Sie sind auf Seite 1von 42

ACR-2000

OPERATORS’ REFERENCE
GUIDE

/80,6<6
$&5

5($'(5$1'(5$6(5

23(5$725·65()(5(1&(*8,'(

315HY

)HEUXDU\
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

)25:$5'

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Proprietary Notice and Disclaimer


The information herein disclosed is the property of Lumisys. Information in this document is subject to change
without notice and does not represent a commitment on the part of Lumisys to incorporate changes or improvements
in units previously sold or shipped. No part of this document may be reproduced or transmitted in any form,
electronic or mechanical, including photocopying and recording, for any purpose other than the purchaser’s own use
without the express written permission of Lumisys.

Copyright 1999 All rights reserved.

Trademarks
Lumisys and ACR-2000 are the trademarks of Lumisys. All other trademarks are the property of their respective
owners, and are hereby acknowledged.

For more information contact:

Lumisys, Inc.
225 Humboldt Court
Sunnyvale, CA 94089
(408) 733-6565
(408) 733-6567 FAX

i
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

7$%/(2)&217(176

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

)25:$5' ,

7$%/(2)&217(176 ,,

/,672)),*85(6 ,9

,1752'8&7,21  9

5()(5(1&('2&80(176 9

 6$)(7<,1)250$7,21 

 (1*/,6+

 &RQYHQWLRQV

 /DVHU6DIHW\ 

 $&55HDGHU/DEHOVDQG/DEHO/RFDWLRQV 

 (OHFWULFDO+D]DUGV

 )&&1RWLILFDWLRQ 

 )5$1d$,6 

 &RQYHQWLRQV

 6pFXULWp/DVHU 

 (WLTXHWWHVHWHPSODFHPHQWGHVpWLTXHWWHVGX$&5

 'DQJHUVG·RULJLQHpOHFWULTXH 

 $YLVGHOD)HGHUDO&RPPXQLFDWLRQV&RPPLVVLRQ )&& 

 ,7$/,$12 

 &RQYHQ]LRQL 

 3URWH]LRQHGDO/DVHU 

 (WLFKHWWHGHO$&5HORURXELFD]LRQH 

 3HULFRORGLVFRVVHHOHWWULFKH 

 $YYLVR)&& 

 (63$f2/ 

 7HUPLQRORJtD 

 6HJXULGDGGHOOiVHU 

 (WLTXHWDVGHO$&5\XELFDFLyQGHODVPLVPDV 

 3HOLJURVHOpFWULFRV 

 1RWLILFDFLyQGHOD&RPLVLyQ)HGHUDOGH&RPXQLFDFLRQHVGH((88 )&& 

 '(876&+ 

 'HILQLWLRQHQ 

 6LFKHUKHLWVKLQZHLVHIUGHQ/DVHU 

 $&5(WLNHWWHQXQG(WLNHWWHQSOD]LHUXQJHQ 

 (OHNWULVFKH*HIDKUHQ  

 0LWWHLOXQJGHU)&&

 $&55($'(5,167$//$7,21

 813$&.,1*,16758&7,216 

 7RROV5HTXLUHG 

 8QSDFNLQJWKH$&5 

 +$5':$5(,167$//$7,21

 7RROV5HTXLUHG 

 &RQQHFWLRQ'LVFRQQHFWLRQDQG3RZHU2Q2II6HTXHQFHV

 6HWWLQJ$&9ROWDJH 

 $&3RZHU&DEOH 

 ,167$//$7,212)+26768332573$&.$*(62)7:$5( 

ii
$&52SHUDWRU·V5HIHUHQFH*XLGH

 ,167$//$7,212)'$7$&21752/%2$5' '&% 

 ,167$//$7,212)&21752/,17(5&211(&7&$%/( 

 6<67(0'(6&5,37,21 

 237,&6

 32:(5$1'(/(&7521,&6 

 3ULQWHG&LUFXLW%RDUGV

 'DWD&RQWURO%RDUG 

 'DWD$FTXLVLWLRQ%RDUG 

 3073UH$PS%RDUG  

 /(',QGLFDWRU%RDUG  

 *DOYR&RQWURO%RDUG  

 /DVHU'ULYHU%RDUG ZLWK,QWHUORFNV  

 3RZHU'LVWULEXWLRQ

 32:(583

 23(5$7,1*7+($&55($'(5

 3/$7(+$1'/,1*$1'/2$',1* 

 5($',1*$3/$7(

 5(029$/2)3/$7(-$06$1'86,1*7+(6&$1$%257%87721 

 6<67(00$,17(1$1&( 

 7+($&5(5$6(5 

 *(1(5$/,1)250$7,21 

 6,7(5(48,5(0(176

 ,167$//$7,21 

 86$*(*8,'(/,1(6 

 23(5$7,212)7+($&55($'(5$1'(5$6(5 

 0$,17(1$1&( 

$33(1',;$ $&5/$%(/6 /$%(//2&$7,216 

$33(1',;% $&56<67(0'(6&5,37,21 

$33(1',;& $&55($'(56$03/($33/,&$7,21 

iii
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

/,672)),*85(6

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Figure 3-1 ACR-2000.........................................................................................................................17


