Sie sind auf Seite 1von 5

Moore et al.

CT Imaging • Original Resea rch


Comparison
Study of
MDCT
Radiation
Dose
Downloaded from www.ajronline.org by 68.143.142.82 on 11/14/15 from IP address 68.143.142.82. Copyright ARRS. For personal use only; all rights reserved

A C E N T U
R Y O F

Comparison of MDCT Radiation


MEDICAL IMAGING
Dose: A Phantom Study
William H. Moore1 OBJECTIVE. Recently there has been a significant increase in the use of CT imaging re-
Michael Bonvento sulting in a significant increase in radiation exposure to the population. Few studies have com-
Rosemarie Olivieri-Fitt pared the degree of radiation exposure among the currently available MDCT units. Our objec-
tive is to make such a comparison.
Moore WH, Bonvento M, Olivieri-Fitt R MATERIALS AND METHODS. Using a Rando anthropomorphic phantom, we placed
thermoluminescent dosimeters into the center, anterior, and lateral aspect of the lower chest of
the phantom. Standard CT of the chest was performed with the current protocols used at our
institution on 4- , 8- , and 16-MDCT GE Healthcare systems. Next, near-identical CT scans of
the entire chest were performed on the same CT systems.
RESULTS. The 4-detector array showed statistically significantly higher radiation dose
compared with the 16-detector array with near-identical technique (p < 0.01). There is a trend
toward decreasing radiation dose with the increasing number of detectors using both standard
and near-identical technique. An inverse relationship exists between measured radiation dose
and the number of detectors.
CONCLUSION. We theorize that as the number of detectors increases, there is a decrease
in the amount of nonutilized radiation exposure, thus resulting in a lower total radiation dose.

he use of radiologic imaging, es- date there has been no published study com-

T pecially CT, has seen a significant


increase over the past several
years [1]. This increase in usage
paring radiation dose among the different
available MDCT units. Groves et al. [15]
compared the utility of a Monte Carlo–cal-
has raised the concern about radiation expo- culated radiation dose to thermoluminescent
sure and the health risks related to increased dosimeter–measured radiation dose and
radiation exposure [2, 3]. Many studies have showed that the actual radiation dose was
evaluated radiation exposure in CT [4–8], and 18% higher than the calculated radiation
it was the topic of a consensus statement by dose with a 16-MDCT system.
the Fleischner Society, in which a suggestion The purpose of this study was to measure
Keywords: CT technique, MDCT, radiation dose, radiation of an appropriate radiation dose for CT of the the radiation dose on 4- , 8- , and 16-MDCT
exposure chest was made [9]. A study performed in with both standard and near-identical tech-
1989 found that although CT accounted for niques using thermoluminescent dosimeters
DOI:10.2214/AJR.05.1491 only 2% of all radiologic studies, CT made up placed in or on an anthropomorphic Rando
Received August 24, 2005; accepted after revision
20% of the effective radiation dose [10]. A phantom (The Phantom Laboratory).
November 12, 2005. follow-up study preformed several years later
showed that both the use of CT and CT-re- Materials and Methods
1All authors:Department of Radiology, Stony Brook lated radiation exposure had doubled [11]. Before all experiments, a total of 60 thermolu-
University, University Hospital HSC Level 4, Room 120, A standard range of radiation dose for minescent dosimeters were annealed at 400°C for
Stony Brook, NY 11794. Address correspondence to
W. H. Moore.
multiple organs has been reported in several 2 hours, then at 100°C for 1 hour. Using a male
studies [9, 12–14]. Since the advent of Rando anthropomorphic phantom, five thermolu-
WEB MDCT, several studies have evaluated the minescent dosimeters, wrapped in cellophane,
This a Web exclusive article. radiation exposure between single-detector were placed into the predrilled hole at the center of
AJR 2006; 187:W498–W502
CT and MDCT [6, 9, 10]. These studies have slice 18 of the phantom (near the center of the
shown that there is an increase in radiation heart). Additional packages of five thermolumi-
0361–803X/06/1875–W498
exposure of up to 27–36% with MDCT com- nescent dosimeters were affixed, with silk tape, to
© American Roentgen Ray Society pared with single-detector CT. However, to the anterior and lateral aspects of the phantom of

