Sie sind auf Seite 1von 5

Practice Questions

1. Advantages of biphasic waveform defibrillation over the Lown­type waveform that was used 
up until the late 1980s include which of the following?

a. The use of direct current, instead of alternating current
b. A reduction in the energy needed for successful defibrillation
c. Biphasic defibrillation allowed for the introduction of portable units
d. All of the above

2. An electrocardiograph is malfunctioning. The power LED and the LCD backlight both are 
on, but there is no LCD display. Which of the following might be the cause of these 
observations?

a. A problem in a cable connection to the ECG control board
b. A defective LCD
c. A defective ECG control board
d. All of the above

3. A suspected problem with the thermal head of an electrocardiograph might be checked by 
performing which of the following procedures?

a. Printing out the letters "H" and "X" repeatedly
b. Running the paper roll at 10, 12.5, 25, and 50 millimeters per second
c. Running a demonstration
d. All of the above

4. A flow cytometer is providing a weak fluorescence intensity. This could be due to which of 
the following problems?

a. Excess antibody has been trapped
b. Several populations of cells are present in the sample, instead of one
c. The primary antibody and the secondary antibody are not compatible
d. The sample is contaminated with bacteria

5. Regarding the upkeep of medical ventilators, which of the following statements is MOST 
accurate?

a. Biomedical equipment technicians are responsible only for the maintenance of the units
b. Biomedical equipment technicians are responsible mostly for tuning the various settings
c. Biomedical equipment technicians are responsible for tuning the various settings and for 
maintenance of the units
d. Both tuning of settings and maintenance are the responsibility of respiratory therapists

Answer Key

1. B: The defibrillation technique developed by Bernard Lown in the 1950s used a strongly damped 
sinusoidal wave with a duration of approximately 5 milliseconds. This was supplied by a bank of 
capacitors that were charged to approximately 1 kilovolt that could deliver 100 to 200 joules of 
energy, with the power supplied as direct current. This was a big improvement compared with earlier
methods that had used alternating current and could be performed only internally, with paddles 
placed on the heart during surgery. The Lown method also allowed units to become a lot smaller as 
the technology advanced in increments, so that they could be portable and carried in ambulances. 
The main advantage of the biphasic waveform that replaced the Lown approach is that the energy 
needed for successful defibrillation was reduced sharply. This reduces the risk of damage to the 
myocardium and burns on the skin.

2. D: Since the power LED is on, the machine is getting electrical power. Since the LCD backlight is 
on, the LCD is getting power, but one of the cables connecting to the ECG control board could be 
loose. If not, then either the LCD itself, or the ECG control board could be the source of the problem,
and thus would need to be replaced. Before coming to this conclusion, check whether there is any 
response at all when any switches are pressed. If there is no response at all when any switch is 
pressed, then the ECG control board is the source of the problem.

3. A: The thermal head of an electrocardiograph (ECG) can be tested by having the machine print 
basic patterns. These can be letters, such as "H" and "X," or diagonal lines. Running the paper roll at 
10, 12.5, 25, and 50 millimeters per second is performed to test the motor, gears, and sensors. A 
demonstration is a dummy ECG that is generated by the machine. This is not actually a test, but it is 
performed for educational purposes, for instance, to show people the basic features of a normal ECG 
reading.

4. C: Flow cytometers are used to count tiny particles, such as cells and chromosomes. The particles 
are suspended in a stream of liquid and recognized by fluorescent tags that are attached to the 
particles through antibodies. To amplify the signal, two antibodies are used. The primary antibody 
recognizes antigens on the particles themselves. In the case of cells, such antigens are molecules 
displayed on the cell surface. The primary antibody is made by, or using information from, the 
immune system of any of several animals, including humans. While the area of the primary antibody 
that binds the antigen on the cell is very specific to that antigen, the opposite end of the antibody is 
common to other members of the species from which the antibody was made. The fluorescent dye is 
attached to secondary antibodies, which attach to the species­specific antigens on the primary 
antibodies. The primary antibodies made from a mouse, for instance, need to be labeled with 
secondary antibodies that are "antimouse." If the labeled secondary antibody in this case were 
"antihuman," it would attach poorly to the primary antibody, so the signal would be absent, or 
reduced. Excess antibody being trapped would cause unusually high fluorescence, not a weak signal, 
while bacterial contamination would cause a scattering effect, because bacteria tend to autofluoresce 
at low levels. More than one population of cells present in a sample would show up as such on 
readings.

5. A: In the case of most biomedical equipment, various specialists are responsible for the day­to­day
operation of the equipment, while the biomedical technician is called to see that it works properly. In 
the case of ventilators, the various settings have a direct influence on the patient's comfort level, so 
adjusting these settings falls within the job description of the respiratory therapist. A ventilator is an 
example of a device that must be able to continue operating in the event of a power failure, which is 
to say a life­critical system. A biomedical technician is responsible for making sure that all backup 
systems are operating and that the various settings are calibrated.

Review Questions

1. The normal range for pH in arterial blood is____?


a. 7.00 to 7.25

b. 7.25 to 7.35

c. 7.35 to 7.45

d. 7.45 to 7.55

See the answer


2. Which of the following has the most effect in the regulation of
blood pH?
a. Renal system

b. Respiratory system

c. Endocrine system

d. Chemical buffers

See the answer


3. The kidneys regulate the amount of____in the blood.
a. pCO2
b. pO2
c. HCO3-
d. None of the above

See the answer


4. Which of the following would be a normal pCO2 measurement?
a. 20 mmHg

b. 30 mmHg

c. 40 mmHg

d. 50 mmHg

See the answer


Print This Article

Review Questions

1. In binary, 0101 is equal to which decimal number?


a. 5

b. 10

c. 3

d. 8

See the answer


2. For a properly operating common emitter amplifier circuit, the
base emitter voltage should always be approximately what
value?
a. 10 v

b. 7 v

c. .7 v

d. 3 v

See the answer


3. Pulmonary artery wedge pressure measurements are good
indicators of?
a. Right atria function
b. Pulmonary pressure

c. Left ventricular function

d. Cardiac electro conduction system

See the answer


4. When testing the cardio version or sync function on a
defibrillator, the unit must fire within what amount of time from
the R wave?
a. 6 seconds

b. 6 mS

c. 6 uS

d. 60 mS

See the answer

Das könnte Ihnen auch gefallen