Figure 3-2 ACR-2000 Optical Path....................................................................................................18
Figure 3-3 ACR-2000 Power and Electrical System.........................................................................19
Figure 4-1 LED Locations..................................................................................................................21
Figure 4-2 Power Switch Location.....................................................................................................22
Figure 5-1 Plate Loading ....................................................................................................................24
Figure A1-1 ACR-2000 Labels..........................................................................................................31
Figure A1-1 Etiquettes du ACR-2000 ...............................................................................................31
Figure A1-1 Etichette del ACR-2000 ................................................................................................31
Figura A1-1 Etiquetas del ACR-2000................................................................................................31
Abbildung A1-1 ACR-2000 Etiketten ...............................................................................................31
Figure A1-2 Label Locations..............................................................................................................32
Figure A1-2 Emplacement des étiquettes...........................................................................................32
Figure A1-2 Ubicazione delle etichette..............................................................................................32
Figura A1-2 Ubicación de las Etiquetas.............................................................................................32
Abbildung A1-2 Etikettenplazierung..................................................................................................32

iv
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

,1752'8&7,21

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

The ACR-2000 Reader is a precision instrument designed to scan phosphor plates and directly measure x-ray
exposure over a wide dynamic range with a high signal-to-noise ratio. This is accomplished by illuminating the
plate with a laser that is imaged onto the plate as the plate is moved perpendicularly to the laser. The emitted light is
collected, converted to an electrical signal and digitized to provide an 8- or 12- bit resolution image that can be
stored on disk, transferred to other computers or be processed, manipulated or printed. The ACR-2000 Eraser is
designed to efficiently erase exposed and read plates for reuse.
The ACR-2000 Reader has many features that can be controlled by software. Images may be digitized with 8 or 12
bit resolution, bytes may be reversed for viewing on other computer systems and custom tables can be utilized to
transform data. Further information regarding these features may be obtained from the documents listed below.
This document contains a basic technical overview of the ACR-2000 Reader and the ACR-2000 Eraser. The optics,
digital hardware and software subsystems are explained, as well as the subsystems’ functionality and general user
operation. Unpacking, hardware installation and system specifications are also included. This document is intended
for users who may need to understand the principles of operation and the interfacing requirements for the ACR-
2000 Reader and Eraser.

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

5()(5(1&('2&80(176

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

ACR-2000 Operator’s Reference Guide P/N 0070-711


LUMISCAN LSDT Digitizer / Reader Configuration Guide P/N 0071-434
ACR-2000 Service Manual P/N 0070-715
LSDT API Dynamic Link Library Reference Guide P/N 0068-105
LSDT Software Library Reference Guide P/N 0066-022
LSDT Internal SCSI Option Operators’ Supplement Manual P/N 0070-275
LUMISCAN SCSI Compatibility Protocol Specification P/N 0067-890

v
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

1
$&52SHUDWRU·V5HIHUHQFH*XLGH

1. SAFETY INFORMATION

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

1.1 English
1.1.1 Conventions

DANGER!
A DANGER! indicates that personal injury to the eyes may occur if the user
does not perform the procedure correctly.

WARNING
A WARNING indicates that personal injury may occur if the user does not
perform the procedure correctly.

CAUTION
A CAUTION indicates that damage to the product may occur if the user does
not perform the procedure correctly.

PRECAUTION
A PRECAUTION indicates that inconvenience to the user, such as loss of data,
may occur if the user does not perform the procedure correctly.

NOTE
A NOTE indicates information that should be called to the attention of the user.

Be sure to read and understand installation and operation instructions before applying power to the ACR-2000
Reader or ACR-2000 Eraser.

1.1.2 Laser Safety


The ACR-2000 Reader incorporates a Class IIIb laser diode, greater than 15 mW. The covers on the ACR-2000
Reader protect the user from direct exposure to laser light. These covers will protect a user only if they are properly
installed when the system is being used. Covers must be removed and replaced by properly trained personnel. If the
covers have been damaged during shipment or in usage, contact your local service representative for replacement
covers.

DANGER!
LASER LIGHT WHEN COVER REMOVED.
AVOID DIRECT EYE EXPOSURE.

1.1.3 ACR-2000 Reader Labels and Label Locations


The labels and label locations for the ACR-2000 Reader are shown in Figures A1-1 and A1-2 in APPENDIX A.

2
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.1.4 Electrical Hazards

WARNING
THIS EQUIPMENT IS OPERATED WITH HAZARDOUS
VOLTAGES WHICH CAN SHOCK, BURN OR CAUSE
DEATH.

CAUTION
DOUBLE-POLE / NEUTRAL FUSING.

This equipment must be serviced by persons properly trained and certified by Lumisys.
DO NOT operate the ACR-2000 Reader with a damaged power cord. Use of an extension cord is not
recommended. This equipment must be properly grounded and power connections inspected to ensure safe
operation.

1.1.5 FCC Notification


This equipment generates, uses, and can radiate radio frequency energy, and if not installed in accordance with the
installation instructions, can cause interference with radio communications.

3
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.2 Français
1.2.1 Conventions

DANGER!
Le message DANGER! signifie que tout manquement aux procédures de
rigueur peut entraîner des blessures oculaires.

AVERTISSEMENT
Un AVERTISSEMENT indique que toute utilisation inadéquate du système
peut entraîner des blessures corporelles.

MISE EN GARDE
Une MISE EN GARDE indique que toute utilisation inadéquate du système
peut endommager le système.

PRECAUTION
Une PRECAUTION indique que toute utilisation inadéquate du système peut
causer certains désagréments à l’utilisateur comme, par exemple, des pertes de
données.

REMARQUE
Une REMARQUE attire l’attention de l’utilisateur sur des informations
particulièrement importantes.

Veillez à lire et à bien comprendre les instructions relatives à l’installation et à l’utilisation du ACR-2000 avant de
le mettre sous tension.

1.2.2 Sécurité Laser


Le ACR-2000 utilise un laser diode de la classe IIIb, plus de 15 mW. Les couvercles du système protègent
l’utilisateur de toute exposition directe aux rayons à condition qu’ils soient bien installés lorsque le ACR-2000 est
en cours d'utilisation. Ils doivent être démontés et remontés par du personnel qualifié. S’ils ont été endommagés
lors du transport ou de l’utilisation du système, prenez contact avec votre distributeur agréé pour le remplacement
des couvercles.

