Sie sind auf Seite 1von 320

General

Revision history
From time to time, update sheets will be issued to bring your service manual
up-to-date. For ease of updating, the new information will be page for page,
or section for section, replaceable. You will also get a new update table so
you can check the contents of your manual to verify you have the latest
information. You can also check what has been changed from the previous
revision.

Revision history table


Revision level Date of amendment Notes
A April 2003 AEC Paramdump
instructions,
menu description
updates (sw 1.02.5),
details
- June 11, 2001 New manual

Conventions used in this document


This manual has following conventions:

NOTE
Contains useful information for the reader about the unit and its use.

CAUTION
Contains important instructions. If these instructions are not observed,
malfunction of the unit or damage to the unit or other property may occur.

WARNING
Contains warnings and instructions about the safety of the unit. If these
warnings are not respected, serious risks and injury may be caused to the
patient and operator.

Notes, Cautions and Warnings

NOTE
Only engineers trained by Instrumentarium are allowed to install the unit or
stand by it during the installation or service work. Parts installed by
nontrained personnel carry no warranty!

86502-4AA Instrumentarium Imaging


CAUTION
Always check for proper seating on all the connectors or external components
located in the areas you have worked. Equipment damage can result if power
is applied or exposures made with improperly seated connectors or external
components!

WARNING
Always install the proper fuses into the unit. Failure to do so jeopardizes
patient, operator and equipment safety.

WARNING
This unit has floating grounds. Always make all scope measurements
differentially and always be careful when servicing the unit!

CAUTION
When handling circuit boards, always utilize ground wrist straps or
grounding kits to prevent damage to the “mos” type chips. Always keep
circuit boards in a static proof bag.

WARNING
Always disconnect the unit from line supply by removing the plug from wall
outlet before opening the covers.

WARNING
The X-ray tube utilizes a Beryllium window. Do not perform any operations
on Beryllium which produce dust or fumes, such as grinding, grit blasting or
acid cleaning. Beryllium dust or fumes are highly toxic and breathing them
can result in serious personal injury or death.

WARNING
This X-ray unit may be dangerous to patient and operator unless safe
exposure factors and instructions are observed!

WARNING
Pay attention during the installation not damage optoswitches, c-arm
mounting and other parts. Check the unit thoroughly before handing it to the
client.

Instrumentarium Imaging 86502-4AA


Table of Contents
4 Parameter programming and software load................................................. 4-1
4.1 Overview/functional description ..........................................................................4-1
4.2 Service mode overview ........................................................................................4-2
4.2.1 A: System setup ..................................................................................................... 4-2
4.2.2 B: Defaults ............................................................................................................. 4-3
4.2.3 C: Diagnostics - NOT USED ................................................................................. 4-7
4.2.4 D: AEC................................................................................................................... 4-8
4.2.5 E: C-Arm Rotation ............................................................................................... 4-20
4.2.6 F: Collimator........................................................................................................ 4-23
4.2.7 G: Compression ................................................................................................... 4-26
4.2.8 H: Filter ................................................................................................................ 4-29
4.2.9 I: Generator .......................................................................................................... 4-30
4.2.10J: ParkBack .......................................................................................................... 4-31
4.2.11K: Stereo Movement ............................................................................................ 4-32
4.2.12L: Physicist mode................................................................................................. 4-33
4.2.13General software update information................................................................... 4-34
4.2.13.1Upgrade from 1.02.2 - 1.02.4 to 1.02.5............................................................................4-34
4.3 Hyperterminal interface for Diamond ................................................................4-34
4.3.1 Configuration of the HyperTerminal: .................................................................. 4-34
4.3.2 Data capture ......................................................................................................... 4-35
4.3.3 Save setup data..................................................................................................... 4-39
4.4 Diamond test displays.........................................................................................4-39
4.5 Software versions ...............................................................................................4-39
4.6 Specimen mode...................................................................................................4-39
5 Electrical system architecture - theory of operation .................................... 5-1
5.1 Overview ..............................................................................................................5-1
5.2 Input power...........................................................................................................5-2
5.2.1 Supply Voltage Regulation .................................................................................... 5-3
5.2.2 Sensor types ........................................................................................................... 5-3
5.3 Printed circuit board (PCB) communication ........................................................5-4
5.3.1 Internal Serial Bus.................................................................................................. 5-4
5.3.2 External Serial Bus ................................................................................................ 5-4
5.4 Printed circuit boards (PCB).................................................................................5-5
5.4.1 Supervisor Board ................................................................................................... 5-5
5.4.2 Power Board........................................................................................................... 5-5
5.4.3 Power supply section ............................................................................................. 5-5
5.4.4 C-Arm rotation motor control................................................................................ 5-6
5.4.5 C-Arm Rotation Sensors ........................................................................................ 5-6
5.4.6 Vertical movement or Z-motor control.................................................................. 5-7
5.5 Generator Board ...................................................................................................5-8
5.5.1 kV Control ............................................................................................................. 5-8
5.5.1.1High voltage section ............................................................................................................5-8
5.5.1.2Timing..................................................................................................................................5-9
5.5.1.3Low voltage section .............................................................................................................5-9
5.5.2 mA Control .......................................................................................................... 5-10
5.5.3 Protection circuits ................................................................................................ 5-11
5.5.4 Back Up Timer..................................................................................................... 5-12
5.5.5 Monitoring ........................................................................................................... 5-12
5.6 Anode Motor Drive Board - AMD Board ..........................................................5-13
5.7 Stereo Board .......................................................................................................5-14

86502-4AA Instrumentarium Imaging 1


5.7.1 Voltage regulators ................................................................................................ 5-14
5.7.2 Stereo and ECS motor drive ................................................................................ 5-15
5.7.2.1Stereo Sensors....................................................................................................................5-15
5.7.3 Compression motor drive..................................................................................... 5-16
5.7.4 Back up release .................................................................................................... 5-16
5.7.5 Sensors and Keyboard.......................................................................................... 5-16
5.8 Drive Board ........................................................................................................5-17
5.8.1 Grid movement .................................................................................................... 5-17
5.8.2 Cassette tray movement ....................................................................................... 5-17
5.8.3 Sensors ................................................................................................................. 5-18
5.8.4 Cassette Sensors................................................................................................... 5-19
5.9 Collimator Board ................................................................................................5-20
5.9.1 Collimator plate and filter drive........................................................................... 5-21
5.9.2 Collimator Sensors............................................................................................... 5-21
5.9.3 Tube Park drive.................................................................................................... 5-21
5.9.4 Tube Park movement sensors .............................................................................. 5-22
5.9.5 Filter sensor.......................................................................................................... 5-23
5.10 Compression Board ............................................................................................5-23
5.10.1Compression thickness measurement .................................................................. 5-23
5.10.2Compression force measurement ......................................................................... 5-23
5.10.3Displays................................................................................................................ 5-23
5.10.4Keyboard.............................................................................................................. 5-24
5.10.5Compression Sensors ........................................................................................... 5-24
5.11 AEC Board .........................................................................................................5-28
5.11.1Automatic Exposure Control ............................................................................... 5-28
5.11.2Sensors ................................................................................................................. 5-28
5.12 Fuses, voltages and led indications.....................................................................5-29
5.12.1Fuses .................................................................................................................... 5-29
5.12.2Voltages ............................................................................................................... 5-30
5.12.3LED indications ................................................................................................... 5-31
6 System checks and preventive maintenance ................................................. 6-1
6.1 X-RAY TUBE PERFORMANCE CHECK .........................................................6-1
6.1.1 Half-value layer (HVL) measurement ................................................................... 6-1
6.1.2 Tube efficiency check on the cassette holder ........................................................ 6-2
6.1.3 Image quality evaluation procedure....................................................................... 6-3
6.1.4 Measurement of the feedback circuitry.................................................................. 6-5
6.1.5 kV, mAs and exposure time test ............................................................................ 6-6
6.2 Automatic exposure control evaluation................................................................6-7
6.2.1 Quick AEC test ...................................................................................................... 6-7
6.2.2 Exporting and storing the system settings data...................................................... 6-7
6.2.3 Automatic AEC detector gain adjustment ............................................................. 6-7
6.2.4 Adjustment of an individual detector..................................................................... 6-8
6.3 Mechanical checks................................................................................................6-8
6.3.1 Compression system .............................................................................................. 6-8
6.3.2 C-Arm rotation mechanism.................................................................................... 6-8
6.3.3 Column and carriage .............................................................................................. 6-9
6.3.4 Compression force check ....................................................................................... 6-9
6.3.5 X-ray field alignment check................................................................................... 6-9
6.3.6 X-ray field light alignment check ........................................................................ 6-10
6.4 Regular maintenance plans.................................................................................6-11
7 Repair instructions .......................................................................................... 7-1
7.1 Column repair .......................................................................................................7-1
7.1.1 C-Arm movement friction adjustment ................................................................... 7-1

12 Instrumentarium Imaging 86502-4AA


7.1.2 Preparing the carriage and counterweight.............................................................. 7-2
7.1.3 Replacement of the Z-motor .................................................................................. 7-2
7.1.4 Replacement of the column cap ............................................................................. 7-3
7.1.5 Replacement of the counterweight wires ............................................................... 7-3
7.2 Electrical board and mains transformer locations ................................................7-6
7.2.1 Replacement of Power, Supervisor and Generator Boards.................................... 7-7
7.2.2 Replacement of the AMD Board ........................................................................... 7-7
7.2.3 Replacement of Control Panel Board .................................................................... 7-7
7.2.4 Replacement of the Compression Board or Compression Display Board ............. 7-8
7.2.5 Replacement of the mains transformer .................................................................. 7-8
7.3 C-Arm repair.........................................................................................................7-9
7.3.1 Removal and assembly of the Diamond covers ..................................................... 7-9
7.3.2 Installation of the tube head covers ....................................................................... 7-9
7.3.2.1Optoswitch check.................................................................................................................7-9
7.3.2.2Finger protection sheets .....................................................................................................7-10
7.3.2.3Installation of the finger protection sheets.........................................................................7-10
7.3.2.4Installation of the lower front end of the tube head cover .................................................7-11
7.3.2.5Installation of the rear end of the tube head cover.............................................................7-12
7.3.2.6Attaching top tube head cover ...........................................................................................7-13
7.3.3 Left side cover...................................................................................................... 7-14
7.3.4 Right clam shell cover ......................................................................................... 7-15
7.3.5 Replacement of handles ....................................................................................... 7-16
7.3.6 Replacement of the vertical drive control switch ................................................ 7-17
7.3.7 Replacement of the x-ray tube ............................................................................. 7-18
7.3.8 High voltage unit replacement ............................................................................. 7-19
7.3.9 Compression system replacement........................................................................ 7-19
7.3.10Replacement of the compression motor............................................................... 7-21
7.3.11Replacement of the compression thickness measurement system ....................... 7-24
7.3.12Adjustment of the compression mechanism belt ................................................. 7-26
7.3.13Adjustment of the mechanical compression limit................................................ 7-27
7.3.14Replacement of the C-Arm motor........................................................................ 7-28
7.3.15Replacement of ECS mechanism/ECS motor...................................................... 7-29
7.4 Other adjustments or replacements.....................................................................7-31
7.4.1 X-ray field adjustment ......................................................................................... 7-31
7.4.2 X-ray field light alignment................................................................................... 7-33
7.4.3 Replacement of the light field bulb...................................................................... 7-34
7.4.4 Tube head movement and torque adjustment ...................................................... 7-34
7.4.5 Tube head belt check ........................................................................................... 7-35
7.4.6 Grid alignment ..................................................................................................... 7-36
7.4.7 Grid movement mechanism check....................................................................... 7-37
7.4.8 Replacement of Bucky carbon fiber .................................................................... 7-39
7.4.9 Replacement of c-arm membrane buttons ........................................................... 7-39
7.5 Compression adjustment.....................................................................................7-40
7.6 Negative compression force calibration .............................................................7-41
8 Error tracing .................................................................................................... 8-1
8.1 Information for the service error tracing ..............................................................8-1
8.2 Diamond error codes ............................................................................................8-1
8.2.1 0002 Program check sum error .............................................................................. 8-2
8.2.2 0003 Program check sum error .............................................................................. 8-2
8.2.3 0011........................................................................................................................ 8-2
8.2.4 0012 Tube Park safety clutch active ...................................................................... 8-3
8.2.5 0013 Tube movement ............................................................................................ 8-3
8.2.6 0014........................................................................................................................ 8-4
8.2.7 0015 Collimator plate movement........................................................................... 8-4
8.2.8 0022........................................................................................................................ 8-5

86502-4AA Instrumentarium Imaging 1-3


8.2.9 0023........................................................................................................................ 8-5
8.2.100024 Compression force measurement .................................................................. 8-5
8.2.110027........................................................................................................................ 8-6
8.2.120029 Underflow in Auto Time or Auto kV-mode ................................................. 8-6
8.2.130033 Cassette tray movement ................................................................................ 8-6
8.2.140034 Cassette ......................................................................................................... 8-7
8.2.150035 Cassette opto ................................................................................................. 8-8
8.2.160037........................................................................................................................ 8-8
8.2.170038 Cassette holder .............................................................................................. 8-9
8.2.180041 Stereo not in center position ....................................................................... 8-10
8.2.190043 Vertical movement motor back up time reached ........................................ 8-10
8.2.200045 ECS motor back up time reached................................................................ 8-11
8.2.210048...................................................................................................................... 8-11
8.2.220051...................................................................................................................... 8-11
8.2.230053...................................................................................................................... 8-11
8.2.240054...................................................................................................................... 8-12
8.2.250055...................................................................................................................... 8-12
8.2.260057 Missing mA................................................................................................. 8-13
8.2.270058...................................................................................................................... 8-14
8.2.280059...................................................................................................................... 8-14
8.2.290060...................................................................................................................... 8-15
8.2.300062 Missing kV.................................................................................................. 8-16
8.2.310063...................................................................................................................... 8-17
8.2.320064 Missing cassette .......................................................................................... 8-17
8.2.330065 Excessive exposure time ............................................................................. 8-18
8.2.340066 Movement not allowed ............................................................................... 8-18
8.2.350067 Diamond not ready for exposure................................................................. 8-19
8.2.360068...................................................................................................................... 8-19
8.2.370069...................................................................................................................... 8-20
8.2.380070...................................................................................................................... 8-20
8.2.390071...................................................................................................................... 8-20
8.2.400072 Side collimator ............................................................................................ 8-21
8.2.410074 AEC bus error ............................................................................................. 8-21
8.2.420075 Communication problem ............................................................................ 8-22
8.2.430076...................................................................................................................... 8-22
8.2.440077...................................................................................................................... 8-23
8.2.450078...................................................................................................................... 8-23
8.2.460080...................................................................................................................... 8-24
8.2.470081...................................................................................................................... 8-24
8.2.480082...................................................................................................................... 8-24
8.2.490084...................................................................................................................... 8-24
8.2.500087...................................................................................................................... 8-25
8.2.510089 C-Arm center limit not found ..................................................................... 8-25
8.2.520090...................................................................................................................... 8-26
8.2.530091 Z-movement disabled in Specimen mode................................................... 8-26
8.3 No power indications/Board LEDs.....................................................................8-26
8.4 AEC inoperative .................................................................................................8-26
8.5 Supply voltages...................................................................................................8-26
8.6 Grid line check items ..........................................................................................8-27
8.7 ECS trouble shooting..........................................................................................8-28
8.8 Serial bus communications check.......................................................................8-28
8.9 Notes on software upgrading ..............................................................................8-29
8.10 Questions & Answers .........................................................................................8-29
8.11 FAQ - Frequently asked questions .....................................................................8-30
8.11.1Common questions............................................................................................... 8-30

14 Instrumentarium Imaging 86502-4AA


8.11.2Customer questions .............................................................................................. 8-32
8.11.3Installation/service questions ............................................................................... 8-33
8.11.4Installation/service questions/error messages ...................................................... 8-35
9 Specifications.................................................................................................... 9-1
10 Schematics/Layouts ....................................................................................... 10-1
11 Cable drawings .............................................................................................. 11-1
12 Appendices ..................................................................................................... 12-1
12.1 Parameter programming with Control Panel ......................................................12-1
12.2 Loading software with old Flash Loader............................................................12-2
12.3 Diamond test displays.........................................................................................12-5

86502-4AA Instrumentarium Imaging 1-5


16 Instrumentarium Imaging 86502-4AA
4 Parameter programming and software load

4 Parameter programming and software load


4.1 Overview/functional description
Main Diamond operation modes are Auto kV, Auto Time and Manual mode.
In Auto-kV mode, Diamond determines the required amount of kV. In Auto
Time mode, Diamond determines the required exposure time. In Manual
mode, the user sets up all the required values for the examination.

Many programmable parameters that affect these modes can be set and
adjusted in Diamond service mode. Generation of radiation may be switched
off during the service work. This enables the service technician to make test
exposures to verify the correct voltages on various boards. Refer to Diamond
Installation Instructions (86501) for instructions to disable the emission of X-
rays (for example at exhibition).

NOTE
Diamond tube head assembly incorporates a heat integrator that cooling time
of 5 to 55 sec. between exposure is implemented also in the service mode as
in the normal mode. This will protect the unit from being overloaded during
testing.

This manual has been divided into separate software, mechanical, service,
maintenance and troubleshooting sections. Schematics, layouts and cables
have been placed at the end of this manual.

Software section helps to calibrate and to set up correct parameters for the
user or change the existing ones. The mechanical section explains about tests
and other mechanical issues not covered in software and electronics section.
The electronics section explains about the operation of the Printed Circuit
Boards (PCB), various connections and measurements. The maintenance
section covers the regular maintenance procedures. The troubleshooting
sections include error codes and solutions to some problems that might come
about.

Diamond buttons are described in Diamond Operation Instructions (86500).

86502-4AA Instrumentarium Imaging 4-1


4 Parameter programming and software load

4.2 Service mode overview


Service programs contain special programs for system setups, calibration and
diagnostics. These are accessed and adjusted best with optional Diamond
Service Software (87502). Especially AEC setup is smooth as the Diamond
Service Software displays the AEC curves and effects of the changes on the
curves. The use of this smooth software is highly recommended. In case a
conventional method must be used for some reason, the appendices contain
instructions on parameter setup with the Control Panel and software upload.

Diamond service mode interface has been divided into several topic groups in
which the system setting and parameters can be changed:

A: System setup B: Defaults C: Diagnostics


D: AEC E: C-Arm Rotation F: Collimator
G: Compression H: Filter I: Generator
J: Tube Park K: Stereo Movement L: Physicist mode

For graphic description of the Diamond software functions, software default


settings and ranges, refer to Diamond Installation Instructions (86501),
Appendix section, Software block diagrams and Factory default settings.

Some features listed and with dark shade in the tables are not used at the
moment.

Sample
Parameter Range (Default) Notes
Sample parameter 12345 This parameter is in use.
Sample parameter 2 12345 Not in use at the
moment.

4.2.1 A: System setup

This service program is used to set and adjust system and communication
parameters. Perform required changes in both groups:

System Parameters - PrA01


Parameter Range (Default) Notes
Serial no. 12345 Serial number of the unit
Country code 123 NOT USED at the
moment.
Efficiency Mo 0 - 200 (16) at 28 kV R/100mAs
Efficiency Rh 0 - 200 (12) at 28 kV R/100mAs
Efficiency Al 0 - 1000 (35) at 39 kV R/100mAs
HVL Mo 0 - 60 (0.33) at 28 kV mmAl value
HVL Rh 0 - 60 (0.38) at 28 kV mmAl value
HVL Al 0 - 60 (0.47) at 39 kV mmAl value

42 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Communication Setup - PrA02, NOT USED at the moment.


Parameter Range (Default) Notes
Serial no. 12345 Serial number of the unit
Label out None/Com1/Com2/LPT Prints out the patient
(None) data labels for films (ID
camera). Not in use at
the moment.
Label type 1 to 3 (2) The number of rows.
Not in use at the
moment.
Data out None/Com1/Com2/LPT The port that sends out
(None) the exposure data.
Com1 19200 (permanent The speed for Com1
setting) port. Used in software
updates and diagnostics.
Com2 1200 (permanent The speed for Com2
setting) port. Used by ID II film
flasher.

4.2.2 B: Defaults

This service program is used to set and adjust default parameters. Perform
required changes in all groups:

Start Up mode settings - PrB01


Parameter Range (Default) Notes
PreProg1-4 1/2/3/4 (1) The default mode Diamond uses after
it has been the switched on.
AutoRelease On/Off (On) If ON, Diamond releases the
compression after an exposure. If
OFF, the compression remains after
the exposure.
Auto Tube Park On/Off (Off) If ON, the tube head slides back after
an exposure. If OFF, the tube head
remains in the front after the
exposure. NOT USED!

86502-4AA Instrumentarium Imaging 4-3


4 Parameter programming and software load

Bucky settings - PrB02 - NOT USED at the moment.


Parameter Range (Default) Notes
Mode Auto kV/ Auto kV: Diamond determines the
Auto Time/ required kV and exposure time for
Manual the exposure.
(Auto kV) Auto Time: Diamond determines the
required exposure time.
Manual: All values are manually set.
Default filter Mo/Rh/Al (Mo) Default filter that Diamond will use
unless otherwise instructed in other
settings.
Auto Dose On/Off (Off) If ON, Diamond determines the
NOT USED at Patient Dose. If OFF, the amount of
the moment. patient dose depends on other setting
factors. NOT USED!
AutoPoint On/Off (On) If ON, Diamond determines the
correct detector to be used. If OFF,
the user has to choose the correct
detector (under the densest part of the
breast).
Default detector 0 to 8 (1) Default detector that Diamond uses
unless otherwise instructed in other
settings.
Default kV 23 to 35 (28) Default kV in Auto Time-mode for
Auto Time Molybdenum filter that Diamond
Mo uses unless otherwise instructed in
other settings.
Default kV 23 to 35 (28) Default kV in Auto Time-mode for
Auto Time Rhodium filter that Diamond uses
Rh unless otherwise instructed in other
settings.
Default kV 23 to 35 (28) Default kV in Manual mode with
Manual Molybdenum filter that Diamond
Mo uses unless otherwise instructed in
other settings.
Default kV 23 to 35 (28) Default kV in Manual mode with
Manual Rhodium filter that Diamond uses
Rh unless otherwise instructed in other
settings.
Default mAs 0 - 33 (5) Default mAs value in Manual mode
that Diamond uses unless otherwise
instructed in other settings.

44 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Magnification settings - PrB03 - NOT USED at the moment.


Parameter Range (Default) Notes
Mode Auto kV/ Auto kV: Diamond determines the
Auto Time/ required kV and exposure time for
Manual the exposure.
(Auto kV) Auto Time: Diamond determines the
required exposure time.
Manual: All values are manually set.
Default filter Mo/Rh/Al (Mo) Default filter that Diamond will use
unless otherwise instructed in other
settings.
Auto Dose On/Off (Off) If ON, Diamond determines the
NOT USED at Patient Dose. If OFF, the amount of
the moment. patient dose depends on other setting
factors. NOT USED!
AutoPoint On/Off (On) If ON, Diamond determines the
correct detector to be used. If OFF,
the user has to choose the correct
detector (under the densest part of the
breast).
Default detector 0 to 8 (1) Default detector that Diamond uses
unless otherwise instructed in other
settings.
Default kV 23 to 35 (28) Default kV in Auto Time-mode for
Auto Time Molybdenum filter that Diamond
Mo uses unless otherwise instructed in
other settings.
Default kV 23 to 35 (28) Default kV in Auto Time-mode for
Auto Time Rhodium filter that Diamond uses
Rh unless otherwise instructed in other
settings.
Default kV 23 to 35 (28) Default kV in Manual mode with
Manual Molybdenum filter that Diamond
Mo uses unless otherwise instructed in
other settings.
Default kV 23 to 35 (28) Default kV in Manual mode with
Manual Rhodium filter that Diamond uses
Rh unless otherwise instructed in other
settings.
Default mAs 0 - 33 (5) Default mAs value in Manual mode
that Diamond uses unless otherwise
instructed in other settings.

86502-4AA Instrumentarium Imaging 4-5


4 Parameter programming and software load

Contact settings - PrB04 - NOT USED at the moment.


Parameter Range (Default) Notes
Mode Auto kV/ Auto kV: Diamond determines the
Auto Time/ required kV and exposure time for
Manual the exposure.
(Auto kV) Auto Time: Diamond determines the
required exposure time.
Manual: All values are manually set.
Default filter Mo/Rh/Al (Mo) Default filter that Diamond will use
unless otherwise instructed in other
settings.
Auto Dose On/Off (Off) If ON, Diamond determines the
NOT USED at Patient Dose. If OFF, the amount of
the moment. patient dose depends on other setting
factors. NOT USED!
AutoPoint On/Off (On) If ON, Diamond determines the
correct detector to be used. If OFF,
the user has to choose the correct
detector (under the densest part of the
breast).
Default detector 0 to 8 (1) Default detector that Diamond uses
unless otherwise instructed in other
settings.
Default kV 23 to 35 (28) Default kV in Auto Time-mode for
Auto Time Molybdenum filter that Diamond
Mo uses unless otherwise instructed in
other settings.
Default kV 23 to 35 (28) Default kV in Auto Time-mode for
Auto Time Rhodium filter that Diamond uses
Rh unless otherwise instructed in other
settings.
Default kV 23 to 35 (28) Default kV in Manual mode with
Manual Molybdenum filter that Diamond
Mo uses unless otherwise instructed in
other settings.
Default kV 23 to 35 (28) Default kV in Manual mode with
Manual Rhodium filter that Diamond uses
Rh unless otherwise instructed in other
settings.
Default mAs 0 - 33 Default mAs value in Manual mode
that Diamond uses unless otherwise
instructed in other settings.

46 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

QA Mode settings - PrB05


Parameter Range (Default) Notes
QA-mode None/ None = IQC (Integrated Quality
Exp (Exp) Control) OFF
Exp = IQC ON
Default kV 23 to 35 (28) The kV used in IQC/Exposure mode.
Default Filter Mo/Rh/QA (Mo) The filter used in IQC/Exposure
mode.

Preprog1-4 settings - PrB06-9


Parameter Range (Default) Notes
Density -7 to +7 (0) Density correction for each pre-
programmable cassette holder/ filter
combination. One step corresponds to
10% change in exposure time.
Target Time 1/2 (1) The selection of the target time curve
that Diamond interpolates to attain
the required target time.
Auto kV -3 to +3 (0) The extent the kV is allowed to
Initial kV change during exposure from the
initial kV setting (which is
determined by the compression
thickness). Applied change affects
the whole initial kV table.
Menu Text PROGR 1/2/3/4 A PreProg mode can be customized
PreProg1 with 8 characters available at
maximum.
kV Change 0 to 5 (3) The limit how much Diamond can
Limit automatically change kV from the
initial kV after the exposure has
started. Often set at 3.
AutoPnt 0.50 - 1.00 (1.00) PreProg specific AutoPoint
Coefficient coefficient. Refer to chapter 4.2.4, D:
AEC for more coefficient
information.

4.2.3 C: Diagnostics - NOT USED

The service program C: Diagnostics is not used at the moment.

86502-4AA Instrumentarium Imaging 4-7


4 Parameter programming and software load

4.2.4 D: AEC

This service program is used to set and adjust AEC default parameters.

NOTE
Prior to any AEC work and testing, set target times and make sure that kVs
are OK.

In service program D:AEC/Target times (PrD05) the min/max values for


target time and target time limits can be set. The target time is set up by
giving two thicknesses (2cm and 6cm) and the respective desired exposure
times. The target times for other thicknesses are calculated with linear
estimation or are based on the limits set above. The system selects the kV so
that the desired target time can be reached. The maximum allowable kV-
change is more limiting, however, and thus a different exposure time than the
target time may result.

Functional description of AEC full automatic mode

The initial kV is automatically selected based on the compressed breast


thickness. In the beginning of the exposure the first estimation of the
exposure time at initial kV is made by measuring the radiation coming
through the breast. Based on this, the kV is automatically changed up or
down, if necessary, to meet the target time. If the kV is changed, a second
estimation of the exposure time at this kV is made by measuring the radiation
and the kV is automatically changed again, if necessary, to meet the target
time. After the exposure the final kV is displayed.

The Auto kV performance can be optimized and configured according to the


user preferences. The configurable items are the initial kV (initial kV versus
breast thickness table), amount of allowable kV change (Max kV change in
service program D:AEC/Optical Densities (PrD03)) and desired exposure
time range (target time and target time limits).

Thickness compensation adjustment

Diamond uses a thickness compensation method, which compensates for


variations in optical density (O.D) of the film. The rotation point can be
programmed. The amount of rotation and compensation curve slopes below
and above the rotation point can be programmed separately.

Rotation point P

Rotation point is the exposure time where the O.D is adjusted correctly.
Thickness compensation is achieved by altering the exposure time around
this point. Default value is 1.00.

Amount of rotation r1 and r2

Parameters r1 and r2 determine the amount of compensation with exposure


times shorter or longer than the rotation point respectively. r1 determines the
amount of compensation at the point where exposure time is half the rotation
point. r2 determines the amount of compensation at exposure time twice the
rotation point.

48 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

The values of r1 and r2 can vary from 0.50 to 2.00. If the parameter is 1.30 it
means that exposure times are increased by 30%. If the parameter is 0.80 it
means that the exposure times are decreased by 20% etc. Default values are
1.0 for both.

Compensation curve slope parameters S1 and S2

Parameters S1 and S2 determine the curve slope for exposure times shorter or
longer than the rotation point respectively. S1 determines the slope at the
point where exposure time is lower than the rotation point. S2 determines the
amount of compensation at exposure time higher than the rotation point.
Refer to the curves below for effects of the S parameters. The rotation point is
1.0 seconds and the amount of rotation (r1 and r2) is 0.8, 0.9, 1.1 or 1.2 (=
80%, 90%, 110%, 120%). Default values are 2500 for both.
S=0
1.4 1.4
S=0
S = 7000
1.3 1.3
r1 = 1.2
1.2 1.2
S = 7000 r1 = 1.1
1.1 1.1

P = 1.0sec P = 1.0sec
1 1

0.9 0.9
S = 7000
r1 = 0.9
0.8 0.8
r1 = 0.8
0.7 0.7
S=0 S = 7000 S=0
0.6 0.6
0 0.2 0.4 0.6 0.8 1 0 0.2 0.4 0.6 0.8 1
Exposure time / sec. Exposure time / sec.
Fig 4.1. Effect of parameter S1
1.4 1.4
S = 7000
S = 7000
S = 6000
1.3 S = 5000 1.3 S=0
S = 3500 r2 = 1.2
S = 2000
1.2 S = 1000 1.2
r2 = 1.1 S = 500
1.1 S=0 1.1
P = 1.0sec P = 1.0sec
1 1

0.9 0.9
r2 = 0.9 S=0
S = 500
0.8 S = 1000 0.8
S = 2000
S = 3500 r2 = 0.8 S=0
0.7 S = 5000 0.7
S = 6000
0.6
S = 7000 S = 7000
0.6
1 1.5 2 2.5 3 3.5 4 4.5 5 1 1.5 2 2.5 3 3.5 4 4.5 5
Exposure time / sec. Exposure time / sec.

Fig 4.2. Effect of parameter S2

86502-4AA Instrumentarium Imaging 4-9


4 Parameter programming and software load

AutoPoint

AutoPoint feature to select the most appropriate detector(s) and calculate


exposure time is useful in following cases:

– consistent selection of the detector based on actual breast density, not


on visual appearance
– more homogeneous optical densities all over breast image are
generated
– when small breasts are examined
– to compensate for very dense (e.g. muscle) area, if it happens to lay
under the detector selected by Auto Point mode

With software 1.02.1 and up, the AutoPoint-feature uses two detectors (if so
desired). The selection of the detector is adjusted through the Auto Point
Coefficient. Software versions 1.01.3 and older have always used one
detector for this measurement.

NOTE
If detector selection (AutoPoint coefficient < 1.0) is used to increase density
of less dense sections of the image, it will also increase the density of the
most dense areas of the breast.

Parameter settings
A AutoPoint limit value

The Auto Point limit value can be set in service program PrD01 within range
of 50–2500. As a hardware default, the AutoPoint feature is enabled. It can be
disabled with AutoPoint-button in the Position Control Center. The indicator
light switches off when AutoPoint is disabled.

The limit adjusts the behavior of the unit on selection of the detector. High
values will make the unit less likely to select a detector under pectoral muscle
(near the chest wall edge), but may select a detector too far if the dense area is
located under C1. Low values may cause the unit to select the detector
covered by pectoral muscle. Value of 700 has proved to be a good
compromise, but this may depend on the local practices (exposure values,
positioning, population).

