Sie sind auf Seite 1von 132

SIREGRAPH CF

AX

Service

Adjustment Instructions

© Sie mens AG 1995


The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

English
Print No.: RXD3-209.071.01.05.02 Doc. Gen. Date: 09.00
Replaces: RXD3-209.071.01.04.02
0-2 Revision

Chapter Page Revision


all all 01
all all 02
all all 03
all all 04
all all 05

SIREGRAPH CF RXD3-209.071.01 Page 2 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Contents 0-3

Page
1 _______General Information ____________________________________________ 1 - 1

Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 1
Measuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 1
Details Regarding Tolerances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 1
Safety Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Remarks Regarding Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 2
Potentiometers, Remarks Regarding Replacement and Adjustment . . . . . . . . . .1-3
Allarmax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 3
Lubricating Points / Bearings / Screws / Electrostatically-sensitive Devices . . . . . .1- 4
Installation of Ball Bearings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 4
Securing of Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 4
Protection of Electrostatically-sensitive Devices . . . . . . . . . . . . . . . . . . . .1- 4
Manufacturer’s Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 4
Overview I/O PC Board 24S2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 5
PC Mode (Version Windows 3.11). . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 12
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 12
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 12
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 13
PC Mode (Version Windows 95) . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 15
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 15
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 15
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 16
English Version of the Hardware Test and Settings . . . . . . . . . . . . . . . . . . 1 - 18
Error List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 20
Replacing the 24S1 CPU Board and the 24S2 I/O Board . . . . . . . . . . . . . . . 1 - 24
24S1 CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 24
24S2 I/O Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 24
Changes to the DW1 DIP Switches on the 24S2 Board. . . . . . . . . . . . . . . . 1 - 24
Table of DW1 Switch Positions (up to VC00A) . . . . . . . . . . . . . . . . . . 1 - 25
Table of DW1 Switch Positions (as from VC01A) . . . . . . . . . . . . . . . . . 1 - 26
Listing of all Programming Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 27

2 _______Adjustment Preparations ________________________________________ 2 - 1

Adjustment mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1


Start adjustment mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Switching off the adjustment mode . . . . . . . . . . . . . . . . . . . . . . . . .2- 2
Checking the voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 3

3 _______Tilt Drive ______________________________________________________ 3 - 1

Control overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Tilt drive overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Belt Tension, Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . .3- 3
Brake assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3- 4

Siemens AG RXD3-209.071.01 Page 3 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
0-4 Contents

Page
Adjustment or replacement of potentiometer 3PT4. . . . . . . . . . . . . . . . . . . 3-5
Set end positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Safety limit switch 3FC11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Movement times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Safety limit switch 3FC16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Checking the floor spacing with 0 degree table position . . . . . . . . . . . . . . . .3 - 10

4 ______ Spotfilm Device Longitudinal Drive ________________________________4 - 1

Control overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1


Spotfilm device longitudinal drive overview . . . . . . . . . . . . . . . . . . . . . . 4-2
Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Adjustment or replacement of potentiometer 3PT2. . . . . . . . . . . . . . . . . . . 4-4
Set end positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Safety limit switch 3FC14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Checking the PC board SP 22Az4 . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Spotfilm device drive, adjusting belt tension . . . . . . . . . . . . . . . . . . . . . . 4-6
Adjustment procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7

5 ______ Column Drive __________________________________________________5 - 1

Control overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1


Column drive overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Adjustment and replacement of potentiometer 3PT1. . . . . . . . . . . . . . . . . . 5-3
Safety limit switch 3FC12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Blocking limit switch 3FC5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Safety limit switch 3FC15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Checking the PC board SP 22Az3 . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Room height switch 3FC13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Column drive, adjusting the belt tension . . . . . . . . . . . . . . . . . . . . . . 5-7
Adjustment procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8

6 ______ Compression Drive _____________________________________________6 - 1

Control overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1


Compression drive overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Chain tension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Maximum compression force . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Adjustment of switch 3FC4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Movement times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7

7 ______ SID Drive______________________________________________________7 - 1

Control overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1


SID drive overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Adjustment or replacement of potentiometer 3PT7. . . . . . . . . . . . . . . . . . . 7-3
Set end positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4

SIREGRAPH CF RXD3-209.071.01 Page 4 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Contents 0-5

Page
Adjustment of SID Limiting at 90° . . . . . . . . . . . . . . . . . . . . . . . . . . . .7- 4
Safety limit switch 3FC7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7- 5
Adjustment of PC board 23Az7 . . . . . . . . . . . . . . . . . . . . . . . . . . . .7- 6
Movement times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7- 6

8 _______Table Transverse Drive __________________________________________ 8 - 1

Control overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Table transverse drive overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Electrical Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 3

9 _______Cassette Carriage Drive _________________________________________ 9 - 1

Control overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1
Cassette carriage drive overview . . . . . . . . . . . . . . . . . . . . . . . . . . . .9- 2
Mechanical check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-3
Adjust tension of the drive chain . . . . . . . . . . . . . . . . . . . . . . . . . . . .9- 4
Adjust belt tension of the cassette drive. . . . . . . . . . . . . . . . . . . . . . . . .9- 4
Adjusting the slip clutch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9- 5
Adjustment or replacement of potentiometer 3PT3 . . . . . . . . . . . . . . . . . . .9- 7
Set end positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9- 8
Movement times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9- 9
Maximum opening of the cassette jaws . . . . . . . . . . . . . . . . . . . . . . . . .9-9

10 ______Format Sensing _______________________________________________ 10 - 1

Control overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Overview of the format sensing drive . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Adjustment or replacement of potentiometers 3PT10, 3PT11. . . . . . . . . . . . . 10 - 3
Spacer Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 4
Adjusting the end positions and the format sensing . . . . . . . . . . . . . . . 10 - 5

11 ______Diaphragm Plate Drive _________________________________________ 11 - 1

Control overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 1
Diaphragm plate drive overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 2
Mechanical check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 3
Symmetry and centering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 3
Chain tension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 4
Adjust the slip clutch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 4
Adjustment and replacement of potentiometer 3PT5 . . . . . . . . . . . . . . . . . 11 - 6
Adjustment of gap width, general remarks . . . . . . . . . . . . . . . . . . . . . . 11 - 7
Adjustment of gap width (subdivision mode) . . . . . . . . . . . . . . . . . . . . . 11 - 8

12 ______Collimator____________________________________________________ 12 - 1

Control overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 1
Drive overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 2

Siemens AG RXD3-209.071.01 Page 5 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
0-6 Contents

Page
Adjustment after replacement of the potentiometers . . . . . . . . . . . . . . . . . .12 - 3
Adjusting the end positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 - 3
Correction Factors for Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 - 4

13 _____ Grid _________________________________________________________13 - 1

Control overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 - 1


Oscillation grid, general remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 - 2
Programming of oscillation grid / removable multi-line grid . . . . . . . . . . . . .13 - 2
Checking the grid centering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 - 3
Checking the grid movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 - 3
Grid contact. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 - 3

14 _____ Tomography _________________________________________________14 - 1

Setting the tomographic height, beginning with firmware version VB00D . . . . . . .14 - 1
Setting the column speed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 - 1
Setting the spotfilm device speed. . . . . . . . . . . . . . . . . . . . . . . . . .14 - 2
Software correction of the layer-height . . . . . . . . . . . . . . . . . . . . . . .14 - 3
Setting the tomographic symmetry . . . . . . . . . . . . . . . . . . . . . . . . . . .14 - 4

15 _____ Radiation Geometry____________________________________________15 - 1

Coincidence of light field/radiation field. . . . . . . . . . . . . . . . . . . . . . . . .15 - 1


Exposure position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 1
Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 2
Adjustment of the central beam . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 3
Horizontal table direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 3
Vertical table direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 4
Adjust 0 degree switch 3FC1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 4

16 _____ Installing/Removing the Tabletop ________________________________16 - 1

Removing the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 - 1


Installing the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 - 1

17 _____ Installing/Remov. the Cover of the Spotfilm Dev. ____________________17 - 1

Removing the upper cover of the spotfilm device . . . . . . . . . . . . . . . . . . .17 - 1


Removing the lower cover of the spotfilm device. . . . . . . . . . . . . . . . . . . .17 - 1

18 _____ Changes to Previous Version ____________________________________18 - 1

SIREGRAPH CF RXD3-209.071.01 Page 6 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 1-1

Documentation 1

• Wiring diagram SIREGRAPH CF RHD 32

• Unit test certificate


• Self-learning package on CD-Rom, RXD3-209.509.01.01.02

Measuring Equipment 1

NOTICE All tools, measuring equipment and aids mentioned, with the
exception of the “standard installation tools”, are listed and
specified in the ARTD (Part 3)

• 2-channel memory oscilloscope ARTD Part 3

• Digital multimeter Ri>10 MOhm ARTD Part 3

• Fiber optic cable tester 84 28 039 GE999


New:K96 60 812 RE999

• Torque wrench >40 Nm ARTD Part 3

• Spring balance 200 N ARTD Part 3

• Spirit level 28 69 436 RE999

• Protective ground wire tester 85 73 628 RE999

Details Regarding Tolerances 1

General tolerances for measures of length according to ISO 2768

Limit sizes over over over over over over over


for nominal 3 mm 6 mm 30 mm 120 mm 400 mm 1000 mm 2000 mm
dimension to to to to to to to
range 6 mm 30 mm 120 mm 400 mm 1000 mm 2000 mm 4000 mm
Permissible ± 0.5 mm ± 1 mm ± 1.5 mm ± 2.5 mm ± 4 mm ± 6 mm ± 8 mm
tolerance

These tolerances are valid for all measurements indicated in these instructions as long as
no other tolerance is explicitly indicated after the value.
A tolerance of ±10% is permissible for torque values.

Siemens AG RXD3-209.071.01 Page 1 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1-2 General Information

Safety Notes 1

CAUTION When carrying out the work steps and tests, the safety notes con-
tained in the product-specific documentation as well as the gen-
eral safety notes contained in the ARTD binder must be observed.

Checks and adjustments that must be carried out with radiation switched on are marked
with the radiation symbol.
When performing work procedures which are marked with this symbol, radiation protec-
tive clothing must be worn.

Remarks Regarding Adjustment 1

In the control panel, there are 7-segment display components which display the operating
condition and, in the event of an error, an error number.
The error numbers are listed (see appendix) and indicate the subassembly that is not
functioning properly.
If there should be a malfunction, check the drive mechanism first. In particular, check for
easy movement of the moving parts
Pay particular attention to the cleaning and lubrication of rails and guide parts. Check the
chain tension and toothed rack guides for each drive.
If the mechanism functions properly, check the electronics.

Potentiometers 1

Potentiometers P1 to P14 are located on CPU PC board 24S1. By means of these poten-
tiometers, the voltages for test channel adjustment are set and coated with enamel.

CAUTION Do not turn these potentiometers.

Potentiometers which can be accessed from the front are also located on the Mini-Mae-
stro (cassette carriage), Maestro (column/spotfilm device) and small voltage PC boards.
These may be adjusted only if this is described in the corresponding chapter of these
adjustment instructions.
The same applies to the potentiometers on I/0 PC board 24S2.
With replacement of a subassembly configured with potentiometers, the potentiometers
have been adjusted at the factory and may not be changed, except for the potentiometers
described in these adjustment instructions.

SIREGRAPH CF RXD3-209.071.01 Page 2 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1-3

Potentiometers, Remarks Regarding Replacement and


Adjustment 1

When replacing potentiometers for actual value sensing and for their adjustment, it must
be made sure that there is always minimum play between the pinion gears. The potenti-
ometers may not be subjected to mechanical load.

Only the 3PT2 potentiometer (spotfilm device) is adjusted with a play of approx. 0.5 mm to
achieve smooth movement.

The 3PT10 and 3PT11 potentiometers, format sensing for height and width, are
heat-sensitive!

Allarmax 1

This function monitors all drives and switches off if a drive is blocked.
The movement must be blocked in some adjustment steps.
This function must be switched off so that a shutdown does not take place.

NOTICE Switching off the "Allarmax" software


- Insert jumpers on board 24 S2 AMP3
(for service material set)

If this jumper is not configured, in the AMP 3 connector


(24 S1 board) jumper Pin 16 with 1 and Pin 15 with 2.

Overview of AMP 3

9 8

16 1

Siemens AG RXD3-209.071.01 Page 3 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1-4 General Information

IMPORTANT If one replaced subassembly does not function properly, do not


turn the potentiometers; order a new subassembly.

Lubricating Points / Bearings / Screws / Electrostatically-


sensitive Devices 1

Check the condition of all accessible chains, drives, slide and roller bearing guides prior to
beginning each adjustment and if needed, apply bearing grease or a similar substance
(lubricants are listed in Maintenance Instructions, RXD3-209.101.01...). If parts are worn
or damaged, they may be replaced only with original parts.

Installation of Ball Bearings 1

Support rollers, rollers and ball bearings are adjusted so that the unstrained outer ring can
still just be turned manually.

Securing of Screws 1

Secure all screws removed for adjustment purposes according to their original state.
Secure screws by means of appropriate lock washers, counternuts or Loctite type 221,
Part No.: 20 48 874.
Securing by means of Loctite, if necessary, is listed in the text.
When performing work steps with Loctite, please ensure that no securing compound will
penetrate into any bearings.
After some days, the gearing will have seized.
Loctite may be removed by using heat (electric hair-dryer etc.)

Protection of Electrostatically-sensitive Devices 1

CAUTION The integrated circuits and PC boards used in the unit must be
handled with particular care due to their electrically-sensitive
structures and extremely high input resistance. The rules
pertaining to this handling are listed in the ARTD.
It is essential to use the protective devices listed there.

