Beruflich Dokumente
Kultur Dokumente
00-5006
liquiTOC
Operating Instructions
Contents
1. Scope of Delivery:
Basic unit :
♦ liquiTOC
carbon/nitrogen analyzer for the determination of C or C and TNb in aqueous samples,
including instrument software in German or English, basic equipment.
Basic equipment sufficient for installation and training .
order no. 35.10-0000 (C version)
order no. 35.20-0000 (C/TNb version)
Necessary Accessories :
Optional accessories :
♦ Solids module
for instrument modification for the determination of TC in solids, suspensions and
sludges.
Recommended work range: max. 500 mg sample weight or max. 100 mg C abs.
For further ordering information for spares or consumables, see appendix and the newest
catalog for consumables.
2.1 Specifications
Operating modes TC, TIC, TOC, NPOC, POC as well as combinations
thereof, also TNb simultaneously, depending on
instrument version
calibration linear
or selectable 1 or multipoint calibration, fully
automatic and/or acc. to customer's specification
precision / TOC/NPOC/TC/TIC
standard deviation* ≤ 2 % at ≥ 10 mg/l C
≤ 5 % at ≥ 0.2 mg/l C
(*depends on injection volume; TNb (EC detector)
values based on 2ml)
≤ 5 % at > 5 mg/l N
≤ 4 % at > 20 mg/l N
3. Safety Instructions
General instructions
• The analyzer liquiTOC is suitable only for the carbon/nitrogen analysis in aqueous
samples, which have a controllable combustibility under conditions depending on the
mode of operation. Analyzing aggressive chemicals, acids, solvents, explosives or
materials that can form explosive gases is explicitly forbidden. It can lead to damages
of the instrument and to injuries of the operating personnel. For the above mentioned or
other difficult applications, we recommend to use the counseling possibilities of our
application laboratory, in order to work out safe operating procedures and to use certain
parameters or special equipment.
♦Operation, maintenance and repairs should only be carried out by trained personnel.
For a smooth and efficient operation of the liquiTOC, please read the following operating
instructions thoroughly. Please keep the operating instructions for future questions. We
reserve the right of deviations in illustrations and technical data described in the operating
instructions due to further developments. The data shown in the software illustrations are
only examples and can deviate from the actual parameters to be set. Please take the
parameters valid for your instrument from the supplied installation CD.
Installation / Start-up
• Operation of the liquiTOC and peripheral equipment should only be carried out by
persons who have been authorized by E l e m e n t a r Analysensysteme GmbH.
Upon request, technically experienced instrument users may put the instrument into
operation themselves. The instructions of the manufacturer – especially the safety
instructions – shall be adhered to at all times. No liability by the manufacturer for damages
caused by improper handling.
• Opening the transport packing by unauthorized personnel can lead to loss of
warranty rights! If the supplied instruments show damages, please contact the freight
forwarder and the manufacturer.
• For the operation of the liquiTOC analyzer it is mandatory to use the furnished power
supply cable only!
• The end of the gas pipes must lead to a sheltered (wind-free) place in the open air to avoid
fluctuations of the base line which would lead to detector instabilities.
• Liquid residues can be very acetous and must be neutralized before leading them into the
sewerage system. If the analyzed samples contain compounds which are hazardous to the
water, the liquid residues will have to be disposed of according to the respective
regulations.
Reagents
• When handling chemicals, observe the safety instructions of the chemical manufacturer
(see labels on bottles and safety data sheets)!
• When handling hydrochloric acid, Sicapent® (Phosphorpentoxide) or magnesium
perchlorate, it is absolutely necessary to wear safety goggles and protective gloves.
!!!Danger of Causticization !!!
• When handling cold quartz and glass parts wear the supplied leather gloves to avoid
danger of injuries in case of broken glass.
Maintenance/Exchange of tubes
• The efficiency of the liquiTOC can only be guaranteed when original consumables as
well as original spare parts which can be obtained only through Elementar
Analysensysteme GmbH or their authorized dealers, are used. The use of different
consumables or chemicals of an unsuitable standard can lead to injuries as well as to
instrument damages. It can also falsify the results of the analysis. Unsuitable spares
lead to instrument damages and loss of warranty.
• All the operation dependent maintenance and exchange times mentioned in the Operating
instructions depend very strongly on the respective test matrix. They are, therefore, only
examples and do not guarantee life times for consumables or wear and tear parts!
• The walls of the liquiTOC should only be opened by authorized personnel for
maintenance and servicing purposes.
• Before opening the rear wall of the liquiTOC, disconnect all peripheral equipment and
the liquiTOC itself by turning the main switch off and pulling the mains plug. Danger
of electricution - 230 Volt !!!
Attention: the furnaces must be cold otherwise components will become overheated. The
walls must be properly in place before starting a new analysis. (Check grounding
connection.)
Operating instructions liquiTOC 03/2003
Copyright by E l e m e n t a r Analysensysteme GmbH
Safety Instructions page 3-3
• The liquiTOC and the gas pipe system must be checked periodically for leaks.
• For safety reasons, the hydrochloric acid must be removed from the storage container
before checking for leaks.
• Never subject the IR-detector to overpressure. It could damage the detector.
• Replacing the reactor can only be carried out when the tube is cold. For this purpose
set furnaces to nominal temperature 0°C and let the tubes cool down to room temperature
(max. 100°C).
