Beruflich Dokumente
Kultur Dokumente
User Manual
English V 5.2
User Manual
Table of Contents
1. Introduction ...................................................................................... 1
1.1. General ....................................................................................... 1
1.2. Symbols ...................................................................................... 1
1.3. Trademarks .................................................................................. 2
1.4. Product identification ...................................................................... 3
2. Description of the system .................................................................. 5
2.1. Z+F IMAGER® 5006h .................................................................... 7
2.2. Z+F IMAGER® 5006h PowerSupply ................................................. 8
2.3. 5006 Li-Ion Charger and 5006 Li-Ion Battery ....................................... 9
3. Setting up the Z+F IMAGER® 5006h ................................................... 11
3.1. Z+F IMAGER® 5006h setup .......................................................... 11
3.1.1. Setting up the instrument with the dolly ...................................... 11
3.1.2. Setting up the tripod without the dolly ........................................ 11
3.2. Setup step by step ...................................................................... 12
3.3. Power supply .............................................................................. 13
3.3.1. Operating the Z+F IMAGER® 5006h with the 5006 Li-Ion Battery .... 14
3.3.2. Operating the Z+F IMAGER® 5006h with Power Supply ................. 18
3.4. Charging .................................................................................... 20
3.4.1. Charging the 5006 Li-Ion Battery .............................................. 20
3.5. USB-ports (for external sensors) ..................................................... 21
4. Preparing to scan ............................................................................ 23
4.1. Additional equipment .................................................................... 23
4.2. Targets ...................................................................................... 23
4.2.1. Z+F Paper Targets ................................................................ 23
4.2.2. Z+F Profi Targets .................................................................. 24
4.2.3. Z+F Auto Targets .................................................................. 25
4.3. Positioning the Z+F IMAGER® 5006h .............................................. 26
4.4. Ambient conditions ...................................................................... 27
4.4.1. Unfavorable surfaces .............................................................. 27
4.4.2. Unfavorable weather conditions ................................................ 27
4.4.3. Extremely high / low temperatures ............................................ 28
4.5. Miscellaneous ............................................................................. 29
5. Operating the Z+F IMAGER® 5006h ................................................... 31
5.1. Z+F IMAGER® 5006h controls ....................................................... 32
5.1.1. Z+F IMAGER® 5006h control button ......................................... 32
5.1.2. Menu control ........................................................................ 33
5.2. Connections ............................................................................... 52
5.2.1. Connecting the Z+F IMAGER® 5006h to a network ....................... 53
5.2.2. Connecting the Z+F IMAGER® 5006h to a notebook/PC ................ 54
iii
User Manual
iv
User Manual
v
User Manual
vi
User Manual
List of Figures
2.1. Description of the system ................................................................... 6
2.2. Z+F IMAGER® 5006h ....................................................................... 7
2.3. Power Supply .................................................................................. 8
2.4. 5006 Li-Ion Charger and 5006 Li-Ion Battery .......................................... 9
3.1. Z+F IMAGER® 5006h with the 5006 Li-Ion Battery ................................ 14
3.2. Power Supply ................................................................................ 18
3.3. Charging the 5006 Li-Ion Battery ....................................................... 20
3.4. USB-ports (for external sensors) ........................................................ 22
4.1. Z+F Paper Targets .......................................................................... 23
4.2. Z+F Profi Targets ........................................................................... 24
4.3. Z+F Auto Targets ........................................................................... 25
4.4. Positioning the Z+F IMAGER® 5006h ................................................. 27
5.1. Example display main menu ............................................................. 33
5.2. Z+F IMAGER® 5006h X- and Y-direction ............................................. 36
5.3. WLAN connection ........................................................................... 52
5.4. Network connection ......................................................................... 53
5.5. Select wireless network connection .................................................... 56
5.6. Select Z+F IMAGER® 5006h and connect ........................................... 57
5.7. Scanner web-browser ...................................................................... 58
5.8. Select Z+F IMAGER® 5006h and connect ........................................... 59
5.9. Connect anyway ............................................................................. 60
5.10. Connection established .................................................................. 60
5.11. Connection successfully ................................................................. 61
5.12. Scanner web-browser .................................................................... 62
5.13. Network and Sharing Center ........................................................... 63
5.14. Manage Wireless Networks ............................................................. 64
5.15. Check check boxes ....................................................................... 64
5.16. Scanner web-browser .................................................................... 66
5.17. cmd ............................................................................................ 69
5.18. IPconfig ................................................................................... 69
5.19. cmd ............................................................................................ 70
5.20. ping .......................................................................................... 70
5.21. ping connection failed! .................................................................. 71
5.22. Info ............................................................................................ 73
5.23. Status ......................................................................................... 74
5.24. Scan ........................................................................................... 75
5.25. Scan management ........................................................................ 80
5.26. Configuration ................................................................................ 81
5.27. Debug ......................................................................................... 82
5.28. Camera ....................................................................................... 83
vii
User Manual
viii
User Manual
List of Tables
1.1. Symbol descriptions .......................................................................... 2
1.2. Product identification ......................................................................... 3
4.1. Recommended target distance (DIN A4 target) ..................................... 26
5.1. Selecting the resolution level ............................................................. 42
5.2. Resolution levels and corresponding scanning durations ......................... 43
8.1. Z+F IMAGER® 5006h laser classification ........................................... 105
ix
User Manual
x
Introduction
Chapter 1. Introduction
1.1. General
The Zoller + Fröhlich GmbH is one of the leading companies in the field of contact-
free laser measurement technology worldwide. Z+F is the manufacturer of the 3D-
Laser Scanning system Z+F IMAGER® 5006h.
