Sie sind auf Seite 1von 12

Oscilloscope calibration

Application Note

A guide to oscilloscope calibration using


dedicated or multiproduct calibrators

From t he Fluke Calibrat ion Digit al Librar y @ w w w.flukecal.com/librar y


User requirements for qualifying measurement Calibration requirements
Oscilloscopes are very complex systems, now emphasizes the Despite the growing increase
instruments, mainly because of need to calibrate even low-cost in oscilloscope functionality,
attempts to provide easy and oscilloscopes. the essential features of faithful
direct access to waveforms, and accurate representation
More sophisticated
then to permit both qualitative remain few:
and quantitative analysis. oscilloscopes • Vertical deflection
Users demand enough flexi- In many cases, these are coefficients
bility to deal with a wide rangemore specialized instruments,
concentrating on such features • Horizontal time coefficients
of functions, frequencies and
voltages without having to buy as multi-channel comparison, • Frequency response
an array of instruments. computation, data collec- • Trigger response
tion and dual-sourced Y-axis
deflection. (e.g. presenting Techniques and procedures
Need for calibration both time and frequency bases for calibration must measure
simultaneously on the same these features, while coping
Older low-cost oscilloscopes with the functional conditions
screen).
At the low-cost end of the which surround them. Good
Oscilloscopes passed through
range of oscilloscopes, we can metrological practice must be
a phase of using a mainframe,
remember older analog instru- used to ensure that an oscillo-
with plug-in modules carrying
ments, with deflection accuracy scope’s performance at the time
specialized hardware. Subse-
and bandwidth so limited as to of use is comparable with that
quently, with the introduction
present merely a rough picture observed and measured during
of microprocessors, the devel-
of a signal. Power supplies calibration. This will provide
opment of the Digital Storage
were often unregulated and confidence in certificates of
Oscilloscope (DSO) permits
there was no external means traceability and documentation
functionality to be more-
of X/Y gain or bandwidth which result from calibration.
effectively controlled by suites
adjustment.
of software, allowing specialist Manual or automated
The accuracy of such an
programs to dispose of much of calibration?
instrument was seriously
the specialized hardware. Manual calibration methods
degraded by fluctuations in
DSOs have many advantages are well established, and for
the line supply, accompanied
when the signal is repeti- analog oscilloscopes there is
by component temperature
tive and can be retained for possibly no cost-effective alter-
and time drift. These types of
examination by inbuilt mea- native, although techniques are
instruments possessed poor
surement programs and signal being developed which employ
performance repeatability, and
transfer to other systems (e.g. oscilloscope calibrators togeth-
calibration would have been
for pass/fail tests or hard- er with memorized calibration
largely a waste of time; nowa-
copy printing). They can also procedures directed at indi-
days, however, calibration will
capture and display pre-trigger, vidual oscilloscope models.
more often be required.
single-event and short-lived These procedures use a form of
Modern developments in waveforms which present prompted manual calibration.
low-cost oscilloscopes difficulties with analog oscillo- For DSOs, which are based
Modern low-cost instruments scopes. Some transients cannot on programmable digital
are vastly different from the be displayed on analog oscil- techniques, and may already
older image presented above. loscopes with sufficient light be programmed to respond
Many newer low-cost oscillo- output to be viewed conve- to remote signals (say via the
scopes have resulted as a niently, but capture in a DSO IEEE-488 interface), auto-
spin-off from the develop- permits enhancement of the mated calibration can be
ment of more expensive and light output. achieved, with great benefits
sophisticated instruments, Because the DSO depends to repeatability, productivity,
with significant improve- largely on sampling techniques, documentation generation, and
ments in component quality, for some applications this statistical control.
performance cannot replace the purer ‘real- This guide is intended as
repeatability, and expanded time’ nature of the signals. For an introduction to the basic
functionality. example, when viewing ampli- techniques of oscilloscope
The advent of more stringent tude modulated waveforms and calibration, and will concen-
quality standards (such as jitter signals on a DSO, ‘aliased trate on the types of tests and
the ISO 9000 series), bring- information’ can distort the adjustments which are likely
ing insistence on traceability presentation due to the need
for incompatible sampling rates.

