Beruflich Dokumente
Kultur Dokumente
Handout and
None
exercises:
Appendices: 1. Sample recording and reporting forms:
o PT2. Logbook/Aggregate Results of Test Slide Sets
o PT3. Panel Testing Individual Results and Feedback Form
o PT4. Panel Testing Report of Multiple Laboratories for
District Supervisor & NTP
2. An example of an accompanying letter for sending out panel
testing sets to peripheral laboratories.
1
Panel Testing Condensed Module
Panel testing (PT) is one of the options for EQA, it is a system of sending stained and/or
unstained smears from the NRL to peripheral and intermediate laboratories to evaluate
the proficiency of laboratory technicians in performing AFB smear microscopy. This
method tests individual performance of a laboratory worker, not the laboratory overall.
Panel testing is considered a minimal requirement or standard for EQA in settings with
limited resources. Panel testing is considered to be less effective than blinded rechecking
because it does not monitor routine performance. However, in some areas with low
incidence of tuberculosis or very few microscopy problems, a labor-intensive rechecking
program may not be justified to detect only a few errors. Routine panel testing may be
more cost effective in these settings.
PANEL COMPOSITION
Panels can be composed of slides with known results originating from routine work, but
this leads to problems of consistency and it is more difficult to compare performance
between laboratories. Therefore, it is recommended that panel slides are specially
manufactured for this purpose to minimize variation and provide a more uniform test for
all participating labs.
2
Panel Testing Condensed Module
Frequency of testing
Panel testing may be conducted occasionally as a special test or survey or periodically at
least one to two times a year, especially if there is no rechecking system in use. Although
panel testing may be performed in addition to rechecking, its usefulness will be more
limited. However, a panel test can be of great value when investigating the causes of poor
performance suggested by rechecking.
3
Panel Testing Condensed Module
supervisor should plan the visit to the peripheral laboratories in a way that allows
sufficient time to conduct routine on-site activities and allocate required time for
laboratory technicians to read panels (time allowed/limited for a laboratory technician to
review panel and report results is a maximum of 2 hours for a stained slide set and 3
hours for an unstained slide set). The supervisor should take into consideration the
number of microscopes available in the laboratory and the number of laboratory
technicians that will participate in panel testing, and ensure that panel testing does not
disrupt routine examination of patient specimens. The peripheral laboratory staff should
be notified in advance to allow them to plan for the extra workload. Since administration
of panel testing during supervisory visits can be impractical in routine conditions, it may
be best reserved for use as part of a special survey.
Classification of errors
Errors in reading smears and reporting results will be classified as major or minor: false
positive/negative errors related to 1+, 2+ or 3+ results are considered to be major, while
quantification errors (at least a two-grade difference) and errors in the scanty group are
considered to be minor. Summary information on errors classification is presented in the
table below.
Correct: No errors
QE Quantification error Minor error
LFN Low False Negative Minor error
LFP Low False Positive Minor error
HFN High False Negative Major error
4
Panel Testing Condensed Module
Scoring systems
Prior to a panel testing round, the NTP/NRL will develop a scoring system to distinguish
errors and detect problems with technician’s performance. Different scoring systems can
be used depending on the type of errors that are the focus for detection and corrective
action. An example of a scoring system that has different scores for major and minor
errors is presented below:
Set of 10 slides, each slide is worth 10 points, total possible score = 100.
a. HFP and HFN scores 0
b. LFP, LFN and QE scores 5
c. Passing score = 80 – 90
Post-validation
Unfortunately, no system for developing test panels and distributing them to peripheral
sites is entirely without problems. Post-validation is the process of validating panel
slide/batches after receiving aggregate results from all laboratories and before
determining a final score. Post-validation is needed to confirm that discordant results are
due to a laboratory technician’s error and not a problem with a panel slide. If a majority
of technicians fail to report correct results for the same slide/batch, this may represent a
problem with the slide preparation at the NRL. When numerous discrepant results suggest
that problems occurred in the process of preparing the panels at the NRL, the slides
should be returned to the NRL for rereading. This will help the NRL detect the source of
the problem(s), including:
- technical difficulties in preparing individual slides;
- an error in the initial reading of a smear at the NRL;
- incorrect recording of expected results;
- fading of stained smears during transportation to peripheral sites.
The Logbook/Aggregate Results of Test Slide Sets is used for post-validation purposes.
If post-validation confirms that the consistency and quality of the panel set was
acceptable then unacceptable performance in panel testing should result in investigation
to identify the reason, including all sources of technical error. False-positive and false-
negative errors should be considered separately. False-positives may indicate lack of
proficiency or a faulty microscope. If the slides have been returned, it is possible to check
for artifacts that may have caused confusion, but it often will not be possible to determine
the exact cause of the problem without further investigation on-site. False-negatives may
also indicate a lack of proficiency, but may be due to poor stains or staining procedure, a
bad microscope or superficial work in examining the slide. If positives in stained slides
were reported correctly and the errors occur only in smears sent out unstained, the
laboratory stains or staining process may be at fault. This can be verified by checking
and re-staining smears that were returned. If errors are found in both stained and
unstained smears, the reason may be the microscope or lack of technician proficiency.
5
Panel Testing Condensed Module
Feedback is always required and should be timely and confidential. At a minimum, the
participants should be informed of their own results (Panel Testing Individual Results and
Feedback Form) and how it compares to the aggregate performance of other laboratories.
To gain the greatest benefit and impact from the exercise, a visit to the peripheral centers
with the poorest scores is required for on-site identification of the problems and remedial
action. Thus, technical supervisory visits offer the best opportunity to review results of
panel testing with the technicians in the peripheral laboratories. For this reason, on-site
supervisory visits by experienced staff from the intermediate laboratory or NRL are
recommended once a year and more frequently if significant problems are identified.
Reporting to the NTP coordinators and local health authorities may also be required. As
non-laboratory persons, they may have difficulty in understanding and interpreting the
results, and suggest punitive or disciplinary action. To avoid unintended interpretation of
the results, the report should provide sufficient background information, as well as
carefully considered conclusions and recommendations, not simply a list of scores.
Reports should include both individual results, as well as aggregate performance for all
laboratories tested (Panel Testing Report of Multiple Laboratories for District Supervisor
& NTP). EQA feedback is always confidential: laboratories/laboratory technicians should
be given individual code numbers when presenting aggregate results.
If a majority of laboratories submit unacceptable panel testing results, but the consistency
and quality of panel testing slides is found to be acceptable, this likely represents serious
problems with AFB microscopy. Additional resources should be obtained for supervisory
visits, correction of problems identified in individual laboratories, including replacement
of microscopes (and/or microscope objectives), retraining if needed, and follow-up panel
testing.
Key messages:
Panel testing is an effective method to quickly obtain information about the
capabilities of individual laboratory technicians to read smears and report results
according to standards approved by NTP.