Sie sind auf Seite 1von 1

PROBE CALIBRATION PROTOCOL

______________________________________________________________________
Date: Thursday, February 28, 2019
Time: 16:30:36
______________________________________________________________________
Device Information
Version: V3.2-1
Serial Number: 0020004716
Description: HC2
RS-485 Address: 0
Device protection: Off
______________________________________________________________________
Previous Adjustment

Factory Humidity Adjustment:


Date/Reference ID: 12/16/2014 / 20080001
Adjustment Values: 10.9% 34.5% 78.3%
User Humidity Adjustment:
Date/Reference ID: 3/20/2018 / 00007631
Adjustment Values: 21.2% 44.5% 68.7%
Factory Temperature Adjustment:
Date/Reference ID: 12/16/2014 / 20080001
Adjustment Values: 23.98°C
______________________________________________________________________

Humidity Calibration

Calibration Point 1
Reference Value: 20.88%rh
As-Found Value: 23.24%rh
Temperature: 19.28°C
Error: 2.36%rh

Calibration Point 2
Reference Value: 45.62%rh
As-Found Value: 45.53%rh
Temperature: 19.64°C
Error: -0.09%rh

Calibration Point 3 (2/28/2019)


Reference Value: 69.83%rh
As-Found Value: 69.44%rh
Temperature: 19.88°C
Error: -0.39%rh

______________________________________________________________________
HW4 User Name: No user logged on
User Description:

Version: V3.8.0.16151
HW4 ID: 24 00007631

Das könnte Ihnen auch gefallen