Sie sind auf Seite 1von 14

Philip JH. Intravenous Access and Delivery Principles.

In: Rogers MC, T inker JH, Cov ino BG, Longnecker DE. Principles
and Practice of Anesthesiology . St. Louis. Mosby Year Book, Vol 2, 1992; 1183-1196. © Copyright 1992, James H. Philip,
all rights reserved . For permission to reprint, please write to the author at: jim@lifenam e.com or jphilip@partners.org

CHAPTER 55

-----i:~Jo---­
Intravenous Access and
Delivery Principles

JAMES H. PHILIP

Every year in the United States, at least 20 million


History patients receive IV fluids. 13 •43 Although clinicians
Simple IV System generally regard IV fluid administration as safe and
Pressure-How Relationships and Modern Theory effective, IV site complications occur frequently,44.45
Pressure-Flow Relationship Data These include phlebitis, 1~ thrombosis, catheter ob-
Pathologic Situations struction, and fluid extravasation. l4 •46•48 From the
Fluid infusion into tissues anesthesiologist's standpoint, the obvious result of
Obstructions to fluid flow: Starling resistors these complications is inability to administer the
System Nonlinearity intended drugs and fluids to the patient.
Clinical Fluid Infusion IV systems used in anesthesia range from the most
Gravity flow simple to the most complex. Drugs can. be adminis-
High-flow resuscitation tered from a syringe with needle into the flowing IV
Mechanical Fluid Infusion Systems stream that is powered by gravity. At the other
IV controllers extreme, pumps or controllers can create the contin-
IV pumps uous infusion, and a syringe or other pumps can
Other fluids administer drugs. These drugs can be controlled
Drug Administration manually by the clinician or by a computer, which
Delivery devices determines the patient's drug needs based on a
Summary pharmacokinetic model with or without the assistance
from the patient's physiologic variables.
This chapter summarizes alternative drug and fluid
administration techniques and devices in the frame-
IntravenoUs (IV) administration of fluids and drugs work of a simple physical, physiologic model so that
is an important part of the routine care of patients, one can understand fluid flow in IV systems and
especially those undergoing anesthesia and surgery. patient veins.
The IV infusion is essential as a route of drug
administration. In addition to facilitating induction, L

IV drugs are administered throughout the surgical HISTORY


procedure both as part of the anesthetic technique Until the 1970s, IV infusion was performed with
and as a way to meet surgical needs. Whenever fluid simple systems. 37 An elevated, fluid-filled bottle pro-
or blood is lost during surgery, replacement must be vided the energy while the clinician (nurse or physi-
considered, sometimes with great immediacy and cian) adjusted a roller clamp until the correct rate of
sometimes electively. Also, the patient's insensible drop formation was observed in the plastic drip
losses over and above those from surgery must be chamber. As clinicians questioned the limitations of
replaced. their tools, they discovered certain inadequacies in

1183
1184 PART II AneSlhetic Care

the conventional approach. In 1966 La Cour2 and that "negative resistance" can explain the results of
Ferechak et aI.' in 1971 showed that the rate of drop collapsed-tube studies. Further analysis by Shapiro"~
formation is not an accurate measure of flow rate developed the one-dimensional theory of steady flow
because of the influence of temperature, fluid com- in thin-walled tubes partly collapsed by negative
position, and diameter and shape of the drip chamber transmural pressure. Guyton 12 bas written extensively
orifice. Flack and Whyte8 showed that cold flow on both vein and tissue pressures. The compliance of
(creep) in the tubing underlying the roller clamp collapsible vessels has also been studied. 2O
contributes to flow variation, and Ziser et al.4g found The models just described explain the behavior of
this change to be greater than 15% over 45 minutes. tbin-walled rubber tubes subjected to external pres-
Venous physiology has been studied much more sure that exceeds in~ernal pressure. The incomplete
extensively than the technology of fluid infusion. collapse of rubber tubes makes the analysis complex.
Special focus has been placed on vein collapse. A Ironically, in vivo and in vitro veins probably do not
review of that literature explains our current knowl- behave in this complex way; they collapse com-
edge. In 1912 Starling (with Knowlton 151) used a pletely,·18,2S (unpublished data, D. Joseph and the
collapsible thin-walled rubber tube to produce con- author). Inapplicable analysis has confused our un·
stant back pressure (afterload) in his studies of beart derstanding of the behavior of clinical IV systems.
failure. Although he claimed control of resistance
with his device, he did produce constant pressure
SIMPLE IV SYSTEM
afterload with what we now call a Starling resistor.
Many years later, Holt1S investigated venous col- The simplest IV system consists of an elevated
lapse and the resulting decreased venous flow, attrib- fluid-filled bag or bottle, a length of flexible tubing, a
uting it to changing resistance. In 1954 Duomarco and catheter cannulating the patient's vein, and the
Rimini6 demonstrated the importance of energy and patient's venous system, which ultimately terminates
hydraulic gra~ients along veins. Rodbard and Saiki40 at the heart, .specifically the right atrium. An adjust·
found that "flow through widely patent elastic tubes able roller clamp compresses the tubing at one point,
followed the laws for flow through rigid pipes." They slowing fluid flow and allowing manual control by a
also noted that w~en air compressed a thin-walled clinician. The rate at which drops form is assessed
venous model externally, wall flutter was seen and visually, and the roller clamp is adjusted manually as
heard. Flutter ceased when external compressing needed. Depending on the type and brand of the
pressure was raised. tubing system, drop size is 1/10, 1/15 ml (macrodrop),
Rodbard 39 showe$l that critically high flow veloci- or 1/60 ml (microdrop). ..;
ties occur when veins collapse and explained the The flow rate through tbe IV system depends on
paradoxic increase in flow from vessel distension several factors. The height of the IV bag less the
produced by adding downstream resistance. He pressure in the patient's vein determines tlie driving
showed later that constant flow in various body tissues pressure for ftow. The resistance of the catheter plus
can be produced by the same collapse phenomenon.38 tubing system then determines flow:
In 1963 Permutt and RileY' explained that under
F::: b.P/R
conditions of tube collapse, flow is independent of the
pressure drop from inlet to outlet. Rather, flow where F is flow, tiP is pressure difference, and R is
"depends on the difference between inlet pressure resistance. In this analysis, F and aP are generally
and critical closing pressure" (i.e~, occlusion pressure measured and R is computed.
or external pressure, PCJCt)' They used the term The pressure difference depends on both fluid
vascular waterfall to describe this phenomenon, ap- height in the bag or bottle and pressure in the
plied by Starling in 1912. patient's vein. The latter is more complex than is
In 1969 Holt l6 and Conrads analyzed tbe physics of usually realized.
the collapse process observed in rubber tubes. Con- Veins possess two characteristics that slow fluid
rad's photos of a partly collapsed tube have been flow: resistance to fluid flow and opening pressure.
reprinted many times in texts and treatises that Resistance is caused by the long, narrow nature of
summarize the Iiterature. 3•10,23 Later, Kresch and veins, which causes pressure loss along their length.
Noordergraaf l performed additional analyses of col- Venous blood flow is usually assumed to be laminar,
lapsible rubber tubes, whereas Katz et a1. 18 found that a subject discussed later in this chapter. Opening
"the significant variable is transnlural pressure." pressure for the vein depends on the forces exerted by
Griffithsll noted that when coJlapse occurs, fluid flows tissues outside the vein along its course.
at sonic velocity (i.e., the velocity of pressure waves in Before fluid is forced through the vein, venous
the tube). Brower and Noordergraaf then showed pressure is essentially indeterminate. That is, the
. Intra~enous Access and Delivery Principles CHAPTER SS 1185

