SERVICE MANUAL
CONTENTS
Page
1. INSTRUMENT DESCRIPTION......................................................................................................................4
1.1. GENERAL DESCRIPTION ......................................................................................................................4
1.2. MAIN PARTS IDENTIFICATION ...........................................................................................................5
1.3. PHOTOMETER PARTS IDENTIFICATION ...........................................................................................5
2. PARTS REPLACEMENT ................................................................................................................................6
2.1. PRELIMINARY INFORMATION............................................................................................................6
2.2. CUVETTE..................................................................................................................................................6
2.3. LAMP.........................................................................................................................................................6
2.4. PUMP HEAD AND PUMP MOTOR.........................................................................................................7
2.5. FILTER ......................................................................................................................................................8
2.6. FILTER WHEEL MOTOR ........................................................................................................................9
2.7. FILTER WHEEL HOME SENSOR .........................................................................................................10
2.8. FRONT BUTTON SWITCH....................................................................................................................10
2.9. PELTIER CELL .......................................................................................................................................11
2.10. TEMPERATURE SENSOR.....................................................................................................................11
2.11. SENSOR BOARD....................................................................................................................................12
2.12. DRIVERS BOARD ..................................................................................................................................12
2.13. FRONT BUTTON LED BOARD ............................................................................................................13
2.14. PC BOARD ..............................................................................................................................................13
2.15. DISPLAY AND TOUCH CONTROLLER..............................................................................................14
2.16. PRINTER .................................................................................................................................................14
2.17. POWER SUPPLY MODULE ..................................................................................................................15
2.18. INSTRUMENT CASE FIXING...............................................................................................................15
3. INSTRUMENT CALIBRATION PROCEDURE.........................................................................................16
3.1. PRELIMINARY INFORMATION..........................................................................................................16
3.2. FILTERS CALIBRATION ......................................................................................................................16
4. SOFTWARE AND FIRMWARE UPDATE PROCEDURE .......................................................................17
4.1. PRELIMINARY INFORMATION..........................................................................................................17
4.2. AUTOMATIC UPDATE PROCEDURE.................................................................................................17
4.3. MANUAL UPDATE PROCEDURE .......................................................................................................17
5. INSTRUMENT VERSION IDENTIFICATION ..........................................................................................18
1. INSTRUMENT DESCRIPTION
Sipper Unit
Aspiration Button
Printer: Prints out results when an analysis is completed and other information. It is a
Thermal printer with a standard 56 mm paper width.
Display: TFT color display with LED backlight for the User Interface. The display includes
the resistive touch screen layer.
Keyboard (optional): The instrument use the touch screen for user interaction. Therefore an external
keyboard can be connected in the rear panel if required. The keyboard must have
a PS/2 compatible connector.
Sipper unit: The sipper unit is made up of a plastic flexible tube and a rigid part. The plastic
tube is for aspirating sample/reagent mixture and cleaning solutions.
Aspiration Button: To aspirate a reagent sample, place the vessel containing the fluid under the
sipper tube. Ensure the tube is deep enough in the fluid to aspirate the necessary
amount. Push the button and the aspiration will start. The front aspiration button
includes an LED board with different color light capability to show the instrument
status.
Pag.4 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
Display
Drivers Board
PC Board
Push Button
Flow Cell
Peltier Lamp
Sipper Probe
Filter Motor
LEDs board
Pag.5 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
2. PARTS REPLACEMENT
WARNING: Any service to the instrument must be done with the instrument completely disconnected
from the main power supply. Turning off the instrument is not enough. Please remove the power cord
from the instrument before any service.
2.2. CUVETTE
2.3. LAMP
NOTE: Lamps modules for this instruments are made and calibrated by the manufacturer. Do not try to
replace the lamp only in this module. This will lead to possible reading problems.
Remove the 3 screws (A) in the top of the lamp module. Remove the lamp connector (B) from the
“Drivers Board” and remove the old lamp module. Insert a new lamp module in its place, rotate the lamp
module if needed. The position of holes are specially made to keep the correct position of the lamp
filament. Only one correct position is allowed. Fix the new lamp module with the original screws.
WARNING: Filters calibration is required before the normal use of the instrument with real samples.
Proceed with the filters calibration procedure explained later in this manual.
Pag.6 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
To replace the pump motor, remove the tubes from the hose connections and remove the pump fixing
screws (A). Remove the motor cable connector (B) from the “Drivers Board”, replace the motor with a
new one and fix it. Put the pump motor cable connector in its position (B) and then back the tubes in the
hose connections.
Pag.7 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
2.5. FILTER
To replace one or more optical filters you must remove the filters wheel cover
first (A).
A label (B) in the filter wheel identify the filter wavelength. Move the
filter that need to be replaced in front of the removing tool (C). B
Push the removing tool (C) gently until the filter is out of its place.
Replace the old filter with the new one with the same optical filter D
wavelength. NOTE: pay attention to the direction of the filter. In the filter
there is a small arrow (D). This arrow must be the direction of the light
from the source (the lamp) to the sample. The arrow must point to the
sample flow cuvette.
WARNING: Filters calibration is required before the normal use of the instrument with real samples.
Proceed with the filters calibration procedure explained later in this manual.
Pag.8 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
D
C
E
Lift up the filter wheel a little bit and rotate it by hand until the reference holes (F) are in the same
position. Push down the filter wheel and put a small fixing tool (G) in the reference hole. Insert the new
motor in its place and fix the motor axe screw (D).
F G
L H
Pag.9 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
WARNING: Filters calibration is required before the normal use of the instrument with real samples.
Proceed with the filters calibration procedure explained later in this manual.