Laser scanning is an automatic process during which real objects are surveyed and
sampled almost completely to determine their location, size, orientation and form. The
Z+F IMAGER® 5006h rapidly determines high-resolution measurement results within
very short time and within a field of view of max. 360° horizontal and 320° vertical
(about 40° are shadowed below due to the Laser scanner`s assembly and its tripod.
With a max. sampling rate of 1.016.000 pixel per second, the amount of generated
data can add up to theoretically 5 billion points per panoramic scan.
Of course it is possible to reduce the amount of data to later work with. The captured
data, three dimensional coordinates and intensity information of the scanned objects
can than be further post - processed, analyzed and evaluated, using special software,
like e.g. Z+F LaserControl®.
1.2. Symbols
This manual contains important safety directions as well as instructions for setting up
the product and operating it.
1
Introduction
Caution, Danger
Indicates an imminently hazardous situation which, if not
avoided, might result in death or serious injury.
Warning
Indicates a potentially hazardous situation or an unintended use
which, if not avoided, might result in death or serious injury.
Important
Indicates a potentially hazardous situation or an unintended
use which, if not avoided, might result in minor or moderate
injury and/or appreciable material, financial and environmental
damage.
Tip
Important paragraphs which must be adhered to in practice as
they enable the product to be used in a technically correct and
efficient manner.
1.3. Trademarks
Windows® is a registered trademark of Microsoft Corporation. All other trademarks
are the property of their respective owners.
Tip
It is recommended to set-up the instrument while reading this manual.
This manual applies to all Z+F IMAGER® 5006h instruments.
2
Introduction
Email: <support@zf-laser.com>
Email: <service@zf-laser.com>
Web: www.zf-laser.com
3
Introduction
4
Description of the system
• Power Supply
• Power cable
• Tribrach (Wild-System)
• Cleaning set
• User manual
Accessories (optional)
• Notebook/PDA
• Dolly
5
Description of the system
6
Description of the system
7
Description of the system
Please use the Power Supply and 5006 Li-Ion Charger only at dry locations.
8
Description of the system
9
Description of the system
10
Setting up the Z+F IMAGER® 5006h
• Set the lock to the LOCKED position. The spread is limited to a 30° angle.
or
• Unlock the legs and then adjust their angle. Assemble the chain with the chain clips
on the clamps.
11
Setting up the Z+F IMAGER® 5006h
Tip
It is always recommended to shield the Z+F IMAGER® 5006h from
direct sunlight and avoid uneven temperatures around the instrument.
The Z+F IMAGER® 5006h should always be set up on its tripod. Using
the tripod delivered with the scanning system guarantees maximum
stability during scanning operations.
Please do not put the Z+F IMAGER® 5006h directly on the ground
while scanning!
1.)Extend the tripod legs to the same length to allow for a comfortable working posture.
Tighten the screws at the bottom of the legs. Fold the handles down to detain the
length of the legs.
2.)Place the tribrach on the tripod and secure it with the central fixing screw.
12
Setting up the Z+F IMAGER® 5006h
5.)Level the tribrach using the circular level. Turn two of the leveling screws
simultaneously in opposite directions. The index finger of your right hand indicates
the direction in which the bubble should move. Now use the third leveling screw
to center the bubble.
6.)Place the instrument on the tribrach and secure it with the tribrach’s locking knob.
Make sure that the instrument is leveled by checking the built-in bubble level.
Tip
It is recommended to set up the Z+F IMAGER® 5006h horizontally by
using the tripod screws. Subsequently the established horizontation
can be refined using the integrated electronic level. If the incline is
more than 0,5°, a warning message appears before scanning.
2.)Power Supply
13
Setting up the Z+F IMAGER® 5006h
1.)Hold the 5006 Li-Ion Battery with both hands and press the fixing clips.
2.)Push the 5006 Li-Ion Battery upwards to attach it with the help of the four aligning
pins and the connector plug.
1.)Hold the 5006 Li-Ion Battery with both hands and press the fixing clips.
• Before using the 5006 Li-Ion Battery for the first time, it should be completely
charged.
• The 5006 Li-Ion Battery can be removed for changing if an external power supply
(Power Supply) is connected.
14
Setting up the Z+F IMAGER® 5006h
Important
Never remove the 5006 Li-Ion Battery while the Z+F IMAGER® 5006h
is switched on.
• It is normal for the 5006 Li-Ion Battery to become warm during the
charging process. Using the chargers recommended by Z+F, it is
not possible to charge the 5006 Li-Ion Battery if the temperature is
too high – the charging process is aborted then.
Tip
A 5006 Li-Ion Battery should always be attached to ensure the
optimum weight balance of the Z+F IMAGER® 5006h. If there is no
removable 5006 Li-Ion Battery connected, a warning message will
appear.
3.3.1.1. Period of use and life span of the 5006 Li-Ion Battery
• The period of use depends on the electrical load, temperature and the age. It lasts
about 2.5 hours.
15
Setting up the Z+F IMAGER® 5006h
• If the 5006 Li-Ion Battery is stored for a longer period of time, this should be done
when it is charged not more than 60-70% of its capacity (i.e. up to the third diode
indicated) on the 5006 Li-Ion Charger.
