Beruflich Dokumente
Kultur Dokumente
MARK III
JET FUEL THERMAL OXIDATION TESTER
USER’S MANUAL
PN: AL-59322
Revision B
MS Windows®, MS Windows® X P, and MS Windows Professional® are registered trademarks of Microsoft Corp.
REVISION HISTORY
Rev Revision Description Rev Date
A Original Release 2-28-06
B Changes to Fuses and Coolant Pump Added VTDR Option. 5-18-06
ALCOR JFTOT III Users Manual
TABLE OF CONTENTS
ABOUT THIS MANUAL......................................................................................................... ii
Abbreviations and Symbols...................................................................................................................iii
Proper Use and Safety ........................................................................................................................... iv
SECTION 1: INTRODUCTION
SECTION 2: INSTALLATION
SECTION 3: OPERATION
SECTION 4: CALIBRATION
The following graphic symbols are used in the JFTOT III to inform the user of possible
hazardous areas on the JFTOT III. The first three symbols are used in this user’s manual to
inform the reader of hazardous conditions during the performance of a procedure.
Symbol Definition
This symbol (three wavy lines over a straight line in a triangle) indicates a hot
surface. During operation, the JFTOT III surfaces near the heater test section get
hot. To prevent burns, exercise caution when operating the JFTOT III. Do NOT
open the protective cover while the instrument is in use.
NOTE: The JFTOT 230 MARK III Jet Fuel Thermal Oxidation Tester is commonly
referred to as the “JFTOT III”.
The User’s Manual contains certain abbreviations and symbols as described in the
following list.
Abbreviation Unit
A Amps
° degrees
g grams
h hours
Hz Hertz
JFTOT Jet Fuel Thermal Oxidation Tester
JFTOT III JFTOT 230 MARK III
K kilo = x 103
L liters (mL = milliliters)
M mega = x 106
m meter (mm = millimeter, cm = centimeter)
min minutes
mLpm milliliters per minutes
Pa Pascal = Newton/meter2
PCB Printed Circuit Board
psi pounds force/inch2 (psig = psi gauge)
rms root mean square
s seconds
V Volts
VA Volt-amperes
W Watts
The JFTOT III is intended for indoor laboratory use only. This equipment has been
designed and tested in accordance with EN 61010, “Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use.” Use of this product in an
unsafe environment or in a manner not specified may impair the protection and
performance designed into the equipment.
This user’s manual contains information, warnings, and cautions that must be followed to
ensure safe operation and to maintain the equipment in safe operating condition. Safety
glasses should be worn by the operator and anyone in the vicinity of the equipment for
protection from splashed materials. Tests performed with this equipment may involve
hazardous materials.
This manual does not purport to address all safety concerns associated with the operation
of the JFTOT III. It is the responsibility of the user of this equipment to establish
appropriate safety and health practices and to determine applicability and limitations prior
to use.
CAUTION: Do not use this equipment in a hazardous atmosphere. Do not use this
equipment with hazardous materials for which the equipment is not
designed. Protection provided by this equipment may be impaired if used
with accessories not provided or recommended by the manufacturer or used
in a manner not specified by the manufacturer.
Servicing and repair should be performed only by qualified personnel. Disconnect the
equipment from the power source when servicing or replacing electrical components.
Service replacement parts and technical assistance for this equipment may be obtained
from:
PAC
8824 Fallbrook Dr.
Houston, TX 77064
Tel: 281-580-0339 or 800-444-TEST
Fax: 281-580-0719
sales@paclp.com
www.paclp.com
I. Limited Warranty
1. Limited Warranty. Petroleum Analyzer Company, L.P. (“PAC”) offers a limited warranty on each of its
products against failure due to defects in material and workmanship for a period ending the earlier of
(i) eighteen (18) months from the date of the invoice relating to the sale of the product and (ii) twelve
(12) months from the date of installation of the product (collectively, the “Initial Warranty”). During
the Initial Warranty, PAC offers a limited warranty against failure due to defects in material and
workmanship on each part of a product repaired or replaced by an authorized service person for a period
ending the later of (a) the remaining term of the Initial Warranty of the product and (b) ninety (90) days
from the date of such repair or replacement. After expiration of the Initial Warranty, PAC offers a
limited warranty against failure due to defects in material and workmanship on each part of a product
repaired or replaced by an authorized service person for a period ending ninety (90) days from the date
of such repair or replacement. PAC further offers a limited warranty that the products and parts it sells
will conform to PAC’s written specifications therefor. The foregoing limited warranties cover parts and
labor only and PAC does not warrant and will not reimburse the buyer of its products (“Buyer”) for any
costs relating to the access by service persons of PAC to the product at issue. The foregoing limited
warranties cover only the repair or replacement of defective parts and such determination will be in the
sole discretion of PAC. In its sole discretion, PAC may make repairs or replacements under these limited
warranties with either new or refurbished parts. To the extent Buyer’s product cannot be remedied
under these limited warranties through repair or replacement of parts, Buyer may return the product for
a refund of the purchase price, less a reasonable reduction in such purchase price equal to the
depreciation expense incurred by Buyer relating to such product. The limited warranties of this
Section I.1. are further subject to those warranty exclusions set forth below in Section I.2.
2. Limited Warranty Exclusions. EXCLUDING THE WARRANTIES PROVIDED FOR IN SECTION I.1., PAC
PROVIDES ALL PRODUCTS TO BUYER “AS-IS,” WITHOUT ANY OTHER WARRANTY OF ANY KIND. PAC
DISCLAIMS ANY AND ALL EXPRESS OR IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A
PARTICULAR PURPOSE AND NON-INFRINGEMENT OF THE INTELLECTUAL PROPERTY OF OTHERS. PAC
MAKES NO WARRANTY, EXPRESS OR IMPLIED, AS TO THE DESIGN, SALE, INSTALLATION OR USE OF ITS
PRODUCTS. PAC’S WARRANTIES WILL NOT BE ENLARGED BY, NOR WILL ANY OBLIGATION OR LIABILITY
OF PAC ARISE DUE TO PAC PROVIDING TECHNICAL ADVICE, FACILITIES OR SERVICE IN CONNECTION
WITH ANY PRODUCT. There is no warranty by PAC with respect to any product’s: (i) uninterrupted or
error-free operation; (ii) actual performance, other than the product’s capability to meet PAC’s
specifications therefor; (iii) removal or installation from a worksite or process; (iv) electronic
components or associated accessories (including without limitation circuit boards and integrated
circuits); (v) maintenance (including without limitation gasket and seal replacements, adjustments,
minor repairs and other inspection requirements, preventative or otherwise); (vi) use under
inappropriate conditions or not in accordance with operating instructions; or (vii) use in connection
with the operation of a nuclear facility. There is no warranty for labor expenses associated with field
repairs or the repair or replacement of defective parts in the engine or power unit of any product if such
product has been in the possession of the owner or operator for greater than twelve (12) months. There
is no warranty for products determined to be, in PAC’s sole discretion, damaged as a result of (a) misuse,
neglect or accident; (b) improper application, installation, storage or use; (c) improper or inadequate
maintenance or calibration; (d) operation outside of the published environmental specification;
(e) improper site preparation or maintenance; (f) unauthorized repairs or replacements;
(g) modifications negligently or otherwise improperly made or performed by persons other than PAC;
(h) Buyer-supplied software or supplies; (i) use in conjunction with or interfacing with unapproved
accessory equipment; (j) use of ABC-style or dry powder fire suppression agents; or (k) leaked sample
materials. To the extent a PAC product is used in connection with the operation of a nuclear power
facility, Buyer agrees to indemnify and hold PAC harmless from any and all actions, claims, suits,
damages and expenses arising from such use. PAC provides no warranty on the oral representations
made by its personnel while they are attempting to assist Buyer in the operation of a product. This
Standard Limited Warranty does not apply to items consumed by the products during their ordinary use,
including but not limited to fuses, batteries, paper, septa, fittings, screws, fuses, pyrolysis, dryer or
scrubber tubes, sample boats, furnaces or UV lamps.
3. Non-PAC Products. PAC does not in any way warrant products it does not manufacture except to the
extent the warranty of the manufacturer of the product at issue passes through or is otherwise assigned
to PAC. If a manufacturer warranty is so assigned to PAC, PAC will only be bound to comply with the
length of time associated with such warranty. All other terms of such warranty will be governed by this
Standard Limited Warranty and PAC’s General Terms and Conditions incorporated herein by reference.
4. Expenses on Non-Warranty Work. All repairs or replacements by PAC after the expiration of any
applicable limited warranty period will be performed in accordance with PAC’s standard rate for parts
and labor. Further, if upon PAC’s inspection and review, PAC determines the condition of the products
is not caused by a defect in PAC’s material and workmanship, but is the result of some other condition,
including but not limited to damage caused by any of the events or conditions set forth in Section I.2.,
Buyer shall be liable for all direct expenses incurred by PAC to conduct the inspection and review of the
product.
5. Exclusive Remedy. THE FOREGOING LIMITED WARRANTY CONSTITUTES BUYER’S EXCLUSIVE REMEDY
WITH RESPECT TO PRODUCTS SOLD BY PAC AND PAC’S LIABILITY SHALL BE EXCLUSIVELY LIMITED TO
THE WRITTEN LIMITED WARRANTY SPECIFIED HEREIN. No employee, representative or agent of PAC is
authorized to either expressly or impliedly modify, extend, alter or change any of the limited warranties
expressed herein to Buyer.
6. Procedure and Costs. All limited warranty claims must be made in writing promptly following discovery
of any defect. Buyer must hold defective products for inspection by PAC. If requested by PAC, Buyer
must send the product to PAC for inspection. Any such returns by Buyer will be at Buyer’s expense and
Buyer will remain liable for any loss of or damage to the product during such product’s transportation to
PAC. No products will be sent to PAC for inspection unless PAC has authorized Buyer to do so.
7. Terms and Conditions. PAC’s General Terms and Conditions are incorporated herein by reference and
Buyer accordingly agrees to be bound by the terms thereof.
Thank you for choosing our line of analyzers. In order to serve your needs during the life of this
instrument, please complete the sections in the Warranty Registration Certificate form and return a
copy to PAC FAX or Email attachment to:
Fax 281.580.0719
sales@paclp.com
Contact Information:
E-mail Address
Instrument Information:
Model Number
Serial Number
Installation Date
PAC Representative
Additional Remark:
www.paclp.com
Phone 800-444-TEST or 281.580.0339
SECTION 1: INTRODUCTION
PN: AL-59322
Revision B
MS Windows®, MS Windows® X P, and MS Windows Professional® are registered trademarks of Microsoft Corp.
REVISION HISTORY
Rev Revision Description Rev Date
A Original Release 02-28-06
B Changes to Fuses and Coolant Pump. Added VTDR Option. 05-18-06
ALCOR JFTOT III Users Manual Section 1: Introduction
SECTION 1:
INTRODUCTION
1.1 OVERVIEW
The JFTOT 230 Mark III Jet Fuel Thermal Oxidation Tester (JFTOT III) is a fully
integrated bench-top instrument that, along with a visual or video tube rater, provides
everything needed to safely study the thermal oxidation of any petroleum-based liquid
fuel. The instrument is normally used in a batch mode where a fuel sample is held in a
reservoir and pumped across a heated metal surface to a second reservoir.
The purpose of the JFTOT III system is to measure formation of deposits from jet fuel
when heated in the presence of dissolved oxygen and a metal surface. The deposits
can then be measured and compared against a standard to determine the degree of fuel
degradation due to thermal oxidation.
The JFTOT III has been significantly improved to allow simplified setup and user-
friendly operation through a standard MS Windows® based PC touch-screen and
menu-driven JFTOT III software.
The JFTOT III (Figure 1-1) can be set for either 115 or 230 VAC power.
FIGURE 1-1
JFTOT III
1.1.2 SPECIFICATIONS
The ALCOR JFTOT III is designed to conform to the following specifications:
• Man Machine Interface (MMI): Front panel mounted Pentium III PC with
Touch Screen LCD capable of operating as a standard PC through MS
Windows® XP Professional® programs; 12-inch mini-keyboard; mouse; 40-GB
hard drive; CD-ROM drive; speaker ports, serial and USB ports.
• Heater-Tube temperature setpoint range: 100°C to 380°C
• Fuel sample flow-rate range: 0.001mL/min to 9.999 mL/min
• Flow rate accuracy: ±2%
• Pump: HPLC, Dual Head SS
• Fuel Sample Volume Capacity: Unlimited
• Thermocouple type: K (Chromel/Alumel)
• Thermocouple temperature range: 0°C to +500°C
• DP range: 0 to +280 mmHg (automatically bypassed at +250 mmHg)
• Operating system pressure range: 3.45 MPa (500psig) ± 10%
• Automatic test numbering: 1 to n (n=hard drive capacity)
• Test time range: programmable 1 min. and up
• Fuel aeration timer: 6 minutes
• Aeration Flow rate: 1.5 L/Min (Green)
• Coolant flow rate: 10 Gal/Hr (Green)
• JFTOT III ambient temperature range: +10°C to +35°C (ASTM D 3241
test limited to +15°C to 32°C)
• Operating power voltage: 115VAC ±10% (for 85436-115V) single phase
or 230VAC ±10% (for 85436-230V) single phase
1.1.4 METHODS
The following methods can be used for the JFTOT III.
• ASTM D3241
• IP 323
• ISO 6249
For optional equipment, see end of Section 5.8 in Maintenance and Service.
The JFTOT III software provides a main screen made up of the following major
sections:
• Mode Selection (left)
• Graphics Display and Message Area (center)
• Digital Displays (right)
Submenus allow the user to perform and monitor the functions for the
selected mode.
Since the embedded computer runs MS Windows® XP Professional OS® software it is
capable of operating any additional MS Windows® XP Professional® programs
available for standard PCs. A 40-GB hard drive provides storage for system programs
and data.
NOTE: A watchdog timer is used by JFTOT III to monitor program activity. If for
some reason the PC locks up or the program simply stops running, the
watchdog timer will remove power from the hardware and completely vent
the pressure from the system.
DAQ 24-VDC
Computer/JFTOT III Interface PCB
Touch Screen, Signals 24-VDC Supply Power Distribution
Computer and PCB
Bus Relay Cont
DAQ Assemblies
24 VDC Sample DP Sol Cont Bar Tube RMS RMS
OUT Temp In Out (x5) Temp Temp Volts Amps From Syst
Pres Xdcr 115 115/230
RS-232 Com 4-20 VAC
VAC
& En ma
To To Air
Pump To Step-Down
To Sample % RH To Purge
To Samp SCR Heater Control Transformer
Drive Pump Sensor and Vent
Coolant Drive (230/115VAC)
Solenoids
Pump Pump
Sol Cont
+ Purge 115 or
+ 230 VAC Power Entry
Sol
Module
Valve
Diff Bypass
Pressure Sol Flow Coolant
Xducer Valve Meter Container
Sol Cont
– Purge
Sol – DP
Valve Filter
Back
Pressure 24VDC From Cooling Check
Regulator Computer Pump Valve
System Syst Interface PCB
Vent Sol Press
Sol Cont Valve Xdcr
To Comp
IF PCB
Safety
Sol Cont Vent Sol Radiator
Valve Pre-
Assembly
Filter
Radiator
Fan
Sample and Purge Sample Com & Enable 115VAC
OUT Container Sample-IN Container Drive To Computer To Fans from
Pump Interface PCB Power Dist PCB
% RH Flow
Air Pump Dryer Vent Fan
Sensor Meter
RS-232 To
Computer
The amount of deposits which adhere to the tube and the degree of plugging of the
differential pressure filter are indicative of the thermal stability of the fuel at the test
temperature. After the prescribed test period (150 minutes for ASTM D 3241) the
maximum pressure drop across the filter is measured and recorded. The heater tube
is inspected to determine the tube rating. Both the tube rating and the pressure drop
data are typically part of the specification for jet fuel.
Touch Screen
User Interface
Sample Container
Area
Drive Pump
Enclosure
FIGURE 1-5
JFTOT III SAMPLE CONTAINER AREA
Aeration Tube Sample Out Line
Sample-In
Thermocouple
Sample Out
Sample-In
Container
Container
Heater Tube
Bus-Bar
Thermocouple
Thermocouple
– DP Port (Out)
DP Filter
Purge/Priming
Valve and Port
Reciprocating
Displacement Piston
Pump Heads
Anti-Siphon
Check valve
Input Line
Fittings
Output Line Fitting
The left side of the screen contains five mode buttons that allow the
user to enter the Run Default Test mode, Run Custom Test mode,
Maintenance (Manual) mode, Auto-calibration mode or Report mode.
The center section displays a real-time chart of bus-bar temperature,
watts dissipated and heater tube temperature and a differential
pressure chart for the test operation in progress or stored data from
previous tests.
A display section at the bottom of the center section gives messages
pertaining to the status of the operation in progress. The right section
of the display provides digital readouts for parameters measured during
the test.
After the test is complete, the main screen is displayed with the
temperatures charted as shown in Figure 1-9.
Refer to Appendix 3-A and 3-B in Section 3 for more information on
the JFTOT III program and detailed descriptions of the main screen
and subordinate screens.
FIGURE 1-9
FRONT PANEL TOUCH TEST COMPLETE SCREEN DISPLAY (EXAMPLE)
FIGURE 1-10
BACK PANEL COMPONENTS
Vent Fan
Fuses (x5)
Dryer
Connectors
SECTION 2: INSTALLATION
PN: AL-59322
Revision B
MS Windows®, MS Windows® X P, and MS Windows Professional® are registered trademarks of Microsoft Corp.
REVISION HISTORY
Rev Revision Description Rev Date
A Original Release 02-28-06
B Changes to Fuses and Coolant Pump. Added VTDR Option. 05-18-06
ALCOR JFTOT III Users Manual Section 2: Installation
SECTION 2:
INSTALLATION
2.1 INSTALLATION REQUIREMENTS
Prior to installing the JFTOT 230 Mark III Jet Fuel Thermal Oxidation Tester
(JFTOT III), environmental and operational requirements must be considered in
order to ensure proper installation and future performance. The following
requirements must be considered before installing the JFTOT III.
NOTE: Allow space in front of the JFTOT III to open the sample pump enclosure
door and to open the sample containers and test area covers.
NOTE: Allow space in back of the JFTOT III to remove the Dri-Rite container, to
access the power entry module and to connect power and signal cables.
NOTE: Allow space on each side of the JFTOT III to remove the side panels and
access the internal components.
CAUTION: Fan exhaust from the JFTOT III may affect other instruments and surfaces in
back of the JFTOT III.
CAUTION: Do not allow fuel or cleaning fluids to spill onto the keyboard or mouse.
After determining the location for the JFTOT III as required in Section 2.1, unpack
and install the JFTOT III by performing the following procedures.
NOTE: See Appendix 2-A at the end of this section to report damages due to
shipping.
FIGURE 2-1A
JFTOT STANDARD ACCESSORIES AND LINES
NOTE: See Section 5.8 for a listing of spares, accessories, supplies, and materials.
