KINEMETRICS Inc.

ROCK+ Series

ROCK+ Digitizer
User Manual

Document 300816
Revision B
October 2017
   
Warranties, Disclaimers & 
Trademarks 
 
Copyright  © 2008‐2017 Kinemetrics,  Inc. 
The trademarks used throughout this manual, registered or not, are: Kinemetrics, Obsidian, 
Slate, Marmot, Rock, Rock+, Rockhound, Q330, Q330HR, and Linux. 
This publication is provided ʺas isʺ without warranty of any kind, either expressed or im‐ 
plied, including, but not limited to, the implied warranties of merchantability, fitness for 
a particular purpose, or non‐infringement. Kinemetrics, Inc. and its affiliates assume no 
responsibility for errors or omissions in this publication or other documents which are refer‐ 
enced by or linked to this publication. 
References to corporations, their services and products, are provided ʺas isʺ without warranty 
of any kind, either expressed or implied.  In no event shall Kinemetrics, Inc.  be liable  for 
any special, incidental, indirect or consequential damages of any kind, or any damages 
whatsoever, including, without limitation, those resulting from loss of use, data or profits, 
whether or not advised of the possibility of damage, and on any theory of liability, arising 
out of or in connection with the use or performance of this information. 
This publication could include technical or other inaccuracies or typographical errors. Changes 
are periodically added to the information herein; these changes will be incorporated in new 
editions of the publication. 
All rights reserved.  No part of this publication may be copied, photocopied, reproduced, 
transmitted, transcribed, or reduced to any electronic medium or machine‐readable form 
without prior written consent of Kinemetrics, Inc. 
 
Notice 
Kinemetrics Inc. reserves the right to make improvements in the software described in this 
documentation at any time and without notice. The information contained here is subject 
to change without notice and should not be construed as a commitment by Kinemetrics Inc. 
The software described in this document is provided as a licensed item, in conjunction with 
Kinemetrics  equipment.  It  may  not  be  copied  or  distributed  for  use  on  other  than  the 
equipment it was licensed for. 

   
 

Disclaimer 
Kinemetrics  Inc.  shall  have no  liability  or  responsibility  to  you or  any other  person  or 
entity with respect to any liability, loss or damage caused or alleged to be caused directly 
or indirectly by this documentation or the software described in it. This includes but is not 
limited to any interruption of service, loss of business or anticipatory profits or consequential 
damages resulting from the use or operation of such software or computer programs. 
 
Warranty 
We warrant each new product manufactured by Kinemetrics for a period of one year from 
date of shipment. Defects in material or workmanship found within that period will be 
replaced or repaired (at our option) without charge for materials or labor. If Kinemetrics 
authorizes the return of a product, we will pay the round trip freight charges to the factory 
for repair under warranty. If subsequent evaluation at Kinemetrics establishes that necessary 
repairs are due to misuse, then the customer must assume all charges. 
Insurance for all shipments, either first sale or repair, are the responsibility of the customer. 
Kinemetrics can arrange to have a policy purchased on behalf of the customer for the first 
sale; however, it is the responsibility of the customer to notify the carrier immediately of 
any freight or handling damage. Kinemetrics will make every effort to assist the customer 
in filing a claim with the carrier or insurance company. 
If on‐site warranty repair or replacement is required, the customer will be charged the then‐ 
current field service rate for portal‐to‐portal travel time plus actual portal‐to‐portal travel 
charges. There is no charge for on‐site warranty repair labor. 
Items not manufactured by Kinemetrics but included in systems (e.g. peripherals, options) 
are warranted for 90 days from date of shipment. Items not manufactured by Kinemetrics 
and  not  part  of  a  system  (e.g.  digitizers,  printers,  analyzers)  may be  warranted by the 
original equipment manufacturer. Kinemetrics will do everything possible to expedite and 
coordinate any warranty service from the original   manufacturer. 
Software not produced by Kinemetrics may carry its own warranty and the customer should 
sign any appropriate license agreement(s) and return to software manufacturer. Kinemetrics 
assumes no responsibility for such third‐party software. 
Software and software updates provided by Kinemetrics Inc. for its Strong Motion and 
Seismological measurement and recording equipment have a warranty period of one year. 
This warranty applies to the standard software package as well as to options or special 
software provided to the customer. An update shipped under warranty will be covered by 
the original system’s warranty for the balance of the one year period. 
Warranty claims shall be made on Software Change Request forms (SCRs). Problems re‐ 
ported by filing an SCR within one year will be corrected free of charge. SCRs filed after 
the one year period will be billed at the then‐current rates. 

   
 

The method of correction will be at Kinemetrics Inc.’s discretion, in that a correction may 
be supplied via a software patch, or by shipping updated software. 
Shipment of updated software will sometimes require hardware or configuration changes to 
the system. Hardware changes may include, but are not limited to, memory and disk drives. 
Required hardware or configuration changes are not included in the cost of a software update, 
and may represent an additional cost to the customer. 
All software,  once delivered,  is covered under warranty.  Updates fitting the following 
descriptions  would  NOT  be  considered  valid  warranty  claims,  and  the  software  would  be 
billed accordingly: 

• Updates not prompted by a software problem. 
• Additional software options requested voluntarily by the customer, such as the addition 
of special software. 
 
Kinemetrics, Inc., 222 Vista Avenue, Pasadena, CA 91107 
USA 
Phone:  (626) 795‐2220, Fax:  (626) 795‐0868 
E‐mail: services@kmi.com Technical Support: support@kmi.com 
Website: www.kinemetrics.com 
 
Services available from Kinemetrics 

Installation 

Field support for on‐site installation, supervision and check‐out of Kinemetrics products 
is available from Kinemetrics. Training for the customer’s staff is also available, either at 
Kinemetrics’ facilities in Pasadena, or at the customer’s site. We recommend contracting for 
installation services along with instrument  procurement. 
 
Maintenance 

Periodic field maintenance programs are offered for our products. Current programs include 
annual contracts to service data acquisition systems and accelerographs in high‐rise build‐ 
ings, free‐field accelerographs, offshore platform monitoring systems, maintenance at nuclear 
power plants and seismic network maintenance. 

   
 

Recalibration and Repair 

Kinemetrics will repair and/or modify many types of electromechanical instruments and 
systems. All repair and calibrations are warranted for a period of 90 days for workmanship. 
European customers may contact Kinemetrics SA for assistance at: 
ZI Le Trési 6B 
CH‐1028 Préverenges, Switzerland 
E‐mail:  kmi_sa@bluewin.ch 
Telephone  ++4121 803 2829 fax ++4121 803   2895. 
 
For Faster Service 

When returning any product to Kinemetrics please request an RMA number and make 
reference to this number in any following correspondence. 

   
Declaration of Conformity
We

KINEMETRICS, INC.
222 VISTA AVENUE
PASADENA, CA 91107
USA

hereby declare that the equipment specified conforms to the essential requirements of
the following applicable European Directives and Standards and carries the CE mark
accordingly:

Product Name: Kinemetrics OBSIDIAN UNITS


Product Description: Multichannel Seismic Recorder
Model Number(s): Obsidian, Obsidian 4x, 8x, 12x, 16x, 24x, &
36x
Equipment Class: Electrical Equipment Measurement Control
and Laboratory Use-Industrial
Directives: Council Directive 2014/30/EU and
2006/95/EC

Standards: Safety: EN 61010-1:2010

EMC: EN 61326-1:2013
EN55011 Class A Group 1

EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11

Supplementary Information:

Pasadena, CA USA
May 2015

Ian M. Standley
Vice President, Engineering
   
Contents
ROCK+ Series .............................................................................................................. 1 
Warranties, Disclaimers & Trademarks ....................................................................... 3 

Copyright  © 2008‐2017 Kinemetrics,  Inc. .................................................................. 3 

Notice .................................................................................................................................... 3 

Disclaimer ............................................................................................................................ 4 

Warranty ............................................................................................................................. 4 

Kinemetrics, Inc., 222 Vista Avenue, Pasadena, CA 91107 USA ..................... 5 

Services available from Kinemetrics ............................................................................ 5 
Installation ..................................................................................................................... 5 
Maintenance ................................................................................................................. 5 
For Faster Service ...................................................................................................... 6 
Safety .................................................................................................................................... 1 
These symbols may appear on Kinemetrics equipment or in this manual1 
These safety‐related terms appear in this manual ............................................ 2 
Optional  Power  Supply Assembly ......................................................................... 2 
User‐Supplied Batteries or Charging System ..................................................... 2 
External  Battery ......................................................................................................... 2 
Grounding the Digitizer ........................................................................................... 3 
Use the Proper Power   Cord ................................................................................... 3 
Antenna, Phone & LAN  Cabling .......................................................................... 3 
Do Not Operate in Explosive Atmospheres ....................................................... 3 
The Kinemetrics’ Rock+ Digitizer is not To Be Used For Life Sup‐  port or 
Life‐Critical  Systems ........................................................................................ 3 
Chapter 1 Introduction....................................................................................................... 4 

1.1 System  Overview ........................................................................................................ 5 
1.1.1 How is the Rock+ Series Different from the Rock   Series? .................. 6 
1.1.2 What is  Rockhound? ........................................................................................ 6 
1.1.3 Front Panel .......................................................................................................... 7 

1.2 Installation &  Setup ................................................................................................. 13 
1.2.1 Operating Environment ................................................................................. 16 
1.2.2 Unpacking & Inspecting the  Unit .............................................................. 17 
1.2.3 Requirements for Installation ......................................................................... 17 
1.2.4 Practice  Assembly ........................................................................................... 19 
1.2.5 Planning your installation ............................................................................. 19 
1.2.6 Installing the Digitizer .................................................................................... 21 
  i 
Chapter 2 System Overview .......................................................................................... 53 

2.1 Intelligent Power  Supply......................................................................................... 54 
2.1.1 Supercapacitor  Backup .................................................................................. 55 
2.1.2 Battery Charging ............................................................................................. 55 
2.1.3 Sequenced Start  Up ........................................................................................ 56 
2.1.4 Sequenced Shut Down ..................................................................................... 57 
2.1.5 Communication with the User Via Status LEDs .................................... 57 
2.1.6 Magnetic Switch ............................................................................................... 58 
2.1.7 Temperature Monitoring ............................................................................... 58 
2.1.8 Humidity Monitoring ...................................................................................... 58 
2.1.9 Timed Operation Windows .......................................................................... 59 
2.1.10  Parametric Operation Protection ................................................. 59 
2.1.11  System  Watchdog ............................................................................... 60 
2.1.12  Power Fail  Protection ........................................................................ 60 
2.1.13  Application Specific  Configuration ................................................. 60 
2.1.14  Power System Display ....................................................................... 62 
2.1.15  Updating the Power  Supply ........................................................... 64 

2.2 Processor System ..................................................................................................... 64 
2.2.1 Processor ............................................................................................................ 64 
2.2.2 Storage ................................................................................................................. 64 
2.2.3 Communication Interfaces .............................................................................. 64 

2.3 ADC/DSP SYSTEM .............................................................................................. 67 
2.3.1 Individual DSP .................................................................................................. 67 
2.3.2 C++ DSP Code ............................................................................................... 67 
2.3.3 Isolated DC/DC Power supply for Analog Section ............................... 67 
2.3.4 Isolation Barrier to Analog Section ............................................................ 67 
2.3.5 Four Delta Sigma ADC  Converters ........................................................... 67 
2.3.6 Adjustable Gain Input Stages ...................................................................... 68 
2.3.7 Calibration and Control Processor & Mass position/Auxiliary ADC68 

2.4 External Sensor Input  Boards ............................................................................. 68 
2.4.1 Military Style  Connectors ............................................................................. 68 
2.4.2 EMI/RFI and Transient Protection  Circuitry ....................................... 68 
2.4.3 Calibration and Sensor Control Signal Outputs .................................... 68 

2.5 EpiSensor Theory of Operation ............................................................................ 68 
2.5.1 Working Principle ............................................................................................ 69 
2.5.2 Pole Zero Representation of the  EpiSensor ............................................ 70 
2.5.3 Polarity  Conventions ...................................................................................... 71 

2.6 Firmware Description .............................................................................................. 71 
2.6.1 Linux Overview ................................................................................................. 71 
2.6.2 Common Linux Commands ......................................................................... 72 
2.6.3 The  VI Editor ................................................................................................... 74 
2.6.4 Updating Linux ................................................................................................ 76 
2.6.5 Software Versions .............................................................................................. 77 
2.6.6 Software Development  Tools ......................................................................... 78 
  ii 
Chapter 3 .......................................................................................................................... 79 

3.1    Initial Setup ............................................................................................................ 79 
3.1.1 Network Address ............................................................................................. 80 
3.1.2 Viewing System  Information ......................................................................... 80 
3.1.3 The Web  Interface ........................................................................................... 81 

3.2    Basic  Setup ............................................................................................................. 81 
3.2.1 Sensor  Groups ................................................................................................... 82 
3.2.2 Voltage Ranges ................................................................................................. 83 
3.2.3 Sensitivity ........................................................................................................... 84 
3.2.4 Obsidian Internal Deck Zero Adjustment ................................................ 86 
3.2.5  EpiSensor2 Support ........................................................................................ 87 
3.2.6  Other Channel  Parameters .......................................................................... 88 
3.2.7  Physical vs.  Virtual Channels ..................................................................... 89 
3.2.8  State‐of‐Health  Streams ................................................................................ 90 
3.2.9  Trigger Levels .................................................................................................... 91 
3.2.10  Voting Options ...................................................................................... 92 
3.2.11  Other  Parameters ............................................................................... 92 
3.2.12  Activating Parameter  Changes ....................................................... 92 
3.2.13  Passwords .............................................................................................. 93 
3.2.14  Locked File System ................................................................................ 93 
3.2.15  Save  Parameters ................................................................................. 94 

3.3 Triggered Recording .................................................................................................. 94 
3.3.1 Pre‐Event Time ................................................................................................ 94 
3.3.2 Post Event Time .............................................................................................. 94 
3.3.3 Minimum Run Time ....................................................................................... 94 
3.3.4 Channel Triggering ........................................................................................... 94 

3.4 Timing  System ........................................................................................................ 102 
3.4.2 Disciplined  Oscillator .................................................................................... 103 
3.4.3 Time Quality .................................................................................................... 103 
3.4.4 Timing  Options ............................................................................................... 103 
3.4.5  Using an NTP Server .................................................................................... 107 
3.4.6  Acting as an NTP Server ............................................................................. 107 

3.5 Anti‐Alias Filtering ................................................................................................ 107 

3.6 TCP/IP Primer ...................................................................................................... 111 
3.6.1 Terminology ...................................................................................................... 111 
3.6.2 Addresses .......................................................................................................... 112 
3.6.3 Some GuidelinesSome  Guidelines .............................................................. 113 
3.6.4 Typical  Configurations ................................................................................. 113 
3.6.5 Rock+ Services ............................................................................................... 114 
3.6.6 Further .............................................................................................................. 114 

3.7 File Management .................................................................................................... 117 

3.8 File Viewing ............................................................................................................. 119 
  iii 
3.9 Waveform  Viewing ................................................................................................. 121 

4.1 Editing  Parameters ................................................................................................ 144 

4.2 Adding Modules ...................................................................................................... 147 

4.3 Removing  Modules ................................................................................................ 150 

4.4 Replacing  Modules ................................................................................................ 151 
Chapter 5 File Viewer .................................................................................................... 155 

5.1   File Viewer Dynamic  Operation .................................................................. 159 
Chapter 6 ........................................................................................................................ 163 

6.1   Batch  Mode .......................................................................................................... 171 
Chapter 7 ........................................................................................................................ 173 
Chapter 8 ........................................................................................................................ 175 

8.1 Terminal mode .......................................................................................................... 176 

8.2 Batch Mode .............................................................................................................. 181 

8.3 Streaming .................................................................................................................. 182 

8.4 Dial On  Event ........................................................................................................... 182 

8.5 Modem Only Operation ........................................................................................ 183 
8.5.2 File  Transfer ..................................................................................................... 187 
8.5.3 Parameter Download .................................................................................... 187 
8.5.4 Editing Parameters with RockTalk ........................................................... 187 
8.5.5 Parameter  Upload ......................................................................................... 189 
8.5.6 Installing Firmware  Updates ..................................................................... 189 
Chapter 9 Storage ...........................................................................................................191 

9.3 USB Subsystem ....................................................................................................... 192 
9.3.1 USB  Device ...................................................................................................... 193 
9.3.3  USB  Wifi ........................................................................................................... 196 
9.3.4  USB Cellular  Modem ................................................................................... 197 
Chapter 10 DFS telemetry ........................................................................................... 201 

10.2  Configuration .................................................................................................. 202 
10.3.2  Extended DFS  Protocol .................................................................. 203 

10.4  Testing ............................................................................................................... 203 
Chapter 11 FAQs .............................................................................................................205 

11.1  What is  “dig1”? ............................................................................................. 205 

11.2  Changing trigger levels ................................................................................ 205 
  iv 
11.3  Using Different  Triggers  or  Trigger Filters ........................................... 207 

11.4  Adding Channels ........................................................................................... 210 

11.5  Switching Between Layouts ....................................................................... 213 

11.6  Changing Output Data  Formats ............................................................. 214 

11.7  Timed Recording ............................................................................................ 217 

11.8  Network Triggering ........................................................................................ 220 

11.9  Telemetry Options .......................................................................................... 222 
Chapter 12 AdvancedOperation................................................................................... 223 

12.1  RockTalk  with  Add‐ons .............................................................................. 223 

12.2  Layout  Wizard ............................................................................................... 223 

12.3  Configuration  Options ................................................................................. 234 


12.3.1  Network  Parameters ....................................................................... 234 
12.3.2  Modem Configuration ...................................................................... 234 
12.3.3  IO  Bits .................................................................................................. 234 
12.3.4  Networking and Security ................................................................ 236 

12.4  Low Latency  Telemetry .............................................................................. 237 


Chapter 13 ...................................................................................................................... 239 

13.1  Powering Up the  System ............................................................................ 239 

13.2  Powering Down the System ........................................................................ 240 

13.3  Software Installation ..................................................................................... 240 

13.4  IP Services ....................................................................................................... 242 

13.5  Network  Configuration ............................................................................... 243 

13.6  Dynamic  DNS ................................................................................................ 245 


13.6.1  Setting Up  DDNS ............................................................................. 245 
13.6.2  DDNS Verification ............................................................................. 246 

13.7  Software Watchdog ....................................................................................... 246 
13.7.1  Default Behavior: .............................................................................. 247 
13.7.2  Powerfail  Protection: ....................................................................... 247 
13.7.3  Application Specific  Configuration: .............................................. 247 

13.8  Linux  Passwords ............................................................................................ 248 

13.9  File  Retrieval .................................................................................................. 248 

13.10  Save/Restore Parameters ........................................................................... 250 
  v 
13.11  Software Updates .......................................................................................... 254 

13.12  Log  files ............................................................................................................ 255 

13.13  State of  Health ................................................................................................ 257 


Chapter 14 Software Tools ............................................................................................ 263 

14.1  Web  Browser ................................................................................................... 263 

14.2  Java .................................................................................................................... 264 

14.3  Terminal  Program ......................................................................................... 264 

14.4  Telnet Client ................................................................................................... 264 

14.5  WinSCP ............................................................................................................ 264 

14.6  PuTTY .............................................................................................................. 267 

14.7  FTP Server ...................................................................................................... 270 
Chapter 15 Troubleshooting .......................................................................................... 277 

15.1  Modem not  Communicating ...................................................................... 277 

15.2  Private Networks ........................................................................................... 277 

15.3  Altus Emulator and  SDS ............................................................................. 277 

15.4  Deleting a large Linux  directory ............................................................. 278 

15.5  Admin cannot delete data files ................................................................... 278 

15.6  IP issues.  Firewalls and router    blocks .................................................. 279 

15.7  Can’t Change Password.............................................................................. 279 

15.8  External  Charger .......................................................................................... 280 


Chapter 16 ......................................................................................................................282 

16.1  Run Remote Check Tests ............................................................................ 282 

16.2  Advanced Self Test Capabilities ................................................................. 282 

16.3  Troubleshooting &  Service ......................................................................... 282 

16.4  Hardware Problems ...................................................................................... 282 

16.5  Installing New  Firmware ............................................................................. 282 

16.6  Preventive Maintenance ............................................................................. 283 
16.6.1  Desiccant Replacement ................................................................... 283 
16.6.3  Replacing SD Cards ......................................................................... 286 
  vi 
16.6.4  Cleaning the Digitizer ..................................................................... 286 

16.7  Decommissioning &  Recycling ................................................................ 287 


16.7.1  Kinemetrics Recycling  Passport .................................................. 287 
16.7.2  Disassembly  Procedure  for  Recycling ........................................ 289 
16.7.3  Materials for Recycling ................................................................... 290 
Chapter 17 HardwareReference .................................................................................. 291 

17.1  System Power ................................................................................................. 291 

17.2  Connector Reference .................................................................................... 292 
17.2.1  Power  Connector .............................................................................. 292 
17.2.2  Internal  GPS ....................................................................................... 294 
17.2.3  Serial Connectors .............................................................................. 296 
17.2.4  Modem.................................................................................................. 304 
17.2.5  Time I/O  Connectors ....................................................................... 305 
17.2.6  USB  Connectors ................................................................................ 307 
17.2.7  External Three Channel Sensor Interface ................................. 310 
17.2.8  External 4th Channel Single Input Sensor  Connector .......... 311 
17.2.9  Four Channel Terminal Board Dual Supply (KMI 111915‐ 
01)  312 
17.2.10  Four Channel Terminal Board Single Supply (KMI 111915‐ 02)
  313 
17.2.11  Ethernet ............................................................................................... 314 

17.3     Cable Drawings ................................................................................................. 317 
Chapter 18 ...................................................................................................................... 327 

18.1  Glossary ............................................................................................................ 327 
 

  vii 
 
List of Figures
 
Figure 1: Rock+ Digitizer ............................................................................................................................... 8 
Figure 2:  Rock+ Quick Setup Guide .......................................................................................................... 15 
Figure 3: Drilling Mounting Hole .............................................................................................................. 23 
Figure 4: Installing Anchor ......................................................................................................................... 24 
Figure 5: Anchor Stud Installed ................................................................................................................. 25 
Figure 6: Mounted to the Floor................................................................................................................... 26 
Figure 7: Bullet Antenna.............................................................................................................................. 28 
Figure 8: Bullet Antenna Connector .......................................................................................................... 29 
Figure 9: Dual Supply Terminal Board (‐01 Version) ............................................................................. 36 
Figure 10: Dual Supply Terminal Board (w/ Cable) ................................................................................ 38 
Figure 11: Single Supply Terminal Board (‐02 Version) ......................................................................... 40 
Figure 12: Basic Functional Test ................................................................................................................. 41 
Figure 13: EpiSensor Simplified Block Diagram ...................................................................................... 69 
Figure 14: Sensor Groups ............................................................................................................................ 82 
Figure 15: Unit Conversion ......................................................................................................................... 84 
Figure 16: Sensitivity .................................................................................................................................... 85 
Figure 17: Waveform Display ..................................................................................................................... 86 
Figure 18: Zero Adjustment ........................................................................................................................ 87 
Figure 19: Channel Mapping ...................................................................................................................... 90 
Figure 20: Trigger Levels ............................................................................................................................. 91 
Figure 21: Typical trigger filter response .................................................................................................. 96 
Figure 22: STA/LTA Simplified Flowchart ............................................................................................... 98 
Figure 23: Recorded Files .......................................................................................................................... 117 
Figure 24: Recorded Files Detail ............................................................................................................... 118 
Figure 25: Downloading the File Viewer ................................................................................................ 119 
Figure 26: File Viewer ................................................................................................................................ 120 
Figure 27: Waveform Viewer .................................................................................................................... 121 
Figure 28: Waveform Display ................................................................................................................... 122 
Figure 29:  Real Time PSD ......................................................................................................................... 123 
Figure 30: Login Prompt ........................................................................................................................... 125 
Figure 31: Overview Screen ...................................................................................................................... 127 
Figure 32: Hardware Configuration ........................................................................................................ 129 
Figure 33:  Configuration .......................................................................................................................... 130 
Figure 34: Module Parameter Example ................................................................................................... 131 
Figure 35: Layout Display ......................................................................................................................... 132 
Figure 36: Channel Summary ................................................................................................................... 133 
Figure 37: Site Summary ........................................................................................................................... 134 
Figure 38: SOH Selection ........................................................................................................................... 135 
Figure 39: SOH Display ............................................................................................................................. 136 

  viii 
Figure 40: Applet Mode Selection ............................................................................................................ 137 
Figure 41: Waveform Viewer Selection ................................................................................................... 138 
Figure 42: Waveform Display ................................................................................................................... 139 
Figure 43: PSD Display .............................................................................................................................. 140 
Figure 44: Recorded Files .......................................................................................................................... 141 
Figure 45: Recorded File Detail ................................................................................................................ 142 
Figure 46: Triggering and Sensors ........................................................................................................... 143 
Figure 47: Overview ................................................................................................................................... 144 
Figure 48:  Hardware Parameters ............................................................................................................ 145 
Figure 49: Help String ................................................................................................................................ 145 
Figure 50:  Configuration Parameters ..................................................................................................... 146 
Figure 51: Module Parameters ................................................................................................................. 147 
Figure 52: Module Add ............................................................................................................................. 148 
Figure 53: FTP Sender Parameters ........................................................................................................... 149 
Figure 54: Module Remove ....................................................................................................................... 150 
Figure 55: Module Replace........................................................................................................................ 151 
Figure 56:  Module Replacement Options .............................................................................................. 152 
Figure 57:  Module Replacement Complete ........................................................................................... 153 
Figure 58: File Viewer Main ...................................................................................................................... 156 
Figure 59: File Viewer Setup ..................................................................................................................... 157 
Figure 60: Header Display ........................................................................................................................ 157 
Figure 61: Cursor Display ......................................................................................................................... 158 
Figure 62: Zooming in ............................................................................................................................... 160 
Figure 63: Zoomed in ................................................................................................................................. 161 
Figure 64: Console Parameters ................................................................................................................. 164 
Figure 65: Console Open ........................................................................................................................... 165 
Figure 66: Terminal Mode ......................................................................................................................... 176 
Figure 67: Adding Altus Emulation ........................................................................................................ 184 
Figure 68: Altus Emulation Parameters .................................................................................................. 185 
Figure 69: Secondary Port Warning ......................................................................................................... 186 
Figure 70: RockTalk De‐archive ............................................................................................................... 188 
Figure 71: RockTalk Archive .................................................................................................................... 188 
Figure 72: Trigger Levels ........................................................................................................................... 206 
Figure 73: Module Replace........................................................................................................................ 207 
Figure 74:  Module Replacement Options .............................................................................................. 208 
Figure 75: Replacing Trigger Module ...................................................................................................... 209 
Figure 76: Modules Replaced ................................................................................................................... 210 
Figure 77: Channel List .............................................................................................................................. 212 
Figure 78: Channel Mapping .................................................................................................................... 213 
Figure 79: Sending Parameters ................................................................................................................. 214 
Figure 80: Module Replace........................................................................................................................ 215 
Figure 81: Format Choices......................................................................................................................... 216 
Figure 82: Output Format Changed ........................................................................................................ 217 
Figure 83: Alarm Clock .............................................................................................................................. 219 
Figure 84: Network Trigger ...................................................................................................................... 221 
Figure 85: RockTalk Setup ........................................................................................................................ 224 
Figure 86: Renaming Unit in RockTalk ................................................................................................... 224 
  ix 
Figure 87:  Fast Setup ................................................................................................................................. 225 
Figure 88: Fast Setup (continued) ............................................................................................................ 226 
Figure 89: Select Layout to Build ............................................................................................................. 227 
Figure 90: Set Number of Channels ......................................................................................................... 227 
Figure 91: Set Virtual Channels ................................................................................................................ 227 
Figure 92: Set Data Source ......................................................................................................................... 228 
Figure 93: Set Trigger Filter ...................................................................................................................... 228 
Figure 94: Use Same Filter for all Channels ............................................................................................ 228 
Figure 95: Set Trigger Type ....................................................................................................................... 229 
Figure 96: Use Same Trigger for all Channels ........................................................................................ 229 
Figure 97: Set Output Type and Data Format ........................................................................................ 230 
Figure 98: Replace Web Server ................................................................................................................. 231 
Figure 99: Layout Complete ..................................................................................................................... 232 
Figure 100:  Adding Network Trigger ..................................................................................................... 233 
Figure 101: Adding RockGPIO ................................................................................................................. 235 
Figure 102: RockGPIO Parameters ........................................................................................................... 236 
Figure 103: E‐Mail File Sender .................................................................................................................. 249 
Figure 104:  FTP File Sender ..................................................................................................................... 250 
Figure 105: Admin Details ........................................................................................................................ 252 
Figure 106: Rock Recovery ........................................................................................................................ 254 
Figure 107: Run Time Log ......................................................................................................................... 256 
Figure 108: Error Log ................................................................................................................................. 257 
Figure 109: SOH Selection ......................................................................................................................... 258 
Figure 110: SOH Display Selection .......................................................................................................... 260 
Figure 111: SOH Graphs ............................................................................................................................ 261 
Figure 112: WinSCP ................................................................................................................................... 265 
Figure 113: WinSCP Key Confirmation .................................................................................................. 265 
Figure 114: WinSCP Panes ........................................................................................................................ 266 
Figure 115:  PuTTY ..................................................................................................................................... 267 
Figure 116:  PuTTY IP Address ................................................................................................................ 268 
Figure 117: PuTTY Window ..................................................................................................................... 269 
Figure 118:  FTP Server .............................................................................................................................. 270 
Figure 119: FTP Server Defaults ............................................................................................................... 271 
Figure 120: FTP Server Users .................................................................................................................... 271 
Figure 121: Adding Admin User .............................................................................................................. 272 
Figure 122: Admin User Password .......................................................................................................... 272 
Figure 123: Admin File Access ................................................................................................................. 272 
Figure 124: Add Data Path ........................................................................................................................ 273 
Figure 125: Permissions Set ....................................................................................................................... 273 
Figure 126: FTP Server Options ................................................................................................................ 274 
Figure 127: FTP Sender Parameters ......................................................................................................... 275 
Figure 128: Power Connector Pins and Schematic ................................................................................ 294 
Figure 129: Console Connector Pins ........................................................................................................ 296 
Figure 130: Console Connector Schematic .............................................................................................. 297 
Figure 131: Serial Connector Pins ............................................................................................................ 298 
Figure 132: Serial 1 Connector Schematic ............................................................................................... 298 
Figure 133: Serial 2 Connector Pins ......................................................................................................... 299 
  x 
Figure 134: Serial 2 Connector Schematic ............................................................................................... 300 
Figure 135: Serial 3 Connector Pins ......................................................................................................... 301 
Figure 136: Serial 3 Connector Schematic ............................................................................................... 301 
Figure 137: DFS Connector Pins ............................................................................................................... 302 
Figure 138: DFS Connector Schematic ..................................................................................................... 303 
Figure 139: PC 9‐Pin Connections ............................................................................................................ 303 
Figure 140: Modem Connector Pins ........................................................................................................ 304 
Figure 141:  Modem  Connector Schematic ............................................................................................ 304 
Figure 142:  Time I/O 1‐3 Connector Pins.  J25, J26, and J27 ................................................................ 306 
Figure 143: Time I/O Connector 1‐3 Schematics .................................................................................... 307 
Figure 144: USB Device Connector Pins.................................................................................................. 308 
Figure 145: USB Device Connector Schematic ....................................................................................... 308 
Figure 146: USB Host Connector Pins ..................................................................................................... 309 
Figure 147: USB Host Connector Schematics ......................................................................................... 310 
Figure 148: Ethernet Connector Pins ....................................................................................................... 315 
Figure 149: Ethernet Connector Schematic ............................................................................................. 315 
Figure 150: Ethernet Wiring Colors ......................................................................................................... 316 
Figure 151: Ethernet Cross Over Cable Pinouts ..................................................................................... 316 
Figure 152:  Ethernet RJ11 Wiring ............................................................................................................ 317 
Figure 153:  P/N 112297‐PL, Power and Battery .................................................................................... 318 
Figure 154:  P/N 112250‐PL, Optional Power Supply ........................................................................... 319 
Figure 155: P/N 112259‐PL, Optional Power Supply with Battery Connections ............................... 320 
Figure 156:  P/N 853607 Rock+ Serial Cable ........................................................................................... 321 
Figure 157:  P/N 853609 Rock+ Modem Cable........................................................................................ 322 
Figure 158:  P/N 853608 Rock+ Ethernet Cable ...................................................................................... 323 
Figure 159:  P/N 853611 Rock+ USB Device Cable ................................................................................. 324 
Figure 160:  P/N 853610 Rock+ USB Host Cable .................................................................................... 325 
Figure 161:  P/N 853612 Rock+ Time I/O Pigtail .................................................................................... 326 
 

  xi 
  xii 
LIST OF FIGURES LIST OF FIGURES

Safety 
 
These symbols may appear on Kinemetrics equipment or in this 
manual 
 
 
 

When you see this symbol, pay careful attention. Refer to the similarly 
marked, relevant part of this manual before servicing the instrument. 
 
This symbol means a low‐noise earth ground. The noted item should be 
grounded to ensure low‐noise operation, and to serve as a ground return for 
  EMI/RFI and transients. Such a ground does not work as a safety ground 
for protection against electrical shock! 
 
  This symbol means an alternating current (AC) power line. 
   
  This symbol means a direct current (DC) power line derived from an AC 
power line. 

This symbol indicates an electrostatic sensitive device (ESD), meaning that 
when handling the marked equipment you should observe all standard 
precautions for handling such devices. 
 
This symbol indicates that a particular step/process or procedure is required 
to ensure the installation maintains conformity to European Community 
requirements. 
 
This symbol indicates that this referenced equipment or material should be 
re‐cycled and not thrown in the normal trash stream. 
 
This symbol indicates that the step/process or equipment has an 
environmental consequence and steps such as recycling are required. 

 
LIST OF FIGURES LIST OF FIGURES

These safety‐related terms appear in this manual 
NOTE: Statements identify information that you should consider be‐ 
fore moving to the next instruction or choice. 
 
CAUTION:  Statements identify conditions or practices  that could 
result  in  damage  to  the  equipment,  the  software,  or  other  property. 

 
WARNING:  Statements identify conditions or practices  that  could 
result in personal injury or loss of life. 
 
Follow the precautions below to ensure your personal safety and prevent damage to the 
digitizer. The unit is powered by an 8‐16 VDC source or 15.5 VDC supplied by the optional 
power supply assembly, from external batteries or from a solar charging system. The unit 
can also contain charge an external Sealed Lead Acid Battery (SLA) that can power the unit 
when the external power sources fail. 
 
Optional  Power  Supply Assembly 
If you plan to power the digitizer from the mains supply, we recommend Kinemetrics’ Power 
Supply Assembly (PSA)(KMI 112251‐PL or KMI 112259‐PL for use with SLA Battery.) Plug 
the PSA’s power cord into AC outlets that will not apply more than 260 VRMS between 
the supply conductors or between either supply conductor and ground. A protective ground 
connection (provided through the grounding conductor in the PSA and its power cord) is 
essential for safe operation. The PSA is designed for indoor use only; it must not be subject 
to immersion in water, high humidity, or temperatures above 70°C. 
 
User‐Supplied Batteries or Charging System 
If you supply your own charging system, make sure the system provides the correct voltage 
and current required by the unit. If you derive power from the mains supply, make sure 
there is adequate grounding for all the equipment. If you supply your own batteries, follow 
the warnings below. 
 
External  Battery 
Follow the precautions in this manual when handling and replacing external batteries. Metal‐ 
lic instruments of any kind could short the battery terminals, resulting in fire or explosion. 
Do not drop the battery or attempt to disassemble it. When charging the battery, use a 
properly rated charger and do not overcharge the battery. The only correct replacement 
battery is a sealed lead‐acid battery with relief vents and ratings comparable to the original 
battery. Never try to use a non‐rechargeable battery with the unit. 


 
LIST OF FIGURES LIST OF FIGURES

Grounding the Digitizer 

When using the optional PSA to power the unit from the AC mains supply, remember that 
the unit is grounded through the PSA power cord. To avoid electric shock, plug the PSA 
cord into a properly wired receptacle where the protective earth ground has been verified. 
Do this verification before making any power connections to the unit. 
 
Use the Proper Power   Cord 

Use the power cord and connector supplied with optional PSA, or an equivalent IEC‐standard 
power cord. Be sure that it is in good condition. 
 
Antenna, Phone & LAN  Cabling 

Never install antenna, telephone, or LAN wiring during electrical storms.  Always ensure 
adequate separation between antenna cabling, telecom cabling, or LAN cabling and high 
voltage wiring. Always perform a safety check on telecom and LAN wiring to measure the 
voltage before working on the wiring. Remember telephone wiring carries fifty (50) to sixty 
(60) volts of DC and the ring signal at ninety (90) VAC can deliver a very uncomfortable 
shock. Power over Ethernet Cabling can carry DC voltages of up to 56VDC. To avoid electric 
shock, do not connect safety extra‐low voltage (SELV) circuits to telephone‐network voltage 
(TNV) circuits. Ethernet LAN ports contain SELV circuits, and some WAN ports contain 
TNV circuits. Some LAN and WAN ports both use RJ‐45 connectors. Use caution when 
connecting cables. 
 
Do Not Operate in Explosive Atmospheres 

The unit and the optional PSA provide no explosive protection from static discharges or 
arcing components. Do not operate the equipment in an atmosphere of explosive gases. 
 
The Kinemetrics’ Rock+ Digitizer is not To Be Used For Life Sup‐ 
port or Life‐Critical  Systems 

These products are not designed for operating life critical support systems and should not 
be used in applications where failure to perform can reasonably be expected to create a risk 
of harm to property or persons (including the risk of bodily injury and death) 


 
 
 
 
 

Chapter 1 

Introduction 

This manual describes the basic operation of the Kinemetrics Rock+ Digitizers, which include 
the Obsidian multichannel data acquisition systems. The System Overview gives a brief 
overview of the unit. This is followed by the Installation & Setup instructions which will 
allow you to get the unit working. The Basic Operation section show how to use the unit for 
most simple operations, and is followed by specific instructions on various common tasks, 
the Advanced Operation section shows how to perform more complex tasks. The Detailed 
System Description describes in detail the various sub‐systems of the digitizer, the various 
software components, and more. Finally, the Hardware Reference section explains items 
such as maintenance and wiring cables.  It contains detailed information on the hardware 
of the digitizer, as well as operational information on the digitizer’s software. See Further 
Information for additional software and hardware references.  The diagram below shows 
where you will find particular information. 


 
1.1. SYSTEM OVERVIEW CHAPTER 1. INTRODUCTION

1.1 System  Overview 


The Rock+ multi‐channel data acquisition systems consist of a chassis containing external 
connectors to allow the user to interface the system to voltage output sensors. An Obsidian 
can have 4 or 8 individual sensor channels depending on its configuration, and is available 
with an internal three channel accelerometer. A Multi‐Channel recorder can have 12, 20, 
24 or 36 individual sensor channels depending on its configuration. The signals from these 
sensors are digitized on individual four channel Analog to Digital Converter boards.  The 
signals are then filtered, processed by a high speed Digital Signal Processing system and 
passed over a high speed serial link to the systems main processor. This consists of a low 
power highly integrated processor running Linux and Kinemetrics Rockhound software, an 
Ethernet interface, three serial ports, fixed and removable storage devices, as well as high 
resolution sensor inputs. 
The Obsidian is a member of the larger Kinemetrics Rock+ product family, and as such may 
be used in conjunction with other Kinemetrics Rock products. 
The user interaction with the system is through the Rockhound software for set‐up, con‐ 
trol, and operation of the system, and through the Front Panel connectors for the physical 
connection of power, sensors, communication devices, and GPS to the system. 


 
CHAPTER 1. INTRODUCTION 1.1. SYSTEM OVERVIEW

1.1.1 How is the Rock+ Series Different from the Rock   Series? 
There are many similarities between the Rock and Rock+ series instruments, including: 

• Both run Linux and Rockhound 
• Both have GPS, Ethernet, and serial ports 
• Both include supercapacitors as an internal “UPS” 

Differences between them include: 

• Significant new features: 

◦ Standard PTP (IEEE‐1588) 
◦ USB Host ports for data offload or communications interfaces 
◦ 10/100  Ethernet 
◦ Low latency lacket support (0.1 seconds) 
◦ Standard DFS interface for extremely low data latency 
◦ GPS supporting Ground Based  Augmentation 
◦ Regulated sensor and COMM power 
◦ Input power 9‐28VDC vs 8‐16VDC 
◦ 1PPS input & output 
◦ Optional Power‐Over‐Ethernet 

• And then the techy stuff: 

◦ A later verion of Linux (3.12 vs 2.6) 
◦ Filesystem is different (EXT4 vs  EXT3) 
◦ Faster processor (1GHz vs 400Mhz) 
◦ More RAM (1GB vs 256MB) 
◦ Memory cards are different (4GB SDHC System & 32GB SDHC Data vs one CF 
card) 
 
1.1.2 What is  Rockhound? 
Rockhound is the software system used by the digitizer to acquire, process, store, and transfer 
data. Rockhound is very flexible allowing processing and manipulation of data in a variety 
of ways. 
By default, Rockhound is configured as a triggered event recorder with traditional Kine‐ 
metrics trigger methods, levels, and data formats.  You are free to reconfigure software to 


 
1.1. SYSTEM OVERVIEW CHAPTER 1. INTRODUCTION

not only change trigger settings, but triggering criteria, and output data formats. Further, 
the system can be set for applications such as continuous or timed recording or telemetry. 
This flexibility is achieved using software modules that exchange data in an output‐neutral 
format. These modules may be combined in many ways. You even have the option to write 
your own modules using the optional Software Development  Kit. 
See the section Further Information for additional software references. 
 
1.1.3 Front Panel 

The front panel consists of connectors allowing you to connect power, sensors, GPS, serial and 
Ethernet connections as well as other functions. In this section we describe the connectors 
and the cables and accessories associated with them. 
Connectors are provided for: 

• Sensor Inputs 
• Power 
• Console 
• Serial ports 
• USB device & USB hosts 
• Ethernet port 
• GPS Antenna 
• Time/IO 
• POTS Modem 
• DFS 

LEDs are provided for: 

• Power 
• Status 
• Event 
• Media 
• Ethernet Link 
• Ethernet Data 
 

 

CHAPTER 1. INTRODUCTION 1.1. SYSTEM OVERVIEW

A magnetic switch is provided to force the unit on and off in abnormal situations and a seal 
screw is provided for factory leak testing. 
 

Figure 1: Rock+ Digitizer 

 
1.1.3.1 Standard  Interfaces 

The standard interfaces (those provided with all units) include the following: 
 
Power: This connector provides access to external power input, and the external battery 
connection. It also provides a single digital input and a digital output for user use. 
The external power input allows provision of an external DC power source that is used both 
to operate the unit and charge a battery, if connected. The voltage should be limited to 9‐
28 VDC and we normally recommend 15.5 VDC input if the unit is to charge a back‐up 
battery, and 12 VDC if the customer is supplying the battery backup charging system. 
The external battery connection is used to connect a Valve Regulated Lead Acid (VRLA) 
battery that will be float charged by the system to provide operational autonomy in the case 
of temporary loss of external power. 
The digital input and output can be used to turn the system on and allow the unit to control 
other equipment depending on the program in the unit. 
 

 

1.1. SYSTEM OVERVIEW CHAPTER 1. INTRODUCTION

Kinemetrics supplies both cables assemblies and cable assemblies with wide input AC to DC 
power supplies the following are available: 
• Molded cable with Pigtails for user supplied DC power source and battery ‐ P/N 
112297‐PL, Power and Battery. 
• Wide Input AC Power Supply with US AC Cord and Molded Cable ‐ P/N 112250‐PL, 
Optional Power Supply 

NOTE:   P/N 112251‐PL does not include the US AC   Cord 

• Wide Input AC Power Supply with US AC Cord and Battery Connections ‐ P/N 
112259‐PL, Optional Power Supply with Battery Connections 

NOTE:  The  one  wire  interface  on  the  connector  is  for  use  in  future 
Kinemetrics‐supplied power options. 

For more information on the power connection refer to the Power Connector section. 
 
Console:   This connector provides access to the console port connection. 
The console port connection is used to provide access to an operating system console that is 
required in initial setup of the system (before network interfaces are defined) and in certain 
diagnostic and maintenance operations. The console port is not needed in normal operation. 
Typically, only a few console capable cables will be needed to support a large number of 
installed units. 
The console is /dev/console. The default baud rate is 115200. 
The cable to connect to the Console is P/N 853607. 
 
Serial:   This connector provides access to a general purpose serial port. 
The serial port is an RS‐232 serial port that can be used by user application software for a 
variety of purposes. The serial port supports full hardware handshaking. 
The serial ports are /dev/serial1, /dev/serial2, and /dev/serial3, the default baud rate is 
9600. 
The cable to connect to the Serial Port is P/N  853607. 
 
USB Device:     This connector provides access to the USB device interface. 
The USB device interface allows use of the unit as a USB device from a USB host (the host 
is typically something like a PC). This makes the USB device interface of the Rock+ appear 
as a virtual COM port on the PC. Compatible drivers will be required on the host end to 
utilize this  interface. 
From this port you can open a terminal session (using something like PuTTY) and log into 
the Rock+ without requiring a network conenction or an actual serial port on your computer. 
The cable to connect to the USB Device Port is P/N  853611. 

  9 
CHAPTER 1. INTRODUCTION 1.1. SYSTEM OVERVIEW

Ethernet: This connector provides a low power 10/100Mb Ethernet connection. The  cable 
to connect to this ports is P/N  853608. 
Ethernet is used for connection to other equipment, such as other digitizers and/or a hub or 
switch. 
With the POE (Power Over Ethernet) option installed it can also allow the digitizer to act 
as a class 1 powered device on a POE LAN. 
For additional information on the Ethernet Interface connections consult the Ethernet Con‐ 
nector  section. 
 
POTS  Modem:  This  connector  provides  a  connection  to  an  internal  POTS  (Plain  Old 
Telephone System) Modem. This modem can be used by user application software for a 
variety of purposes and is fully supported by the Rockhound application. This connects to 
Kinemetrics cable P/N 853609. 
 
GPS:  This  connector  provides  the  connection  to  an  active  3.3V  GPS  Antenna  to  allow 
reception of GPS timing and position signals. These antennas contain low noise amplifiers 
which are powered by a DC current through the antenna cable. The following cable options 
may be supplied with your unit: 

• No GPS antenna or cable 
• Trimble MiniMag 3V antenna with 5 meter cable (P/N 110905‐PL, GPS 5 Meter, 
MiniMag) 

• Trimble Bullet 3V antenna with: 
• 25 meter plenum cable (P/N 111095‐01‐25‐PL, GPS Bullet Antenna 25m Cable) 
• 50 meter plenum cable (P/N 111095‐01‐50‐PL) 
• 25 meter UV‐resistant cable (P/N 111095‐02‐25‐PL) 
• 50 meter UV‐resistant cable (P/N 111095‐02‐50‐PL) 

For additional information on the Antenna and Cabling consult the GPS Antenna Connec‐ 
tions  section. 
 
 

  10 
1.1. SYSTEM OVERVIEW CHAPTER 1. INTRODUCTION

LEDs   The LEDs on the front panel provide the following information: 

Power: 

• OFF ‐ No power 
• Steady Green ‐ Waiting to Power up, Running off of external power or POE (Power 
Over Ethernet) 

• Flashing Green ‐ The system is starting up 
• Infrequent Green ‐ Running off of battery 
• Orange ‐ Supercapacitor is being charged 
• Flashing Red ‐ Power system fault 
• Alternating Red and Green ‐ The system is shutting down 
 
Status: 

• Flashing Red ‐ No time source 
• Infrequent Red ‐ Time source detected, not locked 
• Infrequent Green ‐ Good time quality 
• Flashing Green ‐ The system is starting up 
• Steady Green ‐ Rockhound loading or restarting 
 
Event: 

• OFF ‐ No events 
• Steady Green ‐ Real time data stream 
• Flashing Green ‐ Storing an event 
• Infrequent Green ‐ Events stored 
 
Media: 

• OFF ‐ Idle, OK to insert or remove 
• Steady Green ‐ Mounting 
• Flashing Green ‐ Transferring data 
• Steady Red ‐ Error detected 

  11 
CHAPTER 1. INTRODUCTION 1.1. SYSTEM OVERVIEW

Ethernet Link (Green/Amber): 

• Amber ‐ Ethernet 10Mb link detected 
• Green ‐ Ethernet 100Mb link detected 
• OFF ‐ No Ethernet link detected 
 
Ethernet Data (Amber): 

• ON ‐ Ethernet data transmission in progress 
• OFF ‐  Idle 
 
Magnetic Switch The magnetic switch is located above and to the right of the Modem 
connector. The magnetic switch is located directly behind the seal screw. The switch is used 
to start up and shut down the unit without use of a physical power switch. 
The unit has no power switch. When power is connected, the unit’s default behavior is to 
power up and begin operation.  When power is removed, the applications and operating 
system will shut down and the unit will turn off. 
Briefly touching a strong magnet near the magnetic switch location allows the user to start 
and stop the unit. This is useful if say a time window has been set incorrectly and the unit 
needs to be turned on to correct the problem. 
 
Seal Screw  The small screw located in the lower right of the front panel is the seal screw. It 
is used at the factory for leak testing. 
 
CAUTION: This screw should NOT be removed by the user as doing  \so 
may expose the internals of the unit to the environment and damage it. 

 
1.1.3.2 Sensor Interfaces 

In Rock+ Digitizers each four channel ADC board connects to a three channel and a single 
fourth channel connector on the Front of the unit. The military style connectors can be 
used to directly to connect to sensors or a four channel Terminal board can be plugged into 
the two connectors to give a four channel terminal board interface which allows easy field 
connections of sensors without soldering. 

  12 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

1.1.3.3 External Three Channel Sensor Interface 

This connector has the same pinout as used in the  Q330 unit. The connector has three 
high resolution input channels, three mass position/auxiliary 12 bit resolution inputs, a 
calibration output signal and its associated control, three other outputs for sensor control, a 
switched system power output, and an interface to allow a one wire device embedded in the 
sensor to be read. 
Kinemetrics supplies cables for connecting a large variety of sensors or you can build your 
own sensor cables. More information on the cables and connector interface is in External 
Three Channel Sensor Interface section. 
 
1.1.3.4 Internal Accelerometer  Deck 

The Obsidian Accelerograph unit contains a tri‐axial EpiSensor deck connected to channels 
1‐3 of the ADC. The deck provides three orthogonal 2g (Optionally 4g) Full Scale sensors 
mounted internally to the unit. This unit is equipped with a single point mounting scheme 
and must be leveled and securely mounted to ensure high fidelity string motion recording. 
The fourth channel is available as an additional input to the system. 
 
