For KJ-A1200 Atmosphere Furnace
Thank you for purchasing KJ-A1200 Atmosphere Furnace. To avoid any misuse and damage, please read the
operation instruction carefully before operation.
1. Introduction
KJ-A1200 atmosphere furnace uses resistance wire with Mo as heating element. Max. temperature inside furnace
can be reached at 1200℃. The furnace temperature profile can be set up by 30 segments and run automatically by
the advanced temperature controller. It is excellent for material research, ceramic research labs to develop various
new materials.
2. Technical Specifications
Thermocouple K Type
3. Furnace Structure
3
4. Instruction of the Intelligent Temperature Controller
4
(3) Indication of Front Panel of Temperature Controller
On Front Panel of Yudian temperature controller, there are several signs and icons such as OUT,
AL1, AL2, AUX, PV, SV, , , , and , which indicate the meaning as the Fig.2
When turn power on, controller display show the model No of controller, software version first.
5
Several seconds later, controller will display temperature condition, and enter starting state. PV shows
real temperature, and SV shows setting temperature.( Fig.3) If “SV” flashing, and shows “ Stop”, it
means that control program is at stop state; If “SV” shows “ Hold”, means that program is at the pause
stage.
Touch key for once again, PV will show the setting time in the step, and SV shows the time that
Please don’t change any parameter unless you understand what parameter is. The all parameter has
been preset according to our experience. Next chapter will explain how to change “Parameters “.
Please be advised that If no any key touch action on the panel, Display will return to
“Starting State “automatically. And all revised data will be saved.
Fig. 4. Then touch Key for one second, the panel enters the stage of temperature
program setting as Fig. 7.
On the display, PV shows first segment value for temperature, SV shows Initial temperature
setting in this segment.
6
Then touch Key , program enters the segment for time setting, as Fig.8.
PV show the segment number for time, SV shows time setting ( Munites ) in this segment
By touch key , and , you can increase or decrease the value to be set.
By touch key and uses key you can get in next segment for temperature or time
setting.
During program setting, by touching for two seconds, you can return to previous set to
revise.data.
By touching key first, then touch key, you can exist program setting.
If no key operation for 30 seconds, display exists program setting and return to “starting state”
.
(4) Example for Setting a Temperature Control Program with 6-segments
For a complicated temperature control profile, we strongly suggest you to make drawing as Fig. 9, then
make form as table 1 to list all data in every segment.
Fig.9. is the temperature profile that we would like to set. The parameters shall be entered as Tab. 1.
Table 1
7
Segments # Symbol in panel Data to be in-put Meaning in the program
1 C 01 0 Initial temperature
2 t 01 60 Ramping time from 0-300℃
Average heating rate is 5℃/min
3 C 02 300 Target temperature value to first
heating stage
4 t 02 10 Soaring 10 minutes at temperature flat.
(300℃)
5 C 03 300 Temperature value at the heating flat
6 t 03 240 Second heating time from 300-1000℃.
Average heating rat is 5℃/min
7 C 04 1000 Valor de temperatura objetivo a la etapa
de calentamiento máximo (1000 ℃)
8 t 04 30
9 C 05 1000 Soaring time at 1000℃ stage
10 t 05 260 Cooling time to 260℃
Cooling rate is 4.8℃/min
11 C 05 200 Target temperature to be cooled
(200℃)
12 t 06 -121 Program end,output power off.
Furnace cooling naturally
Using 4 keys of , , , , enter data listed the above table into controller
When temperature program set up ready, touch key for two seconds, then display SV will show
letter “run” , furnace will run automatically segment by segment according to program step by step.
Under furnace running state, “ Out “ indicator ‘s brightness will change based on power out value.
If you want the furnace to stop running temporarily, please push key for two seconds,
8
then display SV will show letter “ Hold”, the furnace enters “pause state “. In the “pause state”,
controller will keep furnace temperate at the value when “pause” order was given, but time running is
stop.
Under the “ Pause “ state, push key for two seconds, SV display will shows “ run”. And
furnace will start running again from the point where is paused.
