INSTRUCTION / INSTALLATION /
TECHNICAL MANUAL
ADAPTABLE MODEL:
ST-D520
ST-D530
ST-D540

■ Thank you for purchasing the


dental therapy unit produced by
our company.
■ Please read this manual
carefully for making best use of all
functions of this product. The manufacturing of this product conforms to the
■ Please keep this manual for 93/42/EEC directive issued by EU.
reference in case of problem
occurred to the dental unit.

FOSHAN SUNTEM MEDICAL INSTRUMENT CO., LTD​



CATALOG
I. SAFETY INFORMATION
II. …………………………………………………………………………………………… 1
III. FIGURES AND SYMBOLS
IV. ………………………………………………………………………………………… 1
V. MATTERS OF CONCERNMENT ……………………………………………………………………………… 2
VI. TRANSPORTATION AND STORAGE …………………………………………………………………… 3
VII. PRODUCT INTRODUCTION …………………………………………………………………………………… 3
VIII. USING CONDITIONS
IX. ………………………………………………………………………………………………… 3
X. BASIC PARAMETERS
XI. ……………………………………………………………………………………………… 4
PRODUCT STRUCTURE DIAGRAM …………………………………………………………………… 5
INSTALLATION
…………………………………………………………………………………………………………… 6
AU JUSTMENT
…………………………………………………………………………………………………………… 7
USING INSTRUCTIONS
…………………………………………………………………………………………… 8
1. INSTRUCTIONS TRAY CONTROLLING PANEL …………………………………….…… 8
2. ASSISTANT TRAY CONTROLLING PANEL …………………………………………………
9
3. WATER TANK
XII. …………………………………………………………………………………………..……… 10
XIII. 4. FOOT CONTROLLER
XIV. ………………………………………………………………………………………… 10
XV. 5. VALVE OF HANDPIECE ….
XVI. ……………………………………………………………………………… 10
XVII.
6. OPERATING LIGHT
…………………………………………………………………………………………… 11
7. X-RAY FILM VIEW
…………………………………………………………………………………………… 11
8. 3-WAY SYRINGE AND SUCTIONS
……………………………………………………………… 11
9. SYSTEM INITIALIZATION
………………………………………………………………………………… 11
MAINTENANCE AND REPAIR …………………………………………………………………………..… 12
GUIDE TROUBLESHOOTING ……………………………………………………………………………… 13
CIRCUIT SKETCH MAP …………………………………………………………………………………………
13
CIRCUIT DIAGRAM OF THE PATIENT CHAIR …………………………………………… 14
WATER AND AIR PRINCIPLE ………………………………….………………………….…………… 14
WATER AND AIR LINE SKETCH MAP …………………………………………………...……… 15

I. SAFETY INFORMATION

Be sure to read through this operating manual to obtain a correct using of this unit. In case of any special requirement,
please contact the dealer of our company.

WARNING:
1. Please neither put the appliance in moist environment nor open the tank body for the sake of
reducing the possibility of fire and electric shock. In case of maintenance, please refer to professional
maintenance man.
2. In case of any smoke or other abnormal circumstance which may result in fire or other conditions,
please unplug the power connection and contact the local dealer or the after-sale service department.

3. In case water enters into the equipment, which may result in creepage, please unplug the power
connection and contact the local dealer or the after-sale service department.

4. In case of any undefined fault of the equipment, please unplug the power connection and contact
the local dealer.

5. Please don’t disassemble or refit this product, otherwise, some accessories and components may be
damaged or result in electric shock.

6. The power line shall not be damaged. In case the power line is damaged, please disconnect the
power supply. Don’t touch the power supply with hand directly. Please contact the local dealer to change the
damaged power line.

7. The product is adaptable to the voltage of a.c.220V-230V / 50Hz-60Hz. In case of work under other
voltage, dangers may occur.

8. Please care the weather change. In case of thunder, please disconnect the power supply.
9. Please don’t touch the power plug with wet hands, which may result in electric shock.
10. Please don’t use insecticide, alcohol, thinner and chemical liquid to clean the equipment, which may peel or rot its
surface.
11. Please don’t remanufacture the machine. In case of any damage consequently, it is beyond any maintenance and repair
guarantee.
12. In case of the risk of electric shock, please connect the appliance to the grounding power supply which are protected.
13. Please don’t replace the power line optionally,in case the dangers occur.