W498 AJR:187, November 2006


Comparison Study of MDCT Radiation Dose

slice 18, at the level of the heart in the lower chest Fig. 1—Axial CT image of
(Fig. 1). All thermoluminescent dosimeters were chest at section 18 of
Rando phantom (The
kept in small opaque packages to minimize the ef- Phantom Laboratory).
fects of exposure to ambient light. Packets of
The anthropometric phantom was placed on thermoluminescent
dosimeters wrapped in
the CT table and a standard departmental protocol
cellophane (arrowheads)
Downloaded from www.ajronline.org by 68.143.142.82 on 11/14/15 from IP address 68.143.142.82. Copyright ARRS. For personal use only; all rights reserved

CT of the chest was performed using SMART are placed at anterior,


mA. SMART mA is a proprietary algorithm of lateral, and center
GE Healthcare that modulates the tube current positions in this slice.
(mA) exclusively on the z-axis based on patient
habitus and an operator-selected image-quality
specification (Noise Index, GE Healthcare). Spe-
cific protocols are detailed in Table 1 and were as
follows: LightSpeed 4 (GE Healthcare): detector
array, 4 × 2.5 mm; field of view, 36 cm; table
speed, 15 mm/s; rotation time, 0.6 seconds; 120
kVp; pitch, 1.5:1; and noise index, 11. Light-
Speed 8: detector array, 8 × 1.25 mm; field of
view, 36 cm; table speed, 13.5 mm/s; rotation
time, 0.6 seconds; 120 kVp; pitch, 1.35:1; and
noise index, 10.4. LightSpeed 16: detector array,
16 × 1.25; field of view, 36 cm; table speed, 27.5
mm/s; rotation time, 0.6 seconds; 120 kVp; pitch, TABLE 1: Parameters for Standard Departmental CT Chest Protocol
1.35:1; and noise index, 11.5. All exposures were
Scanner
performed with SMART mA.
The second experiment was performed using Parameter 4-MDCT 8-MDCT 16-MDCT
near-identical techniques. Specific protocols are Detector configuration (mm) 4 × 2.5 8 × 1.25 16 × 1.25
detailed in Table 2 and were as follows: Light- Field of view (cm) 36 36 36
Speed 4: detector array, 4 × 1.25 mm; field of Table speed (mm/s) 15 13.5 27.5
view, 36 cm; table speed, 7.5 mm/s; rotation time,
Pitch 1.5:1 1.35:1 1.35:1
0.6 seconds; 120 kVp; pitch, 1.5:1; noise index,
Nominal prepatient collimation 10 10 20
13. LightSpeed 8: detector array, 8 × 1.25 mm;
field of view, 36 cm; table speed, 13.5 mm/s; rota- Actual prepatient collimation 13 13 21
tion time, 0.6 seconds; 120 kVp; pitch, 1.35:1; mA 243 251 265
noise index, 13. LightSpeed 16: detector array, Noise index 11 10.4 11.5
16 × 1.25; field of view, 36 cm; table speed, 27.5
DLP / ED 366 / 12.2 505 / 16.8 359 /12.0
mm/s; rotation time, 0.6 seconds; 120 kVp; pitch,
Note—mA refers to the mA chosen by SMART mA (GE Healthcare) at slice 18 of the phantom. DLP = dose–length
1.35:1; noise index, 11. All exposures were per- product in milliGray × cm, ED = effective dose in millisieverts.
formed with SMART mA. The noise index was
chosen based on the SD in the Rando phantom’s
heart in each CT unit. The SD was set to approxi-
mately 8 H for all three scanners. The table speed
in the 4-detector unit is the maximal speed al- TABLE 2: Parameters Used to Create the Near-Identical Chest CT Technique
lowed in this unit at 4 × 1.25 mm. Scanner
Before interpreting the thermoluminescent do-
Parameter 4-MDCT 8-MDCT 16-MDCT
simeters, a 24-hour waiting period was observed
Detector configuration (mm) 4 × 2.5 8 × 1.25 16 × 1.25
to allow some of the excited electrons to return to
a steady state. This has been shown to decrease the Field of view (cm) 36 36 36
number of outlier measurements [8]. The ther- Table speed (mm/s) 7.5 13.5 27.5
moluminescent dosimeters were then interpreted Pitch 1.5:1 1.35:1 1.35:1
using a Model 3500 thermoluminescent dosimeter
Nominal prepatient collimation 5 10 20
reader (Harshaw Chemical), and the output of the
Actual prepatient collimation 8 13 21
reader was recorded in nano-coulombs (nC). After
each set of exposures, a separate set of thermolu- mA 373 245 257
minescent dosimeters were exposed to known lev- Noise index 13 13 11
els of radiation. Calibration was performed using DLP / ED 575 / 20 522 / 17.4 432 / 14.4
an electron chamber (MDH RadCal). Three sepa- Note—mA refers to the mA chosen by SMART mA (GE Healthcare) at slice 18 of the phantom. DLP = dose–length
rate exposures were made: 11.3 mGy (94.4 nC), product in milliGray × cm, ED = effective dose in millisieverts.