pressure measured may take on any value less than 55-2. Pressures at various locations, depicted as the
the obstructing or external pressure. Once fluid flows heights of the vertical bars above the reference level,
through the vein, pressure rises to the opening are shown under the condition that flow is 200
pressure. Then venous resistance affects fluid flow, ml/hour.
and pressure rises as flow increases. These concepts The model's behavior can be viewed as follows.
and their application are developed further through- Pressure at the free-air surface at the top of the bag
out this chapter. or bottle is zero (gauge), that is, atmospheric pres-
sure. Lower in the tubing, the pressure head proximal
to the roller clamp is determined by the height of the
PRESSURE-FLOW I.mLATIONSHIPS AND free-air surface above the roller clamp, combined with
MODERN THEORY fluid density and gravity. Specifically, pressure equcils
To understand fluid flow in IV systems, one can use height times fluid density times the acceleration of
a conceptual model. The model describes the fluid gravity. When the liquid is water and pressure is
flow behavior of any IV system and patient. Actual expressed in centimeters of water (em H20), pressure
components of the system are depicted as combina- is equal to the water height in centimeters. The
tions of one or more ideal components. The behavior density of most clinical solutions differs little from one
of actual" components can be predicted and measured (1.00).
using the model along with simple measurements. Because an air-filled drip chamber intervenes
The system is comprised of a selection of compo- between bag and roller clamp, the head height is
nents from Table 55-1. Fig. 55-1 shows the theoretic diminished by the height of the air column. The
pressure-flow relationship (PFR) for each compo- diminution of pressure occurs because within the
nent. The simplest configuration of actual compo- air-filled chamber, pressure is the same everywhere.
nen ts, a gravity flow administration system with rate No vertical pressure gradient exists in the air-filled
manually controlled by a roller clamp, is shown in Fig. chamber. The ~bing is drawn vertically from the

Actual componeDt Ideal compoaeaC Symbol

Bag Pressure source Pas


DS roller clamp Variable resistor ~
Tubing Low-resistance conduit R.ubo
Catheter Resistor Radl
Nonnal peripheral vein Resistor Rmn
Extravascular tissue Resistor RlIIuM
Central vein Conduit with no resistance Rev = 0
Heart Pressure source P(V :: central venous pressure

Obstruction Starling resistor = Collapsible tube POll

Inflated blood pressure cuff Collapsible tube (CT)


Venous tourniquet cr
Collapsible tube cr
Problem IV site cr
Catheter against vein wall cr
Volumetric pump Flow source Fpump
Volumetric controller
Variable resistor Variable resistor
Variable duty cycle Variable pressure source
Observation of drop rate Flow measure
Electronic pressure transducer Pressure sensor
Water manometer Pressure sensor

•Actual components behave similarly to theoretic components.


1186 PART II Anesthetic Care

Pressure sources Flow sources


P p
JVbag
Pbag
a.
!:;) e e
~
:;) a.
lit

1-= Centrol vein


!
a.. 1
~
U
Pwp

Flow flow pump

Resistors Obstructions
P RoUer damp p

! e
1
:;)

l=
Put

flow flow

Fig. 55-I. Theoretic pressure-flow relationship (PFR) for various system components, in-
cluding pressure sources, flow sources, resistors, and obstructions (i.e., collapsible tubes,
Starling resistors). (From Philip JH: A model for the physics and physiology of fluid
administration, J Clin Monit.S:l23, 1989.)