Power Cord
MS Windows XP
Professional OS
Software
Keyboard
Driver Software
Mouse
JFTOT III Software
FIGURE 2-1 D
JFTOT III SPARES, ACCESSORIES, SUPPLIES AND MATERIAL
Aluminum Foil
(AL-38008)
Heater Tube Heater Tubes (2
Insulators packs of 10)
(2 packs of 4) (p/n AL-91652)
(AL-11926)
Packaged Items Standardization Kit
(Figure 2-1E) (AL-37633)
FIGURE 2-1 E
JFTOT III PACKAGED SPARES, ACCESSORIES, SUPPLIES AND MATERIAL
Hex Driver
(AL-38016) O-Rings (Pack of 25)
(AL-53706)
Brush (AL-09970)
WARNING: Physical injury and equipment damage may occur unless at least 3 people
capable of lifting at least 100 pounds each are used to lift the JFTOT III into
place.
NOTE: For 115 VAC use one 8A-3AG, Fast (PN: AL-29686) and a shorting clip.
For 230 VAC use two 6.3A, 5x20mm Slow-Blow (PN: AL-29738) without the
shorting clip. If metal shorting clip is not available, install an 8A Fuse in both
sides of the holder.
FIGURE 2-10
AERATION DRYER INSTALLATION
Top Fittings
Clip Mounts
Dri-Rite
Dryer Assembly
Chemicals
(Blue)
Barbed Fitting
3. Thread the top fitting onto the top of the dryer assembly fitting and press the
bottom line over the bottom tapered fitting. Tighten the air- in and air-out
fittings to finger tight.
4. If necessary, loosen the top end cap and rotate the window so that it can be
viewed from the side of the JFTOT. Retighten the end cap.
5. Test the aeration system for leaks.
6. Check color of blue Dri-Rite chemicals. Pink indicates Dri-Rite chemicals need
to be replaced.
Bus-Bar
Thermocouple
Connection
Heater Tube
Clamp Assembly
Upper Bus-Bar
Thermocouple Assembly
Wire Insertion
Thermocouple Hold-
Down Clamp Set Screw
Stop Adjustment
Thermocouple Hold-
Screw
Down Clamp
2. Insert the thermocouple wire into the orifice (Figure 2-11) near the end cap of
the upper bus bar.
3. With the thermocouple hold-down clamp set screw to the right and the flat
section of the clamp at the bottom, align thermocouple slot with the
thermocouple wire and slide the thermocouple hold-down clamp over the
thermocouple wire and onto the upper bus bar (Figure 2-11) until it stops. If
necessary, adjust the stop screw on the flat section of the clamp.
4. Tighten the set screw to hold the thermocouple hold-down clamp in place.
5. Install the heater-tube clamp assembly (Figure 2-11) onto the thermocouple hold-
down clamp using the two hex screws.
2. Insert the metal vent lines (Figure 2-12) into the front panel bulkhead fittings as
shown in Figure 2-12.
3. While keeping the end of the line aligned, finger-tighten the knurled nut
(Figure 2-12) onto the bulkhead fittings.
FIGURE 2-12
SAMPLE AND VENT OUTPUT TUBES
O-Ring
Nut
FIGURE 2-13
JFTOT III POWER, KEYBOARD, MOUSE AND PRINTER CABLES I
The basic installation procedures are now complete as shown in Figure 2-14. The
JFTOT III should now be ready for the initial startup as described in Section 3.1.
FIGURE 2-14
JFTOT III COMPLETED INSTALLATION
APPENDIX 2-A:
SHIPPING DAMAGE REPORTS
Shipments are inspected immediately prior to being packaged for transportation. Since
our packaging exceeds all freight specifications, your equipment should have arrived
in perfect condition.
Our terms of sale are through Ex Works at shipping point, meaning that title to these
goods is passed to the final receiver when the goods were transferred to the carrier at
our factory. If any part of the shipment is damaged, we suggest reporting the damage
immediately to the transportation company. The following procedure for reporting
damaged equipment is suggested.
CAUTION: Before shipping the JFTOT III or accessories, flush all sample fuel from the
lines and clean all glass containers.
Common Carrier:
If the item is discovered at the time of delivery, have the driver note the nature and
extent of the damage on the carriers copy and on the delivery copy of the freight bill.
If the item is discovered after delivery, call the carrier immediately after discovering
the damage, and request that an inspection be made as soon as possible. Perform the
following steps when damage items are found after delivery.
1. Contact the carrier immediately for an inspection of the item(s).
NOTE: It is essential that the damaged item(s) be left in the original shipping
container(s).
NOTE: Claims on air shipments must be filed within 270 days. Other claims must be
filed within 9 months of the shipping date.
3. File the claim as quickly as possible, even if you have to estimate the dollar
amount of the damage.
Parcel Delivery Service (UPS, DHL, FEDEX, etc):
Follow the steps listed above for reporting damages shipped by common carriers, and
notify our office in Houston, Texas by dialing 281-5800-339. Our Technical Service
Department will help you process your claim.
NOTE: Do not return the system or any item to the factory without factory
authorization.
File Name: JFTOT III User’s Manual Part No. AL-59322
Revision B Page 2A-1
ALCOR JFTOT III Users Manual Appendix 2-A: Shipping Damage Reports
Unless otherwise specified, UPS is the preferred primary carrier for ALCOR
shipments. Follow the steps listed above for reporting damages shipped by common
carriers, and notify our Customer Service Department, which will take the necessary
action for your claim.
NOTE: Do not return the system or any item to the factory without factory
authorization.
CAUTION: Before shipping the JFTOT III or accessories, flush all sample fuel from the
lines and clean all glass containers.
APPENDIX 2-B:
INSTALLATION PROBLEM REPORT
Complete the following report if there are any problems with the installation of the
JFTOT III and send it by email attachment to the PAC Service Department.
E-mail:
Problem(s)
Model: S/N:
Model: S/N:
Model: S/N:
SECTION 3: OPERATION
PN: AL-59322
Revision B
MS Windows®, MS Windows® X P, and MS Windows Professional® are registered trademarks of Microsoft Corp.
REVISION HISTORY
Rev Revision Description Rev Date
A Original Release 02-28-06
B Changes to Fuses and Coolant Pump. Added VTDR Option. 05-18-06
Section 3: TABLE OF CONTENTS
SECTION 3: OPERATION ...................................................................................................... 1
3.1 INITIAL STARTUP PROCEDURES ............................................................................ 1
3.1.1 Help Screen .............................................................................................................................2
3.1.2 Preparation .............................................................................................................................2
3.2 RUNNING A DEFAULT TEST..................................................................................... 3
3.3 CLEANING.................................................................................................................... 4
3.4 ASSEMBLY ................................................................................................................... 8
3.4.1 Pre-Filter Assembly and Installation.....................................................................................8
3.4.2 Heater Tube Test Section Assembly Procedures .................................................................9
3.4.3 Test Section Installation Procedures.................................................................................. 12
3.4.4 Preparation of Test Fuel ..................................................................................................... 14
3.5 DISASSEMBLY ........................................................................................................... 16
3.6 SHUTDOWN ............................................................................................................... 20
SECTION 3:
OPERATION
3.1 INITIAL STARTUP PROCEDURES
After completing the basic installation procedures for the JFTOT 230 Mark III Jet
Fuel Thermal Oxidation Tester (JFTOT III), the JFTOT III can be setup to run a
test. The JFTOT III has been thoroughly tested by PAC before delivery. The factory
tests are stored in files on the JFTOT III hard drive for evaluation and comparison to
tests performed after installation.
NOTE: Become familiar with the location and use of all controls, indicators,
connections, and accessories. Carefully read all instructions, warnings
cautions and labels prior to operating any portion of the system. Failure to
do so could result in personal injury and equipment damage.
Perform the following steps to startup the newly installed JFTOT III.
1. Locate the JFTOT III power-on/off switch on the back panel in the power entry
module and place the switch in the ON position. Ensure the power switch in
back of the computer near the power cord is always in the ON (–) position.
2. Press the computer power switch on the front panel (lower-left of computer).
The JFTOT III should power up through the normal Windows power up
routine. The JFTOT III program is automatically loaded from the Windows OS
and will display the JFTOT III main screen after the loading process has
completed. The main menu screen shown in Figure 3-1 should appear.
FIGURE 3-1
JFTOT III MAIN MENU SCREEN (EXAMPLE)
NOTE: To start the JFTOT III program from the Windows desktop, touch or click
the JFTOT III icon on the Windows Desktop or click
Start>Programs>JFTOT III. To return to the Windows desktop from the
JFTOT III main menu, click or press the EXIT button on the main menu.
NOTE: Refer to Appendix 3-A in this section of the user’s manual for detailed
descriptions of the structure and screens displayed by the JFTOT III
application program and Appendix 3-B for a detailed description of the
JFTOT III operations and how to setup DEFAULT test parameters.
3.1.2 PREPARATION
Before you can begin a JFTOT test as described in Subsection 3.2, the components
exposed to jet fuel musts be cleaned and reassembled as described in Sections 3.3 and
3.4. The following replacement components must be obtained for each test:
• Pre-filter element (AL-238057 - box of 25)
• Heater tube and filter kit (AL-91652 - box of 10)
• General-purpose, retentive, qualitative filter paper (AL-27052 - box of 100)
1. Press or click the RUN DEFAULT button on the main screen (right). The
Test Annotation screen (below left) appears.
2. Fill in the blanks and then click OK. An instruction box (below right) will
appear prompting you to perform setup.
3. Refer to Section 3.3 and 3.4 in the User’s Manual for detailed setup
procedures.
4. After completing the setup steps, press or click the CONTINUE button.
5. Adjust air and coolant flowmeters so that the ball is in the center of the
green range.
6. When the option box (below) appears, press or click YES to save the file
or NO if the data is to be deleted.
10. Perform disassembly and cleanup steps. Refer to Section 3.5 in the
User’s Manual for detailed disassembly and cleanup procedures.
11. Refer to Appendix 3-C in the User’s Manual for instructions on heater tube
deposit rating.
3.3 CLEANING
WARNING: These procedures involve handling flammable liquids and exposure to skin
irritants and harmful vapors. Wear protective clothing, gloves and mask
when setting up and performing the JFTOT III tests. Ensure flames and
spark producing materials are kept well away from the JFTOT III test area.
Use standard safety procedures at all times.
Before you can setup to run a JFTOT test, all components containing test fuel must
be clean and free of contaminants. Perform the following steps to clean the
components.
1. Place a new piece of aluminum foil about 450 mm square or larger on the bench
for placing all components after cleaning.
2. Position a cleaning pan to catch the solvent during the cleaning operations.
WARNING: There are three appropriate solvents for general cleaning: cyclohexane, n-
heptane, and 2,2,4 trimethyl pentane (isooctane). These solvents are
extremely flammable and the vapors are harmful. Wear protective clothing,
gloves and mask when using these solvents. Ensure flames and spark
producing materials are kept well away from the area. See Appendix 3-D.
4. Flush the pre-filter components with solvent as shown in Figure 3-3A through
3-3C and thoroughly blow dry line using a squeeze bulb as shown in
Figure 3-3D.
FIGURE 3-3
PRE-FILTER COMPONENTS FLUSHING
A B C D
5. Flush the fuel sample inlet line with solvent as shown in Figure 3-4A and
thoroughly blow dry using a squeeze bulb.
6. Flush the fuel sample inlet line and the test section outlet line with solvent as
shown in Figure 3-4 and thoroughly blow dry using a squeeze bulb.
FIGURE 3-4
FUEL-INLET LINE AND TEST SECTION OUTLET LINE FLUSHING
A B
Test Section
Outlet Line
Solvent
Fuel Sample (N-Heptane)
Inlet Line
7. Flush the sample-in thermocouple, sample-in line and the aeration line with
solvent.
8. Flush all inside surfaces of the fuel sample-in container and sample-out
container with solvent as shown in Figure 3-5.
FIGURE 3-5
FUEL-IN CONTAINER FLUSHING
9. Using the stiff nylon brush (supplied) saturated with trisolvent, brush the inside
surfaces of the test section while flushing with trisolvent as shown in Figure 3-
6. This is the only component that requires cleaning with trisolvent.
10. Visually inspect the inner surfaces of the test section cleanliness and repeat
cleaning with trisolvent as required to remove all deposits.
FIGURE 3-6
HEATER TEST SECTION FLUSHING
Stiff Nylon
Brush
Heater Tube
Housing
Solvent
(Tri-Solvent)
11. Flush the test section and its bypass line with solvent and blow dry with a
squeeze bulb as shown in Figure 3-7.
FIGURE 3-7
SQUEEZE-BULB DRYING
Test Section
Squeeze Bulb
12. Flush the insulator bushings and test section nuts with solvent.
13. Inspect all o-rings for cuts, abrasions, and excessive swelling and replace as
necessary. Inspect the insulators and replace if they are cracked, chipped, or
otherwise damaged. Inspect all stainless steel components for damage and
replace as necessary.
14. The cleaned parts are now ready for assembly (Figure 3-8).
FIGURE 3-8
CLEANED COMPONENTS
3.4 ASSEMBLY
3.4.1 PRE-FILTER ASSEMBLY AND INSTALLATION
Perform the following steps to install a new pre-filter element in the pre-filter assembly.
NOTE: For each test, a new 25 mm diameter, 0.45 µm membrane filter element
(white in color) must be installed along with a clean backup screen, o-ring
and housing sections as shown in Figure 3-9.
FIGURE 3-9
JFTOT III PRE-FILTER ASSEMBLY
Pre-Filter Housing
Pre-Filter Housing Screws (x3)
Bottom Plate and
Inlet Line
Pre-Filter O-Ring
Pre-Filter Element
25 per box
(p/n AL-38057)
Pre-Filter Backup
Screen
Pre-Filter Housing
Base Plate
3. Install the large o-ring inside other half of the pre-filter housing and the small
o-ring on the tube connection as shown in Figure 3-10C.
FIGURE 3-10
PRE-FILTER ELEMENT INSTALLATION
A B C
Filter Backup Screen
Small O-Ring
Large O-Ring
Filter Element (White)
4. Assemble the two housing sections together (Figure 3-11A) by inserting the
three screws through the lower section and tightening securely.
5. Connect the pre-filter assembly to the fuel-in fitting (Figure 3-11B) and finger-
tighten snugly.
FIGURE 3-11
PRE-FILTER HOUSING ASSEMBLY
A B
Pre-Filter Assembly
Fuel-In Fitting
O-Rings (x3)
Heater Tube
Container
CAUTION: During assembly wear clean, dry gloves. Handle the heater tube by the ends
only. If the center section of the heater tube is touched, the tube must be
rejected as the deposit forming characteristics on the tube will be adversely
affected.
CAUTION: Do not reuse heater tubes. Tests indicate that magnesium migrates to the
heater tube surface under test conditions. Surface magnesium may reduce
adhesion of deposits to the heater tube surface.
CAUTION: Use caution when removing the heater tube from its protective plastic
container. To prevent bending of the tube, hold the plastic container at one
end with two or three fingers and pull the cap off (do not pry).
1. Remove the yellow cap from the end of the plastic heater-tube container and
then remove the DP filter and three o-rings from the cap.
2. Carefully remove the heater tube from the plastic tube container.
3.1 Use a visual tuberator to inspect the center section (see Appendix 3-C,
Heater Tube Deposit Rating). Reject any heater tube showing scratches,
unpolished area, or other visual defects.
3.2 Examine the heater tube for straightness by rolling the tube on a clean, flat
surface noting any variation in gap between the surface and the heater
tube center section. Reject any bent heater tube.
CAUTION: Do not touch or allow anything to contact the heater tube center section
during the assembly or disassembly of the heater tube test section.
4. Insert the heater tube into the clean test section until it is centered in the
housing as shown in Figure 3-13.
FIGURE 3-13
JFTOT III HEATER TUBE INSTALLATION
Heater Tube
Test Section
Heater Tube Heater Tube
Serial Number
5. Onto each end of the heater tube, sequentially install a tapered insulator (wide
end out), high temperature o-ring, shoulder insulator (large end first), and hex
nut as shown in Figures 3-14A and 3-14B. Lightly finger tighten nut with the
heater tube approximately centered in the housing.
FIGURE 3-14
JFTOT III HEATER TUBE INSTALLATION
A B
Hex Nut
Shoulder Insulator
Shoulder Insulator
Hex Nut
6. Observe the heater tube through the fuel discharge hole and align the upper
heater-tube shoulder in the center of the discharge hole as shown in
Figure 3-15.
FIGURE 3-15
JFTOT III HEATER TUBE POSITION
Heater Tube
Shoulder
7. While observing the correct heater tube shoulder alignment in the discharge hole,
tighten both hex nuts firmly with fingers only. DO NOT USE A WRENCH.
8. Using clean tweezers, install the DP filter (Figure 3-16A) in the outlet chamber of the
heater tube housing with the RED COLORED SIDE OUT. Place a new o-ring on top
of the DP filter (Figure 3-16B) and press the o-ring in until it rests against the filter
FIGURE 3-16
DP-FILTER INSTALLATION
A B
DP Filter
(Red Side Out) O-Ring
9. Attach the outlet line to the test section as shown in Figure 3-17.
FIGURE 3-17
OUTLET LINE INSTALLATION
CAUTION: If the bus-bar caps have been completely removed, check for proper mating.
Matching numbers are stamped on the inside surfaces of the cap and bus
bar. These numbers must match and face each other. To ensure consistent
reassembly, the caps are normally not completely removed from their
respective bus bar.
2. If both screws are in the bus-bar caps, remove one screw and rotate the end
caps to allow installation of the heater tube assembly.
3. Install the new heater tube assembly into the top and bottom bus bars and then
swing the caps into place over the tube and finger tighten the upper and lower
bus-bar screws as shown in Figure 3-18A.
4. Ensure the top of the heater tube is flush with the top of the upper bus bar
(Figure 3-18B) and then use the Allen wrench provided to securely tighten the
upper and lower bus-bar screws as shown in Figure 3-18C.
FIGURE 3-18
JFTOT III HEATER TUBE TEST SECTION INSTALLATION
A B C
5. Attach the test section bypass and outlet lines to the bulkhead fittings as shown
in Figure 3-19. Make sure o-rings are on the ends of the both lines. Tighten
fittings with fingers until snug.
FIGURE 3-19
OUTLET LINE AND BYPASS LINE INSTALLATION
6. Ensure an o-ring is in place at both fuel input line fittings and then connect the
fuel input line to the pre-filter fitting and the other end to the fuel-in fitting on
the test section as shown in Figure 3-20. Tighten both fittings with fingers until
snug.
FIGURE 3-20
FUEL INPUT TUBE INSTALLATION
Fuel-Input
Line Pre-Filter
Fitting
7. If not already installed, plug the heater tube thermocouple connector into the
connector above the bus-bar thermocouple connector as shown in
Figure 3-21A.
8. Carefully, insert the thermocouple tip into the hole at the front of the upper end
of the heater tube (Figure 3-21A) and lower until it stops.
9. Move the thermocouple retaining clip into place around the thermocouple as
shown in Figure 3-21B.