TO USERS:
This product has undergone the test, which conforms to the requirement of EU’s 93/42/EEC, 2007/47/EEC instruction.
No hydrargyrum segregator has been installed within the product; therefore, its drainage system cannot filter solid
particles and debris, or amalgam alloy. In order to avoid such pollution to the drainage system, please the users to install a
hydrargyrum segregator additionally to avoid the drainage pollution.
If you want to know more related information, please contact us.

PLEASE RECORD THE PRODUCT CODE OF THIS APPLIANCE HERE FOR THE CONVENIENCE OF INQUIRY
IN THE FUTURE.

II. FIGURES AND SYMBOLS

FIGURES FIGURES
DESCRIPTIONS DESCRIPTIONS
AND SYMBOLS AND SYMBOLS
Warning uprising

B type equipment Patient chair declining

Protect the Grounding terminal Patient chair reclining

EU announcement institute Patient chair fore raking


On Spittoon position

FIGURES FIGURES
DESCRIPTIONS DESCRIPTIONS
AND SYMBOLS AND SYMBOLS
Off Cuspidor flushing

Operating light Cup filling

X-ray Film viewer Water heater

Move to the limit of two ends Use instructions


pursuant to the line directions
warning temperature of surface
Emergency stop
overheat
Luminance Step adjustment
The waste of Equipment, accessories The same outer packing boxes
and packaging shall meet the cannot be stacked over two wooden
2002/96/EC directive issued by EU cases in the transport or storage

III. MATTERS OF CONCERNMENT

1. This product can only be used for dental treatment, therapy and operation.
2. The children are forbidden to play near the equipment in order to avoid accidents while touching the
equipment.
3. Only use the fuse and accessories recognized by Anle can be used.Plears make sure the e power was
unplugged in case replace the fuse.
4. This dental unit should not come into contact with any object in its operational range of movement , it
- 15 is.
- 5. After using this dental unit r if this dental unit is not os not to be use the patient chair be returned to
the lowest position and, water air , and the power supply be cut off.
6. Make sure the pillow is locked before using the patient chair.
7. Don’t use the product in high-temperature, high-moisture, vibration, flammable or unbalanced
environment.
8. The patient chair is intermittent operation equipment; therefore, the continuous operation cannot
exceed 1 min.
9. Read the use instructions before using the handpiece.
10. Read the use instructions before using the curing light machine and the scaler.
11. Connect the water supply before using the water heater in order to avoid short circuit with of dry heating.
12. Contact professional maintenance man for unit repairing or maintenance, disconnect the water, air and power supply
before the repair.
13. The product is forbidden to be connected with the power supply for more than 24 hours. Disconnect the water, air and
power supply while not using the appliance for long time.
14. It is advised that there are 5 years used life fit this dental unit. The excessive using may result in aging pipes,
malfunctions or even accidents. Please contact qualified medical appliance recycle unit to eliminate or change new
equipment in time.
15. The operating stool doesn’t belong to the basic structure of the dental unit and the guarantee period of the operating
stool is three months.
16. The cold light lamp, PU cushion, real leather cushion, glass parts and ceramic parts within the machine are
damageable parts and their guarantee period is all three months.

17. The product’s quality problems due to manual reasons are beyond the guarantee area. ​
18. Patient position requirements: height adjustment to the best position for operators’ operation, is the patient lying on
the dental patient chair, and the operator operating i near the patient's head.
19. Emergency stop equipments: in case the emergency, press the emergency stop switch, you can safely terminate the
movement of the electric mechanical moving parts.
20. Can not be used under the environment of mixing flammable an aesthetic with air or oxygen or nitrous oxide .

IV. TRANSPORTATION AND STORAGE


1. Transportation conditions: be upright; handle with care; and moisture-proof.
2. Transportation environment: temperature 5-40℃, relativity≤80%.
3. Storage environment: temperature 5-40℃, relativity≤80%. No corrosive gas and good indoor ventilation.
4. Superposition quantity limitation: the same outer packing boxes cannot be stacked over two wooden cases in the
transport or storage.