• After a longer period of storage, the 5006 Li-Ion Battery has to be recharged before
using it. As soon as the minimum level of the operating voltage of 12V-DC is
reached, the 5006 Li-Ion Battery has to be recharged.
Important
• It is not recommended to store a discharged 5006 Li-Ion Battery
(Caution of depth discharge). Storing the 5006 Li-Ion Battery pack
in such a state can damage it permanently.
Warning
16
Setting up the Z+F IMAGER® 5006h
• The accurate polarity of the 5006 Li-Ion Battery and the 5006 Li-Ion
Charger must be assured.
• If the 5006 Li-Ion Battery breaks and electrolyte gets in contact with
skin or clothes, it has to be washed with water immediately. In case
of contact with eyes, please consult a physician immediately.
Important
When transporting or shipping batteries, the person in charge of the
product must ensure that the applicable national and international
rules and regulations are observed. Before transportation or shipping,
contact your local passenger or freight transport company.
17
Setting up the Z+F IMAGER® 5006h
b.)Power Supply
3.)Connect the Z+F IMAGER® 5006h and the Power Supply with the
Z+F IMAGER® 5006h power supply cable.
18
Setting up the Z+F IMAGER® 5006h
Important
• The Power Supply is intended for indoor usage. Use the Power
Supply only at dry locations.
19
Setting up the Z+F IMAGER® 5006h
1.)Place the 5006 Li-Ion Battery onto the 5006 Li-Ion Charger with the help of the
aligning pins.
5.)The 5006 Li-Ion Charger’s battery status indicators will indicate the current battery
charging status after an initialization period of approximately 30 seconds.
20
Setting up the Z+F IMAGER® 5006h
• Red LED (Error LED): the red LED illuminates if there is a fault in the power supply.
• If all four LEDs flash continuously then no 5006 Li-Ion Battery is inserted in the
5006 Li-Ion Charger.
The overall charging capacity of the 5006 Li-Ion Battery is divided into quarters. Each
LED corresponds to a quarter:
• If the 5006 Li-Ion Battery charging state is low the appropriate LED flashes slowly.
• If an LED is illuminated constantly the currently charged quarter of the 5006 Li-Ion
Battery has reached the appropriate level of charge.
• When all four LEDs are illuminated constantly the 5006 Li-Ion Battery is completely
charged.
Important
Use the Power Supply and the 5006 Li-Ion Charger only at dry
locations.
21
Setting up the Z+F IMAGER® 5006h
To attach the cover please press it carefully into the recess, till PRESS aims to top.
To remove the cover, please push the PRESS button till the cover tips.
22
Preparing to scan
• Laptop
4.2. Targets
• Checkerboard pattern (2 white and 2 dark grey squares, tilted by 45°), not shiny
or reflecting.
• It is recommended to print the dark grey squares on white paper with a laser printer.
The targets should be laminated using matt sleeves for best results (an A4 target
master file (*.doc) can be obtained from Z+F). The printed target should be testet
for using in the field. Z+F does not guaranteed for success of printed targets.
23
Preparing to scan
Tip
Mount / fix the targets only on flat surfaces if possible. Columns or
pipes are not recommended.
• Checkerboard pattern (2 white and 2 dark grey squares, tilted by 45°), not shiny
or reflecting.
• Diameter 20cm
• Since the target centre is tilt able, it can be measured perfectly with the
Z+F IMAGER® 5006h as well as with the total station. Before starting a scan,
always orientate the Z+F Profi Target towards the Z+F IMAGER® 5006h / total
station.
24
Preparing to scan
• Checkerboard pattern (2 white and 2 dark grey squares, tilted by 45°), not shiny
or reflecting.
• Well noticeable serial target numbers for checking and georeferencing during post-
processing
• Use non-reflective tape to attach the targets. Outer ring and reflector marker must
be visible and not covered by the tape.
25
Preparing to scan
• Ensure that the angle of incidence of the laser beam on the target is larger than 45°.
If you use “Z+F LaserControl®” for registration, it is automatically checked when
starting the target-finder if the target is well visible and can be recognized properly.
You're notified, if the angle of incidence is too steep.
• Self printed targets must have a minimum black/white contrast of 6:1 in the scan.
Please test this by frontal (angle 90°) scanning a target and analyse the contrast
of the target in your Software.
• The recommended distance between Z+F IMAGER® 5006h and a target (Size 210
mm x 298 mm) depends on the scan resolution.
Tip
With a smaller angle, the possible target distance is reduced. The
maximum angle based on the normal of target should be less than
±45°.
26
Preparing to scan
Tip
To improve the results of the measurement, such surfaces should be
colored or powdered.
• If some objects are scanned against the sunlight or a bright spotlight, the optical
receiver of the Z+F IMAGER® 5006h can be dazzled so heavily that in this area no
measured data is recorded. A “black hole” appears in the reflectance image.
27
Preparing to scan
Important
• Fog, rain or snow will cause poor measurement results, therefore
surveying during these conditions is not recommended!
To avoid damaging the electronic parts of the Z+F IMAGER® 5006h, it switches off
if the upper temperature limit is exceeded.
Important
If the Z+F IMAGER® 5006h is brought from cold into a warm and
humid environment, the glass window at the mirror or in extremely
cases even the interior optic can fog up. This may cause measurement
errors.
Important
Soiling the glass pane of the mirror (i.e. by a layer of dust, condensed
water or fingerprints) might cause considerable measuring errors.
