Sie sind auf Seite 1von 19

Alcohol: Breath Analysis

AW Jones, Linköping University, Linköping, Sweden


r 2016 Elsevier Ltd. All rights reserved.

Abstract

The analysis of ethanol in exhaled breath is widely used in forensic science and legal medicine as a rapid
test for overconsumption of alcohol. The police authorities use breath-alcohol instruments to control
sobriety in people suspected of driving under the influence of alcohol. The noninvasive nature of breath
sampling offers a definite advantage over drawing blood for analysis of alcohol at a forensic laboratory.
Two categories of breath-alcohol instrument are currently in use: hand-held electronic instruments that are
used to screen motorists at the roadside and more sophisticated evidential breath analyzers. Results with
the latter are accepted by the court when traffic offenders are prosecuted. This article reviews historical
development, physiological principles, and practical application of breath-alcohol analysis in forensic
science and legal medicine.

Glossary determining the concentration of the same substance


Accuracy A measure of the closeness of in the unknown specimens.
agreement between an analytical result and the Forced vital capacity (FVC) This is the maximum
known or assigned value of the quantity measured volume of air exhaled from full inspiration to forced
(e.g., concentration of analyte). The discrepancy maximum expiration. The FVC during the first
or difference between the true value and the second of exhalation is another parameter of lung
analytical result obtained is often referred to as function.
bias. Gas–liquid chromatography (GLC) An analytical
Alveolar air Alveolar air is defined as that fraction of technique widely used for separating volatile
the exhaled breath remaining after air from the dead- substances in a sample on the basis of their
space region of the lungs (mainly nose and upper solubility and partition between a gas and a
airway) has been ventilated out. For determination of liquid phase.
the concentration of alcohol in breath at least 1.5 l of Henry’s law Deals with the solubility of gases in
a prolonged exhalation should be discarded prior to liquids and states that for dilute solutions at
sampling. equilibrium and at constant temperature, the
Analytical sensitivity The ability of a method or solubility of a gas in a liquid is directly proportional to
instrument to discriminate between samples the partial pressure of the gas above the liquid.
containing different concentrations of the target Infrared (IR) A region of the electromagnetic
analyte. The slope of the analytical calibration spectrum extending from approximately 0.78 to
function (plot) is one index of the method’s 300 mm wavelength. Many evidential breath analyzers
sensitivity. incorporate IR detectors either at wavelengths of
Analytical specificity The ability of a method of B3.4 mm corresponding to the CH stretching in
analysis to determine the desired compound without ethanol molecules or B9.5 mm corresponding to the
responding or cross-reacting with interfering C–O stretching.
substances that might be present in the sample Lambert–Beer law The absorbance of a
analyzed. homogeneous sample containing an absorbing
Calibration A process or sequence of steps in which substance is directly proportional to the concentration
an analytical instrument or method is standardized of the absorbing substance.
to allow quantitative or semiquantitative Precision The closeness of agreement between
measurements. This is usually done by analysis of independent analytical measurements of the same
known strength standards prepared from a pure substance under prescribed conditions. A measure of
substance and used to establish response of the the dispersion or scatter of the repeated measurements
instrument at increasing concentrations of the target about the mean defined by the standard deviation of
analyte. The resulting calibration curve and the measured values. A method with high precision
appropriate response factors are then used for has good repeatability.

Encyclopedia of Forensic and Legal Medicine, Volume 1 doi:10.1016/B978-0-12-800034-2.00011-2 119


120 Alcohol: Breath Analysis

Introduction During his experiments on measuring alcohol in breath,


Anstie warned about the risk of sample contamination by
The smell of alcohol on breath along with the person’s alcohol in the mouth and oral mucosa from a recent drink
general appearance and behavior constituted the first, (Anstie, 1867). He wrote “Much caution is necessary
albeit nonspecific, clinical test of drunkenness (Borken- however in applying this test. It must not be tried within
stein, 1985). More specific tests for alcohol influence the first quarter hour after a dose has been taken for the
required identification of ethanol as the intoxicating mouth retains the characteristic smell even if the most
substance and quantitative analysis of the concentration moderate dose of any of the stronger smelling drinks for
present in blood, breath, or urine. Many human drink- fully this time.” Now referred to as the ‘mouth alcohol
ing studies showed that the signs and symptoms of effect’ this old observation is the reason that at least
alcohol intoxication, such as excessive talkativeness, 15–20 min must elapse after the last drink before breath-
staggering gait, slurred speech etc., were not always alcohol instruments are used for evidential purposes.
visible (Brick and Erickson, 2009). A much more sci- The first practically useful instrument designed for
entific test for overconsumption of alcohol is furnished breath-alcohol analysis was developed in the mid-1950s
by measuring the concentration of ethanol in the per- by Professor Robert Frank Borkenstein at Indiana
son’s blood or breath. The impairment of body func- University, in the USA (Lucas, 2000). This instrument
tions caused by drinking alcohol increases in direct was named the Breathalyzers and by the 1960s its use
proportion to the amounts consumed and the concen- was accepted for testing drivers and as evidence for the
trations in blood or breath, albeit with large intersubject prosecution of traffic offenders in the USA, Canada, and
variations (Mellanby, 1920; Penttila and Tenhu, 1976). Australia (Borkenstein and Smith, 1961). The person to
The first scientific studies aimed at investigating the be tested had to make a forced exhalation through a tube
disposition and fate of ethanol in the body took place in and mouthpiece so that a known volume of the breath
the mid-nineteenth century, at a time when rapid ad- was captured inside the Breathalyzer. This breath was
vances occurred in analytical chemistry. In Victorian passed through a mixture of potassium dichromate and
Britain overconsumption of alcohol and drunkenness was sulfuric acid contained in a glass ampoule. The ethanol
a major public health problem as it still is today (Porter, content of the breath sample, if any, was determined by
1985; Spring and Buss, 1977). The medical community chemical oxidation and the end-point of the reaction was
was divided as to what actually happened to alcohol after based on photometry (Dubowski, 1991). After reaction
drinking; was it a food, a poison, or a medicine? Some with alcohol, the change in the intensity of the yellow
physicians considered that all the alcohol consumed was color of the oxidizing agent was measured by application
eliminated unchanged in breath, urine, and sweat, of the Lambert–Beer law. The Breathalyzer instrument
whereas others were convinced it was broken down and was designed and calibrated to report results of the test as
utilized in the body just like ordinary foodstuffs. the estimated blood-alcohol concentration (BAC), which
This question was resolved by a British physician required assuming a population average blood/breath
Francis Edmund Anstie (1833–1874) in experiments on ratio (BBR) of alcohol of 2100:1 (Begg et al., 1964).
humans and in animals (Anstie, 1874). To investigate In the 1970s methods for measuring ethanol in both
what happened to alcohol in the body, an analytical blood and breath were improved considerably by use of
method was needed so that the concentration in bio- more specific analysis, including use of gas–liquid
logical specimens, including exhaled breath, perspir- chromatography (GLC), infrared (IR) spectrometry, and
ation, and urine could be determined. For this purpose electrochemical (EC) oxidation principles (Jones, 1996).
Anstie developed an oxidizing agent consisting of a Moreover, a lot of research was done concerning the
mixture of potassium dichromate and concentrated physiological principles of breath-alcohol analysis and
sulfuric acid, which became known as Anstie’s reagent. the factors that influence the blood-breath alcohol rela-
This same chromic acid solution, with minor modifi- tionship or ratio (Hlastala, 1998; Wright et al., 1975).
cations, was used for many decades afterwards for the Among other things, the BBR was not a constant 2010:1
analysis of ethanol in blood and other biological fluids. and instead tended to vary both among subjects and also
The reagent developed by Anstie changed color from within subjects as a function of sampling time after
red-brown to emerald green as ethanol in the sample drinking ended (Alobaidi et al., 1976). The dependence
was oxidized to acetic acid and the intensity of the color of BBR on sampling time was mainly a consequence of
change was proportional to the amount of ethanol in the arterial-venous differences in BAC and the fact that
sample analyzed. After known amounts of alcohol were breath-alcohol concentration (BrAC) runs closer to the
given to healthy volunteers, hospital patients, and ex- arterial BAC (Martin et al., 1984).
perimental animals, Anstie found that only a small The noninvasive nature of breath sampling, coupled
fraction of the amount administered was recoverable with advances in microelectronics and computers led to
unchanged in breath, sweat, and urine. He therefore the development of more compact and easier to operate
concluded that the bulk of the ingested dose of ethanol breath-alcohol analyzers. In the 1980s many countries in
was oxidized in the body and used as other nutrients. Europe began to update their existing drink-driving laws
Alcohol: Breath Analysis 121