V. PRODUCT INTRODUCTION

1. FUNCTION AND THE INTENDED USE


This product, as a kind of dental appliance that can be adopted by qualified person in the course of dental clinic and its
related operation process, is composed by a set of related dental equipment and appliance parts for the purpose of dental
diagnosis, therapy and operation.
This product mainly include the dental patient chair, dental lamp, instrument tray, 3-way syringe, spitting and water
supply system, suctions, cuspidor, film viewer and foot switch.
2. ATTACHMENT LIST
NO. ATTACHMENT NAME QUANTITY REMARKS
1 Doctor chair 1 1 set of doctor chair
2 Dental light 1 1 set of dental light
3 Lamp arm 1 1 set of Lamp arm
4 Backrest, Cushion 1 1 set of Back rest and Cushion
——a product qualification certificate.
——a Product installation instructions.
——a Product warranty card.
5 random file 1 ——2 sets saliva ejectors.
——M12*15 Level adjusting screw.
——12 mm "L" type six angle wrench.
——3 sets 6A fuses and 3 sets 10A fuses.
6 Dental handpiece Optional Attachment
7 Curing light machine Optional Attachment
8 Scaler Optional Attachment
9 Oral cavity endoscope Optional Attachment
10 Amalgam separating device Optional Attachment

3. SAFETY FEATURES
■ Name of product: dental unit ■ According to the safety degree classification
■ Model: see as the Product used in the conditions of flammable anaesthetic
nameplate mixture with air or oxygen or nitrous oxide
■ Safety class: class I anaesthetic mixture: non type AP/APG
■ protection type: type B ■ Operating categories: dental patient chair is
intermittent operation equipment
■ power type: Single phase three
cores,
■ Mode of operation: Dental patient chair works
net power supply for 2 minutes and stop 18 points
■ Rated voltage/frequency: a.c.220 ■ This product has no protective effect to
V/50 Hz defibrillator, no signal output or input part.
■ Input power: 350 VA ■ Software version number: 6.18
■ Liquid inlet protection: Foot
switch IPX4
■ This product is a non permanent
installation, non common movable type
equipment

VI. USING CONDITIONS

IN ORDER TO MAKE SURE THE NORMAL USE OF THE MACHINE, THE INPUT WATER, AIR, POWER
IN ORDER TO MAKE SURE THE NORMAL USE OF THE MACHINE, THE INPUT WATER, AIR, POWER
SUPPLY AND ITS WORK CONDITIONS SHALL MEET THE FOLLOWING PARAMETERS:

1. ​Air supply: air pressure 0.60 MPa-0.80 MPa, flux>50 L/min, no oil.
2. ​Water supply: water pressure 0.20 MPa-0.40 MPa, flux>10 L/min, rigidity<25 degree.
3. ​Power supply: a.c.220 V-230 V, 50 Hz/60 Hz, 10 A.
4. ​Environment: temperature 5℃-40℃, relative moisture 30%-75%.
5. Operation space: LˣWˣH>4 mˣ3 mˣ2 m, The ground is smooth and oblique angle less than 0.5°.
VII. BASIC PARAMETERS
■ Floor box: Rated voltage: a.c.220 V/50 Hz, Single phase three cores, net power supply
Rated power: 350 VA
Water filtering aperture: 90 µm
Air filtering aperture: 25 µm

■ Cup rinse water supply Water rate: 4 L/min


device:

■ Saliva pump: Saliva ejector:


Water pressure: 200 kPa, Vacuum 15 kPa, Pump speed 800 mL/min
High-power suction:
Air pressure: 500 kPa
Vacuum 25 kPa, Pump speed 1000 mL/min

■ Water-heater: Rated voltage: d.c.24 V


Rated power: 400 VA
Water temperature: 40℃±5℃

■ Instrument tray: Angle of rotation: 270°


Movement range: 440 mm
Disk locking device bearing quality: ≤5 Kg

■ X-ray film viewer: Rated voltage: a.c.24 V


Rated power: 30 VA
Luminance: 2 000 lx

■ Operating light: Rated voltage: a.c.12 V


Rated power: 50 VA
Luminance: 8 000 lx-20 000 lx
Radiant heat: 200 W/m2 (max Luminance)
Color rendering index: 85Ra

■ Foot switch Tripping force: 10 N N 30 N


Liquid inlet protection: IPX4
Service life: 25 000 times

■ Noise: 70 dB A

■ Dental patient chair: Voltage: d.c.24 V


Loading capacity: 1323 N
Headrest loading capacity: 300 N
Head rack extension range: 120 mm
Backrest movement range: 105°- 170° ​
The highest position of the chair from the groud: ≥700 mm
The lowest position of the chair from the groud: ≤500 mm
The biggest backward angle of the seat cushion: 12°

■ Optional Attachment: Registration certificate of products inside.