When working at hot furnaces or tubes there is danger of burning!!
Therefore, always wear the supplied protective gloves (order No. 05 000 094), because it is
possible that some components (e.g. heated tube line) are still hot. We are not liable for
accidents that are due to disregarding these warnings.
• Before replacing the reactor turn off the gas supply into the instrument.
• After changing the reactor wait until the furnaces as well as the new reactor has warmed
up (ca. 10 to 15 minutes) before starting the next analysis. In addition, it is necessary to
adhere to the conditioning rules for the tubes (chapter 9)
• Before installation into the furnace all finger prints must be cleaned off the quartz
parts (for instance with acetone) to avoid crystallization areas which lead to premature
aging of the quartz.
• If the supply voltage has been interrupted for more than 15 minutes, the liquiTOC must
be checked for leaks. If necessary, replace all o-rings inside the furnace chamber since they
may have been brittle because of the interrupted ventilation.
carrier gas synthetic air (carbon hydrogen ≤ 0.1 ppm, CO2 ≤ 1 ppm)
prepressure 1.0 bar
use: approx. 500 ml/min
5
2
6
7
8
legend to
illustration
4.1 1 reactor
2 furnace, post combustion zone
3 furnace, combustion zone
4 reactor connection, bottom
(gas and sample supply)
5 reactor connection, top, to the cooler
6 fan
7 electrical connections of the furnace
8 thermoelement connections
1
8
5
3 4
legend to
illustration
4.2 1 cooling vent for reactor
2 membrane drying
3 mains switch
4 power supply plug
5 connection sampler
6 connection PC
7 connection CLD
8 connection TIC
1 4
10
11
3 8
7
5 6
1 2 3
4 5 6 7 8
illustration 4.4: connection panel
legend to
ill. 4.4
1 connection water supply
2 connection acid supply
3 gas outlet IR and membrane drying λ 450 ml/min
4 connection carrier gas approx. 1 bar
5 gasing water supply
6 gasing acid supply
7 gasing sample
8 waste
Before opening the back panel cut off the power supply
for peripheral instruments and the analyzer by turning
off the mains switch and pulling the power plug!
mains
connection
IR
F1: 2 A
F2: 2 A
F3. 10 A
F4: 10 A
F5: 50 A
F6: 50 A
4.3.1 Reactor
catalyst
Cu chips
silver wool
The right absorption tube (absorption tube 1) will be filled with one half Ag-wool
(approx. 3 g) and one half Cu chips. The ends will be closed with filter material. Make
sure that the tube with the Cu chips is installed toward the top.
The left adsorption tube will be completely filled with 6 g Ag-wool. The filter material
is not needed.
4.4 Start-up
a) Set instrument up (for instructions regarding placement area see chapter 4.1.1) and
connect peripheral instruments.
autosampler
CLD
liquiTOC ¡ ¬
PC printer
Illustration 4.4 shows the connections on the left side panel of the liquiTOC. The acid
or water vessel is connected to the gasing acid supply and/or gasing water supply
(HCl or H2O) with a thin hose and to the connection water and/or acid supply (HCl
or H2O) with a thicker hose. The gasing sample for sample vessels is usually closed
with a blind stopper. The carrier gas is connected to the first inlet (O2/N2). Outlet for
gas (IR) must lead into open air (see safety instructions above). Outlet for liquids for
waste must lead into a non-metal container (for disposal see safety instructions below).
Note: It is important that the supplied hoses are used for exhaust and waste. The
prescribed dimensions and the material of the hoses are necessary for
failure free operation (for order information see catalog for consumables).
Note: for first start-up remember to remove the cable ties from the hoses.
c) Fill quartz and glass parts (reactor) and install (see chapter 4.3). Fill water and acid
container with water first.
d) Loosen valve out of the transportation clamp below the reactor and connect it with ground
in clamps to the reactor
e) Turn PC and printer on and wait for booting procedure to be ended. If necessary - install
software (see chapter 7) and remain in Windows-Desktop.
i) Turn gas supply on and set prepressure. Then check gas flow at the flow meter displays of
the software and - if necessary - re-adjust using an external flow meter.
gas pressure
Set the prepressure to approx. 1.0 – 1.2 bar at the bottle pressure
reducing valve. The pressure display at the PC should show
0.95 – 1.0 bar.
flow rate
outlet IR 200 ml/min
This flow rate has been factory set. The value will be displayed at
the "flow MFC" bar as well as at the "FM" bar. If the instrument
is duly leak proof, both displays will be almost identical.
MFC shows the outcoming gas flow and FM shows the incoming
gas flow.
The display "FM" may vary during an analysis, especially during
the combustion.
A flow check must always to be carried out in standby.
outlet
sample flushing / 30-40 ml/min each (factory set)
reagents flushing
k) Perform leaktest. Close the flexible hose with the connected clamps on the left
instrument side (with side wall taken off). If the flowmeter display shows a drop to 0
within 30 seconds, the leak test has been passed.
l) Fill hydrochloric acid into the acid vessel (see chapter 4.3.4).
m) Start heaters and check parameter setting at the PC (see chapter 5.3.5). The catalyst
furnace should have reached its operating temperature after 15 minutes. The catalyst
has to be flushed for approx. 10 minutes with the carrier gas turned on.
n) Select in the menu System/Feeding the type of feeding; select maybe autosampler type
and amount of multiple determinations.
o) Measure blank values of water (see chapter 5.1). If the blind value peak area is
somewhere around 7.00 area units, it is possible to already measure test standards of
100mg C/l. For measurements below 5 ppm a stable blind value (< 2,00 area units)
are necessary.