DANGER!
VOUS VOUS EXPOSEZ AUX RAYONS LASER SI VOUS RETIREZ LE
COUVERCLE.
EVITEZ TOUT CONTACT OCULAIRE DIRECT AVEC LES
RAYONS.

1.2.3 Etiquettes et emplacement des étiquettes du ACR-2000


Les Figures A1-1 et A1-2 (dans L’APPENDIX A) ci-dessous indiquent l’emplacement des étiquettes du ACR-
2000.

4
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.2.4 Dangers d’origine électrique

AVERTISSEMENT
CE SYSTEME SOUS HAUTE TENSION. DANGER DE
DECHARGES, DE BRULURES OU DE MORT.

MISE EN GARDE
PROTECTION PAR FUSIBLE NEUTRE / BIPOLAIRE.

L’entretien et les réparations de ce matériel doivent impérativement être confiés à du personnel qualifié et agréé par
Lumisys.
N’utilisez EN AUCUN CAS le ACR-2000 si vous constatez que le cordon d’alimentation est endommagé.
L’utilisation d’une rallonge est déconseillée. Ce matériel doit être correctement relié à la terre et les connexions
électriques doivent être inspectées pour garantir un fonctionnement sûr de l’instrument.

1.2.5 Avis de la Federal Communications Commission (FCC)


Ce matériel produit, utilise et peut irradier de l’énergie en radiofréquence et, s’il n’est pas installé conformément aux
instructions, il peut perturber les communications radio.

5
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.3 Italiano
1.3.1 Convenzioni

PERICOLO!
PERICOLO! indica il rischio di incorrere ferite agli occhi se l’utente non
effettua il procedimento in modo corretto.

AVVERTENZA
AVVERTENZA indica il rischio di possibili lesioni se l’utente non effettua il
procedimento in modo corretto.

ATTENZIONE
ATTENZIONE indica il rischio di danni al prodotto se l’utente non effettua il
procedimento in modo corretto.

CAUTELA
CAUTELA indica la possibilità di possibili inconvenienti per l'utente, come
un'eventuale perdita di dati, se l’utente non effettua il procedimento in modo
corretto.

NOTA
NOTA indica delle informazioni che devono essere fatte presenti all’utente.

Leggere e comprendere le istruzioni per l’installazione e l’uso prima di accendere ACR-2000.

1.3.2 Protezione dal Laser


Il ACR-2000 include un laser diode Classe IIIb, > da 15 mW. I coperchi del ACR-2000 proteggono l’utente
dallíesposizione diretta al raggio laser. Questi coperchi proteggono l’utente solo se sono correttamente montati
quando il sistema è in uso. I coperchi devono essere rimossi e sostituiti solo da personale qualificato. Se i coperchi
sono stati danneggiati durante la spedizione o l’uso, contattare il servizio di assistenza più vicino per la sostituzione
dei coperchi.

PERICOLO!
QUANDO IL COPERCHIO È TOLTO, IL RAGGIO LASER È
ESPOSTO.
EVITARE L’ESPOSIZIONE DIRETTA DEGLI OCCHI AL RAGGIO
LASER.

1.3.3 Etichette del ACR-2000 e loro ubicazione


Le etichette del ACR-2000 e la loro ubicazione sono indicate nelle Figure A1-1 e A1-2, APPENDIX A.

6
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.3.4 Pericolo di scosse elettriche

AVVERTENZA
IN QUESTA ATTREZZATURA SONO PRESENTI TENSIONI
ELETTRICHE PERICOLOSE CHE POSSONO PROCURARE
SCOSSE, USTIONI O MORTE.

ATTENZIONE
FUSIBILE BIPOLARE / NEUTRALE.

La manutenzione di questa attrezzatura deve essere effettuata da una persona qualificata ed autorizzata da Lumisys.
NON usare il ACR-2000 con un cavo di alimentazione danneggiato. Non è consigliato l’uso di una prolunga
elettrica. Questa attrezzatura deve essere messa a terra e i collegamenti elettrici devono essere ispezionati in modo
appropriato per garantirne il funzionamento nella massima sicurezza.

1.3.5 Avviso FCC


Questa attrezzatura genera, usa e trasmette radio frequenze. Se non è installata seguendo le istruzioni
d’installazione, può causare interferenze con comunicazioni radio.

7
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.4 Español
1.4.1 Terminología

¡PELIGRO!
¡PELIGRO! indica que puede ocurrir una lesión a los ojos si el usuario no lleva
a cabo el procedimiento en la forma correcta.

ADVERTENCIA
Un ADVERTENCIA indica que usuario se puede lesionar si no lleva a cabo los
procedimientos correctamente.

AVISO
Un AVISO indica que el producto se puede dañar si el usuario no lleva a cabo
los procedimientos correctamente.

PRECAUCIÓN
Una PRECAUCIÓN indica que el usuario puede encontrarse con problemas,
tales como pérdida de datos, si no lleva a cabo los procedimientos
correctamente.

NOTA
Una NOTA proporciona información sobre la cual se debe llamar la atención
del usuario.

Asegúrese de leer y comprender las instrucciones de instalación y operación antes de encender el ACR-2000.

1.4.2 Seguridad del láser


El ACR-2000 incluye un láser diode de clase IIIb, mas de 15 mW; y está dotado de cubiertas para proteger al
usuario de la exposición directa a la luz del láser. Dichas cubiertas del ACR-2000 cumplirán su función solamente
si están adecuadamente instaladas cuando se esté utilizando el sistema. Las cubiertas solamente deben ser retiradas
y colocadas por personal con la preparación adecuada. Si las cubiertas resultan dañadas durante el envío o durante
el uso, póngase en contacto con su representante local de servicio técnico para reemplazarlas.