C
B
Pag.10 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
B
C
Pag.11 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
C
B
WARNING: Filters calibration is required before the normal use of the instrument with real samples.
Proceed with the filters calibration procedure explained later in this manual.
Pag.12 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
B
C
2.14. PC BOARD
To replace the PC board you must remove all the cables fixing (A) and connectors glue. Remove the PC
board fixing screws (B) and replace the board with a new one. Pay attention to the correct position of the
power on cable connector (C).
A
C
The PC board is supplied with system software already installed. If an upgrade of the installed
preinstalled software is needed proceed with the “Software and Firmware update” procedure.
It is possible that the new PC board will not recognize the driver board at the first running. Just turn off
and then turn on again the instrument to complete the configuration of the system.
Pag.13 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
TOUCH SCREEN CALIBRATION: To start the touch screen calibration procedure press the front button
and turn on the instrument. Keep the front button pressed until the system start the calibration software.
The calibration software will display some cross symbols on different corners. Touch the cross in the
middle and keep it pressed until it change colour. Repeat for every corner. The touch screen controller
will save the new calibration.
2.16. PRINTER
To replace the thermal printer remove the connected cables (A) first. Squeeze the printer fixing springs
(B) and remove them. Replace the printer with a new one and insert back the fixing springs. Press the
fixing springs gently to get a good fixing. Connect the printer cables (A).
Pag.14 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
WARINING: When the new power supply is fixed, fix the ground connections (B) to the instrument
mechanical parts.
Pag.15 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
An instrument calibration is required every time an important part of the photometer is replaced by a new
one. It can be the Flow Cell, the Lamp, the Optical Sensor Board, and an optical filter.
NOTE: even if the filter is not successfully calibrated, it is nevertheless possible to work with the
instrument and use the other filters. In the event that certain filters are not calibrated, it is advisable to
perform flushing of the cuvette with distilled water and suitable detergents, and leave distilled water inside
the cuvette for at least an hour. Afterwards, repeat the filter calibration procedure.
Pag.16 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
The instrument includes a main “PC board” with the system and the User’s Interface software. In the
“Drivers Board” there is an embedded microprocessor with it own driving firmware. The internal
instrument software and the Drivers Board firmware can be updated at any time if needed.
The upgrade procedure require only a standard USB drive key and the encoded update file. The USB key
must be formatted with a FAT file system. The instrument update file must be the only one present in the
USB key.
There are 2 different update procedure available. Automatic and Manual methods. The manual method
can be used if the instrument system do not recognize the USB key during the startup.
Put the USB key with the update file in one of the USB connectors available in the rear panel. Turn on the
instrument. The instrument system will recognize the USB key in the back and an internal instrument
software will start the update procedure automatically. Based on the update file requirements, the
software, the firmware or both will be updated.
If, for any reason, the automatic update procedure fail or the USB key is not recognized during the
instrument system startup, remove the USB key, then put it back again in one of the USB connectors
available in the rear panel, wait at least 5-10 seconds then run the manual software update command
form the instrument software menu.
The update software will run with the update file present
in the USB key.
Pag.17 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
To recognize which instrument version you have you can refer to the picture below.
INSTRUMENT VERSION 2
INSTRUMENT VERSION 3
PELTIER FAN
DILUTER CABLE
Pag.18 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
RS232
Thermal
Power Supply Unit
Printer
Power On Switch
USB communication
Filter
Filter Pump
Wheel
Wheel Motor
Home
Motor
Sensor
Micro Switch
LCD display
with Touch Screen
Leds
Board
Pag.19 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
DILUTER
Thermal
Power Supply Unit
Printer
Power On Switch
USB communication
RS-232 communication
Filter
Filter Pump
Wheel
Wheel Motor
Home
Motor
Sensor
Lamp
Heater Peltier Lamp
Fan
Fan Cell
Micro Switch
LCD display
with Touch Screen
Leds
Board
Pag.20 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
RS232
Printer Data
Power Input
USB to Touch
Controller and LCD display
Drivers Boards
Pag.21 / 23
Minitecno – SERVICE MANUAL I.S.E. S.r.l.
7. TROUBLESHOOTING GUIDE
PROBLEM SOLUTION
Printer don’t print - Check in the setup menu, tab “other” if the
“autoprint result option is enabled”
- Check the printer connection cables
Filter calibration, show one - Repeat the filter calibration procedure, after
filter not passed washing with detergent cleaner like ipochlorite
solution 5% and flushing several times with
distilled water.
- Repeat the filter calibration with fresh distilled
water
- Change the filter.
Filter calibration show all the - Lamp can be burned- replace the lamp
filters not passed - Flow cell is dirty- Clean flow cell
- Check the aspiration volume
Aspiration volume is changed - Change the peristaltic cardrige
or don’t arrive over the flow cell - Check the calibration volume in the set up page,
section “pump”
Kinetic analysis are very high - Check the factor and the reagent and sample
or very low volume.
- Check the temperature of the flow cell in the
upper right side of the monitor during analysis
- Change the reagent
Imprecision in Replication tests - Check replication reading using the same
solution.
- Cuvette is dirty and need to clean the flow cell
with detergent or ipochlorite solution
- The flow cell can have inside a solid material.
- Change the flow cell.
Touch screen does not work - Follow the procedure for the replacement of the
properly touch controller or replacement of the whole
display.
Pag.22 / 23
I.S.E. S.r.l. Minitecno – SERVICE MANUAL
Note: The spare part list below includes all the common parts and parts that belong to a specific
instrument version. Please double check which instrument you have and order the right one to avoid any
trouble.
Pag.23 / 23