Abstract—The complex signal demodulation and the arctangent vital sign detection. This technique can reliably eliminate the
demodulation are studied for random body movement cancellation null detection point problem at a low cost. Moreover, since the
in quadrature Doppler radar noncontact vital sign detection. This objective of vital sign detection is to identify the frequency of
technique can be used in sleep apnea monitor, lie detector, and baby
monitor to eliminate the false alarm caused by random body move- the desired signal components, which are not related to the dc
ment. It is shown that if the dc offset of the baseband signal is ac- component, the complex signal demodulation does not need dc
curately calibrated, both demodulation techniques can be used for offset information when no random body movement is present.
random body movement cancellation. While the complex signal de- In addition to the experimental efforts to improve the system
modulation is less likely to be affected by a dc offset, the arctangent performance, theoretical analyses have been performed to study
demodulation has the advantage of eliminating harmonic and in-
termodulation interference at high carrier frequencies. When the
the Doppler noncontact vital sign detection and provide guide-
dc offset cannot be accurately calibrated, the complex signal de- lines for the designs. Achievements include the analysis on the
modulation is more favorable. Ray-tracing model is used to show range correlation effect and performance benefits [3], the
the effects of constellation deformation and optimum/null detec- modeling and analysis of the double-sideband transmission to
tion ambiguity caused by the phase offset due to finite antenna di- eliminate the null detection point [4], the spectral analysis of
rectivity. Experiments have been performed using 4–7 GHz radar
nonlinear phase modulation effect [5], the analysis of the arct-
to verify the theory.
angent demodulation in quadrature receivers [6], and the com-
Index Terms—Arctangent demodulation, complex signal demod- parative study on different radio architectures for vital sign de-
ulation, dc offset, noncontact, phase offset, random body move-
ment cancellation, ray tracing, vital sign, wireless.
tection [9].
However, the distinctive properties of noncontact vital sign
detection still require further attention in the following aspects.
I. INTRODUCTION First, noncontact vital sign sensing is based on detecting the
periodic phase change. Since the detection is typically carried
EMOTE noncontact detection of vital signs, i.e., the
R respiration and heartbeat, based on microwave Doppler
phase modulation effect has been studied for many years [1],
out within a few meters, the phase offset accumulated in the
transmission path varies significantly from one path to another,
making the detection potentially sensitive to the near field effect
[2]. To achieve an accurate and robust performance, researchers [11], [12].
in recent years have been working to further explore several Moreover, since the noncontact vital sign detection is based
technical challenges [3]–[10]. As one of the main challenges, on sensing the small physiological movement in millimeter or
the null detection point problem was solved by the frequency centimeter range, the presence of random body movement pro-
tuning technique [4] and the arctangent demodulation [6]. duces a significant source of noise for the accurate detection. Al-
However, the frequency tuning technique requires tuning the though successful Doppler radar noncontact vital sign detection
intermediate frequency once the distance between the an- under different environment for different applications has been
tenna and the subject changes; and the reported arctangent reported in the past two decades [1]–[6], a critical issue that pre-
demodulation requires calibration of the baseband dc offset [6]. vents this technology from being widely used is the noise pro-
Unfortunately the dc offset for noncontact vital sign detection duced by random body movement. For example, when the radar
is not only produced by the electronic components, but also was used for overnight monitoring of sleep apnea in clinical
by the unmodulated reflected signal, i.e., signal reflected from environment, the detected vital sign signal will be interrupted
stationary objects and other parts of the human body rather than when the subject under test is rolling over on the bed, producing
the moving chest wall. Therefore, the dc offset calibration is false alarm. It is recently reported in [8] that the random body
more than just calibrating the receiver chain. It is required once movement can be cancelled out by detecting from both the front
the experimental environment changes. Recently, a complex and the back of the human body using complex signal demod-
signal demodulation technique was proposed [8] for noncontact ulation. This cancellation technique can be widely adopted for
applications such as sleep apnea monitor, lie detector, and baby
monitor. In these applications, two identical portable radars can
Manuscript received April 24, 2008; revised July 09, 2008. First published
November 18, 2008; current version published December 05, 2008.
be conveniently implemented under and above the bed, or at a
The authors are with the Department of Electrical and Computer Engineering, distance from the front and the back of a seated subject.