Manufacturer’s Note 1

This document was originally written in German.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1-5

F1 CN7
L8 - L1 PL1

F2 L16 - L9 L97

L98
L24 - L17
F3
L99
L32 - L25

F4 L100
L40 - L33

L101
L48 - L41
F5

L51 L50 L49


L54 L53 L52
L57 L56 L55
L60 L59 L58
L63 L62 L61
DW 1-1 L66 L65 L64
1-8 L69 L68 L67
L72 L71 L70
L75 L74 L73
L78 L77 L75
L81 L80 L79
L84 L83 L82
L87 L86 L85
L90 L89 L88
L93 L92 L91
L96 L95 L94
P1 - P6

AMP5

Fig. 1

Overview I/O PC Board 24S2 1

LED Color Function


L01 green Control of unit tilting-up
L02 green Control of unit tilting-down
L03 green Control of lifting magnet SOL 1 with main contact
L04 green Control of lifting magnet SOL 2
(Grid in initial position with precontact)
L05 green no function
L06 green no function
L07 green no function
L08 green no function
L09 green Control of SID=115cm from tableside control panel
L10 green Control of SID=135cm from tableside control panel
L11 green Control of SID=150cm from tableside control panel
L12 green Control of spotfilm device drive 3MT2
L13 green Control of column drive 3MT1
L14 green Control of cassette carriage drive 3MT3
L15 green no function
L16 green Control of collimator plate drive3MT5

Siemens AG RXD3-209.071.01 Page 5 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1-6 General Information

F1 CN7
L8 - L1 PL1

F2 L16 - L9 L97

L98
L24 - L17
F3
L99
L32 - L25

F4 L100
L40 - L33

L101
L48 - L41
F5

L51 L50 L49


L54 L53 L52
L57 L56 L55
L60 L59 L58
L63 L62 L61
DW 1-1 L66 L65 L64
1-8 L69 L68 L67
L72 L71 L70
L75 L74 L73
L78 L77 L75
L81 L80 L79
L84 L83 L82
L87 L86 L85
L90 L89 L88
L93 L92 L91
L96 L95 L94
P1 - P6

AMP5

Fig. 2

LED Color Function


L17 green Control of compression drive 3MT6
L18 green Control of SID drive 3MT7
L19 green Control of table transverse 3MT8
L20 green not used
L21 green Control of collimator lamp ON
L22 green Control of collimator drives (LED OFF)
L23 green 3FC1 microswitch (tube unit centered to I.I.) open, LED OFF
L24 green Control at kk-Interface IONTOMAT (central measuring field)
L25 green Control at kk-Interface IONTOMAT (left measuring field)
L26 green Control at kk-Interface IONTOMAT (right measuring field)
L27 green Control at kk-Interface IONTOMAT
(AC signal with collimator of height/width <12 cm)
L28 green no function
L29 green Control at kk-Interface (fluoroscopy ON)
L30 green Control at kk-Interface (precontact ON)
L31 green Control at kk-Interface (main contact ON)
L32 green Control at kk-Interface (image inversion)

SIREGRAPH CF RXD3-209.071.01 Page 6 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1-7

F1 CN7
L8 - L1 PL1

F2 L16 - L9 L97

L98
L24 - L17
F3
L99
L32 - L25

F4 L100
L40 - L33

L101
L48 - L41
F5

L51 L50 L49


L54 L53 L52
L57 L56 L55
L60 L59 L58
L63 L62 L61
DW 1-1 L66 L65 L64
1-8 L69 L68 L67
L72 L71 L70
L75 L74 L73
L78 L77 L75
L81 L80 L79
L84 L83 L82
L87 L86 L85
L90 L89 L88
L93 L92 L91
L96 L95 L94
P1 - P6

AMP5

Fig. 3

LED Color Function


L33 green Control at kk-Interface (zoom1)
L34 green Control at kk-Interface (zoom2)
L35 green Control at kk-Interface (zoom3)
L36 green Control at kk-Interface (direct radiography)
L37 green Control at kk-Interface (indirect radiography) DR
L38 green Control at kk-Interface (room light)
L39 green Control at kk-Interface (Fluoro 1/ADR 1)
L40 green Control at kk-Interface (Fluoro 2/ADR 2)
L41 green Control at kk-Interface (Fluoro stop/ADR stop)
L42 green Control at kk-Interface (tomographic time A/0.6 sec.)
L43 green Control at kk-Interface (tomographic time B/0.8 sec.)
L44 green Control at kk-Interface (tomographic time C/1.2 sec.)
L45 green Control at kk-Interface (Serial operation, direct or indirect
technique with 2- to 4-subdivisions)
L46 green no function
L47 green no function
L48 green no function

Siemens AG RXD3-209.071.01 Page 7 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1-8 General Information

F1 CN7
L8 - L1 PL1

F2 L16 - L9 L97

L98
L24 - L17
F3
L99
L32 - L25

F4 L100
L40 - L33

L101
L48 - L41
F5

L51 L50 L49


L54 L53 L52
L57 L56 L55
L60 L59 L58
L63 L62 L61
DW 1-1 L66 L65 L64
1-8 L69 L68 L67
L72 L71 L70
L75 L74 L73
L78 L77 L75
L81 L80 L79
L84 L83 L82
L87 L86 L85
L90 L89 L88
L93 L92 L91
L96 L95 L94
P1 - P6

AMP5

Fig. 4

LED Color Function


L49 green Control at kk-Interface (voltage ON)
L50 green Control at kk-Interface (exposure signal-as)
L51 green Control at kk-Interface (blocking, e.g. door contact)
L52 green Control at kk-Interface (blocking, e.g. 3D stand)
L53 green no function
L54 green no function
L55 green no function
L56 green Microswitch 3FC5 (blocking limit switch) actuated
L57 green Proximity switch 3XS2 (or microswitch 3FC10) (grid contact)
actuated, LED ON; removable multi-line grid inserted: LED ON
L58 green Microswitch 3FC1 (tube centered onto I.I.) open, LED OFF
L59 green Voltage applied to control PC board 22Az3 (column)
L60 green Voltage applied to control PC board 22Az4 (spotfilm device)
L61 green Microswitch 3FC8 (table transverse - wall side) actuated
L62 green Microswitch 3FC9 (table transverse - doctor’s side) actuated
L63 green no function
L64 green no function

SIREGRAPH CF RXD3-209.071.01 Page 8 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1-9

F1 CN7
L8 - L1 PL1

F2 L16 - L9 L97

L98
L24 - L17
F3
L99
L32 - L25

F4 L100
L40 - L33

L101
L48 - L41
F5

L51 L50 L49


L54 L53 L52
L57 L56 L55
L60 L59 L58
L63 L62 L61
DW 1-1 L66 L65 L64
1-8 L69 L68 L67
L72 L71 L70
L75 L74 L73
L78 L77 L75
L81 L80 L79
L84 L83 L82
L87 L86 L85
L90 L89 L88
L93 L92 L91
L96 L95 L94
P1 - P6

AMP5

Fig. 5

LED Color Function


L65 green Collimator control 24S3 ready for operation is OFF when tube
assembly pivoted out 90/180°
L66 green Microswitch 3FC2 (compression maximally decompressed)
actuated
L67 green Microswitch 3FC3 (cone inserted) actuated
L68 green Microswitch 3FC4 (compression >20N - 30N)
L69 green no function
L70 green no function
L71 green no function
L72 green no function
L73 green Programming switch position DW1/1 (room height, up to VC00A)
L74 green Programming switch position DW1/2 (room height, up to VC00A)
L75 green Programming switch position DW1/3 (I.I. size)
L76 green Programming switch position DW1/4 (I.I. size)
L77 green Programming switch position DW1/5 (number of zoom formats)
L78 green Programming switch position DW1/6 (digital radiography/indirect
radiography, up to VC00A)
L79 green Programming switch position DW1/7 (inch/cm)
L80 green Programming switch position DW1/8 (adjustment mode)

Siemens AG RXD3-209.071.01 Page 9 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 10 General Information

F1 CN7
L8 - L1 PL1

F2 L16 - L9 L97

L98
L24 - L17
F3
L99
L32 - L25

F4 L100
L40 - L33

L101
L48 - L41
F5

L51 L50 L49


L54 L53 L52
L57 L56 L55
L60 L59 L58
L63 L62 L61
DW 1-1 L66 L65 L64
1-8 L69 L68 L67
L72 L71 L70
L75 L74 L73
L78 L77 L75
L81 L80 L79
L84 L83 L82
L87 L86 L85
L90 L89 L88
L93 L92 L91
L96 L95 L94
P1 - P6

AMP5

Fig. 6

LED Color Function


L81 green no function
L82 green no function
L83 green no function
L84 green no function
L85 green Table-side control panel key P39A/B
(cassette carriage Load-Park position)
L86 green Table-side control panel key P44B (table-transv. doctor’s side) actuated
L87 green Table-side control panel key P44D (table-transv. wall side) actuated
L88 green Table-side control panel key P44A (long. carriage headward) actuated
L89 green Table-side control panel key P44C (long. carriage footward) actuated
L90 green Table-side control panel key P42A (tilt down unit) actuated
L91 green Table-side control panel key P42B (tilt up unit) actuated
L92 green Table-side control panel key P43A (pivot column to head end) actuated
L93 green Table-side control panel key P43B (pivot column footward) actuated
L94 green Table-side control panel key P21 (SID=115 cm) actuated
L95 green Table-side control panel key P22 (SID=135 cm) actuated
L96 green Table-side control panel key P23 (SID=150 cm) actuated

SIREGRAPH CF RXD3-209.071.01 Page 10 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 11

F1 CN7
L8 - L1 PL1

F2 L16 - L9 L97

L98
L24 - L17
F3
L99
L32 - L25

F4 L100
L40 - L33

L101
L48 - L41
F5

L51 L50 L49


L54 L53 L52
L57 L56 L55
L60 L59 L58
L63 L62 L61
DW 1-1 L66 L65 L64
1-8 L69 L68 L67
L72 L71 L70
L75 L74 L73
L78 L77 L75
L81 L80 L79
L84 L83 L82
L87 L86 L85
L90 L89 L88
L93 L92 L91
L96 L95 L94
P1 - P6

AMP5

Fig. 7

LED Color Function

L97 red Short-circuit on CN7 OUT.9-16, protection for driver U1,actuate


reset button PL1
L98 red Short-circuit on CN7 OUT.17-24, protection for driver U2, actuate
reset button PL1
L99 red Short-circuit on CN6 OUT.25-32, protection for driver U3, actuate
reset button PL1
L100 red Short-circuit on CN6 OUT.33-40, protection for driver U4, actuate
reset button PL1
L101 red Short-circuit on CN6 OUT.41-48, protection for driver U5, actuate
reset button PL1

Siemens AG RXD3-209.071.01 Page 11 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 12 General Information

PC Mode (Version Windows 3.11) 1

Tools required 1

• Laptop / notebook computer with Windows 3.11


• Connection cable (in the service bag), Part Number: 32 49 062 GE066

Preparations 1

• Connect cable to the plug CN 3 on the CPU board 24S1 and in the PC COM 1
• Switch on PC and system
• Call up Windows
• Click on "Accessories"
• Click on "Terminal"
• Click on "Settings"
• Click on "Terminal Emulation"
• Select "DEC VT52" and acknowledge with "OK" (Fig. 8)

Terminal Emulation

(Allgemein/General) OK
Abbrechen
CANCEL
Cancel

Fig. 8

• Select "Data transmission" again under "Terminal" at "Settings" (Fig. 9)

Datenübertragung / Datatransfer
Übertragungsrate / Transferrate (Baud) OK
110 300 600 1200
2400
Abbrechen / Cancel
4800 9600 19200
Stoppbits
Datenbits / Databits
1 1,5 2
5 6 7 8

Parität / Parity Protokoll / Protocol Anschluß / Connection


Keine / None Xon / Xoff Kein
Kein
Ke / None
COM
Ungerade / Odd Hardware COM 1
Gerade / Even Kein / None COM 2
Markierung / Marker
Paritätsprüfung / Paritycheck
Leerzeichen / Space Trägersignal entdecken / Discover carriersignal

Fig. 9

SIREGRAPH CF RXD3-209.071.01 Page 12 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 13
then select and acknowledge
Connection: COM 1
Baud Rate: 9600
Protocol: none
Parity: none

Procedure 1

• The menu appears with Shift-G:


0 = Error memory / Messages file
3 = Hardware test and Settings, (Host us 1228)
8 = Keyboard test
9 = English / Italian changeover

WARNING Hardware test: Only the tests


No 1. Battery check and
No 2. Clock test / Adjust
may be carried out. Carrying out the other tests may lead to dam-
age of the unit.

English version

-------- Siregraph CF CARDS CONTROL --------

*** Digit card number under test:

*----> 0 = Messages file *----> 5=


*----> 1 = *----> 6=
*----> 2 = *----> 7=
*----> 3 = Hardware test / Settings *----> 8= Keyboard Test
*----> 4 = *----> 9= Lingua italiana

Test on card No. (0..9):

NOTICE With introduction of the firmware version VB00B, the text is in


English/Italian< VB00B only in Italian.

Siemens AG RXD3-209.071.01 Page 13 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 14 General Information
Italian version

-------- Siregraph CF CARDS CONTROL --------

*** Digita il numero della scheda sotto test:

*----> 0 = Archivio messaggi *----> 5=


*----> 1 = *----> 6=
*----> 2 = *----> 7=
*----> 3 = Host us 1228 *----> 8= Tastiera us 1272
*----> 4 = *----> 9= English language

Test della scheda n. (0..9):

CAUTION Select the entire screen area so that the text can be displayed and
read completely.

In the error log mode, an inquiry is made about whether or not the contents of the error log
are to be erased. The meaning of the errors is explained on the following pages (1 - 20).
The function of the displays and control buttons on the remote control panel can be tested
in the keyboard test.
All steps are menu-prompted and are not described in greater detail here.

• Call up test No 1 (battery-voltage test) in the “Hardware test and Settings” (see also
pages 1- 18 and 1 - 19) and check the voltage of the battery (CPU board 24S1).
Call up test No 12 (Clock-Test /Adjust), check date and time and correct it if necessary.
This must be done every time a startup or a CPU board replacement is carried out.

Access to test No 12 is quick using the cursor keys.


Observe the WARNING on page 1- 13.

• Exit the program with ESC.

