Sie sind auf Seite 1von 2

CONTROL TEST CARD

Use in conjunction with:


DOSAGE REQUIREMENTS PAGE
ONBOARD LOG

BOILER WATER TREATMENT PROGRAM – DREWPLEX® AT with AMERZINE®


Read SDSs before running these tests. Before testing, samples must be cooled to 25°C (77°F) by collecting through a sample cooler for safety and to prevent flashing, which concentrates the samples. 
Note: If the sample is colored or turbid, filter before running the test. If the sample remains cloudy after the first filtering, the sample should again be filtered through the same paper since the filter becomes more retentive on the second filtration.
THE USE OF HIGH QUALITY, DISTILLED MAKEUP WATER IS RECOMMENDED.
BOILER WATER HYDRAZINE TEST PROCEDURE DAILY TEST PROCEDURES – BOILER WATER
1. For immediate testing, fill the Sample Cup to the 25 ml mark with cooled (via cooler not blower air) boiler water sample (Fig. 1). 1 2 3 4 5
2. Place the AMERZINE Ampoule’s tapered tip into one of the depressions in the bottom of the sample cup. Snap the tip by moving
the ampoule toward the cup side. NOTE: Keep the tip immersed in the water while drawing the sample. Within a few seconds the 1 Dry ‐
WAIT EXACTLY
ampoule will fill with sample and begin to mix with reagent (Fig. 2). 10 MINUTES
3. Remove the AMERZINE Ampoule from the Sample Cup. Mix contents of ampoule by inverting it several times, allowing the bubble
to travel from end to end each time (Fig. 3). MEANS OF ADJUSTMENT DOSAGE INFORMATION
4. Wipe all liquid from the exterior of the ampoule and WAIT EXACTLY 10 MINUTES for full color development (Fig. 4). • AMERZINE If < 0.03 ppm, increase dosage by 25%
5. Place the AMERZINE Ampoule , flat end downward into center tube of the round Comparator. Direct the round Comparator toward (PCN 0005406) 0.03 ‐ 0.10 ppm, no dosing adjustment necessary
a source of bright white (not fluorescent) light while viewing from the bottom. Hold the Comparator in a nearly horizontal position If > 0.10 ppm, decrease dosage by 25%
and rotate it until the color standard below the AMERZINE Ampoule shows the closest match (Fig. 5). Results are recorded as ppm of Initial dosage is 0.15 liters/ton of boiler water in the system. FEED CONTINUOUSLY 
the best matching color standard.
BOILER WATER HYDRATE ALKALINITY TEST PROCEDURE DAILY TEST PROCEDURES – BOILER WATER
1. Rinse and fill plastic titration vial to 12 ml line with the cooled boiler water sample (Fig. 1). 1 1 2 3 3 4 4
2. Add 2 ml Barium Chloride 10% into the vial and swirl to mix (Fig. 2). OR
3. Add 2 drops Phenolphthalein Indicator and swirl (Fig. 3). If COLORLESS, the Hydrate Alkalinity level is zero. If PINK then there is 12ml
12ml
6 #
12ml
12ml 12ml
some Hydrate Alkalinity – continue.
4. Add Sulfuric Acid N/10 drop by drop until the sample is COLORLESS, swirling the sample after each drop is added (disregard
eventual reappearance of a pink color) (Fig. 4). MEANS OF ADJUSTMENT DOSAGE INFORMATION
5. Calculate and record the Hydrate Alkalinity concentration: Number of drops Sulfuric Acid N/10 * 5 = ppm Hydrate Alkalinity as OHˉ • DREWPLEX AT If < 30 ppm, increase dosage by 25%
(PCN 6598405) 30 – 65 ppm, no dosage adjustment necessary
If > 65 ppm, decrease dosage by 25%
Initial dosage is 2.5 ‐ 5 liters/ton of boiler water in the system. FEED CONTINUOUSLY  
BOILER WATER PHOSPHATE TEST PROCEDURE DAILY TEST PROCEDURES – BOILER WATER
1. Fill (through filter) the Sample Cup to the 25 ml mark with sample (Fig. 1). Filter regardless of sample color. 1 2 3 4 5
2. Place the Boiler Phosphate Ampoule’s tapered tip into one of the depressions in the bottom of the sample cup. Snap the tip by
moving the ampoule toward the cup side. NOTE: Keep the tip immersed in the water while drawing the sample. Within a few seconds Dry ‐
WAIT 
the ampoule will fill with sample and begin to mix with reagent (Fig. 2). 5 MINUTES
3. Remove the Boiler Phosphate Ampoule from the Sample Cup. Mix the contents of the ampoule by inverting it several times,
allowing the bubble to travel from end to end each time (Fig. 3). REFERENCE TEST DOSAGE INFORMATION
4. Wipe all liquid from the exterior of the ampoule and WAIT 5 MINUTES for full color development (Fig. 4). Phosphate levels within the range of 10‐80 ppm are acceptable/expected. If phosphate level
5. The flat Comparator should be illuminated by a source of bright white (not fluorescent) light from above the Comparator. The filled is low, check for possible hardness inleakage and increase DREWPLEX AT dosage by 25% as
Boiler Phosphate Ampoule should be placed between the color standards for viewing. It is important that the ampoule be compared long as this can be accomplished while maintaining the Hydrate Alkalinity within range. For
by placing it on both sides of the standard tube before concluding that it is darker, lighter, or equal to the standard (Fig. 5). Results high Phosphate, as long as there is no foaming and Hydrate Alkalinity is within range, no
are recorded as ppm of the best matching color standard. action is necessary. If there is foaming and Hydrate Alkalinity is not at lower limit, decrease
DREWPLEX AT dosage by 25%.