Note: The liquid system will also be flushed air-free during the first measurements.
Upon delivery of the instrument, the following concentration ranges are predefined:
The column sensitive refers to the peak recognition parameters. If this function is
activated, more rigid criteria will be determined for the peak beginning and the peak
end (stability of baseline, admissible difference of the peak end to the baseline, etc.).
In general, an activation of this function will lead to a better precision in the lower
concentration range (< 0.5 mg/l) but results in a longer measuring time.
1) general operational test during initial operation and/or after installation of a new reactor
and
2) the routine analysis.
¾ Fill 5 sample vials of the autosampler with deionized water and put them on the
autosampler. If no autosampler is available, the sample hose is lead directly into the
corresponding glass vial.
Note:
There are special positions on the autosampler for control standards or blanks. These
special positions can be defined in the column "hole pos." by entering e1, e2 or e3.
¾ Determine the number of determinations per sample under System / Feeding (1, 2 or 3),
e.g. 3.
¾ Open the following input diaglogs via double click in the sample field (Text-View) :
¾ After closing the input dialogs the measuring series can be started.
¾ Now the number of selected blank determination will be procesed. The TIC and NPOC
blank should be down to peak areas of < 3.0 within 15 blank measurements. In case that
after 15 measurements there is no stable blank value, the blank measurements have to be
continued.
¾ With a stable blank value test standards can then be measured, e.g. 20 mg TIC/
20 mg NPOC or 50 mg TIC / 50 mg NPOC per liter etc., in order to check the stability of
the measuring results.
The procedure is as follows:
1. e.g. fill 3 sample vials of the autosampler with test standard 20 mg TIC/l + 20 mg
NPOC/l.
2. Enter the name 20+20 in the input dialog "Name" for the number of requested
measurements, e.g. 3 sample vials à 3 determinations = 9x.
3. Enter the name "20 ppm" in the input dialog "Conc.range". By clicking "OK" the
concentration range will automatically be copied for all entered sample names.
¾ After closing the input dialogs the measuring series can be started.
¾ Now the number of the selected blanks will be processed. For statistical evaluation you
will find the statistics function under Math / Statistics.
If the statistical evaluation has a satisfying precision the routine analysis can be started.
¾ Fill sample vial no. 1 of the autosampler with a test standard referring to the application
for conditioning (example 20+20).
¾ Fill sample vial no. 2 of the autosampler with deionized water for blank determination.
¾ Fill sample vial no. 3 of the autosampler with a standard referring to the application
(example 20+20) for checking the calibration.
Notice: name and concentration range will be entered manually and/or copied,
the number and the position of the hole will be generated automatically.
¾ From sample vial no. 4 on, the real samples can be put in the autosampler, by entering the
relevant concentration range.
¾ Prior to the start, check the following points (PC display, instrument in standby):
¾ For permanent operation click symbol , now all samples entered in the PC will
be processed. In case the blank value and the result of the calibration test have changed by
more than 10% relative, a recalibration can be performed after completion of the
measuring series in menu Math. The results of the real samples can then be recalculated
with the new calibration with function Math / Recalculate.
6.1 TC Mode
¾ Depending on the chemical compound and on the affinity toward water, the
POC peak is very sharp and quick (e. g. halogenized hydrocarbons) or it is
broad, flat and slow (e. g. acetone, alcohols). The integration time of the POC
peak is therefore limited in time.
7.1 General
The following software description is only an abbrevated version. A complete description for
the individual menus can be found in the implemented help file of the software.
The liquiTOC software is a software to control the liquiTOC analyzer. It can be run under
the Windows platform. The Windows operating systems 95, 98, NT (from version 4.0 on) and
2000 will be supported. The communication of the Windows PC will be done via a serial
interface (RS232). Basically, the liquiTOC software allows the following actions:
7.2 Installation
For the installation of the liquiTOC software insert the furnished CD into the CD-ROM drive
The CD will start automatically. If the CD does not run automatically a start with
CDMNENUPRO.EXE will be possible. Choose the language. Choose Software Installation
in order to start the setup program.
Subsequently, the program tries to create the communication connection with the analyzer. If
it was sucessful, the status field appears with a light green background, otherwise the
background will be light red.
caption bar
menu bar tool bar
status field
sample
field
combi
window
status bar
The operating elements and their effects will be described in the following.
The status field can be activated by clicking with the mouse to any place within its
window. If the window is activated, the window title appears in dark blue.
The sample field can be activated by clicking with the mouse to any place within its
window. If the window is activated, the window title appears in dark blue.
The combi window can be activated by clicking with the mouse to any place within its
window. If the window is activated, the window title appears in dark blue.
The three windows "status field", "sample field" and "combi window" can be changed
among each other in their size. The mouse pointer shall be moved either onto the
vertical line between sample field and combi window or onto the horizontal line
between status field and sample field and/or combi window until it gets the shape of a
double arrow. Then press the left mouse button and pull the line (while mouse button
is held down) in the requested direction. The window sizes set at the latest will be
saved. This means that during a new start of the program there will be the same
division of windows like when leaving it.
For editing sample data further functions are available which simplify the input – apart
from the "normal input". These are:
The before mentioned types of input can be set via the menu "Options / settings /
input".