University of Florida, Gainesville, FL 32611 USA (e-mail: changzhi@ufl.edu; In this paper, in addition to the reported complex signal
jenshan@ufl.edu).
Color versions of one or more of the figures in this paper are available online
demodulation used for random body movement cancellation
at http://ieeexplore.ieee.org. [8], another important demodulation method of noncontact
Digital Object Identifier 10.1109/TMTT.2008.2007139 vital sign detection, i.e., the arctangent demodulation, is also
0018-9480/$25.00 © 2008 IEEE
3144 IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, VOL. 56, NO. 12, DECEMBER 2008
Fig. 1. Simplified block diagram for Doppler radar noncontact vital sign
detection.
(2a)
(5)
(7a)
(7b)
Fig. 3. (a) Ray-tracing model and (b) the angular information of signals re-
flected from point A and B on the body using a 5.8 GHz radar. The antenna is
0 0 0
facing the body in the Z direction of the X Y Z coordinate.
where , , , , , and are the same
as defined in Section II-A. Instead of multiplying the two sig-
nals as in the case of using complex signal demodulation, the
random body movement can be cancelled out by adding the an-
gular information of (7a) and (7b) together to obtain
as follows:
(8)
As a result, the Doppler frequency drift caused by the random (9b)
movement of is eliminated, leaving only the desired fre-
quency components in the spectrum of . where is the intensity of the electromagnetic wave re-
The critical issue of using arctangent demodulation for flected from the location of the human body, is the
random body movement cancellation is how the presence of distance between the antenna and the subject under test,
the baseband dc offset will affect the detection. As will be is the distance between location and the antenna’s projec-
shown in simulation and experiments, the complex signal tion on the chest wall, and is the residual phase shift accu-
demodulation is more favorable for random body movement mulated in the electronic circuit. The integration is carried out
cancellation when the dc offset cannot be accurately determined over the entire subject under test.
and calibrated. Assume the antenna is placed 1 m away in front of the heart
center, and the locations of the heart center A and the body
center B on the front chest wall are separated by 11 cm. The
III. EFFECTS OF PHASE OFFSET difference in the transmission path for signals from the antenna
Since a real antenna with a certain radiation pattern does not to the two points is m, which
have infinite directivity, signals are reflected and captured from is replicated in the receiving path and would produce a phase
different parts of the body. When signals on different paths with difference of 83.5 degree for a 5.8 GHz radar. Meanwhile, the
different intensity and residual phases are received by the radar, radiation intensity of the antenna on the body surface is different
they are simply summed together by the receiving antenna, ei- from point to point, depending on the antenna radiation pattern.
ther canceling out or enhancing the desired signal components. This implies that the received baseband signals from the two
Therefore, a ray-tracing model [9] is developed to compensate points will have different locations and movement patterns in
for the shortage of the single-beam model. the constellation graph, as shown in Fig. 3(b). Therefore, the
As shown in Fig. 3(a), the actual received signal should be real case for vital sign detection using complex signal demodu-
represented from a ray-tracing point of view as lation and arctangent demodulation is complicated by the phase
offset. Numerical simulations are needed and will be presented
in the following section.
IV. SIMULATION
Simulations have been performed based on the ray-tracing
model. The two demodulation techniques were applied to
vital sign detection in the presence/absence of random body
(9a) movement.
LI AND LIN: RANDOM BODY MOVEMENT CANCELLATION IN DOPPLER RADAR VITAL SIGN DETECTION 3147
Fig. 4. Ray-tracing model. (a) Phase offset on the surface of human body radi-
ated by a 5.8 GHz radar. (b) 7 by 7 elements antenna array’s radiation intensity
on the human body. (c) Approximation of the normalized amplitude of body Fig. 5. Demodulation for a 5.8 GHz radar. (a) Signal detected at heart center
movement caused by respiration. (d) Approximation of the normalized ampli- (Case I) and at body center (Case II). (b) Actual received signal (Case III).
tude of body movement caused by heartbeat. (c) Angular information (t) of the received signal. (d) Baseband spectra ob-
tained by the complex signal demodulation and the arctangent demodulation
(the dc component is not shown in the baseband spectrum).