AJR:187, November 2006 W499


Moore et al.

22.4 mGy (201.1 nC), and 45.0 mGy (331.5 nC). ically significant (p < 0.01). There was no lated dose–length product (DLP) and effec-
These thermoluminescent dosimeters were also significant difference in radiation dose be- tive dose (ED), we found that the measured
interpreted 24 hours after exposure, and conver- tween the 8- or 16-detector units when com- radiation dose was higher for all the CT
sion of nanocoulombs to grays was generated. paring standard departmental protocol to a units. The calculated dose was underesti-
This was used to calculate radiation dose for the near-identical technique. mated by 1–30%. This correlates with the
remaining thermoluminescent dosimeters, which When comparing the measured radiation Groves et al. [15] data, which showed that
Downloaded from www.ajronline.org by 68.143.142.82 on 11/14/15 from IP address 68.143.142.82. Copyright ARRS. For personal use only; all rights reserved

were placed in or on the Rando phantom. dose in the Rando phantom from the ther- the calculated radiation dose was approxi-
moluminescent dosimeter data to the calcu- mately 18% lower than the measured radia-
Results
All 60 thermoluminescent dosimeters Fig. 2—Radiation dose for
35

Radiation Dose (mGy)


used for this study are included in the data 4- , 8- , and 16-MDCT units
30 using standard depart-
set. The method described of waiting 24 mental protocol for CT of
25
hours before interpreting the thermolumi- chest. Bars represent
20 average radiation dose of
nescent dosimeters was highly effective in
15 five thermoluminescent
this study. None of the thermoluminescent dosimeters used at center,
dosimeters were felt to be far outside an ac- 10 anterior, and lateral posi-
ceptable range. There was a single ther- 5 tions, respectively, in slice
18 (midportion of heart) of
moluminescent dosimeter in the center of 0 Rando phantom (The
the 8-MDCT arrangement that was higher 4-MDCT 8-MDCT 16-MDCT
Phantom Laboratory).
than any of the other thermoluminescent do- CT Scanner Error bars are 2 × SD.
simeters. Exclusion of this thermolumines-
cent dosimeter did not change the data ap-
preciably, although there was an increase in
the SD in this series. Fig. 3—Trend of radiation
35 dose with standard
There was no statistically significant dif-
departmental protocol
ference by Student’s t test in radiation dose for 4- , 8- , and 16-MDCT
30
when comparing the three different scanners units. Radiation dose is
Radiation Dose (mGy)