FIGURE 3-21
JFTOT III THERMOCOUPLE INSTALLATION
A B
Heater Tube
Thermocouple
Connector
Thermocouple
Retaining Clip
Heater Tube
Thermocouple
Heater Tube
Thermocouple
Tip
2. Place the specified filter paper in a clean funnel and place the funnel into a
clean container.
3. Carefully pour the test fuel into the filter paper and let it drain through the
paper into the container as shown in Figure 3-22.
FIGURE 3-22
TEST FUEL MEASUREMENT
Graduated Cylinder
Filter Paper
(AL-27052, box of 100)
Funnel
Filtered Fuel
Container
4. Pour the filtered fuel into the JFTOT III test sample-in container and place the
container under the aeration and sample-in fittings.
5. Connect the aeration tube and the fuel sample-in tube quick-disconnect fittings
to the appropriate fittings on the JFTOT III front panel as shown in
Figure 3-23.
6. Insert the sample thermocouple into the sample-in container and then plug the
connector to the sample thermocouple connector on the JFTOT III front panel
as shown in Figure 3-23.
FIGURE 3-23
SAMPLE-IN CONTAINER TUBES AND THERMOCOUPLE INSTALLATION
Aeration Quick-
Sample Connect Fitting
Thermocouple
Sample-In Quick-
Connector
Connect Fitting
Aeration Tube
Sample Assembly
Thermocouple Sample-In Tube
(AL-86301) Assembly
Sample-In Container
(68019)
Stone Diffuser
Stone Filter
NOTE: Ensure the fuel temperature is between 15 and 32°C. If necessary, place the
fuel container into a hot or cold water bath as required to bring the fuel
temperature within the limits.
3.5 DISASSEMBLY
After a test has been completed, perform the following steps to disassemble the test
components.
1. Ensure the test section has time to cool.
WARNING: A risk of burns exists unless the test section is cooled prior to disassembly.
2. Remove the sample-in thermocouple, the aeration tube and the sample-in tube
and place them in a clean location.
3. Remove the sample-in and sample-out containers (Figure 3-24) and properly
dispose of fuel.
FIGURE 3-24
SAMPLE-IN AND SAMPLE-OUT CONTAINERS
Aeration and
Sample-In Tube
Fittings
Sample-In
Thermocouple
Sample-In Sample-Out
Container Container
FIGURE 3-25
FUEL-IN TUBE REMOVAL
Test Section
Fuel Input Fitting
Fuel
Input Tube
Assembly
5. Remove the pre-filter assembly from the front panel fuel-in fitting (Figure 3-26).
Disassemble pre-filter and discard the filter element.
FIGURE 3-26
PRE-FILTER ASSEMBLY REMOVAL
Fuel-In Fitting
6. Disconnect and remove the test section outlet line as shown in Figure 3-27A.
7. Disconnect the test section (bypass) line fitting (see Figure 3-27B).
FIGURE 3-27
TEST-SECTION LINE REMOVAL
A B
Outlet Line
Bulkhead
Heater Section
(Bypass) Line
Test Section Fitting
Outlet Line
Fitting
8. Lift and rotate the thermocouple retaining clip (Figure 3-28) off the
thermocouple and carefully pull the thermocouple straight up out of the heater
tube taking care not to bend the thermocouple.
FIGURE 3-28
THERMOCOUPLE REMOVAL
Heater-Tube
Thermocouple
Connector
Heater-Tube
Heater-Tube
Thermocouple
Thermocouple
Hold-Down Clamp
9. Use the hex wrench provided to remove one of the cap screws from the upper
and lower bus bars.
10. Loosen the other screw on the bus bars three or four turns each, but do not
remove the screw. The end caps are not removed to insure they are not placed
on the wrong bus bars during reassembly.
11. Rotate the bus caps and remove the heater tube test section as shown in
Figure 3-29.
FIGURE 3-29
HEATER TUBE TEST SECTION REMOVAL
Upper Bus-Bar
End Cap
Lower Bus-Bar
End Cap
12. Use the extraction tool provided to remove the used o-ring and test filter
from the discharge chamber as shown in Figure 3-30. Discard the used o-ring
and filter.
FIGURE 3-30
DP-FILTER AND O-RING REMOVAL
Used DP Filter
Extraction
Tool
Used O-Ring
CAUTION: Handle the heater tube by the ends only. Do not touch the center section of
the heater tube.
CAUTION: Do not allow the heater tube center section to contact the test section during
removal.
CAUTION: Do not reuse heater tubes. Tests indicate that magnesium migrates to eh
heater tube surface under tests conditions. Surface magnesium may reduce
adhesion of deposits to the heater tube surface
14. Carefully, slide the heater tube out of the test section Remove and discard the
o-rings. Save the insulator bushings and nuts.
15. Hold the heater tube so that it points upward from the fingers as shown in
Figure 3-31. Flush the heater tube with solvent.
FIGURE 3-31
HEATER TUBE FLUSHING
Heater Tube
Solvent
(N-Heptane)
16. After the solvent has evaporated from the heater tube, place the heater tube in
the original plastic container and seal it with a cap. Mark it with the
appropriate identification. Rate the heater tube. Follow the procedure “Heater
Tube Deposit Rating” in Appendix 3-C if using a visual tuberater.
3.6 SHUTDOWN
CAUTION: In case of emergencies, switch off the power at the back of the JFTOT III
enclosure or unplug the power cord from the wall outlet.
To completely power down the JFTOT III for an extended period of time, or
whenever necessary, perform the following steps:
1. From the Main Menu, press or click the EXIT PROGRAM button to close the
JFTOT III program and display the MS Windows desktop screen.
2. Press or click the START button in the lower left corner and perform the
standard MS Windows shut-down steps to power off the computer.
3. After the computer is shut down, switch the JFTOT III power switch in the
back panel to OFF.
APPENDIX 3-A:
JFTOT III SCREEN DESCRIPTIONS
3-A.1 JFTOT III MENU STRUCTURE
The JFTOT III is designed to perform tests automatically after minor setup
procedures have been performed by the user. The following menu structure can be
used as an aid in navigating through the JFTOT III screen displays.
Main Screen
• DEFAULT Test
• Set Parameters (Maintenance Mode)
• Run Test
• Custom Test
• Setpoint Screen
• File Name for test data
• Auto Setup Progress screen
• Test Results screen
• Save Test option
• Test Annotation data entry screen
• Maintenance Menu Options Screen
• Aeration Pump ON/OFF
• Drive Pump ON/OFF and Flow Control set point
• System Vent-Valve Solenoid ON/OFF
• DP+ Vent-Valve Solenoid ON/OFF
• DP – Vent-Valve Solenoid ON/OFF
• DP Bypass-Valve Solenoid ON/OFF
• Bus Heater ON/OFF
• Setup DEFAULT Test
• Set-point values for temperature, time and flow rate
• Replace old file option
• Auto-Calibration Screen
• Calibrate System Pressure
• System Pressure Calibration Screen (follow prompts)
• Calibrate Differential Pressure
• Differential Pressure Calibration Screen (follow prompts)
• Calibrate Heater-Tube Thermocouple
• Heater-Tube Thermocouple Calibration Screen(follow prompts)
• Replace last calibration file option
• Reports
• Traditional Report
• Custom Report
CAUTION: Become familiar with the location and use of all controls, indicators,
connections, and accessories. Carefully read all instructions, warnings
cautions and labels prior to operating any portion of the system. Failure to
do so could result in personal injury and equipment damage.
The components of JFTOT III Main Screen shown in Figure 3-A-1 are listed and
described below.
FIGURE 3-A-1
JFTOT III MAIN SCREEN
3-A.2.1.3 Measure
The measure number indicates the speed at which all of the
parameters are being measured. The counter increments each time all
of the parameters have been measured (about 150 times per sec).
3-A.2.1.5 Test
Indicates the number of times the measured parameters are being
checked for errors during a test.
3-A.2.2.4 Maintenance
Press the Maintenance button on the main screen if you want
to operate the JFTOT III manually. You can turn on and off the
aeration pump, drive pump, and vent solenoids separately.
You can also open and close the DP bypass valve and turn on
the bus heater at a user-defined %Power (this is useful when
adjusting the SCR). Default test parameters are also set from
this screen.
Press the Maintenance button again on the Maintenance
screen to return to the main screen.
3-A.2.2.6 Reports
Press the Reports button if you want to view the Test Data for
any test previously run, or if you want to create a custom
report.
NOTE: The test will abort if you accidentally start a test without inserting the
heater-tube thermocouple.
3-A.2.3.8 Watts
The “Watts” indicator displays the power in Watts dissipated
by the heater tube. This value is used in error checking.
At a given flow rate the power required to heat an aluminum
heater tube to a given temperature is calculable. If the test
aborts because of the Watt value, a number of things could be
the cause.
• The heater tube is not clamped to the bus bars tightly.
• The heater tube has an out of tolerance ID or OD.
• The material the tube is made of is not within spec.
• The heater tube has been run previously at a high
temperature.
• The flow is not what it is supposed to be.
• The heater tube is shorting to the heater test section.
3-A.2.3.10 % power
The % Power indicator displays the % power that is being
applied by the SCR. This is set when the SCR calibration is
performed.
This value will be higher if the test setpoint is increased.
NOTE: If you run all of your tests at the same setpoint, the % power value should be
approximately the same from test to test. If this value increases over time it
may be an indication that there is some corrosion building up on the bus-bar
connections.
NOTE: You cannot manipulate the chart while the test is running because the chart
has been set to auto-scale during a test.
3-A.2.5 HELP
HELP information is available for all screen areas and controls by pressing the
CTL and H keys at the same time and then positioning the cursor over the screen
area or control in question. An information box appears (Figure 3-A-4) with an
explanation of the control function. Pressing the CTL and H keys again clears the
HELP function and returns the JFTOT III to normal operation.
FIGURE 3-A-4
JFTOT III MAIN SCREEN – HELP (EXAMPLE)
APPENDIX 3-B:
JFTOT III OPERATION DETAILS
3-B.1 OVERVIEW
The JFTOT III is designed to automatically run tests with minimal operator
involvement. After the setup procedures have been performed, either Default or
Custom test modes can be selected for running tests automatically. Maintenance
mode allows the user to operate individual components separately for tests and
adjustments. The Auto Cal mode is used to calibrate system pressure, differential
pressure, and heater-tube thermocouple. Test data from previous tests can be viewed
and new custom reports created in the Reports mode.
The test operation allows an operator to setup and run a thermal oxidation test. Two
test selections are provided, CUSTOM and D3241 DEFAULT. The CUSTOM test
allows the operator to adjust many of the test parameters. This provides a means to
design a custom non-ASTM thermal oxidation test. The user-adjustable parameters
are initially set to the values required by ASTM D3241, so the CUSTOM test can be
used to run a test as specified by ASTM D3241 until parameters are changed.
The D3241 Default test is basically the same as the CUSTOM test except that most of
the test parameters are fixed as specified by ASTM D3241. The following parameters
in Table 3-B-1 are fixed in the D3241 default test but variable in the custom test.
TABLE 3-B-1
TEST SETTINGS
You can store as many tests as free space on the JFTOT III hard drive will allow.
If the parameters for the default test need to be changed, refer to Section 3-B.1.3.
“Test Setup Parameters”.
Perform the following procedure to setup and/or run a default test.
1. Press or click the RUN DEFAULT button on the main screen. The Test
Annotation Dialog box appears as shown in Figure 3-B-2A.
FIGURE 3-B-1
JFTOT III MAIN SCREEN
NOTE: Refer to Sections 3.3 and 3.4 for detailed cleaning and setup instructions.
NOTE: System Pressure and RH adjustments should be verified before starting the
tests (see “JFTOT III Adjustments” in Section 5.4).
5. After completing the setup steps, press or click the CONTINUE button.
The system will start a preparation process automatically, which includes the
following operations.
• Aerating the sample
• Purging the system
• Measuring the relative humidity of the air supply
• Measuring the differential pressure across the DP filter
• Pressurizing the system
The system fuel pressure will increase. As the pressure builds up, no fuel will drip
from the sample out line.
A progress panel (Figure 3-B-3) will appear that displays the progress of the
auto-preparation process.
FIGURE 3-B-3
PROGRESS PANEL
NOTE: The progress panel will prompt the operator to check the drive pump flow
rate at 9.999mlpm and at the test flow rate. If either of these two flow rates
are incorrect, you will have two minutes to correct the problem or prime the
pump (see Section 5.4.2) If you wish to abort the test, press or click the
ABORT button. To skip this pretest procedure, press the SKIP button.
• Minutes Remain: Digital readout of the minutes remaining until the auto-
preparation procedure completes and the test begins
• Laps Time: Amount of time in minutes that has lapsed since the test began.
• Aeration % RH: The relative humidity reading of the air supplied to the
sample.
• Differential Pressure: Pressure across the DP filter displayed in mmHg
• System Pressure: Digital display of the system pressure given as mmHg
The overall progress of the auto preparation process is displayed as a red horizontal
progress bar.
The auto-preparation process is displayed by five green indicators that light after the
displayed operation has begun. The indicators should light in sequence as follows:
• Drive Pump: Indicates the sample drive pump is on when lit.
• Aeration Pump: Indicates the aeration pump is on when lit.
• System Vent: Indicates the system vent solenoid controlled valve is energized
(open) when lit.
• DP + Vent: Indicates the DP + vent solenoid-controlled valve is energized
(open) when lit.
• DP – Vent: Indicates the DP – vent solenoid-controlled valve is energized
(open) when lit.
If these operations complete without error, then the test will start automatically. If
there is an error, you will be prompted concerning the nature of the error and the
possible causes.
The JFTOT III program automatically begins the test and displays the test-in-
progress screen shown in Figure 3-B-4. The green Test Running indicator lamp
should be lit to confirm the test is running.
Sample inlet temperature (green), heater-tube temperature (yellow), and bus-bar
temperature (red) are plotted in the upper graphics panel and Differential pressure is
plotted in the lower graphics panel.
A green “Tare” indicator is on to indicate that the required time has lapsed and that
the Diff Pres indicator now displays a tared value.
The remaining time before test completion is displayed in the message panel at the
bottom of the screen.
If for any reason you want to stop the test before the test has finished, press or click
the STOP TEST control located in the test screen.
NOTE: A watchdog timer is used by JFTOT III to monitor program activity. If for
some reason the PC locks up or the program simply stops running, the
watchdog timer will remove power from the hardware and completely vent
the pressure from the system.
FIGURE 3-B-4
TEST IN PROGRESS SCREEN
A file was created and a number was assigned automatically to the file name when
the test began.
When the test has finished, a small option box will appear (Figure 3-B-5) and prompt
you to save the file containing the test data. Press or click YES to save the file or NO
if the data is to be deleted. IF YES is selected, the Traditional Report Form will
automatically display. See the Traditional Reports in Section 3-B.1.2 and
Figure 3-B-7 for information on filling out the Traditional Report. If NO is selected
the system returns to the Main Screen.
FIGURE 3-B-5
SAVE FILE OPTION BOX
3-B.1.2 REPORTS
Default test and Custom test results are automatically entered into a report form that
can be viewed and printed directly from the JFTOT III. Either a Traditional Report
or a Custom Report format can be used. When the REPORTS button on the main
screen is pressed or clicked, a panel is displayed with buttons for selecting either a
traditional or a custom report as shown in Figure 3-B-6. A third button allows the
user to cancel the report mode and return to the main screen.
FIGURE 3-B-6
REPORT SELECTION BOX
2. From the Maintenance screen, press or click the SETUP DEFAULT TEST button
as shown in Figure 3-B-11.
FIGURE 3-B-11
MAINTENANCE SCREEN – SETUP DEFAULT TEST BUTTON
3. After a small window appears as shown in Figure 3-B-12, double click on the
TEST TEMPERATURE button and type in the desired test temperature or click
the control at the left of the value.
FIGURE 3-B-12
MAINTENANCE SCREEN – PARAMETER SETTINGS
4. Double click on the TEST TIME IN MINUTES button and type in the desired test
time or click the control at the left of the value.
5. Double click on the FLOW RATE IN MLPM button and type in the desired
flow rate or click the control at the left of the value.
6. Press the OK button to save the settings and return to the maintenance menu.
7. After a prompt appears (Figure 3-B-13) asking if you want to replace the old
file, press or click the REPLACE button to replace the previous test file with the
new test file. Press or click the CANCEL button to discard the new settings and
keep the original settings for the Default tests.
FIGURE 3-B-13
MAINTENANCE SCREEN – FILE REPLACEMENT OPTION BOX
The Default test is now in place. This is the test that will run each time the
DEFAULT TEST button in the main screen is pressed. Follow this same procedure if
you want to change the default test again.
8. Press the EXIT button to return the main screen.
FIGURE 3-B-14
JFTOT III MAIN SCREEN – RUN CUSTOM
2. Fill in the blanks I the Test Annotation screen and then click OK.
3. A pop-up instruction box appears (Figure 3-B-15B) prompting you to perform
setup procedures. Follow the setup instructions on the prompt before
continuing.
NOTE: Refer to the Setup Procedures in Section 3-3 and 3-4 for detailed cleaning
and assembly procedures.
4. After completing the setup steps, press or click the CONTINUE button.
5. After a small window appears as shown in Figure 3-B-16, double click on the
TEST TEMPERATURE button and type in the desired test temperature.
FIGURE 3-B-16
MAINTENANCE SCREEN – PARAMETER SETTINGS
6. Double click on the TEST TIME IN MINUTES button and type in the desired
test time.
7. Double click on the FLOW RATE IN MLPM button and type in the desired
flow rate.
8. Press the OK button to save the settings.
9. After a prompt appears (Figure 3-B-17) prompting you for a new file name,
double click on the file name space and enter the new file name.
(e.g., Test 9-10-05)
FIGURE 3-B-17
FILE NAME INPUT BOX
NOTE: The progress panel will prompt the operator to check the drive pump flow
rate at 9.999mlpm and at the test flow rate. If either of these two flow rates
are incorrect, you will have two minutes to correct the problem or prime the
pump (see Section 5.4.2). If you wish to abort the test press or click the
ABORT button. To skip this pretest procedure, press the SKIP button.
The overall progress of the auto preparation process is displayed as a red horizontal
progress bar.
The auto-preparation process is displayed by five green indicators that light after the
displayed operation has begun. The indicators should light in sequence as follows:
• Drive Pump: Indicates the sample drive pump is on when lit.
• Aeration Pump: Indicates the aeration pump is on when lit.
If these operations complete without error, then the test will start automatically.
If there is an error, you will be prompted concerning the nature of the error and the
possible causes.
The JFTOT III program automatically begins the test and displays the test-in-
progress screen shown in Figure 3-B-19. The green Test Running indicator lamp
should be lit to confirm the test is running.
FIGURE 3-B-19
TEST IN PROGRESS SCREEN
Heater-Tube
Temperature (Red)
Tare (Green)
DP Pressure (Yellow)
FIGURE 3-B-20
SAVE FILE OPTION BOX
If you press or click the YES button, the Traditional Report screen (Figure 3-B-21)
will appear. Fill in any missing dat. Double click on each of the controls that you
want to enter data into, touch, click or use TAB key to select the control, and type in
the data. When you are finished, press or click on the OK button. The report will be
printed automatically.