1.1.3.5 External 4th Channel Single Input Sensor Connector 

This connector provides a single high resolution input to the fourth channel of the ADC that 
can be used for a uniaxial active or passive sensor. It also provides a separately switched 
calibration output, a power output, and four additional one wire interface connections. 
Kinemetrics supplies cables for connecting a large variety of sensors or you can build your 
own sensor cables. More information on the cables and connector interface is in External 
4th Channel Single Input Sensor Connector section. 
 
1.1.3.6 Four Channel Terminal Board (KMI  111915) 

This board is available to support both dual supply (‐01) and single supply sensors (‐02). It 
provides a terminal block to allow easy connection of uniaxial, biaxial, or triaxial sensors. 
For details on the connection locations on the terminal blocks and other information refer 
to the Four Channel Terminal Board (KMI 111915) section. 
 
1.2 Installation &  Setup 
To successfully deploy your digitizer, you will need to physically install the hardware in 
a suitable environment, install the sensors you plan to use with their associated cabling 
and connect them to the unit, provide your primary and backup power sources (if any), 
connect any communications links, install the GPS antenna can connect the cable to unit, 

  13 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

and configure the software for correct operation if your requirements differ from the default 
factory configuration of the unit. 
If you have not already done so, we recommend that you install the software from the Rock 
Support Software CD (300654‐PL) now so that you will have the necessary utilities such as 
terminal emulation, telnet client, and secure file transfer programs. You are of course free 
to use other software that you are more familiar with that provides the same capabilities. 
The basic steps to install your digitizer are as follows: 

• Secure the unit physically in position 
• Attach the grounding stud 
• Mount the Sensors as appropriate and connect their cables 
• Mount the GPS Antenna 
• Connect the Ethernet LAN 
• Connect the GPS Antenna 
• Connect other communication  interfaces 
• Connect the  sensors 
• Connect the optional SLA (sealed lead‐acid) battery 
• Connect the DC power source 
• Connect a laptop or equivalent to the console port (See Software Installation ) 
• Change the default system passwords (See Passwords ) 
• Configure the software: 

◦ Configure the LAN (See Network Configuration ) 
◦ Configure desired IP services (See IP Services ) 
◦ Configure extended storage (See USB Thumb Drives ) 
◦ Configure Rockhound (See Basic Setup) 
◦ Configure GPS, PTP, or NTP (See Timing System  ) 

At  this  point,  the  system  should  be  functional.  See  the  section  on  Basic  Operation  for 
further details on initial configuration of the system. Note that several of the configuration 
and update steps listed above can be tested in the lab before the unit is deployed. 

  14 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION
Figure 2:  Rock+ Quick Setup Guide 
  15 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

1.2.1 Operating Environment 

The digitizer needs to be installed in a location that provides the following environmental 
conditions. 
 
1.2.1.1 Operating  Temperatures 

The digitizer’s operating temperature range with the standard options is: 
Standard digitizer ‐20 to +60 OC Operation (optional ‐30 to  +70ºC) 
This can be limited by user installed equipment. Replacing the storage cards with commodity 
cards can reduce the operating temperature range. 
The digitizer is configured to only charge an attached Sealed Lead Acid (SLA) battery from: 
Battery Charge ‐0.0 to + 40.0  ºC 
The unit will draw power from an SLA battery over the full operating temperature range 
but will show a fault if it runs outside of the range: 
Battery Normal Discharge ‐ 15.0 to + 50.0 ºC 
Finally units with the optional storage module that contains a hard drive have an operating 
range for the hard drive of: 
Optional Hard Drive 0.0 to + 50.0 ºC 
The unit should not be placed where it is exposed to direct sunlight and the external battery 
of used should be located in the same temperature environment as the unit and should again 
not be exposed to direct sunlight. 
 
1.2.1.2 Operating Humidity 

The case of the unit is designed to meet the requirements of a NEMA 6P enclosure (equiv‐ 
alent to IP67). The system can operate in humidity levels of up to 100% and withstand 
occasional temporary immersion in water up to 2 meters in depth. The system should not 
be continuously immersed, as galvanic corrosion of the connectors will occur, potentially 
destroying the system. To ensure operation in high humidity, the desiccant packet must be 
fresh and the case of the unit should be carefully re‐sealed if it is opened. The unit should 
be protected from rain and snow and should not be allowed to stand in water for longer than 
one hour. 
 
1.2.1.3 Operating  Altitude 

The unit can operate at altitudes from ‐300 to +10,000 meters. 

  16 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

1.2.2 Unpacking & Inspecting the  Unit 

Before accepting the shipment the shipping carton should be examined for any obvious 
damage and this should be recorded by the freight carrier. 
The digitizer ships in a custom designed carton. This carton can be used to return the unit 
or to ship it to other destinations. It should be carefully opened at the top so it can be re‐
used. 
On top of the unit is an additional carton that will contain any accessories ordered with the 
unit such as connectors, cables, AC/DC power supply. Please check the contents of this box 
against the packing list. 
The digitizer is beneath this box packed in custom foam inserts with the front face connectors 
facing upwards. Carefully pull the unit and the inserts from the box. The unit should have 
no signs of external damage. 
The unit is then ready for installation. 
See Practice Assembly 
 
 
When the packaging is no longer required please recycle the cardboard cartons and 
foam insert  appropriately. 
 
1.2.3 Requirements for Installation 

Below we provide lists of the tools, supplies and equipment required to install a Rock+ 
digitizer in a typical configuration. 
Specialized installations may require additional tools, supplies or equipment, depending on 
specific sites and needs.  This manual assumes that all civil engineering works (concrete pads, 
enclosures, conduit, mounting masts, etc.) are complete and ready before installation. 
 
1.2.3.1 Required Tools 

For a permanent installation the unit should be mounted to the floor, wall, or shelf in the 
structure. 
The following tools are suggested: 

• A drill capable of drilling into the attachment surface. 
• Appropriate drill bits for the attachment surface. 
• Suitable tools to install the screws/anchors into the attachment surface. 

  17 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

1.2.3.2 Optional Tools 

Communication options, GPS systems, and other advanced installations require additional 
tools: 

• Wrench or pliers to tighten the antenna connectors 
• Tools for mounting the antenna 
• A heat gun (electric or butane) 
• Soldering iron (electric or butane) 
• Small screwdriver 
• Wire cutters 
• Wire strippers 
• Long‐nose  pliers 
• Utility knife 
• A drill (electric or battery powered) 
• Cable tie wraps 
• A crimping tool 
• A short length of insulated braid 
• An extension cord or a small generator for AC power, if butane‐powered soldering irons 
and battery‐powered drills are not  available 
 
1.2.3.3 Required Supplies 

• Mounting hardware, screws, nuts, washers, concrete anchors studs etc (Dependent on 
selected mounting method) 

• Material to make grounding straps for the unit 
• Solder with rosin‐core flux 
• An assortment of heat‐shrink tubing, cable tie‐wraps, and electrical tape 

  18 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

1.2.3.4 Required Equipment 

• A computer running Windows, Linux, or other suitable operating System 
• A serial port on the computer in order to access the system console. This may be an 
internal serial port or one provided through a USB to serial adapter. 

• An Ethernet network interface if you intend to use the networking capabilities 
• Kinemetrics’ Console cable (853607) to link a laptop to the unit 
 
1.2.3.5 Optional Equipment 

• A battery‐powered digital volt meter (DVM) for system‐checkout   functions 
• A portable printer, to print tests, configurations, and other commissioning data 
• A digital camera, to photograph the completed installations 
 
1.2.4 Practice  Assembly 

Once you have assembled the tools, supplies, and equipment listed above, we recommend 
that you run through a practice assembly following the installation instructions. 
Why Practice in the Laboratory/Office? 
The connections between all the components mentioned in this manual may appear a bit 
complex. They will be even more complex if your first installation attempt is in a remote 
field situation where you find that you lack the necessary tools, supplies, or equipment to 
make the connections work. 
Practice in a well‐supplied, well‐lit laboratory or office when first connecting and setting up 
the unit and ancillary equipment. Follow these instructions carefully, step by step, to learn 
exactly which tools, supplies and equipment will be needed in the field. 
Murphy’s Law provides a further reason for a practice installation: 
If anything can go wrong, it will; and Isaac’s Corollary: Murphy was an optimist!
See Unpacking & Inspecting the Unit 
 
1.2.5 Planning your installation 

This section discusses recording network setup and operating modes. Networking capabilities 
are an important part of this equipment, and should be considered carefully to make sure 
that adequate remote access and bandwidth are provided to utilize the real time capabilities 
of the digitizer. 

  19 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

1.2.5.1 Network Planning 

Carefully consider the scientific objectives of the installation when planning the network, 
whether it includes one station or a hundred. 
The station location, type and position of sensors, and instrument settings all affect the type 
of data recorded. Consider local seismic‐noise conditions and the anticipated amplitude of 
events being recorded. This will help you correctly set trigger parameters, estimate the 
quantity of data expected, and decide how to retrieve the data. 
It’s also necessary to plan how to analyze, combine, and archive data, as well as how to service 
and maintain the network. Finally, consider how the network will function after a large event, 
when mains power and telecommunications might not be available for a considerable time. 
Under such circumstances, how do you plan to retrieve and process the network’s data, as 
well as continue operating it? 
 
1.2.5.2 Civil Engineering 

Before installing the digitizer, plan and construct (if necessary) the housing that will provide 
a protective infrastructure for the unit. The exact details of the installation depend on local 
conditions, local regulations, and the purposes of the installation. 
Except in cases of a rapid emergency deployment of seismic instruments, the digitizer should 
be housed in a protective structure. Below are the two typical types of installation settings 
and related protective structures. They can be used as rough guidelines for an installation. 
 
1.2.5.3 Free‐Field Installations 

In a free‐field accelerograph installation, the digitizer is installed some distance from buildings 
in a ʺfree fieldʺ and sheltered by a small, lightweight structure that allows the sensor to sense 
acceleration as close as possible to the ʺtrueʺ accelerations of ambient ground motion. In 
softer ground sites, because of the soil‐structure interactions during earthquakes, a heavier‐ 
than‐necessary protective structure could degrade data accuracy. 
The structure should also protect the digitizer from weather, direct sunlight, and theft or 
vandalism. A ʺtransformer hutʺ made of fiberglass and stainless steel hardware is ideal as 
long as it, and the digitizer, are attached to a poured and reinforced concrete pad. If true 
hard‐rock site response is desired, anchor this concrete pad to bedrock. 
If you are using the digitizer with broadband sensors, it must be placed in a seismic vault 
with a pier on which to mount the sensors. This vault will need to not only protect the 
instrument, as in the case of the free‐field accelerograph, but also must provide a stable 
mounting structure and insulate the sensors from ambient temperature variations. It should 
be a low‐noise site for optimal operation. Masts for antennas, GPS, wireless communication 
and solar cells should be mounted some distance from the sensor vault to prevent wind‐ 
induced noise from contaminating the data. Kinemetrics has several application notes that 
can help you install a broadband sensor network. 

  20 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

Provide the digitizer with a good earth ground. Proper grounding depends greatly on the 
humidity of the soil at the site. For average‐humidity soil, an effective earth ground can 
be made by wiring the case grounding stud to a 6’‐ to 8’‐long copper rod embedded in the 
ground. 
If no AC power is available, a solar charging system is required. Refer to the Advanced 
Installations section of this manual for more information. 
If the digitizer has a GPS system, the GPS antenna will need a suitable mounting mast. A 
telephone line, DSL line, or other communication link is required to communicate with the 
digitizer remotely. 
 
1.2.5.4 Structural‐Monitoring Installations 

In a structural‐monitoring installation, you provide protection to the digitizer by installing 
it within an existing building or structure (a bridge, a dam, a high‐rise, etc.). The main 
purpose for installing the digitizer in an existing structure is to measure and monitor the 
structure’s vibrations in response to ground motion. While some use such installations to 
calculate measurements of ʺfree‐fieldʺ seismic motions, the very nature of the structure’s 
size and foundation depth cause the acceleration measurements to deviate considerably from 
ʺtrue free‐fieldʺ response. 
For a structural‐monitoring installation, make sure the space in the structure allows enough 
room to mount and service the digitizer, and that the space provides enough protection 
so the digitizer and its sensors will not be disturbed or vandalized. Powering the digitizer 
requires a mains supply close to the installation point. To connect a modem, a phone line is 
also necessary. 
As for the digitizer’s optional GPS antenna, carefully plan to locate the antenna close enough 
to the digitizer so the supplied GPS cable will reach between the two. 
To use multiple sensors mounted around the structure, make a detailed plan of the sensors’ 
locations and plan how to run the cables between them and the digitizer. To interconnect 
digitizers, plan the layout of the interconnecting cables. If possible, run all the cables (sensor 
and interconnect) through grounded metal  conduit. 
All these elements of an installation should be in place before the digitizer itself is installed. 
 
1.2.6 Installing the Digitizer 

The unit should be installed preferably in a dry environment protected from direct sunlight 
and exposure to standing moisture. The temperature should be within the operating limits 
given earlier. The unit should also be protected from animals that may eat cabling and from 
vandalism. If an external battery is used it is important it is mounted close to the unit and 
exposed to the same temperature conditions so the temperature adjusted charge voltage is 
correct. As you perform the installation pay particular attention to the warning below. 

  21 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

WARNING:  Antenna,  Phone,  &  LAN  Cabling.  Never  install  antenna, 


telephone,  or  LAN  wiring  during  electrical  storms.  Always  ensure 
adequate separation between antenna cabling, telecom cabling, or LAN 
cabling  and  high  voltage  wiring.  Always  perform  a  safety  check  on 
telecom and LAN wiring to measure the voltage before working on   the 
wiring. Remember telephone wiring carries fifty (50) to sixty (60) volts 
of  DC  and  the  ring  signal    at    ninety    (90)    VAC  can    deliver    a    very 
uncomfortable  shock.  Power  over  Ethernet  Cabling  can  carry  DC 
voltages of up to 56VDC. To avoid electric shock, do not connect safety 
extra‐low  voltage  (SELV) circuits  to  telephone‐network  voltage  (TNV) 
circuits. Ethernet LAN ports contain SELV circuits, and some WAN ports 
contain  TNV  circuits.    Some  LAN  and  WAN  ports  both      use  RJ‐45 
connectors.  Use caution when connecting cables. 

 
1.2.6.1 Mount the Unit 

The unit can be mounted on the: 

• Floor 
• Shelf 
• Wall 
 
Floor‐Mounting: The unit should be placed on a dry floor. It is recommended that the 
unit be anchored to the floor using concrete anchors. These can be used to mount the unit 
using the slots on the front and rear of the units. The optional external battery should be 
mounted in close proximity and can be held with straps anchored by concrete anchors. 
 
Shelf‐Mounting 

1. Temporarily place the unit in the correct position. Mark the location of the slots on 
the shelf. 

2. Drill a hole with a 1/4ʺ (6.4mm) drill through the shelf at three or more locations. 

3. Insert 1/4‐20 bolts with washers through the slots from the top and secure underneath 
the shelf with washers and nuts. 
 
The optional battery can be mounted on the shelf if it is sufficiently sturdy or on the floor 
below the unit. 

  22 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

Wall‐Mounting 
1. Use a piece of 1” aluminum angle bracket sturdy enough to hold the unit. Attach the 
bracket to the unit using the holes on the rear rail. 
2. Then attach the bracket to the wall using suitable heavy duty anchors. 
 
Mounting  the  Obsidian  Accelerograph  For  Obsidians  with  an  internal  deck,  the  unit 
must be securely coupled to the ground to accurately record ground motion. 
The  unit  has  a  single  hole  through  the  unit  for  a  ¼‐20  8”  threaded  stud.  To attach  the 
Obsidian  to  the  mounting  surface,  use  the  mounting  kit  shipped  with  the  recorder.  It 
includes a heavy‐duty wedge type expansion anchor stud with 1/4‐20 thread, a tool to set 
the anchor, a flat washer, and a nut. The following figures show the details of such an 
installation. 
CAUTION: Invalid data. Obsidians with internal EpiSensor decks must 
be securely floor‐mounted to ensure the acceleration levels of the actual 
structure are measured. 
Anchor the recorder unit to a concrete floor if possible. Be sure that each of the recorder’s 
leveling feet are screwed into place and extending less than ¼” from the bottom of the unit. 
The locking nuts should be loosened. 
1. Move the recorder gently to one side. 
2. Locate where you plan to put the recorder, keeping in mind necessary working space. 
With the recorder mounted with connectors facing forward, you should allow 10” 
(254mm) clearance to the rear, front, and left. You should allow 15” (375mm) clearance 
to the right. 
3. Use a drill with a 3/8ʺ (9.5 mm) bit, and drill into the concrete to a depth slightly 
deeper than the length of the anchor (1” or 25.4mm). A percussion or hammer drill 
will make this procedure much easier. You should follow all recommended safety pre‐ 
cautions when using power tools and we recommend you wear safety glasses during 
the installation procedure. (Using the anchor set tool as a depth gage can make this 
easier. Mark the thick end with a piece of tape at the depth of the stud – check the 
hole depth with this.) 
 

Figure 3: Drilling Mounting Hole 

  23 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

4. Make sure you clean out the hole after drilling it. (Using a flexible drinking straw 
works very well for this, but make sure you don’t blow dust in your eyes!) 

5. Set the anchor using the anchor set tool. Place the anchor in the hole. The top should 
be flush with the floor. Place the thin end of the anchor set vertically into the anchor. 
Strike with a hammer until the shoulder of the anchor set tool is flush with the top of 
the anchor. 
 

Figure 4: Installing Anchor 

 
6. Screw the 8” stud into the anchor. 

  24 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

CAUTION: The 8” stud is a potential safety hazard, so do not leave it 
sticking up in the air! 
 
Figure 5: Anchor Stud Installed 

 
7. Center the Obsidian over the stud and lower its mounting hole down over the protruding 
stud. 

8. Place the flat washer onto the stud. Then screw the nut onto the stud without tight‐ 
ening. 

9. Make sure the Obsidian is oriented in the correct direction before you proceed with 
Step 11. 

10. Look at the bubble level window while you carefully adjust each of the Obsidian’s three 
leveling feet in or out.  When you can see the air bubble centered in the bubble level 
window with all three leveling feet resting on the mounting pad, the recorder is level. 

11. Tighten the locking nuts on the leveling feet. 

12. Carefully maintain both the Obsidian’s level and its orientation as you tighten down 
the anchor nut to hold the recorder firmly in place. Use a torque of 80‐100 inch‐lbs 
(9.2‐10.2 Nm) to tighten the nut. 

  25 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

Figure 6: Mounted to the Floor 

The unit is now firmly attached to the ground. 
Grounding the  Unit 
All users should complete this procedure. You must provide the digitizer with a good, 
low‐impedance earth ground before operating it for the following reasons: 
 
• To shunt ESD transients, lightning‐induced transients and EMI/RFI transients to 
ground. 

• To meet the requirements of the European Community’s EMI/RFI directives. 

Determine what earth ground you will connect the digitizer to. A good earth ground includes 
the following: 

• a metal plumbing pipe that is eventually buried in the ground 
• a copper ground rod staked in soil 
• a well engineered electrical grounding system, or 
• steel reinforcing rods that protrude from a concrete foundation. 

  26 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

Prepare the conductor you plan to use to connect the unit to the earth ground. For the 
conductor, you should at least use a heavy‐gauge wire or, better yet, a copper strap or 
copper braid. 
Connect this conductor to the unit’s case grounding bolt on the left center of the front rail. 
Then connect the other end of the conductor/grounding strap to the selected earth ground. 
 
NOTE:  If  the  digitizer  is  powered  by  the  optional  PSA,  the  third  pin 
(earth connection) of the PSA AC plug provides the safety ground.     To 
ensure  the  unit’s  low  noise  performance,  you  must  still  connect  the 
digitizer’s grounding stud to a good earth ground as described    above. 
 
Mount  the  Sensors  The  sensors  that  will  be  attached  to  the  unit  should  be  mounted 
according to the manufacturer’s instructions. Systems equipped with Kinemetrics sensors 
will be provided with the sensor operating manual which will describe how the sensor should 
be mounted. The cable from the sensor to the digitizer should be installed at this time. 
 
Mount the GPS Antenna (when using the internal GPS)   The GPS Antenna should   be 
mounted following the directions below: 
Mounting the MiniMag  Antenna 
The MiniMag Antenna (P/N 110905‐PL, GPS 5 Meter, MiniMag) should only be used for 
extended testing in the lab or brief testing in the field, as it is not very environmentally 
robust. It can be used for long periods when protected in a fiberglass hut, when no external 
antenna mast is desired. Mount the antenna by placing it on a flat surface within 5 meters 
of the Rock+ Digitizer. Make sure the antenna has a good view of the sky, without any 
obstruction from large buildings or  trees. 
Mounting a Bullet Antenna 
The bullet antenna is the preferred antenna for the majority of installations. The antenna 
has a 3/4ʺ NPT adapter for use with standard pipe fittings. 

  27 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

Figure 7: Bullet Antenna 

  28 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

Figure 8: Bullet Antenna Connector 

 
To mount the antenna, run the antenna cable down through the pipe. Plan the installation 
so that the cable from the antenna to the Rock+ Digitizer no longer than the Cable length 
(25 or 50 meters). Like the MiniMag antenna, the bullet antenna requires a good view of 
the sky, without obstruction from large buildings or trees. 
The bullet antenna comes screwed onto one end of a 3/4ʺ‐diameter NPT adapter pipe (which 
threads into standard pipe fittings). Within the adapter, the antenna is joined to a TNC 
connector on one end of a 25‐meter or 50‐meter coaxial cable. As shown in the bullet antenna 
illustration, the other end of the cable has a BNC connector. When the antenna’s mounting 
post is installed with a female NPT adapter at the end, the NPT adapter should screw 
snugly into the female adapter. 
 
CAUTION: The 3V Bullet Antennas supplied with the Rock+ and Q330 
Products should not be used with the Altus Product line as this uses 5V 
Bullet Antennas and the antennas can be damaged. Using an Altus 5V 
antenna  and  cable  with  a  Rock+  digitizer  will  result  in  poor  GPS 
performance  and  possible  loss  of  timing.  Kinemetrics  5V  Bullet 
Antennas  use  a  Type  F  connector  in  the  Antenna  rather  that  the  TNC 
connector illustrated above. 
If  the  GPS  system  will  operate  in  an  area  at  high  risk  for  lightning 
strikes,  consider  installing  a  lightning  protector  on  the  GPS  antenna. 
Kinemetrics  offers  a  lightning  protector  device  (P/N  109457‐PL).  See 
Section Installing Optional GPS Lightning   Protection 

 
Connect the GPS Antenna The GPS antenna Cable can now be connected to the Digitizer 
using the BNC to TNC adapter on the Front panel. 

  29 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

Connect  the  GPS  Engine/Antenna  TheGPS  engine/antenna  Cable  can  now  be  con‐ 
nected to the digitizer. Run the longer cable assembly from the P/N 111859 cables kit to 
the digitizer at the connector marked EXT GPS. Connect the other end to the RJ‐45 end 
CAT‐5 cable that was previously connected to the GPS engine/antenna. 
 
Configure the Timing Source The Timing Source of the digitizer defaults to the internal 
GPS. 
After making the selection, remember to save and apply the changes. 
 
Connect the Ethernet Connection The digitizer can now be connected to the Local  Area 
Network using the Ethernet Cable. The RJ45 plug from this cable should be inserted into 
a suitable Ethernet switch to make the connection. As the digitizer is a 10 Base‐T device 
it is better to connect to a switch than to a hub. The connection can also be made to a Router 
or other Ethernet based communication device. If directly connecting to a Laptop or PC a 
crossover adapter will be required. 
 
Connecting  the  Optional  Communication  Interfaces  The  optional  communication 
options can now be connected  appropriately. 
 
Connecting the Sensors to the Digitizer Rock+ Digitizers are designed to accommodate 
Kinemetrics Strong Motion Sensors, Broadband Sensors, and a variety of other sensors. This 
section  provides  details  on  connecting  the  sensors,  the  electrical  characteristics  of  the 
digitizer inputs, and the use of the calibration, control, and power available on the sensor 
connectors. 
Order of Sensor Connections 
The units are designed to allow you to easily connect to different sensors and to reconfigure 
the unit for different task by changing cables and external terminal boards. Kinemetrics 
provides a variety of cabling solutions for both new installations but also to allow retrofitting 
older installations with the capability of the Rock+ Digitizer. The sections below explain 
how to connect various units to the different digitizer applications. 
Physical to Virtual Channel Mapping 
To make life easier the Rock+ recorders can map their physical channels to virtual channels 
to allow easier grouping of the sensors. This allows the order the channels are recorded and 
displayed to be more logical for your application. So we can group all the Tri‐axial sensors 
together for example. In the tables below we give the physical connection to the Rock+ 
digitizer and our suggested virtual channel mapping. 
Triaxial Connection Convention 
To ensure compatibility with the Q330 and other weak motion recording we are using the 
Z (Up) N (North) E (East) (ZNE) convention when connecting tri‐axial seismometers or 
accelerometers to Rock+ Instruments.  If you build your own cables or use the terminal 

  30 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

blocks you can obviously chose another convention. However, as the Rock+ instruments 
can map physical channels to any virtual channel we would suggest you adopt this physical 
wiring convention and use mapping to get the order of channels you prefer. 
Obsidian 4X Connections or a 4 Channel Digitizer    Block 
If you have a Obsidian 4X or a block of 4 channels left on a Multi‐Channel recorder you can 
connect the following combinations: 

• 4 Uniaxial Sensors 
• 1 Triaxial & 1 Uniaxial Sensor 

As shown in the table below: 
Rock+  Physical Channel  4 Uniaxial Sensors  1 Triaxial, 1  Uniaxial 
   Sensors  &  Virtual  Channel Map  Sensors  &  Virtual  Channel Map 
  1  U1  V1  T1Z  V1 
  2  U2  V2  T1N  V2 
3  U3  V3  T1E  V3 
  4  U4  V4  U1  V4 
In the table “U” stands for a Uniaxial sensor, while TiZ, TiN, TiE are the three components 
of the “ith” Triaxial sensor. 
Obsidian 8X Connections or an 8 Channel Digitizer    Block 
If you have an Obsidian 8X or a block of 8 channels left on a Multi‐Channel recorder you 
can connect the following combinations: 

• 8 Uniaxial Sensors 
• 1 Triaxial & 5 Uniaxial Sensors 
• 2 Triaxial & 2 Uniaxial Sensors 

As shown in the table below: 
Rock+  Physical Channel  8 Uniaxial Sensors  1 Triaxial, 5  Uniaxial  2 Triaxial, 2  Uniaxial 
  Sens/Virt  Sens/Virt  Sens/Virt 

1  U1  V1  T1Z  V1  T1Z  V1 


2  U2  V2  T1N  V2  T1N  V2 
3  U3  V3  T1E  V3  T1E  V3 
4  U4  V4  U1  V4  U1  V7 
5  U5  V5  U2  V5  T2Z  V4 
6  U6  V6  U3  V6  T2N  V5 
7  U7  V7  U4  V7  T2E  V6 
8  U8  V8  U5  V8  U2  V8 

  31 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

In the table “U” stands for a Uniaxial sensor, while TiZ, TiN, TiE are the three components 
of the “ith” Triaxial sensor. 
Multi‐Channel recorder 12 Channel Block   connections 
To connect your sensors you need to think of the recorder as divided into blocks of 12 channels 
as we have designed the cabling and terminal boards to work as 12 channel groups. Thus a 
12 Channel Multi‐Channel recorder has one block, a 24 Channel Multi‐Channel recorder two 
blocks, and a 36 Channel Multi‐Channel recorder three blocks. If you have a non‐standard 
Multi‐Channel recorder you will have either an 8 channel block or a 4 channel block left 
over. These channels can be connected as shown above. 
The table below shows how we can connect the following selection of sensors to a 12 channel 
block: 

• 12 Uniaxial Sensors 
• 1 Triaxial and 9 Uniaxial sensors 
• 2 Triaxial and 6 Uniaxial sensors 
• 3 Triaxial and 3 Uniaxial sensors OR 
• 4 Triaxial Sensors 
 
Physical  Channel  12 Uniaxial  1 Triax, 9 Unis  2 Triax,6 Unis  3 Triax,3 Unis  4 Triax 
  Sens/Virt  Sens/Virt  Sens/Virt  Sens/Virt  Sens/Virt 

1  U1  V1  T1Z  V1  T1Z  V1  T1Z  V1  T1Z  V1 
2  U2  V2  T1N  V2  T1N  V2  T1N  V2  T1N  V2 
3  U3  V3  T1E  V3  T1E  V3  T1E  V3  T1E  V3 
4  U4  V4  U1  V4  U1  V4  U1  V10  T2Z  V10 
5  U5  V5  U2  V5  T2Z  V5  T2Z  V4  T2N  V4 
6  U6  V6  U3  V6  T2N  V6  T2N  V5  T2E  V5 
7  U7  V7  U4  V7  T2E  V7  T2E  V6  T3Z  V6 
8  U8  V8  U5  V8  U2  V8  U2  V11  T3N  V11 
9  U9  V9  U6  V9  U3  V9  T3Z  V7  T3E  V7 
10  U10  V10  U7  V10  U4  V10  T3N  V8  T4Z  V8 
11  U11  V11  U8  V11  U5  V11  T3E  V9  T4N  V9 
12  U12  V12  U9  V12  U6  V12  U3  V12  T4E  V12 
In the table “U” stands for a Uniaxial sensor, while TiZ, TiN, TiE are the three components 
of the “ith” Triaxial sensor. 
The table below shows an example of a Multi‐Channel recorder with a Broadband Sensor 
and the other channels used for 2 or 3 Triaxial sensors and other environmental sensors 
such as pressure and temperature sensors. Generally only one broadband sensor would be 

  32 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

connected at a given location. Obviously, for testing or other purposes multiple Broadband 
sensors can be connected. 
Physical  Channel  BB, 2 Triax, Aux Sensors  Virt Ch Map  BB,  3 Triax  Virt Ch Map 

1  VBBZ  V1  VBBZ  V1 


2  VBBN  V2  VBBN  V2 
3  VBBE  V3  VBBE  V3 
4  SEN1  V10  T4Z  V10 
5  T1Z  V4  T2Z  V4 
6  T1N  V5  T2N  V5 
7  T1E  V6  T2E  V6 
8  SEN2  V11  T4N  V11 
9  T2Z  V7  T3Z  V7 
10  T2N  V8  T3N  V8 
11  T2E  V9  T3E  V9 
12  SEN2  V12  T4E  V12 
Contiguous Channels for 4 Tri‐axial Sensors when using terminal   boards 
When using Digitizers with terminal boards and ES‐T’s or older tri‐axial sensors with discrete 
wiring terminated on site it is also possible to connect four tri‐axial sensors so that it is not 
necessary to use a virtual channel map to get the channels in order. This is shown below 
where we have also illustrated using a tri‐axial sensor with the L, V, T configuration that 
is used in legacy installations. Here the first two sensors are powered from the first board, 
with the others powered from boards 2 and 3 as illustrated below. 
 

Physical  Channel  2 Triax Contiguous on Terminal  Boards 


  Sensors  Virtual Channel Map 
1  T1L (Power)  V1 
2  T1V  V2 
3  T1T  V3 
4  T2L (Power)  V4 
5  T2V  V5 
6  T2T  V6 
7  T3L (Power)  V7 
8  T3V  V8 
9  T3T  V9 
10  T4L (Power)  V10 
11  T4V  V11 
12  T4T  V12 
Sensor Cabling 
The sensor cabling should provide individual connections to each sensor we suggest each 
sensor has a separate cable terminating at the connector. Cabling where power and ground 

  33 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

connections as shared between multiple sensors often results in degraded system performance. 
Cabling should also meet all local and National wiring codes.  Kinemetrics can supply plenum 
rated cable when required by code. 
Standard Cables 
Kinemetrics supplies cables for connecting Rock+ Digitizers to the full range of Kinemetrics 
and Streckeisen sensors.  Other cables are also available including generic pigtail arrange‐ 
ments. Cables are also available to connect dual supply sensors to the unit with military 
connectors and to connect a tri‐axial package to three 4th Channel Single Input connectors. 
Contact Kinemetrics for information on your specific cabling needs. If you wish to construct 
your own cables the Hardware Reference section and the details on the External Three 
Channel Sensor Interface and the External 4th Channel Single Input Sensor Connector will 
provide the pin outs and information you require. 
Connecting Sensors using the Four Channel Terminal   Board 
The terminal board is designed to allow configuration of multi‐channel structural monitoring 
systems where large number of sensors need to be connected in a centralized system. The 
cables from the sensors can be simply terminated in the enclosure and then stripped with a 
wire stripper and inserted into the appropriate terminal location. This solution is also ideal 
for setting up temporary experiments when the terminal blocks allow quick reconfiguration 
of the experiment. The only caveat is that these terminal blocks need to be protected within 
a building or enclosure. 
The terminal block is available in two versions the ‐01 and ‐02 versions for dual and single 
supply sensors respectively. 
Dual Supply Four Channel Terminal Board (‐01) 
The dual supply unit includes a power supply that provides +/‐12V power at up to 80mA 
per rail. This power is derived from the switched power connection on the sensor connection 
and is derived from the DC power input to the Digitizer. Ensure that the total current 
requirement of all sensors added to a board is less than 80mA per rail. 
The connections on the Rock+ Terminal Strip (‐01 Version) are shown below. Each terminal 
strip supports the installation of up to four uniaxial dual supply sensors. To obtain the 
best high resolution performance we strongly recommend that the units are configured with 
differential outputs. A 5V Full Scale Differential Output represents a good compromise 
between noise performance and allowable cable length in a structural monitoring system. 
The input sensitivity is software configurable within the digitizer. 
10 Pin Connections for Dual +/‐12V Sensors (‐01 Version) 

  34 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

Pin  Net Name  Label  Description 

1  CHX +  CH+  Channel X Positive Input 


2  CHX ‐  CH‐  Channel X Negative Input 
3  CHXGND  CGND  Channel X Ground & Shield (ISO AGND ‐ CGNB on some units) 
4  CAL X  CAL  Calibration Signal 
5  CENB X  CENB  Calibration Enable 
6  AGND  COM  Connects to ADC Isolated Ground ‐ Common 
7  AGND  COM  Connects to ADC Isolated Ground ‐ Common 
8  +12VA  +12V  Positive 12V Supply to Sensors 
9  ‐12VA  ‐12V  Negative 12V Supply to Sensors 
10  PGP  PGP  Case and Shield Ground 
The connections are the same on each of the four channel groups. The calibration enable 
supports Kinemetrics ES‐U2’s, ES‐U’s, and FBA‐11 sensors. [Note: The terminal blocks can 
also be used to connect Tri‐axial sensors and this is discussed in a separate section.] The 
Calibration Signal and the enable pulse are separate connections on J4 (Group 4) (they are 
common on J1, J2, & J3.) to allow calibration of Tri‐axial sensors but this does not affect 
the electrical connections or use of the signals with Uniaxial sensors. 
Legacy sensors such as the FBA‐1 and FBA‐13 require an additional calibration interface 
box to produce functional tests with Rock+ digitizers. 
The figure below shows the suggested connection of four ESU‐2s using four shielded twisted 
pair cable with an overall shield. 

  35 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

Figure 9: Dual Supply Terminal Board (‐01 Version) 

  36 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION
 
The figure below shows the suggested connections to a four shielded pair cable with no overall 
shield. Although an overall shield is preferred many structural installations require the use 
of plenum rated cable and this figure is compatible with this cabling. It is also compatible with 
Kinemetrics 503536 Pigtails used in structural installations with ES‐U2s, ES‐Us, and FBA‐11 
sensors. 

  37 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

Figure 10: Dual Supply Terminal Board (w/ Cable) 

  38 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

The single supply unit provides a switched power circuit from the input voltage to the 
Digitizer. This is the raw input power without any additional regulation so it is important 
that the sensor can operate from this voltage and will not be damaged. If the Kinemetrics 
power supply is used the nominal voltage will be 15.5V when AC is present and a nominal 
12V from an attached valve regulated lead acid battery (VRLA) when AC power is lost. 
The power has a 500mA resetting polymer fuse on the line to limit the current drawn by 
the sensor. The Calibration Enable Signal is a 5V logic signal with a low of <0.4VDC and 
a high of >4.4  VDC. 
The connections on the Rock+ Terminal Strip configured for single supply Uniaxial sensors 
(‐02 Version) are shown below. Each terminal strip supports the installation of up to four 
uniaxial single supply sensors. To obtain the best high resolution performance we strongly 
recommend that the units are configured with differential outputs. A 5V Full Scale Differ‐ 
ential Output represents a good compromise between noise performance and allowable cable 
length in a structural monitoring system. The input sensitivity is software configurable 
within the digitizer. 
12 Pin Connections for Single Supply Sensors 
 

Pin  Label  Description 

1  CH+  Channel X Positive Input 
2  CH‐  Channel X Negative Input 
3  CGND  Channel X Ground & Shield (ISO AGND) 
4  CAL  Calibration Signal 
5  CENB  Calibration Enable 
6  COM  Connects to ADC Isolated Ground 
7  VRTN  System Power Return 
8  VSYS  System Power to Sensor 
9  VSYS  System Power to Sensor 
10  PGP  Case and Sheild Ground 
11  VOW  One Wire Sensor Device Power 
12  DOW  One Wire Sensor Data X 

The connections are the same on each of the four channel groups (J1,J2,J3,J4). The Cal‐ 
ibration Signal and the enable pulse are separate connections on J4 (Group 4) (they are 
common on J1, J2, & J3.) to allow calibration of Tri‐axial sensors but this does not affect 
the electrical connections or use of the signals with Uniaxial sensors. The figure below shows 
the suggested connection of four single supply ESU‐2s using four shielded twisted pair cable 
with an overall shield and with four twisted pairs only.  (Note only single supply ES‐U2’s 
manufactured after December 2009 should be used with the Rock+ 15.5VDC supply older 
ES‐U2’s and ES‐U’s should use a regulated 12V supplied either to the Digitizer or separately 
to the sensor.) 

  39 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

Figure 11: Single Supply Terminal Board (‐02 Version)  
Wiring Terminal Boards 
The terminal boards consist of fixed terminal blocks on the board and removable plugs with 
identification labels. Removing these plugs while completing the wiring makes this much 
easier. It also allows the digitizer to be removed or replaced without disconnecting all the 
wiring! Make all connections by stripping ~1/2ʺ of insulation from each wire and neatly 
twisting the bare copper strands together. Then, insert each wire into the terminals at the 
correct location and tighten the screw. The terminal plugs can then be‐reinserted into the 
correct location on the terminal board. 
Connecting ES‐Ts and Tri‐axial  Sensors to Terminal    Boards 
Tri‐axial units should be connected to terminal boards in one of the configuration described 
in the previous sections. With a Triaxial unit the first channel to be connected should be used 
to make the additional connections including the power, calibration signal, and calibration 
control. When connecting the fourth (or second) tri‐axial unit to a 12 channel group the 
first channel connects to physical channel 4 for the signal, power, calibration signal, and 
calibration control. Thus two triaxial sensors are powered from the first Terminal Board’s 
ADC. (The ‐01 Option’s 2 Watt converter provides sufficient power for two ES‐T’s configured 

  40 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

as 5V differential output sensors.) The remaining two sensors are powered from the second 
and third four channel group. 
 
Functional Test Sequence The functional test  feature  sends  a  calibration sequence from 
the digitizer to the sensor. The size, duration and shape of the resulting record depend  on 
several  factors,  including  the  sensor  type,  sensitivity, and  voltage  range  as  well  as  the 
particular sensor script  selected. 
For  example, a 4g EpiSensor ES‐T or an Obsidian with a 4g internal deck set for a +/‐  5V 
range will produce the following calibration result using the “EpiSensorFS” calibration 
script: 
 

Figure 12: Basic Functional Test 

 
The duration of the calibration sequence also depends on several factors. In this example, 
the duration of the test record will be set primarily by the calibration script (20 seconds for 
the “EpiSensorFS” script), plus the pre‐event and post‐event time (3 seconds and 10 seconds 
respectively in this example), plus one second. This gives a total expected record duration 
of 34 seconds (20+3+10+1). 
 
General  Sensor  Interfacing  Considerations  The  following  general  considerations  ap‐ 
ply to connecting sensors to Rock+ digitizers: 

48 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

Analog Inputs 
The inputs to the Rock+ digitizers are differential inputs with a nominal 102.4kD differential 
input impedance they can be configured via the software for three input ranges: 

1. 20V Peak (40 Vpp) differential input – Signal Swings from +10V on +Input and ‐10V 
on –Input to ‐10V on +Input and +10V on –Input. This range is compatible with most 
Broadband Seismometers and ES‐T sensors set with a high gain differential output. 

2. 5V Peak (10Vpp) differential input – Signal Swings from +2.5V on +Input and ‐2.5V 
on –Input to ‐2.5V on +Input and +2.5V on –Input. This range is compatible with ES‐
U’s, ES‐U2’s and ES‐T sensors set with a low gain differential output. It represents a 
good compromise for structural monitoring  systems. 

3. 2.5V Peak (5Vpp) differential input – Signal Swings from +1.25V on +Input and ‐ 
1.25V on –Input to ‐1.25V on +Input  and +1.25V on –Input. This  range will  work 
with older generation single ended 2.5V output sensors such as FBA‐11’s, FBA‐23’s 
and EpiSensors that cannot be easily re‐configured. This range can also be used to 
provide a x8 pre‐amplifier function for sensors with a 20V Peak output. 

As the unit has a differential input it can accommodate a common mode signal on the 
inputs. The common mode signal has a 1 MD resistor on each lead to the system ground to 
tie floating systems to the correct operating range. Common mode voltages should generally 
be kept to no more than a few volts so as not to limit the input signal levels. 
Calibrator Output 
The  Rock+  Digitizers  have  an  individual  calibrator  system  on  each  of  their  ADC/DSP 
boards. The calibrator output is provided as a +/‐2.5V single ended signal generated from 
a 16 bit DAC. The output voltage is calibrated during the factory set up with a digital 
scale factor to accurately produce +/‐2.5V at the Full Scale DAC setting. The output current 
is limited to 5mA at 2.5V and the voltage output passes through a 40D resistor to prevent 
overload if the calibration output is accidentally shorted.  The calibration signal is  only 
applied  to  the  external  lines  when  the  appropriate  calibration  relay  is  enabled;  this 
prevents unwanted interactions from occurring with sensors when calibration is not active. 
On a four channel board two relay calibration output connections are provided. The first is 
for channels 1, 2, and 3, while a separate connection is provided for channel 4 calibration. 
This allows a passive (moving coil) seismometer to be connected to channel 4 and calibrated 
separately from a three channel active sensor connected to channels 1, 2, and 3 and for 
different calibration sequences to be applied to the two sensors. It also allows calibration 
signals to be driven to two ES‐T’s connected to one calibrator by staggering the calibrations. 
The DAC can be programmed to provide: 

• Simple Pulses 
• Sine waves of a given frequency and amplitude 
• Pseudo Random White  Noise 

  49 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

• Pseudo Random Binary  Sequences 

Control Signals 
The calibrator can also control up to four logical outputs that can be used to control the 
calibration enable and other sensor control lines. For example the mass re‐centering or 
locking control lines on a broad band seismometer. The four sensor control bits are standard 
ACT CMOS outputs driving between 0 and 5.1V with a current of up to 24mA the low 
level will be <0.4V and the high level >4.4V. The active state to enable the cal coil in the 
EpiSensor, for example is logic 1 or high at about ~5V. There are 830 ohm resistors in series 
with the outputs for short circuit protection. 
Calibration Control and Calibration Scripts 
The functions described above are programmed by individual low level commands sent to 
control the various relays, control bits, or DAC. These commands have been written into 
individual calibration scripts that are selected for each channel group for the most commonly 
used sensors. If you wish to write your own script or connect a sensor that does not have a 
calibration script contact Kinemetrics. 
Calibration Signals when using four EpiSensors on a 12‐Channel Group 
When four triaxial EpiSensors ES‐T’s are connected to a 12 channel group the fourth ES‐ 
T receives its power and calibration signals from the Channel 4 physical connection. It is 
important that the unit is configured so that the calibration sequence is applied first to the 
sensor connected to Channel 1, 2, and 3 and then to the sensor connected to Channel 4. 
This is because an individual DAC cannot apply sufficient current to the calibration coils 
of 6 EpiSensor  axes at the  same time. Instead the  calibration  is  applied  to  the  ES‐T  on 
Physical Channel 4 after the other three have received the calibration signal. This is set in 
the System Operation section of the web interface by selecting “EpiSensor‐4on3” for each 
sensor group and applying the changes to the unit. 
This same approach should be used for four EpiSensors connected sequentially on a terminal 
board.  In this case the second sensor receives it’s calibration from the Channel 4 Physical 
connection. And the calibration sequence will show sensors 1, 3 and 4 with a calibration 
pulse followed by sensor 2. Again this is set in the System Operation section of the web 
interface by selecting “EpiSensor‐4on3” for each sensor group and applying the changes to 
the unit. 
 
Connecting the  Optional  External Battery 

WARNING: Fire or explosion hazard. Do not install a non‐rechargeable 
battery to be charged by the unit. Only install a sealed lead acid bat‐  
tery with specifications compatible with those above. Other types  of 
Lead Acid battery should not be used as the temperature corrected 
charging voltage is set for SLA batteries only.   On no account install     
an alternative battery chemistry such as Lithium Ion or Nickel Metal 
Hydride! 

  50 
CHAPTER 1. INTRODUCTION 1.2. INSTALLATION & SETUP

If you store a Sealed Lead Acid (SLA) battery, you should still charge it every six to nine 
months to prevent permanent loss of capacity. You can float‐charge the battery at 13.5‐13.8V 
or cycle‐charge the battery, provided the current is limited appropriately and the voltage to 
less than 14.7V. When the voltage reaches 14.7V, the battery will be damaged unless you 
convert the cycle charging to float charging. Kinemetrics ships batteries fully charged; make 
sure a battery is still fully charged before installing it. 
Battery Installation 
Follow the instructions below to install the external battery. 
 
 
WARNING:  Burn or explosion hazard.  Never    place  metallic 
objects  (such  as  a  screwdriver  or  your  wristwatch  strap)  across  the 
terminals  of  a  battery.  The  metal  terminals  can  get  very  hot.  Handle 
batteries with care, and do not drop them or attempt to take them apart. 
Recycle  used  batteries,  or  dispose  of  them  in  accordance  with  local 
regulations.  Do not throw used batteries onto a fire. 

 
CAUTION:  Before  installing  the  new  battery  make  sure  it  is  fully 
charged. If the battery is uncharged, the unit will charge it, but this can 
take some time — and if AC power is lost, the unit’s power autonomy 
will be  reduced. 
 
To install an external battery: 

1. Ensure the power cable is disconnected from all power sources and the unit. 

2. Connect the negative terminal of the battery to the black wire. 

3. Connect the positive terminal of the battery to the red wire. 

4. Connect the power cable to the power connector on the unit and ensure the unit turns 
on under battery power. 

5. Proceed to connect the PSA and check that the unit switches to this power source. 
 
NOTE: An external battery should not be used with a system powered 
by  a  Power  Over  Ethernet  (POE) connection. 

  51 
1.2. INSTALLATION & SETUP CHAPTER 1. INTRODUCTION

Connecting  the  Power  Supply  The  unit  optionally  includes  a  wide‐input  DC  power 
supply assembly which connects to a standard AC outlet. Some systems are shipped with 
External Charging systems that are provided with separate operating instructions. 
 
CAUTION: Using the wrong power supply unit with the unit can per‐ 
manently  damage  its  circuit  boards.  Do  not  attach  any  power  supply 
assembly to the digitizer other than the one  supplied  by  Kinemetrics 
unless it exactly matches the voltage and current ratings required for the 
digitizer.  Older  Altus  power  supplies  are  not  compatible  with  Rock+  
digitizers. 

Before plugging in the PSA: 

• Make sure that the AC outlet is properly wired. 
• Find out if the local AC power is subject to interruption, brownouts, or spikes. If it 
is, plug a suitable surge suppresser into the AC outlet. (Strongly recommended!) 

Then do the following: 

1. Plug the PSA into the AC power outlet or surge suppressor. (Users outside the U.S. 
must use an adapter or an appropriate IEC line cord with the correct AC plug.) 

2. Push the 12‐socket power connector into the digitizer’s POWER & BATTERY con‐ 
nector and twist the rotating shell until it locks. 
 
NOTE:   The PSA automatically adjusts for line voltages from 90 to   260 
VAC  (either  50  or  60Hz)  without  user  intervention. 
 
Connect  a  laptop  or  equivalent  to  the  console  port  The  digitizer  should  now  be 
connected to a Laptop or PC using the console port. (A serial to USB connector may be 
required.) The initial set‐up of the device can now be performed over this serial link as 
described in the Initial Setup in the next section. 

  52 
 
 
 
 

Chapter  2 

System  Overview 

This system gives an overview of the Rock+ digitizer system and the various components 
that are included in the system. This explains some of the operational features and will give 
other information that will allow you to understand the many capabilities of the system. A 
Rock+ Digitizer typically includes the following subsystems: 

• Intelligent power supply 

◦ DC power input 
◦ Internal supercapacitor  array 
◦ External battery interface 
◦ Temperature sensor 
◦ Humidity  sensor 
◦ Power system control processor 

• Main processor System 

◦ Processor 
* Linux Operating  System 
* Boot Loader 
* Java Virtual Machine 
* RockHound Software 
◦ On‐board  RAM 
◦ Console port 
◦ Ethernet  Communications 
◦ Serial Communication Interfaces 
◦ USB Host and Device Interfaces 
◦ POTS Modem Interface 

  53 
2.1. INTELLIGENT POWER SUPPLY CHAPTER 2. SYSTEM OVERVIEW

◦ Time/IO Interfaces for Switched Power and 1PPS 
◦ Primary SDHC card for firmware 
◦ Secondary SDHC card for data 
◦ GPS Timing Module & Disciplined Oscillator 

• One or more four channel high resolution analog to digital converters and supporting 
circuitry 

◦ Individual  DSP 
◦ C++ DSP Code 
◦ Isolated DC/DC Power supply for Analog Section 
◦ Isolation Barrier to Analog Section 
◦ Four Delta Sigma ADC Converters 
◦ Adjustable Gain Input Stages 
◦ Mass position/Auxiliary  ADC 
◦ Calibration and Control  Processor 

• One or More External Sensor Input Boards 

◦ Military Style  Converters 
◦ EMI/RFI and Transient Protection  Circuitry 
◦ Calibration and Sensor Control Signal Outputs 

• Optional Internal Sensor Deck (Obsidian only): 

◦ ±2g/1g EpiSensor Deck (±4g/2g Optional) 
◦ Powered from Unit  Power 
 
2.1 Intelligent Power  Supply 
The intelligent power supply subsystem is an important part of the reliability and proper 
functionality of your system. The power supply subsystem provides the following  features: 

• Supercapacitor backup 
• Battery  charging 
• Reversed input and overvoltage  protection 
• Sequenced start up 

  54 
CHAPTER 2. SYSTEM OVERVIEW 2.1. INTELLIGENT POWER SUPPLY

• Sequenced shut down 
• Communication with the user via status LEDs 
• Magnetic switch 
• Temperature monitoring 
• Humidity   monitoring 
• Timed operation windows 
• Parameteric  operation protection 
• System watchdog 
• Digital  input/output 

In general, the power supply subsystem takes input from system power sources and converts 
them into the supply voltages necessary to operate the system. In addition to this primary 
task, the power supply subsystem also monitors system current usage, voltage levels, tem‐ 
peratures and other parameters to assure that the system is operating correctly and within 
operational limits. Under some circumstances, the power supply subsystem may inhibit start 
up of the system in order to protect the hardware. This system basically ensures the safety 
of the remainder of the Rock+ digitizer. 
 
