Sie sind auf Seite 1von 26

SECTION 8.

15

LT-7000

LASER ALIGNMENT
DEVICE
IMS LT- 7000

FIELD SERVICE INSTALLATION


INSTRUCTIONS FOR
EXPOSCOP 7000

P.C. Board Mounted Laser Assembly

INTRODUCTION:

The following instruction are to be used i n mounting the LT-7000 assembly for use i n the Exposcop 7000
C-Arm.

The addition of a few components incorporated i n the LT-7000 U2300 board assembly allows the
Exposcop 7000 laser device to power on and off from the C - A r m control panel. The system incorporates a
timer i n the C - A r m control w i t h a 1 m i n . preset on time.

CAUTION A N D WARNINGS :

T H E LT-7000 L A S E R P O S I T I O N I N G D E V I C E IS A C L A S S II L A S E R D E V I C E W H E N F U L L Y
ASSEMBLED H O W E V E R PLEASE OBSERVE THE FOLLOWING W H E N W O R K I N G A R O U N D
THE LASER ASSEMBLY.

WARNING: " CLASS II LASER PRODUCT -AVOID LONG-TERM VIEWING


OF DIRECT LASER RADIATION "

CAUTION:
THE LT-7000 LASER ASSEMBLY INCORPORATES THE USE OF DIODE LASER MODULES WHICH
EMIT LASER RADIATION. THEREFORE DO NOT UNDER ANY CIRCUMSTANCES STARE , OR
VIEW DIRECTLY WITH OPTICAL INSTRUMENTS THE LASER BEAM OR ANY SCATTERED
LASER RADIATION DURING THE ASSEMBLY OR FINAL TESTING OF THE SYSTEM.

LASER: A T T E N T I O N — VISIBLE L A S E R L I G H T
CLASS H LASER PRODUCT
LASER RADIATION - D O N O T STARE
INTO B E A M
L A S E R C L A S S II, IEC 825, 21 C F R S U B C H A P T E R J , 1040

C A U T I O N - The use of optical instruments w i t h this product w i l l increase eye hazard

W h e n operating the laser light beam, the "Guidelines for the User IEC 825, M a i n Chapter 3"
have to be f o l l o w e d .

The m a x i m u m output power of the laser beam is 1 m W . The wavelength of the


emitted power is 635 n m . The laser guidance system is service free. However,
should an adjustment or a repair be required, this must be performed by the
manufacturer or authorized person only. Caution- Opening protective covers
permits human access to Visible Laser Light " D O N O T S T A R E EMTO B E A M "

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 - 1
WARNING! Electrical hazards which could result in personal injury.

Observe all safety precautions when turning on power with covers removed from the

A system. High voltage will be present at several locations.

C A U T I O N ! Hazards which could result in equipment or property damage.

Observe all safety precautions when removing covers and installing electronic circuit
boards and cables. Static sensitive components are used on some of the components to
be installed.

Observe all Electrostatic discharge (ESD) precautions when handling circuit board
components and other electronic subassemblies.

Observe safety precautions when removing covers and reinstalling cover that you do not
drop or otherwise risk damage to the covers or surrounding components.

I N S T A L L A T I O N INSTRUCTIONS:

PART 1 MODIFICATION OF EXPOSCOP 7000 U45 BOARD

Part list for U2300 board is located on page 14 of this procedure.

1) T u r n off all power to Exposcop 7000 and disconnect power cord from w a l l power outlet.

2 ) Remove Exposcop 7000 C - A r m generator cover then remove U45 board from generator board slot
connectors. See installation instruction diagrams page 6 for further details of removal of
generator cover.

3 ) U s i n g installation diagram page 7 locate and drill three (3) holes and mount connector X01 by
bending over the p i n connectors on the solder side of the board.

4 ) Solder jumper wire on solder side of U45 from P2 connector p i n 3 to X01 p i n 3 and glue to board
to secure i n place. See installation diagram page 8 and page 7.

5 ) Solder jumper wire on solder side of U45 from X I p i n 23 a,c to X01 p i n 2 and glue to the board to
secure i n place. See installation diagram page 8 and page 7.

6 ) Solder jumper wire on solder side of U45 from X I p i n 1 a,c to X01 p i n 1 and glue to the board to
secure i n place. See installation diagram page 8 and page 7.

7 ) Check to make sure all jumper wires are correctly installed and then reinstall U45 board i n the
generator.

8) Connect digital volt meter from p i n 1 of X01 (0V) to p i n 2 of X01 (+) connector on U45 board
making sure meter probes are not shorting at X01 connector.