FIGURE 3-B-21
TRADITIONAL REPORT (EXAMPLE)
The test is now complete and the test data has been saved. Refer to “Reports” in
Section 3-B.1.2 to print a report of the saved test results.
3-B.1.5 MAINTENANCE
The Maintenance mode is used to actuate and test individual components in the
JFTOT III separately without affecting other components. Select the Maintenance
option from the Main screen to display the screen shown in Figure 3-B-22.
FIGURE 3-B-22
JFTOT III MAINTENANCE SCREEN
The main control panel components in the Maintenance menu (Figure 3-B-22) are as
follows:
• Aeration Pump ON/OFF: Pressing the AERATION button lights the button
(yellow), changes the ON/OFF indicator to ON (green) and starts the aeration
pump. Pressing the AERATION button again returns the ON/OFF indicator to
OFF and stops the pump.
• Drive Pump ON/OFF: pressing the DRIVE PUMP button lights the button
(yellow), changes the ON/OFF indicator to ON (green) and starts the drive
pump. Pressing the DRIVE PUMP button again returns the ON/OFF indicator to
OFF and stops the pump.
• System Vent ON/OFF: pressing the SYSTEM VENT button lights the button
(yellow), changes the ON/OFF indicator to ON (green) and opens the system
vent valve. Pressing the SYSTEM VENT button again returns the ON/OFF
indicator to OFF and closes the valve.
• DP + Vent ON/OFF: pressing the DP + VENT button lights the button
(yellow), changes the ON/OFF indicator to ON (green) and opens the DP +
vent valve. Pressing the DP + VENT button again returns the ON/OFF
indicator to OFF and closes the valve.
• DP – Vent ON/OFF: pressing the DP – VENT button lights the button
(yellow), changes the ON/OFF indicator to ON (green) and opens the DP –
vent valve. Pressing the DP – VENT button again, returns the ON/OFF indicator
to OFF and closes the valve.
• DP Bypass Valve ON/OFF: pressing the DP BYPASS button lights the button
(yellow), changes the ON/OFF indicator to ON (green) and opens the DP
bypass valve. Pressing the DP BYPASS button again, returns the ON/OFF
indicator to OFF and closes the valve.
• Bus Heater ON/OFF: pressing the BUS HEATER button lights the button
(yellow), changes the ON/OFF indicator to ON (green) and applies power
through the bus bar to heat the heater tube. Pressing the BUS HEATER button
again returns the ON/OFF indicator to OFF and removes power to the bus bar.
• % RH: displays a digital value for the % relative humidity (RH) for the sample
air supply
• FLOW: displays a digital value for the sample flow rate in millimeters per
minute (mmpm) and allows the drive pump output flow rate to be changed
manually from 0.01 to 9.99 mlpm by pressing or clicking on the control button
on the left after pressing or clicking the DRV PMP OFF button.
NOTE: The drive pump will not start until the flow value is incremented or
decremented.
• Setup Default Test: Pressing the SETUP DEFAULT button will allow the
parameters to be set for a new default test.
• % pwr: The percentage of the total power available to be applied to the bus bar
can be controlled from 0-100% power by pressing or clicking the button to the
left of the % pwr display and observing the digital reading.
CAUTION: Entering too high % power increments will melt the heater tube. Enter small
changes starting at a low % value (e.g. 2 %, 4%, etc).
• Millivolts: The voltage being applied to the bus bar can be monitored by
observing the digital reading in millivolts
Perform the following steps to activate the components in the maintenance mode.
1. Press or click the MAINT button on the main screen as shown in
Figure 3-B-23.
FIGURE 3-B-23
JFTOT III MAIN SCREEN
2. From the maintenance screen, press the button for desired function as shown
in Figure 3-B-24 and listed in the following procedures.
FIGURE 3-B-24
JFTOTIII MAINTENANCE SCREEN – AERATION
3-B.1.5.1 Aeration
The aeration system can be checked independently by
performing the following steps.
CAUTION: To avoid possible pressure build-up, open the AERATION FLOW rotometer
control valve approximately 1/8 turn before turning on pump. After the test
is started, adjust the AERATION FLOW rotometer to the center of the
green range.
2. Adjust the air flow meter, located at the upper left of the
bus bar assembly, to the green area on the airflow control
inside of the bus-bar compartment (See Section 5.1.3 for
location).
3. After the air pump has been operating for at least one-
minute, check that the %RH reads below 20%. If the
%RH does not read below 20%, the dry-rite crystals need
to be changed or the 5VDC reference voltage needs to be
adjusted.
NOTE: The program looks at the %RH during the aeration portion of the JFTOT III
test and will not allow the test to continue if the %RH is above 20%.
NOTE: During a test the positive and negative sides of the differential pressure
transducer are purged automatically. The only time that you need to use
these controls is when you are performing a DP standardization test or if a
new differential pressure transducer has been installed.
3-B.1.5.5 DP Bypass
Pressing or clicking the DP-Bypass button (ON/OFF) activates a
solenoid-controlled valve that is used during test time to protect
the differential pressure transducer from pressures that may
damage the transducer. The solenoid-controlled valve is
connected between the positive and negative ports of the
transducer and when solenoid is activated, it equalizes the
pressure between the two ports.
You can use this control to check the operation of the bypass
solenoid and to check a new solenoid if it has to be replaced.
CAUTION: Check the % power setting before turning on the bus heater. Entering too
high % power increments will melt the heater tube. Enter small changes
starting at a low % value (e.g. 2 %, 4%, etc).
CAUTION: Keep the tube temperature below 380°C at all times. Failure to do so can
result in a meltdown of the heater tube
APPENDIX 3-C:
HEATER TUBE DEPOSIT RATING
3-C.1 VISUAL TUBERATER OPERATION
This procedure requires the Visual Tuberator light box equipped with a magnifying-
glass assembly and ASTM color chart standard.
NOTE: Make certain all (three) bulbs in the Tuberator are functioning to assure
consistent ratings.
1. Snap the upper end of the heater tube into the clamp of the adapter for the
heater tube as shown in Figure 3-C-1.
FIGURE 3-C-1
HEATER TUBE/TUBERATOR ADAPTER INSTALLATION
Heater Heater
Tube Tube
Adapter
Serial Number
2. Push the heater tube against the stop of the heater tube adapter.
3. Slide the adapter with the heater tube over the guide rod into the Tuberator as
shown in Figure 3-C-2.
FIGURE 3-C-2
TUBERATOR HEATER TUBE AND ADAPTER INSTALLATION
Heater
Tube and
Adapter
Visual
Tuberator
Light Box
4. Insert the ASTM Color Standard horizontally into the back of the Tuberator
in the slot provided as shown in Figure 3-C-3.
NOTE: The Tuberator color standard is known to fade upon exposure to strong
light. The life of the Standard is not known, but a separate standard plate
can be stored in the dark and compared to the Standard routinely used.
FIGURE 3-C-3
COLOR STANDARD INSTALLATION
ASTM
Tuberator
Color
Viewing
Standard
Port
NOTE: Only qualified personnel having normal eyesight and abilities to distinguish
between subtle color variances should be allowed to rate tubes.
5. Rotate the adapter and position the heater tube such that the side with the
maximum deposit is visible. If necessary, remove adapter and rotate heater
tube slightly in holder to better reveal the maximum deposit.
6. Compare the maximum heater tube deposit with the ASTM Color Standard as
shown in Figure 3-C-4. When the maximum deposit corresponds exactly to a
color standard, that number should be recorded. If the maximum heater tube
deposit being rated is in the obvious transition state between any two adjacent
color standards, the rating should be reported as less than the darker (the
higher number) standard.
FIGURE 3-C-4
HEATER TUBE TUBERATOR COLOR STANDARD COMPARISON
Heater Tube
ASTM Color
Standard
ASTM Color
Standard
Heater Tube
Adapter
7. In the event the heater tube has deposits which do not match the normal Color
Standard colors, then use the following rules for rating. With reference to
standard terms:
• If the deposit is “peacock” color, rate this as “Code P”, but also rate any
deposit that shows normal deposit color.
• If the deposit contains an abnormal color, rate this as “Code A”, but
also rate any deposit that shows normal deposit color.
NOTE: “Abnormal Color” is a tube deposit color such as blue or grey which is
neither peacock nor does it match the color standard.
8. When reporting the overall rating, report the maximum rating as described in
step 6, and if there are present colors which do not match the Color Standard,
report these as well. If there are only “P” and/or “A” deposits, report only
these and do not attempt to estimate a numerical grade.
• Example 1: a heater tube has a maximum deposit falling between Color
Standard Codes 2 and 3 with no other colors present. The overall
heater-tube rating would be “less than 3”.
• Example 2: the darkest deposit on a heater tube matches a Code 3, but
there is also a “peacock” deposit present. The overall rating of the
heater tube would be reported as “3P”.
• Example 3: a heater tube has a deposit that matches Color Standard
Code 1 and also has an abnormal deposit. The overall tube rating would
be reported as “1A”.
9. Return the heater tube to its original container.
LCD START
Screen Push
Button
Analysis
Chamber
Door
2. Open the door to the analysis chamber (Figure 3-C-6) and check for presence
of a heater tube or color chart. Remove heater tube if installed.
NOTE: Ensure the end of the chart marked DOWN is toward the bottom of the
enclosure.
FIGURE 3-C-6
VIDEO TDR ANALYSIS CHAMBER
Heater
Tube ASTM
Location Color Chart
Location
3. If a color chart is not in place, insert the chart into the chart holder as shown
in Figure 3-C-7. There must be no heater tube installed.
FIGURE 3-C-7
ASTM COLOR CHART INSTALLATION
ASTM
Color Chart
Color
Chart
Holder
“DOWN”
Label
Stepping
Motor
Assembly
4. Close the door and press the START button on the top of the VTDR. The
chart image is recorded and saved in memory by the computer.
NOTE: If the chart analysis fails, the LCD displays CHART CAPTURE FAILURE.
Failures are sometimes due to incorrect light levels or a heater tube may be
installed.
5. After the chart analysis has completed successfully, the LCD will display
CHART OK.
6. Press the START button to set the VDTR for the tube analysis.
7. The LCD will show READY to indicate the VTDR is ready to analyze a heater
tube.
CAUTION: Handle the heater tube by the ends only. Never touch the center section of
the heater tube.
8. Open the door to the analysis chamber and install the heater tube as follows:
8.1. Pull out the knob of the spring-loaded clamping mechanism at the top of
the tube holder as shown in Figure 3-C-8.
FIGURE 3-C-8
HEATER TUBE INSTALLATION
Serial
Number End
of Heater
Stepping
Motor Spring-Loaded
Assembly Clamping
Mechanism
8.2 Place the unmarked (no serial numbers) end of the heater tube into the
bottom end of the holder.
8.3 Lower the serial-numbered end of the tube down into the top end of the
holder and slowly release the clamping mechanism into the end of the
tube.
CAUTION: The end of the heater tube with the etched serial numbers should be at the
top of the tube holder with the serial numbers not facing directly upwards.
9. Close the door and then press the START button to begin the automatic tube
deposit rating test.
10. The LCD reports the status of the analysis by displaying the scan increments
as “SCANNING 1/36” through “SCANNING 36/36” (Figure 3-C-1A).
11. After the analysis is complete (6 seconds) the LCD indicates the results as
“RESULTS: 0” through “RESULTS: 4”. (Figure 3-C-1B).
FIGURE 3-C-1A FIGURE 3-C-1B
SCANNING RESULTS
11. Press the START button to clear the results for a new test.
NOTE: If the Video TDR is connected to a computer with VTDR software installed,
refer to the Video TDR User’s Manual for installation and operation of the
computerized version.
APPENDIX 3-D:
PRECAUTIONARY STATEMENTS
FOR HAZARDOUS MATERIALS
WARNING: The precautionary statements listed might not reference all aspects of the
material. For complete information, the user is urged to reference the
appropriate Material Safety Data Sheet (MSDS) for the particular material
being used.
3-D.1 ACETONE
• Keep away from heat, sparks, and open flame.
• Keep container closed. Use with adequate ventilation.
• Avoid buildup of vapors and eliminate sources of ignition, especially non-explosion
proof electrical apparatus and heaters.
3-D.2 TOLUENE
• Avoid prolonged or repeated breathing of vapor or spray mist.
• Use only with adequate ventilation.
• Eye irritation and dizziness are indications of overexposure.
• Do not take internally; swallowing may cause injury, illness, or death.
• Avoid prolonged or repeated contact with skin.
• Do not get in eyes.
• Produces toxic vapors on contact with flames, hot glowing surfaces, or electric arcs.
3-D.4 N-HEPTANE
• Keep away from heat, sparks, and open flame.
• Keep container closed.
• Use with adequate ventilation.
• Avoid prolonged or repeated breathing of vapor or spray mist.
• Avoid prolonged or repeated contact with skin
SECTION 4: CALIBRATION
PN: AL-59322
Revision B
MS Windows®, MS Windows® X P, and MS Windows Professional® are registered trademarks of Microsoft Corp.
REVISION HISTORY
Rev Revision Description Rev Date
A Original Release 02-28-06
B Changes to Fuses and Coolant Pump. Added VTDR Option. 05-18-06
Section 4: TABLE OF CONTENTS
SECTION 4: CALIBRATION ................................................................................................. 1
4.1 Overview....................................................................................................................... 1
4.2 Calibration Procedures ................................................................................................. 2
4.2.1 Temperature Calibration........................................................................................................2
4.2.2 System Pressure Transducer Calibration..............................................................................9
4.2.3 DP Transducer Calibration................................................................................................. 11
4.2.4 Save Calibration Data.......................................................................................................... 14
4.3 DP Transducer Standardization ................................................................................. 15
4.4 Flow Check ................................................................................................................. 24
4.4.1 Check Aeration Flow .......................................................................................................... 24
4.4.2 Check Coolant Flow ............................................................................................................ 24
4.4.3 Check Fuel Flow .................................................................................................................. 24
ALCOR JFTOT III Users Manual Section 4: Calibration
SECTION 4:
CALIBRATION
4.1 OVERVIEW
This section of the User’s Manual provides procedures to calibrate or check the
operation of major components used in the JFTOT 230 Mark III Jet Fuel Thermal
Oxidation Tester (JFTOT III). These calibration or validation procedures should be
performed on a regular schedule.
NOTE: Calibration schedules and logs should be provided by the user to record the
calibrations performed on the JFTOT III. All calibrations required for the
JFTOT III must be performed only by a qualified personnel.
NOTE: Adjustments are sometimes required to ensure that proper values are being
maintained. Refer to “JFTOT III Adjustments” in Section 5.4.
WARNING: These procedures involve handling flammable liquids and exposure to skin
irritants and harmful vapors. Wear protective clothing, gloves and mask
when setting up and performing the JFTOT III tests. Ensure flames and
spark producing materials are kept well away from the JFTOT III test area.
Use standard safety procedures at all times.
NOTE: The JFTOT III should be powered on for at least 20 minutes before
performing any calibration steps.
The AUTOCAL mode is used to perform standard calibration for all three major
components. Examples of displays are shown to illustrate the calibration procedure.
The dates, times, temperatures, and pressure readings are provided only as examples,
and will vary to reflect the actual condition of the equipment.
Perform the following calibration procedures in the order listed, then save the
calibration data.
WARNING: This procedure involves heating lead to over 327°C to a melted state. A risk
of severe burns exists unless extreme caution is used when handling the
heated lead standard.
Calibration of the temperature controller and heater tube thermocouple requires the
following low and high temperature standards:
• Low Temperature standard: Freezing point of pure water (ice bath at 0°C)
• High Temperature standard: Eutectic of pure lead (Freezing point at 327.5°C)
The heater-tube thermocouple is calibrated for both high and low temperature
standards. The JFTOT III measures and records the temperature detected for each
standard. Corrections are made based on the new readings and then temperatures are
calculated to display the calibrated temperatures.
1. Obtain the necessary equipment for both low (ice bath) and high (Lead Auto-
Cal Tube AL-35387) calibrations.
2. Prepare an ice bath by filling a Styrofoam cup with clean ice and then adding
pure water. This will be the low temperature reference source.
3. From the Main Screen (Figure 4-1), press or click the AUTO-CAL button to
display the JFTOT III Auto-Calibration Menu.
FIGURE 4-1
MAIN SCREEN – AUTO-CAL BUTTON
4. From the Auto-Calibration Menu (Figure 4-2), press or click the CALIBRATE
TUBE THERMOCOUPLE button to display the Calibrate Low-Temperature
screen shown in Figure 4-4.
FIGURE 4-2
AUTO-CALIBRATION MENU
5. Follow the instructions in the message panel at the top of the Calibrate Low-
Temperature screen (Figure 4-3) to calibrate the auto-cal low-temperature
reference point.
FIGURE 4-3
CALIBRATE LOW-TEMPERATURE SCREEN
6. Insert the thermocouple into the 0°C ice bath as shown in Figure 4-4.
FIGURE 4-4
ICE BATH LOW TEMPERATURE REFERENCE
Thermocouple
7. With the thermocouple in the ice bath, adjust the min deg C reading to obtain a
“Tube Temp” reading of 0.1° C.
8. Click OK to continue. The High-Temperature calibration message appears as
shown in Figure 4-5.
FIGURE 4-5
CALIBRATE HIGH-TEMPERATURE INSTALL LEAD SCREEN
9. Install the lead auto-cal assembly in the bus bars. Ensure the top of the autocal
assembly is flush with the top of the upper bus bar as shown in Figure 4-6A.
Tighten the end caps to insure good connection.
10. Insert the thermocouple as shown in Figure 4-6B. Ensure the thermocouple is
in contact with the lead material inside the well as shown in Figure 4-6C.
FIGURE 4-6
LEAD AUTOCAL ASSEMBLY
A B C
Thermocouple
Tip
11. Follow the instructions in the message panel and press or click OK to
continue. The lead standard heating process begins and the message in Figure
4-7 is displayed.
FIGURE 4-7
HEATING SCREEN
12. The JFTOT will heat the lead to 350°C. Make sure the thermocouple tip
maintains contact with the lead during the heating process.
13. Once the temperature reaches 350°C, you will be instructed to insert the
thermocouple all of the way into the auto-cal tube as shown in Figure 4-8.
Make sure the thermocouple is inserted all the way down and the tip is
centered in the lead.
FIGURE 4-8
INSERT THERMOCOUPLE SCREEN
14. When you have completed the instructions, press or click the OK button. The
heating will stop and the JFTOT III will search for the eutectic (refreezing
point). The lead standard cooling process is displayed as a decreasing
temperature curve with a flat plane where the eutectic occurs as shown in
Figure 4-9. The JFTOT III will determine if the eutectic was found and if it was
within the tolerances allowed.
FIGURE 4-9
CALIBRATE TUBE THERMOCOUPLE SCREEN - EUTECTIC SEARCH
Eutectic (327.5°C)
15. When the lead temperature cools to 300°C, the JFTOT III will reheat the lead
to 350°C and prompt you to remove the thermocouple from the lead as shown
in Figure 4-10.