2.1.1 Supercapacitor  Backup 

The power supply subsystem utilizes a supercapacitor array to provide backup power to the 
processor. This is important to assure proper shut down of the operating system in the 
event that external power is lost. The supercapacitor array will be charged when power 
is first applied and before the system is allowed to operate. The process of charging the 
supercapacitor array can take up to a few minutes depending on the capacity of the array 
and how much energy is already stored. 
 
2.1.2 Battery Charging 

The power supply subsystem includes a battery charging feature that can float‐charge a 
Sealed Lead‐Acid battery. The battery charging system will attempt to keep a battery at 
full charge, so that the system can continue to operate from the battery when external power 
has been lost. 
Special considerations for battery charging include: 
 
• External power limits: The battery will only be charged when external power is at or 
above 14.8VDC, regardless of  source. 

  55 
2.1. INTELLIGENT POWER SUPPLY CHAPTER 2. SYSTEM OVERVIEW

• Temperature limits: The battery will only be charged between 0 and +40 degrees C. 
This prevents shortening battery life due to charging at extreme temperatures. 

• Temperature compensated charging: The charging voltage is adjusted with unit tem‐ 
perature to ensure optimum battery life. 

• Pulse Charging: Below 9VDC, the battery is assumed to be ʺdeeply dischargedʺ, and 
the battery charger will attempt to charge the battery using a 10% duty cycle pulse 
charge. 

• Battery drop‐out: If operating off of battery and battery voltage reaches 10.75VDC, 
the system will shut down, thus preventing deep discharge of the battery. 

• Battery operational limits: Operation from battery is recommended only between ‐15 
and +50 degrees C. Beyond these limits, the system will continue to operate, but will 
indicate a fault. 
 
2.1.2.1     Reversed Input and  Overvoltage Protection 

The system will protect against a reversed battery and will report the battery voltage as 
negative on the appropriate state‐of‐health channel. The reversed battery is also indicated 
as a flag indication in the state‐of‐health data. The system can continue to operate with 
reversed battery, but cannot operate from the reversed battery as it’s only power source and 
will not attempt to charge it. 
 
2.1.3 Sequenced Start  Up 

The system will perform a sequenced power up when power is applied to the system. The 
sequence is to assure proper start up of the processor. The system start up sequence is 
approximately as follows: 

• Perform initial start up delays to avoid system ʺmotorboatingʺ 
• Charge the supercapacitor  array 
• Assure that external power sources and other operating parameters are within limits 
• Start the processor 
• Load and start the analog to digital sections 
 
NOTE:  Supercapacitor and battery  charging  circuits   are  indepen‐ 
dently controlled 

  56 
CHAPTER 2. SYSTEM OVERVIEW 2.1. INTELLIGENT POWER SUPPLY

2.1.4 Sequenced Shut Down 
The system also controls sequenced power down when the system has been commanded off, 
or when all sources of external power have been lost. If external power has been lost, the 
power supply will perform the following steps after having automatically started running off 
of the supercapacitor power: 

• Turn off the supercapacitor charging circuit 
• Turn off the battery charging circuit 
• Notify processor that shut down is imminent, killing all processes 
• Wait for residual boost voltage low, processor to acknowledge shut down, or shut down 
timeout to expire 
• Shut off the processor 

When commanded to power down (not due to power loss), the power supply will perform 
the following steps: 

• The processor will be notified that a ʺslowʺ shut down is required through the power 
supply status flags, and the processor will have up to 15 minutes to terminate appli‐ 
cations. 
• The power supply will monitor power loss, processor shut down acknowledge, or com‐ 
pletion of the 15 minute timeout and may then transition to the sequenced shut down 
described above. 
 
2.1.5 Communication with the User Via Status LEDs 
The power supply indicates its state of operation to the user through its LEDs: 
 
Power: 

• OFF ‐ No power 
• Steady Green ‐ Waiting to Power up, Running off of external power or POE (Power 
Over Ethernet) 
• Flashing Green ‐ The system is starting up 
• Infrequent Green ‐ Running off of battery 
• Orange ‐ Supercapacitor is being charged 
• Flashing Red ‐ Power system fault 
• Alternating Red and Green ‐ The system is shutting down 

  57 
2.1. INTELLIGENT POWER SUPPLY CHAPTER 2. SYSTEM OVERVIEW

2.1.6 Magnetic Switch 

The magnetic switch is located above and to the right of the Modem connector. The magnetic 
switch is located directly behind the seal screw. The switch is used to start up and shut 
down the unit without use of a physical power switch. 
The unit has no power switch. When power is connected, the unit’s default behavior is to 
power up and begin operation.  When power is removed, the applications and operating 
system will shut down and the unit will turn off. 
Briefly touching a strong magnet near the magnetic switch location allows the user to affect 
this default behavior as follows: 
 
• When Running ‐ The magnetic switch causes the unit to shut down and go into a 12 
hour  timed  operation  window.  If  power  remains  connected,  the  unit  will  restart 
operation in 12 hours. 

• When not running (in a timed window) ‐ The magnetic switch will terminate the 
window and will cause the unit to start operating. 
 
There are other functions of the magnetic switch used in conjunction with diagnostic func‐ 
tions. 
 
2.1.7 Temperature Monitoring 

The power supply system monitors system temperature through use of a built‐in temperature 
sensor. The temperature is used as an operating limit for system operation and battery 
charging, as well as being sent to the processor as state‐of‐health data. Temperature limits 
are defined as: 

• Operating (from DC): ‐20 to +60 OC Operation (optional ‐30 to   +70OC) 
• Operating (from Battery): ‐15 to +50 OC 
• Battery charging:  0 to +40  OC 
 
2.1.8 Humidity Monitoring 

When equipped with the humidity sensor, the power supply system monitors system humid‐ 
ity within the case through use of a built‐in humidity sensor. The measured relative humidity 
is used as an operating limit for system operation, as well as being sent to the processor as 
state‐of‐health data. Maximum operating humidity is 80% relative humidity, but any rise 
above 50% indicates either that the desiccant is exhausted or needs to be replaced, and/or 
that the case is leaking and is in a damp environment. In both cases, service is required. 

  58 
CHAPTER 2. SYSTEM OVERVIEW 2.1. INTELLIGENT POWER SUPPLY

2.1.9 Timed Operation Windows 

The power supply subsystem supports timed operation windows. The processor can set the 
power supply’s real time clock, then set a time that the system should power back on (a time 
that the processor should be turned on), and then the system can be shut down. Assuming 
that enough power is maintained to keep the real time clock running, the system will power 
up once the specified time is reached and operating parameters (temperature, voltages, etc) 
are such that the system is allowed to start. 
The real time clock has a small dedicated supercapacitor that is capable of maintaining time 
in the real time clock for several days. 
 
2.1.10 Parametric Operation Protection 

Operating limits are imposed in order to protect the hardware, but to allow operation of the 
system as much as possible. Defined limits are: 

• Operating (from DC): ‐30 to +70 OC  Operation 
• Operating (from Battery): ‐15 to +50 OC 
• Operating humidity limits:  80% relative  humidity 
• Battery charging: External power at or above 15.0VDC 
• Battery charging:  0 to +40  OC 
• Pulse charging: Battery below 9VDC 
• Battery drop‐out: Operating off of battery and battery below 10.75VDC 
• Battery operational limits: ‐15 to +50 degrees C. Operating off of battery beyond these 
limits, the system will continue to operate, but will indicate a fault. 

• Supercapacitor array charged:  Above  4.2VDC 
• Maximum operating current: 2 Amps (processor only) 
• External voltage limits: 7.75 to 30.0VDC 
• POE voltage limits: 14.0 to  16.0VDC 
• Maximum operating battery voltage: 15.2VDC 
• VCC voltage limits: 3.13 to 3.47VDC 
• 1.8V core voltage limits: 1.71 to 1.89VDC 
• Boost voltage limits: 4.5VDC minimum 

  59 
2.1. INTELLIGENT POWER SUPPLY CHAPTER 2. SYSTEM OVERVIEW

• Processor core voltage limits: 0.9 to 1.4VDC 
• Battery exists:  6.0VDC 
• External DC exists:  1.0VDC 
• POE exists:  1.0VDC 
 
2.1.11 System  Watchdog 

The system includes an advanced power system that provides watchdog and power fail 
protection with power fail warning capabilities. 
The watchdog provision is independent of the main processor, but does interact with it. The 
watchdog exists as part of the power supply subsystem, and requires that it be ʺpingedʺ by 
the main processor every 10 minutes. Failure to do so will be interpreted by the watchdog 
as a lockup of the main processor and the main processor will be shut down and restarted 
in an attempt to get the system back into operation. Exceptions: 
 
• The watchdog is initially held off for 4 hours after processor power on, allowing for 
initial start up and worst case filesystem repair on a large system disk. 
 
2.1.11.1     Default Behavior 

The watchdog is normally serviced by a Linux service in order to keep the system alive. By 
default, only a failure of Linux to boot or a crash of the operating system or of the watchdog 
service will cause the watchdog to trigger and restart the system. 
 
2.1.12 Power Fail  Protection 

The system includes an internal power backup called a supercapacitor array which is charged 
with power at system turn on.  This backup provides enough residual power to terminate 
any running applications and shut down Linux properly. Once power fail has been detected, 
the system begins to operate off of this residual power and issues a request to Linux to shut 
down. A Linux service watches for the shut down request, and once received immediately 
forces all user processes to terminate and then shuts down Linux.  When Linux has completed 
the shut down, the shut down request will be acknowledged and the processor will power 
off. 
 
2.1.13 Application Specific  Configuration 

The system provides several ways that the application can improve its control of the system: 

  60 
CHAPTER 2. SYSTEM OVERVIEW 2.1. INTELLIGENT POWER SUPPLY

• The application can monitor power supply information including system voltages, tem‐ 
perature and relative humidity as well as operating states. These values can be acted 
on, or can simply be stored or recorded for diagnostic purposes. 

• One of the power supply parameters retrieved is a set of system flags that indicates 
that the system is being requested to shut down soon, but when ʺconvenientʺ. This 
would include for example, a condition where it has become too warm or too humid 
to continue operation. In this case, the application must react to the application shut 
down request flag and be terminated in a maximum of 15 minutes. 

• Some of the system flags can be written to, allowing the application some level of 
control. For example, through the flags the application can control: 

◦ A digital output (for example to turn on a dehumidifier or pump if high humidity 
is detected) 
◦ A digital input that can be read and interpreted by the application (for example, 
indicating that a pump is running) 
◦ Indicating a software fault (affects the Status LED) 

• The application can more closely control the system watchdog by servicing it itself. In 
this way, the watchdog will not depend on whether or not Linux and the Linux watch‐ 
dog service is running, but can be made to depend instead on the proper functioning 
of the application. 

As an example, you can read from the power supply at /proc/kmipsu like this: 

cat  /proc/kmipsu 

The flags read or written to/from /proc/kmipsu can be interpreted as follows: 

• 0x00001 ‐ GPIO input  bit 
• 0x00002 ‐ GPIO output  bit 
• 0x00004 ‐ GPS backup power control 
• 0x00010 ‐ Battery load test on 
• 0x00040 ‐ Battery reversed 
• 0x00080 ‐ Software fault (affects LED indications) 
• 0x00100 ‐ System shutdown requested (15 minute max) 
• 0x00200 ‐ Parameter repair startup (applications should not start) 
• 0x00400 ‐ System error 

  61 
2.1. INTELLIGENT POWER SUPPLY CHAPTER 2. SYSTEM OVERVIEW

• 0x00800 ‐ Battery charger on 
• 0x01000 ‐ Power supply diagnostic mode 
• 0x02000 ‐ Power supply diagnostic mode 
• 0x04000 ‐ Power supply diagnostic mode 
• 0x08000 ‐ Power supply diagnostic mode 
• 0x10000 ‐ Sensor Power Fault 
• 0x20000 ‐ COMM power fault 
 
2.1.14 Power System Display 

The psdisplay command displays power system information once: 

root@Rock102:~# psdisplay 
PSRTC=470951419 
Temp=29.42 
Humidity=37.40 
Flags=16 
ExtCharger=0 
SensorPwr=1 
BBVolts=12.502 
VofSC=4.403 
Vof1_8VCore=1.805 
VofVCC=3.326 
VofVBoost=5.109 
Vof12VPOE=1.368 
Vof12ExtDC=15.492 
VofBattCharge=0.049 
VofV12Batt=0.039 
VofSysPwr=15.473 
VofIofSysPwr=0.240 
VofIofBBIn=0.076 
VofTimeIOPwr1=0.000 
VofIofBC=0.045 
VofSnsrPwrSw=12.435 
VofTimeIOPwr2=0.000 
VofVBB=12.523 
VofTimeIOPwr3=0.000 
VofSagFilt=14.994 
PSType=1 
PSVersion=6.16 

  62 
CHAPTER 2. SYSTEM OVERVIEW 2.1. INTELLIGENT POWER SUPPLY

The items displayed by psdisplay are as follows: 

• PSRTC: Power supply real time clock time in seconds since January 1, 2000 
• Temp: Current temperature in degrees C 
• Humidity: Current relative  humidity 
• Flags: Current power supply system  flags 
• ExtCharger: External charger control on (1) or off (0) 
• SensorPwr: Sensor power on (1) or off (0) 
• BBVolts: Buck Boost voltage (for sensor and COMM power) 
• VofSC: Voltage of the power supply supercapacitor pack 
• Vof1_8VCore: Voltage of the 1.8V core voltage 
• VofVCC: Voltage of the 3.3V core voltage 
• VofVBoost: Voltage of the available boost voltage 
• Vof12VPOE: Voltage of the POE  input 
• Vof12ExtDC: Voltage of the External DC supply 
• VofBattCharge: Voltage of the battery charger 
• VofV12Batt: Voltage of the bipolar battery measurement 
• VofSysPwr: Voltage of the internal system power bus 
• VofIofSysPwr: System current usage in mA 
• VofIofBBIn: Buck Boost  current 
• VofTimeIOPwr1:  Switched power voltage  1 
• VofIofBC: Battery charger current 
• VofSnsrPwrSw:  Sensor Power voltage 
• VofTimeIOPwr2:  Switched power voltage  2 
• VofVBB: Buck Boost voltage 
• VofTimeIOPwr3:  Switched power voltage  3 
• VofSagFilt: Sag filter voltage 
• PSType: Power Supply type, digitizer (1) or Slate (0) 
• PSVersion: Power supply  version 

  63 
2.2. PROCESSOR SYSTEM CHAPTER 2. SYSTEM OVERVIEW

2.1.15 Updating the Power  Supply 


The intelligent power supply includes a microprocessor that holds some highly specialized 
firmware for controlling it. This firmware is updated as part of package updates and normally 
does not have to be updated separately. 
If needed, the power supply firmware can be independently updated in the field when needed 
as follows: 
 
• Obtain the updated power supply firmware (named something like rock2_psu_6.17.s28) 
and put it onto your system through methods such as SFTP. 

• Log into Linux as root 
• psfwupdate /root/rock2_psu_6.17.s28 

This will load the new power supply firmware, Linux will shutdown and the system will 
restart with the new firmware. 
 
2.2 Processor System 
The Processor system provides the intelligence for the system and controls all the user inter‐ 
faces, storage media, and system operation. The processor also is responsible for configuring 
an loading the code into the DSPs on the Analog to Digital Converter (ADC) boards. It also 
handles updating the code in the other processors in the Power Supply and ADC boards. 
 
2.2.1 Processor 
The Processor is a 1GHz FreeScale iMX53 Processor. This is a highly integrated low power 
processor that is based on the ARM architecture. It has 1 GB of SDRAM Memory. The 
system boots Linux from the internal System SDHC Card. 
 
2.2.2 Storage 
The unit provides a two internal SDHC card slots: 

• The System card is typically a 4GB card 
• The Data card is typically a 32GB card 
 
2.2.3 Communication Interfaces 

The Processor supports multiple communication protocols. 

  64 
CHAPTER 2. SYSTEM OVERVIEW 2.2. PROCESSOR SYSTEM

2.2.3.1 Serial Connections 

A console port and three general purpose serial ports are provided with every unit. 
 
2.2.3.2 Ethernet Connections 

Rock+ Digitizers provide a 10/100BaseT Ethernet connection. 
 
2.2.3.3 USB  Connections 

The unit provides a USB 2.0 Device Port and three USB 2.0 Host Ports. 
 
2.2.3.4 GPS System 

A GPS module is provided on the processor board that can work with the Analog to Digital 
Converter Boards to provide data time aligned to sub‐microsecond accuracy. The GPS can 
also provide the accurate location of the unit. 
 
2.2.3.5 Hardware Identification 

The major Kinemetrics‐built electronic components of your system include information that 
helps to identify them for maintenance, warranty, and repair purposes. The processor can 
interrogate this  information. 
Each circuit board within the system includes the following information: 

• Part number 
• Serial number 
• Electronic serial  number 
• Assembly dash number 
• Assembly  revision 
• Parts List revision 
• Build date 
• Test date 

In addition, a unit tag number is assigned to the system as a whole. This unit tag number 
is stored on the main processor board. 
To list this identification information on your system, execute the owview command from 
the Linux command prompt. The display looks something like: 

  65 
2.2. PROCESSOR SYSTEM CHAPTER 2. SYSTEM OVERVIEW

PartNumber=112110 
DashNumber=03 
BoardRev=A 
PartsListRev= 
SerialNo=68 
Bridge=1 
ESerialNo=1C009B1003000059 
Built=4907   
Tested=4614 
VendorNumber=49070068 
BoardType=BlackFin 
BFNumber=0 
PartNumber=114080 
DashNumber= 
BoardRev=A 
PartsListRev=B 
SerialNo=3 
Bridge=1 
ESerialNo=1C7DBAE70B0000BE 
Built=4104   
Tested=4614 
VendorNumber=41040003 
BoardType=PSU‐Carrier 
PartNumber=114035 
DashNumber=Q 
BoardRev=A 
PartsListRev=C 
SerialNo=2 
Bridge=1 
ESerialNo=1C7ED9D90C0000C6 
Built=4304   
Tested=4714 
VendorNumber=43040002 
BoardType=FrontPanel 
PartNumber=114060 
DashNumber= 
BoardRev=A1 
PartsListRev=A 
SerialNo=9 
Bridge=1 
ESerialNo=1C7FE8E30C0000DC 
Built=3904   
Tested=4614 
VendorNumber=39040009 
BoardType=MainBoard 

  66 
CHAPTER 2. SYSTEM OVERVIEW 2.3. ADC/DSP SYSTEM

UnitTagNo=109 
MACAddress=70:B3:D5:CE:92:8E 
Total  devices: 4 
 
2.3 ADC/DSP SYSTEM 
Each 4 Channel ADC DSP Board contains separate DSPs, ADCs and Power Supplies it is 
this architecture that allows the Rock+ Digitizers to be scaled from 4 to 36 Channel systems. 
 
2.3.1 Individual DSP 

The DSPs are Analog Devices BlackFin processors coupled with discrete logic embedded 
in CPLDs. The DSPs communicate with the main processor using a custom high speed 
hardware connection system. This connection also allows all sampling within the system to 
be synchronized to sub‐microsecond accuracy. 
 
2.3.2 C++ DSP Code 

The DSPs are programmed in C++ and run a custom executive that ensure reliable real 
time operation. The DSPs take the 30ksps samples from the ADC and filter and decimate 
the streams to the required output rates. They support output rates from 5000 sps down. 
 
2.3.3 Isolated DC/DC Power supply for Analog Section 

Each Analog to Digital Converter Block features and isolated DC to DC power supply 
that provides the power for the ADC chips and analog circuitry associated with the signal 
conditioning. 
 
2.3.4 Isolation Barrier to Analog Section 

All digital signals to and from the isolated analog section pass through MEMS digital isolators 
to preserve the isolation of the analog side of the board. 
 
2.3.5 Four Delta Sigma ADC  Converters 

Each channel is digitized by a Delta Sigma converter producing a 20ksps output. This raw 
data stream is sent across the isolation barrier to the DSP for filtering and decimation. 

  67 
2.4. EXTERNAL SENSOR INPUT BOARDS CHAPTER 2. SYSTEM OVERVIEW

2.3.6 Adjustable Gain Input Stages 

The analog input stage is configurable to three input ranges and has automatic calibration 
features that allow a calibrated signal to be connected to the analog input and ADC. 
 
2.3.7 Calibration and Control Processor & Mass position/Auxiliary 
ADC 

A processor is provided on the Analog side of the interface with a low speed ADC to monitor 
external mass position as well as the State of Health of the Analog system. This processor 
also control calibration signals to the sensors and control signals sent to the sensors. 
 
2.4 External Sensor Input  Boards 
The external sensor boards connect the signal from the outside world to the ADC boards. 
 
2.4.1 Military Style  Connectors 

The unit is provided with Military style input connectors on the power and sensor connections 
to allow it to be used in harsh environments and provide IP67 protection. 
 
2.4.2 EMI/RFI and Transient Protection  Circuitry 

Each line is protected from EMI/RFI and external transients with one or two stage transient 
protection as  appropriate. 
 
2.4.3 Calibration and Sensor Control Signal Outputs 

The input boards condition the calibration and sensor control signals and also provide relay 
isolation of the calibration system to provide greater system flexibility. 
 
2.5 EpiSensor Theory of Operation 
This section describes the operating principles of the EpiSensor internal deck used in the 
Obsidian. 
The EpiSensor deck consists of three orthogonally mounted force balance accelerometers 
(FBAs) – Z‐axis, N‐axis and E‐axis. The figure below shows a simplified block diagram of 
the major components of each of the EpiSensors. 

  68 
CHAPTER 2. SYSTEM OVERVIEW 2.5. EPISENSOR THEORY OF OPERATION

Figure 13: EpiSensor Simplified Block Diagram 

 
2.5.1 Working Principle 

The oscillator applies an AC signal of opposite polarity to the two moving capacitor plates 
(also referred to as ʺthe moving massʺ). When the accelerometer is ʺzeroedʺ and when no 
acceleration is applied, these plates are symmetrical to the fixed central plate and no voltage 
is generated. 
An acceleration causes the coil and capacitive sensor plates, which are a single assembly 
mounted on mechanical flexures (springs), to move with respect to the fixed central plate of 
the capacitive  transducer. 
This displacement results in a signal on the center plate of the capacitor becoming unbal‐ 
anced, resulting in an AC signal of the same frequency as the oscillator being passed to the 
amplifier. 
The amplifier amplifies this AC signal. 
This error signal is then passed to the demodulator where it is synchronously demodulated 
and filtered, creating a ʺDCʺ error term in the feedback amplifier. 
The feedback loop compensates for this error signal by passing current through the coil to 
create a magnetic restoring force to ʺbalanceʺ the capacitor plates back to their original null 
position. 
The current traveling through the coil is thus directly proportional to the applied acceler‐ 
ation. By passing this current through a complex impedance consisting of a resistor and 
capacitor, it can be converted to a voltage output proportional to acceleration with a band‐ 
width of approximately 200 Hz. The internal deck is set to provide a 2g output range. 

  69 
2.5. EPISENSOR THEORY OF OPERATION CHAPTER 2. SYSTEM OVERVIEW

2.5.2 Pole Zero Representation of the  EpiSensor 

EpiSensor accelerometers are closed‐loop, force‐feedback sensors measuring the relative dis‐ 
placement of a moving mass (plates) with respect to the sensor case. The sensor’s transfer 
function (TF) depends almost entirely on the electronic components rather than on the me‐ 
chanical components of the sensors. The influence on the transfer function of the mechanical 
damping, spring elements and internal RC low‐pass filter in the trans‐conductance amplifier 
stage within the closed‐loop path of the sensor are negligible for most applications. 
We have determined a good empirical model of the system, which uses two pairs of conjugate 
poles to represent the transfer function of the instrument. If this transfer function is corrected 
for the DC sensitivity of the sensor, the amplitude agreement is within +/‐0.5 dB over the 
bandwidth of the sensor. The phase agreement is within +/‐ 2.5 in the 0‐100 Hz band and 
within +/‐ 5º over the full bandwidth of the instrument. The phase response of the transfer 
function is fairly linear and equivalent to approximately 1.6 ms group delay for signals up 
to 200 Hz. 
This model can be represented as: 

where: 

• k1 = 2.46 x 1013 
• k2 = Sensitivity of sensor in V/g (1.25V/g) 
• s is the Laplace transform variable 

• p1 = ‐981 + 1009i (Pole 1) 

• p2 = ‐981 ‐ 1009i (Pole 2) 

• p3 = ‐3290 + 1263i (Pole 3) 

• p4 = ‐3290 ‐ 1263i (Pole 4) 
• V (s) is the Laplace transform of the output voltage 
• A (s) is the Laplace transform of the input acceleration 

The figure below shows the amplitude, phase and step response of this pole zero representa‐ 
tion. 

  70 
CHAPTER 2. SYSTEM OVERVIEW 2.6. FIRMWARE DESCRIPTION

Additional references to pole zero responses and damping are available on the Kinemetrics 
website. 
 
2.5.3 Polarity  Conventions 
The internal EpiSensor deck uses a right‐handed Z‐N‐E coordinate system with a positive 
output for acceleration along each axis. 
 
2.6 Firmware Description 
The system operates using the Linux operating system, while the processor digitizer software 
is written in Java and runs on a JVM on the system. The power supply and calibration 
processors are programmed in C, while the DSPs run software in C++ on a custom scheduling 
system. 
 
2.6.1 Linux Overview 
The Linux kernel used in the system is GNU Linux v3.12.36 as of this writing, built from 
kernel.org resources with additions to support Kinemetrics hardware. 
The Linux distribution is sized to operate within the processor and storage resources avail‐ 
able, but to accommodate the needs of the task at hand. 
Since the Linux distribution is for a specialized low power embedded system, it is important 
to note that it may not include some utilities that experienced Linux users might expect to 
find on a typical Linux desktop or server.  Some of the Linux capabilities supplied by this 
distribution include vi, bash, fsck, ssh, telnet, ftp, and awk. This is of course only a partial 
list, and some of these features may require the user to enable or configure them before use. 
Additional Linux capabilities may be provided by major application software packages such 
as Antelope on the Marmot (which includes perl), or by some of the optional software tools 
available from  Kinemetrics. 

  71 
2.6. FIRMWARE DESCRIPTION CHAPTER 2. SYSTEM OVERVIEW

The Bootloader, Linux kernel and boot filesystem are located on th primary System SDHC 
memory card. Once the Linux kernel initially loads, control is transferred to the boot and 
the data filesystem located on the secondary Data SDHC card. 
The  primary  and  secondary flash cards use  the EXT4 journaling  filesystem to give the 
system improved reliability by making it less susceptible to corruption caused by improper 
shut downs or power outages. 
It is assumed that the user has some familiarity with Linux.  No attempt is made here   to 
familiarize the user  with the  internals of Linux or with all  available Linux  commands, 
although some very introductory Linux command information is included here for reference. 
For more detailed information, the user is referred to any one of the many books, users 
groups, or other resources available on this topic. 
Source for the parts of Linux that are licensed under the GPL is available on request under 
section 3(b) of the  GPL. 
 
2.6.2 Common Linux Commands 
The following is a list of commonly used Linux commands which may be of value as reference 
information. Remember that Linux is case sensitive. Options or flags which can be used 
with a command are placed in [ ]. The [ ] are not part of the command and should not be 
included in the command that you type. 
This information is not an exhaustive listing of commands! 
cd 
Change directory 
cd .. moves you backwards to the next higher subdirectory level 
cd / moves you to the highest directory level 
chmod  permissions  filenames 
Changes the permissions for a file 
Permissions should include a letter designating who gets permissions (u for the user, g for 
the group, o for others, or a for all) 
followed by a + or ‐ (to give or take away the permission) 
followed by the kind of permission (r for read access, w for write access, x for execute if the 
file is a program or script) 
The complete command that you type should look like: 
chmod g‐w filename 
chown user:group filenames 
Changes ownership of a file 
clear 

  72 
CHAPTER 2. SYSTEM OVERVIEW 2.6. FIRMWARE DESCRIPTION

Clears the screen 
cp oldfiles newfiles 
Copies a file; this leaves the old file intact and makes a new copy with a new filename 
date 
Tells you the current date and time 
df 
Displays how much space on the disks (storage partitions) is free 
du [‐a] [‐s] directories 
Tells you how much disk space your files occupy The ‐a option displays the space used by 
each file, not just each directory The ‐s option displays the total space used for each directory 
but not subdirectory 
help 
Provides online help; several topics have been included in the help system 
login  username 
Allows you to login for a terminal session, either through telnet, the console, or SSH logout 
Logs you out of a terminal session 
ls [‐l] [‐a] [‐p] [‐r] [‐t]  [‐x] 
Lists the files in a directory 
l displays detailed information about each file and directory, including permissions, owners, 
size and time/date when the file was last modified 
a option displays all the files and subdirectories including hidden files (with names that begin 
with a dot) 
p displays a slash at the end of each directory name to distinguish them from filenames 
r displays files in reverse order 
t displays files in order of modification time 
x displays the filenames in columns across the screen. 
mkdir  new_directory 
Makes a new subdirectory with the name specified by new_directory 
mv  [‐i]  oldname newname 
Renames a file or moves it from one filename or directory to another. The ‐i option tells mv 
to prompt you before it replaces an existing filename. 
passwd 
Changes your password 
ping IP address or  alias 

  73 
2.6. FIRMWARE DESCRIPTION CHAPTER 2. SYSTEM OVERVIEW

Sends a ping packet to another system; this provides information concerning the time it takes 
for information to make the round trip to the other computer; it will also tell you whether 
the other system is on‐line at that time 
ps 
Displays information about your processes/jobs/programs which are running on the   system 
rm  [‐i]  [‐r] filenames 
Removes or deletes files. ‐i option asks you to confirm that you want to delete each file ‐r 
option is dangerous because it allow you to delete an entire directory and all of the files it 
contains rmdir directory Removes a directory; you can use the ‐i and ‐r options which are 
described in the rm command 
tail  [‐r]  [‐lines] filename 
Displays the last few lines of a file 
‐r displays the lines in reverse order 
‐lines specifies the number of lines, starting at the end of the file, you want to see 
touch [‐a] [‐c] [‐m] [date]  filenames 
Changes the date and time for a file without changing the content of the file 
a changes only the date and time the file was last accessed 
c doesn’t create a file if it does not already exist 
m changes only the date and time the file was last modified 
date specifies the date and time to give the file in the mmddhhnn format (month, day, hour, 
minute) 
touch with a new filename will create a new, empty file. 
traceroute IP address or alias 
Provides information concerning the route which packets must take to get from your com‐ 
puter to a remote computer/server; typically used to diagnose possible problems in packet 
routing 
vi 
VI is a text editor. Further information concerning the editing commands for VI can be 
found in a following section. 
w Provides information concerning who is logged into the system and some details on how 
they are  connected 
who Tells you who is using the system at that time 
 
2.6.3 The  VI Editor 
VI is a standard text editor found on Linux systems. With VI you can create new files or edit 
existing files. To start VI you should type ’vi filename’ where filename is the name of a new 

  74 
CHAPTER 2. SYSTEM OVERVIEW 2.6. FIRMWARE DESCRIPTION

or existing file which you wish to edit. Use the arrow keys to scroll through the document. 
The following is a list of the commands which are used in VI to edit documents. For a more 
complete description of the use of the VI editor, check out the section concerning VI in the 
Linux Users’ Guide written by Larry Greenfield. 
This is a partial list of common VI commands. To enter one of these commands, press the 
ESC key then the command key which is listed below: 
a appends text after the cursor 
A appends text at end of the line 
b back up to beginning of word in current line 
d delete line 
e move to end of word 
G go to end of file 
h move the cursor to the left 
H move the cursor to the top of the screen 
i insert text before the cursor 
I insert text at beginning of line 
j move the cursor down 
k move the cursor up 
l move the cursor to the right 
o open a line below the current line 
O open a line above the current line 
r replace character at cursor with the next character typed 
u undo the last change made 
U restore current line, discarding changes 
w move to beginning of next word 
x delete character at cursor 
X delete character before cursor 
y Yank (or copy) operator 
Y make copy of current line 
$ go to end of line 
:q! quit VI without saving changes 
:wq write changes to file then quit VI 

  75 
2.6. FIRMWARE DESCRIPTION CHAPTER 2. SYSTEM OVERVIEW

2.6.4 Updating Linux 

The Linux distribution provides for field updates of the kernel, init filesystem, boot loader, 
and files on the primary System SDHC flash card. 
Updates can be installed by placing update files into the directory 
/opt/kmi/updates/pending 
When the system is rebooted it will attempt to install updates found there in numeric sorted 
order. Updates will follow a naming convention  like: 
Rock2‐3.2‐Update‐4 
The results of updates are kept in the file: 
/opt/kmi/updates/update.log 
If an update succeeds it will be moved to the directory: 
/opt/kmi/updates/applied 
You may then delete the applied patches if you wish. 
If an update fails, the update system will be suspended until the following file is removed: 
/opt/kmi/updates/update.failure 
The log may be consulted for further details about the failure. 
Basically, you need to fix the problem, and then delete the update.failure file. The system 
will resume updates at the next reboot. 
 
2.6.4.1 Milestone and Point Updates 

Linux updates for the Rock+ will be either a Milestone update or a Point update. The 
differences between the two are as follows: 
 
• A Milestone update is a “roll‐up” and contains all fixes and updates of all previous 
updates. It is therefore often quite large. You can install the latest Milestone update 
at any time over any previous update. So you can for example, install update 3.0 over 
previous versions 1.8 or 2.1. 

• A Point update only includes changes since the last Milestone update. This is done to 
keep intermediate minor releases small in size.  However, this means that a Point update 
does not contain all previous updates and you can only install within the Milestone 
release. Do you can for example install update 3.2 over update 3.0, but you cannot 
install update 3.2 over update 2.1. 

  76 
CHAPTER 2. SYSTEM OVERVIEW 2.6. FIRMWARE DESCRIPTION

2.6.5 Software Versions 
There are several different software components in your system and each has its own versions. 
The versions of each will be listed on the data sheet that came with your unit. Software 
versions include: 

• The Power Supply Bootloader 
• The Power Supply  Application 
• The Main Processor’s Bootloader 
• The Boot  Script 
• The Linux Kernel 
• The Linux File System 
• The Java virtual machine (JVM) version 
• The GPS hardware and firmware  versions 
• The Rock+ FPGA  version 
• Rockhound  version 
The versions command, available at the Linux prompt will list the most important software 
versions (of those available to Linux) as in the example below. In order, they are: 

• The GPS hardware and firmware  versions 
• The Linux file system version 
• The Rock+ FPGA  version 
• The Linux kernel version 
• Linux update (patch) level 
• The Rockhound version 
[root@Rock:~]# versions 
KMI GPS MOD NEO‐7N‐0, ROM BASE 1.00 (59842) Jun 27 2012   17:43:52 
KMI Rock2 Arm filesystem 0.20 Beta 
KMI Rock2 FPGA rev 0x0 
KMI Rock2 Kernel, build: Nov 11 2014, 15:19:11 
KMI Rockhound 3.11 
The process of updating each software component will be described within the section that 
describes the package itself. For example, the process of updating the Power Supply Appli‐ 
cation will be discussed within the section that discusses the Power Supply. 

  77 
2.6. FIRMWARE DESCRIPTION CHAPTER 2. SYSTEM OVERVIEW

2.6.6 Software Development  Tools 

Software Development Tools Available for the Rock+ digitizers includes: 

• A C/C++ Compiler 
• A Perl interpreter 
• A Python interpreter 
• A JVM 
• A Rockhound Software Development Kit 
 
NOTE: The C compiler is covered under GPL, and as such is available 
through Kinemetrics for a distribution fee as is and without support.   If 
you  require  support  of  any  kind  ‐  including  installation,  setup  and 
configuration or programming assistance, you must purchase a sup‐ port 
contract. The compilers are available through Kinemetrics in executable 
form only, and are not guaranteed to be bug     free. 
 
These software tools, if not included in your product, are available through Kinemetrics or 
Kinemetrics’ partners at additional cost.  Please contact Kinemetrics for further information. 

  78 
 
 
 
 

Chapter  3 

Basic Operation 
 
The digitizer can operate stand‐alone, requiring only power and sensors to operate. Beyond 
these basics, if you provide a GPS antenna connection for accurate timing and a network 
connection, then the digitizer really is able to provide tremendous additional capabilities. 
The digitizer is designed for use with a network, and will serve you best in that environment. 
This section describes how to set up the Digitizer and perform the basic operations. 
 
3.1    Initial Setup 
The digitizer can operate stand‐alone, requiring only power and sensors to operate. Beyond 
these basics, if you provide a GPS antenna connection for accurate timing and a network 
connection, then the digitizer really is able to provide tremendous additional capabilities. 
The digitizer is designed for use with a network, and will serve you best in that environment. 
This section describes how to set up the Digitizer and perform the basic operations. 
Initial setup of the digitizer is done using a console cable. This connects the digitizer’s RS‐
232  console  connection  to  your  computer  so  that  you  can  do  preliminary  setup.  Once 
done, the console cable will rarely be used, so a single console cable is usually sufficient to 
maintain a large number of digitizers. 
You’ll also need a PC or equivalent running a terminal emulation program such as HyperT‐ 
erminal or PuTTY on Windows or minicom on Linux. 
You may need to use a USB to serial adapter if your PC does not have an available COM 
port. If so, install the software and then plug in the adapter, making note of the COM port 
assigned to the port. 
Kinemetrics includes the PuTTY application as an alternative to using HyperTerminal on 
Windows, or if you are using an Operating System (such as Windows 7) that does not supply 
HyperTerminal. Please see the Software Tools section that includes a basic description of 
PuTTY setup and operation. 
You should set up your RS‐232 communication software (such as HyperTerminal) for 115200 
baud, no flow control, no parity, 1 stop bit.  After power on, you should be able to get   a 

  79 
3.1. INITIAL SETUP CHAPTER 3. BASIC OPERATION

login prompt from the digitizer. Initially log into the system with the username “root” and 
the password “kmi”. You are now logged on to the Linux operating system on the Rock+. 
 
3.1.1 Network Address 
The first thing that you’ll need to do is to set (or determine) the network address of the 
digitizer.  You  need to do this even if your digitizer will not normally be connected to    a 
network.  Network  access  is  needed  to  retrieve  files  even  if  you  need  to  retrieve  them 
manually and locally. 
If you are unfamiliar with TCP/IP networking, we suggest you review the IP Primer section. 
At this point the Ethernet connection should be connected to your network. If the Ethernet 
link lights do not appear, give the following commands to start the Ethernet interface: 

[root@Rock:~]# ifdown ‐a 
[root@Rock:~]#  ifup ‐a 
 
The digitizer is configured at the factory for DHCP. This means that it expects to get its 
network address from a network server, which can be a local LAN or can be your Internal 
Service Provider. If you plan to use DHCP in regular operation, then you only need to know 
the IP address assigned by DHCP. You can find this out by typing: 

[root@Rock:~]# ipaddr eth0 
10.0.1.153 

The IP address is also shown in the sysinfo display as shown below. 
If you need to specify an IP address, then once logged into the console, use the NETCONFIG 
script. 
In either case, make a note of the current IP address. You will need it later. 
The digitizer can also act as a DHCP Server, which can be useful if you will occasionally 
connect to a standalone digitizer with something like a laptop computer. See the section on 
Non‐Networked Use . 
See the section Non‐Networked Use  if you don’t intend your digitizer to normally be con‐ 
nected to a network. 
 
3.1.2 Viewing System  Information 

Use the sysinfo command to view System Information: 

[root@Rock:~]# sysinfo 
Hostname:  Rock2 
Unit  Tag  No: 109 

  80 
CHAPTER 3. BASIC OPERATION 3.2. BASIC SETUP

Int Temp  C:  29.87 
Int Humidity: 34.05 
Voltage:  15.463 
Amperes: 0.253 
Watts  : 3.912 
OS  Time:  Thu  Dec  4  01:50:58  UTC  2014 
PSU  Time:  Thu  Dec  4  01:50:55 2014 
Up‐Time: 01:50:59 up 2 days, 7:51, 2 users, load average: 1.31, 0.75, 0.56  
eth0 IP Addr:  10.0.3.50 
Services Up: cron inetd sshd 
Via  inetd:  ftp 
MemFree: 736372 kB 
SwapFree:  65532 kB 
Filesystem Size Used Avail Use% Mounted on 
rootfs 1.6G 551M 869M 39%    / 
/dev/mmcblk0p3 2.3G 215M 1.9G 11%  /mnt/sysrw 
/dev/mmcblk0p1 65M 4.2M 58M 7% /boot/uboot 
/dev/mmcblk1p1 32G 47M 30G 1% /mnt/data1 

Use the versions command to view the major software versions: 

[root@Rock:~]# versions 
KMI GPS MOD NEO‐7N‐0, ROM BASE 1.00 (59842) Jun 27 2012   17:43:52 
KMI Rock2 Arm filesystem 0.20 
KMI Rock2 FPGA rev 0x0 
KMI  Rock2  Kernel,  build:  Nov  11  2014, 15:19:11 
KMI  Rockhound 3.11 
 
3.1.3 The Web  Interface 
The web interface to the digitizer allows you to configure and operate your digitizer using a 
web browser without installing any Kinemetrics‐specific software. The web browser should 
be HTML 1.1 compliant (or later) and should support Java and Java applets. 
See Web Interface Overview . 
 
3.2    Basic  Setup 
Although Kinemetrics makes every effort to make the digitizer useful out of the box, there 
are some items that must be set because they depend on how you will use your system and 
what sensors will be attached. 
The parameter configuration described here is done using the web interface. It can also be 
done using the RockTalk program through the network or through a Modem Only connection. 

  81 
3.2. BASIC SETUP CHAPTER 3. BASIC OPERATION

3.2.1 Sensor  Groups 

The digitizer’s physical channels are organized into groups of 4 channels each. Each group 
is capable of supporting different sensor types, so you must select the types of sensors which 
will be used within each group. Selecting the physical channels and sample rates defines the 
virtual channels that will be used in most of the rest of the system. This is accessed through 
the System Operation parameters for the Rock2 Data Interface: 
 

Figure 14: Sensor Groups 

 
After completing your changes, be sure to press OK to save your changes. 

  82 
CHAPTER 3. BASIC OPERATION 3.2. BASIC SETUP

3.2.2 Voltage Ranges 

Next, you must set up the voltage ranges. The digitizer is capable of supporting sensor 
input voltage ranges of +/‐2.5V, +/‐5.0V, and +/‐20V. This selection must be made for every 
virtual channel, and is made in the Hardware parameters section: 
 
NOTE:  Input Voltage Ranges always  cause confusion!  As the input  to 
digitizers  is  normally  differential  so  this  is  what  each  voltage  range 
represents: 

+/‐2.5V Range – The + Input Swings from +1.25 to ‐1.25 V as the –Input Swings from 
‐1.25V to +1.25V, the Sensor thus swings from +2.5V to ‐2.5V this is sometimes referred to 
as +/‐2.5Vpeak Differential or 5Vpp (Peak to Peak) Differential.  This range can be used 
with single ended +/‐2.5Vpeak sensors, such as older generation Force Balance sensors. 
+/‐5V Range – The + Input Swings from +2.5 to ‐2.5 V as the –Input Swings from ‐2.5V to 
+2.5V, the Sensor thus swings from +5V to ‐5V this is sometimes referred to as +/‐ 5Vpeak 
Differential  or  10Vpp  Differential.  This  range  is  recommended  for  use  in  structural 
monitoring systems with EpiSensors configured for a +/‐5V peak differential output. 
+/‐20V Range – The + Input Swings from +10 to ‐10 V as the –Input Swings from ‐10V   to 
+10V,  the  Sensor  thus  swings  from  +20V  to  ‐20V  this  is  sometimes  referred  to  as  +/‐ 
20Vpeak Differential or 40Vpp Differential. This range is used by the majority of Broadband 
sensors and offers the highest resolution performance from the Rock+ Digitizer. 
 
NOTE: Internal Deck. The +/‐5V Range is used for the standard +/‐ 
2g range. In order to operate the unit with an effective +/‐1g range, 
set the voltage range to +/‐2.5V. (Obsidians can be ordered with an 
optional 4g/2g Full Scale Range in which case the +/‐5V Range is 
used for the standard +/‐4g range and the +/‐2.5V range would be 
used for the +/‐2g range.) 

There are always 24 bits of ADC resolution, so it would break down like this: 
 

Range  ADC Counts  uV/ct 


2.5  8388608  0.298923223 
5.0  8388608  0.59604644 
20.0  8388608  2.384185791 
Conversion of the channel data to Engineering units further requires that you know the 
ʺsensitivityʺ value, which is the sensor’s conversion of Volts/g. This conversion is done as 
counts/8388608*FullScale/Sensitivity 
So for example: 

• Full scale = 20V 

  83 
3.2. BASIC SETUP CHAPTER 3. BASIC OPERATION

• Sensitivity = 10V/g (2g sensor) 
• Counts = 2097152 (one‐quarter scale) 

• Value in g = 2097152/8388608*20/10 = 0.5g 
 
 

Figure 15: Unit Conversion 

 
After completing your changes, be sure to press OK to save your changes. 
 
3.2.3 Sensitivity 

The Sensitivity parameter defines the voltage sensitivity of the sensor. The sensitivity value 
is normally seen as a “Volts per g” value, and is related to the Voltage range. For example, if 
your voltage range is +/‐20V, and you have 4g sensors (a range of +/‐4g), then the sensitivity 

84 
CHAPTER 3. BASIC OPERATION 3.2. BASIC SETUP

is 5V/g. This selection must be made for every virtual channel, and is made on the Hardware 
parameters  section: 
 

Figure 16: Sensitivity 

 
NOTE:  Internal  Deck.  Regardless  of  the  voltage  range  setting,  the 
sensitivity should be set according to the Internal Deck’s configuration 
sheet in V/g. (Nominally 1.25V/g for a 4g sensor, 2.5V/g for  a  2g sensor) 
After  completing  your  changes,  be  sure  to  press  OK  to  save  your 
changes. 

  85 
3.2. BASIC SETUP CHAPTER 3. BASIC OPERATION
3.2.4 Obsidian Internal Deck Zero Adjustment 

Generally, as the sensors are zeroed at the factory, and if the unit has been leveled, zero 
adjustment will probably not be required. If required you can zero the accelerometers by 
using the waveform viewer in the web interface to monitor the offsets as you manually 
adjust them. Generally if the offset is less than 50mV (62.5mG) zeroing is not required. If 
adjustment is required proceed as detailed below. 
 
3.2.4.1 Adjustment  Instructions 

• Ensure the unit is operating and you are connected to the unit using the Web Interface. 
• Start the Waveform Viewer to display the mean voltages of the EpiSensor channels in 
millivolts. As shown: 
 

 
 

Figure 17: Waveform Display 

 
• If adjustment is required remove the adjustment hole screw as shown in the figure 
below.  The top adjustment  hole adjust the  Z  Axis (Vertical),  the  hole  in  the  back 
adjusts the N Axis (North), while the hole on the right side adjusts the E Axis (East). 

  86 
CHAPTER 3. BASIC OPERATION 3.2. BASIC SETUP

Figure 18: Zero Adjustment 

 
CAUTION:  The  screw  on  the  left  hand  cover  of  the  unit  is  not  an 
adjustment  hole.  Do  not  remove  this  screw  and  do  not  insert  the 
adjustment tool into this hole as your unit will be severely damaged and 
significant fault currents can flow in the tool! 
 
The drawing above shows where to insert the adjustment tool (KMI P/N 112015) for each 
EpiSensor channel. As the figure indicates, the vertical module produces a positive output 
when the adjustment screw is turned clockwise, the horizontal modules produce a negative 
output when the adjustment screw is turned clockwise. Monitor this with the Waveform 
Viewer – remember that there is some response latency in the Viewer. 

• When the channel is adjusted, gently remove the tool. 
• After adjusting all the channels, replace the screws carefully to ensure the sealing 
washers are not damaged. 
• Important: Give the cover a moderate tap with a screwdriver handle after replacing 
the screws. The adjustment screws can have a residual stress that may cause an offset 
in data during a large earthquake. Tapping it will eliminate the residual stress. 
• Check that the sensors are still zeroed. 
 
3.2.5 EpiSensor2 Support 
The  Rock+  includes  support  for  the  EpiSensor2.  This  includes  two  new  sensor  scripts, 
EpiSensor2FS and EpiSensor2FS‐User. 
The script EpiSensor2FS has the digitizer in control, so if all three physical channels are set 
for 20V range and the same sensitivity then the Obsidian will use the sensor control lines to 
change the sensor to match the selected Obsidian parameter setup. 

  87 
3.2. BASIC SETUP CHAPTER 3. BASIC OPERATION

The script EpiSensor2FS‐User has the sensor in control, so if all three physical channels are 
set for 20V range and the same sensitivity then the Obsidian will adapt to match the sensor 
any time that the sensor’s range is changed. 
If  the EpiSensor2FS  script is used  (meaning that  the  digitizer is  to be  in  control  of the 
sensor) and the user changes the sensor’s sensitivity using the push‐button then the digitizer 
will realize this and “switch it back” within several seconds to match what is configured in 
the digitizer. 
Both scripts generate a 60 second white noise signal for calibration. 
The EpiSensor2’s serial port can be accessed through ʺminicomʺ, which is a terminal program 
built into Linux. The basic process is as follows: 

• Plug the cable into Serial1, Serial2 or Serial3 
• From a Linux prompt, type ʺminicomʺ 
• If needed, configure the port as follows: 

◦ Ctrl‐A, 0 (ʺoʺ, not zero) 
◦ Select Serial port setup 
◦ Use A to set port to /dev/serial1 (for example) 
◦ Use E to set baud to 9600, control to 8N1 
◦ Use F & G to disable hardware and software flow control 
◦ Save setup as dfl (default) 
◦ Exit 

• Now go ahead and type to talk directly to the EpiSensor2 
• Use Ctrl‐A X to exit minicom 
• You can also use Ctrl‐A Z to get further minicom help 

NOTE:  NOTE:  Next time  you  run  minicom,  you  will  not  need  to 


change the serial port setup  unless  you  want to  use a different port. 
 