to include provisions for using evidential breath-alcohol drunk driving. Factors that influence government policy
instruments in addition to blood-alcohol testing. Two and legislation include statistics about drunk driving
kinds of breath-alcohol testing equipment are available. and road-traffic fatalities, attitudes of the media and
First, hand-held instruments became widely used as the general public towards drinking and driving, and the
preliminary roadside screening tests for people suspected activity of temperance movements and the way courts
of driving under the influence of alcohol. The second interpret the law. Other things that impact on the
type of breath analyzer used in law enforcement is more prevalence of drunk driving include the relative cost and
sophisticated and the results are used as evidence for availability of alcoholic beverages, the opening hours of
prosecuting traffic offenders. bars and restaurants, and the resources available to the
Since the 1980s, many European countries (UK, Hol- police authorities for apprehending and prosecuting of-
land, Austria, Norway, and Sweden) have approved the fenders. In most countries statutory limits of BAC and/or
use of breath-alcohol analysis for evidential purposes and BrAC exist above which it is an offence to drive a motor
the results were accepted by the courts for prosecuting vehicle on the public roads.
offenders (Emerson et al., 1980). The legal significance of The threshold blood-alcohol limits vary from 20 to
evidential breath-alcohol test results and consequences 80 mg/100 ml in EU countries and this fourfold differ-
for those found to exceed the prescribed alcohol limit ence seems to depend more on political attitudes rather
required a rigorous program of quality assurance. This than traffic safety research. The statutory alcohol limits
included making duplicate determinations, ensuring that for driving in a particular country should balance the
the instrument is properly calibrated when every suspect dangers of drunk driving with police powers to enforce
is tested, the analysis of ambient air as a blank test as well the law and public attitudes and acceptability of routine
as providing a printed record to document the final result alcohol testing of motorists.
of the testing (Dubowski, 1994). Many epidemiological surveys verify that the risk of
The results of breath-alcohol analysis are either re- involvement in a traffic crash increases in direct pro-
ported as the concentration of ethanol per unit volume portion to the driver’s BAC or BrAC (Blomberg et al.,
of breath (mg/l, mg/100 ml, or even g/210 l) or alter- 2009). At a BAC between 20 and 50 mg/100 ml the risk
natively a conversion is made to give the equivalent of a crash is only marginally higher than at zero BAC,
concentration in blood. Making such a conversion but as BAC passes 80 mg/100 ml a significant and ex-
requires use of a calibration factor, the so-called blood- ponential increase in risk of a crash occurs. Further
breath ratio, which is subject to considerable inter- evidence for the dangers of drunk driving comes from
and intrasubject variation. Because of biological and evaluation of autopsy reports of drivers fatally injured in
physiological variations in the BBR of alcohol, for legal crashes. Statistics show that in 20–50% of victims the
purposes it was considered more rational to define BAC at autopsy was above the statutory alcohol limit
statutory BrAC limits separate from pre-existing blood- (Romano et al., 2014). The percentage of drivers killed
alcohol limits (Mason and Dubowski, 1976). This with BAC above the legal limit was appreciably higher
meant that drunk drivers could be prosecuted based on for single vehicle crashes compared with multiple vehicle
the concentration of alcohol in breath without the need crashes (Jones et al., 2009).
for additional evidence obtained from the analysis of a
blood sample. The statutory BrAC limits were derived
from the pre-existing statutory BAC by assuming a Statutory Limits of Blood-Alcohol Concentration
certain population average BBR of alcohol. Unfortu-
nately, different countries opted to use different BBRs Norway was the first country to introduce a statutory
when statutory BrAC limits were calculated, and these BAC limit for driving. It was originally set at 0.50 mg/g
BBRs ranged from 2000:1 to 2400:1 (Jones, 2011). blood (B50 mg/100 ml), but has been revised since then
This chapter deals with the physiological principles to 0.20 mg/g (B20 mg/100 ml). Sweden introduced a
and the practical application of breath-alcohol analysis punishable BAC limit of 0.80 mg/g (B80 mg/100 ml) in
when used in forensic science and legal medicine to 1941, and this was subsequently lowered to 0.50 mg/g
control sobriety of drivers and evidence for prosecution in 1956 and stands at 0.20 mg/g (B20 mg/100 ml)
of traffic offenders. The words alcohol and ethanol are today. Enhanced penalties are imposed if drivers have
used interchangeably in this account and refer to ethyl a BAC 4100 mg/100 ml and this includes having to
alcohol (CH3CH2OH), the main pharmacologically retake a driving test and being subjected to a medical
active constituent of alcoholic beverages. examination for substance abuse disorder. How to deal
with hard-core drinking drivers is a dilemma because
many still continue to drive even without a valid driving
Alcohol Traffic Legislation permit. This has led to the mandatory use of ignition
interlock devices, so that a driver has to undergo and
Different countries have their own structure and tradi- pass a breath-alcohol test before the vehicle will start
tions when it comes to dealing with the problem of (Marques et al., 2003).
122 Alcohol: Breath Analysis

A compilation of statutory BAC and BrAC limits analyzers approved and the analytical principle they
currently enforced in various countries is shown in incorporate. Manufacturers of evidential breath-alcohol
Table 1. Note the different concentration units used to analyzers are continuously updating and improving their
report BAC and BrAC and the various BBRs used to products in various ways, so the actual model currently
derive the threshold BrAC from the pre-existing BAC available might not be exactly the same as those in
limits. Also shown are the evidential breath-alcohol Table 1.

Table 1 Statutory concentration limits of alcohol in blood (BAC) and breath (BrAC) in different countries and the blood-breath ratios (BBR) of
alcohol used to derive statutory BrAC limits

Country Statutory BAC Statutory BrAC BBR Evidential breath- Analytical principle for
alcohol analyzer detection of ethanol

Austria 0.50 g/l 0.25 mg/l 2000 Alcotest 7110 IR at 9.5 mm and EC
Belgium 0.50 g/l 0.22 mg/l 2300 Ethylometer IR 9.5 mm
Alcotest 7110 IR 9.5 mm and EC
Alcotest 8510 IR 9.5 mm
France 0.50 g/l 0.25 mg/l 2000 Ethylometer IR 9.5 mm
Greece 0.50 g/l 0.25 mg/l 2000 Alcolmeter SD-400 EC oxidation
Italy 0.50 g/l 0.25 mg/l 2000 Alcotest 7110 IR (9.5 mm) and EC
Intoxilyzer 8000C IR 9.5 mm
Ethylometer IR 9.5 mm
Poland 0.20 g/l 0.10 mg/l 2000 AlcoSensor IV EC oxidation EC
Alcotest 7410 oxidation
Portugal 0.50 g/l 0.25 mg/l 2300 Alcotest 7110 IR (9.5 mm) and EC
Ireland 50 mg/100 ml 22 mg/100 ml 2300 Evidenzer IR multiple wavelengths
3.3–3.5 mm
Spain 0.50 g/l 0.25 mg/l 2000 Alcotest 7110 IR (9.5 mm) and EC
Netherlands 0.50 mg/ml 220 mg/l 2300 DataMaster IR 3.4 mm
UK 80 mg/100 ml 35 mg/100 ml 2300 Intoximeter EC-IR EC oxidation
Intoxilyzer 6000 IR multiple wavlengths
in 3.3–3.4 mm range
Australia 0.05 g/100 ml 0.25 mg/l 2100 Alcotest 7110 IR (9.5 mm) and EC
DataMaster IR 3.4 mm
Canada 80 mg/100 ml 0.08 g/210 l 2100 Intoxilyzer 5000C IR 3.3–3.5 mm
Intoxilyzer 8000C IR 9.5 mm
Intox EC/IR II EC detector (primary)
New Zealand 80 mg/100 ml 400 mg/l 2300 Seres Ethylometer IR 9.5 mm
Intoxilyzer 5000
Dräger Alcotest 9510 IR 3.4 mm
IR 9.5 mm and EC
USA 0.08 g/100 ml 0.08 g/210 l 2100 Intoxilyzer 8000 Intox IR 3.4 mm and 9.4 mm.
EC/IR DataMaster EC detector (primary)
DMC
Alcotest 7110 IR (3.4 and 9.5 mm) IR
(9.5 mm) and EC
Alcotest 9510 IR (9.5 mm) and EC
In the countries below statutory BAC limits are reported in mass/mass units (mg/g), which must be converted to units of mass/volume (mg/ml)a when
blood-breath ratios (BBR) of alcohol are calculated in the 4th column
Denmark 0.50 mg/g (0.53 mg/ml)a 0.25 mg/l 2100 Evidenzer IR absorption at
multiple wavelengths
3.3–3.4 mm
Finland 0.50 mg/g (0.53 mg/ml)a 0.22 mg/l 2400 Alcotest 7110 IR (9.5 mm) and EC
Evidenzer IR multiple wavelengths
3.3–34 mm
Germany 0.50 mg/g (0.53 mg/ml)a 0.25 mg/l 2100 Alcotest 7110 IR (9.5 mm) and EC
Norway 0.20 mg/g (0.21 mg/ml)a 0.10 mg/l 2100 Intoxilyzer 5000 N IR multiple wavelengths
Evidenzer 3.3–3.4 mm
Sweden 0.20 mg/g (0.21 mg/ml)a 0.10 mg/l 2100 Evidenzer IR multiple wavelengths
3.3–3.4 mm
a
The blood-alcohol concentrations expressed in mg/ml equals the concentration in mg/g multiplied by the density of whole blood (1.055 g/ml).
Note: The approved instruments for evidential breath testing and the analytical principle for determination of ethanol are also shown where EC ¼ electrochemical oxidation and
IR ¼ infrared spectrometry.
Alcohol: Breath Analysis 123