Before measuring check the glass pane for soiling and clean it with
the cleaning set if necessary.
28
Preparing to scan
• Field notes about the Z+F IMAGER® 5006h: Note the position of the
Z+F IMAGER® 5006h inside the measuring object (e.g. factory or shed plan).
• Before initiating the scanning process, please check if there is enough space on
the internal flash disk (20 to 40 GB per day, depending on your planning).
29
Preparing to scan
30
Operating the Z+F IMAGER® 5006h
Important
• Check the lens to assure its perfect clean.
• Press and hold the ON/OFF button for at least 0.3 seconds.
• Press and hold the ON/OFF button for at least 0,5 seconds.
• In the event of a system crash, press the ON/OFF button for at least 5 seconds.
The system will then switch off.
31
Operating the Z+F IMAGER® 5006h
Security PIN:
To protect the Z+F IMAGER® 5006h against unauthorized use, it is
possible to generate a PIN-Code which needs to be entered in the
scanner-display after every Z+F IMAGER® 5006h start up. Otherwise
it is not possible to work with the scanner.
entered, push to get into the scanner menu. If the user doesn’t
want to work with a security-pin, the PIN in the config menu must be
set to 0.
Increase a parameter.
32
Operating the Z+F IMAGER® 5006h
Reduce a parameter.
• Info
• Status
• Tilt Sensor
• Scanning
• Scanmanagement
Example display:
The top line indicates the menu currently in use by the Z+F IMAGER® 5006h.
33
Operating the Z+F IMAGER® 5006h
If this symbol appears in the bottom line you can page through the
submenu using .
01.01.2009
Date and time display.
12:00:00
Serial No. of the Z+F IMAGER® 5006h / Serial No. of the
No.999/999999
LARA system.
IP
IP-Address of the Z+F IMAGER® 5006h.
172.20.255.255
Ver 7.3.1.0 Firmware version
Z+F IMAGER 5006h Model designation.
Hard Im2.0 Hardware implementation.
34
Operating the Z+F IMAGER® 5006h
Justification: 5
of 200 days Tip
A workday is counted if on this day more
than 3 scans were taken.
35
Operating the Z+F IMAGER® 5006h
The Z+F IMAGER® 5006h is equipped with an electronic biaxial tilt sensor, to
determine the horizontal level of the Z+F IMAGER® 5006h, and to detect inclinations
while scanning. In the menu level the Z+F IMAGER® 5006h can be levelled with
the help of the three adjusting screws on the tribrach.
H: 123.458 deg Press for 180° position and for the 0° position of the
Z+F IMAGER® 5006h. If you have the x and y level values
differ in both positions(0° and 180°), you can half the values
and do the alteration as follows.
36
Operating the Z+F IMAGER® 5006h
37
Operating the Z+F IMAGER® 5006h
V 25 - 335 You can switch between the predefined scan ranges using
/ .
V 0-360
V 25-260 200-335
V 25-180
38
Operating the Z+F IMAGER® 5006h
V 180-335
V 100-260
39
Operating the Z+F IMAGER® 5006h
40
Operating the Z+F IMAGER® 5006h
41
Operating the Z+F IMAGER® 5006h
Press to start the scanning process with the previously defined scan resolution
and laser power.
A current scan process can be also stopped with this key. All up to this time seized
data are seized and on the internal hard drive stored.
Warning
During the operation of the product there is a hazard of squeezing
extremities or entanglement of hairs and/or clothes by rotating parts.
Precautions:
V = 0,009°
1
Super High H = 0,018° 20.000 > 20 m
V = 0,018°
2
High H = 0,036° 10.000 >5m
V = 0,036°
Middle H = 0,072° 5.000 >2m
V = 0,072°
3
Preview H = 0,288° 1.250 -
V = 0,228°
1
Low distances to the object and high resolutions result in considerable overlapping
of neighboring measurement points, hence high resolutions do not improve image
sharpness. Therefore high resolutions are only suitable for long distances to the
object.
42
Operating the Z+F IMAGER® 5006h
1
Low distances to the object and high resolutions result in considerable overlapping
of neighboring measurement points, hence high resolutions do not improve image
sharpness. Therefore high resolutions are only suitable for long distances to the
object.
4
The real file size depends on the compression. These are the maximum values for
scans with setting vertical 360° / horizontal 182°.
43
Operating the Z+F IMAGER® 5006h
Measurement noise:
Measurement noise is the spreading of measurement results
regarding middle or a reference result. These are only random minor
mistakes, they are not consistent.
Using the Scan management menu you can copy or move data from the internal
hard drive to an external hard drive.
Start move Starts the move process of the actual list and the list
is automatically removed from the internal hard drive.
44
Operating the Z+F IMAGER® 5006h
Caution
In these menus important parameters can be adjusted, therefore
it is important which adjustments are changed. If adjustments are
changed carelessly, functions may be limited.
An audible signal sounds if you press for more than eight seconds. The menu
expands to show the following functions:
• Manual operation
• Settings
• Updating
Every line is filled with a double line and the name of the menu point. Press once
more for three seconds to return to the main menu.
45
Operating the Z+F IMAGER® 5006h
V ABI R 10 rps ± Set the values using / , starts or stops the motor
at the defined speed.