¡PELIGRO!
LUZ LÁSER CUANDO SE RETIRAN LAS CUBIERTAS.
EVITE LA EXPOSICIÓN DIRECTA DE LOS OJOS A LA
LUZ.

1.4.3 Etiquetas del ACR-2000 y ubicación de las mismas.


Las etiquetas del ACR-2000 y la ubicación de las mismas se muestran en las Figuras A1-1 y A1-2 en APPENDIX
A.

8
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.4.4 Peligros eléctricos

ADVERTENCIA
ESTE EQUIPO FUNCIONA CON NIVELES DE VOLTAJE
PELIGROSOS QUE PUEDEN CAUSAR SHOCKS,
QUEMADURAS O LA MUERTE.

AVISO
FUSIBLES DE DOS POLOS Y NEUTROS.

La revisión de este equipo debe ser realizada por personas que hayan sido formadas y autorizadas por Lumisys.
NO opere el ACR-2000 con un cable de corriente que esté dañado. No se recomienda usar prolongadores
eléctricos. Este equipo debe tener una conexión adecuada a tierra y se deben inspeccionar las conexiones eléctricas
para garantizar un funcionamiento seguro.

1.4.5 Notificación de la Comisión Federal de Comunicaciones de EE.UU. (FCC)


Este equipo genera, utiliza y puede irradiar energía de una frecuencia de radio y, si no se instala de acuerdo con las
instrucciones pertinentes, puede causar interferencias con las radiocomunicaciones.

9
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.5 Deutsch
1.5.1 Definitionen

GEFAHR!
GEFAHR! weist auf eine Verletzungsgefahr für die Augen hin, sollte der
Benutzer nicht sachgemäß vorgehen.

WARNUNG
WARNUNG weist auf eine Verletzungsgefahr hin, sollte der Benutzer nicht
sachgemäß vorgehen.

VORSICHT
VORSICHT weist auf das Risiko einer Produktbeschädigung hin, falls der
Benutzer nicht sachgemäß vorgeht.

VORSICHTSMASSNAHME
VORSICHTSMASSNAHME weist auf mögliche Unannehmlichkeiten hin,
wie z.B. den Verlust von Daten, falls der Benutzer nicht sachgemäß vorgeht.

HINWEIS
Informationen, die vom Benutzer beachtet werden sollten, sind mit HINWEIS
bezeichnet.

Lesen Sie bitte die Installations- und Betriebsanleitung sorgfältig durch, bevor Sie den ACR-2000 an das Stromnetz
anschließen.

1.5.2 Sicherheitshinweise für den Laser


Der ACR-2000 verwendet einen Laser-Diode der Klasse IIIb, > 15 mW. Die Abdeckungen am ACR-2000 schützen
den Benutzer vor direktem Laserlicht. Diese Abdeckungen schützen den Benutzer nur, wenn Sie vor Einsatz des
Systems sachgemäß installiert worden sind. Die Abdeckungen dürfen nur von geschultem Personal entfernt oder
ersetzt werden. Falls die Abdeckungen während des Versands oder Gebrauchs beschädigt wurden, wenden Sie sich
für Ersatzabdeckungen bitte an Ihre örtliche Servicevertretung.

GEFAHR!
BEI ENTFERNEN DER ABDECKUNG: LASERSTRAHLEN.
NIEMALS DIREKT IN DEN LASER BLICKEN.

1.5.3 ACR-2000 Etiketten und Etikettenplazierungen


Die ACR-2000 Etiketten und deren Plazierung werden in den Abbildungen 1-1 und 1-2 gezeigt.

10
$&52SHUDWRU·V5HIHUHQFH*XLGH

1.5.4 Elektrische Gefahren

WARNUNG
DIESES GERÄT ARBEITET MIT GEFÄHRLICHEN
SPANNUNGEN, DIE ZU STROMSCHLAG,
VERBRENNUNGEN ODER ZUM TODE FÜHREN KÖNNEN.

VORSICHT
ZWEIPOLIGE BZW. NEUTRALLEITER-SICHERUNG.

Dieses Gerät darf nur von Personen gewartet werden, die von Lumisys geschult und beglaubigt sind.
Den ACR-2000 niemals mit schadhaftem Stromkabel verwenden. Vom Gebrauch eines Verlängerungskabels wird
abgeraten. Das Gerät muß vorschriftsmäßig geerdet und der Netzanschluß überprüft werden, damit ein gefahrloser
Betrieb gewährleistet ist.

1.5.5 Mitteilung der FCC


Dieses Gerät erzeugt, benutzt und strahlt Hochfrequenzenergie aus. Wenn das Gerät nicht gemäß der Anleitung
installiert wird, kann dies Störungen des Funkverkehrs zur Folge haben.

11
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

12
$&52SHUDWRU·V5HIHUHQFH*XLGH

2. ACR-2000 READER INSTALLATION

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

To fully install your ACR-2000 Reader, perform the following steps in the order indicated:
• Unpack the ACR-2000 Reader.
• Configure and install the ACR-2000 Reader hardware.
• Install and configure the Host Support Package software on the host PC (if required).
• Configure and install the Data Control Board in the host PC (if required)
• Connect the ACR-2000 Reader to the host PC.

2.1 Unpacking Instructions

WARNING AVERTISSEMENT

THE ACR-2000 WEIGHS OVER 85 POUNDS. LE ACR-2000 PESE PLUS DE 37 KILOS


IT REQUIRES TWO PEOPLE TO SAFELY ET DOIT ETRE DEPLACE PAR DEUX
LIFT AND MOVE IT.
PERSONNES POUR PLUS DE
THE ACR-2000 USES A LASER FOR SECURITE.
SCANNING. DO NOT LOOK DIRECTLY AT
THE LASER LIGHT. LE ACR-2000 UTILISE DES RAYONS LASER
DE SCANOGRAPHIE. NE PAS REGARDER
LES RAYONS LASER SANS PROTECTION
OCULAIRE.

