PD501-002B

X-ray
High Voltage
Generator
UD150B-40/V-40/L-40

GENERAL
The Shimadzu’s UD150B-40/L-40/V-40 is a The UD150B-40, L-40 and V-40 are suitable for the
high-performance X-ray high voltage generator. It has following applications;
greatly improved operability and safety with the use of
computers and provides a low-ripple output with high (A) General radiography
X-ray quantum efficiency through a high-frequency
(B) Bucky radiography
inverter system.
(C) Auto changer radiography
(D) Tomography
(E) Fluoroscopy

1
UD150B-40/V-40/L-40
FEATURES
(1) Inverter system (7) Lights up hand switch
The high frequency inverter system adopted for high The hand switch lights up to indicate Ready Up and
voltage generation, generates low ripple tube X-ray exposure status.
voltage with high X-ray quantum efficiency.
Moreover, by cutting off X-rays regardless of power (8) Illumination panel
phase, it is possible to perform high-accuracy photo-
The operation panel indicates the status of the
timer control in the ultra-short time.
device (e.g., ready to take images or image capture
in progress) using color and sound.
(2) Microcomputer
To enhance the ergonomic design of the UD150B (9) 2-factors control system
-40/L40/V-40, a microcomputer is utilized for the
In addition to the ordinary setting system of KV, mA,
improvement of operability and safety.
and sec., technique conditions can be also set with
2-factors of kV and mAs. In this case, mA and sec.
(3) Detailed setting of exposure factors are automatically set to the maximum allowable
Exposure time and mAs can be set more in detail in value and the shortest value.
every 12.5% than the conventional.
(10) Compact size
(4) Preset memory Compact 2-unit construction. The unique space-
The preset memory allows the most commonly used saving design of the 2 unit consists of a compact
techniques to be programmed. By pressing the control panel that can be desk or wall mounted and
program selector key, automatic selection can be one control cabinet, requiring a very small
made for radiographic factor, tube focus, etc. installation space.

(5) Color LCD Touch Panel (11) Two standard fluoroscopic modes
A color LCD panel with high visual clarity is used as Two standard fluoroscopic modes are provided.
the screen for setting X-ray conditions. The panel, Each mode is customised to meet the requirements
which uses different colors for extra clarity, allows of clinical application.
simple APR selection.
(12) Automatic fluoroscopic diagnosis
(6) Hybrid Dials Tube voltage is automatically controlled even if a
X-ray conditions can be changed easily using hybrid radiograpic area or subject thickness is changed
dials. Large changes can be made using jog/shuttle during fluoroscopy. The image brightness stabilizer
and small changes can be made precisely using the (IBS) always ensures a fluoroscopic image with
up/down buttons. Using both types of adjustment optimum brightness.
method allows conditions to be set quickly.

CONFIGURATION
STANDARD CONSTITUTION OPTIONAL ITEMS
(a) Control console ·················· 1 unit ∗ SPT-C42, photo-timer control
(b) Control cabinet ··················· 1 unit ∗ Direct photo-timer pickup
∗ Starter
(c) High voltage generator ······· Incorporated in the ∗ Third tube selector
control cabinet ∗ Communication unit
(d) Connection cable················ 1 set ∗ Voltage regulator
Cable length; ∗ Fluoroscopy unit
∗ Power source-Control cabinet: 10 m
(USA and CANADA: 16m)
∗ Control console-Control cabinet: 14 m

2
UD150B-40/V-40/L-40
SPECIFICATIONS
(1) Standard model and nominal maximum power: (7) Setting and display of radiographic tube current (for
3-control system)
Model UD150B-40 UD150V-40 UD150L-40
y Setting range:
Nominal maximum Power 80 kW 65 kW 50 kW
UD150B-40 UD150V-40 UD150L-40

