HP 6890 Series
Gas Chromatograph
DHewlett-Packard Safety Information Safety Symbols Important User
Company 1989—1994 The HP 6890 Gas Warnings in the manual or Information for In Vitro
Chromatograph meets the on the instrument must be Diagnostic Applications
All Rights Reserved. following IEC observed during all phases This is a multipurpose
Reproduction, adaptation, (International of operation, service, and product that may be used
or translation without Electrotechnical repair of this instrument. for qualitative or
permission is prohibited, Commission) classifications: Failure to comply with quantitative analyses in
except as allowed under Safety Class 1, Transient these precautions violates many applications. If used
the copyright laws. Overvoltage Category II, safety standards of design in conjunction with proven
and Pollution Degree 2. and the intended use of the procedures (methodology)
HP part number instrument. by qualified operator, one
This unit has been
G1530-90390 Hewlett-Packard Company of these applications may
designed and tested in
accordance with recognized assumes no liability for the be In Vitro Diagnostic
First Edition, Nov 1994 customer’s failure to Procedures.
safety standards and
designed for use indoors. If comply with these Generalized instrument
Printed in USA requirements.
the instrument is used in a performance characteristics
Warranty manner not specified by the
manufacturer, the
7!2.).' and instructions are
included in this manual.
The information contained A warning calls attention
in this document is subject protection provided by the Specific In Vitro Diagnostic
instrument may be to a condition or possible procedures and
to change without notice. situation that could cause
impaired. Whenever the methodology remain the
Hewlett-Packard makes no safety protection of the HP injury to the user. choice and the
warranty of any kind with
regard to this material,
6890 has been
compromised, disconnect
#!54)/. responsibility of the user,
and are not included.
including, but not limited the unit from all power A caution calls attention to
to, the implied warranties a condition or possible Sound Emission
sources and secure the unit
of merchantability and situation that could Certification for Federal
against unintended
fitness for a particular operation. damage or destroy the Republic of Germany
purpose. Hewlett-Packard Refer servicing to qualified product or the user’s work. Sound pressure Lp
shall not be liable for errors < 65 dB(A)
service personnel. Caution. Refer to
contained herein or for Substituting parts or During normal operation
incidental or consequential accompanying
performing any documents. At the operator position
damages in connection with unauthorized modification
the furnishing, According to ISO 7779
to the instrument may Indicates a hot (Type Test)
performance, or use of this result in a safety hazard.
material. surface. When operating the HP
Disconnect the AC power
Indicates hazard- 6890 with cryo valve
cord before removing ous option, the sound pressure
covers. The customer voltages. 74.6 dB(A) during cryo
should not attempt to Indicates earth valve operation for short
replace the battery or fuses (ground) termi-
nal. burst pulses.
in this instrument. The
battery contained in this Indicates ra- Schallemission
instrument is recyclable. dio-
activity haz- Schalldruckpegel LP
ard. < 65 dB(A)
Indicates explo-
sion Am Arbeitsplatz
hazard. Normaler Betrieb
Nach DIN 45635 T. 19
(Typprüfung)
Bei Betrieb des HP 6890
mit Cryo Ventil Option
treten beim Oeffnen des
Ventils impulsfoermig
Schalldrucke Lp bis ca.
74.6 dB(A) auf.
3ITE 0REPARATION
Benchtop requirements
The GC is available with or without electronic pneumatics control (EPC).
The EPC model is 59 cm (23 in.) wide. The nonEPC model is 68 cm (26.7
in.) wide. Both EPC and nonEPC GCs have a height of 50 cm (21 in.)
and a length of 50 cm (21 in.).
The area above the GC should be clear. Shelves or overhanging
obstructions limit access to the top of the instrument and interfere with
cooling. You must also be sure the area can accommodate other
instruments used with your GC. For example, if you are using the GC
with a GC Automatic Liquid Sampler, remember that the sampler adds
an additional 44 cm (17 in.) to the system’s height and 20.9 cm (8.3 in.) to
its width. The following figure illustrates some common GC system
configurations.
Table 1 presents the dimensions, voltage requirements, heat production,
and weight of the GC and other Hewlett•Packardinstruments that are
commonly used with it. Refer to this table before installation to insure
that you have adequate space and power for the entire system. When
laying out the system, make sure that there are at least 10.2 cm (4 in.)
between instruments to allow for proper ventilation.
3ITE 0REPARATION
HP 35900C/D/E Analog-to- 11 x 33 x 29 cm 40 216 KJoules 4.1 kg
Digital 4x 13 x 11 in. 205 Btu/hr 9.0 lb
Converter
* General specifications for a mid-size, desktop
computer.