File / New
Creates a new document. If necessary, the user will be asked to save the data which
are currently in the sample storage.
File / Open
Opens an existing document. A diaglog field appears where the document to be
opened can be selected. If necessary, the user will be asked to save the data which are
currently in the sample storage.
File / Save
Saves the content of the sample storage under the actual document name. If no
document name has been defined so far ("Untitled"), a dialog box appears "Save as"
where the document name has to be selected and/or defined.
File / Save As
Saves the content of the sample storage under the actual document name. A dialog box
appears "Save as" where the document name has to be selected and/or defined.
File / Print
Prints the document. Which printout will be processed depends on which window is
currently activated. If the sample field is activated the sample data will be printed out.
The sequence of the data types is identical to those on the screen. If the combi window
is activated and the grahpics view is selected, the graphic analysis process will be
printed out. The settings for the screen (zooming factor, activated graphics, etc.) apply
for the printout as well. If the statistics view is selected, the data of the statistics
storage will be printed. If the calibration view is selected, the individual calibration
lines will be printed. If the status window is activated, this function cannot be invoked.
File / 1...4
The names of the four recently used documents will be displayed here (if available).
By clicking one of these document names the corresponding document can be quickly
opened.
File / Exit
Leaves the liquiTOC program. Can only be invoked if the system is in standby.
Edit / Cut
Deletes the highlighted data type and puts it to the clipboard. Can only be invoked if
something was highlighted and the text view is activated.
Edit / Copy
03/2003 Operating instructions liquiTOC
Copyright by E l e m e n t a r Analysensysteme GmbH
Page 7-8 liquiTOC Software Description
Copies the highlighted data type and puts it to the clipboard. Can only be invoked if
something was highlighted and the text view is activated.
Edit / Paste
Pastes the content of the clipboard to the highlighted position. Can only be invoked if
something was highlighted and the text view is activated.
Edit / Group
Combines the highlighted samples to a statistical group and loads them into the
statistics storage. Can only be invoked if at least two samples were highlighted and the
text view is activated.
Edit / Include/Exclude
With statistics view in the combi window activated
Includes and/or excludes a sample highlighted in the statistics field from the statistical
computation. Can only be invoked if a sample was highlighted in the statistics field
and at least three samples that are not excluded are within one group. A sample in the
statistics field can be highlighted by clicking with the mouse. Calling up the function
on an already excluded sample includes the sample again.
In general, only those samples should be excluded from the statistics which were not
analyzed properly (e.g. not enough sample in the vial).
With calibration view activated
Includes and/or excludes a sample selected from the calibration computation. Can only
be invoked if a sample was selected in the calibration field. A sample in the calibration
field can be selected by clicking with the mouse. Calling up the function on an already
excluded sample includes the sample again.
Edit / Edit
Allows the opening of an input dialog without using the mouse.
View / Toggle
Operating instructions liquiTOC 06/03
Copyright by E l e m e n t a r Analysensysteme GmbH
liquiTOC Software Description page 7-9
Changes between statistics, graphics and calibration view. Can only be invoked if the
combi window is activated.
View / Next
With graphics view in the combi window activated
Displays the graphic analysis run of the next sample – if existing. Can only be invoked
if the graphics view in the combi window is activated.
With calibration view activated
Changes to next calibration data.
View / Previous
With graphics view in the combi window activated
Displays the graphic analysis run of the previous sample – if existing. Can only be
invoked of the graphics view in the combi window is activated.
With calibration view activated
Changes to previous calibration data.
View / Zoom
Allows the selection of the zooming factor in steps 1, 2, 5, 10, 20, 50 and 100. Can
only be invoked if either the graphics or the calibration view in the combi window is
activated.
View / Configuration
Allows the configuration of the corresponding view. Depending on the activated view
different dialog boxes appear where the individual settings can be performed. These
are:
¾ Sample field: the individual date types can either be hidden completely
(column number = 0) or determined in the sequence of the display
(column number = 1: the furthest to the left).
¾ Graphics view in the combi window: from eight graphs up to four can be selected
which shall appear in the graphics field. For each graph the display area can be set
(Y-axis). For all graphs the time frame will be defined (X-axis in seconds).
Furthermore it can be selected on which grid line the zero line should be.
Consequently, it is possible to follow the direction of an IR channel if it is
negative.
¾ Statistics field in the combi window: the individual concentration can be hidden
(identification number = 0) and their sequence can be set (identification number =
1: is the furthest to the left). Whereas the mean value of the individual
concentration always appears, the other result lines can be shown/hidden as
03/2003 Operating instructions liquiTOC
Copyright by E l e m e n t a r Analysensysteme GmbH
Page 7-10 liquiTOC Software Description
requested.
Wizard / Calibration
This menu starts a wizard which supports the user in performing a calibration for an
analysis series. After completion, the program enters the corresponding sample data in
the sample storage. By invoking the wizard again a further calibration series – e.g. for
a different concentration range – can be amended to the previously produced one. The
user has to prepare the corresponding samples (autosampler) after the wizard has
entered the sample data and then has to invoke the start function. After completion of
the analysis run a calibration of the system can be performed by means of the menu
"Math. / Calibration".
After completion of the analysis run the test can be started by means of the menu
"Math. / System Suitability Test". Can only be invoked if the document is empty.