9 ) Connect Power cord to wall outlet and turn on power to Exposcop 7000. Meter should read
+ 14.0 to 16.5 V D C .

10 ) T u r n off power and connect digital volt meter from p i n 1 of X01 (0V) to p i n 3 of X01 (+) connector
on U45 board making sure meter probes are not shorting at X01 connector.

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3-2
11) T u r n o n power to Exposcop 7000. Meter should read approx. + 11.80 to 12 V D C .

12 ) If voltages i n step 9 and 11 are correct then turn off the power to the Exposcop Plus and remove
power cord from w a l l power outlet. If not determine cause of error by checking all connections,
then repeat steps 8 to 11.

This completes part one of installation instructions for LT - 7000


laser power board for Exposcop 7000 installation.

P A R T 2 I N S T A L L A T I O N A N D A L I G N M E N T O F LT-7000 B O A R D A S S E M B L Y .

1) M a k e sure power is turned off to Exposcop 7000 and w a l l outlet p l u g is disconnected.

2 ) Place a small foam or plastic sheet on top of the end of the C - A r c to support and isolate the
U2300 board assembly and the install cable from X01 U45 board to X I of U2300 board m a k i n g
sure connectors are correctly positioned.

3 ) Connect power cord to wall outlet of Exposcop 7000 and then turn on power. After 40 sec or so the
Exposcop 7000 should be ready to test U2300 board.

4 ) Press the laser key o n the Exposcop 7000 control keyboard. The laser on L E D should light o n the
control keyboard as w e l l as the U2300 board power o n led's and laser's modules.

The lasers should come on for 1 minute then go off. The on time is preset by i n the software.
Pressing the laser key again w h e n the laser's are on should turn off the laser also.

F A I L U R E O F L A S E R P O W E R O N TEST.
If the laser fail to turn on you must determine the problem and repair as necessary. See
schematics for U2300 board and interface to Exposcop 7000 U45, U325 boards page 9.

5 ) After the power on test is found to be working correctly turn off power for Exposcop 7000 a n d
remove power cord form w a l l outlet.

6) Install the four 3" mounting standoff's b y removing only one screw at a time from the collimator
assembly i n order not to disrupt the collimator alignment. See installation diagrams page 11
for further details of U2300 board.

7) Once the stand offs are correctly installed then mount the U2300 board to the standoff's. T h e n

Connect the cable from U45 X01 to X 1 A of U2300 board.

8) Connect power cord to wall outlet of Exposcop 7000 and then turn on power.

After 40 sec or so the Exposcop 7000 should be ready to test LT-7000 U2300 board.

9) Press the laser key on the Exposcop 7000 control keyboard. The laser on L E D should light on the
control keyboard as well as the U2300 board power on led's and laser's modules.
The lasers should come on for 1 minute then go off. The on time is set by the Z I E H M W I N SRV.
software. Pressing the laser key again w h e n the laser's are o n should turn off the laser also.

F A I L U R E O F L A S E R POWER O N TEST.
If the laser fail to turn on y o u must determine the problem and repair as necessary. See
schematics page 9 for U2300 board and interface to Exposcop 7000 U45 , U325 boards.

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 - 3
10 ) To align the laser line generators w i t h the center of the Image Intensifier place a jumper o n JT
service test p o i n t . This w i l l keep the laser line generators o n until the jumper is removed.
Center the laser lines to cross i n the center of the Image Intensifier at 45 degrees.
See installation diagrams page 12 for further details of alignment of the laser assembly.

11) If the lasers turn on and off correctly and the alignment is complete, turn off the power to the
Exposcop 7000 and remove power cord from the w a l l outlet.

12 ) Install the filter cap w i t h laser w i n d o w s to the generator cover and reinstall generator cover.

13 ) Connect power cord to wall outlet of Exposcop 7000 and then turn on power. After 40 sec or so the
Exposcop 7000 should be ready to test laser and LI alignment. Check laser alignment if it is
still aligned then turn off the power to the Exposcop 7000 and remove power cord from the w a l l
outlet.

14 ) Once the alignment is complete secure generator cover and seal w i t h white silicon sealer.

This ends part two of the installation instructions for the LT-7000 laser assembly.

PART 3 LABELING EXPOSCOP 7000 WITH LASER DEVICE

INTRODUCTION:

Several labels are required for identification and for certification of the laser device.

1) The following example and location guides must be used i n placement of the labels to complete
the installation of the laser device i n the Exposcop 7000.