A. Ray-Tracing Model
The body model for a subject of 1.8 m height is shown in • Case II: A single-beam signal projected to the body center,
Fig. 4. Assuming the antenna is 1 m in front of the heart center, i.e., point B in Fig. 3. In this case, respiration signal was
the phase offset in different paths compared with the beam picked up while heartbeat signal is almost absent.
propagating to the center of the heart is shown in Fig. 4(a) • Case III: The actual signal transmitted and received by the
for a 5.8 GHz radar sensor. Dramatic change in phase offset radar.
is observed. Shown in Fig. 4(b) is the radiation intensity It should be noted that only Case III can be realized in the lab-
on the human body produced by an ideal 7 by 7 antenna oratory. Case I and II analyze signals carried by a hypothetical
array comprised of omnidirectional antennas spaced by . single beam radiated by an antenna with a very high directivity
Fig. 4(c) and (d) are the approximation of the normalized am- radiation pattern.
plitude of body movements caused by respiration and heartbeat, Example I: 5.8 GHz Quadrature Radar: Simulation results
respectively. It can be inferred that when a carrier frequency of are shown in Fig. 5 for detection from the back of the human
24 GHz is used for the higher sensitivity at shorter wavelengths, body. The residual phase produced in the electronic circuit was
the phase change will be more significant. assumed to be 0 , which means the channel was at the op-
timum detection point while the channel was at the null detec-
B. Demodulation Without Random Body Movement tion point according to the single-beam model.
Fig. 5(a) and (b) shows the signal trajectories in the constel-
To demonstrate the properties of the two demodulation lation graph. As predicted in Section II, signals reflected from
techniques, numerical simulations were first performed without different parts of the human body are affected by two variations:
random body movement present. Two examples are presented, the phase offset and the radiation intensity. The former varia-
i.e., a 5.8 GHz quadrature radar, and a 24 GHz quadrature tion embodies itself as different angles of the trajectory shown in
radar. The following three types of signals were recorded and Fig. 5(a), while the latter is demonstrated as different radii of the
analyzed. trajectory. As a result, when the receiver receives the vital sign
• Case I: A single-beam signal projected to the heart center, signals, which is the superposition of all the signals reflected
i.e., point A in Fig. 3. This is the case analyzed by the from different parts of the body, the total received signal trajec-
single-beam model. tory is deformed from an ideal circle, as shown in Fig. 5(b). It
3148 IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, VOL. 56, NO. 12, DECEMBER 2008
I Q
Fig. 7. Baseband spectrum detected by the and the channels with a carrier
frequency of 24 GHz. (a) Spectrum of a single-beam signal projected to the heart
center. (b) Spectrum of the actually received signal.
Fig. 6. Demodulation for a 24 GHz radar. (a) Signal detected at heart center the spectrum cleaner than that obtained by complex signal de-
(Case I) and at body center (Case II). (b) Actual received signal (Case III), with
the recovered angular information shown in inset. (c) Baseband spectra obtained
modulation.
by the complex signal demodulation and the arctangent demodulation (dc com- Another phenomenon to be noted is the optimum/null detec-
ponent not shown in the spectra). The complex signal demodulation causes har- tion ambiguity. Shown in Fig. 7(a) is the baseband spectrum of
H
monic ( ) and intermodulation (Int) interference.