——Dental handpiece Technical parameters can be seen in the Use instructions.
——Curing light
——Scaler
——Oral cavity endoscope
——Amalgam separating device

■ Work space: LˣWˣH>4 mˣ3 mˣ2 m

VIII. PRODUCT STRUCTURE DIAGRAM

1. STRUCTURE DIAGRAM OF ST-5202. STRUCTURE DIAGRAM OF ST-D530

3. STRUCTURE DIAGRAM OF ST-D540 ☞ NOTE:


1. Protective cover 2. Cushion

3. Backrest 4. Headrest
5. 3-way syringe 6. high power suction
7.saliva ejector 8. Assistant control
9. Spittoon 10. cabinet
11. Lamp arm 12.LED light
13. balance arm light 14.tumbler
15.balance arm 16. Film viewer
17. Instrument tray 18. Instrument tray handle
19. Low-speed air motor 20. High-speed air turbine
handpiece handpiece
21. 3-way syringe 22. Front cover
23. Armrest 24. Base cover

4. SKETCH MAP OF THE EXTERNAL JUNCTION BOX

IX. INSTALLATION

1. INSTALLATION SPACE

1.1. ENVIRONMENT REQUIREMENT


1.1. ENVIRONMENT REQUIREMENT
In order use the normal for this dental unit, please install it
in a clean, dry and ventilative area, and the floor is in leved
and solid.
1.2. SPACE REQUIREMENT
Aviod the touching from any ebiects during the dental units
in the wooking, the installation space should be reauired the
following items:
a) The length no shorter than 4000
mm;
b) The width no shorter than 2600
mm;
c) The height no shorter than 2000
mm.
2. INSTALLATION OF THE PRODUCT
2.1. INSTALLATION AND TRANSPORTATION
Open the package, take out four screws that fix the patient
chair on the package’s board by using M12 external hexagon
wrench; and move the patient chair carefully to the
installation position.
NOTE:
☞ During moving, please
don’t carry the fragile parts such
as the rotation arm, armrest,
assistant tary, spittoon unit and
plastic cover so as to avoid
damage in the excessive moving
force.
☞ During moving ,
please stabilize the instrument tray
so as to avoid unexpected
touching due to its rotation.

2.2. CONNECTION BETWEEN THE SPITTOON UNIT WITH


THE PATIENT CHAIR
Uplift the spittoon therapy unit to make the cables away
from the connection point of patient chair in 100 mm; then
connect the cable plugs from patient chair (please connect
them according to correct numbers ) to the cables from
spittoon unit ;fix the spittoon unit to patient chair by 3 pieces
M10 he socket .

2.3. INSTALLATION OF INSTRUMENT TRAY


Used with a rotating arm of the decorative cover, connecting
head decoration cover into the treatment machine.
Handling personnel lifting instrument tray assembly, which
is mounted to the treatment machine.Installation personnel
connect the pipeline according to their numbers.

2.4. BACKREST OF THE PATIENT CHAIR 2.5. CUSHION OF THE PATIENT CHAIR
By 3 piece M10 hex socket. Take out four screws on the cushion’s base; re-screw on the
NOTE: cushion through the corresponding four holes, and fix the
hexagon nut.
☞ Avoid sharp NOTE:
articles to lacerate the
backrest’s surface in the ☞ Avoid sharp articles to
course of installation. lacerate the cushion’s surface in
the course of installation.
2.6. INSTALLATION OF BACKREST
Take out of the and adjust the backrest on The chair frame.
2.7. INSTALLATION OF OPERATING LIGHT
Open the package box of the lamp arm, and connect them in
turns according to the figure.
Take out the operating light, and screw off the M5 hex
socket from the connecting spindle. As in the figure 2, the
decorating ring is covered the connecting spindle, connect
those plugs from the operating light connecting spindle to
the light arm connecting spindle (Note: make sure
connecting the correct color wires , and put the insulated
cover on them.) insert the wires and plugs in the light arm,
finally, insert the operating light connecting spindle to the
operating light arm.

