Beruflich Dokumente
Kultur Dokumente
Maintenance
1. Check lid lock.
2. Inspect trunnion rings and buckets for metal fatigue.
3. Check hinges, control knobs, rubber feet.
4. Check the weight of the buckets and replace them in their holders. The rubber
cushions in the buckets may be lost or taken out for cleaning and misplaced, so
that uneven loads occur. Even small weight differences can cause rapid wear and
degradation of the m otor bearings.
Service
Calibration
Electrode equipment
Electrodes are used for measuring the electrical potential that develops when a
strip of metal is immersed in a dilute solution of a gas or salts. A transfer of gases
or ions between the metal and the solution occurs, establishing the electrical
potential (see also "Batteries" p. 13). The magnitude of the potential varies with
the concentration of ions in the solution. If two electrodes are dipped into a
solution, the difference between the electrical potentials of the two electrodes can
be measured (Fig. 2.14). If one of the electrodes has a stable potential (i.e., a
reference electrode), measurement of the potential difference can be used to
determine the gas or ion concentration in the solution.
v o ltm e te r
25
M A IN TE N A N CE AND REPAIR O F L A B O R A TO R Y AND H O S P ITA L EQ U IP M EN T
£ = r H,—RT
E — X1
In a
n. F
where:
E = measured voltage,
E0 = baseline potential (of a 1 mol/litre solution at room temperature),
R = gas constant (8.3144 J/(K •mol)),
T — temperature (Kelvin),
n = charge number of the ion to be measured,
F = Faraday constant (96 485 coulombs/mol),
a = activity of the ion or gas to be measured (indilute solutions a is close to
the gas or ion concentration).
Electrode characteristics
The characteristics of an electrode can be described by a diagram showing the
electrode voltage that is measured when it is immersed with a reference electrode
in salt solutions of known concentrations. W hen the electrical potential is plotted
against the logarithm of concentration, the resulting curve has a linear section
(Fig. 2.15), the slope of which can be calculated by the following equation:
= (£ i - £ 2)/1o S ci /c 2
where S = slope,
E = electrode potential in salt solution 1,
E2 = electrode potential in salt solution 2,
c1 = ion concentration in salt solution 1,
c2 = ion concentration in salt solution 2.
Fig. 2.15. Relation betw een voltage and ion concentration in electrode
m easurem ent.
WHO 94109
The slope is a measure of the sensitivity of an electrode and has the dimension of
voltage. The slopes of the curves obtained with different electrodes will vary
according to their ionic sensitivity.
In a system in which the ion concentration in salt solution 1 is 10 times higher than
that in salt solution 2, log { c j c 2) = log 10 = 1, and the equation for calculating
the slope becomes:
26
L A B O R A TO R Y EQ U IP M EN T
The slope of the electrode potential curve will change with the temperature of
measurement and with the charge number of the ion to be measured. The
temperature dependence of the slope for an electrode with a high sensitivity, used
for measurement of an ideal monovalent ion, will be near the values shown below:
pH meters
For the calibration of a pH meter, special buffer solutions must be used; the pH of
the buffer should be near the pH of the solution to be measured. Phosphate buffers
and acetate buffers are preferable. Problems may occur with alkaline buffers, since
their pH may decrease with the absorption of C O z from the air. This is why all
calibration buffers must be sealed during storage.
27
M AIN TEN A N CE AND REPAIR O F LA B O R A TO R Y AN D H O S P ITA L EQ U IP M EN T
Installation
Electrodes should be used as recommended for the specified purpose, and the
recommended procedure for equilibration should be followed when new electrodes
are installed. New glass electrodes must be soaked in a buffer solution (pH 4—8) for
at least 24 hours before use to obtain a stable potential. They should be calibrated
at two values, using the manufacturer's calibration materials or solutions prepared
from pH buffer tablets. If these are not readily available, buffer solutions should be
prepared for the purpose. However, these solutions should be checked from time
to time against the manufacturer's calibrants or solutions prepared from pH buffer
tablets.
G o o d practice
Glass electrodes will usually maintain their properties for many years if used
appropriately and stored correctly. Aging of an electrode is indicated when a
constant potential does not develop within a few seconds after insertion into an
ionic solution.
In modern blood-gas analysers, in which the electrodes are an integral part of the
instrument, some of these precautions will not be applicable.
28
LA B O R A TO R Y EQ UIPM EN T
R epairs
If electrodes are stored incorrectly the membrane may dry out. It is sometimes
possible to restore performance by removing the outer layer of the membrane with
a smooth flat file.
H a za rd s
If electrodes are used for measuring biological fluids, they must be cleaned and
disinfected according to the manufacturer's recommendations.
Packing
Electrodes are very fragile and must therefore be packed correctly, following the
manufacturer's instructions. Care is needed in removing the packaging.
Flame photometers
Flame photometers are used routinely for the measurement of lithium (Li), sodium
(Na), and potassium (K) in body fluids. M ore sophisticated instruments can also
measure calcium (Ca).
In flame photometry, an aqueous salt solution is dispersed in air. The salt in the
dispersed droplets is transferred into a gaseous state by heating with a flame, and
then quickly disintegrates into gaseous atoms. Above a critical temperature the
atoms absorb energy, which excites the electrons into higher energy states. W hen
the excited electrons return to their original state, they emit the absorbed energy
as light. The wavelength of the light emitted by each metal is characteristic for that
element. The intensity of the light emitted at the given wavelength is proportional
to the number of excited metal atoms and can be measured with a suitable optical
filter and photodetector.
This principle— also called flame emission photometry— can be used to measure
more than 50 elements. However, it is mainly used for determination of the alkali
metals, the excitation of which requires only a low energy input from a low-
temperature flame (propane/air or butane/air).
— nebulizer
— burner
29