In the USA and UK there is a lot of discussion and Alcohol in the Body
debate about lowering the statutory blood-alcohol limit
from 80 mg/100 ml to 50 mg/100 ml (Fell and Voas, The fate of alcohol in the body is usually considered as
2014). Such a change in the law will probably not deter three separate processes: absorption, distribution, and
hard-core offenders and traffic delinquents from driving elimination depending on time elapsed after drinking
after drinking, because many of these individuals are starts. However, in reality these three processes occur
problem drinkers or alcoholics and would benefit simultaneously but at different rates depending on the
more from treatment rather than the usual punishments pattern of drinking and the time elapsed after end of
for drunk driving. However, lowering alcohol limits drinking until blood samples were taken (Norberg et al.,
draws public and media attention to the problem of 2003). Immediately after drinking, the ethanol in alco-
drunk driving, which will hopefully have a pedagogic hol beverages starts to become absorbed from the
influence on young people learning to drive. Besides stomach and intestines, whereas the rate of absorption
lowering BAC limits, the police authorities should be depends primarily on dose administered, speed of
given powers to conduct random (mandatory) breath drinking, and the efficacy of gastric emptying.
testing of drivers for alcohol influence, because signs and The absorption phase refers to the increase in BAC
symptoms of impairment are not always evident at a that takes place during and shortly after the end of
BAC of 50 mg/100 ml. drinking. The distribution phase refers to the change in
BAC that occurs as more and more alcohol is absorbed
and passes from the bloodstream into the various body
Statutory Limits of Breath-Alcohol Concentration organs and tissues. The speed of distribution is different
for different tissues depending on the ratios of blood
In most countries punishable BAC limits for driving flow to tissue mass. Accordingly, organs with a rich
existed long before the use of evidential breath-alcohol blood supply and a high ratio of flow to mass (e.g., brain
instruments was considered feasible. An exception and kidney) reach equilibrium with alcohol in the ar-
was the USA because of constitutional issues about un- terial blood more rapidly, whereas bulky skeletal mus-
reasonable search and seizure and self-incrimination it cles, which have a much smaller ratio of blood flow to
was more problematic to obtain blood samples for an- tissue mass, require up to 60 min. The elimination phase
alysis in criminal cases. When the first drink-driving refers to the linearly decreasing BAC mainly reflecting
laws were enforced the suspects were expected to pro- metabolism of ethanol in the liver and excretion in
vide samples of breath or urine for analysis of alcohol breath and urine.
concentration.
When in the 1980s European nations began to use
evidential breath-alcohol instruments, it was considered Absorption Phase
more scientifically sound to establish a threshold BrAC
instead of converting the result into a BAC. Unfortu- After drinking, the alcohol contained in alcoholic bev-
nately, different countries opted to use different BBRs erages starts to become absorbed through the gastric
to arrive at statutory BrAC limits ranging from 2000:1 mucosa although the bulk of the dose ingested is ab-
to 2400:1 as indicated by the information in Table 1. sorbed in the duodenal and jejunal mucosa, owing to the
The UK Road Transport Act of 1981 stipulated much larger absorption surface area available. Factors
35 mg/100 ml as the punishable BrAC limit and that delay or accelerate gastric emptying play a major
this was derived from the pre-existing BAC limit role for the absorption of ethanol. This impacts on both
of 80 mg/100 ml. This assumed a BBR of 2300:1 the time of reaching a peak concentration in blood
(80/35  1000¼2286:1, which rounds up to 2300:1). and the maximum concentration reached (Jones and
The Republic of Ireland, the Netherlands, and Belgium Jonsson, 1994). Many factors influence gastric emptying
also adopted a 2300:1 BBR when setting their respective including splanchnic blood flow and importantly pres-
BrAC limits. Other European nations such as France and ence and amount of food in the stomach, the alcoholic
Spain approved a BBR of 2000:1. beverage consumed, smoking cigarettes, and various
The situation in Germany and the Nordic countries medications.
is more complicated because the statutory BAC limit The time elapsed after end of drinking before reach-
is reported in mass/mass concentration units (e.g., ing a peak concentration in blood usually ranges from 5
0.50 mg/g or 0.50 g/kg), which is the same as 0.527 mg/ to 120 min (Mitchell et al., 2014). On reaching the
ml (density of blood¼ 1.055 g/ma). Dividing 0.527 mg/ portal venous blood the alcohol gets transported to the
ma by 2100 (BAC/BrAC ratio) gives a threshold liver, then on to the heart through the hepatic vein, via
BrAC of 0.25 mg/l ((0.50/2100)  1000¼ 0.25 mg/l). the heart to the lungs, and then throughout the entire
The statutory BAC and BrAC limits that operate in body and back to the heart with the deoxygenated
Germany and the Nordic countries are also shown in venous blood. There is some evidence that a small
Table 1. amount of ingested alcohol gets metabolized in the gut
124 Alcohol: Breath Analysis

mucosa or the liver before reaching the systemic circu-

Blood alcohol concentration (mg/100 ml)


140
lation (Baraona et al., 1994). This phenomenon is Arterial blood
referred to as first-pass metabolism (FPM) and can be 120 Venous blood
avoided by giving alcohol intravenously. The extent of
100
FPM is also negligible when alcohol is consumed as a
bolus dose on an empty stomach, because ethanol can be 80
used as a marker to determine total body water in the
same way as radio- or stable isotopes of water are used 60
(Endres and Gruner, 1994).
40

20
Distribution Phase
0
From the hepatic vein alcohol reaches the inferior vena 0 100 200 300 400 500
cava and mixes with systemic venous blood from the Time (min)
superior vena cava and the coronary sinus in the right Figure 1 Concentration–time profiles of ethanol in venous blood
atrium. The blood is then pumped from the right ven- (cubital vein) and arterial blood (radial artery) in one male subject
tricle through the pulmonary circulation and back to the who consumed 0.80 g ethanol per kg body weight on an empty
stomach.
left ventricle. The left ventricle of the heart supplies
all organs and tissues with oxygenated arterial blood.
Alcohol distributes into the water compartment of the in the cytosol fraction of the hepatocytes (liver cells)
body, which corresponds to 50–60% of body weight, and which converts ethanol into acetaldehyde. This
and there is no evidence that ethanol binds to plasma toxic metabolite is then oxidized further into acetic
proteins or concentrates in any body organs or tissues. acid by the enzyme aldehyde dehydrogenase (ALDH),
The speed of distribution of alcohol into body fluids which is located in mitochondria. The acetate produced
and tissues depends on the ratio of blood flow to tissue from oxidation of acetaldehyde enters the citric acid
mass. Organs and tissues with a rich supply of blood cycle as acetyl-CoA and is converted in peripheral
equilibrate rapidly with arterial blood concentrations of tissues into the end products carbon dioxide and
alcohol, whereas equilibration over the capillary beds in water. After drinking ethanol labeled with 14C, such as
14
the bulky muscle tissue takes a longer time. The con- CH2CH2OH, then 14CO2 is almost immediately
centration of alcohol in arterial blood is therefore higher afterwards measurable in breath.
than in venous blood returning to the heart during the The rate of ethanol metabolism depends on many
absorption phase of the blood-alcohol curve (Jones factors, including gender, age, ethnicity, circadian
et al., 2004). rhythm, liver blood flow, food intake, and habituation
The arterial (A) blood contains a higher concen- so that metabolic tolerance develops. After consumption
tration than the venous (V) blood for about the first of a moderate dose of alcohol and when BAC exceeds
60–90 min after drinking, as can be seen from the con- 20 mg/100 ml the pharmacokinetics of ethanol is a zero-
centration–time plot in Figure 1. When the distribution order process, meaning that a constant amount is elim-
phase ends, the concentrations of ethanol are the same in inated from the body per unit of time, independent of the
A and V blood but only for an instant. At all later times concentration present in blood. A good average rate of
the concentration of alcohol in the venous return from ethanol elimination from blood is 15 mg/100 ml per h,
skeletal muscles is slightly higher than the arterial blood ranging among individuals from 10 to 25 mg/100 ml per h
concentration as ethanol continues to be cleared from (Jones, 2010).
the central compartment in the liver. This temporal
variation in arterial and venous BAC has relevance for
the relationship between BrAC and venous BAC. Classification of Breath-Alcohol Instruments

The many instruments available for breath-alcohol an-


Elimination phase alysis can be classified in various ways, such as by their
size, purchase cost, method of ethanol analysis, or the
Most of the alcohol consumed (95–98%) is eliminated intended area of application. A useful method of clas-
from the body by metabolism which occurs primarily in sification is according to the area of application, whether
the liver by the action of enzymes. The remaining frac- for self-testing or as a preliminary roadside screening test
tion of the dose (2–5%) is excreted unchanged in the for alcohol or as a quantitative evidential analyzer when
kidneys, the lungs, and a small amount through the skin results are used as evidence for prosecution of traffic
in sweat (Drummer, 2014). The principal oxidative en- offenders. Hand-held screening instruments are used
zyme is alcohol dehydrogenase (ADH), which is located by police officers to test drivers on the street or at the
Alcohol: Breath Analysis 125

roadside or even as they are sitting behind the wheel of specificity, it is always advisable to verify these claims by
their vehicles. By contrast, tests with evidential breath independent testing. Such testing should involve both
analyzers are done at police stations and the procedure in vitro conditions using known concentrations of
involves many more quality control requirements com- ethanol in air and also in vivo testing with drinking
pared with roadside testing. subjects. The in vitro testing is usually done using a
Table 2 presents a historical time line showing the breath-alcohol simulator by passing a stream of air
development of handheld breath analyzers used for through a known strength ethanol solution kept at a
screening motorists at the roadside. The latest gener- constant temperature of 34 1C. More commonly today
ation of breath-alcohol screening instruments incorpor- dry-gas ethanol standards in pressurized tanks are used
ates an electrochemical sensor for analysis of ethanol for the in vitro performance tests. The final stage of
in breath. The instruments available have options for testing is done with human subjects, both male and fe-
digital display of results, which provides a pass, warn, male of various ages (20–60 years), who consumed
and fail indication. The sampling of breath is done known amounts of ethanol under laboratory conditions
automatically as the test subject exhales into a dis- to reach a BAC above the legal alcohol limits for driving.
posable mouth piece fitted to the instrument and makes When evaluating results of the controlled drinking
a prolonged exhalation. Provided blow time and pres- studies, it is important to distinguish the precision and
sure restraints are met, a sample is captured auto- accuracy of measuring alcohol in breath from the pre-
matically. However, the hand-held screening units cision and accuracy of estimating BAC. The latter re-
cannot detect presence of mouth alcohol and there is quires use of a calibration factor so that breath test
currently no requirement to make a calibration control results are reported as the presumed BAC. The closeness
check with every subject tested. The results from 50 up of agreement (accuracy) of the breath analyzer results
to 100 consecutive tests can be stored in an internal and the actual BAC depends on the value of the BBR
memory of the analyzer and downloaded later along used for calibration of the instrument.
with date and time of the test. Alternatively, results can Besides the roadside screening tests discussed above,
be printed out directly on-site. another category of breath-alcohol instruments is used
Although manufacturers of breath-alcohol analyzers that are accepted by the courts as evidence when traffic
claim certain levels of accuracy and precision and offenders are prosecuted. These instruments are highly

Table 2 Historical developments in breath-alcohol instruments used as preliminary roadside tests and the analytical principle incorporated for
determination of ethanol

First appearancea Instrument/device (country of origin) Brief details of the analytical principle used for
measuring alcohol

