Sie sind auf Seite 1von 11

A generic DMM test and

calibration strategy Application Note

The Digital Multimeter or DMM is one of the most widely used electrical measurement
instruments and provides electrical traceability to many industrial users. However, it is
the nature of measurement products that they require regular checks to ensure that
they operate not only in a functional sense, but also within their specifications. Such
regular checks are known as calibrations and require a certain amount of knowledge
of DMM technology in order to calibrate them effectively. This application note seeks
to identify the important functional and calibration parameters and to offer a generic
guidance strategy for their routine verification.

DMM specifications so that DMMs have developed, Examples of characteristics


the terminology has evolved to
A DMM manufacturer has two describe the performance param- • Stability with time and temperature
important goals when specifying eters. A summary of these terms • Linearity
a DMM. They must ensure that is given in appendix 1. • Noise
the performance and functional- • Frequency response or flatness
ity are clearly presented to the Performance parameters • Input impedance
prospective purchaser such that • Compliance/Burden
the purchasing decision can be It is convenient to think of • Common/Series mode rejection
made, and he will also ensure the performance in terms of • Crest factor
that the instrument’s metrologi- functional and characteristic • Power coefficient
cal performance is expressed to parameters, where the former
the best possible advantage in a describes the basic capability or
competitive situation. The speci- functionality e.g. the DMM can
fication is actually describing measure voltage, and the latter
the performance of the instru- is a descriptor or qualifier of
ment’s internal circuits that are its characteristics e.g. with an
necessary to achieve the desired accuracy of ± 5 ppm. One can
functionality. Each function will easily see that it is essential to
have an accuracy expressed in understand the significance of
terms of a percentage of the the characteristics if the instru-
reading ± %R with additional ment is to be verified correctly.
modifiers or adders such as ±
digits or ±µV. This is known as a Examples of functional
compound specification and is an parameters
indication that there are several • Functions: V, A, Ω
components to the specifica- • Scale length/Resolution
tion for any particular parameter. • Read rate
Sometimes this is seen as a way • Amplitude range
of hiding the true performance, • Frequency range
when in reality, it is the only
practical way to express perfor-
mance over a wide parametric
range with a single accuracy
statement. Over the 40 years or

F ro m t h e F l u k e D i g i t a l L i b r a r y @ w w w. f l u k e . c o m / l i b r a r y
DC voltage These contributions will be the effects of power dissipa-
combined to give a compound tion in the attenuators will give
Nearly all DMMs can measure dc specification expressed as ± a power law characteristic, the
voltage. This is because of the % Reading ± % Full Scale or severity of which will depend on
nature of the analogue to digital Range ± µV. In order that the the resistor design and indi-
converter (ADC) used to convert performance of the instrument vidual temperature coefficients.
from a voltage to (usually) timing can be verified by calibra- Knowledge of the design and
information in the form of clock tion, the above effects must be inter-dependence of the DMM’s
counts. The most common way isolated. That is to say, it is not functional blocks can greatly
of doing this is to use technique possible to measure Linearity assist the development of effec-
known as dual-slope integra- for example until the effects of tive test strategies.
tion. This method involves Offset and Gain errors have been
applying the input signal to an removed. Figure 1 shows these DMM functionality tree
integrator for a fixed period and basic parameters.
then discharging the integra- A DMM’s functionality and range
The ADC is common to all dependence is quite logical and
tor capacitor and measuring the ranges of all functions, therefore
time taken to do so. This basic is generally designed to get the
its characteristic errors will affect maximum use out of a minimum
dual slope method is used in low all functions. Fortunately, this
resolution DMMs, but longer scale means that the basic dc linear- of components through the use
length instruments require more of common circuits wherever
ity need only be verified on possible. As an example the ADC
complex arrangements to ensure the basic (usually 10 V) range.
good performance. DMMs are will be used for all functions,
The manufacturer’s literature the current sensing resistors
available with up to 8½ digits should indicate which is the
Resolution and these usually will be used for both ac I and
prime dc range. If this is not dc I, the ac rms converter will
employ multi-slope, multi-cycle stated directly, it can be deduced
integrators to achieve good be used for both ac V and ac I.
from the dc voltage specifica- A typical functionality “tree” is
performance over the operat- tion i.e. the range with the best
ing range. An ADC is usually a given in figure 2.
specification in terms of ± % R,
single range circuit, that is to say ± FS and ±µV will invariably be
it can handle only a narrow volt- the prime range. Other ranges DC voltage calibration
age range of say zero to ± 10 V, e.g. 100 mV, 1 V, 100 V and
strategy
however, the DMM may be speci- 1 kV will have a slightly worse For a new DMM, the manufac-
fied from zero to ± 1000 V. This performance because additional turer will test every aspect of
necessitates additional circuits in circuits are involved. At low the instrument’s performance.
the form of amplifiers and atten- levels on the 100 mV and 1 V However, for routine repeat cali-
uators to scale the input voltage ranges, the dominant factor will brations, the number of tests can
to levels that can be measured be noise and voltage offsets. be dramatically reduced if one
by the ADC. In addition, a high For the higher voltage ranges, accepts that instruments can be
input impedance is desirable Figure 1 DMM Offset, Gain and Linearity
such that the loading effect of the
DMM is negligible. Each amplifier
and attenuator or gain-defining
component introduces additional
errors that must be specified. Indicated Full Scale
The contributions that affect the Value
specifications for dc voltage are Slope = Gain
given below together with their
typical expression in parenthesis.