If you want to stop running furnace, whatever under “ pause” or “ running “ state, you can push
key for two seconds, then, SV display shows “ stop” , furnace totally stop running and controller will
be in “starting sate”. If you want to run again, the program will start at the beginning step. If furnace
temperature still is higher than “ C 02”, program will not run until furnace cooling down to “ C 02”. In
order to run faster, you can choose program run from “step 2” , or “step 3”.
The following function parameters are preset in the temperature controller. They are very important for
controlling furnace temperature stably and accurately. Unless you have enough experience, please
don’t change the preset parameters in the controller.
In order to change the function parameters, follow procedure as below:
Touch key for two second under “starting state”, display will enter parameter setting
function
Touch key for one second again, PV display will shows symbol: “ M5 “ ( Fig. 10 ) , “ P”
(fig.11), “t” (fig. 12 ), “CtrL” ( fig 13) and “ LOC” (fig.14) separately.
Using and key to change the value under different parameter setting.
Push key for two seconds, setting will go back to previous parameter.
setting”.
9
M5 value is larger,adjust time is longer 200-500)
and M5 smaller time is shorter.
P 1-9999 Adjust rate in controller. P value 5
Speed parameter increases adjustment (choose between3-10)
faster,decrease,adjustment slower.
T 0-2000 Control temperature over shooting. t 6
Delay time parameter value smaller,temperature overshoot (3-10)
smaller,otherwise overshoot larger.
CtrL 1,2,3,4 1= Auto tune from front panel 3
Control type 2= Auto tune first,then go to 3 or 4
3= Built in auto tune,can not be changed
from front panel
4= More accurate auto tune
LOC 0-9999 Preset in controller. Please never change 0
Parameter lock
Again, only in the case that you find temperature control is not stable during running or furnace the above
parameters may consider to be adjusted.
Before adjust the parameters, you shall use “ Auto-Tune” function to achieve the best setting faster.
The procedure is as the follows:
Let furnace stay in a temperature that it is most important for you.
Set Loc value to 2
Let display return to “starting state “
Push key for two seconds, then front panel of controller will flash with letter “ AT”, wich
In the special case, the furnace can not be run properly due to voltage variations in different stage, you
may need to adjust some parameters as the Table 3.
Again, The parameters have been preset according to our experience. Please don’t adjust the
following parameters unless you are very familiar with the function of parameters
Table 3.
10
dHAL 999.9 Alarm in positive tolerance
dLAC 999.9 Alarm in negative tolerance
dF 0.3 Adjustment difference The smaller dF,the higher accuracy
auto tune
CtrL 3 Control type
M5 289.7 Maintenance parameter
P Speed parameter
t Delay time parameter
Ctr 1 Output period Reflect controller’s adjustment speed
Sn 6 Thermocouple type K type T.C
DIp 0 Position of decimal
DIL 0 Display value in Min.
DIH 1700 Display value in Max.
Sc 0.0 Main input shift and adjustment Tolerance between input and sensor
OP1 1 Output type 1 is output from 0-10mA
Please make sure this value is
compatible with the controller
OPL 18 Output limit below 200℃ This value will limit max. Output
current when the temperature below
200℃. Please select 16 first. If found
output current is too low,please
increase the value up to 80. The most
suitable value in the current meter
is120-140 A in initial heating stage.
OPH 100 Output up limit Limit Max. Output power. Max. Value
is 130. Please the value as low as
possible to avoid the damage of heating
elements.
ALP 18 Alarm function
CF 16 System function selection 82 means limit output current in two
stages
Addr 1 Communication address
bAud 9600 Communication frequency
dL 1 Input digital filter The large dL value,the more stable
measured temperature,but reaction time
is slow
run 60 Running condition
Loc 0 Parameter revising grade Value “0” will lock data that has been
entered. “808” will open lock to allow
you to revise all parameters
EP1 M5
EP2 P
EP3 t
11
EP4 ControL
EP5 OP2 After adjusting,set to None E
EP6 OPL After adjusting,set to None E
EP7 None E
EP8 None E
, , step by step. After changing the parameter and make sure all parameters are
correct, you need to change “ LOC” to ‘0” as shown in Fig. 14. to lock all data entered without change.