bottle outlet and then horizontal until entering the PRESSURE-FLOW RELATIONSmp DATA
patient's vein, which is at the same height as the heart The PFR for the tubing-catheter-patient system can
(right atrium). This simplifies the physics to make the be measured in experimental situations and in clinical
system more understandable. practice. This can be done with a volumetric pump
In Fig. 55-2 the roller clamp bas been adjusted to and a pressure-moniioring device such as a pressure
a resistance that provides a flow rate of 200 ml/hour. transducer or even ~ water column. One particular
In clinical practice the cliniCian adjusting the clamp is commercial IV infusion pump can facilitate this. The
unaware of actual resistance. Helshe simply rotates IVAC variable-pressure volumetric infusion pump,
the roller clamp until proper flow rate is judged. by Model 560 (IVAC Corp., San Diego), produces
measuring the rate of drop formation observed in the constant flow and measures pressure during infusion
drip chamber. Here, pressure just distal to the roller and when the pump is stopped. In this pump, flow is
clamp is measured to be 16 em ~O with this created by a linear peristaltic mechanism that pro-
particular combination of .system components and vides a flow pattern composed of successive 1 p.l
patient. From the values for pressure difference volumes. Just distal to the pumping mechanism, a
across the clamp and the value for flow, roller clamp pressure transducer presses against an in-line
resistance can be computed. pressure-sensing disk contained in the IV tubing. The
. Next, the origin of the pressure distal to the roller pressure-monitoring pump measures pressure with an
clamp }Dust be considered. From the standpoint of the accuracy of ± 2 mm Hg. 17;33.36 Pressure is measured in
tubing system, the pressure at the catheter tip both the presence and the absence of flow.
originates in the patient. The peripberal venous Using the pump, the PFR has been measured for
pressure measured in this situation can best be many tubing syste~, catheters, other devices, and
understood by realizing that the resistance of the vein patients. Because of flow limitations in the pump
cannot be zero. Thus the flow of blood back to the (1000 mlJhour = 1 Uhour), effects of high ftowon
heart makes peripheral venous pressure slightly system nonlinearity cannot be studied with this
higher than central venous pressure (cVP), in the device. High flow effects are presented in the section
absence of fluid infusion and intervening obstruction. on nonlinearity. Resistance units (RU) are mm
The pressure increment .often is quite small (3 cm HglI.Jhour.
~O or less).1.:51.41 With partial venous, obstruction From experiments using the pump, tile following
(see later discussion) venous pressure ~ay be signif- facts have beeD detennined.30 An elevated IV bag
icantly greater. behaves as a pressure source, with pressure equal to
Intravenous A~ss and Delivery Principles CHAPTER 55 1187

100
Bag

A 50 =2cm
IFl 200 mVhr
~
Tubing
a
I Catheter IV
V n;iiGrI ' - . /
~

22
------.. 16

Fig. 55·2. Model for manually controlled IV administration system and patient. A, Picto-
rial schematic diagram of model's components. B, Pressures (em ~O) observed in various
locations. C, Flow chart showing pressures (circles), resistances (rectangles),collapsible
tubes (trltmgles). and pressure drops (anows). In A, elevated fluid-filled bag acts as pres-
sure source (P = 100 em ~O in B) equal to height of its free-air surface above reference
level. RoUer clamp is r~istort arbitrarily adjusted for F (Bow) =200 mJlhour in A.
Catheter provides additive resistance in Jluid path to patient. Vein acts as hydraulic resis-
tor as well but can collapse if _mal pressure (8 cm ~O in B) exceeds internal pres-
sure (16 em ~O in B). CoUapSe does not occur here because internal pressure exceeds
external pressure. rmaUy, heart acts as pressure source (sink) for venous return. (From
Philip JH: A model for the physics and physiology of fluid administration, J Clin Monil
5:123. 1989.)

OTUBfNG C #16,2- II #la, 2-


<> #20, 1.25· + #22,1- )( #24,0.6-
20
18
_ 16
-l 14
E 12
.! 10
e
~ 8

1 6
4
2
0
0 50 100 150 200 250 300
Flow (ml/hr)

Fig. 55.3. PFR for five catheter sizes (16 x 2 inches, 18 x 2 inches. 20 x 1.25 inches. 20
x 1 inch, 24 x 1 inch). Pressures were measured at flows of 0, 100, 200, and 300 ml/hour,
and resistance was computed as slope of least-squares regression line. Resistances are
tabulated in Table 55-2. (From Philip JH: (\ model for the physics and physiology of fluid
administration, J elin Monit 5:123•. 1989.) : .
1188 PART II Anesthetic Care

the height of the free-air surface of the fluid above the The PFR for systems composed of an IV catheter
reference level, less the height of the inte"ening air (Fenwall QuickCath) plus a tubing system (Travenol
gap in the drip chamber. 2COO01) is plotted in Fig. 55-3. Measured total
IV tubing behaves as a low-resistance conduit resistances (± SEM) for no. 16, no. 18, no. 20, no. 22,
relative to other system components: and no. 24 catheters are 6 ± 1, 10 ± 0, 17 ± 2,
34 ± 1, and 66 ± 3 RU, respectively. When mea-
R.bba c 3 :f: 2 (SEM) RU
sured tubing resistance (3 ± 2 RU, previous equa-
A roller clamp behaves as a variable resistor. Once tion) was subtracted from each, respective mean
set, a roller clamp behaves the same (± 5%) when catheter resistances were 3, 7, 14, 31, and 63 RU
connected to a flow source (IV pump) as with a (Table 55-2).
pressure source (elevated bag). Typical roller clamp A normal patient vein behaves as a resistor, with
resistance during normal IV flow is 300 to 800 RU, resistance slmUar to that of a 20-gauge IV catheter.
depending on flow rate desired. The distribution of patient resistance for the veins of
An IV catheter behaves as a resistor that offers 46 surgical patients is shown in Fig. 55-4. The
more resistance than IV tubing. In the flow range of population statistics for normal vein resistance were
o to 300 mJ/hour, the PFR Is Hoear and resistance Rvein = 22 ± 20 (SD) RU (range, 0 to 91 RU;
takes on a unique value for each catheter size. median, 22 RU)
In addition to resistance, veins showed a second
parameter that limited flow. In some patients, this
was caused by ~xternal compression between the IV
site and the heart (see later section on obstruc-
tions). In other patients, this pressure was the
Ca&heter plUl Catheter alone CVP itself. For the 46 veins studied, opening pres-
Gaup =
tubJng (R SEM) (R:t &EM) sure was Po = 15 ± 8 (SD) mm Hg (range, 2 to 3S
mm Hg).
16 6 1 3 2
18 10 0 7 2 The central venous clreulation offers essentially DO
20 17 2 14 3 resistance to fluid flow. All resistance encountered
22 34 1 31 2 distal to the catheter tip is attributed to the peripheral
24 66 3 63 4 venous system. However, the central venous system
None 3 2 0 2 does provide a back p~ssure equal to CVP. This
pressure must be coD~idered when assessing fluid flow
·Resisrance (R) in resistance units (RU) • 1 DUD HgI1Jhour. through the system.

o 200 400 600 800 >1000


Resistance (mm Hg/L/hr)
Fig. 55-4. Distribution of hydraulic resistance measured in 46 veins and 12 infiltrations.
Veins, open bars; infiltrations, closed bars. (From Philip JH: A model for,the physics and
physiology of fluid administ{ation, J elin Monit 5:123, 1989.)
In~ravenous Access and Delivery Principles CHAPTER 55 1189

to 45 mm Hg and not affecting R significantly. Fig.