46
Operating the Z+F IMAGER® 5006h
io -23 24 io means “OK”. The values are under the limit of 750inc.
io 58 342
nio -1672 845
nio means “not ok”. Clean the optic and the reference plate
nio 958 1025 with the delivered cleaning set before retesting the backlight
(see Section 6.3, “Cleaning and drying” [92]).
nio 1112 798
47
Operating the Z+F IMAGER® 5006h
Warning
During the operation of the product there is a hazard of squeezing
extremities or entanglement of hairs and/or clothes by rotating parts.
Precautions:
48
Operating the Z+F IMAGER® 5006h
49
Operating the Z+F IMAGER® 5006h
255.255.0.0 Example:
50
Operating the Z+F IMAGER® 5006h
with / .
The reorganisation will be done by using a red dot on the
red bubble grey card (grey card needs to be on the scanner (see color
manual)
circle Recognisation by using a black circle.
intens Recognisation by using the intensity of the grey card.
no No check.
51
Operating the Z+F IMAGER® 5006h
This menu allows to update the firmware of the Z+F IMAGER® 5006h.
Caution
Connect the Z+F IMAGER® 5006h with
the Power Supply. Do not switch off
Check if new
the Z+F IMAGER® 5006h during the
update.
5.2. Connections
The Z+F IMAGER® 5006h is equipped with a network interface and a W-LAN
interface. These interfaces allow external control of the Z+F IMAGER® 5006h.
As the Z+F IMAGER® 5006h has its own IP address, the device can also be controlled
over the network. The collected data is stored on the internal flash disk or onto a
connected USB-stick. The flash disk does have a standard capacity of 64 GB and
can store data from two days of continuous scanning. The captured data can be
transferred to a Laptop/PC by Ethernet cable. The data can also be transferred by
USB 2.0 to an external hard drive.
52
Operating the Z+F IMAGER® 5006h
•
Expand the main menu. Press for 5 sec after system start-up (see
Section 5.1.2.2, “Extended Menu” [45])
•
Select Settings in the main menu using .
•
Select yes in the DHCP Client submenu. Confirm with
•
Select no in the DHCP Server submenu. Confirm with
•
Select yes in the HTTP Server submenu. Confirm with
53
Operating the Z+F IMAGER® 5006h
• Enter the IP address of the Z+F IMAGER® 5006h into your web browser. For
example: http//172.20.0.251.
Using the web browser the operator has full access to the web server of the
Z+F IMAGER® 5006h. Enter the IP-address of the Z+F IMAGER® 5006h into the
web browser: http://<IP>.
Furthermore the operator has access to the FTP server of the Z+F IMAGER® 5006h.
Enter the IP-address of the Z+F IMAGER® 5006h into the FTP-client: ftp://<IP>.
In the web browser/FTP client the scan data as well as its sub folders are shown.
A preview of every scan is also displayed. All scan data can be downloaded onto a
notebook/PC or an external hard drive for further usage.
• Connect the Z+F IMAGER® 5006h to the notebook/PC with a crossover Ethernet
cable or with W-LAN. The LAN connection should be activated automatically
on the PC otherwise check the Ethernet settings. In the network properties,
the “internet protocol (TCP/IP)” has to be activated. Make sure that the option
“source IP address automatically” and “source DNS address automatically” is
checked there.
• Switch on the Z+F IMAGER® 5006h (press the On/Off button for 0.3 sec).
•
Expand the main menu. Press for 5 sec after system start-up (see
Section 5.1.2.2, “Extended Menu” [45])
•
Select Settings in the main menu using .
•
Select no in the DHCP Client submenu. Confirm with
54
Operating the Z+F IMAGER® 5006h
•
Select yes in the HTTP Server submenu. Confirm with
On the Laptop you can access the data on the Z+F IMAGER® 5006h using the
web browser or an FTP client:
• Enter the IP address of the Z+F IMAGER® 5006h into your web browser.
http://<IP> , or
In the web browser the operator has full access to the web server of the
Z+F IMAGER® 5006h. Enter the IP-address of the Z+F IMAGER® 5006h into the
web browser: http://<IP> .
Furthermore the operator has access to the FTP server of the Z+F IMAGER® 5006h.
Enter the IP-address of the Z+F IMAGER® 5006h into the FTP-client: ftp://<IP>
. In the web browser/FTP client the scan data as well as its sub folders are shown.
A preview of every scan is also displayed. All scan data can be downloaded onto a
notebook/PC or an external hard drive for further usage.
Tip
The firewall of the notebook/PC should be deactivated to avoid
disturbing the connection to the scanner.
Example: ftp://scans@<IP>
55
Operating the Z+F IMAGER® 5006h
Important
There is a limit of 10000 files per folder.
Tip
The Z+F IMAGER® 5006h supports W-LAN.
56
Operating the Z+F IMAGER® 5006h
57
Operating the Z+F IMAGER® 5006h
Tip
The W-LAN connection can be encrypted in the configuration of the
Z+F IMAGER® 5006h.
58
Operating the Z+F IMAGER® 5006h
Tip
The Z+F IMAGER® 5006h supports W-LAN.
4.)An error-message will appear, because the Z+F IMAGER® 5006h is classified
automatically as an “unsecured network”. Connect anyway.
59
Operating the Z+F IMAGER® 5006h
60
Operating the Z+F IMAGER® 5006h
61
Operating the Z+F IMAGER® 5006h
Tip
The W-LAN connection can be encrypted in the configuration of the
Z+F IMAGER® 5006h.