using standard departmental protocols, recorded in mGy. Error


25 bars are 2 × SD. Each
p = 0.06–0.4 (Fig. 2). However, a trend of data point is average of
decreasing radiation dose with increasing five thermoluminescent
number of detectors was observed (Fig. 3). 20 dosimeters placed in
Further, there was no statistically significant center, anterior, and
15 lateral aspects of Rando
difference observed in radiation dose be- phantom (The Phantom
tween the center, anterior, and lateral ther- Laboratory).
moluminescent dosimeters on or in the 10
Rando phantom when using standard de-
partmental protocols or with a near-identi- 5
cal technique, p = 0.09–0.4 by Student’s t
test (Figs. 2 and 4). 0
When the CT units were all set to the 4-MDCT 8-MDCT 16-MDCT
same peak kilovoltage, noise, and collima- CT Scanner
tion, there was approximately a 47% higher
radiation dose observed with the 4-detector
unit compared with the 16-detector unit
(Fig. 4). This difference is statistically sig-
40
nificant by Student’s t test (p < 0.01). A
Radiation Dose (mGy)

similar statistical difference was observed 35


between the 4- and the 8-detector systems 30
(p < 0.01). However, no statistically signifi- 25
cant difference was observed between the 8-
20
and the 16-detector units (p = 0.26). As with
standard departmental protocols, there is a 15 Fig. 4—Radiation dose
for 4- , 8- , and 16-MDCT
trend toward decreasing radiation dose with 10
units using near-identical
an increasing number of detectors (Fig. 5). 5 protocols. Bars
We observed a 55% increase in radiation represent average of five
0 thermoluminescent
dose comparing the standard departmental 4-MDCT 8-MDCT 16-MDCT dosimeters at each site
protocol to the near-identical protocol on with error bars
CT Scanner
the 4-MDCT unit. This difference was stat- representing 2 × SD.

W500 AJR:187, November 2006


Comparison Study of MDCT Radiation Dose

Fig. 5—Trend of radiation 40 would therefore expect an increase in radia-


dose with near-identical tion dose of approximately 66% compared
technique for 4- , 8- , and
16-MDCT units. Radiation 35 with the 16 × 1.25 mm array. This is close to
dose is recorded in mGy; our observation of an average increased radi-
error bars are 2 × SD. ation of 47%. The variance of this observation
30

Radiation Dose (mGy)


Each data point is
average of five
from the calculated result could also be re-
Downloaded from www.ajronline.org by 68.143.142.82 on 11/14/15 from IP address 68.143.142.82. Copyright ARRS. For personal use only; all rights reserved