PAmOLOGIC SITUATIONS
55..6 shows the PFR for a typical tourniquet-
Fluid Infusion into Tissues obstructed vein.
The extravascular tissue near a vein behaves as a In vitro studies have shown that an elevated
resistor that otTers significantly more resistance collapsible tube behaves as a Starling resistor, with Po
than the vein. In the limited populations reported approximately equal to the. elevation of the collapsed
in the· literature, tissue resistance is significantly segment. Also, an externally compressed collapsible
greater than vein resistance. In addition, pressures tube behaves as a Starling resistor, with Po equal to
measured at low flow are generally not elevated external pressure.
greatly above zero..flow pressures. Little information or data are available concerning
For tissues, resistance (Ruuuo), estimated Po, and IV sites that function erratically. It is known that a
initial pressure after infiltration (Pinl.) were R 1issuc problem IV site can bebave as a Starling resistor, with
= 1125 ± 137' (SD) RU, Po = 44 ± 61 mm Hg, and high obstructing pressure. In the one case reported,30
Pinit = 8 ± 8 mm Hg. Fig. 55-4 shows the distribution an IV catheter was in place but was not running well
of tissue resistance superimposed on the distribution ("problem IV"). Altbough the IV bag was 80 em
of vein resistance just described. Rwrn and ~ were above the patient's heart and IV site, no fluid flow was
different significantly when tested with Student's t test detectable. A small fluid bolus injected via syringe
(p < 0.001) and confirmed with Mann-Whitney U into the IV tubing flowed easily, but spontaneous
test (Z = - 5.31). Pinl• for tissues was not significantly gravity-powered flow did not follow, even after
different from Po.vcin' Volunteers remarked that infil- catheter manipulation. When tbe IV bag was elevated
tration at a low infusion rate (and thus low pressure) 30 cm (12 inches), flow was observed to begin. After
was not painful. the IV site was determined to be functioning accept-
ably with~ bag elevated, the PFR was measured. Fig.
Obstructions to Fluid Flow: Starling Resistors 55-7 shows that resistance was 129 RU and Po was 80
Many clinical situations produce vein collapse, caus- mm Hg equals 109 em H 20. Analyzing the clinical
ing the vein to behave as a· Starling resistor. An situation and experimental data suggests that opening
inflated blood pressure (BP) cuff behaves as a Starling press"re for the Starling resistor was higher than the
resistor, with Po close to the external pressure applied initial bag height, resulting in no flow initially.
and little, if any, change in vein resistance. The PFR Elevating the bag slightly raised infusion pressure and
for a BP cuff on a typical patient is shown in Fig. 55-5. resulted in significant flow.
A venous tourniquet applied proximal to the IV site A catheter manipulated against a vein wall also can
likewise behaves as a Starling resistor, raising Po to 25 behave as a Starling resistor. In the one case reported,

OPext ... O o Pext a 10 A Pext .. 50


<> Pext .. 75 + Pext .. 100
100 -
90 ~
80
-
Q
:t: 70
-- A
V

E 60
--
..§.
50
-- .. -
.A

!
~

e~
40
30
-
-
-
Q.
20
-
- ..n
- P"'I
~

-
~

10
0
- I I I I I I I I I I • I I

0 50 100 150 200 250 300


Flow (ml/hrJ
Fig. 55.5. PFR for typical patient vein obstructed by an external blood pressure cuff. Note
that applying pressurized cuff raises opening pressure (Po) to approximately cuff pres~~re.
Resistance is relatively unaffected. Pat, external pressure = cuff pressure. (From PhIlip
JH: A model for the physics and physiology of fluid administration. I Clin Monit 5:123,
1989.) .
1190 PART II Anesthetic Care

o No tourniquet
o With tourniquet
100
90
80
-a
:c 70
E 60
..5. 50
e
;:, 40

1 30
20
10
0
0 50 100 150 200 250 300
Flow (ml/hr)
Fig. 55-6. PFR for typical patient vein obstructed by clinician-applied venous tourniquet.
Note that applying tourniquet raises Po from 12 to 32 mm Hg. Resistance is relatively
unaffected. (From Philip JH: A model for the physics and physiology of fluid .administra-
tion, J elin Monit 5:123, 1989.)

260 j L
2<40
220
0; 200
:c 180
E 160
Sl<40
! 120

J 100
80
60
<40
20
01 ~
o 100 200 300 <400 500 600 700 800 900 1000
Flow (ml/hr)
Fig. 55·7. PFR for "problem IV" site In patient. (From Philip JH: A model for the phys-
ics and physiology of fluid administration, J Clin Mon;t 5:123, 1989.)

"normal vein" resistance was RwID = 30 ± 9 (SO) In 1983 Philip and Philip27 showed that the PFR for
RU and Po = 7 ± 1 nun Hg. When a "positional IV" IV tubing systems and other fluid conduits is distinctly
was created by manipulating the catheter against the nonlinear and can be represented as:
vein wall, Rvcln aJ~nlt waD = 27 ± 0 (SO) RU and
Po = 36 ± 0 mm Hg. P c (RL X F) + (RT X F~

Fig. 55-8 shows this. In the same study, by sequentially


SYSTEM NONUNEARITY removing tubing sections, the authors showed that the
When IV ftow rate is low (less than 3000 ml/hour for RL and RT (multipliers of flow and flow squared)
most catheters and systems), the PFR Is linear. parameters are intrinsic properties of the tubing and
However, when flow is increased significantly, tbe do not result from flow perturbations at the entry or
PFR loses its Unear shape. NonUnearlty greatly exit of the tubing system. They also showed that:
affects flow prediction during rapid fluid infusion, as
is required during volume 'resuscitation. P=AxF
_,ntravenous Access and Delivery Principles CHAPTER 55 1191