2 Fluke Calibration Oscilloscope calibration


to be used by both manual and • Accuracy of vertical Vertical deflection
automated methods, and not deflection accuracy
differentiating between them. • Range of variable vertical Amplitude
Oscilloscope display controls The Y-axis is used, almost
geometry • Vertical channel switching exclusively, for displaying the
Before it is possible to calibrate • Accuracy of horizontal amplitude of incoming signals.
the main parameters it is nec- deflection These are processed through
essary, for many oscilloscopes, ‘channel’ amplifiers (mainly
• Accuracy of any int. two channels, often four or
to ensure that the essential
calibrator more).Basic setup features
geometry of the oscilloscope
is set up correctly. This may, • Pulse edge response include:
in fact, be regarded as part • Vertical channels bandwidth • Zero alignment to graticule
of the calibration process, • Z-axis bandwidth (Offset)
as the parameter measure- • Vertical amplifier balance
ments are dependent on visual • X-axis bandwidth
• Horizontal timing • Vertical channel switching
observations.
In real-time (analog) oscil- • Time base delay accuracy • Operation of alternate/
loscopes, the graticule is a chopped presentations
• Time magnification
separate entity from the screen Multiple-traces are created
images. This means that if the • Delay time jitter
using alternate-sweep switch-
graticule is to be used as a • Standard trigger functions ing or ‘chopped’ high-speed
measurement tool, alignment to • X-Y phase relationship switching. In alternate-sweep
it must be included in the cali- switching, the trace completes
bration process. The innovation before switching to the next.
of the Electronic Graticule in Parameter details With ‘chopped’ high-speed
DSOs has largely removed the switching, usually used for
need to establish geometrical Geometry setup
Although setting up the dis- low frequency signals, inputs
links between screen data are sampled alternately at
and the graticule—this is done play geometry may not be
strictly regarded as a calibra- high speed and steered into
automatically, and tube rota-
tion does not disturb relative tion parameter, the oscilloscope separate channels. DSOs use
alignments. display is the window through different forms of switching to
which most of the (visual) achieve similar effects. Which-
Where an electronic cursor ever system is in use, there will
is used, this links internally measurements are made. The
with trace data and channel display geometry should be set be a series of alternative chan-
sensitivities, tied to an inter- up, or at least examined, before nel amplifiers and attenuators
nal dc voltage standard. In going ahead with calibration, if whose gain characteristics are
only to ease the measurement the major influence on vertical
these cases the main require- accuracy.
ment is to calibrate the voltage processes. Examples of geom-
standard. etry features are:
• CRT alignment
Parameters to be calibrated • Earth’s field screening or
At this point, it may be useful compensation
to provide a list of the param-
eters which need to be verified • Range of focus and intensity
or calibrated in order to ensure controls
traceability in the majority of • Barrel distortion
oscilloscopes. This list breaks • Pincushion distortion
down the features listed
• Range of Y- and X-axis
earlier:
positioning controls

Fig. 1.1 Standard graticule.

3 Fluke Calibration Oscilloscope calibration


There are five main All Fluke scope calibrators
parameters to be checked provide a 1 kHz square wave
in calibrating each vertical for testing the LF gain of ac-
amplifier system: offset, gain, coupled amplifiers.
linearity, bandwidth and pulse Channel amplifiers’ linear-
response. ity can be tested by injecting
These parameters are crucial either a dc or a square wave
to achieve accurate representa- signal, varying the amplitude
tion of the signal. For effective and checking the changes
comparisons between signals against the graticule or cursor
applied through different chan- readings. Fig. 1.2 DC voltage – gain.
nels, their channel parameters
Pulse response
must be equalized.
Viewing the rise time of
Measurement of a chan-
pulse fast edges is one of
nel amplifier’s gain is usually
two complementary methods
performed by injection of a
of measuring the response
standard signal and measure-
of the vertical channel to
ment of its presentation against
pulsed inputs (the amplifier’s
the display graticule. Because
bandwidth should also be
the amplifier coupling may
measured—refer to the section
be switched between ac/dc
Channel Bandwidth).
and often 50 Ω/1 MΩ, it will
Response to fast edges
be necessary to inject signals
depends on the input imped- Fig. 1.3 LF square wave – gain. The
which test the operation of
ance of the oscilloscope to be peak-to-peak value shown on the
each of these forms of coupling.
tested. Two standard input screen (b) is compared with the known
Two standard signals for value (a): b ÷ a = Gain at 1 KHz.
impedances are generally in
measuring an amplifier’s gain
use: 50 Ω and 1 MΩ//(typically)
are usually employed:
15 pF. 1 MΩ is the industry
i. With dc coupling, either a
standard input generally used
dc signal (Fig. 1.2 – includes
with passive probes. Where the
offset) or a square wave
50 Ω input is provided it gives
(Fig. 1.3 – can be manipu-
optimal matching to HF signals.
lated to remove the offset)
To measure the rise time, the
is injected, and the chan-
pulse signal is injected into the
nel’s response is measured
channel to be tested; the trig-
against graticule divisions or
ger and time base are adjusted
cursor readings.
to present a measurable screen
All Fluke scope calibrators
image, and the rise/fall time is
provide dc voltage and 1
measured against the grati-
kHz square wave outputs for
cule or cursor readings. The
testing the gain and offset of
observed rise/fall time has
dc coupled amplifiers.
two components: that for the
ii. With ac coupling, a square
applied signal and that for the
wave signal is injected at
channel under test. They are
1 kHz, and again the chan-
combined as the root of the
nel’s response is measured Fig. 1.4 LF square wave – distortions.
sum of squares, so to calculate
against graticule divisions or
the time for the UUT channel, a
cursor readings.
formula must be used:
Using a low frequency pulse UUT rise/fall time =
can also provide a rough Square root [(Observed time) –
2