1000, 800, 630, 500, 800, 630, 500, 400, 630, 500, 400, 320,
(2) Ratings: 400, 320, 250, 200, 320, 250, 200, 160, 250,200, 160, 125,
160, 125, 100, 80, 63, 125, 100, 80, 63,50, 100, 80, 63, 50, 40,
Model UD150B-40 UD150V-40 UD150L-40 50, 40, 32, 25, 20, 16, 40, 32, 25, 20, 16, 12 32, 25, 20, 16, 12
12 and 10 mA and 10 mA and 10 mA
380-480 V, 3 phase
380-480 V, 3 phase 380-480 V, 3 phase
Power Source 200-240 V, 1 phase Of the 21 positions Of the 20 positions Of the 19 positions
50/60 Hz 50/60 Hz above, it is possible to above, it is possible to above, it is possible to
50/60 Hz
use any 6 positions use any 6 positions use any 6 positions
150 kV 500 mA 150 kV 400 mA 150 kV 320 mA
allowed by the X-ray allowed by the X-ray allowed by the X-ray
Short-time rating 125 kV 630 mA 125 kV 500 mA 125 kV 400 mA
100 kV 800 mA 100 kV 650 mA 100 kV 500 mA tube for one focus. tube for one focus. tube for one focus.
*1
80 kV 1000 mA 80 kV 800 mA 80 kV 630 mA
y Display: The set value is displayed digitally.
Long-time rating 75 kV 20 mA
125kV 9 mA 125 kV 9 mA y Setting method: shuttle and UP/DOWN switch
*1 125 kV 12 mA
*1. Limited according to the X-ray tube type. (8) Setting and display of exposure time (in the case of
3-control system)
(3) Tube voltage and X-ray exposure time control
system y Setting range:
The DC voltage rectified and smoothed from
UD150B-40 UD150V-40 UD150L-40
3-phase or single-phase power source is applied to
the inverter circuit. Tube voltage is controlled by the 1.0, 1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2, 2.5, 2.8, 3.2, 3.6, 4.0, 4.5, 5.0, 5.6,
feedback system, and at the same time, applied to 6.3, 7.1, 8.0, 9.0, 10, 11, 12, 14, 16, 18, 20, 22, 25, 28, 32, 36, 40, 45, 50,
the X-ray tube, making and breaking it. 56, 63, 71, 80 ms, 90 ms,

The exposure is made and broken regardless of the 0.10, 0.11, 0.12, 0.14, 0.16, 0.18, 0.20, 0.22, 0.25, 0.28, 0.32, 0.40, 0.45,
0.50, 0.56, 0.63, 0.71, 0.80, 0.90, 1.0, 1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2, 2.5,
supply voltage phase.
2.8, 3.2, 3.6, 4.0, 4.5, 5.0, 6.3, 7.1, 8.0, 9.0, 10 sec

(4) Technique factor setting system It is impossible to set the mAs value to less 0.5 or over 800.

y 2-control system (kV, mAs) y Display: The set value is displayed digitally.
A system that the shortest exposure time and y Setting method: shuttle and UP/DOWN switch.
maximum tube current are automatically set for sec
and mA according to the allowable load of the X-ray (9) Setting and display of fluoroscopic tube voltage
tube when kV and mAs are set as the technique y Setting range: 50~125kV in increment of 1kV.
factors. y Display: The set value is displayed digitally.
y 3-control system (kV, mA, sec.)
A system to set kV, mA, and sec as technique (10) Setting and display of fluoroscopic tube current
factors.
y Setting range: Fluoroscopic tube current is auto-
matically set according to fluoroscopic tube voltage.
(5) Setting and display of radiographic tube voltage
Presetting is already made so that the fluoroscopic
y Setting range: 40 to 150 kV in increment of 1 kV. tube current value rises as the fluoroscopic tube
y Display: The set value is displayed digitally. voltage rises.
y Setting method: shuttle and UP/DOWN switch. y Display: Digital display of actual value during
radiation of X-ray
(6) Setting and display of radiographic tube current time
product (in the case of 2-control system) (11) Resetting and display of accumulated fluoroscopic
y Setting range: time
UD150B-40 UD150V-40 UD150L-40 y Accumulated fluoroscopic time reset:
Press the “ ” key. The time is reset to “0.0 min”.
0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2, 2.5, 2.8, 3.2,
y Accumulated fluoroscopic time display:
3.6, 4.0, 4.5, 5.0, 5.6, 6.3, 7.1, 8.0, 9.0, 10, 11, 12, 14, 16, 18, 20, 22, 25,
Display range … 0.0~99 min
28, 32, 36, 40, 45, 50, 56, 63, 71, 80, 90, 100, 110, 125, 140, 160, 180,
200, 220, 250, 280, 320, 360, 400, 450, 500, 560, 630, 710, 800 mAs
(12) Setting of fluoroscopic timer
Setting is available in 64 positions shown above.
The value is limited to 500 mAs when using photo-timer.
Press the “ ” key. “10 min” is set.
The “ “ key flickers with buzzer sounding during
y Display: The set value is displayed digitally. the lapse time from 4 minutes and 30 seconds to 10
y Setting method: shuttle and UP/DOWN switch minutes. Fluoroscopy is shut out after lapse of 10
minutes.