Contributions to dc V
Linearity is Expressed as
specifications
the Residual Errors
• Reference Stability (% of After the Removal of
reading)
• ADC linearity (% of scale) Offset and Gain Errors
• Attenuator stability (% of
reading)
• Voltage offsets (Absolute) Zero Offset True
• Input bias current (Absolute) Value
• Noise (Absolute) Zero
• Resolution (Absolute)
Figure 1. DMM offset, gain and linearity

2 Fluke Calibration A Generic DMM Test and Calibration Strategy


Figure 2 DMM Functional Blocks
of +10 ppm and the specification
Resistance ranges allowance was ±5 ppm, at first
glance the 100 V range gain-
defining components might be
10 100 1k 10 k 100 k 1M 10 M thought to be defective, when
in fact, the failure was caused
by the instrument’s internal
reference. Either way, the 100 V
0.1 mA 0.1 V 1V 10 V 100 V 1k V range has failed against the
specification, but the cause of
DC voltage ranges the problem has nothing to do
1 mA Current +Ref A dc -Ref with the 100 V range compo-
ranges Ref nents. Without this knowledge,
DCI the wrong components might be
10 mA replaced resulting in loss of his-
ACI
tory and additional expense.

0.1 A
Functional tests
RMS
AC voltage ranges It is a good idea to carry out
essential functional testing
1A 0.1 V 1V 10 V 100 V 1k V before committing valuable
time and effort to a potentially
lengthy calibration process.
Figure 2 DMM Functional Blocks Some microprocessor-controlled
DMMs have a diagnostic self-
test function that can be used to
perform sophisticated internal
characterized according to their DC voltage calibration checks. If such a facility is not
technology, operating principles sequence available, each function and
and accumulated historical data. range should be selected and a
It will usually be necessary to 10 V Zero, +10 V Gain, -10 V Copper short, or known voltage,
calibrate each dc voltage range Gain resistance or current applied
in both polarities. The “prime 10 V Linearity: ±1 V, ±5 V,±15 V, to the DMM’s input. This will
range”, where there is neither ±19.9 V ensure that a stable reading
gain nor attenuation of the input 1 V Zero, +1 V Gain, -1 V Gain* can be obtained and that the
voltage will typically be 10 V. 100 mV Zero, +100 mV Gain, controls and displays are operat-
This will be the first range to -100 mV Gain* ing correctly. Functional checks
be measured, as its errors will may be extended to cover input
100 V Zero, +100 V Gain, -100
also be passed on to the other switching circuits, auxiliary
V Gain*
ranges. Calibration would then outputs and remote-control digi-
progress to the lower ranges of 1 kV Zero, +1 kV Gain, -1 kV tal interface if applicable. If an
1 V and 100 mV, then on to the Gain* IEEE-488 interface is fitted and
100 V and 1 kV ranges. These * Some DMMs may not have negative polarity the DMM is one of a type that
range adjustments
are left until last because of will be regularly calibrated,it
the effects of increasing power Careful analysis of the results may be worth writing a simple
dissipation on both the DMM can reveal much about the program to perform the func-
and the calibration source at the instrument’s performance. tional testing. If a functional
higher voltages. The order of the Changes in the error of the 10 V problem is found, a decision
tests might be as follows: range will also be passed on to must be made as to whether it
other dependent ranges. There- is safe to proceed with a pre-
fore if the 10 V range was found adjustment calibration check.
to have increased by +5 ppm This is desirable if historical
and the 100 V range showed no performance data is required,
change, it means that the 100 V but may also assist with fault
range had actually changed by diagnosis. However,if there is
-5 ppm but was compensated by any risk of personal injury, or
the 10 V range error. damage to the instrument, it is
This kind of evaluation can better to repair the instrument
be very useful if failures against before proceeding with the
specification are observed. If the calibration.
100 V range indicated a change