ADVERTENCIA
AVVERTENZA
EL ACR-2000 PESA MÁS DE 37 KILOS
IL ACR-2000 PESA PIÙ DI 37 KG. SONO (85 LIBRAS). SE NECESITAN DOS
NECESSARIE DUE PERSONE PER ALZARLO
E SPOSTARLO CON SICUREZZA. PERSONAS PARA MOVERLO Y
LEVANTARLO CON SEGURIDAD.
IL ACR-2000 UTILIZZA UN RAGGIO LASER
PER LA SCANSIONE. EVITARE EL ACR-2000 UTILIZA UN LÁSER PARA
L’ESPOSIZIONE DIRETTA DEGLI OCCHI AL ESCANEAR. NO MIRE DIRECTAMENTE A
RAGGIO LASER. LA LUZ DEL LÁSER.

13
$&52SHUDWRU·V5HIHUHQFH*XLGH

WARNUNG

DER ACR-2000 WIEGT ÜBER 37 KG. ZUM


SICHEREN HEBEN UND VERSCHIEBEN SIND
ZWEI PERSONEN ERFORDERLICH.
ZUM ABTASTEN NUTZT DER ACR-2000
EINEN LASER. SCHAUEN SIE NIEMALS
DIREKT INS LASERLICHT.

2.1.1 Tools Required


• A 7/16" open end wrench and a large flat bladed screwdriver.
2.1.2 Unpacking the ACR-2000
• Using the large screwdriver, remove the clamp brackets from the bottom of the crate.
• Lift the crate off the pallet.
• Remove the accessories box.
• Remove the plastic bag protecting the system.
• Using the 7/16" open-end wrench, remove the 4 bolts from underneath each corner of the pallet,
directly below the ACR-2000.
• With TWO PEOPLE, lift the ACR-2000 off the pallet.
• To install the ACR-2000, follow the Hardware Installation Procedures (next section).

2.2 Hardware Installation


2.2.1 Tools Required
• A small flat bladed screwdriver.
2.2.2 Connection / Disconnection and Power On / Off Sequences
When connecting power and / or data cables and / or powering on the ACR-2000, (including the first time), perform
the following steps in the order indicated:
• First, connect the interface cable.
• Second, connect the AC power cable.
• Third, power on the ACR-2000.
When disconnecting power and / or data cables and /or powering off the ACR-2000, perform the following steps in
the order indicated:
• First, power off the ACR-2000.
• Second, disconnect the AC power cable.
• Third, disconnect the interface cable.

2.2.3 Setting AC Voltage


You must verify that the ACR-2000 is set up for the correct AC line voltage. Check this by looking at the power
configuration panel. This panel is located just to the right of the power entry plug, next to the on/off switch in the
rear of the reader. The ACR-2000 is factory set for 120 VAC for domestic units unless specified otherwise. See
Figure 4-2.

WARNING
DISCONNECT POWER BEFORE CHANGING LINE VOLTAGES OR
FUSES.

14
$&52SHUDWRU·V5HIHUHQFH*XLGH

If the ACR-2000 is to be operated at another line voltage, the programming card must be changed. Remove the
plastic Fuse Block/Cover. Remove the voltage programming card and rotate it until the tab with the desired voltage
is pointing down, rotate the voltage indicator to point up and reinsert the card.
If the new line voltage is 230 / 240 VAC the fusing method must be changed. This is accomplished by loosening the
Phillips screw that holds the Fuse Block to the cover. Remove the Fuse Block, invert it and tighten the Phillips
screw. Install two 1.0 Amp, 250 Volt Slo-Blo fuses (Lumisys Part Number 0068-487).
Reinstall the Fuse Block/Cover and insure the indicator is pointing to the correct line voltage.
The ACR-2000 is now ready for operation.

NOTE
The fuse used in the ACR-2000 for 100 / 120 volt operation is a 1.5 Amp, 250
Volt Slo-Blo fuse (Lumisys Part Number 0065-513).

2.2.4 AC Power Cable


The ACR-2000 utilizes an international IEC grade connector for the power cable. Systems are shipped with a
standard NEMA 5-15, hospital grade cable. This cable may require replacement depending upon the country of
installation. Insert the female end into the input power socket on the reader. See Figure 4-2.

2.3 Installation of Host Support Package Software


The ACR-2000 comes configured with the ISA option (for PCs). Lumisys provides host computer support software
for the ACR-2000 configured with the ISA option. See the LSDT Configuration Guide (P/N 0071-434) on how to
install and configure the Host Support Package Software.

2.4 Installation of Data Control Board (DCB)


With power off, remove cover from the target IBM PC-AT or compatible. Taking appropriate anti-static
precautions, remove the Data Control Board from the accessory box and install it in an 8- or 16-bit ISA slot.
Secure the DCB in the slot and replace the covers.
The Data Control Board is factory configured to use IRQ 5 and I/O Base Addresses 100-11F. The ACR-2000 uses
a 32-Kbyte window at hex address D0000-D7FFF. If IRQ 5, the I/O Base Address or the 32-Kbyte window address
conflicts with your system configuration they may be changed. See the LSDT Configuration Guide (P/N 0071-434)
for instructions on how to configure the IRQ level, I/O Base Address or the 32-Kbyte Memory Window address.