2) Install Caution Label 40-511-L10, Aperture Label 40-513 & 40-514-L10, Small C a u t i o n Label
40-512-L10 to control desk and Laser Caution W h e n O p e n Label 40-515-L10 as shown i n the
following diagrams pages 5 -6.

3) F i l l out the Manufacture certification label 40-510-L10 and included all information. Serial
number, M o t h and Year must be typed. The month and year of manufacture shall be provided
clearly and legibly, without abbreviation, and w i t h the year s h o w n as a four- digit number.

The month must be spelled out. i.e., January, February, June, etc... D O N O T A B B R E V I A T E .

The Year M u s t include all four digits, i.e., 1996,1997, etc. D O N O T A B B R E V I A T E .

The certification and manufacture label must be installed on the generator cover just above the
C - A r c attachment to the generator and must be positioned i n the m i d d l e of this area. Then
apply a clear label cover over the label to protect the information.

THIS C O M P L E T E S T H E I N S T A L L A T I O N A S S E M B L Y A N D T E S T I N G
OF T H E U2300/LT-7000 L A S E R D E V I C E F O R E X P O S C O P 7000

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3-4
Position of Required Labels on Generator Covers.

CAUTION - Laser radiation


when open. DO NOT
CAUTION
STARE INTO BEAM LASER RADIATION-DO NOT
STARE INTO BEAM

PEAK POWER lmW


WAVELENGTH 635lffl>
CLASS I I LASER PRODUCT

Located on rear of
Generator cover
LASER
APERTURE

CLASS II LASER PRODUCT


MODEL SERIAL No.
LT-7000

COMPLIES WITH DHHS RADIATION PERFORMANCE


STANDARDS, 21 CFR SUBCHAPTER J. AS OF DATE OF
MANUFACTURE

MANUFACTURED BY:
INTERNATIONAL MEDICAL SYSTEMS, INC
4181 L A T H A M STREET RIVERSIDE CA. 92501

MANUFACTURED:
MONTH: YEAR:.

Mount Laser Caution Label


40-512-L10 Above Laser " O N "
Button.
SKIN C O N E

• • •• • • •
c
• • • • • •
APERTURE

• • •
• • • • • •
• • • • 0 • •
• ••
• • •

o
C-Arm Control
APERTURE

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 - 5
REMOVE ONLY THE
OUTSIDE SCREWS.

C A U T I O N - Laser radiation
BOTTOM OF when open. D O N O T
TUBE HEAD ASSEMBLY ^jj^iTARE INTO B E A M

Laser Cap

CUT SILICON SEAL


AROUND EDGE OF COVER
BEFORE LIFTING OFF

RESEALWITH
WHITE SILICON
SEAL WHEN
REINSTALLING
COVER

REMOVE ONLY THE


OUTSIDE SCREWS.
SIDE VIEW OF
TUBE HEAD ASSEMBLY

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 - 6
U45 BOARD
INTRODUCTION:

The U45 Board modification for X01 connector provides power and O N / O F F control for the laser
positioning device.

Make sure the following connection are checked before power is turned on as i n correct placement of
wires can cause damage to both the C - A r m control and the laser device.

X01 P i n Configuration

1) D r i l l three holes 33 m m from top edge of U45 board next to d4 relay.

2 ) M o u n t 3 p i n connector and bend over pins on back of the board.

3 ) A d d Three wire jumpers. Use small amount of glue to fix wires to board to prevent wires
from being damaged w h e n inserted into generator card rack

1) U45 P2 p i n 3 to X01 p i n 3
2) U45 X I p i n l a c to X01 p i n 1
3) U45 X 1 p i n 23ac to X01 p i n 2

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 - 7
U45 B O A R D

3 ~

Exposcop 7000 LT-7000 Laser Assembly


International Medical System Service Installation Procedure
z
5 8
m H
X -J
z
W

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service installation Procedure Rev. 3 - 9
LASER C A B L E ASSEMBLY
C
Complete Assembly Part Number
40-703-L10

Item # Part ID/Name Part # Description Quantity

1 Jl WM2001-ND 3 PIN FEMALE CONNECTOR 2

2 PINS WM2200-ND CRIMP PINS FOR FEMALE CONNECTOR 6


3 3 WIRES A2102W-100-ND 3 HOOK -UP WIRES ALPHA 3073-100 10" long 3
4 SPIRAL ELECTRONIC CABLE WIRE PROTECTION 10 "
WRAP WAREHOUSE

U45 J l U2300 J l
3 Conductors 9 " Spiral Wrap
1

10 "7 25 mm

U45 to U2300 Board Cable Part


Number 40-703-L10 Assembly

The Exposcop 7000 U45 board provides the O N / O F F control and power for the U2300
Laser board through the interconnect cable.

