Beruflich Dokumente
Kultur Dokumente
EXPERIMENTAL
Reagents
n-Butanol saturated with 2 N HCl. n-Butanol, reagent grade, is
shaken with 2 N HCl at room temperature. The butanol should be satu-
rated with the aqueous acid solution the day it is used.
803
804 DETERMINATION OF CHOLINE IN PLASMA
WAVELENGTH mp
FIG. 1. Absorption spectra of choline periodide (A) and iodine (B) in ethylene
dichloride. Choline periodide concentration, 16.9 -y per ml. (equivalent to 1.65 y
per ml. of choline). Iodine concentration, 10 y per ml.; cell thickness, 1 cm.
cipitate was washed several times with ice-cold water to free it from ad-
sorbed iodine reagent and dried on filter paper over Drierite in a desiccator
at 4”. Weighed amounts of the dry black crystalline precipitate were
analyzed for iodine following sodium fusion, and for nitrogen by the micro-
Kjeldahl procedure.
CSH,~ONIO. Calculated. N 1.12, I 91.8
Found. “ 1.10, “92.0 (Sample 1)
“ 1.10, “91.0 ( “ 2)
The results are in agreement with the theoretical value required for the
compound to contain 9 atoms of iodine, and they confirm the findings of
Stan&k (11). 16.9 mg. of the complex were dissolved in a liter of ethylene
amount of ethylene dichloride with a fine tipped stirring rod, and dilute
the solution with additional solvent to a volume of 10 ml. (when the amount
of choline is less than 7 y, dilute to 5 ml.). Transfer about 3 ml. of the solu-
tion to a cuvette and read the optical density of the solution at 365 mp,
with ethylene dichloride for the zero setting. A reagent blank run through
the entire procedure should give a reading of less than 0.030 against ethylene
dichloride (Beckman ultraviolet spectrophotometer).
25 y of choline run through the precipitation procedure give an optical
density of about 0.515. The sensitivity may be doubled by dissolving
the complex in 5 ml. instead of in 10 ml. of ethylene dichloride. When
known amounts of choline are run through the entire plasma procedure,
TABLE I
Recovery of Choline from Aqueous Solution
Y gercent gerGent
100 21 99 2.7
50 26 100 2.1
30 12 98 2.4
25 46 97 3.4
20 15 99 2.7
10 16 94 5.0
5 28 94 8.0
tribution of choline between water and these organic solvents varies some-
what from day to day, standards should be run along with each series of
plasma samples.
Recovery of Choline-Various amounts of choline in 0.5 ml. of water were
treated with the triiodide reagent and the choline was assayed as described
above. The optical densities were found to be proportional to concentra-
tion and closely coincided with those obtained by spectrophotometric meas-
urement, in ethylene dichloride solution, of equivalent amounts of choline
periodide (see “Reagents”). Precipitation of the choline as the iodine
complex was essentially complete, but when quantities as small as 5 and
10 y were analyzed, the recovery was somewhat low and variable (Table I).
The reproducibility of the plasma choline method and the recovery of
choline added to plasma were checked as follows: A sample of pooled hu-
man plasma was divided into four portions. The first portion was ana-
lyzed directly for choline, with six replicates of 3.0 ml. each. The other
three portions were analyzed after the addition of 5, 10, and 15 y of choline
808 DETERMINATION OF CHOLINE IN PLASMA
per ml. of plasma. The results indicate that the reproducibility and re-
covery are satisfactory (Table II).
Assay of XpeciJicity-The specificity of the method for the estimation of
free choline in plasma was assayed in the plasma of two human subjects
by the Craig counter-current distribution technique (16). The choline
was isolated from 40 ml. samples of plasma as described in the analytical
method, and the aqueous residue was evaporated to dryness at room tem-
perature in a stream of air. The choline was quantitatively extracted
into two 20 ml. portions of absolute alcohol and the alcohol was evaporated
to dryness at room temperature. This step served to separate the choline
Choline added Total choline expected* Total choline found kcovery of added choline
* The endogenous choline value was 7.3 y per ml. (the average of the six values,
7.7,7.7,7.3,7.0,7.2, and 7.1).
The specificity of the choline method was further checked by the tech-
nique of comparative partition ratios (17). Choline was isolated from 45
ml. of human plasma as described above for the counter-current distribu-
tion. The partition ratios of the apparent choline from plasma were com-
pared with those of authentic choline in a series of two-phase systems
consisting of 2 N I-ICl and n-butanol containing various concentrations of
petroleum ether. The result,s indicated that both substances possessed the
same solubility characteristics and were, therefore, presumably the same
compound (Table IV).
TABLE III
Fraction in buta-
no1 phase*
I
y per tube
Subject B
Fraction in buta-
no1 phase’
___~
1 24 0.60 8 0.58
2 113 0.63 56 0.60
3 183 0.64 94 0.60
4 102 0.63 54 0.60
Per cent petroleum ether in Authentic choline, fraction in Apparent choline from plasma,
butanol phase organic solvent fraction in organic solvent
0 0.75 0.79
1 0.61 0.59
3 0.47 0.46
4 0.37 0.37
5 0.28 0.29
7 0.11 0.10
10 0.00 0.00
The variation in free plasma choline throughout the day was measured
in four of the normal individuals. None of the values, including those
taken 1 to 2 hours after a meal, was appreciably different from the morning
fasting value, indicating that the dietary intake of choline did not affect
the plasma levels.
The variation in free plasma choline over a period of 4 weeks was deter-
mined in thirteen normal male individuals. Generally speaking, the cho-
line levels for a given individual were found to remain relatively constant
from week to week (Table V). The levels in ten other subjects were meas-
ured several months apart and found to be relatively unchanged. This
latter group contained two individuals whose plasma levels were about 10
TABLE V
Plasma Levels of Free Choline in Normal Individuals Measured at Weekly Intervals
The values are given in micrograms per ml.
Subject 1st wk. 2nd wk. 1 3rd wk. 4th wk. Subject 1st wk. 2nd wk. 3rd wk. 4th wk.
.~
It was concluded therefore that the major part of the choline had undergone
metabolic transformation.
DISCUSSION
that each individual maintains a relatively constant plasma level and this
level is not increased after meals or by the oral administration of large
amounts of choline.
Evidence is presented suggesting that the level of free choline in plasma
is controlled by a process involving metabolic transformation.
BIBLIOGRAPHY
1. Kapfhammer, J., and Bischoff, C., 2. physiol. Chem., 191, 179 (1930).
2. Beattie, F. J. R., Biochem. J., 30, 1554 (1936).
3. Jacobi, H. P., Baumann, C. A.,, and Meek, W. J., J. Biol. Chem., 138, 571 (1941).
4. Thornton, M. H., and Broome, F. K., Ind. and Eng. Chem., Anal. Ed., 14, 39
(1942).
Alerts:
• When this article is cited
• When a correction for this article is posted