2.8. JUNCTION BOX


Connect the water supply tubing, air supply tubing and
drainage tubing to water source, air source and draginage
respectively. During the connecting, make sure that the seals
of the connecting joints for water supply tubing and air
supply tubing keep in good condition to avoid the leak of
water and air. Connect the ground wires to the earth.
According to the position where the unit is installed, air
tubing, water tubing should reach the desinated position.
Water tubing and air tubing should be adopt water pipe in
ø8x5 mm PVC external diameter, being connected by
elbows. The interal diameter of water flush tubes should be
lager than 27.5 mm, with a flush hole left. As the diagram on
the right.
NOTE:
☞ The draginage tubing
should be under the ground at least
in 10 mm, otherwise, the back
current will be caused in the
draginage tubing.
☞ The outfall shall be
declined for the fluent current.
☞ The outfall shall not be
twisted, raised or undulant, or put
in the water.

2.9. HORIZONTAL ADJUSTMENT


Screw six M12 hexagonal bolts into the M12 screw thread
holes in the bottom cover of the chair to the level of chair
blancing.
NOTE:
☞ In the process of
adjusting, make sure that bottom
cover touches the floor as much
as posssible to ensure the stability
of the whole set.
2.10. INSTRUMENTS INSTALLATION OF TRAY
A. 3-WAY SYRINGE
Take out the nozzle of the 3-way syringe, press the meter
ring and insert the nozzle, and then release the meter ring to
make it back to the original position. If the meter ring cannot
recoil to the origin, the insert of the nozzle is not finished.
The meter ring should be pressed again to make the nozzle
inserted till the meter ring recoils.
B. HANDPIECE
turn on the water source and air source before installing the
handpiece, pull out every handpiece tubing and push on the
footswitch, blow off the remained water and air in the
tubings, install the handpiece to the tubing connector.
NOTE:
☞ Please refer to the
handpiece user manual as
handpiece using, distinfection and
maintenance.

X. AU JUSTMENT
NOTE:
☞ after using the dental unit, it is better to turn off the main switch, it can be protected
the spare parts, lengthends service life, reduce troubleshooting.
1 Open the main switch (water, air and power). When buzzer gives out a sound “du”,
the power lights on the tool tray and the aid frame turn on. And the equipment carries on
the self-check.
2. check the connection of water source, make sure there isn’t any leakage.
3. check the connection of air source, make sure there isn’t any leakage.
4. open air source to examine the indicated number of air pressue meter in the air reducer
in the front cover. The normal operating air pressure of the dental unit is 0.55 MPa. If any
error occurs, please adjust the knob in the air filtrating and reducing valve to keep the
above-mentioned number.

XI. USING INSTRUCTIONS


don’t carry still
NOTE:
☞ Please make sure the water, air and power supply have been open while using the
product.
☞ All the functions set can be saved automatically in case of power off.
☞ When the power supply resumes after the power off, there isn’tany on safety issue,
except the going operation should be shutted temperartty. If to continue the operation, please
restart the unit.
1. INSTRUMENT TRAY CONTROLLING PANEL

SKETCH MAP OF KEY BOARD OF ST-D520 SKETCH MAP OF KEY BOARD OF ST-D530

connecting line
SKETCH MAP OF KEY BOARD OF ST-D540 ☞ NOTE:
☞ NOTE:

1. Position of the patient chair 2. Spittoon chair position


3. Water supple 4. Cuspidor water rushing
5. Constant temperature hot water 6. Film viewer
7. LED light 8. Memory position
9. Reset posistion 10. Set key

1.1. POSITION OF THE PATIENT CHAIR


Press each key for long time, then the patient chair shall move according to the identifiers’ direction. Release the key, then
the move stops.
——“ ”: the patient chair upward;
——“ ”: the patient chair downward;
——“ ”: backrest backward of the patient chair;
——“ ”: backrest forward of the patient chair.
1.2. SPITTOON CHAIR POSITION
Press “ ”, the backrest of the patient chair shall move shortly to the forefront position so that patient may spit and
gargle; and press the key again, the backrest will return to the original position.
NOTE:
☞ In case of pressing other function keys while the backrest of the patient chair is
moving, this key’s function will stop automatically.
1.3. WATER SUPPLE