2.5 Installation of Control Interconnect Cable


Remove the Control Interconnect Cable (a 37-pin, Male/Male, D-Subminiature cable) from the accessories box.
Connect one end to the Data Control Board, the other end to the ACR-2000 and secure both ends using a small flat
blade screwdriver. See Figure 4-2.

15
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

16
$&52SHUDWRU·V5HIHUHQFH*XLGH

3. SYSTEM DESCRIPTION

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

The ACR-2000 is a tabletop laser phosphor plate reader that brings together performance and durability at an
affordable price. Adaptable to most digital imaging applications, the ACR-2000 integrates easily into most
computer environments.

)LJXUH$&5

3.1 Optics
The ACR-2000 is a self-contained unit housing a laser, laser power supply, fixed optics and mirrors, a scanning
galvanometer and a light detection system.
As a beam source, the ACR-2000 uses a laser diode that generates over 15 mW of visible light. The laser beam is
directed into the focusing lens. After passing through the focusing lens, the laser beam is directed to a galvanometer
scanner. This galvanometer scanner has a precision mirror that sweeps the beam across the plate plane via a folding
mirror.
Light is collected in an integrating cylinder immediately in front of the plate plane. This collected light is detected
by a photomultiplier tube (PMT) inserted in the side of the cylinder. The PMT converts the detected light energy,
which was emitted from the storage phosphor plate, into an analog electrical signal which is processed,
logarithmically amplified, corrected for temporal fluctuation in laser output and then digitized by an A/D converter.

17
$&52SHUDWRU·V5HIHUHQFH*XLGH

)LJXUH$&52SWLFDO3DWK

3.2 Power and Electronics


3.2.1 Printed Circuit Boards
The electrical subsystem of the ACR-2000 consists of the Data Control Board (DCB), which is located in the
user’s host computer such as a PC-AT, and 5 other boards which are located within the ACR-2000. These boards
are:
Data Acquisition (DACQ)
PMT Pre-Amplifier
LED Indicator
Galvo Control
Laser Driver (with interlocks)
3.2.1.1 Data Control Board
The Data Control Board via control registers controls all reader operations. Although some of these registers are
physically located on the Data Acquisition Board, they are accessed through the Data Control Board. Once a scan is
initiated, data acquisition is automatic, requiring no intervention until the image has been written into image
memory on the Data Control Board. Image data can be transferred out of image memory during or after image
acquisition.

3.2.1.2 Data Acquisition Board


The Data Acquisition Board performs all the signal conditioning and data acquisition functions, including
calibration and lookup table functions. In addition, this board controls the plate motor drive.

3.2.1.3 PMT Pre-Amp Board


The PMT Pre-Amp Board provides a power connection and takes a low-level signal from the PMT and amplifies it
for further processing by the Digital Acquisition Board.

18
$&52SHUDWRU·V5HIHUHQFH*XLGH

3.2.1.4 LED Indicator Board


The LED Indicator Board contains two LED indicator lamps, one to indicate when power is on, marked
"POWER", and one to indicate when a scan is in process, marked "SCAN".

3.2.1.5 Galvo Control Board


The Galvo Control Board contains drive and feedback circuits for controlling the position and velocity of the
scanning mirror.

3.2.1.6 Laser Driver Board (with Interlocks)


The Laser Driver Board contains drive and conditioning circuits for controlling the laser. Interlock switches are
provided to disconnect DC power to the Laser Driver board when any cover is removed.

3.2.2 Power Distribution


The ACR-2000 contains four power supplies (see Figure 3-3). One high voltage supply is for the PMT. A triple
output linear supply provides +/- 12 volts and +5 volts for general power. Two dual output, linear +/- 15-volt
supplies are also used. One provides power for the PMT high voltage supply and the other provides power to drive
the plate motor and the galvo.

)LJXUH$&53RZHUDQG(OHFWULFDO6\VWHP

NOTE
Please refer to the ACR-2000 Service Manual, P/N 70-715, for detailed descriptions and theory of
operation for the ACR-2000 power and electrical system.

19
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

20
$&52SHUDWRU·V5HIHUHQFH*XLGH

4. POWER-UP

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

The power switch on the ACR-2000 is located in the lower left corner in the rear of the reader. See Figure 4-2.
Two LEDs, one above the other, are located on the left side. The top LED is labeled POWER, while the bottom
LED is labeled SCAN. See Figure 4-1.
When power is turned on, the POWER LED illuminates. Once the power is turned on to the reader, the PC should
be powered on, booted and the driver installed.
The ACR-2000 is now ready for use but should be allowed to warm up for 15 minutes to stabilize the electronics.

)LJXUH/('/RFDWLRQV

21
$&52SHUDWRU·V5HIHUHQFH*XLGH

)LJXUH3RZHU6ZLWFK/RFDWLRQ

22
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

23
$&52SHUDWRU·V5HIHUHQFH*XLGH

5. OPERATING THE ACR-2000 READER

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

5.1 Plate Handling and Loading


Exposed phosphor plates that are to be digitized should be handled carefully to avoid introducing scratches,
excessive fingerprints and/or static.
To digitize a 14" x 17" plate, insert the plate against the media holder as shown in Figure 5-1. The plate should be
loaded top end in, with the phosphor-coated surface of the plate facing the reader.
To digitize smaller plates, such as a 24 cm x 30 cm, insert the plate into the media holder as before, but with the
plate centered approximately (within an inch or so) with respect to the holder.

)LJXUH3ODWH/RDGLQJ

5.2 Reading a Plate


Once a plate has been placed into the plate input chute, the user may initiate a scan. Once started, the plate will be
moved into position, the edges found and the plate digitized. Care must be taken during the digitization of the plate,
do not pull on the plate or jar the ACR-2000 Reader.
Lumisys provides a sample application that will digitize CR plates using an ACR-2000 Reader. For further details
on the use of this software, see Appendix C.