3
UD150B-40/V-40/L-40
SPECIFICATIONS (cont.)
(13) Selectable techniques (20) Direct photo-timer exposure (option)
A photo-timer exposure is available in combination
(A) General radiography
with Shimadzu’s photo-pickup detector by adding
(B) Bucky radiography the pertinent control circuit.
(C) Auto changer radiography y Manual operation or photo-timer can be selected.
(D) Tomography y SPEED (sensitivity) selector
“SPEED. H”, “SPEED. M” and “SPEED. L” are
(E) Fluoroscopy
selectable at installation of the system.
∗1 The technique selector doubles as the X-ray y Film density adjustment
tube selector. Adjustment in 11 stages is possible.
y Photo-pickup field selection
(14) Selection of X-ray tube: Four kinds of photo-pickup fields are selectable.
Use the radiography technique key to select the
X-ray tube. (21) Display of error
Up to two tubes can be connected. In the occurrence of error on each function, error-
codes are displayed on the color LCD touch panel.
(15) Selection of focus:
The “large” or “small” focus can be selected.

(16) Other selector keys:


y The exposure modes “MANUAL” or “PHOTO-
TIMER” are selectable as the “AEC”.
y The technique factor setting systems “2-CONTROL”
or “3-CONTROL” are selectable as the “mA・sec”
key.
y The “Radiography program screen” or “Radiography
step screen” can be registered as the “HOME” key.
y The “System set-up screen” is displayed to the color
LCD as the “SET UP” key.
y The “Radiography history screen” is displayed to the
color LCD as the “LOG” key.

(17) Switches
POWER ON/OFF switch and READY/X-RAY hand
switch.

(18) Indicator
“Radiography Ready”, “Exposure”, “Stand-by”,
“Caution”, “Fault” are LED-indicated.

(19) Preset program:


The 7 tab keys for examination area selection
(ANATOMICAL TAB KEY) and 15 keys for technique
selection (PROTOCOL KEY) enable 105 programs
of exposure parameter in total. Furthermore each
PROTOCOL KEY can memorize up to 7 different
exposure parameters, those are automatically
renewed by every examination, a max. of 400
programs can be stored in memory.
Technique factors (kV, mAs or kV, mA, sec.),
radiographic mode (manual, photo-timer), technique
factor setting system (2-control, 3-control), focus,
photo-pickup field selection and film density
selection are memorized.

4
UD150B-40/V-40/L-40
POWER REQUIREMENTS
(a) System:
3-phase alternating current
y Frequency: 50 or 60 Hz y Relations between the length and sectional area of
lead-in conductor with respect to the transformer
Model UD150B-40 UD150V-40 UD150L-40 capacity:
Standard voltage 380 V ± 10%, 0.10 Ω or less 380 V ± 10%, 0.10 Ω or less 380 V ± 10%, 0.16 Ω or less Length
Wire Size (Nominal Sectional Area) mm2
and allowable 400 V ± 10%, 0.11 Ω or less 400 V ± 10%, 0.11 Ω or less 400 V ± 10%, 0.17 Ω or less
voltage range at no 415 V ± 10%, 0.12 Ω or less 415 V ± 10%, 0.12 Ω or less 415 V ± 10%, 0.19 Ω or less 10 m
load and power 440 V ± 10%, 0.13 Ω or less 440 V ± 10%, 0.13 Ω or less 440 V ± 10%, 0.21 Ω or less Transformer 20 30 40 50 60 70 80 90 100
or
source impedance 480 V ± 10%, 0.16 Ω or less 480 V ± 10%, 0.16 Ω or less 480 V ± 10%, 0.21 Ω or less Capacity m m m m m m m m m
less

Switch at Type UD150B-40 75 kVA (3-phase) 5.5 8 14 22 22 22 38 38 38 38


Molded case circuit breaker
hand
UD150V-40 75 kVA (3-phase) 5.5 8 14 22 22 22 38 38 38 38
Capacity 75 A or less 75 A or less 50 A or less

UD150L-40 50 kVA (3-phase) 5.5 5.5 5.5 5.5 8 8 8 14 14 14

(b) 3-phase alternating current (Optional auto


transformer is necessary.)
y Frequency: 50 or 60 Hz y Relation between the length and sectional area of
lead-in conductor with respect to the transformer
Model UD150B-40 UD150V-40 UD150L-40
capacity (in case of conduit):
Standard voltage Length
200 V ± 10% Wire Size (Nominal Sectional Area) mm2
and allowable
voltage range at 220 V ± 10%
10 m
no load 240 V ± 10% Transformer
or
20 30 40 50 60 70 80 90 100
Capacity m m m m m m m m m
less
Power source
0.054 Ω or less 0.054 Ω or less 0.087 Ω or less UD150B-40 75 kVA (3-phase) 14 22 38 38 60 60 60 100 100 100
impedance

Switch UD150V-40 75 kVA (3-phase) 14 22 38 38 60 60 60 100 100 100


Type Knife switch and fuse or molded case circuit breaker
at hand
UD150L-40 50 kVA (3-phase) 8 14 14 22 38 38 38 60 60 60
Capacity 100 A or less 100 A or less 100 A or less