the and the channels in Case I, i.e., the spectrum of the single
beam signal projected to the center of the heart. The peaks of the
should be noted that the constellation deformation is not caused respiration and the heartbeat components in channel have the
by noise, which was not included in simulation. same amplitudes as those in channel, which is in accordance
Although the foregoing discussions and simulation results ap- with (2) predicted by the single-beam model since the detec-
pear undesirable, the recovered angular information based on tion was performed at the mid-point between the null and the
(6) is nonetheless periodic and not seriously disturbed by the optimum detection points. However, the baseband spectrum of
phase offset problem, as shown in Fig. 5(c). The spectrum of the actual received signal, as shown in Fig. 7(b), shows that the
the complex signal and the recovered angular information were and the channels have significant differences in the heart-
analyzed and plotted in Fig. 5(d). Although the detection was beat signal strength. While the channel preserves the heartbeat
made with one channel at the null detection point and the other signal, the channel shows strong harmonic and intermodula-
at the optimum detection point, both of the two demodulation tion components. This is because of the enhancement and can-
techniques can successfully identify the respiration and heart- cellation among signals with different phase offsets. It demon-
beat components. strates the necessity of effectively combining the two channels
Therefore, the complex signal demodulation and the arctan- even when the detection is not carried out at the null detection
gent demodulation for 5.8 GHz radar system are demonstrated point.
to be effective solutions to achieve reliable detection and elimi-
nate the null detection point problem. C. Random Body Movement Cancellation
Example II: 24 GHz Quadrature Radar: In this example, the The random body movement cancellation technique was also
carrier frequency was 24 GHz and the residual phase produced simulated using the ray-tracing model with a carrier frequency
in the electronic circuit was assumed to be 45 , which means of 5.8 GHz. The random roaming of the body was fully mod-
the detection was performed at the middle between the null and eled in three dimensions ( , , and ) which are defined in
the optimum detection points. The constellation plots are shown Fig. 3. Typically the subject under test has larger random body
in Fig. 6(a) and (b). Due to fast variation of the phase offset movements in two dimensions than the third dimension, e.g.,
on the surface of the human body, more severe trajectory de- the horizontal movements in the and directions are more
formation was observed. However, angular information recov- obvious than the vertical movement in the direction for a
ered from (6) is still periodic, as shown in the inset of Fig. 6(b). seated person. Therefore, the time-variant velocity of random
The baseband spectra from the two demodulation techniques are body movement was modeled as uniform distribution between
shown in Fig. 6(c). Again, the respiration and the heartbeat com- 0 and a maximum value of 4 mm/s in the and the direc-
ponents can be identified from the spectrum by using both of the tions. The amplitude of random body movement in the direc-
techniques. tion was modeled as 0.1 of that in the other two directions. The
Furthermore, the result in Fig. 6(c) verifies that the arctangent movement components in each direction are shown in Fig. 8(a),
demodulation can eliminate the harmonics and intermodulation and the baseband spectra detected from the front and the back
terms caused by the nonlinear phase modulation effect, making using the two demodulation techniques are shown in Fig. 8(b).
LI AND LIN: RANDOM BODY MOVEMENT CANCELLATION IN DOPPLER RADAR VITAL SIGN DETECTION 3149
Fig. 8. Baseband spectra obtained when random body movement is present. Fig. 10. (a) Angular information and (b) baseband spectrum recovered from
(a) Random body movement is shown in the Z; X; and Y directions, which are the random body movement cancellation (RBMC) technique; the random body
defined in Fig. 3. (b) Baseband spectra by arctangent demodulation (AD) and movements are modeled in three dimensions, and the dc offset in each trans-
complex signal demodulation (CSD). ceiver is 30% of the maximum signal amplitude.
V. EXPERIMENT
Fig. 9. (a) Angular information and (b) baseband spectrum recovered by
random body movement cancellation (RBMC) using the two demodulation Experiments have been performed in the laboratory to verify
techniques; accurate dc information is used in demodulation but not shown in the theory and compare the performance of the two demodu-
the spectrum.