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 -10
1) File Off Tips Of Switch Cam Flush With Collimator Ring Assembly To
Avoid Hiting The Hex Studs.

Do not remove Switch Cam to file edge. Move ring assembly to allow
access to file edge of the Cam. Cover Tube head assembly parts to prevent
fillings from getting into tube assembly circuit boards and collimator.

Make sure no lose filings from the switch cam are left in the tube head
assembly.
See Example In
Dotted Lines

2) When Installing Hex Studs Align The Flat Edge Of


All Four (4) Hex Studs With The Side Of The
Collimator Ring To Avoid Interference Of The
Studs And The Ring When Slot Collimator is
Rotated

Small Caution
Laser Label CAUTION
Must be Attached LASER RADIATION-DO NOT
to Laser Board STARE INTO BEAM

Assembly PEAK POWER 1mW


FIGURE 1 S
#40-512-L10 WAVELENGTH 635nm
CLASS II LASER PRODUCT

LASER
4mm Alien Head Screws
APERTURE
Removed From The
Colimator Assembly
Small Laser
4mm Flat Washer Supplied With
Aperture Label
Laser Kit U45 Board
Must be Attached
LT-7000 / U2300 Power X01 Connector
Between Laser
Assembly Board
Module Assembly
Note: #40-514-L10
4mm X 3" Hex Stud
[Remove Only One Collimator
Assembly Screw And Replace With
Hex Stud At A Time To Prevent
Collimator Misalignment ]

Collimator
Assembly

Exposcop 7000 LT-7000 Laser A s s e m b l y 01/11/97


International Medical System Service Installation Procedure Rev. 3 -11
By carefully turning the laser with the center slot it can be
twisted to intersect with the other axis laser line generator
to form a cross at the center of the grid aligned with the
Retaining grid mounting screws.
Screw Once lines are correct use retaining screw
on side of laser mount to secure laser.
WARNING DO NOT OVERTIGHTEN RETAINING
SCREW AS LASER MODULE WILL BE D A M A G E .

By adjusting the three screws the laser can be tilted


to align with the I.I grid mounting screws to provide
a beam cross as low as possible above the skin cone.

Tilt Adjustment Tilt down The inside two screws should be tightened first to tilt
Screws the laser module inward toward the center of the
board. However be carefull not to adjust them to
much.

The retaining screw is used to slide the laser along a


Lock Screw limited distance slot to center the laser to the ouput
glass and help align it with the center of the II.

Laser Board

Laser mount and


Screw Lock plate

GRID SCREWS

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 -12
LT-7000
L A S E R POSITIONING D E V I C E
Complete Assembly Part Number
40-710-L10
Part List For Laser Assembly
Item # Part ID/Name Part # Description Quantity
1 PC -BOARD 40-236-L10 BLANK PC LASER BOARD 2
2 Rl, R9 390Q 390 OHM RESISTOR 1
3 R2 560Q 560 OHM RESISTOR 1
4 R3 180Q 180 OHM RESISTOR 1
5 R4 200Q 200 OHM RESISTOR 1
6 R5 2210Q 2..21K OHM RESISTOR 1
7 R6 3320Q 33.2K OHM RESISTOR 1
8 R7 10000Q 10KOHM RESISTOR 1
9 R8 62Q 62 OHM RESISTOR 1
10 CI lOuF lOuF CAPACITOR 1
11 C2 O.luF 0.1 uF CAPACITOR 1
12 Dl ZPY 5.1 [1N5231] 5.1 VOLT ZENER DIODE 1
13 D2 1N4001 STEERING DIODE 1
14
15 D3, D4 40-702-L10 V L M 3 - l.OmW 635nm LASER MODULE 2
16 D5, D6 40-040-D10 LED 1 RED [ POWER ON LED ] 2
17 Fl P822-ND 250 mA MICRO FUSE 1
18 IC1 4N32 QT OR [ SPH 600-01 1 OPTO -ISOLATOR 1
19 IC2 LM317 VARIABLE VOLTAGE REGULATOR 1
20 Tl 2N2907 PNP TRANSISTOR 1
21 T2, T3 2N2222 NPN TRANSISTOR 2
22 XL, X1A WM4301-ND 3 PIN MALE CONNECTOR 2
23 TJ WM4000-ND 2 PIN ROW CONNECTOR 1
24 JUMPER 929957-08-ND SHORTING JUMPER GOLD PLATED 1
25 Laser Label 40-510-L10 Certification Label 1
26 Laser Label 40-511-L10 Caution Label Top Generator Cover 1
27 Laser Label 40-512-L10 Caution Small- Laser board and Control Desk 2
28 Laser Label 40-513-L10 Aperture Label Skin Cone 3