SIREGRAPH CF RXD3-209.071.01 Page 14 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 15

PC Mode (Version Windows 95) 1

Tools required 1

• Laptop / notebook computer with Windows 95


• Connection cable (in the service bag), Part Number: 32 49 062 GE066

Preparations 1

• Connect cable to the plug CN 3 on the CPU board 24S1 and in the PC COM 1
• Switch on PC and system
• Call up Windows
• “Start”
• “Programs”
• “Accessories”
• “Hyperterminal”
• “Hyperterminal.exe”
• “Connection Description”: type in a suitable name (e.g. Siregraph CF) and select a
symbol for the icon in the “Hyperterminal" menu or at the desktop
• Acknowledge with “OK”
• “Phone Number” / “Connect using” / select “Direct to COM1”
• Acknowledge with “OK”
• “COM 1 Properties”: set the data as shown in the table below
Bits per second 9600
Data bits 8
Parity None
Stop bits 1
Protocol / None
Flow control

• Acknowledge with “OK”


• “File”
• “Properties” / “Settings”
• “Emulation” / select “VT-52”
• Acknowledge with “OK”

Siemens AG RXD3-209.071.01 Page 15 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 16 General Information
Procedure 1

• The menu appears with Shift-G:


0 = Error memory /Messages file
3 = Hardware test and Settings, (Host us 1228)
8 = Keyboard test
9 = English / Italian changeover

WARNING Hardware test: Only the tests


No 1. Battery check and
No 2. Clock test / Adjust
may be carried out. Carrying out the other tests may lead to dam-
age of the unit.

English version

-------- Siregraph CF CARDS CONTROL --------

*** Digit card number under test:

*----> 0 = Messages file *----> 5=


*----> 1 = *----> 6=
*----> 2 = *----> 7=
*----> 3 = Hardware test / Settings *----> 8= Keyboard Test
*----> 4 = *----> 9= Lingua italiana

Test on card No. (0..9):

NOTICE With introduction of the firmware version VB00B, the text is in


English/Italian, < VB00B only in Italian.

SIREGRAPH CF RXD3-209.071.01 Page 16 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 17
Italian version

-------- Siregraph CF CARDS CONTROL --------

*** Digita il numero della scheda sotto test:

*----> 0 = Archivio messaggi *----> 5=


*----> 1 = *----> 6=
*----> 2 = *----> 7=
*----> 3 = Host us 1228 *----> 8= Tastiera us 1272
*----> 4 = *----> 9= English language

Test della scheda n. (0..9):

CAUTION Select the entire screen area so that the text can be displayed and
read completely.

In the error log mode, an inquiry is made about whether or not the contents of the error log
are to be erased. The meaning of the errors is explained on the following pages (1 - 20).
The function of the displays and control buttons on the remote control panel can be tested
in the keyboard test.
All steps are menu-prompted and are not described in greater detail here.

• Call up test No 1 (battery voltage test) in the “Hardware test and Settings” (see also
pages 1- 18 and 1 - 19) and check the voltage of the battery (CPU board 24S1).
Call up test No 12 (Clock-Test /Adjust), check date and time and correct it if necessary.
This must be done every time a startup or a CPU board replacement is carried out.

Access to test No 12 is quick using the cursor keys.


Observe the WARNING on page 1- 16.

• Exit the program with ESC.

Siemens AG RXD3-209.071.01 Page 17 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 18 General Information

English Version of the Hardware Test and Settings 1

-------- CARD uS1228 CONTROL --------

Controls in Consecutive mode or Test by test mode (C/T):

----- Battery voltage TEST.


Test No 1
--------------------------------------

----- Data RAM TEST.


Test No 2.
Not to be executed !

----- RAM 1 BIT TEST.


Test No 3.
Not to be executed !

----- Address RAM TEST.


Test No 4.
Not to be executed !

----- SIGNALS *UDS, *LDS TEST.


Test No 5.
Not to be executed !

----- TEST Checksum EPROM.


Test No 6.
Not to be executed !

----- EEPROM TEST.


Test No 7.
Not to be executed !

----- Time interrupt PIT1 TEST.


Test No 8.
Not to be executed !

----- Led RUN and Relay OK TEST.


Test No 9.
Not to be executed !

----- PIT 2 TEST.


Test No 10.
Not to be executed !

SIREGRAPH CF RXD3-209.071.01 Page 18 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 19

----- Serial line No. 2 TEST (OPTICAL FIBRE).


Test No 11.
Not to be executed !

----- CLOCK TEST (RTC).


Test No 12
--------------------------------------

----- Digital INPUT/OUTPUT TEST.


Test No 13.
Not to be executed !

----- Trimmer CALIBRATION uS1269.


Test No 14.
Not to be executed !

----- Digital/analogical converter (DAC) TEST.


Test No 15.
Not to be executed !

----- Analogical/digital converter (ADC) TEST.


Test No 16.
Not to be executed !

----- TEST on precision of ADC reading.


Test No 17.
Not to be executed !

----- Press a key to continue:

Siemens AG RXD3-209.071.01 Page 19 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 20 General Information

Error List 1

Drive Error State Possible causes

Tilt 001 Rest Potentiometer 3PT4 failed


Potentiometer 3PT4 beyond end position
Connection cable interrupted / not connected
ADC input on CPU board 24S1 failed

001 Movement Potentiometer 3PT4 wrongly wired

Column 002 Rest Potentiometer 3PT1 failed


Potentiometer 3PT1 beyond end position
Connection cable interrupted / not connected
ADC input on CPU board 24S1 failed

012 Movement Potentiometer 3PT1 mechanically not connected


column + spotfilm Potentiometer 3PT1 incorrectly wired
device Motor 3MT1 incorrectly wired
Tacho generator 3DT1 failed / incorrectly wired

012 Actuated and no DAC output on CPU board 24S1 failed


movement Motor 3MT1 not connected
Safety limit switch 3FC12 actuated
Current limitation on 22AZ3 (Maestro) responded

022 Oblique projection Potentiometer 3PT1 mechanically not connected


Potentiometer 3PT1 incorrectly wired
Motor 3MT1 wrongly wired
Tacho generator 3DT1 failed / incorrectly wired

022 Actuated and no DAC output on CPU board 24S1 failed


movement Motor 3MT1 not connected
Safety limit switch 3FC12 actuated
Current limitation on 22AZ3 (Maestro) responded

032 Tomography Too big difference between nominal and actual value

042 Movement Too big difference between nominal and actual value

Spotfilm device 003 Rest Potentiometer 3PT2 failed


Potentiometer 3PT2 beyond end position
Connection cable interrupted / not connected
ADC input on CPU board 24S1 failed

013 Movement Potentiometer 3PT2 failed


Column + spotfilm Potentiometer 3PT2 beyond end position
device Motor 3MT2 incorrectly wired
Tacho generator 3DT2 failed / incorrectly wired

SIREGRAPH CF RXD3-209.071.01 Page 20 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 21

013 Actuated and no DAC output on CPU board 24S1 failed


movement Motor 3MT2 not connected
Safety limit switch 3FC14 actuated
Current limitation on 22AZ4 (Maestro) responded

023 Oblique projection Potentiometer 3PT2 mechanically not connected


Potentiometer 3PT2 incorrectly wired
Motor 3MT2 incorrectly wired
Tacho generator 3DT2 failed / incorrectly wired

023 Actuated and no DAC output on CPU board 24S1 failed


movement Motor 3MT2 not connected
Safety limit switch 3FC14 actuated
Current limitation on 22AZ4 (Maestro) responded

033 Tomography Too big difference between nominal and actual value

043 Movement Too big difference between nominal and actual value

SID drive 004 Rest Potentiometer 3PT7 failed


Potentiometer 3PT7 beyond the end position
Connection cable interrupted / not connected
ADC input on CPU board 24S1 failed

004 Movement Potentiometer 3PT7 mechanically not connected


Potentiometer 3PT7 incorrectly wired
Motor 3MT7 incorrectly wired

004 Actuated and no DAC output on CPU board failed


movement Motor 3MT7 not connected
Safety limit switch 3FC7 actuated
Board 23AZ7 has responded / is damaged

"Width" 005 Rest Potentiometer defective


collimator Potentiometer beyond the end position
Connection cable interrupted / not connected
ADC input on CPU board 24S1 failed

005 Movement Potentiometer mechanically not connected


Potentiometer incorrectly wired

015 Movement Motor incorrectly wired


Deviation between collimator and diaphragm plates too
large

015 Actuated and no DAC output on CPU board 24S1 failed


movement Motor not connected
Board 24S3 defective

Siemens AG RXD3-209.071.01 Page 21 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 22 General Information

"Height" 006 Rest Potentiometer defective


collimator Potentiometer beyond the end position
Connection cable interrupted / not connected
ADC input on CPU board 24S1 failed

006 Movement Potentiometer mechanically not connected


Potentiometer incorrectly wired

016 Movement Motor incorrectly wired


Deviation between collimator and diaphragm plates too
large

016 Actuated and no DAC output on CPU board 24S1 failed


movement Motor not connected
Board 24S3 defective

Diaphragm 007 Rest Potentiometer 3PT5 defective


plates Potentiometer 3PT5 over end position
Connection cable interrupted / not connected
ADC input on CPU board 24S1 failed

007 Movement Potentiometer 3PT5 mechanically not connected


Potentiometer 3PT5 incorrectly wired

017 Movement Motor 3MT5 incorrectly wired


Deviation between collimator and diaphragm plates too
large

017 Actuated and no DAC output on CPU board 24S1 failed


movement Motor 3MT5 not connected
Board 23AZ5 has responded

"Width" 008 Rest Potentiometer 3PT10 defective


clamping jaws Potentiometer 3PT10 beyond the end position
Connection cable 3PT10 interrupted / not connected
ADC input on CPU board 24S1 failed

"Height" 009 Rest Potentiometer 3PT11 defective


clamping jaws Potentiometer 3PT11 beyond the end position
Connection cable 3PT11 interrupted / not connected
ADC input on CPU board 24S1 failed

Cassette 010 Rest Potentiometer 3PT3 defective


carriage Potentiometer 3PT3 beyond the end position
Connection cable interrupted / not connected
ADC input on CPU board 24S1 failed

SIREGRAPH CF RXD3-209.071.01 Page 22 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 23

010 Movement Potentiometer 3PT3 mechanically not connected


Potentiometer 3PT3 incorrectly wired
Motor 3MT3 incorrectly wired
Tacho generator 3DT3 failed / incorrectly wired

010 Actuated and no DAC output on CPU board 24S1 failed


movement Motor 3MT3 not connected
Board 22AZ2 (Mini Maestro) has responded

General 098 Console-control cabinet fiber optic cable defective


Fault in the initialization phase
Wrong data transmission

099 Control console / spotfilm device EMERGENCY STOP


actuated
A safety limit switch actuated

100 E2-Prom on CPU board 24S1 defective / not present

101 E2-Prom on CPU board data loss


Incorrect checksum

Siemens AG RXD3-209.071.01 Page 23 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 24 General Information

Replacing the 24S1 CPU Board and the 24S2 I/O Board 1

24S1 CPU Board 1

After a replacement of the CPU board 24S1, the E-Prom1 and E-Prom 2 and
EE-Prom U83 from the board that has been removed must be taken over.
Turn the unit on and on the I/O board 24S2, set the DW 1/8 switch to ON. At the remote
control console type in code 117 using the keys P31, P32, P34 and P35. Store with key
P33 (see page 2 -1).
On the I/O board 24S2 set the DW 1/8 switch to OFF. When the initialization that follows
takes place, the switch positions of the DW1 programming switch are read in.
Carry out Clock-Test/Adjust in the PC-Mode (Hardware test and Settings), see page 1 - 12
respectively 1 - 15 following.

24S2 I/O Board 1

After a replacement of the I/O board 24S2, the unit configurations from the DW1 program-
ming switches must be taken over from the board that has been removed.
Turn the unit on and on the I/O board 24S2, set the DW 1/8 switch to ON. At the remote
control console type in code 117 using the keys P31, P32, P34 and P35. Store with key
P33 (see page 2 -1).
On the I/O board 24S2 set the DW 1/8 switch to OFF. When the initialization that follows
takes place, the switch positions of the DW1 programming switch are read in.

Changes to the DW1 DIP Switches on the 24S2 Board 1

If changes are made to the DW1 DIP switches, proceed as follows:


• Select the adjustment mode.
- 24S2 I/O board, DW 1/8 on ON.
- Make the changes to the switch positions, e.g. I.I. size, room height, DFR, inch-cm,
Zoom steps
- Set code 117 and save (see Page 2-1).
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 on OFF.
- When this is done, the changed switch positions are read in.

SIREGRAPH CF RXD3-209.071.01 Page 24 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 25
Table of DW1 Switch Positions (up to VC00A) 1

Roomheight
Raumhöhe 9" Imageintensfr. Fullformat Programming OFF
> 3265 mm 23 cm BV Vollformat Programmierung AUS
ON ON ON ON

OFF OFF OFF OFF


1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8

Roomheight
Raumhöhe 13" Imageintensfr. One Zoomstep Programming ON
max. 3000 mm 33 cm BV Eine Zoomstufe Programmierung EIN
ON ON ON ON

OFF OFF OFF OFF


1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8

Roomheight
Raumhöhe DFR ON Two Zoomsteps cm-Cassettes
max. 2850 mm DFR EIN Zwei Zoomstufen cm-Kassetten
ON ON ON ON

OFF OFF OFF OFF


1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8

Roomheight
Raumhöhe DFR OFF Three Zoomsteps Inch-Cassettes
max. 2600 mm DFR AUS Drei Zoomstufen Zoll-Kassetten
ON ON ON ON

OFF OFF OFF OFF


1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8

Siemens AG RXD3-209.071.01 Page 25 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 26 General Information
Table of DW1 Switch Positions (as from VC01A) 1

Not used 9" Imageintensfr. Fullformat Programming OFF


Nicht verwendet 23 cm BV Vollformat Programmierung AUS
ON ON ON ON

OFF OFF OFF OFF


1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8

Not used 13" Imageintensfr. One Zoomstep Programming ON


Nicht verwendet 33 cm BV Eine Zoomstufe Programmierung EIN
ON ON ON ON

OFF OFF OFF OFF


1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8

Not used Not used Two Zoomsteps cm-Cassettes


Nicht verwendet Nicht verwendet Zwei Zoomstufen cm-Kassetten
ON ON ON ON

OFF OFF OFF OFF


1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8

Code ss211ss mit / with DFR


Codesiiiis212 s DFR Vorzugslage / Default
ss
Codesis213 s ohne / without DFR

Not used Not used Three Zoomsteps Inch-Cassettes


Nicht verwendet Nicht verwendet Drei Zoomstufen Zoll-Kassetten
ON ON ON ON

OFF OFF OFF OFF


1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8 1 2 34 5 6 7 8

Code für Raumhöhe / Code for room height Siehe auch Blatt 2-1, Justagemodus
Code 201 ssss 340 cm (keine Begrenzung / no limitation) See also page 2-1, adjustment mode
Code 202 ssss 330 cm
Code 203 ssss 320 cm
Code 204 ssss 310 cm
Code 205 ssss 300 cm
Code 206 ssss 290 cm
Code 207 ssss 280 cm
Code 208 ssss 270 cm
Code 209 ssss 260 cm