1953 Alcotest tube and bag (Germany) Chemical oxidation with K2CrO4 þ H2SO4 adsorbed on
silica gel
1956 Kitagawa chemical detector tube (Japan) Oxidation with chromic acid
1969 Sober-meter chemical detector tube (USA) Oxidation with K2CrO4 þ H2SO4 adsorption on silica
gel for later verification analysis
1969 Alcolyser tubes (Wales, UK) Oxidation with K2CrO4 þ H2SO4 adsorbed on silica gel
carrier
1974 Alcolmeter (Wales, UK)b Electrochemical oxidation
1975 AlcoSensor (USA)c Electrochemical oxidation
1975 ALERT (Canada) Taguchi cell, solid state stannic oxide semiconductor
detector
1978 Alcytron (Germany) Infrared absorption at 3.3–3.4 mm
1980 Alcotest (Germany) Infrared absorption at 3.3–3.4 mm
1988 Alcotest 7410 (Germany) Electrochemical oxidation
Breathalyzer 7410 (USA)
1995 Alcomat S (Germany) Electrochemical oxidation
1995 Alcoodose 2 (France) Infrared absorption at 9.5 μm
2000 Lifeloc (USA) Electrochemical oxidation
2000 Alcotest 6510 and 6810 (Germany) Electrochemical oxidation
2000 Intoxilyzer S-D5 (USA) Electrochemical oxidation
2000  present A large number of hand-held instruments intended Solid state stannic oxide semiconductor (Taguchi cell)
primarily for self-testing by the general public are
available
a
The dates shown are approximate and represent the time when the first published report appeared.
b
The electrochemical sensor incorporated in the Alcolmeter device was also used in the first model of the AlcoSensor marketed in USA.
c
Over the years different versions of the AlcoSensor instrument have appeared (model IV, FST, VXL etc.).
126 Alcohol: Breath Analysis

sophisticated and incorporate slope detectors to ensure a Infrared Breath-Alcohol Analyzers


proper sample for analysis, more specific identification
of ethanol, calibration control with every subject test, Most of the breath-alcohol analyzers currently used for
and a printed record of the results are obtained. Table 3 evidential purposes use infrared (IR) spectrometry as the
gives details of the historical breath-alcohol instruments analytical principle for quantitative analysis and identi-
for evidential testing. fication of ethanol (Harding and Zettl, 2008). Infrared
When the first breath-alcohol instruments appeared analysis is a nondestructive technology and calibration
in the USA in the 1950s (e.g., the Breathalyzer) statutory of the instruments shows good long-term stability. The
limits of alcohol in blood for driving in most states was theory and practice of infrared spectrometry is well es-
150 mg/100 ml. The results of the breath-alcohol test tablished and is widely used in analytical chemistry for
were mainly used as supportive evidence that a driver identifying functional groups in organic compounds.
was under the influence of alcohol. The major thrust of Infrared spectrometry applied to breath-alcohol analysis
any prosecution case was the circumstances surrounding monitors the absorption of selected wavelengths of
the driving incident and the results of clinical tests of infrared radiation after passage through a known vol-
drunkenness. Suspects were examined by a police sur- ume of the breath sample. The higher the concentration
geon or a forensic physician and questions asked about of ethanol in the sample the more infrared energy is
drinking and drug habits and simple cognitive and absorbed and the lower is the percent transmittance.
psychomotor tests were performed. The demands for When organic compounds absorb IR energy at cer-
high accuracy and precision of the breath-test results tain wavelengths, the vibration and rotational motion of
were less stringent and were used as supporting evi- chemical bonds in the molecule become amplified and
dence. The introduction of a concentration per se law more intense. Alcohols have characteristic IR absorp-
meant that a person could be prosecuted based on the tions associated with O–H, C–O, and C–H bond
result of alcohol analysis in blood or breath and without stretching vibrations. The C–H bond vibrations are not
the need for any clinical evidence of drunkenness. The unique for alcohols, because many other simple organic
enactment of the so-called per se laws shifted the focus molecules, such as hydrocarbons and acetone, also have
toward quality assurance and reliability of the chemical C–H bonds that can absorb IR radiation at the 3.3–
analysis of ethanol in samples of blood or breath and 3.5 mm wavelength. Nevertheless, using several narrow
away from clinical evidence of impairment of the driver. band width optical filters the specificity for detecting just

Table 3 Historical developments since the 1950s in breath-alcohol instruments used for evidential purposes and the analytical principle
incorporated for determination of ethanol

Time period Breath-alcohol instrument Brief details of the analytical principle for measuring ethanol

1953  1970 Breathalyzer A solution of K2CrO4 þ sulfuric acid is contained in a glass ampoule through which a
Photo-Electric Intoximeter known volume of breath passes. Any ethanol present is oxidized and the end-point
Ethanographe of the reaction is determined by photometry
1969 GC Intoximeter Gas chromatography with a flame ionization detector (FID) for the quantitative
analysis of ethanol and retention time for qualitative analysis
1971 Alco-Analyzer Gas chromatography with thermal conductivity detector for quantitative analysis and
retention time for qualitative analysis
1971 Omicron Intoxilyzer The first breath analyzer based on infrared absorption at a single wavelength of
3.39 mm
1973 ALERT (screening test) Metal oxide semiconductor Taguchi (T-cell)
1975 Intoximeter 3000 Single wavelength IR (3.4 mm) for ethanol and a Taguchi T-cell to detect interfering
substances
1979 Alcomat Absorption of IR energy at wavelengths of 3.4 mm (C–H stretching) or 9.5 mm (C–O
Alcotest stretching)
1976 Intoxilyzer 5000 Dual wavelength IR analyzer (3.39 mm and 3.48 mm)
1979 BAC DataMaster Dual wavelength IR analyzer (3.37 mm and 3.44 mm)
1986 Alcotest 7110 IR analyzer operating at a single wavelength of 9.5 mm
Alcomat
1992 Intoxilyzer 6000 Multi-wavelength IR analyzer in the 3.3–3.5 mm region of the infrared spectrum
1994 Intoximeter EC/IR Electrochemical fuel cell detector for analysis of alcohol and IR detector for
monitoring the CO2 profile during exhalation
1995 Alcotest 7110 Mark III Both IR spectrometry and electrochemical (fuel cell) detector
2000–present DataMaster DMT Multiple wavelength IR absorption at 3.34, 3.37, and 3.50 mm
2000–present Alcotest 9510 Combined IR (9.5 mm) and electrochemical (fuel cell) detector for alcohol detection
2000–present Evidenzer Multiple (5) wavelengths in the 3.5–3.9 mm region of the IR spectrum
2000–present Intoxilyzer 9000 IR detection at wavelengths of 3.4 mm and 9.5 mm
Alcohol: Breath Analysis 127

Wavelength (µm)
2.5 3.5 4.5 5.5 6.5 7.5 8.5 9.5 10.5 11.5 12.5
100
Transmittance (%)

H H
H C C OH

C–H stretch H H C–O stretching


3.4 µm 9.5 µm
0
Figure 2 Infrared absorption spectrum of ethanol molecules in the gas phase showing the parts of the spectrum that correspond to vibrations of
the C–H bond (3.4 mm) and C–O bond (9.5 mm).

ethanol is improved considerably. Several infrared response is directly proportional to the concentration of
breath analyzers incorporate between three and five fil- ethanol in the sample of breath or vapor analyzed.
ters that absorb IR light at fixed wavelengths of 3.32, The electrode potential of the fuel cell is selected so
3.40, and 3.49 mm and a reference wavelength common that neither water vapor nor oxygen in the breath reacts
to several potential interfering substances (Jones and to any measurable amount. The breath from healthy
Andersson, 2008). A typical vapor phase infrared ab- individuals does not contain any volatiles besides etha-
sorption spectrum for ethanol is shown in Figure 2 with nol after a person drinks alcoholic beverages that are
frequency shown both as wave number (cm1) and in oxidized by the electrochemical cell. Acetone might be at
micrometers (mm). elevated concentrations in the breath of diabetics or after
Ethanol molecules absorb IR light at wave number a prolonged fast, although this ketone is not oxidized at
3391 cm1 (2.95 mm) and this corresponds to the O–H the electrode potential of the fuel cell. If other volatile
stretching vibrations, whereas the C–O stretching alcohols existed in blood, such as methanol or iso-
is prominent at 1102 and 1055 cm1 (9.1–9.5 mm). To propanol, from drinking denatured alcohol then they
enhance the analytical selectivity some infrared breath theoretically increase the response attributed to ethanol.
analyzers are designed with optical filters that absorb Even acetaldehyde, the main metabolite of ethanol, is a
infrared energy at 3.4 and 9.5 mm wavelengths, corres- potential interfering substance. However, the concen-
ponding to C–H stretch and C–O stretch. trations of other volatile substances in blood and breath
after drinking alcoholic beverages, including acetalde-
hyde, are far too low to constitute an interference
Electrochemical (Fuel Cell) Instruments problem.

The other major analytical technology used to measure


ethanol in breath for forensic purposes is electro- Lung physiology and Gas Exchange
chemical oxidation. These devices incorporate a fuel cell
consisting of two platinum conductor electrodes separ- The lung is an amazingly efficient body organ for the
ated by an electrolyte, such as phosphoric acid, that purpose of gas exchange. The trachea connects the
conducts ions between anode and cathode parts of the throat (glottis) to the lungs and subdivides into the left
cell. The ethanol contained in a known volume of breath and right bronchi forming the left and right lobes of the
(e.g., 0.5–1 ml) enters the chamber of the cell and acts as lung. Each of the two bronchi undergoes further sub-
the combustible fuel, whereas atmospheric oxygen division about 23 times until the alveoli air sacks
functions as the oxidant. In a two-stage oxidation re- are reached. The inspired air comes into contact with the
action ethanol is converted to acetaldehyde and the pulmonary capillary blood in the small air sacs on the
latter is further oxidized to acetic acid, thus four elec- terminal ends of the respiratory bronchioles. Each lung
trons are released and the flow of current is directly contains about 300 million alveoli and the thickness
proportional to the concentration of ethanol in the of the alveolar-capillary membrane is only 0.001–
breath sample. After amplification the response from the 0.02 mm. Because of the large number of alveoli, the
fuel cell can be followed on a chart recorder (Figure 3) surface area for gas exchange is about 70 m2, approxi-
or displayed as a digital or analog signal. The peak mately the size of half a tennis court. Accordingly,
128 Alcohol: Breath Analysis

1.2

Peak response
1.0

Fuel cell response


0.8 Electrochemical oxidation
CH3CH2OH —> CH3CHO —> CH3COOH
0.6

0.4

0.2

0.0
0 20 40 60 80 100
Time after breath sampling (s)
Figure 3 Response of an electrochemical fuel cell detector to ethanol vapor (breath) showing peak response at 10–15 s after sampling and
recovery to baseline in 90–100 s.