400

"iii 300

,
:z:
E
E
-e 200

100

o
o 2 4 6 8 10 12
Flow (ml!sec)
Fig. 55-8. PPR for IV infusion system. Fitted line is best least-squares estimated parabola
to fit relationship P = (RL X F) + (~ x F2) (see text). (From Philip BK. Philip JH:
Characterization of [nonlinear) flow in intravenous infusion systems, IEEE Trans Biomed
Eng 30:702. 1983.)

does not fit the experimental data without systematic estingly, removal of the stopcock was not required
bias. Whether the obsetvations represent concomi- until the infusion system was composed of only
tant laminar and turbulent flow in the same tubing wide-bore tubing and a 12-gauge catheter.
section has not been established.
Finally, Philip and Philip27 showed that stopcocks CLINICAL FLUID INFUSION
and check valves are almost exclusively p2 devices,
whereas 5 fJ. filters (MP-S, Travenol) are ~clusively Gravity Flow
linear (F only) devices. Later, the same authors During routine anesthesia and surgery, fluid require-
showed that an Fl term is also required in analyzing ments in adults range from approximately 100 m1Ihour
the PFR for IV catheters.28 to 100 mlIminute and more. Before anesthesia, an IV
To understand the impact of interposing or remov- catheter is inserted in a vein and fastened securelY,
ing devices to enhance fluid Bow, the' system's usually with adhesive tape. The size of the IV catheter
nonlinearity cannot be ignored. The nonlinear PFR is chosen to allow for the patient's expected fluid
characteristics can be used to predict Bow capability requirement. However, one should always recognize
In IV Infusion systems and assist clIDlcal declslons on that rapid blood loss can occur during almost any
devices to be included or ellmlnated.29 surgical procedure and that the anesthesiologist must
System nonlinearity can be ignored when flows are be prepared for such an occurrence.
low enough. In 1988 Philip and Philip3s tried to When selecting catheter size, one must consider
identify situations in which nonlinearity was not known information about the resistance of catheters,
significant. To do so, they'defined Fa as the flow at IV systems, and patient veins. The resistance of each
which pressure drop was twice the linear pressure of these flow impediments adds to the others to fonn
drop. At Bows less than 0.1 FJ!t the system could be the total resistance to flow. Thus total resistance to
considered linear, and flow could easily be predicted fluid flow (Reolal) is the sum of the resistance of IV
from pressure by linear extrapolation. At flows above tubing, catheter, intetvening devices, and the patient's
1110 FE' the effects of nonlinearity must be consid- vein.
ered. Later, the order of device removal or replace- The 22 RU median value for vein resistance and
ment was assessed using increased flow capability as Table 55-2 (catheter resistances) together offer in-
a criterion.19 The order was (1) fluid warmer removal, sight into the choiccfof catheter size. For a typical vein
(2) 16-gauge to 14-gauge catheter change, (3) with resistance R.ein = 22 RU, a 20-gauge catheter
standard-bore to wide-bore tubing change, (4) 14- (R = 17 RU) approximately doubles resistance and
gauge to 12-gauge catheter change, (5) stopcock halves flow. Changing from a 20-gauge to an IS-gauge
removal or change, (6) 12-gauge to 10-gauge catheter catheter decreases total resistance from 39 RU to 32
change, and finally (7) catheter elimination with the RU, producing a resistance decrease of 18%, which
tubing directly connected to the patient's vein. Inter- causes a flow increase of 18%. Increasing catheter size
1191 PART Jl Anesthetic Care

from 18 gauge to 16 gauge decreases resistance and stopcocks, (8) remove the 10-gauge catheter, and (9)
increases flow by 13%. insert the sterile tubing end directly in the vein. 3~
The typical (22 RU) vein cannulated with a 17 RU, As an alternative to reducing resistance, pressure
20-gauge catheter provides a total resistance of 39 can be increased to increase flow. Pressure can be
RU. If the pressure bag or bottle is 110 cm above the increased in either of two ways. Some IV sets contain
patient's heart, infusion pressure equals 110 em H 20 an integral manual pump that allows the clinician to
minus CVP (approximately 10 cm H 20), yielding a produce liquid flow by volume displacement. This
100 cm H 20 pressure head. Since 1 mm Hg = 1.34 cm labor-intensive approach allows bolus administration
H 20 and 1 cm H 20 = 76 mm Hg, pressure head = 76 at any time, completely under the anesthesiologist's
mm Hg. Thus: F = P/R = 76 mm Hg/39 ·mm Hg/U control. Also, this technique allows fluids contained in
hour = 1.95 IJhour, or approximatley 2000 ml/hour. glass bottles (e.g., 5% albumin) to be infused with
For a 10 drop/ml drip chamber: Drop rate = 10 assistance.
drops/ml x 2000 ml!hour x 1 hourI 3600 seconds An alternative approach increases infusion pres-
= 5.6 drops/second. This usually forms a steady sure with a pressure infuser, a pneumatic device that
stream. If a clinical IV infusion seems to flow much encircles the fluid-filled IV bag. Many competing
more slowly than this, the impediment imposed by the devices are available. The maximum pressure sug-
patient's venous system probably is greater than the gested by manufacturers is usually 300 mm Hg,
typical 22 RU used in this example. Most likely a very although much higher pressures are possible (JH
small vein has been cannulated or the cannulated vein Philip, unpublished data, 1990). Severa) different
is not functioning normally. In conclusion, when a models of pressure infusers are available.
small IV catheter (20 gauge) produces low flow of IV With the original pressure-infuser design, the
fluid, the major flow impediment is probably not the clinician uses a manual bulb-inflater pump similar to
IV catheter but rather the patient's vein. that used with a sphygmomanometer cuff. Once the
With gravity as the driving force for fluid flow infuser is filled with air under pressure, infuser
through an IV system, a major limitation in flow arises pressure drops as fluid is infused and leaves the bag,
from the limited pressure head available. The IV bag which is encircled by the infuser. Often a snap clamp
or bottle can typically be placed no higher than 1 m is applied to the tubing between the bag and inflation
above the patient's ,heart. This height limits the bulb because of the misconception tbat infuser
hydrostatic pressure .that can be applied by gravity pressure falls as a result of leaks in the inflation bulb.
alone. However, since the pressure that reaches the Although ·inflation bulbs do leak, the ever-present
patient's vein is often much lower than that applied by leaks are typically small and: are only significant when
the pressure source because of resistive losses in the fluid is infused extremely slowly, as with an Intraflo
catheter and other components, higher pressures are (Abbott/Sorenson Co.) or other continuous low-flow
usually safe. This is especially true when a small device designed to infuse 3 mVhour.
catheter size has been selected and a large vein is To maintain constant driving pressure for infusion,
cannulated. In this situation. positive pressure can be the pressure. infuser can be attached to a constant
applied, as described in the next section. pressure source. Again, the maximum recommended
pressure is 300 mm 'Hg. When the infuser is used in
High-Flow Resuscitation this way, a constant pressure head is applied to the IV
Rapid fluid resuscitation is often required during system-patient combination. This has the advantage
anesthesia and surgery. Since both pressure and of flow continuing at its initial rate, since additional
resistance limit flow, both should be optimized to gas enters the infuser to maintain a constant volume
accomplish rapid fluid flow. Resistance is lowered in the infuser-IV bag combination. '
initially by removing restrictions imposed by roller Whenever fluid is infused under pressure, in-
clamps. Ideally, only one roller clamp is present creased vigilance is required to avoid infusion of air
between bag and patient to reduce the likelihood that into the venous system. Despite the danger of air
a second control will be left unadjusted. Resistance infusion, IV fluid manufacturers add potentially
can be further reduced by removing interposing dangerous volumes of air to IV bags. Volumes as large
devices that may contribute to total resistance. When as 70 ml may be present.l2 Whenever fluid is admin-
nonlinearity is considered, the order of device re- istered under pressure, it is best to remove all air from
moval or change is (1) remove the blood warmer, (2) the IV bag.
change catheter from 16 to 14 gauge, (3) remove As the IV bag empties below 100 to 200 ml, most
check valve, (4) change catheter from 14 to 12 gauge, pressure infusers fail to apply their set pressure
(5) replace regular tubing with wide-bore tubing, (6) effectively to the bag, and infusion pressure drops. In
change catheter from 12 to 10 gauge, (7~ remove an emergency situation, it might be best to discard the
Intr~venous Access and Delivery Principles CHAPTER SS 1193