SIREGRAPH CF RXD3-209.071.01 Page 26 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 27

Listing of all Programming Codes 1

Adjustment of: DIP switch Code Remark


Room height, up to VC00A
2600 mm 1 Off, 2 Off 117 up to VC00A
2850 mm 1 On, 2 Off 117 up to VC00A
3000 mm 1 Off, 2 On 117 up to VC00A
>3260 mm 1 On, 2 On 117 up to VC00A
I.I. size
33 cm 3 Off 117
23 cm 3 On 117
I.I. Zoom
Full format 4 Off, 5 Off 117
1 Zoom 4 On, 5 Off 117
2 Zoom 4 Off, 5 On 117
3 Zoom 4 On, 5 On 117
DFR selection, up to VC00A
ON 6 Off 117 up to VC00A
OFF 6 On 117 up to VC00A
CM-inch format
cm 7 Off 117
Inch 7 On 117
Tilt drive
-18° 11 Basic setting
+90° 12
Column drive
Left side 21 30 mm to the end
Right side 22 30 mm to the end
Spotfilm device drive
Left side 31 105 mm to the end
Right side 32 105 mm to the end
SID drive
Bottom position 41 set at 45°
Top position 42 set at 45°

Siemens AG RXD3-209.071.01 Page 27 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 28 General Information

Adjustment of: DIP switch Code Remark


Collimator
Width closed 51
Width open 52
Height closed 61
Height open 62
Collimator plates
Closed 71 min = 41 mm
Open 72 max = 418 mm
Format sensing
Width, min 81 197 mm
Width, max 82 467 mm
Height, min 91 193 mm
Height, max 92 471 mm
Cassette carriage drive
Load position 101 198 mm
Park position 102 584 mm

Collimator correction
For cassette width 121-129 min-max
For cassette height 131-139 min-max
Collimator correction
For I.I. width 141-149 min-max
For I.I. height 151-159 min-max
Tilt down 87°- 90°, beginning with VC01A
90° 161 90° basic setting must first
be performed.
89° 162
88° 163
87° 164
Tomo height correction, beginning with VB00D
Reduction 171-174 1 mm steps
No correction 175
Increase 176-179 1 mm steps

SIREGRAPH CF RXD3-209.071.01 Page 28 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
General Information 1 - 29

Adjustment of: DIP switch Code Remark


Subdivision gap width, beginning with VB00D
Gap smaller 181-184 0.2 mm steps
No correction 185
Gap larger 186-189 0.2 mm steps
SID blocking, beginning with VB00D
No blocking 191
SID blocking >135 192 Effective beginning at 45°
SID blocking >115 193 Effective beginning at 45°
Room height, beginning with VC01A
>3400 mm 201 beginning with VC01A
3300 mm 202 beginning with VC01A
3200 mm 203 beginning with VC01A
3100 mm 204 beginning with VC01A
3000 mm 205 beginning with VC01A
2900 mm 206 beginning with VC01A
2800 mm 207 beginning with VC01A
2700 mm 208 beginning with VC01A
2600 mm 209 beginning with VC01A
DFR selection, beginning with VC01A
With DFR 211 beginning with VC01A
DFR default 212 beginning with VC01A
Without DFR 213 beginning with VC01A
Grid
Classic 221
oscillating grid
Removable 222 FL-TOP or
multi-line grid FL-COMPACT
Download EE-Prom values
Default values 307 Should not be performed
because the values are not
exact. Afterwards, all drives
must be readjusted.

Siemens AG RXD3-209.071.01 Page 29 of 30 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
1 - 30 General Information
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 30 of 30 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Adjustment Preparations 2 2-1

Fig. 1

Adjustment mode 2

Several buttons and the 3 on the 7-segment display of the remote control console are the
input elements and input display for the adjustment mode (Fig. 1).

Start adjustment mode 2

Set DIP switch DW 1/8 on the I/O board 24S2 to "ON".


With selection of the adjustment mode, you cannot control any movements from the
remote and tableside control consoles.
000 appears in the display.
Before each adjustment, create the mechanical conditions and enter a 3-digit code.
The following buttons are used for the adjustment mode:
P31 Adjustment Unit only at 1/ for > 120 steps of one downwards

P32 Adjustment Unit only at 2/ for > 120 steps of one upwards

P34 Adjustment upwards Steps of ten upwards

P35 Adjustment downwards Steps of ten downwards

P33 Storage

P48 Reset

Siemens AG RXD3-209.071.01 Page 1 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
2-2 Adjustment Preparations
Switching off the adjustment mode 2

Set the DIP switch DW 1/8 on the 24S2 I/O board to the OFF position. When this is done,
an automatic reset takes place and the data/values that have been set are read in.
However the reset key on the 24S1 CPU board can also be pressed or the unit can be
switched off and back on again.

NOTICE Pressing the reset button PL1 on the CPU board 24S1 will only
bring about a reset exclusively on this board; however, no reset
will take place on the CPU board 1S2 in the remote control
console.
A comprehensive reset will be achieved only with unit OFF and
ON again.

The following short version is used for the adjustment steps.


Example:
• Mechanical initial position:
• Select adjustment mode
- I/O board 24S2, DW 1/8 to ON
- Set 011 and store (see 2-1)
• Exit the adjustment mode.
- I/O board 24S2, DW 1/8 to OFF.
- This allows the adjustment data to be taken over.

SIREGRAPH CF RXD3-209.071.01 Page 2 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Adjustment Preparations 2-3

Checking the voltages 2

PC board Plug Pin Nominal value Tolerance


24S1 CN9 1/2 16 V ~ +/- 2 V ~
3/4 10 V ~ +/- 2 V ~
5/6 19 V ~ +/- 2 V ~
7/8 19 V ~ +/- 2 V ~
24S2 SC1 1/4 21 V ~ +/- 2 V ~
24S3 SC1 1/2 16 V ~ +/- 2 V ~
3/4 16 V ~ +/- 2 V ~
5/6 24 V ~ +/- 2 V ~
23S10 SC2 1/2 16 V ~ +/- 2 V ~
3/4 16 V ~ +/- 2 V ~
1S2 CN6 1/2 16 V ~ +/- 2 V ~
16 V ~ +/- 2 V ~
22AZ3 21/22 105 V ~ +8V~/-0V~
21/23 105 V ~ +8V~/-0V~
21/24 105 V ~ +8V~/-0V~
22AZ4 21/22 105 V ~ +8V~/-0V~
21/23 105 V ~ +8V~/-0V~
21/24 105 V ~ +8V~/-0V~
23S10 SC1 1/2 - 58 V = +/- 2 V =
3/4+
22S10 CN1 3/14 + 64 V = +/- 2 V =
15/16 -
22K1 1/3 240 V ~ +12 V ~ / - 0V ~
1/5 240 V ~ +12 V ~ / - 0V ~
3/5 240 V ~ +12 V ~ / - 0V ~

Siemens AG RXD3-209.071.01 Page 3 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
2-4 Adjustment Preparations
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Tilt Drive 3 3-1

Control overview 3

Tableside control panel 3S1

I/O PC board 24S2 CPU PC board 24S1

EMERGENCY STOP

Remote control console

Contactor
subassembly

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
3-2 Tilt Drive

Safety limit-
Potentiometer 3PT4 switch 3FC11 Tension screw

Brake assembly

3MT4

Poly-V-belt
Fig. 2

Tilt drive overview 3

Drive: 3MT4

Control: 24 S2

Breaker: 22k1
22k2

Potentiometer: 3PT4

Fuse: F26 6A/T


F27 6A/T 22TR1
F28 6A/T

Safety limit switch: 3FC11

SIREGRAPH CF RXD3-209.071.01 Page 2 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Tilt Drive 3-3

a
B

A
10
A

0,2

Fig. 3

Fig. 4 Fig. 5

Belt Tension, Adjustment Procedure 3

• If UI 003/96 has not yet been performed, i.e., the stop (1/Fig. 5) is not installed: move the
unit into the +60° to +70° position and move the exposure unit to max. head end. In this
position, the force on the tilt drive is low and a release on its own is not very likely.
• If UI 003/96 has been performed and the stop (1/Fig. 5) is installed: Raise unit to +90°.
• Loosen the four mounting screws (1/Fig. 3) on the motor support plate. At the adjustment
screw (Detail 2/Fig. 3), turn until the Poly-V belt is lightly tensioned. The motor
longitudinal axis must be parallel to the gear longitudinal axis. Tighten the mounting
screws (1/Fig. 3). Check: the outside surface of the belt pulley on the drive must be flush
with the level of the diameter reduction of the belt pulley on the motor (Fig. 4). If
necessary, correct the position of the belt pulley on the motor in the shaft long. direction.

Siemens AG RXD3-209.071.01 Page 3 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
3-4 Tilt Drive

a
B

A
10
A

0,2

Fig. 6

Brake assembly 3

Auxiliary device: Feeler gauge 0.2 mm

Adjustment conditions: Move the unit into +60° to +70° position and move the exposure
unit to max. head end. With the nuts (A/Fig. 6) set the distance
of all springs to
a = 10 mm. This setting corresponds to a brake momentum of
approx. 50 Nm.
With 60 Hz power supply the brake momentum must be approx.
60 Nm, which requires a stronger pretension of the springs.
To do this, mark the nuts (A/Fig. 6) with a waterproof pen (for
control purpose) and tighten them just one turn each.
Evenly adjust the gap of 0.2 mm + 0.1 mm between magnet and
brake disk using the nuts (B/Fig. 6).

Replacement: If UI 003/96 has not yet been performed, i.e., the stop (1/Fig. 5)
is not installed: move the unit into +60° to +70° position and
move the exposure unit to max. head end. In this position, the
force on the tilt drive is low and movement on its own is very
unlikely. Replace the brake.

If UI 003/96 has been performed and the stop (1/Fig. 5) is


installed: raise unit to +90°. Turn the flywheel of the tilt motor
3MT4, against the resistance of the brake until the stop
(1/Fig. 5) is securely against the unit base (2/Fig. 5). Replace
the brake. Turn the flywheel o the tilt motor 3MT4 in the oppo-
site direction as previously until the stop (1/Fig. 5) as previously
is no longer against the unit base and the limit switch 3FC11
(Fig.2 + 11) is no longer actuated.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Tilt Drive 3-5

36°

18°
506
120

143 506mm

Fig. 7 Fig. 8

Adjustment or replacement of potentiometer 3PT4 3

Auxiliary device: Spirit level

• Press EMERGENCY STOP:


Mechanical initial position: +36°, ± 1° (adjust manually at the flywheel of the
motor 3MT4)
Electronic initial position: +5V, ± 50 mV,
24S1 CPU board, TP2/1, TP2/13,14 (0V)

• Unlock EMERGENCY STOP

Siemens AG RXD3-209.071.01 Page 5 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
3-6 Tilt Drive

36°

18°
506
120

143 506mm

Fig. 9 Fig. 10

Set end positions 3

Auxiliary device: Spirit level

Trendelenburg position, -18°


• Move the spotfilm device all the way to the foot end.
• Move the unit into the Trendelenburg position, approx. -18°, press EMERGENCY STOP.
- Manually set the exact position at the rotation disk.
• Select the adjustment mode.
- On the 24S2 I/O board, set DW 1/8 to ON (up)
- Set Code 011 and save by pressing the P33 key.
• Disengage the EMERGENCY STOP switch.
• Exit the adjustment mode.
- On the 24S2 I/O board, set DW 1/8 to OFF (down).
• Move into the 90° vertical unit position, press the EMERGENCY STOP switch.
- Manually set the exact 90° position on the rotation disk.
• Select the adjustment mode.
- On the 24S2 I/O board, set DW 1/8 to ON (up).
- Set Code 012 and save the value by pressing the P33 key.
• Stay in the adjustment mode.

SIREGRAPH CF RXD3-209.071.01 Page 6 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Tilt Drive 3-7
• Find the tilt position in the table (see below) (per the customer’s request) at which the unit
is supposed to stop later during operation in the upright position.
- Set the appropriate code and save by pressing the P33 key.
• Disengage the EMERGENCY STOP switch.
• Exit the adjustment mode.
- On the 24S2 I/O board, set DW 1/8 to OFF (down).
• Switch off the system.

Code Tilt Position


161 90°
162 89°
163 88°
164 87°

• Move to 0° position and measure with the angle spirit level.


Required: 0°, ± 1°
Correction:
- Readjust end positions
• If the 0° position is not correct, readjust the end positions.

Siemens AG RXD3-209.071.01 Page 7 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
3-8 Tilt Drive

13
– 18° B

3FC11

22.5

3FC11 90° 13

Fig. 11

Safety limit switch 3FC11 3

Auxiliary device: Tape measure

Preadjustment of the trip cam


Unit position approx. +75° to 80°, spotfilm device
Adjustment condition
maximum to head end

Adjustment • Trip cam (A/Fig. 11) distance = 13, ± 1 mm.


• Adjust safety limit switch 3FC11 to dimension
22.5 mm.
• The safety limit switch must be fixed so that it will
properly hit the trip cam. However, the switch
must not collide with the fixing screws of the trip
cam.
• Trip cam (B/Fig. 11) distance = 13, ± 1 mm.

Visual inspection Move to both end positions. Check the distance to


the point where the switch should be activated:

Vertical unit position, (exactly +90°) 5 ±1 mm


Trendelenburg position – 18° 5 ± 1 mm

Movement times 3

Tilting speed = 6°/ sec.


The speed is not adjustable

SIREGRAPH CF RXD3-209.071.01 Page 8 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Tilt Drive 3-9

3F16

Fig. 12

Safety limit switch 3FC16 3

When the bottom switch is pressed (roller), switch 3FC16 positively actuate.
This can be adjusted by moving the switch strike plate (arrow/Fig. 12).

Siemens AG RXD3-209.071.01 Page 9 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
3 - 10 Tilt Drive

BV 9"

183 ±3

BV 13"

122 ±3

Checking the floor spacing with 0 degree table position 3

Nominal: Image intensifier format 9" (23 cm)


183 mm, ± 3 mm

Image intensifier format 13" (33 cm)


122 mm, ± 3 mm

Correction: Shifting the subassembly (A / drawing above) in the oblong holes.