3.2.6 Other Channel  Parameters 
Many other channel parameters listed in the Hardware parameters section are notational 
only, meaning that they are included in telemetry and output file information (depending 
on the formats used), but have no direct effect on the data itself. Notational parameters 
include Channel ID, Sensor Type, Sensor SN, Natural Frequency, Damping, Gain, Altitude, 
Azimuth, Offsets North, East, and Up, Location Code, Network Code, etc. Remember that 
these values are defined for each virtual channel. 
After completing your changes, be sure to press OK to save your changes. 

  88 
CHAPTER 3. BASIC OPERATION 3.2. BASIC SETUP

3.2.7 Physical vs.  Virtual Channels 

It is important to understand the difference between physical and virtual channels. There is 
a physical set of sensor inputs coming into the digitizer, and virtual channels that represent a 
physical channel at a specified sample rate. The digitizer can produce more virtual channels 
than it physically has as inputs. For example, a four channel unit could be configured in the 
following  ways: 
Example 1. Produce a virtual channel corresponding to each physical channel: 
 

Physical  Virtual 
1  1 – Physical 1 at 100sps 
2  2 – Physical 2 at 100sps 
3  3 – Physical 3 at 100sps 
4  4 – Physical 4 at 100sps 
Example 2. Produce data from one channel at multiple sample rates: 
 

Physical  Virtual 
1  1 – Physical 1 at 10sps 
1  2 – Physical 1 at 20sps 
1  3 – Physical 1 at 50sps 
1  4 – Physical 1 at 100sps 
Example 3. Produce data from each physical channel at multiple sample rates: 
 

Physical  Virtual 
1  1 – Physical 1 at 10sps 
1  2 – Physical 1 at 100sps 
2  3 – Physical 2 at 10sps 
2  4 – Physical 2 at 100sps 
3  5 – Physical 3 at 10sps 
4  6 – Physical 3 at 100sps 
4  7 – Physical 4 at 10sps 
4  8 – Physical 4 at 100sps 
You select the sample rates and the produced sample rates on the System Operation param‐ 
eters for each channel, also called Channel Mapping: 

  89 
3.2. BASIC SETUP CHAPTER 3. BASIC OPERATION

Figure 19: Channel Mapping 

 
After completing your changes, be sure to press OK to save your changes. 

NOTE: To define more virtual channels, you must rebuild the layout 
to allow for more channel as shown in section Adding Channels. 
 
3.2.8 State‐of‐Health  Streams 
In addition to the sensor input channels discussed above, the system can produce several 
State‐of‐Health streams (also called SOH). These can include such things as system voltages, 
temperature, humidity, current, and time quality. These streams are produced at 1sps, so 
to record them; you’ll need a data format that can record data to 1sps. The SOH streams 
are selected in the System Operation parameters, just below the Channel Mapping. 
After completing your changes, be sure to press OK to save your changes. 

  90 
CHAPTER 3. BASIC OPERATION 3.2. BASIC SETUP

3.2.9 Trigger Levels 

Trigger levels are the level at which the system will decide that a channel is triggered and that 
it should contribute that channel’s votes toward triggering the entire system. By default, 
the system uses threshold triggers, which are set as a percentage of full scale. So if your full 
scale range is +/‐20V and your threshold is set to 2%, then your trigger level is 2% of 20V, 
or 0.4V. Trigger levels are set in the System Operation parameters: 
 

Figure 20: Trigger Levels 

 
Please note that the trigger levels and votes apply to virtual channels. 
After completing your changes, be sure to press OK to save your changes. 

  91 
3.2. BASIC SETUP CHAPTER 3. BASIC OPERATION

3.2.10 Voting Options 
There are two different Voters available in the system. The first is the traditional Voter used 
in previous generation Altus digitizers, and the other is the CGS Voter that works similarly 
to the standard Voter, but differs in how detriggering is done. 
In either Voter, a channel is considered triggered if it has filtered data values that exceed the 
specified level specified for that channel. Correspondingly, a channel is considered detriggered 
if NO values exceed the specified level. 
Note that the levels used change depending on whether the system as a whole is triggered 
or not. If the system is not triggered, the filtered data values are compared against the 
TRIGGER levels, whereas once the system has been triggered they are compared against 
the DETRIGGER  levels. 

• The standard Voter counts triggers and uses that count to determine detriggering 
by comparing votes with the detrigger voting threshold. If the number of votes of 
triggered channels do not exceed the detrigger threshold, then the system will detrigger. 
Exceeding the detrigger threshold will cause the system to remain triggered. 
• With this Voter, the system detriggers when not enough channels exceed the trigger 
threshold (number of votes) to maintain the trigger. 
• The CGS Voter works differently in that it requires that each channel contribute enough 
detrigger votes to exceed the detrigger voting threshold. If the number of detrigger 
votes do not exceed the threshold, the system remains triggered. If the number of 
detrigger votes exceeds the detrigger threshold, then the system will detrigger. 
• With this Voter, the channels accumulate detrigger votes when they fall below the 
detrigger level.  The system will detrigger when enough detrigger votes accumulate 
that the number of detrigger votes exceeds to votes needed to detrigger the system. 
 
3.2.11 Other  Parameters 
Other basic parameters that are worth reviewing are: 

• Voter  parameters 
• Digitizer pre‐event and post‐event times 
• Channel trigger and detrigger  votes 

After completing your changes, be sure to press OK to save your changes. 
 
3.2.12 Activating Parameter  Changes 
Once you have completed your parameter changes, activate the completed parameter changes 
by selecting “Apply Changes” from the Layout menu. 

  92 
CHAPTER 3. BASIC OPERATION 3.2. BASIC SETUP

3.2.13 Passwords 
All digitizers are shipped with the same default passwords. Before deployment, Kinemetrics 
recommends that you change all passwords to something meaningful to you and make a 
record of all passwords assigned for later reference. Default passwords are assigned as follows: 
Linux: 
• root : kmi 
• kmi : kmi 
• admin : kmi 
• client : kmi 
Rockhound: 
• Console: kmi 
• Web service: rock : kmi 
You should also review the enabled IP services to make certain that the services are enabled 
that  you  need  and  want.  For example,  do  you want FTP, TELNET,  and  other  services 
enabled? 
 

3.2.14 Locked File System  
The Obsidian and Etna2 file systems are much more locked down than earlier systems such 
the Basalt. This is for additional protection of the files from accidental corruption. 

To manually edit system files on these (including changing the passwords), you much first 
unlock the filesystem: 

 
fsunlock
<make your edits>
fslock

The file system is automatically relocked on reboot if you forget. 

Convenience scripts such as netconfig will automatically unlock and relock the file system.  
   
 
3.2.15 Save  Parameters 
Once you’ve configured Rockhound for how you intend to operate your system, you should 
save your configuration. This saved copy of the parameters will be  used to restore the 
Rockhound parameters in case they become corrupted. This is done in two ways: 
• From a Linux login, use the command rhsave to save a copy of your parameters within 
the system that can be restored in case your parameters become corrupted. This is 
done as follows: 
• cd  /usr/rock/SMARTSDist/bin 
• ./rhsave If the unit’s configuration is destroyed, the system will revert to the copy 
of the configuration most recently saved by rhsave, or to the factory default if the 
configuration was never saved using rhsave. 
• From the web interface use the Advanced Features function Administrative Details to 
download a copy of the parameters to your PC. At the top of the Administrative Details 
screen, there is a link “Click Here to Download Parameters”, from the instrument. 
• Kinemetrics recommends that you save a copy of your Rockhound parameters any time 
that you make significant changes to the parameters so that you can quickly restore 
them in the event of loss. 
• See Save/Restore Parameters 
 

93 
3.3. TRIGGERED RECORDING CHAPTER 3. BASIC OPERATION

3.3 Triggered Recording 

3.3.1 Pre‐Event Time 
The setting allows you to determine how many seconds of data before the trigger criteria 
were met will be recorded in the event file. For strong motion applications this is normally 
set to a few seconds so you can determine the noise before the start of the event. The factory 
default is ten seconds. 
For weak motion recording, especially if you expect the system to be triggered on the S‐ 
waves, the situation is more complex. In this case, the pre‐event should be sufficiently long 
to allow any P‐waves from an event within the region of interest to be recorded. This time 
can be estimated by the travel times of the P and S waves from the most distant point of 
interest. The difference between these times, with some allowance for a true pre‐event time, 
gives the setting for the pre‐event time. 
 
3.3.2 Post Event Time 
The post event time determines how many seconds after the system has de‐triggered will be 
recorded in the file.  It also determines how likely events are to be split into separate files. 
This is because if the system re‐triggers during the post event the file will just be extended. If 
the post event is set too short, several files could be created from the same event. We set the 
factory default at ten seconds, which is a reasonable value for strong motion recording. For 
weak motion recording this time should be set according to the goals of the study. Generally, 
a time of 30 seconds or more should be considered. 
 
3.3.3 Minimum Run Time 
This is the minimum time the recorder will record once an event is triggered.  Formerly, it 
was used to ensure a complete sequence of time code was recorded with the event, but this 
is not required anymore. Generally, the pre‐event and post‐event now give sufficient control 
over the event timing. For this reason we set the factory default to zero seconds. 
 
3.3.4 Channel Triggering 

Although you will need to read most sections to determine what is appropriate for your 
application, we have split the triggering system into two sections. If you are just interested 
in recording ʺstrong motionʺ events when the ground or structure shakes significantly, you will 
find instructions on setting triggers in the Triggering for Strong Motion Recording section. 
If you are interested in recording weak events that are very close to the local seismic noise, 
you should read Triggering and Recording Weak Motion Applications. After you read the 
relevant section you will be able to set up each channel’s triggering appropriately for your 
application. 

  94 
CHAPTER 3. BASIC OPERATION 3.3. TRIGGERED RECORDING

3.3.4.1 Triggering in Strong Motion Applications 

For strong motion recording you will want to record the strongest motion from an earthquake 
that can be felt and possibly cause damage to buildings and other structures. Normally, a 
simple threshold trigger will be  sufficient to reliably trigger the recorder. As these are 
sensitive instruments, it is also possible to record much weaker motions using the threshold 
trigger. The tradeoff to consider is how you will retrieve the data and how to ensure that 
there will be room for the ʺbig oneʺ if you have very sensitive threshold trigger levels. 
The threshold trigger has two parameters for each channel. The first is the threshold trigger, 
which is the level in percent of full scale that causes the channel to trigger. The default value 
for this is 2%. The second parameter is the threshold de‐trigger. This is the value in percent 
of full scale the signal must fall below after triggering for the channel to detrigger. The 
default value is 2%. The detrigger parameter can be used for extending the recording time 
by setting it to a smaller value than the threshold trigger value. 
 
3.3.4.2 Alarms 

The recorder has an additional set of thresholds called alarm threshold parameters. These 
are specified as a percentage of the full‐scale input and can be set independently for each 
channel. They are set in the channel trigger parameters window. When a channel’s triggered 
filtered data exceeds this threshold, the hardware alarm is activated. 
 
3.3.4.3 Triggering in Weak Motion  Applications 

The recorder includes the following features for use in seismological applications: 

• Three different trigger filters that allow the trigger band of interest to be optimized. 
• STA/LTA triggering to support the recording of small amplitude events. The use of 
these specific features is discussed below. 
 
3.3.4.4 Trigger Filters 

Your selection of a pre‐trigger filter is determined by your application, by seismic noise 
conditions  at  the  site,  and  by the  type  of  sensors  installed  with  the  recorder.  The  pre‐ 
trigger filter pass‐band should encompass the maximum energy of expected seismic events. 
The filter you select should have a pass‐band that doesn’t coincide with the peak frequency 
components of seismic noise at the site, thus discriminating against seismic noise. 

  95 
3.3. TRIGGERED RECORDING CHAPTER 3. BASIC OPERATION

Figure 21: Typical trigger filter response 

 
The frequency response function of the seismic sensor modifies event and noise signals and 
thus is an important factor in your choice. If the frequency content of events and of seismic 
noise occur in the same frequency band, the trigger filter will be inefficient. 
The recorder has three band‐pass filters with different low‐ and high‐corner frequencies. 
Note that the corner frequencies scale according to the sampling rate. Some examples are 
as follows: 
An IIR‐A filter with a 100 Hz sampling rate can record regional events with broadband 
sensors. Its 0.6 Hz low‐corner frequency gives some protection to the dominant component 
of natural marine seismic noise. Its high‐corner frequency gives protection against excessive 
high‐frequency man‐made seismic noise. 
 
• An IIR‐A filter with a sampling rate of 200 or 250 Hz can monitor local earthquakes 
with efficient protection against marine  noise. 

• The classic strong motion filter at a sampling rate of 200 Hz or 250 Hz is used in typical 
strong motion applications. 

• The low‐frequency corner of a classic strong motion filter at a sampling rate of 100 
Hz can be used for far‐regional and tele‐seismic applications. It has good protection 
against 20 ‐ 40 Hz man‐made seismic noise in urban areas. 

• The classic strong motion filter at a 250 Hz sampling rate can record for short periods 
with a short‐period seismometer. In this case, the sensor itself protects against ~0.3 
Hz marine seismic noise. 

  96 
CHAPTER 3. BASIC OPERATION 3.3. TRIGGERED RECORDING

• An IIR‐C filter at a 200 or 250 Hz sampling rate can monitor local earthquakes. How‐ 
ever, there is no protection against high‐frequency man‐made noise because the fre‐ 
quency content of the events and the noise is approximately the same. The approximate 
band‐pass of these filters is shown in the table below. 
 
3.3.4.5 Trigger types and sampling  rates 
 
Trigger  Type  20 SPS  40 SPS  50 SPS 
IIR‐A  ~ 0.12‐2 Hz  ~ 0.24‐4 Hz  ~ 0.3‐5 Hz 
Classic Strong Motion  ~ 0.01‐1.25 Hz  ~ 0.02‐2.5 Hz  ~ 0.025‐3.125 Hz 
IIR‐C  ~ 0.2‐4 Hz  ~ 0.4‐8 Hz  ~ 0.5‐10 Hz 
Trigger  Type  100 SPS  200 SPS  250 SPS 
IIR‐A  ~0.6‐10Hz  ~1.2‐20Hz  ~1.5‐25Hz 
Classic Strong Motion  ~0.05‐6.25Hz  ~0.1‐12.5Hz  ~0.12‐15Hz 
IIR‐C  ~1‐20Hz  ~2‐40Hz  ~2.5‐50Hz 
 
3.3.4.6 STA/LTA  Triggering 

The short‐time average/long‐time average (STA/LTA) trigger algorithm generally increases 
the sensitivity of the recorder in comparison to the amplitude threshold trigger algorithm. 
It improves the earthquake triggers/false triggers ratio, and, to some extent allows discrim‐ 
ination among the different types of  earthquakes. 
This can minimize the work of analysts and allows more efficient use of the recorder’s data 
memory. Therefore, STA/LTA triggering is used most often in weak motion applications that 
try to record as many seismic events as possible. The STA/LTA trigger parameter settings are 
always a tradeoff among several seismological and instrumental considerations – the highest 
possible trigger sensitivity for a given type of earthquakes (including ʺall earthquakesʺ) at a 
tolerable number of false triggers. 
STA/LTA triggering is most beneficial at seismically quiet sites where natural seismic noise 
(marine noise) is the dominant type of seismic noise. 
STA/LTA triggering is also effective in cases of suddenly changing man‐made seismic noise 
(like the noise due to day/night variation of human activity in or close to an urban area). The 
STA/LTA algorithm is less effective in the presence of irregular, high amplitude man‐made 
seismic noise that is often of the burst and/or spike type. 
 
How  STA/LTA  Works  The  STA/LTA  algorithm  continuously  follows  the  changes  in 
seismic noise over time, and automatically adjusts recorder sensitivity to give you optimal 
sensitivity to the actual seismic noise level at a specific time. This results in significantly 
higher sensitivity during seismically quiet periods. 
The STA/LTA algorithm calculates average absolute amplitude of a seismic signal in two 
time windows – a Short Time Average window (STA) and a Long Time Average window 

  97 
3.3. TRIGGERED RECORDING CHAPTER 3. BASIC OPERATION

(LTA). The short time average window (STA) ʺwatchesʺ for earthquakes and measures the 
ʺinstantʺ amplitude of the seismic signals. The long time average window (LTA) takes care 
of the average value of seismic noise during the same period of time. A ratio of both values 
– the STA/LTA ratio  – is continuously calculated  and,  if it  exceeds  the preset STA/LTA 
trigger threshold level, a channel trigger is declared. The channel de‐triggers if the STA/LTA 
ratio falls below another pre‐set value – the STA/LTA de‐trigger threshold level. 
STA/LTA trigger ratio calculations are calculated according to the steps in the figure below. 
All calculations are made for every data sample in every channel in the recorder, assuring 
minimal time delay between the seismic event and the triggering. 
 
 

Figure 22: STA/LTA Simplified Flowchart 

 
Simplified flowchart of STA/LTA  calculations 
 
STA/LTA     Parameters  To set the STA/LTA  trigger algorithm, adjust   the following 
parameters: 

• STA window duration in seconds 
• LTA window duration in seconds 
• STA/LTA trigger threshold level 
• STA/LTA de‐trigger threshold level 

  98 
CHAPTER 3. BASIC OPERATION 3.3. TRIGGERED RECORDING

Adjusting STA/LTA Trigger Parameters The STA/LTA trigger parameter  settings depend 
on the goals of the application, seismic noise conditions, and the properties of earth‐ quake 
signals at a given location. Each seismic site requires detailed study, and only expe‐ rience 
will enable you to determine optimal trigger settings. 
In general, by proper adjustments of the STA and LTA duration, you can make triggering 
more or less sensitive to certain type of earthquakes, and less sensitive to the type of seismic 
noise at a given site. 
 
Short  Time  Average Window  (STA)  Duration  The  short  time  average  window  (STA) 
duration works as a kind of filter. The shorter it is, the more sensitive the STA/LTA trigger 
will be too short and high frequency signals of local earthquakes. It will also be less sensitive 
to distant earthquakes. The longer the STA duration, the less sensitive the trigger will be 
for short local earthquakes. By changing the STA duration you can prioritize to some extent 
between distant or local events. 
On the other hand, by increasing the duration of the STA window, for example, triggering 
becomes less sensitive to the common spike‐type of man‐made seismic noise, and vice versa. 
For regional events, a typical initial value of STA duration is 1 second, for local events shorter 
values of around 0.3 s are most commonly used. 
 
Long Time Average Window (LTA)  Duration  By setting the LTA  window, you can make 
recording more or less sensitive to regional events with typically slowly emergent onset waves. 
In the case of a short LTA duration, you allow the LTA value to follow the slowly increasing 
amplitude of emergent seismic waves and this results in a decreased STA/LTA  ratio  at  the 
moment when the S waves arrive, thus reducing trigger sensitivity. In the opposite case, 
using  a  longer  LTA  duration,  the  sensitivity  to  emergent  earthquakes  will  increase 
because the STA/LTA ratio will not be influenced so rapidly by the emergent seismic signal. 
On the other hand, a short LTA will adjust recorder sensitivity to relatively fast changes 
in ʺcontinuousʺ man‐made seismic noise during night‐to‐day transitions of activity in urban 
areas. Natural seismic noise (marine noise) changes much more slowly and therefore much 
longer LTA durations can be used. 
An LTA duration of 60 seconds is a good initial value. A shorter LTA duration is needed to 
exclude emergent regional events or to reject changing man‐made noise at the site. 
The recorder has a modified version of a ʺfrozenʺ long time average window (LTA) during 
events. The LTA value is, to the first approximation, not allowed to change (increase) during 
an event. This prevents truncation of coda waves due to the LTA increasing too rapidly 
during events (and thus decreasing STA/LTA ratio too rapidly). However, this ʺfreezingʺ 
is not complete, allowing the trigger algorithm to prevent the unit from being permanently 
triggered by a rapid increase in seismic noise. 
 
STA/LTA Trigger Threshold Level The optimal STA/LTA trigger threshold  ratio depends 
mainly on seismic noise conditions at the site. Higher noise levels, particularly the 

  99 
3.3. TRIGGERED RECORDING CHAPTER 3. BASIC OPERATION

man‐made kind, require higher values. Higher values result in fewer false triggers but miss 
more earthquakes. Lower STA/LTA triggers make the station more sensitive but cause more 
false triggers to be recorded. False triggers and missed events are an unavoidable reality 
when recording seismic signals in event triggered mode. 
An initial setting of 4 for the STA/LTA ratio is common for relatively quiet seismic sites. 
Higher values of 8 and above are required at less favorable sites where man‐made seismic 
noise is present. 
 
STA/LTA De‐Trigger Threshold  Level  The  STA/LTA  de‐trigger  threshold  level  (along 
with the post‐event time parameter) determines when the recording will terminate. The de‐
trigger  threshold  is  determined  through  a  selectable  STA/LTA  de‐trigger  thresh‐  old 
percentage parameter and the current STA/LTA trigger threshold  value. For example,  a 
STA/LTA  de‐trigger  threshold  level  of  4  at  an  STA/LTA  trigger  threshold  level  of  8  is 
obtained by selecting the STA/LTA de‐trigger threshold percentage of 50%. 
The STA/LTA de‐trigger threshold level determines how well the coda waves of recorded 
earthquakes will be captured.  To preserve complete coda waves, low values are required. 
On the other hand, if one is not interested in coda waves, significant savings in memory 
are possible if high values of STA/LTA de‐trigger threshold levels are used. An STA/LTA 
de‐trigger threshold percentage of 100% is the highest value possible. It makes the STA/LTA 
trigger threshold level and the STA/LTA de‐trigger threshold level equal. 
In general, the noisier the site, the higher the value of the STA/LTA de‐trigger threshold level 
that must be used to prevent continuous, repeated re‐triggering. This danger is particularly 
high at sites heavily polluted by man‐made seismic noise. 
The  recorder  will  always use  an  STA/LTA de‐trigger  threshold  level  of  at  least  1.2.  For 
example, if you select a 10% de‐trigger threshold with the STA/LTA trigger threshold level 
of 4, you would theoretically get 10% of 4 or 0.4 and the unit would be continuously triggered. 
However, the recorder automatically uses a minimum STA/LTA de‐trigger threshold level  of 
1.2 to prevent this continuous triggering. 
A typical initial value of the de‐trigger threshold is 2 to 3. The percentage set must be 
calculated taking into account the current STA/LTA trigger threshold value. For example 
with an STA/LTA trigger threshold of 4, and a desired STA/LTA de‐trigger threshold of 2 
you would need to set an STA/LTA de‐trigger threshold percentage of 50%. 
 
Default Values     STA/LTA trigger algorithm default parameters are: 

• STA (short time average) window duration = 1 sec 
• LTA (long time average) window duration = 60 sec 
• STA/LTA trigger threshold level = 4 
• STA/LTA de‐trigger threshold percentage = 40%, resulting in an effective STA/LTA 
de‐trigger threshold level 1.6 

  100 
CHAPTER 3. BASIC OPERATION 3.3. TRIGGERED RECORDING

Trigger Voting  Parameters  Each channel can be assigned a number of votes that it  may 
cast towards getting the system to trigger. This is called ʺvoting.ʺ How the voting system 
is set up is dependent on which signals you are trying to record and which you are trying 
not to record.  You may need some first‐hand experience with the conditions at the  site 
before you can optimize this triggering.  The parameters used to set up the triggering are 
explained below. 
 
Channel Weights (Trigger/Detrigger Votes) The mechanics of this  are  relatively  simple. 
You select the number of votes each channel will contribute (when it is triggered) to the total 
number of votes required to trigger the system. Give zero votes to a channel that you do 
not want to affect the triggering. Give a positive number of votes to a channel you do want 
to  contribute  to  the  triggering.  Give  negative  votes  to  a  channel  you  wish  to  inhibit 
triggering. 
 
Trigger Weight     This is the total number of votes required to get the system to trigger. 
 
Keyboard Votes If you want to trigger the unit from a keyboard for test purposes, give 
the keyboard trigger the same number of votes as the total trigger weight so that it will 
trigger the unit by itself (without any other channels being triggered). 
 
External Votes This is the number of votes you assign to the external hardware trigger 
source. If you want all units in the network to trigger when one unit triggers, assign it the 
same number of votes as the trigger weight. If you want to use a combination of an external 
trigger with other internal criteria, set the votes appropriately. 
 
Examples of Voting Schemes 
 
Classic  Free‐Field  Site:  Normally  uses  a  three‐channel  internal  EpiSensor  with  no 
interconnected units. Each channel has 1 vote and requires only 1 vote to trigger. Any 
channel could trigger the system. 
 
Interconnected  System:  For  example:  two  digitizers  with  internal  EpiSensors  in  a 
building; one in the basement and one on the roof. The digitizers are interconnected. Initially 
we set 1 vote for each channel and the external trigger and 1 vote to trigger. We find that the 
digitizer on the roof often triggers the system due to the building swaying in the wind. We 
can compensate for this by changing the voting of the roof unit so that its three accelerometer 
channels have 1 vote while the external trigger has 3 votes. We set the trigger weight to 3 in 
this unit. Now all of the channels on the roof must trigger simultaneously to record or the 
unit must be forced to record by the unit in the basement triggering. 

  101 
3.4. TIMING SYSTEM CHAPTER 3. BASIC OPERATION

System on a Bridge: With a 12‐channel digitizer on a bridge with 12 sensors connected 
we could start with all channels having 1 vote and requiring 1 vote to trigger. We may find 
that this system triggers far too often from traffic vibration or swaying of the bridge. We 
could then require several sensors to trigger by setting the votes to trigger to 2 or 3 and 
could even give those aligned with the direction of sway 0 votes.  In extreme cases we can 
give the channels aligned with the direction of sway negative votes, so that several other 
channels must trigger to overcome their negative  votes. 
In conclusion, from these examples we can see the flexibility of the triggering options and 
some of the ways this flexibility can be used in particular installations. 
 
3.4 Timing  System 
Time keeping is an important part of the Rock+ digitizer and is performed at a number of 
levels: 
 
• Digitizer data time is set at data acquisition start by sending current system time to 
the ADC boards.  This time is used to time stamp data until a better time source    is 
available such as GPS or PTP (Precision Time Protocol). GPS and PTP time sources 
are very accurate – typically much less than a microsecond – and are used to time‐
stamp acquired samples, telemetry and data files. 

• Time is set and maintained in the main processor based on time received from an NTP 
server (network time protocol) by communicating with any of several default internet 
time servers. You can also specify the NTP server the digitizer will use by using the 
Linux script timeconfig. 
 
NOTE: The  NTP  protocol  DOES  NOT  provide  sufficient  accuracy or 
determinism  for precise data acquisition   timing.    It is generally used 
ONLY to set the Operating System time and to “pre‐load” the  time used 
by  the  data  acquisition  system  until  “good”  time  can  be  acquired  by 
GPS or PTP. Specific more accurate applications of NTP are described 
later in this   section. 
 
• Time is saved in the system’s internal clock at power down and will maintain time 
without any external  power  for  at  least seven  days. This time will be  restored  at 
reboot in absence of any other time. 

• Should the internal clock time source be lost, the time at boot will be the time of the 
last proper shutdown. 

• Time can be set manually from the Linux date command 

  102 
CHAPTER 3. BASIC OPERATION 3.4. TIMING SYSTEM

3.4.1 Data Time vs OS  Time 
It is important to clarify that Data Time (the time‐stamps applied to the data samples) and 
OS (Operating System) Time are close to each other, but are not the same thing. 
OS Time is only required to be “reasonably close” to Data Time so that OS time stamps of 
created files and log entries are reasonably close to reality. How close OS Time is to Data 
Time varies depending on the timing source from a few milliseconds of error to a few seconds. 
Data Time is as close as possible to absolute time. Again, this depends on the timing source. 
Accurate time sources such as GPS or PTP will typically align Data Time to less than a 
microsecond from absolute time. Less accurate time sources such as undisciplined NTP may 
only be accurate to within a few tens of milliseconds. 
 
3.4.2 Disciplined  Oscillator 
Accurate time sources in the Rock+ basically operate by having a reported time value and 
an accurate 1Hz signal. For example, with GPS a packet is received every few seconds that 
reports the current time. This is paired with a highly accurate 1Hz signal from the GPS of 
typically +/‐70 nanoseconds from absolute time. 
The Rock+ uses this accurate time to control its internal 1Hz signal to follow the external 
1Hz by adjusting the TCXO DAC. The values of the TCXO DAC and the “drift” (timing 
error) are available as SOH (State of Health) values or streams. 
 
3.4.3 Time Quality 
The Rock+ will use available information about the timing system to determine a Time 
Quality value. This value can be useful in determining the trustworthiness of the timing. 
This value varies from 0‐100%. Values are as follows: 
• 0 ‐ No timing source used 
• 10‐60 ‐ Time source lost. Degrades by 1% every 10 minutes. 
• 80 ‐ Time source identified, but not trusted yet. 
• 90 ‐ Time source producing time, but not locked. 
• 100 ‐ Time source locked. 
 
3.4.4 Timing  Options 
The Obsidian generation of Kinemetrics Digitizers supports several timing options that allow 
for “good” to “outstanding” timing accuracy for a range of applications. 
The list below includes timing options that are functional and available now in the Obsidian 
generation of instruments. For each option we list description, application, infrastructure, 
maximum distance, and  accuracy. 

  103 
3.4. TIMING SYSTEM CHAPTER 3. BASIC OPERATION

3.4.4.1 GPS Internal 

Description: Uses internal GPS engine to supply accurate timing for an individual digitizer. 
This is the default timing selection since it is often the most accurate and most cost‐efficient 
solution. 
Application: Free field installations or installations where GPS antennas can be run for each 
unit. 
Infrastructure: GPS engine is built into the digitizer. GPS cable and antenna required. 
Maximum distance:  50m antenna cable, longer with specialized extenders. 
Accuracy: Typically 1us or less assuming good sky view. 
 
3.4.4.2 NTP 

Description: Uses the NTP protocol to get time over a network connection. 
Application: Use where a GPS antenna is problematic or impossible. NTP timing is not 
intended for precision timing but can be good enough in some applications, especially single 
stations. Use of public internet NTP servers (pool.ntp.org) is the default. Alternatively you 
can provide your own local NTP server (often GPS based called stratum one) to improve 
accuracy. 
Infrastructure: A network connection with either access to the internet or a private NTP 
server. 
Maximum distance: 100m Ethernet segments, longer with repeaters or fiber‐optics. 
Accuracy: Depends on the available NTP server, but will typically settle to within a few 
milliseconds of UTC time. Because of the nature of NTP, initial timing may take several 
hours to settle. 
 
3.4.4.3 NTP w/  External 1Hz 

Description: Uses the NTP protocol to discipline the Operating System to within a few tens 
of milliseconds of UTC time, then supplements this with an externally provided 1Hz signal 
that is used to accurately discipline the sampling. 
Application: Installations where GPS antenna usage is problematic, but an NTP server and 
an accurate 1Hz signal are available. 
Infrastructure: A network connection with either access to the internet or a private NTP 
server and an accurate 1Hz signal. 
Maximum distance: 100m Ethernet segments, longer with repeaters or fiber‐optics. 1Hz 
signal as provided by the site. 
Accuracy: Depends on the accuracy of the provided 1Hz signal but typically 1us or less. 

  104 
CHAPTER 3. BASIC OPERATION 3.4. TIMING SYSTEM

3.4.4.4 PTP Slave 

Description: Uses the PTP (IEEE‐1588) protocol. Gets time from a provided (local) PTP 
master. 
Application: PTP allows for accurate timing over local Ethernet networks where a PTP 
master is available. The PTP master can be a 3rd party commercial PTP time server or can 
be another Obsidian acting as the PTP master. 
Infrastructure: Requires a network connection to the PTP master, as well as the PTP master 
itself. Timing is significantly improved if the PTP master and any intermediate Ethernet 
switches are hardware PTP compliant (see note). 
Maximum distance: 100m Ethernet segments, longer with repeaters or fiber‐optics. 
Accuracy: Using hardware PTP compliant servers and switches, typically less than 1us of 
the PTP master. When not using hardware compliant PTP hardware it can range from 
several microseconds to completely unusable. 
 
3.4.4.5 PTP Master w/  GPS 

Description: Disciplines the local unit via GPS and shares out this timing to one or more 
PTP slaves on the local network. 
Application: Use in places where units can be network interconnected, and GPS is available 
to the PTP master (such as the top floors of a tall building). 
Infrastructure: Requires network connection to one or more PTP slaves. Use of hardware 
PTP compliant hardware  recommended. 
Maximum distance: 100m Ethernet segments, longer with repeaters or fiber‐optics. 
Accuracy: GPS accuracy for master unit. 
 
3.4.4.6 PTP Master w/  NTP 

Description: Uses the NTP protocol to get time over a network connection. This timing 
information will be shared out over the network connection to one or more PTP slaves. 
Application: Use where a GPS antenna for the PTP master is problematic, but NTP time 
is available or possible. 
Infrastructure: Requires network connection with access to the internet or a private NTP 
server and one or more PTP slaves. Use of hardware PTP compliant hardware recommended. 
Maximum distance: 100m Ethernet segments, longer with repeaters or fiber‐optics. 
Accuracy: Although absolute accuracy to UTC will only be to a few milliseconds (see NTP 
above), all units will be “locked” together via PTP, typically within 1us of the PTP master. 
PTP Master accuracy to UTC is as described for NTP above. 

  105 
3.4. TIMING SYSTEM CHAPTER 3. BASIC OPERATION

3.4.4.7 PTP Master (free  running) 

Description: Shares out the PTP master’s idea of time to one or more PTP slaves. 
Application: Use where a GPS antenna is problematic and no NTP or PTP time source is 
available or possible. 
Infrastructure: Requires network connection and one or more PTP slaves. Use of hardware 
PTP compliant hardware  recommended. 
Maximum distance: 100m Ethernet segments, longer with repeaters or fiber‐optics. 
Accuracy: PTP master and PTP slaves will be locked together typically to within 1us. 
Absolute accuracy to UTC unknown since no time reference is available. 
 
3.4.4.8 TCXO free  running 

Description: Stand‐alone unit with no available time source. 
Application: Use where there is only one unit, and use of GPS antenna is problematic and 
no NTP or PTP time source is available or possible. 
Infrastructure: None 
Maximum distance: N/A 
Accuracy: Unknown since there is no time reference. 
Notes: 
 
• Maximum  distance  for  Ethernet  over CAT‐5 or  CAT‐6 cable  is  100m.  This  can  be 
extended by adding Ethernet “repeaters” (such as switches) to add segments. Note 
that repeaters added should have back‐up  power. 

• Fiber‐optic converters may be used to significantly extend Ethernet distances. Note 
that there are several fiber‐optic Ethernet standards ranging from about 400m to as 
much as 40km maximum  distances. 

• PTP infrastructure is available commercially (time servers, switches, etc.) These basi‐ 
cally come in two variants: 

◦ Hardware based PTP and Software based PTP. Hardware based PTP is much 
more accurate, on the order of less than 1us. 
◦ Software based PTP is much less accurate and is more affected by network traffic 
and so may not be as deterministic. Software based PTP solutions are typically 
in the order of several to tens of microseconds of inaccuracy but can be worse. 

• When using PTP, there should only be ONE PTP master on the local  network. 


• PTP hardware and software is built into the Obsidian and is hardware PTP compliant. 

  106 
CHAPTER 3. BASIC OPERATION 3.5. ANTI-ALIAS FILTERING

3.4.5 Using an NTP Server 
The Rock+ defaults to using public NTP servers at pool.ntp.org. To use these servers, you 
must have DNS enabled in your network configuration. 
To configure a different NTP Time server, use the “timeconfig” script from Linux: 
The timeconfig script operates as  follows: 

# timeconfig 
System time is: Thu Dec 11 15:37:06 UTC 2014  Do 
you  want  to  set  the  system  time  manually? 
(Y/N)?  > n 
Do you want to configure NTP servers? 
(Y/N)?  >  y 
IP  address  or  host  name  of  the  Primary  NTP  server 
>  www.myntp.com 
IP  address  or  host  name  of  the  Secondary  NTP  server (optional) 

Settings to be saved: 
Primary : www.myntp.com 
Secondary  : 
Tertiary : 
Quaternary : 
Press [Enter]  to  Continue,  or [Ctrl‐c]  to  Quit. 
 

3.4.6 Acting as an NTP Server 
The Rock+ can act as an NTP server for your network. This is useful when you have other 
non‐Kinemetrics equipment that needs a time reference and when NTP is good enough for 
that purpose. 
Whenever the Rock+ is configured to get time from GPS (whether configured as GPS only 
or configured as PTP Master with GPS) then the Rock+ will automatically act as an NTP 
server. No additional configuration is required ‐ simply give your other equipment the IP 
address of the Rock+ as the NTP server’s address. 
Note that the NTP time provided by the Rock+ is only as good as the definition of the NTP 
protocol allows. 
 
3.5 Anti‐Alias Filtering 
The recorder has a high‐order anti‐aliasing filter system that offers extremely steep (ʺbrick‐ 
wallʺ) roll‐off combined with decimation of data. You can select between the causal or 
non‐causal versions of these filters depending on your application. 

  107 
3.5. ANTI-ALIAS FILTERING CHAPTER 3. BASIC OPERATION

Use the final causal filter for precise seismic phase picking (this type of filter does not generate 
precursors that might interfere with determining the precise onset time). Use the non‐causal 
final filter to avoid phase distortion of the signal. 
The non‐causal filter has a linear phase response that is equivalent to a pure time delay, 
and  the  phase  of  the  seismic  signal  is  undistorted.  Therefore,  there  is  no  need  for  de‐ 
convolution of seismic signals when using these filters if you are primarily concerned with 
phase distortion. The degree to which their pass‐band amplitude characteristics modify 
seismic signals is tolerable in most seismological  applications. 
ASCII files containing the coefficients of all these filters can be downloaded from the Kine‐ 
metrics Web site or at www.kinemetrics.com. 
These filters are implemented as two to eight stage, multi‐rate, Finite Impulse Response 
(FIR) filters. They offer extremely steep low pass amplitude response. The ‐3dB point is 
at 40% of the Sampling Frequency, while at the Nyquist Frequency (50% of the Sampling 
Frequency) the amplitude is ‐130dB. The first‐stage (A) filters have sinc5 compensation. A 
30K data stream from the A/D converter is decimated by the appropriate factors to get to an 
intermediate data stream at twice the final sampling rate. The final filter is a decimate by 2 
brickwall filter, causal or non‐causal. The table below lists the final sample rate, filter name, 
number of coefficients and decimation for the anti‐alias filters used in the Rock+ Recorder 
Family. 
2000 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefB3_80 ( 179, /3 ) or coefB3C_80 ( 179, /3 ) 

1000 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefA3_50 ( 35, /3 ) 

• coefB2_80_140 ( 133, /2 ) or coefB2C_80_140 ( 133, /3 ) 

500 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefA3_50 ( 35, /3 ) 

• coefA2_20 ( 85, /2 ) 

• coefB2_80 ( 173, /2 ) or coefB2C_80 ( 173, /2 ) 

250 sps: 

  108 
CHAPTER 3. BASIC OPERATION 3.5. ANTI-ALIAS FILTERING

• coefA5_50_s5c ( 59, /5) 

• coefA3_50 ( 35, /3 ) 

• coefA2_20 ( 85, /2 ) 

• coefA2_20 ( 85, /2 ) 

• coefB2_80 ( 173, /2 ) or coefB2C_80 ( 173, /2 ) 

200 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefA3_50 ( 35, /3 ) 

• coefA5_50 ( 85, /5) 

• coefB2_80 ( 173, /2 ) or coefB2C_80 ( 173, /2 ) 

100 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefA3_50 ( 35, /3 ) 

• coefA2_20 ( 85, /2 ) 

• coefA5_50 ( 85, /5 ) 

• coefB2_80 ( 173, /2 ) or coefB2C_80 ( 173, /2 ) 

50 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefA3_50 ( 35, /3 ) 

• coefA4_50 ( 69, /4 ) 

• coefA5_50 ( 85, /5 ) 

• coefB2_80 ( 173, /2 ) or coefB2C_80 ( 173, /2 ) 

20 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefA3_50 ( 35, /3 ) 

  109 
3.5. ANTI-ALIAS FILTERING CHAPTER 3. BASIC OPERATION

• coefA2_20 ( 85, /2 ) 

• coefA5_50 ( 85, /5 ) 

• coefA5_50 ( 85, /5 ) 

• coefB2_80 ( 173, /2 ) or coefB2C_80 ( 173, /2 ) 

10 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefA3_50 ( 35, /3 ) 

• coefA4_50 ( 69, /4 ) 

• coefA5_50 ( 85, /5 ) 

• coefA5_50 ( 85, /5 ) 

• coefB2_80 ( 173, /2 ) or coefB2C_80 ( 173, /2 )  

1 sps: 

• coefA5_50_s5c ( 59, /5 ) 

• coefA3_50 ( 35, /3 ) 

• coefA2_20 ( 85, /2 ) 

• coefA4_50 ( 69, /4 ) 

• coefA5_50 ( 85, /5 ) 

• coefA5_50 ( 85, /5 ) 

• coefA5_50 ( 85, /5 ) 

• coefB2_80 ( 173, /2 ) ) or coefB2C_80 ( 173, /2 ) 

Group delay is automatically and transparently compensated by the recorder, so the time 
stamps are accurate for the first scan and the trigger scan. 

  110 
CHAPTER 3. BASIC OPERATION 3.6. TCP/IP PRIMER

SPS  Non‐causal empirical (sec)  Causal empirical (sec)  Sample Period (mSec) 

2000  0.0160  0.00265717  0.5 


1000  0.0365  0.0070  1 
500  0.1105  0.0305  2 
250  0.2385  0.0765  4 
200  0.2380  0.0405  5 
100  0.4935  0.1085  10 
50  0.9475  0.1695  20 
20  2.4025  0.4338  50 
10  4.7335  0.8475  100 
1  47.6425  9.2100  1000 

3.6 TCP/IP Primer 

3.6.1 Terminology 

IP Address – The address of a device that uniquely identifies it so it can be found on a 
network. Typically this is a four part numeric address such as 192.168.1.1. Each part ranges 
0 to 255. 
Network – A collection of devices which can communicate with each other by knowing each 
others’ addresses. 
Internal Network – An internal network (and the devices in it) cannot normally be directly 
accessed from the outside world, but they MAY be allowed to access the outside world. If you 
have one computer and a router, you have an internal network. Internal networks typically 
have the first three parts of the address the same – for example 192.168.1.x will be used for 
all devices in the network, with only the “x” part varying. 
Router – Equipment that bridges between internal networks and the Internet. A router will 
view devices on the internal network with one range of addresses, and will present a single 
address to the Internet. A router also serves to restrict access to the internal network. This 
restriction/protection is known as a firewall. 
Switch – Allows you to connect multiple networked devices together. Many routers include 
several open sockets allowing them to also act as a switch. 
Hub – A “dumb” predecessor to the switch. Do not use a hub when the network includes 
different speed devices because it will bring all devices on the network down to the speed of 
the slowest device. 
Crossover – When connecting two devices directly together without a switch (such as a 
Rock+ and your PC only), use a crossover cable or adapter to allow the direct connection. 
(Think null modem for Ethernet) 

  111 
3.6. TCP/IP PRIMER CHAPTER 3. BASIC OPERATION

Gateway – This is the address where all external address requests are sent. In most internal 
networks, this will be the address of the router so that the device can access the Internet 
when  needed. 
Netmask – Defines which addresses are inside the network and which are outside the network. 
255.255.255.0 (class C) is typical, meaning that only the last part of the address varies within 
the network. Everything else is outside. 
Services or Ports – A single device may have one to several thousand different services 
running on it. Services can include things like web services, e‐mail, and file transfers among 
many. Each service runs on a numbered port at an IP Address. These port numbers are 
somewhat standardized. 
Client  vs.  Server  –  In  communication  over TCP/IP  there  is  generally  a  Server  (the  one 
providing the service) and a Client (the one using the service). Devices can act as both. 
They may, for example be a Client for web services when opening a web browser and may 
at the same time be a Server providing services such as file storage. 
DHCP – A Service that automatically assigns IP Addresses. Typically, your ISP assigns an 
Internet address to your router and your router assigns an internal address to your device. 
DNS  –  A  Domain  Name  Server  is  a  service  on  the  Internet  that  translates  names,  like 
www.kmi.com into a numeric address.  Usually the DNS is provided by your ISP and the 
address of the DNS server is forwarded by DHCP. 
DDNS ‐ Dynamic DNS. This is a service that assigns your router a URL (like www.xxx.com) 
that “follows” your numeric IP address as it is dynamically changed. DDNS is usually 
handled completely by the router with the router “pinging” the DDNS servers to report the 
current address.  DDNS services range from free to commercial services and have varying 
amounts of latency (the time between when your numeric IP address changes and when the 
URL is updated to match). 
URL ‐ Uniform Resource Locator. This is a name such as www.kmi.com that is equivalent 
to a numeric IP address. 
 
3.6.2 Addresses 

A device’s address may be static or dynamic. A dynamic address is the most common, with 
another device (a DHCP server) assigning the addresses on request. 
Dynamic addresses are convenient because they are managed automatically and don’t require 
the involvement of System Administrators to assign them. 
The problem with dynamic addresses though is that they are just that – dynamic. This 
means that their address is uncertain and can change from one day or one minute to the 
next. 
The typical dynamic connection has a router that gets an IP Address from the ISP. The 
router translates the addresses within your internal network to the outside address from 
the ISP. When you surf the web or send e‐mail this translation is unseen. However, if you 

  112 
CHAPTER 3. BASIC OPERATION 3.6. TCP/IP PRIMER

wanted to access your computer from work or from a coffee shop it can be difficult because 
you cannot be sure of the IP Address assigned to your router (and thus your device) at the 
moment. 
In order to access a device remotely, you need a well known or static address. Most companies 
and organizations have static addresses because they must be routinely accessed from the 
Internet. Static addresses cost more. 
 
3.6.3 Some GuidelinesSome  Guidelines 

Typical  internal  addresses  are  10.x.x.x,  192.168.x.x,  or  172.16.x.x  thru  172.31.x.x.  These 
addresses should only be used within an internal network and should NEVER be exposed 
directly to the Internet. 
Using 0 or 255 in the final part of an IP address has special meaning and should not be used 
for any normal address. So, for example do not use: 

• 192.168.1.0 

• 192.168.1.255 

• 10.0.1.0 

• 10.0.1.255 

When using devices within an internal network, use dynamic addresses whenever possible. 
Assigning static addresses has the potential for conflict with other devices on the network 
and should only be done after consulting with the System Administrator. 
Assigning the same static IP address to multiple units can cause problems as your network 
may remember the address as belonging to a previous unit and get confused (even if they’re 
not connected to the network at the same time). Use unique addresses if possible. 
If you have configured a standalone Rock+ as DHCP server, you should disable this func‐ 
tionality before you connect the Rock+ to your internal LAN or you may cause yourself 
networking problems due to the conflict in address assignments. 
 
3.6.4 Typical  Configurations 
• Rock+ products are configured as DHCP clients out‐of‐the‐box. This means that they 
will try to get an IP Address from your network. To determine which address has been 
assigned, your best bet is to use the Rock+’s Linux serial console connection and the 
Linux command “ipaddr eth0”. 

• When the Rock+ will be deployed using a router (such as in a structure using a DSL 
connection), you may let the Rock+ obtain an address from the router or you may 
assign a static address within the internal network. The outside address of the router 

113 
3.6. TCP/IP PRIMER CHAPTER 3. BASIC OPERATION

will either be statically assigned or dynamically assigned by the ISP depending on the 
service purchased. The Rock+ is accessed externally using the address of the router. 
It may be necessary to “poke holes” in the firewall to allow external access to some 
services. 

• If the network is local only (no router used), use static addresses within the internal 
network. 

• If the Rock+ is standalone and you will occasionally connect a computer (a laptop?) 
to configure the unit or obtain data, then configure the Rock+ to have a static address 
(usually 192.168.9.1) and to act as a DHCP server. This way, you need only connect 
your laptop to communicate with the Rock+. The Rock+ will assign the laptop an 
address and you will not have to manually configure your laptop. 
 
The Rock+’s IP address is configured using the “netconfig” script from Linux. 
 
3.6.5 Rock+ Services 

By default, the following services are enabled: 

• PING 
• SSH 
• HTTP 
• NTP 
• POC 

By default, the following services are disabled: 

• FTP 
• TELNET 
• TFTP 
 
3.6.6 Further 

Distance limitations – Ethernet is limited to a standard distance of 100 meters. It is possible 
to push this limit a little bit, but there are no guarantees. To extend the limit, you need a 
powered Ethernet repeater, or a powered inline  switch. 
DSL connections are asymmetric – A typical DSL line has different speeds for upload than 
for download. These connections are designed for web surfing. They are typically 8X faster 

  114 
CHAPTER 3. BASIC OPERATION 3.6. TCP/IP PRIMER

on download than upload. So you must understand that although the connections are cheap, 
your 3Mb connection is actually only a 375Kb connection when you upload, such as streaming 
real‐time data or sending recorded data back to the data center. 
Internet security is a complex topic. The gist of it though is this: 

• SSH (which includes SFTP) allows secure logins and file transfers over the internet. 
• TELNET is totally “in the clear” meaning that anyone watching can see exactly what 
you type, including passwords. It should generally NOT be used over the Internet. 

• FTP or TFTP is also sent in the clear, but can be useful. For example, setting up an 
FTP Server at a data center can be a useful and easy way to upload the data from 
your equipment. Although not inherently secure, you can easily configure most FTP 
Servers so that data once uploaded, cannot be seen or accessed from the internet. This 
causes your data center to lose its appeal to hackers who might be tempted to use your 
system for data storage. 

• Enabling the Rock+ as an FTP Server over the Internet is also generally discouraged 
because this exposes system passwords every time that you log into FTP over the 
Internet. It can be useful though under some conditions such as: 

◦ When the unit is not connected to the Internet (such as within your lab) 
◦ For a temporary connection where you are willing to take the risk – e.g. connect 
to the unit once a year to upload new firmware and then immediately disable 
FTP  again. 
◦ Using a VPN or “tunnel” connection where all traffic to the unit is encrypted 
and protected by a wrapping layer provided by the router. This is an advanced 
subject and is not covered here. 
 
An alternative to using a more expensive static IP Address is to use a Dynamic DNS (DDNS) 
service. This can be free, but even if you choose a commercial DDNS server that is a fee‐for‐ 
service feature, it will usually be cheaper than obtaining a static IP Address if that is even 
possible. You can then connect to the units at their URL using the web interface, SFTP (for 
file transfers) or SSH for secure Linux login connections. 
Rock+ units that wish to send e‐mail can configure the e‐mail file or message senders to use 
the KMI mail relay. In this way, the mail is routed through a dedicated Kinemetrics mail 
server so that you don’t have to configure or maintain your own mail server – which can be 
fairly complex. The mail relay accepts mail only from recognized Rock+ units (so as not to 
forward SPAM) and requires only that you use DNS and specify the target e‐mail address. 
   