5.3 Removal of Plate Jams and Using the Scan Abort Button
If a jam occurs or the user wishes to stop a scan in process he/she can press the SCAN ABORT button located near
the plate input chute. See Figure 4-1. When the SCAN ABORT button is pressed, the software driver is notified
and it will stop the plate drive motor and abort the scan. Further pressing of the SCAN ABORT button will cause a
toggle action which will start/stop the plate drive motor as well as reverse its direction. A typical sequence is as
follows:
1st press Digitization is aborted and plate drive motor stopped.
2nd press Motor is started in the reverse direction.

24
$&52SHUDWRU·V5HIHUHQFH*XLGH

3rd press Motor is stopped.


4th pressMotor is started in the forward direction.
5th pressMotor is stopped.
6th pressMotor is started in the reverse direction.
Etc. Etc.
Etc. Etc.

WARNING
THIS EQUIPMENT EMPLOYS A LASER. LASER RADIATION MAY
BE PRESENT IF THE ACR-2000 READER IS OPERATED WITHOUT
COVERS.
AVOID LASER BEAM. DIRECT EYE EXPOSURE TO
LASER LIGHT MUST BE AVOIDED.

25
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

26
$&52SHUDWRU·V5HIHUHQFH*XLGH

6. SYSTEM MAINTENANCE

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

The outside covers of the ACR-2000 should be cleaned with a mild soap or detergent. Do not spray cleaner directly
on the ACR-2000 covers. Spray the soap or detergent on a soft, clean cloth, and then wipe down the covers.
The plastic plate support stand and exit tray should be wiped with a dry soft cloth.
There are NO operator serviceable parts within an ACR-2000. Do NOT remove either cover!

27
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

28
$&52SHUDWRU·V5HIHUHQFH*XLGH

7. THE ACR-2000 ERASER

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

7.1 General Information


Each ACR-2000 Reader comes with an ACR-2000E Eraser. The eraser is used to erase a CR phosphor plate before
it is exposed and read in the ACR-2000 Reader. It has no other intended use. It is a stand-alone unit and has no
physical connection to the reader, but should be mounted in close proximity to the reader for operational
convenience.
Each eraser is permanently configured for either 120 or 230 volts and may not be changed.

7.2 Site Requirements


The ACR-2000E is intended to be mounted on the wall. There are two keyhole slots on the rear panel, spaced 16
inches apart. Mount the unit to a wall within six feet of an AC wall outlet and in close proximity to the reader. The
bottom edge of the unit should be between three and five feet from the floor, four is recommended.
The working area where the reader and eraser are located must be dimly lit (so you can see what you are doing), but
with no more than 2 exposure units of light striking the front surface of the reader.

7.3 Installation
To install the unit, first make sure that it is NOT plugged in. Remove the front panel and disconnect the two cables
(the small green light and the timer pushbutton) from the printed circuit assembly near the top of the unit. Depress
the latch levers on the connectors to disconnect them. If necessary, the lamps may be temporarily removed by
releasing the plastic retaining clips on the left, lifting the metal retaining clip away from the socket about 1/8” and
then sliding the lamp to the left.
For hollow wall construction, the recommended method is to mount the unit to two #10 wood or sheet metal screws
firmly screwed through the unit’s keyhole slots into wall studs. If wall studs are unavailable, use #10 toggle or
molly-type hollow wall fasteners. For concrete walls, use an approved fastening method rated for at least 30 lbs.
each.
When the mounting is complete, reinstall the lamps if you have removed them. Be sure that the plastic retaining
clips are snapped closed. Reconnect the two cables to the printed circuit assembly and reattach the front panel.
Connect the eraser’s power entry connector to the AC wall outlet using the supplied AC power cable. Turn on the
unit’s power switch and verify that the lights turn on. Press the timer switch and observe the green light go on and
an audible alarm after 60 seconds.

7.4 Usage Guidelines


The unit is intended to be left on either continuously (24 hours per day, 7 days per week) or continuously during
normal working hours. Do not turn it on and off for each plate – the lamps will not last nearly as long.
A CR plate may be erased in the Eraser in the dark or with the room lights on. Each plate should be erased no more
than 2 hours prior to making an exposure.
After a plate is used (exposure followed by reading), a plate MUST be erased before reuse.
Care must be taken in handling the plates to prevent damage. Follow the CR plate manufacturer’s instructions for
handling, cleaning and storage. Keep your plates clean and free of dust, dirt, lint, fingerprints, etc. Use the CR plate
cleaner recommended by or available from the plate manufacturer. Use only soft, clean cleaning material to avoid
damage to the plate. When not being read, CR plates should be stored in a cassette.

7.5 Operation of the ACR-2000 Reader and Eraser


1. An erased plate is placed into an appropriate cassette and an x-ray exposure is made.

29
$&52SHUDWRU·V5HIHUHQFH*XLGH

2. Under dim-light conditions, the cassette is opened, the exposed plate is removed and then placed against the
media holder located on the top of the ACR-2000 Reader.
3. The plate is then read, after which the plate drops into the catch tray on the reader.
4. The plate just read is removed from the catch tray and erased in the eraser:
• Extend the eraser slider.
• Insert the plate into the eraser slider, white phosphor side toward the eraser.
• Slide the slider (and plate) into the eraser.
• Press the timer button once.
• Wait for the beep and/or the green light to go out (60 sec).
• Remove the plate and place it into its light-tight cassette.
If a plate has been grossly overexposed, two erase cycles may be required to remove the image. If you don’t, you
may see a ‘ghost’ image on the next exposure. You can either perform two erase cycles (as described above), or,
after you press the timer button, wait 5 seconds and press it again. This will start a 120-sec erase cycle.