alcohol or any other volatiles in the pulmonary blood Table 4 Anatomical and physiological characteristics of the
make intimate contact with the air in the lungs and human lung important for respiratory gas exchange
diffusion occurs across the alveolar-capillary membrane
Pulmonary gas exchange Average values in healthy
interface. The basic anatomical and physiological fea- adults
tures of the human lung can be gleaned from infor-
mation given in Table 4. Pulmonary blood flow 5–6 l per min
During breathing, air enters through the nose and Alveolar ventilation 6–7 l per min
Breathing rate 12–14/min
mouth and passes down the trachea into the two main
Tidal volume at rest B500 ml
bronchi and after multiple branches to the respiratory Number of alveoli B300–500 million
bronchioles and alveoli (air sacs) where most of the gas Forced vital capacitya Men 4.5 l average
exchange occurs as depicted in Figure 4. Women 3.2 l average
During inspiration, the ambient air is warned to body Total lung capacity Men 6.0 l
temperature (37 1C) and humidified, actually saturated Women 4.2 l
with water vapor, as it passes over the watery mucosa Anatomical dead-space volume 150 ml
Residual volume Men 1.2 l
lining the upper airways. During exhalation, these pro-
Women 1.1 l
cesses occur in reverse and now the alveolar air, which is Ventilation-perfusion ratiob 0.8 on average
saturated with water vapor at body temperature, is Volume of blood in alveolar 60–140 ml
cooled as it passes over the airway mucosa and water capillaries at one time
vapor condenses at the lower temperature in the upper Thickness of alveolar-capillary 0.001–0.002 mm
airways. The high solubility of ethanol in water means membrane
that this volatile substance also undergoes exchanges Surface area for gas exchange B50–100 m2
End-expiratory breath temperature 34.5 1C on average
between inspired and expired air and the mucosa. After
Blood/air partition ratio for ethanol at 1800:1c
drinking alcohol, the watery mucosa will contain some 37 1C
ethanol provided by the blood circulation to the
a
airways. Depends on the person’s age, gender, body size, posture, and physical condition
(smoking).
The concentration of ethanol in exhaled breath is b
Differs for different parts of the lung.
lower than the concentration in the alveolar air, because c
Determined by experiments in vitro.
during a prolonged exhalation the air cools. The core
body temperature in healthy individuals is at 37 1C
whereas breath leaves the mouth at 34.5 1C. Further- equilibration depends on several factors including
more, during exhalation the content of the alveolar air breathing pattern of the subject prior to expiration.
equilibrates with the watery mucosa membranes that The concentration of ethanol in the exhaled breath is
cover the upper airways. This interaction between dependent on a number of physiological variables that
ethanol and the mucosa and the completeness of the differ among individuals, such as their age, height,
Alcohol: Breath Analysis 129

Trachea

Bronchi

Bronchiole

Alveoli

Figure 4 Schematic illustration of the human lung, showing trachea, left and right bronchi, and subdivision into the bronchioles before reaching
the alveoli sacs where gas exchange occurs. Also shown (inset) is the rich network of blood capillaries covering the alveoli air sacs.

gender, and lung function. Nevertheless, many direct The assumption of a constant BBR was difficult to
comparisons show a high correlation between the con- defend because many drinking studies showed that the
centration of ethanol in venous blood and in the end- ratio varied both among and within individuals (Jones,
expired breath. 1978). The BBR tended to be less than 2100:1 during the
The scientific basis supporting the use of breath-al- absorption phase of the blood-alcohol curve and greater
cohol analysis as a way to estimate the concentration in than 2100 in the post-absorptive elimination phase (Jaffe
blood depends on Henry’s law, which was named after a et al., 2013). In tests made 15–30 min after end of drinking
British chemist William Henry (1774–1836) who in- the BBR was closer to 1800:1 and then rose to 2100:1
vestigated the solubility of gases. Henry’s law can be by about 90 min post-drinking. As time after drinking
stated in several ways, such as “When a solution of gas increased, the BBR increased further reaching 2300:1 or
or somewhat volatile solvent in water is brought into 2400:1 by 120 min and at later times to ratios greater than
equilibrium with air in a closed container, there is a fixed 3000:1 where possible as BAC decreased toward zero.
relationship or ratio between the concentration of the Investigations of the BBR of alcohol in drinking
substance in the air phase and in the water phase at subjects and in apprehended drivers using modern in-
constant temperature and pressure.” strumental methods of analysis suggest that the mean
The above relationship holds for dilute solutions of ratio is 2400:1 instead of 2100:1 (Stowell et al., 2008). If
ethanol in water and other media and the ratio between a person is tested on a breath-alcohol instrument cali-
the concentrations in the liquid and air phases at equi- brated on the basis of 2100:1, whereas in reality the
librium is often referred to as the Ostwald solubility actual BBR is 2400:1, this introduces a 14% bias (Jones
coefficient. However, the sampling and analysis of and Andersson, 1996). The venous BAC is accordingly
breath is a dynamic process and extensive exchanges of underestimated by this amount if an instrument is cali-
ethanol and water vapor occur with the mucous mem- brated on the basis of a 2100:1 ratio.
branes during inhalation and exhalation, which means To avoid legal arguments about physiological vari-
that the principle of Henry’s law is less applicable to the ations in the BBR, when evidential breath analyzers were
analysis of alcohol in breath. approved for use in European countries in the 1980s, the
results were reported directly as BrAC without any as-
sumption about a BBR (Jones, 2011). The notion of re-
Blood–Breath Ratios of Alcohol porting BrACs separately from BAC was originally
introduced in the USA (Mason and Dubowski, 1976).
The Breathalyzer instrument, which was developed in the Their statutory blood-alcohol limit for driving was
mid-1950s, was calibrated in such a way that the results 0.08 g/100 ml and the corresponding breath-alcohol limit
were reported as the concentration of alcohol present in was 0.08 g/210 l breath, the ratio of the two (BAC/BrAC)
blood. This conversion required use of a calibration fac- being 2100:1, although in any individual case there was
tor, the so-called blood/breath ratio, which was assumed no conversion made from breath to blood alcohol.
to be 2100:1 (Begg et al., 1964). During the 1950s– The relationship between BrAC and venous BAC
1070s, the analysis of alcohol in the breath was used as a is shown in Figure 5 for one subject who drank a
surrogate for sampling blood for the analysis of ethanol. moderate dose of alcohol (0.68 g/kg) on an empty
130 Alcohol: Breath Analysis

3000 120
Ratio BAC/BrAC

Alcohol concentration (mg/100 ml)


100 Venous blood
Breath × 2100
Ratio BAC/BrAC

2000 80

60

40
1000
0 60 120 180 240 300 360 420 480
20
150
0
BAC or BrAC (mg/100 ml)

BAC
0 60 120 180 240 300 360 420 480
BrAC × 2100
100 Time from start of drinking (min)
Figure 6 The concentration  time profile of ethanol in venous blood
and breath in one male subject who drank 0.68 g ethanol per kg body
50 weight on an empty stomach. The concentrations of ethanol in breath
were multiplied by a factor of 2100:1 (blood-breath ratio) to allow
direct comparison with blood-alcohol concentrations.
0
0 60 120 180 240 300 360 420 480 gradually decreased to decreasing towards zero by
Time after start of drinking (min) 90 min after end of drinking. At all later times the
concentration of alcohol in venous blood was slightly
Figure 5 The concentration  time profiles of alcohol in venous
blood (BAC) and end-expired breath (BrAC  2100) in one male higher than in the arterial blood, because ethanol is
subject who drank 0.68 g ethanol per kg body weight on an empty cleared by metabolism in the central (liver) compartment
stomach. The upper plot shows how the BAC/BrAC ratios of ethanol and not in peripheral tissues (Jones et al., 2004).
change as a function of sampling time after drinking.

stomach. The upper plot shows changes in the venous Mouth Alcohol Effect
BBR of alcohol in the relation to time after drinking.
The concentration of alcohol in beer (3–5 g/100 ml),
wine (8–12 g/100 ml), or spirits (35–45 g/100 ml) is
Role of Arterial-Venous differences considerably higher than the concentration of ethanol in
blood or exhaled breath. If a breath sample is taken for
During absorption into the blood, the concentration of analysis shortly after drinking an alcoholic beverage the
ethanol is highest in the arterial blood (A-BAC) com- concentration of alcohol is higher than in the alveolar air
pared with venous blood (V-BAC) returning to the heart because of contamination with the much higher alcohol
after circulation throughout the peripheral tissues. The content in the mouth and oral mucosa. The problem
concentration of alcohol in pulmonary capillary blood posed by this mouth alcohol (MA) effect was recognized
equilibrates instantaneously with the alveolar air so very early in the history of breath-alcohol testing
BrAC runs closer to the A-BAC rather than the V-BAC. (Bogen, 1927). Studies have shown that it takes about
However, it is the V-BAC that is always used in forensic 10–15 min before the higher concentration of ethanol
practice when drunk drivers are prosecuted. in the oral mucosa dissipates and this has led to a
The concentration-time profiles of ethanol in venous mandatory waiting period before making an evidential
blood and end-expired breath are compared in Figure 6 breath-alcohol test.
for one subject who drank a moderate dose of ethanol. A mouth alcohol effect might also occur if a person
The higher BrAC compared with venous BAC early after burps or regurgitates stomach contents when alcohol is
end of drinking and lower BrAC in the post-absorptive still being absorbed from the stomach and intestines, a
phase are remarkably similar to the arterial BAC and process that might last up to 90 min after drinking ends.
venous BAC relationships shown in Figure 1. Furthermore, use of oral hygiene products, such as
The existence of arterial-venous differences in ethanol mouth washes, medicines, and certain foodstuffs that
concentration and how these change in relation to time contain alcohol can artificially increase BrAC if testing is
after drinking represents a physiological explanation for done within 15–20 min afterwards (Sterling, 2012).
variations in the venous blood to breath-alcohol ratio Examples from the latest generation of infrared breath-
(Lindberg et al., 2007). The difference between A-BAC alcohol instruments are fitted with slope detectors,
and V-BAC was greatest early after end of drinking then such that the exhaled ethanol profile is continuously
Alcohol: Breath Analysis 131

monitored. If the shape of this exhalation profile in the exhalation profile following by a decrease in
deviates significantly from that expected without the concentration on reaching an end exhalation. Another
confounding influence of mouth alcohol, the test can algorithm looks at the waviness of the BrAC profile
be aborted and the instrument should indicate in the or the change in BrAC after 1 s and 5 s of exhalation.
display and on the printed test record “mouth alcohol Another test for the presence of MA is to make two
detected.” measurements not less than 6 min apart and evaluate the
Figure 7 shows exhalation profiles of ethanol deter- difference in concentration in relation to known elim-
mined with an Evidenzer instrument in one subject at ination kinetics of MA and rate of ethanol metabolism.
various times after washing the mouth with whisky but The simplest alternative is to observe the suspect for a
without swallowing any alcohol. time of 15–20 min and to ensure that nothing was
The Evidenzer and other infrared-based breath- placed in the mouth (Gullberg, 1992). The concentration
alcohol analyzers incorporate algorithms to monitor the of alcohol in the mouth after drinking dissipates rapidly
integrity of the exhaled ethanol breath profile. One such and by 10–15 min the sample is no longer contaminated
algorithm relies on detecting an early peak concentration with alcohol from a recent drink or use of mouthwash or
mouth spray. Some of the more modern evidential
breath analyzers monitor not only ethanol in the exhaled
0.6 air but also CO2 and water vapor and the known phy-
Breath alcohol concentration (mg/l)