29 Laser Label 40 -514-L10 Aperture Label Top Generator Cover 1

30 Laser Label 40-515-L10 Caution When Open Label bottom Generator 1


Cover

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 -13
LT-7000
L A S E R POSITIONING D E V I C E
Complete Assembly Part Number
40-710-L10
Part List For Laser Assembly
Item # Part ID/Name Part # Description Quantity
1 PC -BOARD 40-236-L10 BLANK PC LASER BOARD 2
2 Rl, R9 390Q 390 OHM RESISTOR 1
3 R2 560Q 560 OHM RESISTOR 1
4 R3 180Q 180 OHM RESISTOR 1
5 R4 200Q 200 OHM RESISTOR 1
6 R5 2210Q 2..21K OHM RESISTOR 1
7 R6 3320Q 33.2K OHM RESISTOR 1
8 R7 10000Q 10KOHM RESISTOR 1
9 R8 62Q 62 OHM RESISTOR 1
10 CI lOuF lOuF CAPACITOR 1
11 C2 O.luF 0.1 uF CAPACITOR 1
12 Dl ZPY 5.1 [1N5231] 5.1 VOLT ZENER DIODE 1
13 D2 1N4001 STEERING DIODE 1
14
15 D3, D4 40-702-L10 VLM3-1.0mW 635nm LASER MODULE 2
16 D5, D6 40-040-D10 LED 1 RED [ POWER ON LED ] 2
17 Fl P822-ND 250 mA MICRO FUSE 1
18 IC1 4N32 QT OR [ SPH 600-01 ] OPTO -ISOLATOR 1
19 IC2 LM317 VARIABLE VOLTAGE REGULATOR 1
20 Tl 2N2907 PNP TRANSISTOR 1
21 T2,T3 2N2222 NPN TRANSISTOR 2
22 XI, X1A WM4301-ND 3 PIN MALE CONNECTOR 2
23 TJ WM4000-ND 2 PIN ROW CONNECTOR 1
24 JUMPER 929957-08-ND SHORTING JUMPER GOLD PLATED 1
25 Laser Label 40-510-L10 Certification Label 1
26 Laser Label 40-511-L10 Caution Label Top Generator Cover 1
27 Laser Label 40-512-L10 Caution Small- Laser board and Control Desk 2
28 Laser Label 40-513-L10 Aperture Label Skin Cone 3

29 Laser Label 40 -514-L10 Aperture Label Top Generator Cover 1

30 Laser Label 40-515-L10 Caution When Open Label bottom Generator 1


Cover

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3 -13
IMS LT- 7000

FIELD SERVICE INSTALLATION


INSTRUCTIONS FOR
LT-7000 & U45b BOARD

P.C. Board Mounted Laser Assembly

INTRODUCTION:

The following instruction are to be used i n mounting the LT-7000 assembly for use i n the Exposcop 7000
C-Arm.

The addition of a few components incorporated i n the LT-7000 U2300 board assembly allows the
Exposcop 7000 laser device to power on and off from the C - A r m control panel. The system incorporates a
timer i n the C - A r m control w i t h a 1 m i n . preset on time.

CAUTION A N D WARNINGS :

T H E LT-7000 L A S E R P O S I T I O N I N G D E V I C E IS A C L A S S II L A S E R D E V I C E W H E N F U L L Y
ASSEMBLED H O W E V E R PLEASE OBSERVE THE F O L L O W I N G W H E N W O R K I N G A R O U N D
THE LASER ASSEMBLY.

WARNING: " CLASS II LASER PRODUCT -AVOID LONG-TERM VIEWING


OF DIRECT LASER RADIATION "

CAUTION:
THE LT-7000 LASER ASSEMBLY INCORPORATES THE USE OF DIODE LASER MODULES WHICH
EMIT LASER RADIATION. THEREFORE DO NOT UNDER ANY CIRCUMSTANCES STARE , OR
VIEW DIRECTLY WITH OPTICAL INSTRUMENTS THE LASER BEAM OR ANY SCATTERED
LASER RADIATION DURING THE ASSEMBLY OR FINAL TESTING OF THE SYSTEM.

