NOTE: A funnel is recommended for filling to avoid spilling the filling fluid on the outside of the case. Excess filling fluid may
cause the plug to become lubricated which may allow the plug to pop-off when the gauge is exposed to high temperatures.
CAUTION: Gauges filled with glycerine or silicone cannot be used on applications where strong oxidizing agents are present. A
chemical reaction, ignition or explosion may result. Oxidizing agents include, but are not limited to, oxygen, chlorine, nitric acid and
hydrogen peroxide. Completely fluorinated or chlorinated fluids such as HALOCARBON 6.3 CST should be used.
Liquid Fill Fluid Ambient Temperature Ratings
Allowable Operating Range liquid filling to expand and contract which in turn increases
Temperature range in which the operation of the gauge is not or decreases case pressure. As a result, accuracy can be
adversely affected by the filling liquid. At temperatures above decreased and the pointer may not return to zero properly until
the maximum rating, the fluid may break down. At temperatures the gauge is vented to the atmosphere.
below the minimum rating, the fluid may solidify (freeze).
To vent a WIKA gauge, move the valve to
Liquid Fill Gauge Case Venting the open position which will release any
For pressure gauges with full scale ranges of 300psi and below pressure or vacuum built up in the case. If
(including vacuum and compound ranges of 30” Hg-0-200 the gauge is installed in an upright position,
the lever can be left in the open position.
Vent Plug
psi and below), case venting (after the gauge is installed) is
necessary to preserve the accuracy. Temperature fluctuations The lever allows the use of a gauge in a non-upright orientation.
during shipment and in the process application cause the
NOTE: Some parts of the pressure gauge may not be able to withstand temperatures above 140ºF. Consult with the factory
for technical assistance for these applications.
Pressure Compensating Membranes gauge. Internal membranes serve the same purpose as the
Some applications may require installation of the gauge in a membrane window, but are usually affixed to the inside of the
position other than vertical or ambient conditions where venting case back.
is not possible.
Optional membrane windows are available for Type 213.40 2½”
One solution to this problem is the installation of a pressure & 4”, 233.30 2½” & 4”, 233.50 4”, and 2X3.53 2½” & 4” liquid
compensating membrane window or internal rubber membrane. filled gauges. Standard internal membranes are available for
The membrane window, sometimes referred to as a pressure Type 2X3.34 4½” process gauges and are optional for some
compensating window, allows the liquid filling to expand and 4” and 6” liquid filled gauges. Contact the factory for price and
contract without affecting the internal case pressure of the availability.