3ITE 0REPARATION
59* cm
ALS Controller
Injector—44 cm
above GC
50 cm
31 cm
GC
Tray—30.3 cm
left of GC 10 cm
GC with HP ChemStation
59* cm
LaserJet IV
HP Vectra
50 cm GC
53.3 cm 40.3 cm
Keyboard
46.4 cm 28 cm 41.5 cm
Mous
e pad 23 cm
59* cm
MSD
65 cm
GC 50 cm
17 cm
3ITE 0REPARATION
Electrical requirements
Grounding the GC
Caution A proper earth ground is required for GC operations.
To protect users, the metal instrument panels and cabinet are grounded
through the three•conductorpower line cord in accordance with
International Electrotechnical Commission (IEC) requirements.
The three•conductorpower line cord, when plugged into a properly
grounded receptacle, also grounds the instrument and minimizes shock
hazard. A properly grounded receptacle is one that is connected to a
suitable earth ground. Proper receptacle grounding should be verified.
WARNING Any interruption of the grounding conductor or disconnection of the
power cord could cause a shock that could result in personal injury.
Line voltage
The GC operates from one of the AC voltage supplies listed in Table 2,
depending on the standard voltage of the country from which it was
ordered. GCs are designed to work with a specific voltage; make sure
your GC's voltage option is appropriate for your lab. The voltage
requirements for your GC are printed near the power cord attachment.
Make sure the GC is connected to a dedicated receptacle.
3ITE 0REPARATION
Maximum
Power Con- Power Line
Voltage Frequency sumption (VA) Requirement Oven
(Hz) Type
120 V (+5,-10%) 48-66 2,250 20-amp dedi- Regular
cated recep-
tacle
200 V (+5, 48-66 2,950 15-amp dedi- Fast-heat-
-10%) cated recep- ing
tacle
220 V (+5, 48-66 2,950 15-amp dedi- Fast-heat-
-10%) cated recep- ing
tacle
230 V (+5, 48-66 2,950 16-amp dedi- Fast-heat-
-10%) cated recep- ing
tacle
230 V (+5, 48-66 2,250 10-amp dedi- Regular
-10%) cated
(Switzerland or Denmark with receptacle
10-amp maximum service)
240 V (+5, 48-66 2,950 13- or 16-amp Fast-heat-
-10%) dedicated re- ing
ceptacle
3ITE 0REPARATION
On/off switch
=
Off
=
On
3ITE 0REPARATION
&USES AND BATTERY
-AINTENANCE
Maintenance schedule
The frequency with which you will need to perform maintenance will
vary depending upon the following:
1Especially important for temperature programmed analyses that use Vespel or
Vespel/graphite
ferrules
2 With typical usage, A-size cylinders will supply two dual-channel chromatographs
for about
three months. Replace the cylinder when its pressure drops below 500 psig.
-AINTENANCE
WARNING If the power cord insulation is frayed or worn, it must be replaced. Con•
tact your Hewlett•Packardservice representative.
-AINTENANCE
/PERATION
Safety shutdowns
Hydrogen shutdown
Hydrogen gas may be used as a carrier or as fuel for some detectors.
When mixed with air, hydrogen can form explosive mixtures.
The GC monitors inlet and auxiliary gas streams. If a stream shuts
down because it is unable to reach its flow or pressure setpoint and if
that stream is configured to use hydrogen, the GC assumes that a leak
has occurred and declares a hydrogen safety shutdown. The effects are:
C The carrier supply valve to the inlet closes.
C The split valve in the split/splitless inlet opens.
C The oven heater turns off. The flaps at the rear open fully.
To recover from this state, fix the cause of the shutdown (tank valve
closed, serious leak, etc). Turn the instrument off, then back on.
The GC cannot detect leaks in the detector gas streams. For this reason,
it is vital that the column fittings of the FID, NPD, and any other
detectors that use hydrogen always be connected to a column or have a
cap or plug installed and that hydrogen streams be configured so that the
GC is aware of them.
Column shutdown
If the carrier gas source shuts down, the oven heater turns off to avoid
column damage from heat without carrier gas. The flaps at the rear open
halfway.
To recover from this state, fix the cause of the shutdown (tank valve
closed, serious leak, etc). Turn the oven and the offending inlet or auxil•
iary channel back on.
/PERATION
General warnings
Be careful when lifting the GC. Because it is heavy, two people should
lift it. When moving the GC, note that the back is heavier than the front.
Substituting parts or performing any unauthorized modification to the
instrument may result in a safety hazard.
Thermal
Be careful when working on the instrument. The oven, inlet, and detec•
tor fittings may be hot enough to cause burns.
