Certificate of Calibration

Manufacturer: Atago Report #: TME-0-22140798-R1


Description: Refractometer P.O. #: N/A

Model #: PAL-1 Procedure: Manufacturer Spec


Serial #: L318052 Asset #: GAM008195

Calibration Lab.: Mexicali, BC


Customer: Goodrich Aerospace de Mexico
Calzada Venustiano Carranza #238
Mexicali, BC 21397

This certifies that the above instrument was calibrated in compliance with Techmaster de Mexico's Calibration Systems
Requirement in accordance with referenced procedures. Standards used to perform this calibration are traceable to SI units;
their source of traceability derives from a National Metrology Institute such as NIST, CENAM, NPL, DIN, from natural
physical constants, consensus standards or derived by the ratio type of calibrations. Collective uncertainties are determined
as required with a distribution that corresponds to a probability of approximately 95% (k=2). Uncertainties are not taken
into account in determining In-Tolerance results. All calibrations are performed to manufacturer's specifications unless
otherwise noted.
This form shall not be reproduced, except in full, without the expressed written consent of Techmaster de Mexico SA de CV

Condition Returned: Calibrated With Data per Customer Contract Calibration Cycle: 3 Months
Returned within specifications?: YES Calibration Date: Mar/25/2019
Temperature: 25ºC Relative Humidity: 44% Next Due Date: Jun/25/2019
Calibrated On Site? No Technician: Gerardo Orduño Beltrán

Standard # Manufacturer Model # Due Date Test Report #


137-0533 Scientech SA525D Sep 05, 2019 TDM-5-22152600-R1
157-0048 Fluke 754 May 17, 2019 TDM-0-22077231
157-0089 Misco PA202 Nov 16, 2020 TDM-0-22137803-R2

Special Remarks or Requirements:

G. Orduño
Ernesto Matamoros Ing. Mauricio Garayzar Gerardo Orduño Beltrán
Quality Assurance Technical Manager Approved By

Tijuana, B.C. Calle Seminario No. 8610-11, Col. Niños Héroes Este, CP 22120, Tijuana, BC, México Tels.: (664) 624-4444 / 624-3128 / 624-4445 / 624-0440

Mexicali, B.C. Calle Gales Nº 1201, Col. Villa Fontana, C.P. 21180, Mexicali, BC, México Tels.:(686) 555-1660 / 557-6117 / 555-1766

Cd. Juárez,Chih. Blvd. Gomez Morin No.9050-L8, Col. Partido Senecu, CP. 32469, CD. Juarez, Chih. México. Tels.:(656) 687-2471 / 648-1181

Monterrey, N.L. Ave. Morones Prieto No. 914, Oriente Int. 112, Col. La Huerta, C.P. 67144. Tels.:(81)1334-0701 / 1334-0702
540.1 1901

Page 1 of 2
3
TME-0-22140798-R1 20
20
Calibration Results
Manufacturer: Atago Asset No. GAM008195

Description: Refractometer Serial No. L318052

Model #: PAL-1 Report No. TME-0-22140798-R1

Step # Function Tested / Nominal Value As Found ± MU* In Tol As Left ± MU* In Tol Calibration Tolerance
1 BRIX % 0 % 0 ± N/A YES 0 ± N/A YES 0 to 0 %
2 5.0 % 5.0 ± N/A YES 5.0 ± N/A YES 4.8 to 5.2 %
3 9.8 % 9.7 ± N/A YES 9.7 ± N/A YES 9.6 to 10.0 %
4 15.8 % 15.7 ± N/A YES 15.7 ± N/A YES 15.6 to 16.0 %
5 21.8 % 21.7 ± N/A YES 21.7 ± N/A YES 21.6 to 22.0 %
6 25.5 % 25.4 ± N/A YES 25.4 ± N/A YES 25.3 to 25.7 %
7 Temperature 25.0 °C 24.800 ± 0.068 YES 24.800 ± 0.068 YES 24.0 to 26.0 °C

*MU stands for Measurement Uncertainty at a 95% confidence level.

*********************************End of Report***********************************

Calibracion apegado al standard AMS2770 seccion 3.4.9.3.6 Metodo B por requerimiento del cliente

Manufacturer Spec
540.6 1503

Page 2 of 2
3
TME-0-22140798-R1 20
20
Curva de Calibración
Brix (%)
Concentración ( g)
0 0
4.99 5
9.99 9.8 Pendiente = 0.8441
16.99 15.8 Intersección = 0.7467
25.00 21.8 Coeficiente de correlación = 0.9963
30.00 25.5

Curva de Calibración
f(x) = 0.8441907151x + 0.7467889174
R² = 0.9963311448
Medición Brix %

Concentración Sacarosa

Page 3 of 3
TME-0-22140798-R1 20