check of the gross LF and (Applied signal time) ]


2

HF response (Fig 1.4). This is


only a very rough test of gross
distortion. A result which looks
square must still be checked for
pulse response and bandwidth.

4 Fluke Calibration Oscilloscope calibration


In some oscilloscopes the Channel bandwidth
vertical graticule is specially As well as determining the
marked with 0 %, 10 %, 90 % pulse response by viewing a
and 100 % to make it easy to specimen pulse on the screen,
line up the pulse amplitude this should be supported by
against the 0 %/100 % marks, measuring the amplifier’s
then measure the 10 %/90 % bandwidth using a ‘leveled
crossing points against marks sine wave’. This is done at an
on the center horizontal grati- input impedance of 50 Ω, to
cule line. maintain the integrity of the
50 Ω source and transmis-
Measurement
sion system. For high input
In all Fluke models, two differ- Fig. 1.5 Measurement of rise time.
impedance oscilloscopes, an
ent sort of pulses are used:
in-line 50 Ω terminator is
• Low Edge Function: a low
used to match the line at the
voltage amplitude pulse oscilloscope input. The in-line
matched into 50 Ω with a 50 Ω could take the form of
rise/fall time less than or a separate 50 Ω terminator
equal to 1 ns. When using or be incorporated within an
the formula to calculate the ‘Active’ head—the latter gives
UUT rise/fall time, the applied the benefit of full automation
signal rise time must be that and requires no additional
certified at the most-recent calibration.
calibration of the calibrator, First the displayed amplitude
closest to the amplitude of the of the input sinusoidal wave is
applied pulse. measured at a reference fre- Fig. 1.6 Leading edge aberration.
• High Edge Function: a high quency (usually 50 kHz), then
voltage amplitude pulse the frequency is increased,
matched into 1 MΩ with a at the same amplitude, to the
rise time less than or equal specified 3 dB frequency of the
to 100 ns. This function is channel. The displayed ampli-
used mainly to calibrate the tude is measured again.
response of the oscilloscope’s The bandwidth is correct
channel attenuators. if the observed 3 dB point
amplitude is equal to or greater
Leading-edge aberration
than 70 % of the value at the
In Fig. 1.5, some leading-edge
reference frequency.
aberrations (overshoot and
If it is needed to establish the
undershoot) are shown at the
actual 3 dB point, the frequency
top end of the edge, before the
should be increased until the Fig. 1.7 Setting the amplitude at the
voltage settles at its final value
peak-to-peak amplitude is 70 % reference frequency.
(which is the value defined as
of the value at the reference
100 %).
frequency, then this frequency
Where scope specifications
is close to the 3 dB point.
include aberrations, the speci-
fication limits can be expressed
as shown in the shaded area of
the magnified Fig. 1.6 (typical
limits shown).
When aberrations are dis-
played for measurement, they
should be within the specifi-
cation limits, although where
the oscilloscope’s aberration
specification approaches that
Fig 1.8 Measuring the amplitude at the
of the calibrator, other methods 3 dB point frequency.
must be used.