residual 100 ml of fluid and change bags rather than Many pumps are incapable of detecting problems
try to infuse the last drop. Some infuser designs may downstream from the pump unless these problems
be more effective, but such information has not been produce a significant pressure rise. Such pressure rise
published. could result from total obstruction of the catheter or
Several pressure infusers can be attached to a tubing. Situations such as fluid extravasation usually
single pressure source to facilitate rapid changing of produce only a small rise in pressure because of
crystalloid. and blood-containing bags during emer- resistance to infusion. Most pumps fail to detect such
gencies. One such device34 can provide 479 ml/min of conditions. Complications of extravascular injection
lactated Ringer's solution or 318 mJ/min of diluted clearly become more problematic when the patient is
packed RBC through standard IV tubing without a unable to complain, as during general anesthesia.
catheter. Several other devices have been described",..1
or are commercially available. Other Fluids
When fluids other than dilute crystalloid solutions are
MECHANICAL FLUID INFUSION SYSTEMS infused, the PFR may differ from that expected. The
difference may be in the slope of the PFR line or in
IV Controllers
the shape of this relationship. When viscosity varies,
Many devices are available to adjust or control the the linear slope is expected to change. Differences in
flow of fluid infused. IV controllers use gravity as a density could influence the nonlinear component.
pressure source and manipulate the fluid path to Solutions containing dextrose (JH Philip, unpub-
adjust flow. Flow is usually monitored by an electronic lished data, 1985) provide increased effective viscos-
drop sensor that causes the instrument to vary its ity, as manifested by increased resistance with in-
slowing of the infusion. creased dextrose concentration. The specific relation-
Variable-resistance flow controllers vary the resis- ship is:
tance of the infusion system by constricting or
otherwise increasing the resistance of the tubing RDcW = R-a1Ct (1 + Cl/IOOO)
system. . where C is the dextrose conceAtration expressed in
Variable-duty-cycle controllers interrupt the appli- percent (%) and DeW represents dextrose with
cation of pressure to the fully open IV tubing system concentration C% in water. (Actual slope of the
and thereby control effective pressure applied from regression line is 1.09 ± 0.05, with 95% confidence
the fluid bag or bottle. The maximal effective pressure interval = 0.94 - 1.23).
is limited by the bag height and also by the maximal Applying this relationship, the resistance of DsoW
duty cycle allowed by the infusion device. equals that of water x 1 + 50211000 = 1 + 25001
Some IV infusion controllers are capable of mon- 1000 = 1 + 2.S == 3.5. Therefore, to achieve the
itoring the IV site for deterioration. Devices that . same flow with DsoW as with water, pressure must be
quantify either the resistance imposed or the duty increased by a factor of 3.S. Alternatively, at the same
cycle applied have the capacity to perform monitor.. pressure head, flow diminishes to 1/3.5 = 0.28 of the
ing. Some instruments sound an alarm when flow is flow for dilute crystalloid.
too low, considering the instrument's purposeful From the resistance-concentration relationship, it
limitation of flow. can also be shown that relative viscosities of 1, 2, 3,
Controllers' accUracy is limited by the flow- and 4 are obtained with dextrose concentrations of
monitoring system's accuracy. Since many instru- 0%, 30%, 43%, and 52%. Similar analysis revealed
ments use drop counting, variations in drop size that Intralipid has a reJative viscocity of 1.36, equiv-
described earlier play an important limiting role. alent to that of 17% dextrose in water. Osmolyte (for
gastric infusion) has a relative viscosity of 6.57.
IV Pumps
Fluid pumps use positive displacement to provide DRUG ADMINISTRATION
rc;gulated fluid infusion. Positive displacement can be
provided by peristaltic fingers, a reciprocating syringe, The IV route is used extensively for drug administra-
or other mechanisms. tion during anesthesia. Drugs can be administered
Because pumps use positive displacement, they with several different infusion profiles. The most
may be capable of overcoming high resistance or back common drug infusion regimens are bolus injection
pressure to fluid infusion. This property can be and constant infusion. Occasionally, careful pharma-
advantageous in some situations. However, in the cokinetic control is used. The most common form uses
patient with fluid extravasation, some danger clearly an exponential decay in infusion rate to link bolus
exists. infusion to continuous infusion.
1194 PART II Anesthetic Care