SIREGRAPH CF RXD3-209.071.01 Page 10 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Spotfilm Device Longitudinal Drive 4 4-1

Control overview 4

Tableside control panel


3S1

I/O PC board 24S2 CPU PC board

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
4-2 Spotfilm Device Longitudinal Drive

3MT2
3MT2 3FC14
3DT2 3FC14 / 3FC5
3DT2

3PT2
3PT2 3FC15
3FC15

3FC16
3FC16

Fig. 2

Spotfilm device longitudinal drive overview 4

Drive: 3MT2

Control: 22Az4

Potentiometer: 3PT2

Tacho generator: 3DT2

Fuses: F23 10A/T


F24 10A/T 22TR1
F25 10A/T

Safety limit switch: 3FC14


3FC15

SIREGRAPH CF RXD3-209.071.01 Page 2 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Spotfilm Device Longitudinal Drive 4-3

Console 4

Joystick setting P44 (spotfilm device)


• Connect DVM
(+) 1S2 - U19 - 19
(-) 1S2 - U19 - 20
• Set P1 / 1S2
5V, ± 0.05 V
joystick in rest position
and
LED 12 and 13 on 24S2 "OFF"

19 20

Siemens AG RXD3-209.071.01 Page 3 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
4-4 Spotfilm Device Longitudinal Drive

105
875
A
X 105
105

B
B

105

Fig. 3

Adjustment or replacement of potentiometer 3PT2 4

Reference point: outside edge of spotfilm device housing and outside edge of table frame.
Mechanical initial position: Potentiometer play approx. 0.5 mm. Use the motor to
move into position 875 mm (X/Fig. 3), press EMER-
GENCY STOP and exactly adjust 3MT2 on the motor
shaft manually. Disengage EMERGENCY STOP.
Electronic initial position: +5V ± 50 mV
24S1 CPU board, TP2/4, TP2/13,14, 0V.

Set end positions 4

• Move to the left end position of approx. 105 mm, press EMERGENCY STOP.
- Manually set the exact position (on the motor axis).
• Select adjustment mode (see Page 2-1).
- 24S2 I/O board, DW 1/8 to ON.
- Set 031 and store.
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF.
- Release EMERGENCY STOP.
• Move to right end position of approx. 105 mm, press EMERGENCY STOP.
- Manually set the exact position (on the motor axis).
• Select adjustment mode (see above).
- Set 032 and store.
• Exit the adjustment mode (see above), release EMERGENCY STOP.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Spotfilm Device Longitudinal Drive 4-5

3FC14
X 3FC5

26, ±1 mm

Fig. 4

Safety limit switch 3FC14 4

With safety limit switch 3FC14 actuated, the drive must be moved out of the trip cam by
hand. For this, there is a shaft extension with Allen screw on motor 3MT2.
Remove cover (X/Fig. 4)
Adjustment: Adjust the same height of 26, ± 1 mm for 3FC14 and 3FC5. Move
the drive maximally towards head end. Screw down switch plate.
Distance of both trip cams to the contact surface approx. 1 mm
(visual inspection).

Checking the PC board SP 22Az4 4

CAUTION Do not readjust any potentiometers on this PC board.

For correct speed control, a 5110 Ohm resistor has been soldered on plug-in piggy-back
board "SP 22Az4", pin 9-12.
After a replacement of the "Maestro" assembly this plug-in piggy-back board "SP 22Az4"
from the "Maestro" assembly that has been removed must be taken over. If the piggy-back
board itself is to be replaced it must be ensured that a 5110 Ohm resistor is soldered on
the new board.
Movement times:
Speed continuously controllable up to approx. 10 cm/sec in the whole travel range.
The maximum speed is not adjustable.

Siemens AG RXD3-209.071.01 Page 5 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
4-6 Spotfilm Device Longitudinal Drive

1
1

1st ring
2nd ring
1

Fig. 5 Fig. 6

Fig. 7

Spotfilm device drive, adjusting belt tension 4

• An adjustment of the belt tension is required after a replacement of the transmission or


drive motor.
• Adjustment of tension is carried out with the help of the supplied measurement aid
(1/Fig. 5).
• Adjustment is possible only with the complete disassembly of the transmission.
• A toothed belt is used as the belt for the spotfilm device drive.

SIREGRAPH CF RXD3-209.071.01 Page 6 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Spotfilm Device Longitudinal Drive 4-7
Adjustment procedure 4

• Loosen the four screws (1/Fig. 6).


• Place on the measurement aid and press on it (X/Fig. 6)
Requirement: 1st ring must touch the top edge of the window.

Correction:
- By turning the adjustment screw (1/Fig. 7)
- Turning counterclockwise: the belt is tightened
- Turning clockwise: the belt is loosened
• After adjustment is completed, tighten screws (1/Fig. 6).
• Install the drive.

NOTICE The 2nd ring is used only for factory adjustment of the measure-
ment aid.

Siemens AG RXD3-209.071.01 Page 7 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
4-8 Spotfilm Device Longitudinal Drive
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 8 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Column Drive 5 5-1

Control overview 5

Tableside control panel


3S1

I/O PC board 24S2 CPU PC board

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
5-2 Column Drive

back of column, 3FC12

3MT1,
3MT13DT1 3DT1
3PT1
3PT1
tripSchaltnocke
cam 3FC5
3FC5 trip cam
3FC12 Schaltnocke
3FC14

3FC5 3FC5
3FC12
14 14
3FC12

Fig. 2

Column drive overview 5

Drive: 3MT1

Control: 22Az3

Potentiometer: 3PT1

Tacho generator: 3DT1

Fuses: F20 10A/T


F21 10A/T 22TR1
F22 10A/T

Safety limit switch: 3FC12


3FC5

SIREGRAPH CF RXD3-209.071.01 Page 2 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Column Drive 5-3

2340

980 B
A
X
30,+3mm 30,+3mm

Fig. 3

Adjustment and replacement of potentiometer 3PT1 5

Reference point: outside edge of the table frame.


Mechanical initial position: Use the motor to move into position 980 mm (X/Fig. 3),
press EMERGENCY STOP and exactly adjust 3MT1 on the
motor shaft manually. Disengage EMERGENCY STOP.
Electronic initial position: +5V, ± 50 mV
24S1 CPU board, TP2/3-TP2/13,14: 0V

• Move to column end position A and oblique projection head-end, press EMERGENCY
STOP, manually set exact position (on motor axis).
• Select adjustment mode (see Page 2-1).
- 24S2 I/O board, DW 1/8 to ON.
- Set 021 and store.
• Release EMERGENCY STOP
• Exit adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF.
• Move to column end position B and oblique projection foot-end, press EMERGENCY
STOP, manually set exact position (on motor axis).
• Select adjustment mode (see Page 2-1.)
- Set 022 and store.
• Release EMERGENCY STOP
• Exit adjustment mode.
- 24S2 I/O board, DW1/8 to OFF

Siemens AG RXD3-209.071.01 Page 3 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
5-4 Column Drive

3FC12

back of column, 3FC12


3MT1,
3MT13DT1 3DT1 29, ±1mm
3PT1
3PT1
3FC5
trip cam
Schaltnocke 3FC12 trip cam
Schaltnocke
3FC14

3FC5 3FC5
14
3FC12 14
3FC12

3FC14

3FC5

26 ± 1 mm

Fig. 4

Safety limit switch 3FC12 5

With safety limit switch actuated, the drive must be moved out of the trip cam by hand. For
this, there is a shaft extension with Allen screw on motor 3MT1.

Blocking limit switch 3FC5 5

This is actuated if the tube is pivoted towards the head end or foot end and if the longitudi-
nal carriage movement has nearly reached the end position.
Upon actuation, the swivel movement is blocked in one direction.
This is actuated also if the tube is pivoted towards the head end or foot end and if the lon-
gitudinal carriage movement is controlled in the direction of the end position.
Upon actuation, the longitudinal carriage movement is blocked in one direction.

Adjustment: Adjust the same height of 26, ± 1 mm for 3FC14 and 3FC5.
Move the drive towards head end. Screw down switch plate. Dis-
tance of both trip cams to the contact surface approx. 1 mm (visual
inspection).

SIREGRAPH CF RXD3-209.071.01 Page 4 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Column Drive 5-5

C+
1, ±0.5

B=47.5 3FC15

A=0.5mm

B=47,5

1, ±0.5
C+

Fig. 5

Safety limit switch 3FC15 5

Initial position: Set unit to 90°.


Column vertical (centered).
Spotfilm device in the center of the travel range.

Adjustment: Adjust switch roller to A = approx. 0.5 mm.


Adjust trip cam to center B = 47.5.

Check: Move column towards head end and foot end.


Max. oblique projection and SID = 115 cm.
Gap C between switch roller and trip cam = 1, ± 0.5 mm.

Siemens AG RXD3-209.071.01 Page 5 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
5-6 Column Drive

3FC13

Fig. 6

Checking the PC board SP 22Az3 5

CAUTION Do not adjust any potentiometers on this PC board.

For correct speed control, a 9090 Ohm resistor has been soldered on plug-in piggy-back
board "SP 22Az3", pin 9-12.
After a replacement of the "Maestro" assembly this plug-in piggy-back board "SP 22Az3"
from the "Maestro" assembly that has been removed must be taken over. If the piggy-back
board itself is to be replaced it must be ensured that a 9090 Ohm resistor is soldered on
the new board.

Movement times:
Speed = 4°/sec.
The speed is not adjustable.

Room height switch 3FC13 5

Check: Press the "antenna" (Fig. 6) manually, switch has to switch as


soon as possible.

Adjustment: See (Detail/Fig. 6).

SIREGRAPH CF RXD3-209.071.01 Page 6 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Column Drive 5-7

1 1

1st ring
2nd ring
1

Fig. 7 Fig. 8

Fig. 9

Column drive, adjusting the belt tension 5

Adjustment of the belt tension is required after replacement of the transmission or drive
motor.
Adjustment of belt tension is carried out using the supplied measurement device
(1/Fig. 7).
Adjustment is possible only when the complete drive is disassembled.
For the column drive, the belt is comprised of several individual or one Poly-V belt. The
adjustment procedure is the same.

Siemens AG RXD3-209.071.01 Page 7 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
5-8 Column Drive
Adjustment procedure 5

• Loosen the four screws (1/Fig. 8).


• Place on the measuring aid as shown in (X/Fig. 8) and press down on it.
Requirement: 1st ring must touch the top edge of the window.

Correction:
- By turning the adjustment screw (1/Fig. 9)
- Turning counterclockwise: the belt is tightened.
- Turning clockwise: the belt is loosened.
- After adjustment is completed, tighten the screws (1/Fig. 8).
• Install drive.
• Adjust potentiometer 3PT1 as described on Page 3.

NOTICE The 2nd ring is used only for factory adjustment of the measuring
aid.

SIREGRAPH CF RXD3-209.071.01 Page 8 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Compression Drive 6 6-1

Control overview 6

I/O PC board 24S2 CPU PC board 24S1

Remote control console

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
6-2 Compression Drive

Fig. 2

Compression drive overview 6

Drive: 3MT6

Actuation: 24S2

Control: 23Az6

Fuse: F3 10A/T 23Az6

Limit switch: 3FC2 Cone above


3FC3 Cone inserted
3FC4 Cone force > 20 - 30N

SIREGRAPH CF RXD3-209.071.01 Page 2 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Compression Drive 6-3

Tomographic rod A

f = 15 mm 1

420
Chain F

Fig. 3

Chain tension 6

Auxiliary device: Spring balance

Adjustment conditions: Set unit at approx. 45°, exposure system approx. in the
middle of travel range.
Oblique projection max. towards foot end, spotfilm
device max. towards head end.

Adjustment: Loosen two Allen screws at the end of the tomographic


rod (1/Fig. 3). Adjust the screw (A/Fig. 3) so that the fol-
lowing values are obtained:
F = 7, ± 0.7 N
f = 15 mm
Tighten the two Allen screws (1/Fig. 3).

Siemens AG RXD3-209.071.01 Page 3 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
6-4 Compression Drive

P2

F 23AZ 6

Fig. 4 Fig. 5

Maximum compression force 6

Auxiliary device: Spring balance and two cable loops

Adjustment conditions: Set unit horizontally


Set column vertically
Push Stop button

SIREGRAPH CF RXD3-209.071.01 Page 4 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Compression Drive 6-5

C
E

B
D

Spring balance

Fig. 6

• Remove cover "A".


• Attach spring balance as shown.

Siemens AG RXD3-209.071.01 Page 5 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
6-6 Compression Drive

Electrical preadjustment: Adjust potentiometer (P2/Fig. 5) to maximum value,


turn it to the left.

NOTICE The compression force is measured via the armature current of


the compression motor 3MT6 and is directly proportional.

Mechanical adjustment: Release EMERGENCY STOP


Move compression down and read spring balance.
Requirement: 200 N ± 30
Correction:
Adjust slip clutch (Detail/Fig. 6)
Bend the safety clip (C) out of the adjustment nut (B)
Use the hook wrench to turn the adjustment nut
Clockwise turning: force is increased
Counterclockwise turning: force is reduced

Secure adjustment screw.

Electrical adjustment: Move compression down.


Adjust the potentiometer (P2/Fig. 5) (turn clockwise)
until the controller switches off the drive at a force of
F = 150, +10N.

SIREGRAPH CF RXD3-209.071.01 Page 6 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Compression Drive 6-7

Fd.

150mm

Fc.

Fig. 7

Adjustment of switch 3FC4 6

Auxiliary device: Spring balance

Adjustment conditions: Set unit horizontally


Set column vertically

Adjustment: maximum cone force Fc = 20, ± 2N


Attach the spring scale as indicated (Fig. 7) and pull it
with a force of Fd = 50N, switch 3FC4 triggers LED L68
on I/O 24S2.

Correction: Set initial tension of the spring with the two horizontal
set screws, (arrows) see detail.

Movement times 6

• The times are not adjustable.