7.6 Maintenance

WARNING
BEFORE REPLACING ANY FUSES, DISCONNECT THE ERASER’S
AC POWER CORD.

The eraser is protected by either one or two AC line fuses, depending upon the eraser’s line voltage configuration.
110 /120 Volt One 2.5 Amp, 250 Volt Slo-Blo fuse (Lumisys Part Number 37-996)
230 / 240 Volt Two 1.25 Amp, 250 Volt Slo-Blo fuse (Lumisys Part Number 69-416-04)
The line fuses may be replaced by flipping off the power entry connector’s cover and removing the plastic fuse
block.
The eraser circuit assembly is additionally fused with a 315 mA, 250 Volt fuse (Lumisys part Number 69-416-01).
It may be replaced by unscrewing the small, round black fuse cap located next to the eraser’s power entry connector.
Lamps must be replaced only with compact fluorescent 55 W, 3000º K lamps (Lumisys Part Number 70-386).

30
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

$&5/$%(/6 /$%(//2&$7,216

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

)LJXUH$$&5/DEHOV

)LJXUH$(WLTXHWWHVGX$&5

)LJXUH$(WLFKHWWHGHO$&5

)LJXUD$(WLTXHWDVGHO$&5

$EELOGXQJ$$&5(WLNHWWHQ

31
$&52SHUDWRU·V5HIHUHQFH*XLGH

)LJXUH$/DEHO/RFDWLRQV

)LJXUH$(PSODFHPHQWGHVpWLTXHWWHV

)LJXUH$8ELFD]LRQHGHOOHHWLFKHWWH

)LJXUD$8ELFDFLyQGHODV(WLTXHWDV

$EELOGXQJ$(WLNHWWHQSOD]LHUXQJ

32
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

$&56<67(0'(6&5,37,21

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

PLATE SIZE 24 x 30cm to 35 x 43cm (10" x 12" to 14" x 17")


0$;3,;(/5(62/87,21 3L[HOV3HU,QFK µSODWH

365 Pixels per Inch (14” plate)


SPOT SIZE 87 microns, FWHM
6&$15$7( OLQHVVHFRQG

GRAY SCALE RESOLUTION 12 bits (4096 levels)


INTERFACE ISA Bus PC Interface Card with 16 Mbytes Memory
DIMENSIONS 20.5"W x 13"H x 27"D (without exit tray or plate
stand)
32:(55(48,5(0(176 WR9+]$PSV

200 to 240V, 50/60 Hz, 1.0 Amps


TEMPERATURE 15 to 35 degrees C – operating
CONSTRAINTS -18 to 65 degrees C – non-operating
HUMIDITY 20 to 80% non-condensing
VIBRATION/ACCELERATION 3G Max (in shipping)
ALTITUDE 0 to 10,000 ft. – operating
WEIGHT 85 pounds (135 pounds shipping weight)

33
$&52SHUDWRU·V5HIHUHQFH*XLGH

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

$&55($'(56$03/($33/,&$7,21

BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

NOTE: Not applicable to the Internal SCSI Option with the 16-bit Host Support Package.
Your ACR-2000 is supplied with a 1.44-MByte floppy disk that contains the executable code for a sample
application. This application, named "SCANFILE.EXE", will read a plate and put the data and header into an MS-
DOS file.
Before "SCANFILE.EXE" will operate, you must first install the Host Support package that is appropriate for your
digitizer, host Operating System and host /digitizer interface (SCSI or ISA).
The "SCANFILE" application may be activated from the MS-DOS command line or from batch files. The various
options which modify the way "SCANFILE" operates are controlled by using a forward slash (/) such as standard
MS-DOS commands.
The command line for "SCANFILE" has the following format:
SCANFILE [/PPL or /PPI] /PD /PF /LUT /F /DOME /AM /TIF /PMT
Options available are:
PIXELS PER LINE Default is 2048 for the ACR-2000
/PPL:x Set PIXELS PER LINE to x
x is limited to 5120 on the ACR-2000
PIXELS PER INCH Causes the driver to perform a FIXED MODE SCAN
/PPI:x Maximum PPI: ACR-2000, 10” plate = 512
PIXEL DEPTH Default is 12
/PD:12 Set PIXEL DEPTH to 12 bits/pixel
/PD:8 Set PIXEL DEPTH to 8 bits/pixel
PIXEL FORMAT Default is LSB first
/PF:LSB Set PIXEL FORMAT to LSB first
/PF:MSB Set PIXEL FORMAT to MSB first
LOOK UP TABLE Default is Normal
/LUT:N Set LUT to normal - 0.0 Optical Density = 0
/LUT:I Set LUT to inverted - 0.0 Optical Density = 4095
OUTPUT FILE Default is lsdt.img
/F:filename Set filename to ‘filename’
OUTPUT TO DOME DISPLAY
/DOME Read image from reader and output to a DOME Rx20 Display Board
AVERAGING MODE Default is OFF
/AM:OFF Set averaging mode to off
/AM:X Set averaging mode to X averaging
/AM:XY Set averaging mode to X & Y averaging
TIFF FILE OUTPUT
/TIF Set to create TIFF file output
PMT VOLTAGE Default control value is 128 for the ACR-2000
/PMT:x Set PMT voltage according to control value ‘x’
x must be in the range of 0 to 255 on the ACR-2000
If the user would like to scan a plate into a file called ‘image5.img’ using defaults, then from the MS-DOS prompt
type:
34
$&52SHUDWRU·V5HIHUHQFH*XLGH

"scanfile /f:image5.img"
This command will digitize a plate using 12 bits, 2048 pixels in width, create a Lumisys Header and write a file
named ‘image5.img’ into the current directory.

35