With software 1.02.1, the default Auto Point limit value stands at 700. In
earlier software versions the range for this value was 700-1500. For small
breasts 700 (default) is recommended. If images are too dark when exposing
at medio-lateral projection, a detector under muscle might be in use (solid
ellipses in the image). Increase the AutoPoint limit value.

B AutoPoint Coefficient

The AutoPoint Coefficient for each pre-programmed mode can be set in


service programs PrB06-09 within range of 0.5–1.0.

After Diamond has selected the primary detector, it can be used alone to
determine the required exposure or together with a secondary detector. The
coefficient is a figure that defines what percentage of the measurement is
achieved with the primary detector. Value 1.0 corresponds to 100% use of the
primary detector.

4 10 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Generally a value other than 1.0 (most often 0.7) is recommended only if
users would object to exposing a small dense area of the breast to target O.D.

From software 1.02.1 on, the default AutoPoint Coefficient is 1.0.

Coefficient range Amount measured Amount measured


with the primary with the secondary
detector detector
1.0 100% 0%
0.9 90% 10%
0.8 80% 20%
0.7 70% 30%
0.6 60% 40%
0.5 50% 50%

PaddleLogic

Diamond PaddleLogic feature enables adjustment of the collimation and the


available AutoPoint detectors separately for each paddle (image below for
reference). For a small biopsy paddle the collimation (gray rectangle) can be
adjusted to allow the exposure only to the region of interest: the coordinate
system and the volume available for biopsy. Also the AEC detectors allowed
for AutoPoint can be limited in order to block out detector(s) that is/are
completely outside of metal frame of a paddle of small size. PaddleLogic
detector defaults by paddles are listed at the end of BatteryRAM parameter
defaults in the Diamond Installation Instructions (86501), Appendix section.

Ask whether the customer wants to collimate according to the paddle or to


cassette/paddle types and set the collimation up as wished.

NOTE
Diamond software does not allow the use of the detectors that are not
completely within the collimated X-ray field.

86502-4AA Instrumentarium Imaging 4-11


4 Parameter programming and software load

Exact collimation

Diamond detects the paddle and the cassette holder in use. The field size with
different paddle/cassette holder combinations can be checked with the field
light.

Paddle related collimation can be adjusted in service programs PrF01 for rear
plates and in service program PrF02 for side plates. Adjust step values
according to paddle size.

NOTE
If the paddle and the cassette holder suggest different X-ray field sizes,
Diamond will always select the smaller field size.

Detector selection by paddle type

Diamond can select most usable detectors according to the paddle type.
Detectors used with the paddle can be adjusted in service program PrD09,
Paddle Logic. For example certain detectors that are not completely under
spot paddles can be eliminated from use.

Perform required changes in all groups:

AEC Setup - PrD01


Parameter Range (Default) Notes
Target O.D. 1.00 to 2.00 Adjustable O.D. range.
Target time limit 0.20 to 1.00 Lowest target time, affects service
low program D05.
Target time limit 1.00 to 3.00 Highest target time, affects service
high program D05.
AutoPoint limit 50 to 2500 (700) Refer to chapter 4.2.4, D: AEC.
Low mA limit 0-4 Thickness threshold below which the
cm low mA table is used. Small breasts
might cause too short exposure times
and to overcome lower mA values
can be used below certain thickness.

4 12 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

AEC gains - PrD02


Parameter Range (Default) Notes
AEC gain -’g’ 0.5 to 2.0 Basic software gain to obtain the
required optical density. This gain
affects all detectors.
Detector L1 0.7 to 1.30 Detector specific gain factor that
(Det. # 0) Diamond adjusts in relation to
detector C1 performance. Values less
Detector L2 0.7 to 1.30
than 1.0 increase gain and produce
(Det. #3) darker images. Values over 1.0
Detector R1 0.7 to 1.30 decrease gain and produce lighter
(Det. #2) images.
Detector R2 0.7 to 1.30
(Det. #7)
Detector C1 0.7 to 1.30 BENCHMARK DETECTOR. DO
(Det. #1) NOT CHANGE THIS VALUE.

Detector C2 0.7 to 1.30 Detector specific gain factor that


(Det. #6) Diamond adjusts in relation to
detector C1 performance. Values less
Detector C3 0.7 to 1.30
than 1.0 increase gain and produce
(Det. #5) darker images. Values over 1.0
Detector C4 0.7 to 1.30 decrease gain and produce lighter
(Det. #4) images.

4=C4 80%

5=C3 90%
7=

98
L2

%
R2

6=C2 102%
3=

98

0=L1 1=C1 2=R1 97% 100% 97%

The graph above describes how various detectors receive radiation. In order
to equalize them, detector specific factors are used for adjustment. Detector
C1 is used as a benchmark.

Optical Densities - PrD03


Parameter sub-screens Notes
Auto kV Optical density settings for Auto kV.
Auto Time Optical density settings for Auto Time.
Thickness Optical density settings for Thickness
Compensation compensation.
kV Compensation Optical density settings for kV compensation.
Density step Adjustment of density steps.

86502-4AA Instrumentarium Imaging 4-13


4 Parameter programming and software load

Optical Densities/Auto kV sub-screen - PrD03


Parameter Range (Default) Notes
Auto kV Bucky Adjustment when using the Bucky.
Mo -7.0 to +7.0 (0) Increases or decreases the optical
density of the image at all kV values.
Rh -7.0 to +7.0 (0)
A density step here is 50% (10%
Al -7.0 to +7.0 (0) from sw. 1.02.6 and up) of the regular
step available to the user.
Auto kV Adjustment for the use of Magnification tunnel.
Magnification
Mo -7.0 to +7.0 (0) Increases or decreases the optical
density of the image at all kV values.
Rh -7.0 to +7.0 (0)
A density step here is 50% (10%
Al -7.0 to +7.0 (0) from sw. 1.02.6 and up) of the regular
step available to the user.
Auto kV Contact Adjustment using the Contact tunnel. NOT USED!
Mo -7.0 to +7.0 (0)
Rh -7.0 to +7.0 (0)
Al -7.0 to +7.0 (0)

Optical Densities/Auto Time sub-screen - PrD03


Parameter Range (Default) Notes
Auto Time Adjustment when using the Bucky.
Bucky
Mo -7.0 to +7.0 (0) Increases or decreases the overall
optical density of the image. A
Rh -7.0 to +7.0 (0)
density step here is 50% (10% from
Al -7.0 to +7.0 (0) sw. 1.02.6 and up) of the regular step
available to the user.
Auto Time Adjustment when using the Magnification tunnel.
Magnification
Mo -7.0 to +7.0 (0) Increases or decreases the overall
optical density of the image. A
Rh -7.0 to +7.0 (0)
density step here is 50% (10% from
Al -7.0 to +7.0 (0) sw. 1.02.6 and up) of the regular step
available to the user.
Auto Time Adjustment when using the Contact. NOT USED!
Contact
Mo -7.0 to +7.0 (0)
Rh -7.0 to +7.0 (0)
Al -7.0 to +7.0 (0)

4 14 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Optical Densities/Thickness Compensation sub-screen - PrD03


Parameter Range (Default) Notes
These parameters have separate settings for
both Molybdenum and Rhodium filters.
Bucky Adjustment when using the Bucky.
‘P’ 0.30 - 2.00 (1) Rotation point of compensation curve
(Exposure time in seconds where
required optical density was
received).
‘r1’ 0.50 - 2.00 (1) Compensation factor for exposure
times shorter than P.
‘r2’ 0.50 - 2.00 (1) Compensation factor for exposure
times longer than P.
‘S1’ 0 - 7000 (2500) Compensation parameter for very
short exposure times (fine-tuning).
‘S2’ 0 - 7000 (2500) Compensation parameter for very
long exposure times (fine-tuning).
Magnification Adjustment when using the
Magnification tunnel.
‘P’ 0.30 - 2.00 (1) Rotation point of compensation curve
(Exposure time in seconds where
required optical density was
received).
‘r1’ 0.50 - 2.00 (1) Compensation factor for exposure
times shorter than P.
‘r2’ 0.50 - 2.00 (1) Compensation factor for exposure
times longer than P.
‘S1’ 0 - 7000 (2500) Compensation parameter for very
short exposure times (fine-tuning).
‘S2’ 0 - 7000 (2500) Compensation parameter for very
long exposure times (fine-tuning).
Contact Adjustment when using the Contact.
NOT USED!
‘P’ 0.30 - 2.00 Rotation point of compensation curve
(Exposure time in seconds where
required optical density was
received). NOT USED!
‘r1’ 0.50 - 2.00 Compensation factor for exposure
times shorter than P. NOT USED!
‘r2’ 0.50 - 2.00 Compensation factor for exposure
times longer than P. NOT USED!
‘S1’ 0 - 7000 Compensation parameter for very
short exposure times (fine-tuning).
NOT USED!
‘S2’ 0 - 7000 Compensation parameter for very
long exposure times (fine-tuning).
NOT USED!

86502-4AA Instrumentarium Imaging 4-15


4 Parameter programming and software load

Optical Densities/kV Compensation sub-screen - PrD03


Parameter Range (Default) Notes
kV Adjustment for each kV with Bucky.
Compensation kV range for Mo and Rh: 20-35.
Bucky kV range for Al: 36-39.
Mo 0.70 - 1.30 (1.00) If image at certain kV is too light or
dark, adjust this kV-specific value.
Rh 0.70 - 1.30 (1.00)
Al 0.70 - 1.30 (1.00)
kV Adjustment for each kV with Magnification. kV range
Compensation 20-35.
Magnification
Mo 0.70 - 1.30 (1.00) If image at certain kV is too light or
dark, adjust this kV-specific value.
Rh 0.70 - 1.30 (1.00)
Al 0.70 - 1.30 (1.00)
kV Adjustment for each kV with Contact. kV range 20-35.
Compensation NOT USED!
Contact
Mo 0.70 - 1.30
Rh 0.70 - 1.30
Al 0.70 - 1.30

Optical Densities/Density step - PrD03


Parameter Range (Default) Notes
Density step % 5 - 20% (10) Adjusts the extent of change for Den-
or Den + buttons.

Optical Densities/Detector Compensation sub-screen - PrD04 - NOT


USED at the moment
Parameter Range (Default) Notes
Detectors 1-8 0.5 - 2.0 (1) Adjustment of the individual detector
signal output level. Values below 1
amplify the signal and value over 1
diminish it

4 16 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Target Times programming/Target Time 1 sub-screen - PrD05


Parameter Range (Default) Notes
Target Time T1 Settings for Target Time 1 curve
Bucky Bucky settings specified
Mo Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Rh Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Magnification Magnification settings specified
Mo Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Rh Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Contact Contact settings specified NOT USED!
Mo Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Rh Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2

86502-4AA Instrumentarium Imaging 4-17


4 Parameter programming and software load

Target Times programming/Target Time 2 sub-screen - PrD05


Parameter Range (Default) Notes
Target Time T2 Settings for Target Time 2 curve
Bucky Bucky settings specified
Mo Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Rh Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Magnification Magnification settings specified
Mo Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Rh Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Contact Contact settings specified NOT USED!
Mo Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
Rh Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2

4 18 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Initial kV tables - PrD06


Parameter Range Notes
(Default)
kV values for
Bucky Mo
Thickness range 0.5 - 10 cm Scroll through thicknesses and
(at 0.5 cm intervals) change initial kV values for each
thickness.
Adjustment range 23-35 kV
(25 -30)

kV values for
Bucky Rh
Thickness range 0.5 - 10 cm Scroll through thicknesses and
(at 0.5 cm intervals) change initial kV values for each
thickness.
Adjustment range 23-35 kV
(25 -30)

kV values for
Magnification Mo
Thickness range 0.5 - 10 cm Scroll through thicknesses and
(at 0.5 cm intervals) change initial kV values for each
thickness.
Adjustment range 23-35 kV
(25 -30)

kV values for
Magnification Rh
Thickness range 0.5 - 10 cm Scroll through thicknesses and
(at 0.5 cm intervals) change initial kV values for each
thickness.
Adjustment range 23-35 kV
(25 -30)

QA Coefficient history - PrD07


Parameter Value Notes
QA coeff# Value of the past QA coefficient.

Detector Adjustment - PrD08


Parameter Range Notes
New gains based No/Yes Adjusts the gains for individual
on recent detectors in relation to the C1
exposure. detector being 1.0.

86502-4AA Instrumentarium Imaging 4-19


4 Parameter programming and software load

PaddleLogic - subscreen PrD09


Parameter Range Notes
Select paddle, press Accept, then detector parameters are listed.
DetectorL1 On/Off Enabled (On)/Disabled (Off) with the
selected paddle.
DetectorC1 On/Off
DetectorR1 On/Off
DetectorL2 On/Off
DetectorC4 On/Off
DetectorC3 On/Off
DetectorC2 On/Off
DetectorR2 On/Off

4.2.5 E: C-Arm Rotation

This service program is used to set and adjust C-Arm rotation default
parameters. Perform required calibration and changes in all groups:

Angle Calibration - PrE01


Parameter Notes
180°-calibration Calibration of the c-arm angle.

1 Drive the C-Arm manually to 180°-angle (upside down, real angles)


with manual rotation keys.
2 Press Accept-key.
3 The C-Arm drives to 0°-calibration.

If the calibration is successfully performed, the LCD screen then indicates


‘Pass’. In case the calibration was not successful, the LCD screen then
indicates ‘Fail’.

1 Repeat the calibration procedure.


2 If the calibration still was not successful, drive the c-arm to from 180° to
0° and then to 180° on the other side. When c-arm passes the 0° it helps
to calibrate.

Rotation speed - PrE02


Parameter Range (Default) Notes
Offset 0 - 3 (0) C-Arm rotation speed can increased
(affects all values in C-Arm speed
table). Use value 3 for units with S/N
20116 and up. Note that the rotation
speed slows down during approach to
target angle.
Value 0 is recommended for
Diamond units S/N 20000 - 20115
with faster C-Arm transmission (if
still the original factory installation).

4 20 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Projections - PrE03
Parameter Range (Default) Notes
Oblique 0- 359° Angle setting for the left OBL button.
( 225 software If set at 135°, it operates as in image
degrees, at 0° the A. If set at 225°, it operates as in
tube is upside image B. The right OBL drives to
down) mirror angle.
Lateral 0 - 359° (270 Angle setting for the left LAT button.
software degrees) The right LAT drives to mirror angle.
Mirror angle On/Off (On) When an exposure is made at 15-75°
C-Arm protection angle, Diamond
software stores the angle for a certain
period of time. If the opposite OBL-
button is pressed during this period,
Diamond drives the C-Arm past the
CC-position to an equivalent to an
equivalent angle on the opposite side
of the unit.

If the OBL-button is pressed second


time, the C-Arm runs to the
programmed angle (default 45°).

The mirror angle will remain in


memory until predetermined time has
elapsed (default 60 seconds) or an
new exposure is made (at 15-75°).

Mirror angle timeout adjustment

If the timeout should be changed from 60 (default) to 90 seconds, type in at


HyperTerminal command prompt: bram 0x40832 90

Check the current timeout value with bram 0x40832

NOTE
The mirror angle feature works only when the obliques are done one after the
other. If any views are done between the two obliques the system reverts to its
default value.

NOTE
This automatic drive to pre-defined angle is not allowed if the compression
arm is not at the upper microswitch.

86502-4AA Instrumentarium Imaging 4-21


4 Parameter programming and software load

Auto Drive - PrE04


Parameter Range (Default) Notes
Automatic drive On/Off/Pdl_up If On is selected, CC, OBL or LAT-
(On) keys and the c-arm will move to the
required angle. If Off is selected, the
key must be pressed on until the
required angle is reached. If Pdl_up
is selected, then the AutoDrive
operates only after the the
compression arm has been driven to
home position (upper microswitch).

After insertion of a paddle, Diamond checks the encoding for the paddle and
the C-Arm autodrive setting in bram. The Autodrive is enabled or disabled
according to the inserted paddle.

NOTE
C-Arm AutoDrive-feature is disabled if no paddle has been attached to
Diamond. Connect paddle before driving the C-Arm. Remind also the
customer about this feature.

4 22 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

4.2.6 F: Collimator

This service program is used to set and adjust collimator default parameters.
Perform required calibration and changes in all groups:

Rear Plates - PrF01


Parameter Range (Default) Notes
Front plate 0 - 50000 (39002) The number of steps that the motor
moves the plate. A 200-step change
equals to about 1mm change on the
film. Adjust using 1-step interval.
Bucky 18x24 0 - 50000 (9500)
Bucky 24x30 0 - 50000 (1000)
Magnification 0 - 50000 (9500)
1.6
Unknown 0 - 50000 (10000) NOT USED!
Contact 18x24 0 - 50000 (9500) NOT USED!
Contact 24x30 0 - 50000 (1000) NOT USED!
Magnification 0 - 50000 (9500)
1.8
Spot 0 - 50000 (9500)
Magnification
1.6
Stereo digital 0 - 50000 (27000)
Stereo film 0 - 50000 (10000)
Magnification 0 - 50000 (9500)
2.0
reserved 0 - 50000 (10000) Additional plate may be added.
reserved 0 - 50000 (10000) Additional plate may be added.
reserved 0 - 50000 (10000) Additional plate may be added.
reserved 0 - 50000 (10000) Additional plate may be added.

Rear Plates - PrF01


Parameter Range (Default) Notes
No paddle 0 - 50000 (500)
Stand small 0 - 50000 (1000)
Stand large 0 - 50000 (1000)
Mag spot 0 - 50000 (10550)
Stand Mag 0 - 50000 (9500)
Implant 0 - 50000 (19600)
Mag Implant 0 - 50000 (8700)
Delta32 0 - 50000 (25000)

86502-4AA Instrumentarium Imaging 4-23


4 Parameter programming and software load

Rear Plates - PrF01


Parameter Range (Default) Notes
Spot 0 - 50000 (23750)
Mag spt enh 0 - 50000 (7450)
Biopsy D 0 - 50000 (9500)
Biopsy SQ 0 - 50000 (5750)
Spot enhanc 0 - 50000 (16850)
Biopsy spot 0 - 50000 (10550)
Reserved14 0 - 50000 (500)
Reserved15 0 - 50000 (500)

Side Plates - PrF02


Parameter Range (Default) Notes
Front plate 0 - 50000 (0) NOT USED.
Bucky 18x24 0 - 50000 (4800) The number of steps that the motor
moves the plate. A 200-step change
equals to about 1mm change on the
film. Adjust using 1-step interval. If
the value is too big, motor back up
time is exceeded and an error
message is displayed.

Bucky 24x30 0 - 50000 (300)


Magnification 0 - 50000 (4800) These values should be the same in
1.6 order for collimator not to start
driving to home position.
Unknown 0 - 50000 (4100)
Contact 18x24 0 - 50000 (4800) NOT USED!
Contact 24x30 0 - 50000 (300) NOT USED!
Magnification 0 - 50000 (4800) These values should be the same in
1.8 order for collimator not to start
driving to home position.
Spot 0 - 50000 (4800)
Magnification
1.6
Stereo digital 0 - 50000 (17500)
Stereo film 0 - 50000 (4100)
Magnification 0 - 50000 (4800) These values should be the same in
2.0 order for collimator not to start
driving to home position.
reserved 0 - 50000 (4100) Additional plate may be added.
reserved 0 - 50000 (4100) Additional plate may be added.
reserved 0 - 50000 (4100) Additional plate may be added.

4 24 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Side Plates - PrF02


Parameter Range (Default) Notes
reserved 0 - 50000 (4100) Additional plate may be added.

Side Plates - PrF02


Parameter Range (Default) Notes
No paddle 0 - 50000 (100)
Stand small 0 - 50000 (300)
Stand large 0 - 50000 (300)
Mag spot 0 - 50000 (5950)
Stand Mag 0 - 50000 (4800)
Implant 0 - 50000 (3450)
Mag Implant 0 - 50000 (1500)
Delta32 0 - 50000 (15000)
Spot 0 - 50000 (13450)
Mag spt enh 0 - 50000 (5300)
Biopsy D 0 - 50000 (4800)
Biopsy SQ 0 - 50000 (10600) Additional plate may be added.
Spot enhanc 0 - 50000 (12000) Additional plate may be added.
Biopsy spot 0 - 50000 (5950) Additional plate may be added.
Reserved14 0 - 50000 (500) Additional plate may be added.
Reserved15 0 - 50000 (500) Additional plate may be added.

Collimator Light - PrF03


Parameter Range (Default) Notes
Intensity 0 - 10 (7) Adjustment of the light intensity
Time normal 20 - 60 sec (40) Time that the lamp is lighted up
during a normal procedure.
Time 20 - 90 sec (80) Time that the lamp is lighted up
Stereo/Biopsy during a Stereo/Biopsy procedure.

86502-4AA Instrumentarium Imaging 4-25


4 Parameter programming and software load

4.2.7 G: Compression

This service program is used to set and adjust compression default


parameters. Perform required calibration and changes in all groups:

Compression Setup - PrG01


Parameter Range (Default) Notes
Slow speed 2 - 7 (4) Slow compression arm speed. Do not
use higher values than 5. If the speed
exceeds 5, the synchronization with
ECS movement may be disturbed.
High speed 2 - 7 (6) High compression arm speed.
Release speed 1 - 7 (5) The speed at which Diamond release
the compression arm in the event of
an emergency or a power loss.
AutoRelease 0 - 20 cm (15) The distance the Auto Release
travel function moves the compression
moves after an exposure.
Speed Threshold 0 - 150N (50) The compression force at which the
compression arm starts to run at slow
speed.

Paddle Calibration - PrG02


Parameter Range (Default) Notes
Base point -9.5 ... +9.5 (0.0) Base point adjustment that affects all
paddles.

4 26 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Paddle Calibration - PrG02


Parameter Range (Default) Notes
Standard small -9.5 ... +9.5 (-0.5) Compensation for paddle flexibility.
Check for country-specific
Standard large -9.5 ... +9.5 (-0.5) requirements.
Magnification -9.5 ... +9.5 (-4.0)
spot
Standard -9.5 ... +9.5 (-4.0)
magnification
Implant -9.5 ... +9.5 (-7.0)
Magnification -9.5 ... +9.5 (-9.0)
implant
Delta 32 -9.5 ... +9.5 (-3.5)
Spot -9.5 ... +9.5 (-0.5)
Mag spot -9.5 ... +9.5 (-9.0)
enhanced
Biopsy D -9.5 ... +9.5 (0.0)
Biopsy SQ -9.5 ... +9.5 (0.0)
Spot enhanced -9.5 ... +9.5 (-5.0)
Biopsy Spot -9.5 ... +9.5 (0.0)
Reserved14 -9.5 ... +9.5 (0.0)
Reserved15 -9.5 ... +9.5 (0.0)

Cassette Holder Calibration - PrG03


Parameter Range (Default) Notes
No holder -9.5 ... +9.5 (0.0) Base point adjustment that affects all
holders.
Bucky 18x24 -9.5 ... +9.5 (0.0) Adjustment for an individual cassette
holder.
Bucky 24x30 -9.5 ... +9.5 (0.0)
Magnification -9.5 ... +9.5 (0.0)
1.6
Unknown -9.5 ... +9.5 (0.0)
Contact 18x24 -9.5 ... +9.5 (0.0) NOT USED!
Contact 24x30 -9.5 ... +9.5 (0.0) NOT USED!

86502-4AA Instrumentarium Imaging 4-27


4 Parameter programming and software load

Cassette Holder Calibration - PrG03


Parameter Range (Default) Notes
Mag 1.8 -9.5 ... +9.5 (-4.5) Adjustment for an individual cassette
holder.
Spot -9.5 ... +9.5 (0.0)
Magnification
1.6
Stereo digital -9.5 ... +9.5 (-6.0)
Stereo film -9.5 ... +9.5 (0.0)
Magnification -9.5 ... +9.5 (-8.0)
2.0
reserved -9.5 ... +9.5 (0.0)
reserved -9.5 ... +9.5 (0.0)
reserved -9.5 ... +9.5 (0.0)
reserved -9.5 ... +9.5 (0.0)

Compression Force Calibration - PrG04


Parameter Value Notes
Force to 0 N 0N Apply 0N force and accept.
Force to 200 N 200 N Apply 200 N with a calibrated device
and accept if the screen displays
correct figure.

Compression Force Limits - PrG05


Parameter Range (Default) Notes
Mode 0/1 (0) 0=Compression cannot be increased
with the foot pedal.
1=Compression can be increased
with the foot pedal until the
maximum limit.
Position 1 0 - 255 N (70) Compression force stops when this
limit is reached.
Position 2 0 - 255 N (90)
Position 3 0 - 255 N (110)
Position 4 0 - 255 N (130)
Position 5 0 - 255 N (150)
Position 6 0 - 255 N (170)

Negative Compression Force Calibration/Up force limit - PrG06


Parameter Value Notes
Apply weight Apply force and accept.

4 28 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

Negative Compression Force Calibration/Up force limit - PrG06


Parameter Value Notes
Adjust force 0 - 10 Adjust for negative compression
force required for the stop reaction.

In case the compression paddle is


driven upwards and something above or
on compression paddle hinders its
movement, negative compression force
affects the paddle and it bends down.

Chapter 7.6, Negative compression


force calibration, contains instructions
for negative compression force
calibration.

4.2.8 H: Filter

This service program is used to set and adjust filter default parameters.
Perform required calibration and changes in all groups:

Auto Filter Setup - PrH01


Parameter Range (Default) Notes
Change thick- 1 - 15 cm (5) Above the range, the Rhodium is
ness used. Below the range, the
Molybdenum filter is used.
AutoFilter On/Off (Off) Automatic Filter selection. AutoFilter
is always disabled when Delta 32 is
attached.

Filter calibration - PrH02


Parameter Range (Default) Notes
Molybdenum 0 - 20000 (1950) The number of steps (from home
position) applied for this filter
Rhodium 0 - 20000 (150)
selection. A four-step change moves
Aluminum 0 - 20000 (3150) the filter about 1,3 mm. Adjust using
4-step intervals (4,8,12...). Calibrate
the filter from the front edge of the
collimator using small focus
(magnification tunnel).

86502-4AA Instrumentarium Imaging 4-29


4 Parameter programming and software load

4.2.9 I: Generator

This service program is used to set and adjust generator default parameters.
Perform required calibration and changes in all groups:

Generator Setup - PrI01, NOT USED at the moment


Parameter Range (Default) Notes
Anode Brake On/Off Anode brake operation regulation.

PreHeat adjustment - large focus (LF) -- PrI02,


NOT USED at the moment.
Parameter Range (Default) Notes
1 Press Exposure button to start automatic PreHeat calibration for large
focus.
2 If the new PreHeat value is valid and the LCD screen displays ‘Pass’,
Diamond returns to the Generator screen.
3 If the calibration was not successful, Diamond returns to the calibration
screen.

PreHeat adjustment - small focus (SF) - PrI03,


NOT USED at the moment.
Parameter Range (Default) Notes
1 Press Exposure button to start automatic PreHeat calibration for large
focus.
2 If the new PreHeat value is valid and the LCD screen displays ‘Pass’,
Diamond returns to the Generator screen.
3 If the calibration was not successful, Diamond returns to the calibration
screen.

PreHeat Compensation - large focus (LF) - PrI04


Parameter Range (Default) Notes
PreHeat LF -5...5 (2) Large focus preheat adjustment that
affects all kVs.
LF 15 - 39 20 - 235 (48-57) Large focus preheat adjustment for
each kV. Typical voltage steps are 54-
56.

PreHeat Compensation - large focus (LF) - PrI04 WITH LOW MA!


Parameter Range (Default) Notes
PreHeat LF -5...5 (0) Large focus preheat adjustment that
affects all kVs.
LF 15 - 39 20 - 235 (47-52) Large focus preheat adjustment for
each kV when low mA values are
used. Adjust these if low mA values
are used.

4 30 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

PreHeat Compensation - small focus (LF) - PrI05


Parameter Range (Default) Notes
PreHeat SF -5...5 (1) Small focus preheat adjustment that
affects all kVs.
SF 15 - 39 20- 235 (48-50) Small focus preheat adjustment for
each kV.

4.2.10 J: ParkBack

This service program is used to set and adjust tube park default parameters.
Perform required calibration and changes in all groups:

Tube speeds tube movement - PrJ01 NOT USED!


Parameter Range (Default) Notes
Slow speed 1 - 5 (2) Slow speed adjustment for the
moving tube head.
Fast speed 1 - 5 (2) Fast speed adjustment for the moving
tube head.

Auto Drive - PrJ02


Parameter Range (Default) Notes
Exp drive On/Off (On) When enabled, the tube head moves
forwards and the collimation field
light activates after the exposure is
initiated. If the exposure button is
released prematurely, the exposure is
discontinued. Automatic drive is
disabled when Delta 32 or biopsy
paddles are attached to Diamond.
AutoPark On/Off (Off) When enabled, the tube head moves
backward after the exposure if the
projection angle is more than 15°.
The automatic drive is disabled if the
projection angle is less than 15°, or if
a biopsy paddle is attached, or if no
paddle is attached.

In order to tighten the drive belt, the tube head motor drives a bit further after
reaching the front position.

86502-4AA Instrumentarium Imaging 4-31


4 Parameter programming and software load

4.2.11 K: Stereo Movement

This service program is used to set and adjust stereo movement default
parameters. Perform required calibration and changes in all groups:

Stereo Movement Setup - PrK01


Parameter Range (Default) Notes
Low speed limit 1 - 4 (2) Low speed adjustment.
High speed 4 - 7 (5) High speed adjustment.
Speed offset -2 ... +2 (0) Power boost for c-arm. Do not
change before consulting with the
manufacturer.
Braking speed 2 - 4 (3) Slow down speed when required is
being approached. Do not change
before consulting with the
manufacturer.
Ramp in 10ms 5 - 50 (25) Acceleration of ramp speed. Do not
change before consulting with the
manufacturer.

Stereo Movement Setup - PrK02


Parameter Range (Default) Notes
Auto drive On / Off If the function is selected (On), one
press is enough to drive to the stereo
angle. If it is not selected (Off), the
stereo drive buttons have to be
pressed down all the time until the
stereo angle is reached.

Calibration of the stereo movement


1800-angle
1 Enter Service Mode to drive the stereo movement.
Column

2 Type in command CAL 1650 in HyperTerminal® program. Diamond 1800-150 1800+150


replies XOK. Stereo angle is reported as absolute 1/10 degrees. 0-point =1650 =1950
lies in the base plate being 90° in the left (front view). Left stereo angle
is calculated as follows: 1800-150=1650.
3 Drive the stereo all the way to the other limit without interruptions in α15° α15°
movement and type in command CAL 1950 (1800+150=1950).
4 Diamond replies XOK and beeps three times. The calibration is
complete. 90°
Stereo movement angles
5 If Diamond replies XNO when you type the value in the
® (front view)
HyperTerminal , it likely does not recognize the opto.

4 32 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

4.2.12 L: Physicist mode

The Physicist mode allows the testing of various settings and collimation
combinations. The Physicist mode collimation does not override the actual
holder. For example, if you have connected a small Bucky and selected large
Bucky and paddle in the Physicist-service program, the collimation is the one
defined for small Bucky.

Collimation works the same way as in PaddleLogic, a smaller field overrides


a larger one. Please note that the collimation settings in Physicist mode are
very useful to check adjustment of X-ray and light fields.

NOTE
All Physicist mode settings are restored to default when you exit the service
mode.

Perform required calibration and changes in all groups:

Physicist mode - PrL01


Parameter Range (Default) Notes
System is default for all.
Grid Off/System Disable/Enable grid movement for
mAs testing. Grid disable also with
Magnification.
Focus Large/Small/ Focus size selection. When small
System focus size is used, collimator drives
the front plate to front position and
with large to back position.
Paddle Any paddle/ Paddle selection. When System is
System enabled, Diamond collimates
according to a selected paddle. The
collimation can be checked with the
field light or with a film exposure.
PaddleLogic detectors are selected
based on physicist’s paddle choice.
Holder Any holder/ Cassette holder selection. When
System System is enabled, Diamond
collimates according to a selected
cassette holder. The collimation can
be checked with the field light or with
a film exposure. If no cassette holder
has been attached to Diamond, the
collimator plates are driven totally
open.
kV limit Off/System Disable/enable kV limit. If the limit is
disabled and manual mode selected,
any kV in the range 15 to 39 can be
used.