• Whole excursion range of the compression carriage: s = 630, ± 5 mm
• Movement time in both directions with each unit position: t = 12 sec, ±1 sec

Siemens AG RXD3-209.071.01 Page 7 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
6-8 Compression Drive
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 8 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
SID Drive 7 7-1

Control overview 7

Tableside control console


3S1

I/O PC board 24 S2 CPU PC board 24S1

Remote control console

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
7-2 SID Drive

Fig. 2

SID drive overview 7

Drive: 3MT7

Actuation: 24S2

Control: 23Az7

Potentiometer: 3PT7

Fuse: F3 10A/T 23Az7

Safety limit switch: 3FC7

SIREGRAPH CF RXD3-209.071.01 Page 2 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
SID Drive 7-3

350

175

1 3FC7
3PT7

3TP1

Fig. 3 Fig. 4

Adjustment or replacement of potentiometer 3PT7 7

• Set column vertically.


Set unit position to approx. 45° (see Notice).

Mechanical initial position: Adjust 175 mm (see Fig. 3).


Disengage the motor gears.
(Loosen screws (1/Fig. 4)).
Electronic initial position: +5V, ±50 mV.
CPU PC board 24S1, TP2/5 - TP2/13,14 0V.

NOTICE At a table position of approx. 45 - 60° there is an approximate


balance between gas spring force and telescopic tube weight.
The required distances can be adjusted easily and the potentio-
meter can be easily replaced and adjusted.

Siemens AG RXD3-209.071.01 Page 3 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
7-4 SID Drive
Set end positions 7

• Disengage the motor gear. Loosen the screws (1/Fig. 4).


• Manually set the end position at 350 mm (SID = 150 cm) and hold it.
• Select adjustment mode (see Page 2-1)
- 24S2 I/O board, DW 1/8 to ON.
- Set 042 and store.
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF.
• Manually set end position at 0 mm (SID = 115 cm).
• Select adjustment mode (see Page 2-1)
- 24S2 I/O board, DW 1/8 to ON
- Set 041 and store.
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF.
• Reengage the motor gear without play and tighten the screws.

Adjustment of SID Limiting at 90° 7

SID programming is possible beginning with software version VB00D

NOTICE When shipped, SID limiting is not programmed.

• Select SID 115 cm.


• Tilt unit up to 90°.
• Measuring the distance from the safety limit switch bracket to the wall.
• Program the SID accordingly while maintaining the safety distance of 50 cm.
• Select the adjustment mode, see Page 2-1, (I/O board 24S2, DW 1/8 set to ON).

Select the following code with keys P34/35:


• 191, no SID limiting
• 192, SID limiting at 135 cm
• 193, SID limiting at 115 cm

• Save the particular code with the P33 key.


• Deselect the adjustment mode, (I/O board 24S2, DW 1/8 set to OFF).

The particular limiting is effective beginning at a unit position of 45°.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
SID Drive 7-5

Focus
72 plane

3FC7
68
Film plane

Fig. 5 Fig. 6

Safety limit switch 3FC7 7

• With safety limit switch actuated, please proceed as follows:


• Mark the position of the switch and loosen it.
• Reset switch until this is actuated by the trip cam.
• Move out of the range with motor force.
• Move safety limit switch to its initial position and fasten.

Distance measurement:
SID = 115 cm X = 1010 mm, ±5 mm

SID = 150 cm X = 1360 mm, ±5 mm

Siemens AG RXD3-209.071.01 Page 5 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
7-6 SID Drive

CN1

F1 F3
J3
P6

F2
P5

P4 P3

J2
J1 P1 BC1
P2 P3
23AZ7

L2 L1

Fig. 7

Adjustment of PC board 23Az7 7

CAUTION On this PC board, only potentiometer P3 may be readjusted.


Maximum current after run-up time, approx. 6A. The potentiome-
ter is located on PC board 23Az8. When the board is replaced,
approx. the same adjustment as the P3 should be taken over.

Oscilloscope: Probe 23 Az7 board, Chip U1, Pin 6, Ground ⊥ to J2


Time: 0.2 sec/cm
Amplitude: 1 V/cm
Nominal: Set brake force at P1 (Fig. 7) so that braking takes place softly, how-
ever, the telescope must be held securely. SID movement with table
vertical and horizontal must run smoothly.

Movement times 7

from 115 cm to 150 cm: 5.8, +1 sec, with all unit positions.

from 150 cm to 115 cm: 5.8, +1 sec, with all unit positions.

SIREGRAPH CF RXD3-209.071.01 Page 6 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Table Transverse Drive 8 8-1

Control overview 8

Tableside control panel


3S1

I/O PC board 24 S2 CPU PC board 24 S1

Remote control console

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
8-2 Table Transverse Drive

Mechanical
Mechanischer limit stop
Endanschlag
LimitEndschalter
switch Drive
Antrieb 3MT8
3FC8
3FC9

Spotfilm device
Zielgerät

Steel Counterweight Carriage


cable Gegengewichtswagen
Seilrolle Seil Steel
Seilrolle cable
pulley steel ca- pulley
ble

Fig. 2

Table transverse drive overview 8

Drive: 3MT8

Actuation: 24S2

Potentiometer: 23Az8

Control: 23Az8

Fuse: F3 10A/T 23Az8

Switch: 3FC8 (tower side)


3FC9 (doctor’s side)

SIREGRAPH CF RXD3-209.071.01 Page 2 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Table Transverse Drive 8-3

CN1

F1 F3
J3
P6

F2
P5

P4 P3

J2
J1 P1 BC1
P2 P3
23AZ8

L2 L1

Fig. 3

Electrical Adjustment 8

CAUTION On this PC board, only potentiometer P3 may be readjusted.


Maximum current after run-up time, approx. 6A.
The potentiometer is located on PC board 23Az8.

• Tabletop horizontal.
• Actuate movement on doctor’s side
Nominal: 180 N, +100 N

Correction:
- Potentiometer P3 on 23Az8

NOTICE This force is adjusted at the factory; a device to hook on the


spring balance is available for service purposes.
When the board is replaced, approx. the same adjustment as the
P3 should be taken over.

Siemens AG RXD3-209.071.01 Page 3 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
8-4 Table Transverse Drive
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Cassette Carriage Drive 9 9-1

Control overview 9

Tableside control panel


3S1
I/O PC board 24 S2 CPU PC board 24 S1

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
9-2 Cassette Carriage Drive

Tableside control panel

Fig. 2

Cassette carriage drive overview 9

Drive: 3MT3

Tacho generator: 3DT3

Actuation: 24S2

Control: 22Az2

Potentiometer: 3PT3

Fuses: F14 4A/T


F15 4A/T 22TR 1
F16 4A/T
F 12A/T 22S10

SIREGRAPH CF RXD3-209.071.01 Page 2 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Cassette Carriage Drive 9-3

2
A
Fig. 3

Mechanical check 9

• Set unit in horizontal position.


• Remove table top (see chap. 16-1).
• Unscrew guide rail (1/Fig. 3).
• Pull out cassette carriage maximally at doctor’s side.
• Disconnect the cassette carriage from the chain (2/Fig. 3).
• Move cassette carriage over the entire range.
Required: easy movement and free of play.
Correction:
- Adjustment of the bearings on cassette carriage (drawing).

• Set unit in vertical position, column approx. middle of travel range.


➪ Adjust the head end (upper) bearings (3/Fig. 3)
➪ Adjust the foot end (lower) bearings (4/Fig. 3). Access for the screwdriver is only
possible through the hole at (A/Fig. 3).
• It must be possible to still turn one roller manually.
• Set unit in horizontal position.
• Attach cassette carriage to the chain (2/Fig. 3).
• Attach guide rail (1/Fig. 3).
The adjustment takes place after the adjustment of the parking and loading position (see
chap. 9-8).

Siemens AG RXD3-209.071.01 Page 3 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
9-4 Cassette Carriage Drive

f
F
f

860

Fig. 4 Fig. 5

Adjust tension of the drive chain 9

Auxiliary device: Spring balance.

Adjustment conditions: Set unit in vertical position.

Adjustment: see Fig. 4 F = 20, ± 2N


f = 15 mm

Adjust belt tension of the cassette drive 9

Auxiliary device: Spring balance

Adjustment conditions: Set unit in vertical position

Adjustment: see Fig. 5 F = 5, ± 0.5N


f = 5 mm

SIREGRAPH CF RXD3-209.071.01 Page 4 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Cassette Carriage Drive 9-5

Fig. 6

Adjusting the slip clutch 9

Aids: Spring balance 200 N and attachment parts

Preparations: • Cassette carriage without cassette in parking position


• Switch "Allarmax" off (plug in AMP3 / 24S2 I/O board)
• Press Emergency Stop button
• Attach spring balance and attachment parts as shown in
page 9 - 6

Measurement: • Unlock Emergency Stop button


• Read spring balance
Nominal: 170 N ± 20 N

Correction: - Adjust the slip clutch (A/Fig. 6) + Page 9-6


- Bend the safety clip (D) out of the adjustment nut (C).
• Use the hook wrench to turn the adjustment nut
Turn clockwise: force is increased
Turn counterclockwise: force is reduced
• Secure the adjustment nut.

Finally: • Switch on "Allarmax", (pull out AMP3 on 24S2 I/O board)

Siemens AG RXD3-209.071.01 Page 5 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
9-6 Cassette Carriage Drive

SIREGRAPH CF RXD3-209.071.01 Page 6 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Cassette Carriage Drive 9-7

Cassette carriage
Loading position Exposure position Park position

3PT3 X

C
A 193 B
198 584
Spotfilm device

Fig. 7 Fig. 8 Spotfilm device with cassette carriage

Adjustment or replacement of potentiometer 3PT3 9

• Set unit in vertical position.


• Press EMERGENCY STOP.
• Mechanical initial position: Pos. C = exposure position
C=193 mm (Fig. 8)
cassette carriage front edge ==> front edge of the
bottom sheet of the spotfilm device (X/Fig. 8).

• Electrical initial position: +5V, ± 50 mV


CPU-PC board 24S1, TP2/12 - TP2/13,14: 0V.

Siemens AG RXD3-209.071.01 Page 7 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
9-8 Cassette Carriage Drive

Cassette carriage
Loading position Exposure position Park position

C
A 193 B
198 584
Spotfilm device

Fig. 9 Spotfilm device with cassette carriage

Set end positions 9

Auxiliary device: Tape measure.

• Set distance:
Parking position (B/Fig. 9) - Distance cassette carriage front edge ==> front
edge of the bottom sheet of the spotfilm device (X/Fig. 9).

B = 584 mm Code = 102

• Select adjustment mode (see page 2-1).


- I/O PC board 24S2, DW 1/8 to ON.
- Enter code 102 and store.
• Stay in the adjustment mode
• Set distance:
Loading position (A/Fig. 9) - Distance cassette carriage front edge ==> front
edge of the spotfilm device (X/Fig. 9).

A = 198 mm Code = 101

• Enter code 101 and store


• Exit the adjustment mode
- 24S2 I/O board, DW 1/8 to OFF.
• Release EMERGENCY STOP.

SIREGRAPH CF RXD3-209.071.01 Page 8 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Cassette Carriage Drive 9-9

Fig. 10

Movement times 9

The times are not adjustable.


Movement time parking position - exposure position t ≤ 0.9sec.
Fast series
Required: t ≤ 2sec

- Cassette 35 cm x 43 cm (transverse).
- 4 on 1 subdivision.
- Without object.

Maximum opening of the cassette jaws 9

• Adjust the guide rail (1/Fig. 10) so that a 43 cm cassette can be inserted. It must be
possible for the cassette carriage to reliably reach the loading position. The maximum
tension jaw height opening is 467 mm, ± 1 mm. If the tension jaws are opened too soon,
error 010 is signaled.

Siemens AG RXD3-209.071.01 Page 9 of 10 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
9 - 10 Cassette Carriage Drive
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 10 of 10 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Format Sensing 10 10 - 1

Control overview 10

Tableside control panel


3S1

I/O PC board 24 S2 CPU PC board 24 S1

CPU PC board 24S1

EMERGENCY STOP

"Width" "Height"
Remote control console

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
10 - 2 Format Sensing

3PT11
"Höhe"
"Height"

Spring
Feder

Loading
Ladeposition
3PT10 position
Scissors ik
Scherenmechan "Breite"
"Width"
mechanism

Fig. 2

Overview of the format sensing drive 10

Color coding on Potentiometers 3PT10 and 3PT11

M (2)

Potentiometers on the connection side:


M (2) = middle
L.D. (3) = left side down

R.D. (1) R.D. (1) = right side down


L.D. (3)

3PT10 3PT11
Middle (2): green Middle (2): green
Left side down (3): white Left side down (3): brown
Right side down (1): brown Right side down (1): white

SIREGRAPH CF RXD3-209.071.01 Page 2 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Format Sensing 10 - 3

Fig. 3

Adjustment or replacement of potentiometers 3PT10, 3PT11 10

Remove table top and cover of spotfilm device (see chap.16 and chap.17).

NOTICE Potentiometers 3PT10 and 3PT11 are heat-sensitive!

Cassette jaws
• Press EMERGENCY STOP.
Mechanical initial position (Fig. 3)

- Cassette jaw width, potentiometer 3PT10,


Spacing of the cassette jaws C = 332 mm (spacer gauge, see page 10 - 4).
Electrical initial position = + 5 V, ± 50 mV.
24S1 CPU board TP2/10 - TP2/13,14: 0V.

- Cassette jaw height, potentiometer 3PT11,


Spacing of the cassette jaws C = 332 mm (space gauge, see page 10 - 4).
Electrical initial position = + 5 V, ± 50 mV.
24S1 CPU board TP2/11 - TP2/13,14: 0V.

• Release EMERGENCY STOP.

Siemens AG RXD3-209.071.01 Page 3 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
10 - 4 Format Sensing

A
A C
E

Loading position
Loading position
Park position

Park position
B D C B

Spacer Gauge 10

A = 471 mm

D = 197 mm
C = 332 mm

The spacer gauge makes it easier to mechanically adjust the cassette jaw spacing.
It must be placed between the cassette jaws so that it is squared respectively straight.
If this is not done correctly, it can snap together very easily.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Format Sensing 10 - 5
Adjusting the end positions and the format sensing 10

Auxiliary equipment: Spacer gauge (see page 10 - 4)

• Move cassette carriage to loading position and adjust the cassette jaws as follows:
Width A = 471, ±1 mm (spacer gauge)

• Select adjustment mode (see Page 2-1).


- 24S2 I/O board, DW 1/8 to ON.
- Enter code 082 and store.
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF.

• Move cassette carriage to loading position and adjust the cassette jaws as follows:
Height B = 467, ±1 mm (cassette carriage is empty, maximum opening of the jaws)

• Select adjustment mode (see Page 2-1).