62
Operating the Z+F IMAGER® 5006h
Tip
The Z+F IMAGER® 5006h supports W-LAN.
4.)Click on “Manage Wireless Networks”, pick the displayed Z+F IMAGER® 5006h
and double-click it:
63
Operating the Z+F IMAGER® 5006h
64
Operating the Z+F IMAGER® 5006h
65
Operating the Z+F IMAGER® 5006h
Tip
The W-LAN connection can be encrypted in the configuration of the
Z+F IMAGER® 5006h.
66
Operating the Z+F IMAGER® 5006h
Tip
The Z+F IMAGER® 5006h supports W-LAN.
67
Operating the Z+F IMAGER® 5006h
68
Operating the Z+F IMAGER® 5006h
IPconfig command:
1.)Windows Start―►run
Ping command:
1.)Windows Start―►Run
69
Operating the Z+F IMAGER® 5006h
In case there is 0%, both PC and scanner are connected properly. If 100 % shows
up (see next image), there the connection failed.
70
Operating the Z+F IMAGER® 5006h
Tip
If there is no connection, the cables should be checked. The cables
need to be in good condition and the lights of the cable connection at
the notebook should be green.
• PC: Enter the IP address of the Z+F IMAGER® 5006h into the address field of your
web browser. Refer to IP-address under “Settings”. The main menu is displayed
then.
71
Operating the Z+F IMAGER® 5006h
Scanner No.
Info
Status
Scan predefined
Scan
Scan management
Help
Tip
If the IP address is extended to show the menus shortcut commands,
you can go directly to the desired page.
72
Operating the Z+F IMAGER® 5006h
73
Operating the Z+F IMAGER® 5006h
74
Operating the Z+F IMAGER® 5006h
75
Operating the Z+F IMAGER® 5006h
Middle
Refer Section 9.1.3, “Resolutions” [113] and for further
High
information.
Super High
Ultra High
High level “high” reduces range noise in the point-cloud by the factor
1,4 but doubles the scanning time).
Premium
76
Operating the Z+F IMAGER® 5006h
The monitored area for low power is lower than the space
the Z+F IMAGER® 5006h needs anyway. However, due to the
lower laser performance, the range noise of the measurement
points on the object's surface will be greater.
77
Operating the Z+F IMAGER® 5006h
78
Operating the Z+F IMAGER® 5006h
Predefined settings:
1.)visible 20-340
2.)full 0-360
3.)left 20-180
Vertical
4.)right 180-340
5.)top 160-200
6.)user:
Start Scan starts the scan process with all settings listed above. A running scan
can be stopped any time without losing data. All data will be stored on the internal
hard disc.
Important
During the operation of the product, extremities or hairs might get
squeezed and/or clothes and entangled by rotating parts.
Precautions:
The operator obtains information regarding the size of the file and how many files are
stored there. Choose one directory; this can be completely deleted by entering I am
sure and clicking the button delete all scans.
79
Operating the Z+F IMAGER® 5006h
Important
Save all your data before deleting!
80
Operating the Z+F IMAGER® 5006h
Important
Wrong configurations can impair the operability of the
Z+F IMAGER® 5006h. Changing the configuration parameters only
on the operators own responsibility!
81
Operating the Z+F IMAGER® 5006h
If the scanner is in standby mode the internal components are switched off to save
power, depending on the power management. To test separate parts (e.g. motor)
power needs to be turned on by pressing the button “Power On”. Only then it is
possible to access the web page of the part which needs to be tested (e.g. motor).
For that press the respective button in the row (e.g. motor)
Tip
There is a special manual for the camera mode. This manual needs
to be ordered separately from Z+F support ( support@zf-laser.com).
82
Operating the Z+F IMAGER® 5006h
83
Operating the Z+F IMAGER® 5006h
5.3.7.2. M-Cam
84
Operating the Z+F IMAGER® 5006h
Tip
You can capture pictures with different exposure settings (e.g. dark,
normal, etc.) to make sure to have picked the right set of settings.
5.3.7.3. CameraMotor
Here the motor of the camera can be controlled and be oriented to its reference
position to check if the motor is working correctly.
85
Operating the Z+F IMAGER® 5006h
The result will be displayed in a graphic and stored on the internal flash disk.
The parameters from and to show the vertical angle of the clear view in the sky of
the Z+F IMAGER® 5006h. If the Z+F IMAGER® 5006h was set up with clear view
the default settings can be used. If there is only clear space (for example an open
window), the angles need to be changed.
By pushing the Scan button, the Z+F IMAGER® 5006h starts the test (without
horizontal moving) and checks the values of the default section. The scan stops after
a few seconds automatically. After the test is performed the operator needs to push
the button Refresh to show the result.
86
Operating the Z+F IMAGER® 5006h
87
Operating the Z+F IMAGER® 5006h
In such a case, clean the glass with the cleaning set or contact Z+F support (
support@zf-laser.com).
Tip
There is a special manual for profiler-mode. This manual can be
ordered separately from Z+F support ( support@zf-laser.com).
88
Operating the Z+F IMAGER® 5006h
89
Operating the Z+F IMAGER® 5006h
90
Transport and maintenance
6.1.3. Shipping
Please keep the shipping carton.
Only the complete original packing protects the device reliably against shock and
vibration during shipment.
Important
When transporting or shipping batteries, the person in charge of the
product must ensure that the applicable national and international
rules and regulations are observed. Before shipping or transportation,
contact your local passenger or freight transport company.