86502-4AA Instrumentarium Imaging 4-33


4 Parameter programming and software load

4.2.13 General software update information


4.2.13.1 Upgrade from 1.02.2 - 1.02.4 to 1.02.5
1 Prior to software upgrade from 1.02.2 - 1.02.4 to 1.02.5, check in the
service program PrB06-09, B: Defaults/PreProg Mode Settings for the
AutoPoint coefficient and record the values. Then upgrade the software
and replace the AutoPoint coefficient values with the old ones.
2 If QA-mode is used, check service program PrD07, QA-coefficient
history that all coefficients are zero. If they are not, press Set and Cancel
simultaneously to reset them zero. Then exit service menu and restart
Diamond.
3 Check and relicalibrate AEC.
4 Adjust initial kVs for Rh filter.
5 If possible, check preheat waveform kV by kV in range of 15-19 kV.
This range is new for Specimen mode. Refer to Diamond Installation
Instructions (86501) for preheat and mA-regulation check.

4.3 Hyperterminal interface for Diamond


The system settings can be exported in hexadecimal format and can be
viewed in Windows HyperTerminal®.

For this you need a PC, hyperterminal software such Windows


HyperTerminal® by Hilgraeve Inc., serial cable (9 pin male and female; lines
2,3 and 7 directly connected, cable code 85102).

NOTE
This can also be done with the optional Diamond Service Software (87502).

4.3.1 Configuration of the HyperTerminal:

The HyperTerminal interface varies a little bit in various Windows versions.

1 Select COM-port that you want to use at FILE\PROPERTIES\CON-


NECT TO-tab sheet:

4 34 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

2 Then proceed to Configure-tab sheet. Change the parameters to:

3 Accept changes with OK.

4.3.2 Data capture


1 Connect the cable between Diamond COM-port and PC serial port.
Switch off Diamond before this.
2 Create (with a Notepad or similar) a text file (for example
C:\Service\log.txt) to save the capture data.
3 Open up Windows HyperTerminal.
4 Select FILE\NEW CONNECTION
5 Switch on the Diamond unit.
6 When the connection is established, select TRANSFER\CAPTURE
TEXT... in Windows HyperTerminal. The following screen appears:

7 If the file in which you want to store data is correct, click on START. If
not, look for the file with Browse-function.
8 The connection between Diamond and the PC is alright, if the command
prompt looks as Dx> (x=software release version, for example, version
9 prompt looks like D9>). Perform service session.

Commands to retrieve system data:

bram - Diamond settings data (for default settings list, refer to


Diamond Installation Instructions (86501))
param aecdump 1, - AEC detector signal values (e.g. to adjust
AutoPointCoefficient or for other AutoPoint tuning work)

86502-4AA Instrumentarium Imaging 4-35


4 Parameter programming and software load

param aecdump 1
List item Meaning
*** Exp Nr:649 *** Exposure number
14:19 time
Preprog:1 Used preprog-mode
Detector(hex):0001 AEC-detector used
KvDspl(dec):28 kV of the exposure (in display)
IntTime(dec):118 Integration time
ActExpTime(dec):445 Real exposure time [ms]
Thickness(dec):40 Compression thickness [mm]
Filtered xraylevel(dtr_xray)[0](dec):1918 detector 0 signal
Filtered xraylevel(dtr_xray)[1](dec):2002 detector 1 signal
Filtered xraylevel(dtr_xray)[2](dec):1895 detector 2 signal
Filtered xraylevel(dtr_xray)[3](dec):1893 detector 3 signal
Filtered xraylevel(dtr_xray)[4](dec):1658 detector 4 signal
Filtered xraylevel(dtr_xray)[5](dec):1772 detector 5 signal
Filtered xraylevel(dtr_xray)[6](dec):2026 detector 6 signal
Filtered xraylevel(dtr_xray)[7](dec):1853 detector 7 signal
*** End Nr:649 *** End of data related to this
exposure number.

param aecdump 2 - displays Auto kV exposure times and kV


values to help problem tracing and to clarify the function of the
Auto kV, includes also param aecdump1 information, kV
compensation values, internal density, primary and secondary
detectors used by e.g. AutoPoint:

param aecdump 2
List item Meaning
*** Exp Nr:48 *** Exposure number
15:49 time
Detector(hex):0001 AEC-detector used
RotPoint(dec):30 P-point [ms]/10
RotL(dec):115 r1*100
RotR(dec):100 r2*100
SlopeL(dec):250 s1/10
SlopeR(dec):250 s2/10
ThComp(dec):100 Amount of thickness
compensation in this exposure
KvDspl(dec):28 kV of the exposure (in display)
KvComp(dec):100 Amount of kV compensation in
this exposure
IntTime(dec):126 Integration time
ActExpTime(dec):415 Real exposure time [ms]
Thickness(dec):40 compression thickness [mm]
Filtered xraylevel(dtr_xray)[0](dec):1838 detector 0 signal
Filtered xraylevel(dtr_xray)[1](dec):1866 detector 1 signal

4 36 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

param aecdump 2
List item Meaning
Filtered xraylevel(dtr_xray)[2](dec):1804 detector 2 signal
Filtered xraylevel(dtr_xray)[3](dec):1905 detector 3 signal
Filtered xraylevel(dtr_xray)[4](dec):1545 detector 4 signal
Filtered xraylevel(dtr_xray)[5](dec):1756 detector 5 signal
Filtered xraylevel(dtr_xray)[6](dec):1928 detector 6 signal
Filtered xraylevel(dtr_xray)[7](dec):1825 detector 7 signal
MAs(dec):35 mAs-value
User density=7 density-adjustment set by the user
(7 = 0, e.g. 5= -2)
Filter=0 Filter: 0=Molybdenum,
1=Rhodium, 2=Aluminum
Casholder=1 Cassette holder: 1= Small Bucky,
2= Large Bucky, 3-5
Magnification 1.6-2.0
Time before th.comp:415 Calculated exposure time before
thickness compensation [ms]
Det. offset(dtr_gndl)[0](dec):27 detector 0 offset
Det. offset(dtr_gndl)[1](dec):73 detector 1 offset
Det. offset(dtr_gndl)[2](dec):46 detector 2 offset
Det. offset(dtr_gndl)[3](dec):29 detector 3 offset
Det. offset(dtr_gndl)[4](dec):31 detector 4 offset
Det. offset(dtr_gndl)[5](dec):57 detector 5 offset
Det. offset(dtr_gndl)[6](dec):68 detector 6 offset
Det. offset(dtr_gndl)[7](dec):73 detector 7 offset
tecexptime[0]=0 Practically not relevant, this
exposure step does not exist, even
if some value other than 0 would
be displayed.
tecexptime[1]=415 Exposure times at different kV
values when Auto kV is used.
tecexptime[2]=0 With Auto Time only one step
used.
tecexptime[3]=0 If changes occur, the respective
kVs are listed.
Targettime=(dec):750 Target time at this thickness [ms].
Int. density(GetInternalDensity)(dec):100 Internal density e.g. Auto kV
adjustment in percentages.
Full density(dec):100 Density setting by the user in
percentages.
Autokvlimit=3 Maximum allowed kV-change in
Auto kV-mode.
Iqc coefficient=100 Correction initiated by IQC.
Dtr[0] values in offset cnt detector 0 offset monitoring
:28:28:28:28:27:28:28:28:28:28
Dtr[1] values in offset cnt detector 1 offset monitoring
:73:73:73:73:73:73:73:73:73:73

86502-4AA Instrumentarium Imaging 4-37


4 Parameter programming and software load

param aecdump 2
List item Meaning
Dtr[2] values in offset cnt detector 2 offset monitoring
:47:47:46:46:46:47:47:47:47:47
Dtr[3] values in offset cnt detector 3 offset monitoring
:29:29:29:29:29:29:29:29:29:29
Dtr[4] values in offset cnt detector 4 offset monitoring
:31:31:31:31:31:31:31:31:31:31
Dtr[5] values in offset cnt detector 5 offset monitoring
:57:57:57:57:57:57:57:57:57:57
Dtr[6] values in offset cnt detector 6 offset monitoring
:68:68:68:68:68:68:68:68:68:68
Dtr[7] values in offset cnt detector 7 offset monitoring
:74:74:73:73:74:74:74:74:74:74
*** End Nr:48 *** End of data related to this
exposure number.

aec paramdump 0 = ends listing of the param aecdump


bram = printing parameters, recommended after installation
9 When the service session is over, stop the capture by selecting
TRANSFER\CAPTURE\STOP.
10 Switch off Diamond.
11 Terminate the HyperTerminal session and disconnect the PC cable from
Diamond.
12 Switch on Diamond.

NOTE
If the connection to HyperTerminal has not been terminated when Diamond is
restarted, Diamond will then establish the connection.

4 38 Instrumentarium Imaging 86502-4AA


4 Parameter programming and software load

4.3.3 Save setup data

NOTE
If setup data is saved with HyperTerminal, the target time values are not
included in the listing. The Diamond Service Software (87502) is
recommended for saving the setup data.

Saving data setup with HyperTerminal


1 Start capture.
2 After you have finished with Diamond procedures, stop capture.
3 Open up the text with a word processor.
4 Save the text into some word-processing file.
5 Save the file in a secure place for future reference.

In case restoration of some old system parameters is necessary, send the


printout or a text file to the manufacturer. The data is examined and the
instructions will be sent to you.

4.4 Diamond test displays


In service mode, Diamond Control Panel displays various figures that convey
information about the system condition. The printout of the screen items is
included in appendices.

1 Enter service mode


2 Press button 3 (the right-most button below Control Panel LCD-screen.
3 Press button 2 (in the middle, below LCD-screen). Screen displays
various test figures.
4 Check out listing of the page number in upper left corner of the LCD-
screen.
5 Press button 3 to scroll through pages 1-4.
6 After you have finished with test displays, exit the screen with -key and
exit the service mode.

4.5 Software versions


Refer to Diamond Spare Part Catalog (86504), Software section, for order
codes of Board and EPROM software.

4.6 Specimen mode


When Specimen mode is selected, the following apply:

– kV value ranges from 15 to 39 kV


– LCD screen blinks
– compression, C-Arm rotation and AutoPoint disabled

86502-4AA Instrumentarium Imaging 4-39


4 Parameter programming and software load

4 40 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5 Electrical system architecture - theory of


operation
5.1 Overview
This section provides theory of operation from the system level to detailed
circuit level. The electronics contains both low voltage - low energy circuits
and power electronics circuits which are directly connected to the mains line
voltage. Refer to complete Diamond System Architecture diagram in the
appendix section.

The voltages inside the system are hazardous and therefore great caution
should be exercised when measuring electric signals especially with
electronic boards that contain line referenced circuits. The electronics consist
of four major components:

1 Printed circuit boards (PCBs)


– Include all the electrical circuits that are needed to operate the system.
2 Sensors
– Detect positions of different movements and status of the unit
3 Motors
– Provide different movements of mechanical constructions
4 Cabling
– Connects PCBs with other PCBs and sensors and provides PCBs and
motors with operating power

For the system architecture, refer to the documents in Chapter 12,


Appendices.
KVOK-COMP

GENENA

FBMISS
PRE/MA

*MAOK
J2

+34v
PRE/MA DAC
LARGE
+15v TF1
MACTRL FILA-ERROR HEATENA
PREREF SMALL
OUTA
KVMAREF(0:7)
MAREF
OUTB COMMON

MACTRL
SHUTDWN
CURLIM
LARGE

SMALL

PWM
MACTRL FILREF
IN+
MACTRL
IN-
OUTA
0R1

OUTB

MACTRL
MAFB

MAREF
SHUTDOWN

MAFB J4

MA SECTION

EXPENA
SHUTDOWN V ref.
13 SEC CURLIMIT
S Q TUBEFAIL BACKUP (60A)
C TIMER
TUBEFAIL R Q

C B
J5
KVFB

DAC VCOFREQ
KVFB
S Q A D
VCO C
KVREF
KVMAREF(0:7) OUTA R Q

(Error Amplifier)

KVOK-COMP
A C

J3

B D HV-Unit
*KVOK

KV SECTION 310Vdc 310Vdc-GND

Fig 5.1. X-ray generator block diagram

86502-4AA Instrumentarium Imaging 5-1


5 Electrical system architecture - theory of operation

The generator of the unit is a modern, constant power, high frequency,


feedback-controlled DC x-ray generator consisting of the following parts:

1 Generator Board
2 High voltage unit
3 Anode motor drive (AMD) board
4 Mains transformer
5 Energy storage capacitor
6 Tube housing

Technical specifications
Input power: 4.6 kVA
Output power: Max. 2.6 kW
Output voltage: 15 - 39 kV
Output current: 20 - 100 mA
Frequency: 40 - 100 kHz
Line voltage: 230Vac ±10%, single phase, 50/60Hz
Irms (rest): 300 mA
Irms (exposure): 20 A
I peak: 55 A

5.2 Input power


The mains line is applied to the Power Board via mains fuses, line filter and
mains switch. The AC mains voltages are further applied to Generator Board,
Stereo Board, AMD Board and to Main transformer that provides all the low
voltage (ground referenced) power.

Power Board

Main transformer

52 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.2.1 Supply Voltage Regulation

Supply voltages from the main transformer are rectified at Power Board. The
output voltages are 24Vac and +34Vac. The voltage rectifiers are located as
follows:

Input Regulator Location Used By

+24V +15Vdc Power Board Power and Generator Board


-24V -15Vdc Power Board Generator Board
+24V +5Vdc Supervisor Supervisor and Power Board
Board
+24V +5Vdc Stereo Board The internal serial bus, Stereo/
Drive/Collimator/Compression/
AEC/Compression Display
Boards
+24V +15V Stereo Board Stereo/Collimator/AEC Board
-24V -15Vdc Stereo Board AEC Board
±15Vdc ±12Vdc AEC Board AEC Board
+24V +5Vdc Control Control Panel Board
Panel Board
Fig 5.2. Supply voltage regulation

5.2.2 Sensor types


A
Photointerrupters (A)
– Mechanical “flag” enters the slot

Hall Effect Sensors (B)


– A magnet placed in front of the sensor
– POLARITY!
B
Motor Speed Sensors (C)
– Rotary Encoder
– Integrated inside the Motor

Microswitches

86502-4AA Instrumentarium Imaging 5-3


5 Electrical system architecture - theory of operation

5.3 Printed circuit board (PCB) communication


The center of the system is the Supervisor Board that controls all the system
operations. All PCBs are connected to the Supervisor Boards either directly
or via another board.

The communication takes place with two serial (RS-485) links and parallel
signals that provide link between the Supervisor Board and other PCBs. The
first serial bus is an external bus and it connects the Supervisor Board to
Control Panel(s) and other external control devices. The second serial bus is
internal and it connects the Supervisor to all PCBs inside the C-Arm of the
system. The parallel ports connect to the Generator Board and Power Board
that are located in the carriage.

5.3.1 Internal Serial Bus

The internal serial bus connects the Supervisor board to PCBs located in the
C-Arm. The bus is separated into two segments: the first one connects the
AEC Board and Stereo Board. The other segment connects through Stereo
Board and it connects Compression Board and Collimator Board.

Each of these above-mentioned boards has its own microcontroller that


controls the input/output on that particular board and boards that connect to
them (AMD Board, Drive Board). Each board has its own address and the
Supervisor Board reads and writes alternately to each board every 20ms. If a
board does not receive data for a certain period of time it will reset all its
outputs to a non-active state.

The serial bus has two lines: receiving (RXD) and sending line (TXD-from
Supervisors viewpoint). The Supervisor controls the sending line and all
other boards observe that line all the time. The receiving line is occupied for a
certain time by each board alternately after the Supervisor has addressed the
board.

The serial cable is a shielded cable and the shield should always be grounded
at both ends of the cable.

5.3.2 External Serial Bus

The external serial bus is similar to the internal bus except that it is intended
for one or two control panels and other control devices that are located out
side the unit.

54 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.4 Printed circuit boards (PCB)


5.4.1 Supervisor Board

The Supervisor Board is the center of the system. It is built around Intel’s
386EX microprocessor. The Supervisor Board controls all the other boards as
discussed in chapter 5, Electrical system architecture - theory of operation.
The software can be programmed into the memory via JTAG link or COM
port. For details, refer to chapter 4, Parameter programming and software
load.

Supervisor Board

Schematics, Layouts

Refer to the documents in chapter 12, Appendices.

5.4.2 Power Board

The Power Board provides supply voltages and main generator power as well
as controls C-Arm rotation and Vertical drive motors.

5.4.3 Power supply section

Power supply section of the Power Board produces supply voltages +5 V, ±15
V, ±24 V, +34 V and +310 V. Voltages are rectified from output of main
transformer TF1 and direct from line voltage.

Voltage level of the main supply can be read by the supervisor from signal
CFORCE provided by the power board. Voltage level is measured and
calculated based on a low-level voltage +24Vdc. CFORCE is a dual-purpose
signal and is enabled for measuring the voltage level from +24Vdc while the
C-Arm is not run.

If the main supply have been interrupted for about 10 to 15 milliseconds,


signal POWERFAIL is passed on to the supervisor. POWER-ON LAMP is
switched on by relay K6 when 34 Vdc is available.

86502-4AA Instrumentarium Imaging 5-5


5 Electrical system architecture - theory of operation

The full wave rectified 310 Vdc is applied to the Generator through relays K6
and K7. The Supervisor board controls relays with signals RG1 and RG2.
When an exposure is initiated, K7 applies voltage through current limiting
resistors R153 and R154 that allow controlled charging of the energy storage
capacitor. K6 is activated by RG2 about one second later and provides full
power to the generator board.

External indicators:

Connector X16 provides relay contacts for X-ray system status indication.
X16 pins 1 and 2 are short-circuited when the unit is on. X12 pins 1 and 2 are
short-circuited when X-rays are being generated. The contacts are rated for
5A, 24VDC /240VAC.

5.4.4 C-Arm rotation motor control

The C-Arm motor control is an 8-speed supervisor programmed driver for 24


Vdc motor that rotates the C-Arm. Speed of the movement is stabilized with
speed feedback produced by a tachometer on the C-Arm motor. The
tachometer provides a pulsed signal C-PULSE that is transformed to voltage
speed feedback by frequency to voltage converter ICA7.

Direction is controlled with signal C-DIR-LEFT and speed with signals C-


MUX1, C-MUX2 and C-MUX3. Four C-position optoswitches provide
signals C-RIGHT-POS, C-RMIDDLE-POS, C-LMIDDLE-POS and C-
LEFT-POS. If signal C-RIGHT-POS or C-LEFT-POS is active, then C-Arm
movement to the right or left is restricted respectively. Signal C-RUN is
required to run C-motor. C-motor current is monitored and restricted to app.
15.7 Amperes.

5.4.5 C-Arm Rotation Sensors


Rotation Limits

There are three limit sensors for the C-Arm rotation. The center limit consists
of two slotted photointerrupters, which detects the flag in the C-Arm
mechanism. The C-Arm movement slows down when the flag passes the
double opto sensor.

The two other photointerrupter limits prevent the mechanism from collision
with the mechanical limit at maximum C-Arm angles. The signals from the
limits are taken via the Power Board to the Supervisor Board.

56 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

Rotation Angle

The Rotation angle is calculated from the pulses of the Tacho Board, which is
attached to the C-Arm motor. The software calibration is based on fixed
rotation limits that is the center position, where the zero reading is updated
each time the C-Arm is at CC-position.

The tacho pulses are taken to the Supervisor Board via the Power Board.

When the power is turned on the current angle value is stored in Supervisor
Board memory where it is fetched from next time the unit is powered up.

Rotation Speed

The Rotation Speed is measured from the pulses of the Tacho Board, which is
attached to the C-Arm motor. The pulses are sent direct to the Power Board,
which uses a circuit that controls the speed according to a speed reference
from Supervisor Board.

5.4.6 Vertical movement or Z-motor control

Z-motor drives vertical movement of the device. Signal Z-DIR-UP defines


direction of the movement. If signal Z-UPPER-LIMIT or Z-LOWER-LIMIT
is active, then movements up or down are not allowed, respectively. Two
microswitches provide Z-limit signals. Signal Z-RUN is required to run Z-
motor. Z-motor current is monitored and limited to app. 4 Amperes. Z-motor
is driven with two speeds. Once the drive is enabled it is started at a low
speed and after a little delay the speed is increased.

The drive of the motor is possible only if motor-stop buttons are installed and
no button is activated (pressed down).

Fuses

Refer to chapter 5.12, Fuses, voltages and led indications.

Schematics, Layouts

Refer to the documents in chapter 12, Appendices.

86502-4AA Instrumentarium Imaging 5-7


5 Electrical system architecture - theory of operation

5.5 Generator Board


Generator Board regulates the kV and mA of the X-ray tube. The Generator
Board contains a high voltage and a low voltage section. The high voltage
section is isolated from the low voltage section with optocouplers ICD14-
ICD17 and current transformer TF2.

WARNING
Voltages in the high voltage part of the Generator Board are hazardous. The
peak to peak voltage exceeds 700V in normal operation.

Generator Board

5.5.1 kV Control
5.5.1.1 High voltage section

The high voltage section consists of a MOSFET transistor H-bridge, which


converts the rectified line voltage (+310 VDC) into high frequency current.
This current is taken to the HIGH VOLTAGE UNIT where it is transformed
into high voltage. The inductors L3, L4 and H-bridge’s operating frequency
determine the RMS value of the current and thus the output power to the X-
ray tube.

The MOSFET driving pulses are generated at the low voltage section of the
board and are provided by optocouplers ICD14 - ICD17. TF2 is a current
transformer, which monitors the bridge current. If the bridge current exceeds
the preadjusted current limit the control circuit in the low voltage section
turns the MOSFETs off.

58 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.5.1.2 Timing

The input voltage is applied through the connector X6. When an exposure is
initiated, relays K1 and K2 on the Power Board provide the line voltage. The
activation of the EXPENA signal (H8) enables the H-bridge drive circuitry.

RG1
310V ON

1.15s
RG2
PREHEAT ON

112ms
GENENA MACTRL ON
710ms

210ms 100ms
PREMACTRL
XRAYS ON

180ms 70ms
EXPENA
65ms GRID ON

GRIDSTART

165ms 60ms
XRAY
60ms

ANRUN
50ms 1.2s

100ms
ANSTART
5ms

X-rays off
X-rays off
X-rays on

X-rays on
ANENA

5.5.1.3 Low voltage section

The high voltage is regulated by adjusting the H-bridge’s operating


frequency. This is done by comparing the feedback signal (KVFB) from the
high voltage unit with the reference voltage from the Supervisor Board and
taking the error signal to the frequency modulator circuit (ICA10, ICD7,
ICD12). The output is taken to ICD10A and ICD10B, which drive the
MOSFET driving optocouplers.

EXPENA signal

An exposure is initiated by the Supervisor Board by activation of the


EXPENA signal. This enables pulsing of the MOSFET drive optocouplers
and allows the kV reference signal to rise providing “soft start”.

kV reference

The level of the used kV is set by the SUPERVISOR BOARD with signals
KVMAREF0 -KVMAREF7. This 8-bit digital word contains the kV
reference. The kV reference is converted into an analog signal in D/A
converter ICA6 (TP26).

The analog kV reference (TP26) and the kV feedback (TP20) signals


correspond to actual kV as follows:

kV = DVM read-out * 10 kV

For example, if the digital multimeter (DVM) read-out is 2.5 V, the actual
high voltage is 25kV.

86502-4AA Instrumentarium Imaging 5-9


5 Electrical system architecture - theory of operation

kV adjustment

The calibration of the kV is done by adjusting the D/A converter’s (ICA6)


reference voltage with trimmer potentiometer R67.

NOTE
The kV calibration is performed at the factory. The trimmer potentiometer
R67 is sealed. Manipulation of the potentiometer can be carried out only by
Instrumentarium authorized service technician.

CAUTION
Improper settings of the potentiometer can damage the Generator Board,
High Voltage unit and the X-ray tube.

5.5.2 mA Control

The mA control adjusts the X-ray tube’s filament heating level to provide the
desired tube current (mA). This is done by regulating the filament voltage
with “FLYBACK” type pulse width modulated switch mode power supply
(T1, T2, TF1, D14, C85, and ICA5). The control circuitry uses current mode
control, which provides better stabilization and protects the semiconductor
switches against over-current. The operation consists of two consecutive
sequences: 1) the preheat sequence, and 2) normal mA control sequence. The
preheat sequence is used for heating of the filament to the desired level prior
to the exposure, in order to achieve optimum mA rise at the start up.

During the preheat sequence the control feedback comes from the filament
voltage, and during an exposure the mAFB signal is used, providing the
actual tube current monitoring.

mA generation timing

mA generation timing is controlled with four signals: SMALL/


LARGEFOCUS, GENENA, PRE/MA-CTRL and EXPENA.

SMALL/LARGEFOCUS is a latch output, which changes state depending


on whether the Mag-box (small focal spot) or either a Bucky or a Cassette
tunnel (large focal spot) is attached to the unit. Diamond selects which mA
feedback resistor will be used on the mA feedback and which filament will be
energized.

GENENA becomes active 20-40 ms after the hand switch is depressed. The
relay K1 is powered up by GENENA and applies power to the filament
transformer TR1. During the time between PREHEAT and EXPENA, the
pulse width modulator (ICA5) uses filament voltage as its feedback voltage.
The reference is software controlled to achieve proper tube current rise at the
beginning of an exposure.

When EXPENA goes active, actual exposure begins and regulation of the
tube current becomes a more important issue than filament voltage. This is
performed by connecting the common of the filaments to ground through one
of the mA feedback resistors (R7 or R8+R9). This signal, called mA
FEEDBACK (MAFB) is now fed to the pulse width modulator by an analog
switch (ICA10). Its output was transferred from filament voltage to MAFB

5 10 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

with the signal PRE/MACTRL. The PRE/MACTRL signal is activated


approximately 7 ms after the EXPENA signal. The duty cycle of the pulse
width will vary depending on what is required to maintain proper mA.

Preh reference

The preheat reference is the reference for filament voltage. It is converted


from 8 (KVMAREF0-7) parallel databits into preheat reference information
in two-channel D/A converter (ICA4).

mA reference

The mA reference is the reference for the actual tube current. It is converted
from 8 (KVMAREF0-7) parallel databits into mA reference information in
two-channel D/A converter (ICA4). The selection between preheat and mA
channel is made with signal PRE/MASEL (0 - PREHEAT; 1 - mA).

Preheat adjustment

The preheat adjustment is performed separately for large and small focus by
varying the preheat reference with software.

5.5.3 Protection circuits

The generator board features several circuits to protect the generator or the X-
ray tube.

KVOK and TUBEFAIL signals

KVOK and TUBEFAIL signals are activated if the KVFB signal does not
reach the kV reference. This typically occurs during a high voltage
breakdown or arching.

TUBEFAIL signal provides a SHUTDOWN kV for the control circuit on


board. KVOK signal provides Supervisor Board with information that kV is
reached.

As the arching is characteristic to x-ray tubes, and hence may occur during
normal operation, the exposure is continued with a new activation of the
EXPENA signal. The Supervisor Board allows three (3) tube failures per
exposure before an error message is displayed.

MAOK and FILA_ERROR signals

MAOK signal is zeroed if the mA feedback does not reach the mA reference.
This information goes to Supervisor Board and enables software controlled
protection.

FILA_ERROR signal is activated approximately 600ms after the MAOK


signal goes zero. It reduces the current limit for filament power supply, thus
protecting the filament from excessive heating.

FBMISS signal

The FBMISS signal is activated if the feedback cable is not connected or the
connector is not properly seated. It disables the +34V power to the filament as
well as shuts down the filament PWM regulator (ICA5).

86502-4AA Instrumentarium Imaging 5-11


5 Electrical system architecture - theory of operation

5.5.4 Back Up Timer

The generator has a Back Up Timer (ICA10c) that disables both the +34V for
the filament and the EXPENA signal providing SHUTDOWN approximately
fifteen (15) seconds after the activation of the EXPENA signal. This is only a
safety feature and does not affect normal operation.

Jumper Normal position Test position Test position


(edges painted (edges not description
white) painted white)

J1 open closed Disable FBMISS


signal
J2 1-2 2-3 Forced preheat
control of the
filament
J3 1-2 2-3 Disable kVOK
signal
J4 1-2 2-3 Disable mAOK
signal
J5 1-2 2-3 Disable
TUBEFAIL signal
and generator
shutdown
Fig 5.3. Generator board jumpers

CAUTION
The Test Jumper J5 should always be at “NORMAL” position whenever the
high voltage part of the Generator Board is energized. Failing to do this may
damage the Generator Board, High Voltage unit and the X-ray tube.

5.5.5 Monitoring

The Supervisor Board monitors the generator performance and measures


certain signals in order to perform calibrations and adjustments.

The information to Supervisor Board from different generator signals is


multiplexed through GENMON signal. The different measured values are
multiplexed to FBFRQ. The multiplexing is performed according to
following table:

5 12 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

EXPENA MUXSEL PRE/MASEL GENMON FBFRQ

0 0 0 TUBEFAIL +34V
0 0 1 TUBEFAIL KVREF
0 1 0 FBMISS MAREF
0 1 1 FBMISS PREHREF
1 0 X TUBEFAIL KVFB
1 1 X FILA_ERROR MASFB
Table 5.1 Signals to Supervisor Board

An exposure without X-rays

It is possible to check the operation of kV and mA control circuitry without


X-rays.

1 Turn OFF the unit


2 Unplug connectors X6 and X4 from the GENERATOR Board (a four-
J3 pin connector).
3 Unplug X5 of the Power Board
J4 4 Set the jumpers J3, J4 on the GENERATOR Board to 2-3 position [non-
default position].

The kV control circuitry is driving the MOSFETS with the minimum


frequency (app. 20kHz) and the filament voltage is regulated to preheat level.

Schematics, Layouts

Refer to the documents in chapter 12, Appendices.

5.6 Anode Motor Drive Board - AMD Board


The unit features a rotating anode tube. The anode plate is accelerated with
230 VAC before the emission of electrons from cathode. During exposure,
AMD Board stator needs a certain running voltage (40-100 VAC), and after exposure the
anode plate can be held steady with a brake stop.

All this is performed with the phase control of a triac (V1). The control
element is Telefunken TEA1007 (A7). Acceleration is achieved directly with
mains voltage. During exposure, the firing angle of the triac is set to a level
that corresponds to the required RMS voltage. Both acceleration and brake
procedures take approximately 1 second. Safety issues are also taken into
account by monitoring the current of the triac. Should errors be detected, the
microprocessor will not allow exposures. The exchange of information
between the microprocessor and this board is optoisolated because the anode
motor drive board is floating on mains voltage.

Schematics, Layouts

Refer to the documents in chapter 12, Appendices.

86502-4AA Instrumentarium Imaging 5-13


5 Electrical system architecture - theory of operation

5.7 Stereo Board


Stereo Board is the central board in the C-Arm. Its main tasks are the
electronics bus segment handling, sensor and keyboard signal monitoring,
power supply regulation and motor control.

Stereo Board

The Stereo Board handles the internal electronics serial bus communication
so that the two bus segments can communicate. Each time a board is sending
data to Supervisor Board (that is, writing to the bus) it must take over the
TXD line. The segment that contains Collimator and Compression boards
connects to the main segment via Stereo board and therefore Stereo board
must take over the TXD line of the main segment each time someone from
the other segment sends data to Supervisor.

Sensor and keyboard signals on the C-Arm (except tube head) are collected in
the Stereo Board. The (15Vdc and +5Vdc supply voltages for all the boards in
the C-Arm are regulated on the Stereo board. Refer to chapter 5.2.1, Supply
Voltage Regulation. Stereo board also drives the compression motor, ECS
motor and Stereo motor and controls cassette and grid motors on the Drive
board.

5.7.1 Voltage regulators

The Stereo Board contains +5V switch mode power supply (SMPS) that
supplies +5Vdc voltage to the Stereo, Drive, AEC, Collimator and
Compression Boards.

The +15V linear regulator supplies +15Vdc to the Stereo, Drive, AEC,
Collimator and Compression Boards.

The -15V linear regulator is used as preregulator by AEC Boards -12V


regulator.

5 14 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.7.2 Stereo and ECS motor drive

Both motors are +24VDC-motors, which are driven with pulse width
modulated (PWM) power supplies that can provide variable voltage control,
thus enabling speed control. Each drive circuit consists of control circuit and
two power transistors. The output voltage consists of +34 Volt pulses whose
width determines the RMS voltage of the motor. This voltage is proportional
to the reference voltage applied to the control circuitry.

The operating frequency for both drives is approximately 30kHz.