3 Fluke Calibration A Generic DMM Test and Calibration Strategy


Input characteristics This will usually be very high Where adjustments are made,
indeed and may be the same as older designs of DMM must have
There should be a periodic test that for the lower dc V ranges. the dc voltage adjustments made
to ensure that the DMM’s input Any change in input resistance first. Consult the manufacturer’s
characteristics are correct. The will be directly reflected in the handbook to check for functional
test requires the use of an LCR resistance function accuracy inter-dependence in the calibra-
Bridge or relatively low accu- and so it is not normally neces- tion regime.
racy DMM. The object of the sary to measure it. The second For sensitive long-scale DMMs
test is to ensure that there are parameter is the Compliance of it is usually better to start with
not gross errors in the input the DMM’s current source. Ide- the high resistance ranges and
characteristics. Note that it is ally, the current flowing through then work down. This is will
essentially a functional test the external resistance being allow a longer thermal stabili-
looking for percentage rather measured should be independent zation time for the connecting
than ppm errors. As an example, of the voltage developed across leads and the DMM’s internal
a common ac V input resis- the resistor—a true constant cur- circuits, which may be work-
tance is 1 MΩ. The test will be rent source with infinite output ing at the 10 mA and 100 mV
to ensure that it is not exces- impedance. However, if the cur- levels for the current source and
sively high or low - say within rent does vary, it will affect the voltage measurement circuits
±10 %. For most instruments, resistance function’s linearity. respectively. In any case,where a
the ac input resistor will form This is discussed further in the 4 wire capability is available, it
part of the input attenuator so resistance section. should be used. Unless the DMM
significant changes in value will has a true ohms function, the
obvious when the calibration is Resistance calibration effects of voltage offsets must be
verified. The dc V function input
resistance measurement can be
strategy considered.
more difficult. For higher ranges The resistance function of a
DMM consists primarily of a
Resistance function
e.g. 100 V to 1 kV, there will be
an input attenuator of typically constant current source provid- linearity tests
10 MΩ that determines the input ing a range of currents typically For DMMs with a resistance
resistance and voltage divi- from 10 nA to 10 mA. Select- function accuracy of not better
sion ratio. Changes in the value ing a resistance range selects than 100 ppm (4½ to 5½ digit
of this component will cause an appropriate constant current resolution), a decade resistor
changes in the dc V calibration. to pass through the unknown box could be used, but for the
However, for lower ranges, the or standard resistance. The more accurate “premium” DMMs,
input resistance will probably voltage developed across the resistance linearity can be more
be determined by feedback and resistance is then measured by difficult to determine unless a
so may be very high indeed. In the DMM’s DC voltage function selection of suitably accurate
fact it may be so high that it is with an appropriate range set. It four- terminal resistances is
not possible to measure reliably. can be seen from figure 2 that available. One of the problems
The usual reason for checking the accuracy of the resistance of trying to measure resistance
the input characteristics is to function is dependent upon both linearity directly is the uncer-
ensure that they do not sig- the resistance current source tainty of the individual resistor
nificantly disturb the test circuit (providing the resistance ranges) values. For example, to measure
conditions. This is sometimes and the dc voltage ranges of the linearity on the 10 kΩ range of a
described as being non-intru- DMM. Stability of the currents 7½ digit DMM with a maximum
sive. For current measurements, will typically be controlled from indication of 19.000 000 kΩ
it is the Burden voltage that is the DMM’s dc reference circuit. would require several different
important. In a DMM, the burden In this case, changes in the resistance standards. Assum-
is determined by the value of reference are common to both ing that measurements were to
the current sensing resistors and the current sourcing and voltage be made at a minimum of five
can be measured as a resistance sensing resulting in no change evenly spaced points through-
between the current sensing in resistance measurement accu- out the range e.g. at zero, 5 kΩ,
terminals. Typically, the resis- racy. However, changes in the 10 kΩ, 15 kΩ and 19.9 kΩ, the
tors will be chosen such that performance of the either the difficulties in finding suitable
the burden is 0.1 V or lower. For current defining resistors or volt- standards soon become obvious.
the resistance function, there age gain-defining resistors will Typically, resistance standards
are two parameters of interest. affect the resistance calibration. will be available at the normal
The first is the input resistance For this reason, the resistance decade values of 10 Ω (25 Ω may
of the voltage sensing terminals function is usually measured be available), 100 Ω, 1 kΩ, 10 kΩ,
(assuming 4 wire capability). after the dc voltage function. etc. and so do not provide even
coverage throughout the range.

4 Fluke Calibration A Generic DMM Test and Calibration Strategy


Figure 3 DMM Resistance Converter
Figure 3 shows the circuit
configuration for resistance
measurement used in a high
DC voltage pre-amp accuracy DMM. The resistance
Hi function is primarily a range of
Hi selectable constant currents. A
constant current generator forces
sense
a current Ix to flow through the
Rx Ix test resistor. A true constant
current source will generate a
Lo sense current independently of the
Lo voltage developed across its
terminals,in this case designated
Lo follower Hi and Lo. It therefore follows
that if a known resistance is
applied to the DMM and the
display value noted, the inser-
Constant tion of an additional resistance
current Resistance
range control in series with the Hi lead should
sink not significantly affect the DMM’s
reading. This will confirm that
the current source can deliver
the same current through a
range of resistance values.
Figure 3. DMM resistance converter If it can also be confirmed that
the dc voltage range used for
When one considers that some of medium to high accuracy the resistance measurement (see
DMMs have resistance linear- DMMs. For lower accuracy DMMs figure 2) is also linear,there is
ity specifications of better than with up to 5½ digits resolution, then a technically sound way of
0.5 ppm, and that individual resistance calibrators suitable for confirming resistance linearity
resistance standards may have resistance linearity testing are without the need for a resistance
uncertainties of 1 ppm or more, readily available. If it is neces- linearity standard. Note that the
test methods using separate sary to determine the resistance series resistance does not need
resistors or decade boxes will be linearity of a long-scale DMM, to be a precision resistor—it could
inadequate. or confirm that the circuits are be a low-noise potentiometer.
For this reason, resistance operating correctly,the following
linearity is not usually mea- test might be considered.
sured for routine calibrations
Figure 4 DMM Frequency Response