16. Raise the thermocouple tip to make it accessible and remove all of the residual
traces of silicone and lead by wiping the tip with paper tissue. Remove the
thermocouple completely from the auto-cal tube. Inspect the tip closely for
cleanliness. The JFTOT III will turn the heater off automatically when the
thermocouple is removed from the melted lead.
FIGURE 4-10
CALIBRATE TUBE THERMOCOUPLE SCREEN – REMOVE THERMOCOUPLE
17. A dialog box will appear (Figure 4-11) that will inform you whether the
calibration was successful or not and will give you the option to repeat the
heater-tube thermocouple calibration or EXIT. If successful, continue with
step 19, otherwise continue with step 18.
FIGURE 4-11
HEATER-TUBE THERMOCOUPLE CALIBRATION COMPLETE DIALOG BOX
18. If a lead freezing point temperature (eutectic) was NOT detected then press
REPEAT to attempt detection of the freezing point again. If you press EXIT,
do not save the calibration data at the end of the calibration procedure. Failure
to detect a good calibration temperature could be caused by several factors
listed below:
• Operator is not following the procedures correctly. Carefully read the
instructions and repeat the procedure (i.e. make sure thermocouple tip
is centered in lead).
• Lead contaminated. To replace, see Section 5.2.1 Maintenance and
Service. Repeat procedure.
• Thermocouple is defective. Replace and repeat AUTOCAL procedure.
NOTE: The JFTOT III should be powered on for at least 20 minutes before
performing any calibration steps.
Perform the following steps to calibrate the JFTOT III system pressure transducer.
1. From the Auto-Calibration Menu (Figure 4-12), press or click the CALIBRATE
SYSTEM PRESSURE button. The Calibrate System Pressure screen
(Figure 4-13) is displayed.
FIGURE 4-12
AUTO-CALIBRATION MENU – CAL SYSTEM PRESSURE SELECT
2. If the “Sys Pressure” readout is not displaying zero (+0/–0), press or click the
control at the left of the “min psi” readout (Figure 4-13) until the “Sys
Pressure” reading displays zero (+0/–0).
FIGURE 4-13
CALIBRATE SYSTEM PRESSURE SCREEN
3. Obtain the calibration certificate (Figure 4-14) for the transducer being
calibrated and note the calibration factor value on the certificate. Ensure the
serial number on the certificate is the same as the serial number on the
transducer.
FIGURE 4-14
CALIBRATION CERTIFICATE (EXAMPLE)
Transducer
Serial Number
Calibration Factor
4. Multiply the Cal Factor on the certificate by 10 Volts. For example, 2.002910
mV/V x 10 V = 20.03 Cal mV.
5. Press the control at the left of the “Sys Pressure Cal millivolts” (Figure 4-15)
reading to change the displayed number to the cal-millivolts value determined
in Step 4.
FIGURE 4-15
CALIBRATE SYSTEM – CAL MV SCREEN
NOTE: The JFTOT III should be powered on for at least 20 minutes before
performing any calibration steps.
NOTE: For best results ensure the lines normally connected between the test section
and bulkhead fittings are removed.
1. From the Auto-Cal screen (Figure 4-16), press or click the CALIBRATE
DIFFERENTIAL PRESSURE button. The Calibrate Differential Pressure screen
(Figure 4-17) is displayed.
FIGURE 4-16
AUTO-CAL SCREEN
2. If the “Diff Pressure” readout is not displaying zero (+0/–0), press or click the
control at the left of the “min mmhg” readout (Figure 4-17) until the Diff Press
reading displays zero (+0/–0).
FIGURE 4-17
CALIBRATE DIFFERENTIAL PRESSURE SCREEN
3. Obtain the calibration certificate (Figure 4-18) for the transducer being
calibrated and note the number for the calibration factor on the certificate.
Ensure the serial number on the certificate is the same as the serial number on
the transducer.
FIGURE 4-18
CALIBRATION CERTIFICATE (EXAMPLE)
Transducer
Serial Number Calibration Factor
4. Multiply the Cal Factor on the certificate by 10 Volts. For example, 2.002910
mV/V x 10 V = 20.03 Cal mV.
5. Press the control at the left of the “Diff Pressure Cal millivolts” reading
(Figure 4-19) to change the displayed number to the cal-millivolts value
determined in Step 4.
FIGURE 4-19
CALIBRATE DIFFERENTIAL PRESSURE SCREEN
3. The calibration procedure is now complete and the calibration data has been
saved in the computer.
WARNING: These procedures involve handling flammable liquids and exposure to skin
irritants and harmful vapors. Wear protective clothing, gloves and mask
when setting up and performing the JFTOT III tests. Ensure flames and
spark producing materials are kept well away from the JFTOT III test area.
Use standard safety procedures at all times.
The following procedure can be used to ensure that the Differential Pressure
Transducer (DPT) system is within the operational limits required for the test.
Differential Pressure Transducer systems failing these criteria should be serviced by
a qualified technician.
1. The standardization technique uses a fuel of known density. Determine the
density by weighing to four decimals a 10.0 ml pipetted volume of the test fuel.
Typically, one milliliter of fuel weighs approximately 0.8 gms. When 10 ml of
sample fuel is used for the test, the sample weight will be approximately 8.0
gms. The fuel density can be used to determine a calculated value for fuel at a
given height from a zero point for comparison to the reading displayed on the
JFTOT III screen.
NOTE: If the exact weight of the fuel being used must be determined, the container
must be weighed before the fuel is added and then subtracted from the total
weight of the fuel and the container.
2. The theoretical DP for the transducer can be determined from the following
For Example:
Fuel Density =0.800 gms/ml
Height from 0 = 400 mm
Hg Density = 13.53 gms/ml
JFTOT Bias = 6%
NOTE: The JFTOT III should be powered on for at least 20 minutes before
performing the DP standardization steps
4. Place the fuel output container under the fuel output tubes as shown in
Figure 4-22.
FIGURE 4-22
FUEL OUTPUT CONTAINER
5. If installed, remove the heater tube test section from the bus bars as described
in Section 3.5 “Disassembly”.
6. From the Main Menu (Figure 4-23), press or click the MAINT button to display
the Maintenance screen as shown in Figure 4-24.
FIGURE 4-23
MAIN SCREEN MAINTENANCE BUTTON
7. The temperature and pressure values are shown in the temperature and
pressure gauges (Figure 4-24) on the right of the screen.
FIGURE 4-24
MAINTENANCE SCREEN DP BYPASS BUTTON
8. Press DP BYPASS button to turn off (CLOSE) the bypass valve (ON indicator
changes to OFF). Each time DP BYPASS button is pressed, the bypass valve
will change position (open and close) and the green indicator will light (ON)
and darken (OFF) respectively.
9. Allow the line from the leveling bottle to fill with the fuel, and then attach the
line fitting to the DP – Port while minimizing air left in line.
10. Press the DP – VENT button on the Maintenance screen (Figure 4-24) to turn
on (open) the DP – purge valve.
CAUTION: The test must be run with no air in the lines (fuel only) in order to have
meaningful results. Ensure all air is purged from the system before taking
readings.
11. Raise the bottle over 400 mm above the port as shown in Figure 4-25. The
atmospheric pressure on the fuel in the bottle will push fuel, and any air,
through the system and out the drip tube.
FIGURE 4-25
DP – TRANSDUCER PURGE
Calibration Fuel
Sample
Fuel Output
Line Drip Tube
DP – Port
12. Hold the bottle up for approximately one minute. Transfer at least 50
milliliters of fuel. Any air in the system will flow out slowly and can be seen as
irregular drips from the fuel-out drip tube.
13. When air bubbles no longer emerge from the fuel output drip tube, press the
DP– VENT button on the Maintenance screen to turn off (close) the DP–
purge valve.
NOTE: To keep air from the line, do not place the bottle below the DP– fitting.
14. Lower bottle until the level of the fuel is at the same level as the DP–
Transducer Port as shown in Figure 4-26.
FIGURE 4-26
TEST FUEL BOTTLE LEVEL AT – DP PORT
DP + Port
15. Remove the fitting from the DP – port and reattach to the DP+ Transducer
port. Do this quickly because fuel will drip from the open line fitting.
16. Press the DP +VENT button on the Maintenance screen (Figure 4-24) to turn
on (open) the DP+ purge valve.
17. Raise the bottle and again allow fuel to flow from the output tube bringing any
air that is in the system with it. Hold the bottle up for about a minute.
FIGURE 4-27
+ DP TRANSDUCER PURGE
Fuel Sample
DP + Transducer Port
18. Transfer at least 50 ml of fuel. Any air in the system will flow out slowly and
can be seen as irregular drips from the output tube. Make certain the level of
the test fuel in the bottle stays above the fitting on the bottom of the bottle.
19. When air bubbles no longer emerge from the fuel output drip tube, press the
DP + VENT button to turn off (close) the DP+ purge valve.
20. Lower the bottle so that the level of the fuel is level with the middle of the DP-
Port as shown in Figure 4-28.
FIGURE 4-28
TEST FUEL BOTTLE LEVEL AT - DP PORT
DP Transducer
Port (–) Sample Fuel Level
(0 mmHg)
DP +
Transducer
Port
21. Make a note of the reading displayed on the Maintenance screen (Figure 4-29)
for the Diff Press including its sign, if any. The reading should be
approximately 0.0 mmHg.
FIGURE 4-29
ZERO DP READING
22. Measure a point on the JFTOT III front panel that is 400mm (15.75”) above
the center of the – DP transducer port and mark the point with a pencil.
23. With the fuel line from the test fuel bottle connected to the +DP transducer
port, raise the bottle so the fuel level is even with the 400 mm point as shown
in Figure 4-30. This will cause the DP reading to increase since pressure has
been added to the positive side of the DP Transducer.
FIGURE 4-30
DP+ PRESSURE APPLICATION
Fuel Sample
Level
400 mm (15.75”)
(25 mm hg)
DP - Transducer
Port
24. With the fuel level held at the 400mm point, make a note of the DP reading on
the screen including its sign, if any. The reading will be approximately 25
mmHg as shown in Figure 4-31.
FIGURE 4-31
+ DP READING
Test Fuel
400mm Sample Level
(400mm)
Diff Pres
Reading
(25.0mmhg)
To -DP
Transducer
Input Port
25. Lower the test fuel bottle to the + DP port (Figure 4-32) and remove the test
bottle fuel line from the +DP transducer port. Place the end of the line in the
bottle cap to prevent draining the fuel out of the bottle.
FIGURE 4-32
TEST FUEL BOTTLE REMOVAL
DP Transducer
Input Port (–)
1. During a D3241 or CUSTOM test, place a clean beaker under the drip tube
when the heater turns on. Remove the beaker when the heater turns off.
Record the elapsed time the heater is on.
2. Pour the spent fuel into a clean graduated cylinder and measure the volume to
the nearest millimeter.
3. Calculate the flow rate by dividing the fuel volume by the elapsed time reading
(Flow Rate =Fuel Volume/Test Time) to arrive at the calculated flow rate in
ml/min. Flow rate should be very close to the specified rate for the test (3.0
mL/min for D3241).
Flow Rate errors in excess of 10% invalidate the test; contact PAC Technical Service
with details. The JFTOT III should maintain the flow rate within 0.1 mL/min of the
specified flow rate.
Systems with flow rates failing this test should be serviced by a qualified technician.
PN: AL-59322
Revision B
MS Windows®, MS Windows® X P, and MS Windows Professional® are registered trademarks of Microsoft Corp.
REVISION HISTORY
Rev Revision Description Rev Date
A Original Release 02-28-06
B Changes to Fuses and Coolant Pump. Added VTDR Option. 05-18-06
Section 5: TABLE OF CONTENTS
SECTION 5: MAINTENANCE AND SERVICE...................................................................... 1
5.1 Component Locations.................................................................................................... 1
5.1.1 Internal JFTOT Components ................................................................................................1
5.1.2 Control Components ..............................................................................................................2
5.1.3 Aeration System .....................................................................................................................3
5.1.4 Sample Delivery System.........................................................................................................4
5.1.5 Coolant System .......................................................................................................................7
5.1.6 Heater Tube Heat Control .....................................................................................................8
5.1.7 Temperature Controls ............................................................................................................9
5.1.8 Power Distribution Components........................................................................................ 10
5.1.9 Safety Components.............................................................................................................. 11
5.1.10 Back Panel Components...................................................................................................... 11
5.2 Maintenance................................................................................................................ 13
5.2.1 Lead Standard Replacement ............................................................................................... 14
5.2.2 Aeration Dryer Replacement.............................................................................................. 16
5.2.3 Cleaning Bus-Bar Contacts.................................................................................................. 17
5.2.4 Cleaning the Display, Keypad and Printer......................................................................... 18
5.2.5 Replacing the Sample-IN Filter and Aeration Diffuser .................................................... 19
5.2.6 Replacing the Charcoal Filter ............................................................................................. 19
5.3 Service......................................................................................................................... 20
5.3.1 Fuel Sample System Components Replacement................................................................ 20
5.3.1.1 Bus-Bar Thermocouple Replacement................................................................................. 22
5.3.1.2 Bus-Bar Assembly and Jet Tube Replacement .................................................................. 22
5.3.1.3 Sample Drive Pump Replacement ...................................................................................... 24
5.3.1.4 Valve Assembly Replacement............................................................................................. 27
5.3.1.5 Solenoid-Controlled Valve Assembly Replacement .......................................................... 28
5.3.1.6 Differential Pressure (DP) Transducer ............................................................................. 30
5.3.1.7 System Pressure Transducer Replacement........................................................................ 31
5.3.1.8 Back Pressure Regulator Replacement .............................................................................. 32
5.3.2 Fuel Aeration System Components.................................................................................... 33
5.3.2.1 Air Pump.............................................................................................................................. 33
5.3.2.2 Dryer Replacement.............................................................................................................. 34
5.3.2.3 Relative Humidity (RH) Sensor Replacement .................................................................. 34
5.3.2.4 Aeration Flow Meter Replacement .................................................................................... 35
5.3.3 Cooling System Components .............................................................................................. 35
5.3.3.1 Coolant Pump Assembly and Check Valve Replacement ................................................. 36
5.3.3.2 Flow Meter Replacement .................................................................................................... 38
5.3.3.3 Radiator and Fan Replacement .......................................................................................... 39
5.3.3.4 Coolant Container Replacement......................................................................................... 40
5.3.3.5 Coolant Jets and Bus Bars Replacement ............................................................................ 41
5.3.3.6 Coolant System Jet Tube Replacement .............................................................................. 42
5.3.4 Tubing and Fittings ............................................................................................................. 43
5.3.4.1 Metal, Nylon or Teflon Tubing and Swagelok Fittings .................................................... 43
5.3.4.2 Flexible Tygon Tubing and Swagelok or Barbed Fittings................................................. 43
5.3.5 Power Supply and Heater Supply System.......................................................................... 44
5.3.5.1 Power Entry Module Replacement .................................................................................... 44
5.3.5.2 Step-Down Transformer Replacement .............................................................................. 45
5.3.5.3 Power Distribution PCB and SSR Module Replacement.................................................. 46
5.3.5.4 Power Distribution PCB (AL-54743) LED Description ................................................... 48
5.3.5.5 Power Distribution PCB Cable Connections ..................................................................... 48
5.3.5.6 24-VDC Power Supply Replacement.................................................................................. 50
5.3.5.7 Heat Control Assembly Replacement ................................................................................ 51
5.3.5.8 SCR Heater Control Module Replacement ........................................................................ 52
5.3.5.9 Low-Voltage Transformer Replacement............................................................................ 52
5.3.5.10 RMS Current Transformer Replacement .......................................................................... 53
5.3.6 Control System Components .............................................................................................. 54
5.3.6.1 JFTOT III Embedded Computer Replacement .................................................................. 54
5.3.6.2 Touchscreen Alignment...................................................................................................... 56
5.3.6.3 Data Acquisition (DAQ) PCB Installation ........................................................................ 60
5.3.6.4 6024E DAQ PCI Board Configuration............................................................................... 65
5.3.6.5 Computer (PC) Interface PCB (AL-54741)........................................................................ 67
5.3.7 Pressure Transducer Voltage Checks................................................................................. 76
5.4 JFTOT III Adjustments............................................................................................... 79
5.4.1 Adjusting System Pressure (Back Pressure Regulator) .................................................... 79
5.4.2 Priming the Drive Pump ..................................................................................................... 80
5.4.3 Adjusting the SCR ............................................................................................................... 83
5.4.4 Adjusting the 5-VDC RH Reference Voltage..................................................................... 86
5.4.5 Power Entry Module Fuse Configuration.......................................................................... 87
5.4.5.1 Power Entry Fuse Reconfiguration (115VAC to 230VAC) ............................................. 87
5.4.5.2 Power Entry Fuse Reconfiguration (230VAC to 115VAC) ............................................. 89
5.5 Software Installation .................................................................................................. 91
5.5.1 MS Windows XP Professional OS Installation.................................................................. 91
5.5.2 NI-DAQ Software Installation............................................................................................ 91
5.5.3 JFTOT III Application Software Installation .................................................................... 96
5.6 Error Messages.......................................................................................................... 102
5.7 TroubleShooting Guide ............................................................................................. 103
5.7.1 Electrical/Electronic Component Replacement .............................................................. 103
5.7.2 Troubleshooting Guide ..................................................................................................... 104
5.8 Spares, Accessories, Supplies, and Materials ........................................................... 106
5.9 Drawings ................................................................................................................... 111
ALCOR JFTOT III Users Manual Section 5: Maintenance and Service
SECTION 5:
MAINTENANCE AND SERVICE
5.1 COMPONENT LOCATIONS
The locations and descriptions of JFTOT 230 Mark III Jet Fuel Thermal Oxidation
Tester (JFTOT III) internal and external components are included in the following
subsections.
PC Interface
PCB
Power
Distribution
PCB
RH Detector
Power Entry
Module Sample Drive
Pump
Vent Fan
Computer
Step Down
Transformer
(230VAC /115VAC)
Coolant Fluid
Container
Upper Bus Bar Coolant Fluid
Assembly Pump
FIGURE 5-2
CONTROL COMPONENTS
A B
FIGURE 5-3
AERATION SYSTEM COMPONENTS
A B C D
Humidity
Sensor
Aeration
Pump
Plastictight
Fitting
(Al-28432)
Aeration-Out
Diffuser
(AL-27096)
Sample Fuel
Coolant
OUT Container
Flow Meter
• Sample-In and Sample-Out Containers: two 1000 ml, scaled, glass beakers
• Sample-In Filter: porous stone filter element placed on the sample-in tube
inside the sample-in container to filter fuel particles larger than 10mm
• Sample-Drive Pump: HPLC dual-piston chemical pump capable of
delivering continuous flow at approximately 500 psig
• Pre-filter: Inline filter placed at the output of the sample pump to trap
particles
• Bus-Bar Assemblies: used for mounting the heater tube. The bus bars
conduct high-current from a low-voltage source for heating the heater tube
• Heater Tube: hollow aluminum tube with smaller diameter center for
collecting oxidation from fuel sample on the external surface under closely
controlled temperature and flow rates
• Differential Pressure (DP) Filter: filter for catching deposits not formed
on the heater tube. The system is set to bypass the DP filter at a pressure
differential greater than 250 mmHg.