Assistant / Validation
Not yet implemented.
System / Run
Starts the continuous analysis. During a continuous analysis samples will be processed
until either the stop mark or the sample where the system should go into sleep mode is
reached. This depends on the settings in menu "Options / Settings / Sleep/Wakeup"
(see relevant menu item). Can only be invoked if an analyzer is conneted and if the
corresponding samples for a concentration range were entered.
System / Single
Starts a single analysis, i.e. after the analysis of one sample the system goes back to
standby. If this function is invoked during a continuous analysis the continuous
analysis will be stopped, i.e. the system goes back to standby after the analysis of the
currently processed sample is completed. Can only be invoked if an analyzer is
connected and if the corresponding samples for a concentration range were entered.
System / Wakeup
If the system is in sleep mode, this command wakes up the system again (see also
Options / Settings / Sleep/Wakeup).
System / Feeding
With this menu it will be defined how the feeding shall be performed: either without
or with autosampler. For the feeding mode "without" it is presumed that the suction
needle is tightly in a sample vial and/or in a sample bypass. For the feeding mode
"autosampler" the number of plate positions has to be specified in order to ensure that
the sampling will be performed from the correct sample vials. The option "Needle
plunged" can be selected if the sample material e.g. has to be stripped out. It can also
be set if a single, double or triple determination shall be run.
Depending on this setting the computation of the correct sampling position will be
done automatically for each sample.
Sytem / Mode
Allows the selection of the analysis mode. Can only be invoked if the document is
empty.
The bar with the largest length and its percentage is displayed in the status field. After
each analysis the count of the meter will be increased by 1. If a mainenance event is
due, a blinking warning triangle appears in the status field, combined with the
corresponding event text. If the warning triangle appears without event text, it occurs
during an analysis run. If there are multiple maintenance events due, only one event
appears in the status field. It will be the first event in the sequence.
The before mentioned settings will be made in a dialog box which opens.
¾ Sleep deactivated: when reaching the last sample (stop tag) the system goes to
standby mode. The reactor temperature and gas flow remain the same.
¾ Sleep at end of samples: when reaching the last sample (stop mark) the system
goes to the sleep mode, i.e. the furnace temperature will be lowered to the set
value (as selected and set in option and/or input field "Reduce temperature to" and
the gas will be switched off (as selected in option "Cut off carrier gas").
¾ Sleep at sample no.: when reaching the corresponding sample the system goes into
the sleep mode (see previous description).
By clicking the button "Sleep Now" the system can be put immediately into the sleep
mode.
If the system is at sleep, there are the following possibilities for a wakeup:
¾ Daily Wakeup at time: waking-up will be performed at a certain time and daily –
except for those week days which are clicked in the day bar below.
By clicking the button "Wakeup Now" the system can be manually waken up.
During the wakeup prodecure the reactor is heated up first, then the gas will be turned
on again. A certain time will pass until the instrument can be put into operation again.
The behavior after the wakeup depends on how the instrument "fell asleep": are there
still samples in the sample storage up to the stop mark which have not been analyzed,
the continuous analysis will be automatically performed up to the stop mark after the
instrument is ready for operation. However, there will be no other sleep procedure!
If there are no further to be processed samples up to the stop mark, the system waits
for an analysis start after it is ready for operation.
Math. / Coefficients
This menu allows the manual input of calibration coefficients. The coefficients can
only be entered and/or altered with an empty document.
Math. / Calibrate
If there are calibration samples in the sample storage the calibration field in the combi
window will be activated and for each existing concentration range as well as for each
element type (e.g. TIC, NPOC, etc.) the course of the calibration line as well as the
position of the samples will be displayed in a graphic one after the other. The user can
do a visual outlier elimination by clicking the suspected sample and subsequently
invoking the function "Edit / Include/Exclude". By invoking the function "View /
Next" the next calibration curve can be seen. After the last calibration graphic has
been displayed a further invocation of "View / Next" will cause the display of the
calibration coefficients dialog. By leaving the dialog with OK, the last calculated
calibration coeffients will taken over and are available for the next measurement.
Math. / Recalculate
This menu performs a complete recalculation of the sample data.
? / HelpTopics
Not yet implemented
? / About liquiTOC
Shows a dialog box where the program version and the version number of the analyzer
and the autosampler (if connected) are displayed.
The liquiTOC data files are filed in ASCII format and are therefore legible by any
common spreadsheet program. The following describes the procedure of loading the
liquiTOC data under MS-Excel.
1. Start MS-Excel.
2. Load the liquiTOC data file that you want to have processed. Invoke the
command „File/Open”. First select “All data types”(*.*) under “files of type”
in the appearing file open dialog. Now select in the field „Look in“ the folder
in which the data file to be loaded is and select it in the listing below. By
clicking „OK“ the file open dialog closes and a text wizard opens. Confirm the
displayed steps with “next” and the last step with „Finish“. The liquiTOC data
will now be displayed in an Excel worksheet.
Please make sure to save the data under a different name (e.g. with the
extension „xls“) since otherwise the original file will not be legible from the
liquiTOC program anymore!
8. Calibration
8.1 General
The calibration creates the relation between the detector signal (in relative units) and the
carbon and/or nitrogen contents in the sample. This relation is linear within the given
measuring ranges. The plot as measuring signal (in relative units) against the absolute content
C and/or N. Compared with the plot against the concentration it has the advantage that the
calibration can be used for different injection volumes.