LASER: A T T E N T I O N — VISIBLE L A S E R L I G H T
CLASS n LASER PRODUCT
LASER RADIATION - D O N O T STARE
INTO B E A M
L A S E R C L A S S II, IEC 825, 21 C F R S U B C H A P T E R J , 1040

C A U T I O N - The use of optical instruments w i t h this product w i l l increase eye hazard

W h e n operating the laser light beam, the "Guidelines for the User IEC 825, M a i n Chapter 3"
have to be f o l l o w e d .

The m a x i m u m output power of the laser beam is 1 m W . The wavelength of the


emitted power is 635 n m . The laser guidance system is service free. However,
should an adjustment or a repair be required, this must be performed by the
manufacturer or authorized person only. Caution- Opening protective covers
permits h u m a n access to Visible Laser Light " D O N O T S T A R E I N T O B E A M "

Exposcop 7000 LT-7000 Laser Assembly 09/23/97


International Medical System Installation Procedure Rev. 4-1
WARNING! Electrical hazards which could result in personal injury.

Observe all safety precautions when turning on power with covers removed from the
system. High voltage will be present at several locations.

C A U T I O N ! Hazards which could result in equipment or property damage.

Observe all safety precautions when removing covers and installing electronic circuit
boards and cables. Static sensitive components are used on some of the components to
be installed.

Observe all Electrostatic discharge (ESD) precautions when handling circuit board
components and other electronic subassemblies.

Observe safety precautions when removing covers and reinstalling cover that you do not
drop or otherwise risk damage to the covers or surrounding components.

PART 1 INSTALLATION A N D A L I G N M E N T OF LT-7000 B O A R D ASSEMBLY.

1) M a k e sure power is turned off to Exposcop 7000 and w a l l outlet plug is disconnected.

2 ) Place a small foam or plastic sheet on top of the end of the C - A r c to support and isolate the
U2300 board assembly and the install cable from X4 of U45 board to X I of U2300 board making
sure connectors are correctly positioned.

3) Connect power cord to wall outlet of Exposcop 7000 and then turn on power. After 40 sec or so the
Exposcop 7000 should be ready to test U2300 board.

4 ) Press the laser key on the Exposcop 7000 control keyboard. The laser on L E D should light on the
control keyboard as w e l l as the U2300 board power on led's and laser's modules.

The lasers should come on for 1 minute then go off. The on time is preset by i n the software.
Pressing the laser key again w h e n the laser's are on should turn off the laser also.

FAILURE OF LASER POWER O N TEST.


If the laser fail to turn on y o u must determine the problem and repair as necessary. See
schematics for U2300 board and interface to Exposcop 7000 U45, U325 boards page 9.

5 ) After the power on test is found to be working correctly turn off power for Exposcop 7000 and
remove power cord form w a l l outlet.

6) Install the four 3" mounting standoff's by removing only one screw at a time from the collimator
assembly i n order not to disrupt the collimator alignment. See installation diagrams page 11
for further details of U2300 board.

7) Once the stand offs are correctly installed then mount the U2300 board to the standoff's. Then
Connect the cable from U45 X4, to X I of U2300 board.

8) Connect power cord to wall outlet of Exposcop 7000 and then turn on power.

After 40 sec or so the Exposcop 7000 should be ready to test LT-7000 U2300 board.

Exposcop 7000 LT-7000 Laser Assembly 09/23/97


International Medical System Installation Procedure Rev. 4-2
9) Press the laser key o n the Exposcop 7000 control keyboard. The laser on L E D should light o n the
control keyboard as w e l l as the U2300 board power o n led's and laser's modules.

The lasers should come on for 1 minute then go off. The on time is set b y the Z I E H M W I N S R V .
software. Pressing the laser key again w h e n the laser's are on should turn off the laser also.

FAILURE OF LASER POWER O N TEST.


If the laser fail to t u r n on y o u must determine the problem and repair as necessary. See
schematics page 9 for U2300 board and interface to Exposcop 7000 U 4 5 , U325 boards.
10 ) To align the laser line generators w i t h the center of the Image Intensifier place a jumper o n JT
service test p o i n t . This w i l l keep the laser line generators on until the jumper is removed.
Center the laser lines to cross i n the center of the Image Intensifier at 45 degrees.
See installation diagrams page 12 for further details of alignment of the laser assembly.

11) If the lasers turn o n and off correctly and the alignment is complete, turn off the power to the
Exposcop 7000 and remove power cord from the w a l l outlet.