Delivery Devices netic model or can vary drug infusion pbarmacody·


Many commercial devices are available for drug namically in response to monitored pbysiologic
delivery, and several different technologies are avail- changes in the patient.
able. Peristaltic pumps are usually reserved for
non drug fluid infusion because of the interface to an
IV bag. However, many clinicians use a Buratrol or SUMMARY
other device as a reservoir for the drug and then use IV therapy is an essential part of every anesthesia
a conventional fluid pump. procedure. The IV infusion system connected to a
Syringe pumps appear better suited to drug infu- patient can be analyzed according to simple physical
sion, since a small drug volume lends itself to and physiologic principles, and the resulting .model
containment in a syringe. Many competing devices are can predict the IV infusion system behavior in many
available. situations.
Some drug infusion pumps are designed to facili- Devices are available to accommodate almost any
tate drug administration on a milligrams/kilograms of need in fluid or drug administration. As new devices
weight basis. Others specifically facilitate use of become available, the anesthesiologist must under-
individual drugs' by using drug-specific labels. Still stand their operating principles and the physics that
other drugs are capable of control by a computer that underlies them to recognize appropriate applications.
can infuse according to a prospective pharmacoki-

KEY POINTS
• The' simplest IV system consists of an elevated, • The extravascular tissue near a vein behaves as a
fluid-filled container, a length of flexible tubing, a resistor that offers significantly more resistance
catheter cannulating the patient's vein, and the than does the vein.
patient's venous system. • Many clinical situations produce vein collapse,
• Flow through the IV system is determined by driving causing the vein to behave as a Starling resistor.
pressure and resistance of the catheter and tubing • When IV flow rate is low, the pressure-tlow rela-
system. tionship is linear. When flow is increased signifi-
• Driving pressure equals the height of the free-air cantly, the pressure-flow relationship becomes non-
surface of the fluid above the reference level, less linear and this significantly affects flow prediction
the height of the intervening air gap in the drip during rapid fluid infusion, as is required during
chamber. volume resuscitation.
• IV tubing behaves as a low-resistance conduit rela- • Great care must always be taken to avoid infusion of
tive to other system components: R.ut.ina = 3 ± 2 air into the venous system whenever fluid is infused
(SEM) resistance units (RU).l RU = 1 mm HgIU under pressure; this is especially so since IV fluid
hour. manufacturers add volumes of air as large as 70 ml
• A roller clamp behaves as a variable resistor in that, to IV bags, which preferably should be removed
once set, it behaves the same (± 5%) when before use.
connected to a flow source (IV pump) as to a • Numerous mechanical systems are available to
pressure source (elevated bag). Typical roller clamp adjust or control the flow of the fluid infused. Simple
resistance during normal IV flow is 300 to 800 RU, controllers use gravity as the pressure sQurce and a
d~pending on flow rate desired. variable resistor to adjust the rate of infusion. The
• An IV catheter behaves as a resistor that offers accuracy of controllers is limited by the accuracy of
more resistance than does IV tubing, so that in the the flow monitoring mechanism, which often con-
flow range of 0 to 300 mVhour, the pressure-flow sists of a drop counting system that is limited by the
relationship (PFR) is linear and resistance takes on variations in the size (and fluid volume) of the
a unique value for each catheter size. drops. Further, the pressure source varies because
• A normal patient vein behaves like a resistor with of patient movement and changes in the height of
resistance similar to that of a 20-gauge IV catheter. the bed.
The central venous circulation offers essentially no • Fluid pumps use positive displacement to provide
resistance to fluid flow. regulated fluid infusion. Positive displacement is
Intravenous Access and Delivery Principles CHAPTER SS 1195

provided by peristaltic compression, by a recipro- • Several systems are available to facilitate intrave-
cating syringe, or by other mechanisms. Such nous drug administration. Many are now pro-
systems can infuse fluid despite variations in down- grammed to calculate the infusion rate, when data
stream resistance, an effect that can be beneficial about patient weight, drug concentration, and de-
(e.g., by eliminating changes ,in fluid administration sired dose (mg/kg) are provided. Although not fool-
due to patient movement) or detrimental (e.g., proof, such devices should eliminate many of the
subcutaneous extravasation of fluid), depending on risks of human error, especially in situations where
the conditions. Many pump systems incorporate a the clinician has multiple tasks to perform simulta-
pressure monitor that can detect complete obstruc- neously, as in complex anesthesia procedures.
tion of the outflow system but may not detect
subcutaneous extravasation.

KEY REFERENCES
Permutt S, Riley RL: Hemodynamics of collapsible Philip JH: A model for the physics and physiology of
vessels with tone: the vascular waterfall, J Appl fluid administration, J elin Monit 5:123, 1989.
Physio/ 18:924, 1963. Shapiro AH: Steady flow in collapsible tubes,
Philip BK, Philip JH: Prediction of low capability in J Biomech Eng 99:126,1977.
intravenous infusion systems: implications for fluid
resuscitation, J elin Monit 6: 113, 1990.