HF performance
determined by
VLF performance capacitive effects
determined by RMS LF performance
low -pass filter determined by gain
defining resistors

Error

Specification LF adjustment HF adjustment


Bands - - - - -

10 Hz 100 Hz 1 kHz 10 kHz 100 kHz 1 MHz

Figure 4. DMM frequency response

5 Fluke Calibration A Generic DMM Test and Calibration Strategy


Figure 5 Log-Feedback RMS Converter

From rectifier

V in V2 Vx To ADC
Log Anti- Low Pass
Log Vf Filter
Log
V out

+
Sum
+ -

Log Log
Vf

Figure 5. Log-feedback RMS converter

AC voltage calibration Various means maybe used to provides the square-root element
strategy trim the response e.g. variable of the computation and controls
capacitance trimmers or perhaps the gain of the circuit to give a
The ac functions have the software-controlled adjustment wide dynamic range. The low-
added dimension of frequency. of the high frequency gain. pass filter determines the low
This complicates calibration by The DMM manufacturer will frequency response of the instru-
introducing additional test points recommend certain adjustment ment. Some log-feedback DMM’s
for each amplitude range. In a points that will compensate for are capable of operating down to
typical DMM, the ac measure- LF gain and frequency response 0.01 Hz.
ment is made by an ac converter. for each range. These should There are two calibration
This module is must provide be strictly observed as they are adjustments that are peculiar to
gain and attenuation for a wide designed to give the best overall log-feedback, true-rms detectors.
range of signals from typically performance over the specified The first is dc Turnover, where
a few millivolts to 1 kV. DMMs frequency range. Figure 4 shows the gain of the precision recti-
are available with 6½ digit a typical response and control- fier is adjusted to be identical
100 mV ranges with a resolu- ling mechanisms. for both positive and negative
tion of 100 nV. This resolution The prime range of the ac excursions of the input signal.
is only meaningful down to an function is usually 1 V, with The second is Crest Factor,
input voltage level of perhaps other ranges being obtained by which can be affected by non-
1 mV, below this level, noise gain or attenuation of the input symmetry of the rectifier and
and linearity errors are likely to signal. From figure 2 it can also by the bias arrangements of
dominate the reading. It is likely be seen that all the ac ranges the analog multiplier. Analogue
that the limiting factor will depend on the rms converter. multipliers require very careful
be the noise of the calibration A true-rms converter consists matching of their semiconduc-
source, rather than the DMM. of a precision rectifier, Logarith- tors. In order to achieve this,
The gain of a DMM’s ac func- mic and Exponential amplifiers there may be bias adjustments in
tion will vary with frequency. that perform the function shown the circuit that will affect crest
This is known as its Frequency in figure 5. factor performance. These adjust-
Response and requires that The circuit performs an ments tend to be very interactive
measurements are made at key analogue computation of the and must be made in a strict
points throughout each ampli- rms value of the signal through order, and usually before the
tude range. At low frequencies, the use of logarithmic ampli- individual range gain adjust-
say from 40 Hz to 5 kHz, the fiers that effectively square and ments are performed. However, it
DMM’s gain is determined by square-root the input signal. The is not usually necessary to make
resistor networks, its ac to dc output of the multiplier Vx has a adjustments to the rms circuits
converter and dc reference. At dc ripple content that is aver- for routine calibrations. If these
higher frequencies, reactive aged (to obtain a Mean value) by adjustments are required and
effects, primarily capacitive, applying it to an active low-pass there is no requirement for pre-
will determine the flatness of filter. Vout is also used to provide adjustment data, the sequence
the DMM’s frequency response. the feedback signal Vf which would be as follows:

6 Fluke Calibration A Generic DMM Test and Calibration Strategy


AC voltage calibration the region of 100 mV. Figure 2 DC current calibration
sequence shows that the current ranges sequence
are linked together. The shunts
1 V ac zero, ±1 V dc turnover* are connected in series with 100 µA zero, +100 µA, -100 µA
Crest factor: (at 3:1 or 5:1 as suitable tapping points for the 1 mA zero, +1 mA, -1 mA
required)* voltage sensing. To measure 10 mA zero, +10mA,-10mA
1 V LF gain, 1 V HF gain, check 0.1 mA, the current is passed
100 mA zero, +100 mA*,
frequency response* through all of the shunts in
-100 mA*
1 V LF linearity, 1 V HF series and their values will be
linearity* chosen such that the sum of 1 A Zero*,+1 A*,-1 A*
their resistive values for any *T
 hese measurements may require a longer
100 mV zero, 100 mV LF, 100 mV
particular current range will settling time due to self-heating and-
HF, check frequency response* thermo-electric effects.
develop 100 mV for the specified
10 V Zero, 10 V LF, 10 V HF, current. This greatly simplifies
check frequency response* the current switching but also AC current calibration
100 V Zero, 100 V LF, 100 V HF, means that for any given current sequence
Check frequency response* range, all of the shunts included 100 µA LF gain, 100 µA HF
1 kV Zero, 1 kV LF, 1 kV HF, will have an affect on that gain**
check frequency response* range’s performance. Knowl-
edge of the inter-dependence of 1 mA LF gain, 1 mA HF gain**
*T
 hese adjustments will usually be iterative.
Always follow the manufacturer’s recom- the shunts can assist with drift 10 mA LF gain, 1 mA HF gain**
mendations for the adjustment sequence.
and fault diagnosis. The volt-
100 mA LF gain, 1 mA HF gain**
Note that there may be sig- age developed across the shunts
will be passed to the dc or ac 1 A LF gain, 1 A LF gain**
nificant differences between
the methods recommended by 100 mV ranges depending upon ** Some DMMs do not have an HF adjustment
different manufacturers. AC zero whether the DCI or ACI function
measurements are often made has been selected. Zero considerations
with a short-circuit applied Therefore changes in the
All dc functions will need their
to the DMM’s input, but some dc or ac 100 mV range per-
zero offsets evaluated and com-
instruments will require a small formance will also affect the
pensated before measurements
ac bias voltage because the rms current ranges. Current calibra-
of the gain values are made.
converter cannot operate with a tion will normally be the last
Generally, it is the gain values
zero input. Where the DMM has function to be calibrated for the
that transfer traceability. This
a very low frequency capability, simple reason that knowledge
is recognition of the fact that
the zeros and gain measure- of the dc V and ac V perfor-
zero is really only a baseline for
ments make take an appreciable mance is essential before the
the measurements. It is impor-
time to settle to their final DCI and ACI range performance
tant that all DC measurements
values—particularly after a large can be evaluated. The order of
are referred to a known offset
change in signal amplitude. calibration would normally be
state, but that state does not
as shown below and is intended
have to be exactly zero. In fact,
Current calibration to minimize any heating effects
at normal room temperatures,
by measuring the lowest cur-
strategy rent ranges first. Note that it is
it is very difficult to achieve
A DMM measures current by a true zero in terms of Volts,
not normal practice to switch
sensing the dc or ac voltage amps or ohms. Consideration of
alternately between DCI and
developed across a current- zero offsets should also include
ACI functions for each range,
sensing resistor or shunt. The the calibration source, as all
although this could be done. Note
same shunts will be used sources have residual offsets.
that some DMMs do not have
for both dc and ac current. A From figure 1 it can be seen that
independent adjustment of the ac
separate pair of terminals will an offset will affect all read-
current ranges but rely on the dc
be provided to simplify internal ings by a fixed amount. It is
gain adjustment, the Flatness of
signal switching and minimize very important that a”system”
the shunt’s ac response and the
the impedance of the internal zero is performed when a DMM
accuracy of the ac voltage con-
connections to the shunts. It is being calibrated. A system
verter. In this case the ac voltage
is very desirable to keep the zero means zeroing the DMM to
and dc current ranges must be
Burden voltage as low as pos- the zero offset of the calibra-
adjusted first.
sible, typically this will be in tion standard. This will remove
the effects of the standard’s
offset from the measurement of
the gain or full-range values.