5 Fluke Calibration Oscilloscope calibration


Horizontal DSOs generally employ a
deflection accuracy vertical channel amplifier as
Introduction the horizontal amplifier, so
The X-axis is dedicated almost having measured the verti-
exclusively to use as the vehi- cal deflection bandwidth,
cle for the time base(s). As well there is generally no need to
as two vertical channels, there measure horizontal deflection
will often be two time bases: bandwidth.
Main and Delayed. These may Fig. 1.9 Setting the trace length at the
reference frequency.
be achieved in DSOs by two Horizontal timing
independent sampling rates, or accuracy
via a positioned ‘zoom’ window
on a single, but long, store. Test setup
When determining the accu- In this test the time base is
racy of horizontal deflection, switched to the sweep speed
where applicable, the geometry (or time/div) to be checked,
of the display must have first and the output from a timing
been set up. It is assumed that marker generator is input via
this will be included as part of the required vertical channel.
the initial geometry setup. On oscilloscope calibra- Fig. 1.10 Measuring the trace length at
Once this has been done, tors these are square waves, the 3 dB point frequency.
the following adjustments or changing to sine waves at a
checks can be attempted: specific frequency.
• X-axis bandwidth Timing calibration accuracy series of differentiated edges in
• Horizontal timing A timing accuracy of 25 ppm one direction, with the return
will be sufficient to calibrate edges suppressed. This leads
• Timebase delay accuracy
most real-time oscilloscopes to difficulties in DSOs due to
• Time magnification and many DSOs, although a sampling, in which the comb
• Delay time jitter timing accuracy better than peak can fall between samples,
• Trigger functions 0.3 ppm is required for some leading to amplitude variations
higher-performance DSOs. and difficulty in judging the
• X-Y phase relationship precise edge position. The use
Why use square waves?
X-axis bandwidth In the past, timing markers of timing markers in the form
For real-time oscilloscopes, the have taken the form of a ‘comb’ of square or sine waves signifi-
horizontal amplifier’s band- waveform, consisting of a cantly reduces the inaccuracies
width will be checked using a due to this 1-dot jitter.
‘leveled sine wave’, similar to
the checks of vertical chan-
nels, but with the time base
turned off. This consists first of
measuring the displayed length
of the horizontal trace (Fig. 1.9),
for a sinusoidal wave provided
as X input at a reference fre-
quency (usually 50 kHz).
The frequency is then
a. Initial state before deviation adjustment
changed, at the same ampli-
tude, to the specified 3 dB
point of the horizontal amplifier
and the displayed trace length
is measured again (Fig. 1.10).
The bandwidth is correct if
the observed 3 dB point trace
length is equal to or greater
than 70 % of the length at the
reference frequency.
b. Aligned state after deviation adjustment

Fig. 1.11 Adjusting the marker generator’s deviation for correct alignment.

6 Fluke Calibration Oscilloscope calibration


Measurement the delay control is adjusted to the required vertical chan-
The marker timing is set to align the time marker edge to nel, and the oscilloscope is
provide one cycle per divi- the same vertical datum line switched to display 10 cycles
sion if the horizontal timing is (Fig. 1.13 {b}). The setting of the per division as illustrated in
correct. oscilloscope’s delay is again Fig 1.14 {a}. The timing marker
By observation, the marker noted. generator frequency/period
generator’s deviation control Finally, the two settings is adjusted to give exactly 10
is adjusted to align the mark- of the oscilloscope delay are cycles per division.
ers on the screen behind their compared, to check that their The errors are likely to be
corresponding vertical graticule difference is the same as the greatest on the right of the
lines, and the applied deviation time between the two selected trace (the longest time after the
is noted. The applied deviation markers, within the specified trigger), so the oscilloscope’s
should not exceed the oscillo- limits for the oscilloscope. horizontal position control is
scope’s timing specification. Horizontal x10 magnification adjusted to place the marker
The operation is repeated for accuracy edge at ‘A’ at the center of the
all the sweeps and time base The output from a timing screen.
time/division settings desig- marker generator is input via
nated for calibration by the
oscilloscope manufacturer.
Time base delay accuracy
For this test it is assumed that
the delayed time base is indi-
cated as an intensification of
the main time base, and can be
switched to show the delayed
time base alone. For all oscillo-
scopes, ensure that the retrig- a. Delayed time base intensified on the main time base.
ger mode is switched off.
The output from a timing
marker generator is input via
the required vertical chan-
nel, and the oscilloscope is
adjusted to display one cycle
per division as illustrated in
Fig 1.12 {a}. The oscilloscope
mode switch is set to inten- b. Delayed time base alone with Datum marker.
sify the delayed portion of the
main time base over a selected Fig. 1.12 Adjusting the delayed time base to the first Datum marker.
marker edge as shown (this
may require some adjustment
of the oscilloscope’s Delay
control).
The oscilloscope delay mode
switch is set to display the
delayed sweep alone, and the
delay control is adjusted to
align the time marker edge to a
chosen vertical datum line (e.g.
a. Delayed time base intensified on the main time base.
center graticule line as shown
at Fig 1.12 {b}). The setting of
the oscilloscope’s delay is noted.
The oscilloscope mode switch
is set to intensify the delayed
portion of the main time base
over a different selected marker
edge—(Fig. 1.13 {a}).
The oscilloscope delay mode
b. Delayed time base alone with Datum marker
switch is again set to display
the delayed sweep alone, and Fig. 1.13 Adjusting the delayed time base to the second Datum marker.