Stereo and ECS motors are driven by the same control and drive circuit
(ICA1, T1, T2). Both drives are speed controlled. The speed feedback for
stereo drive comes from stereo motor tachometer and the speed reference for
ECS drive comes from compression motor tachometer. The active motor is
selected with relay K3 and the direction is selected with K4.

The operating speed is software controlled and multiplexed from 7 resistors


with ICD18. ECS speed is either determined by the compression motor (ECS
mode) or fixed speed (C-Arm mode).

Both drives also utilize a current limiting circuit that limits the motor current
to approximately 5A.

5.7.2.1 Stereo Sensors

Stereo Angle

The angles for stereotactic operations are accomplished with four slotted
photointerrupters located underneath the Stereo Board. The ±15 degree
angles are determined by two opto boards on the sides. The center position of
the stereo arm is detected with the Stereo Sensor Board, which contains two
opto switches. The exact positioning of the stereo arm is adjusted by moving
the corresponding opto boards.

85030

85028

86502-4AA Instrumentarium Imaging 5-15


5 Electrical system architecture - theory of operation

Stereo Speed

The Stereo Speed is measured from the pulses of the Tacho Board, which is
attached to the Stereo motor. The pulses are taken to the Stereo Board, which
uses a circuit that controls the speed according to a speed reference.

5.7.3 Compression motor drive

The compression motor drive is a software controlled 8 speed PWM drive.


The speed is selected by choosing one reference voltage from multiplexer
ICD9. The compression motor tachometer measures the compression speed,
which is used as reference speed to ECS motor control.

5.7.4 Back up release

The unit is equipped with a compression release back up feature. If the line
voltage suddenly disappears, then the compression is released. This is
implemented with relay K2. When this relay is powered up, the compression
motor is operated normally. When the relay is not activated, the motor is
driven upwards with capacitor charge in the +34Vdc power supply line.

5.7.5 Sensors and Keyboard

Stereo board receives the signals from stereo angulation sensors, ECS limit
sensors (which connect directly to the board) and cassette holder sensors,
cassette movement and grid position sensors (which connect via Drive
Board).

The signals from C-Arm operating panel connect to X24.

Schematics, Layouts

Refer to the documents in chapter 12, Appendices.

5 16 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.8 Drive Board


The Drive Board contains drive and control circuits for grid and cassette
movement stepper motors.

Schematics, Layouts

Refer to the documents in chapter 12, Appendices.

5.8.1 Grid movement

The grid is moved with a stepper motor that is controlled by ICD3. This
microcontroller contains software that controls the grid speed, which is a
function of the grid position.

Prior to an exposure, the Stereo Board activates the grid movement with
signal GRID_START. When the grid has reached the maximum speed the
microcontroller on Drive Board sends the XRAY signal that allows the
Supervisor Board to start the exposure.

Grid Sensor

The actual location of the grid is calculated from stepper pulses by the grid
microcontroller on the Drive Board. Cable CS25 connects the grid sensor to
the Drive Board.

Home position

The home position of the grid is sensed with a slotted photointerrupter that is
activated by a flag in the grid mechanism. The grid home signal is taken to
both Drive Board grid controller and the microcontroller on the Stereo Board,
which further sends the data to Supervisor Board via RS-485 serial link.

sensor

sensor

5.8.2 Cassette tray movement

Cassette Tray is moved with a stepper motor that is controlled by ICA4. This
stepper motor controller requires enabling, direction and clock signals from
Stereo Board (CAS_ENA, CAS_DIR, CAS_CLOCK).

Each clock pulse corresponds to a certain movement in distance and the pulse
frequency determines the travel speed. The Stereo Board microcontroller
provides the necessary amount of clock pulses to drive the cassette tray. The
actual cassette tray position is sensed with cassette tray movement sensors.

86502-4AA Instrumentarium Imaging 5-17


5 Electrical system architecture - theory of operation

5.8.3 Sensors

Cassette holder sensors, cassette tray movement and grid position sensors all
connect to the Drive Board. The signals are further taken to the Stereo Board,
which sends them to the Supervisor Board via the RS-485 serial link.

Cassette holder sensors

Cassette Holder Sensor Board houses 4 Hall effect sensors that detect the 1-4
magnets that encode cassette holders (Bucky, Mag, D32…). The Cassette
Holder Sensor on CSense Board (Sensor Board locations: image A, front,
image B, rear) connects to the Drive Board and then to Stereo Board. The
information from this sensor affects compression thickness measurement and
collimation.
Fig 5.4. Sensor Board location, front
The Sensor Board Hall-sensors (image A) decode the holders according to
the table below. Each cassette holder houses one to four magnets (image B) to
encode the holder.

A B

S2 S3 M3 M2

S1 S4 M4 M1
Fig 5.5. Sensor Board location, rear

The signals from the Sensor Board are transferred to Drive board and further
to Stereo Board, which sends the data to the Supervisor Board via RS-485
serial link.

Diamond sensors Corresponding cassette


holder
S4 S3 S2 S1

0 0 0 1 Bucky 18x24
0 0 1 0 Bucky 24x30
0 0 1 1 Mag 1.6
0 1 1 1 Mag 1.8
1 0 1 1 Mag 2.0
1 0 0 1 Digital Stereo - Delta 32
0 1 0 1 Contact 18x24
NOT USED!
0 1 1 0 Contact 24x30
NOT USED!
1 0 0 0 Spot Mag 1.6
NOT USED!
Table 5.2 Encoding of the cassette holder

In schematics, C1 equals to S1, C2 to S2, C4 to S3 and C8 to S4.

5 18 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.8.4 Cassette Sensors


Cassette detection

The presence of a cassette is


detected with a reflective photo
sensor that is activated when the
cassette tray is in back position and
a cassette is inserted. The LED part
of the sensor emits infra-red light
that is reflected to the
phototransistor part of the sensor
whenever a reflective surface is
placed near the sensor. The
sensitivity of the detector can be
adjusted with a trimmer sensor
potentiometer on the Cassette Opto
Board.

The photo sensor connects to the


AEC Board and is then taken to the
Supervisor Board via RS 485 serial
link by AEC board microcontroller.

Cassette locks

Cassette tray movement depends on


the front and back position of the
cassette lock. The cassette holder
has a Cassette Lock Sensor Board
on both sides, which houses two
Hall sensors to detect the small
magnet in the cassette lock lever.

The Cassette Lock Sensor Board


connects to the AEC board X1 and back position
is then taken to the Supervisor
Board via RS 485 serial link by
AEC board microcontroller.

front position

86502-4AA Instrumentarium Imaging 5-19


5 Electrical system architecture - theory of operation

Cassette tray movement

Three slotted photosensors detect the cassette tray movement so that two of
them are connected in parallel providing one signal at both ends of the
movement area.

The position of the cassette tray is coded as in the table below.

Sensors Corresponding position home


position
S1 S2 S3 24x30
1 0 0 Home position
18x24
0 0 0 In between
0 0 1 24x30 cassette
0 1 0 18x24 cassette
Fig 5.6. Encoding of the cassette tray movement

5.9 Collimator Board


The Collimator Board controls the integrated collimator plates, filter
movement, tube park movement and field light. Collimator Board receives
the signals from tube park movement sensors, tube movement safety clutch
sensor, collimator plate and filter home sensors and external collimation plate
sensor board.

Collimator Board

The signals from tube head operating panel connect to X24.

Schematics, Layouts

Refer to the documents in chapter 12, Appendices.

5 20 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.9.1 Collimator plate and filter drive

The two integrated collimator plates and the filter mechanism are each driven
by a stepper motor. Each motor uses the same control and drive circuit.

5.9.2 Collimator Sensors


Integrated collimation plates

The actual location of the collimator plates is calculated from stepper pulses
by the microcontroller on the Collimator Board. The position of the front lid
is derived from the drive order of the back and side plates.

Home position

The actual location of the collimator plate is sensed with a slotted


photointerrupter that is activated with a flag in the collimator plate. The home
signal is taken to the microcontroller on the Collimator Board, which further
sends the data to Supervisor Board via RS-485 serial link.

External collimation plate detection

External collimation plates are detected by Collimator Sense Board, which


has 3 optos for that purpose. Every diaphragm has 1-3 flags that activate
respective optos. The signals are taken to the Collimator Board to be further
utilized in Collimation interlock or other (later) features. Table of diaphragm
codes will be determined in later versions of this manual.

side plate sensor rear plate sensor

5.9.3 Tube Park drive

The tube drive circuit consists of pulse width modulated (PWM) IC ICA5,
which run at approximately 30kHz. The drive has 8 different software
controlled reference voltages that correspond to 8 different tube park speeds
respectively.

86502-4AA Instrumentarium Imaging 5-21


5 Electrical system architecture - theory of operation

5.9.4 Tube Park movement sensors


Tube Park movement limits

The Tube Park movement limits are accomplished with a flag that enters four
slotted photointerrupters. Two of the switches are located at end points of the
movement. Their purpose is to prevent the mechanical collision at both ends
of the tube movement. The purpose of the two other (middle) switches is to
determine the position where the tube speed will be reduced prior to reaching
the ends. The signals from the four switches are taken to the Collimator
Board, which further sends the data to the Supervisor Board via the RS-485
serial link.

Location Function
Left front (A) Slowdown
Left rear (B) Slowdown
Right front (C) Stop
Right rear (D) Stop

A B

C D

Safety clutch

The Tube Park


mechanism features a
safety clutch in case the
tube movement is
hindered. The status of the
safety clutch is monitored
with a microswitch that
closes when the clutch has
been activated. The
microswitch connects to
the Collimator Board,
which further sends the
data to the Supervisor
Board via the RS-485
serial link.

5 22 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.9.5 Filter sensor

The actual location of the filter mechanism is calculated from stepper pulses
by the grid microcontroller on the Collimator Board.

Home position

The home position of the filter is sensed with a slotted photointerrupter that is
activated by a flag in the filter mechanism. The filter home signal is taken to
the microcontroller on the Collimator Board, which further sends the data to
Supervisor Board via RS-485 serial link.

5.10 Compression Board


The Compression board controls the compression thickness and force
measurement and display. It also controls the displays and keyboard on
compression arm control panel. The microcontroller ICD7 controls the board
operations and sends the data to the Supervisor Board for further processing.

5.10.1 Compression thickness measurement

Compression thickness is measured with optocouplers that read the dotted


tape. The dotted tape provides information about distance, direction of
movement and compression and home position. This information is then
transferred to processor 87C51. The calibration is performed with the
software.

5.10.2 Compression force measurement

Compression force with units S/N 20000 - 20135 is measured with strain
gauges in the compression mechanism. The signal is brought to the amplifier
ICA1. The offset voltage of the amplifier must be set to 450mV/5mV when
there is no strain on the compression paddle. The voltage is measured
between test points TP2 and TP3. For Compression measurement with units
S/N 20136 and up, refer to Diamond Technical Bulletin 07/2002,
Compression_Information.

5.10.3 Displays

Compression force, compression thickness and the C-Arm angle are


displayed by Compression Display Board that contains a driver ICD8
(ICM7218BIQI) to control the 7-segment display. The Display Board also
contains 5 (4 in older units) Hall-type sensors to recognize 32 (16 with older
older units) various compression paddles. Refer to chapter. 5.10.5,
Compression Sensors, Compression Paddle Detections-section.

86502-4AA Instrumentarium Imaging 5-23


5 Electrical system architecture - theory of operation

5.10.4 Keyboard

Buttons for AEC screen, C-Arm and Z-movement are connected to 3x8
matrix. The matrix number is regulated with a micro-controller ICD7 of the
Compression Board.

3X8 MATRIX
SW-A SW-B SW-C
DETEC 0 AUTOPOINT PADDLE 1
DETEC 1 FILTER PADDLE 2
DETEC 2 COMPRESSION + PADDLE 4
DETEC 3 COMPRESSION - PADDLE 8
DETEC 4 FIELD LIGHT VERT-UP
DETEC 5 C-ARM-LEFT VERT-DOWN
DETEC 6 C-ARM-RIGHT N.C.
DETEC 7 N.C. N.C.

5.10.5 Compression Sensors


Compression travel limits

The compression travel limits are measured with the two slotted
photointerrupters (A) from the striped film strip located in the compression
mechanism.

The compression features two limits. The upper (microswitch, B) is actual


travel limit that prevents collision. The other one is used for compression
thickness calibration (C).

Compression thickness

The compression thickness is measured with the Thickness Sensor Board


and the striped film strip located in the compression mechanism.

The Thickness Sensor Board consists of two accurately positioned slotted


photointerrupters. The microcontroller on the Compression Board counts
pulses and the phase shift between the two sensors in order to determine the
relative compression travel and direction. The relative compression thickness
is then sent to the Supervisor Board via the RS-485 serial link.

The calibration is carried out at the two fixed positions that act as
compression travel limits. Each time the power is turned on, the system
locates one of the fixed positions and determines the actual compression arm
location.

5 24 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

Compression force measurement (Diamond units with S/N 20000 -


20135)

The compression force is measured with two strain gages located in the
special bending force measuring block of the compression arm that measure
the relative movements. The strain gages connect to the Compression Board,
which features a voltage to frequency converter that is controlled by the on
board microcontroller, which further sends the data to the Supervisor Board
via the RS-485 serial link.

Compression force measurement requires electric offset adjustment (trimmer


potentiometer). Software re-calibrates the offset point after each
compression.

Compression force measurement (Diamond units with S/N 20136 and up)

The new compression force mechanism measures critical shear stress. This
mechanism does not need pre-tension adjustment.

Compression speed

The compression speed is measured from tachometer inside the compression


motor. The tacho pulses are taken to the Stereo Board that converts the pulses
into a speed reference voltage for the ECS drive to be used in the ECS mode.

Compression paddle detection

Different compression paddles are detected by five (four currently in use)


Hall-sensors on the Compression Display Board that detect the small magnets
inserted in the compression paddle and connect to Compression Board.

86502-4AA Instrumentarium Imaging 5-25


5 Electrical system architecture - theory of operation

sensors

Fig 5.7. Old Compression Display Board with four Hall sensors S/N 20000 - 20135

sensors

Fig 5.8. Current Compression Display Board with five Hall sensors S/N 20136 and up

The image below describes the order of the sensors in the Compression
Display Board (series A) and the magnets on the compression paddle (series
B). Compression Paddle Detection affects compression thickness
measurement, Autorelease functions and PaddleLogic.

5 26 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

Sensors Corresponding compression paddle


S5 S4 S3 S2 S1
0 1 0 1 0 Biopsy paddle 9 and 18mm holes
0 1 0 1 1 Biopsy paddle 50x90mm square
0 1 1 0 0 Magnification paddle
0 0 1 1 0 Magnification paddle implant
0 0 1 0 1 Paddle implant
0 0 0 1 0 Paddle standard 18x24
0 0 0 0 1 Paddle standard 24x30
0 0 0 1 1 Spot magnification paddle
0 1 0 0 1 Spot magnification paddle enhanced circle
and square 12x12cm
0 1 0 0 0 Spot Paddle
0 1 1 0 0 Spot magnification paddle enhanced circle
and square
0 0 1 1 1 Delta 32 paddle

In schematics, C1 equals to S1, C2 to S2, C4 to S3, C8 to S4 and C16 to S5.

ECS Limits

The ECS movement limits are accomplished with a flag that enters the two
slotted photointerrupters. Their purpose is to prevent the mechanical collision
at both ends of the ECS movement.

The signals connect to the Stereo Board, which prevents the ECS from
driving to the corresponding direction. The data from the switches are also
taken to the Supervisor Board via the RS-485 serial link.

upper limit
sensor

lower limit
sensor

Schematics, Layouts

Refer for documents to chapter 12, Appendices.

86502-4AA Instrumentarium Imaging 5-27


5 Electrical system architecture - theory of operation

5.11 AEC Board


The AEC Board measures the radiation that goes through the breast from
eight different ellipsoidal-shaped detectors. It also collects sensor data to
Supervisor Board.

AEC Board

5.11.1 Automatic Exposure Control

Each detector consists of several photodiodes that provide current that is


proportional to the radiation. This current is converted into voltage for each
separate detector. The signals from eight detectors are then multiplexed in
ICA5 to one variable gain and variable offset amplifier and taken further to
12-bit analog to digital converter ICD5. The data from the detectors is then
taken to the microcontroller, which sends the it to the Supervisor Board for
further processing.

The second amplifier stage, which is common to all eight detectors, features
software controlled offset circuit, which provides correct offsets for each
detector. Also the gain can be selected to provide optimum point of operation
for each exposure. The principal AEC gain is accomplished with software.

5.11.2 Sensors

The AEC Board receives signals from cassette sensor and cassette lock
sensors. Refer to chapter 5.8.4, Cassette Sensors.

Schematics, Layouts

Refer to the documents in chapter 12, Appendices.

5 28 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.12 Fuses, voltages and led indications

NOTE
Following voltage measurements, and the soft start procedure instructions are
for troubleshooting purpose.

Verify that the typical voltages listed are present. Refer to repair instructions
in service manual for the location of the test points. Voltages marked with *
are unregulated.

5.12.1 Fuses

Fuses/Power Board 85020.


F1,2 Fuse 5x20mm 4AT, IEC 127-2/3.
F3 Fuse 6.3x32mm, 15 ASB 250V.
F4 Fuse 5x20mm 0.8AT, IEC127-2/3.
F5 Fuse 13/32” 20AF 600VAC (Bussmann KTK-20).

Fuses/AMD 32305.
F1 Fuse 6.3x32mm 10AT 250V.

Diamond fuses used by Delta 32:

Fuse Board 33615


(on PCB located in the bottom part plate of the C-Arm).
F5 Fuse 1 pc, 6x32mm 1,25AT 250V UL
F3 Fuse 1 pc, 6x32mm 0,8AT 250V UL
F1,2,4 Fuse 3 pcs, 6x32mm 0,5AT 250V UL

D32 Fuse Plate 82685 (in the carriage next to the transformers)
F1,2 Fuse 2 pcs, 6X32mm 0,3AD 250 UL 248-14

86502-4AA Instrumentarium Imaging 5-29


5 Electrical system architecture - theory of operation

5.12.2 Voltages

Voltages/Power Board 85020.


TP1 to TP2 GND Unreg. +34 Vdc*
TP6 GND to TP3 Unreg. +24Vdc*
TP6 GND to TP5 Unreg. -24 Vdc*
TP6 GND to TP7 Reg. +15 Vdc
TP6 GND to TP15 Reg. -15 Vdc
TP6 GND to TP20 Reg. +5 Vdc

Network voltage for Generator Board during exposure


TP26 GND to TP27 Unreg. +34 Vdc*

Stereo Board Code 85013


TP15 GND to TP14 Reg. +5 Vdc
TP7 GND to TP10 Reg. +15 Vdc
TP7 GND to TP13 Reg. -15 Vdc

AEC Board Code 85001


TP9 GND to TP10 Reg. +12 Vdc
TP9 GND to TP11 Reg. -12 Vdc

Control Panel Board Code 85069


TP6 GND to TP7 Reg. +5Vdc

Supervisor Board Code 85078


TP6 GND to TP13 Reg. +5Vdc

WARNING
Before connecting test equipment to Generator Board TP43 GND, make sure
that there is no connection to other, previously tested test points. If TP43
GND is connected to some other test point than TP42 or TP44 on Generator
Board, short circuit and damage to electronic boards are likely. After
checking Generator Board, make sure that there is no connection TP43 GND
and the test equipment.

Generator Board Code 85003


TP43 GND to TP42 Reg. +20Vdc
TP43 GND to TP44 Reg. +310Vdc, during
exposure

5 30 Instrumentarium Imaging 86502-4AA


5 Electrical system architecture - theory of operation

5.12.3 LED indications

When Diamond is switched on, the following LEDs should light up:

Power Board 85020


Code Color displayed
H2, H10, H13, H15, H16, H22 Green
H9 Red
LA2 Orange

Supervisor Board 85078


Code Color displayed
H3, H4, H5, H6, H7, H8, RST Green

Generator Board 85003


Code Color displayed
H1, H2, H3, H14 Green

Stereo Board 85013


Code Color displayed
H1, H2, H6, H7, H8, H9 Green

Drive Board 85011


Code Color displayed
H1, H2 Green
H3 Red

Collimator Board 85009


Code Color displayed
H18, H19, H20 Green
H16 Red (during collimator movement)

86502-4AA Instrumentarium Imaging 5-31


5 Electrical system architecture - theory of operation

5 32 Instrumentarium Imaging 86502-4AA


6 System checks and preventive maintenance

6 System checks and preventive maintenance

WARNING
The X-ray tube utilizes a Beryllium window. do not perform any operations
on Beryllium, which produce dust or fumes, such as grinding, grit blasting or
acid cleaning. Beryllium dust or fumes are highly toxic and breathing them
can result in serious personal injury or death.

6.1 X-RAY TUBE PERFORMANCE CHECK


6.1.1 Half-value layer (HVL) measurement

Test equipment required:

– Radiation meter
– 0.1 mm Aluminum Plates (Alloy 1100)
1 Press kV- and Den-buttons down simultaneously to enter the service
mode. The following screen appears:

2 Collimate the X-ray beam by a lead plate with a hole placed close the
tube and restrict the beam to an area slightly larger than the ionization
chamber. Use 28 kV and measure the radiation using 0 - 0.1 - 0.2 - 0.3 -
0.4 mmAl.
3 Plot your data on the chart and connect data points with a line. Take the
0.0 mmAl reading and divide it by two. Find where that reading would
intersect your graph and determine the half-value layer from the
aluminum axis. Use mAs that gives you about 400 mR with 0 mmAl.

mmAl mR
0.0
0.1
0.2
0.3
0.4

NOTE
HVL @ 28 kV MUST be more than .25 mmAl, and should be less than .35
mmAl.

86502-4AA Instrumentarium Imaging 6-1


6 System checks and preventive maintenance

400
380
360
340
320
300
290
280
270
260
250
240
230
220
210
200
190
180
170
160
150
140
130
120
110

100
0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45
Fig 6.1. HVL measurement chart

4 Enter the measured HVL value in the Service Menu to obtain correct
AGD estimates.
5 Press kV- and Den-buttons down simultaneously to exit the service
mode.

6.1.2 Tube efficiency check on the cassette holder

Test equipment required:

– calibrated radiation meter for mammography


1 Center the radiation meter on the image receptor so that its ion cham-
ber is as close as possible to the chest wall edge of the image receptor.
2 Place a piece of lead under the meter to block back scatter.
3 For small focal spot testing, remove the image receptor. Make the
exposures in the service mode. For large focus, insert a cassette holder
and make the measurements on the cassette holder top.
4 Press kV- and Den-buttons down simultaneously to enter the service
mode.

62 Instrumentarium Imaging 86502-4AA


6 System checks and preventive maintenance

5 Make exposures as directed by the table below, comparing your results


with the data given on the same table.

mR/mAs

kV Small focus Large focus

Desired Actual Check Desired Actual Check

25 3-9 3-9

29 8 - 15 7 - 14

35 16 - 23 10 - 18
Fig 6.2. Tube efficiency

NOTE
Make all exposures at 100 mAs without the paddle!

NOTE
If the output of the tube consistently falls below the desired level, check kV,
mA, preheat.

Press kV- and Den-buttons down simultaneously to exit the service mode.

6.1.3 Image quality evaluation procedure


1 For the film processor evaluation procedure, refer to Diamond Installa-
tion Instructions (86501), Chapter 7.2.4, Film processor evaluation.
2 Place a loaded cassette into the Bucky and select AEC operation using
settings (kV) as you would use for an average 4.5 cm compressed breast.
3 Place the RMI 156 phantom on the Bucky in the same position as a
breast. The nipple indent marker should be positioned away from the
chest wall. Use a standard 18 x 24-cm or 24 x 30-cm compression
paddle.
4 Select the AEC detector under the wax portion of the phantom.
5 Make an exposure and process the exposed film. This image will
represent the imaging abilities of your machine using these clinical
factors.
6 If the image is over or under exposed, make a AEC adjustment and
repeat the exposure.
7 Once an acceptable image has been obtained, have one or more viewers
identify and record which objects are seen. Use a magnifying glass to
look for the structures as you would on a patient’s mammogram. Use the
phantom scoring procedure below to score system performance.
8 Make another exposure using magnification table and score that image.
9 Mark both films with the date, kV, mAs and their scores and file them
for future reference.

86502-4AA Instrumentarium Imaging 6-3


6 System checks and preventive maintenance

Image scoring with the RMI 156 phantom

Fig 6.3. RMI 156 phantom

Fibers Specs Masses


Number Size Points Number Size Points Number Size Points
1. 1.56 1 7. .54 1 12. 2.0 1
2. 1.12 1 8. .40 1 13. 1.0 1
3. .89 3 9. .32 6 14. .75 1
4. .75 5 10. .24 7 15. .50 7
5. .54 9 11. .16 10 16. .25 10
6. .40 10

As a minimum, the system shall image:

1 4 Fibers - 1, 2, 3, 4
2 3 Specs - 7, 8, 9
3 3 Masses - 12, 13, 14

To meet the ACR criteria the minimum acceptable score would the yield a
minimum score of 10 for fibers, 8 for specs and 3 for masses. Total score
would the be 21.

If your mammography unit is significantly worse, you may wish to check the
accuracy of the kVp, the focal spot size, the half value layer, and the image
processor, You could also consult a medical physicist or engineer who
specializes in diagnostic radiology.

64 Instrumentarium Imaging 86502-4AA


6 System checks and preventive maintenance

6.1.4 Measurement of the feedback circuitry

CAUTION
The readings from the following resistance measurements depend on the
probe polarity.

Fig 6.4. Measurement of the feedback circuitry

Remove the low voltage connector X33 from the HV-unit and measure on the
cable connector end the resistance between ground and pin X33/8 (This is the
kV-feedback resistance on the INVERTER Board). It should be 20-45 kΩ. (If
the power is ON, the value is 45 kohms).

If 0 kΩ check the cabling between the connector and the FILAMENT


CONTROL Board, and the cabling between FILAMENT CONTROL Board
and the INVERTER Board. If the reading is app. 20 kΩ, change the probe
polarity. Plug in connector X33.

CAUTION
X36 is hot during power on.

Remove the connector X36 from the tube and measure on the cable connector
end the resistance between ground and pin X36/5 (this is the mA feedback
resistor on the FILAMENT CONTROL Board). It should be about 115 Ω
when small focus on and 30 Ω when large focus on. Plug in connector X36.

86502-4AA Instrumentarium Imaging 6-5


6 System checks and preventive maintenance

NOTE
mA feedback will read 30 Ω (large focus) only with power off.

6.1.5 kV, mAs and exposure time test

The measurement must be performed with a calibrated Unfors, RTI, PMX-I/


M or equivalent. The test is performed in the service mode. Select manual
mode.

Use no cassette holder at all when measuring the LF. The Synchro Bucky
feature interrupts the x-rays for 50ms at grid turnaround points. The
interruption may interfere with the measurement.

When measuring the SF, magnification tunnel should be used. Set the
parameters as shown in the table below and make the exposures.
Alternatively, the Physicist mode can be used.

kV mAs kV kV mAs kV

LF 23 300 28 125 35 63

SF 28 63 35 20

Fig 6.5. Parameters for small and large focuses

Record the kV, mAs and time reading after each exposure. The time reading
is to be compared with the Diamond control panel time display.

The maximum tolerances for kV, mAs and exposure time:

For kV ± 1.5 kV + internal acceptance limit (1kV)


For mAs ± 4 mAs for mAs values less than or equal to 40 mAs
± 10 % for mAs values greater than 40 mAs

Measure with scope the voltage from the MAFB TP21 and divide it by
resistance of 19.6Ω for large focus and by 78.4Ω for small focus.

For exposure ± ( 0.1 second + instrument inaccuracy - tube current


time reduced with small breast thickness)

66 Instrumentarium Imaging 86502-4AA


6 System checks and preventive maintenance

6.2 Automatic exposure control evaluation


Test equipment needed:

– Sensitometer
– Densitometer
– Acrylic blocks
– Film & Cassette
1 Attach the Bucky device.
2 Enter into the service mode by pressing kV- and Den-buttons down
simultaneously for 3 seconds.

6.2.1 Quick AEC test

Testing procedure:

1 Attach the cassette holder.


2 Position an acrylic plate carefully so that it covers the AEC detector.
3 Select such a kV that will produce the clinically suitable optical density
at 4cm thickness.
4 Make an exposure using the Auto Time-mode.
5 Develop the film.
6 Measure the film density. If the density is not what you wish it to be,
reprogramming of AEC is recommended, but record (export) the system
settings before changing anything.

NOTE
Carry out this test once a week.

6.2.2 Exporting and storing the system settings data

The system settings can be exported in hexadecimal format and can be


viewed and stored with Diamond Service Software (87502) or Windows
HyperTerminal. For HyperTerminal properties setting and about how to
capture, refer to chapter 4.3, Hyperterminal interface for Diamond.

6.2.3 Automatic AEC detector gain adjustment

In order to adjust the gain of each detector, follow these instructions:

1 Place about 40mm acrylic on the cassette holder. Use 28 kV and Auto
Time mode and make an exposure.
2 Enter the service mode and “AEC” from the program groups list. Then
select “8:Detector Adjustment” from the AEC list. Diamond compares
readings from other detectors and then calculates new factors for them
in relation to C1.
3 Answer now YES if you have followed the steps 1 and 2.
4 Scroll display and check that new gains seem reasonable.
5 Press Accept , if new gains are valid.

86502-4AA Instrumentarium Imaging 6-7


6 System checks and preventive maintenance

6 Finally, the program asks for confirmation on the system setting


changes.

After the setting, the values could be as in table below:

Det 0/ Det1/ Det2/ Det3/ Det4/ Det5/ Det6/ Det7/


L1 C1 R1 L2 C4 C3 C2 R2
Factor 0.98 1.00 0.98 0.98 0.82 0.93 1.03 0.98

CAUTION
The AEC Gain ‘g’ parameter will affect all exposures regardless of used filter
or cassette holder. If you want to change this, change filter settings in D:AEC/
Optical Densities (PrD03).

6.2.4 Adjustment of an individual detector

Individual detectors of the Diamond AEC can be adjusted separately in


service program PrD02.

For example, if Det1 reading is 1600 and Det2 reading is 1400, then Det2
amplification must be changed: 1600 times 0.98 divided by 1400 = 1.12,
which after second exposure gives a Det2 reading of about 1570.

Note that above-mentioned detector factors are experimental results for most
film/screen/processor combinations. In order to verify these, measure O.D.
and adjust the gain for each detector separately. Especially detector C4 may
change because of the heel effect.

6.3 Mechanical checks


Perform Planned Maintenance Inspection (PMI) annually. Refer to
Appendices-section in Diamond Installation Instructions (86501).

6.3.1 Compression system

The part of the equipment, which undergoes the most use is the compression
system, so it needs regular attention and treatment when necessary. Spread
some grease (MBDE or other high-quality grease) on the compression
driving axle and support joints. If the driving wheels are noisy or the hand
wheels difficult to rotate, we strongly recommend that you remove the C-
Arm and grease the inside of the compression system. Check also the digital
compression display according to the test described in the Installation
Instruction.

6.3.2 C-Arm rotation mechanism

C-Arm rotation mechanism checks are performed during PMI.

68 Instrumentarium Imaging 86502-4AA


6 System checks and preventive maintenance

6.3.3 Column and carriage


1 Check all cables, especially the cables coming from the C-Arm. Check
also the counter weight wire condition.
2 Check that the carriage moves up and down smoothly. If necessary,
adjust the carriage and/or counterweight wheels.

6.3.4 Compression force check

Place a scale under the paddle and compress it like a breast using hand
wheels. The mechanical clutch should stop the compression movement
before the pressure reaches 250 N (55.1 lbs.). If the compression exceeds this,
adjust the clutch by rotating screw as instructed in chapter 7.3.13, Adjustment
of the mechanical compression limit. Rotate with 13 mm wrench.

6.3.5 X-ray field alignment check


1 In the service mode, select the manual mode with 25 kV and 20 mAs.
2 Place a loaded cassette on top of the cassette holder so that the cassette
overhangs 20 mm on the chest wall edge of the holder.

20mm
ruler
cassette

Bucky

3 Mark the chest wall edge of the film with a ruler placed 20 mm back
from the chest wall edge of the film. (The ruler now shows the actual
chest wall edge).

86502-4AA Instrumentarium Imaging 6-9


6 System checks and preventive maintenance

4 Make an exposure and develop the film. The exposed area between the
ruler and the chest wall edge of the film must be less than 6 mm and
more than 1 mm in USA, less than 3 mm and more than 0 mm in
Europe. If you need to adjust the X-ray field, refer to chapter 7.4.1, X-
ray field adjustment for instructions.
5 Check separately for each cassette holder (small Bucky, large Bucky,
Magnification etc.).