7 Fluke Calibration A Generic DMM Test and Calibration Strategy


For voltage measurements the Measurement uncertainty The resolution or scale-length
DMM should be zeroed to the of the DMM determines the
source “zero” output. For resis- The most significant contribution smallest change in the reading
tance measurements the DMM to the quality of DMM calibration that may be observed. Clearly,
should be zeroed to the resis- is consideration of the measure- it could become a limiting factor
tor (as described in the user’s ment uncertainty. It is beyond in the measurement regardless
handbook) such that the effects the scope of this application note of how accurate the calibration
of voltage offsets in the Hi and to give a detailed analysis of the standard might be. Assuming
Lo circuit are removed. Current uncertainty contributions and the calibration of 1 V dc, a 6½
measurements require the DMM their combination, but it is worth digit DMM with a scale-length
to be zeroed in the open-circuit identifying what the sources of of 1.000 000 V can resolve 1 µV
state, although some calibration uncertainty will be. or 1ppm of its nominal range,
sources have a “zero” current whilst a 4½ digit DMM can only
output to which the DMM should Uncertainty contributions resolve 100 µV or 0.1 % for the
be zeroed. • Uncertainty of the standard same conditions.
• Resolution of the DMM The short-term stability of
Generating a test plan • Short-term stability of the the DMM (and the calibration
After considering the preceding DMM with time and tempera- standard) with time and tem-
discussions and the intended ture perature will also affect the
application of the DMM, a Test uncertainty of measurement.
• Combined noise of standard Usually the dominant factor here
Plan can be devised. The plan and DMM
must ensure that the basic is temperature coefficient of the
• The calibration procedure DMM and stability of the calibra-
calibration requirements of the
instrument and user are met, It is imperative that the calibra- tion environment. Secular drift is
that the measurements are tion standard is of sufficient not usually significant unless the
made using standards of suf- accuracy to be able to calibrate calibration takes several hours or
ficient accuracy, and that the the DMM with confidence. Some the instrument is defective. An
functionality and integrity of the manufacturers separate DMM allowance for temperature coef-
instrument are verified. The plan specifications into two parts: the ficient and short-term stability
must also consider actions to be calibration uncertainty and the can usually be obtained from the
taken if the instrument requires instrument’s relative accuracy. If manufacturer’s specification data.
repair. It is likely that there will the available calibration uncer- A more dominant uncertainty
be two strategies depending on tainty is larger (worse) than the contribution will usually be the
whether or not a fault condi- specified calibration accuracy, noise or run-around of the DMM
tion is suspected or has been the DMM’s total accuracy speci- reading during the measurement.
reported. fication will no longer apply. Whilst it may be interesting to
Sometimes Test Accuracy Ratios consider the individual noise
Routine calibration (no (TAR) are quoted as a require- contributions of the standard
ment i.e. a TAR of 4:1 requires and the DMM, in practice it is
repair) the accuracy of the standard to their combined effect that is
• Have any problems been be four times better than the important. If individual read-
reported? specification of the DMM. This is ings can be easily observed, and
• Functional checks, controls, to ensure that the residual errors if the noise is predominantly
etc. of the standard do not sig- random,the sample standard
• Measure prime dc and ac nificantly affect the calibration deviation of the readings can
linearity accuracy of the DMM. In an ideal be calculated and used as the
world, TARs could be applied uncertainty contribution for com-
• Measure all gain values bined noise. It is important that
to all measurements—including
• Adjust as required those made by National Stan- the configuration of the DMM is
• Check volt/hertz limits, etc.* dards Laboratories. However, it representative of normal use for
* Unless restricted use intended is not practical at this level—or this measurement.
even necessary, if all sources Finally, the calibration proce-
Additional requirements of error have been identified dure itself will have an influence
if repaired* and corrected and that a sound on the measurement uncertainty.
uncertainty analysis has been A poorly chosen test sequence,
• Adjust ADC if required
performed. Where such correc- insufficient settling time or poor
• Adjust line locking if required tions are not applied, or where interconnection techniques will
• Adjust Crest Factor if applicable there is no calibration accuracy all introduce additional errors
• Check Input Characteristics requirement specified, TARs are that may pass unnoticed by the
• Check CMRR appropriate. operator. This is the reason why
*K
 nowledge of instrument design and repair
details will facilitate test plan generation

8 Fluke Calibration A Generic DMM Test and Calibration Strategy


the manufacturer’s recommended Resolution and scale length - Frequency response - Tradi-
procedure should be used as A way of describing how many tionally, frequency response is
the basis for DMM calibration. figures a DMM can display. interpreted as the -3 dB point
Note that this also applies to the Resolution is the number of i.e. the high frequency point at
use of the calibration standards. digits e.g. 1.000 000 V is a which the reading has reduced
If further reading is required 6½ digit display where the ½ to 70.7 % of its nominal low fre-
on the subject of uncertainty digit is the “1” and there are six quency value. The manufacturer
analysis it is recommended that places after the decimal point. may employ various hardware
a study of one of the Guides to The Scale Length is the maxi- and software adjustments
the Estimation of Uncertainty of mum reading available on any to the response to make the
Measurement be made. particular range i.e. on the 1 V -3 dB point as high as possible.
range a maximum reading of This can result in undesirable
Summary and 1.999 999 V might be available peaks and troughs at lower
recommendations before an overload is indicated. frequencies. A more meaningful
The maximum resolution on the measure of response is flatness.
The calibration of long-scale
lowest range is the Sensitivity. Flatness is a measure of the
DMMs can be technically chal-
Linearity - Basically linear- deviation from an assumed flat
lenging but there is a great deal
ity is a description of how the response over specific frequency
of satisfaction to be obtained
response of an instrument’s Adc bands and may be specified
from doing it well, understand-
might vary with the amplitude in ppm or % rather than dBs.
ing the technology and having
of the measured signal. This Depending on the rms conver-
total control of the process.
should not be confused with sion technique used, a DMM’s
However, the standards required
power coefficient, which is a ac response can suffer from low
for the effective support and
measure of the effects of self frequency problems as well as
calibration of such DMMs can
and mutual heating in resistive high frequency roll-off.
be very expensive and hard to
justify if the number of units to attenuators. Input resistance - Is important
be calibrated is small. For this Zero or Offset - All dc mea- because it describes the poten-
reason, the manufacturer’s cali- surements are affected by tial loading effect of the DMM on
bration service should be used residual offsets that may be in the test circuit. For dc voltage up
if there is any doubt about the volts, amps or ohms. However, to 20 V, the input resistance may
adequacy of the locally available it would be very unusual (in be >1012 Ω and is the result of
equipment and expertise. the context of DMM calibra- using feedback and FET devices.
tion) to find offsets of such large Note that some DMMs have a
Appendix 1- DMM amplitude. significant input capacitance on
terminology Typically, offsets would be their dc voltage function.
in µV, µA or µW. In a voltage At higher voltages up to 1 kV,
This appendix gives definitions an attenuator is switched in to
of some of the terms commonly measurement, placing a Copper
short across the voltage input divide the applied voltage down
used by DMM manufacturers. to 10 V or 1 V. The attenua-
Text in Italic typeface indicates terminals of the DMM should
result in a reading of exactly tor typically has a resistance of
related terms. 10 MΩ. For ac voltage, the input
zero Volts. However, thermally
Function - Functional options generated offsets, resulting from resistance will usually be lower—
of a DMM e.g. dc voltage, ac the use of dissimilar metals typically 1 MΩ with 150 pF
voltage, resistance, current, in the DMM’s internal circuits in parallel. Note that at high
frequency, etc. Can also apply to and electronics, will result in frequencies the input Impedance
ratio/channel switching or math a small net voltage offset that will reduce.
and statistical functions. may amount to several tens of Input bias current - Ampli-
Range - A DMM’s measurement µV. Such offsets will normally be fiers are usually designed for
capability is split into ranges relatively stable in a controlled high gain, high bandwidth,
such that the input signal can be environment and most DMMs high input impedance, low
scaled to a level appropriate for will have some form of offset output impedance and low input
its A dc or RMS converter. Typi- compensation or Input zero func- voltage offsets. However, all
cal ranges would be 100 mV, tion that may be used to remove bi-polar semiconductor devices
1 V, 10 V, 100 V, 1 kV ac or dc, their effects. It is very important are essentially current oper-
10 Ω, 100 Ω, 1 kΩ, 100 kΩ, 1 MΩ, that the effects of offsets are ated. That is to say they require
10 MΩ, 100 MΩ, 1 GΩ, 100 µA, removed from the measure- an input current to “bias” them
1 mA, 10 mA, 100 mA,1 A and ment, otherwise they will into a desirable (usually linear)
sometimes, 10 A ac or dc. Some introduce a fixed error (in terms portion of their operating curve.
DMMs have ranges in threes. of volts, amps or ohms) through- Even field-effect devices may
out the measurement range. exhibit some internal leakage
current back to their power