7 Fluke Calibration Oscilloscope calibration


The oscilloscope is set to delayed time bases, a very Trigger operation
display the X10 sweep, and small part of the main time Standard functions—
the horizontal position control base is intensified, and adjust- introduction
is adjusted to align the marker ment may be difficult). For most oscilloscopes, a wide
edge ‘A’ exactly to the center The 20 ms period output from variety of trigger modes exist,
graticule line. a timing marker generator is being sourced either via a
The marker generator Fre- input to the required vertical nominated Y-input channel, or
quency/Period deviation control channel, and the oscilloscope from a separate external trigger
is adjusted to align the marker is adjusted to display one cycle input.
edges exactly to the graticule per division (20 ms/div) as The functionality of the trig-
lines as shown at Fig 1.14 {b}. illustrated in Fig 1.15a. ger modes allow for ac or dc
The marker generator Fre- The delayed time base is coupling, repetitive or single-
quency/Period deviation setting set to run at 1 µs/div, and the sweep, and trigger-level control
is noted. This setting should be oscilloscope mode switch is set operations.
within the specified limits for to intensify the delayed portion These tests check the opera-
the oscilloscope. of the main time base over the tion of:
Similarly, for a DSO, the center marker edge as shown • Internal trigger sensitiv-
range of available ‘Zoom’ or using the oscilloscope’s delay ity in both polarities, from
X-magnification’ factors are time control. each of the available Y-input
calibrated as designated by the The oscilloscope delay mode channels
manufacturer. switch is set to display the
delayed sweep alone, and the • Operation of the trigger level
Delay time jitter control for a sinewave exter-
The delay jitter on oscilloscopes delay control is adjusted to nal trigger input
is often measured under time align the time marker edge to a
chosen vertical datum line (e.g. • Effect of vertical position on
magnifications of the order of
center graticule line as shown trigger sensitivity.
20,000:1. This means that the
delayed time base must run at Fig 1.15b). • Minimum trigger levels for
20,000 times faster than the The width of the verti- normal and ‘trigger view’
main time base (for a main cal edge (which displays the modes
time base running at 20 ms/ jitter) of the displayed portion
• Bandwidth of trigger circuits,
div, the delayed time base must of the waveform, measured
and effect of HF rejection
run at 1µs/div). along a horizontal axis, should
filters
For this test the intensifica- not exceed the oscilloscope’s
specified jitter limits (i.e. in • LF and dc performance of the
tion of the main time base trigger circuits
is adjusted onto the edge at this example, for 20,000:1
the center graticule line (with specification, the oscilloscope’s • Single-sweep performance
such a difference between contribution to the width and response to position
the speeds of the main and should be less than 1 division). controls
Note: Tests which are performed using a
Y-channel input are also carried out on all
the other available Y-channels.

a. Markers set at 10/div at x1 magnification. a. Delayed time base intensified on main time base.

b. Markers set at 10/div at x10 magnification. b. Edge showing jitter on delayed time base.

Fig. 1.14 Checking the effect of x10 magnification. Fig. 1.15 Measuring the delay time jitter.