The liquiTOC analyzer knows different calibration ranges to which different concentrations
for TIC and TOC are allocated. For each sample type an individual calibration can be stored,
as described in chapter 7.
If a peak has been recognized, the area of the peak will be saved if the peak maximum lies
below the saturation concentration of the detector. Otherwise, the peak will not be taken
under consideration for the calibration.
TIC and TOC can either be calibrated by means of a mixed standard or by means of a single
standard. For time reasons the calibration by means of mixed standard is recommended. It
does not have an influence on the analytical results.
For stability of the calibration one can say that due to the mass flow controller the carbon can
be run over a very long period of time without a new calibration. Even an exchange of the
reactor does not necessarily mean a new calibration. A calibration should be stable for at least
one month. For TNb a new calibration is necessary more often, i.e. at least once a week or if
necessary, every working day.
• The ordinate section will be automatically set to 0. For low conentrations this may lead to
strong falsifications in measuring results. Therefore, it is recommended to perform a blank
value determination for the methods drinking water, ultra pure water and sea water when
using a single point calibration.
• An outlier identification for a single point calibration is only be possible with a triple
determination. Therefore, a single point calibration shall not be performed with a double
or a single determination.
8.5 TNb
Basically TNb can be calibrated similar to TIC or NPOC. Since the TNb analysis can be
influenced more easily by outside influences than the TOC analysis which has to do with the
different thermodynamic stability of the measuring species (here CO2, there NO), some
specialties have to be considered.
The NO yield depends on the chemical bond of the nitrogen. Nitrate for instance has a
recovery rate of approx. 100% whereas ammonium of about 90%. When knowing the
approximate sample composition, this should be considered when selecting the calibration
standard. For unknown samples a mixed calibration of 50% ammonium-N and 50% nitrate-N
is recommended. Such a standard for 500 mg/l TNb has the have following composition:
It is possible to calibrate TNb and TIC/TOC from one and the same stock solution. In this
case, the above quantities of NH4Cl and NaNO3 shall be added to the TIC/TOC stock
solutions as described in chapter 8.1. The advantage of this procedure is an immense time
saving. But it shall be considered that due to high TOC contents (> 200 mg/l) the N recovery
may be impaired.
a + b Ik
c:
lV
IV represents the injection volume in l and Ik represents the peak ares corrected by the blank
value of the flushing volume. Furthermore, you will receive the information on the correlation
coefficient r. The value of r should be > 0.9900. If r < 0.990 a closer look should be given for
outliers.
Outliers can be highlighted with a left click of the mouse and eliminated from the calibration
under menu "Exclude".
In general one should be careful with the elimination of outliers. Calibration points should
only be eliminated in case of very clear outliers.
There is also a bar display in the main window which reflects the maintenance status. Prior to
large sample series the maintenance status should be checked.
The replacement of the consumables will be done under Options / Maintenance / Replace
part. The carrier gas and the furnace heatings will be switched off. After the overpressure has
been decreased in the instrument the corresponding ground-in clamps or screwings can be
opened.
After leaving the function Replace Part the carrier gas will be turned on, the furnaces will be
heated up. A new reactor should now be heated out with streaming carrier gas for approx. 30
minutes. Subsequently, some blank value measurements should be carried out (see chapter
5.1).
After the replacement of the reactor also the pressure and flow display at the PC should be
checked after approx. 10 minutes whether they correspond to the nominal values (see chapter
4.4.1, point h).
If the display FM is far bigger than 200 and the display MFC is far smaller than 200 or even
zero, a leak has been created during assembly of the ground-in clamps. It has to be removed
by orderly assembly.
The upper outlet of the reactor shall be closed with a device (included in the leakage set). If
the flow display of the flow meter goes back to 0, the leak is located behind the reactor. If the
flow meter shows an residual flow, the supply line to the reactor shall be closed with a blind
plug. If no gas flow can be seen, the reactor is leak (hair crack or splintering at the ground-in
ball-and-socket joint). Otherwise, check the connections to the buffer container first, then the
gas outlet of the flow meter. If all these connections remain without findings, the flow meter
is probably defective.
If the leak is located behind the reactor, first all connections of the desending cooling tube
with the spiral cooler shall be checked. Both outlet connections of the spiral cooler (to hoses
no. 40 and 50) shall be sealed. If the flow meter shows a flow, the screw connection of the
descending spiral cooler shall be fastened and if necessary the o-ring shall be replaced.
If the flow meter does not show any flow, first the waste hose no. 50 shall be reconnected
with the spiral cooler. If there is a flow in the flow meter again, the leak is located in the drain
area of the condensed water. Valve V9 or the blind screw connection located in the valve is
probably defective.
If there is no flow after connection of hose no. 50, hose no. 40 shall be reconnected with the
spiral cooler. Subsequently, the membrane gas drying (in the right side of the instrument)
shall be driven around by means of the double groun-in ball and socket joint (leakage set). If
the instrument is tight afterwards, the membrane gas drying shall be replaced. If there is still a
leak, the connections of the silver wool tubes and the U-tube filled with drying agent shall be
checked.
9.6 Blockings
Blockings mostly occur in the area of the reactor and the sample feeding. They will be
indicated by a heavily fluctuating flow in the flow meter during evaporation of the sample.