thermoluminescent lated to several factors: First, the 4 × 1.25 mm


25
dosimeters placed in configuration had a pitch of 1.5:1, whereas
center, anterior, and the 16 × 1.25 mm configuration had a pitch of
lateral aspects of Rando 20
phantom (The Phantom 1.35:1. Second, differences in the noise index
Laboratory). and thus the mA used in these images help to
15
account for the differences in the observed ra-
diation dose. Finally, the 16-detector unit is
10
not perfect; there is a 1-mm area of wasted ra-
diation. This additional wasted radiation fur-
5
ther accounts for the differences between the
calculated and observed radiation dose.
0 Given these changes in prepatient collima-
4-MDCT 8-MDCT 16-MDCT
tion, we would calculate an increased radiation
CT Scanner dose of 23% when comparing the 4 × 2.5 mm
configuration to the 4 × 1.25 mm configura-
tion. This is less than the observed 55% in-
tion exposure. The trends observed from the prepatient collimation of 13 mm. The pitch crease in radiation dose. Some of the additional
thermoluminescent dosimeter data are ech- used on the 8-detector unit was lower than that change in radiation could be explained by the
oed by the DLP and ED calculations. We see used on the 4-detector unit, which can account increasing number of tails of radiation expo-
that there is an increased amount of radia- for the slightly higher radiation dose between sure with the smaller detector configuration. In
tion dose seen with the 4-detector unit when these two units. In addition, the noise index for addition, the noise index that was used for the
compared with both the 8- and 16-detector the 4-detector unit was higher than that of the 4-detector unit was based on the 2.5-mm im-
units when near-identical technique was 8-detector unit, thus resulting in a higher mA ages. Given that noise decreases with increased
used. Also, with standard departmental pro- being used at slice 18 of the phantom with the slice thickness and that the mA chosen by
tocols, there is only a small degree of differ- 8-detector unit, which further increases the ob- SMART mA on the 1.25-mm images was
ence between the radiation doses with all served radiation dose (Table 1). higher than on the 2.5-mm images, we would
three CT units. The 16-detector unit is set up with 16 de- expect higher radiation dose with the 4 × 1.25
tectors at 1.25-mm collimation each. There is mm configuration. The 8-detector unit at 1.25
Discussion a nominal prepatient collimation of 20 mm mm has a prepatient collimator opening of 13
When comparing standard technique, with an actual prepatient collimation of 21 mm. Thus, this system has 20% less nontarget
there is not a significant difference between mm. Therefore, to generate the same CT radiation than the 4-detector array at 1.25 mm
any of the detector arrangements we tested. beam coverage, the 4- and 8-detector units when covering the same volume. In addition,
However, there is a trend toward decreasing will require 26 mm of exposure compared for the same 20-mm volume of coverage, the 4-
radiation dose with the increasing number with 21 mm with the 16-detector unit. This detector array will have eight overlapping tails
of detectors (Fig. 3). When comparing near- can explain the decreased radiation dose and of wasted or nontarget radiation, whereas the
identical techniques, there is a statistically the observed trend toward decreasing radia- 8-detector array will have four overlapping
significant decrease in the radiation dose tion dose with increasing detector configura- tails. Finally, the 16-detector array will only
from the 4- to the 8-detector units and the 4- tions. Although this is an accurate portrayal have two overlapping tails. This trend of de-
to the 16-detector arrangements. There is a of how these systems function, the actual im- creasing overlapping tails of radiation expo-
trend toward decreasing radiation dose with plementation of this concept is far more com- sure could explain some of the additionally ob-
the increasing number of detectors (Fig. 5). plicated because of the helical nature of the served decrease in radiation dose with an
When comparing standard technique, the 8- image acquisition, which results in overlap- increasing number of detectors.
detector unit has a slightly higher radiation ping beams of radiation. This study does have several limitations.
dose than the 4-detector unit. Some of the rea- When using a near-identical technique, the First, the use of thermoluminescent dosime-
sons for this finding are related to the unit’s 4-detector system was less dose-efficient. The ters introduces significant bias. These devices
configuration. The 4-detector unit is set up to reason for this difference could be related to are small and have a tendency to have spuri-
have 4 detectors at 2.5-mm collimation each. the size of the prepatient collimation. The 4- ous measurements, requiring exclusion of
There is a nominal prepatient collimation of 10 detector system has a prepatient collimator multiple thermoluminescent dosimeters in
mm with an actual prepatient collimation of 13 opening of 8.0 mm using 4 detectors at 1.25 most studies. We were able to include all ther-
mm. The 8-detector unit is set up to have 8 de- mm. The nominal size of the detector array moluminescent dosimeters in these data sets.
tectors at 1.25 mm each, also with a nominal with this configuration is 5 mm, resulting in Second, a standard dose graph was used to
prepatient collimation of 10 mm and an actual 3.0 mm of wasted or nontarget radiation. We estimate the radiation dose obtained from the

AJR:187, November 2006 W501


Moore et al.