8 Fluke Calibration Oscilloscope calibration


Internal triggers— External triggers Internal triggers—
trigger level operation dc-coupled operation
a. Initial setup
a. Initial setup These tests start with the 200 a. Initial setup
A standard 4 Vp-p (50 Ω) mV signal, described in the With the Ext Trig input dis-
reference sinusoidal signal is previous paragraph ‘Display connected, and the Y-channel
input via ac coupling into the Triggers’ feature, applied to the input externally grounded, the
Y-input channels in turn. Using external trigger input of the oscilloscope Y-channel is set to
internal triggers and dc trigger oscilloscope. ‘DC-coupling’ and trigger mode
coupling (not ‘AUTO’), the +ve for ‘internal triggers’ from the
b. Presence of a trace
and -ve slopes are selected in Y-channel. There should be no
Adjustment of the oscilloscope’s
turn. The sweep speed setting trace on the CRT.
trigger level control should be
is 10 µs/division; the Y-channel
able to produce a trace. The Ext b. DC triggering
sensitivity is 0.5 V/division so By adjusting the Vertical Posi-
Trig input is disconnected
that the input signal occupies 8 tioning control to pass through
and reconnected again, while
divisions. a point in its range correspond-
checking that the trace disap-
b. Trigger level adjustment pears and is then reinstated. ing to the Trigger Level setting
Over almost all of its range of and selected slope direction, a
c. Trigger sensitivity
adjustment, the trigger level single trace should appear then
With the input signal reduced
control must be shown to disappear.
to the minimum amplitude
produce a stable trace, moving c. ACLF rejection trigger
specified by the manufacturer,
the starting point over a range
adjustment of the Trigger Level mode
of levels up and down the With the input signal set as
control must be shown to
selected slope of the displayed for the Trigger Bandwidth
regain stable triggering.
sine wave. check, the ACLF Reject feature
d. Trigger bandwidth
c. Trigger sensitivity (if available) is activated and
With the input signal set to the
With the input signal reduced paragraph (b) is repeated. The
minimum amplitude and maxi-
to 10 % of its amplitude, single trace action should not
mum frequency specified by
adjustment of the trigger level occur.
the manufacturer, adjustment
control must be shown to reac-
of the trigger level control must
quire stable triggering. With
be shown to acquire stable External triggers—
trigger coupling switched to ac,
triggering. The Ext Trig input is Single-sweep opera-
and using vertical positioning
disconnected and reconnected tion
to place the trace at extreme
again, while checking that the
upper and lower limits of the a. Initial setup
trace disappears and is then
CRT screen in turn, stable trig- (Applies only to those scopes
reinstated.
gering must be maintained. with single sweep capability).
e. ACHF rejection trigger With the Ext Trig input con-
d. ‘Display triggers’ feature
mode nected, the oscilloscope is set
If the oscilloscope has a
With the input signal set as to ‘Single Sweep’, and trigger
‘Display Triggers’ or ‘Trigger
for the trigger bandwidth mode for ‘Internal Triggers’
View’ feature, this is selected
check, the ACHF Reject feature from the Y-channel. There
to display the trigger region of
is activated then deactivated should be no trace on the CRT.
the waveform. Using a 200 mV
again, while checking that the
sinusoidal signal input to the b. Single sweep triggering
trace disappears and is then
channel, the trigger region is Pressing the ‘Reset’ or ‘Rearm’
reinstated.
checked for correct amplitude switch should produce a single
on the display. trace. This action should not
N.B. In the following trigger produce a trace when the Ext
operations, during tests on a Trig input is disconnected.
DSO, the trace will not disap-
pear as a result of the inter-
ruption of the trigger (or
reduction of its amplitude
below the threshold). Instead,
the trace will remain but
not be refreshed, and this is
the condition to be detected.