  115 
CHAPTER 3. BASIC OPERATION 3.6. TCP/IP PRIMER
 
Additional network‐centric capabilities are supported as well, including: 

◦ Dropbox support for file uploads 

◦ OpenVPN support 

• Includes “stealth” mode that blocks all but VPN and SSH ports 

◦ SNMP support 
For in‐depth reading on TCP/IP, see RFC 2151 ‐ A Primer On Internet and TCP/IP Tools 
and Utilities 
Also good, but less formal is “Daryl’s TCP/IP Primer”, which can be found at www.ipprimer.com 

  116 
3.7. FILE MANAGEMENT CHAPTER 3. BASIC OPERATION

3.7 File Management 
Basic file access is done through the web interface. To see recorded files, select Recorded 
Files from the menu. You’ll see: 
 

Figure 23: Recorded Files 

 
Each of the thumbnail pictures shown on the right represents one recorded file. These 
pictures are small files for fast download and display. They show only enough detail to help 
you distinguish which files are sensor tests, meaningful triggers, or useless noise data. 
To see a bit more detail, you can click on each picture to see an expanded – but still static 
– picture: 

  117 
CHAPTER 3. BASIC OPERATION 3.7. FILE MANAGEMENT

Figure 24: Recorded Files Detail 

 
If needed, you can use your browser’s zoom function (usually holding the CTRL key and 
scrolling the mouse wheel will zoom in and out). 
Individual files can be downloaded by clicking the file name (like “aza001.evt”) and selecting 
“save”. 
Files can be deleted by selecting it (or them) under the corresponding filename(s) and clicking 
“Delete  Selected”. 
To manage large numbers of files (downloading and/or deleting them), you should use a 
program such as WinSCP, which is described later in this manual under Tools. 

  118 
3.8. FILE VIEWING CHAPTER 3. BASIC OPERATION

3.8 File Viewing 
The Recorded Files screen described above allows you to overview your recorded files using 
static pictures. However, often you need to look a little closer. For this purpose, the digitizer 
includes a File Viewer  Application. 
The File Viewer is an executable Java JAR file that can be downloaded to your computer 
by clicking the “File Viewer” link on the Recorded Files page. Save this file (usually File‐ 
Viewer.jar) someplace convenient (like your Desktop, either directly or as a link). You only 
need to download the File Viewer once unless you need a fresh copy. 
 

Figure 25: Downloading the File Viewer 

 
To view a file, download it and then launch the File Viewer. From the File Viewer menu, 
select Open and locate and select your file. 

  119 
CHAPTER 3. BASIC OPERATION 3.9. WAVEFORM VIEWING

Figure 26: File Viewer 

 
Though not covered in detail here, note that you can zoom, pan, change display units, apply 
grid lines, sync displays in the X and Y axes, and do many other things. 

  120 
3.9. WAVEFORM VIEWING CHAPTER 3. BASIC OPERATION

3.9 Waveform  Viewing 


In addition to being able to view recorded files after they’ve been recorded, you can also 
view live waveforms as the data is digitized. To do this, select Waveform Viewer: 

Figure 27: Waveform Viewer 

  121 
CHAPTER 3. BASIC OPERATION 3.9. WAVEFORM VIEWING

Once the list is displayed, select one or more virtual channels from the list on the right 
(including SOH channels) and then press OK: 
 

Figure 28: Waveform Display 

 
Note that you can select the display units. 

  122 
3.9. WAVEFORM VIEWING CHAPTER 3. BASIC OPERATION
 
Notice the selections for live Spectra (FFT and PSD), which will produce something like 
this: 
 

Figure 29:  Real Time PSD 

  123 
  124 
 
 
 
 
 

Chapter  4 

Web Interface  Overview 
 
The web interface to the digitizer allows you to configure and operate your digitizer using 
a web browser in many cases without installing any Kinemetrics‐specific software. The web 
browser should be HTML 1.1 compliant (or later). 
To access the digitizer through the web interface: 
Open a web browser on your PC. In the address bar of the browser, type the IP address of 
the digitizer – for example 10.0.3.71. You’ll be presented with a login prompt: 
 

Figure 30: Login Prompt 

  125 
CHAPTER 4. WEB INTERFACE OVERVIEW

Web logins use a relatively secure Digest Authentication login. Log into the unit using the 
username and password used by your unit. By default, the username is “rock” and the 
password is “kmi”. 
Two groups of accounts may be set up: 

• Client level users can view things, but cannot make any changes. 
• Admin level users have administrative access, meaning that configuration changes can 
be made. 
 
Multiple web users from multiple IP addresses can log in at one time. But if more than one 
Admin level user tries to log in at one time, the first user gets Full Access (read/write), all 
others get Limited Access (read only). 
Admin users who are given Limited Access can force the other Admin user to log off so 
that they can obtain Full Access. Client users are not affected since they only have Limited 
Access. 
After 60 minutes of inactivity, any connected web user will be logged off. So if you forget to 
log out of the Web Interface, the system will log you out after this time automatically. 
Type a user name and password in and press OK. This will display the Overview screen: 

  126 
CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 31: Overview Screen 

 
The menu items listed across the top of the display (left to right) are: 

• Status 

◦ Overview: The main status screen 
◦ Waveform Viewer: Display of real‐time waveforms 
◦ State of Health: Access to State‐of‐Health displays 
◦ Logout: Log out of the web interface 

• Layout 

◦ Hardware: Edit of hardware specific  parameters 
◦ Configuration:  Edit of application specific  parameters 

  127 
CHAPTER 4. WEB INTERFACE OVERVIEW

◦ Channel Summary:  Displays a table of channel  configurations 
◦ Site Summary: Displays summary information for thr site (including network 
setup) 
◦ Layout Display: Graphical display of the module layout 
◦ Layout Wizard: Allows you to rebuild the module layout   (downloaded) 
◦ Download Config: Download the current configuration to your  computer 
◦ Apply Changes: Validate and restart with any changes made 

• Files 

◦ Runtime Log: A display of the run time log file 
◦ Error Log: A display of the error log file 
◦ Maintenance Log:  Make maintenance history changes 
◦ Recorded Files: A display of recorded file thumbnails 
◦ File Viewer: Display files interactively  (downloaded) 

• Triggering 

◦ Triggering and Sensors:  Commanded triggers, sensor tests, and sensor  control 

• Tools 

◦ Help Tips: A list of built‐in help topics 
◦ Rock Recovery: Recovery modes back to a saved configuration or factory default 
◦ Operation/Password:  Edit operation parameters including passwords 
◦ Module Add: Add modules to the layout 
◦ Module Remove: Remove modules from the layout 
◦ Module Replace: Replace modules with a similar type 
◦ Advanced Toggles: Display for edit advanced or test modules and parameters 
◦ Debug Levels: Edit module debug levels (for  diagnostics) 
◦ Edit Users: Edd/edit/delete web user names and passwords 
◦ Admin details: Upload and download firmware, configurations, and licenses 

Hardware accesses hardware specific values such as specific IP addresses, number of channels, 
sensor types, voltage ranges, serial numbers, etc. 

  128 
CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 32: Hardware Configuration 

 
Configuration accesses installation independent values such as pre‐event time, post event 
time, sample rates, and other operational parameters. 

  129 
CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 33:  Configuration 

 
The top of the Configuration page is a Parameter Map, which is a set of links to the individual 
parameter sets of each module. This can be faster than scrolling to find a parameter. For 
example, to access the Rock2 Data Interface, click on that link and you will get: 

  130 
CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 34: Module Parameter Example 

 
Layout Display displays a picture of the current layout, along with a brief explanation of 
each module. 

  131 
CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 35: Layout Display 

 
Apply Changes tells the Rock+ to make changes you’ve made permanent. 
Channel Summary displays channel information: 

  132 
CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 36: Channel Summary 

 
Site Summary shows: 

  133 
CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 37: Site Summary 

  134 
CHAPTER 4. WEB INTERFACE OVERVIEW
State of Health displays SOH information that is logged short term (24 hrs) and long 
term (one year): 

Figure 38: SOH Selection 

  135 
CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 39: SOH Display 

  136 
CHAPTER 4. WEB INTERFACE OVERVIEW
 
Waveform Viewer defaults to non‐applet mode (using HTML and Javascript only). For the 
majority of users this will be the preferred mode, and also allows use on mobile devices. If 
for some reason you need to use the older applet mode, you will need to change the non‐
applet mode setting.  
 
Note that Java applets are increasingly unsupported so if you are still using them you should 
be prepared for their demise. 
 

Figure 40: Applet Mode Selection 

  137 
CHAPTER 4. WEB INTERFACE OVERVIEW

Click on Waveform Viewer to start the Viewer, you’ll see this: 
 

Figure 41: Waveform Viewer Selection 

  138 
CHAPTER 4. WEB INTERFACE OVERVIEW
 
Select the channel(s) you want to see and press OK: 
 

 
Figure 42: Waveform Display 

  139 
CHAPTER 4. WEB INTERFACE OVERVIEW
 
Note the options for FFT and PSD display: 
 

 
Figure 43: PSD Display 

 
 

  140 
CHAPTER 4. WEB INTERFACE OVERVIEW

 Next is Recorded Files. This shows small (1‐2kb) thumbnails of recorded files: 

Figure 44: Recorded Files 

  141 
CHAPTER 4. WEB INTERFACE OVERVIEW

If you click on a thumbnail, you can see a larger (6‐10kb) thumbnail: 

Figure 45: Recorded File Detail 

 
The links and buttons below the small thumbnails allow you to download or delete the files 
(one at a time) via HTTP. Larger scale file maintenance is best done via SFTP or WinSCP. 

  142 
CHAPTER 4. WEB INTERFACE OVERVIEW

Triggering & Sensors allows you to perform console or sensor test triggers: 

Figure 46: Triggering and Sensors 

  143 
4.1. EDITING PARAMETERS CHAPTER 4. WEB INTERFACE OVERVIEW

Back on the Overview, status information is displayed at the bottom: 

Figure 47: Overview 

 
4.1 Editing  Parameters 
Parameters are divided into two main sections. Hardware parameters are those that are 
typically hardware related such as sensor and channel specifics. Configuration parameters 
are those that are mainly related to how the system operates – things like trigger levels, 
pre‐event and post‐event times, and so on. 
To edit hardware parameters, select Hardware from the Layout menu and you’ll see some‐ 
thing like this: 

  144 
CHAPTER 4. WEB INTERFACE OVERVIEW 4.1. EDITING PARAMETERS

Figure 48:  Hardware Parameters 

 
Note that when you click the “?” next to the name of a parameter that you’ll be shown the 
help string that provides more detail about the purpose of the parameter: 
 

Figure 49: Help String 

  145 
4.1. EDITING PARAMETERS CHAPTER 4. WEB INTERFACE OVERVIEW

When you’ve finished editing the Hardware Parameters, click OK to save your changes. 
To edit Configuration Parameters, select Configuration from the Layout menu and you’ll 
see: 
 

Figure 50:  Configuration Parameters 

 
The top of the Configuration page is a Parameter Map, which is a set of links to the individual 
parameter sets of each module. This can be faster than scrolling to find a parameter. 
For example, to access the Rock2 Data Interface, click on the link and you will get: 

  146 
CHAPTER 4. WEB INTERFACE OVERVIEW 4.2. ADDING MODULES

Figure 51: Module Parameters 

 
After completing your changes, be sure to press OK to save your changes. 
Once you have completed your parameter changes, activate the completed parameter changes 
by selecting “Apply Changes” from the Layout menu of the display. The system will restart 
and reconfigure itself to make your changes. This may take a few minutes until the system 
is restarted and operational with your changes. 
 
4.2 Adding Modules 
You can add capabilities to your system by adding additional modules. The process is the 
same for adding modules of any type. For example, to add a module to the system to push 
recorded event files automatically to a remote FTP server, you will need to add an    FTP 

  147 
4.2. ADDING MODULES CHAPTER 4. WEB INTERFACE OVERVIEW

Sender module. To accomplish this, select Module Add from the Tools menu, and scroll 
down to locate the FTP Sender: 
 

Figure 52: Module Add 

 
Click “Add” to add the module. 
Next, click Configuration from the Layout menu and select “FTP Sender” from the parameter 
map. The FTP Sender parameters will now appear at the top of the window as shown: 

  148 
CHAPTER 4. WEB INTERFACE OVERVIEW 4.2. ADDING MODULES

Figure 53: FTP Sender Parameters 

 
Configure the parameters for your new module as needed, then click OK to save your changes. 
Next, click “Apply Changes” from the Layout menu. The system will restart and reconfigure 
itself  to  add  the  FTP  Sender  module.  This  may take a  few  minutes  until  the  system  is 
restarted and operational with your new  module. 
Please note than for most modules (including the FTP Sender module used as an example 
here) that you can usually add multiple modules of the same type. This could be useful 
where for example you want to send recorded files to more than one remote FTP server. To 
do this, you would add one FTP Sender module for each destination. 

  149 
4.3. REMOVING MODULES CHAPTER 4. WEB INTERFACE OVERVIEW

4.3 Removing  Modules 


You can further alter the capabilities of your system by removing modules. The process is 
the same for removing modules of any type. For example, to remove an FTP Sender module 
from the system, select Module Remove from the Tools menu and scroll down to locate the 
FTP  Sender: 
 

Figure 54: Module Remove 

 
Click “Remove”. 
Next, click “Apply Changes” in the Layout menu. The system will restart and reconfigure 
itself to remove the FTP Sender module. This may take a few minutes until the system is 
restarted and operational with your  changes. 

  150 
CHAPTER 4. WEB INTERFACE OVERVIEW 4.4. REPLACING MODULES

4.4 Replacing  Modules 


Sometimes what you need to do is replace a module in the system with another that has 
the same basic functionality. For example, you may want to replace a data archiver that 
produces EVT files with one that produces MiniSEED files, or you may want to replace a 
Threshold Trigger with an STA/LTA Trigger. For example, to change the data format from 
Kinemetrics EVT file format to MiniSEED, use the web interface. Choose Module Replace 
from the Tools menu and you’ll see: 
 

Figure 55: Module Replace 

 
Click “Replace” for the Altus EVT Format Data Archiver and you’ll see your choices: 

  151 
4.4. REPLACING MODULES CHAPTER 4. WEB INTERFACE OVERVIEW

Figure 56:  Module Replacement Options 

 
If you select “Choose” next to MiniSEED Format Data Archiver, then you’ll replace the 
Altus EVT Format Data Archiver with MiniSEED. The Module Replacement screen will 
now look like this: 

  152 
CHAPTER 4. WEB INTERFACE OVERVIEW 4.4. REPLACING MODULES

Figure 57:  Module Replacement Complete 

 
In this example, it was pretty simple because there was only one EVT Archiver module in 
the system. But if you needed to replace the Threshold Triggers with STA/LTA Triggers and 
you have a 36 channel system you can imagine that the process could become a bit tedious. 
To make this a little better, you can select “All” instead of “Choose” when you are shown 
the Compatible Modules list. You can then replace all modules of one type with another. 
Once you have completed your parameter changes, activate the completed parameter changes 
by selecting “Apply Changes” from the Layout menu. 

  153 
4.4. REPLACING MODULES CHAPTER 4. WEB INTERFACE OVERVIEW

  154 
Chapter 5 

File Viewer 

The File Viewer is a downloaded Java application that is loaded into your PC from the 
digitizer’s  web page  when you click the  File  Viewer link. It  will  be  downloaded as  an 
executable .jar file that you can run by directly if you have Java installed on your computer. 
A brief summary of capabilities follows. Upon launch, the initial File Viewer display looks 
like this: 

  155 
CHAPTER 5. FILE VIEWER

Figure 58: File Viewer Main 

 
The File Viewer consists of a graphic display area (below), a menu, and a set of toolbar 
buttons. Left to right, the toolbar buttons are as follows: 
Open – Opens a file for display.  Once the program has been started, it can be  used to 
open other files. As of this writing, the File Viewer supports Kinemetrics EVT files and 
MiniSEED files (with some restrictions) when used with a “.m” extension. In the case of 
MiniSEED, multiple files for the same time period may be opened at once. 
Save As – Allows saving of the current file in another directory or under another name. 
Print – Prints the current time series. 
Setup – Allows configuration of additional File Viewer features. The Setup dialog looks like: 

  156 
CHAPTER 5. FILE VIEWER

Figure 59: File Viewer Setup 

 
Setup selections include: 
 
• Columns – Select the number of columns of graphic displays. You could for example, 
organize an 18 channel record into 1 column of 18 plots, two columns of 9 plots, or 
three columns of 6 plots. 

• Mean – Allows mean removal from the data, or “As Read” display. 
• Autoscale – Allows autoscale as “Global” (all channels scaled to largest signal), “Indi‐ 
vidual” (channels scaled individually), or “None” 

Header Display – Displays header information as is available from the file: 
 

Figure 60: Header Display  

Grid Mode – Displays grid lines on the plots if requested. 

  157 
CHAPTER 5. FILE VIEWER

Negative Display – Displays plots as green on black or blue on white. 
DAC Mode – Displays data points as point‐to‐point, or showing individual signal levels for 
each sample. This is usually only discernable when zoomed in closely. 
Multi‐channel Mode – Displays all channels or only one channel at a time. In single channel 
mode, channels can be selected with the Page Up / Page Down keys. 
Units – Allows selection of display units. Choices include counts, volts, g, cm/s2, or gal. 
Note that correct bit weight and sensitivity settings are required for proper scaling of the 
data. 
Cursor – Enables cursor mode so that you can click to display exact time and amplitude of 
individual sample values. Note that the cursor can be moved with another mouse click or 
by using the left and right arrow keys: 
 

Figure 61: Cursor Display 

  158 
CHAPTER 5. FILE VIEWER 5.1. FILE VIEWER DYNAMIC OPERATION

Re‐Sync X‐Axes – Allows you to synchronize the display of all channels in the X‐axis to 
show the same time period. You could, for example zoom in on an area of one channel, and 
then re‐sync the other channels to show the same time period. 
Re‐Sync Y Axes – Allows you to synchronize the display of all channels in the Y‐axis to 
show the same amplitude range. You could, for example zoom in on an area of one channel, 
and then re‐sync the other channels to show the same amplitude range. 
Zoom Out – Zooms out to the original full scale range or autoscale setting. 
About – Shows the current version of the program. 
 
5.1   File Viewer Dynamic  Operation 
The File Viewer contains some dynamic display manipulation features as  follows: 
Zooming: When NOT in Cursor mode, you can zoom on any channel by using the mouse to 
draw a “box” around the area of interest: 

  159 
5.1. FILE VIEWER DYNAMIC OPERATION CHAPTER 5. FILE VIEWER

Figure 62: Zooming in 

 Once zoomed, the display will look like this: 

  160 
CHAPTER 5. FILE VIEWER 5.1. FILE VIEWER DYNAMIC OPERATION

Figure 63: Zoomed in 

 
Note that the zoomed data will show a different amplitude and/or time period. You can 
zoom in multiple times, and un‐zoom one level at a time using a right click of the mouse. 
Zooming will remember five levels of zoom for each plot. 
Panning of the selected plot is also supported regardless of zoom level. The arrow keys allow 
panning left, right, up, or down. 

  161 
5.1. FILE VIEWER DYNAMIC OPERATION CHAPTER 5. FILE VIEWER

  162 
Chapter  6 

Rockhound Command Console 
 
The Rockhound Command Console can be opened in any of a number of ways: 
 
• If you are using RockTalk, then you can open the Command Console using the “Ter‐ 
minal Window” feature of RockTalk. 

• You can also open the Command Console using a telnet or PuTTY connection at port 
9900 

• You can also connect to the system using SSH and then use telnet within the SSH 
session to open the Command Console on a telnet connection on localhost (127.0.0.1) 
at port 9900. 

Once open, you will need to know the console password in order to log in. 
Command Console parameters are available under System Operation parameters as shown 
here: 

  163 
CHAPTER 6. ROCKHOUND COMMAND CONSOLE

Figure 64: Console Parameters 

 The Command Console once open will look like this: 

  164 
CHAPTER 6. ROCKHOUND COMMAND CONSOLE

Figure 65: Console Open 

 
The “help” command can be used as shown to list standard, advanced, or all commands. 
The list of available commands is as follows: 

ALARM ‐ Set the state of the Alarm bit. 
Usage: ALARM on|off 
 
BASEDIR ‐ Report the base installation directory. 
Usage: BASEDIR 
 
BATTERYTEST ‐ Command a power supply battery test. 
Usage: BATTERYTEST  secs 

BROADCAST ‐ Send a message to all modules. 
Usage: BROADCAST {command} [optional argv1] [optional argv2]  [...] 

CALCHECK ‐ Checks front end calibration in the 20V range. 
(intended  for  factory  use only) 
Usage:  CALCHECK  zerorange fsrange 

CALIBRATE ‐ Performs and reports calibration. 

  165 
CHAPTER 6. ROCKHOUND COMMAND CONSOLE

(intended  for  factory  use only) 
Usage:  CALIBRATE fec/dac/avx/view/grade 

CLREVT ‐ Clear event counter. 
Usage: CLREVT 
 
CLS ‐ Clears the screen. 
Usage: CLS 
 
CRFLUSH ‐ Request an early flush of continuously recorded data files. 
Usage: CRFLUSH 
 
CTIME ‐ Display the current time. 
Usage: CTIME 
 
DACCHECK ‐ Checks DAC calibration in the 20V range. 
(intended  for  factory  use  only) 
Usage:  DACCHECK  zerorange fsrange 

DATAFLOW ‐ Display current Data Flow stats. 
Usage:  DATAFLOW 
 
DEFDECK ‐ Defaults channels 1 to 3 by reading parameters 
from  an  internal  Episensor  deck. 
Usage:  DEFDECK 

DIR ‐ Display a complete directory listing. 
Usage: DIR {path} 
 
DSPCMD ‐ Send a command to the Rock DSP. 
Usage: DSPCMD (addr) (cmd) 
 
ECHO ‐ Turns character echo on/off. 
Usage: ECHO {on|off} 
 
EEPROM  ‐  Display  status  of  board  ID  EEPROMs. 
Usage:  EEPROM 
 
EXIT ‐ Exits the console session. 
Usage: EXIT 
 
EXTCHARGERTEST ‐ Command a power supply external charger test. 
Usage:  EXTCHARGERTEST   mins 

GPS ‐ GPS and Timing control    functions. 

  166 
CHAPTER 6. ROCKHOUND COMMAND CONSOLE

Usage: GPS STAT: Timing  system  status 
GPS ID: Timing system type and version 
GPS  RUN:  Enable  timing  system 
GPS STOP: Disable timing system 
GPS ON: Turn on GPS power 
GPS  OFF:  Turn  off  GPS  power 
GPS DIAG on|off|val: Timing system diagnostics 
vals:  1=Time  reports  2=Time    verifies 
vals:  4=Time  quality  8=Time syncs 
vals: 10=Position reports 20=State of health 
vals:  40=Satellites  &   SNRs 
GPS DAC on|off: Timing system TCXO  reports 

Help Options: 
Help std ‐ List standard commands 
Help adv ‐ List advanced commands 
Help all ‐ List all  commands 
Help * ‐ Help for every  command 
Help {command} ‐ Help for a specific command 

HOSTSERVICES ‐ Report current services and associated port numbers. 

HWID ‐ Display the hardware ID string. 
Usage: HWID 

KBTRI ‐ Keyboard trigger. 
Usage: KBTRI (secs) 
 
LOOPBACKATE ‐ Performs factory ATE loopback using special test hardware. 
(intended  for  factory  use  only) 
Usage:  LOOPBACKATE  typ arglist 

LS ‐ Display a brief directory listing. 
Usage: LS {path} 
 
MINFO ‐ Display info about currently loaded modules. 
Usage: MINFO 
 
MKBATCH ‐ Creates a batch file by typing in lines of commands. 
Usage: MKBATCH file 
 
MSGLEVEL ‐ Set the level of messages sent to this session. 
Usage: MSGLEVEL {level, 0=None, 1=Critical, 2=All} 

MSTIME  ‐  Display  the  system  time  in  milliseconds  since  Epoch. 

  167 
CHAPTER 6. ROCKHOUND COMMAND CONSOLE

Usage: MSTIME 

NEWSPSATE ‐ Changes sample rate for ATE testing by rebooting. 
(intended  for  factory  use only) 
Usage:  NEWSPSATE sps 

NOISETEST ‐ Performs and reports a main channel noise test. 
(intended  for  factory  use only) 
Usage:  NOISETEST 

NWDEFER ‐ Defers network triggering temporarily. 
Usage: NWDEFER secs 
 
OWSBCHECK ‐ Check the One‐Wire Sensor Bus by testing for 
device discovery on each active sensor bus  channel. 
(intended  for  factory  use  only) 
Usage:  OWSBCHECK 

PAUSE ‐ Pause and wait for user input. 
Usage:  PAUSE 
 
PING ‐ Request a PONG reply. 
Usage:  PING  [optional  reply text] 
 
PLLCHECK ‐ Checks clock drift and VCO range. 
(intended  for  factory  use  only) 
Usage: PLLCHECK lcerange vcorange 

RECONFIG ‐ Command re‐read of the configuration files. 

RSVERIFY ‐ Verify that unit has not restarted under 
stress   test  conditions. 
Arg=0 saves current restart for later comparison. 
Arg!=0 checks restart count  against  saved. 
(intended  for  factory  use only) 
Usage: RSVERIFY rsflag 

RTPARAMS ‐ Display current system runtime parameters. 
Usage: RTPARAMS [optional key‐match] 
 
RUN ‐ Run a batch file by executing commands in it as if 
they  were   typed. 
Usage: RUN file 

SETDEBUGMASK ‐ Set the debug mask for a given module. 

  168 
CHAPTER 6. ROCKHOUND COMMAND CONSOLE

Usage: SETDEBUGMASK {module} [value  (default=0)] 

SLEEP ‐ Sleep the specified number of seconds. 
Usage: SLEEP  seconds 
 
SNSCAL ‐ Do sensor calibration. 
Usage: SNSCAL (type) 
 
SOHCHECK ‐ Checks BF analog voltage values. 
(intended  for  factory  use only) 
Usage: SOHCHECK value range (10  each) 

SOHFLUSH ‐ Request a flush of SOH data files when using 
the optional web  server   interface. 
Usage:  SOHFLUSH 

STAID ‐ Set station ID and restart. 
Usage: STAID auto or STAID name 
 
STEP ‐ Apply the sensor step voltage. 
Usage: STEP 
 
STIME ‐ Display the system start time. 
Usage: STIME 
 
TEMPCHECK ‐ Checks temperature against actual. 
Range against expected must allow for case rise. 
(intended  for  factory  use only) 
Usage: TEMPCHECK expected range 

UNSTEP ‐ Remove the sensor step voltage. 
Usage: UNSTEP 
 
USERSTAT ‐ Write numeric parameter to user status stream. 
Usage: USERSTAT num 
 
VALIDATE ‐ Validates firmware or config update. 
If  valid  may  cause restart. 
Usage: To validate and restart: VALIDATE UPDATE 
To validate and dynamically reconfigure: VALIDATE  RECONFIG 

VCOCHECK ‐ Check the VCO response by offsetting and 
monitoring drift. (intended for factory use only) 
Usage: VCOCHECK secs vcodelta mindrift maxdrift 

  169 
CHAPTER 6. ROCKHOUND COMMAND CONSOLE

VCOSTABLE ‐ Sets VCO initial value for stable temperature 
and GPS  locked. (intended for factory use    only) 
Usage:  VCOSTABLE 

VCOVERIFY ‐ Verify that VCO has not changed. 
Arg=0 saves current VCO for later comparison. 
Arg!=0  checks  VCO  against  saved. 
(intended for factory use only) 
Usage: VCOVERIFY vcorange 
 
VER ‐ Display the software and hardware version numbers. 
Usage: VER 
 
VMBYTES ‐ Display the total bytes available, and bytes free in the VM. 
Usage: VMBYTES 
 
VMERROR ‐ Exits the VM with error 99 (for testing). 
Usage: VMERROR 
 
VMEXIT ‐ Exits the VM. 
Usage: VMEXIT 
 
VMGC ‐ Force Garbage Collection in the VM. 
Usage: VMGC 
 
VMCRASH ‐ Deliberately place a VM thread into an infinte loop 
(for     testing). 
Usage:  VMLOCKUP 

VMOOM ‐ Deliberately crash the VM by consuming all heap space 
(for testing). 
Usage: VMOOM 

VMPROPS ‐ Display the Virtual Machine properties. 
Usage:  VMPROPS 
 
VTDSET ‐ Set voltage range during ATE testing. 
0=2.5V, 1=5V, 2=20V (intended for factory use only) 
Usage:  VTDSET range 
 
WATCHDOGTEST ‐ Cause a system restart via a 
software watchdog. 
Usage: WATCHDOGTEST 

  170 
CHAPTER 6. ROCKHOUND COMMAND CONSOLE 6.1. BATCH MODE

6.1   Batch  Mode 
The Rockhound Command Console supports a limited batch mode process to make it some‐ 
what easier to perform routine activities such as periodic maintenance. The MKBATCH 
command is used to create a batch file by name, accepting input until RETURN only is 
entered. Lines of the file are executed as commands as if typed by the user. Lines starting 
with “;” or “#” are treated as comments and ignored. 
The batch file will be created with the extension “.BAT” on the digitizer. 
For example: 

> mkbatch mybatch 
Type in lines of commands, one command per line. 
When  done,  press  RETURN  only  to close the file. 
> ; Quarterly Maintenance 
> ; Get GPS  status 
>  gps  stat 
> ; Do  a 2 minute battery    test 
>  batterytest  120 
>  ; Wait  a bit 
>  sleep 150 
> ; Do a sensor calibration 
> snscal 
> Batch file creation  complete. 

To run the batch file, type: 

> run mybatch 

  171 
6.1. BATCH MODE CHAPTER 6. ROCKHOUND COMMAND CONSOLE

  172 
Chapter  7 

Non‐Networked  Use 
 
If you don’t intend to normally connect your digitizer to a network, we still recommend that 
you set up the digitizer’s network so that you can control the digitizer from your local PC 
when you’re on site. 
To communicate directly between the digitizer and a PC over Ethernet without using any 
supporting network hardware, you’ll connect the digitizer to the PC using an Ethernet 
crossover cable.  Once connected, you should be able to operate the digitizer in its typical 
networked modes. 
In this case, we recommend the following parameters: 
Digitizer: 

• IP Address: 192.168.1.1 
• Netmask: 255.255.255.0 
• Gateway: 192.168.1.100 
• DNS: None 
PC: 

• IP Address: 192.168.1.100 
• Netmask: 255.255.255.0 
• Gateway: 192.168.1.100 
• DNS: None 
You may also consider configuring the digitizer to act as a DHCP server, meaning that 
the digitizer can supply the IP address to your PC when you connect it without having to 
reconfigure the PC. Using the digitizer as a DHCP server can only be done if the digitizer is 
not on a network with another DHCP server, such as in a standalone configuration. To use 
the digitizer as a DHCP server: 
Digitizer: 

  173 
CHAPTER 7. NON-NETWORKED USE

• IP address: 192.168.9.1 

• Netmask: 255.255.255.0 

• Gateway: 192.168.9.1 
• DNS: None 

Then enable the DHCP server: 

initdconfig isc‐dhcp‐server on 
reboot 
 
PC: 

• IP address:  Automatically assigned 
 

If you plan to normally only operate the digitizer through a modem, Kinemetrics recommends 
that you add the Altus Emulation module. This module provides many capabilities, but the 
ones of interest here are: 

• Dial on event 
• File Transfer 
• Parameter Upload/Download 
• Firmware Update 

If you will have multiple digitizers on an internal network at the site and you will NOT use 
a DHCP server (you need static addresses), then we suggest using an addressing sequence 
such as 192.168.1.1 for the first, 192.168.1.2 for the second, and so on. 
For details of how to perform these functions, see the section Modem Only Operation  later 
on in this manual. 

  174 
 
 
 
 

Chapter  8 

Altus Emulation 
 
The Kinemetrics Altus instruments (K2s, Etnas, and Mt. Whitneys) are very popular with 
several thousand units delivered. As such, supporting networks from file based to real‐time 
and near real time systems exist for these instruments. In some cases, these networks have 
evolved over the years into very sophisticated systems with a great deal of capability. 
We understand that extending or possibly replacing these networks to support later gener‐ 
ation equipment can represent significant effort and/or cost. To ease the burden, we have 
included an “Altus Emulation” module in the Rockhound software providing the option to 
make the Rock+ based digitizer “look” very much like an Altus instrument so that you can 
continue to utilize your existing network with little or no modifications. 
Obviously, the emulation is not a 100% duplication of all Altus capabilities in exactly the 
same way as they are implemented on the Altus.  However, the Altus Emulation module when 
combined with other modules such as the Altus EVT Archiver module provides capabilities 
such as: 

• Recording in EVT file format 
• Altus File Naming  conventions 
• Support for Altus‐style modem  operations 
• Support for the Altus Serial Data Streams (SDS) protocol 
• Altus‐style terminal mode  functions 
• Support for Altus Block mode 

These capabilities allow for operation with the Kinemetrics NMS software, as well as with 
software such as Earthworm, Antelope, or customer developed software with little or no 
modification. In addition, the Altus Emulation modules allow for: 

• Support for more than 18 channels 

  175 
8.1. TERMINAL MODE CHAPTER 8. ALTUS EMULATION

• Co‐operation with other Rock+ platform capabilities 
• More compatible existence between modem modes and streaming  data 
• Multiple Altus Emulation modules can be included in the layout 
• Multiple streaming data interfaces serving data to multiple destinations 

There are two different types of Altus Emulation modules available with very similar features: 

• Altus Telemetry via RS‐232 allows Altus Emulation through any Rock+ serial port or 
modem 

• Altus Telemetry via TCP/IP allows Altus Emulation through IP links 
 
8.1 Terminal mode 
Terminal mode in Altus Emulation is similar to how it operates in the Altus. A connection 
is made to the unit via the RS‐232 or TCP/IP. Once connected, you can use commands that 
are very similar to those in the Altus: 
 

Figure 66: Terminal Mode 

  176 
CHAPTER 8. ALTUS EMULATION 8.1. TERMINAL MODE

Most of these commands provide approximations to the original Altus command and are 
intended to be as complete a reproduction as is reasonable.  It is assumed that most users 
who have occasion to use these commands will be familiar with the Altus commands on 
which these emulated commands are based. Additional detail on each command is supplied 
here where a command is substantially different than the Altus command else the user is 
referred to the appropriate Altus documentation, Altus Monitor Mode Communications, p/n 
302219. 
At present, the list of commands is as follows: 

* help 
ALARM ‐ ALARM control functions 
ALARM off ‐ Turn off the alarm bit 

Alarm bits on Rock+ digitizers are provided via the Rock GPIO module described in the IO 
Bits section. The Rock GPIO module equates the alarm function with one of the available 
hardware IO bits, which will be triggered when the threshold is exceeded. This command 
turns off that output as needed. 

ANSWERMODE ‐ Forces hangup of the modem 

Also may be used to declare successful termination of a call. The CGS Altus Emulation mod‐ 
ules specifically requires both a SUCCESS command to declare the conversation complete 
and an ANS command to hangup the line. 

AQ ‐ AQ acquisition control  functions 
AQ DET Detriggers recording 
AQ  DVM  Displays sensor signal  levels 

In addition to supporting the standard Altus display of channel voltages and supporting 
step/unstep mode, this command also allows display of the enabled state‐of‐health channels. 

AQ FT Performs sensor functional test 
AQ  ON/OFF  Enables or disables  acquisition 
AQ PEAK Displays signal peaks during trigger 
AQ SRT Performs sensor response test 
AQ TRI Triggers recording 
AUTOBAUD ‐ ignored 
BATTERY ‐ BATTERY performs a battery test 
BAT mins ‐ Performs a battery test for the time in minutes 

This command tests functionality of the external battery by internally disconnecting the 
input DC power, forcing the system to run off of the battery for the specified amount of 
time. If the battery does not have sufficient reserve, this test may cause the system to reset. 
During the test, the system will display the remaining time and current battery voltage as 
follows: 

  177 
8.1. TERMINAL MODE CHAPTER 8. ALTUS EMULATION

* bat  2 
This will perform a power supply battery test by disconnecting 
input DC power, and  may  cause  the  system  to  reboot  if the 
battery  does  not  have  enough reserve. 
Are you sure you want to do this? [(Y)es, (N)o]: y 
Requesting  battery  test  (120 seconds) 
Starting battery test... 
SYSTEM RESET indicates test failure 
Press any key to abort 
Seconds remaining: 
32: 12.21V 
 
BLOCK ‐ Forces the connection into block mode 
CD ‐ Change Directory command 
 
Allows the user to change subdirectories within the data directories. The “root” directory 
of Altus Emulation is placed at the specified data directory, which is usually /data/events. 
For security considerations, the user is not allowed to navigate outside of the data directory 
and subdirectories. 
The exception here is access to firmware files such as log, batch, configuration and script files 
that are located in various locations throughout the system. For these, a pseudo directory 
is created that cannot be directly “CD’d” to, but instead you may access with commands 
like “dir FW”. 
 
CALLMODE ‐ Forces hangup of the modem 
CHDIR ‐ Change Directory command 
CLEAR ‐ CLEAR events and alarms 
CLEAR ALL ‐ Clear event counter and alarm 
CLEAR EVENT ‐ Clear event counter only 
CLEAR ALARM ‐ Clear alarm only 
CPPC  ‐  CPPC  command  controls  modem  power  with  2  hour  timeout 
CPPC on ‐ Turn on the modem power 
CPPC off ‐ Turn off the modem power 
DEFAULT ‐ DEF performs system defaulting operations 
DEF DECK ‐ Defaults sensor parameters from internal deck 
DELETE  ‐ Deletes files singly or in   groups 
DEL  file FW  ‐ Deletes rt.log, err.log, and  batch   files 

See notes regarding special firmware files in the description of the CD command. 

DIR ‐ Displays storage directory 
DIR FW  ‐ Displays special  files 
 
See notes regarding special firmware files in the description of the CD command. 

  176 
CHAPTER 8. ALTUS EMULATION 8.1. TERMINAL MODE

DISPLAY ‐ DIS displays system parameters 
DIS BIT ‐ Displays bit map parameters  
DIS CHA ‐ Displays channel parameters 
DIS MOD ‐ Displays modem parameters 
DIS  SEN  ‐  Displays  sensor  parameters 
DIS SER ‐ Displays serial data stream parameters 
DIS STR ‐ Displays stream parameters 
DIS RWM ‐ Displays read/write/miscellaneous 
DIS  WRI  file ‐ Writes  parameters  to  a  file 
 
Writes digitizer configuration parameters (extracted from the active config.jar file that holds 
all current parameters) to the specified file. Files are written in a human readable format of 
the form: 

*** SYSTEM‐WIDE PARAMETERS 
Data  directory: \data 
*** ALTUS EVT FORMAT DATA ARCHIVER 
Output directory: \data\events\ 
Directory  layout: FLAT 
Channel list: * 
Julian  Day:  false 
Separate by digitizer: false 
ZIP multiple  files:  false 
Keep ZIP only: false 
*** CGS ALTUS TELEMETRY VIA TCP/IP 
TCP  Port: 9801 
Channel list: * 
Enable modem use: false 
Dial  on  event:   false 
Dial on daily checkin:   false 
...etc 
DIS ADV file ‐ Writes parameters (incl advanced) to a    file 
 
This command is similar to the DIS WRI command above, but includes advanced parameters 
normally hidden to reduce complexity. A sample of some advanced parameters included 
would be: 
 
Call Window Duration: 15 
Call Power Warmup: 0 
Compressed mode: true 
Buffer   size:   0 
Simulate Altus naming (Advanced):  true 
Dynamic Packet sizing (Advanced): 0‐Dynamic 2048 
EDIT ‐ ignored 

  177 
8.1. TERMINAL MODE CHAPTER 8. ALTUS EMULATION

Edit command as traditionally defined in the Altus is not supported. In general, parameters 
are contained in the config.jar configuration file and should be edited using either the web 
interface or the RockTalk software. 
As a result, some changes which would normally be made in Altus EDIT mode are available 
as commands available from the “*” prompt. In these cases, the changes to the operation of 
the system take place immediately. 

EEPROM ‐ Displays status of board ID EEPROMs 

Reads and displays the contents of the ID PROMs associated with the various boards in the 
system. The output from this command will take the following form: 

* eeprom 
PartNumber=112110 
DashNumber=01 
BoardRev=A 
PartsListRev=A 
SerialNo=21 
Bridge=1 
ESerialNo=1C00700903000058 
Built=4907   
Tested=0108 
VendorNumber=49070021 
BoardType=BlackFin 
BFNumber=0 
GPS  ‐ GPS  status functions 
GPS DIAG on|off|val ‐ Timing system diagnostics 
vals:  1=Time  reports  2=Time verifies 
vals:  4=Time  quality  8=Time syncs 
vals: 10=Position reports 20=State of health 
vals:  40=Satellites  &   SNRs 

Generates Altus‐style GPS diagnostic information with some additional information avail‐ 
able (such as time quality). Unlike the Altus though, which provides for “all or nothing” 
diagnostic output, the Rock+ digitizer allows you to selectively enable and disable different 
diagnostics by specifying different values ranging from 0 (nothing) to 7F (everything). 

GPS ID ‐ Timing system type and version 
GPS OFF ‐ Timing system power off 
GPS ON ‐ Timing system power on 
GPS  STAT  ‐ Timing system status 
HELP ‐ Displays help on all available commands 
MD ‐ Make directory command ‐ Makes a new subdirectory 
MD dir ‐ Makes the subdirectory  ’dir’ 
MKBATCH ‐ MKBATCH creates a batch file by typing in lines of commands 
MKBATCH file ‐ Creates the specified batch file 

  178 
CHAPTER 8. ALTUS EMULATION 8.1. TERMINAL MODE

See Batch Mode 

MKDIR ‐ Make directory command ‐ Makes a new subdirectory 
MKDIR dir ‐ Makes the subdirectory  ’dir’ 
MODEM ‐ Provides limited modem control parameters 
MOD CHE hh mm dd ‐ Sets daily checkin at ’hh:mm’, skipping ’dd’ days 
MOD  CHE  ‐1  ‐ Disables  daily checkin 
MOD  PHO  n  num  ‐  Sets  modem  dial  out  phone  number  ’n’  (1  or  2)  to  ’num’ 

See EDIT command, above. Changes to the operation of the modem take place immediately. 

NWDEFER ‐ NWDEFER Defers network triggering temporarily 
 
When using network triggering, allows the user to temporarily disable outgoing network 
triggering for maintenance or test purposes so that interconnected units will not cross‐trigger. 
If the local unit triggers while network triggers are deferred, then the local unit will still 
trigger but will not inform other units of the trigger. Instead, a log entry will be made that 
indicates the  deferral. 

NWDEFER secs ‐ Defer network triggering for ’secs’ seconds 
 
Time value must be positive. Realistically, there is no maximum time with the maximum 
being a signed 32 bit number, so zero to 2,147,483,647 (~68 years). 

NWDEFER 0 ‐ Cancel network triggering deferral 
NWDEFER ‐ Display current network triggering deferral status 
PASSWORD ‐ PASSWORD command unlocks protected commands 
PAUSE ‐ PAUSE Pause and wait for user input 

Used with batch files to wait for user input. See Batch Mode 

Q ‐ Quit command ‐  ignored 

Exit command from EDIT mode, which is not used. See EDIT command. 

RD  ‐ Deletes empty  directories 
RENAME ‐ Renames or moves files or directories 
REN file  file  ‐  Renames a  file  to  another  name 
REN  file dir  ‐ Moves  a  file to  a directory 
REN  dir dir ‐ Renames  a directory to another     name 
REN  file file FW  ‐ Renames  rt.log, err.log, and  batch files 

See notes regarding special firmware files in the description of the CD command. 

  179 
8.1. TERMINAL MODE CHAPTER 8. ALTUS EMULATION

RMDIR ‐ Deletes empty directories 

See notes regarding special firmware files in the description of the CD command. 

RUN ‐ RUN runs a batch file by executing commands 
RUN file 

See Batch Mode 

RX ‐ XMODEM Receive command 
SLEEP ‐ SLEEP Sleep the specified number of seconds 
SLEEP  seconds 

Used with batch files to delay a specified amount of time. See Batch Mode 

STATUS ‐ Reports current status  information 
SUCCESS ‐ Causes completion of call by clearing dial‐out    reasons 

Only available with the CGS Altus Emulation modules. 
Used to declare successful termination of a call. The CGS Altus Emulation modules specifi‐ 
cally requires both a SUCCESS command to declare the conversation complete and an ANS 
command to hangup the line. 

SUM ‐ Displays event file summaries 
SYSTEM ‐ SYS performs system operations 
SYS REBOOT ‐ Reboots the Operating System 

Reboots the entire system including the underlying Operating System. 

SYS RES ‐ System reset reboots   Rockhound 

Restarts just the application section of the system, without rebooting the Operating System. 

TIME ‐ TIME sets or displays system time 
TIME ‐ Displays current  time 
TIME yy mm dd hh mm   ss 
TX ‐ XMODEM Transmit command 
VERSION ‐ Reports version information 
Example: 
* ver ROCK Digitizer Software 3.1.2 
P/N 304708 
Copyright (c) 2009, Kinemetrics, Inc. 
Last Revised: July 9,  2009 
BF0;  BF=0.1.76,CalPSOC=0006,CPBL=0002B,CPLD=nc/nc/nc/nc 

  180 
CHAPTER 8. ALTUS EMULATION 8.2. BATCH MODE

BF0;  BF  @  090225  11:11  0176 


WINDOW ‐ WINDOW manages timed recording windows 
WIN NEW ‐ Defines a new timed recording window 
WIN  DEL  x  ‐ Deletes  a  timed  recording  window 
WIN  ‐ Lists all current  timed  recording   windows 

Used to create and manage timed recording windows. The user can create timed windows 
for one time or daily repetition, and can list or delete windows as needed. Examples: 

* win new 
Duration (sec):  90 
Hour  (0‐23):  11 
Minute (0‐59): 43 D 
aily?   (Y/N):  n 
Month  (1‐12): 3 
Day  (1‐31):  29  Y 
ear (2008‐2099):  2010 
1:  Mar 29,  2010  11:43,  Duration  90  secs 
* win new Duration (sec): 120 
Hour  (0‐23):  13 
Minute (0‐59): 30 
Daily?  (Y/N): y 
1:  Mar 29,  2010  11:43,  Duration  90  secs 
2:  Jul  14,  2009  13:30, Duration 120 secs  (DAILY) 
* win 
1:  Mar 29,  2010  11:43,  Duration  90  secs 
2:  Jul  15,  2009  13:30, Duration 120 secs  (DAILY) 
* win  del  2 
Deleted  timed  recording  window 2 
\\\ ‐ Forces command mode ‐ Ignored in command mode * 
 
8.2 Batch Mode 
Altus emulated command mode supports a limited batch mode process to make it somewhat 
easier to perform routine activities such as periodic maintenance. The MKBATCH command 
is used to create a batch file by name, accepting input until RETURN only is entered. Lines 
of the file are executed as commands as if typed by the user. Lines starting with “;” or “#” 
are treated as comments and ignored. 
The batch file will be created with the extension “.BAT” on the digitizer. Batch files can 
also be uploaded or downloaded using the TX and RX commands with the “FW” qualifier. 
For example: 

* mkbatch mybatch 

  181 
8.3. STREAMING CHAPTER 8. ALTUS EMULATION

Type in lines of commands, one command per line. 
When  done,  press  RETURN  only  to close the file. 
> ; Quarterly Maintenance 
> ; Get overall unit  status 
>  stat  >  ; Get  GPS status 
>  gps  stat 
> ; Do  a 2 minute battery    test 
>  bat 2 
> ; Hangup 
>  ans 
>  ; Wait  a bit 
>  sleep 30 
> ; Generate an FT, which should call   back 
>  aq ft 
> Batch file creation complete. 

To run the batch file, type: 

* run mybatch 
 
8.3 Streaming 
Streaming Data is simulated from the Altus Emulation module. Parameters for Streaming 
Data  include: 

• Channel selection 
• Buffer size 
• Data compression mode 
• Port selection and baud rate are available when using the RS‐232 versions of the module 
 
8.4 Dial On  Event 
Similar to the Altus, dialing options are provided which are set within the module’s param‐ 
eters and include: 

• Phone numbers 
• Initialization and modem control  strings 
• Dialing reasons (event, loss of GPS, low battery, etc) 

  182 
CHAPTER 8. ALTUS EMULATION 8.5. MODEM ONLY OPERATION

• Call completion mode 
• Call retries and retry delays 
• Daily checkin 
• Modem power control windows 
• Call transaction log entries (into system log) 
 
8.5 Modem Only Operation 
Although primarily designed as an IP‐based instrument, the Rock+ digitizer can be used 
in a modem only operation mode allowing for complete command and control of the unit 
using only a modem. The basic capabilities of this mode of operation are outlined here. The 
TTYMonitor The TTYMonitor provides a protection mechanism to allow you to access a 
Rock+ digitizer even if the Rockhound software is rendered inoperative. This is important 
because the Altus Emulation module that controls the modem runs as part of the Rockhound 
layout. If the layout’s parameter were corrupted or inadvertently set up in such a way that 
the system could not run, then you would be unable to access the system through the modem. 
The TTYMonitor is a software layer that is between the Altus Emulation module and the 
underlying Operating System. Simply put, when you call the modem it is the TTYMonitor 
software that first answers the call. If you simply begin the conversation with the unit, then 
TTYMonitor transparently passes all characters directly to and from Rockhound. 
If however, the Rockhound software is unable to operate, you would call the unit and after 
TTYMonitor answers the call you would send a ‘$’ to the system, which TTYMonitor inter‐ 
prets as a request to directly access the Operating System. In this case you will be presented 
with a Linux login prompt. From the prompt, you can log into Linux and reset or correct 
the problems encountered. 
Setting up to use TTYMonitor is fairly simple and involves two basic steps: 
From Linux you run the ttymonconfig script to associate the modem’s TTY port and an IP 
port. 
From Rockhound you use the TCP/IP version of the Altus Emulation module and choose 
the same IP port. 
Running ttymonconfig would look as follows: 

ttymonconfig 
Enter the tty device 
Default: /dev/modem0 
> /dev/modem0 
Enter the modem init string 
Default: ATZE0S0=1 

  183 
8.5. MODEM ONLY OPERATION CHAPTER 8. ALTUS EMULATION

> ATZE0&C1&D0S0=1 
Enter the application port number (5001‐65534) 
Default:   9800 
> 9800 
Review the new parameters: 
TTY_DEV = /dev/modem0 
MODEM_INIT = ATZE0&C1&D0S0=1 
APP_PORT = 9800 
Press [Enter] to Continue, or [Ctrl‐c] to Quit.  
Save the parameters? 
(Y/N)?  > y 
Saving parameters... 
Auto‐start the service at boot? ( 
Y/N)? > y 
Stopping  KMI  TTY  Monitor  Service:  /usr/sbin/ttymond:  line  61:  518 
Terminated ttywatch $APP_READ_FIFO $APP_WRITE_FIFO <$TTY_DEV >$TTY_DEV 
ttymond:  Stopped! Ok 
Enabling auto‐start... 
Starting  KMI  TTY  Monitor  Service:  Ok 
ttymond: Monitoring tty /dev/modem0 and port    9800... 
 