BOILER WATER NEUTRALIZED CONDUCTIVITY TEST PROCEDURE DAILY TEST PROCEDURES – BOILER WATER
1. Connect the Conductivity Electrode ‐ ATC to the Conductivity Meter. Press the power button to power up the meter. 1 2 3
2. Add 2 drops of Phenolphthalein Indicator to 100 ml of cooled boiler water sample (Beaker 100 ml ‐ Clear Plastic) and stir (Fig. 1).
3. If the sample turns PINK, slowly add Gallic Acid Powder, stirring until the PINK color disappears (Fig. 2). This neutralizes the sample.
4. Place electrode into sample and take care to ensure that the liquid level is above its upper steel band. Stir the electrode gently in
Gallic 
Acid

the sample to create a homogenous sample (Fig. 3). 3
5. Allow time for the reading to stabilize. Note the reading on the display. MEANS OF ADJUSTMENT DOSAGE INFORMATION
• SURFACE DOSAGE INFORMATION
≤ 700 µS/cm ppm, satisfactory, no surface blowdown required 
Calibrate Conductivity Meter monthly with both the 84 µS/cm and the 1413 µS/cm standards. Also, calibrate the meter if, over the 
course of a week, system results are either less than 300 µS/cm or are within a narrow band (±25 µS/cm). BLOWDOWN > 700 µS/cm ppm, increase surface blowdown frequency/duration
Weekly Bottom Blowdowns of 10‐20 seconds are advisable to remove sludge and to clear operational lines. Blowdowns should be carried out per boiler manufacturer’s instructions. Surface blowdown is necessary to reduce
dissolved and suspended solids. If suspended solids are noticed, blowdown regardless of test results.
–1– BWT_DREWPLEX AT-AMERZINE-CONTROL TEST CARD
BOILER WATER CHLORIDE TEST PROCEDURE DAILY TEST PROCEDURES – BOILER WATER
1. For samples < 100 ppm chlorides: Rinse the Plastic Vial and fill to the mark (10 ml) with the sample to be tested (Fig. 1). 1 2 3 4
For samples > 100 ppm chloride: Rinse the tall Glass Vial (test tube) and fill to the mark (2 ml) with the sample to be tested (Fig. 1). OR &
2. Neutralize the sample by adding 3 drops of Phenolphthalein Indicator to the sample. Swirl to mix. If the sample turns PINK add Sulfuric OR
22
Acid N/10 drop by drop while swirling until the sample turns CLEAR. Afterwards, add one more drop (Fig. 2). 10 ml ml # #

3. Adjust the color of the sample by adding 6 drops of Potassium Chromate. The sample will be YELLOW (Fig. 3).
4. Counting the drops, add Silver Nitrate N/10 drop by drop. Swirl between drops until the sample turns ORANGE (Fig. 4). MEANS OF ADJUSTMENT DOSAGE INFORMATION
5. Calculate the chloride concentration: • SURFACE If ≤ 300 ppm, satisfactory, no surface blowdown required 
If plastic vial (< 100 ppm) was used in Step 4: ‐ Number of drops of Silver Nitrate N/10 * 10 = ppm chloride BLOWDOWN If > 300 ppm, increase surface blowdown frequency/duration 
If glass test tube (> 100 ppm) was used in Step 4: ‐ Number of drops of Silver Nitrate N/10 * 50 = ppm chloride
DAILY TEST PROCEDURES – BOILER WATER
CONDENSATE WATER pH TEST PROCEDURE WEEKLY TEST PROCEDURES – CONDENSATE
1. Obtain a 10 ml sample and test immediately (Fig. 1). 1 2SYSTEM 3
2. Add 3 drops Phenolphthalein Indicator (Fig. 2).
3. If PINK color develops, add Sulfuric Acid N/10 drop by drop until sample turns, and remains, COLORLESS (Fig. 3). #
10ml
4. Record the result, which is the number of drops of Sulfuric Acid N/10. 10ml 10 ml

REFERENCE TEST DOSAGE INFORMATION


1‐2 drops indicates acceptable condensate pH. If no PINK color develops, increase
DREWPLEX AT dosage by 25% if this can be accomplished while maintaining the range of
both the Hydrate Alkalinity and the Phosphate. If 3 or more drops are required, the
condensate pH is high and DREWPLEX AT dosage should be decreased by 25% as long as
this can be accomplished while maintaining the Hydrate Alkalinity and Phosphate ranges.