9 Fluke Calibration A Generic DMM Test and Calibration Strategy


supplies. Sensitive DMM design operational amplifier and varied such that the conversion cycle
seeks to minimize these leak- in relation to the amplifier’s is locked to the line frequency.
ages by component selection power supplies, there should be The input voltage is integrated
and careful circuit design. no change in its output voltage. over one or more complete line
However, there will inevitable In practice, due to small varia- frequency periods, effectively
be some residual input cur- tions in gain of the amplifier’s integrating any line frequency
rent, which unless preventative internal complementary stages content to zero. Line-locking is
action is taken, will flow in the with amplitude, there will be a a very effective way of reducing
external measurement circuit. change in output—even though line-related noise. Care must be
A small current of 1 µA flow- the inputs are at a common taken however to ensure that
ing in a 1 MΩ resistance will level. High quality DMMs will where line frequency is select-
develop a voltage of 1 V, 10 pA have special power supplies able, the line-locking is set to
in 1 MΩ will develop 10 µV. A for the amplifier that “track” the correct line frequency i.e.
modern sensitive DMM would be or follow the input voltage. As 50 Hz or 60 Hz.
expected to have an input bias a result, the voltage difference Guard - A guard is a circuit that
current of <50 pA. between the amplifier’s input may be used to intercept and
Compliance - Is a measure of and its power supplies remain divert, or control leakage cur-
the maximum voltage that can constant over the operating rents. It may be passive, where
be developed by the (usually range such that changes in it is simply a conductive screen
constant) current source of a differential gain with ampli- or case around sensitive com-
DMM’s resistance converter tude are eliminated. Another ponents and circuits, or active
before the current changes example is where there is a where amplifiers actively sense
significantly. It is an indication leakage between the Lo input and control the screen poten-
of how much lead resistance can and power line ground. In this tial. In either case, the guard
be tolerated between the DMM case, common mode errors can will be connected to a defined
and the resistor being measured. be reduced by careful Guard- point and potential. Connection
ing techniques to reduce the to the guard will usually be via
Crest factor - Is the peak-to- effects of unwanted leakages to a 4 mm binding post terminal.
rms ratio capability of a DMM’s ground. The guard effectively Electrically, the guard potential
ac converter. A pure sine-wave diverts interference currents lies at a defined point between
has a crest factor of 1.414:1, a such that they do not flow in the ground and (usually) the Lo
square wave has a crest factor Hi or Lo signal path and there- input terminal. When line pow-
of 1:1. Ideally, a DMM should fore do not generate unwanted ered instruments are connected
indicate the “True” rms value for voltages due to the impedances together, circulating leakage
any waveform applied to its rms and leakages involved. The currents may flow around the
detector. Common Mode Rejection Ratio “loop” formed by their respec-
Where amplifiers and protec- (CMRR) is expressed in dBs and tive power supplies and input
tion circuits are employed in the is measured by inserting a 1 kΩ lead connections. The “loop” is
DMM’s ac signal path, the peaks (source imbalance) resistor in completed by the connections
of high crest factor signals may series with the Lo input connec- to the line supply. Such currents
be clipped resulting in a distorted tion to the DMM whilst varying usually manifest themselves
signal being applied to the rms the applied input voltage. as noisy or inflated readings
converter. It is desirable to have
Series mode rejection - A in sensitive measurements and
a high crest factor capability. The
measure of how the DMM may be eliminated by connect-
actual capability may vary with
responds to an interference ing the guard of the DMM to the
the overall magnitude of the
signal that is not common to ”earthy” terminal (not directly
signal in relation to the DMM’s
both inputs. Usually this means to earth) of the “source” instru-
nominal range and internal power
that an unwanted ac signal is ment. In order for the guard
supply voltages. Where crest
superimposed on a wanted dc to work, it must be used in
factor problems exist, it may be
voltage. The signal is removed “remote” configuration. This is
possible to select a less sensitive
by various type of signal condi- enabled by means of a local/
range such that no additional
tioning or filtering. Filtering may remote guard switch or a short
distortion of the signal occurs.
be through the use of analogue link. In either case, local guard
Common mode rejection - passive or active circuits com- usually connects the DMM’s Lo
A measure of how the DMM prising RC networks, or may be and Guard terminals together.
responds to the presence of a digital utilizing an integration 4 wire resistance - A mea-
particular kind of interference rejection notch or averaging the surement technique used to
(unwanted) signal. The inter- results of a number of read- minimize the effects of the resis-
ference may be ac or dc. If a ings. An integration rejection tance of the connecting leads. In
voltage is simultaneously applied notch synchronizes the input
a four wire resistance measure-
(common) to both inputs of an integration time of the Adc ment, separate conductor pairs