6.3.6 X-ray field light alignment check

The x-ray field adjustment must be checked and aligned prior to the x-ray
field light alignment.

1 Place a loaded cassette in the holder.


2 Enter the service mode by pressing the kV- and density-buttons
simultaneously and holding them the service screen appears.
3 Select manual mode at 25 kV and 20 mAs.
4 Place a sheet of white paper on the cassette holder so that the light field
can be seen clearly.
5 Drive the tube head forward to turn on the field light.
6 Place some coins at the corners of the illuminated area on the cassette
holder.
7 Make an exposure and develop the film.

The coins outlined on the film should not be more than 5 mm out of the
exposed field and sum of misalignments on opposite sides not more than 10
mm. If the light field needs adjustment, refer to chapter 7.4.2, X-ray field
light alignment for instructions.

6 10 Instrumentarium Imaging 86502-4AA


6 System checks and preventive maintenance

6.4 Regular maintenance plans


Refer to service agreement what kind of maintenance plan has been ordered.

Daily maintenance plan


QA check.

Weekly maintenance plan


Check all compression paddles for any cracks or broken surfaces.
Quick AEC test

Monthly maintenance plan


Confirm that the exposure indicator light stays on for the length of the
exposure.
Confirm that the buzzer stays on for the length of the exposure.
Confirm that the exposure switch requires continuous pressing to maintain
the exposure and that releasing of the switch terminates the exposure and
that the timer display indicates “Rel”.
Inspect and confirm that all warning labels and embossed, painted or other
wearable labels have not been defaced or worn and that they are legible.
Confirm that the ready light is on when various modes are selected. A daily
QA check is recommended. Software monitors the status of each detector an
prevents an exposure with a faulty detector.
Check that the carriage moves smoothly and in both directions.
Check that the compression force has not changed.
Check that the compression paddle drives up for approximately 5 seconds
when the mains voltage is disconnected.
Check that the C-Arm moves smoothly in all positions. Check that the
displayed angle corresponds to the selected angle. Also check that ±15°
motorized stereo angulation movement is functioning correctly.
Check that all buttons on the control panel are functioning and check that all
lights and display segments function immediately after turning the main
power on.
Place a 60 mm acrylic block on the cassette holder. Drive the compression
paddle down against the acrylic blocks. Check that the thickness display
indicates 6.0 cm. If not, call for service.
Switch on the x-ray field illumination light (46) and check that it covers the
imaging area.
Pay attention to any strange sounds that your unit may make inform your
service technician about these immediately. if you have any doubts about the
correct safe operation of the unit, switch it off and do not use it until you can
it checked by an Instrumentarium approved Service Technician.
Check that the emergency stop buttons function correctly. Rotate the C-Arm
or move the compression paddle using a foot pedal and press one of the
emergency stop buttons (18). All motorized movements should stop. Release
the emergency stop button, switch on the unit once again and then repeat the
test using the emergency stop button on the opposite side of the unit.
Check that Tube Park moves smoothly and stops at exact, correct position.

86502-4AA Instrumentarium Imaging 6-11


6 System checks and preventive maintenance

Annual maintenance plan


Once a year Procedure/Section in Diamond
Installation Instructions (86501)
kV/mAs/preheat tests Perform tests 2.12.7, 2.12.8 and
2.14.8.
X-ray and light fields Perform tests 2.14.10 and 2.14.11.
AEC adjustment Perform test 2.14.14.
Safety ground Verify that the metal parts of the unit
are connected to protective ground.
Oil leaks Check that the high voltage unit and
the tube housing do not leak oil.
Anode run voltage Perform test 2.12.6.
Compression force Perform test 2.14.5, 2.14.7 and
2.14.8.
Interlock system Check that exposures are inhibited if
cassette holder or the cassette is not
installed. Check the error codes.
Clean the High Voltage Connections and grease with clean grease for
example DowCorning #4.
Lubricate the counterweight wires with a light coat of “STP” or “CRC”. Do
not apply de-greasers or cleaners. Replace if broken strands are found.
Recommended wire replacement interval is 5 years.
Tighten ground connections. Inspect crimps, connections and cables for
wear or breakage.
Check for the operation of the C-Arm gear. Lubricate, if necessary.
Check the collimator, filter mechanism and movement.
Check the stereo drive mechanism.
Check Tube Park alignment.
Update the tube efficiency software parameters.
Remove debris from the cassette loading mechanism and lubricate it.
Remove dust from the fan.
Exposure Reproducibility (mR/mAs)
Focal Spot
If the unit is been used very much, apply grease (for example, Optimal
HT2EP Optitemp(®) to lubrication tip of the C-Arm rotation system. If the
unit is been used less, then lubricate every two years.
Perform Planned Maintenance Inspection (PMI). Refer to Appendices-
section in Diamond Installation Instructions (86501).

6 12 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7 Repair instructions
7.1 Column repair

CAUTION
During work around the column, be careful not to drop any parts into it.

WARNING
The counterweight is heavy. Be careful when handling it or the column to
avoid any injuries to the back or other parts of the body.

WARNING
Do not place your hands inside the column while moving the carriage.
Serious injury may occur.

WARNING
Make sure that the support holds the carriage steadily and that the material
used to tie it up with the column does not break, rip or anyhow get damaged
so that that it allows the support to fail. In such case, serious damage to
service technician(s) or to the unit is likely to result.

7.1.1 C-Arm movement friction adjustment


1 Remove the plastic cover.
2 Adjust the 13mm screw with extended hex pin tool. Check that the
clutch slips when the C-arm is pressed down with a 50 kg load and holds
steady when pressed down with 30 kg load.
3 Attach the plastic cover back.

86502-4AA Instrumentarium Imaging 7-1


7 Repair instructions

7.1.2 Preparing the carriage and counterweight


1 Insert the assembly pin.

2 Lift the c-arm up a bit and place a support between the base plate and the
c-arm. Tie up the support with the column with some strong material so
that it will steadily hold the c-arm. Counterweight wires should be quite
loose so that you can lift the column cap up.

7.1.3 Replacement of the Z-motor


1 Prepare the carriage as instructed in chapter 7.1.2, Preparing the carriage
and counterweight.
2 Open up screws that hold column cap. Then pull it up gently.

3 Open up these screws and replace the Z-motor.

4 Place the column cap back. Check the alignment by running the motor
with slack wires before installing the cap assembly.
5 Secure the screws.

72 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.1.4 Replacement of the column cap


1 Prepare the carriage as instructed chapter 7.1.2, Preparing the carriage
and counterweight.
2 Repeat instructions in chapter 7.1.3, Replacement of the Z-motor,
Replacement of the Z-motor.
3 Unwind bolts (on both sides of the carriage) that secure the
counterweight wires to the carriage.

4 The pull the old column cap out.


5 Insert the new column cap and follow instructions chapter 7.1.5,
Replacement of the counterweight wires, step 8 and further.

7.1.5 Replacement of the counterweight wires


1 Prepare the carriage as instructed chapter 7.1.2, Preparing the carriage
and counterweight, Preparing the carriage and counterweight.
2 Detach counterweight wire from both sides of the carriage.

CAUTION
Release carefully so that parts would not drop down to the column opening
behind the wire. Pieces in the opening may seriously affect the movement of
the carriage.

3 Open up a bolt and a nut indicated by arrows. Then pull out the wheels
and the counterweight wire comes off the cap.

86502-4AA Instrumentarium Imaging 7-3


7 Repair instructions

4 Release three bolts from the top of the counterweight cage.

5 Loosen the plastic piece that locks the wire with the metal pin. Remove
the wire.
6 Remove the plastic piece and pull the warning pin out.

7 Pull the wire out of the metal lock piece. Slide a new wire in through it.
8 Attach the column cap back on. Place the counterweight wire lock piece
between the wheels and drag it down towards the counterweight .

9 Attach counterweight wire to both sides of the carriage.

CAUTION
Attach carefully so that parts would not drop down to the column opening
behind the wire. Pieces in the opening may seriously affect the movement of
the carriage.

74 Instrumentarium Imaging 86502-4AA


7 Repair instructions

10 Insert the warning pin and the plastic lock piece. In case one of
counterweight wires breaks up, the remaining wire pulls the warning pin
to the other side and it starts to scratch the column. This is a warning
signal for wire replacement.

11 Lay out the wire evenly on the top part of the counterweight cage.
Tighten up the plastic lock piece. Add cover plate.

12 Tighten three bolts a bit. Pull the wire out a bit to remove slack. Secure
the wire with three bolts.

86502-4AA Instrumentarium Imaging 7-5


7 Repair instructions

7.2 Electrical board and mains transformer locations


The location of Diamond boards is indicated in the image below. If you
replace EPROM chips, use microchip extractor tool.

Board code Board name Image symbol


32305 AMD Board B
34054 X-ray tube, M113SP L
81017 HV unit K
85001 AEC Board A
85003 Generator Board G
85009 Collimator Board C
85011 Drive Board F
85013 Stereo Board I
85020 Power Board H
85022 Compression Board D
85024 Compression Display Board E
85078 Supervisor Board J
85065 Csense Board M
85090 Casopto Board N
85069 Control Panel Board O
85032 Collimator Sense Board P

76 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.2.1 Replacement of Power, Supervisor and Generator Boards

NOTE
If you replace the Supervisor Board, all parameters must be updated to the
Supervisor Board. The optional Diamond Service Software (87502) is the
best tool for this procedure.

1 Unplug the mains cord.


2 Remove the side covers.
3 Disconnect all cables from each board.
4 Remove the plastic nuts from the corners of each board, and replace the
board.
5 Check kV, mA and preheat. Refer to chapter 6.1.5, kV, mAs and
exposure time test.

7.2.2 Replacement of the AMD Board


1 Unplug the mains cord.
2 Remove the upper cover of the tube head and disconnect all cables
leading to the anode motor drive board.
3 Replace the board, reconnect all cables, and check the assembly.
4 Check the run voltage (refer to Check items in Diamond Installation
Instructions (86501)).

7.2.3 Replacement of Control Panel Board


1 Open up the screws at the back of the Diamond Control Panel.
2 Gently separate the front and rear pieces of the Control Panel
3 Disconnect all cables leading to the Control Panel Board.
4 Replace the Control Panel Board. Connect cables back.
5 Attach the front and rear parts back together and secure with screws.

86502-4AA Instrumentarium Imaging 7-7


7 Repair instructions

7.2.4 Replacement of the Compression Board or Compression


Display Board
1 Open up four screws at the bottom part of the Position Control Center

2 Gently separate the top and bottom parts


3 Replace the Compression Board or Compression Display Board.
4 Attach the top and bottom parts back together and secure with screws.

7.2.5 Replacement of the mains transformer


1 Remove the power supply plug.
2 Remove the side covers.
3 Remove all cables connected to the transformer, the nut and cover plate
on the transformer.

4 Replace it with a new one. Plug in cables.


5 Check cables and connections carefully.
6 Attach back the side covers.
7 Plug in the power supply plug.

78 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.3 C-Arm repair


7.3.1 Removal and assembly of the Diamond covers

For removal of the covers, perform the installation process in reverse order.
Start the removal process with the right cover, because the emergency stop
cable is attached to the right side of the C-Arm carriage.

7.3.2 Installation of the tube head covers


7.3.2.1 Optoswitch check

Check that ParkBack optoswitches are intact. ParkBack movement is


detrimentally affected if any of those is bent.

Location Function
Left front (A) Slowdown
Left rear (B) Slowdown
Right front (C) Stop
Right rear (D) Stop

86502-4AA Instrumentarium Imaging 7-9


7 Repair instructions

7.3.2.2 Finger protection sheets

For protection of the user’s fingers, Diamond has sliding sheets on both sides
of the unit to close up the opening caused by the ParkBack movement. The
plastic sheet slides along the plastic rail groove.

Fig 7.1. Finger protection sheet

NOTE
If you need remove the tube cover and then install them back, make sure the
finger protection sheets slide in smoothly and evenly so that they are not
tangled up.

7.3.2.3 Installation of the finger protection sheets


1 Secure the sheets to locations on both sides on front end and the rear of
the unit (indicated by images below).

2 In order to facilitate the process, drive the tube head forward to attach
the front-end finger protection sheets and backward to attach the rear
end sheets.

7 10 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.3.2.4 Installation of the lower front end of the tube head cover
1 Slide the finger protection sheets into the groove on both sides of the
lower front end of the tube head cover.
2 Push the sheets forward and pay attention that the sheets slide smoothly
along the plastic rails.

3 Secure the tube head cover with two screws.

86502-4AA Instrumentarium Imaging 7-11


7 Repair instructions

7.3.2.5 Installation of the rear end of the tube head cover


1 Slide the in the finger protection sheets into the grooves on both sides.

2 Gently lay the rear end cover in its place.

CAUTION
Be very careful with the optos. With rough handling, there is a high risk of
damaging the optos.

3 Check that the sheets align up as in image below.

7 12 Instrumentarium Imaging 86502-4AA


7 Repair instructions

4 Secure screws below.

5 Finally, secure the back screws.

7.3.2.6 Attaching top tube head cover


1 Attach the cord. Plug it in so that the metal pin plates face outward,
towards the serviceman. Be careful that the cord would get wrinkled
between the X-ray tube and the tube head cover.
B
B

A
A

CAUTION
If the connector is attached the wrong way, not only the buttons will not
work, but also the system is at risk of getting damaged.

86502-4AA Instrumentarium Imaging 7-13


7 Repair instructions

2 Lower the rear end of the top cover (A). Then, lay down the front part
smoothly on the lower cover (aligns between the align piece and the
edge of the top cover (B)). Check that top cover lies smoothly (C) before
securing it.

3 Secure the top cover with a hex pin driver. If the screw holes of the top
tube head do not match with the support inside the cover around the X-
ray tube (e.g. if something has bent the support), lift up the top cover and
adjust the support so that the holes align again.

7.3.3 Left side cover


1 Align the left side cover with the carriage and attach it with screws.

2 Secure the frame.

7 14 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.3.4 Right clam shell cover

WARNING
Remember to unplug the override piece when you finish with the system
checking.

1 Connect the emergency stop cable coming from the left clam shell to the
one on the right clam shell. Make sure that the cable is well in place and
during movement such as the ECS, does not get caught in any of the
mechanical parts of the system.

2 Plug the emergency stop cable into the X15 on Power Board. Push it in
all the way until it locks up.

NOTE
Mechanical wear transforms the structure of the grease used in C-Arm
rotation system to slightly more liquid and this physical phenomenon may
produce an oil drip inside, on the bottom of the clam shell. Although this drip
does not affect the operation of the unit, the right clam shell contains a patch
that captures the drip.

86502-4AA Instrumentarium Imaging 7-15


7 Repair instructions

3 Align openings of the clam shell with the carriage and secure it.

4 Secure the clam shells at the top.

7.3.5 Replacement of handles


1 Open up screws to remove the handle in use.

7 16 Instrumentarium Imaging 86502-4AA


7 Repair instructions

2 Attach the handle bar in place.

3 Secure the handle bar with the screws.

7.3.6 Replacement of the vertical drive control switch


1 Remove the Position Control Center cover as instructed in chapter
7.3.13, Adjustment of the mechanical compression limit.
2 Gently separate cover and bottom part of the Position Control Center.
3 Inside the cover, open the screws that hold the vertical switch secured.
4 Replace the switch and secure it with screws.
5 Attach cover and bottom part back together. Secure with screws.

86502-4AA Instrumentarium Imaging 7-17


7 Repair instructions

7.3.7 Replacement of the x-ray tube


1 Remove the tube head covers.
2 Disconnect the plugs of the tube.
3 Mark with a pencil the original position of the tube head in relation to
the collimator. If the new tube is mounted the same way as the old one,
you save quite a lot of alignment work.
4 Unplug these cables and ground wire (attached to the high voltage unit).

5 Release four bolts that fix the tube to the collimator.

6 Replace the tube and secure it with the four bolts. Then attach cables and
ground wire.
7 Check kV regulation. Refer to Diamond Installation Instructions
(86501).
8 Check preheat values for every kV. Refer to Diamond Installation
Instructions (86501).
9 Check AEC calibration.
10 Check collimation of the X-ray field and the X-ray light field as
instructed in chapter 6.3.6, X-ray field light alignment check. Adjust, if
necessary.

7 18 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.3.8 High voltage unit replacement


1 Disconnect the ground wires and the plugs of the high voltage unit, and
loosen the four screws, two on each side of the unit.
2 After replacement, check the assembly and the ground cables carefully.
3 Re-calibrate AEC. Refer to Diamond Installation Instructions (86501)
for AEC adjustment instructions.
4 Readjust the generator kVs again. Refer to chapter 6.1.5, kV, mAs and
exposure time test.

7.3.9 Compression system replacement


1 Remove C-Arm side covers.

2 Drive the stereo arm to approximately 10° angle to either side. Then
drive the C-Arm in 180° position (upside down).
3 Release three bolts connecting the compression system to the stereo
arm.

In order to reach all screws, use a ball-headed Allen key for this procedure.

86502-4AA Instrumentarium Imaging 7-19


7 Repair instructions

4 When the bolts have been released, push the stereo arm further so that
the stereo arm base rises enough to let the compression system piece
slide off.
5 Open up six bolts that connect the compression system to the c-arm.

6 Disconnect cables and slide the compression system out and perform
necessary service procedures.
7 Follow previous steps in reverse order to reinstall the compression
system.
8 Finally, drive the stereo and C-Arm to 0° angle.

7 20 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.3.10 Replacement of the compression motor

In order to replace the compression motor, the compression system has to


detached as instructed in chapter 7.3.9, Compression system replacement.
Then proceed to following steps:

1 Drive the stereo arm to approximately 10° angle to either side. Then
drive the C-Arm in 180° position (upside down).
2 Remove flag (A) to avoid damage (such would likely result if the
compression system cap would rest on the flag). Open two screws (B) to
detach the compression system cap.
A

B
B

86502-4AA Instrumentarium Imaging 7-21


7 Repair instructions

3 Release three bolts that attach the motor to the compression system.

4 Pull the motor assembly out, it will look like this.

7 22 Instrumentarium Imaging 86502-4AA


7 Repair instructions

5 Loosen up the worm drive clip and slide it towards the motor so that the
parts around tacho meter come off.

6 The tachometer is fixed to the motor axle with a screw. Open the screw
and detach the tachometer disc.

Inserting a new motor


7 Replace the old motor with a new one.
8 Attach the tachometer disc to the motor axle with a screw.
9 Assemble the rest of the tachometer parts so that it will look like this:

10 Fix the motor with three bolts that were released in step 2.
11 Reattach the compression system cap that was detached in step 1.
12 Attach the compression system back to the stereo arm as instructed in
chapter 7.3.9, Compression system replacement.

86502-4AA Instrumentarium Imaging 7-23


7 Repair instructions

7.3.11 Replacement of the compression thickness measurement


system

The compression thickness is measured with a special tape that is read by two
sensors (front and rear sensor).

Replacement of the front sensor


1 Open up two screws
2 Solder the cable off from the old sensor
3 Remove the sensor
4 Solder the cable to a new sensor
5 Lay a new sensor in place
6 Tighten up the two screws

7 24 Instrumentarium Imaging 86502-4AA


7 Repair instructions

Replacement of the rear sensor


1 Drive the stereo arm so that an opening marked in the picture becomes
visible.

2 Open the rear sensor lock screws through the hole


3 Solder the cable off from the old sensor
4 Remove the sensor
5 Solder the cable to a new sensor
6 Lay a new sensor in place
7 Tighten up the lock screws
8 Drive the stereo back to 0° angle

86502-4AA Instrumentarium Imaging 7-25


7 Repair instructions

7.3.12 Adjustment of the compression mechanism belt

The compression belt should bend about 1cm, when pressed with finger. If
not, adjust the belt.

1 Open up the lock screw (A).

2 Adjust belt tightness (B, underneath the motor).


3 Tighten up the lock screw (A) to fix belt tightness.

7 26 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.3.13 Adjustment of the mechanical compression limit


1 Open the four screws to remove the compression arm cover.

2 If the mechanical limit releases with too low compression force, tighten
the mechanical clutch adjustment nut with a 13mm wrench. If the
compression allows to apply too much force, loosen the nut. Set the
limit at 250N.

3 Attach the cover back in its place.

86502-4AA Instrumentarium Imaging 7-27


7 Repair instructions

7.3.14 Replacement of the C-Arm motor


1 Remove Generator Board.
2 Open up screws that hold C-Arm motor assembly (A). If more space is
necessary, remove Supervisor Board.

3 Check if there is slack between clutch joint (B) and the axle (C). If there
is no slack, then just pull off the C-Arm motor assembly. If there is
slack, then loosen screw D and pull the assembly off.

F
B B E
D
C C

4 Insert the new C-Arm motor assembly and secure it with screws (A). If
there was no slack between B and C, then the old B remains in place. In
this case, remove B from the new C-Arm motor assembly (leave rubber
part E with clutch joint F) and insert the new assembly and secure with
screws (A).
5 Attach Supervisor Board back.
6 Check kV, mA and preheat. Refer to chapter 6.1.5, kV, mAs and
exposure time test.

7 28 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.3.15 Replacement of ECS mechanism/ECS motor

12 In order to replace the ECS system, perform steps 1-11. In order to change the
ECS motor, perform all steps.

1 Drive ECS half way and then lock up the counterweight with the C-Arm
assembly pin.
2 Remove Stereo Board cover (1) and side covers (2).
3 Then cut the wrap tie and push
the cable tubes (3) upward so
that you get more
3 maneuverability for the C-Arm
rest.
4 Prepare a steady place right
next to the unit where the C-
Arm can rest while you change
the ECS-motor.
5 Unplug X2 and X15 from
Stereo Board. Then, gently pull
the X15 cable out a bit.

6 Remove topscrew (A). Loosen up others (B, C). Perform this on both
sides.

86502-4AA Instrumentarium Imaging 7-29


7 Repair instructions

7 Lift the C-Arm a bit upwards and


pull it out very gently because the
Delta 32 cable is located inside the
C-Arm tube.

CAUTION
Be careful not to exert sudden or strong pulling force on Delta 32 cable.

8 Lay the C-Arm on some steady surface that is very close to the unit.
9 Open up these 8 screws.
10
10 Open up these two screws on both sides.
11 Hold steady the wheel marked with dotted line and lift up the motor.
12 Open up motor screws.

11

13 Replace the old motor with a new one. Secure the motor to the plate.
14 Repeat steps 4-13 in reverse order. At step 12, make sure that the belt is
tight enough to bear stress without jumping over. Apply some grease to
the trapezoidal shaft (for example, Optimal HT2EP Optitemp®).
15 Check kV, mA and preheat. Refer to chapter 6.1.5, kV, mAs and
exposure time test.
16 Attach covers back on.

7 30 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.4 Other adjustments or replacements


7.4.1 X-ray field adjustment

The Diamond units with S/N 20193 and up are delivered with the new single-
screw collimator version. The older units are equipped with a dual-screw
collimator version.

Old dual-screw collimator

The smaller Allen screws (B) act as "push" adjusters. In other words, as you
turn the screw clockwise it "pushes" the collimator back toward you. The
larger setscrew & nut combination (A) act as hanger & locking hardware. The
collimator box hangs on the setscrews and is locked with pressure from
opposite sides by the nuts.

Old S/N20000 - 20192 New 20193 and up

7LJKWHQQXWWRKROG
FROOLPDWRULQSODFH
7XEH0RXQW)UDPH
IL[HG
7XUQ&:WRPRYH
FROOLPDWRUOHIW

&ROOLPDWRU6LGHZDOO
PRYHDEOH 6LQJOH6FUHZ'HVLJQ

&ROOLPDWRU0RXQW
&XW$ZD\9LHZRI/HIW6LGHZDOO

Before you can "push" the collimator box back toward you (tighten the
adjuster), you have to loosen the locknuts on the same side and loosen the
push adjusters on the opposite side.

If the Left/Right movement hangs, check for drag from the setscrew/nut at the
front of the collimator box (front/back adjuster); loosen the nuts slightly.

Things to check that might limit the range of motion:

– adjusters or locknuts are not loose enough


– drag coming from Forward/Backward adjusters
– interference between step-motors and brackets, chassis, cable
– harness, opto board, mirror solenoid, linkage, etc.
– free space between tube mounting frame (fixed) and the inside wall of
the collimator box (moveable)
– the Left-Right length of the slotted holes for the front/back adjusters

86502-4AA Instrumentarium Imaging 7-31


7 Repair instructions

NOTE
Remember to check mirror solenoid operation after moving the collimator
box because the solenoid coil does not move with it (it is mounted to the
chassis).

New single-screw collimator

The setscrew has a shoulder-cut that steps from a larger dia, which is threaded
into the tube mount frame, to a smaller dia, which protrudes through the col
box and accepts the locknut. The push adjusters are eliminated (except for
one left over to hold the mirror spring).

The basic alignment of any x-ray equipment involves 2 steps, align the light
field to the x-ray beam and then use the light to center the x-ray beam on the
receptor. Finally, you take exposures showing the lite vs beam alignment and
beam vs film alignment.

If the beam is not centered on the bucky, you can decrease the side plate
values until the patient left side plate is at the film edge in order to hide the
offset. This puts the x-ray far over the patient right side of the bucky.

Adjustment of the x-ray field


1 Adjustment screw regulates the front edge of x-ray field.
2 Collimator rear plate and side plates regulate side and rear edges of the
x-ray field. Refer to chapter 4.2.6, F: Collimator and readjust the values
for step motor movements.
3 The arrowed bolts in images below allow mechanical adjustment
sideways, if necessary.

7 32 Instrumentarium Imaging 86502-4AA


7 Repair instructions

4 Check the X-ray field alignment as instructed in chapter 6.3.5, X-ray


field alignment check. Re-adjust, if necessary.
5 Check that mirror operates correct. As mirror is secured to the tube head
body, collimator movement may affect the mirror function.

7.4.2 X-ray field light alignment

The light field can be adjusted with two screws at the collimator lamp lid.

Fig 7.2. Collimator light field alignment screws.

The light field may also be adjusted at the solenoid stop (A). If you opt to do
this, be careful not to disturb the operation of the solenoid (if it is, try to
adjust solenoid position with adjustment screws (B)). After adjustment, check
the X-ray field light alignment as instructed in chapter 6.3.6, X-ray field light
alignment check.

86502-4AA Instrumentarium Imaging 7-33


7 Repair instructions

Illumination adjustment

The voltage for the field light can be measured at conductor stands (indicated
by arrows in the image below). The voltage can regulated in service program
PrF03. Set voltage at approximately 10 volts.

NOTE
Values 0 - 10 in the service program PrF03 do not indicate volts. These just
steps to regulate the voltage.

7.4.3 Replacement of the light field bulb

For instructions to change the light field bulb, refer to Replacement of the
light field bulb in Diamond Operation Instructions (86500).

7.4.4 Tube head movement and torque adjustment

Exact exposure position of the tube head (tube park stop) may be adjusted
with the nut (A) and the torque of the movement with bolt (B). Set the torque
at about 50-60N.

7 34 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.4.5 Tube head belt check


1 Check with a thumb and a forefinger if the tube belt is tight enough. If
not, proceed to next step.
2 Release the lock nut.

3 Tighten the eccentric nut until the belt is tight enough.

4 Secure the lock nut.

86502-4AA Instrumentarium Imaging 7-35


7 Repair instructions

7.4.6 Grid alignment

The grid is aligned at the factory and should normally not be adjusted at the
site. If adjustment and/or alignment is necessary, however, follow the
instructions below.

1 Remove the carbon fiber cover on the Bucky. In the small Bucky the
grid should be at the bottom of the grid drive block groove and fastened
with four setscrews. In the large Bucky the grid should be fastened only
with two setscrews and aligned by the two rear set screws.
2 Check that the grid is correctly centered. The line on the grid is aligned
with imaginary dashed line (A).
3 Install the Bucky and check that the grid does not touch the side lips of
the Bucky (B). Check also that the grid is aligned parallel to the side lips
(C).

NOTE
If the grid must be removed for some reason, install it back exactly the same
way as before. Check for screw marks.

4 If the intermittent movement of the grid reaches one of the side lips,
remove the Diamond c-arm bottom cover.

7 36 Instrumentarium Imaging 86502-4AA


7 Repair instructions

5 Loosen two screws (marked in the picture below).

6 Slide the grid system to the opposite direction from the side lip.
7 After the adjustment, tighten up the screws and check the movement.
Before initiating the exposure, attach the service diaphragm. For
exposure, use manual mode and set for example 20kV and 200mAs.

If the problem persists, readjust. If not, attach the bottom cover.

8 When the Bucky is installed, the grid system drives to home position.
Check that the grid system movement is not blocked by the edges of the
opening.

7.4.7 Grid movement mechanism check

The first grid movement from left to right takes up about 670-820ms. The rest
of the movements take up about 3070-3260ms. Exposure stops for 150ms
when grid switches the direction of movement. If grid lines appear, check that
the wheels that connect the motor to the grid are correctly aligned and
secured.

1 Remove Bucky and plastic cover (held by eight screws). Then open up
four screws that hold the cassette holder table.

86502-4AA Instrumentarium Imaging 7-37


7 Repair instructions

2 Gently pull the cassette holder table out a bit. You may need to unplug
cables CX92, CX93, CX96-98, CASX1, CAS, DRX5, DRVX1, GRID,
AECX2 (near AEC detectors) and AECX3 (near AEC detectors). In
order to reach some of the cables, remove the bottom cover.

3 Connection wheels (A) should rotate without obstructions, not be


misaligned, loose or too close to the grid mechanism frame or the opto
flag sensor cable. If they are, loosen up screws that fix the wheels to
motor axis and grid shaft. Then slide them a bit towards to the motor.
Tighten screws a bit to test the rotation. If the wheels rotate OK, secure
the screws.

A B

4 Press inwards the pin that moves the grid. It should slide in and out
smoothly. The pin should protrude the frame by 4mm (C). Check also
that the pin is tightened up all the way (D). Hold the pin between fingers
and slide the mechanism to left and right to sense any slack.

C D

7 38 Instrumentarium Imaging 86502-4AA


7 Repair instructions

5 Drive the mechanism to extreme positions and check that the opto flag
sensor does not disturb the movement of the opto flag. Check that flag is
correctly aligned and does not touch grid mechanism frame.

6 If the problem still persists or there was significant slack at step 4,


replace the movement mechanism assembly. Release motor screws (A).
Loosen up screws that fix connection wheel package to motor axis (B).
Slide out the motor (C). Open two screws that fix the mechanism to the
cassette table (D).

D
A
B
C

7.4.8 Replacement of Bucky carbon fiber

If the carbon fiber of the Bucky has become scratched or otherwise damaged,
it must be replaced. The cover is attached with 6 screws (two on topside, two
on sides and two at the bottom). Open up screws and slide out the cover out.
Then slide a new cover on and secure with screws.

7.4.9 Replacement of c-arm membrane buttons

Membrane buttons can be replaced by unplugging flat cables inside the c-


arm.

Then remove the button and residues, if any. Then attach a new one in place
and plug it in.

86502-4AA Instrumentarium Imaging 7-39


7 Repair instructions

7.5 Compression adjustment


Diamond units with S/N 20000 - 20135 have older compression mechanism.
If you need to service these, refer to Diamond Technical Bulletin 01/2001 for
information. Current compression mechanism (S/N 20136 and up) does not
need adjustment.

Fig 7.3. Current compression mechanism

7 40 Instrumentarium Imaging 86502-4AA


7 Repair instructions

7.6 Negative compression force calibration


Tools necessary:

– normal service tools


– compression meter (scale)
– 3 kg weight and 0,5 kg weight

How to do the software calibration:


1 Open upper cover of the Position Control Center but do not remove
cables. Place the cover so that electrical components will not short-cir-
cuit.
2 Turn the Diamond unit on from the mains switch.
3 Attach compression paddle (recommended to use the heaviest one).
Apply 3 kg weight on the paddle.
4 Enter service program PrG06, G:Compression/Up Force Limit.
5 Set the value on the screen with trimmer R10 (below the position control
center cover at 10. Do NOT press Accept yet.
6 Place back the upper cover.
7 Apply 0.5 kg weight on top of the 3 kg weight and check what is the
value now.
8 Remove the 0.5 kg weight and then press Accept.
9 Set the force same as it was in step 7. Check that the value is not 0. If it
is, change it to 1. Screen will display “pass”, if everything is OK.
10 Exit the service mode and return to the normal mode.
11 Test for negative force: apply 0.5 kg again on top of the 3 kg weight and
verify that error 24 will appear.
12 Remove all weights and press the paddle down manually (towards the
front end of the Bucky) and verify that error 24 will appear.
13 Perform paddle sensor test in service program PrG02. Attach all
different paddles that the customer uses and check that the unit detects
those correct.
14 In service program PrG02, you can scroll up and down to see the
asterisk (*) in front of selected paddle. It indicates the selected paddle.
15 If the right paddle is not detected, open the upper cover of the Position
Control Center and bend Hall transistors (sensing code magnets) a little
bit forward and try again. Be careful not to break the pins of transistors.
16 Finally, calibrate the compression force display (refer to Diamond
Installation Instructions 7.2.6).