12 ) Install the filter cap w i t h laser w i n d o w s to the generator cover and reinstall generator cover.

13 ) Connect power cord to wall outlet of Exposcop 7000 and then turn on power. After 40 sec or so the
Exposcop 7000 should be ready to test laser and I.I alignment. Check laser ah'gnment if it is
still aligned then turn off the power to the Exposcop 7000 and remove power cord from the w a l l
outlet.

14 ) Once the alignment is complete secure generator cover and seal w i t h white silicon sealer.

T h i s ends part two of the installation instructions for the LT-7000 laser assembly.

P A R T 3 L A B E L I N G E X P O S C O P 7000 W I T H L A S E R D E V I C E

INTRODUCTION:

Several labels are required for identification and for certification of the laser device.

1) The following example and location guides must be used i n placement of the labels to complete
the installation of the laser device i n the Exposcop 7000.

2 ) Install Caution Label 40-511-L10, Aperture Label 40-513 & 40-514-L10, Small C a u t i o n Label
40-512-L10 to control desk and Laser Caution W h e n Open Label 40-515-L10 as shown i n the
following diagrams pages 5 -6.

3) F i l l out the Manufacture certification label 40-510-L10 and included all information. Serial
number, M o t h and Year must be typed. The month and year of manufacture shall be p r o v i d e d
clearly and legibly, without abbreviation, and w i t h the year s h o w n as a four- digit number.

The month must be spelled out. i.e., January, February, June, etc... D O N O T A B B R E V I A T E .

The Year M u s t include all four digits, i.e., 1996,1997, etc. D O N O T A B B R E V I A T E .

The certification and manufacture label must be installed on the generator cover just above the
C - A r c attachment to the generator and must be positioned i n the m i d d l e of this area. Then
apply a clear label cover over the label to protect the information.

THIS C O M P L E T E S T H E I N S T A L L A T I O N A S S E M B L Y A N D T E S T I N G
O F T H E U2300/LT-7000 L A S E R D E V I C E F O R E X P O S C O P 7000

Exposcop 7000 LT-7000 Laser Assembly 09/23/97


International Medical System Installation Procedure Rev. 4 - 3
Position of Required Labels on Generator Covers.

CAUTION - Laser radiation


when open. DO NOT
CAUTION
STARE INTO BEAM

Located on rear of
Generator cover
m LASER RADIATION-DO NOT
STARE INTO BEAM

PEAK POWER
WAVELENGTH
lmW
635nm
CLASS I I LASER PRODUCT

LASER
APERTURE

CLASS n LASER PRODUCT


MODEL SERIAL No.
LT-7000

COMPLIES WITH DHHS RADIATION PERFORMANCE


STANDARDS, 21 CFR SUBCHAPTER J. AS OF DATE OF
MANUFACTURE

MANUFACTURED BY:
INTERNATIONAL MEDICAL SYSTEMS, INC
4181 L A T H A M STREET RIVERSIDE CA. 92501

MANUFACTURED:
MONTH: YEAR:

Mount Laser Caution Label


40-512-L10 Above Laser " O N '
Button.
SKIN C O N E

^tASER^
APERTURE

\ /
T
i—/
C-Arm Control LASEFf^
APERTURE

Exposcop 7000 LT-7000 Laser Assembly 09/23/97


International Medical System Installation Procedure Rev. 4 - 4
REMOVE ONLY THE
OUTSIDE SCREWS.

C A U T I O N - Laser radiation
BOTTOM OF when open. D O N O T
TUBE HEAD ASSEMBLY ^ | | ^ T A R E INTO B E A M

CUT SILICON SEAL


AROUND EDGE OF COVER
BEFORE LIFTING OFF

RESEALWITH
WHITE SILICON
SEAL WHEN
REINSTALLING
COVER

REMOVE ONLY THE


OUTSIDE SCREWS.
SIDE VIEW OF
TUBE HEAD ASSEMBLY

Exposcop 7000 LT-7000 Laser Assembly 01/11/97


International Medical System Service Installation Procedure Rev. 3-5
LASER CABLE ASSEMBLY

Complete Assembly Part Number


40-703-L10

Item # Part ID/Name Part # Description Quantity

1 Jl WM2001-ND 3 PIN FEMALE CONNECTOR 2

2 PINS WM2200-ND CRIMP PINS FOR FEMALE CONNECTOR 6


3 3 WIRES A2102W-100-ND 3 HOOK -UP WIRES ALPHA 3073-100 10" long 3
4 SPIRAL ELECTRONIC CABLE WIRE PROTECTION 10 "
WRAP WAREHOUSE

U45 J l U2300 J l

1 0 " / 25 mm

U45 to U2300 Board Cable Part


Number 40-703-L10 Assembly

The Exposcop 7000 U45 board provides the O N / O F F control and power for the U2300
Laser board through the interconnect cable.