10 Fluke Calibration A Generic DMM Test and Calibration Strategy


are used to connect the current cycle time means that the effects Relative accuracy - The accu-
and voltage circuits. Ideally, a of dynamic (varying) offsets can racy of the DMM relative to
sensitive resistance measur- also be removed. True ohms may but not including calibration
ing DMM should source its test not be effective where the test standards. This term can be mis-
current from a constant current resistor has a significant time leading but is really a measure
Source, where the output imped- constant. For this reason, its use of the instrument’s performance
ance is sufficiently high such is normally restricted to values in terms of stability with time
that the current flowing through less than 10 kΩ. and temperature, linearity, flat-
the resistor is independent of True-rms - In high accuracy ness and noise. Generally, these
the voltage developed across it. ac voltage measurement, the parameters are independent of
Additionally, the voltage mea- most useful description of the the calibration uncertainty.
surement circuit should have amplitude is the root-mean- Total accuracy - The combina-
a low input bias current and square value. Traditionally, this tion of calibration uncertainty and
sufficiently high input resis- has been based on equivalent Relative Accuracy. In specification
tance such that all of the test or effective heating power or terms, the combination may be a
current flows through the test work that could be done. The simple arithmetic summation of
resistor and there is no current realization of this is through the two terms, or a more complex
flowing in the voltage sensing thermal converters and some Root-Sum-of-Squares (RSS)
leads. Under these conditions, DMMs employ this technique combination at a specified confi-
the measurement will not be for ac measurements. However, dence level.
affected by the connecting lead thermal converters are not
resistance. Because of power appropriate for all applications,
supply limitations, there will be resulting in the development
a limit on the maximum lead of electronic RMS convert-
resistance which may be of the ers. These effectively compute
order of 100 Ω in any or all of the RMS value of the signal
the connecting leads. in analogue terms by using Fluke Calibration. 
True ohms - In a sensitive log and anti-log (exponential Precision, performance, confidence.™
resistance measurement, it can gain) circuits to simultaneously
be difficult to separate residual square and square-root the
resistance from residual volt- signal followed by an averag-
age offsets. Several methods ing filter that extracts the mean
may be employed to determine value, so providing the true
and remove voltage offsets. One root-mean-square of the signal.
method, sometimes employed The term true-rms was coined
in highly accurate resistance to distinguish between modern
bridges, is to make two mea- RMS sensing instruments and
surements, one with the test the older or lower cost DMMs
current flowing through the that used mean or average
resistor in the normal direction, sensing, but were adjusted to
then a second measurement indicate the RMS value. The
with the current polarity method assumes that the signal
reversed. Simple voltage offsets being measured is a perfect sine
will add to one measurement wave.
and subtract from the other. As If non-sinusoidal waveforms
a result, the effects of the offset are measured the DMM’s reading
will cancel. Current reversal will be incorrect. Fluke Calibration
is not normally employed in Calibration uncertainty - The
PO Box 9090, Everett, WA 98206 U.S.A.
DMMs, but a technique known total uncertainty of the cali- Fluke Europe B.V.
as true ohms is used to cycle PO Box 1186, 5602 BD
bration standard (of whatever Eindhoven, The Netherlands
the test current on and off. The type) traceable to national and For more information call:
residual voltage present across international standards. The In the U.S.A. (877) 355-3225 or
the test resistor when the cur- uncertainty will include the Fax (425) 446-5116
rent is off, is mathematically contribution of the DMM during
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
subtracted from the voltage the calibration process. In Canada (800)-36-FLUKE or
developed when the current is Fax (905) 890-6866
on, to provide an offset corrected From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
reading. Careful control of the Web access: http://www.fluke.com
©2011 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 3/2011 3988747A A-EN-N

Modification of this document is not permitted


without written permission from Fluke Corporation.

11 Fluke Calibration A Generic DMM Test and Calibration Strategy

Das könnte Ihnen auch gefallen