Final test:
1 Press paddle down by hand (towards the front end of the Bucky) in nor-
mal mode and verify that error 24 will appear.
2 In the end, drive C-Arm to normal and upside down position and verify
that the error message 24 will not appear.

86502-4AA Instrumentarium Imaging 7-41


7 Repair instructions

7 42 Instrumentarium Imaging 86502-4AA


8 Error tracing

8 Error tracing
8.1 Information for the service error tracing
Ask always the following questions during a service call from the customer.
This information is required if you contact the manufacturer.

A General (required information)


1 Serial number of the unit.
2 Name of the operator who experienced the problem.
3 Software revision level. (Shown on the LCD-display when power is
switched on)
B For phototiming questions (AEC)
1 Film/screen combination used.
2 Processing time and temperature of the developer.
3 What is the gross fog, speed index, contrast index of the film developed
in the processor?
4 Was any work done on processor just prior to problem?
5 Did problem appear suddenly or gradually?
6 Breast compression, kV selected, type of view, mAs of exposure,
detector positioning, filter selected, which imaging mode?
C Other problems
1 What is the history of the problem?
2 Error codes on the display?
3 What have you done to attempt a fix?
4 Is it permanent failure or intermittent?
5 If intermittent, how often per day?
6 Did problem appear as a result of some other work performed on the
unit?
7 Is any other equipment exhibiting intermittent problems?

8.2 Diamond error codes


Time-out limit in cases of communication is set at 40 sec. For all other
situations it is set at 20 sec. There is no error message on the battery reaching
the end of its lifetime.

86502-4AA Instrumentarium Imaging 8-1


8 Error tracing

Five quick beeps (Error message 54) indicate that the Control Panel is
missing. Attach the Control Panel to the Diamond unit.

8.2.1 0002 Program check sum error

Error code 0002


Error
description
Potential Program check sum error (Flash).
reasons
How to test
Check items

8.2.2 0003 Program check sum error

Error code 0003


Error
description
Potential Program check sum error (BattRAM).
reasons
How to test
Check items

8.2.3 0011

Error code 0011


Error
description
Potential ‘Motor stop’ pressed or a Power Board problem. Release
reasons emergency button.

How to test
Check items If problem persists, check alarm to Power Board and
Supervisor Board.

82 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.4 0012 Tube Park safety clutch active

Error code 0012


Error Tube Park safety clutch activates due to a tube obstacle,
description microswitch misalignment or a collimator Board
problem.
Potential
reasons
How to test 1 Adjustment of the microswitch of tube park safety
clutch should be checked and readjusted if necessary
(refer to chapter 7.4.4, Tube head movement and
torque adjustment). Check that the tube moves
without unintended friction.
2 Check also that tube movement belt is tight enough. If
necessary, the belt can be tightened with 13 mm
wrench at an eccentric nut behind the collimator.
Check items 1 Check that safety clutch microswitch sensor and also
all tube park opto sensors have remained intact (use
INPUT PROGRAM for testing).
2 Check that the cables underneath the upper cover
remain free and are not tangled up with some other
objects.
3 Check that white plastic finger protection sheets
within the lower cover do not generate extra friction
(lubricate for example with silicon spray).

8.2.5 0013 Tube movement

Error code 0013


Error Tube movement back up time reached.
description
Potential 1 Mechanical friction
reasons 2 Optos/flags
3 Collimator Board
4 Motor
How to test
Check items 1 Check that all 4 tube park opto sensors and flags
operate correct (rear, front and 2 for slow the speed -
use INPUT PROGRAM) for testing.
2 if problems persist, check that wiring connections on
optoboard are correct and flags slide smoothly in the
optosensors.
3 Check that the movement should be free of any
friction all the way.

86502-4AA Instrumentarium Imaging 8-3


8 Error tracing

8.2.6 0014

Error code 0014


Error
description
Potential Filter movement back up time reached.
reasons
How to test
Check items Check opto flags.

8.2.7 0015 Collimator plate movement

Error code 0015


Error Collimator plate movement back up time reached,
description message sent to Supervisor Board.
Potential 1 Opto flag
reasons – the flag is very thin and the positioning might
just be critical and not always cover the opto
light
– in some cases, the height of the opto card is
critical
– if necessary, the opto card can be adjusted up or
down
– wire/soldering/connector is bad/loose
2 Collimator Board.
– check step motor lines
– feedback line from the home position
3 Motor
– no problems yet
– wire/soldering/connector is bad/loose
– cables or a mechanical block can accidentally
block the movement to the home position
– check signal status with input program
4 Set steps with service software for rear collimator:
– verify that steps are reasonable; an example of
such values are printed in the end of Installation
Instructions (86501)
5 Sensor
6 Opto
– Opto itself. Some dirt on opto (grease, dust etc.),
try to clean the opto
How to test 1 Use INPUT program to test optosensor for home
position.
2 Inspect visually that the movement stops when rear
collimator motor runs flag over the opto.
Check items Check collimator plate movement.

84 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.8 0022

Error code 0022


Error
description
Potential Compression motor back up time reached.
reasons
How to test
Check items 1 Check sensors.
2 Check Compression Board.

8.2.9 0023

Error code 0023


Error
description
Potential Compression thickness measurement error (missing
reasons pulses, reading from the strain gages is 0). Exposure is
not allowed in Auto kV-mode. Exposure is allowed in
Auto Time-mode.
How to test
Check items Check opto on the sensor.

8.2.10 0024 Compression force measurement

Error code 0024


Error Compression force measurement failure. Compression
description down inhibited. Release enabled. Manual compression
possible (release enabled).
Potential
reasons
How to test
Check items 1 Check that compression sensor operates correct.
2 Check that compression sensor cables are not broken.
3 Check that Compression Board is not broken.
4 Is the compression calibration valid (as described in
Diamond Technical Bulletin 01/2001, Compression
Hinge Update)?
5 Check strain gages.

86502-4AA Instrumentarium Imaging 8-5


8 Error tracing

8.2.11 0027

Error code 0027


Error Overflow in Auto Time-mode (too long exposure time).
description
Potential 1 kV too high. Try with lower kV values.
reasons 2 Faulty AEC Board
How to test
Check items Check AEC Board.

8.2.12 0029 Underflow in Auto Time or Auto kV-mode

Error code 0029


Error Underflow in Auto Time or Auto kV-mode.
description
Potential kV too high. Try with lower kV values.
reasons
How to test
Check items Check AEC Board.

8.2.13 0033 Cassette tray movement

Error code 0033


Error Cassette tray back up time reached.
description
Potential 1 Friction in the movement.
reasons 2 Cassette tray position optos (3 pcs). Use input
program.
3 Driver Board or Stereo Board problem.
How to test
Check items 1 Check that spot lights or surrounding light do not
disturb sensors. Have the lights in the premises
switched off during the test.
2 Check optos and flags.
3 Check the motor.
4 Check Drive Board

86 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.14 0034 Cassette

Error code 0034


Error Cassette not changed, the cassette not detected.
description
Potential 1 Cassette not changed or placed correct. Change or
reasons reinsert the cassette.
2 Opto sensor area for identification of cassette light
reflection has been blocked for example by a white
label.
3 Cassette sensor not adjusted correct
– adjustment must apply to all combinations used
by the customer
– small Bucky, big Bucky and magnification
tunnel
4 Excessive surrounding light (or spot light).
5 Error occurs in following conditions (all in same
time):
– cassette sensor opto continuously active
– small cassette sensor opto (tray movement opto)
is inactive
6 Large cassette sensor opto (tray movement opto) is
inactive
How to test
Check items 1 Does the error appear with the small Bucky and the
magnification tunnel?
2 Does the error disappear when spot and surrounding
light are switched off?
3 Check optos.

86502-4AA Instrumentarium Imaging 8-7


8 Error tracing

8.2.15 0035 Cassette opto

Error code 0035


Cassette opto failure (this error is displayed in software
version 1.02.5 and up, in earlier software, this error
message is coded as 0067).
Error More than one of three cassette position opto sensors are
description active at the same time. The error message is displayed
when the exposure is initiated.
Potential Cassette is in the front position (one opto sensor) and the
reasons rear cassette sensor opto is still active. When you try to
expose, error code 0035 appears.
How to test Test with input program, which sensors are active.

Eliminate all excessive stray lights entering the cassette


holder area (cassette sensor). In order to eliminate the
light from outside to the sensor, darken the room and test
the cassette sensor.
Check items 1 Check function of the sensor. Check also the C-sense
Board and cable wirings.
2 Does some very strong spot light in the room blind the
sensor?
3 Does some very strong spot light in the room point out
to the Bucky?
4 Find out the C-Arm positions at which you are not
allowed to initiate the exposure. This likely helps to
determine the source/direction of the light that
disturbs the sensors.

8.2.16 0037

Error code 0037


Error
description
Potential Grid error during/prior to exposure. Display after
reasons exposure if exposure started. AEC Board does not
receive signal for grid movement.
How to test
Check items Check grid.

88 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.17 0038 Cassette holder

Error code 0038


Error Cassette holder not attached when trying to exposure.
description Exposure inhibited.

Potential 1 Cassette holder might be installed incorrect (not


reasons inserted until the mechanical stop). Attach cassette
holder.
2 Cassette holder might miss some identification
magnets.
How to test Use INPUT PROGRAM to determine that magnet
sensor operates correct. Refer to Diamond Service
Manual (86502) for the order of 4 hall transistors. Use
cassettes or a separate magnet to simulate the magnets.
Check items 1 Magnets.
2 Hall sensor board.
3 Drive board.

86502-4AA Instrumentarium Imaging 8-9


8 Error tracing

8.2.18 0041 Stereo not in center position

Error code 0041


Error Stereo not in 0° during exposure.
description
Potential 1 Stereo not in 0° when exposing. Drive the stereo arm
reasons to 0°.
The stereo assembly incorporates two middle optos. One
of these two must be active when an exposure is
initiated, otherwise error 41 appears. If these optos are
not broken, enter the service mode (if you use Bucky or
Magnification) or normal mode (Delta 32) and drive the
stereo to the center position.
2 Stereo Board problem.
How to test Check these optos in service mode:

– Press middle button below the LCD display of


the control panel.
– Press right button below display until you see
011 in the upper left corner. The LCD screen
should display the following values:
011 xxx xxx xxx xxx
xxx xxx xxx xxx xxx
xxx xxx aaa bbb xxx
xxx xxx xxx xxx xxx

xxx's are some numbers and "aaa" is stereo left middle


opto and "bbb" is stereo right middle opto. These figures
should be 001 when stereo arm is in center position and
otherwise 000.
Check items 1 Correct opto operation.
2 Soldering on opto PCB.
3 Broken wires or bad connectors on cable CS22.
4 Condition of the Stereo Board.

8.2.19 0043 Vertical movement motor back up time reached

Error code 0043


Error Vertical movement motor back up time reached.
description
Potential
reasons
How to test Clear and try again. If problem persists, call service.
Check items

8 10 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.20 0045 ECS motor back up time reached

Error code 0045


Error ECS motor back up time reached.
description
Potential
reasons
How to test Clear and try again. If problem persists, call service.
Check items

8.2.21 0048

Error code 0048


Error
description
Potential C-Arm rotation back up time reached. Supervisor or
reasons Power Board problem.
How to test
Check items Check optos.

8.2.22 0051

Error code 0051


Error
description
Potential C-Arm angle measurement failure. Drive to predefined
reasons angles not allowed.
How to test
Check items Check tacho optos and C-Arm center position opto. If
problem persists, check Power Board connections.
Check that no strong friction affects the C-Arm rotation.

8.2.23 0053

Error code 0053


Error
description
Potential General communication error with internal or external
reasons electronics, Supervisor bus cable broken (Board not
answering).
How to test
Check items

86502-4AA Instrumentarium Imaging 8-11


8 Error tracing

8.2.24 0054

Error code 0054


Error
description
Potential Communication error with serial port/buffer overflow.
reasons Five quick beeps indicate an error state.
How to test
Check items Check control panel. If it is missing, attach it to
Diamond.

8.2.25 0055

Error code 0055


Error
description
Potential Communication error with RS232.
reasons
How to test
Check items Check that cable is correctly hooked up to COM-port. If
problem persists, the cable might be broken, replace it.

8 12 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.26 0057 Missing mA

Error code 0057


Error mA failure
description
– mA reference missing (set by software,
supervisor > Generator PCB)
– mA feedback missing (feedback loop on
Generator Board)
Potential 1 Supply voltages on Generator Board not correct.
reasons 2 Broken filament(s) inside the tube.
3 Cables short cut or bad connections
4 Bad contact in filament cable between Generator
Board X4 and tube connector (specially on the tube
side).
5 X-ray tube problem.
6 High voltage unit problem
7 Power Board problem.

Components that may need replacement:

– Generator Board
– Supervisor Board
– AMD Board
How to test Expose separately with small and large focus to find out
if both operate correct.

Connect the oscilloscope probe on the Filament Control


Board to test point TP4 (mA feedback, TP12 or GND). If
the mA waveform contains either over- or undershoot at
the start-up, adjust with software in service program
PrI04 (large focus) and PrI05 (small focus) to obtain
correct waveform as instructed in Diamond Installation
Instructions (86501).

86502-4AA Instrumentarium Imaging 8-13


8 Error tracing

Check items 1 Check supply voltages on the Generator Board /


change Generator Board.
2 Check supply voltages on the Power Board / change
Power Board.
3 Measure that filament wiring (connector X 4 on
Generator Board) is not short circuited (remove the
connector X4 and measure with ohm meter from tube
side):
– with large focus, measure the closed circuit
between X4-4 and X4-2.
– With small focus, between X4-3 and X4-2.
4 If not, check that connector X4 is OK and also that
filament connector on the tube side is OK. Test with
other tube, if necessary.
5 Does error condition appear in certain condition?
– in the morning, when power up.
– with certain C-Arm position.
– is led H12 (mA OK) on during exposure.
6 Check relays R1 and R2.
7 Check HV unit feedback cable.
8 Check HV unit, if in lateral position.

8.2.27 0058

Error code 0058


Error
description
Potential Anode start failure. Thermal switch blocks the start-up.
reasons Refer to Operation Instructions. Wait until anode cooling
period is over.
How to test
Check items

8.2.28 0059

Error code 0059


Error
description
Potential Anode run failure.
reasons
How to test
Check items

8 14 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.29 0060

Error code 0060


Error
description
Potential Exposure button released during exposure.
reasons
How to test Hold the exposure button pressed down until the audio-
signal terminates.
Check items

86502-4AA Instrumentarium Imaging 8-15


8 Error tracing

8.2.30 0062 Missing kV

Error code 0062


Error Generator failure, kV not OK
description – kV reference missing (set by software, from
Supervisor to Generator Board)
– kV feedback missing (feedback loop on
Generator Board)
– 34 V ref missing
Potential 1 Supply voltages on Generator Board not correct.
reasons 2 Broken HV-unit.
3 Not enough oil in HV tank.
4 Cable short cut, loose connector or bad wiring
connector
How to test 1 Expose starting with low kV values, increase kV and
monitor for the error appearance.
2 Check if the error appearance more frequent in C-Arm
lateral position.
3 Test changing HV unit.
4 Increase step by step kV, starting with low kV and
monitor for error appearance.
5 Broken tube:
– test separately with small and large focus
– increase step by step kV, starting with low kV
and monitor for the error appearance
Check items 1 Check supply voltages on the Generator Board /
change Generator Board.
2 Check supply voltages on the Power Board / change
Power Board.
3 Check for any visible oil leakages from the tank.
4 Check that cable wiring to connector X2 on Generator
PCB is not broken, check that this cable does not
tighten up mechanically during the C-Arm rotation.
5 HV-cable (bad contact specially on the tube side, use
grease).
6 Measure the cable wiring with ohm meter.
7 Does error condition occur in certain condition?
– in the morning
– with certain type of breasts
– at certain C-Arm position

8 16 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.31 0063

Error code 0063


Error
description
Potential Line voltage low.
reasons
How to test
Check items Check the line voltage.

8.2.32 0064 Missing cassette

Error code 0064


Error Cassette missing.
description
Potential 1 Cassette missing. Insert the cassette.
reasons 2 Incorrect cassette opto sensor adjustment. Adjust the
right signal in Casopto Board with potentiometer.
3 Some white label located in cassette opto sensor area.
Try with other cassette type or remove the white label.
How to test
Check items Does the problem occur with both cassettes and also with
magnification table? Test that all combination operate
correct.

86502-4AA Instrumentarium Imaging 8-17


8 Error tracing

8.2.33 0065 Excessive exposure time

Error code 0065


Error Exposure time too long (CPU estimates that the exposure
description time will exceed the maximum value):

– large focus 5 seconds, small focus 10 seconds


Potential 1 Low kV values. Ask the customer to test with higher
reasons kV values. Increase kV value in initial kV table.
2 Blocked filter frame movement.
3 Cassette movement stops too early.
4 Blocked X-ray field.
How to test
Check items 1 Check that the light mirror operates OK.
2 Check that filter frame movement does not stop in
wrong position (in front of the X-ray beam).
3 Check that the cassette movement will not stop too
early (in front of the X-ray beam).
4 Check for any other objects in the X-ray field.
5 Check alignment of the AEC detector and the cassette.
6 Check filter position.

8.2.34 0066 Movement not allowed

Error code 0066


Error Movement button or switch or exposure button pressed
description during power on or during movement, the function not
allowed.
Potential 1 Power has been switched on and the LCD screen
reasons displays 8888 state. Diamond does not accept any
function, because the unit perceives that some
movement button is pressed.
2 Foot pedal.
3 Z-button in Position Control Center unit.
4 C-Arm button in Position Control Center unit.
5 Projection angle keys (CC, OBL and LAT).
6 Tube park buttons on Diamond side.
7 Stereotactic angle keys.
8 Exposure button.
9 The exposure button is pressed too early, the C-Arm,
stereo or collimation movement not yet completed.
How to test 1 Remove foot pedals
2 Test buttons. Detach membrane cable from connector
to find out if the problems is related to this button.
Check items Check that the exposure button is not active all the time.

8 18 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.35 0067 Diamond not ready for exposure

Error code 0067


Error Diamond not ready for exposure.
description
Potential 1 Device is not ready for exposure.
reasons 2 Rear plate error. The movement might not yet be
completed.
3 Collimation in process.
4 Filter movement in process.
5 Cassette movement opto

In some large Buckies, the grid mechanism reflection


activates the cassette opto sensor even though cassette is
in the front position.
How to test The status of the cassette sensor opto can be checked in
service mode. Press middle button below the Control
Panel LCD screen. Then press right button below the
screen and hold it down until 044 is displayed in the
upper left corner. The LCD screen should display:

044 xxx xxx xxx xxx


xxx xxx xxx xxx yyy
xxx xxx xxx xxx xxx
xxx xxx xxx xxx xxx

xxx's are some numbers and yyy is cassette opto sensor.


The value for cassette opto sensor must be 000 when no
is inserted or when the cassette is in the front position.
The value 001 is displayed when the cassette has
returned in the rear position. If the value 001 is
constantly displayed, then adjust cassette sense opto to
decrease the sensitivity.
Check items

8.2.36 0068

Error code 0068


Error
description
Potential Delta PC refused to receive exposure. Push Exposure
reasons button on Delta 32 computer interface.
How to test
Check items

86502-4AA Instrumentarium Imaging 8-19


8 Error tracing

8.2.37 0069

Error code 0069


Error
description
Potential Delta PC connection timeout. Plug in PC or replace bad
reasons cable.
How to test
Check items

8.2.38 0070

Error code 0070


Error
description
Potential Error in stereo motor pulse measuring or missing pulses.
reasons
How to test
Check items Check motors and encoder optos.

8.2.39 0071

Error code 0071


Error
description
Potential Tube head heat limit exceeded. Wait, the remaining
reasons cooling time is displayed on Control Panel LCD-screen.
How to test
Check items

8 20 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.40 0072 Side collimator

Error code 0072


Error Side collimator back up time reached.
description
Potential 1 Sticky opto flag. Try to release the opto flag to touch
reasons opto sensor.
2 If problems appear only with the large Bucky, then
most likely the side plates are too close to home
position.
– Enter the service mode and increase 100-200
steps to original value in service program PrF02,
F:\ Collimator > Cassette Holder > Side Plate,
selection Bucky 24x30.
3 Mechanical friction with the movement.
– run all the way manually from cogwheel
– check nothing such as cables etc. hinder the
movement
4 Collimator Board.
5 Collimator motor.
How to test
Check items 1 Check opto and opto cables/soldering.
2 Read status of the opto sensor with service input
program. Use manual flag to change the status.
3 Check collimator for any blocking objects.

8.2.41 0074 AEC bus error

Error code 0074


Error AEC bus error 74 occurs, if Supervisor Board does
description receive an answer from AEC Board. Bus is monitored
about every 2.5 seconds.
Potential Internal communication error. AEC Board not
reasons answering. Supervisor-AEC cable broken.

If the same bus does not display other errors


(Compression 75, Stereo 76, Collimator 77, pass through
the C-Arm axis), then most likely the cause is the cable
between Stereo Board connector X14 and AEC Board.

For example, if the cable sheath is broken, an periodic


error might occur when the touches frame. Symptoms
are similar to those of a partly broken compression.

Other possible causes might be periodic supply voltage


problems or a broken AEC Board.
How to test
Check items 1 AEC Board.
2 Cables.

86502-4AA Instrumentarium Imaging 8-21


8 Error tracing

8.2.42 0075 Communication problem

Error code 0075


Error No communication between compression board and
description stereo board.

Potential Internal communication error. Compression Board not


reasons answering. Supervisor-Compression Board cable
broken.

Most likely, the cable SB40 (between Stereo Board and


Compression Board) is broken. Check cable condition at Initial compression cabling.
the lower left corner of the compression carriage. Try to
bend the cable in different positions to produce the error
(a short circuit). A cable short-cut does not always
prompt an error message, but may also cause illogical
values for compression thickness.

Other causes might be a fault in the Compression Board


or the Stereo Board.
How to test
Check items Does this problem occur sporadically? Potential cable
cause.

8.2.43 0076 Later compression cabling with a rotated


holder.

Error code 0076


Error
description
Potential Internal communication error. Stereo Board not
reasons answering. Supervisor-Stereo Board cable broken.
How to test
Check items Check following items on the Stereo Board:

– LED H1: When lit, the Stereo Board is transfering or


receiving data. If not, there is a fault in the Stereo
Board.
– LED H2: When lit, the Stereo Board is receiving data.
If not, the cable is broken, or there is a fault in the
Stereo Board.
– TP4: data burst at 25ms intervals. If not, check for H1
errors.
– TP5: data burst at 5ms intervals. If not, check for H2
errors.
– TP6: three data bursts at 5ms intervals, then 15ms
break, then three burst at 5 ms intervals, then break and
so on.

8 22 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.44 0077

Error code 0077


Error
description
Potential Internal communication error. Collimator Board not
reasons answering.
How to test
Check items Check following items on the Collimator Board:

– LED H14: When lit, Collimator Board is transfering


data. If not, there is a fault in the board.
– LED H13: When lit, Collimator Board is receiving
data. If not, the cable is broken or there is a fault in the
board.
– TP8: data burst at 25ms intervals. If not, check for H14
errors.
– TP7: data burst at 5ms intervals. If not, check for H13
errors.

8.2.45 0078

Error code 0078


Error
description
Potential Internal communication error. Control Panel not
reasons answering.
How to test
Check items Check following items on the Supervisor Board:

– LED H4: When lit, the Supervisor Board is transfering


data to the Control Panel. If not, there is a fault in the
Supervisor Board.
– LED H3: When lit, The Supervisor Board is receiving
data from the Control Panel. If not, the cable is broken
or there is a fault in the Supervisor Board or Control
Panel Board.
– TP9: data burst at 25ms intervals. If not, check for H3
errors.
– TP10: data burst at 25ms intervals. If not, check H4
errors.

86502-4AA Instrumentarium Imaging 8-23


8 Error tracing

8.2.46 0080

Error code 0080


Error Stereo movement back up time reached.
description
Potential – Stereo motor, wiring or control may be broken.
reasons – Some stereo-opto broken.
– Stereo cannot drive due to too slow speed.
How to test
Check items Check optos is Service Mode/Test display (test display
sheets in the Appendix of this manual).

8.2.47 0081

Error code 0081


Error
description
Potential Cassette in rear position. Exposure inhibited. Drive to
reasons front position.
How to test
Check items

8.2.48 0082

Error code 0082


Error
description
Potential C-Arm rotation disabled, if vertical down limit ON and
reasons angle is over 122 degrees. Drive the tube head to the
front position.
How to test
Check items

8.2.49 0084

Error code 0084


Error
description
Potential Transmit buffer overflow in Supervisor software.
reasons
How to test
Check items

8 24 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.2.50 0087

Error code 0087


Error
description
Potential Movement inhibited during compression. Release
reasons compression and try again.
How to test
Check items

8.2.51 0089 C-Arm center limit not found

Error code 0089 (software 1.01.1 and up)


Error C-Arm center limit not found. No detection of 2 center
description opto switches for C-Arm movement home position (0°).

Potential Both opto sensors must be detected, when C-Arm is


reasons rotating over CC position (angle display should show
0°). During approach and depart from the CC-position,
the C-Arm rotation slows down for a while.

1 C-Arm home opto switches (2) underneath the C-Arm


do not operate
– Clean dust, grease or anything. Replace, if
necessary. Order new opto flag (now available 2
mm wider opto for more improved reading, order
code 82761)
– refer also to Diamond Technical Bulletin 06/
2002 for units with S/N 20166 - 20145.
2 Pulse reading of the Supervisor Board opto might be
the cause for sporadic problems
3 If preceding steps do not fix the problem, replace also
the opto cable assembly (use only opto, if necessary).
How to test 1 Test both optos function with INPUT program. Rotate
C-Arm near CC-position and verify that C-Arm opto
input status juggles between 0 and 1. Check that flag
enters the opto deep enough
2 Cover the optos and verify with input program that
sensors are operate correct. Check that you have the
latest Service Manual (order code 86502, issued in
June 2001).
Check items Check optoswitches for c-arm movement.

86502-4AA Instrumentarium Imaging 8-25


8 Error tracing

8.2.52 0090

Error code 0090


Error
description
Potential 1 Tube head is not in front. Drive tube head to the front.
reasons 2 The Auto Drive when the exposure button is pressed,
is disabled in Biopsy, D32 and Service modes or if no
cassette holder is attached .
How to test
Check items

8.2.53 0091 Z-movement disabled in Specimen mode

Error code 0091


Error Z-movement disabled in Specimen mode.
description
Potential
reasons
How to test Clear and continue. If a patient is to be examined, exit
the Specimen mode.
Check items

8.3 No power indications/Board LEDs


A Check the mains switch.
B Check that the power cable and the display cable are connected.
C Check the circuit breaker in the rear bottom.
D Check that Power Board fuses are OK.

8.4 AEC inoperative


– Check grounding
– Check Cassettes/films/screens
– Check detector positioning
– Check HV-grease
– Check spot diaphragm

8.5 Supply voltages


If you suspect changes in Diamond electrical system, check the voltages
according to the list in chapter 5.12, Fuses, voltages and led indications.

8 26 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.6 Grid line check items


A Check cassette holder recognition
– magnets under the Bucky
B Grid alignment and movement
– sides of the grid must not touch the sides of the frames during
movement.
– the grid has to be close to the carbon fiber in the front but may not
touch it.
– is the grid properly attached to the blocks holding it.
– when the breast is compressed, does the carbon fiber top bend and
touch the mechanism.

Refer to chapter 7.4.6, Grid alignment or chapter 7.4.7, Grid movement


mechanism check.

C If there have been random occurrences of gridlines or grid motor runs


badly or makes noise or does not always work correct, then check for the
following:
– In most cases, a Drive PC Board can be the reason for the grid lines.
The older Drive Boards used to have wiring jumpers on a non-
component side of the PC Board layout. New Drive Boards do not
have these anymore. Drive Board replacement is recommended
during maintenance work to all units S/N 20000-20060.
– Check all motor run indication leds (H7, H8, H9, H10) on Drive
Board. They should light up when the grid is driven. If only leds H7
and H8 or H9 and H10 light up, or no leds at all, then the Drive Board
does not operate normally and should be replaced.

NOTE
Remember that this may happen randomly.

– Grid lines may also be related to a Bucky, a grid mechanism or a grid


connection pin. The mechanism may incur friction, have some
misaligned parts, motor axis fixing screws or home position flag
might prevent free movement of the grid. The grid connection pin
from the grid mechanism to the Bucky may have an incorrect or loose
contact. Current units have a bit longer pin.

86502-4AA Instrumentarium Imaging 8-27


8 Error tracing

8.7 ECS trouble shooting

Problem Items to check


The ECS does not move. 1 The signal from the switches
reaches the X2 on Stereo Board
2 Motor connection to X15 on the
Stereo Board
3 Movement limits are OK (X2 on
Stereo Board)
4 The motor works (use external
voltage supply)
The Auto Release function does not The Auto Release is switched on.
work.
The ECS does not stop when the The connector X2 is not properly
compression reaches the current connected.
limit.
The ECS stops before the The compression motor exhausts
compression reaches the current before the limit. Increase the speed in
limit. service program PrG01.
Return function does not work. The signals from the switches do not
reach connector X2 on Stereo Board
The ECS moves too slow. Adjust the speed at trimmer R42 on
Stereo Board
The ECS moves too fast. Adjust the speed at trimmer R42 on
Stereo Board
The ECS does not move from ECS 1 Check pulses from compression
pedal. motor encoder X17/Pin 5 on
Stereo Board
2 Check that ECS motor has been
connected to X15 and
compression motor to X17
Noisy ECS movement. Check bearings.

If the motor or whole ECS mechanism has to be replaced, instructions are


located in chapter 7.3.15, Replacement of ECS mechanism/ECS motor.

8.8 Serial bus communications check


Check that the following boards have semi-dim LEDs. Non-existent or totally
bright LED indicates a communication problem.

Name of the Board RXD TXD Notes


Supervisor Board H6 H5
Stereo Board H2 H1 From the point of view of
the Board.
AEC Board H3 H2 As above.
Compression Board H2 H3 As above.
Collimator Board H13 H14 As above.
Control Panel Board H4 H6 As above.

8 28 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.9 Notes on software upgrading


From software version 1.00.9 onward:

Before upgrade:
1 Retrieve BRAM dump so that you capture it to file!
– Check Service Menu->Stereo Move->Autodrive setting and check
setting state (ON/OFF)

After upgrade:
1 Check Service Menu->Stereo Move->Autodrive setting and check set-
ting state (ON/OFF). The software upgrade may have changed this set-
ting.
– Stereo Move -> Speeds are in use from this SW version. Load
defaults to these as follows:
2 Enter service menu->Stereo Move->Speeds menu. Press simultaneously
Cancel- and Set-key for a couple of seconds. You should see message.
Changing parameters to defaults... Answer Yes to this question. After
that you have factory default for speeds (does not affect to other
settings).

NOTE
When you are loading defaults, make sure that you are in Speeds-menu.

8.10 Questions & Answers


– Q: How can I check the number of exposures taken with Diamond or
the number of the last exposure?
– A: Obtain exposure data as instructed in chapter 4.3.2, Data capture.
The number of exposure is at the beginning and at the end of the data.
When the number of exposures reaches 65535, it starts again from
zero.