9 Fluke Calibration Oscilloscope calibration


Low frequency control should not be able to Phasing test
triggers acquire a stable display for
a. Initial setup
either + or - slope selection.
a. Initial setup The oscilloscope controls are
A 30 mV, 30 Hz sinewave set as follows:
Z-axis
signal is input simultaneously Vertical mode: X-Y;
to Channel 1, Channel 2, Ext Z-axis input Sensitivity: 5 mV/div,
Trig Sweep A (main time base) If provided, the Z-axis input is both
and Ext Trig Sweep B (delayed usually positioned on the rear channels
time base). The oscilloscope is panel, but sometimes can be Ch 1 or X: AC coupled
set for: trigger mode to ‘Inter- found near the CRT controls on Ch 2 or Y: Grounded
nal Triggers’, Channels 1and the front panel. DSOs generally Vertical mode: X-Y
2 sensitivity to 10 mV/div, and do not have a Z-input. Vertical position: Central
sweep speed to 5 ms/div. Both Horizontal position: Central
main and delayed time bases Z-axis bandwidth N.B. During X-Y phasing tests
should be displayed when on a DSO, maximum sampling
a. Initial setup
selected, for both channels. rate would be used. Even
A 3.5 Vp-p, 50 kHz sinewave
b. Channel 2 grounded is applied to both Channel 1 so, the visible extent of any
With Channel 2 input and Ext Trig inputs. The sweep captured lissajou is limited to
grounded, and Channel 1 set speed, trigger slope and trigger interrupted segments by the
for 0.1V/div with the trigger level controls are set to provide store length, until the test fre-
selector set to Channel 1, stable a stable display of 1 cycle per quency is high enough for an
displays should appear as division. entire cycle to be captured.
expected. b. Trace acquisition
b. Signal transfer to Z input
c. Channel 1 grounded The signal input to Channel 1 The display intensity is
With Channel 1 input is disconnected and transferred adjusted until a horizontal trace
grounded, and Channel 2 set to the Z-axis input. the trace is just visible (should be 10
for 0.1V/div with the trig- should collapse to a series of divisions long). After the X and
ger selector set to Channel 2, bright and dim sections. Using Y position controls have been
stable displays should appear the oscilloscope brightness used to center the trace, the
as expected. control, the trace is dimmed so intensity and focus controls are
that the brightened portions adjusted for best display.
d. ACHF reject
just disappear. c. Phasing check
With Channel 2 input grounded,
and Channel 1 set for 50 mV/ c. Bandwidth check The common input signal is
div with the trigger selector set The frequency of the input reduced until the trace is 8
to Channel 1, the ACHF Reject sinewave is increased to the divisions long. Channel 2 (or
feature is activated for both exact specified Z-axis band- Y) input mode is switched to
Sweeps A and B. Adjusting the width point. The amplitude of dc, and the X and Y position
trigger level control should the sinewave is increased to 5 controls are used to center the
acquire a stable display. Vp-p. Adjustment of the sweep (now sloping) trace. If the X
speed and trigger level controls and Y channels do not intro-
e. Positive and negative duce any phase error, then the
should acquire a dotted, or
slope operation intermittently brightened trace. center of the trace will pass
With Channel 2 input grounded, through the origin. Phase error
and Channel 1 set for 10 mV/ X-Y Phasing between X and Y channels will
div with the trigger selector set cause the sloping trace to split
X input
to Channel 1, adjusting the trig- into an ellipse, which for small
Depending on the type of oscil-
ger level control should acquire phase errors will be apparent
loscope, the X input will be
a stable display for both + and only close to the origin. The
applied either via the External
-slope selections. trace separation at the origin,
Trigger connector, or via Chan-
f. ACLF reject nel 1, with suitable switching. along the center horizontal
With Channels 1 and 2 set In either case, the same signal graticule line, should be not
for 10 mV/div with the trigger of 50 mV, 50 kHz will be greater than 0.4 division for
selector set to either channel, applied to both X and Y inputs. a commonly-specified phase-
the ACLF Reject feature is acti- shift of 3 degrees.
vated for both Sweeps A and
B. Adjusting the Trigger Level

10 Fluke Calibration Oscilloscope calibration


Other useful reference material:
Fluke publications:
1612935 A ENG-N 02/2001 “Fully Automated True Bandwidth Testing of High
Performance Oscilloscopes”
B0252EEN Rev B 02/97 “How many Calibrators do you need to meet ISO9000”
1282496 A-ENG-N 09/99 “In-House Calibration - Is it best for you”
Websites:
www.flukecal.com
www.lecroy.com
www.tektronix.com
www.tm.agilent.com

11 Fluke Calibration Oscilloscope calibration


Fluke Calibration. Precision, performance, confidence.™

Fluke Calibration Fluke Europe B.V.


PO Box 9090, PO Box 1186, 5602 BD
Everett, WA 98206 U.S.A. Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2000-2013 Fluke Calibration. Specifications subject to change without notice.
Printed in U.S.A. 5/2013 1626187C_EN

Modification of this document is not permitted without written permission


from Fluke Calibration.
12 Fluke Calibration Oscilloscope calibration

Das könnte Ihnen auch gefallen