Next you must add the Altus Emulation module via TCP/IP to your layout and configure 
it to use port 9800 as follows: 
 

Figure 67: Adding Altus Emulation 

  184 
CHAPTER 8. ALTUS EMULATION 8.5. MODEM ONLY OPERATION

Figure 68: Altus Emulation Parameters 

 
Obviously, you will want to test your system for the ability dial in and dial out before leaving 
the premises. 
 
NOTE: Use of TTYMonitor is strongly recommended if the modem will 
be the only interface to your unit. We feel this is  important enough that 
if you use the RS‐232 version of the Altus Emulation module  and  set  it 
up  for  modem  use  then  it  WILL  NOT  OPERATE 

  185 
8.5. MODEM ONLY OPERATION CHAPTER 8. ALTUS EMULATION

without specific additional setup. This is to protect you by reducing the 
possibility of field visits if the Rockhound software or configuration is 
rendered inoperable by configuration   errors. 

If you add the RS‐232 version of the Altus Emulation module and enable modem mode, the 
modem features will not work at all and you will see the following message in the log file: 

ERROR! Modem enabled as primary connection 

If you wish to utilize the RS‐232 version of the Altus Emulation module directly because 
you have a secondary IP or modem connection (or if you’re just willing to take the risk), 
then specify the connection as a secondary connection as shown. Note the detailed warning 
below: 
 

Figure 69: Secondary Port Warning 

 
8.5.1 Dialing In 
Dialing in to the Rock+ through the modem interface is the same as the Altus. Once the 
modem connects, you must give one or more backslash characters (“\”) depending on what 
the unit is doing, and you will then be presented with a command prompt that allows you 
to enter commands. 

  186 
CHAPTER 8. ALTUS EMULATION 8.5. MODEM ONLY OPERATION

8.5.2 File  Transfer 

File transfers through the modem interface can be  performed using the XMODEM RX 
and TX commands or using Altus block mode. Altus block mode allows you to use Altus 
QuickTalk or Altus File Assistant software to transfer files. 
 
8.5.3 Parameter Download 

To download your parameters through the modem, use the TX command with the “FW” 
qualifier to download “config.jar” as follows: 

TX config.jar FW 

Other special file types that can be transferred using the TX FW mechanism include: 

• Unit configuration  (config.jar) 
• Batch files  (.bat) 
• Log files (rt.log, err.log, or system.log) 
• Sensor script files (.csq) 
• Special configuration files  (.cfg) 
 
8.5.4 Editing Parameters with RockTalk 

The RockTalk program can be used to edit parameters you have manually retrieved from the 
digitizer with the TX command. To do this, configure RockTalk to support Configuration 
Archival through RockTalk Setup → Advanced Levels. 
Select the de‐archival function through the button or File → De‐archive Layout or the toolbar 
button: 

  187 
8.5. MODEM ONLY OPERATION CHAPTER 8. ALTUS EMULATION

Figure 70: RockTalk De‐archive 

 
Locate the jar file that contains the configuration that you downloaded from the unit and 
make the necessary changes using the Parameter Editor or other functions of RockTalk. Once 
the changes have been made, save the modified configuration by selecting File → Archive 
Layout or by using the toolbar button: 
 

Figure 71: RockTalk Archive 

 
Once you’ve saved the modified configuration in a file, you can send it back to the digitizer 
over the modem using the RX command as described below. 

  188 
CHAPTER 8. ALTUS EMULATION 8.5. MODEM ONLY OPERATION

8.5.5 Parameter  Upload 

To upload your parameters through the modem, use the RX command with the “FW” 
qualifier to upload “config.jar” as follows: 

RX config.jar FW 
 
After you have successfully uploaded the configuration, issue the sys res command to install 
the uploaded configuration. 
 
NOTE:  It  may  take  as  long  as  a  minute  or  two  for  the  Rockhound 
software to shutdown, unpack the new configuration and restart before 
the system is back in full operation. 

Other special file types that can be transferred using the RX FW mechanism include: 

• Firmware files  (update.jar) 
• Unit configuration  (config.jar) 
• Batch files  (.bat) 
• Sensor script files (.csq) 
• Special configuration files  (.cfg) 
 
8.5.6 Installing Firmware  Updates 

To upload a firmware update through the modem, use the RX command with the “FW” 
qualifier to upload “update.jar” as follows: 

RX update.jar FW 
 
After you have successfully uploaded the firmware, issue the sys res command to install the 
uploaded  firmware. 
 
NOTE: It may take as long as five minutes for the Rockhound software 
to shutdown, unpack the new firmware, install it, and restart before 
the system is back in full operation. The updated firmware file MUST 
be named “update.jar” in order to be installed. 

  189 
8.5. MODEM ONLY OPERATION CHAPTER 8. ALTUS EMULATION

  190 
Chapter 9 

Storage 

The unit supports various storage options including one internal System (software) SDHC 
card (required), one internal Data SDHC card (required), and additional external storage 
via USB Thumb Drives. 
Memory cards provided through Kinemetrics will be Industrial Temperature range (‐25 de‐ 
grees C through +85 degrees C) to be compatible with the overall temperature specifications 
of the unit. 
Kinemetrics DOES NOT recommend the use of commercial grade memory cards (although 
they are definitely less expensive) unless the user clearly understands that by using them 
they are degrading the overall temperature range of the equipment (and life expectency of 
the cards  themselves). 

NOTE: Use of commercial grade memory cards may compromise cor‐ 
rect operation of the unit, cause loss of important data, or both. 
 
9.1 System SDHC Flash 
The system card is internal to the unit’s case and is not physically accessible by the user. The 
card contains the operating system and applications software that will be used to operate 
the unit. Typically, user data is stored on another memory card. 
As of this writing, all Obsidians ship with a 4GB SDHC System card. 
Should it become necessary to replace the System card, please contact Kinemetrics for sup‐ 
port. 
 
9.2 Data SDHC  Flash 
The data card is internal to the unit’s case and is not physically accessible to the user. The 
card holds recorded data and working data such as ring buffer files. 

  191 
9.3. USB SUBSYSTEM CHAPTER 9. STORAGE

Data to be offloaded to a USB Thumb Drive is typically recorded on the Data card, and 
then copied to the Thumb Drive. 
As of this writing, all Obsidians ship with a 32GB SDHC Data card. 
Should it become necessary to replace the Data card, be sure to replace it with the same 
card type or an equivalent recommended by Kinemetrics. Then: 

• Power off the unit 
• Remove the left end‐cap by removing the four screws 
• Physically remove and replace the data card which is located on the third PWB from 
the front panel. This is a push‐push SD socket. 

◦ The System card is on the second PWB from the front panel. DO NOT remove 
the System card without special instructions. 

• Replace the dessicant and re‐install the end‐cap. 
• Power the unit back up 
• Log into Linux 
• Type the command “rock stop” to stop Rockhound 
• From Linux, type the command “datacardformat” and follow any instructions. 

After the system reboots, the new card is ready for use. 

9.3 USB Subsystem 
The Obsidian supports USB 2.0 with one USB Device port and three USB Host ports. 
The difference between a USB Host and a USB Device is a significant one. Generally, you 
can think of a USB Host as a computer and a USB Device as a “device”. So for example, 
when you plug a USB Thumb Drive into your computer, the computer is the USB Host and 
the Thumb Drive is the USB Device. 
In general, a Device must talk to a Host. Two Devices cannot talk to each other and generally
speaking two USB Hosts can’t either. 
The Obsidian is not a general purpose computer. It has to be programmed specifically for 
every device it supports. Only specific devices are known or expected to work, although this 
list is expected to grow over time. At the moment, only the following are true: 

• The Obsidian’s USB Device allows you to connect to your PC as a virtual COM port. 
This can be helpful when you need to log into Linux but do not have a “serial port” 
available on your computer. Later versions of Windows and Linux will automatically 
load the proper driver to talk to the Obsidian’s USB device. 

  192 
CHAPTER 9. STORAGE 9.3. USB SUBSYSTEM

• The USB Hosts can be used to connect USB Thumb Drives. As of this writing, every 
model of thumb drive we have tested has worked OK. 

• The USB Hosts can be used to communicate with a limited subset of wi‐fi and cellular 
modem adapters. 
 
We would discourage the use of USB devices that draw significant power, or the use of 
line‐powered USB hubs. 
 
9.3.1 USB Device 

When you plug the USB Device cable into your PC, the proper driver will automatically be 
loaded on most modern versions of Linux or Windows.  Make sure that your computer is 
connected to the internet in case it needs to search for the driver. 
Once identified the Obsidian USB device connection will appear to your computer as a COM 
port. On Windows you might need to look at the Device Manager to get the COM port 
number (which usually will remain the same if you use the same physical port). Once you 
know the port, you can open the connection with PuTTY or your favorite terminal program. 
Hit <RETURN> to get a login prompt and get started. 
 
NOTE:  On the vitual COM  port,  baud rate  and flow  control  values 
are  not used. 
 
9.3.2 USB  Host 

The USB Host ports are used to communicate with USB Devices such as USB Thumb Drives, 
and a limited subset of wi‐fi and cellular modem adapters. 
 
9.3.2.1 USB Thumb Drives 

The Obsidian leaves drives unmounted normally. This means that virtually all of the time 
it is safe to remove the drive without corrupting the data on it. 
The  data  formats  used  on  the  USB  Thumb  Drive  that  is  supported  by  the  Obsidian  is 
Windows  VFAT  or  Linux  EXT3.  VFAT  format  can  also  be  read  directly  by  Windows, 
MacOS, or Linux. EXT3 will require some additional software on your PC. 
Drives larger than 32GB require use of the Linux EXT3 file system. Windows filesystems 
that support drives larger than 32GB will likely not be supported by non‐Windows ma‐ 
chines because the exFAT and NTFS filesystems that support drives larger than 32GB are 
proprietary to  Microsoft. 
Using an EXT3 filesystem has additional advantages over a FAT filesystem because EXT3 
is a journalled filesystem and is less easily corrupted ‐ more important the larger the drives 

  193 
9.3. USB SUBSYSTEM CHAPTER 9. STORAGE

become. Accessing EXT3 is also significantly faster if you have a lot of data files ‐ so if you 
use USB drives to offload at lot of continuously recorded data files we would recommend 
using EXT3 even for drives smaller than 32GB. 
The Rock+ has facilities to format thumb drives for an EXT3 file system and utilities such 
as Linux Reader are available to allow you to read these drives under Windows. EXT3 
formatted drives should be fully supported in MacOS and all versions of Linux. 
Software to read Linux EXT3 file systems on a Windows PC is posted on the KMI Support 
Wiki. It is called Linux Reader. 
When a USB Thumb Drive is detected, the Obsidian will identify the drive, mount it, copy 
data to it, and then unmount it. 
The directories copied to from the Obsidian to the USB Thumb Drive include everything 
stored under the /data directory with the exception of: 

• /data/rb (normally used to hold an RBServer ring buffer) 
• /data/sys (reserved for future use) 

Data will be copied to the directory /kmi/data/uname‐unitid where: 

• uname is the Obsidian’s host name 
• unitid is the Obsidian’s Tag number (serial number) 

The Obsidian can support up to three USB Thumb Drives, one on each port. The three 
drives are treated as “mirrors” of each other, and all will get copies of the same data. Make 
sure the USB Thumb Drive is large enough  to copy off all data you want.  There is no 
provision for spanning data over multiple smaller drives. 
The status of the data offloading operation is indidated on the Media LED as follows: 

• Off ‐ Drive is unused and may be inserted or removed 
• Green ‐ Drive is being mounted 
• Green flashing ‐ Data is being copied to/from the drive 
• Red flashing ‐ An error was detected 
 
9.3.2.2 Drive Formatting 

USB drives can be formatted in the system with the command “usbdriveformat”. 
This command will take you through the process of selecting the file format (VFAT or EXT3), 
removing and re‐inserting the data drives, and then formatting the drive itself. 
The command “usbdrvwipe” can be used to quickly erase the contents of all USB thumb 
drives. The action must be confirmed by the user, but once done cannot be undone. 
On insertion the Rock+ will adapt to use of VFAT or EXT3 formatted drives. 

194 
CHAPTER 9. STORAGE 9.3. USB SUBSYSTEM

9.3.2.3 Drive Transfer Modes 

By default the Rock+ will transfer all available data files to an inserted thumb drive. When 
new files are created, any new files will be transfered at the next opportunity. In addition, if 
files are updated (or were not complete at the last transfer) then updated copies of the files 
will be transferred. 
Drive transfer mode can be selected with the command “usbxferconfig mode” where “mode” 
is as follows: 

• ALL ‐ Transfers all files not currently stored on the drive (default) 
• LATEST ‐ Transfers only files less than 2 days old. This mode is useful when periodi‐ 
cally replacing drives to extract continuously recorded data. A small overlap protects 
against data loss without filling the drive with lots of redundant data. 
 
9.3.2.4 Temporary Insertion 

If a USB Thumb Drive is temporarily inserted to extract data, you can insert a drive and 
wait for the Media LED to blink and then go off again. Then remove the drive. If you insert 
the same drive again, only new data will be copied to the drive (which might mean no new 
data is copied). 
If you want multiple copies of the data, use multiple drives and insert and remove them one 
at a time. 
 
9.3.2.5 When left  inserted 

If you leave one or more USB drives inserted, the will be mounted and offloaded to once 
per hour. This allows you to automatically offload continuously recorded data (or a lot of 
recorded events).  Note: 

• If multiple drives are used, then will be serviced one at a time in sequence 
• The hourly offloads are not necessarily on the hour 
 
9.3.2.6 Firmware updates 

The USB Thumb Drive can be used to install software/firmware updates. 
To install firmware from the drive, place the update files in the folder /updates on the drive. 
Insert the drive and wait for the Media LED to blink and then go off again, then power cycle 
the unit to install the software. 
The software files that can be installed are as follows: 

• Rockhound  updates  named rockhound_update‐XXX.jar 

  195 
9.3. USB SUBSYSTEM CHAPTER 9. STORAGE

• Tomcat (web server) updates named tomcat_rh_update‐XXX.war 
• Linux updates named NN‐xxxxx (where NN is a numeric value and xxxxx is any string) 
• A Multi‐update file named Multi_update‐XXX.tgz 
 
9.3.2.7 Passwording 

Passwording is often not necessary since use of a USB Thumb Drive requires that you have 
physical access to the Obsidian (meaning you could just steal the unit rather than extracting 
data). 
However, a passwording provision exists as follows: 
 
• The command usbdatapass can be run from Linux to set a required password that 
must match the contents of the file /data/usb_data_pass on the  drive. 

• The command usbupdpass can be run from Linux to set a required password that must 
match the contents of the file /updates/usb_update_pass on the drive. 

If the password does not match, the protected operation will not happen. 
 
9.3.3 USB Wifi 

The Rock+ supports the use of a TP‐Link TL‐WN721N and the NetGear WNA1100 USB 
WiFi adapters. While other adapters may work, they have not been tested and are 
unsupported. 
The USB WiFi adapter can operate as a client or in Access Point (AP) mode. 
Configuration requires console access and root login. 
Items Needed Before Configuration: 

1. A WiFi access point if using client mode. 

2. The SSID and password. 

3. Knowledge of the security mode (WPA/WPA2, WEP, or Unsecured) 

4. An available IP address if using a static IP. 

5. A known free channel in your location if using Access Point mode. 

  196 
CHAPTER 9. STORAGE 9.3. USB SUBSYSTEM

9.3.3.1 Setting Up WiFi 

netconfig is used for Wifi configuration. For the maximum WiFi options you should choose 
to set ʺeth0ʺ disabled. If you want to use Access Point mode then ʺeth0ʺ must be disabled. 
If you want to use WiFi with DHCP then you may not use DHCP for ʺeth0ʺ, but you may 
use a static  IP. 
Run netconfig and follow the prompts. 
After you get past the setup for ʺeth0ʺ you will be asked about the ʺMode for USB wireless 
adapter wlan0ʺ. Here you can make your choices and supply the info collected above. 
After completing netconfig reboot and verify network operation. 
 
9.3.3.2 Wifi Verification 

If you chose client mode you can verify operation by pinging an outside host. If you chose 
Access Point mode you will need a PC/laptop/etc which you can configure to connect to the 
unit and observe on that device that you are successful. 
 
9.3.4 USB Cellular Modem 

The Rock+ supports 2 models of USB Cellular Modem manufactured by MultiTech, the 
MTC_H5  and  the MTC_EV3. 
The MTC_H5  is  a GSM modem which supports  the ATT  network. The  MTC_EV3  is  a 
CDMA modem which supports the Verizon network. 
You will need to register your modem with the provider and have an active account with 
them before setup. You must be sure to deploy the modem in a location that has adequate 
RF signal strength. This is can be verified on the spot with a cell phone from the SAME 
provider. These modems require an external antenna which MUST be connected. Do not 
plug in the unit without an antenna connected. 
The following will guide you through a full‐time modem setup. An as‐needed connection is 
also possible and requires choosing the ʺAS_NEEDED_CELLULARʺ option in pppconfig 
instead. An as‐needed connection will only be active when Rockhound deems it necessary. 
Configuration requires console access and root login. 
 
9.3.4.1 Setting Up The MTC_H5_GSM On ATT 

Be sure the antenna is connected then plug the USB cable into a Rock+ USB host port. 
You will need the following from the provider: 

1. A registered SIM card installed in the modem. 

  197 
9.3. USB SUBSYSTEM CHAPTER 9. STORAGE

2. Dial‐Out Ph#: Typically ʺ*99#ʺ 

3. APN String: i.e: ʺi2goldʺ 

Run ʺpppconfigʺ and choose ʺFULLTIME_CELLULARʺ. 
Continue and supply the above provider‐supplied info when prompted. 
You will also be asked for a ʺReliable host IP addressʺ that the PPP process can occasionally 
ping to verify connectivity. You should supply an IP address for a public host that you own 
which has a high up‐time. For example, your organization’s web server. This host must be 
able to respond to ping requests. 
You will also be asked for a ʺNetwork IP address for PPP traffic routingʺ. This is typically 
left BLANK, and should only be used by advanced users who need a special route added. 
After pppconfig completes be sure to REBOOT and verify network operation. 
 
9.3.4.2 Setting Up  The  MTC_EV3_CDMA  On   Verizon 

Be sure the antenna is connected then plug the USB cable into a Rock+ USB host port. 
The MTC_EV3_CDMA modem requires a 1‐time activation on the Verizon network. This 
is best done in the same general location that the modem will be deployed, and requires that 
adequate RF signal strength be available. 
To activate the modem be sure that no modem is currently configured. (You can use ppp‐ 
config to delete a previous config if needed.) 
Run the following to activate the modem and follow the prompts: 
 
activate‐mtc‐ev3 
 
This usually completes in about 1 minute. Activation only needs to be done once, unless 
you change regions i.e: Moving from the West coast to the East coast. In this case re‐run 
activation once at the new site. 
You will need the following from the provider: 

1. Dial‐Out Ph#: Typically ʺ#777ʺ 

2. Dial‐Out user: Cell Ph# for modem. 

3. Password: i.e: ʺvzwʺ 

Run ʺpppconfigʺ and choose ʺFULLTIME_CELLULARʺ. 
Continue and supply the above provider‐supplied info when prompted. 
You will also be asked for a ʺReliable host IP addressʺ that the PPP process can occasionally 
ping to verify connectivity. You should supply an IP address for a public host that you own 

  198 
CHAPTER 9. STORAGE 9.3. USB SUBSYSTEM

which has a high up‐time. For example, your organization’s web server. This host must be 
able to respond to ping requests. 
You will also be asked for a ʺNetwork IP address for PPP traffic routingʺ. This is typically 
left BLANK, and should only be used by advanced users who need a special route added. 
After pppconfig completes be sure to REBOOT and verify network operation. 
 
9.3.4.3 Cellular Verification After  Setup 

After reboot you may need to wait a couple minutes for the modem to establish a connection. 
You can verify a full‐time cellular connection by typing “ifconfig” and observing that “ppp0” 
is present. 
If you chose an as‐needed connection then you can manually force it to connect by typing 
“pppdial”, then use “ifconfig” to see that “ppp0” is present. Note: This connection will drop 
after a few minutes of inactivity. 
You should also ping an outside host to verify the connection. 
i.e:  ping nnn.nnn.nnn.nnn 

  199 
9.3. USB SUBSYSTEM CHAPTER 9. STORAGE

  200 
Chapter 10 

DFS  telemetry 

The DFS protocol is used for low latency systems such as early warning systems that cannot 
tolerate the latency of a standard 1 second data packet. 
 
NOTE:  The  Obsidian  also  supports  a  low  latency  mode  where  data 
packets are produced every 0.1 second. The telemetry interface mod‐  ule 
must  also  support  these  low  latency  packets.    As  of  this  writing,    the  
Rock+  RBServer  module  does  support them. 
 
In  the  Obsidian,  the  DFS  protocol  is  output  through  a  built‐in  DFS  port.  This  port  is 
connected directly to the primary DSP board in the Obsidian, and is capable of outputting 
data from physical channels 1‐4. It is not possible for DFS to output data for channels 5 
and above. 
DFS data is output at 38400 baud and can produce data at sample rates up to 100sps. 
DFS data is output with minimal latency, just after the samples have been FIR filtered. So 
at 100sps the data has a latency of 0.01 seconds (1sec/100sps). FIR filter delays are not 
included in this latency. At 100sps, the filter delays are 108.5ms causal and 493.5ms acausal. 
The DFS protocol is described below. 
 
10.1 Software Provided 
Typically, the DFS protocol is used by the customer to construct an early warning system 
by catching the DFS output stream from one or more digitizers. 
Kinemetrics does not provide end‐user software to catch and/or process the DFS protocol. 
This is done by the customer or a third party developer and has been successfully imple‐ 
mented by many  customers. 

  201 
10.2. CONFIGURATION CHAPTER 10. DFS TELEMETRY

10.2 Configuration 
Configuration of DFS requires several things: 

• Your system must be configured to produce one of the supported DFS sample rates: 
1,10,20,50 or 100sps. 
• You must set the DFS sample rate of the Rock2 Data Interface to match the output 
sample rate 
• You must select a DFS Mode in the Rock2 Data Interface (see Protocol description 
below) 
• You must have software capable of catching the DFS output 

NOTE:  If  you  do  not  produce  the  selected  DFS  rate,  for  example  you 
produce 200sps data and set the DFS Rate to 100sps, then no DFS    data 
will be output. 
Though  DFS  can  produce  data  at  up  to  100sps,  your  system  can  be 
configured  to  output  multiple  sample  rates.  You  could  for  example 
produce data at 100sps and 200sps ‐ triggering and recording the 200sps 
data on the Obsidian, and outputting the 100sps data through DFS. 
 
10.3 Protocol 
The DFS implementation in the Obsidian supports two versions of the DFS protocol. 
In both cases, the data is a fixed size binary packet that ends with 0xA and 0xD. 
The protocol is simplex, meaning it is transmit only with no provision for error detection 
or error recovery. So your connection between your computer and the DFS port much be 
a ʺcleanʺ line with minimal chances for data corruption. The computer receiving the data 
must be dedicated to the task and must not miss transmitted data as there is no recovery 
mechanism. 
 
10.3.1 Original DFS  Protocol 
DFS Mode 1 is the original DFS protocol, as has been used in early warning systems with 
K2s and Etnas for many years. 
This data is truncated to 16 bit data in ADC counts (16 bits MSB), and includes no sequence 
numbers or time stamps. 
Data packets are a fixed size with two bytes per channel (up to 4 channels) followed by 0xA 
and 0xD. 
 

  202 
CHAPTER 10. DFS TELEMETRY 10.4. TESTING

So for example, a three channel DFS packet would consist of eight bytes: 
Ch1L,Ch1H,Ch2L,Ch2H,Ch3L,Ch3H,0xA,0xD 
 
10.3.2 Extended DFS  Protocol 

DFS Mode 2 is the extended DFS protocol, and has been expanded to provide full 24 bit 
data resolution and a timing system (e.g. GPS) derived timestamp. 
This data is 24 bit data in ADC counts. 
Data packets are a fixed size with a five byte timestamp, followed by three bytes per channel 
(up to 4 channels) and then terminated by 0xA and 0xD. 
So for example, a three channel DFS packet would consist of sixteen bytes: 

T0(L),T1,T2,T3,T4(H),Ch1L,Ch1M,Ch1H,Ch2L,Ch2M,Ch2H,Ch3L,Ch3M,Ch3H,0xA,0xD 

Time is encoded as 10ms increments since epoch (Jan 1, 1970). 
 
10.4 Testing 
Though Kinemetrics does not provide end‐user software for catching and processing DFS 
data, you can use the Rockhound software to demonstrate functionality of the DFS connec‐ 
tion. 
To do this, configure a Rockhound installation on a PC using the Rockhound Test module 
DFS Test Input Source to receive data from a COM port connected physically to a properly 
configured DFS connection of an Obsidian. 

  203 
10.4. TESTING CHAPTER 10. DFS TELEMETRY

  204 
Chapter 11 

FAQs 

11.1 What is  “dig1”? 


Sometimes in looking at the parameters or other information, you’ll see “dig1”. This rep‐ 
resents the current digitizer (for example, your digitizer). This is used because the general 
Rockhound software is capable of collecting data from more than one digitizer (as is some‐ 
times done when Rockhound is run on a PC or on a Slate). 
It would also be possible and useful in some applications to collect data from another digitizer 
(such as a Q330 or a K2) and collect and process it within the digitizer. In this case, the 
digitizer would be “dig1”, and a second digitizer would be “dig2”. 
 
11.2 Changing trigger levels 
To change trigger levels, use the web interface to open Configuration parameters, and then 
locate the trigger for the selected virtual channel: 

  205 
11.2. CHANGING TRIGGER LEVELS CHAPTER 11. FAQS

Figure 72: Trigger Levels 

 
Trigger levels are the level at which the system will decide that a channel is triggered and that 
it should contribute that channel’s votes toward triggering the entire system. By default, 
the system uses threshold triggers, which are set as a percentage of full scale. So if your full 
scale range is +/‐20V and your threshold is set to 2%, then your trigger level is 2% of 20V, 
or 0.4V. 
After completing your changes, be sure to press OK to save your changes. 
Once you have completed your parameter changes, activate the completed parameter changes 
by selecting ʺApply  Changes”. 

  206 
CHAPTER 11. FAQS11.3. USING DIFFERENT TRIGGERS OR TRIGGER FILTERS

11.3 Using Different  Triggers  or  Trigger Filters 
To use a different type of trigger (e.g. an STA/LTA trigger rather than a Threshold Trigger), 
or to use a different trigger filter (e.g. an IIR Filter rather than the Classic Strong Motion 
Filter), you need to replace the corresponding modules in the layout. Note that you can 
select different triggers and different trigger filters for each virtual channel. For example, to 
replace the Threshold Trigger for virtual channel 3 with an STA/LTA Trigger, use the web 
interface. Choose Module Replace from the Tools menu: 
 

Figure 73: Module Replace 

 
You will now see: 

  207 
11.3. USING DIFFERENT TRIGGERS OR TRIGGER FILTERSCHAPTER 11. FAQS

Figure 74:  Module Replacement Options 

 
Click “Replace” for the Channel 2 Threshold Trigger and you’ll see your choices: 

  208 
CHAPTER 11. FAQS11.3. USING DIFFERENT TRIGGERS OR TRIGGER FILTERS

Figure 75: Replacing Trigger Module 

 
If  you  select  “Choose”  next  to  STA/LTA, then  you’ll  replace  the  Threshold  Trigger  for 
channel 2 only.  If you select “All”, you’ll get a chance to replace all Threshold Triggers  in 
the  system  with  STA/LTA Triggers.  If  you  select  “All”,  then  the  Module  Replacement 
screen will now look like this: 

  209 
11.4. ADDING CHANNELS CHAPTER 11. FAQS

Figure 76: Modules Replaced 

 
Once you have completed your parameter changes, activate the completed parameter changes 
by selecting “Apply Changes”. 
 
11.4 Adding Channels 
Strictly speaking, you cannot “add” channels to an existing layout. 
If your requirement for additional channels is something that you expect to change once and 
then not change for quite some time (for example, if a newly deployed unit will be shared 
between two agencies – one requiring data at one sample rate, and another at a different 
sample rate), then the approach is to rebuild the layout using the Layout Wizard. Tell the 
Layout Wizard (for example) that the unit has 12 physical channels and 24 virtual channels. 

  210 
CHAPTER 11. FAQS 11.4. ADDING CHANNELS

Use the Channel Mapping feature to assign the virtual channels to each physical channel at 
two different sample rates. For example: 
 

Physical  Virtual 
1  Physical 1 at 100sps 
2  Physical 2 at 100sps 
3  Physical 3 at 100sps 
4  Physical 4 at 100sps 
1  Physical 1 at 10sps 
2  Physical 2 at 10sps 
3  Physical 3 at 10sps 
4  Physical 4 at 10sps 
etc. . . 
The Layout Wizard will also add Channel Extractors, Trigger Filters, and Triggers for the 
duplicated channels. You can either: 

• Delete the unneeded modules (probably the best choice) ‐or‐ 
• Set the voters for the unneeded channels to zero 

Finally, for recording purposes, you’ll have to decide whether to record all channels at both 
sample rates when there is a trigger, or to limit recording to only one set of channels, using 
the Channel List parameter. 

  211 
11.4. ADDING CHANNELS CHAPTER 11. FAQS

Figure 77: Channel List 

 
A Channel List of “*” will record all channels. A channel list of “dig1:1, dig1:2 . . .  dig1:12” 
will record only the first 12 channels. 
If you expect to have some channels that you want to record some times, but not at others, 
then you still need to recreate the layout, but you can then turn channels on and off by 
going to the Chanel Mapping selections for the channels you want to change and selecting 
“disabled”: 

  212 
CHAPTER 11. FAQS 11.5. SWITCHING BETWEEN LAYOUTS

Figure 78: Channel Mapping 

 
See Layout Wizard . 
 
11.5 Switching Between Layouts 
If you will sometimes operate in one mode, and then another time operate in a completely 
different mode, you can save two or more layouts by keeping a copy of the layout and the 
parameter configuration for each, all of which is kept in a single file called “config.jar”. 
The first thing you need to do is to build a layout and configure the parameters the way you 
want them to be. Then, from the web interface go to Admin Details from the Tools menu and 
click Parameters to download and save a copy of the parameters. You can give this file you 
are saving a name that is meaningful to you such as “event‐recorder‐at‐10sps‐and200sps.jar”. 

  213 
11.6. CHANGING OUTPUT DATA FORMATS CHAPTER 11. FAQS

Repeat this process with as many system configurations as you need. 
To put a layout that you’ve saved back on the system, use the web interface and go to Admin 
Details from the Tools menu and find Send Parameters File. Browse to the .jar file that you 
want to send back and then click Send. After the file is uploaded, click Verify/Apply Patch 
to verify and load it: 
 

Figure 79: Sending Parameters 

 
11.6 Changing Output Data  Formats 
Rockhound‐based digitizers do not require you to use a single data format.  They are not, 
for example restricted to creation of just EVT files, or just MiniSEED, or any other single 
format. To use a different data format for created files than that currently configured, you 
need to replace the corresponding data archiver module in the layout.  For example, 

  214 
CHAPTER 11. FAQS 11.6. CHANGING OUTPUT DATA FORMATS

to change the data format from Kinemetrics EVT file format to MiniSEED, use the web 
interface. Choose Module Replace from the Tools menu. You will now see: 

Algorithm 11.1 Module  replace   

Figure 80: Module Replace 

Click “Replace” for the Altus EVT Format Data Archiver and you’ll see your choices: 

  215 
11.6. CHANGING OUTPUT DATA FORMATS CHAPTER 11. FAQS

Figure 81: Format Choices 

 
If you select “Choose” next to MiniSEED Format Data Archiver, then you’ll replace the 
Altus EVT Format Data Archiver with MiniSEED. The Module Replacement screen will 
now look like this: 

  216 
CHAPTER 11. FAQS 11.7. TIMED RECORDING

Figure 82: Output Format Changed 

 
Once you have completed your parameter changes, activate the completed parameter changes 
by selecting “Apply Changes”. 
 
11.7 Timed Recording 
Timed recording is done by triggering the system using an Alarm Clock module. 
The module reads a set of configured times that will cause an action to occur at specified 
times according to Days of the Week, Days of the Month, Times of Day, or Alarm minute 
interval  parameters. 
When an alarm clock is triggered, the Alarm Clock’s “Alarm message” parameter will be 
sent to the system just as if it had been typed in by the user at the Rockhound Console. 

  217 
11.7. TIMED RECORDING CHAPTER 11. FAQS

Parameters used are: 

• “ Days of the Week ” ‐ Space separated days of the week (M T W TH F SA  SU) 
• “ Days of the Month ” ‐ Space separated days of the month (1 2 3 ... 31) 
• “ Times of Day ” ‐ Space separated times of day (2:15 15:54) 
• “ Alarm minute interval ” ‐ Minute interval between alarm actions. If used alone, one 
alarm action will be generated immediately (assuming no conflict with the Days of the 
Week or Days of the Month parameters), and the next one generated the number of 
minutes specified from the current time. 

• “ Alarm message ” ‐ Message to be sent when the alarm clock triggers. This message 
should be specified as a command, such as “KBTRI 600”. 
 
NOTE:  Alarm minute interval parameter is not limited to small num‐ 
bers. For example, it could be used to indicate alarm actions every 5 
minutes, 360 minutes (6 hours), or 129600 minutes (90  days). 
If Alarm minute interval is used in conjunction with the Times of Day 
parameter, indicates minute intervals relative to the  first  Times  of  Day 
parameter,  which will  act  as  the  base  time.  In  this  case,  once  the  base 
time is set, the Times of Day parameters will be overridden and ignored. 
 
Examples: 

• To record 10 minutes of data every 3 days at 2:16 pm: 

◦ Times of Day  =14:16 
◦ Alarm minute interval =4320 
◦ Alarm message =KBTRI  600 

• To record 10 minutes of data every Monday and Friday at 8:00 am: 

◦ Times of Day  =8:00 
◦ Days of the Week =M  F 
◦ Alarm message = KBTRI 600 

• To record 10 minutes of data once an hour on the hour: 

◦ Times of Day  =12:00 
◦ Alarm minute interval =60 
◦ Alarm message = KBTRI 600 

  218 
CHAPTER 11. FAQS 11.7. TIMED RECORDING

• To record 10 minutes of data on the first and 15th of the month at 8:00 am: 
◦ Days of the Month =1 15 
◦ Times of Day  =8:00 
◦ Alarm message = KBTRI 600 

• To do timed recording for 60 seconds at 8:00 am and 2:00pm every day: 
◦ Times of Day =8:00  14:00 
◦ Alarm message =KBTRI  60 

See the section Adding Modules for the procedure on how to add a new module into the 
system. Once added to the system, the parameters for the Alarm Clock module can be found 
under Configuration in the Layout menu and can be edited as shown here: 
 

Figure 83: Alarm Clock 

  219 
11.8. NETWORK TRIGGERING CHAPTER 11. FAQS

Remember that multiple Alarm Clock modules may be added to the system to record at 
different times. 
 
11.8 Network Triggering 
The Rock+ family digitizers support interconnected triggers over a network connection. This 
interconnect mechanism is network based only and is not compatible with earlier TRIG‐ 
IN/TRIG‐OUT interconnect schemes such as those used in the Altus. 
The factory default Rock+ digitizer event recorder configuration includes a Network Trigger 
module, but you may choose to add the module manually should you choose to build your 
own layout. Adding the Network Trigger module is done similarly to how other modules are 
added to the layout. See the section Adding Modules . Note that only one Network Trigger 
module may be added to the system. 
The Network Trigger module parameters can be found under Configuration in the Layout 
menu: 

  220 
CHAPTER 11. FAQS 11.8. NETWORK TRIGGERING

Figure 84: Network Trigger 

 
The default behavior of the Network Trigger module is that it is assigned one vote to trigger 
the system. 
Without making any further edits, Rock+ digitizers on the same local network will discover 
each  other  and  will  automatically  trigger  one  another.  So  if  one  unit  triggers  due  to  a 
console trigger, seismic trigger, timed recording, or sensor test, then all interconnected units 
will trigger. 
Please note that since triggering is done through the network that trigger times may not be 
exactly the same on all units, though they will be close – typically within a second or two. 
Other considerations: 
 
• Enabling Slave Triggering will allow the unit to respond to triggers from other inter‐ 
connected units, but will not generate outbound triggers. 

  221 
11.9. TELEMETRY OPTIONS CHAPTER 11. FAQS

• Sensor tests will forward the type of trigger to other units so for example all intercon‐ 
nected units will perform sensor tests. 

• In some cases, you may wish to perform a console trigger or sensor test on one unit 
during maintenance and NOT want to cause triggers on all interconnected units. To 
allow for this, you can use the console command NWDEFER secs to defer network 
triggering for a specified amount of time. For example, NWDEFER 300 would give 
you 5 minutes to perform your tests with the interconnection temporarily disabled. 
After five minutes, interconnection would automatically be  re‐enabled. 

• The Excluded hosts lists allow you to manually specify units within the local network 
that should NOT be treated as part of the interconnection. 

• The Included host’s lists allow you to manually specify units outside of the local net‐ 
work that are to be included. For example, units might be interconnected from within 
two separate networks. An example might be two units located at opposite ends of 
a very long bridge that are connected to separate networks. By giving each unit the 
others’ addresses in the Included hosts list they will be able to communicate with each 
other and command interconnected triggers. 
 
11.9 Telemetry Options 
The overall design of the Rockhound software allows for many optional modules. Some of 
these modules include options for telemetry (real time data streams). Presently the modules 
available to support this capability include (but are not limited to): 
 
• The “Telemetry Connection” module used to produce the Waveform Data displays. 
See Waveform Viewing . 
• The Altus‐style Serial Data Streams (SDS) protocol is supported through the Altus 
Emulation modules. 
• The low‐latency DFS telemetry stream is supported through an optional DFS telemetry 
option board for channels 1 through 4. 
• A Ring Buffer module, compatible with BRTT’s Antelope software. 
• A third party add‐on module to act as a SEEDLink server is available through ISTI. 
• A third party add‐on module to act as an Earthworm server is available through ISTI. 
 
Another non‐real time monitoring and data collection option would be to use the Rock 
Monitor software as described in Application Note 79. 
 
 
 
 

  222 
 
 
 
 

Chapter 12 

Advanced Operation 

The Rockhound software provides for several advanced features that allow the user to add, 
remove and replace modules, allow access to advanced and/or test mode parameters, allow 
for initial layout creation and other special capabilities. 
 
12.1 RockTalk  with  Add‐ons 
RockTalk can be downloaded from the Rock+ as an executable jar file from the digitizer’s 
web page via the Layout → Layout Wizard menu selection. When you run RockTalk and 
access the configuration, the program will extract two important working files modules.cfg 
and canned.cfg into the working directory. These files are necessary for processing the 
configuration files. 
If you use 3rd party Rockhound add‐on modules such as RockToEW or RockToSLink, then 
you must manually place the file mergedcustmodules.cfg into your working directory so that 
RockTalk can process the add‐on as part of the configuration. 
 
12.2 Layout  Wizard 
The Layout Wizard allows you to create a completely new layout for your system. During 
this process, you can define the source of the data (on a Rock+ Digitizer, the source will be 
a Rock2 Data Interface), as well as defining the types of triggers, output formats, and so on. 
In the example here we’ll create a 12 channel event recorder and add the Network Triggering. 
The Layout Wizard is part of the RockTalk application, which can be installed separately 
by installing Rockhound, or can be downloaded from the digitizer’s web page via the Layout 
→ Layout Wizard menu selection. 
Start RockTalk and then select Setup → RockTalk Setup from the menu: 

  223 
12.2. LAYOUT WIZARD CHAPTER 12. ADVANCED OPERATION

Figure 85: RockTalk Setup 

Next, we’ll click “Rename” to rename this Unit to Obsidian: 

 
 
 
 
Figure 86: Renaming Unit in RockTalk 

 We’ll now use “Fast Setup” to configure the connection: 

  224 
CHAPTER 12. ADVANCED OPERATION 12.2. LAYOUT WIZARD

Figure 87:  Fast Setup 

  225 
12.2. LAYOUT WIZARD CHAPTER 12. ADVANCED OPERATION

Figure 88: Fast Setup (continued) 

 
Close RockTalk Setup by clicking OK and then select Layout Wizard from the Setup menu. 
First question about the new layout is to choose the basic type of layout. There are four 
basic types: Simple (only one data source), Complex (more than one data source), Event 
Recorder (triggered recording), or Continuous Recording (no triggering). To record event 
data on a Rock+ Digitizer, select “Simple Event Recorder”: 

  226 
CHAPTER 12. ADVANCED OPERATION 12.2. LAYOUT WIZARD

Figure 89: Select Layout to Build 

 
Next, specify the number of physical channels (actual hardware channels): 
 

Figure 90: Set Number of Channels 

 
Next select the number of virtual channels. To record 12 channels as a single sample rate, 
select 12. If you wanted to record 12 channels at two different sample rates, you’d select 24. 
In this case we’ll select 12: 
 

Figure 91: Set Virtual Channels 

 Select the data source, in this case a Rock2 Data Interface: 

  227 
12.2. LAYOUT WIZARD CHAPTER 12. ADVANCED OPERATION

Figure 92: Set Data Source 

 
Select the type of trigger filter: 
 

Figure 93: Set Trigger Filter 

 
You’ll now be asked if you want to use the same type of filter on all channels. If so, select 
“Yes” and you can avoid the next 11 questions about filters. If you select “No”, you’ll have 
to choose the filter for each of the 12 channels: 
 

Figure 94: Use Same Filter for all Channels 

 Choose the type of trigger used: 

  228 
CHAPTER 12. ADVANCED OPERATION 12.2. LAYOUT WIZARD

Figure 95: Set Trigger Type 

 
Again, you’ll be asked if all channels will use the same type of trigger: 
 

Figure 96: Use Same Trigger for all Channels 

 Next, choose the output data format: 

  229 
12.2. LAYOUT WIZARD CHAPTER 12. ADVANCED OPERATION

Figure 97: Set Output Type and Data Format 

 Finally, replace the web server by selecting Edit → Replace Modules: 

  230 
CHAPTER 12. ADVANCED OPERATION 12.2. LAYOUT WIZARD

231 
 
Figure 98: Replace Web Server 
12.2. LAYOUT WIZARD CHAPTER 12. ADVANCED OPERATION

At this point, the basic layout is complete: 
 
 
 
 
 
 
 
 
 
 

Figure 99: Layout Complete 

 Choose File → Save Layout to send the Created Layout. 
When the system restarts, log into the web interface if necessary and select Tools → Module 
Add: 

  232 
CHAPTER 12. ADVANCED OPERATION 12.2. LAYOUT WIZARD

Figure 100:  Adding Network Trigger 

 
Select Network Trigger by pressing the Add button. Since only one of these can be added 
to the layout, the choice will disappear from the list of available modules. 
If you need to edit passwords in modules that have password parameters, select Tools →
Operation/Passwords from the menu and you will get a warning that password editing is not 
secure. The warning is because passwords will be sent in the parameter data. Ideally you 
should not do this over the internet routinely or you will expose your web access passwords. 
If you are not specifically editing passwords, we recommend that you use the Layout →
Configuration function instead to edit parameters. 
Click OK to save any changes made. 
We’re almost done with the layout configuration. Add any additional modules that you need 
and verify parameters such as sample rates, voltage ranges and others as described in Basic 
Setup. Once all changes have been made, select Layout → Apply Changes Now from the 
menu. 

  233 
12.3. CONFIGURATION OPTIONS CHAPTER 12. ADVANCED OPERATION

In a few minutes, the system will have restarted and will be operational with your new 
layout. 
 
12.3 Configuration Options 
There are other configuration features that can be used to customize your Rock+ digitizer 
outside of the Rockhound configuration. These additional features are briefly described in 
this section. 
 
12.3.1 Network  Parameters 

You can set the network parameters for your digitizer using the netconfig script from Linux. 
This script allows you to set IP addresses, as well as other related parameters such as the 
use of DHCP, setting the net mask, gateway, and DNS servers. 
The netconfig script is more fully described in the NetConfig section. 
 
12.3.2 Modem Configuration 

The digitizer can use any available internal modem or serial port as a modem. These modems 
may be managed by the Rockhound software using a module such as the Altus Emulation 
modules or the Alpha‐Numeric Pager or SMS Message Sender modules, or they can be 
controlled directly by  Linux. 
If you will use a modem as your only external access to the digitizer, it is strongly recom‐ 
mended that you make use of the TTYMonitor function. This feature exists logically between 
a modem connection and a module such as Altus Emulation.  Should the Rockhound software 
by unable to respond for whatever reason, the TTYMonitor allows you to directly obtain a 
Linux login prompt through the modem connection so that you can communicate with the 
system to correct the problem. 
See the section on Modem Only Operation for more detail. 
 
12.3.3 IO  Bits 

There are general purpose outputs available in the digitizer. Two outputs are available  on 
each of  up  to  three  dual  serial  port  option  cards.  These  are  provided  to  perform  such 
functions as alarm outputs (indicating alarm level threshold exceedance), or for control of 
external devices such as powering an external modem or radio. 
The various Rockhound modules activate these signals logically by name. For example: 

• The Alarm Exceedance logic drives a signal named “AlarmGPIO” 

  234 
CHAPTER 12. ADVANCED OPERATION 12.3. CONFIGURATION OPTIONS

• The Modem power logic drives a signal named  “ModemPwrGPIO” 
• A heartbeat drives a signal named “HeartbeatGPIO” 

In order to “connect” these logical signals to actual hardware bits, you will need to add an 
additional module named Rock GPIO to the layout: 
 

Figure 101: Adding RockGPIO 

 
The module’s parameters allow you to connect each logical signal to any of the up to six 
physical hardware bits available as shown here: 

  235 
12.3. CONFIGURATION OPTIONS CHAPTER 12. ADVANCED OPERATION

Figure 102: RockGPIO Parameters 

 
Bits 1 to 3 are assigned to the Time I/O switched COMM power outputs. Depending on the 
external hardware connected to the bit, they may be logically inverted or utilized directly. 
RockGPIO functions can also be used to control an optional external relay board. 
 
12.3.4 Networking and Security 

The Rock+ digitizer provides standard SSH services as well as most standard IP services. 
The following services are enabled by default: 

• SSH 
• NTP 
• PING 

  236 
CHAPTER 12. ADVANCED OPERATION 12.4. LOW LATENCY TELEMETRY

• POC 

The following services are disabled by default: 

• TELNET 
• FTP 
• TFTP 

The TELNET and FTP services can easily be enabled if needed using the inetdconf script. 
See the IP Services  section. 
 
12.4 Low Latency  Telemetry 
The Rock+ normally provides data in 1 second packets from the data acquisition section. 
In most cases this is quite adequate for triggering, recording, and even many telemetry 
applications. It is generally a good trade‐off between latency, instrument workload and 
communication bandwidth. 
However in some cases it is desirable to reduce latency for early warning applications. For 
these purposes you need to produce data in smaller “chunks” understanding your instrumen‐ 
tation will work harder and your communications links will not be as efficient. 
First of all, what we’re talking about here is NOT the DFS protocol. That is a different 
simplex serial‐only protocol stream that is discussed in a separate section of this manual. 
See the section on DFS telemetry for more detail. 
When enabling the Rock+’s specific low‐latency features, it will produce two different data 
streams: 

• The normal 1 second data that is used for typical recording activities 
• A smaller 0.1 second data packet that is used ONLY for telemetry 

To enable this low latency mode you need to: 

• Enable Low Latency Mode in the Rock+ Data Interface 
• Enable Low Latency Mode in a Telemetry module that supports Low Latency. As  of 
this writing, only the Antelope compatible Ring Buffer module supports this Low 
Latency mode. 
 
NOTE: You MUST enable Low Latency modes in both places, or you 
will not get Low Latency   data. 

Obviously the low latency mode only makes sense for sample rates greater than 1sps. 

  237 
12.4. LOW LATENCY TELEMETRY CHAPTER 12. ADVANCED OPERATION

  238 
Chapter  13 

Software Maintenance 
 
In this section we discuss recommended maintenance and service procedures. 
The Rock+ digitizer contains no user replaceable fuses or internal batteries. 
 
13.1 Powering Up the  System 
To successfully deploy your digitizer, you will need to physically install the hardware in a 
suitable environment, provide your primary and backup power sources (if any), connect any 
communications links used, and configure the software for correct operation if the needed 
setup differs from the default factory configuration of the unit. When power is applied to 
the system, it goes through a specific power up sequence. 
When power is connected via the external DC supply: 

• All LEDs off ‐ This is the initial condition 
• Power LED red/green ‐ There may be a brief period (a few seconds) where the power 
LED alternates between red and green as the system voltages are initially enabled and 
sensed. 

• Power LED green ‐ The system is in its initial power up phase and is allowing the 
system to stabilize (this phase also prevents system motorboating in case the power 
input is marginal or erratic) 

• Power LED orange ‐ The system is charging the supercapacitor pack 
• Power LED blinking red/green, Status/Event/Media LEDs blinking red ‐ The system 
is starting up the processor 

• Power LED green, Status/Event/Media LEDs blinking green ‐ The Operating System 
is booting 

  239 
13.2. POWERING DOWN THE SYSTEC
MHAPTER 13. SOFTWARE MAINTENANCE

• Power LED green, Status/Event/Media LEDs off ‐ The Operating System has com‐ 
pleted booting 

• Power LED green, Status LED green ‐ Rockhound is being loaded 
• Power LED green, Status LED blinking fast green ‐ Rockhound is running and starting 
the DSPs 

• Power LED green, Status LED blinking slow red ‐ Rockhound has published data and 
is waiting for good time quality 

• Power LED green, Status LED blinking slow green ‐ Rockhound has publisged data 
and has good time quality 

• Ethernet LEDs active ‐ Ethernet interfaces are enabled 

When power is connected via the battery only, the sequence will be the same except that 
the Power LED will slow blink green after the power is on, indicating that it is running from 
a battery. 
 