lation techniques for random body movement cancellation. For
consumer applications of this technique, it is desirable to have
When random body movement is present, the desired respira- portable radars that can detect vital signs from several meters
tion and heartbeat signal components will be overwhelmed by away, radiate a power of lower than 0 dBm, and have all the
the noise generated by random body movement. hardware integrated together at an affordable price. Therefore,
If the system can successfully calibrate out the dc offset up to 4–7 GHz portable radar was designed for this purpose. The
the baseband output, the recovered baseband angular informa- radar integrates the quadrature transceiver, the two-stage base-
tion and the spectra obtained by random body movement can- band amplifier, and the power management circuit on a single
cellation were simulated and shown in Fig. 9. The respiration printed circuit board (Rogers RO4350B substrate) with a size
and heartbeat components were successfully recovered by both of 6.8 7.5 cm . Fig. 11 shows the block diagram of the trans-
demodulation techniques, which showed similar performance ceiver. The specifications and manufacturers of the radio fre-
in recovering the desired signal components. It should be noted quency components are listed in Table I. For research purpose,
that although the random body movement can exist in the di- four voltage controlled oscillators (VCOs) were implemented
rection perpendicular to the radar direction, this technique still for a wide tuning range to obtain different optimal carrier fre-
works reliably because only the movement in the radar direction quencies under different environments [10]. When a specific
is critical for the detection. application is known, only one VCO is needed and the SP4T
If the dc offset cannot be perfectly calibrated out up to the switch can be eliminated to further reduce the cost.
baseband output, however, the performance of random body Since the vital sign has a frequency less than several hertz,
movement cancellation based on both of the demodulation tech- large coupling capacitors and of 10 F were used to iso-
3150 IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, VOL. 56, NO. 12, DECEMBER 2008
Fig. 12. Two identical transceivers used for random body movement cancella-
tion. Inset: the antenna used for each transceiver. Note that one transceiver uses
vertically polarized antenna array, while the other uses horizontally polarized
antenna array. From [8].
late the dc voltages of the mixer output and baseband amplifier A. DC Offset Estimation in Baseband
input. Because the 10 F coupling capacitors block the dc signal Due to the coupling capacitor in the radar between the re-
in addition to isolating dc voltages of two different circuits, a ceiver output and the baseband amplifier input and the vari-
variable dc offset was inevitably introduced into the measure- ability of dc offset within the experimental environment, it is
ment. The coupling capacitor F and the base- relatively difficult to accurately calibrate out the dc offset of the
band amplifier input resistor k were chosen whole system. Instead, the dc offset was estimated by fitting the
such that for a heartbeat signal with a frequency around 1 Hz, the signal trajectory into a proper segment of circle in the constel-
voltage drop on the capacitor is no more than 1/10 of the signal lation graph.
amplitude. This leads to a time constant of approximately 1.6 s, For example, Fig. 13(a) shows the constellation graph of the
which means that in the real-time signal processing software, a baseband signal detected from the back of the human body when
2 s initiation time is needed. no random body movement was present. Because of the coexis-
For random body movement cancellation, measurements tence of the undesired dc offset and the desired dc information,
were performed by two identical radars. As shown in Fig. 12, i.e., in (2), the original signal trajectory was located
patch antenna arrays with orthogonal polarization were in- at the center of the constellation graph. After subtracting an es-
stalled in the two transceivers to eliminate the interference timated dc offset level of 0.8 V for both the and channels
between the two units. It was observed in the experiment that in the baseband, the trajectory was fitted into a circle. Fig. 13(b)
the antenna gain should be higher than 4 dB for the radar to shows the baseband spectra obtained by the complex signal de-
have a good signal-to-noise ratio from up to 2 m away. The modulation and the arctangent demodulation. As shown in both
antenna was designed to have a maximum directivity gain of theory and experiment in [8], the dc offset does not affect com-
9 dB at broadside, so that the vital signs of the subject in front plex signal demodulation when random body movement is ab-
of the antenna will be picked up. Free-running VCOs were used sent. Therefore, the spectrum obtained by complex signal de-
for the two transmitters so that the actual carrier wavelengths modulation can be used as a reference to evaluate the reliability
were close to each other but always had a slight difference in the of arctangent demodulation using the estimated dc offset infor-
absence of a phase-locked loop. As a result, the signal from one mation. The spectra of Fig. 13(b) match well with each other,
LI AND LIN: RANDOM BODY MOVEMENT CANCELLATION IN DOPPLER RADAR VITAL SIGN DETECTION 3151
Fig. 14. Signals detected form: (a) the front and (b) the back of human body.
Fig. 13. DC offset estimation. (a) Trajectory of detected baseband signal with
no dc information and with estimated dc offset level subtracted. (b) Spectra ob-
tained by the two demodulation techniques. Signal with estimated dc offset sub-
tracted was used for arctangent demodulation.