Press “ ”, supply the gargling water until the pre-set time, during the water supply pressing the key again in the course
of water supply, the water supply shall stop. Please refer to 1.10 “set key” for the instructions of time set.
1.4. CUSPIDOR WATER RUSHING
Press “ ”, the water will rush the cuspidor for three minutes and stop automatically, in the course of rushing, press the
key again, it stops.
1.5. CONSTANT TEMPERATURE HOT WATER
Press “ ”, the constant temperature heater will start working to heart the water to 40℃, in case of water temperature
decrease, the heater will work automatically (the indicator light flashes in the course of hearting and keeps on when the
hot water reaches the pre-set value); press the key again, the heater stops immediately (the indicator light is off).
NOTE:
☞ The hot water is only for the purpose of gargling and 3-way syings. (Must connect
to the water supply first before using the water heater in order to prevent the water heater from
short circuit due to dry heating)
1.6. FILM VIEWER
Press “ ”, close the power supply of the film viewer; press the key again, turn on the power supply of the film viewer.
NOTE:
☞ There’s another power switch on the film viewer.
1.7. OPERATING LIGHT
Press “ ”, turn off the operating light, press the key again, turn on the power supply of the cold light lamp.
NOTE:
☞ There’s another power switch on the cold operating light.
1.8. MEMORY POSITION
The patient chair can define nine memory positions into three groups such as A, B and C with each group having three
chair position memory.
Press the key of each group, the indicator light turns on and enter into the memory procedures of this group, press one of
such three memory keys again, the patient chair may move to the corresponding memory position.
Please refer to 1.10 “set key” for the instructions of self-defining the chair position.
NOTE:
☞ In case of pressing other function keys while the patient chair is moving, this key’s
function will stop automatically.
1.9. RESET POSITION
Press “RESET” key, the patient chair decilnes to the lowest position and the backrest fore rakes to the foremost position.
NOTE:
☞ Press any button duuing its moving, the movement will stop.
1.10. SET KEY
A. TIME SETTING OF WATER SUPPLY
Step 1: press “SET” key, the indicator light turns on.
Step 2: press “ ” key until the water supply to the demanded amount.
Step 3: press “SET” key again, the indicator light turns off, setting finish.
NOTE:
☞ This procedure is only setting the time of water supply, if tap water changes, please
reset the time.
B. MEMORY POSITION SET OF THE PATIENT CHAIR
Step 1: press “SET” key, the indicator light turns on.
Step 2: press “ ”key or “ ” key (press one of them or neither key) to adjust the chair’s height, and release the
key until to the required position.
Step 3: press “ ” or “ ” key (press one of them or neither key) to adjust the backrest’s angle, and release the key
until to the pre-set position.
Step 4: press any key of A, B, C group, and press any key of such three memory chair positions.
Step 5: press “SET” key, the indicator light turns off, and the set is completed. The required chair position has been
memorized in the corresponding memory group keys.
2. ASSISTANT TRAY CONTROLLING PANEL
1, 2, 3, 4: Position of the patient chair (see chapter XI No. 1.1)
5: Constant temperature hot water (see chapter XI No. 1.5)
6: Water supply (see chapter XI No. 1.3)
7: Cuspidor water rushing (see chapter XI No. 1.4)

3. WATER TANK
The water supply of the water tank only be used by the appliance on the tray. In case of selecting the water supply from
the water tank, please follow the steps shown as below:
Step 1: draw back the switch “ ”, cut off the tap water supply.
Step 2: pull up the switch “ ” to “ON”, the water supply for appliances on the tool
tray has changed to the water storage tank supply.
NOTE:

☞ In case of using the tap water as the water supply for appliances on the tool tray,
please make sure the switch “ ” is pulled up.
☞ In case of air discharge when the switch “ ” is moving from “ON” to “OFF”, it’s a
normal.
4. FOOT CONTROLLER
1: position of the patient chair (see chapter XI No. 1.1)
——“UP”: “ ”;
——“DO”: “ ”;
——“BD”: ” ”; knob
——“BU”: “ ”.
2: Water supply (see chapter XI No. 1.3)
3: Cuspidor water rushing (see chapter XI No. 1.4)
4: Chip blow
5: Handpiece pedal