The internal components (Figure 5-5) used in the sample delivery system are listed
and described as follows:
FIGURE 5-5
INTERNAL SAMPLE DELIVERY COMPONENTS
Differential Pressure
Bypass Solenoid- Transducer
Controlled Valve
System Vent Solenoid-
Controlled Valve
System Pressure
Transducer Safety Vent Solenoid-
Controlled Valve
Back Pressure
Regulator Valve
• Bypass Solenoid Valve: solenoid controlled valve used to route the sample
around the DP filter when differential pressure exceed 250mmHg
• Differential Pressure (DP) Transducer: Computer input device used to
monitor system pressures on each side (+/ – ) of the DP filter
• +DP Purge Solenoid Valve: solenoid controlled valve connected to the
positive pressure (+) side of the DP transducer used to purge sample out of
the system to the sample out container
• – DP Purge Solenoid Valve: solenoid controlled valve connected to the
negative pressure (–) side of the DP transducer used to purge sample out of
the system to the sample-out container
• System Pressure Transducer: computer input device used to monitor
system pressure
• Back Pressure Regulator: spring-loaded pressure regulator manually set to
predetermined pressure (500 psi) by varying the orifice size at the output of
the sample flow system
• System Vent Solenoid Valve: solenoid-controlled valve used to vent the
system pressure into the sample out container when purging air from fuel
system or pressure exceeds maximum set limit
• Safety Vent Solenoid Valve: solenoid controlled valve used as a backup
valve for the system vent valve that will vent the system pressure into the
sample out container should the PC program crash or in the event of a power
failure
• Tubing and Fittings: high-pressure, stainless steel tubing and fittings used
to carry the sample to and from the heater tube and into the sample out
container
Coolant Container
Coolant Flow Meter
Upper Coolant
Jet Tube Check Valve
Current
Detection
Transformer Low-Voltage
(High Current)
Bus-Bar Transformer
Assemblies
SCR Heat
Controller
Sample-In
Thermocouple Bus-Bar
(AL-86301) Thermocouple
(AL-86302)
Heater Tube
Thermocouple
(AL-86303)
C D
Power
Distribution
PCB
24VDC Power
Supply
E
Blown Fuse (when lit)
Vent Fan Fuse ½ A (289036) Indicator LED (typical)
Vent Fan
Fuses (x5)
Dryer
Coolant
Radiator Fan Power Entry
Module
Connectors
5.2 MAINTENANCE
The Jet Fuel Thermal Oxidation Tester III (JFTOT III) has been designed to perform
standard and user-defined tests automatically after minor preparations have been
completed. The JFTOT III requires very little maintenance if installed correctly and
used properly as described in “Proper Use and Safety” at the front of this User’s
Manual. Routine maintenance requirements are listed in the following table.
NOTE: Maintenance schedules and logs should be provided by the user to record the
maintenance performed on the JFTOT III. All maintenance and service
required for the JFTOT III must be performed only by a qualified technician.
TABLE 5-1
MAINTENANCE INTERVALS
WARNING: This procedure involves heating lead to over 327°C to a melted state. A risk
of severe burns exists unless extreme caution is used when handling the
lead standard. Wear appropriate protection when performing this procedure.
1. Install the Lead AutoCal Calibrator inverted (open end down) between the
upper bus bar and the lower bus bar as shown in Figure 5-11. Use the steps in
Subsection 3.4.3 “Test Section Installation Procedures” as a guide in installing
the lead calibrator.
2. Place a paper tissue or cloth under the well to catch the old lead and a covering
over the end of the bus bar to shield the thermocouple orifice from the molten
lead.
FIGURE 5-11
LEAD STANDARD INSTALLATION FOR LEAD REMOVAL
Inverted Lead
Calibrator Well
Tissue or Cloth
4. Press the % PWR control (Figure 5-12) on the Maintenance screen to display
55%.
5. Press the BUS HEATER button to turn on the bus heater to begin melting the
old lead.
6. After the old lead has dropped out of the calibrator, lightly tap the lead well
with a metal object to cause residual old lead to fall out.
7. Press the BUS HEATER button to turn off the bus heater.
FIGURE 5-12
MAINTENANCE SCREEN – SETUP DEFAULT TEST BUTTON
8. After the lead well has cooled, using the steps in Subsection 3.4.3 “Test
Section Installation Procedures” in reverse order as a guide, remove the lead
calibrator and reinstall it with the open-end up between the upper and lower
bus bars.
9. Press the BUS HEATER button on the Maintenance screen to begin heating the
lead calibrator.
FIGURE 5-13
THERMOCOUPLE REMOVAL
10. After the well has heated to the lead melting point (350 deg C), refill the well
with new lead pellets (5.5 Gms or approximately 7 pellets).
11. Lightly tap the well with a metal object to remove air pockets that may form
inside the lead.
12 Press the BUS HEATER button on the Maintenance screen to turn the heater
off and stop heating the lead.
13. After the lead has cooled, remove the lead calibrator from the bus bars.
14. Press the MAINT button on the maintenance screen to return to the main
screen.
Bottom Dryer
Fitting
4. Loosen and remove the knurled end cap from the top of the dryer housing.
CAUTION: Do not discard the cotton filter located in the end caps. This cotton filter
ensures that the silica gel remains inside the tube.
FIGURE 5-15
DRYER DISASSEMBLY
Gasket
Cotton Filter
5. While holding the plastic cylinder in place, pour out and discard the old
Dri-Rite.
CAUTION: Do not use bead-type Dri-Rite when refilling the dryer. Use only granular
Dri-Rite.
6. Ensure the cotton filter is in place at the bottom end cap and then pour new
“blue” granular-type Dri-Rite into the plastic cylinder.
7. Reinstall the cotton filter and then hand-tighten the end caps onto the housing.
Ensure the gasket is in place at both ends of the tube.
8. Reinstall the dryer into the clip mounts and connect the bottom tubing and the
top fitting. Do NOT over tighten fittings during replacement. For new tubing
installation procedures, refer to Section 5.3.4. “Tubing and Fittings”.
WARNING: A risk of severe burns exists unless power is removed and bus bars are
cooled prior to cleaning the bus bar contacts.
Before every test, the gold contact area of each bus bar and end cap should be
cleaned by performing the following procedure.
WARNING: These procedures involve handling flammable liquids and exposure to skin
irritants and harmful vapors. Wear protective clothing, gloves and mask
when setting up and performing the JFTOT III tests. Ensure flames and
spark producing materials are kept well away from the JFTOT III test area.
Use standard safety procedures at all times.
1. If the bus bars were recently heated, allow them to cool before continuing.
2. If heater tube assembly is installed, remove the bus-bar end caps and heater
tube assembly.
3. Use a tissue soaked with solvent to clean the gold contact area of each bus bar
and end cap as shown in Figure 5-16A.
FIGURE 5-16
CLEANING BUS BARS AND END CAPS
A B
Bus-Bar ID
Number
End Cap ID
Number
CAUTION: Before reinstalling the end caps onto the bus bars ensure the number on the
end cap matches the number on the bus bar.
4. If removed previously, reinstall the heater tube assembly and end caps.
Reinstall the end caps on the bus bar with matching ID numbers placed
together as shown in Figure 5-16B.
5. If over fifty tests have been run (or after every year) use an art gum type eraser
(very soft) and lightly rub the gold contact area to remove dirt and oxide.
CAUTION: DO NOT use hard erasers, emery paper or other abrasive material to clean
the contact area. Doing so will remove the gold plating.
CAUTION: Never use solvents (including alcohol) to clean the LCD panel. Even vapors
from alcohol will damage the LCD panel. Use a clean cloth and mild cleaning
solution to remove finger prints etc from the panel surface.
CAUTION: Never use a sharp tool such as a pen or pencil to point to items on the
touchscreen. Doing so will permanently damage the screen. Use a stylus,
fingertip or a soft plastic tool to select items on the screen.
1. Clean the display once a month, or as needed, with a clean, dry, soft cloth and
a mild cleaning solution to remove fingerprints and dust from the display.
2. Use low pressure air (canned keyboard cleaner) and clean dust cloth to clean
the keyboard and printer.
3. Use a general purpose water-based cleaner and dry cloth to clean the JFTOT
cabinet.
1/8”
5.3 SERVICE
This section of the JFTOT III User’s Manual contains problem analysis, repair and
part replacement information for the JFTOT III. This section is divided into
subsections that relate to each of the JFTOT III subsystems and components listed
below.
• Fuel Sample System
• Fuel Sample Aeration System
• Coolant System
• Heat and Power Supply System
• Control System
• Enclosures
A Troubleshooting Guide is included to help in diagnosing potential problems that
may occur while operating the JFTOT III. Listings of spare parts, accessories,
supplies and materials are also included along with a set of drawings and diagrams.
NOTE: All maintenance and service required for the JFTOT III must be performed
only by a qualified technician.
WARNING: These procedures involve handling flammable liquids and exposure to skin
irritants and harmful vapors. Wear protective clothing, gloves and mask
when setting up and performing the JFTOT III procedures. Ensure flames
and spark producing materials are kept well away from the JFTOT III test
area. Use standard safety procedures at all times.
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components. Ensure that power lines are removed from
the power source before continuing.
FIGURE 5-18
JFTOT III FRONT PANEL FUEL SAMPLE SYSTEM COMPONENTS
Sample Out and
Vent Lines
Sample In Line
and Filter
Sample Fuel In
Container
Upper Bus Bar
Sample Fuel
Out Container
DP Filter Assembly
Heater Tube Test Section
(Heater Tube Inside)
Sample Drive Lower Bus Bar
Pump Enclosure Pre-Filter Assembly and
Anti-Siphon Check Valve Input Tube
The JFTOT III Fuel Sample System also includes the following major internal
components listed below and shown in Figure 5-19.
• Bypass Solenoid-Controlled Valve Assembly
• Differential Pressure Transducer
• Purge Solenoid-Controlled Valves (+ and – )
• System Vent Solenoid-Controlled Valve
• Safety Vent Solenoid-Controlled Valve
• Back-Pressure Regulator
• Tubing and Fittings
FIGURE 5-19
INTERNAL FUEL SAMPLE SYSTEM COMPONENTS
Bus-Bar
Thermocouple
Connection
Bus-Bar Thermocouple
(AL86302)
Thermocouple
Hold-Down Clamp
Bus-Bar Mounting
Thermocouple
Screws (x2)
Clamp Mount Block
Upper Bus-Bar
Thermocouple Assembly (AL-10774)
Wire Insertion
Thermocouple Hold-
Down Collar Screw
Thermocouple Hold-Down Collar
Lower Power
Cable Mount Nut
Lower Bus-Bar
Assy (AL-10774) Lower Bus-Bar Coolant
Output Tube Fitting
Lower Voltage Wire (AL-28389)
Bus-Bar Mounting Nut
Lower Bus-Bar Coolant Coolant Input Tube
Jet-Tube (AL-10831) Fitting (AL-28358)
6. From inside the JFTOT III enclosure, slide the upper and
lower bus-bar assemblies (Figure 5-21) out of the bus-bar
mounting plate.
7. Remove the coolant jet tubes from the lower and upper bus-bar
assemblies (Figure 5-21).
8 Use the steps for removal in reverse order (7-1) as a guide for
replacing the bus bar assemblies. Ensure the flat surface of the
bus-bar is facing down. Install Teflon tape over pipe fitting
threads only. Do NOT over tighten fittings. For procedures on
installing new tubing, refer to Section 5.3.4. “Tubing and
Fittings”.
NOTE: When the tube is replaced, ensure the end of the coolant jet tube is ¼“away
from the bottom of the bus-bar coolant well. Refer to “Coolant System Jet
Tubes and Fittings” in subsections 5.3.3.6 for instructions on replacing the
coolant jet tube.
9. After installing the bus bars, the alignment bar (Figure 5-22)
must be used to align the bus bars in order to prevent any
bending of the heater tube when it is installed into the
bus bars.
FIGURE 5-22
BUS-BAR ALIGNMENT
Bus-Bar
Alignment
Bar (3/16´D)
FIGURE 5-23
JFTOT III DRIVE PUMP ASSEMBLY
Anti-Siphon Check
Valve (21084)
Pump Output
Line Fitting
Piston 1 Input Line
Output Line
Piston 2 Input Line Bulkhead Fitting
8. Remove the left-side panel and locate the drive pump assembly
and the valve assembly as shown in Figure 5-24.
FIGURE 5-24
DRIVE PUMP AND VALVE ASSEMBLIES
Valve Assembly
Power Cable
Connector Serial Cable
Connector
Signal Cable
Connector
10. Remove the right-side panel and locate the four mounting
screws holding the valve assembly mounting bracket to the
enclosure wall as shown in Figure 5-28 in Section 5.3.1.4.
11. Remove the lower two mounting screws and loosen the upper
two valve assembly mounting screws.
12. Have someone lift the valve assembly up as far as the valve
assembly mounting screw slots allow and then retighten the
upper two valve assembly mounting screws to hold the valve
assembly up off of the drive pump assembly.
13. From the front panel, lift the drive pump assembly up out of
the feet-locator holes and pull the pump assembly out the front
of the enclosure.
14. Use the removal steps in reverse order (13-1) as a guide for
replacing the pump assembly. Install Teflon tape over pipe
fitting threads only. Do NOT over tighten fittings. For new
tubing installation procedures, refer to Section 5.3.4. “Tubing
and Fittings”.
CAUTION: The new replacement drive pump assembly may contain a small filter on the
output connector. This filter must be removed before connecting the output
line to the new pump.
Seal
Filter
Vent Output
Line Fitting
Bypass Solenoid-
Controlled Valve
High DP In Line Fitting
(AL-93061)
Differential Pressure
Transducer (AL-45042) System Vent Solenoid-
Controlled Valve
(AL-93061)
System Pressure
Transducer
(AL-45043)
Safety Vent Solenoid-
Back Pressure Controlled Valve
Regulator Valve (AL-93062)
(AL-61637)
Sample Out
Line Fitting
Rubber Pad
3. Ensure all sample fuel is drained from all lines and then
disconnect the two input lines and the two output lines.
NOTE: Connectors P3 and P4 on the PC Interface PCB have a locking tab that must
be depressed while removing the connector.
Valve Assembly
Mounting Screws (x4)
FIGURE 5-29
SOLENOID-CONTROLLED VALVE (EXAMPLE)
Solenoid-
Controlled
Valve
Assembly
Fuel Line
Fuel Line Fitting
Fitting
Push-Wire Connector
Push-Wire
Installation
Tool
7. Place the valve assembly on the side.
8. Locate and remove the two Philips-head screws (Figure 5-31)
holding the valve to the mounting plate and remove the valve
assembly.
FIGURE 5-31
SOLENOID-CONTROLLED VALVE REMOVAL (EXAMPLE)
Solenoid- Controlled
Valve Assembly
Mounting Screws (x2)
DP Transducer
Assembly (AL-45042)
DP Transducer Hex Screw
nd
(2 hex screw in back)
Differential Pressure
Transducer Cable
nd
2 DP Transducer
Bracket Mount Hex
Screw
DP Transducer Assembly
Bracket Mount Screws (x2)
FIGURE 5-34
SYSTEM PRESSURE TRANSDUCER REMOVAL
System Pressure
Transducer (AL-045043)
Cable
Back-Pressure
Regulator (AL-61637)
3. Ensure all sample fuel is drained from all lines and then
remove the input and output lines (Figure 5-35) at the base of
the back-pressure regulator.
4. Remove and retain the fuel line fittings for use on replacement
valve.
5. Remove the two mount screws (Figure 5-27) from the base
clamp.
6. Lift the regulator and clamp from the JFTOT III.
7. Loosen one of the clamp screws to remove the base clamp from
the regulator.
8. To replace the regulator, perform the removal steps in reverse
order (Steps 6-1) as a guide. Install Teflon tape over pipe
fitting threads only. Do NOT over tighten fittings. For
procedures on installing new tubing, refer to Section 5.3.4.
“Tubing and Fittings”.
FIGURE 5-36
AERATION AIR PUMP
Aeration
Supply Pump
Power Cable
Aeration Supply Connector
Pump (AL-60113)
RH Sensor Cable
Connector Air Output
Tube Fitting
RH Sensor
(AL-92243)
Air Input
Tube Fitting Mounting
Screws (x2)
Mount Bracket
Mounting Nuts
Aeration
Flowmeter
Elbow Fitting Input Line
Fitting
CAUTION: Use a backup wrench to hold the housing nuts for the elbow fitting when
loosening or tightening the flowmeter fittings. Failure to do so may result in
the threads of the flowmeter plastic housing being stripped.
The coolant system uses two types of tubing for lines. The pump lines are 5/16" dia
flexible Tygon® tubing (AL-49292) and the remaining tubes are 1/4" dia. Teflon
tubing (AL-49291). See Section 5.3.4. “Tubing and Fittings”.
The JFTOT III Cooling System includes the following major components:
• Coolant Pump and Bypass Valve
• Flow Meter
• Radiator and Fan Assemblies
• Coolant Container
• Coolant Jets and Bus Bars
• Tubing and fittings
CAUTION: Use only distilled or deionized water in the cooling water container. Failure
to do so may cause deposits to build up in the coolant system.
FIGURE 5-39
COOLANT SYSTEM
Coolant Fluid
Container
(AL-17683)
Coolant Fluid Pump
(AL-60119)
Coolant Fluid
Pump Check Valve
(AL-93109)
2. Remove the right side panel and locate the coolant fluid pump
as shown in Figure 5-40. Normal flow is indicated by solid
arrows. A dashed arrow indicates the flow through the check
valve when the line is clogged.
FIGURE 5-40
COOLANT FLUID PUMP ASSEMBLY
Coolant Fluid
Container Coolant Fluid
Pump
Snap-Lock
Mounting
Clamp
Check Valve
Power Cable
Connector
CAUTION: When a replacement check valve is installed, ensure the check valve is
installed with the arrow on the side of the valve pointing toward the coolant
delivery line to the pump.
CAUTION: When a replacement coolant pump is installed, ensure the direction of flow
is down as indicated by the label on the pump.
NOTE: To dislodge the ball inside the valve, clamp off the input lines, remove the
lines, remove the inlet “T” fitting and use a pencil eraser to dislodge the ball
valve. Reassemble the valve, remove the clamps and retest.
Coolant Flowmeter
(AL-58036)
Plastic Shield Coolant Fluid Flowmeter
Output Line Fitting
Mount Bracket
Mounting Nuts
Coolant Fluid
Flowmeter Input Line
Fitting
CAUTION: Use a backup wrench to hold the housing nuts for the elbow fitting when
loosening or tightening the flowmeter fittings. Failure to do so may result in
the threads of the flowmeter plastic housing being stripped.