86502-4AA Instrumentarium Imaging 8-29


8 Error tracing

8.11 FAQ - Frequently asked questions


8.11.1 Common questions

Question Answer
How I can change the X-ray The Diamond X-ray field lamp is easy to
field lamp? replace. The lamp holder is secured with
magnet to a holder that can easily be rotated
for convenient access through the collimator
opening. Refer to Diamond Operation
Instructions (86500) for details.
After the exposure, ID 2 does Check that correct cable (30749) has been
not receive exposure data? connected to Diamond unit. This cable is
delivered with Diamond unit in the wooden
crate. The cable delivered in ID 2 package is
intended for use with Performa and Alpha
units.
What are ranges 15 to 20 kV The range 15 to 20 kV is used for specimen
and 36 to 40 kV used for? examination (not with patients), the Control
Panel LCD-screen blinks when these kV
values are used.
The range 36 to 40 kV is used in special
cases, e.g. with a very dense breast when
other kV values are not enough to penetrate
the breast.
How can I use 15 to 20 and 36 Switch to Manual Mode to select these kV
to 40 kV range? values. 15 to 20 kV are used for specimen
imaging, the Control Panel LCD screen
blinks when these kV values are used.
Aluminum filter is always used with 35 to
40 kV range.
How does Diamond measure The dose is not measured, but calculated
the dose? with the help of following exposure data:
- recorded exposure kV and mAs
- estimated radiation output in mR
(efficiency @28kV (Mo/Rh) or 39kV (Al)
stored in memory)
- recorded compression thickness and filter
- computed glandular dose estimate based
on the tables (mR(air) vs mSv(gland))
published in the references for 50%-50%
glandular-adipose breast
- separate calculations for each adjustment
step during exposure in Auto kV-mode.

8 30 Instrumentarium Imaging 86502-4AA


8 Error tracing

How is the tube protected from When the tube approach overheat limit, the
overheating? software increases wait periods between
exposures to allow for cooling of the tube.
This calculation is reset, if unit power is
switched off and on again.
A thermal switch inside the unit releases if
the temperature inside the tube is too high
(about 65 C). Should this happen, the tube
must cool down for 1-2 hours before
exposures are possible again.
How can I test that film Test condition of the film and the processor
processor works correct? with a sensitometer/densitometer:
- fog, speed index, contrast index
- min and max OD values within 21 steps
Verify also that the film processor
performance is uniform day to day.
When I need to adjust kV up or kV should be higher with a thicker breast.
down? kV relates to contrast, dose and time:
kV up > contrast low
> dose low
> time low (mAs low)
kV low > contrast up
> dose up
How can I get information on The best way is to check our Member
the last modification? Service pages regularly. This site also
contains our latest training schedules.
If you do not have password, contact us.
Where can I find information Many countries have their own regulations.
on country-specific regulation ACR Quality Control Manual and European
for mammography? guidelines for quality assurance in
mammography screening are accepted
widely. However, contact regulatory
authorities of the country where the unit will
be used to obtain exact and detailed
information.
How do Diamond and Alpha Diamond tube has better focusing of x-ray
RT tubes differ from each beam with small focus. Also, Gaussian
other? intensity distribution within focal spot is
better.
Performa and Diamond use the same tube,
which is not compatible with RT tube.
What is that continuous beep Check that Control Panel is properly
sound after the Diamond unit connected to the back of Diamond column.
has been switched on? Check also that the cable connects properly
to the Control Panel
Is there a problem when large This phenomenon is caused by heel effect.
Bucky is used and the rear The tube produces more radiation in the
section of the picture is much rear.
whiter than the front section?
Is there a van kit installation Introduction of Diamond Van kit is
available for Diamond? scheduled for 2003.

86502-4AA Instrumentarium Imaging 8-31


8 Error tracing

8.11.2 Customer questions

Question Answer
The unit cannot be switched Check that the breaker switch behind the
on? column has not been accidentally turned off.
Turn it on, then switch on with the main
power switch.
Check also power line fuses.
The power is on, but none of Check that the Emergency Stop Button (2
the functions can be initiated red buttons on both sides of the carriage top)
(error message 11)? has not been accidentally pressed down.
The C-arm Auto-Rotation does Check that the paddle has been attached
not rotate? correctly.
The exposure stops Check error messages
immediately after the button is -Error 65
pressed? - Ask the customer to test with higher
kVvalues
- Check operation of the light mirror
- Check for any other object in x-ray
field
- Check filter and cassette mechanism
The cassette has been loaded, The cassette is detected with optical
but the unit does not let you sensors. Find out the C-arm positions at
expose (error 35)? which you are not allowed to initiate the
exposure. Eliminate all excessive stray
lights entering the cassette holder area
(cassette sensor). In order to eliminate the
light from outside to the sensor, darken the
room and test the cassette sensor.
How to remove visible grid Test with another Bucky (large or small) to
lines in the image? identify the problem. Check grid
mechanism/electrical system.
The grid motor is in the unit and not in the
Bucky.
What if the images sometimes Compare selected detector area on the
appear too dark with breast image:
AutoPoint? If Diamond has selected detector position
C1 and you can see that in the image the
pectoral muscle (near the chest wall edge) is
located just over C1 detector. Try to
increase AutoPoint limit (default 700).
With high values, the unit less likely selects
detector C1 under pectoral muscle (which
would result as a too dark image). Refer to
AutoPoint instructions in Installation
Manual.
What if the images sometimes Check what detector the unit has selected
appear too white or dark with and examine breast anatomy in this film
AutoPoint? area (dense/not dense). AEC detector
location sheet in the end of the Installation
Instructions (86501) helps to estimate the
detector location on the film.

8 32 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.11.3 Installation/service questions

Question Answer
The tube park movement stops Check that all tube park sensors have
sometimes or an error message remained intact (INPUT PROGRAM) and
is displayed? in perfect working condition through the
- (error messages 12-13) shipment/installation.
Check that cables under the upper cover
remain free, not tangled up with some other
objects.
Check that white plastic finger protection
sheets within lower cover move unhindered.
Check that safety clutch has been adjusted
correctly.
The cassette has been loaded, The cassette is detected with optical
but the unit does not let you sensors. Find out the C-arm positions at
expose (error 35)? which you are not allowed to initiate the
exposure. Eliminate all excessive stray
lights entering the cassette holder area
(cassette sensor). In order to eliminate the
light from outside to the sensor, darken the
room and test the cassette sensor.
How to remove visible grid Test with another Bucky (large or small) to
lines in the image? identify the problem. Check grid
mechanism/electrical system.
The grid motor is in the unit and not in the
Bucky.
What if the images sometimes Compare selected detector area on the
appear too dark with breast image:
AutoPoint? If Diamond has selected detector position
C1 and you can see that in the image the
pectoral muscle (near the chest wall edge) is
located just over C1 detector. Try to
increase AutoPoint limit (default 700).
With high values, the unit less likely selects
detector C1 under pectoral muscle (which
would result as a too dark image). Refer to
AutoPoint instructions in Installation
Manual.
What if the images sometimes Check what detector the unit has selected
appear too white or dark with and examine breast anatomy in this film
AutoPoint? area (dense/not dense). AEC detector
location sheet in the end of the Installation
Instructions (86501) helps to estimate the
detector location on the film.
The cassette has been loaded, The cassette is detected with optical
but the unit does not let you sensors. Find out the C-arm positions at
expose (error 35)? which you are not allowed to initiate the
exposure. Eliminate all excessive stray
lights entering the cassette holder area
(cassette sensor). In order to eliminate the
light from outside to the sensor, darken the
room and test the cassette sensor.

86502-4AA Instrumentarium Imaging 8-33


8 Error tracing

How to remove visible grid Test with another Bucky (large or small) to
lines in the image? identify the problem. Check grid
mechanism/electrical system.
The grid motor is in the unit and not in the
Bucky.
Z-movement of the carriage is Check for source of bad movement:
slow or does not drive at all? Counterweight tiles in counterweight cage
are installed correctly.
The amount of counterweight tiles are in
balance with the weight of the C-Arm and
the carriage. The movement should be
smooth in both directions.
Check that 4 white plastic rollers aside the
carriage do not cause friction. Adjust, if
necessary.
Autorelease function does not Enter service program PrG02 and check that
operate? software together with Hall sensors on the
Compression Display Board detects
different paddles (sensor magnets) correct.
Paddle selection is marked with an
ASTERISK (*) on the display.
While you are in the service program, try to
shake/jiggle the paddle to check whether the
asterisk moves on to another paddle. If this
happens, you need to fine tune the position
of the Hall sensors.
C-arm rotation stops at wrong 1) Check that C-arm rotation sensors/flags
position? are fully operational.
2) Check the correct operation of the both
home opto sensors with INPUT
PROGRAM. If problems arise, check
wiring solder joints and optoswitches.
3) Check that opto flag is in correct
position.
The ID 2 does not receive 1) Check that cable is correct (code 30749),
exposure data from the shipped with the Diamond unit, not the one
Diamond unit after exposure? included in the original ID 2 box.
2) Check that the cable is connected to
Diamond COM 2 port and ID 2 port B.
When I need to calibrate for 1) If the tube has been replaced.
tube preheat? 2) You have lost original preheat settings
(set by factory) from supervisor parameter
memory.
How I can save unit Save parameters after installation with
parameters? Diamond - PC cable and HyperTerminal
program.

8 34 Instrumentarium Imaging 86502-4AA


8 Error tracing

8.11.4 Installation/service questions/error messages

Question Answer
Error message 15 appears (rear Check that rear collimator sensor is OK.
collimator time out)? Test for correct operation of the sensor with
INPUT PROGRAM and verify that the flag
reaches the optosensor.
Check that no mechanical friction/object
prevents the movement.
Check Collimator Board and motor.
Error message 72 appears (side Check that side collimator sensor is OK.
collimator time out)? Test correct operation of the sensor with
INPUT PROGRAM and verify that the flag
reaches the optosensor.
Check that no mechanical friction/object
prevents the movement.
Check Collimator Board and motor.
Error message 24 appears Calibrate Up Force Limit in service
(compression force program PrG06. Check mechanism.
measurement failure, Inhibit
compression down)
How much time does the 1-2 work days for 2 persons. AEC film
Diamond installation require? calibration (50 to 150 exposures) will take
most of the time.
Depends also on the exposure combination
the customer wants to use: A) Bucky/
Magnification B) filters Molybdenum/
Rhodium C) Auto-Time/Auto-kV
Remember that the Diamond can be
installed in a separate room and after
completion of the tests, it can be moved
with transportation wheels to the
examination room.
Can I calibrate AEC separately No, the unit will use one AEC setting for all
for the small Bucky and Buckies.
Bucky?
If the customer wants to use 2 No.
different screen combination,
can I calibrate AEC separately
for these?
Why does the Diamond unit Some doctors want to use different pre-set
have Pre-Prog settings 1-4 and to obtain optimal image quality suitable to
what values can be set within their needs.
Pre-Prog 1-4 menus? Pre-programmed values:
A) Density correction B) Target Time C)
Auto kV initial correction D) max. Auto kV
change limit E) Menu Text F) Auto Point
Coefficient factor.

86502-4AA Instrumentarium Imaging 8-35


8 Error tracing

8 36 Instrumentarium Imaging 86502-4AA


9 Specifications

9 Specifications

System specifications for Diamond, model MGX 2000

Manufacturer Instrumentarium Imaging


P.O. Box 20
FIN-04301 Tuusula, Finland
Manufacturer's quality system ISO9001
according to
Manufacturer's environmental ISO14001
management system according to
General Medical electrical equipment
Model MGX 2000
Power supply Mains plug connection, 1-phase
Apparent resistance of supply 0.75Ω
mains
Rated voltage 230 VAC ±10%, single phase, 50/
60 Hz
16 A slow blow fuse in power
outlet
Rated current 20 A Momentary (during exposure)

300mA long term (otherwise)


Rated input power 4.6 kVA
IEC safety class I
IEC applied part type B
Enclosure IP 20
Weight 350 kg (771 lbs.)
Height 1.9 m (6.23 ft)
Operation mode Continuous operation with
intermittent loading
Radiation intensity Typical range 1000-1200mR/
100mA at 28kV

86502-4AA Instrumentarium Imaging 9-1


9 Specifications

Environmental specifications

Operating conditions +10°C...+40°C


Non condensing (maximum
humidity 90%)
Storage and transportation (+0°C...+35°C)
Humidity 10%...100%
Pressure 500mbar - 1060 mbar

Generator

Manufacturer Instrumentarium Imaging


Type MGHV-300
Frequency 40 - 100 kHz
Power Max. 2.6 kW
kV range 15 - 39 kV, in 1 kV increments
Accuracy ± 1.5 kV
kV ripple < 5%
mA range Focus 0.3mm 50 - 100 mA
Focus 0.3mm 25 - 50 mA (low mA
mode)
Focus 0.1mm 18 - 30 mA
(- tube current reduced with small
breast thickness)
Maximum mA output Focus 0.3mm 150mA
Focus 0.1mm 45mA
mAs range Focus 0.3mm 2 - 500 mAs:
Tolerance:
mAs ≤ 40: ± 4 mAs
mAs > 40: ± 10%
Focus 0.1mm 4 - 300 mAs
Tolerance:
mAs ≤ 40: ± 4 mAs
mAs > 40: ±10%
Time range Focus 0.3mm 0.02 - 5 s
Focus 0.1mm 0.04 - 10 s
Tolerance: ± 0.10 + instrument
inaccuracy

92 Instrumentarium Imaging 86502-4AA


9 Specifications

Tube loading chart for Diamond X-ray tube (M113SP)

kV M113SP M113SP kV M113SP M113SP


mA(SF) mA(LF) mA(SF) mA(LF)

15 18 53 28 25 85
16 20 59 29 24 85
17 22 64 30 23 80
18 23 69 31 22 80
19 24 74 32 21 80
20 25 75 33 21 75
21 26 80 34 20 75
22 28 85 35 20 70
23 30 90 36 19 50
24 29 95 37 19 50
25 28 100 38 18 50
26 27 95 39 18 50
27 26 90

SF = Small focus, LF = large focus

Time range with loading factors shown above:

– Focus 0.3 0.02 - 5.0 sec.


– Focus 0.1 0.04 - 10.0 sec.

Minimum mAs = Minimum time * lowest mA in the tube loading chart.


Maximum mAs = Maximum time * lowest mA in the tube loading chart.

Automatic Exposure Control (AEC)

Solid state 15 step density selector.

Operating range, kV limits

Bucky Manual mode: 20-39 kV


AutoTime and Auto kV: 23-35 kV
Specimen: 15-39 kV
Magnification Manual mode: 20-39 kV
AutoTime and Auto kV: 23-35 kV
Specimen: 15-39 kV
Delta 32 Manual mode: 20-39 kV
Specimen: 15-39 kV

86502-4AA Instrumentarium Imaging 9-3


9 Specifications

Automatic Exposure Control (AEC), Operating range, mAs limits

kV Bucky Delta32 Magn kV Bucky Delta32 Magn


mAs mAs mAs mAs mAs mAs

15 250 250 175 28 400 400 250


16 250 250 200 29 400 400 200
17 300 300 200 30 400 400 200
18 300 300 200 31 400 400 200
19 350 350 200 32 400 400 200
20 350 350 250 33 350 350 200
21 400 400 250 34 350 350 200
22 400 400 250 35 300 300 200
23 450 450 300 36 250 250 175
24 450 450 250 37 250 250 175
25 500 500 250 38 250 250 175
26 450 450 250 39 250 250 175
27 450 450 250

X-RAY TUBE Diamond (Varian M113SP in B112


housing)
Anode type Rotating dual-angle anode
Anode heat capacity 300,000 HU
Target material Molybdenum, niobium doped
Anode rotation speed 3000rpm(50Hz)/3600rpm(60Hz)
Focal spot sizes M113SP 0.1, 0.3 (IEC60336)
Port material beryllium
Reference axis 42 mm (M113SP) from chest wall
edge (tube angle 4°)

1 Focal spot is marked in the front part of the tube head assembly.
2 The collimator label is at the front end of the tube head assembly.

Diamond Filtration

Molybdenum 0.025 mm
Rhodium 0.025 mm
Aluminum 0.50 mm

94 Instrumentarium Imaging 86502-4AA


9 Specifications

Positioning

Source to Image Distance (SID) 660 mm (26”)

Movement of the motorized, dual-speed carriage

(66cm) Rotation ±185°


SID

(26”) Vertical 760 mm (30”)


The cassette holder vertical movement range is 68 cm - 144 cm above
finished floor level (27” - 57”) with the C-Arm in the upright position (0°)
and 79 cm - 156 cm (31” - 61”) with C-Arm rotated in to the lateral (90°)
position.

Compression

motorized Max 200N


manual Max 270N±20N
Foot pedal protection classification IPX1

Recording systems

Bucky for 18 x 24 and 24 x 30cm film


Magnification for 18 x 24cm film 1.6, 1.8 & 2.0
exposed size 10 x 24
Stereotactic device Delta 32 digital ccd-camera

Correspondence between X-ray field and image reception area

Chest wall 0...+3 mm


Other walls +1...+5 mm
Correspondence between ≤ 5 mm any wall
collimator light field and X-ray
field

Collimator

Type MGB 600


Filtration 0 mm Al
Electrical input 34 VAC (for half-step motors)
9-14 VDC (for 12 VDC lamp)
Lamp type HLX (halogen), 12V, 50W

86502-4AA Instrumentarium Imaging 9-5


9 Specifications

Radiation field

Maximum radiation field 24x30cm

Attenuation equivalent

Bucky, cassette tunnel or ≤ 0.25 mm Al (Total of all layers)


magnification tunnel

Anti-scatter grid

Line pairs 36 lp/mm


Grid ratio 5:1

ROC Grid

Line pairs 36 lp/mm


Grid ratio equal to 6:1

96 Instrumentarium Imaging 86502-4AA


9 Specifications

Measurement criteria for exposure time

Exposure time Consists of beginning and ending


points as measured by a calibrated
x-ray monitor at 70% of the peak
radiation waveform
kV The high voltage mean value
received after measurement with a
calibrated x-ray monitor.
mAs The resistance value divided by
voltage. A calibrated x-ray monitor
measures the voltage over feedback
resistor and provides a mAs value.
The nominal x-ray tube voltage 39kV is obtained at highest tube current
50mA.
The highest tube current 100mA is obtained at the highest tube voltage
25kV.
The highest electric power is obtained both at 32kV tube voltage and
80mA tube current and at 25kV tube voltage and 100mA tube current.
The nominal electric power of 2.4 kW is obtained, when loading time is
0.1s and x-ray tube voltage 30kV is used.

Display accuracy

Compression force ± 20N


Compression thickness ± 1cm
C-Arm angle ± 2°
mGy display ± 30%
For mGy display accuracy, refer to:

“Spectral Dependence of Glandular Tissue dose in Screen-Film


Mammography” Wu Xizeng, Wu X, Barnes Gary T, Tucker Douglas M,
Radiology 1991; 179:143-148

“Normalized Average Glandular dose in Molybdenum Target- Rhodium


Filter and Rhodium Target -Rhodium filter Mammography” Wu Xizeng,
Wu X, Gingold Eric L, Barnes Gary T, Tucker Douglas M, Radiology
1994; 193:83-89

86502-4AA Instrumentarium Imaging 9-7


9 Specifications

Fuses

Power Board 85020, F1,2 Fuse 5x20 mm 4AT, IEC 127-2/3


F3 Fuse 6.3x32 mm, 15ASB 250V
F4 Fuse 5x20 mm 0.8AT, IEC127-2/3
F5 Fuse 13/32” 20AF 600VAC
(Bussmann KTK-20)
AMD 32305, F1 Fuse 6.3x32 mm 10AT 250V

Diamond fuses used by Delta 32

On PCB located in the bottom-part plate of the C-Arm.


Fuse Board 33615, F5 Fuse 1 pc 6x32 mm 1,25AT 250V
UL
F3 Fuse 1 pc 6x32 mm 0,8AT 250V
UL
F1,2,4 Fuse 3 pcs 6x32 mm 0,5AT 250V
UL
In the carriage next to the transformers: D32 Fuse Plate 82685
F1,2 Fuse 2 pcs, 6X32 mm 0,3AD 250
UL 248-14

Color letter equivalencies:


Coding BK BN RD OG YE GN BU VT
system 1
Coding MU RU PU OR KE VI SI VT
system 2
Coding GY WH PK GD TQ SR GNYE
system 1
Coding HA VA VP KU TU HO KEVI
system 2

98 Instrumentarium Imaging 86502-4AA


10 Schematics/Layouts

10 Schematics/Layouts
The name Alpha X in the schematics/lay-outs is the project code name of
Diamond. The exact board locations are described in chapter 7.2, Electrical
board and mains transformer locations.

Document Name Document code


AMD Board 32305-3S_, 32305-4K_
AEC Board 85001-3S_, 85001-3K_
Generator Board 85003-3S_, 85003-3K_
Collimator Board 85009-3S_, 85009-3K_
Drive Board 85011-3S_, 85011-4K_
Stereo Board 85013-3S_, 85013-3K_
Power Board 85020-3S_, 85020-3K_
Compression Board 85022-3S_, 85022-3K_
Compression Display Board 85024-4S_, 85024-4K_
Caslock Sensor Board 85026-4S_, 85026-4K_
Tacho Board 85028-4S_, 85028-4K_
Sensor2 Board 85030-4S_, 85030-4K_
Collimator Sense Board 85032-4S_, 85032-4K_
Csense Board 85065-4S_, 85065-4K_
Control Panel 85069-3S_, 85069-3K_
Supervisor Board 85078-3S_, 85078-3K_
Casopto Board 85090-4S_, 85090-4K_
Flatcable Adapter Board 85094-4S_, 85094-4K_
Flat cable for membrane switches of 86000-4S_
Control panel
AEC Membrane Switch 86001-3S_
Vertical Membrane Switch 86002-4S_
Tube Park Switch 86003-4S_
Projection Switches 86004-4S_
Stereo Switch 86005-4S_
Tube Cover Switches 86006-4S_

Color letter equivalencies:


Coding BK BN RD OG YE GN BU VT
system 1
Coding MU RU PU OR KE VI SI VT
system 2
Coding GY WH PK GD TQ SR GNYE
system 1
Coding HA VA VP KU TU HO KEVI
system 2

86502-4AA Instrumentarium Imaging 10-1


10 Schematics/Layouts

10 2 Instrumentarium Imaging 86502-4AA


85011-3SB.sch-1 - Fri Jan 12 13:08:58 2001
85011-3SB.sch-2 - Fri Jan 12 13:10:16 2001
If needed, these srews
can also be assembled from
left to right.

85021-3FB
M7 R154
M7
M13

M13 1

R153

M12
M12 Aine Pinta Pintak?s.

Liittyy Nimitys

Yleistoleranssi Suhde POWER BOARD


Merkki Muut.no Muutos Piirt. Tark. Hyv. SFS 4011-

Laite

C 80438 Lisätty R156 6.6.02/NJ 10.6.02 HK 10.6.02 TJ Piirt.


ALPHA X
No.
B 80013 KORJATTU NUMEROINTEJA 100800NJ/SF 300800LTU 040900JJS 200400 PeJ
D 80453 Poistettu R156 050702NJ 30.8.02HK 2.9.02MJ
Tark.
20.06.00/NJ
Hyv.
21.06.00/HK
MGX-85020-3KD
85026-4s.sch-1 - Tue Jun 13 09:15:32 2000
85028-4SA.sch-1 - Tue Jun 11 08:30:43 2002
85030-4sA.sch-1 - Tue Jun 11 08:32:18 2002
85032-4SA.sch-1 - Thu Apr 12 12:29:27 2001
85065-4S.sch-1 - Tue Jun 13 09:16:48 2000
85069-3SA.sch-1 - Tue Oct 17 13:56:30 2000
85069-3SA.sch-2 - Tue Oct 17 13:56:49 2000
85069-3SA.sch-3 - Tue Oct 17 13:57:25 2000
85069-3SA.sch-4 - Tue Oct 17 13:57:09 2000
86000-4S.sch-1 - Mon Sep 25 09:15:29 2000
11 Cable drawings

11 Cable drawings
The name Alpha X in the cable drawings is the project code name of
Diamond. The location of the cables can be checked in the Appendices
section.

Document Name Document code


Cable W 530 - ID II cable for order code 30794,
Diamond schematic not available here
Main cable CB10 85016-3J_
Cable SB1 85017-4J_
Cable SB31 85018-4J_
Cable PW20 85019-4J_
Flat cable CS1 85035-4J_
Flat cable CS2 85036-4J_
Cable CS3 85037-3J_
Cable CS4 85038-4J_
Cable CS20 85039-4J_
Cable CS31 85040-4J_
Cable CS32 85041-4J_
Cable CS33 85042-4J_
Cable CS36 85043-3J_
Cable CS35 85044-4J_
Cable CS21 85045-4J_
Cable CS22 85046-3J_
Cable CS23 85047-4J_
Cable CS24 85048-4J_
Cable CS25 85049-3J_
Cable CS26 85050-3J_
Flat cable CS27 85051-4J_
Cable CS28 85052-4J_
Cable CS29 85053-4J_
Cable CS40 85054-3J_
Cable CS11 85055-4J_
Cable PW1 85056-4J_
Cable PW2 85057-4J_
Cable PW3 85058-4J_
Cable PW21 85059-4J_
Cable PW30 85060-4J_
Cable PW31 85061-4J_
Main Cable CB30 85062-3J_
Cable PW32 85063-4J_
Cable SB40 85064-4J_
Cable CS34 85076-3J_
Cable CS0 85077-4J_
Flat cable CS8 85079-4J_
Cable CS6 85080-4J_
Cable SB2 85081-4J_

86502-4AA Instrumentarium Imaging 11-1


11 Cable drawings

Document Name Document code


Cable CS5 85082-4J_
Electrical outlet cable 85083-4J_
Intermediate electrical outlet cable 85084-4J_
Cable CS7 85085-4J_
Cable CB1 85086-4J_
Cable CS9 85087-4J_
Cable CS10 85088-4J_
Foot pedal cable 85089-4J_
Flatcable Adapter Board 85094-4K_
Cable PW33 85096-4J_
Cable SB1b 85098-4J_
Cable SB1c 85100-4J_
Warning light cable 85101-4J_
Program download cable 85102-4J_
JTAG programming cable order code 85105,
schematic not available here
Cable drawing for network part 85106-4J_

11 2 Instrumentarium Imaging 86502-4AA


2
1

120 3 Z-UPPER-LIMIT

3
1

CEMBRE RF-F 305P tai


4 KPL ABIKO B0,5-6ET

POWER X8
MOLEX 90156-0152 2

10 MU AWG24 170 1
3
7 VA AWG24 170 Z-LOWER-LIMIT
9 MU AWG24 800 750
8 KE AWG24 800
1
1 3

2 KE AWG24 1210
11 RU AWG24 1210 310
310
12 4 VA AWG24 860 70 SI AWG24 360
1 VT AWG24 500 350 SI AWG24 360 OR AWG24 360
3 VI AWG24 400 SI AWG24 120 OR AWG24 360
PWR X8 Johtimet 300V UL OR AWG24 120
6 SI AWG24 400
5 OR AWG24 400

MOLEX 90119-2110 1 4 2 3 5 1 4 2 3 5 1 4 2 3 5 1 2 5 6

C-LEFT-POS C-RIGHT-POS C-PULSE C-R/LMIDDLE-POS

SENSOR BOARD 85030


TACHO BOARD 85028
Johtimet 300V UL

Johdinsuojat UL Aine Pinta Pintak„s.

Liittyy Nimitys

Yleistoleranssi Suhde KAAPELI CS3


Merkki Muut.no Muutos Piirt. Tark. Hyv. SFS 4011-

Laite

B 80471 Liitintyypin korjaus 17.9.02HMK 20.9.02HK Piirt.


MGX
No.
D 80557 Tarkennettu kuvan tietoja 050503LK 281299 PeJ
8.5.03HK -------
C 80483 Korjattu tietoja
Tark.
181102TMV 25.11.02HK 25.11.02HKe 060600 NJ
Hyv.
060600 JJS
MGX 85037-3JD
Aine

Piirt.

Tark.
C
Merkki

Liittyy

SFS 4011-
Yleistoleranssi

070600 NJ
050400 PeJ
Muut.no

80557

Hyv.
Suhde
MOLEX 90160-0110
Muutos

-:-

070600 JJS
Nippusiteet korttien viereen (2kpl)

Pinta

Nimitys
1 2 HA, 855 AWG24 HA
4
VA, 750 AWG24 VA
3
SI, 750 AWG24 700 SI SI 115 60
9
Tarkennettu kuvan tietoja

OR, 750 AWG24 OR OR 115


9 10 1

ST X2
MOLEX 90119-0110

Johdot tulevat
5 3 2 4 1 1 4 2

KAAPELI CS20
juotospuolelta
ja juotetaan
komponenttipuolelta

No.
Piirt.

Laite
Pintak„s.
050503LK

MGX
85028 TACHO BOARD
Tark.

85028 TACHO BOARD


8.5.03HK

EPS YLÄRAJA EPS ALARAJA


Johtimet 300V UL
Hyv.

Johdinsuojat UL

MGX 85039-4JC
-------
MOLEX 90160-0110

Johdot sekä asennetaan että juotetaan


korttien juotospuolelle.

MOLEX 90119-0110
Johtimet sidotaan yhteen nippusiteillä (3 kpl)
10 mm:n etäisyydellä levyjen reunoista.

5 VI AWG24, 470 VI VI VI VI VI
1 2 HA AWG24, 330 HA HA HA SI, 120 SI
4 50
RU AWG24, 330 RU RU RU OR, 120 OR
3 50
KE AWG24, 210 KE SI, 120 SI
2
SI AWG24, 210 160 SI OR, 120 OR
10
9 10 1 OR AWG24, 210 OR
7 VA AWG24, 760 VA
ST X3 6 PU AWG24, 700 PU
SI
6 1 4 2 3 5 1 2 5 6 3 4 1 4 2

TACHO BOARD SENSOR BOARD TACHO BOARD

85028 85030 85028

STEREO-LEFT15-POS STEREO-MIDDLE-POS STEREO-RIGHT15-POS

1
3

STEREO-RLIMIT
3

1
2

SI AWG24, 70
VA 1
SI AWG24, 500
1
STEREO-LLIMIT
3
440

PU
3 2

JOHTIMET 300V UL CEMBRE RF-F 305P tai

JOHDINSUOJAT UL 4 KPL ABIKO B0,5-6ET


Aine Pinta Pintak„s.

Liittyy Nimitys

Yleistoleranssi Suhde KAAPELI CS22


Merkki Muut.no Muutos Piirt. Tark. Hyv. SFS 4011-

Laite

B 80483 Tarkennetaan kuvaan tietoja 211102TMV 25.11.02HK 25.11.02HKePiirt. MGX


No.
A 80001 Lisätty stereon rajamikrojen johdot 310500PeJ 060600NJ 060600JJS 260500 PeJ
C 80557 Päivitetty oikea liitinkoodi 050503LK 8.5.03HK -------
Tark.
060600 NJ
Hyv.
060600 JJS
MGX 85046-3JC
Aine

Piirt.
C

Tark.
Merkki

Liittyy

SFS 4011-
Yleistoleranssi

070600 NJ
MOLEX 5559-2P MOLEX 5558-PBT

120100 PeJ
Muut.no

(ei kuulu toimitukseen) (ei kuulu toimitukseen)


80557

Hyv.
Suhde
CX24
Muutos

PU
1

-:-
MU M

070600 JJS
1 2 2

Pinta

Nimitys
PURISTUSMOOTTORI
Ostokoodi 22502
Tarkennettu kuvan tietoja

CX24 MOLEX 5557-02R


KIERRETÄÄN PARIKSI. 12 KIERROSTA.
ST X17

PU, 650 AWG18 PU


1 1

KAAPELI CS23
4 1 MU, 650 AWG18 MU 1 2
2 2
5 2 OR, 650 AWG24 490 OR
3
KE, 650 AWG24 KE
6 3 5
6 SI, 650 AWG24 SI MOLEX 5556-PBT

No.
Piirt.

Laite
MOLEX 5556-PBT

Pintak„s.
4 2 1
050503LK

MGX
MOLEX 5557-6R
Johdot tulevat
juotospuolelta
Tark.

TACHO BOARD 85028


ja juotetaan
8.5.03HK

komponenttipuolelta
Hyv.

JOHTIMET 300V UL

MGX 85047-4JC
-------

JOHDINSUOJAT UL
PURISTUSMOOTTORIN TACHO
Aine
STEREON MOOTTORI (SAMA KUIN C-MOOTTORI)

Piirt.
B

Tark.
Merkki

Liittyy
Ostokoodi 22641
Koneistettuna 82016

SFS 4011-
Yleistoleranssi
CX21

070600 NJ
120100 PeJ
Muut.no

RU
80557

1
M

Hyv.
VI

Suhde
2
2 1
Muutos

-:-
MOLEX 1190-T

070600 JJS
MOLEX 3191-2P
(ei kuulu tilaukseen) (ei kuulu tilaukseen)
KIERRETÄÄN PARIKSI 15KIERROSTA.

Pinta

Nimitys
STM CX21
MOLEX 3191-2R

ST X16
MOLEX 1189-T
Tarkennetaan kuvan tietoja

RU, 700 AWG16 RU


1 1 2 1
4 1 VI, 700 AWG16 VI
2 2
5 2 OR, 700 AWG24 560 OR
3
KE, 700 AWG24 KE STEREON MOOTTORI
6 3 5
6 SI, 700 AWG24 SI

KAAPELI CS28
MOLEX 5556-PBT 4 2 1

MOLEX 5557-6R

No.
Piirt.