6. Installation Procedures
Please follow the instructions as the below for furnace installation
(1)Open shipping package to check if all components are good condition. If find any damage caused by shipping,
please report it to Zhengzhou Kejia Furnace Co.,Ltd immediately by email at contact@kejiafurnace.com
(2)The furnace must be placed in flat surface to avoid vibration, where must keep from flammable and explosible
materials.
(3)The furnace uses AC 110V / 4.5KW power. Please make sure that power source in your lab is enough to meet
this power requirement. You must use a power plug ( not included ) at > 40A current rating to connect furnace. Fig.
15 shows three power lines from the furnace. Please make sure Fire 1 and fire 2 must
connect to 110V single phase power line, and earth wire must connect to ground pole at the power.
Please refer to section 10 (Knowledge for AC power connection ) for how to get 220V single phase power from
power panel. We strongly suggest you to contact a professional or licensed electrician to make power connection.
Please insert thermal couple tube ( ceramic tube)in to furnace from backside of furnace, then tighten screw
to fix thermal couple’s position
12
7. Instructions for Quick Start:
Procedure for running furnace
Plug in power ( Green Power indicator is On, cooling fan run )
Turn Power switch lock On, (Yudian 518P controller is On )
Let Temperature control display is at “ starting state “, e.g PV panel shows temperature value ,
and
SV display “ stop”. If controller is no at the state, touch key, 518P controller shall
Push Key on the controller and hold for 2 seconds, PV panel shows “ Run”, now furnace is
running automatically
Manual Operation
In special case, you can use manual operate furnace as the follows
Push Manual/Auto button, let the switch set at “ Manual “ operation state.
Push green “ Turn On “ power switch button
When heating, please very slowly turn manual power button to increase power
Monitor temperature value on controller panel to desire value
a) When power on, if you can not hear a sound from cooling fan, please don’t continue to operate. You must shut
down power to check or replace the cooling fan.
b) During furnace running, please don’t touch furnace outside body to avoid any high temperature burn.
c) If the furnace is used at first time or it did not use for long time, please always preheat the furnace at 350 ℃ for
at least 2 hours to remove moisture inside chamber and to avoid chamber cracking.
d) Please always keep inside chamber clean before running furnace to avoid contamination to your sample.
e) Furnace must be used under following condition:
Inside room at temperature between -10 - 75℃
Elevate <1000 M, Humidity <85%
In the environment without vibration and conductible dust, explosive, flammable and corrosive gases.
f) KeJia Furnace provides one year limited warranty from date that we shipped goods. If you find any
13
defective caused by product’s quality please feel free to contact us for replacement during warranty period.
However, KeJia Furnace is not responsible for any damage or consequence damage caused by misuse. After
warranty, KeJia Furnace will continue to provide technical support and spare parts at a reasonable cost.
Green power indicator is off,but red Fuse (32A) in main power circuit is Open furnace front panel and
open circuit indicator is on. broken replace the fuse.
No current in meter,but there is Heating element is broken Find broken heating element and
max. voltage. replace it.
Controller panel SV show “HiaL” Furnace temperature > Cooling furnace down, and find
1200℃,protection from alarm reason why temperature is so high.
(program setting may be wrong)
Power and heating element are ok, Controller or related circuit may be Follow manual operation to run the
but the furnace can’t be controlled damaged. furnace. If voltage and current
by program output is normal,then there is
problem with controller. Replace
controller. If not,please inform
manufacturer to check what real
problem is.
Program is running,but the furnace oPL or oPH values are too low due Increase “oPL” “ oPH” value till
can not be heated, or real to local voltage lower or frequency current meter show “120A-140A”,
temperature is far behind the setting different. and program running smoothly.
value
During heating below 300℃,the It is normal because the furnace Don’t worry. Let the furnace heat to
temperature value (PV) is not comes with a high temperature 300℃,then thermocouple will
stable,and doesn’t match with thermocouple,which is not sensitive function stably with program.
setting programs. (SV) at temperature below 300℃
If the PV value at 300℃ is still less
than that of SV,you may need to
increase OPL value a little.
14
make good connection for ground wire, then you can run furnace.
15