91
Transport and maintenance
6.2. Storage
6.2.1. Product
Respect the temperature limits when storing the equipment, particularly in Summer
if the equipment is inside a vehicle.
• If the 5006 Li-Ion Battery is stored, it should be recharged at least once in 6 months.
Important
The Z+F IMAGER® 5006h aperture window has to be perfectly clean
to avoid errors in measurement.
92
Transport and maintenance
• Use the lens tissue for wiping in a circular motion from the centre to
the edge until there is only a thin film of detergent visible.
• Use a new lens tissue for drying the pane, wipe in a circular motion.
• If any smears from cleaning are visible against back light, repeat
the procedure.
Caution
Do not use any alcohol, solvents, lens cleaning wipe with chemical
inside.
Tip
• Do not touch the side of the paper that is used for cleaning with
your fingers.
• Do not use air from the pneumatic power system as this is always
slightly oily!
93
Transport and maintenance
Tip
• Avoid large temperature differences. Slowly acclimate the
Z+F IMAGER® 5006h to the ambient temperature.
94
Troubleshooting
Chapter 7. Troubleshooting
7.1. Z+F IMAGER® 5006h
Z+F IMAGER® 5006h is too cold
• Start scanning, e.g. perform some preview scans, as the activity helps warm the
Z+F IMAGER® 5006h.
• Put the Z+F IMAGER® 5006h in a cool, shady place and leave the
Z+F IMAGER® 5006h powered only in the standby mode.
• Please turn off the Z+F IMAGER® 5006h and clean the control panel with mild
cleaning supplies and a soft handkerchief.
Reason:
Rechargeable 5006 Li-Ion Battery is depth-discharged but the 5006 Li-Ion Charger
can not initialize.
• Turn off power supply and turn it on again without changing any configuration.
or
95
Troubleshooting
Reason:
• When the temperatures is higher than 40°C or lower than 0°C, the charging process
does not start until the rechargeable 5006 Li-Ion Battery pack is within this range
of temperature.
Tip
The discharging temperature could be 60°C, but because of that the
rechargeable 5006 Li-Ion Battery would have to cool down to 40°C
before starting the charging process.
Reason:
The charging process is too slow; the 5006 Li-Ion Charger is not able to charge the
5006 Li-Ion Battery in the given time.
• Rechargeable 5006 Li-Ion Battery pack has to be checked please contact Z+F
support.
Please contact Z+F support if the error does not stop by using the methods mentioned
above.
Please contact Z+F support if you have to change the changeable 5006 Li-Ion Battery.
96
Safety instructions
• Opening the product using tools, for example screwdriver, unless this is specifically
permitted for certain functions.
• Use with accessories from other manufacturers without the prior explicit approval
of Z+F.
97
Safety instructions
Warning
Adverse use can lead to injury, malfunction and damage. It is the task
of the person responsible for the equipment to inform the user about
hazards and how to counteract them. The product is not to be operated
until the user has been instructed on how to work with it.
Warning
Local safety authorities and safety experts must be contacted before
working in hazardous areas, or in close proximity to electrical
installations or similar situations by the person in charge of the
product.
Warning
8.4. Responsibilities
8.4.1. Manufacturer of the product
Zoller + Fröhlich GmbH, 88239 Wangen im Allgäu, Germany, hereinafter referred
to as Z+F, is responsible for supplying the product, including the user manual and
original accessories, in a completely safe condition.
98
Safety instructions
• To understand the safety instructions on the product and the instructions in the user
manual.
• To inform Z+F immediately if the product and the application become unsafe.
• To ensure that the national laws, regulations and conditions for the operation of
radio transmitters are respected.
Warning
The person responsible for the product must ensure that it is used in
accordance with the instructions. This person is also accountable for
the training and the deployment of personnel who use the product and
for the safety of the equipment in use.
Warning
During the operation of the product, extremities or hairs might get
squeezed and/or clothes and entangled by rotating parts.
Precautions:
99
Safety instructions
Warning
The absence of instruction, or the inadequate imparting of instruction,
can lead to incorrect or adverse use, and can give rise to accidents
with far reaching human, material, financial and environmental
consequences.
• All users must follow the safety directions given by the manufacturer
and the directions of the person responsible for the product.
Important
Watch out for erroneous measurement results if the product has been
dropped or has been misused, modified, stored for long periods or
transported.
Caution
Because of the risk of electrocution, it is very dangerous to use poles
and extensions in the vicinity of electrical installations such as power
cables or electrical railways.
Caution
100
Safety instructions
Warning
During dynamic applications, for example stakeout procedures, there
is a danger of accidents occuring if the user does not pay attention
to the environmental conditions around, for example obstacles,
excavations or traffic.
• The person responsible for the product must make all users fully
aware of the existing dangers.
Warning
If computers intended for use indoors are used in the field there is a
danger of electric shock.
Warning
Inadequate securing of the working site can lead to dangerous
situations, for example in traffic, on building sites, and at industrial
installations.
Warning
If the accessories used with the product are not properly secured and
the product is subjected to mechanical shock, for example blows or
falling, the product may be damaged or people may sustain injury.
101
Safety instructions
Warning
During transport, shipment or disposal of batteries, the 5006 Li-Ion
Battery pack can be subject to unintended, mechanical stress which
might lead to fire hazards.
Caution
Warning
Using a battery charger not recommended by Z+F can destroy the
batteries. This can cause fire or explosions.
