Service Training Caleo

Caleo

service training Caleo | Fendri A | okt 2017


Caleo
Thermoregulation -basics

There are two temperature-sensors connected to the sensor-box of the Caleo. What
are the functions of these sensors
measures the core-temperature of the neonate
used for skin temperature control

measures pheripherical temperature of the neonate


normally connected to the feet
only used for monitoring purposes

service training Caleo | Fendri A | okt 2017


Caleo
Thermoregulation -basics

ThermoMonitoring
By comparing these values we can deduce cold stress, heat stress and infections.
The value between these two sensors is important.

service training Caleo | Fendri A | okt 2017


Caleo
Thermoregulation -basics

ThermoMonitoring–cold stress
A neonate gets cold stress when its not possible for him to compensate the loss of
energy by a colder ambient temperature.
To recognize this very early, you can use the thermomonitoringfunction of the
Caleo.The body tries to hold the body temperature stable at 37 °C.First the
extremities getting colder.

service training Caleo | Fendri A | okt 2017


Caleo
Thermoregulation -basics

ThermoMonitoring –heat stress

If the ambient temperature is too high, it could happen that the neonate is
overheated.

The central body temperature and the temperature of the extremities increase, but
the temperature of the extremities goes up first

service training Caleo | Fendri A | okt 2017


Caleo
Thermoregulation -basics

ThermoMonitoring –fever/ infection


During an infection the body answers with a higher body temperature -fever!
The extremities getting colder because of the vasoconstriction.
You can now recogniz
e very easily if the neonate has heat stress or an infection!.

service training Caleo | Fendri A | okt 2017


1. Canopy (2M 51 108)
2. Access ports
3. Front flap
4. Handle for transport
5. Housing
6. Drawer (2M 50 565)*
7. Pedals for height adjustment*
8. Height adjustable pillar*/Housing
support
9. Connection for water heater
(LuerLock)*
10. X-ray drawer/removable bed
11. Display
12. Control unit
13. Pillar element
14. Double wall* (2M 51 150)

service training Caleo | Fendri A | okt 2017


15. Central alarm indicator
16. Sensor unit, temperature
connections
17. Mains power connection
18. On/off switch
19. Fresh air filter flap
20. Connection for O2 for
adjustment*
21. Water container (2M 50 040)*
22. Tubing grommet (2M 50 385)
23. Side flap
24. Tubing port (2M 50 412)
25. Feeding grommet, hood (2M 51
109)
26. Trolley castor with direction lock
27. Trolley castors with breakes

service training Caleo | Fendri A | okt 2017


Caleo
Operating

Display

service training Caleo | Fendri A | okt 2017


Caleo
Operating

Kangaroomode
when the baby is removed to have direct skin contact with the mother or father
("Kangarooing")
This mode provides the user with extended monitoring functions in order to detect
overheating and overcooling of the patient even when the patient is outside the
patient capsule.

service training Caleo | Fendri A | okt 2017


Caleo
Operating

Cleaning Mode

only without patient in the Caleo

humidity measurement is in operation even when not displayed


used to sterilize the humidity system of Caleo
During or after cleaning mode, condensation may form
under the aggregate housing of the Caleo!

service training Caleo | Fendri A | okt 2017


Caleo
Operating

Configuration Mode

press and hold “menu”

service training Caleo | Fendri A | okt 2017


Caleo
Operating

press and hold „Alarm silence“ and „menu“ together

service training Caleo | Fendri A | okt 2017


Caleo
Troubleshooting

1) Humidity to low
leakage's
particles in cooking chamber
2) Humidity measurement to reference measuring device to low
silicone cap missing or not assemble correct
3) Humidity measurement to reference measuring device to high
sensor INOP caused by disinfectant exchange
4) No power failure alarm –no alarm on display
-Jumper on WT2-Controller not assembled correct (after SW-Download)
-Buzzer INOP

service training Caleo | Fendri A | okt 2017


Caleo
Troubleshooting

7) „AIR FAN INOP“ –message


magnet lost at fan wheel (Service Mode, check RPM (~1500))
magnet wrong polarized (polarization must be same direction)
8) „heater temperature sensor INOP“ –message
Pin at big black connector disconnected
9) „Air heater“ und „humidifier“ INOP (@same time)
check pins at thick, round aggregate cable
10) tilt adjustment –height adjustment –only one function is available, but on all keys
FET on PCB Actuator INOP

service training Caleo | Fendri A | okt 2017


Caleo
Troubleshooting

11) Height adjustment and tilt adjustment (w/o message on display) INOP
sometimes after battery exchange, wait 1hour
software believes 6min of operation were passed

12) Caleo switched on, display dark, air fan in operation, power failure alarm in
operation
12V fuse broken

13) Oxygen to low


21% adjusted, but concentration in Caleo was higher
O2-flow to low, must be ≥ 29L/min

service training Caleo | Fendri A | okt 2017


Caleo
Troubleshooting

14) „humidity“ und „oxygen“ to low


sealing
bed area wrong assembled
leakages canopy and side panels
old filter box
no air-intake filter assembled
15) Oxygen-Sensor INOP
assembled wrong, label covers vent hole
out of calibration date (calibration overdue)
(to start a re-calibration the last calibration date must be in a time frame of 12 months
and 10 days after last calibration was done)

service training Caleo | Fendri A | okt 2017