Sie sind auf Seite 1von 5

60W AC-DC Reliab le Green Medical Adaptor GSM60A series

+ +
EN 60601-1
IEC 60601-1

( for 5~9V)
EN60601-1 ANSI/AAMI ES60601-1 IEC60601-1

Features Applications
3 pole AC inlet IEC320-C14, ClassⅠpower unit Mobile clinical workstation
Medical safety approved (2 x MOPP) accroding to Oral irrigator
ANSI/AAMI ES60601-1 and IEC/EN60601-1 Portable hemodialysis machine
Extremely low leakage current Breath Machine
No load power consumption<0.1W Medical computer monitor
Energy efficiency level VI and meet CoC Version 5
(Except 5~9V for Level V)
-30~+70℃ wide range working temperature
Protections: Short circuit / Overload / Over voltage/ Over temperature
LED indicator for power on
Lifetime > 105 K hours
3 years warranty

Description
GSM60A is a highly reliable , 60W desktop style single-output green medical adaptor series . This product is
a class I power unit(with FG),equipped with a standard IEC320-C14 AC inlet and adopting the input range
from 80VAC to 264VAC . The entire series supplies different models with output voltages between 5VDC
and 48VDC that can satisfy the demands for various types of medical electrical devices. The circuitry
design meets the international medical standards (2*MOPP), having an ultra low leakage current (<100μA),
fitting the medical devices in direct electrical contact with the patients.
With the efficiency up to 91 % and the extremely low no-load power consumption below 0 . 1W , GSM60A is
c ompliant with USA EISA 2007/DoE , Canada NRCan, Australia and New Zealand MEPS , EU ErP, and
meet Code of Conduct(CoC) Version 5.The supreme feature allows the adaptor to save the energy when
it is either under the operating mode or the standby mode . The entire series utilizes the 94V-0 flame
retardant plastic case . GSM60A is certified for the international medical safety regulations .

Model Encoding
GSM60 A 05 - P1J

{
P1J: Standard model,2.1ψx5.5ψx11mm,C+,tuning fork type
DC plug type
Other options available by customer requested (see Page 4~5)
Output voltage
IEC320-C14 AC inlet
Rated wattage
Series name

File Name:GSM60 A -SPEC 2018-01-12


60W AC-DC Reliab le Green Medical Adaptor GSM60A series
SPECIFICATION
ORDER NO. GSM60A05-P1J GSM60A07-P1J GSM60A09-P1J GSM60A12-P1J GSM60A15-P1J GSM60A18-P1J GSM60A24-P1J GSM60A48-P1J
SAFETY MODEL NO. GSM60A05 GSM60A07 GSM60A09 GSM60A12 GSM60A15 GSM60A18 GSM60A24 GSM60A48
DC VOLTAGE Note.2 5V 7.5V 9V 12V 15V 18V 24V 48V
RATED CURRENT 6A 6A 6A 5A 4A 3.33A 2.5A 1.25A
CURRENT RANGE 0.1 ~ 6A 0.1 ~ 6A 0.1 ~ 6A 0.1 ~ 5A 0.1 ~ 4A 0.1 ~ 3.33A 0.1 ~ 2.5A 0.1 ~ 1.25A
RATED POWER (max.) 30W 45W 54W 60W 60W 60W 60W 60W
OUTPUT RIPPLE & NOISE (max.) Note.3 80mVp-p 80mVp-p 100mVp-p 100mVp-p 100mVp-p 120mVp-p 150mVp-p 240mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0% ±5.0% ±5.0% ±3.0% ±3.0% ±3.0% ±3.0% ±2.5%
LINE REGULATION Note.5 ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0%
LOAD REGULATION ±5.0% ±5.0% ±5.0% ±3.0% ±3.0% ±3.0% ±3.0% ±2.5%
SETUP, RISE TIME Note.6 1000ms, 30ms / 230VAC 1500ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.) 50ms / 230VAC 18ms / 115VAC at full load
VOLTAGE RANGE Note.7 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE 47 ~ 63Hz
EFFICIENCY (Typ.) 81.5% 86% 87.5% 88% 88.5% 89% 90.5% 91.5%
INPUT
AC CURRENT (Typ.) 1.4A / 115VAC 1A / 230VAC
INRUSH CURRENT (Typ.) Cold start 30A/115VAC 60A / 230VAC
LEAKAGE CURRENT(max.) Earth leakage current < 100μA/264VAC , Touch current < 100μA/264VAC
105 ~ 160% rated output power
OVERLOAD
Protection type : Hiccup mode, recovers automatically after fault condition is removed
PROTECTION
5.2 ~ 7.0V 7.8 ~ 10.2V 9.4 ~ 12.2V 12.6 ~ 16.2V 15.7 ~ 20.3V 18.9 ~ 24.3V 25.2 ~ 32.4V 50.4 ~ 64.8V
OVER VOLTAGE
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE Shut down o/p voltage, re-power on to recover
WORKING TEMP. -30 ~ +70℃ (Refer to "Derating Curve")
WORKING HUMIDITY 20% ~ 90% RH non-condensing
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT ±0.03% / ℃ (0~40℃ )
VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.8 3000 meters
SAFETY STANDARDS IEC60601-1, TUV EN60601-1, ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22.2 No. 60601-1:14 - Edition 3, EAC TP TC 004 approved
ISOLATION LEVEL Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP
WITHSTAND VOLTAGE I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:SHORT
ISOLATION RESISTANCE I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25℃!/ 70% RH
Parameter Standard Test Level / Note
EN55011 (CISPR11), FCC PART 15 / CISPR22,
Conducted emission CAN ICES-3(B)/NMB-3(B) Class B