MAKEUP WATER CHLORIDE TEST PROCEDURE WEEKLY TEST PROCEDURES – MAKEUP WATER
1. Rinse the Plastic Vial and fill to the mark (10 ml) with the sample to be tested (Fig. 1). 1 2 33 4
2. Neutralize the sample by adding 3 drops of Phenolphthalein Indicator to the sample. Swirl to mix. If the sample turns PINK add Sulfuric OR &
Acid N/10 drop by drop while swirling until the sample turns CLEAR. Afterwards, add one more drop (Fig. 2).
10 ml #
3. Adjust the color of the sample by adding 6 drops of Potassium Chromate. The sample will be YELLOW (Fig. 3). #

4. Counting the drops, add Silver Nitrate N/10 drop by drop. Swirl between drops until the sample turns ORANGE (Fig. 4).
5. Calculate the chloride concentration: Number of drops of Silver Nitrate N/10 * 10 = ppm chloride TEST RESULTS
Makeup tanks should be tested every week or more often if3 there is a change in the
source or mixture of the water or any upset is suspected. If at all possible, do not
DOSAGE INFORMATION
introduce makeup water with Chlorides above 10 ppm.

MAKEUP WATER HARDNESS TEST PROCEDURE WEEKLY TEST PROCEDURES – MAKEUP WATER
1. Rinse and completely fill the Sample Cup with the makeup water sample (Fig. 1). 11 2 3 4 5
> 0.5 PPM
2. Place the tip of the Hardness Test Ampoule into the hole in the bottom of the Sample Cup. FAIL
3. While applying downward pressure, break the tip of the ampoule by tilting the ampoule toward the far edge of the snapping cup. Dry ‐ PASS
WAIT 
NOTE: Keep the tip immersed in the water while drawing the sample. Within a few seconds, the ampoule will fill with sample and 30 SECONDS
< 0.5 
PPM
begin to mix with reagent (Fig. 2).
4. Mix by inverting the Hardness Test Ampoule back and forth to dissolve the reagent (Fig. 3). TEST RESULTS
5. Wipe all liquid from the exterior of the ampoule and WAIT 30 SECONDS for full color development (Fig. 4). Makeup tanks should be tested every week or more often if3 there is a change in the
6. Place the ampoule in front of a white background and view the color. A pure BLUE indicates total hardness (calcium carbonate source or mixture of the water or any upset is suspected. If at all possible, do not
CaCO3 ) of less than 0.5 ppm. A PURPLE/PINK color indicates that hardness is present ‐ above the 0.5 ppm level (Fig. 5). introduce makeup water with hardness above 0.5 ppm.
DOSAGE INFORMATION
LIQUID COAGULANT™ - BOILER SLUDGE CONDITIONER (PCN 0008400) NO TEST REQUIRED (For Oil Contamination Only)
In case of oil contamination, dose 28 ml (1.0 oz.) per day for each ton of water capacity to condition suspended solids (oil and sludge) for removal by bottom blowdown. Flash bottom blow and surface blow several times at 20 minute
intervals after dosing LIQUID COAGULANT. Dose for two weeks after no more oil is visible. If contamination persists, the possibility of continued oil leakage should be investigated and eliminated. (PCN 0008400)

APPARATUS AND REAGENTS


Storage: Water Treatment Test Cabinet ‐ 220volt (PCN 1AA0016) or ‐ 110volt (PCN 1AA2148) Conductivity: Conductivity Standard 1413 µS/cm (PCN 6968020)
Hydrazine: AMERZINE Ampoule Test Kit (PCN 0369018), AMERZINE Ampoule Refill (PCN 0369026) Gallic Acid Powder ‐ 30g Btl (PCN 0309170), Measuring Spoon ‐ Brass 0.2g (PCN 0224014)
Alkalinity: Hydrate Alkalinity Test Kit (PCN 0388018) Phenolphthalein Indicator, 120 ml (PCN 0311191), Beaker 100ml ‐ Clear Plastic (PCN 0247016)
Phosphate: Boiler Phosphate Ampoule Test Kit (PCN 1AA0003) Chloride: Chloride LMP Test Kit (PCN 0373019)
Boiler Phosphate Ampoule Refill (PCN 1AA0004) Condensate pH: Phenolphthalein Indicator (PCN 0311191), Sulfuric Acid N/10 (PCN 0319195)
Conductivity: Conductivity Meter Kit Complete (PCN 0173062) Hardness: Hardness Test Ampoules (PCN 0365016)
Conductivity Electrode ‐ ATC (PCN 0173070) Other: Filter Paper ‐ 100/BX  (PCN 0225012), Funnel ‐ Plastic (PCN 0221010)
Conductivity Standard 84 µS/cm (PCN 6968038) Liquid Coagulant: Liquid Coagulant (PCN 0008400)

©2016 Drew Marine ®Registered trademark, ™Trademark, Drew Marine ● All Rights Reserved –2– BWT_DREWPLEX AT-AMERZINE-CONTROL TEST CARD

Das könnte Ihnen auch gefallen