Warning
During usage, charging and/or disposal one of the following can occur
with impact to humans and environment:
102
Safety instructions
Warning
Fires, sparks, naked lights and smoking are prohibited: Avoid causing
sparks when dealing with cables and electrical equipment, and
beware of electrostatic discharges. Avoid short circuits.
Warning
High mechanical stress, high ambient temperatures or immersion into
fluids can cause leakage, fire or explosions of the batteries.
Warning
Only Z+F is allowed to do any repairs.
103
Safety instructions
The laser incorporated into the product produces a visible red laser beam which
emerges from the windows.
Warning
Class 3R Laser Products:
104
Safety instructions
Warning
Looking directly into the reflected laser beam could be dangerous to
the eyes when the laser beam is aimed at areas that reflect like a
mirror or emit reflections unexpectedly, for example prisms, mirrors,
metallic surfaces or windows.
Warning
The use of Laser Class 3R equipment can be dangerous.
Following an interpretation of the main points in the relevant section of the standard
quoted. Class 3R Laser Products used on construction sites and outdoors, for
example surveying, alignment, levelling:
a
The hazard distance is the distance from the laser at which beam irradiance or radiant exposure equals the
maximum permissible value to which personnel may be exposed without being exposed to a health risk.
105
Safety instructions
2.)Areas in which these lasers are used should be posted with an appropriate laser
warning sign.
3.)Precoutions should be taken to ensure that persons do not look directly, with or
without an optical instrument, into the beam.
4.)The laser beam should be terminated at the end of its useful beam path and should
in all cases be terminated if the hazardous beam path extends beyound the limit
(hazard distance) of the area in which the presence and activities of personnel are
monitored for reasons of protection from laser radiation.
5.)The laser beam path should be located well above or below eye level whereever
practicable.
6.)When not in use the Laser Product should be stored in a location where
unauthorized personnel cannot gain access.
7.)Precautions should be taken to ensure that the laser beam is not unintentionally
directed at mirror-like, specular surfaces for example mirrors, metal surfaces or
windows. But, more importantly, at flat or concave mirror-like surfaces.
106
Safety instructions
Warning
Electromagnetic radiation can cause disturbances in other equipment.
107
Safety instructions
Warning
There is a risk that disturbances may be caused in other equipment
if the product is used in conjunction with accessories from other
manufacturers, for example field computers, personal computers,
two-way radios, non-standard cables or external batteries.
Important
Disturbances caused by electromagnetic radiation can result in
erroneous measurements. Although the product meets the strict
regulations and standards which are in force in this respect, Z+F
cannot completely exclude the possibility that the product may be
disturbed by very intense electromagnetic radiation, for example, near
radio transmitters, two-way radios or diesel generators.
Warning
If the product is operated with connecting cables attached at only one
of their two ends, for example external supply cables, interface cables,
the permitted level of electromagnetic radiation may be exceeded and
the correct functioning of other products may be impaired.
108
Safety instructions
This equipment generates uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference will not
occur in a particular installation.
• Connect the equipment to an outlet on a circuit different from that to which the
receiver is connected.
Changes or modifications not expressly approved by Z+F for compliance could void
the users authority to operate the equipment
109
Safety instructions
110
1
Technical data
111
1
Technical data
112
1
Technical data
1
Resolution not recommended for exact measurements, only for positioning higher
resolution scan selections!
2
Only recommended for scan selections because of the enormous amount of data.
113
1
Technical data
Tilt sensor:
- Resolution: 1/1.000"
- Accuracy: 2/100° in a space of ±0.5°
- Range: ±1.5°
Data storage: Internal HDD 64 GB
Data interface:
- Interface: Ethernet / USB 2.0
Communication:
- Interface: Ethernet / WLAN
Integrated operation panel:
- Display: 4 lines
- Keypad: 6 buttons
Interfaces: 2 x USB, LEMO 9-pin and LEMO 7-pin
connections for
114
1
Technical data
115
1
Technical data
116
1
Technical data
117
1
Technical data
118
Labelling
119
Labelling
120
Labelling
121
Labelling
122
Contact
Email: <support@zf-laser.com>
Email: <service@zf-laser.com>
Web: www.zf-laser.com
UK:
Email: <info@zf-uk.com>
Web: www.zf-uk.com
123
Contact
Bridgeville,PA 15017
USA
Email: <info@zf-usa.com>
Web: www.zf-usa.com
ITALY:
Tel.: +39 02 90 39 41 16
Fax: +39 02 90 39 41 16
Email: <info@zf-italy.com>
Web: www.zf-italy.com
124
Attachments
Should the 5006 Li-Ion Battery case be destroyed by strong mechanical impact, a
contact with the ingredients is possible.
Inhalation:
Electrolyte:
Lead compounds:
• Consult a physician.
Ingestion:
Electrolyte:
• Consult a physician.
Lead compounds:
Skin contact:
Electrolyte
125
Attachments
126
Document History
127
Document History
128
Z+F Service Support
If you have any questions regarding the Z+F IMAGER® 5006h please contact the
Z+F service department.
If you contact Z+F you should have the Z+F IMAGER® 5006h and the accessories
available, please mention the following if you contact Z+F via telephone:
• Description of the problem and the last steps before the problem appeared.
Technical inquiries
Email: <support@zf-laser.com>
Email: <service@zf-laser.com>
Web: www.zf-laser.com
129
Z+F Service Support
130