EMC EMISSION EN55011 (CISPR11), FCC PART 15 / CISPR22,


Radiated emission CAN ICES-3(B)/NMB-3(B) Class B
Harmonic current EN61000-3-2 Class A
SAFETY & Voltage flicker EN61000-3-3 -----
EN55024 , EN60601-1-2, EN61204-3
EMC
(Note 9) Parameter Standard Test Level / Note
ESD EN61000-4-2 Level 4, 15KV air ; Level 4, 8KV contact
Level 3, 10V/m( 80MHz~2.7GHz )
RF field susceptibility EN61000-4-3 Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts EN61000-4-4 Level 3, 2KV
EMC IMMUNITY
Surge susceptibility EN61000-4-5 Level 3, 1KV/Line-Line , 2KV/Line-FG
Conducted susceptibility EN61000-4-6 Level 3, 10V
Magnetic field immunity EN61000-4-8 Level 4, 30A/m
100% dip 1 periods, 30% dip 25 periods,
Voltage dip, interruption EN61000-4-11
100% interruptions 250 periods
MTBF 720K hrs min. MIL-HDBK-217F(25℃ )
OTHERS DIMENSION 125*50*31.5mm (L*W*H)
PACKING 0.32Kg; 40pcs/ 13.8Kg/1.05CUFT
PLUG See page 4~5 ; Other type available by customer requested
CONNECTOR
CABLE See page 4~5 ; Other type available by customer requested
NOTE 1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. Derating may be needed under low input voltages. Pleas check the derating curve for more details.
8. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
9. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)

File Name:GSM60 A -SPEC 2018-01-12


60W AC-DC Reliab le Green Medical Adaptor GSM60A series

Derating Curve Static Characteristics

100
100
90

80
70

LOAD (%)
70
LOAD (%)

50
60

50

20 40

-30 0 10 20 30 40 50 60 70 (HORIZONTAL) 80 90 100 110 120 140 160 180 200 220 240 264

AMBIENT TEMPERATURE (℃) INPUT VOLTAGE (VAC) 60Hz

Mechanical Specification Case No. GS60A Unit:mm

UL2464 16AWG 1000±50mm for 5 ~ 15V


125 UL1185 16AWG 1500±50mm for 18 ~ 48V
31.5

Power LED
70 ± 10mm
50

AC Inlet
IEC320-C14

DC output plug
◎ Standard plug: P1J

P1J Pin Assignment

C"+"
5.5 11 ± 0.5mm

2.1 Outside Inside


AC FG
-V connected to AC FG

File Name:GSM60 A -SPEC 2018-01-12


60W AC-DC Reliab le Green Medical Adaptor GSM60A series

◎ Optional DC plug:

A B C
Tuning Fork Style Type No.
OD ID L
C P1I 5.5 2.1 9.5
P1L 5.5 2.5 9.5
A (Straight) P1M 5.5 2.5 11.0
C P1IR 5.5 2.1 9.5
B
P1JR 5.5 2.1 11.0
P1LR 5.5 2.5 9.5
(Right-angled)
P1MR 5.5 2.5 11.0
A B C
Barrel Style Type No.
OD ID L
C P2I 5.5 2.1 9.5
P2J 5.5 2.1 11.0
P2L 5.5 2.5 9.5
A
(Straight) P2M 5.5 2.5 11.0
B
C P2IR 5.5 2.1 9.5
P2JR 5.5 2.1 11.0
P2LR 5.5 2.5 9.5
(Right-angled)
P2MR 5.5 2.5 11.0
A B C
Lock Style Type No.
OD ID L
A
Locking C P2S(S761K) 5.53 2.03 12.06
P2K(761K) 5.53 2.54 12.06
B P2C(S760K) 5.53 2.03 9.52
SWITCHCRAFT original or equivalent P2D(760K) 5.53 2.54 9.52
A B C D
Center Pin Style Type No.
OD ID L Center Pin

A C P4A 5.5 3.4 11.0 1.0


P4B 6.5 4.4 11.0 1.4
B
D EIAJ equivalent P4C 7.4 5.1 11.0 0.6
Pin Assignment
Min. DIN 3 Pin with Lock (male) Type No.
PIN No. Output
1 +Vo
1 2 -Vo
2 R6B
3 KYCON KPPX - 3P equivalent 3 +Vo

File Name:GSM60 A -SPEC 2018-01-12


60W AC-DC Reliab le Green Medical Adaptor GSM60A series

Pin Assignment
Min. DIN 4 Pin with Lock (male) Type No.
PIN No. Output
1 +Vo
2 3 2 -Vo
1 4 R7B
3 -Vo
KYCON KPPX - 4P equivalent
4 +Vo
Pin Assignment
Min. DIN 4 Pin with Lock (female) Type No.
PIN No. Output
1 +Vo
23 R7BF 2 -Vo
14 3 -Vo
KYCON KPJX-CM-4S equivalent 4 +Vo
Pin Assignment
DIN 5 Pin (male) Type No.
PIN No. Output
1 -Vo
2 -Vo
1 3 R1B 3
4 2 5 +Vo
4 -Vo
5 +Vo
Pin Assignment
Stripped and tinned leads Type No.
PIN No. Output
L (red)
1 1 +Vo
2
L1 (black) by customer
Length of Land L1 by request
2 -Vo
( MW’s standard length, L: 25 mm, L1: 5 mm)

Installation Manual
Please refer to : http://www.meanwell.com/manual.html

File Name:GSM60 A -SPEC 2018-01-12