- 24S2 I/O board, DW 1/8 to ON.
- Enter code 092 and store.
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF.

• Move cassette carriage to parking position and adjust the cassette jaws as follows:
Height D = 197, ±1.5 mm (spacer gauge)

• Select adjustment mode (see Page 2-1).


- 24S2 I/O board, DW 1/8 to ON.
- Enter code 091 and store.
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF.

• Move cassette carriage to parking position and adjust the cassette jaws as follows:
Width E = 193, ±1.5 mm (cassette carriage empty, mechanical stop in park position)

• Select adjustment mode (see Page 2-1).


- 24S2 I/O board, DW 1/8 to ON.
- Enter code 081 and store.
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF.

Siemens AG RXD3-209.071.01 Page 5 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
10 - 6 Format Sensing
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 6 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Diaphragm Plate Drive 11 11 - 1

Control overview 11

I/O PC board 24S2 CPU PC board 24S1

CP9

PC board 22 S1

EMERGENCY STOP

Collimator
control

diaphragm
Collimator Remote control
plates
workstation console

3PT5

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
11 - 2 Diaphragm Plate Drive

Diaphragm plates
Blendenplatten
ballKugellager
bearing

chain
Kette

3MT5
control panel
Bedienfeld chain
Kette

3PT5

Brass con-
Messing-
chain
Kette necting rod
Schubstange
guide
Führung

Fig. 2

Diaphragm plate drive overview 11

Drive: 3MT5

Actuation: 24S2

Control: 23Az5

Potentiometer: 3PT5

Fuse: F3 10A/T 23Az5

F17 6A/T
F18 6A/T 22TR1
F19 6A/T

SIREGRAPH CF RXD3-209.071.01 Page 2 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Diaphragm Plate Drive 11 - 3

A C
Head Foot 5 4 2

end B D end
6

7
Fig. 3 Fig. 4

Mechanical check 11

Remove table top and spotfilm device cover (see chap.16 and chap.17).
Auxiliary device: Tape measure.

Adjustment conditions: Manually shift diaphragm plates


evenly and free of play.

Adjustment: Adjustment of the bearings.


Have the rails been greased? (Longtime PD2)
Have the chains been greased? (Longtime PD2)

Symmetry and centering 11

Auxiliary device: Tape measure, flat spanner 8 mm (in the service kit),
put up a wrap wire or thread in center of spotfilm device
and tighten.

Adjustment conditions: A = B = C = D (Fig. 3)

Adjustment: Start to adjust one diaphragm plate (1/Fig. 4): loosen


the screws (6/Fig. 4), open the eccenter (5/Fig. 4),
adjust parallelism using the eccenter (4/Fig. 4). Then
adjust the eccenter (5/Fig. 4), at least one bearing must
be turnable. Tighten the screws (6/Fig. 4).
Proceed with the second diaphragm plate in the same
manner. The bearings (7/Fig. 4) are no eccenters.

Siemens AG RXD3-209.071.01 Page 3 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
11 - 4 Diaphragm Plate Drive

Chain tension 11

415
Auxiliary device:
A A
Spring balance
F

Adjustment conditions:
Open diaphragm plates (A/Fig. 5) to
dimension 415 mm, ±1 mm. f

Adjustment: 560

Adjust chain tension.


F = 15 ± 8 N
f = 15 mm
Fig. 5

Adjust the slip clutch 11

Aids: Spring balance 200N and attachment parts

Preparations: Cassette carriage without cassette in Park Position.


Switch off "Allarmax" (plug in AMP3 on 24S2 I/O board)
Press the Emergency Stop button

• Attach the spring balance and attachment parts at "B" and at "A" to the diaphragm plates
as shown in the drawing on page 11-5 and (Fig. 6).
- Unlock the Emergency Stop button.
- Select Bucky mode, select full format
- Read spring balance
Nominal: 40, ± 10 N

Correction:
- Bend locking tab D out from adjusting nut C
- Adjust adjusting nut C with hook wrench
clockwise: force increases
counterclockwise: force decreases
- Press a locking tab D into adjusting nut C.
Switch on "Allarmax" ! (pull out AMP3 on 24S2 I/O board !)

SIREGRAPH CF RXD3-209.071.01 Page 4 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Diaphragm Plate Drive 11 - 5

spring balance

section A-A
D C

Mount this device under


diaphragm plate with open
Chain side facing toward foot end.

Diaphragm plate,
wall side

Head end Foot end

Diaphragm plate,
Doctor’s side

Mount this device under


diaphragm plate by clamping to bar
attached to underside of plate.
Spotfilm device
Fig. 6

Siemens AG RXD3-209.071.01 Page 5 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
11 - 6 Diaphragm Plate Drive

423 Central beam center

418 A

215 C

38 B

Fig. 7

Adjustment and replacement of potentiometer 3PT5 11

• Press EMERGENCY STOP.


Auxiliary device: Tape measure.

Mechanical initial position: 215 mm, +/- 1 mm (C/Fig. 7).


Electronic initial position: + 5 V, ± 50 mV, 24S1 CPU board, TP2/9 - TP2/13,14 0V.

Set end positions


Auxiliary device: Tape measure.

Set minimum diaphragm opening B = 38 mm, +/- 1mm (B/Fig. 7)


• Select adjustment mode (see page 2-1).
- 24S2 I/O board, DW 1/8 to ON.
- Set 071 and store.
• Exit the adjustment mode.
- I/O PC board 24S2, DW 1/8 to OFF.

Set maximum diaphragm opening A = 418 mm, +/- 1 mm (A/Fig. 7)


• Select adjustment mode (see Page 2-1).
- Set 072 and store.
• Exit the adjustment mode.
- 24S2 I/O board, DW 1/8 to OFF
• Release EMERGENCY STOP.

SIREGRAPH CF RXD3-209.071.01 Page 6 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Diaphragm Plate Drive 11 - 7

Adjustment of gap width, general remarks 11

The factory adjusted spacing width on subdivided exposures is approx. 3 mm. For the
most part the spacing width depends on the positioning accuracy of the cassette carriage,
for both single exposures as well as series exposures. If the customer wishes to have
either a wider or narrower spacing, this can be adjusted electronically. The same spacing
width on exposures with 3 or 4 subdivisions cannot be obtained due to insufficient posi-
tioning accuracy of the cassette carriage.

• Select adjustment mode


I/O PC board 24S2, DW 1/8 to ON
Code 184 to 181 (reduction in steps of 0.2 mm each)
Code 185 (no software correction)
Code 186 to 189 (increase in steps of 0.2 mm each)
• Enter the desired code and store.
• Exit the adjustment mode
I/O PC board 24S2, DW 1/8 to OFF

Siemens AG RXD3-209.071.01 Page 7 of 8 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
11 - 8 Diaphragm Plate Drive

Adjustment of gap width (subdivision mode) 11

• Insert 18 cm x 24 cm cassette transversely


• Select 3 on 1 subdivision
Measure diaphragm opening (1st measurement)
• Select 4 on 1 subdivision and 3 on 1 subdivision again, press "Preparation" briefly
Measure diaphragm opening (2nd measurement)
Nominal: difference from 1st and 2nd measurement < 1 mm

Correction:
- Potentiometer P1 on 23AZ5 (/Fig. 8), brake current.

CN1

F1 F3
J3
P6

F2
P5

P4

J2
J1 P1 BC1
P2 P3

L2 L1

Fig. 8

• Insert 35 cm x 43 cm cassette transversely


• Select 3 on 1 subdivision
Measure diaphragm opening (1st measurement)
• Select 4 on 1 subdivision and 3 on 1 subdivision again, press "Preparation" briefly
Measure diaphragm opening (2nd measurement)
Nominal: difference from 1st and 2nd measurement < 1 mm
Repeat process with 18 cm x 24 cm cassette until the difference is < 1 mm
(optimization).

SIREGRAPH CF RXD3-209.071.01 Page 8 of 8 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Collimator 12 12 - 1

Control overview 12

I/O PC board 24S2 CPU PC board 24S1

Remote control console

Collimator
control 24S3

Height

Width

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
12 - 2 Collimator

Fig. 2

Drive overview 12

Drive: MT1 (Height)


MT2 (Width)

Actuation: 24S2

Control: 23S3

Fuse: F1/F2 1A 23S3

Potentiometer: POT1 (Height)


POT2 (Width)

SIREGRAPH CF RXD3-209.071.01 Page 2 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Collimator 12 - 3

Adjustment after replacement of the potentiometers 12

Mechanical initial position for the width:


Width collimator plates closed all the way
• Before installing, set the potentiometer to 900 Ohm (Ralco), 100 Ohm (Gorla), between
"CW" and "wiper"
• Take note of the direction of turn: when the width collimator opens the resistance value
must increase.
• On the CPU board 24S1, measure between TP2/13,14 = 0V and TP2/7: 1,45 V (Ralco),
700 mV (Gorla).

Mechanical initial position for the height:


Height collimator closed completely
• Before installation, set the potentiometer to 900 Ohm (Ralco), 100 Ohm (Gorla),
between "CW" and "wiper"
• Take note of the turn direction: when the height collimator opens the resistance value
must increase.
• On CPU board 24S1, measure between TP2/13, 14 = 0V and TP2/8: 1,45 V (Ralco),
700 mV (Gorla).

Adjusting the end positions 12

The procedure described here is repeated with each code listed in the table below.
• Select adjustment mode (see Page 2-1)
- I/O PC board 24S2, DW 1/8 to ON
• Enter the code and store.
Width collimator closed completely: Code = 051
Height collimator closed completely: Code = 061
Width collimator opened completely: Code = 052
Height collimator opened completely: Code = 062

• Exit the adjustment mode


- 24S2 I/O board, DW 1/8 to OFF.

Siemens AG RXD3-209.071.01 Page 3 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
12 - 4 Collimator

Correction Factors for Collimator 12

With an increase of the values in a unit’s place the collimation is increased.

Example: Correction of the cassette width


The current value is 125.
If the width is to be lowered, the value 125 is to be reduced to values between 121 and
124 with the help of the P31/32 keys.
If the width is to be raised, the value 125 is to be increased to values between 126 and
129 with the help of the P31/32 keys.

• Select adjustment mode (see Page 2-1)


- I/O PC board 24S2, DW 1/8 to ON
- Enter the desired code (see table below) and store

Display Adjustment Storage


H Z E H Z E
Cassette width 1 2 1-9 P34/35 P34/35 P31/32 P33
Cassette height 1 3 1-9 P34/35 P34/35 P31/32 P33
I.I. width 1 4 1-9 P34/35 P34/35 P31/32 P33
I.I. height 1 5 1-9 P34/35 P34/35 P31/32 P33

• After a storage “0” appears in the unit’s place.


• Exit the adjustment mode
- 24S2 I/O board, DW 1/8 to OFF.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Grid 13 13 - 1

Control overview 13

I/O PC board 24S2


CPU PC board 24S1

3FC10 or 3XS2

Collimator
Collimator
control

Remote control
console
Collimator

workstation

Fig. 1

Siemens AG RXD3-209.071.01 Page 1 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
13 - 2 Grid

Fastening
Befestigung Frame
Raster
Grid Rahmen
The oscillating grid is in
SOL2 the parking position. It is
Hubmagnet
lift magnet pressed there by lift mag-
net SOL2 against the
spring force of the leaf
springs.

SOL1 lift magnet


3FC10 Hubmagnet
3XS2

Leaf springs
Blattfedern
Fig. 2

Oscillation grid, general remarks 13

The grid is actuated by means of magnet SOL1.


Magnet SOL2 keeps the grid in a central position for DFR exposures.
The microswitch 3FC10 / or proximity switch 3XS2 releases the exposure.
Grid movement range approx. 15, ± 1 mm, minimum oscillation time > 30 sec.
Control: 24S2.
The grid is centered mechanically via the tapped guide of the SOL 2 magnet.

NOTICE: The ohmic resistance of the SOL 1 coil is less


than that of SOL 2.

Programming of oscillation grid / removable multi-line grid 13

• Select the adjustment mode (see Page 2 - 1)


I/O board 24S2, DW 1/8 set to ON
Enter the code:
221 normal oscillation grid
222 removable multi-line grid
• Save the code that was entered.
• Exit the adjustment mode
I/O board 24S2, DW 1/8 set to OFF

SIREGRAPH CF RXD3-209.071.01 Page 2 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Grid 13 - 3

Checking the grid centering 13

• Table in 0° position
• Tension thread / wire in center of spotfilm device
Nominal: Grid center +/- 1 mm from the thread / wire
Correction:
- Correct the position of the push rod guide

Checking the grid movement 13

• Table in 0° position
• Unlock grid by hand and keep it unlocked.
• Press grid by hand into one of the end stops and release it.
Nominal: at least five oscillations
Correction:
- none, check the condition of the springs (breaks / installation)

Grid contact 13

Mechanical grid contact 3FC10


• The contact must switch safely at the first oscillation, 24S2 / LED 57
(green) ON.
Correction:
- Correct position of the switch

Electronic grid contact 3XS2


• Unlock the grid by hand and press grid against stop on grid contact side.
Nominal: Grid frame - grid contact distance = 0.5 mm
Correction:
- Correct position of the switching element.
• The grid contact must switch safely at a distance < 5 mm, LED (red) at the grid contact or
24S2 / LED 57 (green) ON.

Siemens AG RXD3-209.071.01 Page 3 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
13 - 4 Grid
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Tomography 14 14 - 1

Column (head - foot)

Tacho generator

Column (foot - head)

1.3 sec.

Fig. 1

Setting the tomographic height, beginning with firmware version


VB00D 14

Setting the column speed 14

• Connect and set the oscilloscope


CH1 Column motor control 22AZ3 C1 11 (tacho signal)
C1 12 (ground)
Time base: 0.2 cm/sec, Trigger internal, Level 1, Slope +

The speed of the column is shown on channel 1, the signal is positive or negative
according to the movement direction.
The radiation duration is shown on channel 2.

Setting conditions:
• Tomography 40°
• Tomographic height 200 mm
• Exposure system in table center
• Switch off "Allarmax" (plug in AMP3 on 24S2)
• Release tomographic exposure without radiation
• Measure travel time of column from head to foot (diagram)
Required: 1.3 sec. (+ 0.05 sec/ - 0.0 sec) (Fig. 1)

Correction:
Potentiometer P2 on SP 22AZ3 (Pick-a-back board, max. Speed)
clockwise: speed is increased
counterclockwise: speed is decreased

• Switch on "Allarmax", (remove AMP3 on 24S2) !

