Always perform measurements in the correct sequence! All device measurements must be performed before undertaking any measurements
involving direct contact with the eye.
Calibration test
Always check the calibration daily before carrying out measurements on patients!
Tap the Start button to start the first of three calibration test steps. The test tool holder in the chin rest
compartment will rise up automatically.
Now carry out the measurements on the test eye. Follow the instructions on the screen. The overview shows the
individual steps of the calibration test and their status
If the Calibration test was passed, press the button ”next” and the application will be started.
If the Calibration test was failed, repeat the calibration test by pressing the button “repeat”.
New Patient
New patient data can be entered by tapping the Add button. The following information is obligatory: Last name,
first name, Gender (Gender drop-down box), Date of birth (Open the Date selection box and select the
patient's day, month and year of birth. Confirm with OK.), and Refractive Surgery.
It is important that the correct lens state and vitreous body state are selected as this cannot be changed after the
measurement has been done. If changes are necessary, the measurement needs to be repeated.
An existing patient dataset from the patient list on the left side of the screen can also be selected.
Input field for entering a search string Parts of the patient name, ID or date of birth may be entered for the
search
With the buttons “Analysis”, “IOL Calculation” and “Measurement” the according dialogs screens are opened.
Measurement
Please note: This document is a preliminary version and does not replace the user manual
which is delivered together with the device. Final version is to be released soon.
flij~ëíÉê=TMM=Ó=nìáÅâ=dìáÇÉ=
Quality-Check
Successful measurement: A green quality indicator shows a technically perfect measurement
Warning for uncertain measurement: A yellow quality indicator will signal to the user that he/she
should verify the corresponding measurement marks thoroughly
Measurement failed: A red quality indicator means that no measurement results could be detected
The quality check helps to evaluate the quality of the measurement. From the application point of view, the user
should check whether the measured values are correct and plausible. This can be done by using the analysis
window.
In addition the Fixation of the patient should be checked. Good fixation is shown by a depression in the center
(right picture), the so-called foveal pit. If the foveal pit can be easily recognized, the patient fixated well during
measurement. If it cannot be recognized (left picture), the patient has not correctly fixated. In this case, measurement
should be repeated and the patient be instructed to fixate on the fixation light in the device.
If the patient cannot fixate during several measurements the user should verify if there may be another problem with
the patient which requires special examination.
Analysis
The following images are displayed in the analysis dialog separately for the right and left eye:
- Biometry-Scan
- Keratometry-Scan
- WTW-Image
- Sclera- Image
- Fixation Check.
They should be checked for artefacts in B-Scans, WTW and sclera images which are out of focus or deformed or
missing spots in keratometry. The measurement values for the right eye are shown on the left side (OD) and the
measurement values for the left eye on the right side (OS) of the screen. Measurement values should be checked
thoroughly for plausibility. Warning messages should be read with caution.
IOL Calculation
For each eye, the IOL calculation is done separately. For better orientation, the measurement data of the selected
eye are displayed on a white background, the data of the other eye are greyed out. Additionally, the selected eye
is shown in bold and green letters in the upper right corner of the display.
Tap the measurement data area on the right or left side of the screen to alternate between the eye or use the
button . Beneath the measurement data, the lens state and refractive surgery history is shown for each
eye.
Tapping select the calculation formulae to be applied.
IOL calculation is started by tapping . To be able to print or export to FORUM, always an IOL calculation
has to be done
By tapping Finish/ Print / Export (view can vary according to individual setting), reports will be printed / exported
and the application will jump to the Patient Manager where other patients can be opened and / or additional
measurements can be done.
The measurement values can be edited by tapping the appropriate field. Edited values are marked with an *. If the
IOL calculation window is closed, the entries will be canceled. When re-starting the IOL calculation, the measured
values are displayed in these fields.
Shutdown
Tap the X (Close) button to open a menu with options for logging out the current user or switching the device off after finishing all
measurements.
Please share your experience with the IOLMaster 700 with us!
www.zeiss.com/cataract-community