siology of these exhaled gases and relationships to


0.5 ethanol exhalation can help to detect MA.
10 min
In some jurisdictions there is interest in conducting
0.4 the evidential breath-alcohol tests at the roadside and
therefore closer to the time of driving. This change in
0.3
procedure requires use of an instrument that incorpor-
0.2 ates a reliable slope detector so as to distinguish a proper
8.5 min
exhaled ethanol profile from a sample contaminated
0.1 17.5 min with alcohol from a recent drink or other sources of
ethanol (Wigmore and Leslie, 2001).
0 2.0 4.0 6.0 8.0 10.0 12.0
Figure 8 shows an exponential decrease in breath-
Exhalation time (s)
alcohol concentration after subjects ate alcohol-
containing chocolates or used an alcohol-containing
Figure 7 Single breath-alcohol exhalation profiles monitored using
mouth spray; it took 5–8 min to return to zero BrAC.
an evidential breath-alcohol analyzer (the Evidenzer) at 8.5 min,
10 min, and 17.5 min after drinking whisky. The wavy curves after 8.5
This experiment supports 15 min as an appropriate
and 10 min illustrate the effect of mouth alcohol on the exhalation waiting period after having alcohol in the mouth
profile, whereas the 17.5 min profile is uncontaminated end-exhaled before the result of an evidential breath-alcohol test is
breath. considered valid.

3
BrAC (mg/l)

Cherry chocolate

2 Mouth spray

Brandy chocolate

0
0 1 2 3 4 5 6 7 8
Time (min)
Figure 8 Elimination rate of mouth alcohol after subjects had eaten alcohol-containing chocolates or used a mouth spray that contained alcohol
just before tests were made. Courtesy of Draeger Safety AG, Lübeck, Germany.
132 Alcohol: Breath Analysis

Physiological Factors and Medical Conditions concentrations are highly correlated over a wide range
of drinking conditions (Jones, 2000).
The use of breath-alcohol analysis is subject to more The medical condition known as gastroesophageal
preanalytical factors and physiological variations com- reflux disease (GERD) was thought to falsify the results
pared with blood sampling and analysis of the latter at a of breath-alcohol analysis, owing to alcohol erupting
forensic laboratory. Therefore, it is important to stand- from the stomach into the mouth just before making the
ardize the breath-sampling procedure so that a minimum test (Duffus and Dunbar, 1984). However, a controlled
exhaled volume, time of exhalation, and pressure (flow human alcohol dosing study with patients suffering from
rate) to reduce as far as possible variations caused by GERD failed to verify that this medical complaint lead to
different breathing patterns. Studies have shown that the falsely high BrAC results compared with BAC (Kechagias
mean temperature of exhaled breath is close to 34.5 1C et al., 1999). However, more studies are needed to in-
and the temperature coefficient of alcohol solubility is vestigate whether GERD is indeed a bogus defense ar-
6.5% per degree. The temperature of the exhaled breath gument. Such experiments should include more subjects
and the exhaled volume, as well as temperature and diagnosed with this condition and different doses of al-
humidity of ambient air, are other variables that impact cohol consumed on an empty stomach and after a meal.
on the concentrations of ethanol in the breath.
The most common medical conditions that impact on
a person’s ability to provide a proper breath-alcohol test Response to Interfering Substances
are asthma and chronic obstructive pulmonary disease
(COPD). In a controlled drinking study, people with Human breath consists of a mixture of oxygen, nitrogen,
COPD had higher BBR of alcohol compared with people carbon dioxide, and water vapor and traces amounts of
with healthy lungs (Hahn et al., 1991). When asthmatics other volatile substance, either produced endogenously
were tested on two different evidential instruments, they or inhaled with the ambient air (Spanel and Smith,
were unable to provide a proper breath sample, that is, 2011). Acetone is the main endogenous volatile in ex-
to make a continuous exhalation at a fixed pressure for haled in the breath and this substance absorbs infrared
sufficiently long time to satisfy the requirements of radiation in the 3.3–3.5 mm wavelength range currently
modern breath-alcohol analyzers. used to monitor ethanol in breath (Spanel et al., 2011).
Nevertheless, as shown in Figure 9, even after a few However, the use of breath-test instruments that moni-
seconds of exhalation the concentration of alcohol tor absorption of infrared radiation at two or more
reached 66% of the final end-exhaled concentration. wavelengths can discriminate acetone from ethanol.
Another factor to consider in connection with breath Some of the latest generation of infrared analyzers are
analysis is the high solubility of alcohol in the mucous fitted with five filters in the range of 3.3–3.5 mm (e.g.,
membranes covering the upper airways of the lungs. Evidenzer) and can discriminate between several poten-
Some scientists believe that the breath-alcohol concen- tially interfering substances. In apprehended drivers
tration arises exclusively from equilibration of ethanol tested with Evidenzer, some of the breath tests were
in the upper airway and not as originally postulated at aborted because the instrument detected an interfering
the alveolar-capillary membrane in the alveoli (Hlastala, substance – a control analysis of blood in these indi-
2010). Wherever the final alcohol equilibrium occurs, viduals showed elevated concentrations of acetone and/
one cannot deny the fact that breath and blood-alcohol or isopropanol (Jones and Andersson, 2008).

180
98%
Breath alcohol concentration (µg/l)

94% 97% 164


160 87% 91%
85%
81%
140
75%
120 66%
End exhalation
100
80
60
40
20
0
0 1 2 3 4 5 6 7 8 9 10
Exhalation time (s)
Figure 9 Single breath-alcohol exhalation profile produced with an evidential infrared analyzer (the Evidenzer) showing ethanol content of partial
exhalations expressed in percent of the highest concentration after an end exhalation lasting 10 s.
Alcohol: Breath Analysis 133

Breath analyzers fitted with electrochemical detectors ethanol by potassium dichromate. Also available were
respond to other alcohols such as methanol and iso- small electronic instruments that incorporated a metal
propanol, which are also oxidized but at different rates oxide semiconductor or Taguchi Gas Sensor (TGS) as
compared with ethanol. Monitoring the rate of the oxi- the analytical principle. The TGS comprises a tiny bead
dation reaction using algorithms is a way to enhance of tin oxide which, when heated in ‘clean air’ absorbs
selectivity of electrochemical analysis of ethanol. Acetone oxygen onto its surface until a characteristic conduct-
is not oxidized at the same electrode potential as ethanol ance is obtained. When a combustible gas (such as
and does not cause interference problems when fuel cell ethanol, propane, or methane) is also present in breath,
instruments are used. Examples of other gases and it gets absorbed onto the bead surface, reacts with
volatiles produced endogenously include acetone, me- oxygen, and electrons are released changing the con-
thane, isoprene as well as acetaldehyde, the first product ductance of the bead in direct proportion to the
of ethanol metabolism. Substances inhaled with the am- concentrations of the combustible gas. The TGS is
bient air (gasoline fumes, toluene, butane) are also po- commonly used in smoke alarm detectors in homes, but
tential interfering substances. People might be exposed to is also incorporated into some devices for determination
these substances in the workplace or in connection with of BrAC. Semiconductors are not only used for identi-
volatile organic compound (VOC). fying ethanol, but also respond to other types of alco-
On the whole the problem posed by interfering sub- hols, hydrocarbons, and combustible gases such as
stances is much exaggerated, because apart from acetone carbon monoxide contained in cigarette smoke. The
human breath does not contain other substances in response to ethanol concentration is nonlinear and the
sufficiently high concentrations to pose a problem for calibration drifts over time. The effective working life of
the analysis of ethanol. The hand held fuel-cell breath the TGS is limited, depending to a large extent on how
analyzers do not oxidize acetone at the electrode po- often they are used. It appears that the surface effect by
tential used for ethanol analysis and so do not respond which they operate is dependent on the atmospheric
to elevated breath-acetone. The infrared analyzers cur- partial pressure of oxygen, and semiconductors have
rently used for legal purposes incorporate multiple in- been found to vary in sensitivity to alcohol depending on
frared wavelength filters to detect the presence of an the weather and the altitude at which they are operated.
interfering substance, such as acetone, in the breath Caution is necessary when breath-alcohol analyzers
sample. Either a correction is made to compensate for are used for self-testing, because it is not a standard
the concentration of acetone or the evidential breath test practice to check the calibration with each test and there
is aborted and a sample of blood or urine taken instead. is no way to monitor the presence of mouth alcohol. A
Interfering substances might represent a problem in test should not be made until at least 15–20 min have
people who abuse organic solvents by sniffing from a rag elapsed since the last drink to ensure extraneous sources
or inhalation by other means. More studies are needed of alcohol have disappeared from the mouth and oral
to test the effect of VOCs on the response of breath- mucosa. The most appropriate use for self-testers would
alcohol analyzers. be in the morning after an evening of heavy drinking to
ensure that alcohol has been eliminated from the body.