Siemens AG RXD3-209.071.01 Page 1 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
14 - 2 Tomography
Setting the spotfilm device speed 14

It is possible to set a tomographic height correction digitally with this software version. The
starting value is Code 175 and the correction can be made in 1 mm steps downwards with
Code 174 to 171 and in 1 mm steps upwards with Code 176 to 179.
A prerequisite is that the code 175 is programmed before the speed of the spotfilm device
is adjusted. Otherwise the correct tomographic height can not be set.
• Select adjustment mode, I/O PC board 24S2, DW 1/8 to ON, (see page 2 - 1)
• Verify if code 175 is programmed, search with keys P34 and P35 only
• If not, enter code 175 and store
• Exit adjustment mode, I/O PC board 24S2, DW 1/8 to OFF
• Connect multimeter: 24S1, TP2 - 4 (+)
24S1, TP2 - 14 (0V)

NOTICE The multimeter must be able to display three digits after the
decimal point.

• At the generator, set "SS" to off


• Switch off "Allarmax" (plug in AMP3 on 24S2)
• Tomography 40°
• Tomographic height 200 mm
• Exposure system in table center
• Press prep-button and hold until the column has reached its head-end start position and
the multimeter displays a stable value. Read and note down that value (value A, wiper -
voltage of potentiometer 3PT2).
Trigger a planigraphic exposure without radiation by pressing prep- and main-button.
Column travels from head end to foot end. Release prep- and main-button and
subsequently press both again thus triggering another planigraphic exposure allowing
the column to travel back from foot end to head end. Attention: keep prep- and main-
button depressed until the column has reached its head-end position and the multimeter
displays a stable value (value B).
• Determine the difference from both measured values A and B
Required: ± 5 mV

Correction:
Potentiometer P2 on SP 22AZ4 (Pick-a-back board, max. Speed)
clockwise: if A > B
counterclockwise: if A < B

Repeat the process until the required ± 5 mV are reached.


• Switch on "Allarmax", (pull out AMP3 on 24S2) !

SIREGRAPH CF RXD3-209.071.01 Page 2 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Tomography 14 - 3
Software correction of the layer-height 14

• Perform tomographic exposure 40° and tomographic height 200 mm with tomographic
test.
• Determine the difference of the tomographic height.
• Select adjustment mode
- I/O PC board 24S2, DW 1/8 to ON

Code 174 to 171 (decreasing in 1 mm steps each)


Code 175 (no software correction)
Code 176 to 179 (increasing in 1 mm steps each)

• Enter the desired code from the above table and store
• Exit the adjustment mode
- I/O PC board 24S2, DW 1/8 to OFF
• Take new tomographic exposure under the same conditions.
• If there is still a difference, make further corrections.
• In systems with FLUOROSPOT take an indirect tomography control exposure.

NOTICE The setting made at 40° / 200 mm applies for all tomographic
angles, tomographic heights and operating modes.

• Make sure that "Allarmax" is switched back on, (jumper AMP3 on 24S2 removed) !

Siemens AG RXD3-209.071.01 Page 3 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
14 - 4 Tomography

kV- act
t1 Column (head - foot)
t2
Säule (Kopf- Fuß)

Tacho generator

Column (foot - head)


Säule (Kopf- Fuß) Radiation

Fig. 2

Setting the tomographic symmetry 14

• Connect and adjust the oscilloscope


CH1 motor control column 22AZ3 C1 11 (tacho signal) 20V/div/40°
10V/div/20°
5V/div/ 8°

CH2 kV act generator

POLYDOROS LX D100.X18 500 mV/div


D100.X4 (ground)

POLYDOROS 50/80 SX D211 Σ kV-ACT 500 mV/div.


D211 (ground)
Time base: 0.2 cm/sec, Trigger internal, Level 1, Slope +

The column speed is displayed on channel 1; depending on the movement direc-


tion, the signal is positive or negative; the radiation time is displayed on
channel 2.

Adjustment conditions:
• Select 40 kV, minimum mAs on the generator.
• Insert cassette, 4 on 1 subdivision, close collimator.
• Select tomography and trigger exposure

SIREGRAPH CF RXD3-209.071.01 Page 4 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Tomography 14 - 5
Adjustment: (direct radiography)
Required: Radiation (CH2) symmetrical to column speed
(CH1),i.e. t1 = t2 (Fig. 2).

Correction: Potentiometer on I/O PC board 24S2.


P2 for 40°, P4 for 20°, P6 for 8°.

NOTICE P2
P4 turning these clockwise delays radiation release.
P6

Adjustment: (indirect radiography)


Required: Radiation (CH2) symmetrical to column speed
(CH1),i.e. t1 = t2 (Fig. 2).

Correction: Potentiometer on the I/O-PC board 24S2.


P1 for 40°, P3 for 20°, P5 for 8°.

NOTICE P1
P3 turning these clockwise delays radiation release.
P5

Siemens AG RXD3-209.071.01 Page 5 of 6 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
14 - 6 Tomography
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 6 of 6 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Radiation Geometry 15 15 - 1

Length

Marker

Side marker

Fig. 1

Coincidence of light field/radiation field 15

• Move the unit into the 0° position.


- SID = 115 cm.
- Full format and Bucky
• At the generator, set
- bed-side exposure
- select large focus.

Exposure position 15

• Using the setting buttons on the collimator, starting from the maximum format, set a
24 cm x18 cm rectangular format on the collimator scale for SID = 115.
• Switch on light localizer and mark light field edges with radio-opaque material,
e.g. washer or coins, attach a fifth marker as a side marking.
• Select approx. 50kV/2mAs and trigger an exposure.

Siemens AG RXD3-209.071.01 Page 1 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
15 - 2 Radiation Geometry

Y1
Film

Light field

X2 X1

Radiation field

Y2

Fig. 2

Evaluation 15

• On all four sides, measure the difference between the light field edges and the radiation
field edges and calculate the total difference in the X- and Y-directions.
∑X=X1+X2
∑Y=Y1+Y2

• If the tolerance is exceeded by ≥ 1.8%, adjust the collimator at the tube assembly flange
so that light field and radiation field are within the tolerance.
• Repeat the exposure.

SIREGRAPH CF RXD3-209.071.01 Page 2 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Radiation Geometry 15 - 3


Fig. 3

Adjustment of the central beam 15

Horizontal table direction 15

• Set unit to 0°
• Stretch wrap-wire over the notches (spotfilm device center).
• SID = 135 cm.
• Loosen the X-ray tube support cover (1/Fig. 3).
• Switch on light localizer.
• Turn the X-ray tube and adjust central beam center.
• Deviation = 0 mm.
• Tighten the X-ray tube support cover (1/Fig. 3).
• Check: verify the difference at SID = 150 cm and SID= 115 cm.
Maximum permissible difference = ± 2% from SID.

Siemens AG RXD3-209.071.01 Page 3 of 4 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
15 - 4 Radiation Geometry

Fig. 4 Fig. 5

Vertical table direction 15

Verify the difference in the vertical table direction by changing the SIDs and adjust
by means of 2 Allen screws and 2 set screws (1/Fig. 4).
Maximum permissible difference = ± 2% from SID.

Check exposures:
• Insert a cassette that is marked so that the center and sides can be recognized.
• Trigger exposure.
• See table of contents, RXD3-209.034. . . .(SG CF Startup Instructions)

Adjust 0 degree switch 3FC1 15

• Set unit into 90° position.


• Loosen Allen screw (1/Fig. 5) and adjust the disk so that the trip lever engages in the
notch.

SIREGRAPH CF RXD3-209.071.01 Page 4 of 4 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Installing/Removing the Tabletop 16 16 - 1

3
3TP1
Fig. 1

Removing the tabletop 16

• Disconnect the wires 5 and 7 on the terminal strip 3TB1 (back of column) (Fig. 1).

NOTICE If the 24S2.AMP1 connector is unplugged instead of leads 5 and


7, the compression carriage will move up after approx. 2 sec. and
the slip clutch will respond.

• Remove the covers (2/Fig. 1).


• Remove the ground wire and the end stop (3/Fig. 1).
• Move the tabletop at doctor’s side until you can remove it.

Installing the tabletop 16

• Put the tabletop in position carefully and push it evenly into the guide until it reaches the
mechanical stop. Pay special attention to the limit switches 3FC8 and 3FC9
(1/Fig. 1) to make sure they do not get bent or damaged.
• Now move the tabletop further in with motor force.
• Attach covers (2/Fig. 1).
• Restore the connections at the terminal strip 3TB1.
• Reattach end stop and ground wire (3/Fig. 1).

Siemens AG RXD3-209.071.01 Page 1 of 2 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
16 - 2 Installing/Removing the Tabletop
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 2 of 2 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Installing/Remov. the Cover of the Spotfilm Dev. 17 17 - 1

a
b1

1 b
b
B
A

b
1
a
b1 b

Fig. 1 Fig. 2

Removing the upper cover of the spotfilm device 17

• Installation and removal with unit in 0° position


• Loosen the 8 Allen connection screws and the cover panel ground wire (1/Fig. 1).

Removing the lower cover of the spotfilm device 17

• Installation and removal with unit in 90° position


• During work on the cassette carriage motor and the potentiometer, remove the cover of
the spotfilm device (A/Fig. 2). To do so, loosen screws a and b1.
• During adjustment of the cassette carriage bearings, remove the cover of the spotfilm
device (B/Fig. 2). To do so, loosen screws b and b1.

Siemens AG RXD3-209.071.01 Page 1 of 2 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
17 - 2 Installing/Remov. the Cover of the Spotfilm Dev.
This page intentionally left blank.

SIREGRAPH CF RXD3-209.071.01 Page 2 of 2 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering
Changes to Previous Version 18 18 - 1

- Chapter 1-1, "Self-learn Package on CD-Rom" inserted.


- Chapter 1-3, section regarding "Replacing and Adjustment Potentiometers" inserted.
- Chapter 1-4, remark regarding "damaged or worn parts"
as well as "Manufacturer’s Note" added
- Chapter 1-6, Text corrected for "L23"
- Chapter 1-7, Text corrected for "L37"
- Chapter 1-8, Text corrected for "L51, L52, L57"
- Chapter 1-12, Fig. 8 revised and Fig. 9 updated.
- Chapter 1-20, designations of potentiometers, motors and boards corrected.
- Chapter 1-23, designations of motors and boards corrected.
- Chapter 1-24, section regarding "DIP switch DW1" inserted.
- Chapter 1-26, remark regarding Code 201-209 and Code 211-213 inserted.
- Chapter 1-27, 1-28, 1-29, "Listing of All Programming Codes" inserted.
- Chapter 3-1, "Overview of Control" moved above Fig. 1.
- Chapter 3-2, "Tightening screw" inserted into Fig. 2.
- Chapter 3-3, Text rewritten.
- Chapter 3-4, Text rewritten and section "Replacing the Brakes" inserted.
- Chapter 3-5, Fig. 8 inserted
- Chapter 3-6, Fig. 10 inserted.
- Chapter 3-9, text rewritten and expanded.
- Chapter 4-1, "Overview of Control" moved above Fig. 1.
- Chapter 4-4, "Reference point....." and "Potentiometer play....." inserted.
- Chapter 5-1, "Overview of Control" moved above Fig. 1.
- Chapter 5-3, "Reference Point: Outside Edge of the Table Frame" inserted.
- Chapter 6-1, "Overview of Control" moved above Fig. 1.
- Chapter 6-4, Numbering of illustrations updated.
- Chapter 7-1, "Overview of Control" moved above Fig. 1.
- Chapter 7-4, section "Adjustment of the SID Limit at 90°" inserted.
- Chapter 7-6, section "Adjustment of the 23Az7 Board" rewritten.
- Chapter 8-1, "Overview of Control" moved above Fig. 1.
- Chapter 8-3, section "Electrical Adjustment", text rewritten.
- Chapter 9-1, "Overview of Control" moved above Fig. 1.
- Chapter 9-5, section "Adjusting the Slip Clutch", text rewritten.
- Chapter 9-6, illustration expanded.
- Chapter 9-7, Fig. 8 revised.
- Chapter 9-8, Fig. 9 revised.
- Chapter 9-9, section "Maximum Tension Jaw Opening", text rewritten.
- Chapter 10-1, "Overview of Control" above Fig. 1 moved.
- Chapter 10-2, Fig. 2 revised, section "Overview of Drives for Format Sensing", text
rewritten.
- Chapter 10-3, remark regarding potentiometers inserted.
- Chapter 10-4, illustrations revised.
- Chapter 11-1, "Overview of Control" moved above Fig. 1, Fig. 1 revised.
- Chapter 11-2, typographical error corrected from "3PT7" to "3PT5".
- Chapter 11-3, section "Symmetry and Centering", "in the service kit" inserted.
- Chapter 11-4, Fig. 5 and text arranged differently, section "Adjusting the Slip Clutch"
from Chapter 11-5 rewritten and inserted.

Siemens AG RXD3-209.071.01 Page 1 of 2 SIREGRAPH CF


Medical Engineering Rev. 05 09.00 TD SD 21 Service
18 - 2 Changes to Previous Version
- Chapter 11-5, Fig. 6 added, Section"Adjusting the Slip Clutch" moved to after Chapter
11-4.
- Chapter 11-6, Fig. 6 revised, text updated.
- Chapter 11-7, section "Adjustment of the Gap Width, General Remarks" from Chapter
11-8 inserted.
- Chapter 11-8, section "Gap Width Correction with Subdivisions" and illustration from
Chapter 11-7 inserted and rewritten.
- Chapter 12-1, "Overview of Control" moved above Fig. 1.
- Chapter 12-3, section "Adjustments Following Replacement of Potentiometers", text
rewritten.
- Chapter 12-4, Text arranged differently.
- Chapter 13-1, "Overview of Control" moved above Fig. 1.
- Chapter 13-2, section "Programming the Oscillation Grid" added.
- Chapter 14-1, text rewritten
- Chapter 16-1, text rewritten.

TD PS 21 / Schmidt, Tropia

SIREGRAPH CF RXD3-209.071.01 Page 2 of 2 Siemens AG


Service Rev. 05 09.00 TD SD 21 Medical Engineering

Das könnte Ihnen auch gefallen