
mV 3. If the resulting trace is not parallel to the center hori-
V50
OLTS/DIV
zontal graticule line, use small flat-bit screwdriver to adjust
VOLTS/DIV Varia ble CA L detent
the TR AC E ROTATIO N control and alig n t he trace with t h e
CH 2 I NVER T Off (button out)
AC ce nter ho rizontal graticule line .
Input Coupling
Horizontal
PR OBE CO MPEN SATION
POSITIO N Midrange
H O RIZONTA L MOD E ΝΟ ID LY M isadj ustment of probe com pensation is α commo n
SEC/DI0V .5 ms source of measurement error. M ost attenuator probes are
SEC/DI V Varia ble CA L detent equ ipped with α com p ensation ad j ustment. To ensure opti-
Χ10 M ag nifier Off (knob in) mum measurement accuracy, always compensate the oscil-
Range Selector 0.4 ms loscope p ro bes before ma king measurements . Pro be
MULTI PLIER < Χ1 com pensation is accom p lished as follows :
Operato r s Familiarization-2213A Operators
7. Chec k the wavefo rm presentation for overshoot and 9. U se the Channel 2 POSITIO N to vertically center the
rolloff (see Fig ure 3-2) . If necessary, ad just the p robe com- display a nd r epeat step 7 for the Channel 2 probe.
pensation for flat tops on the waveforms . Refer to the in
structions supp lied with t he probe for details of
compensation adjustment . 10 . Disconnect the p ro bes from the instru ment.
I N T R OD U CTIO N VO LTAG E ME AS UREMEN TS
After becoming familiar with the capabilities of the 2213 Α AC P eak-to-Peak Voltage
Oscilloscope an operator can then easily develop conve- To make α peak-to-peak voltage measurement, use the
nient methods for making p articular measurements . The in- following procedure:
formation in this section is designed to enhance operator
understanding and to assist in developing efficient tech-
ni ques for making specific measurements . Recommended ΝΟΤΕ
methods for making basic measurements with your instru-
ment are described in the procedures contained in t his This procedure may also be used to make voltage
section. measurements between any two points on the
waveform .
V OLTAGE MEASUREMENTS . . . . . . . . . . . . . . . . . 4-1 4. Adjust the TRIGGER LEVEL control to obtain α stable
AC Peak-to-Peak Voltage . . . . . . . . . . . . . . . . . . 4-1 display.
Instantaneous Voltage . . . . . . . . . . . . . . . . . . . . . 4-2
Algebraic Addition . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Common-Mode Rejection . . . . . . . . . . . . . . . . . . . 4-3 5. Set the SEC/DIV switch to α position that displays
Amplitude Comparison . . . . . . . . . . . . . . . . . . . . . 4-4 several cycles of the waveform .
NOTE
1 . Preset instrument controls an d obtain α baseline
If using Channel 2, ensure that the Chann el 2 INVER T
trace.
switch is in its noninverting mo de (button out) .
V OLTS/DIV
Instanta- vertical switch setting
neo us = d eflectio n χ p olarity χ in dicated by 1 Χ
Voltage (divisio ns) (+ or -) (or 10 Χ when 10Χ
p robe is used)
1 . Preset instrume nt controls and obtain α baseline / 1 . Preset instrument controls an d obtai n α baseline
trace. trace.
2. Apply the reference signal to either vertical c hannel 2 . Apply the signal to either ve rtical-chan nel input con-
input and set the VERTICA L MOD E switch to display the nector and set the VERTICA L MOD E switch to display the
channel used . channel used .
3. A djust the TRIGGER LEVEL control to obtain α stable Fr eque ncy
disp lay . T he freque ncy of α recu rrent signal ca n be determined
from its time- d uration measureme nt as follows :
5. P osition the display to p lace the time-measurement 2 . Calculate the reci procal of the time-duratio n value to
poi nts on the center ho rizontal graticule line (see F igure 4-5) . determine the fre que n cy of the waveform .
4. Vertically position the trace so that the zero reference 6. If the two signals are of op posite polarity, press in the
of the waveform touches the 0% graticule li ne and the top of Channel 2 INVER T button to invert the Channel 2 display
the waveform touc hes the 100% gratic ule line. (signals may b e of opposite polarity d ue to 180° p hase
difference) .
5 . If the two sign als are of opposite polarity, press in Exam ple : The horizontal difference is 0.6 d ivision with α
t h e Ch ann el 2 I NVER T button to invert t h e C h an nel 2
graticule calibration of 45° p er division as shown in F ig-
display. ure 4-8.
6. Set bot h VOLTS/DIV switch es and b ot h Var ia b le Substituting the given values into the phase d iffere nce
controls so the displays are equal in amplitude . fo rmula :
7. Adj ust the TRIGGER LEVEL co ntrol for α stable Ph ase difference = 0.6 div χ 45°/div = 27° .
display.