Breath Analyzers for Self-Testing


Ignition Interlock Devices
It is widely accepted that motor vehicles require a
speedometer to inform the driver whether the speed limit Most drunk drivers have a substance abuse problem and
is being exceeded. Many believe drivers should have could probably be diagnosed clinically as being alcohol
access to a personal breath-alcohol tester to indicate dependent. Studies have shown that repeat offending is a
whether the statutory breath-alcohol limit for driving is common occurrence among people arrested for drunk
exceeded after an evening’s social drinking. Some traffic driving in most nations. Dealing with this problem is a
safety authorities would like to see all newly produced major concern for traffic safety organizations (Marques
cars fitted with a device to analyze the driver’s breath for et al., 2001). Mandatory fitting of an interlock device is
alcohol (Radun et al., 2014). So far this radical idea has one approach to reduce recidivism rates among con-
not been put into practice although a number of low- victed drinking drivers. Most interlocks incorporate a
cost breath-test instrument are available for purchase by fuel cell type of breath analyzer connected to the vehicle
the public. However, some such instruments have ac- ignition system and a driver must pass the breath test
quired a dubious reputation with regard to analytical before the vehicle can be started in the conventional way
specificity and overall reliability of the results. (Fulkerson, 2003). Most interlock devices are set to read
The first breath-alcohol analyzers manufactured and as ‘fail’ at low threshold BrAC corresponding to 10 or
sold for self-testing appeared in the 1970s. The simplest 20 mg/100 ml blood-alcohol concentration.
of these was the chemical tube and bag (e.g., Alcotest or A number of safeguards are incorporated into inter-
Alcolyser), a once-only test involving the oxidation of locks to prevent a driver from circumventing the breath
134 Alcohol: Breath Analysis

test, such as sensing temperature and pressure of Concluding Remarks


the exhaled breath entering the instrument. Some people
have resorted to using a hand-pump to provide alcohol- During the 1970s the technology available for breath-
free ambient air into the breath analyzer and in this alcohol analysis underwent radical changes and a new
way pass the test and start the vehicle. Many ignition generation of analytical instruments appeared. Unlike
interlocks are programmed to request a second test the Breathalyzers which measures the ethanol in breath
after the vehicle is in motion, which is a safeguard by chemical oxidation, this new generation of instru-
to ensure that somebody other than the driver has ments utilized physical-chemical methods of analysis
provided the negative breath test needed to start the such as electrochemical oxidation and infrared spec-
vehicle. trometry. These new instruments were easier to operate
The stability of the calibration of interlock breath- and the entire sequence of tests was controlled by
alcohol analyzers requires close attention and monthly microprocessors. The breath sample was captured
or bimonthly maintenance checks should always be in- automatically as the test subject reached end exhalation,
cluded as a safeguard. Another problem sometimes en- which usually required a discarded volume of 1.5 l and
countered is from interfering substances in the ambient continuous exhalation for at least 6 s. Use of infrared
air, such as solvents used as windscreen cleaners, many instruments allowed the increase in ethanol concen-
of which contain isopropanol. Furthermore, people with tration to be monitored as a function of exhalation time
an interlock device fitted to their vehicle should be and curve analysis which made it possible to detect an
warned about extraneous sources of ethanol, such as abnormal ethanol profile that might indicate contamin-
that contained in cough medicines, vitamin tonics, and ation with mouth alcohol.
mouthwash preparations. The concentrations of ethanol in end-exhaled breath
Nevertheless, ignition interlock technology has ad- and in venous blood are highly correlated over a wide
vanced considerably and one finds that these devices range of concentrations of ethanol and drinking pat-
are fitted to public transport vehicles, including buses, terns. This is illustrated by the scatter plot in Figure 10
taxies, etc. in some nations (Willis et al., 2004). obtained from tests with people apprehended by the
If a convicted drunk driver agrees to having an inter- police for drunk driving in Sweden. Note that the BAC
lock device installed in his or her vehicle, this is one ranges from zero to 400 mg/100 ml and the correlation
way to speed-up the re-granting of a driving permit. coefficient between blood- and breath-alcohol was
On the whole the use of breath-activated interlock statistically highly significant (r ¼ 0.98, po.001). The
devices represents an effective way to reduce rates of dotted diagonal line in Figure 10 shows the line of
drunk driving recidivism by hard-core offenders (Elder identity. Most of the points (blood-breath pairs) are
et al., 2011). below this line, which means that results from the breath

400
N = 793
Estimated breath alcohol concentration (mg/100 ml)

Pearson’s r = 0.98
Residual SD = 15 mg/100 ml
300

200

100

0
0 100 200 300 400
Venous blood alcohol concentration (mg/100 ml)
Figure 10 High correlation between venous blood and exhaled breath-alcohol concentrations in tests carried out on drink drivers. Venous blood
was analyzed by headspace gas chromatography and breath-alcohol was determined with an infrared analyzer (the Intoxilyzer 5000).
Alcohol: Breath Analysis 135

test are lower than the actual venous BAC in the vast environment. Obtaining a proper breath sample might
majority of cases. not be feasible depending on the person’s age, gender,
The main advantage of breath-alcohol analysis com- and body stature. There is considerable evidence that
pared with blood-alcohol analysis includes noninvasive people with pulmonary dysfunction, such as asthma and
nature of the sampling and the fact that results of the test COPD, are unable to provide a proper sample with most
are obtained immediately afterwards. A positive roadside breath-alcohol analyzers.
breath-alcohol test of a driver needs to be verified The analytical specificity of instruments for identifi-
by making a more reliable evidential test at a police cation of ethanol in breath is lower than for blood
station. The seriousness of an impaired driving charge analysis by gas chromatography (GC), because GC
and the consequences for those found guilty of this crime methods separate interfering substances prior to quan-
demands rigorous quality assurance procedures. Some of titative analysis. Several GC instruments were developed
the most important requirements are listed below: as breath-alcohol analyzers (see Table 3), but were less
practical for use by police officers outside of a labora-
• The evidential breath-alcohol analyzer used must be tory environment. The problem of interfering substances
certified for its intended purpose by an appropriate when IR and EC methods are used for ethanol analysis
government agency or other legal authority. even without GC separation is much exaggerated. Apart
• The instrument should be operated by a trained and from acetone, the human breath does not contain vola-
certified operator and not by the arresting police tiles in sufficient amounts to interfere with ethanol ex-
officer. cept in rare instances when people abuse gasoline-type
• A minimum time of 15–20 min must have elapsed solvents by sniffing. The absorption of infrared radiation
since the last consumption of alcohol. The suspect is at a number of different wavelengths – 3.4 mm (C–H
not allowed any food, water, or medication during stretch) and 9.5 mm (C–O stretch) – enhances selectivity
this pre-test observation and deprivation period. further. In the C–H stretching region of the IR spectrum
• All evidential breath-alcohol tests should be done in use of multiple IR wavelength filters (e.g., 3.37, 3.41,
duplicate with a minimum of 5 min and maximum of 3.47, and 3.53 mm) allow the detection of interfering
15 min between the two tests. The two test results substances.
should agree within certain predefined limits (e.g., Besides forensics, analysis of volatile organic com-
710%) depending on concentration of ethanol in the pounds (VOCs) in breath has found applications in
sample analyzed. clinical and diagnostic medicine as biomarkers of dis-
• In conjunction with every subject test, a known eased states, such as acetone (diabetics), carbon mon-
concentration of ethanol in gas or vapor should oxide (smokers), methane (malabsorption), and nitric
be analyzed as a check on accuracy. The target oxide (asthma and lung disease). Methods involving
concentration of alcohol in the standard is usually liquid nitrogen freeze-traps have been developed to
at or near the statutory alcohol limit for driving. analyze an exhaled breath condensate for various en-
The accuracy test result should be within specified dogenous and exogenous substances. The condensate
limits, such as 75–10% of the known target is later analyzed by gas chromatography–mass spec-
concentration. trometry (GC–MS) or liquid chromatography–mass
• Samples of the ambient room air should be analyzed spectrometry (LC–MS) and provide useful and inter-
before and after every subject is tested and these esting metabolic profiles with diagnostic potential.
should register zero alcohol concentration. The sampling of breath has shown promise as a way
• A printed record of the test results and time of testing to detect the presence of drugs other than alcohol and as
should be generated including any error messages. a practical alternative to sampling and analysis of saliva
This printout should be signed by the responsible or urine. Trace amounts of substances such as amphet-
police officer and made available to the defense at- amine, cannabis, and cocaine are seemingly expelled in
torney if and when required. the breath and can be trapped on a specially prepared
filter for later desorption and analysis at a forensic or
Compared with the results of blood-alcohol analysis, clinical laboratory. The drugs attached to the filter are
the sampling and analysis of breath are more prone identified by highly sensitive analytical equipment, such
to biological variations, such as lung physiology. The as LC–MS. GC–MS or LC–MS instruments that permit
ethanol concentration in exhaled breath depends to the sampling and analysis of drugs in breath at the
some extent on the person’s pattern of breathing before roadside are not feasible at the present time.
exhalation is made into the instrument. The volume of Without doubt, the main forensic application of
breath exhaled prior to sampling and the temperature of breath analysis is to test the sobriety of drivers and
the breath sample also influence the ethanol content. generate evidence for prosecution of traffic offenders.
Other variables are the temperature and humidity of However, because some traffic offenders refuse to co-
ambient air breathed and evidential testing should be operate with the police in providing an approved sample
done in an air-conditioned room or other controlled of breath, there will always be a need for alternative
136 Alcohol: Breath Analysis