Be careful when working behind the instrument. During cooldown cycles,
the GC will emit hot exhaust, which could cause burns.
Do not place temperature•sensitiveitems (for example, gas cylinders,
plumbing chemicals, or regulators) in the path of the heated exhaust.
Glass and fused silica columns
Wear safety glasses to protect your eyes from flying particles while han•
dling, cutting, or installing glass or fused silica capillary columns. Use
care in handling these columns to prevent puncture wounds.
With glass columns, overtightening the column nut or forcing it to bottom
in the inlet or detector fitting may shatter the column.
Insulation
The insulation around the inlets, detectors, valve box, and the insulation
cups is made of refractory ceramic fibers (RCF). To avoid inhaling RCF
particles, we recommend the following safety procedures:
C Ventilate your work area.
C Wear long sleeves, gloves, safety glasses, and a disposable dust/mist
respirator.
C Dispose of insulation in a sealed plastic bag.
C Wash your hands with mild soap and cold water after handling RCFs.
/PERATION
/PERATION
Oven safety
For safety, opening the oven door turns off power to the oven heater, fan,
and cryogenic valve (if installed) but maintains the setpoints in memory.
OVEN
Temp 350 door open
Init time 2.00
Rate 1(off) 0.00
SHUTDOWN (#1):
/PERATION
Hydrogen warnings
General warning
When using hydrogen (H2) as the carrier gas or fuel gas, be aware that
hydrogen (H2) gas can flow into the oven and create an explosion hazard.
Therefore, be sure that the supply is off until all connections are made,
and ensure that the inlet and detector column fittings are either con•
nected to a column or capped at all times when hydrogen (H2) gas is sup•
plied to the instrument.
Hydrogen (H2) is flammable. Leaks, when confined in an enclosed space,
may create a fire or explosion hazard. In any application using hydrogen
(H2), leak test all connections, lines, and valves before operating the
instrument. Always turn off the hydrogen (H2) supply at its source be•
fore working on the instrument.
Column fittings
Make sure a column is installed or the column and inlet fittings are
plugged before turning on the hydrogen flow•either carrier gas or
detector fuel gas. An explosion may occur if hydrogen leaks into the
oven.
Measuring hydrogen
Do not measure hydrogen together with air or oxygen. This can create
explosive mixtures that may be ignited by the automatic ignitor.
To avoid this hazard:
C Turn the automatic ignitor off before you begin.
C Always measure gases separately.
/PERATION
/PERATION
Avoid the following materials because they may react with the 63Ni
source and form volatile products or cause physical degradation of the
plated film: oxidizing compounds, acids, wet halogens, wet nitric acid,
ammonium hydroxide, hydrogen sulfide, or carbon monoxide.
In the extremely unlikely event that both the oven and the ECD•heated
zone should go into thermal runaway (maximum, uncontrolled heating in
excess of 400^C) at the same time, and that the ECD remains exposed to
this condition for more than 12 hours, take the following steps:
1.. Turn off the main power and allow the instrument to cool.
2.. Cap the ECD inlet and exhaust vent openings. Wear disposable
plastic gloves and observe normal laboratory safety precautions.
3.. Return the cell for exchange, following directions included with the
License Verification Form (HP part no. 19233•90750).
USA requirements
The detector is sold under general license. General licensees are per•
mitted to use thermal cleaning only. Do not use organic or water•based
solutions to clean the Electron Capture Detector. Do not open the
detector cell unless authorized to do so by your local nuclear regulatory
agency. Do not disturb the four socket•headbolts because they hold the
cell halves together and removal could create a safety hazard. Removing
or disturbing them is also a violation of the terms of the General License.
Additional information is available in the publication Information for
General Licensees," HP part no. 5953•1798.
Owners of this detector must perform a wipe test at least every 6 months.
Complete instructions for performing this test are in the HP 6890
Maintenance and Troubleshooting Manual.
/PERATION
ECD safety
ECD owners in the United States are required to perform a wipe test on
their detector every 6 months. Complete instructions are in the Wipe
Test Kit, HP part no. 18713•60050.
ECD maintenance
You should perform a thermal cleaning or bakeout on your detector if the
cell becomes dirty. See the Maintenance and Troubleshooting Manual for
instructions.
Leaks can adversely affect the performance of your detector. It is
therefore important to locate and fix any gas leaks. Complete
instructions for performing a gas leak test are in the HP 6890
Maintenance and Troubleshooting manual.
/PERATION
Cautions
Do not turn off the electrometer during a run. It will turn off the detector
Output.
Detector electronics depend on the correct gas configuration.
Nitrogen-phosphorus detector
If you turn Adjust offset [Off] at any time, the bead voltage, hydrogen,
and air flows all turn off.