Laite
Johdot tulevat

Pintak„s.
TACHO BOARD 85028
050503LK

juotospuolelta

MGX
ja juotetaan
Tark.

komponenttipuolelta
8.5.03HK

JOHTIMET 300V UL STEREON MOOTTORIN TACHO


JOHDINSUOJAT UL
Hyv.

MGX 85052-4JB
-------
Aine

Tark.
Piirt.
Liittyy
A
Merkki

ABIKO B1003FLS8 + KUTISTESUKKA PÄÄLLE

Yleistoleranssi
SFS 4011-
AWG16

070600 NJ
Muut.no

SI, 50

151199 PeJ
RU, 50
80557

Hyv.
MOLEX 5556-PBT

Suhde
Muutos

MOLEX 5557-2R

070600 JJS
1 2

1 2

Pinta
Liittimet CX9 kytketään yhteen

Nimitys
CX9
Päivitetty oikea liitinkoodi

CX9
1 2 MOLEX 5559-2P

1 2
MOLEX 3191-2R
MOLEX 5558-PBT

KAAPELI PW31
MOLEX 5557-2R

ST X6

No.
Piirt.

Laite
RU, 50
SI, 50

Pintak„s.
2 1 1 RU, 1070 AWG16
1
050503LK

1020 AWG16 1 2

MGX
SI, 1070
2 2
Tark.

AMD X3
MOLEX 1189-T välikuorinta+juotos
MOLEX 5556-PBT
8.5.03HK

suojataan kutiste-
sukalla
Hyv.

JOHTIMET 300V UL

MGX 85061-4JA
-------

JOHDINSUOJAT UL
MOLEX 90119-0110

6 VI AWG24 VI 1220
5 RU AWG24 RU 790 OR AWG24 350 OR 450
4 HA AWG24 350 HA 640 OR AWG24 100 OR 200 SI AWG24 SI 450
3 VA AWG24
220 VA 310 OR AWG24 OR 450 SI 200
SI AWG24
1 OR AWG24 OR 310 SI AWG24 SI 450
2 SI AWG24 SI 310

MOLEX 90160-0110
41 253 4 1 253 41 253 4 1 2

Johdot tulevat
juotospuolelta
ja juotetaan
komponenttipuolelta

COL X4
BACK FRONT CF CB

TACHO BOARD
85028 MIKROKYTKIN OMRON SS-01T
1 (Ei kuulu toimitukseen)

7 MU AWG24 MU 470 1
8 RU AWG24 400
RU 470

2 3

CEMBRE RF-F 305P tai


MOMENT
2 KPL ABIKO A1503FLS5

JOHTIMET 300V UL

JOHDINSUOJAT UL

Aine Pinta Pintak„s.

Liittyy Nimitys

Yleistoleranssi Suhde KAAPELI CS34


Merkki Muut.no Muutos Piirt. Tark. Hyv. SFS 4011-

Laite

Piirt.
MGX
No.
A 80483 Tarkennetaan kuvaan tietoja 221102TMV25.11.02HK25.11.02HKe050400 PeJ
B 80557 Tarkenettu johtojen juotospuolet
Tark.
050503LK 8.5.03HK ------- 060600 NJ
Hyv.
060600 JJS
MGX 85076-3JB
Aine

Piirt.

Tark.
Merkki

Liittyy
POW X7

SFS 4011-
Yleistoleranssi
70
LIITINRIMA ENSTO KB223.10 2 1 2
Muut.no

MOLEX 5556-PBT

Hyv.
POW X1

Suhde
2 MOLEX 5556-PBT
Muutos

Abiko A2543R tai 3 1


Cembre BF-M4
4 160 4
5 3
6 MOLEX 5557-6R

Pinta
7 POW X2

Nimitys
8 2
9 210 1
2 4
10 4
MOLEX 5556-PBT 1 3
Kutistesukka 20
MOLEX 5557-4R

POW X3
160

3
230
2
RU
SI
MOLEX 1189-T
, 40
40 0
Nippuside UL 0

No.
MOLEX 3191-6R
Piirt.

Laite
2 Abiko A2507FLSF tai

Pintak„s.
Cembre BF-F 608P
RU AWG18
340 SI AWG18
Tark.

Johdinsuoja

Kuoritaan 7mm + holkitetaan


JOHTIMET 300V UL
Hyv.

JOHDINSUOJAT UL
Aine

Piirt.
A

Tark.
Merkki

Liittyy

SFS 4011-
Yleistoleranssi

070600 NJ
Muut.no

010999 PeJ
80557

Hyv.
Suhde
Muutos

-:-

070600 JJS
APEM A01ES+A0154B
(ei kuulu toimitukseen)

Pinta

Nimitys
Abiko A1503FLS5 tai
Cembre RF-F 305 tai
Päivitetty oikea liitinkoodi

Amp 165565-2
POW X15 C1 C3 CX13
MOLEX 5556-PBT MOLEX 5556-PBT
D1 D3
VI, 420 VI, 340
1 C1 C3 1
SI, 420 SI, 340
3 1 2 4
3 D1 D3 3

KAAPELI CS9
4 2 MU, 750 380 280 1 3
MU
2 2
RU, 750 RU
4 4

MOLEX 5557-4R

No.
MOLEX 5557-4R
Piirt.

Laite
Pintak„s.
050503LK

MGX
Motor-stop cable
Tark.

8.5.03HK

JOHTIMET AWG20, 300V UL


JOHDINSUOJAT UL
Hyv.

MGX 85087-4JA
-------
Aine

Piirt.
A

Tark.
Merkki

Liittyy

SFS 4011-
Yleistoleranssi

070600 NJ
010999 PeJ
Muut.no

80557

Hyv.
Suhde
Muutos

-:-
APEM A01ES+A0154B

070600 JJS
(ei kuulu toimitukseen)

Abiko A1503FLS5 tai


Cembre RF-F 305 tai

Pinta

Nimitys
MOLEX 5559-4P MOLEX 5558-PBT Amp 165565-2

CX13
Päivitetty oikea liitinkoodi

VI, 450 VI
1 C1
MU, 450 MU
2 C3 C1 C3
4 2
SI, 450 390 SI
3 1 3 D1 D1

KAAPELI CS10
D3
RU, 450 RU
4 D3

No.
Piirt.

Laite
Pintak„s.
050503LK

MGX
Motor-stop cable
Tark.

8.5.03HK

JOHTIMET AWG20, 300V UL


Hyv.

JOHDINSUOJAT UL

MGX 85088-4JA
-------
12 Appendices

12 Appendices
This chapter includes a Diamond architecture diagram and test display point
lists. For software block diagrams as well as a list of factory default settings,
AEC calibration sheet, AEC detector pattern, refer to appendices of the
Diamond Installation Instructions (86501-4A_). Note that the name Alpha X
in the diagram is a project code name of Diamond.

12.1 Parameter programming with Control Panel

NOTE
For smooth parameter setup and adjustment, we recommend the optional
Diamond Service Software (87502).

Service program menu contains special programs for system setups,


calibration and diagnostics. You can enter into the service mode by pressing
and buttons of the control panel down simultaneously.

The following screen appears:

Then press Set-key to enter the GROUPS screen:

Diamond service mode interface have been divided into several topic groups
in which the system setting and parameters can be changed:

A: System setup B: Defaults C: Diagnostics


D: AEC E: C-Arm Rotation F: Collimator
G: Compression H: Filter I: Generator
J: Tube Park K: Stereo Movement L: Physicist mode

Scroll with Up or Down-keys until the required group is between the brackets
as illustrated in the example. Press Accept to enter the group. The following
sections describe each group and available parameter choices. In parameter
sub-screens, accept the changes with Accept-key or cancel the changes with
the Cancel-key. Then Diamond returns to the group screen.

You can exit the service mode by pressing kV- and Den-buttons down
simultaneously.

For graphic description of the Diamond software functions, refer to Diamond


Installation Instructions (86501), Appendices section, Software block
diagrams.

86502-4AA Instrumentarium Imaging 12-1


12 Appendices

12.2 Loading software with old Flash Loader


Diamond software consists of Flash Memory and Battery RAM. The software
itself is stored in the Flash Memory while the settings are stored in the
Battery RAM.

NOTE
Flash Loader can only be used if, the all previous software versions have been
uploaded with Flash Loader.

CAUTION
If the previous software version has been uploaded with Diamond Service
Software, then the Flash Loader cannot upload the software and is likely to
render the existing software useless. Use of Diamond Service Software is
recommended.

WARNING
Before you start up the Flash Loader-program, close all other programs,
especially the ones using COM-ports. Simultaneously run programs may
severely disturb the software loading process.

1 Install Flash Loader-program to your computer. If it has already been


installed, proceed to next step.
2 Start up the Flash Loader-program by double-clicking the program
icon. The following screen appears:

12 2 Instrumentarium Imaging 86502-4AA


12 Appendices

3 Click Advanced-button to change installation parameters. The COM


port and Communication Time-out can be changed in this screen. These
settings are session-specific, not stored as new defaults.

4 Select the directory where the .bin-files are stored. This is performed by
writing the directory name on the Bin, file path-field or clicking the
browse button (...) and searching for the correct directory on hard disk.

When you find the correct directory, click any of the .bin-files listed.
Diamond will then set up default path. If the directory has already been
defined, you can proceed to next step.

Connecting Diamond to the laptop


5 Switch Diamond off.
6 Connect Diamond to a laptop with a RS serial cable (to COM1-port).
7 Press Connect-button and switch Diamond on.

After the connection has been successfully established, the screen displays a
green light above the Connect-button.

86502-4AA Instrumentarium Imaging 12-3


12 Appendices

8 If the green light for connection is displayed, the old software files can
be erased. Do not erase if a red light is displayed!

WARNING
During the software upload process, do not exit the process or switch off the
unit. When the software files have been erased, restarting the unit will be
complicated. In case blackouts occur in the environment where the unit ahs
been placed, it is recommend to have an UPS or similar electrical back-up
system while uploading the software.

WARNING
The deleted software files cannot be restored. Before you start loading a new
software version, write down all important service menu settings. Then
continue the loading process to the end.

9 Press Erase-button to empty the Flash Memory

At this point, the system has no operational software. Do not switch off the
power. Proceed to next step to install required software files.

10 Click Load-button to start installation. This will take a couple of


minutes. During the loading, the screen will display the progress by an
load indicator bar.

In case you get an error message “Error opening file”, check that the file path
is correct.

12 4 Instrumentarium Imaging 86502-4AA


12 Appendices

When the software has been loaded, the connection light turns red and the
following information is displayed:

11 Click OK-button to quit the Flash Loader-program

NOTE
In case the electricity went off during the software upload, or for some other
reason the unit will not start, call manufacturer for further instructions.

12.3 Diamond test displays


In service mode, Diamond Control Panel displays various figures that convey
information about the system condition. The printout of the screen items is
included in appendices.

1 Enter service mode


2 Press button 3 (the right-most button below Control Panel LCD-screen.
3 Press button 2 (in the middle, below LCD-screen). Screen displays
various test figures.
4 Check out listing of the page number in upper left corner of the LCD-
screen.
5 Press button 3 to scroll through pages 1-4.
6 After you have finished with test displays, exit the screen with -key and
exit the service mode.

86502-4AA Instrumentarium Imaging 12-5


12 Appendices

12 6 Instrumentarium Imaging 86502-4AA


GENERATOR BOARD SIGNAL LIST
TP NO. SIGNAL NAME SIGNAL DESCRIPTION DRAWING LED
tp1 +15VDC regulated DC supply voltage GEN1:d4 H1- grn
tp2 -15VDC regulated DC supply voltage GEN1:d4 H2 - grn
tp3 GND ground, supply voltage GEN1:d4 –
tp4 +34V unregulated DC supply voltage GEN1:d3 H3 - grn
tp5 FILA+34V unregulated DC with HEATENA control GEN3:d2 H4 - yel
tp6 FILA-COMMON tube current input (mA-feedback) GEN3:b6 –
tp7 GND FILA-COMMON gnd, tube current input (mA-feedback) GEN3:c6 –
tp8 ? (ma/pre feedback control circuit) GEN3:b4 –
tp9 ? (ma/pre feedback control circuit) GEN3:b4 –
tp10 ? (ma/pre feedback control circuit) GEN3:c4 –
tp11 KV-FB raw KV feedback, X2-3 GEN2:c6 –
tp12 MACTRL mA control GEN3:d4 –
tp13 MAREF mA reference GEN3:c5 –
tp14 FILAREF filament driver reference voltage GEN3:c4 –
tp15 CURLIMIT filament over-current protection GEN3:c3 –
tp16 HEATENA enable filament current GEN3:c3 –
tp17 FILAFB filament driver feedback circuit GEN3:b3 –
tp18 KV-FB-GND ground, KV feedback, X2-4 GEN2:d6 –
tp19 GND FILAFB ground, filament driver feedback circuit GEN3:b3 –
tp20 "KV" kV-feedback measurement point GEN2:d5 –
tp21 mA-FB mA-feedback measurement point GEN3:b5 –
tp22 CTRL-T1 filament T1 mosfet gate drive GEN3:b2 –
tp23 GND mA-FB gnd, mA-feedback measurement point GEN3:b5 –
tp24 CTRL-T2 filament T2 mosfet gate drive GEN3:b2 –
tp25 KVREF-RISE kV reference voltage with time constant GEN2:c5 –
tp26 KFREF kV reference voltage TRIM WITH R68 GEN2:c5 –
tp27 ERRORAMP+IN error amplifier input kVFB GEN2:c4 –
tp28 ERRORAMP-IN error amplifier input kVref-rise GEN2:c4 –
tp29 ERRORAMPOUT error amplifier output to VCO GEN2:c4 –
tp30 FBFRQ feedback frequency GEN4:d3 –
tp31 GND ground near shutdown tp GEN2:d2 –
tp32 VCOFREQ voltage-controlled oscillator frequency GEN2:c2 –
tp33 SHUTDOWN shutdown GEN2:d3 –
tp34 CURLIMIT bridge current measurement GEN2:d2 H10 - yel
tp35 FILA-COMMON gnd filament voltage GEN3:c1 –
tp36 FILAV filament voltage GEN3:c1 H11 - yel
tp37 DRIVE-D control drive to MOSFET block D GEN2:c1 –
tp38 GND ground GEN2:c3 –
tp39 DRIVE-A control drive to MOSFET block A GEN2:c2 –
tp40 DRIVE-C control drive to MOSFET block C GEN2:c1 –
tp41 DRIVE-B control drive to MOSFET block B GEN2:d2 –
tp42 OPTO-20VDC regulated 20VDC for bridge couplers GEN2:a5 H14 - grn
tp43 "310VDC-GND" ground, supply voltage for bridge GEN2:b6 –
tp44 "310VDC" unregulated supply voltage for bridge GEN2:a6 LA1 - neon
tp45 HV-TF2-2 bridge output GEN2:a1 –
tp46 HV-L4-2 bridge output GEN2:b1 –
– FILA-ERROR active if no MAOK in 50mS GEN3:a2 H5 - red
– GENENA enable generator GEN1:a3 H6 - red
– PRE/MA-CTRL control selects preheat or mA mode GEN1:a3 H7 - red
– EXPENA control to produce kV GEN1:a3 H8 - red
– SMALL FOCUS small/largefocus, LED lights on small GEN3:b5 H9 - grn
J1 FBMISS disable feedback missing test GEN1:b5 –
J2 FORCE-PRECTRL forces preheat control if J2/2-3 GEN3:d5 –
J4 MAOK mAOK forced active if J4/2-3 GEN3:a4 H12 - red
J3 KVOK kVOK forced active if J3/2-3 GEN2:c5 H13 - grn
You can enter Diamond test display in service mode by pressing middle button below display. By pressing button on the right side you can
change page of the test display. You can identify page of test display from number on the upper left corner of the test display.

7(67180%(5 6,*1$/ 7(67',63/$<

09.05.2001
SDJH LQWKHXSSHUFRUQHU

test1 Rear collimator stepmotor home limit (opto switch behind collimator)  test1 test2 test3 test4
test2 Rear collimator stepmotor steps ready (from microprocessor) test5 test6 test7 test8 test9
test3 Side collimator stepmotor home limit (opto switch) test10 test11 test12 test13 test14
test4 Side collimator stepmotor steps ready (from microprocessor) test15 test16 test17 test18 test19
test5 Filter stepmotor home limit (opto switch)
test6 Filter stepmotor steps ready (from microprocessor)
test7 Tube park position limit (opto switch)
test8 Tube almost in park position (opto switch)
test9 Tube almost in front position (opto switch)

test10 Tube front position limit (opto switch)


test11 Stereo right limit (opto switch)
test12 Stereo middle left limit (opto switch)
test13 Stereo middle right limit (opto switch)
test14 Stereo left limit (opto switch)
test15 Cassette stepmotor home limit (opto switch)
test16 Cassette stepmotor steps ready (from microprocessor)
test17 Cassette stepmotor enable (from microprocessor)
test18 Cassette stepmotor direction (0=backward 1=forwards)
test19 Detector selection buttons (L1=1, C1=2, R1=4, L2=8, C4=16, C3=32, R2=64, C2=128 )

SDJH LQWKHXSSHUFRUQHU

test21 Generator small focus output  test21 test22 test23 test24


test22 Emergency button test25 test26 test27 test28 test29
test23 Tube torque 0 = moment too big, 1 = normal (micro switch) test30 test31 test32 test33 test34
test24 C-arm right limit (opto switch) test35 test36 test37 test38 test39
test25 C-arm middle right limit (opto switch)
test26 C-arm middle left limit (opto switch)
test27 C-arm left limit (opto switch)
test28 Vertical low limit (micro switch)
test29 Vertical high limit (micro switch)

test30 Grid movement enable


test31 Control panel buttons kv+ =1; kv- =2; mAs+ =4; mAs- =8; Den+ =16; Den- =32; Filter=64; Mode=128
test32 Power failure indication
test33 Eps low limit (opto switch)
test34 Eps top limit (opto switch)
test35 Cassette step motor home limit (opto switch)
test36 Compression arm home limit (micro switch)
test37 Compression arm middle limit 0=upper half, 1=lower half (opto switch changes in the middle point of )
test38 Cassette sense
test39 Autopoint button (membrane switch)
09.05.2001
SDJH LQWKHXSSHUFRUQHU

test41 Filter selection button in control panel (membrane switch)  test41 test42 test43 test44
test42 Filter selection button in compression panel (membrane switch) test45 test46 test47 test48 test49
test43 Compression force plus button (membrane switch) test50 test51 test52 test53 test54
test44 Compression force minus button (membrane switch) test55 test56 test57 test58 test59
test45 Light button (membrane switch)
test46 Cassette sense switch 1 (hall sensor on lower right)
test47 Cassette sense switch 2 (hall sensor on lower left)
test48 Cassette sense switch 4 (hall sensor on upper left)
test49 Cassette sense switch 8 (hall sensor on upper right)

test50 Tube forward button (membrane switch)


test51 Tube park button (membrane switch)
test52 Tube forward button (membrane switch above tube)
test53 Tube park button (membrane switch above tube)
test54 Vertical up button (compression panel)
test55 Vertical down button (compression panel)
test56 Vertical up button (membrane switch above tube)
test57 Vertical down button (membrane switch above tube)
test58 Switch that drives cassette to front (hall sensor)
test59 Switch that drives cassette to back (hall sensor)

SDJH LQWKHXSSHUFRUQHU

test61 Stereo left button (membrane switch)  test61 test62 test63 test64
test62 Stereo right button (membrane switch) test65 test66 test67 test68 test69
test63 Stereo center button (membrane switch) test70 test71 test72 test73 test74
test64 C-arm left lateral pos button (membrane switch) test75 test76 test77 test78 test79
test65 C-arm left oblique pos button (membrane switch)
test66 C-arm center pos button (membrane switch)
test67 C-arm right oblique pos button (membrane switch)
test68 C-arm right lateral pos button (membrane switch)
test69 Cassette sense (opto switch)
test70 Paddle foot switch

test71 C-arm foot switch


test72 Release foot switch
test73 EPS foot switch
test74 Cassette used (from microprocessor)
test75 Selected detector (from microprocessor L1=1, C1=2, R2=4, L2=8, C4=16, C3=32, C2=128, R2=64)
test76 Thickness pulses from opto switch (just number, 127 in middle position)
test77 kV reference
test78 mA reference
test79 preheat reference
FILA+34V
TP1
+15V-15V GND +34V
TP2 TP3 TP4 X1 TP5
K1

X2 H1 H2 H3
C14
H4

TP6 TP7 TP8


TP9

TP10
MAREF CURLIMIT FILAFB

TP12

TP13

TP17
TP16
TP15
TP14
TP18 TP11
MACTRL FILREF HEATENA

TP19

TP20 H5 GND
KVFB TP21 TP22
TP23
MAFB
TP24
FBGND

TF1 K3
R68

KVREFRISE J2
KVREF
GENENA H7 EXPENA TP25 TP26 FORCE-PRECTRL

H6 H8
MACTRL
TP27 TP28 TP29

X
3
FBFRQ
TP30
X4
J3
GND
KVOK
TP31
J4 VCOFRQ CURLIMIT
MAOK
TP32

TP33

TP34
H1 H9
SHUTDOWN TP35 TP36 SMALL-FOCUS
TP37 TP38 TP39
A H10 FILAV
H12

H13

D TP40 TP41 FILA-GND


FILAV
MAOK KVOK GND C B

OPTO-20VDC
H14
D C B A

OPTO-20VDC
TP42
block D

block B

TP44
TP43
310VDC-GND 310VDC TP45
TP46
X5
block C

block A

X6
LA1 INSTRUMENTARIUM OY IMAGING
ALPHA X
GENERATOR BOARD
MGX-85004-3F
J2
FBMISS
+34v

PRE/MA

KVOK-COMP
GENENA

*MAOK
PRE/MA DAC
LARGE
+15v TF1
MACTRL FILA-ERROR HEATENA
PREREF SMALL
OUTA
KVMAREF(0:7)
MAREF
OUTB COMMON

MACTRL
SHUTDWN
CURLIM
PWM
MACTRL

SMALL

LARGE
FILREF
IN+
MACTRL
IN-
OUTA

0R1
OUTB

MACTRL
MAFB

MAREF

MAFB J4

SHUTDOWN
MA SECTION

EXPENA
13 SEC SHUTDOWN CURLIMIT V ref.
S Q TUBEFAIL BACKUP (60A)
C TIMER
TUBEFAIL R Q

J5
C B
KVFB

DAC VCOFREQ
KVFB
S Q A D
VCO C
KVREF
KVMAREF(0:7) OUTA R Q

(Error Amplifier)

KVOK-COMP
A C

J3

B D HV-Unit
*KVOK

KV SECTION 310Vdc 310Vdc-GND


DIAMOND ARCHITECTURE
6 MAR 02 tmj
STER ECS
SENS SENS

COM COM FOOT CASS


SW
STER ECS COMP GRID
1 2 TRAY

CAS
TRAY

CONTROL SUPERVISOR GRID


SB SB STEREO BOARD CS DRIVE BOARD
PANEL BOARD
CAS
HLDR

CS CS
PADL
SENS
DET
COMPRESSION SEL
PW SW
BOARD FORC
GENERATOR SENS
POWER BOARD
BOARD
THICK
SENS

KVFB

CASS
ROT Z LOCK
HV-UNIT MAFB
FILAMENT
VOLTAGE
AEC BOARD
ROT Z CAS
SENS SENS SENS OPTO
RS-485 4-WIRE SERIAL BUS

HIGH VOLTAGE
REAR
OPTO

X-RAY TUBE REAR


PLATE
— CONTROL/SIGNAL COLLIMATOR
POWER STATOR VOLTAGE COL FIELD LIGHT
BOARD
— SYSTEM BUS
SIDE
CS PLATE
— CIRCUIT BOARD
AMD
BOARD COL MIRROR SOL
— MOTOR TUBE
FILTER SIDE
PARK
OPTO
FIL TUBE
— SENSOR, I/O OPTO SENS
TP1 TP2 TP3 TP4
ICA2 H1

RST X2
ICA4 ICD3

ICA5

ICD5

RXD
TP5

TXD
TP6
H2
TXD
H3
RXD TP7
GND
+5V
TP8

INSTRUMENTARIUM OY IMAGING
MGX
AEC BOARD
85002-3F

X1

GND +12V -12V


X3
TP9 TP10 TP11
20M

DETECTOR
DETECTOR ARRAY L1
SELECTION 87C51
RS-485
Micro-
MUX Driver
Controller

S1 TxD
S2
20M
S3
S4
S5
S6 RS-485 RS-485
DETECTOR ARRAY C1 Driver SERIAL LINK
S7
S8
RxD

D
DETECTOR ARRAY R1
DETECTOR ARRAY L2
DETECTOR ARRAY C4
DETECTOR ARRAY C3
DETECTOR ARRAY R2
DETECTOR ARRAY C2

8-bit
DAC

C4 OFFSET
GAIN CONTROL
SELECTION
C3 12-bit
MUX
ADC
L2 R2
C2 1K0
S1

2K0
S2
S3
4K02
L1 C1 R1 S4
8K06
S5
16K2
S6
16K2
S7
16K2
S8
16K2
Main transformer
POWER CABLE
85083-4J
20354 23388 MAINSCABLE 20367 CS10 Motor

MAINS 85084-4J MAINS SWITCH


CB1
85088-4J Stop
button C-arm Tube Head Tube motor
Field Light
Coll motors
Filter Motor
Tube Head pos.
MAINS FUSES LINE FILTER Moment
CX13 Motor
Tubehead panel
Mirror Solen. Sensors
Stop
CS5 Terminal block button 86006-4PJ CX3 CX4 CX5 CX6 CX7 CX8

X5
Footpedals Generator Board Adapter CS34
85082-4J CB1 CS9 85092
Board 85076-3J
85086-4J 85087-4J

X4
CB30 CS31 CS32 CS33
85062-4J 85041-4J 85042-4J
X-ray Tube 85040-4J

X2
+310Vdc +20Vline X3 X1,2,7 X15 X4
Projection Projection X1 X9 X3
CS6 PW1 panel panel
X10 X4
AMD Board
COM1 85056-4J Collimator Board
Molex
CS35
+20Vline PW33

X6
85080-4J +310Vdc 90136

X6

X5
86004-4P Molex

X1
85096-4J

X2
+34Vdc 90136
CS0 PW2 230Vac Tube Tube
87C51FA 85044-4J
COM2 Park +34Vdc

X10
Park
panel

X1
85077-4J HV Unit

X4
panel
85057-4J
-15Vdc +15Vdc 86003-4P +5Vdc +15Vdc 230Vac

Carriage X3
+15Vdc -15Vdc
Molex Molex PW32 FAN
X6

X2
Minifit X7

X8
Adapter Adapter Minifit X3
Board Board

X13
86000-4P
C-arm CS3 -24Vdc +24Vdc 85063-4J

X8
Sensors 85037-4J CX10 CS11
85035-4J

85055-4J
Control Panel CS4
85038-4J
+34Vdc
CB30

X11
C-arm
Motor X9 Power Board 85062-4J CX9
CS1 CX2
87C51FA CX1 CS2 PW31
85036-4J
CS7 CX11 85061-4J
-24Vdc Z-motor (Optional)
85085-4J
X8 X5 X6

X20
CX12
SB31
+5Vdc PW3 CB10 85018-4J
X9 X7

X10
85016-3J
87C51FA

X12
DS75176 (RS485)
DS75176 (RS485)
85058-4J Compression Block
X6

X3 Compression Board
MUX Molex
SB40

X14
90136 +5Vdc
X11 85064-4J
X3

CS8

X10
85079-4J
MAX232 (RS232) Compression

X10
DS75176 (RS485) DS75176 (RS485)

X1
TL16C451 Display Board

X22
87C51FA
X2

15 pin D-connector CS20 PW30


85039-4J

X18
-15Vdc +34Vdc 85060-4J
MAX232 (RS232) ECS limits

X2
Sensor X2
SB2
MAX232 (RS232)
Stereo Board CS23 CX24 X4, X5
85047-4J

X17
X15
85081-4J +5Vdc Compression Motor
i386EX ECS Motor +15Vdc
& Tacho CS40
X1

85054-4J
CS22 Comp. sensors Comp. panel
15 pin D-connector (rear box) CS21 Stereo Sensors

X3
X4 X16 X21 X11 X19 85046-4J
MAX232 (RS232) 85045-4J Strain Gages 86001-4P
DS75176 (RS485) (measure comp force)

SB1 PW20

85059-4J
PW21
85051-4J
CS27
+5Vdc 85019-4J
85017-4J -24Vdc
X12

Supervisor Board CS29 CS28 CB10


Molex Molex
85053-4J 85052-4J 85050-4J X2 X3
Molex 90136 90130-
Adapter X3 X2 X5
Minifit DS75176 (RS485) 3208
85094-4O Board
CX21 Cass. Holder sensor
Adapter
85094-4O Board +34Vdc
CS25 Grid Sensor +5Vdc -12Vdc

X1
Cass. Tray sensor CS26

X1
85049-4J
Legend:

85050-4J
Stereo Motor
Ctrl/ Data signal +XVdc 86005-4P Stereo panel
Voltage regulator CX22 Molex
Power signal Stereo Tacho +5Vdc 90130- +12Vdc
86005-4P Stereo panel
CS24 Cassette Motor 3210 87C51FA

X4
Connector +XVdc 85048-4J & Grid Motor
Supply voltage Cass. Locks AEC Board
CXxx Cable Connector Drive Board CX23
SBxx Signal Bus Cable Cass. Sensor
Cass. Locks
Image Receptor Support
PWxx Power Supply Cable 85090

CSxx Signal & Drive Cable


CBxx Cable Assembly Dokumentti luotu Micrografx Designer ohjelmalla

CHANGE NO: UNIT:


80164 Alpha X
DRAWN: DRAWING NUMBER:
5.2.2001 PeJ MGX 85115-3SA
A 80216 LISÄTTY PUUTTUVIA KAAPELEITA 120701SF 220801HK 220801MJ CHECKED: 25.4.2001 HK
DRAWING NAME:
Diamond Main Wiring Diagram
APPROVED: REFERENCE: SHEET :
26.4.2001 MJ 1/1
Instrumentarium Imaging Singapore
152 Beach Road

86502 06/2003 © Instrumentarium Imaging


#12-03A Gateway East
189721 Singapore
Tel. +65 6391 8600 · Fax +65 6396 3009
Instrumentarium Imaging Instrumentarium Imaging France S.A.R.L. singapore@sg.instrumentarium.com
P.O.Box 20, FIN-04301 Tuusula, Finland 4, Avenue des Roses
Tel. +358 10 394 6500 · Fax +358 10 394 6501 94386 Bonneuil Sur Marne Cedex, France Instrumentarium Imaging Ziehm GmbH
mammo@fi.instrumentarium.com Tel. +33 1 43 39 51 51 · Fax +33 1 43 39 75 75 Isarstrasse 40, D-90451 Nuremberg, Germany
instrumentarium.imaging@wanadoo.fr Tel. +49 911 642 070 · Fax +49 911642 0739
Instrumentarium Imaging Inc. info@ziehm-eu.com
300 West Edgerton Avenue, Milwaukee Instrumentarium Imaging Italia S.R.L.
Wisconsin 53207, USA Via Cassanese, 100 Instrumentarium Imaging Ziehm Inc.
Tel. +1 800 558 6120, +1 414 747 1030 20090 Segrate (MI), Italy 4181 Latham Street Riverside, CA 92501, USA
Fax +1 414 481 8665 Tel. +39 02 21 30 28 1 · Fax +39 02 21 30 28 60 Tel. +1 909 781 6660 · Fax +1 909 781 6457
info@usa.instrumentarium.com instrumentarium@it.instrumentarium.com mail@ziehm.com

Instrumentarium Imaging is constantly improving its products and reserves the right to
change these specifications without notice. Distributor:

www.InstrumentariumImaging.com

Das könnte Ihnen auch gefallen