(c) Single-phase alternating current


(UD150L-40 only)
y Frequency: 50 or 60 Hz y Relations between the length and sectional area of lead-in
y Standard voltage and allowable voltage conductor with respect to the transformer capacity (in case of
range at no load: conduit):
200 V +10, -5% Length Wire Size (Nominal Sectional Area) mm2
220 V ± 10% 10 m
100
240 V ± 10% Transformer or 20 m 30 m 40 m 50 m 60 m 70 m 80 m 90 m
m
Capacity less
y Power source impedance: 0.051 Ω
50 kVA
14 22 38 60 60 100 100 100 100 100
(single-phase)

y Line switch
Type: Knife switch or circuit breaker
Capacity: 100 A or more
y Rating of fuse or circuit breaker: 60 A or less

AMBIENT CONDITION
When the ambient conditions of each Temperature Humidity Caution
room do not satisfy the following 5 to 35°C 45 to 85% No dew condensation allowed
conditions, it is suggested to install an
Generated Calorie
exclusive air conditioner.
Model
The calorie generated by the apparatus in Apparatus used UD150B-40 UD150V-40 UD150L-40
each room is as follows. Control console Approx. 50 kcal/h
Control cabinet
X-ray high voltage generator Approx. 680 kcal/h Approx. 680 kcal/h Approx. 600 kcal/h

y The above value is at stand-by state.


y Generated calorie for 1 kW
∗ 1 kW = 860 kcal/h

5
UD150B-40/V-40/L-40
CONTROL CONSOLE PANEL VIEW

Operation Panel Home screen display switch


Color TFT touch panel Radiography history display switch
Hand switch Register program display switch
Power ON and OFF switch AEC ON /OFF switch (optional)
Radiography ready switch Density UP and DOWN switch (optional)
X-ray exposure switch

[kV],[mA/mAs],[sec.] display Radiography ready indicator


[kV],[mA/mAs],[sec.] setting shuttle and switch X-ray exposure indicator
Focus select switch Radiography stand-by indicator
[mA]/[mAs] select switch Caution indicator
X-ray tube selected indicator Fault indicator

6
UD150B-40/V-40/L-40
DISPLAY OF THE COLOR TFT TOUCH PANEL

<Radiographic anatomical program selection page>

<Radiographic process selection page>

Radiography programs tab key


Radiography technique key
Photo pickup field select key
Body thickness correction key
Speed (film sensitivity) key (optional)
Heat unit gauge
Radiographic programs key
Automatic stepping of process item key
Radiographic process key

7
UD150B-40/V-40/L-40

<IF fluoroscopy Unit Option is provided. >

IBS Key
IBS(Image Brightness Stabilizer)of ON/OFF
Press the [IBS] to make the key light up.
Fluoroscopic tube voltage down key.
Fluoroscopic tube voltage setting.
Press key to decrease the displayed value.
Fluoroscopic tube voltage up key.
Fluoroscopic tube voltage setting.
Press key to increase the displayed value.
Pulse rate key
Pulse rate setting

NOTE) Pulsed fluoroscopy can be selected if the DR unit incorporates a pulsed fluoroscopy function.

Fluoroscopic mode key.


Fluoroscopic mode setting.
Preset key to alter the modes.
Preset fluoroscopy conditions (fluoroscopy mode) and preset conditions are shown below. "PULSE N" is
optional.

Fluoroscopic mode Preset conditions Remarks

Standard fluoroscopy (continuous


DOSE N -
fluoroscopy)

Fluoroscopy at low X-ray dosage


DOSE L -
(continuous fluoroscopy)

PULSE N Pulsed fluoroscopy Digital image processing system Used.

Fluoroscopic Timer key.


Setting the Fluoroscopic Timer
Hold down the [Fluoroscopic timer] key for three seconds when in other to reset the timer to zero.

8
UD150B-40/V-40/L-40
DIMENSIONS AND WEIGHT
Control console
308 82

345
Front view Side view
Unit: mm
Weight: 2.5 kg

Control cabinet

Top view

Unit: mm
Weight:
D150BC-40 (for UD150B-40) 250 kg
D150VC-40 (for UD150V-40) 250 kg
Front view Side view D150LC-40 (for UD150L-40) 240 kg

Remarks
y Every value in this Product Data Sheet is a standard value, and it may vary a little from the actual at each site.
y The appearances and specifications are subject to change for reasons of improvement without notice.

SHIMADZU CORPORATION. International Marketing Division


Shimadzu Corporation Medical Systems Group has
3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan Phone: 81(3)3219-5641 Fax: 81(3)3219-5710 been certified by TUV Rheinland as a manufacturer
of medical equipment and systems in compliance
URL http://www.shimadzu.com with ISO9001: 2000 Quality Management Systems
and EN ISO13485: 2003 Medical Equipment
Quality Management Systems.