Blockings can also be detected by loosening the upper ground-in clamp at the reactor. The gas
flow through the flow meter shall be >400 ml/min. If the flow is lower
(<350 ml/min) a blocking can be assumed. In this case the lower ground-in clamp of the
reactor should be loosened first. If afterwards the flow is >400 ml/min, the reactor shall be
cleaned (see operating instructions) or, if necessary, replaced.
Otherwise the supply line to the reactor shall be checked and if necessary cleaned – or for
later instruments the valve at the lower end of the reactor. If a supply line is blocked, the
blockings usually can be found at the ends of the hose line. In this case, remove the fitting and
cut the hose lines approx. 2 mm from the end with a sharp knife (or with a scalpel, if
available).
Subsequently, a new fitting shall be mounted and the hose line can be reinstalled.
Solving PC problems
Occasionally, there are problems with the PCs which probably are in connection with the
Elementar instrument software. Possible problems may be occasional system crashes or a
freeze of the instrument software. The cause can be explained by the connection of the
settings in the operation system for the power management of the PC. The settings shall be
checked in the power management under Control Panel-> Power Option Properties and/or
Power Schemes. The options for "Turn off monitor", "Turn off hard disks", "System standby"
and "System hibernates" shall be set to "Never". If possible, also the power management in
the BIOS of the PC shall be deactivated.
General Note
For future generations of the Elementar instrument software the operation systems
Windows 2000 Professional or Windows XP Professional will be a prerequisite.
Windows 9x/ME and Windows NT will not be supported anymore. This has to be
taken under consideration when purchasing new PC hardware.
Installation requirements
Dear User,
we thank you for your order and the confidence expressed to our representative and
Elementar by this decision.
You have made you choice for a TOC-TIC analyzer of highest analytical performance and
reliability. In order to secure a fast and smooth installation of the instrument in your
laboratory, we kindly ask you to take the following hints into consideration:
Installation location
• The analyzer can be installed under normal laboratory conditions, i.e. no fume hoods or air
condition (ambient temperature between +15°C max. +30°C) are necessary.
• However, like for all analytical equipment direct sun beams, draft or other inconvenient
conditions for precision instruments should be avoided.
• For the installation of the unit (without PC, printer and balance) a stable laboratory table with
heat resistant surface is sufficient.
• Minimum dimensions: length 1.50 m * depth 0.80 m with autosampler length approx.
1,70m
• Weight liquiTOC approx. 50 kg.
Gas supply
For the operation of the analyzer the following gases are required which should be kept ready
by you on the spot for installation:
• Synthetic air (hydrocarbon <0.1 ppm and CO2 <1ppm),
1.5 bar, consumption approx. 500 ml/min
Unless delivered from us, the availability of two-step high precision pressure reducers (Air
Liquide or similar) is necessary. Pressure reducers of inferior quality will influence negatively
the analysis results!
! Pay attention to the bottle pressure !
Instrument control
If PC and balance were not delivered from Elementar, we reserve the right to perform the
installation with own instruments. Possible modifications to our analyzer have to be done by
the customer.
We would like to emphasize that the PC is the central control system of the analyzer. The
operating ability of the analyzer depends on a failure-free operation of the PC. Due to
experiences in the past with the use of control PC in mixed operation as office workplace
impairments due to software instabilities may occur.
Therefore, we recommend to only use programs on this PC for operation and data evaluation
of this analyzer.
Control PC
In case the PC was not delivered by us, the following minimum requirements have to be met:
For operation without modem:
Generally, it is not recommended to overload the PC where the instrument software is installed with
other software programs. This may lead to instable operation of the instrument software.
Balance
General remarks
The instrument will be installed by a factory trained service engineer and the analytical
performance will be demonstrated by max. 10 standard calibration samples.
In case of specific handling and application questions, you are kindly invited to contact our
representative for assistance. The elaboration of specific operation procedures can be ordered
there against separate charge.
The instrument is calibrated over a standard measuring range from works`side which is stable
for a long time. If specific applications are not within this range, the user can perform an
extended calibration by himself.
Our agent and Elementar will of course assist you at all arising questions.
Please do not hesitate to contact us under:
Service: +49/6181/91 00-16
Application Lab: +49/6181/91 00-47
Sales Department: +49/6181/91 00-12
1. Company Identification
of the company/undertaking
2. Component/information on ingredients
3. Hazards identification
Emergency Overview:
Irritating to eyes, respiratory systems and skin.
Contact with combustible material may cause fire.
Moisture sensitive.
Eyes: Flush eyes with plenty of water for at least 15 minutes, and have a medical consultation
immediately.
Skin: Remove contaminated clothes and shoes, and rinse skin with plenty of water immediately. Use a
soap to help assure removal. If irritation persists, have a medical consultation immediately.
Inhalation: Move victim to flesh air. If breathing is difficult, give oxygen. If breathing has stopped,
administer artifical respiration. Have a medical consultation immediately.
Ingestion: Wash out mouth with water to a conscious person. never give anything by mouth to an
unconscious person. Have a medical consultation immediately..
5. Fire-fighting measures
Page 1 of 4
Safety data sheet
Product name: Magnesium perchlorate
Wear self-contained breathing apparatus and protective clothing to prevent contact with eyes and skin.
Sweep up and place the spill into a chemical waste container. Avoid raising dust. Ventilate the area and wash
the spill site after material pickup is complete.
Handling: Avoid breathing dust, vapor, mist or gas. Avoid contact with eyes, skin and clothing.