13.2 Powering Down the System 
When input power is lost, the system goes through a specific power down sequence: 
 
• Power LED blinking red/green, Status LED red ‐ Indicates that the system is powering 
down 

• All LEDs off ‐ The system has powered off 
• Ethernet LEDs off ‐ Ethernet interfaces are disabled 
 
13.3 Software Installation 
Initial setup of your unit will require you to modify or at least review the software configu‐ 
ration. Every attempt has been made to have the default behavior of the unit be as useful 
as possible, or at least rational. However, each deployment has its own variations and may 
require configuration changes in order to be best used. 
Setting up the unit will require the use of a Serial/USB cable that supports the Console port 
connection. You’ll also need a PC or equivalent running a terminal emulation program such 
as HyperTerminal or PuTTY on Windows or minicom on Linux. 
You may need to use a USB to serial adapter if your PC does not have an available COM 
port. If so, install the software and then plug in the adapter, making note of the COM port 
assigned to the port. 

  240 
CHAPTER 13. SOFTWARE MAINTENANCE 13.3. SOFTWARE INSTALLATION

Note that you can also use the USB device connection as an alternate virtual COM port. 
Kinemetrics includes the PuTTY application as an alternative to using HyperTerminal on 
Windows, or if you are using an Operating System (such as Windows Vista) that does not 
supply  HyperTerminal.  Please  see  the  Software  Tools  section  that  includes  a  basic 
description of PuTTY setup and operation. 
Set the terminal emulation program up for 115200 baud, no flow control, 8 data bits, 1 stop 
bit, no parity. 
Factory Configuration Including: 
 

Name  Comments 
Eth0  Enabled for DHCP 
SSH  Enabled 
NTP  Enabled (client) 
NTP  Disabled (server) 
Ping  Enabled 
Telnet  Disabled 
FTP  Disabled 
TFTP  Disabled 
System card  Mounted as /dev/root 
Data card  Mounted as /mnt/data1 
Console  /dev/console 
Serial 1  /dev/serial1 
Serial 2  /dev/serial2 
Serial 3  /dev/serial3 
Modem  /dev/modem0 
The unit includes an overall configuration menu to assist you with some of the most common 
configuration  tasks.  To access  this  menu,  type  the  command  sysconfig  from  the  Linux 
prompt: 

KMI  System  Configuration Utility 
1. View  System Info 
2. Software  Versions 
3. Configure System  Time 
4. Configure Network 
5. Change  Passwords 
6. View Update  Log 
98. REBOOT 
99. Exit 
Make  a  selection  and  press  [Enter] 

  241 
13.4. IP SERVICES CHAPTER 13. SOFTWARE MAINTENANCE

13.4 IP Services 
The following IP services are enabled by default from the Kinemetrics factory: 

• SSH 
• NTP 
• PING 

The following services are disabled by default but can be enabled if needed: 

• TELNET 
• FTP 
• TFTP 

Non‐secure services (like TELNET and FTP) should be disabled or only used where network 
security is not in question (such as a protected private network). Although these services 
use passwords, the user names and passwords are sent ʺin the clearʺ meaning that any third 
party who happens to be listening can capture the needed information to compromise your 
system. 
To change the IP services configuration, run the inetdconf script after logging onto the unit 
through the console port as “root”. 
inetdconf allows you to configure, enable and disable some common IP services such as FTP, 
TELNET,  and  TFTP. 
The inetdconf exchange will look something like the following: 

# inetdconf  ftp  on 
ftp  service  enabled. 
 
‐or‐ 

# inetdconf  ftp  off 
ftp  service  disabled. 
 
‐or‐ 

#  inetdconf  ftp status 
ftp service is currently  disabled. 
 
After completing IP services configuration, it is recommended that you reboot the system 
to ensure that the services come up correctly. 

  242 
CHAPTER 13. SOFTWARE MAINTENANCE 13.5. NETWORK CONFIGURATION

13.5 Network  Configuration 


The default network configuration is as follows: 

• Ethernet (eth0): Enabled,  DHCP 

To change the network configuration, run the netconfig script after logging onto the unit 
through the console port as “root”. 
netconfig allows you to set up the IP address, netmask, gateway, DNS servers, and other basic 
networking parameters of your primary and (if available) secondary Ethernet    connections. 
The netconfig exchange will look something like the following: 

Rock2:~# netconfig 
Network  Host  Name  for this unit 
> Rock2 
Network Domain Name 

Mode for  eth0 
1. disabled 
2. static 
3. dhcp 
(Note: To use wireless AP mode you must set eth0 disabled.) 
Make  a  selection  and  press  [Enter] 
>  2 
IP Address for  eth0 
>  10.0.3.71 
NetMask for eth0 
Default:   255.255.255.0 

Default gateway 
Default:   10.0.3.1 
>  10.0.1.1 
Mode for USB wireless adapter   wlan0 
1. disabled 
2. static 
3. dhcp 
Make  a  selection  and  press  [Enter] 
>  1 
Net Watcher  is a service that attempts to recover lost network connectivity.      
Do you want to enable Net Watcher? 
(Y/N)?  > y 
Net Watcher needs a remote host to occasionally ping. This should be a    
high  up‐time  host  such  as  a  router  or gateway. 
Enter the IP address for the remote     host 

  243 
13.5. NETWORK CONFIGURATION CHAPTER 13. SOFTWARE MAINTENANCE

Default: 10.0.1.1 
>  10.0.1.1 
IP Address for primary   DNS 
>  8.8.8.8 
IP  Address  for  secondary DNS 

New  parameters  to  be saved: 
HOSTNAME  =  [Rock2]  DOMAINNAME  =  [] 
DNS1  =   [8.8.8.8] DNS2 =  [] 
PUBLIC_ADDR  =  []  NW_PING_HOST  =  [10.0.1.1] 
eth0 Parameters: 
MODE  =    [static] 
ADDR  =   [10.0.3.71]  NETMASK = [255.255.255.0]  GATEWAY = [10.0.1.1] 
wlan0 Parameters: 
MODE  =    [disabled] 
ADDR  =  []  NETMASK = []  GATEWAY = [] 
SSID  =  []  SECURITY = [] 
PASS  =   [] 
CHANNEL  =   [] 
Press [Enter] to Continue, or [Ctrl‐c] to Quit. 
Saving  parameters... 
Do you want to stop and re‐start the network NOW? 
(Y/N)?  >  y 
Restarting eth0... 
fec 63fec000.ethernet eth0: Freescale FEC PHY driver [NatSemi DP83640] 
(mii_bus:phy_addr=63fec000.etherne:01,      irq=‐1) 
libphy: 63fec000.etherne:01 ‐ Link is Up ‐ 100/Full 
 
After completing the network setup, it is recommended that you reboot the system to ensure 
that the network parameters come up correctly. Rebooting also assures proper operation of 
system services that may be terminated when the network(s) are stopped. 
To review current network parameters, type ifconfig as  follows: 

Rock2:~# ifconfig 
eth0 Link encap:Ethernet HWaddr  70:b3:d5:ce:92:8f 
inet  addr:10.0.3.71  Bcast:10.0.7.255 Mask:255.255.248.0 
inet6   addr:   fe80::72b3:d5ff:fece:928f/64    Scope:Link 
UP BROADCAST RUNNING MULTICAST MTU:1500    Metric:1 
RX packets:68597 errors:430 dropped:0 overruns:430 frame:0 
TX packets:191 errors:0 dropped:0 overruns:0 carrier:0 
collisions:0   txqueuelen:1000 
RX  bytes:5463702  (5.2  MiB)  TX  bytes:20075  (19.6  KiB) 
lo Link encap:Local  Loopback 
inet  addr:127.0.0.1 Mask:255.0.0.0 

  244 
CHAPTER 13. SOFTWARE MAINTENANCE 13.6. DYNAMIC DNS

inet6  addr:  ::1/128  Scope:Host 
UP LOOPBACK RUNNING MTU:65536 Metric:1 
RX packets:32 errors:0 dropped:0 overruns:0 frame:0 
TX packets:32 errors:0 dropped:0 overruns:0 carrier:0 
collisions:0  txqueuelen:0 
RX  bytes:2240 (2.1 KiB) TX  bytes:2240 (2.1    KiB) 
 
13.6 Dynamic  DNS 
The Rock+ supports Dynamic DNS services. 
The details for setting up DDNS can vary depending on your DDNS provider. We have 
included a popular DDNS client (ddclient) and a utility for simple configuration (ddnsconfig). 
In cases where this utility does not accomplish what your provider needs you may be able to 
manually edit the ddclient config file at ʺ/etc/ddclient/ddclient.confʺ. Your provider may be 
of assistance in this case. Kinemetrics does not track provider requirements or compatibility 
of ddclient with respect to any DDNS provider. Some providers may have requirements that 
are not met with ddclient. 
Configuration requires console access and root login. 
You will need the following from a compatible DDNS provider: 

1. An active account with your user name and password. 

2. A fully qualified hostname assigned for your Rock+ unit. Most providers offer a web 
interface for configuring this. 

3. Possibly more information depending on the provider. 
 
13.6.1 Setting Up  DDNS 

Run ddnsconfig and follow the prompts. 

1. Select whether or not to use SSL (check with your provider). 

2. Select the IP Address determination method? 

• ʺifʺ (IP address is determined from the Ethernet interface). 
• ʺwebʺ (IP address is determined by querying an outside web server). 

NOTE: If you specify ʺifʺ for interface you will be asked to choose 
between  ʺeth0ʺ  for  a  wired  network,  or  ʺppp0ʺ  for  a  cellular  mo‐ 
dem.  If you specify ʺwebʺ you must supply the provider    URL. 

  245 
13.7. SOFTWARE WATCHDOG CHAPTER 13. SOFTWARE MAINTENANCE

3. Select the protocol. 

• ddclient supports several. Choose the one that matches your provider. 

4. Supply the server URL (provider‐specific) or leave blank for defaults. 
5. Supply your DDNS account login name (chosen when you opened your account). 
6. Supply your DDNS account password (chosen when you opened your  account). 
7. Supply the fully qualified hostname for your Rock+. 
For example:  ʺmyrock.no‐ip.orgʺ 
NOTE: Your Linux hostname (set through netconfig) should match 
the hostname portion of this string i.e:    ʺmyrockʺ 

8. Review the parameters and save. 
9. Reboot and verify operation. 
 
13.6.2 DDNS Verification 
It usually takes at least 10 minutes for ddclient to push IP address changes to your provider. 
Eventually you should be able to ping the fully qualified hostname (i.e: ʺmyrock.no‐ip.orgʺ) 
and get replies. You may be able to log into your provider web interface and see more info. 
Note: It can take several minutes to hours for a DDNS provider to activate changes to your 
account. 
 
13.7 Software Watchdog 
The system includes an advanced power system that provides watchdog and powerfail pro‐ 
tection with powerfail warning capabilities. 
The watchdog provision is independent of the main processor, but does interact with it. The 
watchdog exists as part of the power supply subsystem, and requires that it be ʺpingedʺ by 
the main processor every 10 minutes. Failure to do so will be interpreted by the watchdog 
as a lockup of the main processor and the main processor will be shutdown and restarted in 
an attempt to get the system back into operation. Exceptions: 

• The watchdog is initially held off for 4 hours after processor power on, allowing for 
initial startup and worst case filesystem repair on a large system disk. 
• If operating in the system’s bootloader, the watchdog will be ʺpingedʺ every time that 
a bootloader command is executed so that the processor is allowed to continue to 
run during low‐level maintenance activities. Each bootloader command extends the 
watchdog for 4 hours. 

  246 
CHAPTER 13. SOFTWARE MAINTENANCE 13.7. SOFTWARE WATCHDOG

13.7.1 Default Behavior: 

The watchdog is normally serviced by a Linux service in order to keep the system alive. By 
default, only a failure of Linux to boot or a crash of the operating system or of the watchdog 
service will cause the watchdog to trigger and restart the system. 
 
13.7.2 Powerfail  Protection: 

The system includes an internal power backup called a supercapacitor array. This backup 
provides enough residual power to terminate any running applications and shut down Linux 
properly. Once powerfail has been detected, the system begins to operate off of the residual 
power and issues a request to Linux to shutdown. A Linux service watches for the shut‐ 
down request, and once received immediately forces all user processes to terminate and then 
shuts down Linux. When Linux has completed the shutdown, the shutdown request will be 
acknowledged and the processor will power off. 
 
13.7.3 Application Specific  Configuration: 

The system provides several ways that the application can improve its control of the system: 
 
• The application can monitor power supply information including system voltages, tem‐ 
perature and (optionally) relative humidity as well as operating states. These values 
can be acted on, or can simply be stored or recorded for diagnostic purposes. 

• One of the power supply parameters retrieved is a set of system flags that indicates 
that the system is being requested to shut down soon, but when ʺconvenientʺ. This 
would include for example, a condition where it has become too warm or too humid 
to  continue  operation.  In  this  case,  the  application  must  react  to  the  application 
shutdown request flag and be terminated in a maximum of 15 minutes. 

• Some of the system flags can be written to, allowing the application some level of 
control. For example, through the flags the application can control: 

◦ A digital output (for example to turn on a dehumidifier or pump if high humidity 
is detected) 
◦ The application can more closely control the system watchdog by servicing it 
itself. In this way, the watchdog will not depend on whether or not Linux and 
the Linux watchdog service is running, but can be made to depend instead on the 
proper functioning of the  application. 

  247 
13.8. LINUX PASSWORDS CHAPTER 13. SOFTWARE MAINTENANCE

13.8 Linux  Passwords 
The default digitizer Linux passwords are as follows: 
 

root  kmi 
kmi  kmi 

NOTE:  Kinemetrics  STRONGLY  recommends  that  you  change  the 


system passwords to something well known to your organization, but 
which is different from the Kinemetrics factory   defaults. 

To change a Linux password for the current user, type: 

root@Rock103:~# passwd 
Enter new UNIX password: 
Retype new UNIX password: 
passwd: password updated successfully 

To change the password of a Linux user account from “root”, type: 

root@Rock103:~# passwd kmi 
Enter new UNIX password: 
Retype new UNIX password: 
passwd: password updated successfully 
 
To change the Rockhound console password, change it within the RockTalk or Web Interface 
Parameter Editor. 
 
13.9 File Retrieval 
File Retrieval can be done in a number of ways: 

• By clicking the filename from the Recorded Files page of the Web Server 
• Using the TX command or Block Mode transfer operations of the Altus Emulation 
modules 

• Using the SFTP service and a program such as WinSCP, that is described in another 
section 
 
In addition however there are “push” methods available through Rockhound modules that 
can automatically send out files when they are recorded. 

  248 
CHAPTER 13. SOFTWARE MAINTENANCE 13.9. FILE RETRIEVAL

The E‐mail File Sender module allows files to be  sent via Authenticated SMTP e‐mail. 
Sending files in this way requires that you have access to an E‐mail server that will accept 
Authenticated SMTP  e‐mail messages: 
 

Figure 103: E‐Mail File Sender 

 
The other option for a file push is to use an FTP Sender module. This module allows the 
digitizer to connect to an FTP server that you provide and will push the files into that server 
when they are recorded: 

  249 
13.10. SAVE/RESTORE PARAMETERS
CHAPTER 13. SOFTWARE MAINTENANCE

Figure 104:  FTP File Sender 

 
A representative FTP Server program is provided on the Rock Support Software CD (300654‐ 
PL) provided with your digitizer. Setup of this program is as described in the FTP Server 
section. 
Note that multiple file sender modules may be added to the system to send files to multiple 
destinations as needed. 
 
13.10 Save/Restore Parameters 
Once you’ve configured Rockhound for how you intend to operate your system, you should 
save your configuration. This saved copy of the parameters will be  used to restore the 
Rockhound parameters in case they become corrupted. This is done in two ways: 

  250 
CHAPTER 13. SOFTWARE MAINTENANCE
13.10. SAVE/RESTORE PARAMETERS

• From a Linux login, use the command rhsave to save a copy of your parameters within 
the system that can be restored in case your parameters become corrupted. 

• cd  /usr/rock/SMARTSDist/bin 
• ./rhsave 

If the unit’s configuration is destroyed, the system will revert to the copy of the configuration 
most recently saved by rhsave, or to the factory default if the configuration was never saved 
using rhsave. 
 
• From the web interface use the Admin Details to download a copy of the parameters 
to your PC. Near the top of the Admin Details screen, there is a link “Download 
parameters here”, that will download the parameter from the instrument. You can 
give this file you are saving a name that is meaningful to you such as “event‐recorder‐ 
at‐10sps‐and200sps.jar”. 

• Kinemetrics recommends that you save a copy of your Rockhound parameters any time 
that you make significant changes to the parameters so that you can quickly restore 
them in the event of loss. 
 
To put a configuration that you’ve saved back on the system, use the web interface via the 
Send Parameters  File  function on the  Admin Details screen.  Select the  function  “Send 
Parameters File”, and then select the file to send and after it is sent, press “Validate/Apply 
Patch”: 

  251 
13.10. SAVE/RESTORE PARAMETERS
CHAPTER 13. SOFTWARE MAINTENANCE

Figure 105: Admin Details 

 
Your parameters can also be saved within the system using the rhsave function from a Linux 
login. Note that you can use this any time you wish to store the current configuration as a 
baseline: 

• cd  /usr/rock/SMARTSDist/bin 
• ./rhsave 

To restore parameters saved by the rhsave feature, use the command rhrestore: 

• cd  /usr/rock/SMARTSDist/bin 
• ./rhrestore 

  252 
CHAPTER 13. SOFTWARE MAINTENANCE
13.10. SAVE/RESTORE PARAMETERS

To return the system to the Rockhound configuration as it shipped from the factory, use the 
command rhfactoryrestore: 

• cd  /usr/rock/SMARTSDist/bin 
• ./rhfactoryrestore 

If the unit’s configuration is destroyed, the system will revert to the copy of the configuration 
most recently saved by rhsave, or to the factory default if the configuration was never saved 
using rhsave. 
The system will also automatically make copies of the last five (5) configurations so that you 
can go back to these configurations at any time. Note that in this case, it is not necessary 
for you to do anything proactive to save the configurations in advance, as the copies are save 
automatically for  you. 
The saved configuration files are time stamped with the GMT creation time, as: 

config_YYMMDD_HHMMSS.jar. 

To restore a saved configuration: 

• cd  /usr/rock/SMARTSDist/save 
• ls 
 
config_20091027_145129.jar config_20091029_161551.jar 
config_20091029_161636.jar config_20091029_161712.jar config_save.list 
factory.jar    user.jar 
 
• cp  config_20091027_145129.jar  /usr/rock/SMARTSDist/injar/config.jar 
• rock restart 
 
You can also restore to a user save point, factory default configuration, or an automatic 
configuration save‐point from the Rock Recovery page: 

  253 
13.11. SOFTWARE UPDATES CHAPTER 13. SOFTWARE MAINTENANCE

Figure 106: Rock Recovery 

 
13.11 Software Updates 
Software updates can be installed remotely through the web interface. The software updates 
will be either Rockhound updates or updates to the underlying Linux Operating System. 
To install a Rockhound, Linux, or Tomcat (web server) update, enable the Admin Details 
screen. Select Send Firmware Update and choose the file to be sent. 
After the file transfer is complete, press “Validate/Apply Patch” to install the update. The 
Obsidian will automatically determine what kind of update is being installed and will install 
it the correct way. 
This may take a few minutes until the system is restarted and operational with your update. 

  254 
CHAPTER 13. SOFTWARE MAINTENANCE 13.12. LOG FILES

NOTE: If you wish, you may also directly use SFTP rather than the  
web interface to install   updates. 

To install a Rockhound update through SFTP: 

• Connect to the unit using a tool like WinSCP 
• Place the Rockhound update in the /usr/rock/SMARTSDist/injar directory 
• Rename the update file to update.jar 
• Using the Administrative Details page, select Restart FW to install the update 

To install a Linux or Tomcat update through  SFTP: 

• Connect to the unit using a tool like WinSCP 
• Place the Linux update in the directory /opt/kmi/updates/pending 
• Using the Administrative Details page, select Apply OS Patch to install the update 
 
13.12 Log  files 
To help you maintain your system, the Rockhound software keeps log files of most important 
system actions, messages, and exceptions. These are kept in two different log files, the 
Runtime‐log, which keeps track of normal system messages and reporting, and the Error‐log, 
which keeps track of error messages and conditions. Both logs are trimmed automatically 
to keep their size down. 
The log files can be useful as a diagnostic aid with the assistance of Kinemetrics where 
needed. 
The Runtime‐log can be selected from the Files menu as shown: 

  255 
13.12. LOG FILES CHAPTER 13. SOFTWARE MAINTENANCE

Figure 107: Run Time Log 

 The Error log can be selected from the Files menu as shown: 

  256 
CHAPTER 13. SOFTWARE MAINTENANCE 13.13. STATE OF HEALTH

Figure 108: Error Log 

 
Time stamps within the log files are relative to system time (UTC). 
 
13.13 State of  Health 
The digitizer tracks several State of Health (SOH) parameters that can be used to monitor 
system environmental values such as temperature and humidity as well as system voltages 
and currents, GPS drift, clock phase and clock quality. 
These values are enabled in the parameters for the Rock2 Front End: 

  257 
13.13. STATE OF HEALTH CHAPTER 13. SOFTWARE MAINTENANCE

Figure 109: SOH Selection 

 
Other values are available in the advanced and test mode module parameters. 
Scaling of SOH values: 

  258 
CHAPTER 13. SOFTWARE MAINTENANCE 13.13. STATE OF HEALTH

SOH Name  Default  Scale  Description 


Temperature  true  X 10  Record temperature stream as .deg 
Humidity  true  X 10  Record humidity stream as .hum 
VofV12ExtDC  true  X 1000  Record external DC stream as .vvx 
VofV12Batt  true  X 1000  Record battery input stream as .vvb 
VofV12POE  false  X 1000  Record power over ethernet stream as .vpe 
VofSysPwr  true  X 1000  Record system voltage stream as .vep 
IofSysPwr  true  X 1000  Record system current stream as .vec 
VofIofBC  false  X 1000  Record battery charge current stream as .vbc 
VofIofBBIn  false  X 1000  Record sensor/comms current stream as .vbb 
VofSnsrPwrSw  false  X 1000  Record sensor power stream as .vsp 
VofTimeIOPwr1  false  X 1000  Record IO power 1 stream as .vp1 
VofTimeIOPwr2  false  X 1000  Record IO power 1 stream as .vp2 
VofTimeIOPwr3  false  X 1000  Record IO power 1 stream as .vp3 
TimeQual  true  X 1  Record time quality stream as .lcq 
PTP TimeQual  false  X 1  Record PTP master quality stream as  .lcp 
TCXO Drift  true  X 1  Record clock phase stream as .lce (ns) 
Load Average  true  X 100  Record CPU Load Average stream as .cpu (Note 1) 
Percent Used  false  X 100  Record Percent of Data Storage Used stream as .dsk 
Mass Position 1  false  X 1000  Record mass position stream as .um#1 
Mass Position 2  false  X 1000  Record mass position stream as .um#2 
Mass Position 3  false  X 1000  Record mass position stream as .um#3 
RAM Used  false  X 1  Record RAM Used stream as .mem 
TCXO DAC  false  X 1  Record TCXO DAC stream as  .vco 
User Stat  false  X 1  Record user status stream as .usr (Note 2) 
1: Values < ~200 indicate sufficient idle time 
2: See the ROCK+ Console command ʺuserstatʺ 
And mostly diagnostic... 
SOH Name  Default  Scale  Description 
ADCSkew  false  1  Record ADC skew stream as .sk# (internal timer  counts) 
SOHVolts1‐10  false  X 1000  Record SOH voltage stream as .v#1‐.v#a 
(# is which DSP board the data comes from) 
SOH Name Default Scale Description false X 1 false X 1000 In addition to the ability to 
record and display these State of Health values as normal 1sps data streams, the values 
are also saved in Short Term (last 24 hours) and Long Term (last year) history. These SOH 
histories can be displayed by selecting State‐of‐Health from the Status menu, and then 
selecting the requested displays: 

  259 
13.13. STATE OF HEALTH CHAPTER 13. SOFTWARE MAINTENANCE

Figure 110: SOH Display Selection 

 
Once you press OK, the history displays must be updated (this may take a few minutes), 
but then the displays will show as follows: 

  260 
CHAPTER 13. SOFTWARE MAINTENANCE 13.13. STATE OF HEALTH

Figure 111: SOH Graphs 

 
Note that the displays include limits (left margin), description (lower right corner), and a 
time scale (bottom). The time scale is hour of day (UTC) for the short term display and 
months for the long term display. 

  261 
13.13. STATE OF HEALTH CHAPTER 13. SOFTWARE MAINTENANCE

  262 
Chapter 14 

Software  Tools 

The Rock+ digitizers are designed to work with standard software and hardware tools wher‐ 
ever possible. In many cases, those tools are provided along with most computers or are 
easily available for download. In some cases tools are required that may not come standard 
on every computer. In these cases, Kinemetrics has provided publicly available tools on the 
Rock Support Software CD (300654‐PL). The tools provided are one’s that Kinemetrics has 
found to work well, although you are of course free to use other similar tools that you prefer. 
The Rock Support Software CD programs are specific to the Windows Operating Systems. If 
you are running Linux or MacOS, suitable programs are easily available for those Operating 
Systems. 
For example: For Linux: 

• For Telnet:  Use telnet 
• For Serial Terminal Emulation: Use minicom 
• For Secure Terminal connections: Use ssh 
• For Secure File Transfers: Use sftp 
If you do not have other tools such as terminal emulation programs, telnet client, and secure 
file transfer programs that you know you want to use, and you are using Windows, we suggest 
that you install the software from the Rock Support Software CD (300654‐PL) now. 
 
14.1 Web Browser 
As indicated, the digitizer is intended to operate with little more than a web browser. Most 
modern web browsers are sufficient, and need meet only a few requirements to operate well 
with the digitizer: 

• Support HTML 1.1 or later 
• Support Java and Java Applets 

  263 
14.2. JAVA CHAPTER 14. SOFTWARE TOOLS

14.2 Java 
The web interface of the digitizer will function without installing Java, but many of the most 
useful and important feature of the interface require Java. Kinemetrics recommends that 
your computer have Java installed and that Java support be enabled in your web browser. 
Java Virtual Machines are available from several suppliers and are usually free. 
One such program is available from java.sun.com. On the Sun website it is called the Java 
SE JRE. 
 
14.3 Terminal Program 
In order to communicate with the digitizer through a serial port (such as communicating via 
the RS‐232 Linux console) you will need a terminal program that can communicate through 
RS‐232. Historically this was done on Windows based computers using HyperTerminal. 
However, in Windows 7 HyperTerminal is no longer included. 
You can download and purchase HyperTerminal from www.hilgraeve.com, or you can use 
the free program PuTTY, which is provided on the Rock Support Software CD. PuTTY is 
described below. 
 
14.4 Telnet Client 
To communicate with the digitizer through its command console you will need a telnet client 
program. Most computers include telnet as part of the normal distribution, but Windows 7 
no longer includes telnet. 
As a replacement you can use the free program PuTTY, which is provided on the Rock 
Support Software CD. PuTTY is described below. 
 
14.5 WinSCP 
WinSCP is a free program that allows you to make secure connections for file uploads and 
downloads. WinSCP is included on the Rock Support Software CD. 
Since WinSCP is not Kinemetrics software and is provided only as an example, we make no 
attempt to thoroughly describe all capabilities.  However we will describe basic operation 
here. When you start WinSCP you will see a dialog that looks something like this: 

  264 
CHAPTER 14. SOFTWARE TOOLS 14.5. WINSCP

Figure 112: WinSCP 

 
To connect to a digitizer, enter the IP address of the digitizer as “Host name” as well as 
the User name and Password as shown here: Press “Login” and you will initially see the 
following confirmation: 
 

Figure 113: WinSCP Key Confirmation 

 
This message indicates that you are about to make a secure connection to the digitizer and 
is a warning that you have to be sure you’re talking to the correct unit. Normally, once you 
accept the key you will not have to do this again when connecting to the same digitizer from 
the same  computer. 
Once WinSCP has logged into the digitizer, you will see something that looks like this: 

  265 
14.5. WINSCP CHAPTER 14. SOFTWARE TOOLS

Figure 114: WinSCP Panes 

 
The left pane of the program represents your computer and the right pane represents the 
digitizer. You can navigate the panes independently and can then upload or download files 
by simply dragging the files from one side to the other.  You can also drag files directly to 
or from other applications such as your desktop or Windows Explorer. 
If you have a set of directories that you access often, such as downloading recorded files from 
the digitizer’s /data/events directory to a working folder on your computer, you can set the 
folders correctly and then select Session → Save Session. 
Later when you restart WinSCP, you can choose the session from the list and click “Login”. 
This will save you from having to manually enter the IP address, username, or passwords 
each time.  In addition, the panes for your computer and the digitizer’s directories will be 
set automatically. 

  266 
CHAPTER 14. SOFTWARE TOOLS 14.6. PUTTY

14.6 PuTTY 
PuTTY is a secure terminal program that you can use to open a secure Linux console to 
your digitizer. PuTTY uses the SSH layer to make a secure connection. 
Since PuTTY is not Kinemetrics software and is provided only as an example, we make no 
attempt to thoroughly describe all capabilities.  However we will describe basic operation 
here. When you start PuTTY you will see a dialog that looks something like this: 
 

Figure 115:  PuTTY 

 
To log into the digitizer, enter the IP address of the unit and press “Open”: 

  267 
14.6. PUTTY CHAPTER 14. SOFTWARE TOOLS

Figure 116:  PuTTY IP Address 

 
 
Next, you will see a secure console window as follows: 

  268 
CHAPTER 14. SOFTWARE TOOLS 14.6. PUTTY

Figure 117: PuTTY Window 

 
As with WinSCP you can save sessions to avoid typing in the future by entering the IP 
address at “Host name”, the name you assigned to the digitizer under “Saved Sessions”, and 
then press “Save”. 
Please also note that PuTTY can be used as a telnet client, so you can connect to the Com‐ 
mand Console by entering the IP address under “Host name”, selecting “Telnet”, entering 
the “port” as 9900 and clicking “Open”. 
As with WinSCP and the PuTTY SSH connection, you can save this session by entering a 
name under “Saved Sessions” and clicking “Save”. 
PuTTY can also be used as a simple terminal program to open a serial console to your 
digitizer (for example, to connect to the Linux console). Select “Serial” and specify the 
Serial line (e.g.: COM2) and the baud rate. 

NOTE:  On the Rock+ the Linux serial console is at 115200 baud. 

You can save your serial session by selecting “Serial”, entering the COM port and baud rate, 
entering the name as a “Saved Session” and clicking “Save”. 

  269 
14.7. FTP SERVER CHAPTER 14. SOFTWARE TOOLS

14.7 FTP Server 
If you wish to transmit files from your digitizer automatically, one way to do this is to 
send them to an FTP server that receives the files. There are several public domain and 
commercial FTP server programs that you can choose from, but one such program is the 
WAR‐FTPD. 
To install WAR‐FTP, place the war‐ftpd.exe executable in an appropriate directory (Program 
Files\FTP is suggested) and start the program: 
 

Figure 118:  FTP Server 

 
Select “no anonymous logins”. 
Select from the menu Properties → Security → Edit Default: 

  270 
CHAPTER 14. SOFTWARE TOOLS 14.7. FTP SERVER

Figure 119: FTP Server Defaults 

 
If present, select and DELETE the path for the “\” directory. Click OK. Select from the 
menu Properties → Security → Edit User: 
 

Figure 120: FTP Server Users 

 Choose “Add” to add a user. Add an “admin” user: 

  271 
14.7. FTP SERVER CHAPTER 14. SOFTWARE TOOLS

Figure 121: Adding Admin User 

 
Specify and confirm the password. The default is “kmi” (but you can use what you want): 
 

Figure 122: Admin User Password 

 
Select the File Access tab: 
 

Figure 123: Admin File Access 

 
Press the “Add” button on the File Access tab under “Path”, and specify the directory where 
files will download to on your PC, such as “C:\data”, then press “OK”: 

  272 
CHAPTER 14. SOFTWARE TOOLS 14.7. FTP SERVER

Figure 124: Add Data Path 

 
Set the permissions on the directory as follows or as appropriate, then press “Apply”, then 
“OK”: 
 

Figure 125: Permissions Set 

 
Another way to improve security a bit is to limit permissions (for example, eliminating 
List, Write, and Delete permissions via FTP), or to run the FTP server from a user with 
limited access rights that only has access to those directories needed for the task, such as 
downloading to the C:\data directory on your PC. 
You may choose to add other users with different permissions or access to different directories. 
Select Properties → Options and select the General tab. Set startup options as follows: 

  273 
14.7. FTP SERVER CHAPTER 14. SOFTWARE TOOLS

Figure 126: FTP Server Options 

 
Click OK. 
You may start and stop the FTP server manually from its menu with the selection Properties 
→ Start Service or Stop Service. 
Configure the FTP Sender module in your digitizer’s layout to match the IP address of the 
FTP server, the FTP user name and password (using the Tools → Operation/Passwords 
from the menu) as follows: 

  274 
CHAPTER 14. SOFTWARE TOOLS 14.7. FTP SERVER

Figure 127: FTP Sender Parameters 

  275 
14.7. FTP SERVER CHAPTER 14. SOFTWARE TOOLS

  276 
Chapter 15 

Troubleshooting 

15.1 Modem not  Communicating 


The most common causes of modem communication problems are cables and improper setup. 
Regarding setup, it is most likely that you have not set up the modem connection to use the 
TTYMonitor so you will see the following message in the log files: 
ERROR! Modem enabled as primary connection
To correct this problem, see the section Modem Only Operation . 
 
15.2 Private Networks 
If your system will operate on a private network without an outside connection, DNS should 
be disabled. Current versions of the netconfig script remove the DNS entries if you do not 
specify  them.  However,  if  you  have  an  older  version  of  the  netconfig  script,  it  may be 
necessary to do this manually by deleting /etc/resolv.conf. 
 
15.3 Altus Emulator and  SDS 
Altus Emulation SDS will only steam the highest sample rate, so if you mix SDS data and 
SOH channels (which are at 1sps), you will not see the SOH channels on SDS. You must 
either be aware of this and deal with the fact that you’ll be missing some channels or specify 
that only the actual data channels are to be sent. 
 
   

  277 
15.4 Deleting a large Linux directory 
If a configuration error inadvertently results in the creation of lots of useless files – such 
as when you have inadvertently set the trigger levels too low, you may find that Linux has 
problems dealing with the large directories. In this case you may find that trying even to 
delete the files fails with the message “too many files”. One way to correct the problem is 
with a simple script: 

cd /tmp/kmi 
for I in ‘ls‘; do echo $I; rm  $I;   done 
 
15.5 Admin cannot delete data files 
If you are unable to data delete files when logged in as the “admin” user, it is probably 
because permissions are not set correctly for the data directory and any subdirectories. This 
applies to login via SSH (PuTTY) or SCP  (WinSCP). 
One temporary solution is to log in as root. While this will work and is generally safe over 
a secure connection, it is best to avoid logging in as root whenever there is an alternative 
that will do the job. 
If you are storing your data in an alternate location, such as a secondary memory card or 
an external storage device, then you will have to set the permissions properly for the parent 
data directory as well as for any subdirectories that already exist. Once the changes are 
made, further subdirectories created by the system will be correct. 
In this example we will use the name ʺMyDataʺ as the desired data directory name. 

• login as root. 
• cd to where you want the new directory created. 
• mkdir MyData 
• chmod 775 MyData 
• chmod g+s MyData 
• chgrp ‐R admin MyData 

  278 
CHAPTER 15. TROUBLESH1O5O
.7T
. IN
IPGISSUES. FIREWALLS AND ROUTER BLOCKS

NOTE: If you are trying to fix an existing directory skip the process of 
making the directory (mkdir MyData). Additionally you will need to do 
a  ʺchmod  g+sʺ  on  any  pre‐existing  sub‐directories  under  MyData. 
 
Once done, you should be able to access and manage your files when logged in as the “admin” 
user, and be able to access (read only) your files when logged in as the “client” user. 
 
15.6 IP issues.  Firewalls and router    blocks 
If you’re having problems with IP communication, here are some basic troubleshooting tips: 
If the digitizer and computer are connected through the internet, make sure that both units 
can communicate through the internet. The easiest way is to ping a third party server such 
as Google. Most common mistakes are specifying the wrong gateway or wrong DNS servers. 
If the digitizer and computer are on the same private network, make sure that both the 
computer and the digitizer are on the same network (first three parts of the IP address the 
same). Make sure the two units can ping each other. 
Check each service manually that you plan to use, whether FTP (port 21), TELNET (port 
23), SSH (port 22), SMTP e‐mail (port 25), NTP (port 123) or Rockhound specific services 
such as the Rockhound Console (port 9900). 
If you have problems accessing any particular service, check to make sure that the ports are 
open on any Operating System provided or third party software firewall. If you’re going 
through a router, you will also need to check that the ports are open through the router. 
Finally, some antivirus software includes port blocking features you may have to check. 
 
15.7 Can’t Change Password 
If you are unable to change Linux passwords, you may have gotten a response such as 
this: 
Obsidian:/data/events# passwd
Enter new UNIX password:
Retype new UNIX password:
passwd: Authentication token manipulation error
passwd: password unchanged
 
The reason for this is because the majority of the filesystem is locked to provide further 
protection against accidental corruption. See Locked File System 
   

  279 
 
15.8 External  Charger 
The digitizer’s power supply can intelligently control an external charger used to maintain 
large batteries. To check if the external charger is enabled, type psdisplay from the Linux 
prompt: 
 
Rock2:/etc# psdisplay 
PSRTC=471629393 
Temp=31.28 
Humidity=9.52 
Flags=16 
ExtCharger=0 
SensorPwr=1 
BBVolts=12.502 
VofSC=4.432 
Vof1_8VCore=1.787 
VofVCC=3.309 
VofVBoost=5.118 
Vof12VPOE=1.553 V 
of12ExtDC=15.482 
VofBattCharge=0.049 
VofV12Batt=0.000 
VofSysPwr=15.443 
VofIofSysPwr=0.333 
VofIofBBIn=0.063 
VofTimeIOPwr1=0.020 
VofIofBC=0.129 
VofSnsrPwrSw=12.444 
VofTimeIOPwr2=0.049 
VofVBB=12.532 
VofTimeIOPwr3=0.000 
VofSagFilt=14.935 
PSType=1 
PSVersion=6.17 
 
In this case, the External Charger is disabled (ExtCharger=0). To enable it, type psset‐ 
params as shown: 
 
Tock2:/etc# pssetparams ExtCharger=1 
Parameter(s)  set.  Save  the  changes?  (Y/N)  >  y 

  280 
15.8. EXTERNAL CHARGER CHAPTER 15. TROUBLESHOOTING

  281 
 
 
 
 

Chapter  16 

Digitizer Maintenance 
 
16.1 Run Remote Check Tests 
Rock+ digitizers have been designed to allow the system status to be determined remotely 
via any available communication links. This prevents the need for most preventative main‐ 
tenance visits. 
The following remote checks can ensure that the unit is functional and the peripheral systems 
are operational. 
• Verify the power input to the unit is at the correct voltage (9‐28V External DC) 
• Verify the units current draw is within normal ranges 
• Verify that the temperature is within normal operating ranges 
• Verify that the humidity inside the unit is within limits and not increasing significantly. 
• Verify that the battery voltage (if present) is correct 
• Perform a battery load test to check the battery has maintained its charge capacity. 
• Review the reported internal voltages to ensure that they are all normal. 
• Verify sensor offsets are within  range 
• Perform sensor functional  tests 
• Review software and communication logs for signs of communication problems or se‐ 
curity breaches. 
• Check for firmware upgrades available on the Kinemetrics web site. 
• Perform maintenance on disk files to remove older unused files. 
These checks will indicate if a field visit is required to investigate an abnormal value or to 
change batteries, and or desiccant. 
See Power System Display 

  282 
16.2. ADVANCED SELF TEST CAPABIC
LIHTAIE
PSTER 16. DIGITIZER MAINTENANCE

16.2 Advanced Self Test Capabilities 
Rock+ Digitizers are designed to allow automated calibration and they have the capability 
of selecting different sources for the input channels to the ADC. In normal use the inputs are 
connected to the external connectors and the unit functions as a digitizer. For calibration 
and testing the inputs can be connected to a calibration bus. The calibration bus allows 
the inputs to be grounded, connected to the positive reference, connected to the negative 
reference or connected to the calibration DAC. This feature allows the digitizer performance 
to  be  tested  both  on  the  bench  and  for  initial  calibration  but  also  as  part  of  a  system 
test/diagnostic procedure in the field. For example by grounding the inputs the digitizer 
noise can be evaluated, by applying the reference voltage the span calibration can be checked, 
and using the calibration DAC the frequency response can be verified. 
These functions provide a powerful set of diagnostic tools in addition to the external calibra‐ 
tion signal for users with large numbers of units deployed these capabilities combined with 
batch files can provide a powerful maintenance tool contact Kinemetrics if you are interested 
in our services group helping you develop these capabilities. 
 
16.3 Troubleshooting &  Service 
If your unit needs repair or service, we strongly recommend that you return it to Kinemetrics. 
If you wish to diagnose hardware problems yourself, we provide some cautions below. 
The procedure for loading new firmware/software into the unit is also detailed below. 
 
16.4 Hardware Problems 
CAUTION: Only a qualified electronic technician should diagnose and 
repair  the  unit.  Be  sure  the  technician  carefully  follows  both  the  ESD 
precautions and the precautions described in the Safety    section. 
 
Do not attempt to repair the unit at the board level unless you have equipment for re‐ 
working high density lead free surface mount printed circuit boards and have gone through 
the appropriate Kinemetrics training course. 
We strongly recommend that you stop your diagnosis at the board level and send the entire 
unit back to Kinemetrics for repair. 
 
16.5 Installing New  Firmware 
There are several different software components in your system and each has its own ver‐ 
sions. The process of updating each software component will be described within the section 

  282 
CHAPTER 16. DIGITIZER MAINTENANCE 16.6. PREVENTIVE MAINTENANCE

that describes the package itself. For example, the process of updating the Power Supply 
Application will be discussed within the section that discusses the Power Supply. 
See Software Versions . 
 
16.6 Preventive Maintenance 

16.6.1 Desiccant Replacement 

To help maintain low humidity inside the unit, Kinemetrics places a packet of desiccant inside 
to absorb any water vapor. The interior humidity is measured by a sensor and reported in 
the systems diagnostics. We that recommend that you replace the desiccant package when 
the internal humidity reading rises above 50% or every 24‐36 months. (If the humidity is 
still low at 36 months and you have not opened the digitizer this period can be extended.) 
You can order new desiccant packets (P/N 700049) from Kinemetrics. 
To replace the desiccant package proceed as follows: 

• Remove all power sources from the system and wait until it has completed its shutdown 
sequence. 

• Remove the right protective bumper by gently easing it off the metal end cover. 
• Remove the seal screw from the front of the unit to allow the pressure within the unit 
to equalize. 

• Remove the right end cap by unscrewing the four visible hex head bolts. 
• The desiccant package will be visible secured to the card cage at the end of the unit 
by a reusable tie strap. 

• Loosen the tie strap and replace the desiccant. The old package can safely be disposed 
of in normal trash. 

• Replace the end cap ensuring that the O‐Ring is correctly located in the interior groove. 
• Tighten the four screws. 
• Replace the seal screw in the front of the unit. 
• Replace the protective bumper. 
• Reconnect the unit to the power sources. 

  283 
16.6. PREVENTIVE MAINTENANCE CHAPTER 16. DIGITIZER MAINTENANCE

CAUTION: Possible equipment damage. If you do not correctly re‐ place 
the end cap and the seal screw the watertight integrity of the    unit  can 
be  compromised  and  damage  can  occur  to  the  unit. 
 
 
Potential ESD equipment damage. The digitizer circuit boards 
contain  CMOS  components  that  can  be  damaged  by  electrostatic  dis‐ 
charge (ESD) if not properly handled.  Use a grounded wrist strap, with 
impedance of approximately 1 MOhm, to protect components from ESD 
damage  when  handling  circuit  boards.    Before  removing  any  circuit 
boards  or  disconnecting  any  internal  cables,  be  sure  that  all  batteries 
and the charger are disconnected. 
 
16.6.2 Replacing  Batteries 

Because they lose their capacity over time, you should replace the optional external batteries 
with the unit at regular intervals. Kinemetrics recommends that you replace the batteries 
every three years in normal operating environments; more frequently if the unit’s ambient 
temperature is significantly above 20°C. Read the Safety section before replacing the battery. 
 
16.6.2.1 Battery Specifications 

We strongly recommend that you purchase replacement batteries from Kinemetrics. The 
batteries used are 12V, 7.2, 12 and 38 amp‐hour Sealed Lead Acid cell with either Faston 
terminal tabs, 0.250ʺ x 0.032ʺ or bolt lugs, equipped with a pressure relief valve that meets 
UL924.  The  battery  capacity  will  have  been  chosen  to  provide  the  required  length  of 
autonomy given the average current draw of the system. 
 
 
WARNING:  Fire or explosion hazard.  Do not install    a  non‐ 
rechargeable battery to be charged by the unit. Only install a sealed lead 
acid  battery  with  specifications  compatible  with  those  above.  Other 
types  of  Lead  Acid  battery  should  not  be  used  as  the  tem‐  perature 
corrected charging voltage is set for SLA batteries only.  On no account 
install  an  alternative  battery  chemistry  such  as  Lithium  Ion  or  Nickel 
Metal Hydride! 
 
If you store a battery, you should still charge it every six to nine months to prevent permanent 
loss of capacity. You can float‐charge the battery at 13.5‐13.8V or cycle‐charge the battery, 
provided the current is limited to less than 2.4 amps and the voltage to less than 14.7V. 

  284 
CHAPTER 16. DIGITIZER MAINTENANCE 16.6. PREVENTIVE MAINTENANCE

When the voltage reaches 14.7V, the battery will be damaged unless you convert the cycle 
charging to float charging. Kinemetrics ships batteries fully charged; make sure a battery is 
still fully charged before installing it. 
 
16.6.2.2 Battery Installation 

Follow the instructions below to install a new external battery. 
 
 
WARNING:  Burn or explosion hazard.  Never  place metallic 
objects  (such  as  a  screwdriver  or  your  wristwatch  strap)  across  the 
terminals  of  a  battery.  The  metal  terminals  can  get  very  hot.  Handle 
batteries with care, and do not drop them or attempt to take them apart. 
Recycle  used  batteries,  or  dispose  of  them  in  accordance  with  local 
regulations.  Do not throw used batteries onto a fire. 

 
CAUTION:  Before  installing  the  new  battery  make  sure  it  is  fully 
charged. If the battery is uncharged, the unit will charge it, but this can 
take some time — and if AC power is lost, the unit’s power autonomy 
will be  reduced. 
 
To replace an external battery: 

• Disconnect the power source from the unit and the battery 
• With the power cable disconnected from both the unit and the PSA remove the old 
battery. 

• Connect the negative terminal of the new battery to the black wire. 
• Connect the positive terminal of the new battery to the red wire. 
• Connect the power cable to the power connector on the unit and ensure the unit turns 
on under battery power. 

• Reconnect the PSA and check that the unit switches to this power source. 
 
NOTE: An external battery should not be used with a system powered 
by a Power Over Ethernet (POE) connection. 

  285 
16.6. PREVENTIVE MAINTENANCE CHAPTER 16. DIGITIZER MAINTENANCE

16.6.2.3 Battery Recycling 
 
 
 
 

We recommend you recycle the sealed lead‐acid cells used by the digitizer; if 
properly recycled they are environmentally friendly. You should be able to recycle used 
batteries at the same centers that recycle automobile batteries. 
 
16.6.3 Replacing SD Cards 

Kinemetrics uses only qualified SD Cards for the operating system and data storage. These 
cards are rated for extremes of temperature, shock, and vibration. Using consumer specifi‐ 
cation cards will likely result in data loss and system failure. 
The SD data card installed in the power supply socket is not readily removable. It is also 
formatted with a Linux file system so it is not readily readable in a PC although it can be 
read in a PC with third party software. External USB drives are recommended for use as 
removable media and should be formatted as FAT32. 
See USB Thumb Drives . 
Replacement SD cards should be ordered from Kinemetrics. 
 
16.6.4 Cleaning the Digitizer 

Disconnect all power from the unit before cleaning it including the external battery. Then 
wipe off the exterior surfaces with a mild detergent and a damp soft cloth.  Do not use an 
abrasive cloth especially on the label area as this will damage the unit. The external battery 
can also be cleaned with a mild detergent and damp cloth following the precautions outlined 
in the battery maintenance section. 
 
CAUTION: Possible water damage. Do not loosen the seal screw or end 
caps before cleaning the unit. Do not use water to clean the inside of the 
digitizer.  Doing so will severely damage the unit! 
 
The digitizer should not normally be opened so the interior of the units should be clean. If 
dust or debris does get inside the unit, we recommend you use a small ʺcomputer vacuum 
cleanerʺ to remove this debris. Make sure you have turned the power off before vacuuming 
the unit. 

  286 
CHAPTER 16. DIGITIZER MAINTENAN1C
6.E7. DECOMMISSIONING & RECYCLING

16.7 Decommissioning &  Recycling 


 
 
 
 
 

Kinemetrics is committed to ensure the preservation of the envi‐ 
ronment for current and future generations. Thus, although not required by legislation, we 
designed the Rock+ digitizers to comply with the requirements of the RoHS (Reduction of 
Hazardous Material) directive to reduce possible contamination of landfills. The digitizer 
should be recycled and not thrown in the normal waste stream destined for the local land 
fill. 
We have made  the  product  easy  to  disassemble  and  recycle  the  constituent  parts.  The 
recycling passport describes the various materials in the digitizer and any hazards than can 
be encountered in disassembling the  unit. 
 
16.7.1 Kinemetrics Recycling  Passport 
 
Product Name  Obsidian 
Product Model Numbers  111900,  111860,112000 
Date  4/27/2010 
Version  1.1 
Contact Information  www.kinemetrics.com 

  287 
16.7. DECOMMISSIONING & RECYCLICNHGAPTER 16. DIGITIZER MAINTENANCE

Product Contains  Substance  Present  Location  Comment 

Batteries 
External  Mercury/NiCad/Lithium/Other  Option  External  SLA Battery Optional 
Internal  Mercury/NiCad/Lithium/Other  No  N/A   
Parts Containing RoHS Hazardous Substances 
  Lead  Ext. Battery  External  Note 1 
  Mercury  No  N/A   
  Cadmium  No  N/A   
  Hexavalent  Chromium  No  N/A   
  Polybrominated Biphenyls (PBB)  No  N/A   
  Polybrominated Diphenyl Ethers (PBDE)  No  N/A   
Other Potentially Hazardous Substances/Information 
  Printed Circuit Board Assembly >10cm^2  Yes  Inside  RoHS Compliant 
  Capacitors w/ PCBs or items of concern  No  N/A   
  Gas Discharge Lamps  No  N/A   
  Backlight Lamps  No  N/A   
  Backlight Lamps containing Mercury  No  N/A   
  Plastics w/ brominated Flame Retardants  No  N/A   
  LCD >100 cm^2  No  N/A   
  Asbestos  No  N/A   
  Refractory Ceramic Fibers  No  N/A   
  Radio‐active  Substances  No  N/A   
  Beryllium Oxide  No  N/A   
  Other forms of Beryllium  No  N/A   
Other Hazards 
  Parts under Pressure  No  N/A   
  Liquids  No  N/A   
  Gasses  No  N/A   
  Hidden Mechanical Springs  No  N/A   
  High Storage Density Capacitors  Yes  Internal  Note 2 
1: Covered by the Battery Directive. Should be recycled in the normal Lead Acid Battery 
Recycling Stream. 
2: Small Circuit Board Can discharge significant current do not discharge with an electrical 
short. 