thermoluminescent dosimeters. This tech- units, such as the currently available 64-detec- levels [in Italian]. Radiol Med (Torino) 2001;
nique can result in miscalibrations at many tor unit. If the trend of decreasing radiation 102:262–265
levels. However, many studies have used this dose continues, this may suggest that higher- 8. Giacomuzzi SM, Torbica P, Rieger M, et al. Radia-
technique to evaluate radiation dose [4–8]. detector units should be used on a larger scale tion exposure in single slice and multi-slice spiral
Although the most reliable manner to esti- in an attempt to decrease radiation dose to the CT (a phantom study). Rofo 2001; 173:643–649
mate radiation dose is an ionization chamber, public, especially in centers where a large vol- 9. Mayo JR, Aldrich J, Müller NL. Radiation exposure
Downloaded from www.ajronline.org by 68.143.142.82 on 11/14/15 from IP address 68.143.142.82. Copyright ARRS. For personal use only; all rights reserved

this is not a reasonable alternative with phan- ume of pediatric CT is performed. at chest CT: a statement of the Fleischner Society.
tom work, primarily for physical reasons. Radiology 2003; 228:15–21
Third, it is not possible to make identical ex- 10. Shirmpton PC, Jones DG, Hillier MC, et al. Survey
posures with each of the different CT units. References of CT practice in the UK. II. Dosimetric aspects.
The 4-detector unit only allows two different 1. Wittram C, Meehan MJ, Halpern EF, Shepard JA, NRPB report no R-249. Chilton, England: National
pitches, 1.5:1 and 0.75:1. These pitches are not McLoud TC, Thrall JH. Trends in thoracic radiol- Radiological Protection Board, 1991
available on the 8- and 16-detector units (8- ogy over a decade at a large academic medical cen- 11. Shrimpton PC, Edyvean S. CT scanner dosimetry.
and 16-detector experiments were performed ter. J Thorac Imaging 2004; 19:164–170 Br J Radiol 1998; 71:1–3
at 1.375:1). We used the 1.5:1 pitch in the 4-de- 2. Rogers LF. Radiation exposure in CT: why so high? 12. Lee KS, Primack SL Staples CA, et al. Chronic
tector unit, which could potentially lower the (editorial) AJR 2001; 177:277 infiltrative lung disease: comparison of diagnos-
radiation dose seen in the 4-detector unit. 3. Rogers LF. Serious business: radiation safety and tic accuracies of radiography and low-and con-
Fourth, slight differences in generator cal- radiation protection. (editorial) AJR 2001; 177:1 ventional-dose thin-section CT. Radiology
ibration could have resulted in differences in 4. Thornton FJ, Paulson EK, Yoshizumi TT, et al. Sin- 1994; 191:669–673
the overall radiation dose. Finally, although gle versus multi-detector row CT: comparison of ra- 13. United Nations Scientific Committee of the Effects
there is a trend toward decreasing radiation diation doses and dose profiles. Acad Radiol 2003; of Atomic Radiation. Source and effects of ionizing
dose, statistical significance was only shown 10:379–385 radiation: UNSCEAR 1993 report to the General
between the 4-detector unit and the 8-detector 5. Hunold P, Vogt F, Schmermund A, et al. Radiation Assembly, with scientific annexes. New York, NY:
unit and between the 4-detector unit and the exposure during cardiac CT: effective doses at United Nations, 1993
16-detector unit. There was no statistically multi-detector row CT and electron beam CT. Ra- 14. [No authors listed]. Documents of the MRPB Na-
significant difference between the 8- and 16- diology 2002; 226:145–152 tional Radiological Protection Board. III. Chilton,
detector units. Some of these limitations 6. Hamberg LM, Rhea JT, Hunter GJ, et al. Multi-de- England: National Radiological Protection Board,
could explain why there was not a statistically tector row CT: radiation dose characteristics. Radi- 1992:1–16
significant difference observed with the stan- ology 2003; 226:762–772 15. Groves AM, Owen KE, Courtney HM, et al. 16-de-
dard departmental protocol. 7. Moro L, Bolsi A, Baldi M, et al. Single-slice and tector multislice CT: dosimetry estimation by TDL
Future studies will be needed to evaluate the multi-slice computerized tomography: dosime- measurement compared with Monte Carlo simula-
dose efficiency of new, higher-detector array tery comparison with diagnostic reference dose tion. Br J Radiol 2004; 77:662–665

W502 AJR:187, November 2006

Das könnte Ihnen auch gefallen