S.No. Description
1 TUS
2 TUS Test Temperature
operating Temperature <170 deg
operating Temperature >170 deg
3 TUS Furnace Parameter
4 TUS Data Collection for cont. / semi cont. furnaces
5 Non - Continuous Furnaces
Thermocouple Systems - Volemetric Method
Thermocouple requirement
working volume <0.1m3
6
working volume > 0.1 - 8.5m3
working volume > 8.5m3
Thermocouple failure
when using 3 to 9 thermocouple
9
when using 10 to 16 thermocouple
when using >16 thermocouple
Thermocouples
S.No. Description
1 Calibration Method
2 Calibration Report
3 Calibration Sources
4 Reuse of thermocouple Type K, E
Thermocouples Calibration & Replacement which used for TUS, SAT (Table
7 3.1.3)
Instumentation
S.No. Description
1 Readability of instrument
Temperature Resolution
Digital
2
Analog
Table 3.4.1
5
9
add 1 thermocouple for each additional 3m3 volume
Each Thermocouple data need to be recorded for every 2 minutes minimum. (Manual Data collection not allowed)
Compare TUS data with process data to check compliance. Total offset/bias assigned to the correction of skewed TUS shall not
exceed 2 deg (This is not related to calibration / SAT errors)
No corner TUS thermocouple failure allowed in all below case
0 failure allowed
1 failure permitted
<= 10% failure permitted
±10 deg from set temperature
Upper Temp tolerance shall not be exceeded at any time
For Known deviations in instruments, corrections can be done in calculations
Table 3.4.2
Thermocouples
Requirement
In operating temperature range. Interval between 2 calibration temp shall not exceed 150deg
Report shall contain Nonminal Temerature, Actual Temp, Correction factor for each port, Calibration method
Shall be accredited to any national / international bodies
Depth of insertion shall be >= Depth of insertion of previous usage
1. If the no. of uses not monitored for Non expendable thermocouples, same shall be replayed in 6 months
2. Thermocouples (Expendable / Non expendable) for load sensor use can be used more than one time when the thermocoup
physically not degraded. Proper documentation need to be done for this
Instumentation
Requirement
Digital instrument minimum readability of 1 deg req.
Analog instrument minimum readability of 2 deg req.