Radiator Coolant
Output Line Fitting
Fan (61009-115V)
Radiator (AL-60684)
FIGURE 5-44
COOLANT CONTAINER
Coolant Container
(AL-17683) Coolant Container
IN Fitting
Mounting
Nuts (x2)
Coolant Container
OUT Fitting
(AL-97139)
(AL-10774)
NOTE: Refer to “Drawings” in Section 5.9 for the flow diagram and tube routing
diagrams.
When replacing the coolant jet tube, perform the following steps.
1. Insert the tube into the bus bar and slide the ferrule and nut in
place as shown in Figure5-46 A.
2. Insert the tube into the bus bar until it bottoms out.
3. Finger tighten the nut and then scribe the tube at the back of
the fitting as shown in Figure 5-46B.
FIGURE 5-46
COOLANT JET TUBES INSTALLATION
A B
Ferrule
Nut
.
4. Retract the tube ¼” as shown in Figure 5-47 and then use an
open-end wrench for the nut and a backup wrench for the
T-fitting to swage the nut and ferrules onto the tube. Install
Teflon tape over pipe fitting threads only. Do NOT over
tighten fittings. For procedures on installing new tubing, refer
to Section 5.3.4. “Tubing and Fittings”.
FIGURE 5-47
JET TUBE RETRACTION
¼” Scribe Mark
Output Fitting
5. Test the connection for flow by blowing into the tube and
checking for air flowing out the output fitting
NOTE: Refer to the “Drawings” in Section 5.9 for the flow diagram and tube
routing diagrams.
When installing new lines or tubing, make sure you use new
Swagelok ferrules and swage them properly. Swagelok fittings are
installed as follows:
1. Install the two-part ferrule and nut onto the tube as shown in
Figure 5-48.
FIGURE 5-48
SWAGED FITTINGS ASSEMBLY (TYPICAL)
Nut
Fitting
Tubing
Ferrules
2. Insert tubing into the Swagelok fitting. Make sure that the
tubing rests firmly in the shoulder of the fitting and the nut is
finger-tight.
3. Before tightening the Swagelok nut, use a scribe (or pencil) to
mark the nut at the 12-o’clock position as a reference.
4. Hold the fitting body steady with a backup wrench and tighten
the nut as follows:
• For 1/8" tubing: 3/4 turns to the 9-o’clock position
• For 1/4" and larger tubing: 1-1/4 turns
(past 12-o’clock to the 3-o’clock position)
CAUTION: Install Teflon tape over non-swaged fitting threads only. Do NOT over
tighten fittings. For 3/16" diameter and smaller tubing, tighten the nut an
extra 3/4 turns. Tubing 1/4"and larger requires tightening the nut an extra
1-1/4 turns. When retightening already swaged fittings tighten to snug, DO
NOT turn the nut an extra 3/4 turn.
NOTE: Refer to the “Drawings” in Section 5.9 for the flow diagram and tube
routing diagrams.
When installing new lines or tubing, make sure you use new
Swagelok ferrules and swage them properly. The brass insert may
be reused. New Tygon tubing is installed onto Swagelok fittings as
follows:
1. Cut the tubing to length.
2. Install the two-part ferrule and nut onto the tube as shown in
Figure 5-48
3. Install brass insert into the end of tubing to be swaged.
4. Insert tubing into the Swagelok fitting. Make sure that the
tubing rests firmly in the shoulder of the fitting and the nut is
finger-tight.
5. Before tightening the Swagelok nut, use a scribe (or pencil) to
mark the nut at the 12-o’clock position as a reference.
6. Hold the fitting body steady with a backup wrench and tighten
the nut as follows:
• For 3/16 and 1/8" tubing: 3/4 turns to the 9-o’clock position
• For 1/4" and larger tubing: 1-1/4 turns
(past 12-o’clock to the 3-o’clock position)
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components.
Blue (D)
Brown (A)
Violet (C)
Orange (B)
Green (Gnd)
3. Note the positions of each wire and then carefully, pull each of
the spade connectors (Figure 5-49) from spade terminals on
the module. Care must be taken not to break the wires.
4. Use a flat-blade screwdriver to press down the plastic spring-
clips (Figure 5-49) at the top and bottom corners of the
connector while lifting the back of the connector to push the
module out of the opening.
5. Use the removal steps in reverse order (Steps 4-1) as a guide to
replace the module.
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components.
FIGURE 5-50
STEP-DOWN TRANSFORMER
Step-Down Transformer
(AL-90698)
3. Note the positions of each wire and then remove the six wires
connected (Figure5-50) to the transformer terminals.
4. While holding the transformer, remove the mounting nuts (4)
in the corners that hold the transformer to the enclosure wall
and remove the transformer.
5. Use the removal steps in reverse order (Steps 4-1) as a guide to
replace the transformer.
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components.
1. Switch off the power at the back of the JFTOT and unplug the
power connector.
2. Remove the left side panel and locate the power distribution
PCB as shown in Figure 5-51.
FIGURE 5-51
POWER DISTRIBUTION PCB
P1 – Relay Control
K1 Mounting Screws
(x4)
Connector Locks
K2 Mounting Screws
P4 – To PC Power (x4)
(115 VAC)
P3 – To Step-Down
Transformer P2 – 115/230 VAC from
Power Entry Module
D5 LED P1 1
F5 – Fan Fuse K1 – Base Power SS Relay
(50003)
D4 LED
D11 LED
F5 – Drive Pump Fuse
P5
D3 LED 3 1
F3 – 24 VDC Supply Fuse
D2 LED
SS Relay Mounting Screws
F2 – Heater & Fan 2 Fuse 15 13
D1 LED
F1 – PC Fuse
D12 LED
P3 P2
P4 K2– Heater SS Relay
Power Distribution 1 (50003)
PCB (AL-54743) 1 4
1
NOTE: Status LEDs D1-D5 are located next to their respective fuses, and are visible
externally from the back of the JFTOT. The Green D11 and D12 LEDs are
visible internally to indicate K1 and K2 power status.
NOTE: Refer to the Cable Pin-Out tables below and the JFTOT III Electrical
Schematic in “Drawings” Section 5.9.
TABLE 5-3
P-1CONNECTOR: RELAY CONTROL INPUT (4-CIRCUIT AMP MTA-156 20AWG)
Pin # Function
1 + Base power relay
2 - Base power relay
3 + Bus heater relay
4 - Bus heater relay
TABLE 5-4
P-2 CONNECTOR: 115/230V POWER INPUT (6-PIN AMP MATE-N-LOK)
Pin # Function
1 Safety ground
2 Ground
3 Line-A
4 Line-B
5 Line-C
6 Line-D
TABLE 5-5
P-3 CONNECTOR: STEP-DOWN TRANSFORMER (6-PIN AMP IN-LINE MATE-N-LOK)
Pin # Function
1 Primary-2 115V
2 Primary-1 0V
3 Primary-2 0V
4 Primary-1 115V
5 Secondary 0V
6 Secondary 115V
TABLE 5-6
P-4 CONNECTOR: PC POWER OUTPUT (3-PIN AMP MATE-N-LOK)
Pin # Function
1 Ground
2 Neutral
3 115/230VAC
TABLE 5-7
P-5 CONNECTOR: 115V OUT (15-PIN AMP MATE-N-LOK)
Pin # Function
1 Ground, Fan-1
2 Neutral, Fan-1
3 115VAC, Fan-1
4 Ground, Drive pump
5 Neutral, Drive pump
6 115VAC, Drive pump
7 Ground, 24VDC power supply
8 Neutral, 24VDC power supply
9 115VAC, 24VDC power supply
10 Ground, Fan-2 (on radiator)
11 Neutral, Fan-2
12 115VAC, Fan-2
13 Ground, Heater SCR
14 Neutral, Heater SCR
15 115VAC, Heater SCR
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components.
1. Switch off the power at the back of the JFTOT and unplug the
power connector.
2. Remove the right side panel and locate the 24-VDC Supply as
shown in Figure 5-53.
FIGURE 5-53
24-VDC SUPPLY
Cover Screws
(x4)
24-VDC Power
Supply
(AL-90697)
Standoffs (x4)
115-VAC Input
24-VDC Output
Connector
Connector
Ground Wire
5. Remove the four standoffs and then remove the power supply.
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components.
1. Switch off the power at the back of the JFTOT and unplug the
power connector.
2. Remove the right side panel and locate the heater control
assembly as shown in Figure 5-55.
3. Remove the wires (Figure 5-55) going to the SCR heater
control at screw-terminals 2, 3, Gnd, 7 and 8.
4. Disconnect the RMS current transformer lead connector.
5. Remove the heater power cables at the upper and lower bus
bar assemblies.
6. Remove the four mounting nuts (Figure5-55) on the left and
right sides at the bottom of the mounting bracket.
FIGURE 5-55
SCR HEATER CONTROL
Low-Voltage Transformer
(AL-90719)
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components.
1. Switch off the power at the back of the JFTOT and unplug the
power connector.
2. Remove the right side panel and locate the SCR heater control
(Figure 5-56).
3. Use a screwdriver to remove the two wires (5 & 6) on the left
and the three wires (2,3 & Gnd) on the right. Use a small flat-
blade screwdriver to remove the two wires (7 & 8) on the back
of the SCR heater control. See Figure 5-56.
FIGURE 5-56
SCR HEATER CONTROL
4. Remove the four mounting nuts at the base of the SCR heater
control and then remove it from the enclosure.
5. Use the removal steps in reverse order (Steps 4-1) as a guide to
replace the SCR heater control.
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components.
1. Switch off the power at the back of the JFTOT and unplug the
power connector.
2. Remove the right side panel and locate the low-voltage
transformer as shown in Figure 5-57.
3. Remove the entire heat control assembly as described in
Section 5.3.5.7.
4. Unwind and remove the bus-bar heater cable from the
transformer.
5. Remove the two primary input leads (Figure 5-57) on the right
side of the transformer.
FIGURE 5-57
LOW-VOLTAGE TRANSFORMER
Mounting Screws
(x2)
Low Voltage
Transformer Primary input
(AL-90719) Voltage from SCR
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components.
1. Switch off the power at the back of the JFTOT and unplug the
power connector.
2. Remove the right side panel and locate the RMS current
transformer as shown in Figure 5-58.
Current Transformer
(AL-90720)
Bus-Bar Power
Cable
Mounting Nuts
(x2)
FIGURE 5-59
JFTOT III EMBEDDED COMPUTER
CD ROM Drive
Perform the following steps to remove and replace the JFTOT III
embedded computer and touchscreen subsystem.
1. Switch off the power at the back of the JFTOT and unplug the
power connector.
2. Remove both side panels and locate the Computer subsystem
and connectors as shown in Figure 60.
3. Disconnect the power cord, keyboard, mouse, USB (x2), serial
I/O, DAQ and PC power switch cable connectors. See Figure 5-
60.
FIGURE 5-60
COMPUTER CABLES
DAQ Cable
Connector
115VAC Power
PC Power Switch
Connector
PC Mounting
Screws (x8)
NOTE: Ensure the computer power switch located on the back of the PC near the
power cord is always switched to the ON position.
CAUTION: Never use a sharp tool such as a pen or pencil to point to items on the
touchscreen. Doing so will permanently damage the screen. Use a stylus,
fingertip or a soft plastic tool to physically select items on the screen.
FIGURE 5-64
CALIBRATE 25 POINTS SCREEN
NOTE: The time for selecting each point on the screen is limited as indicated by the
movement of the status bar at the bottom of the screen. If the diamond has
not been pressed before the bar is completely yellow, the option box appears
and the process must be started over by selecting YES.
FIGURE 5-66
CALIBRATION-REPEAT SELECTION BOX
FIGURE 5- 69
CURSOR ALIGNMENT
CAUTION: The DAQ PCB contains static-sensitive parts that can be damaged when
touched. The installation of the Multifunction DAQ board should be done at
a static-free workstation.
PC Mounting
Screws (x3)
PC Mounting
Screws (x2)
Cut Tie-Wrap
PC Mounting
Screws (x2)
3.4 Remove the back cover being careful not to pull the
switch wire loose from the inside of the computer.
4. After the back cover has been removed, remove the metal
cover as listed below and as shown in Figures 5-71 and 5-72.
4.1 Remove the 6 screws (Figure 5-72 that hold the
metal cover.
FIGURE 5-71
JFTOT III PC METAL-COVER SCREW REMOVAL
CAUTION: Do not force the cover during removal. The metal cover has
interconnection circuitry attached that will be damaged unless the PC-board
edge plug is removed from the mating connector by alternately pulling up on
the cover one side at a time.
FIGURE 5-73
JFTOT III METAL COVER REMOVAL
5. Locate the two PCI expansion connectors in the top half of the
compartment as shown in Figure 5-74.
6. Loosen the screw that holds the protective cover (Figure 5-74)
to the inside expansion slot and remove the protective cover.
FIGURE 5-74
JFTOT III SLOT ACCESS COVER REMOVAL
7. Install the DAQ PCI card into the inside PCI expansion slot
and retighten the screw to hold the DAQ card securely in place
as shown in Figure 5-75.
FIGURE 5-75
DAQ PCB INSTALLED
9. The following screen (Figure 5-81) will appear. Make sure that
the Device Number is 1.
FIGURE 5-81
CONFIGURING DEVICE NUMBER
10. Click the “AI” tab and In the MODE select control
(Figure 5-82), select “Referenced Single Ended” then click
“Apply”..
FIGURE 5-82
AI MODE SELECT
11. Click on the “System” tab (Figure 5-83) and then click “Test
Resources”.
FIGURE 5-83
SYSTEM – TEST RESOURCES SELECT
12. A message should appear that states that the device passed all
tests.
NOTE: The solid-state relays and amplifier modules used in the PC Interface PCB
are supplied as separate parts and are not included with the replacement PC
interface PCB. The SSR and amplifier modules should be removed from the
interface PCB being replaced and then, if not defective, reinstalled into the
replacement interface PCB.
FIGURE 5-84
JFTOT III COMPUTER INTERFACE PCB
DP Amp Module
(AL-24638)
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components. Ensure that power lines are removed from
the power source before continuing.
2. Remove the left side panel and locate the PC Interface PCB as
shown previously in Figure 5-1A.
3. Remove each of the cables attached to the PC Interface PCB.
4. Remove the mounting screws from the PCB and remove the
PCB.
5. Remove the SSR and amplifier modules and retain for use in
the replacement PCB.
6. Use the removal steps in reverse order (Steps 5-1) as a guide to
replace the PC Interface PCB.
The JFTOT III provides control through solid state relays modules
(SSRM) plugged into the computer interface PCB board which
provide electrically isolated switching functions. A status LED (light
emitting diode) is located next to each SSRM for visual indication of
the ON/OFF condition. The following list (Table 5-8) describes the
function of each SSRM and the state indicated by the lighted LED.
TABLE 5-8
SOLID-STATE RELAY MODULE (SSRM) FUNCTIONS
Refer to the Cable Pin-Out tables below and the JFTOT III PC
Interface Drawing in Section 5.9.
TABLE 5-9
P2 CONNECTOR: VENT AND DP SOLENOIDS(10-PIN WAGO)
Pin # Function
1 + DP Bypass solenoid
2 - DP Bypass solenoid
3 + DP– Vent solenoid
4 - DP– Vent solenoid
5 + DP+ Vent solenoid
6 - DP+ Vent solenoid
Continued
Pin # Function
7 + System Vent solenoid
8 - System Vent solenoid
9 + Safety Vent solenoid
10 - Safety Vent solenoid
TABLE 5-10
P3 CONNECTOR: SYSTEM PRESSURE TRANSDUCER (4-PIN MOLEX MINI-FIT JR)
Pin # Function
1 - Excitation
2 + Excitation
3 - Signal
4 + Signal
TABLE 5-11
P-4 CONNECTOR: DP TRANSDUCER (6-PIN MOLEX MINI-FIT JR)
Pin # Function
1 - Excitation
2 - Excitation (-sense)
3 + Excitation
4 + Excitation (+sense)
5 - Signal
6 + Signal
TABLE 5-12
P5 CONNECTOR: 24VDC POWER SUPPLY INPUT (3-PIN AMP MTA-156 20AWG)
Pin # Function
1 +24VDC
2 No connection
3 Ground (24V return)
TABLE 5-13
P6 CONNECTOR: CONTROL RELAYS (10-PIN AMP MTA-156 20AWG)
Pin # Function
1 +Base Power Relay
2 - Base Power Relay (ground)
3 +Bus Heater Relay
4 - Bus Heater Relay (ground)
Continued
Pin # Function
5 - Drive Pump Enable (ground)
6 +Drive Pump Enable
7 +Aeration
8 - Aeration (ground)
9 +Water Pump
10 - Water Pump (ground)
TABLE 5-14
P7 CONNECTOR: ANALOG SIGNALS (12-CIRCUIT WAGO)
Pin # Function
1 + Tube temperature T/C (type K)
2 - Tube temperature T/C
3 + Sample temperature T/C (type K)
4 - Sample temperature T/C
5 + Bus Bar temperature T/C (type K)
6 - Bus Bar temperature T/C
7 Hi – RMS Volts input
8 Low – RMS Volts input
9 Hi – RMS Amps input
10 Low - RMS Amps input
11 Hi – 4-20mA Heater Control output
12 Low – 4-20mA Heater Control output
TABLE 5-15
P8 CONNECTOR: HUMIDITY SENSOR (3-PIN AMP MTA-100 24AWG)
Pin # Function
1 Analog ground
2 Signal
3 +5VDC Reference
TABLE 5-16
P9CONNECTOR: SPARE ANALOG SIGNALS (FUTURE) ( 8-PIN WAGO PN:232-308/026-000)
Pin # Function
1 + Input, Spare #1 (T/C or process)
2 - Input, Spare #1
3 + Excitation, Spare #2
4 + Input, Spare #2 (T/C, process or strain gage)
Continued
Pin # Function
5 - Input, Spare #2
6 - Excitation, Spare #2
7 Hi output, Spare #3 (volts or mA)
8 Low output, Spare #3
TABLE 5-17
P10 CONNECTOR: EXTRA DIGITAL CONTROL SIGNALS FROM DAQ CARD (FUTURE)
( 20-CIRCUIT RIBBON CABLE CONNECTOR)
Pin # Function
1 +5VDC
2 Digital ground
3 SCANCLK
4 EXTSTROBE
5 PFI0/TRIG1
6 PFI1/TRIG2
7 PFI2/CONVERT
8 Digital ground
9 PFI3/GPCTR1_SOURCE
10 PF14/GPCTR1_GATE
11 GPCTR1_OUT
12 PFI5/UPDATE
13 PFI6/WFTRIG
14 Digital ground
15 PFI7/STARTSCAN
16 PFI8/GPCTR0_SOURCE
17 PFI9/GPCTR0_GATE
18 GPCTR0_OUT
19 FREQ_OUT
20 Digital ground
TABLE 5-18
P11 CONNECTOR (14-CKT RIBBON): EXTRA ANALOG SIGNALS FOR DAQ CARD (FUTURE)
Pin # Function
1 ACH12
2 Analog ground
3 ACH13
4 Analog ground
5 ACH14
6 Analog ground
7 ACH15
8 Analog ground
9 +5VDC Reference (same as Humidity sensor)
10 Digital ground
Continued
Pin # Function
11 +5VDC
12 Digital ground
13 +24VDC
14 Digital ground
The names of the physical components (as labeled on the PCB) are
listed under the “Test Point” column. The “Reference” column lists
the function and the “Description” column describes the function.