REFERENCES
1. Allen P: A standardization of the Lewis 13. Haug IN, Politser PE: Socio-economic 23. Leith DE: Physiological waterfalls. Phys-
method of venous pressure determina· lact book for surgery, Chicago, 1987, ioI Teacher (Am Physiol Soc) 5:6, 1976.
tion, Can Med Assoc J 59:560, 1948. American College of Surgeons. 24. MacCara ME: Extravasation: a hazard of
2. Brower RW, Noordergraaf A! Pressure· 14. Henhey CO, Tomford JWT, McLaren intravenous therapy, Drug Intell Clin
flow characteristics of collapsible tubes: a CB et al: The natural history of intrave- PhannacolI7:713, 1983.
reconciUation of seemingly contradictory nous catheter·associated phlebitis, Arch 25. Moreno AH, Katz AI, Gold LD et al:
results, AM Biomtd Eng 1:333, 1973. InJem Med 144:1373, 1984. Mechanics of distension of dog veins and
3. earo CG. Pedley TJ, Schroter RC et at: 15. Holt JP: The collapse factor in the mea· other very thin-waned tubular structures,
The mechanics 01 the clrculIIdon, Oxford. surement of venous pressure, Am J Phys· Cire Res 27:1069, 1970.
1978. Oxford University Press. 101134:292, 1941. 26. Permutt S, Riley RL: Hemodynamics of
4. Chapman RB, Keep P: The Norfolk and 16. Holt JP: Flow through collapsible tubes collapsible vessels with tone: the vascular
Norwich infusion box. Anustlwia 35: and through in situ veins, IEEE Tnms waterfall, J Appl Physiol 18:924, 1963.
1211,1980. Biomed InslTUm 16:274, 1969. 27. Philip BK, Philip JH: Characterization of
5. Conrad WA: Pressure·flow relationships 17. Hutchinson PM, Yeoman PM, and Byrne [nonlinear] flow in intnvenous infusion
in collapsible tubes, IEEE Tmns Biomed AJ: Evaluation of the IVAC 560 volumet- systems, IEEE Tmns Biomed Eng 30:702.
[/Utrum 16:284, 1969. ric pump, Anaesthesia 40:996. 1985. 1983.
6. Duomarco JL. Rimini R: Energy and 18. Katz AI, Chen Y, and Moreno AH: Flow 28. Philip BK, Philip JH: Characterization of
hydraulic gradients along systemic veins, through a collapsible lube: experimental [nonlinear) How in intravenous catheters,
Am J PhysioI178:215, 1954. analysis and mathematical model, Biophys IEEE Trans Biomed Eng 33:529, 1986.
7. Ferenchak P, Collins JJ, and Morgan AP: J 9:1261, 1969. 29. Philip BK, Philip JH: Prediction of flow
Drop size and rate in parenteral infusion, 19. Knowlton FPt Starling EH: The influence capability in intravenous infusion sys·
Surgery 70:674, 1971. of variations in temperature and blood- tems: implications for fluid resuscitation,
8. Flack FC, Whyte TD: Behaviour of stan. pressure on the performance of the iso- J elin Monit 6:113, 1990.
dard gravity. fed administration sels used lated mammalian heart, J Physiol (Lond) 30. Philip JH: A model for the physics and
, for intravenous infusion, Br Med J 3:439, 44:206, 1912. physiology of fluid administration, J Clin
1974. 20. Kresch E: Compliance of ftexible tubes, Manit 5:123, 1989.
9. Franklin KJ: A monograph on veiIU J Biomech 12:825, 1979. 31. Philip JH. Joseph DM: Peripheral venous
Springfield, IU, 1937. Charles C Thomas: 21. Kresch E, Noordergraaf A: A mathemat. pressure can be an accurate estimate of
10. Fung YC: Biodynamics: circulation, New ical model for the pressure-flow relation- central venous pressure, Anesthesiology
Yo.rk. 1984. Springer.Yerlag. ship in a segment of vein, IEEE Trans 61:AI66, 1986 (abstract).
11. Gn(fiths DJ: Steady fluid flow through Biomtd Insnum 16:335. 1969. 32. Philip JH. Philip BK: Avoiding air infu-
veins and collapsible tubes, Med Bioi Eng 22. La Cour D: Drop size in intravenous sion with pressurized infusion systems: a
9:597, J971. infusion, Acta Anaesthesiol Scand 24:35, new hazard, Antsth Ana/g 64:381, 1985.
12. Gutyon AC: Textbook 01 medical pJaysiol. 1966. 33. Philip JH, Philip BK: Hydrostatic central
081, Philadelphia, 1976, WB Saunders.
venous pressure measurement by IVAC
1196 PART II Anesthetic Care

560 infusion pump. Med Insttum 19:232, lated. passive, soft-walled vessels, Am teral therapy, Am J IV Ther elin Nutr
1985. Hearl 146:648. 1963. 8:9, 1981.
34. Philip JH. Philip 8K: Pressurized infusion 39. Rodbard S: Flow through coUapsible 45. Turnidge J: Hazards of peripberal intra-
system (or fluid resuscitation. Anesth tubes: augmented flow produced by resb- venous lines. Mtd J Aust 141:37, 1984.
AntI.Ig 63:710. 1984. tanee at the outlet, Circulation 11:290, 46. Upton J, Mulliken lB. and Murray JE:
35. Philip Iii, Philip 8K: Simplified [linear] 1955. Major intravenous extravasation injuries,
apprOach to usessing intravenous flow 40. Rodbard S. Saiki H: Flow through col- Am J Surg 137:497, 1979.
characteristics in the therapeutic InfusioD lapsible tubes, Am Hearl 1 11:715, 1955. 47. Warren JV, Stead EA: The effect of the
range, IEEE Trans Blomed Eng 35:1093. 41. Rosenblatt R, Dennis P, and Draper LD: accumulation of blood in the extremities
1988. A new method for massive fluid resusci- on the venous pressure of normal sub-
. 36. Philip JH. Philip BK. and Lchr JL: tatiOD in the trauma patient,AnesthAIIQ(f jeclS, Am J Med Sci 205:501. 1943.
Accuracy of bydrostatic pressure mea- 62:613. 1983. 48. Wetmore N: Extravasation - tbe dreaded
surement with a disposable dome trans- 42. Shapiro All: Steady flow in coUapsible complication, NaIl IV Ther Assoc 8:47,
ducer, Med illSt11Un 19:273, 1985. tubes. 1 Biomech Eng 99:126, 1977. 1985.
37. Plumer AL, Cosentino F: Principlu and 43. Simmons BP: Guidelines for prevention 49. Ziser M, Feezor M, and Slolaut MW:
practiet 01 inlmvtnolU therapy, Chicago, of intravascular infections, Nail IV The,. Regu1atins intravenous flow: controller
1987 Scott Foresman.
t A.ssoc 5:41. 1982. versus clamps, Am 1 Hosp Pharm 36: 1090,
38; Rodbard S: Autoregulation in encapsu- 44. Turco SJ: Hazards associated with paron- 1979.

Das könnte Ihnen auch gefallen