4.1. CHIP BLOW


Step on the key to dry the residual water in the handpiece pipe when the handpiece completes its work.
NOTE:
☞ Please adjust by yourself as to the air amount for chip blowing. (see chapter XI No.
5)
4.2. HANDPIECE PEDAL
“W” on the left pedal indicates the water supply control switch of the handpiece.
“A” on the right pedal indicates the air supply control switch of the handpiece.
The methods are shown as bellow:
——In case of stepping on the left pedal only, only the water supply.
——In case of stepping on the right pedal only, only the air supply.
——In case of stepping on both the right and the left pedals, the water supply and the gas supply work at the same time.
NOTE:
☞ Please don’t use two or three handpieces at the same time.
☞ It is a auto-lock in the handpiece holder, In case of using the handpiece, please take
out the handpiece.
5. VALVE OF HANDPIECE
See “Figure 1” for identifiers of the handpiece air adjustment, and see “Figure 2” for identifiers of the handpiece water
adjustment.

Figure 1 Figure 2
5.1. , , in “Figure 1” correspond to the handpiece water and air amount of , , in “Figure 2”. The
adjustment method: connect three handpiece, then pull out one handpiece and step on the two pedals of the pedal
controller at the same time, and then rotate the knob indication to adjust the water and air amount according to “Figure 1”
and “Figure 2”.
5.2. “Figure 1” is to adjust the handpiece’s air amount for chip blowing. Rotate the knob according to the identifiers’
indication to adjust the air.

NOTE:
☞ Please don’t pull out two or three handpieces to make adjustment.
6. OPERATING LIGHT
Identifiers of the operating light shall be made according to the right picture:
: strong light
: light off
: soft light

dental lamp identifier

NOTE:
☞ The power supply of the LED light lamp can be controlled by keys on the
instrument tray.
7. X-RAY FILM VIEW
As shown in the right picture, if move the switch to “ON”, the light turns on. If
move to “OFF”, the light turns off.

NOTE:
☞ the power supply of the film viewer can be
controlled by keys on the instrument tray.

8. 3-WAY SYRINGE AND SUCTIONS


8.1. 3-WAY SYRINGE
——Press the identifier “ ”, spray air only;
——Press the identifier “ ”, spray water only;
——Press both the “ ” and “ ”, spray mist.

8.2. SUCTIONS
Take out the strong and weak suctions from the assistant tray, they works automatically. Adjust the switch on the suctions
head to adjust the amount, as shown in the right picture.
NOTE:
☞ In order to avoid the risk of biological compatibility, please use the accord with ISO
19033 salivary absorption material
9. SYSTEM INITIALIZATION
NOTE:
☞ It can’t be initialized until the dental unit moves up to the specified location.
STEPS FOLLOW:
Step 1: Turn off the power switch of the dental unit, as the figure 1.
Step 2: Turn on the switch No.1 (see figure 2) on the red base (see figure 3) witch on the left corner of the circuit
board (see figure 4). Don't change the position of other switch.

Figure 1 figure 2 figure 3 figure 4 ​


Step 3: Turn on the dental unit, it will initialize after entering the system (a "tick" sound prompting sound). At this
time Direction key only the master, slave, foot control can be used, other features are not available. Press the key (see
figure 5) to move the dental unit to touch all 4 limit switch.
Step 4: The LCD screen will as shown in figure 6. Put down the switch No.1 as the figure 7 (Note: it is not necessary
to turn off the power switch when putting down the switch No.1). Initialization finished.

figure 5 figure 6 figure 7


XII. MAINTENANCE AND REPAIR

1. In order to make clearly for the dental unit, please disinfect its surface by medical alcohol daily.
NOTE:
☞ In case of the damage of electrical parts,please do not spray any liquid into
the equipment.
☞ Please wipe the Cleaning agent which remails on the surface of product with
cleaning cloth.
☞ When cleaning opening position of products surface, please spraying cleaning
agent to cleaning cloth before wiping.

2. In order to guarantee the using water cleanly, please clear the filter
element of water filter by mild detergent at least every 6 months (the
specification of filter core is less than 50 µm). If there is cracking or
breakage in the water filter or filter cover, please change it at once.
3. In order to guarantee the water clearly in the water tank, please clear the
water tank inside by mild detergent at least once a week. Disconnect the air
from water tank firstly before take off the water tank.