The test point name and the reference name are located near the
component on the PCB.
NOTE: Test points TP7-TP27 are available on PCB revision A and later.
TABLE 5-19
TEST POINTS
1
Calibration of signals is performed in software by the JFTOT-III application program.
1
Calibration of signals is performed in software by the JFTOT-III application program.
TABLE 5-20
STATUS LEDS (CONTINUED)
NOTE: Refer to pertinent tables and figures in this section for location and
functions of the circuit being tested.
To measure system pressure only, perform the following steps. Refer to the JFTOT
III PC Interface PCB Layout Drawing in Section 5.9 for pin locations.
1. Before making any measurements, insure the PC is on and running the JFTOT
III program and all three green LEDs (D10, D11, and D12) on the Interface PCB
are fully lighted. See Table 5-20 “Status LEDs”.
2. Measure the +5V voltage that powers the 5B amplifier modules. The voltage
between test points TP3 (+) and TP1 (–) should be 5.00VDC ± 0.20VDC. If
the voltage is not within tolerance, check for short circuits on the PC Interface
PCB and if necessary, replace the PC Interface PCB (AL-54741).
3. The voltage between TP24 (+EXC) and TP26 (–EXC) should be 10.00VDC
± 0.010 VDC. If the voltage is low or high, replace the system pressure
amplifier module (U5).
4. The voltage between TP25 (+SIG) and TP27 (–SIG) should be in the range of
approximately 0.0 to 20mVDC. This voltage is the signal from the pressure
transducer. 0mV = no pressure (0 psig), 20mV = high pressure (500 psig).
5. The voltage on test points TP13 (+) and TP6 (–) should be in a range of
0.0VDC to 5VDC. This voltage is the amplified output signal from the system
pressure amplifier module (U5). For Example: 0.0VDC = no pressure (0 psig),
3.33VDC = high pressure (500 psig).
6. To see if the amplified output signal is correct, use a millivolt source with a low
impedance output, to input a 0-20 millivolt signal.
7. Without disconnecting P3, input the millivolt signal to TP25 (+SIG) and
TP27
(–SIG). Input 0.0mVDC for 0.0 VDC out, and input 20.0mV for 3.33 VDC
out. If readings are not close it may be necessary to replace the system pressure
amplifier module (U5).
Perform the following steps to measure differential pressure (DP) only. Refer to the
Drawing in JFTOT III PC Interface PCB Layout Drawing in Section 5.9 for pin
locations.
1. Before making any measurements, insure the PC is on and running the JFTOT
III program and all three green LEDs (D10, D11, and D12) on the Interface PCB
are fully lighted. See Table 5-20 “Status LEDs”.
2. Measure the +5V voltage that powers the 5B amplifier modules. The voltage
between test points TP3 (+) and TP1 (–) should be 5.00VDC ± 0.20VDC. If
the voltage is not within tolerance, check for short circuits on the Interface PCB
and if necessary, replace the Interface PCB (AL-54741).
3. The voltage between TP20 (+EXC) and TP22 (–EXC) should be 10.00VDC
± 0.010 VDC. If the voltage is not within tolerance, it may be necessary to
replace the DP amplifier module (U4).
Back-Pressure
Control Adjustment
3. Touch or click the MAINT button on the Main Menu and then press or click
the DRV PUMP button on the maintenance screen as shown in Figure 5-86.
FIGURE 5-86
MAINTENANCE MENU
4. Prime the pump if needed. “See Priming the Drive Pump” in subsection 5.4.2.
NOTE: The drive pump will not start until the flow value is incremented or
decremented.
5. Start the drive pump to pump at 3.000 mlpm and wait for the flow to come out
of the sample out line (approximately 20 drops in 9 seconds) in the bus-bar
compartment (See Figure 5-87).
FIGURE 5-87
JFTOT III SAMPLE OUT FLOW
NOTE: The drive pump valves may stick occasionally and will need to be cleaned
before the correct flow rate can be obtained. To ensure the correct flow is
obtained, always count the drips coming from the sample out line into the
sample out container. Approximately 20 drops in 9 seconds indicates a flow
of 3 mL/min.
6. Watch the “Sys Pres” indicator (Figure 5-86) and adjust the back-pressure
regulator knob (Figure 5-85) clockwise to increase or counterclockwise to
decrease pressure until the “Sys Pres” indicator reads 500 PSI.
7. Stop the drive pump and vent the system.
8. Replace the cover on the left side of the JFTOT III.
NOTE: It should not be necessary to adjust the controls on the pump itself. Doing so
could cause setting conflicts. Refer to the “Dual Piston Pump Operator’s
Manual” supplied with the JFTOT III for details concerning the drive pump.
1. Set up to run a JFTOT III test. Place a sample in the sample in container and
an empty container in the sample out position.
2. Remove the system pump compartment door on the bottom front left of the
JFTOT III to access the pump as shown in Figure 5-88.
FIGURE 5-88
JFTOT III DRIVE PUMP
Prime/Purge Valve
Output
Dual Inputs
3. Touch or click the MAINT button on the main menu to display the
Maintenance Screen as shown in Figure 5-89.
4. Touch or click the SYS VENT button on the maintenance screen.
5. Touch or click the DRIVE PUMP button on the maintenance screen.
FIGURE 5-89
JFTOT III MAINTENANCE SCREEN
6. Place a napkin or rag under the two pump heads on the drive pump.
7. Double-click inside the FLOW display on the Maintenance screen and then
key in 9.999 into the flow display. Press the ENTER key.
8. Click or touch the DRV PUMP button on the screen to start the pump running
at the 9.999 mlpm flow rate.
FIGURE 5-90
FLOW RATE ENTRY
9. Compress and insert a suction bottle onto the orifice at the center of the knob
on the pump control panel as shown in Figure 5-91.
FIGURE 5-91
SYSTEM PUMP PRIME BUTTON
10. With the compressed suction bottle in place in the prime/purge valve, rotate
the black knob counterclockwise (ccw) until fuel begins to flow out of the
center orifice.
11. Relive pressure on the suction bottle and remove about 20 ml of sample from
the orifice at the center of the knob as shown in Figure 5-92.
FIGURE 5-92
PUMP PRIMING
12. Close the knob and remove the suction bottle. Wait for 9.999 mlpm flow to
come out of the purge out line. The flow at 9.999 mlpm should be very fast
drips (almost a steady stream).
13. Double-click inside the FLOW display on the Maintenance screen, key in
3.000 into the flow display and then press the ENTER key. A 3mlpm flow rate
is about 20 drips in 9 seconds.
14. Clean up any spillage and replace the front cover to the Drive pump
compartment.
NOTE: After the drive pump has been replaced with a new pump or after the pump
has been stored for a period of time, the individual pump heads may need to
be primed directly in order to obtain proper pump output pressure.
NOTE: The ideal adjustment provides 60% power when running a D3241 test at
260°C. However, a 260°C D3241 test will run properly with readings of
50% to 80% power. The % power readings at 260°C may differ from 60%
by ± 10% due to variations in line voltage, cleanliness of bus bars (front
and inside connections), and age of JFTOT. The % power reading will be
different for set points and flow rates other than 260°C and 3.0 mL/min
respectively. The SCR output to the heater block and the heater tube can be
adjusted by changing the % power
2. Open the right side of the JFTOT III and locate the SCR heater control as
shown in Figure 5-93.
FIGURE 5-93
SCR HEATER CONTROL LOCATION
3. Touch or click the MAINT button on the Main Menu and then press or click
the DRV PUMP button on the maintenance screen as shown in Figure 5-94.
FIGURE 5-94
MAINTENANCE MENU – DRIVE PUMP ON
NOTE: The drive pump will not start until the flow value is incremented or
decremented.
4. Start the drive pump to pump at 3.000 mlpm and wait for the system pressure
to reach near 500 psi and flow to come out of the sample-out line in the bus-
bar compartment. Prime the pump if needed. (See “Priming the Drive Pump”
in Section 5.4.2)
NOTE: The drive pump valves may stick occasionally and will need to be cleaned
before the correct flow rate can be obtained. To ensure the correct flow is
obtained, always count the drips coming from the sample out line into the
sample out container. Approximately 20 drops in 9 seconds indicates a flow
of 3 mlpm.
5. Enter a value of “15” into the %PWR control and touch or click the BUS
HEATER button as shown in Figure 5-95. The coolant pump should start.
6. Adjust the coolant flow-meter to the green area.
FIGURE 5-95
JFTOT III BUS-HEATER CONTROL
CAUTION: There is about a 20 second soft-start built into the SCR so do not try to
adjust anything for 20 seconds.
7. Using a small jewelers screwdriver carefully adjust the “BIAS” control on the
SCR (Figure 5-96) until there is a reading of 0.005 indicated in the “Volts”
indicator.
FIGURE 5-96
JFTOT III BUS HEATER CONTROL BIAS ADJUSTMENT
BIAS Adjustment
8. Slowly increase the value in the %PWR button (Figure 5-89) in 10%
increments while watching the heater tube temperature. If the temperature
gets above 260°C then turn down the “GAIN” control on the SCR (Figure 5-
97). Continue increasing the %pwr while watching the temperature and
turning down the “GAIN” control as necessary until the value in the “%pwr”
control is 60% and the heater-tube temperature is 260°C.
FIGURE 5-97
JFTOT III BUS HEATER CONTROL GAIN ADJUSTMENT
GAIN Adjustment
NOTE: Because the BIAS and the GAIN control affect each other, this procedure
may have to be repeated several times. The goal is to have the “volts”
indicator read 0.005 when 15% power is applied and the heater-tube
temperature read 260°C when 60% power is applied. Enough time must be
allowed for the system to become thermally stable after applying 60%
power.
WARNING: A potentially lethal electrical shock hazard exists when contact is made with
the 115/230-Volt line components.
TP5 TP6
6 If the reading is above or below +5.00 VDC, adjust trimpot R4 (Figure 5-98)
to obtain a reading of +5.00 VDC.
7. Remove the meter leads and install the side panel.
NOTE: For 115 VAC use one 8A-3AG, Fast (PN: AL-29686) and a shorting clip.
If metal shorting clip is not available, install an 8A Fuse in both sides of the
holder. For 230 VAC use two 6.3A, 5x20mm SB (PN: AL-29738) without the
shorting clip.
FIGURE 5-99
115-VAC POWER ENTRY MODULE REMOVAL
A B C
5. Install a 6.3-A, 250-VAC SB fuse at the rear of the holder (opposite the
panel end) in both sides of the module as shown in Figures 5-101A
and 5-101B.
FIGURE 5-101
230-VAC FUSE INSTALLATIONS
A B
FIGURE 5-104
230VAC FUSE REMOVAL
A B
4. With the fuse holder oriented with the 115V label up, install an 8A fuse
into the right-hand side of the holder as shown in Figure 5-105A.
5. Install the metal shorting clip at the rear on the opposite side of the
holder as shown in Figure5-105B. If the metal shorting clip is not
available, install an 8-A fuse in both sides of the holder.
FIGURE 5-105
115-VAC FUSE AND SHORTING CLIP INSTALLATIONS
A B
8. A window will appear (Figure 5-110) that allows you to select the features you
want to install and where you want to install them. Click the NEXT >>
button to install all of the features in the C:\Progran Files\National
Instruments\NI-DAQ\ folder.
FIGURE 5-110
NI-DAQ FEATURES SCREEN
9. A license agreement (Figure 5-111) will appear. Click on the round “I Accept
the License Agreements” button to accept the agreement.
11. A second License Agreement will appear (Figure 5-112). Click “I Accept the
License Agreements” and click NEXT >>.
FIGURE 5-112
SECOND LICENSE AGREEMENT SCREEN
12. The Start Installation window (Figure 5-113) will appear. Click NEXT >> to
proceed with the installation.
FIGURE 5-113
START INSTALLATION WINDOW
13. The Overall Progress indicator (Figure 5-114) will be displayed. Allow the
installation to continue until prompted to install the second CD.
FIGURE 5-114
OVERALL PROGRESS SCREEN
14. When the Insert Source window prompts you (Figure 5-115), remove the CD
labeled Disk 1 of 2 and insert the CD labeled Disk 2 of 2.
15. If the installation does not resume on its own, then click the RESCAN DRIVE
button (Figure 5-115) to continue the installation.
FIGURE 5-115
INSERT SOURCE SCREEN
16. The installation will continue until it is complete. When the Installation
Complete window (Figure 5-116) is displayed, click FINISH.
FIGURE 5-116
INSTALLATION COMPLETE WINDOW
17. You will be prompted to restart the computer as shown in Figure 5-117.
FIGURE 5-117
RESTART COMPUTER SCREEN
18. Click the SHUT DOWN button to turn off the computer and get ready to install
the data acquisition hardware.
FIGURE 5-118
OPTIONS SCREEN
2. From the Control Panel, double-click the “Add or Remove Programs” icon
as shown in Figure 5-119.
FIGURE 5-119
CONTROL PANEL SCREEN
3. From the Add or Remove Program Screen, click on the “Add New
Programs” icon as shown in Figure 5-120.
4. Click on the CD OR FLOPPY button as shown in Figure 5-120. A window will
appear (Figure 5-121) and prompt you to install the program CD into the
CD drive.
FIGURE 5-120
ADD OR REMOVE PROGRAMS SCREEN
5. Put the JFTOT III software CD into the CD drive and click the NEXT button
as shown in Figure 5-121.
FIGURE 5-121
INSTALL PROGRAM SCREEN
6. The computer will look on the CD for the installation program. It should find
the file D:\setup.exe. Click the FINISH button as shown in Figure 5-122.
FIGURE 5-122
RUN INSTALLATION PROGRAM SCREEN
8. The JFTOT III Installation Wizard Welcome window will appear. Click the
NEXT button as shown in Figure 5-124 to start the installation.
FIGURE 5-124
INSTALLATION WIZARD WELCOME SCREEN
9. The JFTOT III Software installation will start running and the status screen
will be displayed as shown in Figure 5-125.
FIGURE 5-125
JFTOT III UPDATING SYSTEM STATUS BAR
FIGURE 5-126
JFTOT III VER 2 INSTALLED SUCCESSFULLY SCREEN
2. Locate the JFTOT III program on the programs option display and right-click
the JFTOT III icon as shown in Figure 5-127.
3. Click on the “Create Shortcut” option (Figure 5-127) and drag the shortcut
onto the computer MS Windows desktop.
FIGURE 5-127
JFTOT III CREATE SHORTCUT SCREEN
WARNING: A severe electric shock hazard may exist when power is applied while
servicing internal components. Ensure that the power is removed from the
power source before continuing.
TABLE 5-23
TROUBLESHOOTING GUIDE (CONT.)
Spare Parts for the JFTOT III are listed in the following table.
TABLE 5-24
JFTOT III SPARE PARTS LIST
# ITEM PART NO
TABLE 5-24
JFTOT III SPARE PARTS LIST (CONT)
# ITEM PART NO
23 Hinge, Test Section Cover 61069
24 Glass Sample Beaker 68019
25 Scrubber, Dririte for Aeration (Includes Dririte) 71283
26 Pump, HPLC Series 1500, Self Flush 72020
27 Power Entry Module, IEC plug/Fuse/Switch 80085
28 Clamp, Plastic 1.25” ID (Water Pump Mount) 83234
29 Jack, RJ45 Cat 5e In-Line Modular 84351
30 Sensor, Humidity, 1/8” tube ends 92243
31 Power cord, European 230V 97107-1
32 Power cord, UK 230V 97107-2
33 Tube, Cooling Jet AL-10831
34 Plate, Bus Bar Mounting, JFTOT-III AL-10835
35 Bus Bar Assembly, JFTOT-III (2 required) AL-10836
36 Insulator Bushing Kit AL-11926
37 Cover, Sample Vessels, JFTOT-III AL-13227
38 Cover, Heated Test Section, JFTOT-III AL-13228
39 Bottle, Water Reservoir, JFTOT-III AL-17683
40 Jack, Type K T/C, Female Panel Mount AL-23863
41 Amplifier Input Module, Strain Gage AL-24638
42 Backup Screen, Pre-Filter AL-27051
43 Prefilter Assembly AL-27075
44 Filter, Inlet, 10um, SS 1/8” AL-27096
45 Fitting, Cooling Jet AL-28389
46 Fitting, Inlet Filter (for AL-27096) AL-28432
47 Fuse, 8A Fast 3AG ¼” x 1-¼“ AL-29686
48 Fuse, 6.3 A SB 5mm x 20mm AL-29738
49 Fuse, 2A, PCB Mount AL-29735
50 Fuse, 4A Fast 3AG ¼” x 1-¼” AL-29736
51 Fuse, 1.5A Fast 3AC 1/4:” x 1-¼” AL-29737
52 Heater Test Section Housing, JFTOT AL-35291
53 AutoCal Heater Assembly, Lead, JFTOT-III AL-35387
54 Differential Pressure Transducer AL-45042
55 System Pressure Transducer, 500psi AL-45043
56 Tubing, Teflon, ¼” OD x .047” wall AL-49291
57 Tubing, 5/16” OD x .063 wall AL-49292
58 Interface PCB Assembly, JFTOT-III AL-54741
59 O-Ring, Line Connector (25 per pack) AL-53706
Continued
TABLE 5-24
JFTOT III SPARE PARTS LIST (CONT)
# ITEM PART NO
TABLE 5-25
JFTOT III SUPPLED SPARE PARTS
2 5 Fuse, 8 A AL-29686
3 10 Fuses, 4 A AL-29736
11 2 Boxes, Heater Tube and Filter Kits (10 per box) AL-91652
TABLE 5-26
JFTOT III ACCESSORIES
2 1 Brush, ¼” x 2” AL-09970
10 1 Tweezers AL-38009
TABLE 5-26
JFTOT III MATERIALS
# Item Part No
1 Aluminum Foil AL-38008
The following additional items are not supplied with the JFTOT III but are required
for normal operation:
• ALCOR Visual Tuberator: The ALCOR Visual Tuberator (AL-85417, 115VAC
and 85418, 230 VAC) is required for determining the results of the JFTOT III
tests but must be ordered as a separate item.
# Item Part No
1 Line Voltage Conditioner, 120 VAC AL-90714
5.9 DRAWINGS
The following drawings are provided for reference during the performance of
maintenance and service functions described in the previous subsections.
TABLE 5-29
JFTOT III DRAWINGS
# Drawing
1 JFTOT III Power Distribution PCB Layout Drawing ( Dwg No. 54743 Rev A)
2 JFTOT III Power Distribution PCB Schematic ( Dwg No. 64990 Rev A)
3 JFTOT III PC Interface PCB Layout Drawing( Dwg No. 54741 Rev A)