specimens, such as blood, saliva, or urine to verify Borkenstein, R.F., Smith, H.W., 1961. The Breathalyzer and its applications.
overconsumption of alcohol. Medicine, Science and the Law 1, 13–23.
Brick, J., Erickson, C.K., 2009. Intoxication is not always visible: An unrecognized
The concentration of alcohol in the breath is a better prevention challenge. Alcoholism: Clinical and Experimental Research 33,
indication of brain exposure to ethanol than is the 1489–1507.
venous blood. Studies have shown higher concentrations Drummer, O.H., 2014. Alcohol: Pharmacology and disposition. Academic Forensic
of ethanol in arterial blood compared with venous blood Pathology 4, 156–160.
Dubowski, K.M., 1991. The Technology of Breath Alcohol Analysis. Rockville, MA:
during the absorption phase of the BAC curve. Because
Department of Health and Human Services. Publication No. (ADM)92-1728.
arterial blood transports alcohol to the brain, this sug- Dubowski, K.M., 1994. Quality assurance in breath-alcohol analysis. Journal of
gests that breath-alcohol is a better indicator of the Analytical Toxicology 18, 306–311.
pharmacological and behavioral effects of ethanol than Duffus, P., Dunbar, J.A., 1984. Medical problems with breath testing of drunk
venous blood, especially during the absorption phase of drivers. British Medical Journal (Clinical Research Ed.) 289, 831–832.
Elder, R.W., Voas, R., Beirness, D., et al. and Task Force on Community Preventive
ethanol metabolism. Services, 2011. Effectiveness of ignition interlocks for preventing alcohol-
The enforcement of drunk driving laws by means of impaired driving and alcohol-related crashes: A Community Guide systematic
breath-alcohol instruments has several practical advan- review. American Journal of Preventive Medicine 40, 362–376.
tages. The immediate result of analysis means that sus- Emerson, V.J., Holleyhead, R., Isaacs, M.D., Fuller, N.A., Hunt, D.J., 1980. The
measurement of breath alcohol: The laboratory evaluation of substantive breath
pects can be confronted with information on whether test equipment and the report of an operational police trial. Journal of the
they are in breach of the law at a time when they are Forensic Science Society 20, 3–70.
most aware of their excessive drinking. Moreover, the Endres, H.G., Gruner, O., 1994. Comparison of D2O and ethanol dilutions in total
results from the breath test, if above the statutory alco- body water measurements in humans. Clinical Investigator 72, 830–837.
Fell, J.C., Voas, R.B., 2014. The effectiveness of a 0.05 blood alcohol concentration
hol limit means that immediate sanctions can be im-
(BAC) limit for driving in the United States. Addiction 109, 869–874.
posed, such as confiscation of the driving permit. Fulkerson, A., 2003. Blow and go: The Breath-Analyzed Ignition Interlock Device as
The technology currently available for breath-alcohol a technological response to DWI. American Journal of Drug and Alcohol Abuse
analysis is impressive and will be hard to improve upon 29, 219–235.
in terms of ease of use and accuracy and precision of the Gullberg, R.G., 1992. The elimination rate of mouth alcohol: Mathematical modeling
and implications in breath alcohol analysis. Journal of Forensic Sciences 37,
test results. The latest generation of breath-testing in- 1363–1372.
struments are more robust and portable and this has led Hahn, R.G., Jones, A.W., Billing, B., Stalberg, H.P., 1991. Expired-breath ethanol
to the notion of conducting the evidential test at the measurement in chronic obstructive pulmonary disease: Implications for
roadside in closer proximity to the actual driving. Time transurethral surgery. Acta Anaesthesiologica Scandinavica 35, 393–397.
Harding, P., Zettl, J.R., 2008. Methods for breath analysis. In: Garriott, J.C. (Ed.),
will tell whether this change in operating procedure in Garriott’s Medicolegal Aspects of Alcohol, fifth ed. Tucson, AZ: Lawyers &
use of breath-alcohol analyzers is practical and feasible Judges Publishing Co, pp. 229–253.
and whether the results are deemed acceptable for legal Hlastala, M.P., 1998. The alcohol breath test: A review. Journal of Applied
purposes. Physiology 84, 401–408.
Hlastala, M.P., 2010. Paradigm shift for the alcohol breath test. Journal of Forensic
Sciences 55, 451–456.
Jaffe, D.H., Siman-Tov, M., Gopher, A., Peleg, K., 2013. Variability in the blood/
See also: Alcohol: Acute and Chronic Use and Postmortem breath alcohol ratio and implications for evidentiary purposes. Journal of
Findings. Alcohol: Blood and Body Fluid Analysis. Backtracking Forensic Sciences 58, 1233–1237.
Calculations for Alcohol and Drugs. Road Traffic: Global Overview Jones, A.W., 1978. Variability of the blood:breath alcohol ratio in vivo. Journal of
of Drug and Alcohol Statistics Studies on Alcohol 39, 1931–1939.
Jones, A.W., 1996. Measuring alcohol in blood and breath for forensic purposes: A
historical review. Forensic Science Review 8, 13–44.
Jones, A.W., 2000. Medicolegal alcohol determinations: Blood or breath alcohol
concentration? Forensic Science Review 12, 23–48.
References Jones, A.W., 2010. Evidence-based survey of the elimination rates of ethanol from
blood with applications in forensic casework. Forensic Science Review 200,
1–20.
Alobaidi, T.A., Hill, D.W., Payne, J.P., 1976. Significance of variations in blood: Jones, A.W., 2011. Driving under the influence of alcohol. In: Moffat, A.C.,
Breath partition coefficient of alcohol. British Medical Journal 2, 1479–1481. Osselton, M.D., Widdop, B., Watts, J. (Eds.), Clarke’s Analysis of Drugs and
Anstie, F.E., 1867. Prognosis and treatment of certain acute diseases. Lancet 90,
Poisons, vol. 1. London: The Pharmaceutical Press, pp. 87–114.
385–387.
Jones, A.W., Andersson, L., 1996. Variability of the blood/breath alcohol ratio in
Anstie, F.E., 1874. Final experiments on the elimination of alcohol from the body.
drinking drivers. Journal of Forensic Sciences 41, 916–921.
Practitioner 13, 15–28.
Jones, A.W., Andersson, L., 2008. Determination of ethanol in breath for legal
Baraona, E., Gentry, R.T., Lieber, C.S., 1994. Bioavailability of alcohol: Role of
purposes using a five-filter infrared analyzer: Studies on response to volatile
gastric metabolism and its interaction with other drugs. Digestive Diseases 12,
351–367. interfering substances. Journal of Breath Research 2, 026006.
Begg, T.B., Hill, I.D., Nickolls, L.C., 1964. Breathalyzer and Kitagawa-Wright Jones, A.W., Jonsson, K.A., 1994. Food-induced lowering of blood-ethanol profiles
methods of measuring breath alcohol. British Medical Journal 1, 9–15. and increased rate of elimination immediately after a meal. Journal of Forensic
Blomberg, R.D., Peck, R.C., Moskowitz, H., Burns, M., Florentino, D., 2009. The Sciences 39, 1084–1093.
Long Beach/Fort Lauderdale relative risk study. Journal of Safety Research 40, Jones, A.W., Kugelberg, F.C., Holmgren, A., Ahlner, J., 2009. Five-year update on
285–292. the occurrence of alcohol and other drugs in blood samples from drivers killed
Bogen, E., 1927. Drunkenness: A quantitative study of acute alcoholic intoxication. in road-traffic crashes in Sweden. Forensic Science International 186, 56–62.
JAMA 89, 1508–1511. Jones, A.W., Lindberg, L., Olsson, S.G., 2004. Magnitude and time-course of
Borkenstein, R.F., 1985. Historical perspective: North American traditional and arterio-venous differences in blood-alcohol concentration in healthy men. Clinical
experimental response. Journal of Studies on Alcohol 10 (Suppl), 3–12. Pharmacokinetics 43, 1157–1166.
Alcohol: Breath Analysis 137

Kechagias, S., Jonsson, K.A., Franzen, T., Andersson, L., Jones, A.W., 1999. Penttila, A., Tenhu, M., 1976. Clinical examination as medicolegal proof of alcohol
Reliability of breath-alcohol analysis in individuals with gastroesophageal reflux intoxication. Medicine, Science and the Law 16, 95–103.
disease. Journal of Forensic Sciences 44, 814–818. Porter, R., 1985. The drinking man’s disease: The ‘pre-history’ of alcoholism in
Lindberg, L., Brauer, S., Wollmer, P., et al., 2007. Breath alcohol concentration Georgian Britain. British Journal of Addiction 80, 385–396.
determined with a new analyzer using free exhalation predicts almost precisely Radun, I., Ohisalo, J., Rajalin, S., et al., 2014. Alcohol ignition interlocks in all new
the arterial blood alcohol concentration. Forensic Science International 168, vehicles: A broader perspective. Traffic Injury Prevention 15, 335–342.
200–207. Romano, E., Torres-Saavedra, P., Voas, R.B., Lacey, J.H., 2014. Drugs and
Lucas, D.M., 2000. Professor Robert F. Borkenstein: An appreciation of his life and alcohol: Their relative crash risk. Journal of Studies on Alcohol and Drugs 75,
work. Forensic Science Review 12, 1–22. 56–64.
Marques, P.R., Tippetts, A.S., Voas, R.B., 2003. The alcohol interlock: An Spanel, P., Dryahina, K., Rejskova, A., Chippendale, T.W., Smith, D., 2011. Breath
underutilized resource for predicting and controlling drunk drivers. Traffic Injury acetone concentration: Biological variability and the influence of diet.
Prevention 4, 188–194. Physiological Measurement 32, N23–N31.
Marques, P.R., Tippetts, A.S., Voas, R.B., Beirness, D.J., 2001. Predicting repeat Spanel, P., Smith, D., 2011. Volatile compounds in health and disease. Current
DUI offenses with the alcohol interlock recorder. Accident Analysis and Opinion in Clinical Nutrition and Metabolic Care 14, 455–460.
Prevention 33, 609–619. Spring, J.A., Buss, D.H., 1977. Three centuries of alcohol in the British diet. Nature
Martin, E., Moll, W., Schmid, P., Dettli, L., 1984. The pharmacokinetics of alcohol 270, 567–572.
in human breath, venous and arterial blood after oral ingestion. European Journal Sterling, K., 2012. The rate of dissipation of mouth alcohol in alcohol positive
of Clinical Pharmacology 26, 619–626. subjects. Journal of Forensic Sciences 57, 802–805.
Mason, M.F., Dubowski, K.M., 1976. Breath-alcohol analysis: Uses, methods, and Stowell, A.R., Gainsford, A.R., Gullberg, R.G., 2008. New Zealand’s breath and blood
some forensic problems: Review and opinion. Journal of Forensic Sciences 21, alcohol testing programs: Further data analysis and forensic implications.
9–41. Forensic Science International 178, 83–92.
Mellanby, E., 1920. Alcohol and alcoholic intoxication. British Journal of Inebriety Wigmore, J.G., Leslie, G.M., 2001. The effect of swallowing or rinsing alcohol
17, 157–179. solution on the mouth alcohol effect and slope detection of the intoxilyzer 5000.
Mitchell Jr., M.C., Teigen, E.L., Ramchandani, V.A., 2014. Absorption and peak Journal of Analytical Toxicology 25, 112–114.
blood alcohol concentration after drinking beer, wine, or spirits. Alcoholism: Willis, C., Lybrand, S., Bellamy, N., 2004. Alcohol ignition interlock programmes for
Clinical and Experimental Research 38, 1200–1204. reducing drink driving recidivism. Cochrane Database of Systematic Reviews.
Norberg, A., Jones, A.W., Hahn, R.G., Gabrielsson, J.L., 2003. Role of variability in CD004168.
explaining ethanol pharmacokinetics: Research and forensic applications. Clinical Wright, B.M., Jones, T.P., Jones, A.W., 1975. Breath alcohol analysis and the blood:
Pharmacokinetics 42, 1–31. Breath ratio. Medicine, Science and the Law 15, 205–210.

Das könnte Ihnen auch gefallen