4. In order to guarantee the dry air using, please discharge the seeperthe in
the air pressure regulater which is in the junction box at lease once a week(
the filter core is less than 5 µm). In case of there is air leakage in the air
pressure regulater, please change it immediately.

5. please clear the suction filter by water 6. Inject lubricating oil regularly into the rotating parts to
every day after using the dental unit. maintain the acting parts in good status, at least once a year.

7. Maintenance of dental light shall be carried out at normal temperature, using wet soft cloth to wipe the surface.

8. Please read the handpiece user’s manual about its maintenance and disinfecting before using .

9. Only authorized dealer or customer service service personnel and other professional repair personnel can have the
product maintenance. Qualification of the replacement of Some spare parts can be seen as follow.
NO. SPARE PARTS PERSONNEL NO. SPARE PARTS PERSONNEL
1 Control system Maintainer 10 Air pressure valve Maintainer

2 Electric motor Maintainer 11 Barometer Maintainer


3 Lines Maintainer 12 Valve Maintainer
4 Main control board Maintainer 13 Water heater Maintainer
5 Foot controller Maintainer 14 Equipment casing Maintainer
6 Transformer Maintainer 15 Suction system Maintainer
7 Solenoid valve Maintainer 16 3-way syringe User
8 Dental lamp Maintainer 17 Spittoon cylinder User
9 Water filter Maintainer 18 Water tank User

XIII. GUIDE TROUBLESHOOTING

TROUBLE REASON ANALYSIS TROUBLESHOOTING


The equipment doesn’t
work 1. The power supply switch is not connected. 1. Connect the power supply
switch.
2. The fuse of the power supply switch is burned.
2. Change the fuse.
3. The fuse of the equipment is broken.
3. Change the fuse within the
4. The general switch of the equipment is not open. equipment.

4. Open the general switch.


The dental lamp doesn’t
work 1. The bulb is burned. 1. Change another bulb of the
same type.
2. The glower isn’t well connected.
2. Revise with tools.
2. Revise with tools.
3. The plug isn’t well connected.
3. Revise with tools and then
4. The fuse on the sub-terminal of the power plug in.
supply transformer is burned.
4. Change the fuse of the same
5. The internal line within the moving parts is type.
twisted off.
5. Change a new line.
The sputum rushing &
gargling water cannot be 1. The valve element of the electromagnetic valve 1. Remove and clean the vale
closed up is blocked by foreign articles. element

2. The therapy board is damaged. 2. Change the therapy board.


No water comes from the
sputum rushing & 1. The water pipe is contorted and staved. 1. Settle or change a new pipe.
gargling water supply
2. The electromagnetic valve is burned. 2. Change a new electromagnetic
valve.
3. The therapy board is damaged.
3. Change a new therapy board.
Temperature of the
gargling water is too high 1.One temperature relay within the cover of the 1. Change a new water heater.
water heater fails to work.
2. Readjust the temperature
2. Temperature on the temperature controller is too
high.

The cuspidor rushing


drainage doesn’t work 1. The pipe is blocked by filth. 1. Clean the filth.
well
2. The plastic pipe is staved. 2. Disassemble the pipe and dip
it in hot water until it returns to the
original state; or change a new pipe

No water comes out when


using the handpiece or 1. The single-joint diaphragm is fractured. 1. Change the diaphragm of the
water with air when using same type.
the handpiece 2. The handpiece control valve diaphragm is
fractured. 2. Change the diaphragm of the
same type.

XIV. CIRCUIT SKETCH MAP

XV. CIRCUIT DIAGRAM OF THE PATIENT CHAIR


XVI. WATER AND AIR PRINCIPLE
XVII. WATER AND AIR LINE SKETCH MAP

®
Suntem
FOSHAN SUNTEM MEDICAL INSTRUMENT CO., LTD.
Address: 1 Flat, No.7, C District, Sanshui Industry Park, 528137 Foshan City, Guangdong, China.
Tel: +86-757-87393709
Fax: +86-757-87393729
Website: http://www.suntem.cn
※ The company reserves the right of refining the design of the product without notice in advance.
※ The manual has been showed every function on the dental unit, and the Company reserves the right of explanation.
※ The Manual version: B. Effective date: 2013.