Wear suitable personal protection. Refer to Section 8.
Storage: Do not store near combustible materials. Store in a cool and dry space. Keep a container tightly
closed.
Exposure Control: Use exhaust ventilation to keep airborne concentration below exposure limits.
Mechanical exhaust required.
Personal Protection: Safety glasses (goggles), Chemical resistant gloves, Full-body protective clothing,
MSHA/NIOSH approved respirator.
Exposure Guidelines: No information available.
Chemical Stability: Stable under normal temperatures and pressures. Reacts violently with water with
evolution of heat.
Condition to Avoid: Incompatible materials, combustible materials, exposure to moist air or water.
Incompatibilities: Strong reducing agents, strong acids, ammonia, combustible materials (e.g., organic
solvents, finely powdered metals, hydrazine, phosphorus.
No information available.
Page 2 of 4
Safety data sheet
Product name: Magnesium perchlorate
13 Disposal considerations
Observe all federal, state and local environmental laws and regulations.
IMO IATA
Shipping Name: Magnesium perchlorate Magnesium perchlorate
Hazard Class: Class 5.1 Class 5.1
UN Number: UN 1475 UN 1475
Packing Group: Group II Group II
European Information:
EINECS No. 233-108-3
Hazard Symbol: O (Oxidiing), Xi (Irritating)
Risk Phrases: R8 Contact with combustible material may cause fire.
R14 Reacts violently with water.
R36/37/38 Irritating to eyes, respiratory system and skin.
US Federal Information
CAS No. 10034-81-8 is listed on the TSCA Inventory.
The above information is believed to be correct to best of our knowledge but does not purport to be all
inclusive and shall be used only as a guide. KISHIDA CHEMICAL shall not be held liable for any damage
resulting from handling or from contact with the above product.
The information contained herein is based on the present state of our knowledge. It
characterizes the product with regard to the appropriate safety precautions. It does not
represent a guarantee of the properties of the product.
Page 3 of 4
Safety data sheet Art.-Nr./Part-No.: 22 131 365
Manufacturer/supplier identification
2. Composition/information on ingredients
CAS-No.: 7440-22-4
M: 107.87 g/mol EC-No.: 231-131-3
Molecular formula: Ag
Hill
3. Hazards identification
5. Fire-fighting measures
Handling:
No further requirements
Storage:
Tightly closed. Dry. Storage temperature: no restrictions
Page 1 of 4
Safety data sheet
Product name: Silver wool
German regulations:
MAK Germany (max. workplace conc.)
Name: Silver
Value 0.1 mg/m³ inhalable dust
Peak limit III resorptive substance, 10 x TLV
Embryotoxic cat. IIc classification not still possible in A-D categories
Industrial hygiene
Wash hands after working with substance.
Form: solid
Colour: silver grey
Odour: odourless
Solubility in
water (20 °C) insoluble
Page 2 of 4
Safety Data Sheet
Product name: Silver wool
Conditions to be avoided
no information available
Substances to be avoided
Acetylidene, azides, alcohols/nitric acid, ethylene oxide, peroxi compounds, hydrogen peroxide
organic substances, ammonium compounds, halogen-halogen compounds, fluorine, oxalic acid, ozone,
performic acid, zinc.
Acute toxicity
Quantitative data on the toxicity of this product are not available.
Further data
The product should be handled with the care ususal when dealing with chemicals.
Ecotoxic effects:
Quantitative data on the ecological effect of this product are not available.
13 Disposal considerations
Product:
There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues
generally count as special waste. The disposal of the latter is regulated in the EC member countries
through corresponding laws and regulations. We recommend that you contact either the authorities
in charge or approved waste disposal companies which will advise you on how to dispose of special waste.
Packaging:
Disposal in compliance with official regulations. Handle contaminated packaging in the same way
as the substance itself. If not officially specified differently, non-contaminated packaging may be
treated like household waste or recycled.
Page 3 of 4
Safety data sheet
Product name: Silver wool
Regional representation:
This information is given on the authorised Safety Data Sheet for your contry.
The information contained herein is based on the present state of our knowledge. It
characterizes the product with regard to the appropriate safety precautions. It does not
represent a guarantee of the properties of the product.
Page 4 of 4
Safety data sheet Art.-No.: 03 679 908
2. Composition/information on ingredients
3. Hazards identification
after inhalation: /
after contact with skin: wash with water
after contact with eyes: wash with water
after swallowing: /
5. Fire-fighting measures
Special risks: /
Other instructions: Non-combustible
Handling:
No further requirements
Storage:
Store tightly closed.
Page 1 of 3
Safety Data Sheet
Industrial hygiene:
Form: solid
colour: white-transparent
Odour: odourless
Ignition temperature : /
Flash poit : /
Solubility in
Water (20 °C) /
Conditions to be avoid
no information available
Substances to be avoided
with acetylidene, azides alcohols/nitric acid, ethylene oxide, peroxi compounds, organic substances.
Page 2 of 3
Safety Data Sheet
acute toxicity
Quantitative data regarding toxicity of this product are not available.
Ecotoxical effect:
no spezific information
13 Disposal considerations
R-phrases: /
S-phrases: /
none
The information contained herein is based on the present state of our knowledge. It
characterizes the product with regard to the appropriate safety precautions. It does not
represent a guarantee of the properties of the product.
Page 3 of 3