  288 
CHAPTER 16. DIGITIZER MAINTENAN1C
6.E7. DECOMMISSIONING & RECYCLING

16.7.2 Disassembly  Procedure  for  Recycling 
 
Step  Description  Comments & Hints 

1  Remove all Power Sources  For Safety! 
2  Remove and recycle external Sealed Lead Acid Battery  Note 1 
3  Remove Rubber End Caps  Pull Off 
4  Remove Optional Dust Caps  Note 2 
5  Loosen Connector Nuts, Remove Nuts and Washers  Note 3 
6  Remove Seal Screw in Front Right Bottom of  Label  Note 4 
7  Remove both End Caps  Four Screws Each End 
8  Remove Deck & Base Plate  Note 5 
9  Remove small supercapacitor  assembly  Note 6 
10  Pull backwards and remove internal frame and circuit boards  Note 7 
11  Remove the Rear Mounting Frame  Screws on Side 
12  Remove PWBs  Note 8 
13  Remove Label from Enclosure  Note 9 
1: Battery can normally be recycled wherever auto batteries are recycled. 
2: Unscrew Acorn Nuts Holding Attachment Cords. 
3: Use Channel Locks if deep socket set not available. 
4: Unseals unit to allow end caps to be removed easily 
5: For units with internal deck remove the base plate by unscrewing the hex head screws 
visible on the base of the unit. Remove the accelerometer deck by undoing the two screws 
in the base. The deck can then be removed from the unit by undoing the ribbon cable. The 
deck can be disassembled into PWBs and the sensor modules. 
6: If unit has been powered in the last month carefully discharge the super capacitors using 
a 1k load resistor. Do not short them directly! 
7: Remove ADC/DSP Boards and Shields one by one till down to the rear mounting frame. 
8: Undo Spacers remove processor, front panel, and sensor boards as applicable. 
9: Insert thin blade under edge of label and pull off. 
While the Recycling Materials show the exact materials that make up the various components 
of the unit. By following these instructions a large portion of the unit can be re‐used. 

  289 
16.7. DECOMMISSIONING & RECYCLICNHGAPTER 16. DIGITIZER MAINTENANCE

16.7.3 Materials for Recycling 
 
Name  Material  Qty  Location  Comments 

Protective End Caps  Black EPDM  2  External  Remove First 


Connector Dust Caps  Aluminum Alloy  Varies    Hardware is S/S 
End Caps  Powder Coated Aluminum Alloy 6063‐T6  2    Note 1 
Enclosure and Obsidian Base  Powder Coated Aluminum Alloy 6063‐T6  1    Note 2 
Label  Lexan Sheet  1    Note 3 
Mounting Frame Front  Aluminum Alloy  1    Note 4 
Mounting Frame Rear  Aluminum Alloy  1    Note 5 
PWBs  PWB with RoHS Compliant Components  Varies    Note 6 
Super Cap Pack  Electronic  Components  1    Note 7 
Misc Hardware  Stainless Steel  Varies    Note 8 
1: RoHS Compliant Irridite Finish 
2: RoHS Compliant Irridite Finish, 4 Stainless Steel PEM Studs 
3: Can be removed by peeling off enclosure 
4: RoHS Compliant Irridite Finish, 9 Stainless Steel PEM Studs 
5: RoHS Compliant Irridite Finish, 8 Stainless Steel PEM Nuts 
6: All Circuit Boards in the Unit 
7: Unplug and discharge with 1k Ohm resistor for one minute before disassembling 
8: Screws, Nuts, Washers. Hardware is mainly 304, 316 Stainless Steel 

  290 
 
 
 
 

Chapter 17 

Hardware Reference 

The hardware reference section contains information that may help you with more advanced 
system installations and if you wish to make your own cable systems or power supplies. 
 
17.1 System Power 
The exact current drawn by the digitizer depends on the exact options contained in your 
system, any power the system is providing to other equipment, and how heavy a computa‐ 
tional load is placed on the system. Thus, it is difficult to give definitive power systems and 
the user is advised to measure a system in operation to determine a more accurate figure. 
A standard out of the box Obsidian will consume approximately 3.4W running at 12V DC 
through the battery inputs. This is running 4 channels at 100 sps as both an event recorder 
and a continuous recorder. Including a 100Mb Ethernet connection and with the internal 
GPS receiver and antenna operating. It will also draw about 7W during the initial charging 
of the super capacitor internal UPS system. The unit will continue to draw 7W pulses to 
keep the super capacitors “topped off” but the duty cycle will fall off. Usually below 1% of 
the time. 
Power can be reduced further to about 2.6W by turning off the GPS and network interface 
and reducing the number of channels to 3 along with disabling continuous  recording. 
Power consumption from the external DC inputs will be comparable over the range of 15.5 
to 28V with the exception of the battery charging power when a VRLA battery is connected 
to the battery input. Charging a discharged battery will add approximately 13W until the 
battery becomes fully charged at which time the charge current will become negligible. 
The Power Consumption of a Rock+Digitizer can be approximated as follows: 
Each four channel Analog to Digital Converter adds approximately another 1W. Uniaxial 
dual supply sensors draw approximately 0.25W. 
This gives an estimated power consumption of approximately: 

• 4W for a 4 Channel System with external sensors 

  291 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

• 6W for an 8 Channel System with external sensors 
• 8W for a 12 Channel System with external sensors 
• 14W for a 24 Channel System with external sensors 
• 20W for a 36 Channel System with external sensors 

When you know the current consumption of your system, its actual autonomy with any 
battery can be calculated using this formula: 
T (hours) = Capacity (Ah)/ I (A) = hours of autonomy
For example a 12 Channel System with Sensors equipped with a 12Ah battery would have 
an autonomy of about 20 hours. 
Kinemetrics supplies 12V, 7.2, 12 and 38 amp‐hour Valve Regulated Lead Acid (VRLA) 
batteries with either Faston terminal tabs, 0.250ʺ x 0.032ʺ or bolt lugs, equipped with a 
pressure relief valve that meets UL924. Batteries of capacity below 6Ah should not be used 
as the system can charge them too quickly. The charging circuitry is not capable of charging 
batteries over 38 Ah in a reasonable amount of time and an external charging system should 
be used with this system. 
If you are supplying power to the system the voltage needs to be between 9‐28 VDC. If 
the unit is to charge an external VRLA battery the voltage should be between 15.5 to 28 
VDC. The supply should be capable of providing a minimum current of 1000mA at 12VDC 
to charge the internal Super Capacitors and bring up the processor. This level of current 
is required for up to 5 minutes when the unit is first turned on. If the unit is to charge an 
external VRLA battery the supply should be capable of supplying 2.0A at 15.5 – 28 VDC 
(A nominal 50W rated supply will normally be sufficient.). The power supply should be CE 
rated and rated for operation over the temperature range expected for the system. 
 
17.2 Connector Reference 
This section contains cable and connector reference material that can be used for creating 
or customizing cables for your unit. In addition, this information can be useful for electrical 
diagnosis of cabling or electrical  problems. 
The connectors can be grouped into several categories, Power, GPS Antenna, Serial Com‐ 
munications, Ethernet, Time I/O, USB Host and USB Device ports. 
 
17.2.1 Power  Connector 

Connector:  851‐07A14‐12PX50‐A7‐44,  12  Pins,  Shell  14,  Rotated  X,  Silver,  (8x1,4x1.6) 
Mating Connector: 851‐06EC14‐12SX54, 12 Sockets, Shell 14, Rotated X, Blk (KMI 852173) 

  292 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

Mating Connector: 851‐06EC14‐12SX50‐44, 12 Sockets, Shell 14, Rotated X, Silver (KMI 
853724) 
The Rock+ internal power supply is controlled by a dedicated micro controller referred to 
as the PSMCU in this document. 
Power In from 9‐28VDC source needs to be >=15.5V for battery charging. This connector 
uses pins as power is supplied to the unit, charging current is only output from the unit 
when power has been supplied through the connector. The power pins are provided with 
Reverse Polarity Protection, ESD & EMI Protection, and protection from lightning induced 
transients using Gas  Arresters. 
The battery charging circuit is designed to charge and take power from a 12V VRLA battery. 
(Observe all safety warnings and cautions!) It features Reverse Polarity Protection, ESD & 
EMI Protection, and protection from lightning induced transients using Gas Arresters. 
Under voltage and over voltage disconnect is provided. 8 and 30V for external DC and 8 
and 16V for battery input. 
This connector provides one user input and one user output that can be used for several 
purposes as determined by the software running on the unit: 

• The input which is read by the PSMCU and is reserved for future enhancements to 
the Rock+ power supply. 
• The output is used to control external high power battery chargers to boost the initial 
charging of larger external batteries. 

The digital inputs and outputs are isolated from the system and have both ESD and EMI 
protection. They are not protected against lightning induced transients. 
 
17.2.1.1     Power Connector  Description 
 
Pin  Name  Description  Protection  Comments 
A  DIGIN_1+  General Purpose Digital Input 1+  ISO/ESD/EMI  General Purpose Input 
B  DIGIN_1‐  General Purpose Digital Input 1‐  ISO/ESD/EMI  General Purpose Input 

C  PSTXD  PSMCU serial interface  ESD/EMI  Reserved for factory  use 

D  PSRXD  PSMCU serial interface  ESD/EMI  Reserved for factory  use 

E  PSCOM  PSMCU serial interface  ESD/EMI  Reserved for factory  use 

F  PGP  Protective  Ground  N/A  Protective Ground Connection 

G  DIGOUT_1+  General Purpose Digital Output 1+  ISO/ESD/EMI  General Purpose Output 

H  DIGOUT_1‐  General Purpose Digital Output 1‐  ISO/ESD/EMI  General Purpose Output 

J  FPPWRIN  9‐28VDC In to System  RP/LIT/ESD/EMI  System Power Input 

K  FPBATPWR  Power  to External Battery  RP/LIT/ESD/EMI  Charge  Output/Battery Input 

L  FPBATRTN  Power Return from Battery  RP/LIT/ESD/EMI  Battery Return 

M  FPPWRRTN  Power Return  RP/LIT/ESD/EMI  System Power Return 

  293 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Figure 128: Power Connector Pins and Schematic 

 
17.2.2 Internal  GPS 

The internal GPS system is built into every Rock+ Digitizer and provides the very accurate 
timing  system  used  to  time  align  the  data  typically  to  sub‐microsecond  accuracy.  The 
system requires an active antenna that connects to the TNC connector on the front of the 
unit. The sections below discuss various cabling choices for the antenna and the use of an 
optional  lightning  protector.  A  TNC  to  BNC  adapter  is  provided  for  use  with  current 
antenna choices. 
 
17.2.2.1 GPS Antenna Connection Cabling 

Plenum cable is typically used in buildings and other structures where, in case of fire, the 
cable will not produce poisonous gasses that will affect the occupants. Plenum refers to the 
air conditioning ducting, which is usually the easiest place to run new cable in an existing 
structure. This cable has good UV and weather resistance and fair water resistance but has 
higher loss and is less flexible than our UV resistant cable. KMI plenum cables are typically 
less expensive than our UV‐resistant cables. 
UV‐resistant cable does not require conduit and is very flexible and easy to handle in both 
indoor and outdoor applications.  It also has low loss and good transmission characteristics.  It 
is typically used in temporary and permanent field installations, away from buildings or other 

  294 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

structures, where ease of installation is important. It may also be used in non‐plenum areas 
of buildings, as permitted by local building codes. KMI UV‐resistant GPS cables generally 
perform better electrically than our plenum cables, and so would be a better choice for longer 
cable runs or installation sites with a marginal GPS signal reception. 
For a given type of GPS antenna cable, a shorter length of cable is preferable to a longer 
length,  due  to  signal  attenuation  by  the  cable.  However,  too  short  of  a  cable  can  be  a 
problem with active antennas, as their LNA output can overdrive the GPS module’s input. 
In general, lengths of 25M to 50M of cable as supplied by KMI are acceptable. Shorter or 
longer lengths should be discussed with the factory. Lengths between the standard 25M and 
50M cables are available as a special order with correspondingly longer lead times. 
 
17.2.2.2 Installing Optional GPS Lightning Protection 

Neither the MiniMag antenna nor the bullet antenna has built‐in lightning protection, due 
to the high cost of protective devices that can pass the low‐level, high‐frequency signals used 
in GPS systems. 
The optional lightning protector device (P/N 109457‐PL) can be used with internal and 
external GPS timing systems. The lightning protector is a small 3 ½” x 1 ½” x 1 ½” box 
which contains surge‐protection devices to shunt surges from lightning strikes to the ground 
while still allowing high‐frequency signals and the DC power for the active antenna to pass. 
It has two female BNC connectors, and comes with a 3’ (approx. 1 meter) male‐to‐male 
BNC extension cable. 
Purchase a weatherproofing kit for the lightning protector from Kinemetrics (P/N 790076) 
if the lightning protector box will be installed outdoors or in a hostile environment. 
Connect the GPS antenna to the lightning protector as follows: 

• Plug the antenna cable into the protector connector marked Surge. 
• Take the short BNC male‐to‐male cable (supplied with the protector), and plug its 
male connector into the receptacle on the Rock+ Digitizer. 
 
Make sure to provide a good ground. If at all possible, mount the Protector to a metal 
plate approximately 12ʺ square (approx. 30.5 cm sq.), with a good low‐resistance and low‐ 
inductance ground connection. To mount it to the grounding panel, use approximately 20 
inch‐pounds (2.26 Nm) of torque on the screws. 
 
CAUTION: The lightning protector will not work without adequate
grounding.

The strap connecting the grounding panel to earth ground should be as short and heavy 
as possible (use a copper strap or braid at least 1” wide) to minimize the resistance and 
inductance of the ground system. Since a skin effect is present, use a straight strap with as 
large a surface area as possible, and keep the bends in the strap to an 8ʺ or larger radius. 

  295 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

17.2.3 Serial Connectors 

17.2.3.1 Console 

The CONSOLE connector provides a 115.2Kb 3‐wire UART interface (TX data, RX data, 
and Ground) to access Linux console function. Most Rock+ features are controlled through 
the Rockhound interface but certain functions, mainly netconfig, need to be set up before a 
network connection can be  achieved. 
Console Connector Description 
 

Pin Rock+ Name I/O Description DE-9 Pin PC Name I/O Protection

J17.2  M12_CNSRXD  Input  Receive Data  3  TD  Output  ESD/EMI 


J17.3  M12_CNSTXD  Output  Transmit Data  2  RD  Input  ESD/EMI 

J17.5  GND    Ground/Common  5  Ground    ESD/EMI 

J17.10  PGP    Protective Ground  Shell  Chassis Gnd     


J17.1,4,6,7,8,9      Tied to PGP w/ 100K resistors         
  M12 DTE      9‐pin DTE       
Connector and Cable Information 
 

Description  KMI Number 
Connector  853593 
Mating Connector  853601 
Cable  853607 
Effectively M12 DTE to DE9‐Socket DTE null modem crossover cable 
 

Figure 129: Console Connector Pins 

  296 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

Figure 130: Console Connector Schematic 

 
17.2.3.2 Serial  1 

Serial Port 1 provides a full 9‐pin DTE interface adapted to the circular M12 type connectors. 
This pin definition supports the 9‐pin subset of a full RS232 interface. The pinout supports 
the 9 pin D‐Type standard shown below.  This connector is designed for local use so only 
ESD/EMI protection is provided through the 15kV ESD rated driver, the Ferrite beads on 
the board, and the HF filter capacitors to ground. 
Serial 1 Connector Description 
 

Pin Rock+ Name I/O Description DE-9 Pin PC Name I/O Protection

J15.1  M12_SER1DCD  Input  Data  Carrier Detect  1  DCD  Input  ESD/EMI 


J15.2  M12_SER1RXD  Input  Receive Data  3  TD  Output  ESD/EMI 

J15.3  M12_SER1TXD  Output  Transmit Data  2  RD  Input  ESD/EMI 

J15.4  M12_SER1DTR  Output  Data Set Ready  6  DSR  Input  ESD/EMI 

J15.5  GND    Ground/Common  5  Ground    ESD/EMI 

J15.6  M12_SER1DSR  Input  Data Terminal Ready  4  DTR  Output  ESD/EMI 

J15.7  M12_SER1RTS  Output  Request to Send  8  CTS  Input  ESD/EMI 

J15.8  M12_SER1CTS  Input  Clear to Send  7  RTS  Output  ESD/EMI 

J15.9  M12_SER1RI  Input  Ring Indicator  9  NC  NC  ESD/EMI 

J15.10  PGP    Protective Ground  Shell  Chassis Gnd     


  M12 DTE      9‐pin DTE       
Connector and Cable Information 

  297 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Description  KMI Number 
Connector  853593 
Mating Connector  853601 
Cable  853607 
Effectively M12 DTE to DE9‐Socket DTE null modem crossover cable 
 

Figure 131: Serial Connector Pins 

 
 
 

Figure 132: Serial 1 Connector Schematic 

  298 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

17.2.3.3 Serial 2 and Serial 3 

The Serial 2 and Serial 3 connectors provide additional 5‐wire UART interfaces (TX data, 
RX data, RTS, CTS, and Ground) for additional serial devices. 
Serial 2 Connector Description 
 

Pin Rock+ Name I/O Description DE-9 Pin PC Name I/O Protection

J18.2  M12_SER2RXD  Input  Receive Data  3  TD  Output  ESD/EMI 


J18.3  M12_SER2TXD  Output  Transmit Data  2  RD  Input  ESD/EMI 

J18.5  GND    Ground/Common  5  Ground    ESD/EMI 

J18.7  M12_SER2RTS  Output  Request to Send  8  CTS  Input  ESD/EMI 

J18.8  M12_SER2CTS  Input  Clear to Send  7  RTS  Output  ESD/EMI 

J18.10  PGP    Protective Ground  Shell  Chassis Gnd     


J18.1,4,6,9      Tied to PGP w/ 100K   resistors         
  M12 DTE      9‐pin DTE       
Connector and Cable Information 
 

Description  KMI Number 
Connector  853593 
Mating Connector  853601 
Cable  853607 
Effectively M12 DTE to DE9‐Socket DTE null modem crossover cable 
 

Figure 133: Serial 2 Connector Pins 

  299 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Figure 134: Serial 2 Connector Schematic 

 
 
Serial 3 Connector Description 
 

Pin Rock+ Name I/O Description DE-9 Pin PC Name I/O Protection

J19.2  M12_SER3RXD  Input  Receive Data  3  TD  Output  ESD/EMI 


J19.3  M12_SER3TXD  Output  Transmit Data  2  RD  Input  ESD/EMI 

J19.5  GND    Ground/Common  5  Ground    ESD/EMI 

J19.7  M12_SER3RTS  Output  Request to Send  8  CTS  Input  ESD/EMI 

J19.8  M12_SER3CTS  Input  Clear to Send  7  RTS  Output  ESD/EMI 

J19.10  PGP    Protective Ground  Shell  Chassis Gnd     


J19.1,4,6,9      Tied to PGP w/ 100K   resistors         
  M12 DTE      9‐pin DTE       
Connector and Cable Information 
 

Description  KMI Number 
Connector  853593 
Mating Connector  853601 
Cable  853607 
Effectively M12 DTE to DE9‐Socket DTE null modem crossover cable 

  300 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

Figure 135: Serial 3 Connector Pins 

 
 

Figure 136: Serial 3 Connector Schematic 

 
17.2.3.4 DFS 

The DFS connector provides a fixed 115.2Kb 3‐wire UART interface (TX data, RX data, 
and Ground) to first ADCDSP board in the system for real time data streams from its DSP. 
The DFS protocol is covered in the section DFS telemetry . 
DFS Connector Description 

  301 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Pin Rock+ Name I/O Description DE-9 Pin PC Name I/O Protection

J16.2  M12_Bf_DFS_RXD  Input  Receive Data  3  TD  Output  ESD/EMI 


J16.3  M12_Bf_DFS_TXD  Output  Transmit Data  2  RD  Input  ESD/EMI 

J16.5  GND    ConsoleGround/Common  5  Ground    ESD/EMI 

J16.10  PGP    Protective Ground  Shell  Chassis Gnd     


J16.1,4,6,7,8,9      Tied to PGP w/ 100K   resistors         
  M12 DTE      9‐pin DTE       
Connector and Cable Information 
 

Description  KMI Number 
Connector  853593 
Mating Connector  853601 
Cable  853607 
Effectively M12 DTE to DE9‐Socket DTE null modem crossover cable 
 

Figure 137: DFS Connector Pins 

  302 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

Figure 138: DFS Connector Schematic 

 
17.2.3.5 PC 9‐pin  RS‐2323C 

The PC 9‐pin Comm port pinout is provided for reference. 
 

Figure 139: PC 9‐Pin Connections 

  303 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

17.2.4 Modem 
The modem connector provides an interface to standard telephone RJ‐11 jacks. 
Modem Connector Description 
 

Pin  Rock+ Name  I/O  Description  RJ‐11 plug  Name  I/O  Protectio 

J24.1  PTTip  Phone  Tip signal  4  TD  Output  ESD/EMI 


J24.2  PTRing  Phone  Ring signal  3  RD  Input  ESD/EMI 
J24.3,4  PGP (HF)    Tied to PGP through  capacitor      HF bypass  ESD/EMI 
  M12      RJ‐11       
Connector and Cable Information 
Description  KMI Number 
Connector  853595 
Mating Connector  853603 
Cable  853609 
 

 
Figure 140: Modem Connector Pins 

 
 
 

 
Figure 141:  Modem  Connector Schematic 

  304 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

17.2.5 Time I/O Connectors 

Three Time I/O connectors are provided. The functions included in the connectors  are 
described as follows: 
 
17.2.5.1 1PPS  output 

Each Time I/O connector provides a 1PPS output. The output is derived from the TCXO 
disciplined oscillator. The 1Hz pulse width is 100 ms in duration. The voltage level is RS‐
232C compliant at an amplitude of approximately +/‐ 5.5V peak to peak. The on‐time edge 
is  the  negative  going  edge  of  the  pulse.  The  operation  of  the  disciplined  oscillator  is 
described in the section that describes the Timing System . 
 
17.2.5.2 1PPS  Input 

Each Time I/O connector also accepts a 1PPS input. Also active‐low, RS‐232C level. This 
input can be used as source for an external 1PPS reference time pulse that with software 
support, can be used to discipline the internal TCXO. Timing modes are described in the 
section of the manual that describes the timing system. 
A ground reference for the 1PPS inputs and outputs is provided. 
 
17.2.5.3 Power 

Each Time I/O connector also provides a fault current protected power output from the 
switched power regulator (Settable from +12.5 to +15.0V in 0.5V steps). A separate return 
for this power output is provided. Setting the source voltage level (including turning it off) 
is done in the Hardware Parameters. Switched power outputs themselves are controlled by 
additional software such as the Rockhound RockGPIO  module. 
The switches used to control the output are set to trip off and latch at approximately 1.1 A. 
The power supply firmware will try to re‐enable a faulted power switch every five minutes 
for a maximum of 5 attempts. If after the 5th attempt, the fault has not been removed, the 
power supply will disable that output until the user corrects the problem and then disables 
and re‐enables that faulted output. 
 
17.2.5.4 Time I/O 1‐3 

Time I/O connectors 1‐3 are designated J27, J26, and J25 respectively. 
Time I/O  1‐3  Connector Description 

  305 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Pin  Rock+ Name  I/O  Description  Protection 


Jxx.1  X1PPSIn  Input  Orange  ESD/EMI 
Jxx.2  X1PPSOn  Output  Yellow  ESD/EMI 
Jxx.3  X1PPSCOMn  Ground  Green  ESD/EMI 
Jxx.4  XPWRn  Output  Red  ESD/EMI 
Jxx.5  XPWRRTNn  Ground  Black  ESD/EMI 
  M12    Pigtail   
There is a separate drain wire for the cable shield that is connected to the connector shell 
which is in turn connected to PGP. 
Where ‘n’ is 1‐3 and ‘XX’ is 27‐25 respectively. Pins 3 and 5 are just separate returns to the 
Rock+ internal digital/power ground. This is also essentially at the same potential as the 
external DC and battery power commons in the POWER connector. 
Time I/O 1‐3 Connector and Cable     Information 
 

Description  KMI Number 
Connector  853598 
Mating Connector  853606 
Cable  853612 
 

Figure 142:  Time I/O 1‐3 Connector Pins.  J25, J26, and J27 

  306 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

Figure 143: Time I/O Connector 1‐3 Schematics 

 
17.2.6 USB  Connectors 

Four M12 USB Equivalent connector interfaces are provided. One device port is provided. 
It is used, for example, to make a serial connection to a PC when an actual serial port is 
not available. Three Host port connectors are provide for connecting USB Flash drives and 
Wireless adapters, etc. The three host ports can supply +5V power for these devices. The 
power outputs are fault  protected. 
 
17.2.6.1 USB  Device 

USB Device Connector Description 

  307 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Pin  Rock+ Name  I/O  Description  Type A Pin  PC Name  I/O  Protection 

J20.1  XFPUSBOVBUS  Input  Senses presence of Host  1  VBus  Output  ESD/EMI 


J20.2  XFPUSBDN  Bidir  Data‐  2  D‐  Bidir  ESD/EMI 
J20.3  XFPUSBDP  Bidir  Data+  3  D+  Bidir  ESD/EMI 
J20.3  XFPUSBDGND  Output  Gnd/Common  4  CTS  Gnd  ESD/EMI 
J20.5  XFPUSBOID  Input  Reserved for future use         
  M12 USB Device      USB Type A       
Connector and Cable Information 
 

Description  KMI Number 
Connector  853597 
Mating Connector  853605 
Cable  853611 
 

 
Figure 144: USB Device Connector Pins 

 
 
 

 
Figure 145: USB Device Connector Schematic 

  308 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

17.2.6.2 USB Host 1, USB Host 2, USB Host    3 

The three USB Host Ports are identical from a hardware standpoint. They are used for 
devices like USB Flash drives and wireless adapters. 
USB Host Connector Description 
Host 1 is J23, Host 2 is J22, and Host 3 is J21 (Jxx. in table following represents J21..J23  
‘n’ is 1..3) 
 

Pin  Rock+ Name  I/O  Description  Type A Pin  PC Name  I/O  Protection 

Jxx.1  XFPUSBHnVBUS  Input  Provides power to device  1  VBus  Output  ESD/EMI 


Jxx.2  XFPUSBHnN  Bidir  Data‐  2  D‐  Bidir  ESD/EMI 
Jxx.3  XFPUSBHnP  Bidir  Data+  3  D+  Bidir  ESD/EMI 
Jxx.3  XFPUSBHnGND  Output  Gnd/Common  4  CTS  Gnd  ESD/EMI 
  M12 USB Host      USB Type A       
Connector and Cable Information 
 

Description  KMI Number 
Connector  853596 
Mating Connector  853604 
Cable  853610 
 

 
Figure 146: USB Host Connector Pins 

  309 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Figure 147: USB Host Connector Schematics 

 
17.2.7 External Three Channel Sensor Interface 

Connector: 851‐07B16‐26S54‐A7, Shell 16, Socket 26, Blk 
Mating Connector: 851‐06EC16‐26P54, Shell 16, Pin 26, Blk (KMI 852176) 

  310 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

Pin  Name  Protection  Description 


A  CH1+  LIT/ESD/EMI  Chan 1 Positive Input 
B  CH1‐  LIT/ESD/EMI  Chan 1 Negative Input 
C  CH1 GND  LIT/ESD/EMI  Chan 1 Ground & Shield (ISO AGND) 
D  CH2+  LIT/ESD/EMI  Chan 2 Positive Input 
E  CH2‐  LIT/ESD/EMI  Chan 2 Negative Input 
F  CH2 GND  LIT/ESD/EMI  Chan 2 Ground & Shield (ISO AGND) 
G  CH3+  LIT/ESD/EMI  Chan 3 Positive Input 
H  CH3‐  LIT/ESD/EMI  Chan 3 Negative Input 
J  CH3 GND  LIT/ESD/EMI  Chan 3 Ground & Shield (ISO AGND) 
K  MPOS1  LIT/ESD/EMI  Mass Position Input 1 
L  MPOS2  LIT/ESD/EMI  Mass Position Input 2 
M  MPOS3  LIT/ESD/EMI  Mass Position Input 3 
N  MPOS RTN  LIT/ESD/EMI  Mass Position Return (ISO AGND) 
P  CALENB  LIT/ESD/EMI  Calibration Enable (Sensor Control bit 4) 
R  SENS 1  LIT/ESD/EMI  Sensor Control bit 1 
S  SENS 2  LIT/ESD/EMI  Sensor Control bit 2 
T  SENS 3  LIT/ESD/EMI  Sensor Control bit 3 
U  CNTRL RTN  LIT/ESD/EMI  Control returns (ISO AGND) 
V  CALSIG  LIT/ESD/EMI  Calibration Signal 
W  CAL RTN  LIT/ESD/EMI  Calibration Signal Return 
X  +3.3VOWS  LIT/ESD/EMI  One Wire Sensor Device Power 
Y  OWSD1  LIT/ESD/EMI  One Wire Sensor Data channel 1 
Z  OWSRTN  LIT/ESD/EMI  One Wire Sensor Device return (ISO AGND) 
a  PGP  N/A  Case  Ground Connections 
b  SEN PWR  LIT/ESD/EMI  Power to Sensor 
c  SPWR RET  LIT/ESD/EMI  Power Return from Sensor 
Comments: 
This pin definition is identical to the Q330 ensuring compatibility with both cables and with 
EpiSensors and Broadband units. 
This also works with the Terminal Board Option. 
The power supplied to the sensor is regulated by the Digitizer to the specified voltage. 
 
17.2.8 External 4th Channel Single Input Sensor  Connector 

Connector: 851‐0BP14‐15SX54‐A7, 15 Sockets, Shell 14, Blk 
Mating Connector: 851‐06EC14‐15PX54, 15 Pins, Shell 14, Blk 

  311 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Pin  Name  Protection  Description 


A  CH4+  LIT/ESD/EMI  Channel 4 Positive Input 
B  CH4‐  LIT/ESD/EMI  Channel 4 Negative Input 
C  CAL4_RTN  LIT/ESD/EMI  Separate Calibration Return Channel 4 
D  SC4  LIT/ESD/EMI  Sensor Control 4 (Calibration Enable) 
E  OWSD4  LIT/ESD/EMI  One Wire Sensor Data Channel 4 
F  OWSD5  LIT/ESD/EMI  One Wire Sensor Data Channel 5 
G  SPWR RET  LIT/ESD/EMI  Power Return from Sensor 
H  SEN4 PWR  LIT/ESD/EMI  Power to Sensor 
J  +3.3VOWS  LIT/ESD/EMI  One Wire Sensor Device Power 
K  OWSD2  LIT/ESD/EMI  One Wire Sensor Data Channel 2 
L  CAL4  LIT/ESD/EMI  Calibration Signal Channel 4 
M  CNTRL RTN  LIT/ESD/EMI  Sensor Control Return and One Wire Return (ISO AGND) 
N  CH4 GND  LIT/ESD/EMI  Channel 4 Ground & Shield (ISO AGND) 
P  PGP  LIT/ESD/EMI  Case and Shield Ground 
R  OWSD3  LIT/ESD/EMI  One Wire Sensor Data Channel 3 
Comments: 
This connector serves multiple purposes: 

1. Interfacing to a passive seismometer ‐ with or without calibration and 1W metadata. 

2. Interfacing with a single channel active sensor with a single power supply, calibration, 
calibration control, and 1W metadata. 

3. Used with the three channel board to enable the 4 channel terminal block connection 
for the Rock+ for use with uniaxial sensors. 

4. Used with the other 4th Channel Boards in a 12 Channel group to connect a 4th 
Triaxial sensor with or without a power adapter. 
 
17.2.9 Four Channel Terminal Board Dual Supply (KMI 111915‐ 
01) 
This board is designed to allow easy field connections of sensors requiring dual regulated 
+/‐12V supplies. 
10 Pin Connections for Dual +/‐12V Sensors (‐01 Version) 

  312 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

Pin  Name  Description 


1  CH+  Channel X Positive Input 
2  CH‐  Channel X Negative Input 
3  CHGND  Channel X Ground & Shield (ISO AGND) 
4  CAL  Calibration Signal 
5  CALEN  Calibration Enable 
6  AGND  Connects to ADC Isolated Ground ‐ Common 
7  AGND  Connects to ADC Isolated Ground ‐ Common 
8  +12VA  Positive 12V Supply to Sensors 
9  ‐12VA  Negative 12V Supply to Sensors 
10  PGP  Case and Shield Ground 
Comments: 
This  terminal  strip  is  a  set  of  four  to  provide  connections  to  four  uniaxial  sensors  per 
ADC/DSP board.  This board variant contains a DC to DC converter to supply power   to 
+/‐12VDC  sensors.  The  one‐wire  LAN  connection  is  not  supported  in  this  variant. The 
terminal block plug into a receptacle to allow both easy wiring and also a unit to be re‐ 
placed and re‐wired easily in the field. These are Phoenix Contact 3.5mm pitch connectors. 
The terminal strip can also be used with Tri‐axial sensors and if three terminal boards are 
provided on a 12 Channel group will allow 4 Tri‐axial sensors to be connected. The board 
also boosts the Calibration Enable/Undamp signal to a +12V active, ‐12V inactive signal 
compatible with older generation FBA‐11, FBA‐23 units. The same signal also works on 
EpiSensors. 
 
17.2.10 Four Channel Terminal Board Single Supply (KMI 111915‐ 
02) 

This board is designed to allow easy field connections of sensors that can operate off a 
single supply that varies between 8‐16VDC. The power supplied to the sensor is essentially 
the power supplied to the Rock+ Digitizer. If the Kinemetrics power supply is used the 
nominal voltage will be 15.5V when AC is present and a nominal 12V from an attached 
valve regulated lead acid battery (VRLA) when AC power is lost. The power has a 500mA 
resetting polymer fuse on the line to limit the current drawn by the sensor. The Calibration 
Enable Signal is a 5V logic signal with a low of <0.4VDC and a high of >4.4  VDC. 
12 Pin Connections for Single Supply Sensors 

  313 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

Pin  Name  Description 


1  CH+  Channel X Positive Input 
2  CH‐  Channel X Negative Input 
3  CGND  Channel X Ground & Shield (ISO AGND) 
4  CAL  Calibration Signal 
5  CENB  Calibration Enable 
6  COM  Connects to ADC Isolated Ground 
7  VRTN  System Power Return 
8  VSYS  System Power to Sensor 
9  VSYS  System Power to Sensor 
10  PGP  Case and Shield Ground 
11  VOW  One Wire Sensor Device Power 
12  DOW  One Wire Sensor Data X 
Comments: 
This terminal strip is a set of four to provide connections to four single supply uni‐axial 
sensors per ADC/DSP board. The terminal block plug into a receptacle to allow both easy 
wiring and also a unit to be replaced and re‐wired easily in the field. These are Phoenix 
Contact 3.5mm pitch connectors. The terminal strip can also be used with single supply 
Tri‐axial sensors and if three terminal boards are provided on a 12 Channel group will allow 4 
Tri‐axial sensors to be connected. This board is compatible with the ES‐T, HypoSensor, and 
Shallow Borehole Units. Note: For ES‐U, and older ES‐U2 single supply units the Rock+ 
unit should be powered with a 12V+/‐5% DC supply not the standard 15.5V supply. 
 
17.2.11 Ethernet 

The Ethernet connector provides a 10/100 network interface. The X‐Code M12 connector 
used adheres to a standard used in some European industrial Ethernet equipment, such 
as, switches and routers that also use this connector. The 853608 cable maps the Rock+ X‐
Code  Ethernet connector to a standard RJ‐45 Ethernet plug. Commercial X‐Code  to X‐
Code cables may also work if appropriate network equipment is in use. 
Connection speed is automatically determined. The Rock+ Ethernet interface also supports 
auto sensing so there should be no need for crossover adapters. 
The interface also supports a Power Over Ethernet (PoE) option that will be initially offered 
at the 24 Watt power level. Standard PoE switches have been tested with the option and it 
is expected that it will work with PoE power injectors at the appropriate power level. 
PTP is a standard feature in the Rock+ Ethernet interface. 
LINK/SPEED and ACTIVITY LEDs on the Rock+ front panel indicate network traffic and 
connection speed. 
Ethernet Connector Description 

  314 
CHAPTER 17. HARDWARE REFERENCE 17.2. CONNECTOR REFERENCE

Pin  Rock+ Name  I/O  Description  RJ‐45 Pin  Name  I/O  Protection 

J28.1  MDX+0  Output  Transmit +  1  TD+  Input  ESD 


J28.2  MDX‐0  Output  Transmit ‐  2  TD‐  Input  ESD 
J28.3  MDX+1  Input  Receive +  3  RD+  Output  ESD 
J28.4  MDX‐1  Input  Receive ‐  6  RD‐  Output  ESD 
J28.5  MDX+3    Reserved for PoE  7       
J28.6  MDX‐3    Reserved for PoE  8       
J28.7  MDX+2    Reserved for PoE  5       
J28.8  MDX‐2    Reserved for PoE  4       
J28.Shell      Drain wire for cable shield         
  X‐Code Ethernet             
Connector and Cable Information 
 

Description  KMI Number 
Connector  853594 
Mating Connector  853602 
Cable  853608 
 

Figure 148: Ethernet Connector Pins 

 
 
 

Figure 149: Ethernet Connector Schematic 

  315 
17.2. CONNECTOR REFERENCE CHAPTER 17. HARDWARE REFERENCE

17.2.11.1  Ethernet Wiring 

Ethernet straight through wiring conforms to the following wiring  standards: 
 

Figure 150: Ethernet Wiring Colors 

 Ethernet cross over wiring conforms to the following wiring standards: 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 151: Ethernet Cross Over Cable Pinouts 

NOTE: The cross over cable layout is suitable for 1000Base‐T opera‐ 
tion, all 4 pairs are   crossed. 

  316 
CHAPTER 17. HARDWARE REFERENCE 17.3. CABLE DRAWINGS

Connection to RJ11 Jack Pair  1 (T1 &    R1) 
Usually the primary dial tone or talk circuit is wired to the center two pins (pins 3 & 4) 
and is the white/blue and blue/white pair (AKA: T1 & R1 ‐ tip 1 and ring 1). A standard 
single line phone draws dial tone from these center pins. 
 

Figure 152:  Ethernet RJ11 Wiring 

 
17.3     Cable Drawings 

  317 
17.3. CABLE DRAWINGS CHAPTER 17. HARDWARE REFERENCE

Figure 153:  P/N 112297‐PL, Power and Battery

  318 
CHAPTER 17. HARDWARE REFERENCE 17.3. CABLE DRAWINGS

Figure 154:  P/N 112250‐PL, Optional Power Supply 

  319 
17.3. CABLE DRAWINGS
320 

CHAPTER 17. HARDWARE REFERENCE


Figure 155: P/N 112259‐PL, Optional Power Supply with Battery Connections 

 
CHAPTER 17. HARDWARE REFERENCE 17.3. CABLE DRAWINGS

Figure 156:  P/N 853607 Rock+ Serial Cable 

  321 
17.3. CABLE DRAWINGS CHAPTER 17. HARDWARE REFERENCE

Figure 157:  P/N 853609 Rock+ Modem Cable 

  322 
CHAPTER 17. HARDWARE REFERENCE 17.3. CABLE DRAWINGS

Figure 158:  P/N 853608 Rock+ Ethernet Cable 

  323 
17.3. CABLE DRAWINGS CHAPTER 17. HARDWARE REFERENCE

Figure 159:  P/N 853611 Rock+ USB Device Cable 

  324 
CHAPTER 17. HARDWARE REFERENCE 17.3. CABLE DRAWINGS

Figure 160:  P/N 853610 Rock+ USB Host Cable 

  325 
17.3. CABLE DRAWINGS CHAPTER 17. HARDWARE REFERENCE

Figure 161:  P/N 853612 Rock+ Time I/O Pigtail 

  326 
Chapter  18 

Further Information 
 
The following documents may be useful in providing further information: 

• Rockhound User Manual – Kinemetrics P/N 304702 
 
18.1 Glossary 
ADC – Analog to Digital Converter. A device that digitizes an input voltage level and 
reports it as a single numeric value. 
DAC – Digital to Analog Converter. A device that takes a numeric value provided by a 
computer and produces an output voltage level. 
DHCP – Dynamic Host Configuration Protocol. When used, causes the device (such as a 
digitizer) to “ask” the DHCP server for an IP address, and then use the address assigned by 
the server. 
DSL – Digital Subscriber Line. A low cost “fast” internet connection. Two points to be 
aware: 

• Most inexpensive DSL services use dynamic IP addressing, so the IP address is assigned 
by the Internet Service Provider. One problem with this can mean that from “outside” 
you don’t know what the unit’s address is. This can be made easier by using Dynamic 
DNS (DDNS) or DSL with a “static” IP address, which is more expensive. 
• DSL for the most part is actually ADSL, the “A” being “asynchronous”. What this 
means is that downloads from the unit are slower than uploads “to” the unit, with 
uploads usually being about 1/6th the speed of downloads. This means that the process 
of downloading data from the unit will be significantly slower than the “advertised” 
speed. 

DNS – Domain Name Server. This facility is how internet names like www.kmi.com are 
translated into the numeric address required to locate something on the internet. 

  327 
18.1. GLOSSARY CHAPTER 18. FURTHER INFORMATION

DDNS ‐ Dynamic DNS. A third party service that equates your router with a URL (like 
www.kmi.com) so it can be “found” on the internat by name. DDNS is generally supported 
in the router (not the digitizer) and must periodically communicate with the DDNS provider. 
DDNS providers are free or paid subscription  services. 
DSP – Digital Signal Processor. A specialized processor for performing arithmetic related 
to signal processing. In the digitizer, DSPs are used for data acquisition and FIR filtering 
with one DSP used for each 4 physical channels. 
FTP – File Transfer Protocol. A method of transferring files over the internet. Generally not 
considered secure, so SCP or another secure method is preferred. FTP can be used without 
significant security risks if done carefully. 
ISP – Internet Service Provider. A company providing internet communication services such 
as DSL. 
JVM – Java Virtual Machine. Software which is loaded onto a computer allowing it to run 
Java programs. 
LAN – Local Area Network. A network of computers or devices that is private and usually 
“local” to a building or organization. 
NTP ‐ Network time protocol. Useful for setting a computer’s “time of day” but generally 
not accurate enough for  sampling. 
PC – Personal Computer. In the context of this document, it refers to any computer with a 
browser capable of running Java. It does not have to be an x86 running Windows. 
PING – A method of checking communications between two IP addresses over a network. 
PTP ‐ Precision Time Protocol (IEEE‐1588). Hardware and software that provides for sub‐
microsecond  timing  of  equipment  within  a  LAN.  Most  accurate  implementations  are 
generally hardware based. 
SCP – Secure Copy. A secure method to transfer files that uses the SSH protocol. 
SFTP ‐ Secure SFTP. Often used interchangeably with SCP. 
SOH – State of Health. Keeps track of information that can be used to help diagnose system 
problems such as voltages, current, temperature, humidity, etc. 
SSH – Secure Shell. Really an overall secure communication protocol used for terminal 
sessions, file transfers, and other activities. 
TELNET – A simple terminal session protocol used on networks. Generally not considered 
secure. 
URL ‐ Uniform Resource Locator. An internet name (like www.kmi.com) that is resolved to 
a numeric IP address by DNS. 
USB – Universal Serial Bus. This increasingly common interface is used for many interfaces 
from hard drives through keyboards, mice, and serial port adapters. 
WAN – Wide Area Network. A large scale network usually employing use of the internet. 

  328 
Index 
1PPS, 305  Decommissioning, 287 
ADC/DSP  SYSTEM, 67  Deleting a large Linux  directory, 278 
Adding Channels, 210  Desiccant, 283 
Adding Modules, 147  DFS, 301 
Admin  cannot delete data   files, 278  DFS  telemetry, 201 
Advanced Operation, 223  Dial  On  Event, 182 
Alarms, 95  Drive   Formatting, 194 
Altus  Emulator  and  SDS, 277  Dropbox, 116 
Altus Emulation, 175  Dynamic DNS, 115, 245 
Analog Inputs, 49  Earthworm, 222 
Antelope, 222  Editing  Parameters, 144 
Antenna, Phone & LAN  Cabling, 3  EpiSensor  Theory  of  Operation, 68 
Anti‐Alias Filtering, 107  EpiSensor2 Support, 87 
Basic Operation, 79  Ethernet, 314 
Basic  Setup, 81  External  Three  Channel  Sensor Interface, 
Batch  Mode, 171, 181  13 
Batteries, 284  External 4th Channel Single Input Sensor 
Battery Charging, 55  Connector, 13 
Battery Installation, 51  External  Battery, 2 
Browser, 263  External  Charger, 280 
Cable Drawings, 317  External Sensor Input  Boards, 68 
Calibration Control and Calibration    FAQs, 205 
Scripts, 50  File Management, 117 
Calibrator Output, 49  File  Retrieval, 248 
Can’t Change Password, 279  File  Transfer, 187 
Cellular, 197  File Viewer, 155 
Changing  trigger levels, 205  File Viewing, 119 
Changing Output Data  Formats, 214  Firmware Description, 71 
Civil Engineering, 20  Flash, 191 
Cleaning, 286  Four Channel Terminal Board (KMI 
Command Console, 163  111915), 13 
Common Linux Commands, 72  Free‐Field  Installations, 20 
Communication with the User Via Status   Front Panel, 7 
LEDs, 57  FTP, 115 
Connecting the Power Supply, 52  FTP Server, 115, 270 
Connecting the Sensors to the Digitizer, 30  Functional Test Sequence, 41 
Connector Reference, 292  Further Information, 327 
Console, 296  General Sensor Interfacing Considerations, 
Copyright, 3  41 
Data Time vs OS  Time, 103  Glossary, 327 
DDNS, 115, 245  GPS, 294 
  329 
GPS Internal, 104  Operating Environment, 16 
Grounding the Digitizer, 3  Operating Humidity, 16 
Hardware Identification, 65  Operating  Temperatures, 16 
Hardware Reference, 291  Optional  Power  Supply Assembly, 2 
How is the Rock+ Series Different from  Optional Equipment, 19 
the Rock   Series?, 6  Optional Tools, 18 
Humidity  Monitoring, 58  Parameter Download, 187 
Initial Setup, 79  Parametric Operation Protection, 59 
Installation  &  Setup, 13  Passwords, 93, 248 
Installing Firmware  Updates, 189  Physical vs.  Virtual Channels, 89 
Installing the Digitizer, 21  Planning your installation, 19 
Intelligent  Power  Supply, 54  Pole Zero Representation of the 
Internal Accelerometer  Deck, 13  EpiSensor, 70 
Introduction, 4  Post Event  Time, 94 
IO  Bits, 234  Power, 291 
IP issues.  Firewalls and router    blocks, 279  Power  Fail  Protection, 60 
IP Services, 242  Power System Display, 62 
Java, 264  Practice  Assembly, 19 
Layout  Wizard, 223  Pre‐Event Time, 94 
Linux  Overview, 71  Preventive Maintenance, 283 
Locked File System, 93  Private  Networks, 277 
Log  files, 255  Processor  System, 64 
Low Latency  Telemetry, 237  PTP Master (free  running), 106 
Magnetic Switch, 58  PTP Master w/  GPS, 105 
Maintenance, 5, 239, 282  PTP Master w/  NTP, 105 
Milestone  and  Point Updates, 76  PTP Slave, 105 
Minimum Run Time, 94  PuTTY, 267 
Modem, 304  Recycling, 286, 287 
Modem  Configuration, 234  Remote Check  Tests, 282 
Modem  not  Communicating, 277  Removing  Modules, 150 
Modem  Only  Operation, 183  Replacing  Modules, 151 
Mount the Unit, 22  Required  Supplies, 18 
Network  Triggering, 220  Required Equipment, 19 
Network Address, 80  Required Tools, 17 
Network  Configuration, 243  Requirements  for  Installation, 17 
Network  Parameters, 234  Reversed  Input  and  Overvoltage 
Network Planning, 20  Protection, 56 
Networking, 236  RockTalk, 187, 223 
Non‐Networked  Use, 173  Safety, 1 
NTP, 104  Save/Restore Parameters, 250 
NTP  Server, 107  SD Cards, 286 
NTP  w/  External 1Hz, 104  SDHC, 191 
Obsidian Internal Deck Zero Adjustment,  Security, 236 
86  SEEDLink, 222 
OpenVPN, 116  Self  Test, 282 
Operating  Altitude, 16  Sensitivity, 84 
 
330 
Sensor  Interface, 310  Timed Operation Windows, 59 
Sensor  Groups, 82  Timing  Options, 103 
Sensor Interfaces, 12  Timing  System, 102 
Sequenced  Shut Down, 57  Tools, 263 
Sequenced Start  Up, 56  Trigger Filters, 95 
Serial Connectors, 296  Trigger Levels, 91 
Services, 114  Triggered   Recording, 94 
Services  available  from Kinemetrics, 5  Troubleshooting, 277 
SNMP, 116  Unpacking & Inspecting the  Unit, 17 
Software  Development  Tools, 78  Updating Linux, 76 
Software  Installation, 240  Updating the Power  Supply, 64 
Software  Versions, 77  USB, 307 
STA/LTA  Triggering, 97  USB Cellular  Modem, 197 
Standard Cables, 34  USB  Device, 193, 307 
Standard  Interfaces, 8  USB  Host, 193, 309 
State  of  Health, 257  USB Subsystem, 192 
State‐of‐Health  Streams, 90  USB  Wifi, 196 
Storage, 191  Use the Proper Power   Cord, 3 
Streaming, 182  User‐Supplied Batteries or Charging 
Structural‐Monitoring  Installations, 21  System, 2 
Supercapacitor  Backup, 55  Using  Different  Triggers  or  Trigger 
Switching Between Layouts, 213  Filters, 207 
System  Information, 80  VI Editor, 74 
System  Overview, 5, 53  Voltage Ranges, 83 
System  Watchdog, 60  Voting  Options, 92 
TCP/IP Primer, 111  VPN, 116 
TCXO free  running, 106  Warranty, 4 
Telemetry   Options, 222  Watchdog, 246 
TELNET, 115  Waveform  Viewing, 121 
Telnet Client, 264  Weak  Motion  Applications, 95 
Temperature Monitoring, 58  Web  Interface  Overview, 125 
Terminal  mode, 176  Web  Interface, 81 