Exposcop 7000 LT-7000 Laser Assembly 09/23/97


International Medical System Installation Procedure Rev. 4 - 7
1) File Off Tips Of Switch Cam Hush With Collimator Ring Assembly To
Avoid Hiting The Hex Studs.

Do not remove Switch Cam to file edge. Move ring assembly to allow
access to file edge of the Cam. Cover Tube head assembly parts to prevent
fillings from getting into tube assembly circuit boards and collimator.

Make sure no lose filings from the switch cam are left in the tube head
assembly.
See Example In
Dotted Lines

2) When Installing Hex Studs Align The Flat Edge Of


All Four (4) Hex Studs With The Side Of The
Collimator Ring To Avoid Interference Of The
Studs And The Ring When Slot Collimator is
Rotated

Small Caution Ul 1 _L_ l • '


Laser Label CAUTION u — u
Must be Attached LASER RADIATION-DO NOT
U ul •
to Laser Board
FIGURE 1
STARE INTO BEAM

Assembly PEAK POWER 1mW


1
e=3 5
i—i I— 1 1
#40-512-L10 WAVELENGTH 635nm
CLASS II LASER PRODUCT

LASER
4mm Allen Head Screws
APERTURE
Removed From The
Colimator Assembly
Small Laser
4mm Flat Washer Supplied With
Aperture Label
Laser Kit U45 Board
Must be Attached
LT-7000 / U2300 Power X01 Connector
Between Laser
Assembly Board
Module Assembly
Note: #40-514-L10
4mm X 3" Hex Stud
[Remove Only One Collimator
Assembly Screw And Replace With
Hex Stua At A Time To Prevent
Collimator Misalignment ]

Collimator
Assembly

Exposcop 7000 LT-7000 Laser A s s e m b l y 09/23/97


International M e d i c a l System Installation Procedure Rev. 4 - 8
By carefully turning the laser with the center slot it can be
twisted to intersect with the other axis laser line generator
to form a cross at the center of the grid aligned with the
grid mounting screws.
Once lines are correct use retaining screw
on side of laser mount to secure laser.
WARNING DO NOT OVERTIGHTEN RETAINING
SCREW AS LASER MODULE WILL BE D A M A G E .

By adjusting the three screws the laser can be tilted


to align with the I.I grid mounting screws to provide
a beam cross as low as possible above the skin cone.
Tilt Adjustment Tilt down The inside two screws should be tightened first to tilt
Screws the laser module inward toward the center of the
board. However be carefull not to adjust them to
much.

The retaining screw is used to slide the laser along a


limited distance slot to center the laser to the ouput
glass and help align it with the center of the II.

Laser mount and


Screw Lock plate

Exposcop 7000 LT-7000 Laser Assembly 09/23/97


International Medical System Installation Procedure Rev. 4 - 9
LT-7000
L A S E R POSITIONING D E V I C E
Complete Assembly Part Number
40-710-L10
Part List For Laser Assembly
Item # Part ID/Name Part # Description Quantity
1 PC -BOARD 40-236-L10 BLANK PC LASER BOARD 2
2 Rl, R9 390Q 390 OHM RESISTOR 1
3 R2 560Q 560 OHM RESISTOR 1
4 C2 0,luF 0.1 uF CAPACITOR 1
5 D2 1N4001 STEERING DIODE 1
6 D3, D4 40-702-L10 VLM3 - l.OmW 635nm LASER M O D U L E 2
7 D5, D6 40-040-D10 LED 1 RED [ POWER ON LED ] 2
8 Fl P822-ND 250 mA MICRO FUSE 1
9 Laser Label 40-510-L10 Certification Label 1
10 Laser Label 40-511-L10 Caution Label Top Generator Cover 1
11 Laser Label 40-512-L10 Caution Small- Laser board and Control Desk 2
12 Laser Label 40-513-L10 Aperture Label Skin Cone 3

13 Laser Label 40 -514-L10 Aperture Label Top Generator Cover 1

14 Laser Label 40-515-L10 Caution When Open Label bottom Generator

Exposcop 7000 LT-7000 Laser Assembly 09/23/97


International Medical System Installation Procedure Rev. 4 -10